Krohne H250 M40 User Manual

Page 1
Handbook
Handbook
H250 M40
H250 M40
H250 M40 H250 M40
Variable area flowmeter
HandbookHandbook
© KROHNE 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 2
:
IMPRINT
:::::::::::::::::::::::::::::::::::::::
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.
Copyright 2015 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
2
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 3
H250 M40
CONTENTS
1 Safety instructions 5
1.1 Intended use ..................................................................................................................... 5
1.2 Certifications .................................................................................................................... 6
1.3 Pressure equipment directive.......................................................................................... 6
1.4 Safety instructions from the manufacturer ..................................................................... 7
1.4.1 Copyright and data protection ................................................................................................ 7
1.4.2 Disclaimer ............................................................................................................................... 7
1.4.3 Product liability and warranty ................................................................................................ 8
1.4.4 Information concerning the documentation........................................................................... 8
1.4.5 Warnings and symbols used................................................................................................... 9
1.5 Safety instructions for the operator................................................................................. 9
2 Device description 10
2.1 Scope of delivery............................................................................................................. 10
2.2 Device version................................................................................................................. 11
2.2.1 Indicator versions ................................................................................................................. 12
2.2.2 Float damping ....................................................................................................................... 14
2.2.3 Pointer damping.................................................................................................................... 14
2.3 Nameplate ...................................................................................................................... 15
2.4 Description code............................................................................................................. 16
2.5 Electronics revision........................................................................................................ 17
3 Installation 18
3.1 Notes on installation ......................................................................................................18
3.2 Storage ........................................................................................................................... 18
3.3 Installation conditions ....................................................................................................19
3.3.1 Tightening torques................................................................................................................ 21
3.3.2 Magnetic filters ..................................................................................................................... 21
3.3.3 Heat insulation ...................................................................................................................... 22
4 Electrical connections 23
4.1 Security information.......................................................................................................23
4.2 Electrical connection indicator M40............................................................................... 24
4.2.1 Limit switches K1/K2 ............................................................................................................ 24
4.2.2 Current output ESK4 / ESK4A............................................................................................... 27
4.2.3 ESK4-T limit outputs............................................................................................................. 30
4.2.4 Connection Harting HAN
®
7D............................................................................................... 32
4.3 Grounding connections................................................................................................... 33
4.4 Protection category ........................................................................................................33
5 Start-up 34
5.1 Standard device .............................................................................................................. 34
5.2 Indicator ESK4-T ............................................................................................................ 34
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
3
Page 4
CONTENTS
H250 M40
6 Operation 35
6.1 ESK4 / ESK4A - Loop Check Mode................................................................................. 35
6.2 Operating elements ESK4-T........................................................................................... 36
6.3 Basic principles of operation ESK4-T ............................................................................ 37
6.3.1 Functional description of the keys........................................................................................ 37
6.3.2 Navigation within the menu structure.................................................................................. 37
6.3.3 Change the settings in the menu..........................................................................................38
6.4 Overview of the units ESK4-T......................................................................................... 39
6.5 Error messages ESK4-T................................................................................................. 40
6.6 Menu indicator ESK4-T................................................................................................... 43
6.6.1 Factory settings .................................................................................................................... 43
6.6.2 Menu structure ..................................................................................................................... 44
6.6.3 Menu explanations................................................................................................................ 46
7 Service 54
7.1 Maintenance ................................................................................................................... 54
7.2 Replacement and retrofitting......................................................................................... 54
7.2.1 Replacing floats .................................................................................................................... 54
7.2.2 Retrofitting float damping..................................................................................................... 55
7.2.3 Retrofitting limit switch ........................................................................................................ 55
7.2.4 Replacement - Retrofitting ESK4 / ESK4A ........................................................................... 56
7.2.5 Replacement - Retrofitting add-on module ESK4-T / PA / FF............................................. 57
7.3 Spare parts availability...................................................................................................57
7.3.1 List of spare parts................................................................................................................. 57
7.4 Availability of services .................................................................................................... 61
7.5 Returning the device to the manufacturer..................................................................... 61
7.5.1 General information.............................................................................................................. 61
7.5.2 Form (for copying) to accompany a returned device............................................................ 62
7.6 Disposal .......................................................................................................................... 62
8 Technical data 63
8.1 Functional principle........................................................................................................ 63
8.2 Technical data................................................................................................................. 64
8.3 Dimensions and weights ................................................................................................ 71
8.4 Measuring ranges...........................................................................................................74
9 Notes 82
4
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 5
H250 M40
1.1 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator.
INFORMATION!
This device is a Group 1, Class A device as specified within CISPR11:2009. It is intended for use in industrial environment. There may be potential difficulties in ensuring electromagnetic compatibility in other environments, due to conducted as well as radiated disturbances.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose.
The variable area flowmeters are suitable for measuring clean gases, vapours and liquids.
Intended use:
The product may not contain any ferromagnetic particles or solids. It may be necessary to install magnetic filters or mechanical filters.
The product must be sufficiently liquid and free of deposits.
Avoid pressure surges and pulsing flows.
Open valves slowly. Do not use solenoid valves.
SAFETY INSTRUCTIONS
1
Use suitable measures to eliminate compression vibrations during gas measurements:
Short pipeline lengths to next restriction
Nominal pipe size not greater than nominal device size
Use of floats with damping
Increase in operating pressure (while taking into account the resulting change in density and
thus change in scale)
Observe installation conditions according to VDI/VDE 3513-3.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
CAUTION!
Do not use any abrasive media containing solid particles or highly viscous media.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
5
Page 6
1
SAFETY INSTRUCTIONS
1.2 Certifications
CE marking
The device fulfils all applicable statutory requirements of the following EC directives:
Pressure equipment directive
For devices with electrical installations: EMC directive
Devices for use in hazardous areas: ATEX directive
as well as
NAMUR recommendations NE 21, NE 43 and NE 107
The manufacturer certifies successful testing of the product by applying the CE marking. A CE declaration of conformity regarding the directives in question and the associated harmonised standards can be downloaded from our internet site.
H250 M40
1.3 Pressure equipment directive
A conformity assessment in accordance with pressure equipment directive has been carried out for the devices described. Conformity is certified by applying the CE mark. The number of the notified body is also stated.
The PED key describes rating of devices:
Example: PED/G1/III/H G Gases and vapours 1Fluid group 1 III Category III H Conformity assessment method according to Module H
The PED key identification can be found on the nameplate of the device (for details refer to
Nameplate
INFORMATION!
The stated pressures (PS) and temperatures (TS) only apply as refers to the pressure resistance of the sensor body. As regards the functionality of the entire device, further restrictions of the maximum temperature may need to be observed (e.g. ATEX approval). Devices rated below Category I due to their size, do not receive the CE mark in the scope of the PED. These devices are subject to applicable Sound Engineering Practice SEP.
on page 15).
6
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 7
H250 M40
1.4 Safety instructions from the manufacturer
1.4.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
SAFETY INSTRUCTIONS
1
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.4.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
7
Page 8
1
SAFETY INSTRUCTIONS
1.4.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation or operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
1.4.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of icons as shown below.
H250 M40
8
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 9
H250 M40
1.4.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This warning refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
SAFETY INSTRUCTIONS
1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.5 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
9
Page 10
2
DEVICE DESCRIPTION
2.1 Scope of delivery
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
H250 M40
Figure 2-1: Scope of delivery
1 Measuring device in ordered version 2 For the ESK4-T version - bar magnet 3 Documentation 4 Certificates, calibration report (supplied to order only) 5 Wrench
10
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 11
H250 M40
2.2 Device version
H250 with indicator M40
H250 with M40 indicator with display cut-out for ESK4-T
1. H250/RR/M40
Local indicator without auxiliary power
max. 2 limit switches, type NAMUR, NAMUR safety-oriented or transistor (3-wire)
Electrical signal output 4...20 mA, HART
Intrinsically safe [Ex i] or in explosion-proof enclosure [Ex d]
DEVICE DESCRIPTION
®
or Fieldbus communication
2
2. H250/RR/M40
additional LCD, measured value and/or flow counter
2 configurable binary outputs, limit value or pulse output
1 binary input, Start / Stop / Reset flow counter
2-wire current output 420 mA, HART
Intrinsically safe [Ex i] or in explosion-proof enclosure [Ex d]
®
communication
The following designs are available as options:
H250 with indicator M40 as high-temperature version HT
H250 with indicator M40 with increased corrosion protection (special paint finish)
H250H for use in horizontal pipelines
H250U for use in vertical fall pipes
H250F with hygienic measuring tube design for Food & Pharma
H250C with PTFE / TFM liner for aggressive media
Display options
M40 - Aluminium, single-layer powder coating (polyester)
M40S - Aluminium, two-layer powder coating (epoxy / polyester)
M40R - Stainless steel without coating
Offshore wet coating for aluminium or stainless steel on request
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
11
Page 12
2
DEVICE DESCRIPTION
2.2.1 Indicator versions
The M40 indicator can be fitted with various modules.
Basic version
H250 M40
1 Pointer module 2 Bolts for ESK4 / ESK4A attachment 3 Baseplate 4 Module profile 5 Pressure piece for ESK4 / ESK4A attachment 6 Housing cover locking device 7 Earth terminal external
Versions K1 / K2 and ESK4 / ESK4A
1 Indicator with K2 contact module 2 Indicator with ESK4 / ESK4A current output 4...20 mA
12
Both versions can be combined with one another.
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 13
H250 M40
Version ESK4-T
1 ESK4 / ESK4A connection 2 Module cover 3 Display 4 Display module ESK4-IO 5 Operating keys ^ 6 Connection binary outputs and reset input 7 Module connection cable
DEVICE DESCRIPTION
2
Version Fieldbus ESK4-FF / ESK4-PA
1 Basic module with electronic magnet sensors ESK4 / ESK4A 2 Connection bus module 3 DIP switch for bus settings
For more details see the supplementary instructions "H250 M40 Foundation Fieldbus" and "H250 M40 Profibus PA"
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
13
Page 14
2
DEVICE DESCRIPTION
2.2.2 Float damping
Float damping is characterised by high standstill times and self-centering. The damping sleeve is made of high performance ceramic or PEEK, depending on the medium and the application. Float damping can also be retrofitted for the user (refer to "Service").
Use of damping
Generally when CIV and DIV floats are used for gas measurement.
For TIV floats (H250/RR and H250/HC only) with an operating primary pressure:
Nominal size acc. to Operating primary pressure
EN 1092-1 ASME B16.5 [bar] [psig]
DN 50 ½" 0.3 4.4
DN25 1" 0.3 4.4 DN50 2" 0.2 2.9 DN80 3" 0.2 2.9
DN 100 4" 0.2 2.9
H250 M40
2.2.3 Pointer damping
In principle, the indicating element with its magnetic system contains indicator damping. An additional eddy current brake is advantageous in the event of fluctuating or pulsing flows. The magnets on the eddy current brake surround the pointer vane without touching it, damping its movement. The result is a much steadier pointer position, without distorting the measured value. A turnbuckle ensures a proper fit. The eddy current brake can be retrofitted during operation without recalibrating (refer to "Service").
1 Eddy current brake 2 Pointer vane 3 Bracket 4 Pointer cylinder
14
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 15
H250 M40
2.3 Nameplate
INFORMATION!
Check on the device nameplates, that the device is supplied according to your order.
DEVICE DESCRIPTION
2
Figure 2-2: Nameplate on the indicator
1 Device type 2 Manufacturer 3 Notified ATEX & PED body 4 Rating data: temperature & pressure rating 5 PED data 6 Ex data 7 Electrical connection data 8 Internet site
Additional markings on the indicator
SN - serial number
SO - sales order / item
PA - Production order
Vx - product configurator code
AC - article code
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
15
Page 16
2
DEVICE DESCRIPTION
2.4 Description code
The description code* consists of the following elements:
1 Device type H250 - standard version H250H - horizontal flow direction H250U - flow direction from top to bottom
2 Materials / versions RR - stainless steel C - PTFE or PTFE/ceramics HC - Hastelloy Ti - Titanium Mo - Monel F - aseptic version (food)
H250 M40
3 Heating jacket version B - with heating jacket
4 Series of indicators M40 - Indicator M40 M40S - Indicator with increased corrosion protection M40R - Indicator in stainless steel housing
5 High temperature version HT - Version with HT extension
6 Electrical signal output ESK - Electrical signal output 4...20mA (ESK4 / ESK4A)
- optionally available with counter, I/O module and display (ESK4-T) or
- Foundation Fieldbus (ESK4-FF) or
- Profibus PA (ESK4-PA)
7 Limit switches K1 - One limit switch K2 - Two limit switches
8 Explosion protection Ex - Explosion-protected equipment
9 SIL Version SE - SIL compliant electronic signal output SK - SIL compliant limit switch
* positions which are not needed are omitted (no blank positions)
16
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 17
H250 M40
2.5 Electronics revision
The Electronic Revision (sticker on the base module ESK4 / ESK4A) indicates the respective hardware/software status of the electronics. All add-on modules (ESK4-T, ESK4-FF and ESK4­PA) have an additional sticker indicating their respective firmware version.
