All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without
the prior written authorisation of KROHNE Messtechnik GmbH.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 5
H250 M40
1.1 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and
corrosion resistance of the used materials against the measured fluid lies solely with the
operator.
INFORMATION!
This device is a Group 1, Class A device as specified within CISPR11:2009. It is intended for use in
industrial environment. There may be potential difficulties in ensuring electromagnetic
compatibility in other environments, due to conducted as well as radiated disturbances.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than
the intended purpose.
The variable area flowmeters are suitable for measuring clean gases, vapours and liquids.
Intended use:
• The product may not contain any ferromagnetic particles or solids. It may be necessary to
install magnetic filters or mechanical filters.
• The product must be sufficiently liquid and free of deposits.
• Avoid pressure surges and pulsing flows.
• Open valves slowly. Do not use solenoid valves.
SAFETY INSTRUCTIONS
1
Use suitable measures to eliminate compression vibrations during gas
measurements:
• Short pipeline lengths to next restriction
• Nominal pipe size not greater than nominal device size
• Use of floats with damping
• Increase in operating pressure (while taking into account the resulting change in density and
thus change in scale)
Observe installation conditions according to VDI/VDE 3513-3.
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
CAUTION!
Do not use any abrasive media containing solid particles or highly viscous media.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
5
Page 6
1
SAFETY INSTRUCTIONS
1.2 Certifications
CE marking
The device fulfils all applicable statutory requirements of the following EC directives:
• Pressure equipment directive
• For devices with electrical installations: EMC directive
• Devices for use in hazardous areas: ATEX directive
as well as
• NAMUR recommendations NE 21, NE 43 and NE 107
The manufacturer certifies successful testing of the product by applying the CE marking.
A CE declaration of conformity regarding the directives in question and the associated
harmonised standards can be downloaded from our internet site.
H250 M40
1.3 Pressure equipment directive
A conformity assessment in accordance with pressure equipment directive has been carried out
for the devices described. Conformity is certified by applying the CE mark. The number of the
notified body is also stated.
The PED key describes rating of devices:
Example: PED/G1/III/H
GGases and vapours
1Fluid group 1
IIICategory III
HConformity assessment method according to Module H
The PED key identification can be found on the nameplate of the device (for details refer to
Nameplate
INFORMATION!
The stated pressures (PS) and temperatures (TS) only apply as refers to the pressure resistance
of the sensor body. As regards the functionality of the entire device, further restrictions of the
maximum temperature may need to be observed (e.g. ATEX approval). Devices rated below
Category I due to their size, do not receive the CE mark in the scope of the PED. These devices
are subject to applicable Sound Engineering Practice SEP.
on page 15).
6
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 7
H250 M40
1.4 Safety instructions from the manufacturer
1.4.1 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no
guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third
parties are identified as such. Reproduction, processing, dissemination and any type of use
beyond what is permitted under copyright requires written authorisation from the respective
author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created
in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the
manufacturer's documents is always on a voluntary basis whenever possible. Whenever
feasible, it is always possible to make use of the offerings and services without providing any
personal data.
SAFETY INSTRUCTIONS
1
We draw your attention to the fact that data transmission over the Internet (e.g. when
communicating by e-mail) may involve gaps in security. It is not possible to protect such data
completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish
an imprint for the purpose of sending us any advertising or informational materials that we have
not expressly requested.
1.4.2 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including,
but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross
negligence. In the event any applicable law does not allow such limitations on implied warranties
or the exclusion of limitation of certain damages, you may, if such law applies to you, not be
subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant
product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this
disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable
in any way for possible consequences of such changes.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
7
Page 8
1
SAFETY INSTRUCTIONS
1.4.3 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose.
The manufacturer accepts no liability for the consequences of misuse by the operator. Improper
installation or operation of the devices (systems) will cause the warranty to be void. The
respective "Standard Terms and Conditions" which form the basis for the sales contract shall
also apply.
1.4.4 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the
information in this document and observe applicable national standards, safety requirements
and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the
text, we advise you to contact your local office for assistance. The manufacturer can not accept
responsibility for any damage or injury caused by misunderstanding of the information in this
document.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device. Special considerations and precautions are also described in the
document, which appear in the form of icons as shown below.
H250 M40
8
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 9
H250 M40
1.4.5 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This warning refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to
serious health problems and even death. There is also the risk of seriously damaging the device
or parts of the operator's plant.
SAFETY INSTRUCTIONS
1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems.
There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's
plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
• HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the
specified sequence.
iRESULT
RESULT
RESULTRESULT
This symbol refers to all important consequences of the previous actions.
1.5 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and
maintained by properly trained and authorized personnel.
This document is provided to help you establish operating conditions, which will permit safe and
efficient use of this device.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
9
Page 10
2
DEVICE DESCRIPTION
2.1 Scope of delivery
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
H250 M40
Figure 2-1: Scope of delivery
1 Measuring device in ordered version
2 For the ESK4-T version - bar magnet
3 Documentation
4 Certificates, calibration report (supplied to order only)
5 Wrench
10
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 11
H250 M40
2.2 Device version
• H250 with indicator M40
• H250 with M40 indicator with display cut-out for ESK4-T
1. H250/RR/M40
• Local indicator without auxiliary power
• max. 2 limit switches, type NAMUR, NAMUR safety-oriented or transistor (3-wire)
• Electrical signal output 4...20 mA, HART
• Intrinsically safe [Ex i] or in explosion-proof enclosure [Ex d]
DEVICE DESCRIPTION
®
or Fieldbus communication
2
2. H250/RR/M40
• additional LCD, measured value and/or flow counter
• 2 configurable binary outputs, limit value or pulse output
1 Basic module with electronic magnet sensors ESK4 / ESK4A
2 Connection bus module
3 DIP switch for bus settings
For more details see the supplementary instructions "H250 M40 Foundation Fieldbus"
and "H250 M40 Profibus PA"
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
13
Page 14
2
DEVICE DESCRIPTION
2.2.2 Float damping
Float damping is characterised by high standstill times and self-centering. The damping sleeve
is made of high performance ceramic or PEEK, depending on the medium and the application.
Float damping can also be retrofitted for the user (refer to "Service").
Use of damping
• Generally when CIV and DIV floats are used for gas measurement.
• For TIV floats (H250/RR and H250/HC only) with an operating primary pressure:
Nominal size acc. toOperating primary pressure
EN 1092-1ASME B16.5[bar][psig]
DN 50½"≤0.3≤4.4
DN251"≤0.3≤4.4
DN502"≤0.2≤2.9
DN803"≤0.2≤2.9
DN 1004"≤0.2≤2.9
H250 M40
2.2.3 Pointer damping
In principle, the indicating element with its magnetic system contains indicator damping. An
additional eddy current brake is advantageous in the event of fluctuating or pulsing flows. The
magnets on the eddy current brake surround the pointer vane without touching it, damping its
movement. The result is a much steadier pointer position, without distorting the measured
value. A turnbuckle ensures a proper fit. The eddy current brake can be retrofitted during
operation without recalibrating (refer to "Service").
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 15
H250 M40
2.3 Nameplate
INFORMATION!
Check on the device nameplates, that the device is supplied according to your order.
DEVICE DESCRIPTION
2
Figure 2-2: Nameplate on the indicator
1 Device type
2 Manufacturer
3 Notified ATEX & PED body
4 Rating data: temperature & pressure rating
5 PED data
6 Ex data
7 Electrical connection data
8 Internet site
Additional markings on the indicator
• SN - serial number
• SO - sales order / item
• PA - Production order
• Vx - product configurator code
• AC - article code
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
15
Page 16
2
DEVICE DESCRIPTION
2.4 Description code
The description code* consists of the following elements:
1 Device type
H250 - standard version
H250H - horizontal flow direction
H250U - flow direction from top to bottom
2 Materials / versions
RR - stainless steel
C - PTFE or PTFE/ceramics
HC - Hastelloy
Ti - Titanium
Mo - Monel
F - aseptic version (food)
H250 M40
3 Heating jacket version
B - with heating jacket
4 Series of indicators
M40 - Indicator M40
M40S - Indicator with increased corrosion protection
M40R - Indicator in stainless steel housing
5 High temperature version
HT - Version with HT extension
6 Electrical signal output
ESK - Electrical signal output 4...20mA (ESK4 / ESK4A)
- optionally available with counter, I/O module and display (ESK4-T) or
- Foundation Fieldbus (ESK4-FF) or
- Profibus PA (ESK4-PA)
7 Limit switches
K1 - One limit switch
K2 - Two limit switches
8 Explosion protection
Ex - Explosion-protected equipment
9 SIL Version
SE - SIL compliant electronic signal output
SK - SIL compliant limit switch
* positions which are not needed are omitted (no blank positions)
16
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 17
H250 M40
2.5 Electronics revision
The Electronic Revision (sticker on the base module ESK4 / ESK4A) indicates the respective
hardware/software status of the electronics. All add-on modules (ESK4-T, ESK4-FF and ESK4PA) have an additional sticker indicating their respective firmware version.
