Teachware
Metrohm AG
CH-9100 Herisau
teachware@metrohm.com
This documentation is protected by copyright. All rights reserved.
Although all the information given in this documentation has been
checked with great care, errors cannot be entirely excluded. Should you
notice any mistakes please send us your comments using the address
given above.
Documentation in additional languages can be found on
The NH3-selective gas membrane electrodes are combined electrodes, i.e.
they contain one measuring electrode and one reference electrode.
They enable the rapid, simple, inexpensive and precise determination of
dissolved ammonia (NH3) in aqueous systems, e.g. in natural water, sewage, boiler feed water, beer, etc.
In addition, the NH3-selective gas membrane electrodes can detect the following substances:
■Ammonium ions (NH
■Organic nitrogen (N) after a Kjeldahl digestion
Discolored or turbid samples do not compromise the measurement, which
means that, generally speaking, no pre-distillation is necessary.
The membrane of the NH3-selective electrode is gas-permeable and
water-repellent, i.e. water cannot dampen the membrane and cannot permeate the pores. In the case of samples that contain surface-active substances, or in the case of nonaqueous systems, fluid makes its way into
the membrane. This leads to difficulties, e.g. in the case of sewage samples (detergents) or with nylon or dye samples (nonaqueous systems). An
ammonia determination of such samples should therefore be carried out
by suspending the electrode above the solution instead of immersing it.
+
) after their conversion into ammonia
4
1 Introduction
NH3-selective electrodes
■■■■■■■■
1
2 Mode of operation
Ammonia (NH3) reacts in water with the formation of ammonium ions
+
(NH
):
4
The equilibrium of this reaction, i.e. the relative proportion of ammonia
and ammonium ions, is determined by the pH value of the solution. In
acidic solutions, ammonia reacts almost completely to ammonium ions. At
a pH value of 9.2, the ratio of ammonia to ammonium ions is approximately 1 : 1. If the pH value of the solution is known, then the ratio of
ammonia to ammonium ions can also be calculated directly. The following
applies:
■■■■■■■■■■■■■■■■■■■■■■
(1)
(2)
The value of the equilibrium constant K is also dependent on the temperature.
Table 1
Temperature dependency of the ammonia – ammonium ions
equilibrium
Temperature in °C– log(K)
54.13
254.78
355.08
The addition of an excessive amount of a strong base, e.g. concentrated
caustic soda (NaOH), causes the ammonium ions to be transformed into
ammonia in their entirety. Determination of the ammonium concentration
is thus indirectly possible by means of the determination of ammonia. In
the case of the NH3-selective electrode, a gas-permeable, hydrophobic
membrane separates the measuring solution from the measuring electrolyte. Ammonia now diffuses through the membrane for as long as the
partial pressure on the two sides is identical in size. The partial pressure of
ammonia is always proportional to its concentration. The measuring electrolyte is comprised of a sufficiently concentrated ammonium chloride
solution, which means that the ammonium concentration can be assumed
to be constant.
■■■■■■■■
2
The potential U of the measuring electrode is proportional to the hydroxide concentration of the measuring electrolyte:
NH3-selective electrodes
Loading...
+ 17 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.