The information contained in this publication is derived in part from proprietary data and trade secrets.
This information has been prepared for the expressed purpose of assisting operating and
maintenance personnel in the efficient use of the instrument described herein. Publication of this
information does not convey any rights to use or reproduce it or to use for any purpose other than in
connection with the installation, operation and maintenance of the equipment described herein.
Copyright 2008
Printed in the USA. All Rights Reserved.
SAFETY INSTRUCTIONS
The following instructions must be observed.
• Every effort has been made to design and manufacture this instrument to be safe for its intended
use. A hazardous situation may occur if this instrument is not used for its intended purpose or is
used incorrectly. Please note operating instructions provided in this manual.
• The instrument must be installed, operated and maintained by personnel who have been properly
trained. Personnel must read and understand this manual prior to installation and operation of the
instrument.
• An auto-resettabl e fuse internally protects this instrument. To reset the fuse, remove all power
from the unit for one minute.
• The manufacturer assumes no liability for damage caused by incorrect use of the instrument or for
modifications or changes made to the instrument.
Technical Improvements
Liquid Controls Sponsler, Inc. may modify the technical data herein without notice.
Warning: Do not open the enclosure when an explosive gas atmosphere is present.
Power .............................................................................................................................................. 13
Appendix 1: Menu Quick Reference ................................................................................................ 25
Appendix 2: Coefficient of Thermal Expansion for Common Fluids ................................................. 26
Appendix 3: Menu Flow Chart ......................................................................................................... 27
Appendix 4: Manual Revision History .............................................................................................. 34
Liquid Controls
Sponsler, Inc.
IT400 Remote Totalizer & Rate Indicator
Page: 1 DOC#: MN-IT400-R2d.doc
Description
General Unit Operation
The IT400 Remote Totalizer and Rate Indicator is a microcontroller based flow instrument capable of
translating flow in format ion and conditions to the built-in display and various outputs.
Features
• Pulse input supports turbine as well as many other pulse-type flowm et ers
• All features/configuration settings are available via field programmability
(Max values at 25°C. Max combined Pulse/Alarm
output power: 190mW @ 25°C, 32mW @) 65°C)
• Rate (4-20mA)
• Factored rate to current loop output
• Programmable low and high
• Fully isolated (unless loop powered)
• Loop Voltage limit: 7-36VDC
Accuracy
• Display: ±0.01% reading (rate) or ± 1 count (total)
• Analog output: ±0.025% of fs @20°C
• Digital output: ±1 count
Environmental
• Operating: -30 to 65°C (-22 to 150°F)
• Storage: -40 to 85°C (-40 to 185°F)
Enclosure
• Explosion Proof Aluminum
• FM Approved, CSA Certified
• Class I, Division1, Groups B, C, & D
• Class II, Division 1, Groups E, F, & G
• NEMA 4X, IP66
• All openings ¾” FNPT
• Weight: 2.5lbs.
• Mount: Directly on flowmeter or Wall mount
Compliance
• CE 0575
• ATEX Ex II 2 G Ex d IIC T6 Gb (X and WX
enclosure types only)
•DNV-2008-OSL-ATEX-19241X
Communications
• RS-232
Datalogger
• Up to four triggers (including timer)
• Up to seven system values
• Real-time clock
• RS-232 or Internal SD/MMC Flash media card
Other features
• EEPROM setup storage (>100 year retention)
• All features/settings are field programmable.
Convenient number input via 0 -9 rotary switch and
two pushbuttons
Liquid Controls
Sponsler, Inc.
IT400 Remote Totalizer & Rate Indicator
Page: 4 DOC#: MN-IT400-R2d.doc
Wiring and Hookup Diagrams
The IT400 has many connections all made through the bottom/mounting board. It is
recommended to remove the battery when connecting wires to the connectors and when plugging
the connectors into their sockets.
Wiring
In order to maintain the rated CE marking, the following practices must be followed:
• All wires connected to the IT400 must be shielded (exception: SI supplied signal cable). o All shields must be connected to one of the internal mounting screws to achieve their
ground connection.
o It is recommended that the non-IT400 end of these shields is left unconnected. This
will help avoid ground loops and unexpected behavior.
o It is recommended that only one spade type terminal be used on each of the screws.
