KROHNE PAC-050 User Manual

Page 1
07/2008
Installation and Operating Instructions
OPTISENS PAC 050
Measuring and control device for pH and ORP measurements
Page 2
Manual OPTISENS PAC 050
Content
0. Introduction......................................................................................................................................................................................... 4
0.1 General..................................................................................................................................................................................................................4
0.2 Legal matters.........................................................................................................................................................................................................4
0.3 Safety..................................................................................................................................................................................................................... 5
0.3.1 Documentation symbols......................................................................................................................................................................................... 5
0.4 Features................................................................................................................................................................................................................. 6
0.4.1 Device.................................................................................................................................................................................................................... 6
0.4.2 Controller................................................................................................................................................................................................................ 6
0.4.3 Connections........................................................................................................................................................................................................... 6
1. Mechanical installation......................................................................................................................................................................7
1.1 Installation of panel-mounting converters.............................................................................................................................................................. 7
1.2 Installation of wall-mounting converters.................................................................................................................................................................8
2. Electrical connection..........................................................................................................................................................................9
2.1 Connection diagram panel-mounting converter................................................................................................................................................... 10
2.2 Connection diagram wall-mounting enclosure..................................................................................................................................................... 11
3. Operation of the device.................................................................................................................................................................... 12
3.1 How to adjust parameters.................................................................................................................................................................................... 12
3.1.1 Selection between alternatives............................................................................................................................................................................ 12
3.1.2 Adjustment of numerical parameters...................................................................................................................................................................13
3.2 Menu Overview.................................................................................................................................................................................................... 13
3.2.1 Main menu and basic settings .............................................................................................................................................................................13
3.3 Password and language ......................................................................................................................................................................................14
3.3.1 Enter password.................................................................................................................................................................................................... 14
3.3.2 Language............................................................................................................................................................................................................. 14
4. Adjustments for the measurement.................................................................................................................................................14
4.1.1 Measurement....................................................................................................................................................................................................... 15
4.2 pH measurement.................................................................................................................................................................................................. 15
4.3 ORP measurement.............................................................................................................................................................................................. 15
4.4 Calibration............................................................................................................................................................................................................ 16
4.4.1 Calibration procedure........................................................................................................................................................................................... 16
4.4.2 Buffers..................................................................................................................................................................................................................16
4.5 Temperature compensation................................................................................................................................................................................. 17
4.5.1 Calibration of the temperature measurement ...................................................................................................................................................... 17
5. Adjustments of the controller..........................................................................................................................................................17
5.1 ON/OFF controller................................................................................................................................................................................................ 18
5.2 P / PI controller as impulse-frequency controller .................................................................................................................................................19
5.3 P / PI controller as pulse-pause controller...........................................................................................................................................................19
5.4 Activation and deactivation of the controller ........................................................................................................................................................20
5.5 Turn-on delay....................................................................................................................................................................................................... 20
5.6 External controller stop........................................................................................................................................................................................20
5.7 Manual operation of the relays.............................................................................................................................................................................21
5.8 Limit values.......................................................................................................................................................................................................... 22
5.8.1 Turn-on delay....................................................................................................................................................................................................... 22
5.9 Dosage check......................................................................................................................................................................................................22
6. Alarm..................................................................................................................................................................................................23
6.1 Error messages.................................................................................................................................................................................................... 24
7. Output................................................................................................................................................................................................24
7.1 Current output...................................................................................................................................................................................................... 25
7.2 Serial interface RS485 (option)............................................................................................................................................................................ 25
8. Operation and maintenance............................................................................................................................................................25
8.1 Maintenance of the device................................................................................................................................................................................... 25
8.2 Display contrast.................................................................................................................................................................................................... 25
8.3 Exchange fuse.....................................................................................................................................................................................................25
8.4 Cleaning............................................................................................................................................................................................................... 25
8.5 Maintenance of the pH or ORP measurement.....................................................................................................................................................26
8.6 Disposal...............................................................................................................................................................................................................26
8.7 Service................................................................................................................................................................................................................. 26
8.7.1 Product info.......................................................................................................................................................................................................... 26
8.7.2 Analog inputs.......................................................................................................................................................................................................26
8.7.3 Erase settings (reset)...........................................................................................................................................................................................26
OPTISENS PAC 050 2
Page 3
Manual OPTISENS PAC 050
9. Technical data................................................................................................................................................................................... 27
9.1 Technical data......................................................................................................................................................................................................27
9.2 Dimensions..........................................................................................................................................................................................................28
9.2.1 Panel-mounting enclosure...................................................................................................................................................................................28
9.2.2 Wall-mounting enclosure ..................................................................................................................................................................................... 28
10. Device return form............................................................................................................................................................................ 29
11. Customer settings - for reference...................................................................................................................................................30
3 OPTISENS PAC 050
Page 4
Manual OPTISENS PAC 050
0. Introduction
0.1 General This manual applies to the following devices:
It contains technical information for the installation, start-up and maintenance. If you have any questions not
0.2 Legal matters
Device and type Revision date
AC 050 pH/ORP P 06 / 05 AC 050 pH/ORP W 06 / 05
included in this manual please contact your supplier or the official representative of KROHNE Water Solutions in your country.
Authorized personnel
Installation, connection, adjustment, start-up, and maintenance of the device are carried out by authorized personnel with adequate qualification.