DEVICE DESCRIPTION
2
Electronics
Explanations
revision
ER 1.1.xBasic version (cannot be combined with other indicator versions):
can be combined with indicator version ESK4 FF / Foundation Fieldbus;
(Firmware Version FF module from 1.0.2)
can be combined with indicator version ESK4-PA / Profibus PA;
(Firmware Version PA module from 1.0.0)
can be combined with indicator version ESK4-T / LCD, binary inputs/outputs;
(Firmware Version T module from 1.1.0)
Support of failure signal (low) according to NE43
for the ESK 4 current output module
®
Update for HART
can be combined with ESK4-FF (Firmware Version FF module from 1.0.2)
Firmware Version ESK4-PA (... PA module from 1.0.0)
Firmware Version ESK4-T (... T module from 1.2.0)
communication from 5.9 to 7.4 including new DD/DTM
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
17
Page 18
3
INSTALLATION
3.1 Notes on installation
INFORMATION!
Inspect the packaging carefully for damages or signs of rough handling. Report damage to the
carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
3.2 Storage
• Store the device in a dry and dust-free location.
• Avoid lasting direct exposure to the sun.
• Store the device in its original packing.
• The permissible storage temperature for standard devices is -40...+80°C / -40...+176°F.
H250 M40
18
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 19
H250 M40
3.3 Installation conditions
CAUTION!
When installing the device in the piping, the following points must be observed:
•
The variable area flowmeter must be installed vertically (measuring principle). Flow direction
from bottom to top. For installation recommendations please refer also to directive
VDI/VDE 3513, sheet 3.
H250Hs are installed horizontally and H250U devices are installed vertically with the flow
direction from top to bottom.
•
A straight unimpeded inlet run of ≥ 5x DN upstream of the device and a straight outlet run of
≥ 3x DN downstream of the device are recommended.
•
Screws, bolts and gaskets are to be provided by the customer and must be selected in
accordance with the pressure rating of the connection or the operating pressure.
•
The inside diameter of the flange deviates from the standard dimensions. Flange seal
standard DIN 2690 can be applied without any limitation.
•
Align the gaskets. Tighten the nuts with the tightening torques of the appropriate pressure
rating.
For devices with PTFE liner or ceramic liner and PTFE raised faces, refer to chapter
"Tightening torques".
•
Control devices are to be positioned downstream of the measuring device.
•
Shut-off devices are preferably to be positioned upstream of the measuring device.
•
Before connecting, blow or flush out the pipes leading to the device.
•
Pipes for gas flow need to be dried before the device is installed.
•
Use connectors suitable for the particular device version.
•
Align the pipes centrically with the connection bores on the measuring device so they are free
of stresses.
•
If necessary, the piping has to be supported to reduce the vibrations transmitted to the
measuring device.
•
Do not lay signal cables directly next to cables for the power supply.
INSTALLATION
3
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
19
Page 20
3
INSTALLATION
Minimum distance between devices
When several devices are installed next to one another, a minimum distance of a > 300 mm /
11.8" between the devices is necessary.
H250 M40
Take special note of the installation position for the H250H with horizontal flow
direction:
In order to comply with thermal parameters and measuring accuracy, H250H flowmeters for
horizontal installation are to be installed in the pipeline so that the display is located on the side
of the measuring tube. The maximum product and ambient temperatures indicated as well as
the measuring accuracy are based on lateral installation of the display.
20
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 21
H250 M40
3.3.1 Tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange
threads with the following torques:
The use of magnetic filters is recommended when the medium contains particles which can be
influenced magnetically. The magnetic filter is to be installed in the flow direction upstream of
the flowmeter. Bar magnets are positioned helically in the filter to provide optimal efficiency at
low pressure loss. All of the magnets are coated individually with PTFE to protect against
corrosion. Material: 1.4404/316L
Magnetic filters
1 Type F - fitting part with flange - overall length 100 mm / 4"
2 Type FS - fitting part without flange - overall length 50 mm / 2"
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
21
Page 22
3
INSTALLATION
3.3.3 Heat insulation
CAUTION!
The indicator housing may not be heat-insulated.
The heat insulation 3 may only reach as far as the housing fastening 4.
1 Standard indicator M40
2 Indicator with HT extension
H250 M40
CAUTION!
The heat insulation 1 may only reach to the rear of the housing 2. The area around the cable
entries 3 must be freely accessible.
Figure 3-1: Insulation - cross section
22
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 23
H250 M40
4.1 Security information
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take
note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex
documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the
electrical components of the measuring device may only be carried out by properly trained
specialists.
ELECTRICAL CONNECTIONS
4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order.
Check for the correct supply voltage printed on the nameplate.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
23
Page 24
4
ELECTRICAL CONNECTIONS
4.2 Electrical connection indicator M40
4.2.1 Limit switches K1/K2
The M40 indicator can be fitted with a maximum of two limit switches. The limit switch works as
a slot sensor which is inductively activated via the semi-circular metal vane of the pointer. The
switching points are set using the contact pointer. The position of the contact pointer is indicated
on the scale.
Limit switch module
H250 M40
1 Min. contact
2 Max. contact
3 Locking screw
4 Maximum pointer
5 Connection terminal
The connection terminals feature a pluggable design and can be removed to connect the cables.
The built-in limit switch types are shown on the indicator.
Electrical connection of the limit switches
ContactMINMAX
Terminal no.123456
Connection 2-wire NAMUR-+-+
Connection 3-wire+-+-
Connection Reed SPST+-+-
24
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
1 Contact pointer MAX
2 Contact pointer MIN
3 Locking screw
Setting is carried out directly via contact pointers 1 and 2:
• Slide the scale away.
• Loosen the locking screw 3 slightly.
• Slide the scale back to the latching point.
• Set contact pointers 1 and 2 to the desired switching point.
After setting has been carried out, fix the contact pointers with the locking screw 3.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
25
Page 26
4
ELECTRICAL CONNECTIONS
Switch contact definition
1 MIN contact
2 MAX contact
3 Pointer vane with switching vane
If the measuring pointer vane goes into the slot an alarm is triggered. If the pointer vane is
outside the slot sensor, a wire break in a NAMUR contact also triggers the alarm.
The 3-wire limit switch does not have any wire break detection.
H250 M40
Definition MinMin - MaxMax
1 MIN 2 contact or MAX 1 contact
2 MIN 1 contact or MAX 2 contact
Current consumption in the position shown:
ContactTypecurrent
MIN 1NAMUR≤ 1 mA
MIN 2NAMUR≤ 1 mA
MAX 1NAMUR≥ 3 mA
MAX 2NAMUR≥ 3 mA
26
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 27
H250 M40
4.2.2 Current output ESK4 / ESK4A
The connecting terminals of the ESK4 / ESK4A have a pluggable design and can be removed in
order to connect the cables.
ESK4 / ESK4A connection
1 Current output ESK4 /ESK4A
2 Power supply 14...30 VDC
3 Measuring signal 4...20 mA
4 External load, HART
®
communication
ELECTRICAL CONNECTIONS
4
Power supply M40 with galvanic isolation
Power supply M40 with galvanic isolation
Power supply M40 with galvanic isolationPower supply M40 with galvanic isolation
Wiring must be planned with great care when it comes to connecting other devices such as
evaluation units or process control. Internal connections in these devices (e.g. GND with PE,
mass loops) may lead to non-permitted voltage potentials which could negatively affect the
function of the converter itself or that of a device connected to it. In such cases a protected extralow voltage (PELV) is recommended.
1 Terminal connection
2 Converter supply isolator with electrical isolation
3 Power supply (see supply isolator information)
4 Measuring signal 4...20 mA
5 External load, HART
®
communication
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
27
Page 28
4
ELECTRICAL CONNECTIONS
Power supply
Power supply
Power supplyPower supply
INFORMATION!
The supply voltage has to be between 14 VDC and 30 VDC. This is based on the total resistance of
the measuring loop. To calculate this, the resistance of each component in the measuring loop
(not including the device) must be added up.
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
where
U
ext.
= the total measuring loop resistance.
R
L
24 mA + 14 V
L
= the minimum supply voltage and
INFORMATION!
The power supply has to be able to supply a minimum of 30 mA.
H250 M40
HART
HART® communication
HARTHART
When HART® communication is carried out with the ESK4, the analogue measured data
transmission (4...20 mA) is not impaired in any way.
Exception for multidrop operation. In multidrop mode, a maximum of 15 devices with HART
function can be operated in parallel, whereby their current outputs are switched inactive
(I approx. 4.5 mA per device).
®
28
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 29
H250 M40
ELECTRICAL CONNECTIONS
Load for HART® communication
INFORMATION!
®
A load of at least 230 Ohm is required for HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output
signal.
In some cases a shielded cable may be necessary, if noise levels higher than the NE21
specification are anticipated.
communication.
4
Configuration
The ESK can be configured via HART® communication. DD (Device Description) for AMS and PDM
TM
as well as a DTM (Device Type Manager) for PACTware
can be downloaded free of charge from our website.
The current flow rate can be transmitted using the integrated HART
counter can be configured. Two limit values can be monitored. The limit values are assigned
either to flow values or to the counter overflow.
are available for configuration. They
®
communication. A flow
Self monitoring - Diagnostics
During both start-up and operation, a wide variety of diagnostic functions are performed
cyclically in the ESK4 / ESK4A in order to guarantee function reliability. When an error is
detected, a failure signal (high) is activated (current > 21 mA, typically 22 mA) via the analogue
®
output. In addition more detailed information can be requested via HART
signal is not activated for information and warnings.
