Safety instructions
This product is designed for use in accordance with EN IEC 61010-1 for Installation Category 2 and Pollution
Degree 2. Hazardous voltages are present within this product during normal operation. The product is designed
for Protection Class I and should never be operated without protective earthing. The product shall also never be
operated with covers removed unless equivalent protection of the operator and its environment from accidental
contact with hazardous internal voltages is provided. Always follow basic and local safety precautions when
using this product to reduce risk of injury from electrical shock, spread of fire or other dangerous situations.
1. Do not install, maintain or operate this flow meter without reading, understanding and following the factory-supplied
instructions and manuals, otherwise injury or damage may result.
2. Read these instructions carefully before starting installation. Save these instructions for future reference.
3. Observe all warnings and instructions marked on the product
4. Use only mains supply with protective earthing connected
5. Do not use the product with removed covers under wet conditions
6. Consider handling and lifting instructions to avoid damage
7. Install the product securely and stable
8. Install and connect cabling proper to exclude damage or harmful situations
9. If the product does not operate normally refer to qualified KROHNE service technicians
10. Ther are no operator serviceable components inside the product
The following symbols may appear in this manual or on the product :
Attention : refer to the manual
DANGER : risk of electrical shock
PROTECTIVE conductor terminal
WARNING
Warning statements identify conditions or practice that could result in injury loss of live
CAUTION
Caution statements identify conditions or practice that could result in damage to the product
or other property
Declarations of Conformity / Statement of Compliance
Low Voltage directive
The product complies with the requirements of the Low Voltage Directive 73/23/EEC and is designed in accordance with
EN IEC 61010-1 first and second edition. (safety requirements for electrical equipment for measurement, control and
laboratory use part 1) Local safety regulations shall be observed in combination with the measures special to this
product to avoid dangerous situations
2
Page 3
Unpacking and inspection
This product has been thoroughly inspected and tested before shipment and is ready for operation.
After carefully unpacking, inspect for shipping damage before attempting to operate. If any indication of mechanical damage is
found contact immediate the responsible transport service and your local KROHNE representative.
A simple operating check of the electronics after unpacking and before permanent installation, is advisable to ascertain whether it
has suffered damage during shipment. Confirm for the correct mains voltage printed on the nameplate. If it differs from the
ordered product please contact your local KROHNE representative.
After connecting to the mains, check if there is any indication on the display and if the backlight of the display is lighted. If not
contact your local KROHNE representative for advise.
Disclamer
This document contains important information of the product. KROHNE attempts to be as accurate and up-to-date as possible but
assumes no responsibility for errors or omissions. Nor does KROHNE make any commitment to update the information contained
herin. This manual and all other documents are subject to change without notice. KROHNE will not be liable for any damage of
any kind by using its product, including, but not limited to direct, indirect, incidental, punitive and consequential damages. This
disclamer does not apply in case KROHNE has acted on purpose or with gross negligence. In the event any applicable law does
not allow such limitations on implied warranties or the exclusion of limitation of certain damages, you may, if such law applies to
you, not be subject to some or all of the above disclamer, exclusions or limitations. Any product purchased from KROHNE is
warranted in accordance with the relevant product documentation and our Terms and Conditions of Sale. KROHNE reserves the
right to alter the content of its documents, including this disclamer in any way, at any time, for any reason, without prior
notification, and will not be liable in any way for possible consequences of such changes.
CAUTION
There are no operator servicable parts or adjustments within the product enclosure
Refer all servicing to trained service technicians
Source power shall be removed from the product to perform any servicing
This product is prepared for the nominal AC mains voltage indicated on its tagplate. To reconfigure
the product power input for other mains voltages as listed in the chapter Technical data, the product
must be modified including the fuses by a trained service technician. After a power input
reconfiguration the tagplate needs to be adapted.
This product is a Class 1 device, (earthed which utilizes a proper connection to the protective earthing
for protection from electrical shock to ensure operator and its environmental safety.
The protective conductor terminal of the product shall be properly connected to the protective earth
wiring of the source power to ensure safety from electrically shock. See instruction manual
Protective earthing conductor terminal.
This symbol indicates the point on the product to which the protective earthing conductor shall be
connected. It is positioned near the terminal block, inside the terminal compartmen. The diameter of
the protective earthing conductor shall be in accordance with the Low Voltage Directive.