Liability
Responsibility as to suitability and intended use of these devices rests solely with the user. Improper installation and operation may lead to loss of warranty. In addition, KROHNE Group's Standard General Conditions of Sale and Delivery, found on the back of the invoice and forming the basis of the purchasing contract, are applicable.
General limitation on liability
Unless otherwise expressly set forth in the Standard Terms and Conditions of Sale and Delivery, the Seller is only liable for damages, whatever their legal basis is, in case they are based on willful action or gross negligence. This limitation on liability does not apply in the event the Buyer raises claims relating to personal injury or damages to property according to the product liability law based on a defect of the delivered goods.
Any advice given by the Seller, in particular regarding the application of the delivered goods, shall only commit the Seller if given or confirmed in writing.
Returning the device
If you need to return the level gauge to the manufacturer or supplier, please read to the instructions and complete the form given in the appendix.
Warranty
Please consult KROHNE Water Solutions General Terms and Conditions for information on guarantee and liability.
OPTISENS PAC 050 4
Page 5
Manual OPTISENS PAC 050
0.3 Safety Please check for damages immediately after receiving the devices and report any damages within 24 hours to
the delivering company. Never work with a damaged device. Keep this manual at a safe place where you can always look up the safety instructions and the information on
handling and usage. This device was designed and built according to the safety measurements for electronic devices and has left
our company in perfect working condition. To preserve this condition and to ensure safe usage follow all instructions carefully and pay special attention to all warnings issued in this manual. If the device is visibly damaged or has been stored inappropriately or if there are any doubts concerning safe usage, shut it down and make sure it cannot be restarted by accident.
0.3.1 Documentation symbols
A set of symbols is used to give warnings or information relevant to particular applications. These are defined below:
Caution / Attention
Information that, if not followed, may lead to actions resulting in incorrect functioning of the device.
Warning
Information that, if not followed, may lead to actions resulting in measurement error, personal injury and/or
NOTE
damage to the device.
Is used to highlight interesting details.
5 OPTISENS PAC 050
Page 6
0.4 Features
0.4.1 Device Measuring ranges
0.4.2 Controller
0.4.3 Connections
Temperature measuring ranges -30.0…+140.0°C Display measured value with dimension
Temperature compensation manually or automatically with Pt100 or Pt1000 Calibration automatic recognition of the calibration solutions in arbitrary order
Set points 2 set points with adjustable direction Controller types ON/OFF controller with hysteresis
Hysteresis 0.00…14.00 pH or 0…1500 mV P range X Integral time T
Least pulse 0.1…9.9 s Pulse+ pause time 02…99 s Impulse frequency 00…72 equiv. to 0...7200 pulses/h Turn-on delay 0…200 s Alarm function min. and max. limit and onset delay, power failure alarm,
Dosage control 0…2000 s
Relays 3 potential-free contacts (2x controller, 1x alarm) Analog output 0/4…20 mA galvanically isolated Analog inputs 1 measuring input for pH or ORP Digital input external controller stop or lack-of-water indication
Serial interface (Option) RS 485, Baud rate 9600, data format 8 Bit,
0.00…14.00 pH or 0…1500 mV
P
N
Manual OPTISENS PAC 050
0.00…14.00 pH
-1500…+1500 mV
temperature with dimension status display sensor, calibration, controller & alarm
P controller as Pulse-Pause-, Impulse-Frequency- or steady controller PI controller as Pulse-Pause-, Impulse-Frequency- or steady controller
0…2000 s
impulse or permanent signal
6 A, 250 V, max. 550 VA max. burden 500 Ohm 1 measuring input for Pt100 or Pt1000
1 start and 1 stop bit, no parity
OPTISENS PAC 050 6
Page 7
1. Mechanical installation
On chapter 9.2 you will find detailed instructions for the installation.
For panel-type meters you have to prepare an opening of 92 x 92 mm / 3.6” x 3.6”. Install the device and fix it with the two mounting clips which were part of the delivery.
You can install devices in wall-mounting enclosures either by hanging them upon the center slot or by sliding the slot under a screw, which is an alternative for limited space. Either way you have to fix it additionally with two screws.
Install the device in a place where it is not put under mechanical or chemical strain!
ATTENTION
1.1 Installation of panel-mounting converters
Mind the protection class:
Panel mounting version: IP 54 (front), IP 30 (Housing)
Wall mounting version: IP 65 (closed terminal cover)
Prepare an opening of 92 x 92 mm / 3.6” x 3.6”.
Manual OPTISENS PAC 050
Install the device from the front side and fix it with the two mounting clips which were part of the delivery.
Screw tight until the device is fixed perfectly.
7 OPTISENS PAC 050
Page 8
1.2 Installation of wall-mounting converters
Manual OPTISENS PAC 050
Unscrew the terminal cover.
Drill three holes (max. M5) according to the drawing. Mind that there are two ways for installation: (1) You can hang the device upon the upper screw. In that case drill the upper hole 120mm / 4.7” above the lower two. (2) Or you can slip the fixture on the back of the device under the upper screw. In that case the upper hole has to be another 15mm / 0.6” higher.
Insert the upper screw and make sure to leave at least 3 mm between wall and screw head for the fixture.
Mount the device and fix it with the two lower screws. Close the terminal cover or start with the connections.