DEVICE DESCRIPTION
2
Electronics
Explanations
revision
ER 1.1.x Basic version (cannot be combined with other indicator versions):
ER 2.0.x Functional add-on to ER 1.1.x:
ER 2.1.x Functional add-on to ER 2.0.x
ER 2.2.x Functional add-on to ER 2.1.x:
ER 3.0.x Functional add-on:
ESK4 / Current output 420mA with HART® communication; (ESK4 HART DD 01.01. AMS10x AMS11x ESK4 HART DD 01.01. PDM6.0 ESK4 HART DTM 1.0.3 FDT1.2)
can be combined with indicator version ESK4 FF / Foundation Fieldbus; (Firmware Version FF module from 1.0.2)
can be combined with indicator version ESK4-PA / Profibus PA; (Firmware Version PA module from 1.0.0) can be combined with indicator version ESK4-T / LCD, binary inputs/outputs; (Firmware Version T module from 1.1.0)
Support of failure signal (low) according to NE43 for the ESK 4 current output module
®
Update for HART can be combined with ESK4-FF (Firmware Version FF module from 1.0.2) Firmware Version ESK4-PA (... PA module from 1.0.0) Firmware Version ESK4-T (... T module from 1.2.0)
communication from 5.9 to 7.4 including new DD/DTM
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
17
Page 18
3
INSTALLATION
3.1 Notes on installation
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
3.2 Storage
Store the device in a dry and dust-free location.
Avoid lasting direct exposure to the sun.
Store the device in its original packing.
The permissible storage temperature for standard devices is -40...+80°C / -40...+176°F.
H250 M40
18
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 19
H250 M40
3.3 Installation conditions
CAUTION!
When installing the device in the piping, the following points must be observed:
The variable area flowmeter must be installed vertically (measuring principle). Flow direction from bottom to top. For installation recommendations please refer also to directive VDI/VDE 3513, sheet 3.
H250Hs are installed horizontally and H250U devices are installed vertically with the flow direction from top to bottom.
A straight unimpeded inlet run of 5x DN upstream of the device and a straight outlet run of 3x DN downstream of the device are recommended.
Screws, bolts and gaskets are to be provided by the customer and must be selected in accordance with the pressure rating of the connection or the operating pressure.
The inside diameter of the flange deviates from the standard dimensions. Flange seal standard DIN 2690 can be applied without any limitation.
Align the gaskets. Tighten the nuts with the tightening torques of the appropriate pressure rating.
For devices with PTFE liner or ceramic liner and PTFE raised faces, refer to chapter "Tightening torques".
Control devices are to be positioned downstream of the measuring device.
Shut-off devices are preferably to be positioned upstream of the measuring device.
Before connecting, blow or flush out the pipes leading to the device.
Pipes for gas flow need to be dried before the device is installed.
Use connectors suitable for the particular device version.
Align the pipes centrically with the connection bores on the measuring device so they are free of stresses.
If necessary, the piping has to be supported to reduce the vibrations transmitted to the measuring device.
Do not lay signal cables directly next to cables for the power supply.
INSTALLATION
3
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
19
Page 20
3
INSTALLATION
Minimum distance between devices
When several devices are installed next to one another, a minimum distance of a > 300 mm /
11.8" between the devices is necessary.
H250 M40
Take special note of the installation position for the H250H with horizontal flow direction:
In order to comply with thermal parameters and measuring accuracy, H250H flowmeters for horizontal installation are to be installed in the pipeline so that the display is located on the side of the measuring tube. The maximum product and ambient temperatures indicated as well as the measuring accuracy are based on lateral installation of the display.
20
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 21
H250 M40
3.3.1 Tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange threads with the following torques:
Nominal size according to Stud bolts Max. torque
EN 1092-1 ASME B 16.5 EN ASME EN 1092-1 ASME 150 lb
DN PN Inch lb 150 lb 300 lb Nm ft*lbf Nm ft*lbf
15 40 ½“ 150/300 4x M12 4x ½“ 4x ½“ 9.8 7.1 5.2 3.8 25 40 1 150/300 4x M12 4x ½“ 4x 5/8 21 15 10 7.2 50 40 2 150/300 4x M16 4x 5/8 8x 5/8 57 41 41 30 80 16 3 150/300 8x M16 4x 5/8 8x ¾“ 47 34 70 51
100 16 4 150/300 8x M16 8x 5/8 8x ¾“ 67 48 50 36
INSTALLATION
3
3.3.2 Magnetic filters
The use of magnetic filters is recommended when the medium contains particles which can be influenced magnetically. The magnetic filter is to be installed in the flow direction upstream of the flowmeter. Bar magnets are positioned helically in the filter to provide optimal efficiency at low pressure loss. All of the magnets are coated individually with PTFE to protect against corrosion. Material: 1.4404/316L
Magnetic filters
1 Type F - fitting part with flange - overall length 100 mm / 4" 2 Type FS - fitting part without flange - overall length 50 mm / 2"
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
21
Page 22
3
INSTALLATION
3.3.3 Heat insulation
CAUTION!
The indicator housing may not be heat-insulated. The heat insulation 3 may only reach as far as the housing fastening 4.
1 Standard indicator M40 2 Indicator with HT extension
H250 M40
CAUTION!
The heat insulation 1 may only reach to the rear of the housing 2. The area around the cable entries 3 must be freely accessible.
Figure 3-1: Insulation - cross section
22
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 23
H250 M40
4.1 Security information
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
ELECTRICAL CONNECTIONS
4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
23
Page 24
4
ELECTRICAL CONNECTIONS
4.2 Electrical connection indicator M40
4.2.1 Limit switches K1/K2
The M40 indicator can be fitted with a maximum of two limit switches. The limit switch works as a slot sensor which is inductively activated via the semi-circular metal vane of the pointer. The switching points are set using the contact pointer. The position of the contact pointer is indicated on the scale.
Limit switch module
H250 M40
1 Min. contact 2 Max. contact 3 Locking screw 4 Maximum pointer 5 Connection terminal
The connection terminals feature a pluggable design and can be removed to connect the cables. The built-in limit switch types are shown on the indicator.
Electrical connection of the limit switches
Contact MIN MAX
Terminal no. 1 2 3 4 5 6
Connection 2-wire NAMUR - + - +
Connection 3-wire + - + -
Connection Reed SPST + - + -
24
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 25
H250 M40
Limit switch connection terminals
1 2-wire limit switch NAMUR 2 3-wire limit switch 3 Reed SPST Limit switch 4 Terminal connection min contact 5 Terminal connection max contact 6 3-wire load 7 NAMUR isolated switching amplifier 8 3-wire power supply
ELECTRICAL CONNECTIONS
4
Limit setting
Figure 4-1: Limit switch settings
1 Contact pointer MAX 2 Contact pointer MIN 3 Locking screw
Setting is carried out directly via contact pointers 1 and 2:
Slide the scale away.
Loosen the locking screw 3 slightly.
Slide the scale back to the latching point.
Set contact pointers 1 and 2 to the desired switching point.
After setting has been carried out, fix the contact pointers with the locking screw 3.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
25
Page 26
4
ELECTRICAL CONNECTIONS
Switch contact definition
1 MIN contact 2 MAX contact 3 Pointer vane with switching vane
If the measuring pointer vane goes into the slot an alarm is triggered. If the pointer vane is outside the slot sensor, a wire break in a NAMUR contact also triggers the alarm.
The 3-wire limit switch does not have any wire break detection.
H250 M40
Definition MinMin - MaxMax
1 MIN 2 contact or MAX 1 contact 2 MIN 1 contact or MAX 2 contact
Current consumption in the position shown:
Contact Type current
MIN 1 NAMUR 1 mA MIN 2 NAMUR 1 mA MAX 1 NAMUR 3 mA MAX 2 NAMUR 3 mA
26
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 27
H250 M40
4.2.2 Current output ESK4 / ESK4A
The connecting terminals of the ESK4 / ESK4A have a pluggable design and can be removed in order to connect the cables.
ESK4 / ESK4A connection
1 Current output ESK4 /ESK4A 2 Power supply 14...30 VDC 3 Measuring signal 4...20 mA
4 External load, HART
®
communication
ELECTRICAL CONNECTIONS
4
Power supply M40 with galvanic isolation
Power supply M40 with galvanic isolation
Power supply M40 with galvanic isolationPower supply M40 with galvanic isolation
Wiring must be planned with great care when it comes to connecting other devices such as evaluation units or process control. Internal connections in these devices (e.g. GND with PE, mass loops) may lead to non-permitted voltage potentials which could negatively affect the function of the converter itself or that of a device connected to it. In such cases a protected extra­low voltage (PELV) is recommended.
1 Terminal connection 2 Converter supply isolator with electrical isolation 3 Power supply (see supply isolator information) 4 Measuring signal 4...20 mA
5 External load, HART
®
communication
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
27
Page 28
4
ELECTRICAL CONNECTIONS
Power supply
Power supply
Power supplyPower supply
INFORMATION!
The supply voltage has to be between 14 VDC and 30 VDC. This is based on the total resistance of the measuring loop. To calculate this, the resistance of each component in the measuring loop (not including the device) must be added up.
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
where U
ext.
= the total measuring loop resistance.
R
L
24 mA + 14 V
L
= the minimum supply voltage and
INFORMATION!
The power supply has to be able to supply a minimum of 30 mA.
H250 M40
HART
HART® communication
HARTHART
When HART® communication is carried out with the ESK4, the analogue measured data transmission (4...20 mA) is not impaired in any way.
Exception for multidrop operation. In multidrop mode, a maximum of 15 devices with HART function can be operated in parallel, whereby their current outputs are switched inactive (I approx. 4.5 mA per device).
®
28
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 29
H250 M40
ELECTRICAL CONNECTIONS
Load for HART® communication
INFORMATION!
®
A load of at least 230 Ohm is required for HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal. In some cases a shielded cable may be necessary, if noise levels higher than the NE21 specification are anticipated.
communication.
4
Configuration
The ESK can be configured via HART® communication. DD (Device Description) for AMS and PDM
TM
as well as a DTM (Device Type Manager) for PACTware can be downloaded free of charge from our website.
The current flow rate can be transmitted using the integrated HART counter can be configured. Two limit values can be monitored. The limit values are assigned either to flow values or to the counter overflow.
are available for configuration. They
®
communication. A flow
Self monitoring - Diagnostics
During both start-up and operation, a wide variety of diagnostic functions are performed cyclically in the ESK4 / ESK4A in order to guarantee function reliability. When an error is detected, a failure signal (high) is activated (current > 21 mA, typically 22 mA) via the analogue
®
output. In addition more detailed information can be requested via HART signal is not activated for information and warnings.
(CMD#48). The failure
Diagnostic functions (Monitoring):
Plausibility of FRAM data
Plausibility of ROM data
Working range of internal reference voltages
Signal detection of the measuring range of the internal sensors
Temperature compensation of the internal sensors
Calibration based on the application
Plausibility of counting value
Plausibility of physical unit, system and selected unit
For ESK4A (HART 7) the diagnosis is reported in compliance with NE107.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
29
Page 30
4
ELECTRICAL CONNECTIONS
4.2.3 ESK4-T limit outputs
Once the housing cover has been unscrewed, the scale can be removed. The connection terminals feature a pluggable design and can be removed to connect the cables.
H250 M40
1 Binary output 1 2 Power supply / current output ESK4 / ESK4A 3 Binary output 2 4 Binary input
The binary inputs/outputs are electrically isolated from each other and from the ESK4 / ESK4A current output.
INFORMATION!
The binary inputs/outputs can only be operated if the power supply is applied to ESK4 / ESK4A terminal 11+ and 12-. The binary inputs/outputs come inactive by default and must thus be activated prior to first use (refer to section "Menu ESK4-T").
Connection binary outputs
Connection binary outputs
Connection binary outputsConnection binary outputs In accordance with the desired signal transmission, select one of the following connection types for binary outputs B1 and B2:
NAMUR (DC interface in accordance with EN 60947-5-6)
Transistor output (passive, open collector)
Binary output B1 B2
Terminal no. 1 2 3 4 5 6
30
Connection NAMUR + - + -
Connection transistor output + B
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
OC
+ B
OC
Page 31
H250 M40
1 NAMUR terminal connection 2 Isolated switching amplifier 3 OC switch output terminal connection 4 Power supply U
5 Load R
L
ext.
Value range NAMUR
ELECTRICAL CONNECTIONS
4
Normally closed Normally open
Switching value reached 1 mA > 3 mA Switching value not reached > 3 mA 1 mA
Value range applies only when connected to a switching amplifier with the following reference values:
Open-circuit voltage Uo = 8.2 VDC
Internal resistance R
= 1 kOhm
i
Value range for transistor output
Signal voltages UL [V] UH [V]
lower limit upper limit lower limit upper limit
via load R
Signal currents IL [mA] IH [mA]
Category 2 0 2 20 110
L
To ensure the value ranges, a load RL between 250 Ohm and 1 kOhm is recommended for the passive transistor output with a nominal voltage of 24 VDC.
0 2 16 30
lower limit upper limit lower limit upper limit
If other loads are used, caution is advised as the range of values of the signal voltages then no longer corresponds to the range of values for the inputs of process control systems and controls (DIN IEC 946).
