KROHNE H250-M40 User Manual

Page 1
Handbook
Handbook
H250 M40
H250 M40
H250 M40 H250 M40
Variable area flowmeter
HandbookHandbook
© KROHNE 11/2012 - 4000640702 MA H250 M40 R02 en
Page 2
: IMPRINT :::::::::::::::::::::::::::::::::::::::
All rights reserved. It is prohibited to reproduce this documentation, or any part thereof, without the prior written authorisation of KROHNE Messtechnik GmbH.
Subject to change without notice.
Copyright 2012 by KROHNE Messtechnik GmbH - Ludwig-Krohne-Str. 5 - 47058 Duisburg (Germany)
2
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 3
H250 M40
CONTENTS
1 Safety instructions 5
1.1 Intended use ..................................................................................................................... 5
1.2 Certifications .................................................................................................................... 6
1.3 Pressure equipment directive.......................................................................................... 6
1.4 Safety instructions from the manufacturer ..................................................................... 7
1.5 Copyright and data protection.......................................................................................... 7
1.5.1 Disclaimer ............................................................................................................................... 7
1.5.2 Product liability and warranty ................................................................................................ 8
1.5.3 Information concerning the documentation........................................................................... 8
1.5.4 Warnings and symbols used................................................................................................... 9
1.6 Safety instructions for the operator................................................................................. 9
2 Device description 10
2.1 Scope of delivery............................................................................................................. 10
2.2 Device version................................................................................................................. 11
2.2.1 Indicator versions ................................................................................................................. 12
2.2.2 Float damping ....................................................................................................................... 14
2.2.3 Pointer damping.................................................................................................................... 14
2.3 Nameplate ...................................................................................................................... 15
2.4 Description code............................................................................................................. 16
2.5 Electronics revision........................................................................................................ 17
3 Installation 18
3.1 Notes on installation ......................................................................................................18
3.2 Storage ........................................................................................................................... 18
3.3 Installation conditions ....................................................................................................19
3.3.1 Tightening torques................................................................................................................ 21
3.3.2 Magnetic filters ..................................................................................................................... 21
3.3.3 Heat insulation ...................................................................................................................... 22
4 Electrical connections 23
4.1 Security information.......................................................................................................23
4.2 Electrical connection indicator M40............................................................................... 24
4.2.1 Indicator M40 - limit switches .............................................................................................. 24
4.2.2 Current output ESK4............................................................................................................. 27
4.2.3 Binary inputs/outputs ESK4-T .............................................................................................. 30
4.2.4 Connection Harting HAN® 7D.............................................................................................. 33
4.3 Grounding connections................................................................................................... 34
4.4 Protection category ........................................................................................................34
5 Start-up 35
5.1 Standard device .............................................................................................................. 35
5.2 Indicator ESK4-T ............................................................................................................ 35
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
3
Page 4
CONTENTS
H250 M40
6 Operation 36
6.1 ESK4- Loop Check Modus .............................................................................................. 36
6.2 Operating elements ESK4-T........................................................................................... 37
6.3 Basic principles of operation ESK4-T ............................................................................ 38
6.3.1 Functional description of the buttons................................................................................... 38
6.3.2 Navigation within the menu structure.................................................................................. 38
6.3.3 Changing the settings in the menu....................................................................................... 39
6.4 Overview of the units ESK4-T......................................................................................... 40
6.5 Error messages ESK4-T................................................................................................. 40
6.6 Menu indicator ESK4-T................................................................................................... 43
6.6.1 factory settings ..................................................................................................................... 43
6.6.2 Menu structure .................................................................................................................... 44
6.6.3 Menu explanations................................................................................................................ 46
7 Service 54
7.1 Maintenance ................................................................................................................... 54
7.2 Replacement and retrofitting......................................................................................... 54
7.2.1 Replacing floats .................................................................................................................... 54
7.2.2 Retrofitting float damping..................................................................................................... 55
7.2.3 Retrofitting limit switch ........................................................................................................ 55
7.2.4 Replacement - Retrofitting ESK4 ......................................................................................... 56
7.3 Spare parts availability...................................................................................................57
7.3.1 List of spare parts................................................................................................................. 57
7.4 Availability of services .................................................................................................... 59
7.5 Returning the device to the manufacturer..................................................................... 59
7.5.1 General information.............................................................................................................. 59
7.5.2 Form (for copying) to accompany a returned device............................................................ 60
7.6 Disposal .......................................................................................................................... 60
8 Technical data 61
8.1 Functional principle........................................................................................................ 61
8.2 Technical data................................................................................................................. 62
8.3 Dimensions and weights ................................................................................................ 68
8.4 Measuring ranges...........................................................................................................71
9 Notes 79
4
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 5
H250 M40
1.1 Intended use
CAUTION!
Responsibility for the use of the measuring devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator.
INFORMATION!
The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose.
The variable area flowmeters are suitable for measuring clean gases, vapours and liquids.
Intended use:
The product may not contain any ferromagnetic particles or solids. It may be necessary to install magnetic filters or mechanical filters.
The product must be sufficiently liquid and free of deposits.
Avoid pressure surges and pulsing flows.
Open valves slowly. Do not use solenoid valves.
SAFETY INSTRUCTIONS 1
Use suitable measures to eliminate compression vibrations during gas measurements:
Short pipeline lengths to next restriction
Nominal pipe size not greater than nominal device size
Use of floats with damping
Increase in operating pressure (while taking into account the resulting change in density and
thus change in scale)
Observe installation conditions according to VDI/VDE 3513-3
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Responsibility for the use of the measurement devices with regard to suitability, intended use and corrosion resistance of the used materials against the measured fluid lies solely with the operator. The manufacturer is not liable for any damage resulting from improper use or use for other than the intended purpose. Do not use any abrasive media containing solid particles or highly viscous media.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
5
Page 6
1 SAFETY INSTRUCTIONS
1.2 Certifications
CE marking
The device fulfils all applicable statutory requirements of the following EC directives:
Pressure equipment directive 97/23/EC
For devices with electrical installations: EMC Directive 2004/108/EC
Devices for use in hazardous areas: ATEX Directive 94/9/EC
as well as
NAMUR recommendations NE 21 and NE 43 The manufacturer certifies successful testing of the product by applying the CE mark.
H250 M40
1.3 Pressure equipment directive
A conformity assessment in accordance with pressure equipment directive 97/23/EC has been carried out for the devices described. Conformity is certified by applying the CE mark. The number of the notified body is also stated.
The PED – Key describes rating of devices:
PED/G1/III/H G Gases and steams 1 Fluid group 1 III Category III H Conformity assessment method according to Module H
The PED key identification can be found on the nameplate of the device (see section 2.3 Nameplate)
INFORMATION!
The stated pressures (PS) and temperatures (TS) only apply as refers to the pressure resistance of the sensor body. As regards the functionality of the entire device, further restrictions of the maximum temperature may need to be observed (e.g. ATEX approval). Devices rated below Category I (Article 3/3) due to their size, do not receive the CE mark in the scope of the PED. These devices are subject to applicable Sound Engineering Practice SEP.
6
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 7
H250 M40
1.4 Safety instructions from the manufacturer
1.5 Copyright and data protection
The contents of this document have been created with great care. Nevertheless, we provide no guarantee that the contents are correct, complete or up-to-date.
The contents and works in this document are subject to copyright. Contributions from third parties are identified as such. Reproduction, processing, dissemination and any type of use beyond what is permitted under copyright requires written authorisation from the respective author and/or the manufacturer.
The manufacturer tries always to observe the copyrights of others, and to draw on works created in-house or works in the public domain.
The collection of personal data (such as names, street addresses or e-mail addresses) in the manufacturer's documents is always on a voluntary basis whenever possible. Whenever feasible, it is always possible to make use of the offerings and services without providing any personal data.
SAFETY INSTRUCTIONS 1
We draw your attention to the fact that data transmission over the Internet (e.g. when communicating by e-mail) may involve gaps in security. It is not possible to protect such data completely against access by third parties.
We hereby expressly prohibit the use of the contact data published as part of our duty to publish an imprint for the purpose of sending us any advertising or informational materials that we have not expressly requested.
1.5.1 Disclaimer
The manufacturer will not be liable for any damage of any kind by using its product, including, but not limited to direct, indirect or incidental and consequential damages.
This disclaimer does not apply in case the manufacturer has acted on purpose or with gross negligence. In the event any applicable law does not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to you, not be subject to some or all of the above disclaimer, exclusions or limitations.
Any product purchased from the manufacturer is warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale.
The manufacturer reserves the right to alter the content of its documents, including this disclaimer in any way, at any time, for any reason, without prior notification, and will not be liable in any way for possible consequences of such changes.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
7
Page 8
1 SAFETY INSTRUCTIONS
1.5.2 Product liability and warranty
The operator shall bear responsibility for the suitability of the device for the specific purpose. The manufacturer accepts no liability for the consequences of misuse by the operator. Improper installation and operation of the devices (systems) will cause the warranty to be void. The respective "Standard Terms and Conditions" which form the basis for the sales contract shall also apply.
1.5.3 Information concerning the documentation
To prevent any injury to the user or damage to the device it is essential that you read the information in this document and observe applicable national standards, safety requirements and accident prevention regulations.
If this document is not in your native language and if you have any problems understanding the text, we advise you to contact your local office for assistance. The manufacturer can not accept responsibility for any damage or injury caused by misunderstanding of the information in this document.
This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device. Special considerations and precautions are also described in the document, which appear in the form of underneath icons.
H250 M40
8
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 9
H250 M40
1.5.4 Warnings and symbols used
Safety warnings are indicated by the following symbols.
DANGER!
This information refers to the immediate danger when working with electricity.
DANGER!
This warning refers to the immediate danger of burns caused by heat or hot surfaces.
DANGER!
This warning refers to the immediate danger when using this device in a hazardous atmosphere.
DANGER!
These warnings must be observed without fail. Even partial disregard of this warning can lead to serious health problems and even death. There is also the risk of seriously damaging the device or parts of the operator's plant.
SAFETY INSTRUCTIONS 1
WARNING!
Disregarding this safety warning, even if only in part, poses the risk of serious health problems. There is also the risk of damaging the device or parts of the operator's plant.
CAUTION!
Disregarding these instructions can result in damage to the device or to parts of the operator's plant.
INFORMATION!
These instructions contain important information for the handling of the device.
LEGAL NOTICE!
This note contains information on statutory directives and standards.
HANDLING
HANDLING
HANDLINGHANDLING
This symbol designates all instructions for actions to be carried out by the operator in the specified sequence.
i RESULT
RESULT
RESULTRESULT This symbol refers to all important consequences of the previous actions.
1.6 Safety instructions for the operator
WARNING!
In general, devices from the manufacturer may only be installed, commissioned, operated and maintained by properly trained and authorized personnel. This document is provided to help you establish operating conditions, which will permit safe and efficient use of this device.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
9
Page 10
2 DEVICE DESCRIPTION
2.1 Scope of delivery
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
H250 M40
Figure 2-1: Scope of delivery
1 Measuring device in ordered version 2 For the ESK4T version - bar magnet 3 Documentation 4 Certificates, calibration report (supplied to order only) 5 Key
10
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 11
H250 M40
2.2 Device version
H250 with indicator M40
H250 with M40 indicator with display cut-out for ESK4-T
1. H250/RR/M40
Local indicator without auxiliary power
max. 2 limit switches, type NAMUR, NAMUR safety-oriented or transistor (3-wire)
Electrical signal output 4...20 mA, HART
Optionally intrinsically safe (Ex i) or in explosion-proof enclosure (Ex d)
DEVICE DESCRIPTION 2
®
or Fieldbus communication
2. H250/RR/M40
additional LCD, measured value and/or flow counter
2 configurable binary outputs, limit value or 10Hz pulse output
1 binary input, Start / Stop / Reset flow counter
®
2-wire current output 420 mA, HART
Optionally intrinsically safe [Ex i] or in explosion-proof enclosure [Ex d]
communication
The following designs are available as options:
H250 with indicator M40 as high-temperature version HT
H250 with indicator M40 with added impact and corrosion protection (special paint finish)
H250H for use in horizontal pipelines
H250U for use in vertical fall pipes
H250F featuring hygienic measuring tube design for Food & Pharma
H250C with PTFE / TFM liner for aggressive media
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
11
Page 12
2 DEVICE DESCRIPTION
2.2.1 Indicator versions
The M40 indicator can be fitted with various modules.
Basic version
H250 M40
1 Pointer module 2 Bolts for ESK4 attachment 3 Baseplate 4 Module profile 5 Pressure piece for ESK4 attachment 6 Housing cover locking device 7 Earth terminal external
Versions K1 / K2 and ESK4
1 Indicator with K2 contact module 2 Indicator with ESK4 current output 4...20 mA
12
Both versions can be combined with one another.