OPTISENS PAC 050 8
Page 9
2. Electrical connection
ATTENTION
You will find a detailed connection diagram on the following pages. Before connecting the power supply check the information on the nameplate of the device! Input, output and control lines must be installed separate from each other and separate from power lines!
For inputs and outputs use screened lines and connect the screen on one side only. The pH measurement is interference-sensitive. Use a special screened cable and over long distances or in
humid atmosphere additionally an impedance converter. When using the relays, mind that with inductive loads, interference must be suppressed. If that is not
possible, the relay must be protected at the terminal block in the converter by a resistance-capacitance filter or, in case of direct current, by a free- wheeling diode.
Manual OPTISENS PAC 050
Current up to Capacitor C Resistance R
60 mA 10 nF 260 V 390 Ohm 2 Watt 70 mA 47 nF 260 V 22 Ohm 2 Watt 150 mA 100 nF 260 V 47 Ohm 2 Watt
1.0 A 220 nF 260 V 47 Ohm 2 Watt
9 OPTISENS PAC 050
Page 10
2.1 Connection diagram panel-mounting converter
Manual OPTISENS PAC 050
Connection Terminals Notes
1 = reference electrode = screen pH electrode 1 - 2 2 = measuring electrode = core 1 = measuring electrode = core ORP electrode 1 - 2 2 = reference electrode = screen
Impedance converter 1 - 4
Pt 100 5 + 6 Analog output 9 + 10 9 = +, 10 = - max. burden 500 Ohm Digital input 11 + 12 11 = +, 12 = - External controller stop or lack of water ind. Relay 1 14 + 15 Relay 2 16 + 17 Relay 3 18 + 19 Alarm relay Power supply 21+ 22+23 check information given on instrument label RS485 (option) Sub-D
Sub-D 4/7
Measurement = brown & white 1 = brown, 2 = white Supplied voltage = yellow & green 3 = yellow = -6V, 4 = green = +6V
3 = +, 8 = ­bridged = terminating impedance activated
OPTISENS PAC 050 10
Page 11
2.2 Connection diagram wall-mounting enclosure
Manual OPTISENS PAC 050
Connection Terminals Notes
1 = reference electrode = screen pH electrode 1 - 2 2 = measuring electrode = core 1 = measuring electrode = core ORP electrode 1 - 2 2 = reference electrode = screen
Impedance converter 1 - 4
Pt 100 6 - 7 Display contrast Display Potentiometer to adjust the display contrast
Analog output 11 + 12 11 = +, 12 = - max. burden 500 Ohm Relay 1 14 + 15 Relay 2 16 + 17 Relay 3 18 + 19 Alarm relay Power supply 20+21+22 check information given on instrument label RS 485 (option) 23 + 24 23 = 1, 24 = + Digital input 26 + 27 26 = +, 27 = -, external controller stop or lack-of water indication
Measurement = brown & white 1 = brown, 2 = white Supplied voltage, yellow & green 3 = yellow = -6V, 4 = green = +6V
11 OPTISENS PAC 050
Page 12
3. Operation of the device
3.1 How to adjust parameters
3.1.1 Selection between alternatives
When turned on the device shows the measured values of conductivity and temperature together with the controller mode (Man) and the status of the relays S1 and S2 (both OFF).
With five membrane keys you can move within the menu:
With key you enter the main menu.
With keys and you move up and down in the menu.
With key you address a menu or parameter.
With key you leave a menu without storing
With key “Enter“, changes are stored.
For your convenience triangles in the display indicate the directions you can take from your position in the menu.
Temp. comp. Manual.
Temp. comp. Auto
Temp. comp. Manual.
For many parameters you have the choice between two or more alternatives, for e. g. between manual and automatic temperature compensation. For these parameters you need only key. Switch from one alternative to the next until you either come back to where you started or until you reach the alternative you were looking for.
With these parameters any changes are immediately valid - there is no need to store the change.
Manual OPTISENS PAC 050
1 Measured value 2 Status relay 1 3 Status relay 2
F relay OFF J relay ON
4 Temperature 5 Controller AUTO: controller ON MAN : controller OFF (manual operation of the relays) 6 Orientation aids 7 Key left () 8 Key up () 9 Key down () 10 Key right () 11 Key “Enter“
1) When you address a parameter the actual setting is displayed.
2) Switch to the next alternative setting with key.
3) When you have come to the last alternative, pressing the key once more will bring you back to the start.
OPTISENS PAC 050 12
Page 13
Manual OPTISENS PAC 050
3.1.2 Adjustment of numerical parameters Numerical parameters can only be altered when a double arrow is visible behind the number. This double
arrow appears when you address the parameter with key. Adjust the parameter with keys and . A short pressure on the key changes the last decimal by 1. If you
keep the key pressed, the value will continue changing until the pressure is released.
Store the changes with key „Enter“. The double triangle disappears.
Enter password 058
Enter password 058
Enter password 062
3.2 Menu Overview
6.98 pH 25.0°C
F S1 F S2 Auto
Main menu
Calibrate Temp. comp. Temperature compensation (only for pH measurements)
Enter password Password function Set points Controller settings - set points, P ranges, I functions Limit values
Basic settings Service Contr. Settings
Turn-on delay Analog outputs Language
Bus address pH/ORP
3.2.1 Main menu and basic settings
The parameters are sorted into two menus: In the main menu you will find all functions which are used regularly. The menu basic settings contain all parameters which are set just once during start-up.
On the following pages you will find information on how to adjust parameters and which parameters you need for which application.