(CMD#48). The failure
Diagnostic functions (Monitoring):
• Plausibility of FRAM data
• Plausibility of ROM data
• Working range of internal reference voltages
• Signal detection of the measuring range of the internal sensors
• Temperature compensation of the internal sensors
• Calibration based on the application
• Plausibility of counting value
• Plausibility of physical unit, system and selected unit
For ESK4A (HART 7) the diagnosis is reported in compliance with NE107.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
29
Page 30
4
ELECTRICAL CONNECTIONS
4.2.3 ESK4-T limit outputs
Once the housing cover has been unscrewed, the scale can be removed. The connection
terminals feature a pluggable design and can be removed to connect the cables.
H250 M40
1 Binary output 1
2 Power supply / current output ESK4 / ESK4A
3 Binary output 2
4 Binary input
The binary inputs/outputs are electrically isolated from each other and from the ESK4 / ESK4A
current output.
INFORMATION!
The binary inputs/outputs can only be operated if the power supply is applied to ESK4 / ESK4A
terminal 11+ and 12-. The binary inputs/outputs come inactive by default and must thus be
activated prior to first use (refer to section "Menu ESK4-T").
Connection binary outputs
Connection binary outputs
Connection binary outputsConnection binary outputs
In accordance with the desired signal transmission, select one of the following connection types
for binary outputs B1 and B2:
• NAMUR (DC interface in accordance with EN 60947-5-6)
• Transistor output (passive, open collector)
Binary outputB1B2
Terminal no.123456
30
Connection NAMUR+-+-
Connection transistor output+B
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
OC
+B
OC
Page 31
H250 M40
1 NAMUR terminal connection
2 Isolated switching amplifier
3 OC switch output terminal connection
4 Power supply U
5 Load R
L
ext.
Value range NAMUR
ELECTRICAL CONNECTIONS
4
Normally closedNormally open
Switching value reached≤1 mA> 3 mA
Switching value not reached> 3 mA≤1 mA
Value range applies only when connected to a switching amplifier with the following
reference values:
• Open-circuit voltage Uo = 8.2 VDC
• Internal resistance R
= 1 kOhm
i
Value range for transistor output
Signal voltagesUL [V]UH [V]
lower limitupper limitlower limitupper limit
via load R
Signal currentsIL [mA]IH [mA]
Category 20220110
L
To ensure the value ranges, a load RL between 250 Ohm and 1 kOhm is recommended for the
passive transistor output with a nominal voltage of 24 VDC.
021630
lower limitupper limitlower limitupper limit
If other loads are used, caution is advised as the range of values of the signal voltages then no
longer corresponds to the range of values for the inputs of process control systems and controls
(DIN IEC 946).
CAUTION!
The upper limit of the signal current must not be exceeded as this may damage the transistor
output.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
12 NAMUR MIN (-)2 Reed MIN-11
22 NAMUR MIN (+)2 Reed MIN-23
33 NAMUR MAX (-)3 Reed MAX-44
43 NAMUR MAX (+)3 Reed MAX-56
5--4 4...20mA (+)11
6--4 4...20mA (-)12
7---
8---
Combinations of K1 / K2 and ESK4 / ESK4A are possible.
1 Pin No. HAN® 7DESK4-TTerminal no.
15 Binary output B1 open coll. (+)1
25 Binary output B1 open coll. (-)3
37 Binary output B2 open coll. (+)4
47 Binary output B2 open coll. (-)6
56 4...20mA (+)11
66 4...20mA (-)12
7-
8-
32
INFORMATION!
No Harting connection for module ESK4-FF/PA is provided and is only available on request.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 33
H250 M40
4.3 Grounding connections
1 Grounding connection in the indicator
2 Grounding connection outside
DANGER!
The grounding wire may not transfer any interference voltage.
Do not use this grounding cable to ground any other electrical devices.
ELECTRICAL CONNECTIONS
4
4.4 Protection category
The measuring device meets all requirements of protection category IP66/68.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection class
must be ensured again.
Therefore it is essential to observe the following points:
• Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
• The electrical cables must be undamaged and must comply with regulations.
• The cables must be laid with a loop 3 upstream of the measuring device to prevent water
from getting into the housing.
• The cable feedthroughs 2 must be tightened.
• Close the unused cable feedthroughs using blanking plugs 1.
1 Use blanking plugs if no cable is routed through
2 Tighten cable feedthrough firmly
3 Lay the cable in a loop
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
33
Page 34
5
START-UP
5.1 Standard device
CAUTION!
When starting up the device, the following points must be observed:
•
Compare the actual operating pressure and the product temperature of the system with the
specifications on the nameplate (PS and TS). These specifications may not be exceeded.
•
Make sure materials are compatible.
•
Slowly open the shut-off valve.
•
When measuring liquids, vent the pipes carefully.
•
When measuring gases, increase pressure slowly.
•
Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring
unit or float.
A minimum operating pressure (primary pressure) is necessary to operate the device:
MediumPressure loss : operating pressure
H250 M40
Liquids1 : 2
Gases without float damping1 : 5
Gases with float damping1 : 2
5.2 Indicator ESK4-T
INFORMATION!
The device is always preset for the user and his application.
Start
After the device is switched on, the display shows
• "INITIALISING"
• Firmwareversion IO-Modul
The device first performs a self-test. Here, all of the parameters preset for the customer are
analysed and checked for plausibility. The device then switches to measuring mode and indicates
the current measured value.
Operation
INFORMATION!
The device is low-maintenance
Comply with the application limits with regard to temperature of the medium and ambient
temperature.
34
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 35
H250 M40
6.1 ESK4 / ESK4A - Loop Check Mode
The ESK4 / ESK4A is equipped with a loop check function, enabling a simple test of the entire
4…20 mA current loop.
It is activated and operated using a microswitch 1.
CAUTION!
When activating the loop check mode, ensure that no alarms are unintentionally triggered in the
downstream system components.
OPERATION
6
• Press and hold the microswitch 1 for more than 6 seconds to activate loop check mode 3.
The current output jumps to constant 4mA.
• Change the current output from constant 4 to constant 20 mA as often as you like by briefly
pressing it (less than 6 seconds) to check the function of the measuring circuit.
• Exit loop check mode by holding down the microswitch (longer than 6 seconds). The current
output jumps back to measuring mode 2.
INFORMATION!
If the microswitch has not been pressed for longer than 60 seconds, the ESK4 / ESK4A
automatically reverts to measuring mode 2.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
35
Page 36
6
OPERATION
6.2 Operating elements ESK4-T
H250 M40
The device is operated with the cover on the front open, using the mechanical keys
cover closed using a bar magnet
bar magnet.
bar magnetbar magnet
keys, or with the
keyskeys
CAUTION!
The switching point of the magnetic sensors is directly at the height of the corresponding circle
(see figure). Only touch the circle vertically and from the front using the bar magnet. Touching it
from the side may cause an error in the measurement as the float position is recorded by the
magnetic field sensors.
Figure 6-1: Display and operating elements
1 Enter button (circuit for bar magnet)
2 Up button (circuit for bar magnet)
3 Right button (circuit for bar magnet)
The mechanical keys and keys for the bar magnet have the same functionality. In this
documentation the keys are represented as symbols to describe the operating functions:
KeySymbol
1Enter^^^^
2up↑
3right→
Table 6-1: ESK4-T operation keys
36
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 37
H250 M40
6.3 Basic principles of operation ESK4-T
6.3.1 Functional description of the keys
OPERATION
6
→
↑
^^^^
Table 6-2: Description of the operating keys
Switch from measuring mode to menu mode
Switch to one menu level lower
Open menu item and activate change mode
Confirm query whether data should be accepted.
In change mode:
In change mode: Move the input cursor one position to the right. After the last digit the
In change mode:In change mode:
input cursor jumps back to the beginning.
Change between the menu items within a menu level
In measuring mode:
In measuring mode: Switch between measured values and error messages
In measuring mode:In measuring mode:
In change mode:
In change mode: Changing parameters or settings. Run through the available characters
In change mode:In change mode:
(including decimal point.
In measuring mode:
In measuring mode: Switch between measured value display and error messages /
In measuring mode:In measuring mode:
warnings
Switch to one menu level higher
Return to measuring mode with a query whether the data should be accepted
Cancel the query, whether data should be accepted.
6.3.2 Navigation within the menu structure
Navigate through the menu using the →, ↑ and ^^^^ keys. Pressing the → key takes you one menu
lower. Use the ↑ key to go one menu point higher (e.g. from 1 to 2). Pressing the ^^^^ key takes you
one menu higher.
If you are already at the lowest level (function level), you can use the → button to go to the
change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^^^^ key to exit the menu mode and
return to the measuring mode.
If settings were changed, the query whether they should be saved will appear. Confirm this query
with the → key or cancel it with the ^^^^ key.
Measuring
mode
Table 6-3: Navigation menu structure
→Main menu↑→Submenu↑→Function↑→Edit
^^^^^^^^^^^^^^^^
→ ↑ ^^^^
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
37
Page 38
6
OPERATION
6.3.3 Change the settings in the menu
Starting operation
Operation is started using the → key
If a control lockout is set, the code (→ → → ^ ^ ^ ↑ ↑ ↑) must be entered. The code can be set
in menu 3.13. The code shown here is set at the factory but not activated. A warning is displayed
and the indicator returns to measuring mode if no key is pressed within 5 seconds or if an
incorrect code is entered.
Exiting operator input
Operation is exited by pressing the ^^^^ key several times.
If data have been changed:
H250 M40
Save Yes→Changes are accepted. An update is carried out and the display returns to
Save No^^^^Changes are discarded and the indicator jumps back to measuring operation.
measuring mode.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal
plausibility check.