Manufacturer
The GFM 700 is developed and manufactured by
KROHNE Altometer
Kerkeplaat 12
3313 LC Dordrecht, The Netherlands
3
Page 4
Content
1 GFM 700 Ultrasonic gas flowmeter 5
2 Size, flow velocity, flowrate 5
3 Measuring principle 6
4 Technical data 7 – 10
4.1 Versions, full-scale ranges, accuracies 7
4.2 GFS 700 Primary head 7
4.3 GFC 700 Signal converter 8 – 9
4.4 Environmental conditions 9
4.5 Dimensions and weights 10
5 Installation notes 11
6 Electrical 11
6.1 Electrical connections 11
6.2 Electrical installation to the main supply voltage 11
7 Operation of the signal converter 12 – 20
7.1 Table of settable functions 12 – 18
7.2 Error messages 19 – 20
8
Addition for the GFM 700 F-EEx ATEX and
20 – 30
GFM 700 HT EEx-ATEX
8.1 General 21
8.2 System components 21 – 22
8.3 Electrical connection 23 – 28
8.4 Connecting cables 29
8.5 Connection diagram 30
9 ATEX approvals 31 – 36
9.1 GFC 700 F - EEx Signal converter 31 – 33
9.2 GFS 700 F - EEx Primary head (flowmeter) 34 – 36
4
Page 5
1 GFM 700 Ultrasonic gas flowmeter
First ultrasonic gas flowmeter
with two beams
Non-contact ultrasonic gas flow measurement
and volume counting, suitable for
all gases available meter sizes DN 50-600
and 2”-24”.
Wide range of application
The large range of products that can be
measured accurately and easily make this a
truly universal device. Just a few examples:
Natural gas
• Air
• Methane
• Nitrogen
• Determination of the molecular weight of
gases
• Measurement in hazardous locations, Zones 1 and 2
2 Size, flow velocity, flowrate
Selection of meter size / nominal pipe size
Flow velocitys in m/s or ft/s and the flowrate in m3/h or
US Gal / min can be determinded for each meter size
(DN) using the Tables below.
Example: v in m/s Example: v in ft/s
Nominal pipe size DN 150
Required measuring range 1000 m3/h
For a flow velocity of 1 m/s at DN 150,
the table gives a flowrate of 63.617 m3/h.
For 1000 m
3
/hr, the flow velocity v is thus
1000 m3/h x 1 m/s
v = -------------------------- = 15.72 m/s
63.617 m3/h
Flowtable for v = 1 m/s
Meter size Flow Meter size Flow
in mm in inches m3/h in inches in mm US Gal/min
DN 50 2 7.0686 2 DN 50 31.13
Calibrated on EN 17025 accredited calibration rigs,
calibration accuracy better than 99.97% of the
measured value.
Precise determination of flow velocities
For range setting, use the flow table below to determine
the precise flow velocity for each nominal pipe size.
Nominal pipe size 6”
Required measuring range 5000 US Gal / min
For a flow velocity of 3.3 ft/s at 6’’,
the table gives a flowrate of 280.11 gal/min.
For 5000 Gal / min, the flow velocity v is thus
5000 US / min x 3.3 ft/s
v = ---------------------------------- = 58.91 ft/s
280.11 US Gal / min
Flowtable for v = 10 ft/s
5
Page 6
3 Measuring principle
All KROHNE ultrasonic flowmeters operate using the
transit-time differential method.
Transit-time differential measurement is based on a
simple physical fact.
Imagine two canoes crossing a river on the same
diagonal line, one with the flow and the other against the
flow. The canoe moving with the flow needs much less
time to reach the opposite bank.
Ultrasonic waves behave exactly the same way. A sound
wave travelling in the direction of flow of the product is
propagated at a faster rate than one travelling against the
flow (v
> vBA).
AB
Transit times t
difference (t
waves is directly proportional to the mean flow velocity
(vm) of the product.
The volumetric flowrate per unit time is the product of the
mean flow velocity (v
section.
A liquid product is identified by direct measurement of the
transit time of ultrasonic waves. Assuming the same path
length (L), the transit time in water is shorter than in crude
oil, for example.
and tBA are measured continuously. The
AB
- tAB) in time travelled by the two ultrasonic
BA
) multiplied by the pipe cross-
m
Propagation rate
of ultrasonic waves . . .