The information follows the following scheme:
1) General adjustments: password and language
2) Adjustments for measurement: calibration and temperature compensation for pH or selection of ORP measurement
3) Adjustments of the controller: selection of the controller version and corresponding parameters
4) Adjustments to read out data: analog, digitally and/or as alarm
1) Address the parameter with key .
2) A double arrow appears behind the number indicating that the number can be changed now with keys and .
3) Store the new value with key „Enter“. The double arrow disappears ­the new value is stored.
Display of the measured values (example)
Press key to enter the main menu.
Calibration function (only for pH measurements)
Alarm function
Basic settings
Cal. Pt 100 Buffers used
All parameters which are set just once at the beginning can be found in the menu “basic settings“
13 OPTISENS PAC 050
Page 14
3.3 Password and language
3.3.1 Enter password
Main menu
. .
Enter password 056 Code
To get access to the various parameters you have to enter the correct password:
Code 11 gives access to the parameters „calibration“, „temperature compensation“, and „set points“.
Code 86 gives access to all parameters and functions.
With any other number it is impossible to select, view or change any parameter.
Main menu
. .
Basic settings
3.3.2 Language
For the communication with the device you can choose from a variety of languages. Since choosing a language is part of the basic settings, it requires code 86. If a different code is set, you will
be asked to enter the correct password.
4. Adjustments for the measurement
Main menu
Calibrate Calibration of the pH measurement, display of slope and zero-point. Temp. comp.
. . .
Basic settings Buffers used
The instruments can be used for pH or ORP measurements.
Manual OPTISENS PAC 050
Enter password
Basic settings
.
.
Language
Selection between automatic and manual temperature compensation, adjustment of temperature for manual compensation.
Basic settings
Cal. Temp. Calibration of the temperature measurement.
Language
We speak English
Adjustment of the calibration solutions and the inner buffer of the electrode.
OPTISENS PAC 050 14
Page 15
4.1.1 Measurement
4.2 pH measurement
4.3 ORP measurement
NOTE
Manual OPTISENS PAC 050
Configuration
pH/Redox Measurement
Displayed is always the actual active configuration. With keyyou can switch from pH measurement to ORP measurement. Press the key again to go back to the original setting.
pH measurement requires a calibration. By means of calibration the electrode’s characteristics are determined and stored. The procedure is a
subsequent measurement of two solutions with a known pH. The pH values of the buffers used as calibration solutions are stored in the menu basic settings.
You have to adjust the calibration data only if you want to use special calibration solutions or electrodes with a special buffer solution.
The electrode’s slope is influenced by temperature. This influence can be compensated by manual or automatic compensation. Manual compensation means that the temperature is entered manually. Automatic compensation requires measuring the temperature with a Pt100 or Pt1000.
ORP measurements do not require calibration or temperature compensation. Therefore the parameters „calibration“ and „temperature compensation“ do not appear in the menu if ORP measurement has been selected.
The ORP measurement needs no calibration and no compensation for temperature effects. Therefore these parameters do not appear in the menu if ORP measurement is selected.
pH/Redox Measurement
pH/Redox Measurement
15 OPTISENS PAC 050
Page 16
4.4 Calibration
4.4.1 Calibration procedure
NOTE
4.4.2 Buffers
Main menu
Calibrate
.
Cal. press &
.
Slope
.
058 mV/pH Sensor zero 001 mV
Basic settings Buffers used  Buffer 1
7.00 pH Buffer 2
4.00 pH Sensor buffer
7.00 pH
1) Switch the controller OFF and select manual temperature compensation. Enter the temperature of the calibration solutions.
2) Immerse the electrode in one of the calibration solutions. Wait until the measured value is stable, then calibrate by pressing keys and - while still applying pressure - additionally key . The pH of the calibration solution is now displayed as measured value.
3) Rinse the electrode and repeat step 2 with the second calibration solution.
4) Check the slope and sensor zero-point, then put the electrode back into the armature. Select automatic temperature compensation and switch ON the controller.
The slope should be close to 59 mV/pH, the zero-point close to 0 mV. The slope decreases and the zero error increases with time. When either value exceeds certain limits, the instrument displays an error message indicating that the electrode has to be replaced.
At works the following buffers are stored:
Calibration solutions: buffer solutions pH 4 and pH 7
Inner buffer of the electrode: pH 7
If you want to use other calibration solutions, adjust the values - it does not matter if you start with the higher or the lower pH.
If you are using electrodes with a special inner buffer, adjust the pH value of the sensor buffer, since this value is used as zero for the calculation of the measured values. You will find the pH of the sensor buffer printed on the electrode.
Calibrate
Meas. val. 6.99pH
Basic settings
.
Manual OPTISENS PAC 050
The calibration solutions stored as „buffers used“ are recognised automatically. It does not matter whether you start with buffer 1 or 2.
Buffers used
OPTISENS PAC 050 16
Page 17
4.5 Temperature compensation
Main menu
.
Temp. comp. Auto comp. Manual comp.
. . .
Basic settings
Select between two ways of compensation:
1) Automatic compensation with temperature sensor Mind that the temperature sensor should always measure the medium temperature around the sensor. When pH sensor and temperature sensor are not immersed in the same solution, better switch to manual compensation.
2) Manual compensation If the temperature can be regarded as constant, you can enter the temperature manually instead of measuring it continuously. The device will then compensate the temperature error of this temperature.