CAUTION!
The upper limit of the signal current must not be exceeded as this may damage the transistor output.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
31
Page 32
4
ELECTRICAL CONNECTIONS
4.2.4 Connection Harting HAN® 7D
1 Terminal pin assignment HAN® 7D - View of plug connection
H250 M40
1 Pin No. HAN® 7D K1/K2: NAMUR
contacts
R1/R2 Reed contacts
ESK4/ESK4A Terminal no.
Namur Reed
1 2 NAMUR MIN (-) 2 Reed MIN - 1 1 2 2 NAMUR MIN (+) 2 Reed MIN - 2 3 3 3 NAMUR MAX (-) 3 Reed MAX - 4 4 4 3 NAMUR MAX (+) 3 Reed MAX - 5 6 5 - - 4 4...20mA (+) 11 6 - - 4 4...20mA (-) 12
7 - - -
8 - - -
Combinations of K1 / K2 and ESK4 / ESK4A are possible.
1 Pin No. HAN® 7D ESK4-T Terminal no.
1 5 Binary output B1 open coll. (+) 1 2 5 Binary output B1 open coll. (-) 3 3 7 Binary output B2 open coll. (+) 4 4 7 Binary output B2 open coll. (-) 6 5 6 4...20mA (+) 11 6 6 4...20mA (-) 12
7 -
8 -
32
INFORMATION!
No Harting connection for module ESK4-FF/PA is provided and is only available on request.
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 33
H250 M40
4.3 Grounding connections
1 Grounding connection in the indicator 2 Grounding connection outside
DANGER!
The grounding wire may not transfer any interference voltage. Do not use this grounding cable to ground any other electrical devices.
ELECTRICAL CONNECTIONS
4
4.4 Protection category
The measuring device meets all requirements of protection category IP66/68.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection class must be ensured again.
Therefore it is essential to observe the following points:
Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
The electrical cables must be undamaged and must comply with regulations.
The cables must be laid with a loop 3 upstream of the measuring device to prevent water
from getting into the housing.
The cable feedthroughs 2 must be tightened.
Close the unused cable feedthroughs using blanking plugs 1.
1 Use blanking plugs if no cable is routed through 2 Tighten cable feedthrough firmly 3 Lay the cable in a loop
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
33
Page 34
5
START-UP
5.1 Standard device
CAUTION!
When starting up the device, the following points must be observed:
Compare the actual operating pressure and the product temperature of the system with the specifications on the nameplate (PS and TS). These specifications may not be exceeded.
Make sure materials are compatible.
Slowly open the shut-off valve.
When measuring liquids, vent the pipes carefully.
When measuring gases, increase pressure slowly.
Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring unit or float.
A minimum operating pressure (primary pressure) is necessary to operate the device:
Medium Pressure loss : operating pressure
H250 M40
Liquids 1 : 2
Gases without float damping 1 : 5
Gases with float damping 1 : 2
5.2 Indicator ESK4-T
INFORMATION!
The device is always preset for the user and his application.
Start After the device is switched on, the display shows
"INITIALISING"
Firmwareversion IO-Modul
The device first performs a self-test. Here, all of the parameters preset for the customer are analysed and checked for plausibility. The device then switches to measuring mode and indicates the current measured value.
Operation
INFORMATION!
The device is low-maintenance Comply with the application limits with regard to temperature of the medium and ambient temperature.
34
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 35
H250 M40
6.1 ESK4 / ESK4A - Loop Check Mode
The ESK4 / ESK4A is equipped with a loop check function, enabling a simple test of the entire
420 mA current loop.
It is activated and operated using a microswitch 1.
CAUTION!
When activating the loop check mode, ensure that no alarms are unintentionally triggered in the downstream system components.
OPERATION
6
Press and hold the microswitch 1 for more than 6 seconds to activate loop check mode 3.
The current output jumps to constant 4mA.
Change the current output from constant 4 to constant 20 mA as often as you like by briefly
pressing it (less than 6 seconds) to check the function of the measuring circuit.
Exit loop check mode by holding down the microswitch (longer than 6 seconds). The current
output jumps back to measuring mode 2.
INFORMATION!
If the microswitch has not been pressed for longer than 60 seconds, the ESK4 / ESK4A automatically reverts to measuring mode 2.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
35
Page 36
6
OPERATION
6.2 Operating elements ESK4-T
H250 M40
The device is operated with the cover on the front open, using the mechanical keys cover closed using a bar magnet
bar magnet.
bar magnetbar magnet
keys, or with the
keyskeys
CAUTION!
The switching point of the magnetic sensors is directly at the height of the corresponding circle (see figure). Only touch the circle vertically and from the front using the bar magnet. Touching it from the side may cause an error in the measurement as the float position is recorded by the magnetic field sensors.
Figure 6-1: Display and operating elements
1 Enter button (circuit for bar magnet) 2 Up button (circuit for bar magnet) 3 Right button (circuit for bar magnet)
The mechanical keys and keys for the bar magnet have the same functionality. In this documentation the keys are represented as symbols to describe the operating functions:
Key Symbol
1 Enter ^^^^ 2 up 3 right
Table 6-1: ESK4-T operation keys
36
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 37
H250 M40
6.3 Basic principles of operation ESK4-T
6.3.1 Functional description of the keys
OPERATION
6
^^^^
Table 6-2: Description of the operating keys
Switch from measuring mode to menu mode
Switch to one menu level lower
Open menu item and activate change mode
Confirm query whether data should be accepted.
In change mode:
In change mode: Move the input cursor one position to the right. After the last digit the
In change mode:In change mode: input cursor jumps back to the beginning.
Change between the menu items within a menu level
In measuring mode:
In measuring mode: Switch between measured values and error messages
In measuring mode:In measuring mode:
In change mode:
In change mode: Changing parameters or settings. Run through the available characters
In change mode:In change mode: (including decimal point.
In measuring mode:
In measuring mode: Switch between measured value display and error messages /
In measuring mode:In measuring mode: warnings
Switch to one menu level higher
Return to measuring mode with a query whether the data should be accepted
Cancel the query, whether data should be accepted.
6.3.2 Navigation within the menu structure
Navigate through the menu using the →, ↑ and ^^^^ keys. Pressing the → key takes you one menu lower. Use the key to go one menu point higher (e.g. from 1 to 2). Pressing the ^^^^ key takes you one menu higher.
If you are already at the lowest level (function level), you can use the button to go to the change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^^^^ key to exit the menu mode and return to the measuring mode.
If settings were changed, the query whether they should be saved will appear. Confirm this query with the key or cancel it with the ^^^^ key.
Measuring
mode
Table 6-3: Navigation menu structure
Main menu↑→ Submenu↑→ Function↑→ Edit
^^^^ ^^^^ ^^^^ ^^^^
→ ↑ ^^^^
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
37
Page 38
6
OPERATION
6.3.3 Change the settings in the menu
Starting operation
Operation is started using the key
If a control lockout is set, the code (→ → → ^ ^ ^ ↑ ↑ ↑) must be entered. The code can be set in menu 3.13. The code shown here is set at the factory but not activated. A warning is displayed and the indicator returns to measuring mode if no key is pressed within 5 seconds or if an incorrect code is entered.
Exiting operator input
Operation is exited by pressing the ^^^^ key several times.
If data have been changed:
H250 M40
Save Yes Changes are accepted. An update is carried out and the display returns to
Save No ^^^^ Changes are discarded and the indicator jumps back to measuring operation.
measuring mode.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal plausibility check. If inplausible entries were made, a warning message is shown. If this warning is confirmed with the ^^^^ key, the display returns to the respective menu point without saving the relevant change. A new entry can now be made.
Example: Changing the flow unit from m3/h to l/h
Display Display
Example: 7.2
m
1x Fct. 1111
OPERATION
2x Fct. 3333
INSTALLATION
1x Fct 3.1111
LANGUAGE
10x Fct 3.11
FS&UNIT
3
/h
11
1111
1x Fct. 3.11.1111
1x 10.0000
4x 10000
3x ^^^^ 7200
FLOW
3
/h
m
l/h
Confirm with
deny ^
l/h
38
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 39
H250 M40
6.4 Overview of the units ESK4-T
Volume units can either represent real operating volumes (no prefix before the unit) or standard volumes virtually converted to reference statuses.
Prefix Volume definition
None
N
S
Operating volume flow e.g. m3/h or ft3/h Volume flow in normal state (0°C - 1.013 bara) according to DIN 1343 e.g. Nm Volume flow in standard state (15°C - 1.013 bara) according to ISO 13443 e.g. Sft
OPERATION
3
/h
3
/h
6
Measured
Units
variables
Operating volume flow
Standard volume flow
Standard volume flow
Mass flow g/s g/min g/h -
Operating volume totalizer
Standard volume totalizer
Standard volume totalizer
Mass totalizer kg g t lb
Temperature °C °F K
m3/s m3/min m3/h m3/d
L/s L/min L/h -
ft3/s ft3/min ft3/h ft3/d
gal/s gal/min gal/h gal/d
bbl/s bbl/min bbl/h bbl/d
ImpGal/s ImpGal/min ImpGal/h ImpGal/d
Nm3/s Nm3/min Nm3/h Nm3/d
NL/s NL/min NL/h -
Sm3/s Sm3/min Sm3/h Sm3/d
SL/s SL/min SL/h -
Sft3/s Sft3/min Sft3/h Sft3/d
kg/s kg/min kg/h kg/d
- t/min t/h t/d
lb/s lb/min lb/h lb/d
- STon/min STon/h STon/d
- - LTon/h LTon/d
3
m
ImpGal gallon bbl bbl (liq)
3
Nm
3
Sft
STon LTon
l hl
NL
SL
Sm
3
3
ft
In addition to the predefined units shown here, a user-defined unit can be activated in menu 3.12 by entering a conversion factor and a unit designation as free text.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
39
Page 40
6
OPERATION
6.5 Error messages ESK4-T
Error messages and warnings are indicated by one of the following symbols in the bottom left corner of the display. The ^^^^ key switches from the measured value display to the display of current errors / warnings. The table below contains a description of possible error messages.
H250 M40
Symbol NE107
Description Consequence
Category
FFFF Failure No measured value available
S Out of specification Measurement available but increasing
M Maintenance required Measurement still sufficiently accurate
C Function check Device is in test or linearization mode.
I Information No influence on the measurement,
Output signal invalid Error current signal is output.
measurement uncertainty. Device must be checked.
but maintenance required.
The output signal temporarily does not correspond to the measurement.
information only.
Error message Description Category Remedy
NOT LINEARIZED Linearization faulty or
NEW LINEARI. TABLE BAD
LINEARIZATIO UNDER CONFIG
UNIT SYSTEM CONFLICT
TOO FEW ENTRIES
NOT MONOTONOUS
FIRST NOT 0 % The first flow value if
LAST NOT 100 % The last flow value if
not activated = measuring error
Faulty or missing data in the linearization table = measuring error
The device is in linearization mode = measuring error
The unit for the linearization flow is incompatible with the selected flow type (mass/volume)
The linearization table has too few data points
The sequence of the linearization values is not strictly monotonic increasing
the linearization table is not 0%
the linearization table is not 100%
S Activate linearization or carry it out again
S
1
S Complete the linearization and activate it
S Correct error, carry out linearization again if
S Carry out linearization at at least 5 points
S Check linearization and/or carry it out again
®
(HART software are required; the original calibration values must be known), or send the device back to the manufacturer for linearization.
®
(HART software are required), or send the device back to the manufacturer for linearization.
necessary (HART linearization software are required), or send the device back to the manufacturer for linearization.
®
(HART software are required), or send the device back to the manufacturer for linearization.
®
(HART software are required), or send the device back to the manufacturer for linearization.
communication and linearization
communication and linearization
®
communication and
communication and linearization
communication and linearization
40
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 41
H250 M40
Error message Description Category Remedy
OPERATION
6
NO ZERO CAL OF AOThe analog output
NO F.SC. CAL OF AOThe analog output
NO TEMP. COMPENSATION
WRONG ELEC.REV.
OUTPUT NOT LINEARIZED
COUNTER LOST Totalizer value was
FRAM WRITE FAULT
ROM/FLASH ERROR
RESTART OF DEVICE
MULTIDROP MODE
CRYSTAL OSC FAULT
REF VOLTAGE FAULT
SENSOR A FAULT Internal error in
SENSOR B FAULT
zero point 4.00mA is not calibrated. = possible measuring error in process control.
100% = 20.00mA is not calibrated. = possible measuring error in process control.
The sensor temperature compensation of the device is faulty or was not carried out = possible measuring error
The ER revision of the ESK4 / ESK4A is not compatible with the add-on module, or the ribbon cable connection is not connected properly.
Linearization is not activated = measuring error
reset by error/overflow
Internal communication error
Memory error detected during self­test.
A device restart has taken place
The HART® multidrop mode is activated. The current output is set to a fixed value of
4.5 mA.
Internal error in device
Internal error in device
device
Internal error in device
S Perform calibration using ammeter and
menu 3.10 or using standard HART tools/process control system and poss. external ammeter. Caution: during calibration, switch the measuring point to manual control.
