a Machine lid
b Door handle
c Conductivity display meter
456
d Power cord
e "Entrance" hose (blue) connects to
tap water
f Hose "H
labwasher
O pur" (green) connects to
2
3
Guide to the machine
78
EINGANG
ENTREE
ENTRANCE
9
g Water demineralization cartridge
h Connection "Exit"
i Connection "Entrance"
4
IMPORTANT SAFETY INSTRUCTIONS
This machine conforms to current
safety requirements. However,
inappropriate use can lead to
personal injury and property
damage. To prevent accidents and
machine damage, read all
instructions before using this
machine. Keep these instructions in
a safe place, and make them
available to future users.
Use
This machine is designed for use
as described in these Operating
Instructions. Using it for purposes other
than those for which it was designed
would be unauthorized and could
cause harm. The manufacturer cannot
accept responsibility for damage
resulting from improper use or
operation.
The installation of this unit at non
stationary locations (e.g. ships)
must be performed by a qualified
installer or service agency, in strict
accordance with national and local
safety regulations and standards.
Please pay attention to the following
notes, to maintain safe procedures.
The machine should be
commissioned and then
maintained only by a Miele Authorized
Service Technician.
Do not install the machine in an
area where a danger of explosion
or freezing may be present.
Be certain this appliance is
properly installed and grounded by
an authorized technician. To guarantee
the electrical safety of this appliance,
continuity must exist between the
appliance and an effective grounding
system. It is imperative that this basic
safety requirement be met. If there is
any doubt, have the electrical system
checked by a qualified technician. The
manufacturer cannot be held
responsible for damage or injury
caused by missing or defective
grounding, such as electric shock.
A damaged machine is dangerous.
Turn off the power immediately and
call an authorized Miele Service
Technician.
Repair work may only be carried
out by a Miele Authorized Service
Technician. Repairs by unqualified
persons could be dangerous.
Personnel operating the machine
should be trained regularly.
Untrained personnel must not be
allowed access to the machine or its
controls.
Water in the machine is not suitable
for drinking.
5
IMPORTANT SAFETY INSTRUCTIONS
Never clean the machine near or
with a water or high pressure hose.
Before servicing, disconnect the
power supply by either removing
the fuse, unplugging the unit or
manually "tripping" the circuit breaker.
The following points should be
observed to avoid damage to the
machine and the loads being
cleaned.
Use only resins which are tested by
Miele. If in doubt contact the Miele
Applications Specialist.
The resins are made for one time
use only. Dispose of them
accordingly.
Please follow the advice on
installation in this manual and the
separate Installation Instructions.
To avoid any corrosion damage
ensure that solutions or steam
containing hydrochloric acid do not
come into contact with the steel outer
casing of the machine.
Using accessories
Only specific equipment made by
Miele should be connected to this
machine. Consult Miele on the type and
application of such equipment.
Disposal of an old appliance
Cut off the cable directly behind
the appliance and at the plug to
prevent misuse. For environmental and
safety reasons ensure the machine is
completely emptied of any residue,
water and resins. (Observe safety
regulations and wear safety goggles
and gloves). Make the door lock
inoperative, so that children cannot
accidentally shut themselves in. Make
appropriate arrangements for the safe
disposal of the machine.
The manufacturer cannot be held
responsible for damages caused by
not observing the Important Safety
Instructions in this manual.
6
Description of the machine
The Aqua Purificator G 7895 is
operated using the E 318 water
demineralization cartridge:
The cartridge is empty when delivered
and must be filled with 20 l of one-time
use resins (see "Filling the Cartridge E
318 with one-time resins E 315").
When the cartridge is empty refill it with
new one-time resins.
The resins are made for one time
use only. Dispose them accordingly.
To ease the emptying and refilling of the
water demineralization cartridge E 318,
we recommend as additional
equipment to the cartridge E 318, the
"Original Miele resin replace set E 316"
which includes:
30 l Plastic keg with special lid and
–
funnel
2 bags one-time use resins E 315, 10
–
l each, with filter bag for the used
resins
You can order 20 l of additional
one-time resins with the part number
E 315.
7
Filling Cartridge E 318 with one-time resins E 315
Opening the lid
^ Hold the lid, to keep it from falling
into the cartridge and unscrew the
winged nut.