If inplausible entries were made, a warning message is shown. If this warning is confirmed with
the ^^^^ key, the display returns to the respective menu point without saving the relevant change. A
new entry can now be made.
Example: Changing the flow unit from m3/h to l/h
DisplayDisplay
Example:7.2
m
1x →Fct. 1111
OPERATION
2x ↑Fct. 3333
INSTALLATION
1x →Fct 3.1111
LANGUAGE
10x ↑Fct 3.11
FS&UNIT
3
/h
11
1111
1x →Fct. 3.11.1111
1x →10.0000
4x ↑10000
3x ^^^^7200
FLOW
3
/h
m
l/h
Confirm with →
deny ^
l/h
38
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 39
H250 M40
6.4 Overview of the units ESK4-T
Volume units can either represent real operating volumes (no prefix before the unit) or standard
volumes virtually converted to reference statuses.
Prefix Volume definition
None
N
S
Operating volume flow e.g. m3/h or ft3/h
Volume flow in normal state (0°C - 1.013 bara) according to DIN 1343 e.g. Nm
Volume flow in standard state (15°C - 1.013 bara) according to ISO 13443 e.g. Sft
OPERATION
3
/h
3
/h
6
Measured
Units
variables
Operating
volume flow
Standard
volume flow
Standard
volume flow
Mass flowg/sg/ming/h-
Operating
volume totalizer
Standard
volume totalizer
Standard
volume totalizer
Mass totalizerkggtlb
Temperature°C°FK
m3/sm3/minm3/hm3/d
L/sL/minL/h-
ft3/sft3/minft3/hft3/d
gal/sgal/mingal/hgal/d
bbl/sbbl/minbbl/hbbl/d
ImpGal/sImpGal/minImpGal/hImpGal/d
Nm3/sNm3/minNm3/hNm3/d
NL/sNL/minNL/h-
Sm3/sSm3/minSm3/hSm3/d
SL/sSL/minSL/h-
Sft3/sSft3/minSft3/hSft3/d
kg/skg/minkg/hkg/d
-t/mint/ht/d
lb/slb/minlb/hlb/d
-STon/minSTon/hSTon/d
--LTon/hLTon/d
3
m
ImpGalgallonbblbbl (liq)
3
Nm
3
Sft
STonLTon
lhl
NL
SL
Sm
3
3
ft
In addition to the predefined units shown here, a user-defined unit can be activated in menu 3.12
by entering a conversion factor and a unit designation as free text.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
39
Page 40
6
OPERATION
6.5 Error messages ESK4-T
Error messages and warnings are indicated by one of the following symbols in the bottom left
corner of the display. The ^^^^ key switches from the measured value display to the display of
current errors / warnings. The table below contains a description of possible error messages.
H250 M40
SymbolNE107
DescriptionConsequence
Category
FFFFFailureNo measured value available
SOut of specificationMeasurement available but increasing
MMaintenance requiredMeasurement still sufficiently accurate
CFunction checkDevice is in test or linearization mode.
IInformationNo influence on the measurement,
Output signal invalid
Error current signal is output.
measurement uncertainty. Device must
be checked.
but maintenance required.
The output signal temporarily does not
correspond to the measurement.
information only.
Error messageDescriptionCategoryRemedy
NOT LINEARIZED Linearization faulty or
NEW LINEARI.
TABLE BAD
LINEARIZATIO
UNDER CONFIG
UNIT SYSTEM
CONFLICT
TOO FEW
ENTRIES
NOT
MONOTONOUS
FIRST NOT 0 %The first flow value if
LAST NOT 100 %The last flow value if
not activated =
measuring error
Faulty or missing
data in the
linearization table =
measuring error
The device is in
linearization mode =
measuring error
The unit for the
linearization flow is
incompatible with the
selected flow type
(mass/volume)
The linearization
table has too few data
points
The sequence of the
linearization values is
not strictly monotonic
increasing
the linearization table
is not 0%
the linearization table
is not 100%
SActivate linearization or carry it out again
S
1
SComplete the linearization and activate it
SCorrect error, carry out linearization again if
SCarry out linearization at at least 5 points
SCheck linearization and/or carry it out again
®
(HART
software are required; the original
calibration values must be known), or send
the device back to the manufacturer for
linearization.
®
(HART
software are required), or send the device
back to the manufacturer for linearization.
necessary (HART
linearization software are required), or send
the device back to the manufacturer for
linearization.
®
(HART
software are required), or send the device
back to the manufacturer for linearization.
®
(HART
software are required), or send the device
back to the manufacturer for linearization.
communication and linearization
communication and linearization
®
communication and
communication and linearization
communication and linearization
40
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 41
H250 M40
Error messageDescriptionCategoryRemedy
OPERATION
6
NO ZERO CAL OF AOThe analog output
NO F.SC. CAL OF AOThe analog output
NO TEMP.
COMPENSATION
WRONG
ELEC.REV.
OUTPUT NOT
LINEARIZED
COUNTER LOSTTotalizer value was
FRAM WRITE
FAULT
ROM/FLASH
ERROR
RESTART OF
DEVICE
MULTIDROP
MODE
CRYSTAL OSC
FAULT
REF VOLTAGE
FAULT
SENSOR A FAULT Internal error in
SENSOR B
FAULT
zero point 4.00mA is
not calibrated. =
possible measuring
error in process
control.
100% = 20.00mA is
not calibrated. =
possible measuring
error in process
control.
The sensor
temperature
compensation of the
device is faulty or was
not carried out =
possible measuring
error
The ER revision of the
ESK4 / ESK4A is not
compatible with the
add-on module, or
the ribbon cable
connection is not
connected properly.
Linearization is not
activated = measuring
error
reset by
error/overflow
Internal
communication error
Memory error
detected during selftest.
A device restart has
taken place
The HART® multidrop
mode is activated.
The current output is
set to a fixed value of
4.5 mA.
Internal error in
device
Internal error in
device
device
Internal error in
device
SPerform calibration using ammeter and
menu 3.10 or using standard HART
tools/process control system and poss.
external ammeter. Caution: during
calibration, switch the measuring point to
manual control.
SPerform calibration using ammeter and
menu item 3.11 or using standard HART
tools and external ammeter if necessary.
Caution: during calibration, switch the
measuring point to manual control.
SThe device, together with an indication of the
SCheck that the ribbon cable fits properly.
SActivate linearization or carry it out again
S
1
FCheck whether the display is plugged in
FRestart device. If the error occurs again:
IThe device has been restarted using menu
I
FThe device must be sent back to the
F
1
error, must be sent back to the
manufacturer for checking.
Check module compatibility (see chapter
2.5).
®
(HART
software are required; the original
calibration values must be known), or send
the device back to the manufacturer for
linearization.
Because the reset time is not known:
Controlled reset of the counter using menu
item 1.6.1 or using HART
control system.
correctly and restart the device. If the error
occurs again: send the device back to the
manufacturer with an indication of the error.
send the device back to the manufacturer
with an indication of the error.
item 1.6.2 since the last time the error
messages were reset.
The HART® multidrop mode is activated with
selection of a polling address not equal to 0
using menu item 3.7. Polling address 0
activates the analogue output again.
manufacturer with an indication of the error.
communication and linearization
®
tools/process
®
®
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
41
Page 42
6
OPERATION
Error messageDescriptionCategoryRemedy
H250 M40
MEMORY
CORRUPTION
AO FIXEDThe current output is
AO SATURATEDCurrent output
ERROR TIMEOUTData not transferred,
WARNING
TIMEOUT
1 Category can be changed by user
Internal memory
error, caused by a
hardware or software
problem
set to a fixed value.
saturated
or transferred
incorrectly from the
ESK to the counter
module
FRestart the device; if the error occurs again
IThe current output is fixed and does not
IThe current output is saturated at >20.4 mA
FConfirm menu item 1.6.3 WRITE INFO I/O.
I
the device must be sent back to the
manufacturer with an indication of the error.
reflect the measured value. This is the case
in multidrop mode, with current output
test/calibration using the menu or HART
and is no longer linked to the measurement.
®
.
42
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 43
H250 M40
6.6 Menu indicator ESK4-T
6.6.1 Factory settings
MenuFunctionSetting
1.1.1OUTPUT B1INACTIVE
1.2.1OUTPUT B2INACTIVE
1.3.1PULSE WDTH100 ms
1.3.2PULSE/UNIT1 pulse / unit
1.4.1DISPLAYMeasured value
1.4.2ROTATION0°
1.5TIME CONST.1.0 s
1.6.1RESET COUNT.NO
1.6.2RESET ERRORNO
1.6.3RE-INIT I/ONO
3.1LANGUAGEENGLISH
3.2FUNCTION B1INACTIVE
3.3CONTACT B1NO CONTACT
3.4FUNCTION B2INACTIVE
3.5CONTACT B2NO CONTACT
3.6FUNCTION B3INACTIVE
3.7MULTIDROPPOLLING ADD: 00
3.84mA CALIBR.4.000 mA
3.920mA CALIBR.20.000 mA
3.10ALARM CURRENTALARM HIGH
3.11.1FS&UNITApplication specific
3.11.2COUNTERApplication specific
3.12EDIT UNITUser defined unit / factor
3.13LFC4% ON, 6% OFF
3.14DESCRIPTORFree text
3.15ENTRY CODEOFF
3.16BASIC SETTINGNO
OPERATION
6
INFORMATION!
The measuring device has been preset at the factory in accordance with the customer order.