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 13
H250 M40
DEVICE DESCRIPTION 2
Version ESK4-T
1 ESK4 connection 2 Module cover 3 Display 4 Display module 5 Operating keys ^ 6 Connection binary input/outputs 7 Connection cable module
Version Fieldbus ESK4-FF / ESK4-PA
1 Basic module with electronic magnet sensors 2 Connection bus module 3 hardware settings
For more details see the supplementary instructions "H250 M40 Foundation Fieldbus"
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
13
Page 14
2 DEVICE DESCRIPTION
2.2.2 Float damping
Float damping is characterised by high standstill times and self-centering. The damping sleeve is made of high performance ceramic or PEEK, depending on the medium and the application. Float damping can also be retrofitted for the user (see Service).
Use of damping
Generally when CIV and DIV floats are used for gas measurement.
For TIV floats (H250/RR and H250/HC only) with an operating primary pressure:
Nominal size acc. to Operating primary pressure
EN 1092-1 ASME B16.5 [bar] [psig]
DN 50 ½" 0.3 4.4
DN25 1" 0.3 4.4 DN50 2" 0.2 2.9 DN80 3" 0.2 2.9
DN 100 4" 0.2 2.9
H250 M40
2.2.3 Pointer damping
In principle, the indicating element with its magnetic system contains indicator damping. An additional eddy current brake is advantageous in the event of fluctuating or pulsing flows. The magnets on the eddy current brake surround the pointer vane without touching it, damping its movement. The result is a much steadier pointer position, without distorting the measured value. A turnbuckle ensures a proper fit. The eddy current brake can be retrofitted during operation without recalibrating (see Service).
1 Eddy current brake 2 Pointer vane 3 Bracket 4 Pointer cylinder
14
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 15
H250 M40
2.3 Nameplate
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
DEVICE DESCRIPTION 2
Figure 2-2: Nameplate on the indicator
1 Device type 2 Manufacturer 3 Notified ATEX & PED body 4 Sizing data: temperature & pressure rating 5 PED data 6 Ex data 7 Electrical connection data 8 Internet site
Additional markings on the indicator
SN - serial number
SO - sales order / item
PA - Production order
Vx - product configurator code
AC - article code
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
15
Page 16
2 DEVICE DESCRIPTION
2.4 Description code
The description code* consists of the following elements:
1 Device type H250 - standard version H250H - horizontal flow direction H250U - flow direction from top to bottom
2 Materials / versions RR - Stainless Steel C - PTFE or PTFE/ceramics HC - Hastelloy Ti - Titanium F - aseptic version (food)
3 Heating jacket version B - with heating jacket
H250 M40
4 Series of indicators M40 - Indicator M40 M40S - Indicator with added corrosion protection M40R - Indicator in Stainless Steel housing
5 High temperature version HT - Version with HT extension
6 Electrical signal output ESK - Electrical signal output 4...20mA (ESK4)
- optionally available with counter, I/O module and display (ESK4-T) or
- Foundation Fieldbus (ESK4-FF) or
- Profibus PA (ESK4-PA)
7 Limit switches K1 - One limit switch K2 - Two limit switches
8 Explosion protection EX - Explosion-protected equipment
9 SIL Version SE - SIL compliant electronic signal output SK - SIL compliant limit switch
* positions which are not needed are omitted (no blank positions)
16
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 17
H250 M40
2.5 Electronics revision
The Electronic Revision (sticker on the base module ESK4) indicates the respective hardware/software status of the electronics. All add-on modules (ESK4-T, ESK4-FF and ESK4­PA) have an additional sticker indicating their respective firmware version.
DEVICE DESCRIPTION 2
Electronics
Explanations
revision
ER 1.1.x Basic version (cannot be combined with other indicator versions):
ER 2.0.x Functional add-on to ER 1.1.x:
ER 2.1.x Functional add-on to ER 2.0.x
ER 2.2.x (in preparation)
ESK4 / Current output 420mA with HART® communication; (ESK4 HART DD 01.01. AMS10x AMS11x ESK4 HART DD 01.01. PDM6.0 ESK4 HART DTM 1.0.3 FDT1.2)
can be combined with indicator version ESK4 FF / Foundation Fieldbus; (Firmware Version FF module from 1.0.2)
can be combined with indicator version ESK4-PA / Profibus PA; (Firmware Version PA module from 1.0.0) can be combined with indicator version ESK4-T / LCD, binary inputs/outputs; (Firmware Version T module from 1.1.0)
Functional add-on to ER 2.1.x (in preparation): Support of failure signal (low) according to NE43 for the ESK 4 current output module
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
17
Page 18
3 INSTALLATION
3.1 Notes on installation
INFORMATION!
Inspect the cartons carefully for damages or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer.
INFORMATION!
Do a check of the packing list to make sure that you have all the elements given in the order.
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
3.2 Storage
Store the device in a dry and dust-free location.
Avoid lasting direct exposure to the sun.
Store the device in its original packing.
The permissible storage temperature for standard devices is -40...+80°C / -40...+176°F.
H250 M40
18
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 19
H250 M40
3.3 Installation conditions
CAUTION!
When installing the device in the piping, the following points must be observed:
The variable area flowmeter must be installed vertically (measuring principle). Flow direction from bottom to top. For installation recommendations please refer also to VDI/VDE 3513 Sheet 3.
H250Hs are installed horizontally and H250U devices are installed vertically with the flow direction from top to bottom.
A straight unimpeded inlet run of 5x DN upstream of the device and a straight outlet run of 3x DN downstream of the device are recommended.
Screws, bolts and gaskets are to be provided by the customer and must be selected in accordance with the pressure rating of the connection or the operating pressure.
The inside diameter of the flange deviates from the standard dimensions. Flange seal standard DIN 2690 can be applied without any limitation.
Align the gaskets. Tighten the nuts with the tightening torques of the appropriate pressure rating.
For devices with PTFE liner or ceramic liner and PTFE raised faces, see chapter "Tightening torques".
Control devices are to be positioned downstream of the measuring device.
Shutoff devices are preferably to be positioned upstream of the measuring device.
Before connecting, blow or flush out the pipes leading to the device.
Pipes for gas flow need to be dried before the device is installed.
Use connectors suitable for the particular device version.
Align the pipes centrically with the connection bores on the measuring device so they are free of stresses.
If necessary, the piping has to be supported to reduce the vibrations transmitted to the measuring device.
Do not lay signal cables directly next to cables for the power supply.
INSTALLATION 3
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
19
Page 20
3 INSTALLATION
Minimum distance between the divices
If several instruments are installed side by side, a minimum distance a < 300mm between these divices is required.
H250 M40
Take special note of the installation position for the H250H with horizontal flow direction:
In order to comply with thermal parameters and measuring accuracy, H250H flowmeters for horizontal installation are to be installed in the pipeline so that the display is located on the side of the measuring tube. The maximum medium and ambient temperatures indicated as well as the measuring accuracy are based on lateral installation of the display.
20
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 21
H250 M40
3.3.1 Tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange threads with the following torques:
Nominal size according to Stud bolts Max. torque
EN 1092-1 ASME B 16.5 EN ASME EN 1092-1 ASME 150 lb
DN PN inch lb 150 lb 300 lb Nm ft*lbf Nm ft*lbf
15 40 ½“ 150/300 4x M12 4x ½" 4x ½" 9.8 7.1 5.2 3.8 25 40 1 150/300 4x M12 4x ½" 4x 5/8 21 15 10 7.2 50 40 2 150/300 4x M16 4x 5/8 8x 5/8 57 41 41 30 80 16 3 150/300 8x M16 4x 5/8 8x ¾" 47 34 70 51
100 16 4 150/300 8x M16 8x 5/8 8x ¾" 67 48 50 36
INSTALLATION 3
3.3.2 Magnetic filters
The use of magnetic filters is recommended when the medium contains particles which can be influenced magnetically. The magnetic filter is to be installed in the flow direction upstream of the flowmeter. Bar magnets are positioned helically in the filter to provide optimal efficiency at low pressure loss. All of the magnets are coated individually with PTFE to protect against corrosion. Material: 1.4571
Magnetic filters
1 Type F - fitting part with flange - overall length 100 mm 2 Type FS - fitting part without flange - overall length 50 mm
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
21
Page 22
3 INSTALLATION
3.3.3 Heat insulation
CAUTION!
The indicator housing may not be heat-insulated. The heat insulation 3 may only reach as far as the housing fastening 4.
1 Standard indicator M40 2 Indicator with HT extension
H250 M40
CAUTION!
The heat insulation 1 may only reach to the rear of the housing 2. The area of the cable entries 3 must be freely accessible.
Figure 3-1: Insulation - cross section
22
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 23
H250 M40
4.1 Security information
DANGER!
All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate!
DANGER!
Observe the national regulations for electrical installations!
DANGER!
For devices used in hazardous areas, additional safety notes apply; please refer to the Ex documentation.
WARNING!
Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists.
ELECTRICAL CONNECTIONS 4
INFORMATION!
Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
23
Page 24
4 ELECTRICAL CONNECTIONS
4.2 Electrical connection indicator M40
4.2.1 Indicator M40 - limit switches
The M40 indicator can be fitted with a maximum of two limit switches. The limit switch works as a slot sensor which is inductively activated via the semi-circular metal vane of the pointer. The switching points are set using the contact pointer. The position of the contact pointer is displayed on the scale.
Limit switch module
H250 M40
1 Min. contact 2 Max. contact 3 Locking screw 4 Maximum pointer 5 Connection terminal
The connecting terminals have a pluggable design and can be removed in order to connect the cables. The built-in limit switch types are shown on the indicator.
Electrical connection of the limit switches
Contact MIN MAX
Terminal no. 1 2 3 4 5 6
Connection 2-wire NAMUR - + - +
Connection 3-wire + - + -
Connection Reed SPST + - + -
24
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 25
H250 M40
ELECTRICAL CONNECTIONS 4
Limit switch connection terminals
1 2-wire limit switch NAMUR 2 3-wire limit switch 3 Limit switch Reed SPST 4 Terminal connection min contact 5 Terminal connection max contact 6 3-wire load 7 NAMUR isolated switching amplifier 8 3-wire power supply
Limit setting
Figure 4-1: Limit switch settings
1 Contact pointer MAX 2 Contact pointer MIN 3 Locking screw
Setting is carried out directly via contact pointers 1 and 2:
Slide the scale away
Loosen the locking screw 3 slightly
Slide the scale back to the latching point
Set contact pointers 1 and 2 to the desired switching point
After setting has been carried out: Fix the contact pointers with the locking screw 3.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
25
Page 26
4 ELECTRICAL CONNECTIONS
Switch contact definition
1 MIN contact 2 MAX contact 3 Pointer vane with switching vane
If the measuring pointer vane goes into the slot an alarm is triggered. If the pointer vane is outside the slot sensor, a wire break in a NAMUR contact also triggers the alarm.
The 3-wire limit switch does not have any wire break detection.
H250 M40
Definition MinMin - MaxMax
1 MIN 2 contact or MAX 1 contact 2 MIN 1 contact or MAX 2 contact
Current consumption in the position shown:
Contact Type current
MIN 1 NAMUR 1 mA MIN 2 NAMUR 1 mA MAX 1 NAMUR 3 mA MAX 2 NAMUR 3 mA
26
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 27
H250 M40
4.2.2 Current output ESK4
The connecting terminals of the ESK4 have a pluggable design and can be removed in order to connect the cables.
ESK4 connection
1 ESK4A current output 2 Power supply 14...30 VDC 3 Measuring signal 4...20 mA
4 External load, HART
®
communication
ELECTRICAL CONNECTIONS 4
Power supply M40 with electrical isolation
Power supply M40 with electrical isolation
Power supply M40 with electrical isolationPower supply M40 with electrical isolation
Wiring must be planned with great care when it comes to connecting other devices such as evaluation units or process control. Internal connections in these devices (e.g. GND with PE, mass loops) may lead to non-permitted voltage potentials which could impair the function of the converter itself or that of a device connected to it. In such cases a protected extra-low voltage (PELV) is recommended.
1 Terminal connection 2 Converter supply isolator with electrical isolation 3 Power supply (see supply isolator information) 4 Measuring signal 4...20 mA
5 External load, HART
®
communication
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
27
Page 28
4 ELECTRICAL CONNECTIONS
Power supply
Power supply
Power supplyPower supply
INFORMATION!
The feed voltage must be between 14 VDC and 30 VDC. This is based on the total resistance of the measuring loop. To determine this, add up the resistances of each component in the measuring loop (not including the level meter).
The required supply voltage can be calculated using the formula below:
.
= R
U
ext.
where U
ext.
R
= the total measuring loop resistance.
L
INFORMATION!
The power supply has to be able to supply a minimum of 30 mA.
24 mA + 14 V
L
= the minimum supply voltage and
H250 M40
HART
HART® communication
HARTHART
When HART® communication is carried out with the ESK4, the analogue measured data transmission (4...20 mA) is not impaired in any way.
Exception for multidrop operation. In multidrop mode, a maximum of 15 devices with HART function can be operated in parallel, whereby their current outputs are switched inactive (I approx. 4.5 mA per device).
®
28
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 29
H250 M40
ELECTRICAL CONNECTIONS 4
Load for HART® communication
INFORMATION!