S Perform calibration using ammeter and
menu item 3.11 or using standard HART tools and external ammeter if necessary. Caution: during calibration, switch the measuring point to manual control.
S The device, together with an indication of the
S Check that the ribbon cable fits properly.
S Activate linearization or carry it out again
S
1
F Check whether the display is plugged in
F Restart device. If the error occurs again:
I The device has been restarted using menu
I
F The device must be sent back to the
F
1
error, must be sent back to the manufacturer for checking.
Check module compatibility (see chapter
2.5).
®
(HART software are required; the original calibration values must be known), or send the device back to the manufacturer for linearization.
Because the reset time is not known: Controlled reset of the counter using menu
item 1.6.1 or using HART control system.
correctly and restart the device. If the error occurs again: send the device back to the manufacturer with an indication of the error.
send the device back to the manufacturer with an indication of the error.
item 1.6.2 since the last time the error messages were reset.
The HART® multidrop mode is activated with selection of a polling address not equal to 0 using menu item 3.7. Polling address 0 activates the analogue output again.
manufacturer with an indication of the error.
communication and linearization
®
tools/process
®
®
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
41
Page 42
6
OPERATION
Error message Description Category Remedy
H250 M40
MEMORY CORRUPTION
AO FIXED The current output is
AO SATURATED Current output
ERROR TIMEOUT Data not transferred,
WARNING TIMEOUT
1 Category can be changed by user
Internal memory error, caused by a hardware or software problem
set to a fixed value.
saturated
or transferred incorrectly from the ESK to the counter module
F Restart the device; if the error occurs again
I The current output is fixed and does not
I The current output is saturated at >20.4 mA
F Confirm menu item 1.6.3 WRITE INFO I/O.
I
the device must be sent back to the manufacturer with an indication of the error.
reflect the measured value. This is the case in multidrop mode, with current output
test/calibration using the menu or HART
and is no longer linked to the measurement.
®
.
42
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 43
H250 M40
6.6 Menu indicator ESK4-T
6.6.1 Factory settings
Menu Function Setting
1.1.1 OUTPUT B1 INACTIVE
1.2.1 OUTPUT B2 INACTIVE
1.3.1 PULSE WDTH 100 ms
1.3.2 PULSE/UNIT 1 pulse / unit
1.4.1 DISPLAY Measured value
1.4.2 ROTATION 0°
1.5 TIME CONST. 1.0 s
1.6.1 RESET COUNT. NO
1.6.2 RESET ERROR NO
1.6.3 RE-INIT I/O NO
3.1 LANGUAGE ENGLISH
3.2 FUNCTION B1 INACTIVE
3.3 CONTACT B1 NO CONTACT
3.4 FUNCTION B2 INACTIVE
3.5 CONTACT B2 NO CONTACT
3.6 FUNCTION B3 INACTIVE
3.7 MULTIDROP POLLING ADD: 00
3.8 4mA CALIBR. 4.000 mA
3.9 20mA CALIBR. 20.000 mA
3.10 ALARM CURRENT ALARM HIGH
3.11.1 FS&UNIT Application specific
3.11.2 COUNTER Application specific
3.12 EDIT UNIT User defined unit / factor
3.13 LFC 4% ON, 6% OFF
3.14 DESCRIPTOR Free text
3.15 ENTRY CODE OFF
3.16 BASIC SETTING NO
OPERATION
6
INFORMATION!
The measuring device has been preset at the factory in accordance with the customer order. Therefore subsequent configuration via the menu is only necessary if the intended use of the device is changed.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
43
Page 44
6
OPERATION
6.6.2 Menu structure
Main menu Submenu 1 Submenu 2
H250 M40
1111 OPERATION 1.1 OUTPUT B1 1.1.1 INACTIV, MEAS.VALUE B1, COUNTER.VAL B1, PULSE
1.2 OUTPUT B2 1.2.1 INACTIV, MEAS.VALUE B2, COUNTER.VAL B2, PULSE
1.3 PULSEOUTP. 1.3.1 PULSE WDTH
1.4 DISPLAY 1.4.1 FLOW, COUNTER; FLOW/CNT; FLOW&CNT, PERCENT
1.5 TIME CONST. -
1.6 RESET 1.6.1 COUNTER
2222 TEST & INFO 2.1 4-20mA OUTPUT 2.1.1 NORMAL OP
2.2 ALARM CURR. Test of ALARM HIGH/LOW
2.3 OUTPUT B1 2.3.1 NORMAL OP
2.4 OUTPUT B2 2.4.1 NORMAL OP
2.5 INPUT B3 ACTIV HI INPUT; ACTIV LO, ON, OFF
2.6 DEVICE ID 2.6.1 ELEC. REV.
WDTH
1.1.2 HYST. B1, PULSE/UNIT
WDTH
1.2.2 HYST. B2, PULSE/UNIT
1.3.2 PULSE/UNIT
1.4.2 ROTATION
1.6.2 ERROR
1.6.3 WRITE INFO IO
2.1.2 4.0 mA
2.1.3 5.6 mA
2.1.4 7.2 mA
2.1.5 8.8 mA
2.1.6 10.4 mA
2.1.7 12.0 mA
2.1.8 21.6 mA
2.1.9 15.2 mA
2.1.10 16.8 mA
2.1.11 18.4 mA
2.1.12 20.0 mA
2.1.13 21.6 mA
2.3.2 OPEN
2.3.3 CLOSED
2.4.2 OPEN
2.4.3 CLOSED
2.6.2 SN ESK4
2.6.3 PA ORDER
2.6.4 DEVICE SN.
44
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 45
H250 M40
Main menu Submenu 1 Submenu 2
2222 TEST & INFO 2.7 SOFT.VERSION 2.7.1 FW. ESK4
2.7.2 FW. ESK4 I/O
2.8 TAG No. 8 characters
2.9 LONG TAG 32 characters
3333 INSTALLATION
3.1 LANGUAGE 3.1.1 ENGLISH
3.1.2 DEUTSCH
3.1.3 FRANCAIS
3.1.4 ITALIANO
3.1.5 ESPANOL
3.1.6 CESKY
3.1.7 POLSKI
3.1.8 NEDERLANDS
3.1.9 DANSK
3.2 FUNCTION B1 INACTIV, SWITCHPOINT, CONTERLIM., PULSEOUTP.
3.3 CONTACT B1 NO CONTACT, NC CONTACT
3.4 FUNCTION B2 INACTIV, SWITCHPOINT, CONTERLIM., PULSEOUTP.
3.5 CONTACT B2 NO CONTACT, NC CONTACT
3.6 FUNCTION B3 INACTIV, ACTIV HI, ACTIV LO, STARTH STOPL, STARTL STOPH
3.7 MULTIDROP POLLING ADR.
3.8 4mA CALIBR. 4.000 mA
3.9 20mA CALIBR. 20.000 mA
3.10 ALARM CURR. OFF, ALARM HIGH, ALARM LOW
3.11 FS&UNIT 3.11.1 MEAS.VALUE
3.11.2 COUNTER
3.12 EDIT UNIT 3.12.1 MEAS.VALUE
3.12.2 COUNTER
3.13 LFC 3.13.1 CONTROL ON, OFF
3.13.2 LFC ON VAL
3.13.3 LFC OFF VAL
3.14 DESCRIPTOR Free text
3.15 ENTRY CODE OFF, ON
3.16 BASIC SETTING SET ALL NO, SET ALL YES
OPERATION
6
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
45
Page 46
6
OPERATION
6.6.3 Menu explanations
1 OPERATION
H250 M40
Designation Level Select /
Explanation
Input
OUTPUT B1 1.1 Output B1 is a binary switching output. Under Fct. 3.2, one of the
One of the following functions can be selected as contact type under Fct. 3.3: NO CONTACT 1 / NC CONTACT 2
1.1.1 INACTIVE
MEAS.VALUE B1 Flow value switching point
COUNTER.VAL B1 Switching point totaliser value
PULSE WDTH Pulse value (pulse/unit)
1.1.2 HYST.B1 Hysteresis for the flow value switching point
following functions can be attributed to this output: INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.
Value range: 0.0measuring range end value The switching point is entered in flow units. If the current flow value exceeds this predetermined switching point, output B1 changes its binary state 3. Under Fct. 1.1.2 a hysteresis can also be specified.
Value range 0.0counter limit The switching point is entered in volume or mass units. If the the current counter value exceeds this predetermined switching point, output B1 changes its binary state 3. There is no hysteresis setting for the counter value switching point.
The value is now displayed here. Configuration takes place under Fct. 1.3.1 (PULSE WDTH), Fct. 1.3.2 (PULSE/UNIT) and Fct. 3.11.2 (COUNTER).
Value range 0.0switching point If the current flow value exceeds the predetermined switching point from Fct. 1.1.1, output B1 changes its binary state 3. In order for output B1 to change its binary state back to the initial setting, the switching point made smaller by the hysteresis must be undershot 4.
46
Example:
Example:
Example:Example: Under 1.1.1, a switching point of 200 L/h is set. The possible value range for the hysteresis is then 0.0200 L/h. With a hysteresis value of 0, the switching point has no hysteresis (3=4). If a hysteresis value of 20 L/h is entered, output B1 changes its binary state to the initial setting if 180 L/h is undershot 4.
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 47
H250 M40
OPERATION
6
Designation Level Select /
Explanation
Input
OUTPUT B2 1.2 Output B2 is a binary switching output. Under Fct. 3.4, one of the
One of the following functions can be selected as contact type under Fct. 3.5: NO CONTACT 1 / NC CONTACT 2
1.2.1 INACTIVE
MEAS.VALUE B2 See FLOW.VAL B1
COUNTER.VAL B2 See COUNTER.VAL B1
PULSE WDTH B2 see PULSE WDTH Fct. 1.1.1
1.2.2 HYST.B2 See HYST. B1
PULSEOUTP. 1.3
1.3.1 PULSE WDTH
50ms Ti = 50 ms; f
100ms Ti = 100 ms; f
200ms Ti = 200 ms; f
500ms Ti = 500 ms; f
1.3.2 PULSE/UNIT
0.001... 1000
following functions can be attributed to this output: INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.
Under Fct. 1.2.2 a hysteresis can also be specified.
Configuration takes place under Fct. 1.3.1 (PULSE WDTH), Fct. 1.3.2 (PULSE/UNIT) and Fct. 3.11.2 (COUNTER).
= 10 Hz
max. pulse / h = 36000
max. pulse / h = 18000
max. pulse / h = 9000
max. pulse / h = 3600
Pulse per volume or mass unit of counter (setting under Fct. 3.11.2), which can be output via one of the binary outputs. The maximum frequency for the pulse output (see Fct 1.3.1) cannot be exceeded even at maximum flow rate (final value).
max
max
max
max
= 5 Hz
= 2.5 Hz
= 1 Hz
Example:
Example:
Example:Example: Final value Qmax = 1200 l/h; volume unit counter = Litre; pulse duration = 100 ms; If the factor 1 is entered, 1 pulse/litre = 1200 pulses are generated in one hour at maximum flow rate. Number of max. permissible pulses:
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
47
Page 48
6
OPERATION
H250 M40
Designation Level Select /
Explanation
Input
DISPLAY 1.4 Different measured values can be selected for permanent or
1.4.1 Measured value permanent display of flow in flow units
COUNTER permanent display totalizer
MEAS.VAL/CTN alternating display flow value in flow units and totalizer
MEAS.VAL&CTN simultaneous display of flow value and totaliser
PERCENT permanent display flow value in percent
1.4.2 0° The reading is not rotated. 90° The reading on the display is rotated by 90°. 180° The reading on the display is rotated by 180°. 270° The reading on the display is rotated by 270°.
TIME CONST. 1.5 0.0...20.0s Given in seconds
RESET 1.6 Local resetting of totalizer and acknowledgement of warnings.
alternating display. The display can be rotated.
The output variables (value of the current loop and indicated flow value) follow the current process by the value set here (in seconds) with a time delay.
Note:
Note:
Note:Note: If the current flow is polled via HART
measured value is also played back with a delay.
There is always a safety prompt (yes/no), to avoid accidental resetting.
®
communication, the transferred
Note:
Note:
Note:Note: External resetting of the totalizer can be installed with the binary input B3.
1.6.1 COUNTER Confirming with YES resets the value of the totalizer to 0.0.
1.6.2 ERROR Confirming with yes acknowledges all existing warnings.
Note:
Note:
Note:Note: Confirming with YES acknowledges existing errors and warnings.
1.6.3 RE-INIT IO As a general rule, data is transferred from the counter module to the
ESK4 and vice versa when the device starts. To make sure, selecting this menu point when confirming with YES can perform another data transmission.
48
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 49
H250 M40
2 TEST & INFO
OPERATION
6
Designation Level Select /
Explanation
Input
4-20mA OUT 2.1 Testing the current loop by setting various current values
Note:
Note:
2.1.1 NORMAL OP The value of the current loop follows the current process.
2.1.2 4.0 mA The value of the current loop no longer follows the current process.