4.5.1 Calibration of the temperature measurement
If you connect the temperature sensor with a twin-core cable, slight deviations might occur between measured and real temperature. These deviations can be eliminated by calibration.
During start-up, measure the temperature manually and enter a correction term so that the display shows the exact temperature.
5. Adjustments of the controller
Main menu
. .
Selection of ON/OFF, P or PI controller, adjustment of P ranges and I Set points
. .
Basic settings
functions for 2 set points
Contr. settings For any type of controller you have to enter one or two set points, and you have to tell the device whether
these set points are reached by increasing or decreasing the measured value.
Automatic compensation
Temp. comp.
Basic settings
Cal. temp.
Manual compensation
Temp. comp. Temperature
25.0°C
Calibration temperature
Cal. temp.
0.0°C
Basic settings
. .
Adjustment of frequencies and pulse/pause times, hysteresis, direction of the control action, delay.
Manual OPTISENS PAC 050
17 OPTISENS PAC 050
Page 18
5.1 ON/OFF controller
Manual OPTISENS PAC 050
You can choose between three different controller versions:
ON/OFF controller
The ON/OFF controller switches ON if the measured value exceeds the set point and OFF if it drops back below it or vice versa. Dosage is always carried out with 100% (ON) or 0% (OFF). The parameter for an ON/OFF controller is the hysteresis.
P controller
The P controller or proportional controller reduces the dosage in the vicinity of the set point proportional to the control deviation. This is easily achieved if the analog output is used as steady control output. If the relays are used, the proportional reduction is achieved by either reducing the switch frequency (Impulse­frequency controller) or reducing the time within a given period of time in which the relay is ON (pulse-pause controller). The parameters for a P controller are the P range and the impulse- frequency or the pulse+ pause time and the minimum pulse.
PI controller
The PI controller is a P controller with an additional I function. Adjustments and parameters are the same as for a P controller. Additionally the integral action time has to be adjusted which determines the I function. The I function eliminates the P controller’s disadvantage of a remaining steady-state deviation.
Main menu
. .
Set points
.
7.50 pH controller versions for S1
.
P range S1 and S2
.
0.20 pH
Set points and controller version
Set point S1 You can choose different
Integral Time S1 000 sec. Set point S2
6.50 pH P range S2 Controller settings
0.20 pH
.
Integral Time S2 . 000 sec. Direction S1 raise
Basic settings
Basic settings
.
Lower
.
Contr. settings
Direction S2
Hysteresis
0.01 pH
For an ON/OFF controller you have to set the following parameters:
1) Set points S1 and S2
Set point S1 refers to relay 1, set point S2 refers to relay 2.
2) P range and integral action time for S1 and S2
For an ON/OFF controller set P range = 0 and integral time = 0.
3) Acting direction for S1 and S2
Select “raise“ if the dosage raises the measured value. Select “lower“ if the dosage lowers the measured value.
4) Optionally a hysteresis
The hysteresis prevents fast switching in the vicinity of the set point. If hysteresis is activated (by setting a value > 0) the relay switches only when the set point is exceeded by half the hysteresis.
OPTISENS PAC 050 18
Page 19
Manual OPTISENS PAC 050
5.2 P / PI controller as impulse-frequency controller
Main menu
. .
Set points
.
7.50 pH controller versions for S1
.
P range S1 and S2
.
0.20 pH
Set points and controller version
Setpoint S1 You can choose different
Integral time S1 000 sec Set points S2
6.50 pH P range S2 Controller settings
0.20 pH Pulse-Freq. S1 Integral Time S2 00*100/4 000 sec. Pulse-Freq. S2 00*100/h
Basic settings
Basic settings
.
.
Direction S2
Contr. settings
Direction S1
Raise
Lower
For an impulse-frequency controller you have to set the following parameters:
1) Set points S1 and S2
S1 refers to relay 1, S2 refers to relay 2.
2) P range and integral action time for S1 and S2
Adjust a P range > 0. For a P controller set integral time = 0, for a PI controller set an integral time > 0.
3) The acting direction for S1 and S2
Select „raise“ if the dosage raises the measured value. Select „lower“ if the dosage lowers the measured value.
4) Pulse-frequencies for S1 and S2
Enter the maximum pulse-frequency that corresponds to 100% dosage.
5.3 P / PI controller as pulse-pause controller
Main menu
. .
Set points
.
7.50 pH Different controller
.
P range S1 Controller Settings Versions for S1+S2
.
0.20 pH Pulse-Freq. S1
.
Integral time S1 00*100/h
Set points and controller versions
Set point S1
You can choose
000 sec Pulse-Freq. S.2 Set point S2 00*100/h
6.50 pH Direction S1 P range S2 Raise
0.20 pH Direction S2 Integral time S2 Lower 000 sec.
Basic settings
Basic settings
.
10 sec
.
Contr. settings
. .
Pulse+Pause The pulse+pause time
must be at least twice
Least Pulse
0.1 sec pulse time
as long as the least
19 OPTISENS PAC 050
Page 20
5.4 Activation and deactivation of the controller
For a pulse-pause controller you have to set the following parameters:
1) Set points S1 and S2
S1 refers to relay 1, S2 refers to relay 2.
2) P range and integral action time
Adjust a P range > 0. For a P controller set integral time = 0, for a PI controller set an integral time > 0.