… in direction of flow
of product from sensor A to B
+ vm x cosϕ
= c
v
AB
o
… counter to direction of flow
of product
v
BA
= c
– vm × cosϕ
o
from sensor A to B
with the given product
flow direction
mean flow velocity v
of liquid product
m
v
AB
> v
BA
(t
– tAB)
BA
vm = GK x ---------------
(tAB x tBA)
Transit time
of ultrasonic waves … L
= ---------------------
t
AB
co + vm x cosϕ
L
t
= --------------------
AB
c
- vm x cosϕ
o
tAB < tBA
A (A’)
B (B’)
co
GK
L
ϕ
Sensor A, transmitter and receiver
Sensor B, transmitter and receiver
Sound velocity in the product
A calibration constant
Length of measuring beam,
distance between sensors A and B
Angle between pipe axis and
measuring beam
t
Transit time of ultrasonic waves
AB
from sensor A to sensor B
t
Transit time of ultrasonic waves
BA
from sensor B to sensor A
v
Propagation rate of ultrasonic waves between
AB
sensor A and sensor B
v
Propagation rate of ultrasonic waves between
BA
sensor B and sensor A
v
Mean flow velocity of liquid product
m
6
Page 7
4
Technical data
The responsibility as to the suitability, intended use and corrosion-resistanc
of the materials used in their construction rests solely with the purchaser.
4.1 Versions, full-scale ranges, accuracies
Versions Primary head (S) Signal converter (C)
GFM 700 K integral system GFS 700 K GFC 700 K
GFM 700 F remote system GFS 700 F GFC 700 F
GFM 700 F-EEx GFS 700 F-EEx GFC 700 F - EEx
(Ex versions (remote system)) Approval to ATEX (Ex II 2 G) signal converter is installed
EEX de IIC T6 oder EEx d IIC T6 in non hazardous area
Flange connections Pressure rating (standard)
to DIN 2501 DN 50, DN 80: PN 40 40 bar / 580 psig
DN 65, DN 100 – 150: PN 16 16 bar / 230 psig
DN 200 – 600: PN 10 10 bar / 150 psig
to ANSI B 16.5 2” – 24”: Class 150 lb / RF 19 bar / 275 psig
Special versions max. 100 bar / 900 lb
Max. operating data Gas-Temperature Operating pressure
Integral systems standard option
Ambient temperature ≤ 40°C / ≤ 104°F
Ambient temperature ≤ 60°C / ≤ 140°F
Remote systems
Hazardous-duty versions
≤ 140°C / 284 °F ≤ 25 bar / 360 psig ≤ 40 bar / 580 psig
≤ 60°C / 140 °F ≤ 25 bar / 360 psig ≤ 40 bar / 580 psig
≤ 180°C / 356 °F ≤ 25 bar / 360 psig ≤ 40 bar / 580 psig
≤ 180°C / 356 °F ≤ 20 bar / 300 psig
–
Max. allowable flow velocity
Max. allowable meter size (DN)
as a function of process product
Protection category (IEC 529 / EN 60529)
Materials
≤ 25 m/s / ≤ 80 ft/s, optionally ≤ 30 m/s / ≤ 100 ft/s
DN
[mm] = 200 x
max
ensity ρ
D
in kg/m3 or in [lb/ft3]
Gas
[kg/m3] or [inches] = 0.47 x ρ
ρ
Gas
IP 65 equivalent to NEMA 4 and 4X
Measuring tube and flanges
Standard DN 50-300 2“-12“ SS 1.4301 (measuring tube) and steel (flanges)
DN 350-600 14“-24“ steel
Option DN 50-600 2“-24“ SS 1.4404
DN 50-600 (only DIN) SS 1.4571
Ultrasonic sensors
Gaskets
Viton
SS 1.4301
7
[lb/ft3]
gas
Page 8
4.3 GFC 700 Signal converter
Versions
Integral systems (K)
Remote systems (F)
Option MP
Power supply 1. AC Version AC / DC Version
Standard Option
1. Rated voltage 115 / 230 V 24 V AC 24 V DC
Tolerance band +/- 13 % 20 – 27 V AC 18 – 32 V DC
Frequency 48 – 63 Hz 48 – 63 Hz –
Power consumption Max. 13 VA approx. 10 VA approx. 8 W
(incl. primary head)
to VDE 0100 / VDE 0106 and IEC 364 / IEC 536 or equivalent national standards.
