Miele G 7804 Installation diagram

Installationsplan / Installation plan
Installatietekening Plano de instalación Asennusohje Plan d`installation Plano de instalação Installasjonsplan
US/CDN
G 7804
Materialnummer
/
Mat. no.
9180761 Änderungsstand / Version 01 Änderungsnr. / Alteration number A031079 Datum Zeichnung / Drawing Date 01.05.2010 Datum Legende / Legend Date 01.05.2012
Installationsplan G 7804 Stand: 01.05.2012 Seite 1
Technical Data Sheet
Laboratory Glassware Washer
Legend:
Highlighted circled abbreviations: Connection required
G 7804
CDNEN
Broken line circled abbreviations: Connection optional or required depending on model version
Electrical
connection
Voltage
Voltage
Country variations:
1. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
Convertible
3N AC 400/50
9.7 3 × 16 5 × 2.5
1.7
2N AC 400/50
6.7 2 × 16 4 × 2.5
1.7
AC 230/50
3.7 1 × 16 3 × 2.5
1.7
Voltage
Voltage
2. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
Convertible
3N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
AC 230/50
5.1 1 × 25 3 × 4
2.1
Installationsplan G 7804 Stand: 01.05.2012 Seite 2
3. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 230/50
7.3 3 × 20 5 × 2.5
1.7
Voltage
Voltage
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
4. Voltage Connected load Fuse rating (integrated in plug) Min. cross-sectional area of supply lead Connection cable length with plug
5. Voltage Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length with plug
6. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
Convertible
V/Hz
kW A mm² m
V/Hz kW A mm² m
V/Hz kW A AWG m
3 N AC 400/50
7.3 3 × 16 5 × 2.5
1.7
AC 230/50
5.1 1 × 25 3 × 4
2.1
AC 230/50
2.5 1 × 13 1 × 1.5
2.1
AC 230-240/50
3.2 – 3.4 1 × 15 3 × 1.5
1.8
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
Voltage
Voltage
Voltage
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
7. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length without plug
8. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A AWG m
Convertible
V/Hz kW A AWG m
V/Hz kW A AWG m
Convertible
V/Hz kW A mm² m
V/Hz kW A mm² m
Convertible
AC 208/60
6.0 2 × 30 3 × 10
1.9
AC 208/60
6.0 2 × 30 3 × 10
1.9
3 AC 208/60
6.0 3 × 20 4 × 12
1.9
3 AC 200/50
5.7 3 × 20 5 × 2.5
2.3
AC 200/50
5.7 2 × 30 3 × 4
2.1
Installationsplan G 7804 Stand: 01.05.2012 Seite 3
9. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
3 AC 200/60
5.7 3 × 20 5 × 2.5
2.3
Voltage
Voltage
Voltage
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
10. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
11. Standard voltage (at delivery) Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
AC 200/60
5.7 2 × 30 3 × 4
2.1
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
AC 220/60
4.7 1 × 25 3 × 4
2.1
3 AC 220/60
6.7 3 × 20 5 × 2.5
2.3
Voltage
Voltage
It is recommended to make electrical connection via a plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
Connected load Fuse rating Min. cross-sectional area of supply lead Connection cable length (H05(07)RN-F) without plug
and socket so that electrical safety checks can be carried out easily. The socket must be accessible after machine installation. With a fixed installation, this must be via a main switch to be provided on site, which must totally isolate the machine from the mains with a contact gap of at least 3 mm. The fitting of an earth leakage circuit breaker is strongly recommended. If necessary, an equipotential bond with good contact connection must be provided in accordance with all appropriate national and local regulations.
V/Hz kW A mm² m
Convertible
V/Hz kW A mm² m
Convertible
3N AC 380/60
6.7 3 × 16 5 × 2.5
2.3
AC 220/60
4.7 1 × 25 3 × 4
2.1
Installationsplan G 7804 Stand: 01.05.2012 Seite 4
Cold water or hot water
Max. temperature Max. water hardness Min. flow pressure Max. pressure Flow rate On-site threaded union
Connection hose length (supplied)
Machine is DVGW certified. Direct connection to mains water supply without special safety stopcock permitted.
Water connection Australia: To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
°C °dH kPa kPa l/min inch
mm
70 60 250 1000 8 ¾ male thread (USA: 11.5 NH) 1500
Demineralised water
Drain water Drain hose (inner dia. × thickness × l)
External dispenser (optional)
Machine data Height including lid
Max. temperature Standard version: Min. flow pressure Min. flow pressure with extended water intake Max. pressure Flow rate Variant with demineralised water pump (aqua destillata pump) (ADP): Min. flow pressure Max. pressure On-site threaded union
Connection hose length (supplied)
Water connection Australia: To prevent back-syphonage, the water connection must be via an approved Dual-Check-Valve supplied with the machine.
Max. drain pump head height from bottom edge of machine Max. brief flow rate per hose Hose connector on site for drain hose (outer dia. × l)
Connection for one of each of dispensers DOS G 20 and DOS G 60 on rear panel
Height without lid Width Depth Net weight Floor load in operation Min. installation location access width including transport pallet Min. installation location access height including transport pallet
°C
kPa kPa kPa l/min
kPa kPa inch
mm
mm m l/min mm
mm mm mm mm kg N mm
mm
70
150 100 1000 8
8.5 50 ¾ male thread (USA: 11.5 NH) 1500
22 × 6 × 1500
1.0 50 22 × 30
850 820 598 600 70 1500 700
950
When using plinths (UM 30-60 and/or UE 30-60) these must be secured to the floor. It is essential to fasten the machine to the plinth. The electrical, water, drain and exhaust connections must be made in accordance with all appropriate local and national safety regulations and applicable standards. During installation it is essential to follow the installation instructions provided! Modification rights reserved! Dimensions in mm.
Installationsplan G 7804 Stand: 01.05.2012 Seite 5