Therefore subsequent configuration via the menu is only necessary if the intended use of the
device is changed.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
45
Page 46
6
OPERATION
6.6.3 Menu explanations
1 OPERATION
H250 M40
DesignationLevel Select /
Explanation
Input
OUTPUT B11.1Output B1 is a binary switching output. Under Fct. 3.2, one of the
One of the following functions can be selected as contact type under Fct. 3.3:
NO CONTACT 1 / NC CONTACT 2
1.1.1INACTIVE
MEAS.VALUE B1Flow value switching point
COUNTER.VAL B1Switching point totaliser value
PULSE WDTHPulse value (pulse/unit)
1.1.2HYST.B1Hysteresis for the flow value switching point
following functions can be attributed to this output:
INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.
Value range: 0.0…measuring range end value
The switching point is entered in flow units.
If the current flow value exceeds this predetermined switching point,
output B1 changes its binary state 3.
Under Fct. 1.1.2 a hysteresis can also be specified.
Value range 0.0…counter limit
The switching point is entered in volume or mass units.
If the the current counter value exceeds this predetermined switching
point, output B1 changes its binary state 3.
There is no hysteresis setting for the counter value switching point.
The value is now displayed here.
Configuration takes place under Fct. 1.3.1 (PULSE WDTH), Fct. 1.3.2
(PULSE/UNIT) and Fct. 3.11.2 (COUNTER).
Value range 0.0…switching point
If the current flow value exceeds the predetermined switching point
from Fct. 1.1.1, output B1 changes its binary state 3.
In order for output B1 to change its binary state back to the initial
setting, the switching point made smaller by the hysteresis must be
undershot 4.
46
Example:
Example:
Example:Example:
Under 1.1.1, a switching point of 200 L/h is set.
The possible value range for the hysteresis is then 0.0…200 L/h.
With a hysteresis value of 0, the switching point has no hysteresis
(3=4).
If a hysteresis value of 20 L/h is entered, output B1 changes its binary
state to the initial setting if 180 L/h is undershot 4.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 47
H250 M40
OPERATION
6
DesignationLevel Select /
Explanation
Input
OUTPUT B21.2Output B2 is a binary switching output. Under Fct. 3.4, one of the
One of the following functions can be selected as contact type under Fct. 3.5:
NO CONTACT 1 / NC CONTACT 2
1.2.1INACTIVE
MEAS.VALUE B2See FLOW.VAL B1
COUNTER.VAL B2See COUNTER.VAL B1
PULSE WDTH B2see PULSE WDTH Fct. 1.1.1
1.2.2HYST.B2See HYST. B1
PULSEOUTP.1.3
1.3.1PULSE WDTH
50msTi = 50 ms; f
100msTi = 100 ms; f
200msTi = 200 ms; f
500msTi = 500 ms; f
1.3.2PULSE/UNIT
0.001...
1000
following functions can be attributed to this output:
INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.
Under Fct. 1.2.2 a hysteresis can also be specified.
Configuration takes place under Fct. 1.3.1 (PULSE WDTH), Fct. 1.3.2
(PULSE/UNIT) and Fct. 3.11.2 (COUNTER).
= 10 Hz
max. pulse / h = 36000
max. pulse / h = 18000
max. pulse / h = 9000
max. pulse / h = 3600
Pulse per volume or mass unit of counter (setting under Fct. 3.11.2),
which can be output via one of the binary outputs. The maximum
frequency for the pulse output (see Fct 1.3.1) cannot be exceeded even
at maximum flow rate (final value).
max
max
max
max
= 5 Hz
= 2.5 Hz
= 1 Hz
Example:
Example:
Example:Example:
Final value Qmax = 1200 l/h; volume unit counter = Litre; pulse
duration = 100 ms;
If the factor 1 is entered, 1 pulse/litre = 1200 pulses are generated in
one hour at maximum flow rate.
Number of max. permissible pulses:
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
47
Page 48
6
OPERATION
H250 M40
DesignationLevel Select /
Explanation
Input
DISPLAY1.4Different measured values can be selected for permanent or
1.4.1Measured valuepermanent display of flow in flow units
COUNTERpermanent display totalizer
MEAS.VAL/CTNalternating display flow value in flow units and totalizer
MEAS.VAL&CTNsimultaneous display of flow value and totaliser
PERCENTpermanent display flow value in percent
1.4.20°The reading is not rotated.
90°The reading on the display is rotated by 90°.
180°The reading on the display is rotated by 180°.
270°The reading on the display is rotated by 270°.
TIME CONST.1.50.0...20.0sGiven in seconds
RESET1.6Local resetting of totalizer and acknowledgement of warnings.
alternating display.
The display can be rotated.
The output variables (value of the current loop and indicated flow
value) follow the current process by the value set here (in seconds)
with a time delay.
Note:
Note:
Note:Note:
If the current flow is polled via HART
measured value is also played back with a delay.
There is always a safety prompt (yes/no), to avoid accidental resetting.
®
communication, the transferred
Note:
Note:
Note:Note:
External resetting of the totalizer can be installed with the binary
input B3.
1.6.1COUNTERConfirming with YES resets the value of the totalizer to 0.0.
1.6.2ERRORConfirming with yes acknowledges all existing warnings.
Note:
Note:
Note:Note:
Confirming with YES acknowledges existing errors and warnings.
1.6.3RE-INIT IOAs a general rule, data is transferred from the counter module to the
ESK4 and vice versa when the device starts. To make sure, selecting
this menu point when confirming with YES can perform another data
transmission.
48
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 49
H250 M40
2 TEST & INFO
OPERATION
6
DesignationLevel Select /
Explanation
Input
4-20mA OUT2.1Testing the current loop by setting various current values
Note:
Note:
2.1.1NORMAL OPThe value of the current loop follows the current process.
2.1.24.0 mAThe value of the current loop no longer follows the current process.
2.1.35.6 mA
2.1.47.2 mA
2.1.58.8 mA
2.1.610.4 mA
2.1.712.0 mA
2.1.821.6 mA
2.1.915.2 mA
2.1.1016.8 mA
2.1.1118.4 mA
2.1.1220.0 mA
2.1.1321.6 mA
OUTPUT B12.2ALARM
CURRENT
2.3Testing the binary switching output B1 by changing its binary status.
Note:Note:
The simulation is not available in HART
Attention:
Attention:
Attention:Attention:
During the test, the value of the current loop does not follow the current
process.
It is set to the selected current value.
<3.6 / >21 mA
Testing the alarm current HIGH /LOW according to setting Fct. 3.10
®
multi-drop mode (see Fct. 3.7)
Attention:
Attention:
Attention:Attention:
During the test, the binary status does not correspond to the current
process.
2.3.1NORMAL OPThe binary status of the switching output corresponds to the current
2.3.2OPENThe binary status of the switching output no longer corresponds to the
2.3.3CLOSED
OUTPUT B22.4see test OUTPUT B1
2.4.1NORMAL OP
2.4.2OPEN
2.4.3CLOSED
INPUT B32.5TESTThe current binary status of the binary input B3 is displayed. External
DEVICE ID2.6Information for device identification
2.6.1ELEC. REV.Electronics revision
2.6.2SN ESK4Serial number ESK4
2.6.3PA ORDERProduction order for the complete flowmeter
2.6.4DEVICE SN.Serial number for the complete flowmeter
process.
current process. The selected status is tested.
switching of the voltage applied to input B3 activates a change of the
displayed binary status.
Note:
Note:
Note:Note:
A reaction to the switching of the voltage applied to input B3 only occurs
when the function of B3 is activated (see Fct. 3.6).
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
49
Page 50
6
OPERATION
H250 M40
DesignationLevel Select /
Explanation
Input
SOFT.VERSION2.7Information about software revision status
2.7.1FW. ESK4Software revision status for device insert ESK4-Basic
2.7.2FW. ESK4 I/OSoftware revision status for optional add-on module ESK4-T
TAG NO.2.88 digitsTAG Designation
LONG TAG2.932 digitsTAG Designation
Alphanumerical characters. Up to eight digits possible
Alphanumerical characters. Up to 32 digits possible
3 INSTALLATION
DesignationLevel Select /
Input
LANGUAGE3.1Language selection for the menu texts displayed
3.1.1ENGLISHThe menu texts are displayed in the selected language.
3.1.2DEUTSCH
3.1.3FRANCAIS
3.1.4ITALIANO
3.1.5ESPANOL
3.1.6CESKY
3.1.7POLSKI
3.1.8NEDERLANDS
3.1.9DANSK
FUNCTION B13.2INACTIVEThe binary switching output B1 has no function.
SWITCHING
POINT
COUNTERLIM.The binary switching point B1 functions as a limit switch depending on
PULSEOUTP.The binary switching point B1 functions as a pulse output depending on
Explanation
The binary switching point B1 functions as a limit switch depending on
the current flow value.
Configuration of the switching point takes place under Fct. 1.1.1
(MEAS.VALUE B1)
the current counter value.
Configuration of the switching point takes place under Fct. 1.1.1
(MEAS.VALUE B1)
the current flow value.
Pulses can be generated up to a maximum of 10 Hz. Configuration
occurs under Fct. 1.3.1 (PULSE WDTH) and Fct. 1.3.2 (PULSE/UNIT).
Note:
Note:
Note:Note:
Configuration under Fct 1.3.1 and Fct. 1.3.2 applies to both pulse
outputs. If both B1 and B2 are configured as pulse outputs, both binary
outputs behave in exactly the same way.