3) Pulse-frequencies for S1 and S2
Both frequencies must be set to 00, otherwise the controller will act as an impulse-frequency controller.
4) The acting direction for S1 and S2
Select „raise“ if the dosage raises the measured value. Select „lower“ if the dosage lowers the measured value.
5) Pulse + pause time
Define a period of time during which the relay is proportionally to the control deviation ON (pulse) or OFF (pause), respectively.
6) Least pulse time
Set a minimum pulse time that the relay has to at least remain open to allow the actuator to react.
The controller is activated and deactivated from the main display without any menus. Press key to switch from Manual Mode (controller OFF) to Automatic Mode (controller ON) and vice versa. The actual mode is indicated in the display.
Make sure that the controller is deactivated before connecting dosing pumps or other actuators!
Manual OPTISENS PAC 050
WARNING
5.5 Turn-on delay
5.6 External controller stop
NOTE
Main menu
. .
Basic settings
180 s
Set a delay time which has to pass before the controller is activated after start-up or power interrupt. This allows the measurement to settle and prevents inappropriate dosage of chemicals.
You can activate or deactivate the controller with an external switch by using the digital input. This feature requires no settings or adjustments. If the digital input is short- circuited, the controller stops, and the message „external controller stop“ appears in the display.
This feature can also be used as lack-of water indication. Just connect a level sensor to the digital input.
Basic settings
.
.
Turn-on delay Turn-on delay ON Turn-on delay
Turn-on delay
OPTISENS PAC 050 20
Page 21
5.7 Manual operation of the relays
6.99 pH 25.0°C
F S1 F S2 Auto
6.99 pH 25.0°C
F S1 F S2 Man
6.99 pH 25.0°C
 S1 F S2 Man
6.99 pH 25.0°C
J S1 F S2 Man
6.99 pH 25.0°C
 S1 F S2 Man
6.99 pH 25.0°C
F S1 S2 Man
6.99 pH 25.0°C
F S1 J S2 Man
6.99 pH 25.0°C
F S1 S2 Man
6.99 pH 25.0°C
F S1 F S2 Man
For manual operation you need no menu.
With key you switch OFF the controller.
With key you switch between Manual operation <> operation mode S1 <> operation mode S2 <>
manual operation.
In the operation mode you can switch ON and OFF the selected relay with key  .
A flashing square indicates that the relay is in operation mode.
A dark square indicates that the relay is switched ON.
A light square indicates that the relay is switched OFF.
If you switch ON a relay it stays ON until you switch it OFF again manually!
Manual OPTISENS PAC 050
1) If the controller is ON, switch it OFF with key  . Instead of „Auto“ the display shows „Man“.
2) Switch to the operation mode of S1 with key. The square to the left of S1 starts to flash.
3) Switch ON relay 1 with key  . The square to the left of S1 gets dark.
4) Switch OFF relay 1 again with key  . The square gets light.
5) Switch to the operation mode of S2 with key. The square to the left of S2 starts to flash.
6) Switch ON relay 2 with key  . The square to the left of S2 gets dark.
7) Switch OFF relay 2 again with key  . The square gets light.
8) Leave the operation mode of relay 2 with key . Both squares appear light, none flashes - You have left the operation mode.
WARNING
21 OPTISENS PAC 050
Page 22
5.8 Limit values
NOTE
5.8.1 Turn-on delay
5.9 Dosage check
NOTE NOTE
Manual OPTISENS PAC 050
Main menu
. .
Limit values
Limit values
Limit S1
8.00 pH Limit S2
6.00 pH Turn-on delay 005 sec.
For the alarm, you can adjust two limits:
Limit 1 is an upper limit. If the measured value exceeds limit 1, an alarm is issued.
Limit 2 is a lower limit. The alarm is issued if the measured value drops below limit 2.
In case of alarm the display shows the message „limit 1“ (or 2, respectively) and relay 3 is switched ON. This relay can be used to activate an external horn or lamp.
The limit function is active only if the controller is set on automatic mode. When you switch the controller to MAN, the alarm is extinguished.
In some applications it happens regularly that the measured value exceeds a limit for a short period of time. To avoid having an alarm issued under these circumstances you can adjust a turn-on delay which has to pass before an alarm is issued. If the turn- on delay time is >0 then the alarm is issued only if the cause of alarm remains longer than the specified turn-on delay time.
Main menu
. .
Basic settings
Contr. settings
Basic settings
.
.
Controller settings
Dosage check S1 00 min Dosage check S2 00 min
In the basic settings of the controller you can define, how long a controller is supposed to dose with 100% without raising alarm.
If the controller output is 100% for more than the specified time, this is interpreted as an indication of failure, and the device issues an alarm and deactivates the controller, thus stopping further dosage.
The dosage check is a safety catch to prevent hazardous chemicals to be set free in case of a defective dosing tube or tube connection.
In case of an alarm due to dosage check, only the controller concerned is deactivated. If you set the dosage check time to 0 min, the dosage check function is deactivated.
OPTISENS PAC 050 22
Page 23
6. Alarm
Manual OPTISENS PAC 050
Additional to the limit function the device provides various check functions that raise alarm. In case of an alarm, relay 3 switches, undelayed, and the cause of the alarm is indicated in the display. If the cause of alarm is such that control is no longer possible or might even be dangerous, the controller is
automatically deactivated until the alarm is switched off. Switching off the alarm is done automatically by the device as soon as the cause of alarm is eliminated.