®
A load of at least 230 Ohm is required for HART
The maximum load resistance is calculated as follows:
DANGER!
Use a twisted two-core cable to prevent electrical interference from impeding the DC output signal. In some cases a shielded cable may be necessary. The cable shield may only be earthed (grounded) at one place (on the power supply unit).
communication.
Configuration
The ESK can be configured via HART® communication. DD (Device Descriptions) for AMS 10x,
AMS 11x and PDM 6.0 as well as a DTM (Device Type Manager) for PACTware 3.0.2.28(3.0 SP5),
3.6.0.3(3.6 SP2) and 4.0.0.6 are available for configuration. They can be downloaded free of
charge from our website.
®
The current flow rate can be transmitted using the integrated HART counter can be configured. Two limit values can be monitored. The limit values are assigned either to flow values or to the counter overflow.
communication. A flow
Self monitoring - Diagnostics
During both start-up and operation, a wide variety of diagnostic functions are performed cyclically in the ESK4, in order to guarantee function reliability. When an error is detected, a failure signal (high) is activated (current > 21 mA, typically 22 mA) via the analog output. More
®
detailed information can also be obtained via HART in the event of information and warnings.
(CMD#48). The failure signal is not activated
Diagnostic functions (Monitoring):
Plausibility of FRAM data
Plausibility of ROM data
Working range of internal reference voltages
Signal detection of the measuring range of the internal sensors
Temperature compensation of the internal sensors
Calibration based on the application
Plausibility of counting value
Plausibility of physical unit, system and selected unit
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
29
Page 30
4 ELECTRICAL CONNECTIONS
4.2.3 Binary inputs/outputs ESK4-T
Once the housing cover has been unscrewed, the scale can be removed. The connection terminals feature a pluggable design and can be removed to connect the cables.
H250 M40
1 Binary output 1 2 Power supply / durrent output ESK4 3 Binary output 2 4 Binary input
The binary inputs/outputs are electrically isolated from each other and from the ESK4 current output.
INFORMATION!
The binary inputs/outputs can only be operated if the power supply is applied to ESK4 terminals 11+ and 12-. The binary inputs/outputs come inactive by default and must thus be activated prior to first use (see Section 6.6 Menu ESK4-T)
Connection binary outputs
Connection binary outputs
Connection binary outputsConnection binary outputs In accordance with the desired signal transmission, select one of the following connection types for binary outputs B1 and B2:
NAMUR (DC interface in accordance with EN 60947-5-6)
Transistor output (passive, open collector)
Binary output B1 B2
30
Terminal no. 1 2 3 4 5 6
Connection NAMUR + - + -
Connection transistor output + B
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
OC
+ B
OC
Page 31
H250 M40
1 NAMUR terminal connection 2 Isolated switching amplifier 3 Terminal connection transistor output 4 Power supply U
5 Load R
L
ext.
Value range NAMUR
ELECTRICAL CONNECTIONS 4
Normally closed Normally open
Switching value reached 1 mA > 3mA Switching value not reached > 3mA 1 mA
Range of values applies only when connected to a switching amplifier with the following reference values:
Open-circuit voltage Uo = 8,2 V DC
Internal resistance Ri = 1kOhm
Range of values transistor output
Signal voltages UL [V] UH [V]
lower Limit upper Limit lower Limit upper Limit
over load R
Signal currents IL [mA] IH [mA]
Category 2 0 2 20 110
L
To safeguard the range of values, a load RL between 250 Ohm and 1kOhm is recommended for the passive transistor output with a nominal voltage of 24V DC.
0 2 16 30
lower Limit upper Limit lower Limit upper Limit
If other loads are used, caution is advised as the range of values of the signal voltages then no longer corresponds to the range of values for the inputs of process control systems and controls (DIN IEC 946).
CAUTION!
The upper limit of the signal current may not be exceeded as this may damage the transistor output.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
31
Page 32
4 ELECTRICAL CONNECTIONS
Pulse output mode
INFORMATION!
The binary outputs can also be operated as pulse outputs.
Figure 4-2: Data pulse output
1 f
= 10 Hz
max
2 t
on
3 t
off
The pulse width ton can be configured from 50...500 ms in the indicator menu.
H250 M40
Connection binary input
The binary input can be used to control the internal flow counter (start/stop/reset).
Reset input
1 Function active HI 2 Function active LO
As standard, the binary input is inactive and can be activated in menu point 3.6.
Range of values
Input voltage UL [V] UH [V]
lower Limit upper Limit lower Limit upper Limit
Terminal (7) (8) 0 2 16 30
32
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 33
H250 M40
4.2.4 Connection Harting HAN® 7D
1 Terminal pin assignment HAN® 7D - View of plug connection
ELECTRICAL CONNECTIONS 4
1 Pin No. HAN® 7D K1/K2: NAMUR
contacts
R1/R2 Reed contacts
ESK4 Terminal no.
NAMUR Reed
1 2 NAMUR MIN (-) 2 Reed MIN - 1 1 2 2 NAMUR MIN (+) 2 Reed MIN - 2 3 3 3 NAMUR MAX (-) 3 Reed MAX - 4 4 4 3 NAMUR MAX (+) 3 Reed MAX - 5 6 5 - - 4 4...20mA (+) 11 6 - - 4 4...20mA (-) 12
7 - - -
8 - - -
Combinations of K1 / K2 and ESK4 are possible.
1 Pin No. HAN® 7D ESK4-T Terminal no.
1 5 Binary output B1 open coll. (+) 1 2 5 Binary output B1 open coll. (-) 3 3 7 Binary output B2 open coll. (+) 4 4 7 Binary output B2 open coll. (-) 6 5 6 4...20mA (+) 11 6 6 4...20mA (-) 12
7 -
8 -
INFORMATION!
No Harting connection for module ESK4-FF/PA is provided.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
33
Page 34
4 ELECTRICAL CONNECTIONS
4.3 Grounding connections
1 Grounding connection in the indicator 2 External grounding connection
DANGER!
The grounding wire may not transfer any interference voltage. Do not use this grounding cable to ground any other electrical devices.
H250 M40
4.4 Protection category
The measuring device meets all requirements of protection category IP66/68.
DANGER!
After all servicing and maintenance work on the measuring device, the specified protection class must be ensured again.
Therefore it is essential to observe the following points:
Use only original gaskets. They must be clean and free of any damage. Defective gaskets must
be replaced.
The electrical cables used must be undamaged and must comply with regulations.
The cables must be laid with a loop 3 upstream of the measuring device to prevent water
from getting into the housing.
The cable feedthroughs 2 must be tightened.
Close the unused cable feedthroughs using blanking plugs 1.
34
1 Use blanking plugs if no cable is routed through 2 Tighten cable feedthrough firmly 3 Lay the cable in a loop
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 35
H250 M40
5.1 Standard device
CAUTION!
When starting up the device, the following points must be observed:
Compare the actual operating pressure and the product temperature of the system with the specifications on the nameplate (PS and TS). These specifications may not be exceeded.
Make sure materials are compatible.
Slowly open the shut-off valve.
When measuring liquids, vent the pipes carefully.
When measuring gases, increase pressure slowly.
Avoid float impact (e.g. caused by solenoid valves), as this is likely to damage the measuring unit or float.
A minimum operating pressure (primary pressure) is necessary to operate the device:
Medium Pressure loss : operating pressure
START-UP 5
Liquids 1 : 2
Gases without float damping 1 : 5
Gases with float damping 1 : 2
5.2 Indicator ESK4-T
INFORMATION!
The device is always preset for the user and his application.
Start After the device is switched on, the display shows
"INITIALISING"
Firmwareversion IO-Modul
The device first performs a self-test. Here, all of the parameters preset for the customer are analysed and checked for plausibility. The device then switches to measuring mode and indicates the current measured value.
Operation
INFORMATION!
The device is low-maintenance Comply with the application limits with regard to temperature of the medium and ambient temperature.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
35
Page 36
6 OPERATION
6.1 ESK4- Loop Check Modus
The ESK4 is equipped with a loop check function, enabling a simple test of the entire 4…20mA
current loop.
It is activated and operated using a microswitch 1.
CAUTION!
When activating the loop check mode, ensure that no alarms are unintentionally triggered in the downstream system components.
H250 M40
36
Press and hold the microswitch 1 for more than 6 seconds to activate loop check mode 3.
The current output jumps to constant 4mA.
Change the current output from constant 4 to constant 20mA as often as you like by briefly
pressing it (less than 6 seconds) to check the function of the measuring circuit.
Exit loop check mode by holding down the microswitch (longer than 6 seconds). The current
output jumps back to measuring mode 2.
INFORMATION!
If the microswitch has not been pressed for longer than 60 seconds, the ESK4 automatically reverts to measuring mode 2.
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 37
H250 M40
6.2 Operating elements ESK4-T
OPERATION 6
The device is operated with the cover on the front open, using the mechanical keys cover closed using a bar magnet
bar magnet.
bar magnetbar magnet
keys, or with the
keyskeys
CAUTION!
The switching point of the magnetic sensors is directly at the height of the corresponding circle (see figure). Only touch the circle vertically and from the front using the bar magnet. Touching it from the side may cause a malfunction.
Figure 6-1: Display and operating elements
1 Enter button (circuit for bar magnet) 2 Up button (circuit for bar magnet) 3 Right button (circuit for bar magnet)
The mechanical keys and keys for the bar magnet have the same functionality. In this documentation the keys are represented as symbols to describe the operating functions:
Button Symbol
1 Enter ^^^^ 2 up 3 right
Table 6-1: ESK4-T operation keys
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
37
Page 38
6 OPERATION
6.3 Basic principles of operation ESK4-T
6.3.1 Functional description of the buttons
H250 M40
^^^^
Table 6-2: Description of the operating keys
Switch from measuring mode to menu mode
Switch to one menu level lower
Open menu item and activate change mode
Confirm query whether data should be accepted.
In change mode:
In change mode: Move the input cursor one position to the right. After the last digit the
In change mode:In change mode: input cursor jumps back to the beginning.
Change between the menu items within a menu level
In measuring mode:
In measuring mode: Switch between measured values and error messages
In measuring mode:In measuring mode:
In change mode:
In change mode: Changing parameters or settings. Run through the available characters
In change mode:In change mode: (including decimal point.
In measuring mode:
In measuring mode: Switch between measured value display and error messages /
In measuring mode:In measuring mode: warnings
Switch to one menu level higher
Return to measuring mode with a query whether the data should be accepted
Cancel the query, whether data should be accepted.
6.3.2 Navigation within the menu structure
Navigate through the menu using the →, ↑ and ^^^^ keys. Pressing the → key takes you one menu lower. Use the key to go one menu point higher (e.g. from 1 to 2). Pressing the ^^^^ key takes you one menu higher.
If you are already at the lowest level (function level), you can use the button to go to the change mode, which can be used to set data and values.
If you are located at the first level (main menu), you can use the ^^^^ key to exit the menu mode and return to the measuring mode.
If settings were changed, the query whether they should be saved will appear. Confirm this query with the key or cancel it with the ^^^^ key.
measuring
mode
Table 6-3: Navigation menu structure
Main menu↑→ Submenu↑→ Function↑→ Edit
^^^^ ^^^^ ^^^^ ^^^^
→ ↑ ^^^^
38
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 39
H250 M40
6.3.3 Changing the settings in the menu
Starting operation
Operation is started using the key
If a control lockout is set, the set code → → → ^ ^ ^ ↑ ↑ ↑ must be entered. The code can be set in Menu 3.13. The code shown here is set at the factory but not activated. If no key is pressed within 5 seconds or if an incorrect code is entered, a warning message is displayed and the display returns to measuring mode.
Exiting operator input
Operation is exited by pressing the ^^^^ key several times.
If data have been changed:
OPERATION 6
Save Yes Changes saved. An update occurs and the indicator jumps back to measuring
Save No ^^^^ Changes are discarded and the indicator jumps back to measuring operation.
mode.
CAUTION!
Each time parameters or settings are changed, the measuring device carries out an internal plausibility check. If inplausible entries were made, a warning message is shown. If this warning is confirmed with the ^^^^ key, the display returns to the respective menu point without saving the relevant change. A new entry can now be made.
Example: Changing the flow unit from m3/h to l/h
Screen Screen
Example: 7.2
m
1x Fct. 1111
OPERATION
2x Fct. 3333
INSTALLATION
1x Fct 3.1111
LANGUAGE
10x Fct 3.11
FS&UNIT
3
/h
11
1111
1x Fct. 3.11.1111
1x 10.0000
4x 10000
3x ^^^^ 7200
FLOW RATE
3
/h
m
l/hr
Confirm with
deny ^
l/hr
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
39
Page 40
6 OPERATION
6.4 Overview of the units ESK4-T
H250 M40
Measured
Units
variables
volume
mass g/s g/min g/h -
Volume Totalizer
mass totalizer kg g t lb
Temperature °C °F K
m3/s m3/min m3/h m3/d
L/s L/min L/h -
ft3/s ft3/min ft3/h ft3/d
gal/s gal/min gal/h gal/d
bbl/s bbl/min bbl/h bbl/d
ImpGal/s ImpGal/min ImpGal/h ImpGal/d
kg/s kg/min kg/h kg/d
- t/min t/h t/d
lb/s lb/min lb/h lb/d
- STon/min STon/h STon/d
- - LTon/h LTon/d
3
m
ImpGal gallon bbl bbl (liq)
STon LTon
l hl
3
ft
6.5 Error messages ESK4-T
Error messages and warnings are indicated by one of the following symbols in the bottom left corner of the display. The ^^^^ key switches from the measured value display to the display of current errors / warnings. The table below contains a description of possible error messages.