2.1.3 5.6 mA
2.1.4 7.2 mA
2.1.5 8.8 mA
2.1.6 10.4 mA
2.1.7 12.0 mA
2.1.8 21.6 mA
2.1.9 15.2 mA
2.1.10 16.8 mA
2.1.11 18.4 mA
2.1.12 20.0 mA
2.1.13 21.6 mA
OUTPUT B1 2.2 ALARM
CURRENT
2.3 Testing the binary switching output B1 by changing its binary status.
Note:Note: The simulation is not available in HART
Attention:
Attention:
Attention:Attention: During the test, the value of the current loop does not follow the current process.
It is set to the selected current value.
<3.6 / >21 mA Testing the alarm current HIGH /LOW according to setting Fct. 3.10
®
multi-drop mode (see Fct. 3.7)
Attention:
Attention:
Attention:Attention: During the test, the binary status does not correspond to the current process.
2.3.1 NORMAL OP The binary status of the switching output corresponds to the current
2.3.2 OPEN The binary status of the switching output no longer corresponds to the
2.3.3 CLOSED
OUTPUT B2 2.4 see test OUTPUT B1
2.4.1 NORMAL OP
2.4.2 OPEN
2.4.3 CLOSED
INPUT B3 2.5 TEST The current binary status of the binary input B3 is displayed. External
DEVICE ID 2.6 Information for device identification
2.6.1 ELEC. REV. Electronics revision
2.6.2 SN ESK4 Serial number ESK4
2.6.3 PA ORDER Production order for the complete flowmeter
2.6.4 DEVICE SN. Serial number for the complete flowmeter
process.
current process. The selected status is tested.
switching of the voltage applied to input B3 activates a change of the displayed binary status.
Note:
Note:
Note:Note: A reaction to the switching of the voltage applied to input B3 only occurs when the function of B3 is activated (see Fct. 3.6).
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
49
Page 50
6
OPERATION
H250 M40
Designation Level Select /
Explanation
Input
SOFT.VERSION 2.7 Information about software revision status
2.7.1 FW. ESK4 Software revision status for device insert ESK4-Basic
2.7.2 FW. ESK4 I/O Software revision status for optional add-on module ESK4-T
TAG NO. 2.8 8 digits TAG Designation
LONG TAG 2.9 32 digits TAG Designation
Alphanumerical characters. Up to eight digits possible
Alphanumerical characters. Up to 32 digits possible
3 INSTALLATION
Designation Level Select /
Input
LANGUAGE 3.1 Language selection for the menu texts displayed
3.1.1 ENGLISH The menu texts are displayed in the selected language.
3.1.2 DEUTSCH
3.1.3 FRANCAIS
3.1.4 ITALIANO
3.1.5 ESPANOL
3.1.6 CESKY
3.1.7 POLSKI
3.1.8 NEDERLANDS
3.1.9 DANSK
FUNCTION B1 3.2 INACTIVE The binary switching output B1 has no function.
SWITCHING POINT
COUNTERLIM. The binary switching point B1 functions as a limit switch depending on
PULSEOUTP. The binary switching point B1 functions as a pulse output depending on
Explanation
The binary switching point B1 functions as a limit switch depending on the current flow value. Configuration of the switching point takes place under Fct. 1.1.1 (MEAS.VALUE B1)
the current counter value. Configuration of the switching point takes place under Fct. 1.1.1 (MEAS.VALUE B1)
the current flow value. Pulses can be generated up to a maximum of 10 Hz. Configuration occurs under Fct. 1.3.1 (PULSE WDTH) and Fct. 1.3.2 (PULSE/UNIT).
Note:
Note:
Note:Note: Configuration under Fct 1.3.1 and Fct. 1.3.2 applies to both pulse outputs. If both B1 and B2 are configured as pulse outputs, both binary outputs behave in exactly the same way.
CONTACT B1 3.3 NO CONTACT Binary switching output B1 is an NO contact.
NC CONTACT Binary switching output B1 is an NC contact.
50
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 51
H250 M40
OPERATION
6
Designation Level Select /
Explanation
Input
FUNCTION B2 3.4 INACTIVE See FUNCTION B1
SWITCHING POINT
COUNTERLIM. See FUNCTION B1
PULSEOUTP. See FUNCTION B1
CONTACT B2 3.5 NC CONTACT See CONTACT B1
NO CONTACT See CONTACT B1
FUNCTION B3 3.6 INACTIVE The binary switching input B3 has no function.
ACTIV H The internal flow counter is reset to 0.0 if there is an H-level at input B3
ACTIV L The internal flow counter is reset to 0.0 if there is an H-level at input B3
STARTH STOPL The counter is started by creating an H level at the B3 input and stopped
STARTL STOPH The counter is started by creating an L level at the B3 input and stopped
MULTIDROP 3.7 015
See FUNCTION B1 Configuration of switching point occurs under Fct. 1.2.1 (MEAS. VALUE B2).
Configuration of switching point occurs under Fct. 1.2.1 (MEAS. VALUE B2).
Configuration occurs under Fct. 1.3.1 (PULSE WDTH) and Fct. 1.3.2 (PULSE/UNIT).
for at least 100 ms.
for at least 100 ms.
by creating an L level at the B3 input.
by creating an H level at the B3 input.
Polling adress for HART® multi-drop mode When the address is 0, the HART
®
multi-drop mode is disabled.
Attention:
Attention:
Attention:Attention: When the HART
current loop is inactive (fixed current value of 4.5 mA) and no longer follows the current process.
4mA CALIBR. 3.8 D/A calibration for beginning of measuring range (4 mA)
Note:
Note:
Note:Note: This function is not available in HART
Attention:
Attention:
Attention:Attention: During calibration, the value of the current loop does not follow the current process. If a current measuring device in the 4...20 mA loop detects a deviation from the desired value 4.000 mA, the measured value must be entered. The correction value is accepted following the query SAVE and Confirm with YES.
20mA CALIBR. 3.9 D/A calibration for the final value of the measuring range (20 mA)
Note:
Note:
Note:Note: This function is not available in HART
Attention:
Attention:
Attention:Attention: During calibration, the value of the current loop does not follow the current process. If a current measuring device in the 4...20 mA loop detects a deviation from the desired value 20.000 mA, the measured value must be entered. The correction value is accepted following the query SAVE and Confirm with YES.
®
multi-drop mode (Address 115) is activated, the
®
multi-drop mode.
®
multi-drop mode.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
51
Page 52
6
OPERATION
H250 M40
Designation Level Select /
Explanation
Input
ALARM CURRENT
END&UNIT 3.11 By changing the unit, the respective final value is scaled accordingly.
EDIT UNIT 3.12 User-defined unit with conversion to calibrated unit
LFC 3.13 LFC stands for LLLLow FFFFlow CCCCutoff
3.10 OFF Error indication via the current loop is deactivated. The current loop
ALARM HIGH Error indication via the current loop is activated (failure signal "high"
ALARM LOW Error indication via the current loop is activated (failure signal "low" as
3.11.1 Measured value For units for volume flow or mass flow, refer to
3.11.2 COUNTER For units for volume counter or mass counter and pulse output refer to
follows the current process.
Note:
Note:
Note:Note: This function is not available in HART
acc. to NE43).
Note:
Note:
Note:Note: This function is not available in HART
per NE43).
Note:
Note:
Note:Note:
1) This function is not available in HART
2) This function is supported as of Electronic Revision 2.2.x.
Depending on calibration, either units for volume or mass flow measurement are available for selection.
®
multi-drop mode.
®
multi-drop mode.
®
multi-drop mode.
Overview of the units
ESK4-T
Overview of the units ESK4-T
To ensure a stable zero point of the current output, the current output can be stably set to 4.00 mA 3 in a range to be selected.
on page 39
on page 39
3.13.1 CONTROL OFF LFC function is not active.
CONTROL ON LFC function is active.
3.13.2 LFC ON_VALUE Switch-on value 1:
3.13.3 LFC OFF_VALUE Switch-off value 2:
DESCRIPTOR 3.14 12 digits Free text input for header in LCD
Value range 119% (from final value of measuring range) Flow is greater than the switch-on value. The current output corresponds to this. If the flow rate falls, the current output follows until the switch-on value
1. If the flow value continues to fall, the current output is switched to
4.00 mA 3.
Note:
Note:
Note:Note: The switch-on value to be set must be smaller than the previously selected switch-off value.
Value range 220% (from the measuring range final value) The flow rate is 0. The current output is 4.00 mA 3. If the flow rate increases, the current output remains at 4.00 mA until it has reached the switch-off value 2.
Note:
Note:
Note:Note: The switch-off value to be set must be greater than the previously selected switch-on value.
52
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 53
H250 M40
OPERATION
6
Designation Level Select /
Explanation
Input
INP. CODE 3.15 Input code for the local operating menu
3.15.1 OFF Use of an input code is not activated.
3.15.2 ON If YES is selected, the last code that was entered has to be typed in.
BASIC SETTING 3.16 Resetting parameters to factory default
In the factory setting, the input code is not active.
Factory code: → → → ^ ^ ^ ↑ ↑ ↑ If, after confirmation with YES, the → button is also pressed, then a new, individual, nine-element code can be typed in. The desired button combination is visualized on the display.
There is always a safety prompt (yes/no), to avoid accidental resetting.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
53
Page 54
7
SERVICE
7.1 Maintenance
Within the scope of routine maintenance of the system and pipelines, the flowmeter should also be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to the measuring tube and indicator. We advise that inspections be carried out at least once per year. The device must be removed from the piping before cleaning.
CAUTION!
Pressurized pipes must be depressurized before removing the device. Empty pipes as completely as possible. In the case of devices used for measuring aggressive or hazardous media, appropriate safety precautions must be taken with regard to residual liquids in the measuring unit. Always use new gaskets when reinstalling the device in the pipeline. Avoid electrostatic charges when cleaning the surfaces (e.g. sight window)!
7.2 Replacement and retrofitting
H250 M40
Some of the variable area flowmeter components can be retrofitted:
Float damping
Indicator M40:
Indicator M40:
Indicator M40:Indicator M40:
Limit switch module K1 / K2
Current output 4..20mA ESK4 / ESK4A
Counter module with LCD and I/O ESK4-T
Fieldbus interface ESK4-PA / FF
7.2.1 Replacing floats
Remove the device from the piping.
Take the upper snap ring out of the measuring unit.
Take the upper float stop and float out of the measuring unit.
Insert the new float into the centre hole of the lower float stop and push into the measuring
unit along with the upper float catcher. While doing this, the float's upper guide rod must be guided through the middle hole of the float stop.
Insert the snap ring into the measuring unit.
Fit the device back into the piping.
CAUTION!
An additional measuring error is to be expected if recalibration is not conducted.
54
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 55
H250 M40
7.2.2 Retrofitting float damping
Take the upper snap ring 1 out of the measuring unit.
Take the upper float stop 2 and float 5 out of the measuring unit.
Fasten the locking ring 3 into the lower slot of the float's guide rod.
Slide ceramic sleeve 4 on to the float's guide rod and attach it to the top slot using the locking
ring 3.
Insert float into the lower float guide in the measuring unit.
Retrofit the supplied damping cylinder with the integrated float stop 2 into the measuring
unit.
Insert upper snap ring 1.
SERVICE
7
1 Snap ring 2 Float stop 3 Locking ring 4 Ceramic sleeve 5 Float
7.2.3 Retrofitting limit switch
Remove ESK4 / ESK4A add-on module (if available).
Merge contact pointer 2 in the middle.
Loosen the locking screw 1 on the contact pointer.
Insert the contact module into slot 3 of the bracket until the semi-circle 1 on the contact
board surrounds the pointer cylinder.
The contact module connecting terminals feature a pluggable design and can be removed in order to connect the cables.
WARNING!
Do not damage the indicating element!
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
55
Page 56
7
SERVICE
7.2.4 Replacement - Retrofitting ESK4 / ESK4A
When replacing or retrofitting an ESK4 / ESK4A, the following are required at the time of ordering:
SN - serial number or
SO - sales order
This information can be found on the indicator nameplate.
The ESK4 / ESK4A is factory calibrated, making it possible to replace it or retrofit it without recalibrating.
De-energise the ESK4 / ESK4A.
Lift and remove the ESK4 / ESK4A with a screwdriver.
WARNING!
Do not damage the indicating element!
H250 M40
The ESK4 / ESK4A plug-in tongues 1 are plugged in under the two bolts 2 on the baseplate.
Slight pressure is used to press the ESK4 / ESK4A onto the spring pins 3 until it stops, firmly
attaching the ESK4 / ESK4A.
If a change in measuring range, product temperature, product, density, viscosity or pressure is
®
desired, this can be done with the KroVaCal program and with a HART measuring unit is subject to its own physical limits, which the KroVaCal program correctly calculates, and may thus reject the desired change. If a change is performed using the program, the new data is also transmitted to the ESK4 / ESK4A.
Device identification
Device address
Serial Number
Measuring point identifier
Digital measured value query in flow units, % and mA
Test / setting functions
Calibration 4.00 and 20.00 mA
Set current output to any desired value
modem. However, each
56
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 57
H250 M40
7.2.5 Replacement - Retrofitting add-on module ESK4-T / PA / FF
The add-on module for the ESK4 / ESK4A can be replaced or retrofitted on-site without removing the device from the process control.
