Installationsplan / Installation plan
Installatietekening Plano de instalación Asennusohje
Plan d`installation Plano de instalação Installasjonsplan
Pianta di installazione Σχέδιο εγκατάστασης Installationsplan
USA
G 7889
Materialnummer / Mat. no. 9510090
Änderungsstand / Version 00
Änderungsnr. / Alteration number A24364
Datum Zeichnung / Drawing Date 01.01.2013
Datum Legende / Legend Date 01.01.2013
Freigabe
/
Approval Name: Leifeld
Name: Götza
Installationsplan G 7889
Stand: 01.01.2013
Seite 1

Technical Data Sheet
Washer-Disinfector
Legend:
Highlighted circled abbreviations:
Connection required
G 7889
USA
Broken line circled abbreviations:
Connection optional or required depending on model version
Electrical
connection
Voltage
It is recommended to make electrical connection via
Standard voltage (at delivery)
Connected load
Fuse rating
Min. cross-sectional area of supply lead
Length of electrical cord without plug
Connected load
Fuse rating
Min. cross-sectional area of supply lead
Length of electrical cord without plug
a plug and socket so that electrical safety checks
can be carried out easily. The socket must be
accessible after machine installation.
With a fixed installation, this must be via a main
switch to be provided on site, which must totally
isolate the machine from the mains with a contact
gap of at least 3 mm.
The use of a grounded leakage circuit breaker
(ELCB) is strongly recommended.
If necessary, an equipotential bond with good contact
connection must be provided in accordance with all
appropriate national and local regulations.
AC 208/60 V/Hz
6.0 kW
2 × 30 A
3 × 10 AWG
6’
3 AC 208/60 V/Hz
6.0 kW
3 × 20 A
4 × 12 AWG
6’
Convertible
AC 208/60 V/Hz
6.0 kW
2 × 30 A
3 × 10 AWG
1.9 m
3 AC 208/60 V/Hz
6.0 kW
3 × 20 A
4 × 12 AWG
1.9 m
Interface (SST) Serial interface for output of process data provided as
Cold water Max. temperature
Installationsplan G 7889
Stand: 01.01.2013
Seite 2
standard.
A 9-pole sub D plug (male) is located on the rear of the
machine.
Max. water hardness
Min. flow pressure
Min. flow pressure with extended water intake
Max. pressure
Flow rate
On-site threaded union
Connection hose length (supplied)
68 °F
64 gpg
36 PSI
15 PSI
150 PSI
2.1 gal/min
11.5 NH
5’
20 °C
60 °dH
250 kPa
100 kPa
1000 kPa
8 l/min
¾” male thread
1500 mm

Demineralized
water
Min. flow pressure
Min. flow pressure with extended water intake
Max. pressure
Flow rate
On-site threaded union
Connection hose length (supplied)
If demineralised water is not available do not connect
the hose to cold or hot water.
36 PSI
15 PSI
150 PSI
2.1 gal/min
11.5 NH
5’
250 kPa
100 kPa
1000 kPa
8 l/min
¾” male thread
1500 mm
Drain water Drain hose (inner dia. × thickness × lenght)
Drainage
Cool-Down
(optional)
External
dispenser
(optional)
Machine data Height with lid
When fitting the machine in a housing unit, a suitable opening min. 5 7/8” x 4” (150 x 100 mm) must be provided for the supply
and drainage hoses.
If a new installation is being planned, the preferred niche depth is 27 9/16” (700 mm) in order to be able to lay the hoses
easily.
Installation should only be carried out by authorized installers and must comply with local and national building codes.
Observe installation instructions when installing this machine. Alteration rights reserved.
Max. drain pump head height from bottom edge of
machine
Max. brief flow rate per hose
Hose connector on site for drain hose
(outer dia. × lenght)
Max. temperature of drain water
Cool-Down Solution to reduce drain temperature to
140°F.
Please allow 7” additional space in the back of the
machine.
Connection for up to 2 dispensers (DOS K 60) on rear
panel
Height without lid
Width
Depth
Net weight
Floor load in operation
Min. installation location access width including
transport pallet
Min. installation location access height including
transport pallet
7/8“ × 1/4“ × 5‘
3‘
13.2 gal/min
7/8‘ × 1 3/16“
199°F
33.5”
32.3”
23.6”
23.6”
154 lb
1500 N
28”
38”
22 mm × 6 mm ×
1500 mm
1.0 m
50 l/min
22 mm × 30 mm
93 °C
850 mm
820 mm
598 mm
600 mm
70 kg
1500 N
700 mm
950 mm
Änderungsnr. Datum Bezeichnung:
A24364 01.01.2013 Einführung
Installationsplan G 7889
Stand: 01.01.2013
Seite 3