Symbols
1 Error 2 Warning - out of specification 3 Function check 4 Maintenance necessary.
40
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 41
H250 M40
OPERATION 6
Error Message Description Category Remedy
NOT LINEARIZED Linearization faulty or
not activated = measuring error
NEW LINEARI. TABLE BAD
Faulty or missing data in the linearization table = measuring error
LINEARIZATIO UNDER CONFIG
The device is in linearization mode = measuring error
UNIT SYSTEM CONFLICT
The unit for the linearization flow is incompatible with the selected flow type (mass/volume)
TOO FEW ENTRIES
The linearization table has too few data points
NOT MONOTONOUS
The sequence of the linearization values is not strictly monotonic increasing
FIRST NOT 0 % The first flow value if
the linearization table is not 0%
LAST NOT 100 % The last flow value if
the linearization table is not 100%
NO ZERO CAL OF AO
The current output zero point 4.00mA is not calibrated. = possible measuring error in process control.
NO F.SC. CAL OF AO
The current output 100% = 20.00mA is not calibrated. = possible measuring error in process control.
NO TEMP. COMPENSATION
The sensor temperature compensation of the device is faulty or was not carried out. = possible measuring error
OUTPUT NOT LINEARIZED
Linearization is not activated = measuring error
COUNTER LOST Totalizer value was
reset by error/overflow
Maintenance Activate linearization or carry it out again
(HART
®
communication and linearization
software are required; the original
Warning
calibration values must be known), or send the device back to the manufacturer for linearization.
Warning Complete the linearization and activate it
(HART
®
communication and linearization software are required), or send the device back to the manufacturer for linearization.
Warning Correct error, carry out linearization again if
necessary (HART
®
communication and linearization software are required), or send the device back to the manufacturer for linearization.
Warning Carry out linearization at at least 5 points
(HART
®
communication and linearization software are required), or send the device back to the manufacturer for linearization.
Warning Check linearization and/or carry it out again
(HART
®
communication and linearization software are required), or send the device back to the manufacturer for linearization.
Maintenance Perform calibration using ammeter and
menu 3.10 or using standard HART
®
tools/process control system and poss. external ammeter. Caution: during calibration, switch the measuring point to manual control.
Maintenance Perform calibration using ammeter and
menu item 3.11 or using standard HART
®
tools and external ammeter if necessary. Caution: during calibration, switch the measuring point to manual control.
Maintenance The device, together with an indication of the
error, must be sent back to the manufacturer for checking.
Maintenance Activate linearization or carry it out again
(HART
®
communication and linearization software are required; the original calibration values must be known), or send the device back to the manufacturer for linearization.
Warning Because the reset time is not known:
Controlled reset of the counter using menu item 1.6.1 or using HART
®
tools/process
control system.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
41
Page 42
6 OPERATION
Error Message Description Category Remedy
H250 M40
FRAM WRITE FAULT
ROM/FLASH ERROR
RESTART OF DEVICE
MULTIDROP MODE
CRYSTAL OSC FAULT
REF VOLTAGE FAULT
SENSOR A FAULT Internal error in
SENSOR B FAULT
MEMORY CORRUPTION
AO FIXED The current output is
AO SATURATED Current output
ERROR TIMEOUT Data not transferred,
WARNING TIMEOUT
Internal communication error
Memory error detected during self­test.
A device restart has taken place
The HART® multidrop mode is activated. The current output is set to a fixed value of
4.5 mA.
Internal error in device
Internal error in device
device
Internal error in device
Internal memory error, caused by a hardware or software problem
set to a fixed value.
saturated
or transferred incorrectly from the ESK to the counter module
Error Check whether the display is plugged in
Error Restart device. If the error occurs again:
information The device has been restarted using menu
information
Error The device must be sent back to the
Error Restart the device; if the error occurs again
information The current output is fixed and does not
information The current output is saturated at 20.4 or
Error Confirm menu point 1.6.3 WRITE INFO I/O.
Warning
correctly and restart the device. If the error occurs again: send the device back to the manufacturer with an indication of the error.
send the device back to the manufacturer with an indication of the error.
item 1.6.2 since the last time the error messages were reset.
The HART® multidrop mode is activated with selection of a polling address not equal to 0 using menu item 3.7. . The polling address 0 reactivates the analog output.
manufacturer with an indication of the error.
the device must be sent back to the manufacturer with an indication of the error.
reflect the measured value. This is the case in multidrop mode, with current output
test/calibration using the menu or HART
22.0 mA (depending on whether the alarm current is activated or deactivated in menu item 3.10), and is no longer coupled with the measured value.
®
.
42
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 43
H250 M40
6.6 Menu indicator ESK4-T
6.6.1 factory settings
Menu Function Adjustment
1.1.1 OUTPUT B1 INACTIVE
1.2.1 OUTPUT B2 INACTIVE
1.3.1 Pulse width 100 ms100ms
1.3.2 Pulses / unit 1,000000 Pulses / L
1.4.1 Screen FLOW RATE
1.4.2 DISPLAY ROTATION 0°
1.5 TIME CONST. 001.0 s
1.6.1 RESET COUNT. NO
1.6.2 RESET ERROR NO
1.6.3 WRITE INFO I/O NO
3.1 LANGUAGE ENGLISH
3.2 FUNCTION B1 INACTIVE
3.3 CONTACT B1 NO CONTACT
3.4 FUNCTION B2 INACTIVE
3.5 CONTACT B2 NO CONTACT
3.6 FUNCTION B3 INACTIVE
3.7 MULTIDROP POLLING ADD: 00
3.8 4mA CALIBR. 4.000 mA
3.9 20mA CALIBR. 20.000 mA
3.10 ALARM CURRENT ALARM HIGH
3.11.1 FS&UNIT l/hr
3.11.2 COUNTER l
3.12 Low flow cutoff treshold 4% ON 6% OFF
3.13 ENTRY CODE OFF
3.14 BASIC SETTING NO
OPERATION 6
INFORMATION!
The measuring device has been preset at the factory in accordance with the customer order. Therefore subsequent configuration via the menu is only necessary if the intended use of the device is changed.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
43
Page 44
6 OPERATION
6.6.2 Menu structure
Main menu Submenu 1 Submenu 2
H250 M40
1111 OPERATION 1.1 OUTPUT B1 1.1.1 INACTIVE, FLW VAL B1. CNT VAL B1, PULSE
1.2 OUTPUT B2 1.2.1 INACTIVE, FLW VAL B2. CNT VAL B2, PULSE
1.3 PULSEOUTP. 1.3.1 PULSEWDTH
1.4 DISPLAY 1.4.1 FLOW, COUNTER; FLOW&CNT; FLOW&CNT 1,
1.5 TIME CONST. -
1.6 RESET 1.6.1 COUNTER
2222 TEST & INFO 2.1 4-20mA OUTPUT 2.1.1 NORMAL OP
2.2 OUTPUT B1 2.2.1 NORMAL OP
2.3 OUTPUT B2 2.3.1 NORMAL OP
2.4 INPUT B3 ACTIV HI INPUT; ACTIV LO, ON, OFF
2.5 DEVICE ID 2.5.1 ELEC. REV.
2.6 SOFT.VERSION 2.6.1 FW. ESK4
2.7 TAG No.
WDTH
1.1.2 HYST. B1, PULSE/UNIT
WDTH
1.2.2 HYST. B2, PULSE/UNIT
1.3.2 PULSE/UNIT
PERCENT
1.4.2 ROTATION
1.6.2 ERROR
1.6.3 WRITE INFO IO
2.1.2 4.0 mA
2.1.3 5.6 mA
2.1.4 7.2 mA
2.1.5 8.8 mA
2.1.6 10.4 mA
2.1.7 12.0 mA
2.1.8 13.6 mA
2.1.9 15.2 mA
2.1.10 16.8 mA
2.1.11 18.4 mA
2.1.12 20.0 mA
2.1.13 21.6 mA
2.2.2 OPEN
2.2.3 CLOSED
2.3.2 OPEN
2.3.3 CLOSED
2.5.2 SN ESK4
2.5.3 PA ORDER
2.5.4 DEVICE SN.
2.6.2 FW. ESK4 I/O
44
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 45
H250 M40
OPERATION 6
Main menu Submenu 1 Submenu 2
3333 INSTALLATION
3.1 LANGUAGE 3.1.1 English
3.1.2 Deutsch
3.1.3 Francais
3.1.4 Italiano
3.1.5 Espanol
3.1.6 Cesky
3.1.7 Polski
3.1.8 Nederlands
3.1.9 Dansk
3.2 FUNCTION B1 INACTIV, SWITCHPOINT, CONTERLIM., PULSEOUTP.
3.3 CONTACT B1 NO contact, NC contact,
3.4 FUNCTION B2 INACTIV, SWITCHPOINT, CONTERLIM., PULSEOUTP.
3.5 CONTACT B2 NO contact, NC contact,
3.6 FUNCTION B3 INACTIV, ACTIV HI, ACTIV LO, STARTH STOPL, STARTL STOPH
3.7 MULTIDROP POLLING ADR.
3.8 4mA CALIBR. 4.000 mA
3.9 20mA CALIBR. 20.000 mA
3.10 ALARM CURR. OFF, ALARM HIGH, ALARM LOW
3.11 FS&UNIT 3.11.1 FLOW
3.11.2 COUNTER
3.12 LFC 3.12.1 CONTROL ON, OFF
3.12.2 LFC ON VAL
3.12.3 LFC OFF VAL
3.13 ENTRY CODE OFF, ON
3.14 BASIC SETTING SET ALL NO, SET ALL YES
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
45
Page 46
6 OPERATION
6.6.3 Menu explanations
1 Operation
H250 M40
Label Level Select /
Explanation
Input
OUTPUT B1 1.1 Output B1 is a binary switching output. Under Fct. 3.2, one of the
One of the following functions can be selected as contact type under Fct. 3.3:
NO contact 1 / NC contact 2
1.1.1 INACTIVE
FLOW.VAL B1 Flow value switching point
COUNTER.VAL B1 Switching point totaliser value
Pulse width Pulse value (pulse/unit)
following functions can be attributed to this output INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.
Value range: 0.0 measuring range end value The switching point is entered in flow units. If the current flow value exceeds this predetermined switching point, output B1 changes its binary state 3. Under Fct. 1.1.2 a hysteresis can also be specified.
Value range 0.0 counter limit The switching point is entered in volume or mass units. If the the current counter value exceeds this predetermined switching point, output B1 changes its binary state 3. There is no hysteresis setting for the counter value switching point.
The value is now displayed here. Configuration takes place under Fct. 1.3.1 (pulse width) and Fct. 1.3.2 pulse/unit).
46
1.1.2 HYST.B1 Hysteresis for the flow value switching point
Value range 0.0 switching point If the current flow value exceeds the predetermined switching point from Fct. 1.1.1, output B1 changes its binary state 3. In order for output B1 to change its binary state back to the initial setting, the switching point made smaller by the hysteresis must be undershot 4.
Example:
Example:
Example:Example: Under 1.1.1, a switching point of 200 L/h is set. The possible value range for the hysteresis is then 0.0 200 L/h. With a hysteresis value of 0, the switching point has no hysteresis (3=4). If a hysteresis value of 20 L/h is entered, output B1 changes its binary state to the initial setting if 180 L/h is undershot 4.
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 47
H250 M40
OPERATION 6
Label Level Select /
Explanation
Input
OUTPUT B2 1.2 Output B2 is a binary switching output. Under Fct. 3.4, one of the
One of the following functions can be selected as contact type under Fct. 3.5:
NO contact 1 / NC contact 2
1.2.1 INACTIVE
FLOW.VAL B2 see FL.R. VALUE B1
COUNTER.VAL B2 See COUNTER.VAL B1
PULSE WDTH B2 see PULSE WDTH Fct. 1.1.1
1.2.2 HYST.B2 See HYST. B1
PULSEOUTP. 1.3
1.3.1 Pulse width
50ms Ti = 50 ms; f
4ms Ti = 100 ms; f
200ms Ti = 200 ms; f
500ms Ti = 500 ms; f
1.3.2 PULSE/UNIT
0.001 ... 1000
following functions can be attributed to this output INACTIVE, SWITCHPOINT, COUNTERLIM. or PULSEOUTP.
Under Fct. 1.2.2 a hysteresis can also be specified.
Configuration takes place under Fct. 1.3.1 (pulse width) and Fct. 1.3.2 (pulse/unit).