When connected to functional extra-low voltage (24 V) safety separation (PELV) is essential
Analog Output
Function
Current for Q = 0%: 0 – 16 mA
for Q = 100%: 4 – 20 mA
Active mode load max. 680 Ohm
Passive mode
Low-flow cutoff cutoff “on” value 1 – 19%
cutoff “off” value 2 – 20%
Time constant 0.04 – 3600 s, setting in increments of 1, 0.1 or 0.01 s
Forward / reverse measurement direction identified via status output (or pulse output)
Internal power supply for binary outputs
GFC 700 K signal converter mounted on primary head
GFC 700 F signal converter with wall mount (rotating design) and
additional terminal box
signal converter equipped with magnet sensors, to set the signal
converter by means of hand held bar magnet without opening the
housing
• continuous flowrate measurement or measurement of sound velocity
to determine (composition of) liquid product
• all operating data configurable
• galvanically isolated
• for active and passive mode
• useable as internal power supply for the binary outputs
settings in 1 mA increments
(I
= 22 mA)
max
external voltage ≤ 18 V DC
load ≤ 680 Ohm
100%
,
of Q
setting in 1% increments
= 19 – 32 V DC / I ≤ 50 mA
U
int
Pulse output
Function
– continuous flow counting or measurement of sound velocity
to determine (composition of) liquid product
– all operating data configurable
– galvanically isolated
– active and passive mode
Pulse rate for Q = 100% 10 – 3600000 pulses/h
0,167 – 60000 pulses/min
0.0028 – 1000 pulses/s (= Hz)
optionally in pulse per liter, m
US gallons or user-defined unit
Time constant 0.04 – 3600 s, setting in increments of 1, 0.1 or 0.01 s
direction identified via status output or current output (status output only
Forward / reverse measurement
available in non-EEx version)
8
Page 9
Local display
Display function
Units: actual flowrate
totalizer
Language of plain texts English, French, German, Dutch, other on request
Display 1st line (top)
2nd line (middle) 10-character, 14-segment text display
3rd line (bottom) 5 markers to identify display in measuring mode
Housing
Material die-cast aluminium with polyurethane finish
Protection category IP 67 (IEC 529 / EN 60529) equivalent to NEMA 4 and 4X
Signal cable
only for remote systems (F)
3-field LCD
actual flowrate, measurement of sound velocity to determine (composition of) liquid product, forward, reverse and sum totalizer (7-digit) and
status messages; each can be set for continuous or sequential display
m3, liter, US gallons per second, minute or hour,
or in user-defined unit, e.g. liter/h or US gallon/day
3
, liter, US gallons or in user-defined unit, e.g. hecto liter or US million
m
gallon (min. 1 year overflow time)
8-character, 7-segment numeral and sign display,
and symbols for key acknowledgement
length up to 10 m / 30 ft (max. 20 m / 60 ft, option)
4.4 Environmental conditions
In accordance with EN IEC 61010-1 the following enviromental conditions have to be observed .
The GFM 700 is designed to operate safe under the following conditions:
a) suitable for indoor and outdoor use, the housing of the signal converter is usable up to Protection Category IP67
b) IP 67 is only warranted when using suitable cabling in the cable glands
c) use up to an altitude of 2000 m above see level
d) suitable for an operation ambient temperature range - 40 …. +60 °C
e) suitable for an storage temperature range -40 …. + 80 °C
f) suitable for use in atmospheres with a relative humidity up op 80%
g) mains supply voltage fluctuations up to + 13 % of the nominal voltage range
h) withstands over voltages up to category II on the main supply voltage ( IEC 60364-4-443)
i) connected to protective earth conductor ( Protection Class I)
j) rated pollution degree 2
9
Page 10
4.5 Dimensions and weights
Flange connections to DIN 2501 / pressure PN, s. Tabelle: dimensions b
GFM 700 K Compact flowmeter GFS 700 S Primary head
GFC 700 F Signal converter GFC 700 F – EEx Signal converter
wall mounting (rotatable) wall mounting (rotatable)
weight approx. 4.2 kg weight approx. 4.2 kg
10
Page 11
p
5 Installation notes
Inlet run:
• downstream of a fan
• downstream of fully open control valve
• downstream of a 90° bend (elbow)
Outlet run:
Installation
conditions
• downstream of a compressure or nozzle
≥ 40 x DN
≥ 30 x DN
≥ 20 x DN
≥ 20 x DN
• downstream of a reducer (α/2 < 4°)
no additional inlet run required
≥ 10 x DN (DN = meter size)
Select position such that the measuring beam is approximately horizontal.