CONTACT B13.3NO CONTACTBinary switching output B1 is an NO contact.
NC CONTACTBinary switching output B1 is an NC contact.
50
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 51
H250 M40
OPERATION
6
DesignationLevel Select /
Explanation
Input
FUNCTION B23.4INACTIVESee FUNCTION B1
SWITCHING
POINT
COUNTERLIM.See FUNCTION B1
PULSEOUTP.See FUNCTION B1
CONTACT B23.5NC CONTACTSee CONTACT B1
NO CONTACTSee CONTACT B1
FUNCTION B33.6INACTIVEThe binary switching input B3 has no function.
ACTIV HThe internal flow counter is reset to 0.0 if there is an H-level at input B3
ACTIV LThe internal flow counter is reset to 0.0 if there is an H-level at input B3
STARTH STOPLThe counter is started by creating an H level at the B3 input and stopped
STARTL STOPHThe counter is started by creating an L level at the B3 input and stopped
MULTIDROP3.70…15
See FUNCTION B1
Configuration of switching point occurs under Fct. 1.2.1 (MEAS. VALUE
B2).
Configuration of switching point occurs under Fct. 1.2.1 (MEAS. VALUE
B2).
Configuration occurs under Fct. 1.3.1 (PULSE WDTH) and Fct. 1.3.2
(PULSE/UNIT).
for at least 100 ms.
for at least 100 ms.
by creating an L level at the B3 input.
by creating an H level at the B3 input.
Polling adress for HART® multi-drop mode
When the address is 0, the HART
®
multi-drop mode is disabled.
Attention:
Attention:
Attention:Attention:
When the HART
current loop is inactive (fixed current value of 4.5 mA) and no longer
follows the current process.
4mA CALIBR.3.8D/A calibration for beginning of measuring range (4 mA)
Note:
Note:
Note:Note:
This function is not available in HART
Attention:
Attention:
Attention:Attention:
During calibration, the value of the current loop does not follow the
current process.
If a current measuring device in the 4...20 mA loop detects a deviation
from the desired value 4.000 mA, the measured value must be entered.
The correction value is accepted following the query SAVE and Confirm
with YES.
20mA CALIBR.3.9D/A calibration for the final value of the measuring range (20 mA)
Note:
Note:
Note:Note:
This function is not available in HART
Attention:
Attention:
Attention:Attention:
During calibration, the value of the current loop does not follow the
current process.
If a current measuring device in the 4...20 mA loop detects a deviation
from the desired value 20.000 mA, the measured value must be entered.
The correction value is accepted following the query SAVE and Confirm
with YES.
®
multi-drop mode (Address 1…15) is activated, the
®
multi-drop mode.
®
multi-drop mode.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
51
Page 52
6
OPERATION
H250 M40
DesignationLevel Select /
Explanation
Input
ALARM
CURRENT
END&UNIT3.11By changing the unit, the respective final value is scaled accordingly.
EDIT UNIT3.12User-defined unit with conversion to calibrated unit
LFC3.13LFC stands for LLLLow FFFFlow CCCCutoff
3.10OFFError indication via the current loop is deactivated. The current loop
ALARM HIGHError indication via the current loop is activated (failure signal "high"
ALARM LOWError indication via the current loop is activated (failure signal "low" as
3.11.1Measured valueFor units for volume flow or mass flow, refer to
3.11.2COUNTERFor units for volume counter or mass counter and pulse output refer to
follows the current process.
Note:
Note:
Note:Note:
This function is not available in HART
acc. to NE43).
Note:
Note:
Note:Note:
This function is not available in HART
per NE43).
Note:
Note:
Note:Note:
1) This function is not available in HART
2) This function is supported as of Electronic Revision 2.2.x.
Depending on calibration, either units for volume or mass flow
measurement are available for selection.
®
multi-drop mode.
®
multi-drop mode.
®
multi-drop mode.
Overview of the units
ESK4-T
Overview of the units ESK4-T
To ensure a stable zero point of the current output, the current output
can be stably set to 4.00 mA 3 in a range to be selected.
on page 39
on page 39
3.13.1CONTROL OFFLFC function is not active.
CONTROL ONLFC function is active.
3.13.2LFC ON_VALUESwitch-on value 1:
3.13.3LFC OFF_VALUESwitch-off value 2:
DESCRIPTOR3.1412 digitsFree text input for header in LCD
Value range 1…19% (from final value of measuring range)
Flow is greater than the switch-on value. The current output
corresponds to this.
If the flow rate falls, the current output follows until the switch-on value
1.
If the flow value continues to fall, the current output is switched to
4.00 mA 3.
Note:
Note:
Note:Note:
The switch-on value to be set must be smaller than the previously
selected switch-off value.
Value range 2…20% (from the measuring range final value)
The flow rate is 0. The current output is 4.00 mA 3.
If the flow rate increases, the current output remains at 4.00 mA until it
has reached the switch-off value 2.
Note:
Note:
Note:Note:
The switch-off value to be set must be greater than the previously
selected switch-on value.
52
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 53
H250 M40
OPERATION
6
DesignationLevel Select /
Explanation
Input
INP. CODE3.15Input code for the local operating menu
3.15.1OFFUse of an input code is not activated.
3.15.2ONIf YES is selected, the last code that was entered has to be typed in.
BASIC SETTING3.16Resetting parameters to factory default
In the factory setting, the input code is not active.
Factory code: → → → ^^^ ↑ ↑ ↑
If, after confirmation with YES, the → button is also pressed, then a new,
individual, nine-element code can be typed in.
The desired button combination is visualized on the display.
There is always a safety prompt (yes/no), to avoid accidental resetting.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
53
Page 54
7
SERVICE
7.1 Maintenance
Within the scope of routine maintenance of the system and pipelines, the flowmeter should also
be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to
the measuring tube and indicator.
We advise that inspections be carried out at least once per year.
The device must be removed from the piping before cleaning.
CAUTION!
Pressurized pipes must be depressurized before removing the device.
Empty pipes as completely as possible.
In the case of devices used for measuring aggressive or hazardous media, appropriate safety
precautions must be taken with regard to residual liquids in the measuring unit.
Always use new gaskets when reinstalling the device in the pipeline.
Avoid electrostatic charges when cleaning the surfaces (e.g. sight window)!
7.2 Replacement and retrofitting
H250 M40
Some of the variable area flowmeter components can be retrofitted:
• Float damping
Indicator M40:
Indicator M40:
Indicator M40:Indicator M40:
• Limit switch module K1 / K2
• Current output 4..20mA ESK4 / ESK4A
• Counter module with LCD and I/O ESK4-T
• Fieldbus interface ESK4-PA / FF
7.2.1 Replacing floats
• Remove the device from the piping.
• Take the upper snap ring out of the measuring unit.
• Take the upper float stop and float out of the measuring unit.
• Insert the new float into the centre hole of the lower float stop and push into the measuring
unit along with the upper float catcher. While doing this, the float's upper guide rod must be
guided through the middle hole of the float stop.
• Insert the snap ring into the measuring unit.
• Fit the device back into the piping.
CAUTION!
An additional measuring error is to be expected if recalibration is not conducted.
54
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 55
H250 M40
7.2.2 Retrofitting float damping
• Take the upper snap ring 1 out of the measuring unit.
• Take the upper float stop 2 and float 5 out of the measuring unit.
• Fasten the locking ring 3 into the lower slot of the float's guide rod.
• Slide ceramic sleeve 4 on to the float's guide rod and attach it to the top slot using the locking
ring 3.
• Insert float into the lower float guide in the measuring unit.
• Retrofit the supplied damping cylinder with the integrated float stop 2 into the measuring
unit.
• Insert upper snap ring 1.
SERVICE
7
1 Snap ring
2 Float stop
3 Locking ring
4 Ceramic sleeve
5 Float
• Loosen the locking screw 1 on the contact pointer.
• Insert the contact module into slot 3 of the bracket until the semi-circle 1 on the contact
board surrounds the pointer cylinder.
The contact module connecting terminals feature a pluggable design and can be removed in
order to connect the cables.
WARNING!
Do not damage the indicating element!
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
55
Page 56
7
SERVICE
7.2.4 Replacement - Retrofitting ESK4 / ESK4A
When replacing or retrofitting an ESK4 / ESK4A, the following are required at the time
of ordering:
• SN - serial number or
• SO - sales order
This information can be found on the indicator nameplate.
The ESK4 / ESK4A is factory calibrated, making it possible to replace it or retrofit it without
recalibrating.
• De-energise the ESK4 / ESK4A.
• Lift and remove the ESK4 / ESK4A with a screwdriver.
WARNING!
Do not damage the indicating element!
H250 M40
• The ESK4 / ESK4A plug-in tongues 1 are plugged in under the two bolts 2 on the baseplate.
• Slight pressure is used to press the ESK4 / ESK4A onto the spring pins 3 until it stops, firmly
attaching the ESK4 / ESK4A.
If a change in measuring range, product temperature, product, density, viscosity or pressure is
®
desired, this can be done with the KroVaCal program and with a HART
measuring unit is subject to its own physical limits, which the KroVaCal program correctly
calculates, and may thus reject the desired change. If a change is performed using the program,
the new data is also transmitted to the ESK4 / ESK4A.
• Device identification
• Device address
• Serial Number
• Measuring point identifier
• Digital measured value query in flow units, % and mA
• Test / setting functions
• Calibration 4.00 and 20.00 mA
• Set current output to any desired value
modem. However, each
56
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The add-on module for the ESK4 / ESK4A can be replaced or retrofitted on-site without removing
the device from the process control.