= 10 Hz
max. pulse / h = 36000
max. pulse / h = 18000
max. pulse / h = 9000
max. pulse / h = 3600
Pulses per volume or mass unit, which can be output via one of the binary outputs. The maximum frequency for the pulse output (see Fct 1.3.1) cannot be exceeded, even at maximum flow rate (final value).
max
max
max
max
= 5 Hz
= 2,5 Hz
= 1 Hz
Example:
Example:
Example:Example: Final value Qmax = 1200 l/h; volume unit counter = Litre; pulse duration = 100 ms; If the factor 1 is entered, 1 pulse/litre = 1200 pulses are generated in one hour at maximum flow rate. Number of max. permissible pulses:
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
47
Page 48
6 OPERATION
H250 M40
Label Level Select /
Explanation
Input
DISPLAY 1.4 Different measured values can be selected for permanent or
1.4.1 FLOW RATE permanent display of flow in flow units
COUNTER permanent display totalizer
FLOW&COUNT alternating display flow value in flow units and totalizer
FLOW&COUNT 1 simultaneous display of flow value and totaliser
PERCENT permanent display flow value in percent
1.4.2 0° The reading is not rotated. 90° The reading on the display is rotated by 90°. 180° The reading on the display is rotated by 180°. 270° The reading on the display is rotated by 270°.
TIME CONST. 1.5 0.0...20.0s Given in seconds
CONFIG/RESET 1.6 Local resetting of totalizer and acknowledgement of warnings.
alternating display. The display can be rotated.
The output variables (value of the current loop and indicated flow value) follow the current process by the value set here (in seconds) with a time delay.
Note:
Note:
Note:Note: If the current flow is polled via HART
measured value is also played back with a delay.
There is always a safety prompt (yes/no), to avoid accidental resetting.
®
communication, the transferred
Note:
Note:
Note:Note: External resetting of the totalizer can be installed with the binary input B3.
1.6.1 COUNTER Confirming with YES resets the value of the totalizer to 0.0.
1.6.2 ERROR Confirming with yes acknowledges all existing warnings.
Note:
Note:
Note:Note: Confirming with YES acknowledges existing errors and warnings.
1.6.3 WRITE INFO IO As a general rule, data is transferred from the counter module to the
ESK4 and vice versa when the device starts. To make sure, selecting this menu point when confirming with YES can perform another data transmission.
48
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 49
H250 M40
2 TEST & INFO
OPERATION 6
Label Level Select /
Explanation
Input
4-20mA OUT 2.1 Testing the current loop by setting various current values
Note:
Note:
Note:Note: The simulation is not available in HART
Attention:
Attention:
Attention:Attention: During the test, the value of the current loop does not follow the current process.
2.1.1 NORMAL OP The value of the current loop follows the current process.
2.1.2 4.0 mA The value of the current loop no longer follows the current process.
2.1.3 5.6 mA
2.1.4 7.2 mA
2.1.5 8.8 mA
2.1.6 10.4mA
2.1.7 12.0 mA
2.1.8 21.6 mA
2.1.9 15.2 mA
2.1.10 16.8 mA
2.1.11 18.4 mA
2.1.12 20.0 mA
2.1.13 21.6 mA
OUTPUT B1 2.2 Testing the binary switching output B1 by changing its binary status.
It is set to the selected current value.
®
multi-drop mode (see Fct. 3.7).
Attention:
Attention:
Attention:Attention: During the test, the binary status does not correspond to the current process.
2.2.1 NORMAL OP The binary status of the switching output corresponds to the current
2.2.2 OPEN The binary status of the switching output no longer corresponds to the
2.2.3 CLOSED
OUTPUT B2 2.3 see test OUTPUT B1
2.3.1 NORMAL OP
2.3.2 OPEN
2.3.3 CLOSED
INPUT B3 2.4 ACTIVE HI
ACTIVE LO ON OFF
DEVICE ID 2.5 Information for device identification
2.5.1 ELEC. REV. Electronics revision
2.5.2 SN ESK4 Serial number ESK4
2.5.3 PA ORDER Production order for the complete flowmeter
2.5.4 DEVICE SN. Serial number for the complete flowmeter
process.
current process. The selected status is tested.
The current binary status of the binary input B3 is displayed. External switching of the voltage applied to input B3 activates a change of the displayed binary status.
Note:
Note:
Note:Note: A reaction to the switching of the voltage applied to input B3 only occurs when the function of B3 is activated (see Fct. 3.6).
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
49
Page 50
6 OPERATION
H250 M40
Label Level Select /
Explanation
Input
SOFT.VERSION 2.6 Information about software revision status
2.6.1 FW. ESK4 Software revision status for device insert ESK4-Basic
2.6.2 FW. ESK4 I/O Software revision status for optional add-on module ESK4-T
TAG NO. 2.7 xxxxxxxx Measuring point identifier
Alphanumerical characters. Up to eight digits possible
50
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 51
H250 M40
3 INSTALLATION
OPERATION 6
Label Level Select /
Explanation
Input
LANGUAGE 3.1 Language selection for the menu texts displayed
3.1.1 ENGLISH The menu texts are displayed in the selected language.
3.1.2 DEUTSCH
3.1.3 FRANCAIS
3.1.4 ITALIANO
3.1.5 ESPANOL
3.1.6 CESKY
3.1.7 POLSKI
3.1.8 NEDERLANDS
3.1.9 DANSK
FUNCTION B1 3.2 INACTIVE The binary switching output B1 has no function.
SWITCHING POINT
COUNTERLIM. The binary switching point B1 functions as a limit switch depending on
The binary switching point B1 functions as a limit switch depending on the current flow value. Configuration of the switching point takes place under Fct. 1.1.1 (FLOW. VAL B1).
the current counter value. Configuration of the switching point takes place under Fct. 1.1.1 (ZHL. WERT B1)
PULSEOUTP. The binary switching point B1 functions as a pulse output depending on
CONTACT B1 3.3 NO CONTACT Binary switching output B1 is an NO contact.
NC contact Binary switching output B1 is an NC contact.
FUNCTION B2 3.4 INACTIVE See FUNCTION B1
SWITCHING POINT
COUNTERLIM. See FUNCTION B1
PULSEOUTP. See FUNCTION B1
CONTACT B2 3.5 NC contact See CONTACT B1
NO CONTACT See CONTACT B1
the current flow value. Pulses can be generated up to a maximum of 10 Hz. Configuration occurs under Fct. 1.3.1 (pulse duration) and Fct. 1.3.2 pulse/unit).
Note:
Note:
Note:Note: Configuration under Fct 3.6 / Fct. 1.3.1 and Fct. 1.3.2 applies to both pulse outputs. If both B1 and B2 are configured as pulse outputs, both binary outputs behave in exactly the same way.
See FUNCTION B1 Configuration of the switching point takes place under Fct. 1.2.1 (FLOW. VALUE B2).
Configuration of the switching point takes place under Fct. 1.2.1 (COUNTER. VAL B2)
Configuration occurs under Fct. 1.3.1 (pulse width) and Fct. 1.3.2 (pulse/unit).
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
51
Page 52
6 OPERATION
H250 M40
Label Level Select /
Explanation
Input
FUNCTION B3 3.6 INACTIVE The binary switching input B3 has no function.
ACTIV H The internal flow counter is reset to 0.0 if there is an H-level at input B3
ACTIV L The internal flow counter is reset to 0.0 if there is an L-level at input B3
STARTH STOPL The counter is started by creating an H level at the B3 input and stopped
STARTL STOPH The counter is started by creating an L level at the B3 input and stopped
MULTIDROP 3.7 015
4mA CALIBR. 3.8 D/A calibration for beginning of measuring range (4mA)
20mA CALIBR. 3.9 D/A calibration for the final value of the measuring range (20mA)
for at least 100 ms.
for at least 100 ms.
by creating an L level at the B3 input.
by creating an H level at the B3 input.
Polling adress for HART® Multi-Drop Mode When the address is 0, the HART
Attention:
Attention:
Attention:Attention: When the HART
®
multi-drop mode (Address 115) is activated, the
®
multi-drop mode is disabled.
current loop is inactive (fixed current value of 4.5 mA) and no longer follows the current process.
Note:
Note:
Note:Note: This function is not available in HART
Attention:
Attention:
Attention:Attention:
®
multi-drop mode.
During calibration, the value of the current loop does not follow the current process. If a current measuring device in the 4...20mA loop detects a deviation from the desired value 4,000mA, the measured value must be entered. The correction value is accepted following the query Save and Confirm with YES.
ALARM CURRENT
Note:
Note:
Note:Note: This function is not available in HART
Attention:
Attention:
Attention:Attention:
®
multi-drop mode.
During calibration, the value of the current loop does not follow the current process. If a current measuring device in the 4...20mA loop detects a deviation from the desired value 20,000mA, the measured value must be entered. The correction value is accepted following the query Save and Confirm with YES.
3.10 OFF Error indication via the current loop is deactivated. The current loop
follows the current process.
Note:
Note:
Note:Note: This function is not available in HART
®
multi-drop mode.
ALARM HIGH Error indication via the current loop is activated (failure signal "high" as
per NE43).
Note:
Note:
Note:Note: This function is not available in HART
®
multi-drop mode.
ALARM LOW Error indication via the current loop is activated (failure signal "low" as
per NE43).
Note:
Note:
Note:Note:
1) This function is not available in HART
®
multi-drop mode.
2) This function is supported as of Electronic Revision 2.2.x.
52
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 53
H250 M40
OPERATION 6
Label Level Select /
Explanation
Input
END&UNIT 3.11 By changing the unit, the respective final value is scaled accordingly.
3.11.1 FLOW RATE For units for volume flow or mass flow, refer to
3.11.2 COUNTER For units for volume counter or mass counter refer to
Low flow cutoff treshold
3.12 LFC stands for LLLLow FFFFlow CCCCutoff
3.12.1 CONTROL OFF LFC function is not active.
CONTROL ON LFC function is active.
3.12.2 LFC ON_VALUE Switch-on value 1:
Depending on calibration, either units for volume or mass flow measurement are available for selection.
Overview of the units
ESK4-T
on page 40
Overview of the
units ESK4-T
To ensure a stable zero point of the current output, the current output can be stably set to 4.00 mA 3 in a range to be selected.
Value range 1 19% (from final value of measuring range) Flow is greater than the switch-on value. The current output corresponds to this. If the flow rate falls, the current output follows until the switch-on value
1. If the flow value continues to fall, the current output is switched to
4.00 mA 3.
on page 40
Note:
Note:
Note:Note: The switch-on value to be set must be smaller than the previously selected switch-off value.
3.12.3 LFC OFF_VALUE Switch-off value 2:
INP. CODE 3.13 Input code for the local operating menu
3.13.1 OFF Use of an input code is not activated.
3.13.2 ON If YES is chosen, the most recently input code must be typed in.
BASIC SETTING 3.14 Resetting parameters to factory default
Value range 2 20% (from the measuring range final value) The flow rate is 0. The current output is 4.00 mA 3. If the flow rate increases, the current output remains at 4.00 mA until it has reached the switch-off value 2 .
Note:
Note:
Note:Note: The switch-off value to be set must be greater than the previously selected switch-on value.
The input code is not active by default.
Factory code: → → → ^ ^ ^ ↑ ↑ ↑ If, after confirmation with YES, the → button is also pressed, then a new, individual, nine-element code can be typed in. The display shows the required key combination.
There is always a safety prompt (yes/no), to avoid accidental resetting.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
53
Page 54
7 SERVICE
7.1 Maintenance
Within the scope of routine maintenance of the system and pipelines, the flowmeter should also be inspected for signs of fouling, corrosion, mechanical wear and leaks, as well as damage to the measuring tube and indicator. We advise that inspections be carried out at least once per year. The device must be removed from the piping before cleaning.
CAUTION!
Pressurized pipes must be depressurized before removing the device. Empty pipes as completely as possible. In the case of devices used for measuring aggressive or hazardous media, appropriate safety precautions must be taken with regard to residual liquids in the measuring unit. Always use new gaskets when reinstalling the device in the pipeline. Avoid electrostatic charges when cleaning the surfaces (e.g. sight window)!
7.2 Replacement and retrofitting
H250 M40
Some of the variable area flowmeter components can be retrofitted:
Float damping
Indicator M40:
Indicator M40:
Indicator M40:Indicator M40:
Limit switch module
Current output 4-20 mA
Counter module with LCD and I/O
Fieldbus interface
7.2.1 Replacing floats
Remove the device from the piping.
Take the upper snap ring out of the measuring unit.
Take the upper float stop and float out of the measuring unit.
Insert the new float into the centre hole of the lower float stop and push into the measuring
unit along with the upper float catcher. While doing this, the float's upper guide rod must be guided through the middle hole of the float stop.
Insert the snap ring into the measuring unit.
Fit the device back into the piping.
CAUTION!
An additional measuring error is to be expected if recalibration is not conducted.
54
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 55
H250 M40
7.2.2 Retrofitting float damping
Take the upper snap ring 1 out of the measuring unit.