Sensor check during measurement
During measurement all connected sensors are checked. If an analog input does not receive a correct signal, for e. g. if a cable is broken or a sensor damaged, an alarm is issued, and the controller deactivated. Alarm and controller stop remain until the analog input receives correct signals again.
Sensor check during measurement
During measurement all connected sensors are checked. If an analog input does not receive a correct signal, for e. g. if a cable is broken or a sensor damaged, an alarm is issued, and the controller deactivated. Alarm and controller stop remain until the analog input receives correct signals again.
Dosage control
If a controller output is 100% longer than the defined dosage time, an alarm is issued, and the corresponding controller is deactivated. The alarm remains until the controller output drops below 100% - it can also be extinguished by setting the controller to manual mode.
Cause of alarm only active in AUTO mode deactivates controller
Slope error no no Sensor zero no no Error input 1 no yes Error input 2 no yes Limit yes no Dosage yes yes
23 OPTISENS PAC 050
Page 24
6.1 Error messages
7. Output
Manual OPTISENS PAC 050
Error message Cause Action
Slope error The slope determined by
calibration was higher than 65 mV/pH or lower than 50 mV/pH.
Please check the sensor connection and cable, and make sure that the calibration data stored in the instrument corresponds to the calibration solutions used. Also make sure that the temperature settings are correct. Then repeat the calibration. If the error message remains, the sensor has to be replaced or regenerated.
Sensor zero The zero deviation determined by
calibration was higher than 60 mV or lower than -60 mV.
Please follow the procedure described above. Make sure that the sensor buffer stored in the calibration data corresponds to the sensor used. If the error remains, at least the reference electrode has to be replaced or regenerated.
Error input 1 The pH or ORP electrode gives no
signal.
Error input 2 The temperature sensor gives no
signal.
Please check the connections, the cable and the sensor for signs of damage. Please check the connections, the cable and the sensor for signs of damage. This message also appears if automatic temperature compensation was selected although no temperature sensor was used.
Limit 1 / 2 The measured value exceeded
limit 1 (or dropped below limit 2,
Please check the dosing and readjust the control parameters, if necessary.
respectively).
Dosage Pump 1 (or 2, resp.) feeds with
100% for more than the defined period of time.
Ext. controller stop The digital input has been short-
circuited.
Please check the dosing, especially the feeding tubes and connections. Caution! Carefully check for leaking chemicals! This only indicates the external controller stop. If, however, you have connected a level sensor, this message appears due to the „lack of water“ alarm.
Main menu
. .
Basic settings
Analog output
Basic settings
.
.
.
Start 0/4mA
.
End 20 mA
.
Analog output
Analog output
0…20mA/4…20mA
Meas. value
Bus address Nr. 00
Serial interface
RS485 (option)
Bus address
Set a number between 0 and 31 to address the device via data bus.
OPTISENS PAC 050 24
Page 25
7.1 Current output
7.2 Serial interface RS485 (option)
You can read out the measured value as 0/4…20 mA signals via the current output. With the setting 4…20 mA the resolution is lower, but defective cable connections are immediately evident.
With the parameters Begin and End you define which part of the measuring range you want to read out. Alternatively you can use the current output as steady-state controller output. In that case assign the current
output to controller S1 or S2, respectively.
The devices are available with serial interface RS485 by means of which they can be integrated in a data bus system. Via the interface, all settings, measured and control values as well as any error messages can be read out digitally.
Devices with RS485 are automatically delivered with the leaflet „Information on the RS485“ which contains instructions on the communication and a complete list of the functions available via interface.
8. Operation and maintenance
8.1 Maintenance of the device
8.2 Display contrast
8.3 Exchange fuse
WARNING
The device does not require any maintenance. There is no need for readjustment. If you want to have the device checked regularly, you are welcome to send it to KROHNE Water Solutions. Alternatively the device can be checked on site by one of our engineers.
With devices in wall-mounting enclosures the display contrast can be adjusted to the actual light conditions by means of a potentiometer. It is indicated in the connection diagram with the word „display“.
Devices in wall-mounting enclosures have an internal fuse which has to be replaced at need. You will find a spare fuse fixed to the inside of the terminal cover. Information on the fuse can be found in the chapter „Technical data“.
To exchange the fuse, open the front carefully. The fuse is located on the right hand side. It is kept in place by a Bayonet lock. Turn the lock to the left until the fuse pops up. Exchange it and fix the new fuse by turning the lock to the right. Put the front back on and fix it tightly.
Disconnect the power supply before opening the device!
Mind that the cable connections to the front are not damaged, broken or torn during the process!
Manual OPTISENS PAC 050
ATTENTION
8.4 Cleaning
25 OPTISENS PAC 050
The front and the display should not get in touch with organic solutions such as methanol. Never let water get inside the device. We suggest to simply use a damp cloth for cleaning.
Page 26
8.5 Maintenance of the pH or ORP measurement
The metallic surface of the ORP electrode must be cleaned regularly. pH and reference electrodes change with time, so their characteristics have to be determined regularly by means of calibration.
NOTE
8.6 Disposal
8.7 Service
8.7.1 Product info
8.7.2 Analog inputs
8.7.3 Erase settings (reset)
The instrument checks the sensor’s characteristic parameters after each calibration and displays an error message if the sensor has to be replaced. See chapter „Error messages“.