^ Remove the bracket.
Filling the resins into the cartridge
Insert the funnel into the cartridge
^
opening.
^ Empty both bags of resins into the
cartridge.
^ Clean the cartridge opening of any
resins.
^
Open the lid.
8
^ Insert the lid, hold it at the venting
valve. Tighten the bracket and
winged nut.
Open the machine front.
^
Unscrew the caps from the cartridge
^
connection.
E
IN
G
A
N
G
ENTREE
EN
TR
AN
C
E
^ Screw the hoses onto:
"Entrance" (blue) to "Entrance",
"H
O pur" (green) to "Exit".
2
^ See graphic for hose location.
Connection of E 318 cartridge
9
Connection
Connect the "H
^
O pur" hose (green)
2
to the DI water intake hose of the
respective machine.
Connect the "Entrance" hose (blue)
^
with the on-site cold water valve.
See the Installation manual
Important
The water pressure (flow rate) must be
between 37 and 145 kPa.
The water temperature should not
exceed 68 °F (20 °C).
^ Plug the machine into the outlet to
operate the conductance display.
Priming the cartridge
E
IN
G
A
N
G
E
N
T
R
E
E
E
N
T
R
A
N
C
E
^ Open the venting valve (arrow) by
turning it counterclockwise.
^ Open the cold water valve until water
appears from the venting valve.
^ Close the venting valve
10
Replace the cartridge
^
Place the connected cartridge in the
rear of the machine. The front area
can be used for an additional
cartridge.
^
Open the cold water valve. The
machine is ready for use.
Conductance display
The conductance display shows the
conductance of the demineralized
water, measured in ZS/cm (micro
siemens per centimeter) and is a
quality control for the demineralized
water.
If the water demineralization is below
the minimum value, the cartridge needs
to be refilled. A manual is supplied with
the "Original Miele resins replace set
E 316".
Uninstall the cartridge E 318
Turn off the water valves.
^
If available, open the water valve for
^
manual DI water extraction.
Open the priming valve the cartridge.
^
Dry the water exiting from the
cartridge..
Unscrew the hoses.
^
Turn the cartridge over to let the
^
residual water run out.
11
Conductance measure module
The conductance measure module has
the same function as the built-in
conductance measure module in the
Aqua Purificator G 7895.
The water conductance is measured
and a quality control of the
demineralized water enabled.
^ The connection of the conductance
measure module is always after the
water demineralization cartridge:
a water valve
b non-return valve with pipe venting
c water demineralized cartridge
d conductance measure module
f machine.
12
^ Install the module at a wall or in a
cabinet. Do not install the module
on the floor or underneath water
valves.
^
Electrical connection e: Plug the
cord into an outlet.
Please also read the separate
Installation Instructions supplied
with this machine.
All electrical work must be
,
carried out by a suitably qualified
electrician in accordance with local
and national safety regulations.
The unit is supplied without a plug. A
^
proper plug for the receptacle
available is user supplied.
The electrical requirement is 220V.
^
This is needed only to power the
meter.
Pease contact Miele Technical
^
Service at the phone numbers
located on the back cover of this
manual is there are any questions.
^ To increase security it is
recommended to install a protective
switch (30 mA) to the machine.
^ When replacing the power supply
line use an original Miele spare part
or a respective line with conductor
sleeves.
Electrical connection
^
For Technical Data see data plate.
The machine must only be operated
with the voltage, frequency and fusing
shown on the data plate.
13
After sales service
Should you have any difficulties please
contact your authorized Service Dealer
or the Miele Service Department. See
back cover for addresses. Please quote
the model and serial number of your
machine. This information can be found
on the data plate (see also “Electrical
connection”).
14
Caring for the environment
Disposal of the packing
material
The protective packing materials are
environmentally safe and can be
recycled.
Ensure that any plastic wrappings,
bags, etc. are disposed of safely and
kept out of the reach of children.
Danger of suffocation!
Disposal of the appliance
Appliances contain materials which can
be recycled. Please contact your local
authorities about recycling in your area.
Ensure that the machine presents no
danger to children while being stored
for disposal. See "Important Safety
Instructions".
15
All rights reserved / 1104
This paper has been bleached without the use of chlorine.
M.-Nr. 06 159 860 / 00
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.