Miele G 7893 Installation diagram

Installationsplan / Installation plan
Installatietekening Plano de instalación Asennusohje Plan d`installation Plano de instalação Installasjonsplan
Pianta di installazione Σχέδιο εγκατάστασης Installation plan
CDN
G 7893
Materialnr. / Mat. no. 9180780 Modification
/ Version 00 index Change no. / Alteration number A17814 Datum Zeichnung / Drawing Date 01.12.2007 Datum Legende / Legend Date 01.12.2007 Freigabe
/
Approval Name: Schröder Name: Götza
Installation plan G 7893 As per: 01.12.2012
Page 1
Technical Datasheet
Laboratory Glassware Washer
Legend:
Circled, bold-type abbreviations: Connection required
G 7893
EN
Abbreviations surrounded by broken circle: Connection optional or required, depending on model
Electrical
connection
Country variations:
Voltage
1. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
2. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
V/Hz kW A mm² m
V/Hz kW A mm² m
V/Hz kW A mm² m
convertible
3N AC 400/50
9.7 3 × 16 5 × 2.5
1.7
3N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
Voltage
Voltage
3. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
4. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead with plug
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
V/Hz kW A mm² m
V/Hz kW A mm² m
convertible
V/Hz
kW A mm² m
V/Hz kW A mm² m
convertible
3 AC 230/50 7,3 3 × 20 5 × 2,5 1,7
3 N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
3N AC 400/50 9,2 3 × 15/16 5 × 2.5
1.7 AC 230/50
6.7 1 × 30 3 × 4
2.1
Installation plan G 7893 As per: 01.12.2012
Page 2
5. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead with plug
V/Hz
kW A mm² m
AC 230/50
6.7 1 × 30 3 × 4
2.1
Voltage
Voltage
Voltage
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
6. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead with plug
Rated load Fuse rating Min. cable cross-section Length of supply lead with plug
7. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead with plug
Rated load Fuse rating Min. cable cross-section Length of supply lead with plug
V/Hz kW A mm² m
convertible
V/Hz
kW A mm² m
V/Hz kW A mm² m
convertible
V/Hz
kW A mm² m
V/Hz kW A mm² m
convertible
3N AC 400/50
9.2 3 × 15/16 5 × 2.5
1.7
3N AC 400-415/50
9.2-9.9 3 × 15 5 × 2.5
2.1 AC 230-240/50
6.7-7.2 1 × 30 3 × 6
2.1
AC 230-240/50
6.7-7.2 1 × 30 3 × 6
2.1 3N AC 400-415/50
9.2-9.9 3 × 15 5 × 2.5
2.1
Voltage
Voltage
8. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead without plug
9. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead without plug
10. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
V/Hz kW A AWG m
V/Hz kW A AWG m
convertible
V/Hz kW A AWG m
V/Hz kW A AWG m
convertible
V/Hz kW A mm² m
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
AC 208/60
6.0 2 × 30 3 × 10
1.9
AC 208/60
6.0 2 × 30 3 × 10
1.9
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
3 AC 200/50
5.7 3 × 20 5 × 2.5
2.3
Voltage
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
Installation plan G 7893 As per: 01.12.2012
Page 3
V/Hz kW A mm² m
convertible
AC 200/50 5,7 2 × 30 3 × 4
2.1
11. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 200/60
5.7 3 × 20 5 × 2.5
2.3
Voltage
Voltage
Voltage
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
12. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
13. Voltage (as supplied) Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
V/Hz kW A mm² m
convertible
V/Hz kW A mm² m
V/Hz kW A mm² m
convertible
V/Hz kW A mm² m
convertible
V/Hz kW A mm² m
AC 200/60
5.7 2 × 30 3 × 4
2.1
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
AC 220/60
4.7 1 × 25 3 × 4
2.1
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
Voltage
Voltage
Plug and socket connection recommended to facilitate
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
Rated load Fuse rating Min. cable cross-section Length of supply lead (H05(07)RN-F) without plug
electrical safety tests. Socket must be accessible after installation. On hard-wired machines, a multi-pole circuit breaker with a minimum contact gap of 3 mm must be provided on site. The use of an earth leakage circuit breaker (ELCB) is strongly recommended. If necessary, equipotential bonding with good galvanic contact must be provided in accordance with all appropriate national and local regulations.
V/Hz kW A mm² m
convertible
V/Hz kW A mm² m
convertible
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
AC 220/60
4.7 1 × 25 3 × 4
2.1
Installation plan G 7893 As per: 01.12.2012
Page 4