For disposal please notice that the device contains electrolyte capacitors which have to be disposed separately.
Main menu
.
Unit no.
.
No. 041
.
Software date
.
M/Y 1.00
.
Service Analog inputs
Input 1 003 Analog inputs Input 2 25.0°C
Erase settings Press & 
In this menu you will find information which is especially important for any inquiries, updates or problems.
These figures allow a precise identification of the device (hardware and software).
Here you can see the raw data the device obtains from the sensors. They are not influenced by compensations or calibration and offer valuable information in case of problems with the measurement or the device.
If you have problems interpreting this data, send them to your supplier together with the device data and ask for support.
With this function you can erase all customer settings and restore the original at-works data. The process takes some 30 seconds. When it is finished the display will show the measured value, and the
controller will switch off.
Manual OPTISENS PAC 050
Product. date
M/Y 1.00
Analog inputs
Erase settings
Service
Product info
OPTISENS PAC 050 26
Page 27
9. Technical data
9.1 Technical data
Manual OPTISENS PAC 050
Feature PAC 050 R PAC 050 W
Version panel-type converter wall-mounting enclosure Dimensions (w x h x d) 96 x 96 x 127 mm / 3.8 x 3.8 x 5” 165 x 160 x 80 mm / 6.5 x 6.3 x
3.1” Weight 0.8 kg / 1.8 lbs 1.0 kg / 2.2 lbs Terminals screw terminals max. 1.5 mm2 Spring clamp max. 1,5 mm2 Protection class IP 54 (front), IP 30 (housing) IP 65 Power supply 230 V AC +6/-10 %, 50/60 Hz, alternativ 110 V AC oder 24 V AC Internal fuse none 230 V : 63 mA slow
110 V : 125 mA slow
24 V : 800 mA semi-slow Power consumption 10 VA 10 VA Display LCD, 2-line, 2x16 characters, illuminated background
measured value and temperature with dimensions indication of relay
status Current output 0/4...20 mA, galvanically isolated, max. burden 500 Ohm Interface (option) RS485, Baud rate 9600, data format 8 Bit, 1 Start and 1 Stop bit Controller ON/OFF controller with hysteresis, P or PI controller as Pulse- Pause-
or Impulse-frequency controller, steady controller, bidirectional PI
control action, adjustable onset-delay, dosage control function, manual
operation of the relays, controller stop via external switch or level
sensor (lack-of-water sensor) Set points 2 set points adjustable within the measuring range Alarm function with min. and max. limit and turn-on delay Contact rating 6 A/ 250 V, max. 550 VA resistive load (with RC protective circuit) Operation temperature 0…+50°C / 32…+120°F Storage temperature -20…+65°C / -5…+150°F Humidity max. 90% at 40°C /105°F non-condensing
27 OPTISENS PAC 050
Page 28
9.2 Dimensions
9.2.1 Panel-mounting enclosure
Manual OPTISENS PAC 050
9.2.2 Wall-mounting enclosure
OPTISENS PAC 050 28
Page 29
Manual OPTISENS PAC 050
10. Device return form
Company:………………………………………………. Address:…………………………………………….. Department:…………………………………………… Name:……………………………………………….. Tel.-No.:……………………………………………….. Fax No.:……………………………………………... Email:…………………………………………………… The enclosed device Type:…………………………………………………………………………………………………………………………… KROHNE Water Solutions Order No. or Series No.:………….………………………………………………………….. has been operated with the following liquid:……………………………………………………………………………….
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems. Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, KROHNE Water Solutions may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that KROHNE Water Solutions can only service this device if it is accompanied by the following certificate confirming that the device is safe to handle.
S P E C I M E N certificate
Because this liquid is
water-hazardous toxic caustic flammable
we have
checked that all cavities in the device are free from such substances flushed out and neutralized all cavities in the device
We confirm that there is no risk to humans or environment through any residual liquid contained in this device.
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralizing, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
We cannot service this device unless accompanied by such a form.
29 OPTISENS PAC 050
Page 30
Manual OPTISENS PAC 050
11. Customer settings - for reference
Type ………………. Date of installation ………………. Serial number ………………. Software version ……………….
Cal. solution 1 ………………...pH Cal. solution 2 ………………...pH Sensor buffer ………………...pH
Temperature …………….°C Correction ……………………..°C
Output 1
Start …………………… End ……………………
Controller S1 Controller S2 Acting direction Acting direction
Set point …………………… Set point ………………………. Hysteresis …………………… Hysteresis ………………………. P range …………………… P range ………………………. Integral time …………………..s Integral time ……………………….s Pulse + pause time …………………..s Pulse+ pause time ……………………….s Minimum pulse …………………..s Minimum pulse ……………………….s Pulse frequency …………….*100/h Pulse frequency ………………...*100/h
Delay time ……………….min.
Limit S1 …………………… Limit S2 ……………………….. Delay time ……………….min.
Bus address ……………………
Device
Measurement
Temperature compensation
Analog output
Controller
Turn-on delay
Alarm
Serial interface RS 485
Identification / location ………………………………………………….
F pH F ORP
F Manual F Automatic
F Pt100
F 0…20mA F 4…20mA
for F pH/ORP F Controller S1 F Controller S2
F raise F lower F raise F lower
F normally closed F normally open F pulse contact F perm. contact
OPTISENS PAC 050 30
Loading...