Care should be taken to avoid shorting any spade with other circuitry on the mounting
board. If a longer screw is needed to attach the shielding spade, then the screw
must be chosen to ensure at least one extra thread of space behind the screw. The
hole depth is approximately 0.250 inches.
o Shields should not cover more than one pair of signal wires at a time.
• The IT400’s chassis must be connected to an earth ground via the screw attached to the
outside of the unit. At least a 12 AWG preferably green wire is recommended.
• All connections are to be made to the unit via rigid conduit.
• All connected wires must not exceed 98’ length.
• DC powered devices requiring the CE marking must be 4-20mA loop powered.
If no wires are required for a particular installation; the only requirement is the chassis ground
connection.
Battery Removal and Installation
Battery Replacement:
- E8010 from Liquid Control s Sponsler, Inc.
- TL-5930/F from Tadiran Batteries, Ltd.
Battery Removal:
reset the totalizer appears, the totalizer/grand totalizer/service timer will be saved to memory.
Using two fingers inserted in the slots on the top and bottom of the main board overlay, gently rock
the board top to bottom while pulling straight out. If this is done too quickly or unevenly, the main
board will be damaged. After the main board is removed, the battery is accessible by opening the
hook-loop strap. Once the battery is loose, the battery clip is removed by compressing the locking
tab on the connector (locking tab is on the top of the connector {J3}) and pulling it out.
Battery installation: Reverse the pr ocess for installation. Ensure the battery connector key is in the
proper orientation pri or to inser ti on.
The internal battery may only be replaced with one of two batteries:
Unscrew the IT400 cover, press and hold the menu button until the option to
Liquid Controls
Sponsler, Inc.
Diagrams:
IT400 Remote Totalizer & Rate Indicator
Page: 5 DOC#: MN-IT400-R2d.doc
Figure 1 IT400 Connections
Liquid Controls
Sponsler, Inc.
IT400 Remote Totalizer & Rate Indicator
Page: 6 DOC#: MN-IT400-R2d.doc
Grand
TB1
J3
1
2
TB6
1
J1
TB5
4
3
2
1
TB4
4
3
2
1
TB3
3
2
1
TB5
1
2
1
2
1
TB2
JP3
Grand
LIQUID CONTROLS GROUP
SPONSLER
Precision Turbine Flowmeters
A UNIT OF
IDEX
CORPORATION
2363 SANDIFER BLVD. WESTMINSTER, SC 29693
1.864.647.2065 WWW.SPONSLER.COM
I
T
4
0
0
b
e
r
m
u
N
LIQUID CONTROLS GROUP
SPONSLER
2363 SANDIFER BLVD. WESTMINSTER, SC 29693
I
T
4
0
0
Precision Turbine Flowmeters
A UNIT OF
IDEX
CORPORATION
1.864.647.2065 WWW.SPONSLER.COM
b
e
r
m
u
N
Figure 2 IT400 Board Stack Assembly and Battery Removal
Liquid Controls
Sponsler, Inc.
IT400 Remote Totalizer & Rate Indicator
Grand
I
T
4
0
0
r
e
b
m
u
N
2363 SANDIFER BLVD. WESTMINSTER, SC 29693
1.864.647.2065 WWW.SPONSLER.COM
LIQUID CONTROLS GROUP
SPONSLER
Precision Turbine Flowmeters
A UNIT OF
IDEX
CORPORATION
Page: 7 DOC#: MN-IT400-R2d.doc
Figure 3 Meter Mounting and Dimensional
Liquid Controls
Sponsler, Inc.
IT400 Remote Totalizer & Rate Indicator
Page: 8 DOC#: MN-IT400-R2d.doc
IT400 Operational Overview
The IT400 flowmeter system will perform the necessary calculations to provide the user with an
optional temperature compensated and optional linearized output of the following values:
• Rate
• Total
• Grand total
• Errors/warnings
• Various outputs
Calibration
Calibration of the IT400 is accomplished using either a linear K-Factor (pulses per gallon) or a
linearized K-Factor linearization table (when the linearizer is enabled). Independent unit conversion
factors and decimal points are available for the rate and total displays. Calibration and configuration
data are stored in an onboard EEPROM for permanent storage. Calibration of the IT400’s reference
circuitry is factory set and is not user accessible. Calibration of the IT400’s reference circuitry should
be performed by factory trained service personnel.