• ESK4-T (Display module with LCD and I/O)
• ESK4-PA (Profibus PA interface)
• ESK4-FF (Foundation Fieldbus interface)
More detailed information can be found in the service manual included with each retrofit or
replacement set.
7.3 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each
device or each important accessory part will be kept available for a period of 3 years after
delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal
operating conditions.
SERVICE
7
7.3.1 List of spare parts
Spare partOrder no.
DN 15
DN 15
DN 15DN 15
Float CIV 15, 1.4404X251041000
Float DIV 15, 1.4404X251042000
Float TIV 15, 1.4404X251043000
Float DIVT 15, 1.4404X251044000
Float TIV 15, AluminiumX251043100
Float TIV 15, TitaniumX251043200
Set float stop; standard (1 float stop, 1 span ring)X251050100
Set float stop; gas damping (ZrO2)X251050200
Set of float stop; gas damping (PEEK)X251050300
Damping bush (7x8) ZrO2 incl. 2 locking ringsX251053100
Damping bush (7x8) PEEK incl. 2x locking ringsX251053200
DN 25
DN 25
DN 25DN 25
Float CIV 15, 1.4404X252041000
Float DIV 25, 1.4404X252042000
Float TIV 25, 1.4404X252043000
Float DIVT 25, 1.4404X252044000
Set float stop; standard (1 float stop, 1 span ring)X252050100
Set float stop; gas damping (ZrO2)X252050200
Set of float stop; gas damping (PEEK)X252050300
Damping bush (7x8) ZrO2 incl. 2x locking ringsX252053100
Damping bush (12x8) PEEK incl. 2x locking ringsX252053200
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
57
Page 58
7
SERVICE
H250 M40
Spare partOrder no.
DN 50
DN 50
DN 50DN 50
Float CIV 55, 1.4404X253041000
Float DIV 55, 1.4404X253042000
Float TIV55, 1.4404X253043000
Float DIVT 55, 1.4404X253044000
Set float stop; standard (1 float stop, 1 span ring)X253050100
Set float stop; gas damping (ZrO2)X253050200
Set of float stop; gas damping (PEEK)X253050300
Damping bush (7x8) ZrO2 incl. 2x locking ringsX253053100
Damping bush (14x10) PEEK incl. 2x locking ringsX253053200
DN 80
DN 80
DN 80DN 80
Float CIV 85, 1.4404X254041000
Float DIV 85, 1.4404X254042000
Float TIV 85, 1.4404X254043000
Float DIVT 85, 1.4404X254044000
Set float stop; standard (1 float stop, 1 span ring)X254050100
Set float stop; gas damping (ZrO2)X254050200
Set of float stop; gas damping (PEEK)X254050300
Damping bush (18x14) ZrO2 incl. 2x locking ringsX254053100
Damping bush (18x14) PEEK incl. 2x locking ringsX254053200
DN 100
DN 100
DN 100DN 100
Float CIV 105, 1.4404X255041000
Float DIV 105, 1.4404X255042000
Float DIVT 105, 1.4404X255044000
Set float stop; standard (1 float stop, 1 span ring) only for bottom !X255050100
Set float stop; gas damping (ZrO2)X255050200
Set of float stop; gas damping (PEEK)X255050300
Damping bush (18x14) ZrO2 incl. 2x locking ringsX254053100
Damping bush (18x14) PEEK incl. 2x locking ringsX254053200
Indicator M40:
Indicator M40:
Indicator M40:Indicator M40:
Housing components
Housing components
Housing componentsHousing components
Standard indicator housing M40 complete no scaleX251110000
Standard indicator housing M40R complete, no scale* (stainless steel, non-coated) X251111000
Standard indicator housing M40S complete, no scale * (2-layer coating)X251112000
Cover M40 complete, standardX251110100
Standard cover M40S (2-layer coating)X251110200
Standard cover M40R (stainless steel, non-coated)X251110400
Cover sealX251112100
M40 baseplate standardX251120100
Standard M40S baseplate (2-layer coating) *X251120200
Standard M40R baseplate (stainless steel, non-coated) *X251120300
Standard = Not pressure tight
Standard = Not pressure tight
Standard = Not pressure tightStandard = Not pressure tight
58
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Contact module K1 min SB3,5 E2 - act.LowX251133900
Contact module K1 max SB3,5 E2 - act.LowX251134000
Contact module K1 min / max SB3,5 E2 - act.LowX251134100
3-wire limit module normally open
3-wire limit module normally open
3-wire limit module normally open3-wire limit module normally open
Contact module K1 min SB3,5 E2 - act.HighX251134200
Contact module K1 max SB3,5 E2 - act.HighX251134300
Contact module K1 min / max SB3,5 E2 - act.HighX251134400
Electronic modules
Electronic modules
Electronic modulesElectronic modules
ESK4, ESK4-FF, ESK4-PA, ESK4-FF (serial number required)
Cover for add-on modulesX251121500
Connection cable ESK4 to add-on modulesX251121600
Plugs (10 pcs) for ESK4 connection for bus modulesX251132500
Replacement connector
Replacement connector
Replacement connectorReplacement connector
Replacement connector 11/12 for ESK4X251121700
Replacement connector D/D+ for ESK4-FF or ESK4-PAX251121800
Replacement connectors 1/2/3, 4/5/6, 7/8 for ESK4-TX251121900
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
59
Page 60
7
SERVICE
Spare partOrder no.
Replacement cable gland
Replacement cable gland
Replacement cable glandReplacement cable gland
Single cable gland M20x1.5 black, plastic - for non-Ex / Ex-iX251150300
Single cable gland M20x1.5 blue, plastic - for non-Ex / Ex-iX251150100
Single cable gland M20x1.5 brass - for non-Ex / Ex-i / Ex-nAX251151000
Single cable gland M20x1.5 brass Ex-d/t - for Ex-d / Ex-tX251152000
Single plug M20x1.5 brass Ex-d/t - for Ex-d / Ex-tX251153000
Single cable gland M20x1.5 stainless steel Ex-d/t - for Ex-d / Ex-tX251154000
Single plug M20x1.5 stainless steel Ex-d/t - for Ex-d / Ex-tX251155000
Other spare parts on request
H250 M40
60
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 61
H250 M40
7.4 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the
warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
7.5 Returning the device to the manufacturer
7.5.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance
with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict
attention to the following points:
•
Due to statutory regulations on environmental protection and safeguarding the health and
safety of the personnel, the manufacturer may only handle, test and repair returned devices
that have been in contact with products without risk to personnel and environment.
•
This means that the manufacturer can only service this device if it is accompanied by the
following certificate (see next section) confirming that the device is safe to handle.
SERVICE
7
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products,
you are kindly requested:
•
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from
such dangerous substances,
•
to enclose a certificate with the device confirming that is safe to handle and stating the
product used.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
61
Page 62
7
SERVICE
7.5.2 Form (for copying) to accompany a returned device
CAUTION!
To avoid any risk for our service personel, this form has to be accessible from outside of the
packaging with the returned device.
Company:Address:
Department:Name:
Tel. no.:Fax no. and/or Email address:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
This medium is:radioactive
water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
H250 M40
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the
device when it is returned.
Date:Signature:
Stamp:
7.6 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:Separate collection of WEEE (Waste Electrical and Electronic Equipment) in the European Union:
According to the directive 2012/19/EU, the monitoring and control instruments marked with the
WEEE symbol and reaching their end-of-life must not be disposed of with other waste
The user must dispose of the WEEE to a designated collection point for the recycling of WEEE or
send them back to our local organisation or authorised representative.
must not be disposed of with other waste.
must not be disposed of with other wastemust not be disposed of with other waste
62
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 63
H250 M40
8.1 Functional principle
The H250 flowmeter operates in accordance with the float measuring principle. The measuring
unit consists of a metal cone in which a float can move freely up and down. The flow goes from
bottom to top. The float changes position so that the lifting force acting on it F1 is in equilibrium
with the form drag F2 and its weight F3: F3 = F1 + F2
TECHNICAL DATA
8
1 Indication principle M40 magnetic coupling
2 Magnetic coupling sensors
1 For the indicator, the flow-dependent height of the float in the measuring unit is transmitted
by means of a magnetic coupling and displayed on a scale.
2 For a built-in signal converter (ESK4 / ESK4A), the flow-dependent height of the float in the
measuring unit is detected by the S1 and S2 magnetic field sensors and electronically
processed.
Operating principle of H250H and H250U
Figure 8-1: Operating principle H250H and H250U
1 H250H - horizontal flow direction
2 H250U - flow direction from top to bottom
The flowmeter operates in accordance with a modified float measuring principle.
The guided float adjusts itself so that the flow force acting on it is in equilibrium with the
opposing spring force. The flow-dependent position of the float in the measuring unit is
displayed on a scale by means of a magnetic coupling.
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
63
Page 64
8
TECHNICAL DATA
8.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Downloadcenter).