Take the upper float stop 2 and float 5 out of the measuring unit.
Fasten the snap ring 3 into the lower slot of the float's guide rod.
Slide ceramic sleeve 4 on to the float's guide rod and attach it to the top slot using the span
ring 3.
Insert float into the lower float guide in the measuring unit.
Retrofit the supplied damping cylinder with the integrated float stop 2 into the measuring
unit.
Insert upper snap ring 1.
SERVICE 7
1 Snap ring 2 Float catcher 3 Spring washer 4 Ceramic sleeve 5 Float
7.2.3 Retrofitting limit switch
Remove ESK4 add-on module (if available).
Merge contact pointer 2 in the middle.
Loosen the locking screw 1 on the contact pointer.
Insert the contact module into slot 3 of the bracket until the semi-circle 1 on the contact
board surrounds the pointer cylinder.
The contact module connecting terminals feature a pluggable design and can be removed in order to connect the cables.
WARNING!
Do not damage the indicating element!
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
55
Page 56
7 SERVICE
7.2.4 Replacement - Retrofitting ESK4
When replacing or retrofitting an ESK4, the following are required at the time of ordering:
SN - serial number or
SO - sales order
This information can be found on the indicator nameplate.
The ESK4 is factory calibrated, making it possible to replace it or retrofit it without recalibrating.
De-energise the ESK4.
Lift and remove the ESK4 with a screwdriver.
WARNING!
Do not damage the indicating element!
H250 M40
The ESK4 plug-in tongues 1 are plugged in under the two bolts 2 on the baseplate.
Slight pressure is used to press the ESK4 onto the spring pins 3 until it stops, firmly attaching
the ESK4.
If a change in measuring range, product temperature, product, density, viscosity or pressure is
®
desired, this can be done with the KroVaCal program and with a HART measuring unit is subject to its own physical limits, which the DroVaCal program correctly calculates, and may thus reject the desired change. If a change is performed using the program, the new data is also transmitted to the ESK4.
Device identification
Device address
Serial Number
Measuring point designation
Digital measured value query in flow units, % and mA
Test / setting functions
Calibration 4.00 and 20.00 mA
Set current output to any desired value
modem. However, each
56
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 57
H250 M40
7.3 Spare parts availability
The manufacturer adheres to the basic principle that functionally adequate spare parts for each device or each important accessory part will be kept available for a period of 3 years after delivery of the last production run for the device.
This regulation only applies to spare parts which are subject to wear and tear under normal operating conditions.
7.3.1 List of spare parts
Spare part Order no.
DN 15
DN 15
DN 15DN 15
Float CIV 15, 1.4404 X251041000
Float DIV 15, 1.4404 X251042000
Float TIV 15, 1.4404 X251043000
Float DIVT 15, 1.4404 X251044000
Float TIV 15, Aluminium X251043100
Float TIV 15, Titanium X251043200
Set float stop; standard (1 float stop, 1 span ring) X251050100
Set float stop; gas damping (ZrO2) X251050200
Set float stop; gas damping (PEEK) X251050300
Damping bush (7x8) ZrO2 incl. 2 span rings X251053100
Damping bush (7x8) PEEK incl. 2 span rings X251053200
DN 25
DN 25
DN 25DN 25
Float CIV 15, 1.4404 X252041000
Float DIV 25, 1.4404 X252042000
Float TIV 25, 1.4404 X252043000
Float DIVT 25, 1.4404 X252044000
Set float stop; standard (1 float stop, 1 span ring) X252050100
Set float stop; gas damping (ZrO2) X252050200
Set float stop; gas damping (PEEK) X252050300
Damping bush (12x8) ZrO2 incl. 2 span rings X252053100
Damping bush (12x8) PEEK incl. 2 span rings X252053200
DN 50
DN 50
DN 50DN 50
Float CIV 55, 1.4404 X253041000
Float DIV 55, 1.4404 X253042000
Float TIV55, 1.4404 X253043000
Float DIVT 55, 1.4404 X253044000
Set float stop; standard (1 float stop, 1 span ring) X253050100
Set float stop; gas damping (ZrO2) X253050200
Set float stop; gas damping (PEEK) X253050300
Damping bush (14x10) ZrO2 incl. 2 span rings X253053100
Damping bush (14x10) PEEK incl. 2 span rings X253053200
SERVICE 7
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
57
Page 58
7 SERVICE
DN 80
DN 80
DN 80DN 80
Float CIV 85, 1.4404 X254041000
Float DIV 85, 1.4404 X254042000
Float TIV 85, 1.4404 X254043000
Float DIVT 85, 1.4404 X254044000
Set float stop; standard (1 float stop, 1 span ring) X254050100
Set float stop; gas damping (ZrO2) X254050200
Set float stop; gas damping (PEEK) X254050300
Damping bush (18x14) ZrO2 incl. 2 span rings X254053100
Damping bush (18x14) PEEK incl. 2 span rings X254053200
DN 100
DN 100
DN 100DN 100
Float CIV 105, 1.4404 X255041000
Float DIV 105, 1.4404 X255042000
Float DIVT 105, 1.4404 X255044000
Set float stop; stand. (1 float stop, 1 span ring) only for bottom ! X255050100
Set float stop; gas damping (ZrO2) X255050200
Set float stop; gas damping (PEEK) X255050300
Damping bush (18x14) ZrO2 incl. 2 span rings X254053100
Damping bush (18x14) PEEK incl. 2 span rings X254053200
Indicator M40
Indicator M40
Indicator M40Indicator M40
Indicator housing complete no scale X251110000
Cover M40 complete, standard X251110100
Cover M40 complete, salt water resistant X251110200
M40 baseplate standard X251120100
M40 baseplate salt water resistant X251120200
Retrofit kit HT extension X251021000
Module carrier (profile track) X251121100
Pointer system, complete X251122100
Printed scale (serial number required) On request
H250 M40
58
Other spare parts on request
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 59
H250 M40
7.4 Availability of services
The manufacturer offers a range of services to support the customer after expiration of the warranty. These include repair, maintenance, technical support and training.
INFORMATION!
For more precise information, please contact your local sales office.
7.5 Returning the device to the manufacturer
7.5.1 General information
This device has been carefully manufactured and tested. If installed and operated in accordance with these operating instructions, it will rarely present any problems.
CAUTION!
Should you nevertheless need to return a device for inspection or repair, please pay strict attention to the following points:
Due to statutory regulations on environmental protection and safeguarding the health and safety of our personnel, manufacturer may only handle, test and repair returned devices that have been in contact with products without risk to personnel and environment.
This means that the manufacturer can only service this device if it is accompanied by the following certificate (see next section) confirming that the device is safe to handle.
SERVICE 7
CAUTION!
If the device has been operated with toxic, caustic, flammable or water-endangering products, you are kindly requested:
to check and ensure, if necessary by rinsing or neutralising, that all cavities are free from such dangerous substances,
to enclose a certificate with the device confirming that is safe to handle and stating the product used.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
59
Page 60
7 SERVICE
7.5.2 Form (for copying) to accompany a returned device
Company: Address:
Department: Name:
Tel. no.: Fax no.:
Manufacturer's order no. or serial no.:
The device has been operated with the following medium:
H250 M40
This medium is: water-hazardous
toxic
caustic
flammable
We checked that all cavities in the device are free from such substances.
We have flushed out and neutralized all cavities in the device.
We hereby confirm that there is no risk to persons or the environment through any residual media contained in the device when it is returned.
Date: Signature:
Stamp:
7.6 Disposal
CAUTION!
Disposal must be carried out in accordance with legislation applicable in your country.
60
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 61
H250 M40
8.1 Functional principle
The H250 flowmeter operates on the float measuring principle. The measuring unit consists of a metal cone, inside of which a float can move up and down freely. The flow goes from bottom to top. The float changes position so that the lifting force acting on it F1 is in equilibrium with the form drag F2 and its weight F3: F3 = F1 + F2
TECHNICAL DATA 8
1 Indication principle M40 magnetic coupling 2 Magnetic coupling sensors
1 For the indicator, the flow-dependent height of the float in the measuring unit is transmitted
by means of a magnetic coupling and displayed on a scale.
2 For a built-in signal converter (ESK4), the flow-dependent height of the float in the measuring unit is detected by the S1 and S2 magnetic field sensors and electronically processed.
Operating principle of H250H and H250U
Figure 8-1: Operating principle H250H and H250U
1 H250H - horizontal flow direction 2 H250U - flow direction from top to bottom
The flowmeter operates based on a modified float measuring principle. The guided float adjusts itself so that the flow force acting on it is in equilibrium with the opposing spring force. The flow­dependent position of the float in the measuring unit is displayed on a scale by means of a magnetic coupling.
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
61
Page 62
8 TECHNICAL DATA
8.2 Technical data
INFORMATION!
The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local sales office.
Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Application range Flow measurement of liquids, gases and vapors
Function / Measuring principle Variable area measuring principle
Measured value
Primary measured value Float position
Secondary measured value Operating and standard volumetric flow
Measuring accuracy
Directive VDI / VDE 3513, sheet 2 (qG = 50%)
H250 /RR /HC /F 1.6%
H250/C (Ceramic, PTFE) H250H, H250U, H250 (100 : 1)
H250 M40
2.5%
Operating conditions
Temperature
Temperature
TemperatureTemperature Max. operating temperature TS -196..+300°C / -321...+572°F
Pressure
Pressure
PressurePressure
Max. operating pressure PS Depending on the version, up to 400 bar / 5802 psig 1
Max. test pressure PT Pressure equipment directive 97/23/EC or AD 2000-HP30
Min. required operating pressure 2 times greater than pressure loss (see measuring ranges)
Protection category
Protection category
Protection categoryProtection category
M40, M40S, M40R IP 66/68 acc. to EN60529, NEMA 4/4X/6 acc. to NEMA 250
M40R IP69K acc. to DIN 40050-9
Float damping during gas measurement recommended:
Float damping during gas measurement recommended:
Float damping during gas measurement recommended:Float damping during gas measurement recommended: DN15...25 / ½"...1" Operating pressure <0.3 bar / 4.4 psig
DN50...100 / 2"...4" Operating pressure <0.2 bar / 2.9 psig
Installation conditions acc. to VDI/VDE 3513 Sheet 3
Inlet run 5 x DN Outlet run 3 x DN
1 higher operating pressures on request
62
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 63
H250 M40
TECHNICAL DATA 8
Materials
Item Flange / raised face Measurin
g tube
H250/RR stainless steel
H250/HC Hastelloy
H250/C Ceramics/PTFE
2
H250/F - Food CrNi-Stahl 1.4435 -
1 CrNi steel 1.4571 on request, for clamp connection CrNi steel 1.4435 2 DN100/4" only PTFE 3 TFM/PTFE liner (electrically non-conductive)
®
CrNi steel 1.4404 massive 1
CrNi steel 1.4571 with plated Hastelloy
(2.4610) 1
CrNi-Stahl 1.4571 with TFM/PTFE liner 3
®
C4
Float Float stop /
guide
CrNi steel 1.4404 1 -
Hastelloy® C-22 (2.4602)
PTFE or Al2O3 with FFKM gasket
Al2O
3
and PTFE
Ring orifice
-
Al2O
Other options:
Special materials on request: e.g. SMO 254, titanium, 1.4435
Float damping: ceramic or PEEK
Gasket for devices with female thread as insert: O-ring FPM / FKM
Temperatures
Temperatures
TemperaturesTemperatures
For devices to be used in hazardous areas, special temperature ranges apply. These can be found in the separate instructions
3
Temperatures H250/M40 - mechanical indicator without power supply
Material Product temperature Ambient temperature
Float Liner [°C] [°F] [°C] [°F]
H250/RR stainless steel -196...+300 -321...+572 -40...+120 -40+248 H250/RR screw fitting -196...+300 -321...+572 -20+120 -4+248
H250/HC H250/C PTFE -196...+70 -321...+158 -40...+70 -40+158
H250/C Ceramic PTFE -196...+150 -321...+302 -40...+70 -40+158 H250/C Ceramic TFM / Ceramic -196...+250 -321...+482 -40...+120 -40+248
H250 H/U Spring material Stainless Steel
Ambient temperatures T
Hastelloy® C4
316 Spring material Hastelloy -40...+200 -40...+392 -40+120 -40+248
with electrical components
amb.
Version [°C] [°F]
ESK4, ESK4-FF, ESK4-PA -40...+70 -40...+158
ESK4-T 1 -40...+70 -40...+158
Limit switches SJ3,5-SN / I7S23,5-N / Reed SPST -40...+70 -40...+158
Limit switches SC3,5-N0 / SJ3,5-S1N / SB3,5-E2 -25...+70 -13...+158
1 Display contrast out of the temperature range 0...60 °C / 32..140°F decreasing.
-196...+300 -321...+572 -40...+120 -40+248
-40...+100 -40+212 -40+120 -40+248
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
63
Page 64
8 TECHNICAL DATA
Temperatures H250/M40 - with electrical components [°C]
EN ASME Version with Standard HT Standard HT
H250 M40
Tamb. < +40 °C Tamb. < +60 °C
DN15, DN25
DN50 2 ESK4, ESK4-FF, ESK4-PA +200 +300 +165 +300
DN80, DN100
½", 1" ESK4, ESK4-FF, ESK4-PA +200 +300 +180 +300
ESK4-T +200 +300 +80 +130
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +200 +300 +130 +295
ESK4-T +180 +300 +75 +100
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +200 +300 +120 +195
3, 4" ESK4, ESK4-FF, ESK4-PA +200 +300 +150 +250
ESK4-T +150 +270 +70 +85
Limit switch NAMUR +200 +300 +200 +300
3-wire limit switch +190 +300 +110 +160
Maximum product temperatures H250/M40 - with electrical components [°F]
T
< +104 °F T
amb.