ESK4-T (Display module with LCD and I/O)
ESK4-PA (Profibus PA interface)
ESK4-FF (Foundation Fieldbus interface)
More detailed information can be found in the service manual included with each retrofit or replacement set.
7.3 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal operating conditions.
SERVICE
7
7.3.1 List of spare parts
Spare part Order no.
DN 15
DN 15
DN 15DN 15
Float CIV 15, 1.4404 X251041000
Float DIV 15, 1.4404 X251042000
Float TIV 15, 1.4404 X251043000
Float DIVT 15, 1.4404 X251044000
Float TIV 15, Aluminium X251043100
Float TIV 15, Titanium X251043200
Set float stop; standard (1 float stop, 1 span ring) X251050100
Set float stop; gas damping (ZrO2) X251050200
Set of float stop; gas damping (PEEK) X251050300
Damping bush (7x8) ZrO2 incl. 2 locking rings X251053100
Damping bush (7x8) PEEK incl. 2x locking rings X251053200
DN 25
DN 25
DN 25DN 25
Float CIV 15, 1.4404 X252041000
Float DIV 25, 1.4404 X252042000
Float TIV 25, 1.4404 X252043000
Float DIVT 25, 1.4404 X252044000
Set float stop; standard (1 float stop, 1 span ring) X252050100
Set float stop; gas damping (ZrO2) X252050200
Set of float stop; gas damping (PEEK) X252050300
Damping bush (7x8) ZrO2 incl. 2x locking rings X252053100
Damping bush (12x8) PEEK incl. 2x locking rings X252053200
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
57
Page 58
7
SERVICE
H250 M40
Spare part Order no.
DN 50
DN 50
DN 50DN 50
Float CIV 55, 1.4404 X253041000
Float DIV 55, 1.4404 X253042000
Float TIV55, 1.4404 X253043000
Float DIVT 55, 1.4404 X253044000
Set float stop; standard (1 float stop, 1 span ring) X253050100
Set float stop; gas damping (ZrO2) X253050200
Set of float stop; gas damping (PEEK) X253050300
Damping bush (7x8) ZrO2 incl. 2x locking rings X253053100
Damping bush (14x10) PEEK incl. 2x locking rings X253053200
DN 80
DN 80
DN 80DN 80
Float CIV 85, 1.4404 X254041000
Float DIV 85, 1.4404 X254042000
Float TIV 85, 1.4404 X254043000
Float DIVT 85, 1.4404 X254044000
Set float stop; standard (1 float stop, 1 span ring) X254050100
Set float stop; gas damping (ZrO2) X254050200
Set of float stop; gas damping (PEEK) X254050300
Damping bush (18x14) ZrO2 incl. 2x locking rings X254053100
Damping bush (18x14) PEEK incl. 2x locking rings X254053200
DN 100
DN 100
DN 100DN 100
Float CIV 105, 1.4404 X255041000
Float DIV 105, 1.4404 X255042000
Float DIVT 105, 1.4404 X255044000
Set float stop; standard (1 float stop, 1 span ring) only for bottom ! X255050100
Set float stop; gas damping (ZrO2) X255050200
Set of float stop; gas damping (PEEK) X255050300
Damping bush (18x14) ZrO2 incl. 2x locking rings X254053100
Damping bush (18x14) PEEK incl. 2x locking rings X254053200
Indicator M40:
Indicator M40:
Indicator M40:Indicator M40:
Housing components
Housing components
Housing componentsHousing components
Standard indicator housing M40 complete no scale X251110000
Standard indicator housing M40R complete, no scale* (stainless steel, non-coated) X251111000
Standard indicator housing M40S complete, no scale * (2-layer coating) X251112000
Cover M40 complete, standard X251110100
Standard cover M40S (2-layer coating) X251110200
Standard cover M40R (stainless steel, non-coated) X251110400
Cover seal X251112100
M40 baseplate standard X251120100
Standard M40S baseplate (2-layer coating) * X251120200
Standard M40R baseplate (stainless steel, non-coated) * X251120300
Standard = Not pressure tight
Standard = Not pressure tight
Standard = Not pressure tightStandard = Not pressure tight
58
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 59
H250 M40
SERVICE
Spare part Order no.
Housing components indicator M40
Housing components indicator M40
Housing components indicator M40Housing components indicator M40
Retrofit kit HT extension X251021000
Module carrier (profile track) X251121100
Set housing attachment parts X251121300
Pointer system, complete * X251122100
Magnetic brake for indicating element X251122200
* Accuracy loss without recalibration!
Namur 2-wire limit module
Namur 2-wire limit module
Namur 2-wire limit moduleNamur 2-wire limit module
Contact module K1 min I7S23,5-N X251135100
Contact module K1 max I7S23,5-N X251135200
Contact module K2 min / max I7S23,5-N X251135300
Contact module K1 min SC3,5 N0 X251133100
Contact module K1 max SC3,5 N0 X251133200
Contact module K2 min / max SC3,5 N0 X251133300
Contact module K2 min / min min - SJ3,5 S1N / SJ 3,5 SN X251133400
Contact module K2 max / max max - SJ3,5 S1N / SJ 3,5 SN X251133500
Contact module K1 min SJ3,5 SN X251133600
Contact module K1 max SJ3,5 SN X251133700
Contact module K1 min / max SJ3,5 SN X251133800
7
3-wire limit module NC contact
3-wire limit module NC contact
3-wire limit module NC contact3-wire limit module NC contact
Contact module K1 min SB3,5 E2 - act.Low X251133900
Contact module K1 max SB3,5 E2 - act.Low X251134000
Contact module K1 min / max SB3,5 E2 - act.Low X251134100
3-wire limit module normally open
3-wire limit module normally open
3-wire limit module normally open3-wire limit module normally open
Contact module K1 min SB3,5 E2 - act.High X251134200
Contact module K1 max SB3,5 E2 - act.High X251134300
Contact module K1 min / max SB3,5 E2 - act.High X251134400
Electronic modules
Electronic modules
Electronic modulesElectronic modules
ESK4, ESK4-FF, ESK4-PA, ESK4-FF (serial number required)
Cover for add-on modules X251121500
Connection cable ESK4 to add-on modules X251121600
Plugs (10 pcs) for ESK4 connection for bus modules X251132500
Replacement connector
Replacement connector
Replacement connectorReplacement connector
Replacement connector 11/12 for ESK4 X251121700
Replacement connector D/D+ for ESK4-FF or ESK4-PA X251121800
Replacement connectors 1/2/3, 4/5/6, 7/8 for ESK4-T X251121900
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
59
Page 60
7
SERVICE
Spare part Order no.
Replacement cable gland
Replacement cable gland
Replacement cable glandReplacement cable gland
Single cable gland M20x1.5 black, plastic - for non-Ex / Ex-i X251150300
Single cable gland M20x1.5 blue, plastic - for non-Ex / Ex-i X251150100
Single cable gland M20x1.5 brass - for non-Ex / Ex-i / Ex-nA X251151000
Single cable gland M20x1.5 brass Ex-d/t - for Ex-d / Ex-t X251152000
Single plug M20x1.5 brass Ex-d/t - for Ex-d / Ex-t X251153000
Single cable gland M20x1.5 stainless steel Ex-d/t - for Ex-d / Ex-t X251154000
Single plug M20x1.5 stainless steel Ex-d/t - for Ex-d / Ex-t X251155000
Other spare parts on request
H250 M40
60
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 61
H250 M40
7.4 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
7.5 Returning the device to the manufacturer
7.5.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of the personnel, the manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
SERVICE
7
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
61
Page 62
7
SERVICE
7.5.2 Form (for copying) to accompany a returned device
CAUTION!
To avoid any risk for our service personel, this form has to be accessible from outside of the packaging with the returned device.
Company: Address:
Department: Name:
Tel. no.: Fax no. and/or Email address:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
This medium is: radioactive
water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
H250 M40
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
7.6 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
According to the directive 2012/19/EU, the monitoring and control instruments marked with the WEEE symbol and reaching their end-of-life must not be disposed of with other waste The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or send them back to our local organisation or authorised representative.
must not be disposed of with other waste.
must not be disposed of with other wastemust not be disposed of with other waste
62
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 63
H250 M40
8.1 Functional principle
The H250 flowmeter operates in accordance with the float measuring principle. The measuring unit consists of a metal cone in which a float can move freely up and down. The flow goes from bottom to top. The float changes position so that the lifting force acting on it F1 is in equilibrium with the form drag F2 and its weight F3: F3 = F1 + F2
TECHNICAL DATA
8
1 Indication principle M40 magnetic coupling 2 Magnetic coupling sensors
1 For the indicator, the flow-dependent height of the float in the measuring unit is transmitted
by means of a magnetic coupling and displayed on a scale.
2 For a built-in signal converter (ESK4 / ESK4A), the flow-dependent height of the float in the measuring unit is detected by the S1 and S2 magnetic field sensors and electronically processed.
Operating principle of H250H and H250U
Figure 8-1: Operating principle H250H and H250U
1 H250H - horizontal flow direction 2 H250U - flow direction from top to bottom
The flowmeter operates in accordance with a modified float measuring principle. The guided float adjusts itself so that the flow force acting on it is in equilibrium with the opposing spring force. The flow-dependent position of the float in the measuring unit is displayed on a scale by means of a magnetic coupling.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
63
Page 64
8
TECHNICAL DATA
8.2 Technical data
INFORMATION!
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Application range Flow measurement of liquids, gases and vapours
Function / Measuring principle Variable area measuring principle
Measured value
Primary measured value Float position
Secondary measured value Operating and standard volumetric flow
Measuring accuracy
Directive VDI / VDE 3513, sheet 2 (qG = 50%)
H250 /RR /HC /F 1.6%
H250/C (Ceramic, PTFE) H250H, H250U, H250 (100 : 1)
H250 M40
2.5%
Operating conditions
Temperature
Temperature
TemperatureTemperature Max. operating temperature TS -196..+300°C / -321...+572°F
Pressure
Pressure
PressurePressure
Max. operating pressure PS Depending on the version (refer to nameplate)
Max. test pressure PT Depending on the version (refer to nameplate)
Min. required operating pressure 2 times greater than pressure loss (refer to measuring
Protection category
Protection category
Protection categoryProtection category
M40, M40S, M40R IP66/68 acc. to EN 60529, NEMA 4/4X/6 acc. to NEMA 250
M40R IP69K acc. to DIN 40050-9
Float damping during gas measurement recommended
Float damping during gas measurement recommended
Float damping during gas measurement recommendedFloat damping during gas measurement recommended DN15...25 / ½...1" Operating pressure: <0.3 bar / 4.4 psig
DN50...100 / 2...4" Operating pressure: <0.2 bar / 2.9 psig
Depending on the version (refer to nameplate)
ranges)
Installation conditions acc. to VDI/VDE 3513 Sheet 3
Inlet section 5 x DN Outlet section 3 x DN
64
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 65
H250 M40
Materials
TECHNICAL DATA
8
Device Flange
H250/RR Stainless steel CrNi 1.4404 / 316L
H250/HC
H250/F - Food CrNi steel 1.4435 CrNi steel 1.4435 / 1.4404
H250/C Ceramics/PTFE
1
1 DN100/4" only PTFE 2 TFM/PTFE liner (electrically non-conductive), conductive PTFE on request
Hastelloy® C-22 (2.4602) solid or plated
CrNi steel 1.4571 with TFM/PTFE
2
Measuring tube
Hastelloy® C4
Float Float guide Ring orifice
PTFE or Al2O3 with FFKM gasket
Al2O
3
and PTFE
Al2O
3
Other options on request.
Special materials: e.g. SMO 254/6Mo, Titan, Hastelloy® C276/2.4819, Monel / 2.4360 a.o.
Float damping: ceramic or PEEK
Standard gasket for devices with female thread as insert: O-Ring FPM / FKM, others optional
e.g. FFKM, EPDM
M40 Aluminium, single-layer powder coating (polyester)
M40S Aluminium, two-layer powder coating (epoxy / polyester)
M40R Stainless steel without coating 1.4408 / CF8M
Offshore Wet coating on request
Temperatures
Temperatures
TemperaturesTemperatures
For devices to be used in hazardous areas, special temperature ranges apply. These can be found in the separate instructions.
Temperatures H250/M40 - mechanical indicator without power supply
Material Product temperature Ambient temperature
Float Liner [°C] [°F] [°C] [°F]
H250/RR Stainless steel -196...+300 -321...+572 -40...+120 -40+248 H250/RR screw fitting FPM/FKM -20...+200 -4...+392 -20+120 -4+248
H250/HC H250/C PTFE -196...+70 -321...+158 -40...+70 -40+158
H250/C Ceramic PTFE -196...+150 -321...+302 -40...+70 -40+158 H250/C Ceramic TFM / Ceramic -196...+250 -321...+482 -40...+120 -40+248
H250 H/U Spring material: stainless steel
Hastelloy
1.4301
Spring material: Hastelloy® 2.4610
®
-196...+300 -321...+572 -40...+120 -40+248
-40...+100 -40+212 -40+120 -40+248
-40...+200 -40...+392 -40+120 -40+248
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
65
Page 66
8
TECHNICAL DATA
H250 M40
Minimum ambient temperatures T
with electrical components
amb.