Temperature Compensation
The temperature compensator utilizes a platinum RTD to detect the process temperature. Process
temperature is used in conjunction with the reference temperature and Coefficient of Thermal
Expansion of the measured liquid (menu items) to adjust the rate and total values. When the
temperature compensation is enabled, the temperature warning annunciator (thermometer on the
LCD) will indicate when the process temperature is out of a programmable range. Optionally, an
alarm output as well as a totalizer inhibit function can also be tied to this temperature warning.
Various Coefficient of Thermal Expansions and their reference temperatures can be found in the
appendix. The temperature circuit and associated algorithm, when enabled, increase the IT400
current consumption fr om the power source and diminish battery life if running solely from the battery.
A programmable delay function has been provided in order to minimize the impact of the temperature
sampling on the battery life. The delay feature specifies the number of seconds between temperature
value samples. The higher the delay, the longer the battery will last.
Linearization
The linearizer utilizes a linear interpolation algorithm to calculate the rate and total based on a set of
calibration data points pr ogrammed by the user. The linearizer has21 points, the first point (lowest) is
automatically set at zero hertz, the remaining 2 to 20 are user programmable. The linearizer table
must have at least two data points in addition to the first (fixed) point and must be ascending in
frequency from the first point to the last. To have fewer than 20 points in the linearizer table, after the
last desired linearization point add one additional point and set the frequency at zero. Zero is an
invalid frequency; this will indicate the end of the table. The IT400 will not allow frequencies to be
entered out numerical order. Frequencies between zero and the first programmed point will use the
K-Factor of the first programmed point.
Liquid Controls
Sponsler, Inc.
8 digit
IT400 Remote Totalizer & Rate Indicator
Page: 9 DOC#: MN-IT400-R2d.doc
LCD Display
The LCD display has three main areas: The 8 digit totalizer,
5 digit rate indicator, and the annunciators.
• The 8-digit totalizer shows the total number of units
of volume through put of the flowmeter since the last
time the totalizer was reset. The totalizer is
resettable by 2 methods, via a magnetic reed sw it ch
in the lower left side of the unit and by menu. Th e
grand total is displayed as long as a magnet is
placed near the top of the face of the unit. The word
"GRAND" indicates the displayed value is the grand
total. Grand should be activated prior to battery
replacement to save total/grand total values to nonvolatile memory. Engineering unit annunciators
display the units and decimal place selected to compute total and grand total values. The
grand totalizer uses the same engineering unit as programmed for the totalizer.
• The 5-digit rate display shows the rate of the flowmeter throughput. If the rate exceeds the
high flow warning setpoint, "-oL-" is displayed. “-UL-“ is displayed when the rate is less than
the low flow warning setpoint. The warning is displayed until flowrate within these setpoints
resumes. The totalizer is not affected by this warning unless programmed to inhibit totalizer
when out of range. The range limits are established by the flow warning setpoints (Warning
menu) in the menu system. Indicators on the display also show the engineering units of
volume, time and decimal place used to compute the rate value.
Note: When the Grand Total magnetic reed switch is activated (and GRAND is displayed on the
screen), the rate display will display the current rate regardless of the range limits. This is
helpful when troubleshooting flow condition problems when the low rate setpoint is set above
zero.
• Refresh rate of the display of the unit happens at pre-determined intervals depending on the
selected power mode (see “Power”). Once the power mode is programmed (in the menu), the
refresh rate is fixed. This prevents a low refresh rate from being automatically selected which
could cause measurement delay.
• Other display annunciators. These are warning indicators and do not inhibit the operation of
the unit (unless explicitly enabled to do so by the alarm opti o n). These indicators may
however indicate the condition causing an inaccurate reporting of rate or total information.
o Wrench – The service annunciator is user programmable to a specific amount of
“turbine hours” (activ e flowmeter hours).
o Thermometer – The thermometer annunciator is displayed continuously any time the
temperature compensator is enabled and the process temperature is out of range.
o Battery – The battery annunciator is composed of two segments. The segments
indicate the level of the battery charge, like an automotive gas gauge.
Two segments – The battery is within normal operating voltage range.
One segment – A replacement battery should be acquired and installed. This will
also cause high temperature range on the temperature circuit to be reduced.
Total
5 digit
Rate
Loading...
+ 28 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.