EN ASME Version with Standard HT Standard HT
DN15, DN25
DN 50 2 ESK4, ESK4-FF, ESK4-PA 392 572 165 572
DN 80, DN100
½", 1" ESK4, ESK4-FF, ESK4-PA 392 572 356 572
ESK4-T 392 572 176 266
Limit switch NAMUR 392 572 392 572
3-wire limit switch 392 572 266 563
ESK4-T 356 572 167 212
Limit switch NAMUR 392 572 392 572
3-wire limit switch 392 572 248 383
3, 4" ESK4, ESK4-FF, ESK4-PA 392 572 302 482
ESK4-T 302 518 158 185
Limit switch NAMUR 392 572 392 572
3-wire limit switch 374 572 230 320
< +140 °F 1
amb.
64
1 if there are no heat insulation measures, a heat-resistant cable is necessary (continuous operating temperature of the cable to be used: +100°C)
Abbreviation
HT High-Temperature version ESK4 Current output 2-wire 420 mA
ESK4-T ESK4 with LCD, binary status outputs, digital counter and pulse output.
ESK4-FF FOUNDATION FIELDBUS interface
ESK4-PA PROFIBUS PA interface
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 65
H250 M40
TECHNICAL DATA 8
Cable glands
Cable gland Material Cable diameter
M 20x1.5 Standard PA 8...13 mm 0.315...0.512"
M20 x 1.5 Nickel-plated brass 10...14 mm 0.394...0.552"
Limit switches
Terminal connection
Limit switches I7S23,5-N
2.5 mm
2
SJ3,5-SN 1 SJ3,5-S1N 1 SB3,5-E2 Reed
SC3,5-N0
NAMUR (IEC60947-5-6) Yes Yes Yes no no
Type of connection 2-wire 2-wire 2-wire 3-wire 2-wire
Switching element function
Nominal voltage U
0
Pointer vane not detected 3 mA 3 mA 1 mA 0.3 VDC U
Normally closed
Normally closed
Normally open
PNP NO contact
NC SPST
8,2 VDC 8,2 VDC 8,2 VDC 10...30 VDC max. 32 VDC
0
Pointer vane detected 1 mA 1 mA 3 mA UB - 3 VDC 0 VDC
Continuous current - - - max. 100 mA max. 100mA
No load current I
0
- - - 15 mA -
Operating cycles - - - - 100.000
1 safety oriented
Current output ESK4
Terminal connection
Power supply 14...30 VDC
Min. power supply for HART
®
Measuring signal 4.00...20.00 mA = 0...100% flow value in 2-wire technology
Power supply influence <0.1%
Dependence on external resistance <0.1% Temperature influence 5 µA / K
Max. external resistance / load 650 Ohm at 30 VDC
Min. load for HART
®
2
2.5 mm
20 VDC at 250 Ohm load
250 Ohm
ESK4 HART® configuration
Manufacturer name (code) KROHNE Messtechnik (69 = 45h)
Model name ESK4 (214 = 0xD6)
HART® protocol revision
5.9
Device revision 1
Physical layer FSK
Device category Transmitter without galvanic isolation
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
65
Page 66
8 TECHNICAL DATA
ESK4 process variable
Over range +102.5 (±1%) 20.24...20.56
Device error identification > 106.25 >21.00
Max. current consumption 131.25 25
Multi-drop operation 4.5
ESK4-FF
Physical layer IEC 61158-2 and FISCO Modell
Communication standard H1 FOUNDATION Fieldbus protokol
ITK version 5.2
Power supply Bus supply
Nominal current 16 mA
error current 23 mA
Starting current after 10 ms < Nominal current
Values [%] from full-scale range
H250 M40
Signal output [mA]
For more details see the supplementary instructions "H250 M40 Foundation Fieldbus"
ESK4-PA
Physical layer IEC 61158-2 and FISCO Modell
Communication standard Profibus PA Profil 3.02
PNO ID 4531 HEX
Power supply Bus supply
Nominal current 16 mA
error current 23 mA
Starting current after 10 ms < Nominal current
For more details see the supplementary instructions "H250 M40 Profibus PA"
66
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 67
H250 M40
TECHNICAL DATA 8
ESK4-T with LCD, binary inputs and outputs and digital counter
Binary output
Two binary outputs Galvanically isolated, passive
Mode Switch output NAMUR or transistor (open
Configurable as Switch contact or
NAMUR switch output
Power supply 8,2 VDC
Signal current > 3 mA switching value not
Switch output transistor (open collector)
Power supply Nominal 24 VDC, maximal 30 VDC
P
max
Continuous current max. 100 mA
No load current I
0
Pulse output
reached;
500 mW
2mA
collector)
Opener/NO contact or max. 10 pulses / s
< 1 mA switching value reached
Pulse output
T
on
T
off
Pulse value
configurable from 50...500 ms
depending on flow rate
configurable in flow units e.g. 5 pulses / m
Binary input
Input Galvanically isolated
Mode Reset counters or run/stop
Configurable as active Hi / active Lo
H- signal 16...30 VDC
Internal resistance R
Ton (active) 500 ms
i
typ. 20 kOhm
Approvals
Standard Screen Marking
ATEX / IECEx
FM (USA) FM (Canada)
NEPSI M40 pending
M40 mechanical II2GD IIC
II3GD IIC
M40 electrical III2G Ex ia IIC T6 Gb
II2G Ex d IIC T6 Gb II3G Ex nA IIC T6 Gc II2D Ex t IIIC T70°C Db
M40 pending
3
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
67
Page 68
8 TECHNICAL DATA
8.3 Dimensions and weights
Dimensions H250/M40
Front view Side view with heating High temperature
H250 M40
a b d h
[mm] ["] [mm] ["] [mm] ["] [mm] ["]
All nominal sizes 138 5.44 250 9.85 160 6.30 150 5.91
ISO 228 300 11.82
H250/C - 3"/300 lb 300 11.82
EN ASME c e Ø f g j
[mm] ["] [mm] ["] [mm] ["] [mm] ["] [mm] ["]
DN15 ½" 94 3.70 114 4.49 20 0.79 100 3.94 197 7.76
DN25 1" 94 3.70 125 4.92 32 1.26 106 4.18 208 8.19
DN50 2" 107 4.22 139 5.48 65 2.56 120 4.73 222 8.75
DN80 3" 107 4.22 155 6.11 89 3.51 145 5.71 238 9.38
DN100 4" 107 4.22 164 6.46 114 4.49 150 5.91 247 9.73
ISO 228
Female thread
screwed
ISO 228
Female thread
welded
H250/F
Clamp connection
H250/F
Screw connection
DIN 11851
1
1 Stainless steel 1.4435 - EHEDG tested - wetted surfaces Ra 0.8 / 0.6 µm
68
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 69
H250 M40
TECHNICAL DATA 8
Weights
Weights
WeightsWeights
H250 with heating
Nominal meter size EN 1092-1 Flange connection Ermeto connection
EN ASME [kg] [lb] [kg] [lb] [kg] [lb]
DN15 ½“ 3.5 7.7 5.6 12.6 3.9 8.6 DN25 1 5 11 7.5 16.5 5.8 12.8 DN50 2 8.2 18.1 11.2 24.7 9.5 21 DN80 3 12.2 26.9 14.8 32.6 13.1 28.9 DN100 4 14 30.9 17.4 38.4 15.7 34.6
H250/C [Ceramic / PTFE] Screw connec.
Nominal meter size EN 1092-1 ASME 150 lb ASME 300 lb DIN 11864-1
EN ASME [kg] [lb] [kg] [lb] [kg] [lb] [kg] [lb]
DN15 ½“ 3.5 7.7 3.2 7.1 3.5 7.7 2 4.4 DN25 1 5 11 5.2 11.5 6.8 15 3.5 7.7 DN50 2 10 22.1 10 22.1 11 24.3 5 11 DN80 3 13 28.7 13 28.7 15 33.1 7.6 16.8 DN100 4 15 33.1 16 35.3 17 37.5 10.3 22.7
Process connections
Process connections
Process connectionsProcess connections
Standard Conn. dim. Pressure rating
Flanges (H250/RR /HC /C) EN 1092-1 DN15...150 PN16...250
ASME B16.5 ½...6" 150...2500 lb JIS B 2220 15100 10...20K
Clamp connections (H250/RR /F) DlN 32676 DN15...100 10...16 bar
ISO 2852 Size 25...139.7 10...16 bar
Screw connections (H250/RR /HC /F) DIN 11851 DN15...100 25...40 bar
SMS 1146 1...4" 6 bar / 88.2 psig
Female thread welded (H250/RR /HC) ISO 228 G½...G2" 50 bar / 735
ASME B1.20.1 ½…2" NPT
Female thread (H250/RR /HC) with insert, FPM gasket and union nut
Thread connection aseptic (H250/F) DIN 11864 - 1 DN1550 PN40
Flange aseptic (H250/F) DIN 11864 - 2 DN1550 PN40
Meters (H250/RR /HC) with heating:
Meters (H250/RR /HC) with heating:
Meters (H250/RR /HC) with heating:Meters (H250/RR /HC) with heating:
Heating with flange connection EN 1092-1 DN15 PN40
Heating pipe connection for Ermeto - E12 PN40
ISO 228 G½…2" 50 bar ASME B1.20.1 ½…2" NPT
DN80100 PN 16
DN80DN100 PN 16
ASME B16.5 ½" 150 lb / RF
psig
735 psig
Higher pressure ratings and other connections on request
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
69
Page 70
8 TECHNICAL DATA
Bolts and tightening torques
For measuring devices with PTFE liner or ceramic liner and PTFE raised face, tighten the flange threads with the following torques:
Nominal sizes EN
Nominal size acc. to EN 1092-1 Quantity x size [Nm] [lb-ft]
DN15 PN40 1 4x M12 9.8 7.1
DN25 PN40 1 4x M12 21 15
DN50 PN40 1 4x M16 57 41
DN80 PN16 1 8x M16 47 34
DN100 PN16 1 8x M16 67 48
1 standard connections; other connections on request
H250 M40
Stud bolts Tightening torques
Nominal size ASME
Stud bolts Tightening torques
Nominal sizes acc. to ASME B
Quantity x size [Nm] [lb-ft]
16.5 150 lb 300 lb
½" 150 lb / 300 lb 1 4x ½" 4x ½" 5.2 3.8 1" 150 lb / 300 lb 1 4x ½" 4x 5/8" 10 7.2
2" 150 lb / 300 lb 1 4x 5/8" 8x 5/8" 41 30 3" 150 lb / 300 lb 1 4x 5/8" 8x ¾" 70 51 4" 150 lb / 300 lb 1 8x 5/8" 8x ¾" 50 36
1 standard connections; other connections on request
Pressure tightness (vacuum) H250/C
Max. process temperature  +70°C (+158°F) +150°C (*302°F) +250°C (+482°F)
Min. operating pressure
Nominal meter size
Float Lin
er
[mbar abs.]
[psia] [mbar
abs.]
[psia] [mbar
abs.]