Design [°C] [°F]
ESK4, ESK4A, ESK4-FF, ESK4-PA
Limit switches SJ3,5-SN / I7S23,5-N / Reed SPST -40...+70 -40...+158
Limit switches SC3,5-N0 / SJ3,5-S1N / SB3,5-E2 -25...+70 -13...+158
1 Display contrast out of the temperature range 0...+60°C / +32..+140°F decreasing.
1
-40...+70 -40...+158
Temperatures H250/M40 - with electrical components [°C]
T
< +40°C T
amb.
EN ASME Version with Standard HT Standard HT
DN15, DN25
DN 50 2 ESK4 / ESK4A, -FF, -PA +200 +300 +165 +300
DN 80, DN100
½", 1" ESK4 / ESK4A, -FF, -PA +200 +300 +180 +300
ESK4-T +200 +300 +80 +130
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +200 +300 +130 +295
ESK4-T +180 +300 +75 +100
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +200 +300 +120 +195
3, 4" ESK4 / ESK4A, -FF, -PA +200 +300 +150 +250
ESK4-T +150 +270 +70 +85
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +190 +300 +110 +160
< +60°C 1
amb.
Maximum product temperatures H250/M40 - with electrical components [°F]
T
< +104°F T
amb.
EN ASME Version with Standard HT Standard HT
DN15, DN25
DN 50 2 ESK4 / ESK4A, -FF, -PA 392 572 165 572
DN 80, DN100
½", 1" ESK4 / ESK4A, -FF, -PA 392 572 356 572
ESK4-T 392 572 176 266
Limit switch NAMUR 392 572 392 572
3-wire limit switch 392 572 266 563
ESK4-T 356 572 167 212
Limit switch NAMUR 392 572 392 572
3-wire limit switch 392 572 248 383
3, 4" ESK4 / ESK4A, -FF, -PA 392 572 302 482
ESK4-T 302 518 158 185
Limit switch NAMUR 392 572 392 572
3-wire limit switch 374 572 230 320
1 if there are no heat insulation measures, a heat-resistant cable is necessary (continuous operating temperature of the cable to be used: +100°C / +212°F)
< +140°F 1
amb.
66
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 67
H250 M40
TECHNICAL DATA
Abbreviation
HT High-Temperature version ESK4 / ESK4A 2-wire current output 420 mA with HART 5 / HART 7
ESK4-T ESK4 with LCD, binary status outputs, digital counter and pulse output.
ESK4-FF FOUNDATION FIELDBUS interface
ESK4-PA PROFIBUS PA interface
Cable glands
Cable gland Material Cable diameter
M 20x1.5 Standard PA 8...13 mm 0.315...0.512"
M 20x1.5 Nickel-plated brass 10...14 mm 0.394...0.552"
M 20x1.5 Stainless steel 10...14 mm 0.394...0.552"
Limit switches K1/K2
8
Terminal connection
Limit switches I7S23,5-N
NAMUR (IEC 60947-5-6) Yes Yes Yes No No
Type of connection 2-wire 2-wire 2-wire 3-wire 2-wire
Switching element function
Nominal voltage U
Pointer vane not detected 3 mA 3 mA 1 mA 0.3 VDC U
Pointer vane detected 1 mA 1 mA 3 mA UB - 3 VDC 0 VDC
Continuous current - - - max. 100 mA max. 100 mA
No load current I
Switching cycles - - - - 100000
1 safety oriented 2 no inductances
0
0
2.5 mm
SC3,5-N0
NC contact
8.2 VDC 8.2 VDC 8.2 VDC 10...30 VDC max. 32 VDC
- - - 15 mA -
2
SJ3,5-SN 1SJ3,5-S1N
NC contact NO contact PNP NO
1
SB3,5-E2 Reed
contact
NC contact SPST
0
Current output ESK4 / ESK4A
Terminal connection
Power supply 14...30 VDC (12...30 VDC without ESK4-T)
Min. power supply for HART
Measuring signal 4.00...20.00 mA = 0...100% flow value in 2-wire technology
Power supply influence <0.1%
External resistance dependency <0.1% Temperature influence <5 µA / K
Max. external resistance / load 650 Ohm (30 VDC)
Min. load for HART
®
®
2
2.5 mm
20 VDC, load 250 Ohm
250 Ohm
2
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
67
Page 68
8
TECHNICAL DATA
ESK4 HART® configuration
Manufacturer name (code) KROHNE Messtechnik (0x45 = 69)
Model name / HART® Revision
Physical layer FSK
ESK4 / ESK4A process variable
Over range +102.5 (±1%) 20.24...20.56
Device error identification > 106.25 >21.00
Max. current consumption 131.25 25
Multi-drop mode 4.5
ESK4 (214 = 0xD6) / HART 5.9
ESK4A (17854 = 0x45BE) / HART 7.4
Values [%] from full scale range
Signal output [mA]
H250 M40
ESK4-FF
Physical layer IEC 61158-2 and FISCO model
Communication standard H1 FOUNDATION Fieldbus protocol
ITK version 5.2
Power supply Bus supply
Nominal current 16 mA
Error current 23 mA
Boot-up current after 10 ms < Nominal current
For more details refer to the supplementary instructions "H250 M40 Foundation Fieldbus".
ESK4-PA
Physical layer IEC 61158-2 and FISCO model
Communication standard Profibus PA profile 3.02
PNO ID 4531 HEX
Power supply Bus supply
Nominal current 16 mA
Error current 23 mA
Boot-up current after 10 ms < Nominal current
68
For more details refer to the supplementary instructions "H250 M40 Profibus PA".
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 69
H250 M40
TECHNICAL DATA
ESK4-T with LCD, binary inputs and outputs and digital counter
Binary output
Two binary outputs Galvanically isolated, passive
Mode Switch output NAMUR or transistor (open
Configurable as Switch contact or
NAMUR switch output
Power supply 8.2 VDC
Signal current > 3 mA switching value not reached < 1 mA switching value reached
Switch output transistor (open collector)
Power supply Nominal 24 VDC, maximal 30 VDC
P
max
Continuous current Max. 100 mA
No load current I
0
pulse output
500 mW
2mA
collector)
NC/NO contact or max. 10 pulses/s
8
Pulse output
T
on
T
off
Pulse value
Configurable from 50...500 ms
Depending on flow rate
Configurable in flow units e.g. 5 pulses/m
3
Binary input
Input Galvanically isolated
Mode Reset counters or run/stop
Configurable as active HI / active LO
H signal 16...30 VDC
Internal resistance R
Ton (active) 500 ms
i
Typically 20 kΩ
LC display
Technology Passive graphic LCD
Display Measurement with units and/or counter reading with units. Counter
Configuration Local plain text menu navigation via microswitch or bar magnet or using
reading max. 11-digit with power failure proof storage. Binary flags for limit value status. 0...100% bar graph for measurement. NE 107 diagnostic status symbols. Plain text menu for configuration.
DD/DTM software
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
69
Page 70
8
TECHNICAL DATA
Approvals
Standard Display Marking
ATEX / IECEx
FM (USA) NEC500 FM (Kanada) NEC505
NEPSI M40 electrical Ex ia, Ex d, Ex nA, Ex t
CCOE/PESO M40 electrical Ex ia, Ex d
EAC/GOST M40 electrical Ex ia, Ex d
INMETRO M40 electrical Ex ia, Ex d, Ex nA, Ex t
KGS M40 electrical Ex ia, Ex d, Ex nA, Ex t
H250 M40
M40 mechanical II2GD IIC
II3GD IIC
M40 electrical II2G Ex ia IIC T6 Gb
II2G Ex d IIC T6 Gb II3G Ex nA IIC T6 Gc II2D Ex t IIIC T70°C Db II2D Ex ia IIIC T85°C Db
M40 electrical IS Class I Div 1
XP Class I Div 1 NI Class I Div 2 DIP Class II / III Div 1
70
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 71
H250 M40
TECHNICAL DATA
8.3 Dimensions and weights
Dimensions H250/M40
Front view Side view with heating High temperature
8
a b d h
[mm] ["] [mm] ["] [mm] ["] [mm] ["]
H250/RR Flange H250/F Clamp connection
H250/RR from 2" 600 lb ISO 228 / ASME B1.20.1 / SMS
141 5.56 250 9.85 150 5.91 150 5.91
300 11.82
EN ASME c 1 e 2 Ø f g j
[mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]
DN15 ½" 94 3.70 114 4.49 20 0.80 97 3.82 197 7.76
DN25 1" 94 3.70 127 5.00 32 1.28 109 4.27 209 8.23
DN50 2" 107 4.22 141 5.55 65 2.57 125 4.90 222 8.74
DN80 3" 107 4.22 157 6.18 89 3.51 143 5.61 238 9.37
DN100 4" 107 4.22 167 6.57 114 4.50 150 5.91 248 9.76
1 without cable gland 2 Ex d, Ex t, Ex nA +10mm [0,39"]
ISO 228 / ASME B1.20.1
Female thread screwed
ISO 228 / ASME B1.20.1
Female thread welded
H250/F 1 Clamp
connection
H250/F Screw conn.
DIN 11851
1 Stainless steel 1.4435 - EHEDG tested - wetted surfaces Ra ≤ 0.8 / 0.6 µm
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
71
Page 72
8
TECHNICAL DATA
Weights
Weights
WeightsWeights
Nominal meter size EN 1092-1 Flange connection Ermeto connection
EN ASME [kg] [lb] [kg] [lb] [kg] [lb]
DN15 ½“ 3.5 7.7 5.6 12.6 3.9 8.6 DN25 1 5 11 7.5 16.5 5.8 12.8 DN50 2 8.2 18.1 11.2 24.7 9.5 21 DN80 3 12.2 26.9 14.8 32.6 13.1 28.9 DN100 4 14 30.9 17.4 38.4 15.7 34.6
Nominal meter size EN 1092-1 ASME 150 lb ASME 300 lb DIN 11864-1
EN ASME [kg] [lb] [kg] [lb] [kg] [lb] [kg] [lb]
DN15 ½“ 3.5 7.7 3.2 7.1 3.5 7.7 2 4.4 DN25 1 5 11 5.2 11.5 6.8 15 3.5 7.7 DN50 2 10 22.1 10 22.1 11 24.3 5 11 DN80 3 13 28.7 13 28.7 15 33.1 7.6 16.8 DN100 4 15 33.1 16 35.3 17 37.5 10.3 22.7
H250 M40
H250 with heating
H250/C [Ceramic / PTFE] Screw connec.
Process connections
Process connections
Process connectionsProcess connections
Standard Conn. dim. Pressure rating
Flanges (H250/RR /HC /C) EN 1092-1 DN15...150 PN16...250
ASME B16.5 ½...6" 150...2500 lb JIS B 2220 15100 10...20K
Clamp connections (H250/RR /F) DIN 32676 DN15...100 10...16 bar
ISO 2852 Size 25...139.7 10...16 bar
Screw connections (H250/RR /HC /F) DIN 11851 DN15...100 25...40 bar
SMS 1146 1...4" 6 bar / 88.2 psig
Female thread welded (H250/RR /HC) ISO 228 G½...G2" 50 bar / 735
ASME B1.20.1 ½…2" NPT
Female thread (H250/RR /HC) with insert, FPM gasket and union nut
Thread connection aseptic (H250/F) DIN 11864 - 1 DN1550 PN40
Flange aseptic (H250/F) DIN 11864 - 2 DN1550 PN40
Meters (H250/RR /HC) with heating:
Meters (H250/RR /HC) with heating:
Meters (H250/RR /HC) with heating:Meters (H250/RR /HC) with heating:
Heating with flange connection EN 1092-1 DN15 PN40
Heating pipe connection for Ermeto - E12 PN40
ISO 228 G½…2" 50 bar ASME B1.20.1 ½…2" NPT
DN80100 PN16
DN80DN100 PN16
ASME B16.5 ½" 150 lb / RF
psig
735 psig
72
Higher pressure ratings and other connections on request
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 73
H250 M40
TECHNICAL DATA
8
Bolts and tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange threads with the following torques:
Nominal sizes EN
Stud bolts Tightening torques
Nominal size acc. to EN 1092-1 Quantity x size [Nm] [lb-ft]
DN15 PN40
DN25 PN40
DN50 PN40
DN80 PN16
DN100 PN16
1 standard connections; other connections on request
1
1
1
1
1
Nominal size ASME
Nominal sizes acc. to ASME B
16.5
½" 150 lb / 300 lb
1" 150 lb / 300 lb
2" 150 lb / 300 lb
3" 150 lb / 300 lb
4" 150 lb / 300 lb
1 standard connections; other connections on request
1
1
1
1
1
Quantity x size [Nm] [lb-ft]
150 lb 300 lb
4x ½" 4x ½" 5.2 3.8 4x ½" 4x 5/8" 10 7.2
4x 5/8" 8x 5/8" 41 30 4x 5/8" 8x ¾" 70 51 8x 5/8" 8x ¾" 50 36
4x M12 9.8 7.1
4x M12 21 15
4x M16 57 41
8x M16 47 34
8x M16 67 48
Stud bolts Tightening torques
Pressure tightness (vacuum) H250/C
Max. process temperature  +70°C / +158°F +150°C / +302°F +250°C / +482°F
Min. operating pressure
Size Float Liner [mbar
abs.]