Measuring system
Application rangeFlow measurement of liquids, gases and vapours
Function / Measuring principleVariable area measuring principle
Measured value
Primary measured valueFloat position
Secondary measured valueOperating and standard volumetric flow
TemperatureTemperature
Max. operating temperature TS-196..+300°C / -321...+572°F
Pressure
Pressure
PressurePressure
Max. operating pressure PSDepending on the version (refer to nameplate)
Max. test pressure PTDepending on the version (refer to nameplate)
Min. required operating pressure2 times greater than pressure loss (refer to measuring
Protection category
Protection category
Protection categoryProtection category
M40, M40S, M40RIP66/68 acc. to EN 60529, NEMA 4/4X/6 acc. to NEMA 250
M40RIP69K acc. to DIN 40050-9
Float damping during gas measurement recommended
Float damping during gas measurement recommended
Float damping during gas measurement recommendedFloat damping during gas measurement recommended
DN15...25 / ½...1"Operating pressure: <0.3 bar / 4.4 psig
DN50...100 / 2...4"Operating pressure: <0.2 bar / 2.9 psig
Depending on the version (refer to nameplate)
ranges)
Installation conditions acc. to VDI/VDE 3513 Sheet 3
Inlet section≥ 5 x DN
Outlet section≥ 3 x DN
64
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
1 Display contrast out of the temperature range 0...+60°C / +32..+140°F decreasing.
1
-40...+70-40...+158
Temperatures H250/M40 - with electrical components [°C]
T
< +40°CT
amb.
ENASMEVersion withStandard HTStandard HT
DN15,
DN25
DN 502“ESK4 / ESK4A, -FF, -PA+200+300+165+300
DN 80,
DN100
½", 1"ESK4 / ESK4A, -FF, -PA+200+300+180+300
ESK4-T+200+300+80+130
Limit switch NAMUR+200+300+200+300
3-wire limit switch+200+300+130+295
ESK4-T+180+300+75+100
Limit switch NAMUR+200+300+200+300
3-wire limit switch+200+300+120+195
3“, 4"ESK4 / ESK4A, -FF, -PA+200+300+150+250
ESK4-T+150+270+70+85
Limit switch NAMUR+200+300+200+300
3-wire limit switch+190+300+110+160
< +60°C 1
amb.
Maximum product temperatures H250/M40 - with electrical components [°F]
T
< +104°FT
amb.
ENASMEVersion withStandard HTStandard HT
DN15,
DN25
DN 502“ESK4 / ESK4A, -FF, -PA392572165572
DN 80,
DN100
½", 1"ESK4 / ESK4A, -FF, -PA392572356572
ESK4-T392572176266
Limit switch NAMUR392572392572
3-wire limit switch392572266563
ESK4-T356572167212
Limit switch NAMUR392572392572
3-wire limit switch392572248383
3“, 4"ESK4 / ESK4A, -FF, -PA392572302482
ESK4-T302518158185
Limit switch NAMUR392572392572
3-wire limit switch374572230320
1 if there are no heat insulation measures, a heat-resistant cable is necessary (continuous
operating temperature of the cable to be used: +100°C / +212°F)
< +140°F 1
amb.
66
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 67
H250 M40
TECHNICAL DATA
Abbreviation
HTHigh-Temperature version
ESK4 / ESK4A 2-wire current output 4…20 mA with HART 5 / HART 7
ESK4-TESK4 with LCD, binary status outputs, digital counter and pulse output.
ESK4-FFFOUNDATION FIELDBUS interface
ESK4-PAPROFIBUS PA interface
Cable glands
Cable glandMaterialCable diameter
M 20x1.5 StandardPA8...13 mm0.315...0.512"
M 20x1.5Nickel-plated brass10...14 mm0.394...0.552"
Power supply14...30 VDC (12...30 VDC without ESK4-T)
Min. power supply for HART
Measuring signal4.00...20.00 mA = 0...100% flow value in 2-wire technology
Power supply influence<0.1%
External resistance dependency<0.1%
Temperature influence<5 µA / K
Max. external resistance / load650 Ohm (30 VDC)
Min. load for HART
®
®
2
2.5 mm
20 VDC, load 250 Ohm
250 Ohm
2
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
67
Page 68
8
TECHNICAL DATA
ESK4 HART® configuration
Manufacturer name (code)KROHNE Messtechnik (0x45 = 69)
Model name / HART® Revision
Physical layerFSK
ESK4 / ESK4A process variable
Over range+102.5 (±1%)20.24...20.56
Device error identification> 106.25>21.00
Max. current consumption131.2525
Multi-drop mode4.5
ESK4 (214 = 0xD6) / HART 5.9
ESK4A (17854 = 0x45BE) / HART 7.4
Values [%] from full scale
range
Signal output [mA]
H250 M40
ESK4-FF
Physical layerIEC 61158-2 and FISCO model
Communication standardH1 FOUNDATION Fieldbus protocol
ITK version5.2
Power supplyBus supply
Nominal current16 mA
Error current23 mA
Boot-up current after 10 ms< Nominal current
For more details refer to the supplementary instructions "H250 M40 Foundation Fieldbus".
ESK4-PA
Physical layerIEC 61158-2 and FISCO model
Communication standardProfibus PA profile 3.02
PNO ID4531 HEX
Power supplyBus supply
Nominal current16 mA
Error current23 mA
Boot-up current after 10 ms< Nominal current
68
For more details refer to the supplementary instructions "H250 M40 Profibus PA".
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 69
H250 M40
TECHNICAL DATA
ESK4-T with LCD, binary inputs and outputs and digital counter
Binary output
Two binary outputsGalvanically isolated, passive
ModeSwitch outputNAMUR or transistor (open
Configurable asSwitch contact or
NAMUR switch output
Power supply8.2 VDC
Signal current> 3 mA switching value not reached < 1 mA switching value reached
Switch output transistor (open collector)
Power supplyNominal 24 VDC, maximal 30 VDC
P
max
Continuous currentMax. 100 mA
No load current I
0
pulse output
500 mW
≤ 2mA
collector)
NC/NO contact or
max. 10 pulses/s
8
Pulse output
T
on
T
off
Pulse value
Configurable from 50...500 ms
Depending on flow rate
Configurable in flow units e.g. 5 pulses/m
3
Binary input
InputGalvanically isolated
ModeReset counters or run/stop
Configurable asactive HI / active LO
H signal16...30 VDC
Internal resistance R
Ton (active)≥ 500 ms
i
Typically 20 kΩ
LC display
TechnologyPassive graphic LCD
DisplayMeasurement with units and/or counter reading with units. Counter
ConfigurationLocal plain text menu navigation via microswitch or bar magnet or using
reading max. 11-digit with power failure proof storage.
Binary flags for limit value status. 0...100% bar graph for measurement.
NE 107 diagnostic status symbols. Plain text menu for configuration.
DD/DTM software
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
69
Page 70
8
TECHNICAL DATA
Approvals
StandardDisplayMarking
ATEX /
IECEx
FM (USA) NEC500
FM (Kanada)
NEC505
NEPSIM40 electricalEx ia, Ex d, Ex nA, Ex t
CCOE/PESOM40 electricalEx ia, Ex d
EAC/GOSTM40 electricalEx ia, Ex d
INMETROM40 electricalEx ia, Ex d, Ex nA, Ex t
KGSM40 electricalEx ia, Ex d, Ex nA, Ex t
H250 M40
M40 mechanicalII2GD IIC
II3GD IIC
M40 electricalII2G Ex ia IIC T6 Gb
II2G Ex d IIC T6 Gb
II3G Ex nA IIC T6 Gc
II2D Ex t IIIC T70°C Db
II2D Ex ia IIIC T85°C Db
M40 electricalIS Class I Div 1
XP Class I Div 1
NI Class I Div 2
DIP Class II / III Div 1
70
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 71
H250 M40
TECHNICAL DATA
8.3 Dimensions and weights
Dimensions H250/M40
Front viewSide viewwith heatingHigh temperature
8
abdh
[mm]["][mm]["][mm]["][mm]["]
H250/RR Flange
H250/F Clamp connection
H250/RR from 2" 600 lb
ISO 228 / ASME B1.20.1 / SMS
1415.562509.851505.911505.91
30011.82
ENASMEc 1e 2Ø fgj
[mm]["][mm]["][mm]["][mm]["][mm]["]
DN15½"943.701144.49200.80973.821977.76
DN251"943.701275.00321.281094.272098.23
DN502"1074.221415.55652.571254.902228.74
DN803"1074.221576.18893.511435.612389.37
DN1004"1074.221676.571144.501505.912489.76
1 without cable gland 2 Ex d, Ex t, Ex nA +10mm [0,39"]
1 P >0.5 bar
2 with TR float
3 300 mbar with damping (gas measurement)
2
--50
1
1
2
1
1
1
1
1
1
1
1
12212838
13223250
---85
12263878
133045103
8137460
8137769
91384104
8166895
91689125
--120220
3
74
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
1 P >7.4 psig
2 with TR float
3 4.4 psig with damping (gas measurement)
2
--31.0
1
1
2
1
1
1
1
1
1
1
1
0.180.310.410.56
0.190.320.470.74
---1.25
0.180.380.561.15
0.190.440.661.51
0.120.191.090.88
0.120.191.131.01
0.130.191.231.53
0.120.241.001.40
0.130.241.311.84
-1.763.23
3
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bar abs. (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bara (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bara (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
80
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bara (DIN 1343)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
The operating pressure should be at least double the pressure loss for liquids and five times for
gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other
flow ranges on request. Conversion of other media or operating data is performed using the
calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements:
Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bara (ISO 13443)
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
81
Page 82
9
NOTES
H250 M40
82
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
Page 83
H250 M40
NOTES
9
www.krohne.com08/2015 - 4000640703 - MA H250 M40 R03 en
83
Page 84
KROHNE – Process instrumentation and measurement solutions