[psia]
70
DN15...DN100 PTFE PTFE 100 1.45 - - - --
DN15...DN80 Ceramic PTFE 100 1.45 250 3.63 - -
DN15...DN80 Ceramic TFM /
Ceramic
100 1.45 100 1.45 100 1.45
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 71
H250 M40
8.4 Measuring ranges
TECHNICAL DATA 8
H250/RR - Stainless Steel, H250/HC - Hastelloy
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
®
Water Air Max. pressure loss
Float TIV CIV DIV TIV
Nominal
Cone [l/h]
TIV DIV TIV
Alu
[Nm3/h]
TIV CIV DIV
Alu
[mbar]
meter size
DN15, ½" K 15.1 18 25 - 0.42 0.65 - 12 21 26 -
K 15.2 30 40 - 0.7 1 - 12 21 26 -
K 15.3 55 63 - 1 1.5 - 12 21 26 -
K 15.4 80 100 - 1.7 2.2 - 12 21 26 -
K 15.5 120 160 - 2.5 3.6 - 12 21 26 -
K 15.6 200 250 - 4.2 5.5 - 12 21 26 -
K 15.7 350 400 700 6.7 10 18 1 12 21 28 38
K 15.8 500 630 1000 10 14 28 2 13 22 32 50
K 15.8 - - 1600 3 - - 50 3 - - - 85
DN25, 1" K 25.1 480 630 1000 9.5 14 - 11 24 32 72
K 25.2 820 1000 1600 15 23 - 11 24 33 74
K 25.3 1200 1600 2500 22 35 - 11 25 34 75
K 25.4 1700 2500 4000 37 50 110 2 12 26 38 78
K 25.5 3200 4000 6300 62 95 180 2 13 30 45 103 4
DN50, 2" K 55.1 2700 6300 8400 58 80 230 2 8 13 74 60
K 55.2 3600 10000 14000 77 110 350 2 8 13 77 69
K 55.3 5100 16000 25000 110 150 700 2 9 13 84 104
DN80, 3" K 85.1 12000 25000 37000 245 350 1000 2 8 16 68 95
K 85.2 16000 40000 64000 280 400 1800 2 9 16 89 125
DN100, 4" K105.1 19000 63000 100 000 - 550 2800 2 - - 120 220
1 P >0.5 bar 2 P >0,5 bar 3 with TR float 4 300 mbar with damping (gas measurement)
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bar abs. (DIN 1343)
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
71
Page 72
8 TECHNICAL DATA
H250 M40
H250/RR - Stainless Steel, H250/HC - Hastelloy
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
®
Water Air Max. pressure loss
Float TIV CIV DIV TIV
Nominal
Cone [GPH] [SCFM] [psig]
Alu
TIV DIV TIV
Alu
TIV CIV DIV
meter size
DN15, ½" K 15.1 4.76 6.60 - 0.26 0.40 - 0.18 0.31 0.38 -
K 15.2 7.93 10.6 - 0.43 0.62 - 0.18 0.31 0.38 -
K 15.3 14.5 16.6 - 0.62 0.93 - 0.18 0.31 0.38 -
K 15.4 21.1 26.4 - 1.05 1.36 - 0.18 0.31 0.38 -
K 15.5 31.7 42.3 - 1.55 2.23 - 0.18 0.31 0.38 -
K 15.6 52.8 66.0 - 2.60 3.41 - 0.18 0.31 0.38 -
K 15.7 92.5 106 185 4.15 6.20 11.2 1 0.18 0.31 0.41 0.56
K 15.8 132 166 264 6.20 8.68 17,4 2 0.19 0.32 0.47 0.74
K 15.8 - - 423 3 - - 31.0 3 - - - 1.25
DN25, 1" K 25.1 127 166 264 5.89 8.68 - 0.16 0.35 0.47 1.06
K 25.2 217 264 423 9.30 14.3 - 0.16 0.35 0.49 1.09
K 25.3 317 423 660 13.6 21.7 - 0.16 0.37 0.50 1.10
K 25.4 449 660 1057 22.9 31.0 68,2 2 0.18 0.38 0.56 1.15
K 25.5 845 1057 1664 38.4 58.9 111 1 0.19 0.44 0.66 1.51 4
DN50 2" K 55.1 713 1664 2219 36.0 49.6 143 2 0.12 0.19 1.09 0.88
K 55.2 951 2642 3698 47.7 68.2 217 2 0.12 0.19 1.13 1.01
K 55.3 1347 4227 6604 68.2 93.0 434 2 0.13 0.19 1.23 1.53
DN80 3" K 85.1 3170 6604 9774 152 217 620 2 0.12 0.24 1.00 1.40
K 85.2 4227 10567 16907 174 248 1116 2 0.13 0.24 1.31 1.84
DN100 4" K105.1 5019 16643 26418 - 341 1736 2 - 1.76 3.23
1 P >7.4 psig 2 P >7,4 psig 3 with TR float 4 4.4 psig with damping (gas measurement)
72
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition during gas measurements:
Reference condition during gas measurements:
Reference condition during gas measurements:Reference condition during gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bar abs. (ISO 13443)
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 73
H250 M40
TECHNICAL DATA 8
H250/C - Ceramic/PTFE
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
Flow Max. pressure loss
Water Air Water Air
Liner /
PTFE Ceramic PTFE Ceramic PTFE Ceramic PTFE Ceramic
Float
Nominal
Cone [l/h]
[Nm3/h]
[mbar]
meter size
DN15, ½" E 17.2 25 30 0,7 - 65 62 65 62
E 17.3 40 50 1,1 1.8 66 64 66 64
E 17.4 63 70 1,8 2.4 66 66 66 66
E 17.5 100 130 2,8 4 68 68 68 68
E 17.6 160 200 4,8 6.5 72 70 72 70
E 17.7 250 250 7 9 86 72 86 72
E 17.8 400 - 10 - 111 - 111 -
DN25, 1" E 27.1 630 500 16 18 70 55 70 55
E 27.2 1000 700 30 22 80 60 80 60
E 27.3 1600 1100 45 30 108 70 108 70
E 27.4 2500 1600 70 50 158 82 158 82
E 27.5 4000 1 2500 120 75 290 100 194 100
DN50, 2" E 57.1 4000 4500 110 140 81 70 81 70
E 57.2 6300 6300 180 200 110 80 110 80
E 57.3 10000 11000 250 350 170 110 170 110
E 57.4 16000 1 - - - 284 - - -
DN80, 3" E 87.1 16000 16000 - - 81 70 - -
E 87.2 25000 25000 - - 95 85 - -
E 87.3 40000 1 - - - 243 - - -
DN100, 4" E 107.1 40000 - - - 100 - - -
E 107.2 60000 1 - - - 225 - - -
1 special float
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition during gas measurements:
Reference condition during gas measurements:
Reference condition during gas measurements:Reference condition during gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bar abs. (DIN 1343)
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
73
Page 74
8 TECHNICAL DATA
H250/C - Ceramic/PTFE
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
Flow Max. pressure loss
Water Air Water Air
Liner / Float PTFE Ceramic PTFE Ceramic PTFE Ceramic PTFE Ceramic
H250 M40
Nominal
Cone [GPH] [SCFM] [psig]
meter size
DN15, ½" E 17.2 6.60 7.93 - 0,94 0,90 0,94 0,90
E 17.3 10.6 13.2 1.12 0,96 0,93 0,96 0,93
E 17.4 16.6 18.5 1.49 0,96 0,96 0,96 0,96
E 17.5 26.4 34.3 2.48 0,99 0,99 0,99 0,99
E 17.6 42.3 52.8 4.03 1,04 1,02 1,02 1,02
E 17.7 66.0 66.0 5.58 1,25 1,04 1,25 1,04
E 17.8 106 - - 1,61 - 1,61 -
DN25, 1" E 27.1 166 132 11.2 1,02 0,80 1,02 0,80
E 27.2 264 185 13.6 1,16 0,87 1,16 0,87
E 27.3 423 291 18.6 1,57 1,02 1,57 1,02
E 27.4 660 423 31.0 2,29 1,19 2,29 1,19
E 27.5 1056 1 660 46.5 4,21 1,45 2,81 1,45
DN50, 2" E 57.1 1057 1189 86.8 1,18 1,02 1,18 1,02
E 57.2 1664 1664 124 1,60 1,16 1,60 1,16
E 57.3 2642 2906 217 2,47 1,60 2,47 1,60
E 57.4 4226 1 - - 4,12 - - -
DN80, 3" E 87.1 4227 4227 - 1,18 1,02 - -
E 87.2 6604 6604 - 1,38 1,23 -
E 87.3 10567 1 - - 3,55 - -
DN100, 4" E 107.1 10567 - - 1,45 - -
E 107.2 15850 1 - - 3,29 - -
1 special float
74
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bar abs. (ISO 13443)
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 75
H250 M40
TECHNICAL DATA 8
H250H - Horizontal installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
EN ASME Cone Water [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15 ½ K 15.1 70 1.8 195
K 15.2 120 3 204
K 15.3 180 4.5 195
K 15.4 280 7.5 225
K 15.5 450 12 250
K 15.6 700 18 325
K 15.7 1200 30 590
K 15.8 1600 40 950
K 15.8 2400 60 1600
DN25 1" K 25.1 1300 35 122
K 25.2 2000 50 105
K 25.3 3000 80 116
K 25.4 5000 130 145
K 25.5 8500 220 217
K 25.5 10000 260 336
DN50 2" K 55.1 10000 260 240
K 55.2 16000 420 230
K 55.3 22000 580 220
K 55.3 34000 900 420
DN80 3" K 85.1 25000 650 130
K 85.2 35000 950 130
K 85.2 60000 1600 290
DN100 4" K 105.1 80000 2200 250
K 105.1 120000 3200 340
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bar abs. (DIN 1343)
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
75
Page 76
8 TECHNICAL DATA
H250H - Horizontal installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
EN ASME Cone Wasser [GPH] Luft [SCFM] Pressure loss
DN15 1/2" K 15.1 18.5 1.12 2.87
K 15.2 31.7 1.86 3.00
K 15.3 47.6 2.79 2.87
K 15.4 74.0 4.65 3.31
K 15.5 119 7.44 3.68
K 15.6 185 11.2 4.78
K 15.7 317 18.6 8.68
K 15.8 423 24.8 14.0
K 15.8 634 37.2 23.5
DN25 1" K 25.1 343 21.7 1.79
K 25.2 528 31.0 1.54
K 25.3 793 49.6 1.71
K 25.4 1321 80.6 2.13
K 25.5 2245 136 3.19
K 25.5 2642 161 4.94
DN50 2" K 55.1 2642 161 3.53
K 55.2 4227 260 3.38
K 55.3 5812 360 3.23
K 55.3 8982 558 6.17
DN80 3" K 85.1 6604 403 1.91
K 85.2 9246 589 1.91
K 85.2 15851 992 4.26
DN100 4" K 105.1 21134 1364 3.68
K 105.1 31701 1984 5.00
H250 M40
[psig]
76
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bar abs. (ISO 13443)
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 77
H250 M40
TECHNICAL DATA 8
H250U - Vertical installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
Flow direction vertical downwards
EN ASME Cone Water [l/h]
Air [Nm3/h]
Pressure loss
[mbar]
DN15 ½" K 15.1 65 1.6 175
K 15.2 110 2.5 178
K 15.3 170 4 180
K 15.4 260 6 200
K 15.5 420 10 220
K 15.6 650 16 290
K 15.7 1100 28 520
K 15.8 1500 40 840
DN25 1" K 25.1 1150 30 97
K 25.2 1800 45 85
K 25.3 2700 70 92
K 25.4 4500 120 115
K 25.5 7600 200 172
DN50 2" K 55.1 9000 240 220
K 55.2 15000 400 230
K 55.3 21000 550 240
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
3
Nl/h or Nm
/h: Volume current in standard state 0°C - 1.013 bar abs. (DIN 1343)
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
77
Page 78
8 TECHNICAL DATA
H250U - Vertical installation position
Measuring span: 10 : 1 Flow values: Values = 100% Water 20°C / 68°F Air: 20°C [68°F], 1,013 bar abs. [14,7 psia]
Flow direction vertical downwards
EN ASME Cone Wasser [GPH] Luft [SCFM] Pressure loss
DN15 ½" K 15.1 17.2 0.99 2.57
K 15.2 29.1 1.55 2.62
K 15.3 44.9 2.48 2.65
K 15.4 68.7 3.72 2.94
K 15.5 111 6.20 3.23
K 15.6 172 9.92 4.26
K 15.7 291 17.4 7.64
K 15.8 396 24.8 12.3
DN25 1" K 25.1 304 18.6 1.42
K 25.2 476 27.9 1.25
K 25.3 713 43.4 1.35
K 25.4 1189 74.4 1.69
K 25.5 2008 124 2.53
DN50 2" K 55.1 2378 149 3.23
K 55.2 3963 248 3.38
K 55.3 5548 341 3.53
H250 M40
[psig]
INFORMATION!
The operating pressure should be at least double the pressure loss for liquids and five times for gases. The indicated pressure losses are valid for water and air at maximum flow rate. Other flow ranges on request. Conversion of other media or operating data is performed using the calculation method in accordance with VDI /VDE Directive 3513.
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:Reference condition for gas measurements: Flow measurements for gases are attributed to
SCFM or SCFH: Volume current in standard state 15°C - 1.013 bar abs. (ISO 13443)
78
www.krohne.com 11/2012 - 4000640702 MA H250 M40 R02 en
Page 79
H250 M40
NOTES 9
www.krohne.com11/2012 - 4000640702 MA H250 M40 R02 en
79
Page 80
KROHNE product overview
Electromagnetic flowmeters
Variable area flowmeters
Ultrasonic flowmeters
Mass flowmeters
Vortex flowmeters
Flow controllers
Level meters
Temperature meters
Pressure meters
Analysis products
Products and systems for the oil & gas industry
Measuring systems for the marine industry
Head Office KROHNE Messtechnik GmbH Ludwig-Krohne-Str. 5 47058 Duisburg (Germany) Tel.:+49 (0)203 301 0 Fax:+49 (0)203 301 10389 info@krohne.de
The current list of all KROHNE contacts and addresses can be found at:
© KROHNE 11/2012 - 4000640702 MA H250 M40 R02 en - Subject to change without notice.
www.krohne.com
Loading...