DN15...DN100 PTFE PTFE 100 1.45 - - - -
DN15...DN80 Ceramic PTFE 100 1.45 250 3.63 - -
DN15...DN80 Ceramic TFM /
Ceramic
100 1.45 100 1.45 100 1.45
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
[psia] [mbar
abs.]
[psia] [mbar
abs.]
[psia]
73
Page 74
8
TECHNICAL DATA
8.4 Measuring ranges
H250 M40
H250/RR - Stainless Steel, H250/HC - Hastelloy
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
®
Water Air Max. pressure loss
Float TIV CIV DIV TIV
Nominal
Cone [l/h]
TIV DIV TIV
Alu
[Nm3/h]
TIV CIV DIV
Alu
[mbar]
meter size
DN15, ½" K 15.1 18 25 - 0.42 0.65 - 12 21 26 -
K 15.2 30 40 - 0.7 1 - 12 21 26 -
K 15.3 55 63 - 1 1.5 - 12 21 26 -
K 15.4 80 100 - 1.7 2.2 - 12 21 26 -
K 15.5 120 160 - 2.5 3.6 - 12 21 26 -
K 15.6 200 250 - 4.2 5.5 - 12 21 26 -
K 15.7 350 400 700 6.7 10 18
K 15.8 500 630 1000 10 14 28
K 15.8 - - 1600
DN25, 1" K 25.1 480 630 1000 9.5 14 - 11 24 32 72
K 25.2 820 1000 1600 15 23 - 11 24 33 74
K 25.3 1200 1600 2500 22 35 - 11 25 34 75
K 25.4 1700 2500 4000 37 50 110
K 25.5 3200 4000 6300 62 95 180
DN50, 2" K 55.1 2700 6300 8400 58 80 230
K 55.2 3600 10000 14000 77 110 350
K 55.3 5100 16000 25000 110 150 700
DN80, 3" K 85.1 12000 25000 37000 245 350 1000
K 85.2 16000 40000 64000 280 400 1800
DN100, 4" K105.1 19000 63000 100 000 - 550 2800
1 P >0.5 bar 2 with TR float 3 300 mbar with damping (gas measurement)
2
- - 50
1
1
2
1
1
1
1
1
1
1
1
12 21 28 38
13 22 32 50
- - - 85
12 26 38 78
13 30 45 103
8 13 74 60
8 13 77 69
9 13 84 104
8 16 68 95
9 16 89 125
- - 120 220
3
74
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 75
H250 M40
TECHNICAL DATA
8
H250/RR - Stainless steel, H250/HC - Hastelloy
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
®
Water Air Max. pressure loss
Float TIV CIV DIV TIV
Nominal
Cone [GPH] [SCFM] [psig]
Alu
TIV DIV TIV
Alu
TIV CIV DIV
meter size
DN15, ½" K 15.1 4.76 6.60 - 0.26 0.40 - 0.18 0.31 0.38 -
K 15.2 7.93 10.6 - 0.43 0.62 - 0.18 0.31 0.38 -
K 15.3 14.5 16.6 - 0.62 0.93 - 0.18 0.31 0.38 -
K 15.4 21.1 26.4 - 1.05 1.36 - 0.18 0.31 0.38 -
K 15.5 31.7 42.3 - 1.55 2.23 - 0.18 0.31 0.38 -
K 15.6 52.8 66.0 - 2.60 3.41 - 0.18 0.31 0.38 -
K 15.7 92.5 106 185 4.15 6.20 11.2
K 15.8 132 166 264 6.20 8.68 17.4
K 15.8 - - 423
DN25, 1" K 25.1 127 166 264 5.89 8.68 - 0.16 0.35 0.47 1.06
K 25.2 217 264 423 9.30 14.3 - 0.16 0.35 0.49 1.09
K 25.3 317 423 660 13.6 21.7 - 0.16 0.37 0.50 1.10
K 25.4 449 660 1057 22.9 31.0 68.2
K 25.5 845 1057 1664 38.4 58.9 111
DN50 2" K 55.1 713 1664 2219 36.0 49.6 143
K 55.2 951 2642 3698 47.7 68.2 217
K 55.3 1347 4227 6604 68.2 93.0 434
DN80 3" K 85.1 3170 6604 9774 152 217 620
K 85.2 4227 10567 16907 174 248 1116
DN100 4" K105.1 5019 16643 26418 - 341 1736
1 P >7.4 psig 2 with TR float 3 4.4 psig with damping (gas measurement)
2
- - 31.0
1
1
2
1
1
1
1
1
1
1
1
0.18 0.31 0.41 0.56
0.19 0.32 0.47 0.74
- - - 1.25
0.18 0.38 0.56 1.15
0.19 0.44 0.66 1.51
0.12 0.19 1.09 0.88
0.12 0.19 1.13 1.01
0.13 0.19 1.23 1.53
0.12 0.24 1.00 1.40
0.13 0.24 1.31 1.84
- 1.76 3.23
3
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bar abs. (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
75
Page 76
8
TECHNICAL DATA
H250/C - Ceramic/PTFE
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
Flow rate Max. pressure loss
Water Air Water Air
H250 M40
Liner /
PTFE Ceramic PTFE Ceramic PTFE Ceramic PTFE Ceramic
Float
Size Cone [l/h]
DN15, ½" E 17.2 25 30 0.7 - 65 62 65 62
E 17.3 40 50 1.1 1.8 66 64 66 64
E 17.4 63 70 1.8 2.4 66 66 66 66
E 17.5 100 130 2.8 4 68 68 68 68
E 17.6 160 200 4.8 6.5 72 70 72 70
E 17.7 250 250 7 9 86 72 86 72
E 17.8 400 - 10 - 111 - 111 -
DN25, 1" E 27.1 630 500 16 18 70 55 70 55
E 27.2 1000 700 30 22 80 60 80 60
E 27.3 1600 1100 45 30 108 70 108 70
E 27.4 2500 1600 70 50 158 82 158 82
E 27.5 4000
DN50, 2" E 57.1 4000 4500 110 140 81 70 81 70
E 57.2 6300 6300 180 200 110 80 110 80
E 57.3 10000 11000 250 350 170 110 170 110
E 57.4 16000
DN80, 3" E 87.1 16000 16000 - - 81 70 - -
E 87.2 25000 25000 - - 95 85 - -
E 87.3 40000
DN100, 4" E 107.1 40000 - - - 100 - - -
E 107.2 60000
1 special float
1
1
1
1
2500 120 75 290 100 194 100
- - - 284 - - -
- - - 243 - - -
- - - 225 - - -
[Nm3/h]
[mbar]
76
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 77
H250 M40
TECHNICAL DATA
H250/C - Ceramic/PTFE
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
Flow rate Max. pressure loss
Water Air Water Air
Liner / Float PTFE Ceramic PTFE Ceramic PTFE Ceramic PTFE Ceramic
Size Cone [GPH] [SCFM] [psig]
DN15, ½" E 17.2 6.60 7.93 0.43 - 0.94 0.90 0.94 0.90
E 17.3 10.6 13.2 0.68 1.12 0.96 0.93 0.96 0.93
E 17.4 16.6 18.5 1.12 1.49 0.96 0.96 0.96 0.96
E 17.5 26.4 34.3 1.74 2.48 0.99 0.99 0.99 0.99
E 17.6 42.3 52.8 2.98 4.03 1.04 1.02 1.02 1.02
E 17.7 66.0 66.0 4.34 5.58 1.25 1.04 1.25 1.04
E 17.8 106 - 6.2 - 1.61 - 1.61 -
DN25, 1" E 27.1 166 132 9.92 11.2 1.02 0.80 1.02 0.80
E 27.2 264 185 18.6 13.6 1.16 0.87 1.16 0.87
E 27.3 423 291 27.9 18.6 1.57 1.02 1.57 1.02
E 27.4 660 423 43.4 31.0 2.29 1.19 2.29 1.19
E 27.5 1056
DN50, 2" E 57.1 1057 1189 68.2 86.8 1.18 1.02 1.18 1.02
E 57.2 1664 1664 111.6 124 1.60 1.16 1.60 1.16
E 57.3 2642 2906 155 217 2.47 1.60 2.47 1.60
E 57.4 4226
DN80, 3" E 87.1 4227 4227 - - 1.18 1.02 - -
E 87.2 6604 6604 - - 1.38 1.23 -
E 87.3 10567
DN100, 4" E 107.1 10567 - - - 1.45 - -
E 107.2 15850
1 special float
1
1
1
1
660 74.4 46.5 4.21 1.45 2.81 1.45
- - - 4.12 - - -
- - - 3.55 - -
- - - 3.29 - -
8
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bara (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
77
Page 78
8
TECHNICAL DATA
H250H - Horizontal installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
H250 M40
EN ASME Cone Water [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15 ½ K 15.1 70 1.8 195
K 15.2 120 3 204
K 15.3 180 4.5 195
K 15.4 280 7.5 225
K 15.5 450 12 250
K 15.6 700 18 325
K 15.7 1200 30 590
K 15.8 1600 40 950
K 15.8 2400 60 1600
DN25 1" K 25.1 1300 35 122
K 25.2 2000 50 105
K 25.3 3000 80 116
K 25.4 5000 130 145
K 25.5 8500 220 217
K 25.5 10000 260 336
DN50 2" K 55.1 10000 260 240
K 55.2 16000 420 230
K 55.3 22000 580 220
K 55.3 34000 900 420
DN80 3" K 85.1 25000 650 130
K 85.2 35000 950 130
K 85.2 60000 1600 290
DN100 4" K 105.1 80000 2200 250
K 105.1 120000 3200 340
78
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 79
H250 M40
TECHNICAL DATA
H250H - Horizontal installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
EN ASME Cone Water [GPH] Air [SCFM] Pressure loss
DN15 1/2" K 15.1 18.5 1.12 2.87
K 15.2 31.7 1.86 3.00
K 15.3 47.6 2.79 2.87
K 15.4 74.0 4.65 3.31
K 15.5 119 7.44 3.68
K 15.6 185 11.2 4.78
K 15.7 317 18.6 8.68
K 15.8 423 24.8 14.0
K 15.8 634 37.2 23.5
DN25 1" K 25.1 343 21.7 1.79
K 25.2 528 31.0 1.54
K 25.3 793 49.6 1.71
K 25.4 1321 80.6 2.13
K 25.5 2245 136 3.19
K 25.5 2642 161 4.94
DN50 2" K 55.1 2642 161 3.53
K 55.2 4227 260 3.38
K 55.3 5812 360 3.23
K 55.3 8982 558 6.17
DN80 3" K 85.1 6604 403 1.91
K 85.2 9246 589 1.91
K 85.2 15851 992 4.26
DN100 4" K 105.1 21134 1364 3.68
K 105.1 31701 1984 5.00
8
[psig]
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bara (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
79
Page 80
8
TECHNICAL DATA
H250U - Vertical installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
Flow direction vertical downwards
H250 M40
EN ASME Cone Water [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15 ½" K 15.1 65 1.6 175
K 15.2 110 2.5 178
K 15.3 170 4 180
K 15.4 260 6 200
K 15.5 420 10 220
K 15.6 650 16 290
K 15.7 1100 28 520
K 15.8 1500 40 840
DN25 1" K 25.1 1150 30 97
K 25.2 1800 45 85
K 25.3 2700 70 92
K 25.4 4500 120 115
K 25.5 7600 200 172
DN50 2" K 55.1 9000 240 220
K 55.2 15000 400 230
K 55.3 21000 550 240
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
80
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 81
H250 M40
TECHNICAL DATA
H250U - Vertical installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water: 20°C / 68°F Air: 20°C / 68°F, 1.013 bara / 14.7 psia
Flow direction vertical downwards
EN ASME Cone Water [GPH] Air [SCFM] Pressure loss
DN15 ½" K 15.1 17.2 0.99 2.57
K 15.2 29.1 1.55 2.62
K 15.3 44.9 2.48 2.65
K 15.4 68.7 3.72 2.94
K 15.5 111 6.20 3.23
K 15.6 172 9.92 4.26
K 15.7 291 17.4 7.64
K 15.8 396 24.8 12.3
DN25 1" K 25.1 304 18.6 1.42
K 25.2 476 27.9 1.25
K 25.3 713 43.4 1.35
K 25.4 1189 74.4 1.69
K 25.5 2008 124 2.53
DN50 2" K 55.1 2378 149 3.23
K 55.2 3963 248 3.38
K 55.3 5548 341 3.53
8
[psig]
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bara (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
81
Page 82
9
NOTES
H250 M40
82
www.krohne.com 08/2015 - 4000640703 - MA H250 M40 R03 en
Page 83
H250 M40
NOTES
9
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
83
Page 84
KROHNE – Process instrumentation and measurement solutions
Flow
Level
Temperature
Pressure
Process Analysis
Services
Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.: +49 203 301 0 Fax: +49 203 301 10389 info@krohne.com
© KROHNE 08/2015 - 4000640703 - MA H250 M40 R03 en - Subject to change without notice.
The current list of all KROHNE contacts and addresses can be found at: www.krohne.com
Loading...