Fit the mating flanges precisely at right angles to the pipeline.
6.1 Electrical connection
• Power supply, power consumption and load rating
of outputs, see “Technical data”
• Current and pulse outputs (I + P) are galvanically
isolated from all input and output circuits.
• Selection of connection diagrams Standard operation EEx operation
Current output 2) passive/active 1) active (also standard)
Pulse output 3) passive 4) passive (also standard)
Power supply and connection terminals 5) 6)
1) Current output active
Standard and EEx
2) Current output passive
Standard
3) Pulse output active
Standard
4) Pulse output passive
Standard and EEx
≤ 680 Ω
R
i
≤ 18 V DC
U
ext.
≤ 680 Ω
R
i
5) Power supply connection and all outputs
standard
binary
LN
oututs
L~
current
output
L~
only internal protective conductor
connection terminal
6.2 Electrical installation to the mains supply voltage
The GFM 700 is intended for permanent connection to the mains. It is required (for example for service) to mount an
external switch or circuitbreaker near the product for disconnection from the mains. It must be reached easily by the
operator and marked as the disconnecting device for this product. The switch or circuitbreaker has to be suitable for the
application and shall also be in accordance with local (safety) requirements and requirements of the building installation.
The protective conductor clamp terminal size M5, press-fitted, in the terminal compartment (Inear the power connection
terminals) shall always be connected to the protective earth conductor of the main supply. Conductors up to 4mm2 can be
connected.
The diameter of the conductors of the main supply, including the protective earth conductor shall to be in accordance with
the general requirements.
= 19 V − 32 V
U
Output
R1 ≥ 650 Ω
I ≤ 50 mA
EC electronic totalizer
≤ 32 V DC ≤ 24 V AC
U
ext.
I ≤ 150 mA
EC electronic totalizer
EMC electromechanical
totalizer
6) Power supply connection and all outputs
EEx version
binary
L N
out
L~
uts
current
output
L~
power supply
protective conductor
terminal
11
Page 12
7 Operation of the signal converter
7.1 Table of settable functions
Fct. Text Description and settings
1.0 OPERATION Main menu 1.0 Operation
1.1.0 BASIC.
PARAM
1.1.1 FULL SCALE Full-scale range for flowrate Q
1.1.2 REV. SCALE Different range for reverse flow
1.1.3 REV. VALUE Full-scale range for reverse flow
1.1.4 ZERO SET. Zero setting, see Fct. 3.1.4
1.2.0 DISPLAY Submenu 1.2.0 Display
1.2.1 DISP. FLOW Unit for flowrate display,
1.2.2 DISP. TOTAL. Function of totalizer display,
1.2.3 UNIT TOTAL. Unit for totalizer display required?
1.2.4 DISP. SP. S. Display of sound velocity required?
1.2.5 CYCL. DISP. Cyclic display required?
1.3.0 CUR.
OUTP. I
1.3.1 TIMECONST.I Time constant of current output I,
1.3.2 L.F.CUTOFF I Low-flow cutoff (SMU) for current
1.3.3 CUTOFF
ON
1.3.4 CUTOFF OFF Cutoff “off” value SMU-I,
1.4.0 FREQ. OUTP.F Submenu 1.4.0
1.4.1 PULSRATE
or
PULSE/UNIT
1.4.2 L.F.CUTOFF F Low-flow cutoff (SMU) for frequency
1.4.3 CUTOFF ON Cutoff “on” value SMU-F,
1.4.4 CUTOFF OFF Cutoff “off” value SMU-F,
2.0 TEST Main menu 2.0 Test functions
2.1 TEST DISP. Carry out display test (Sect. 7.1.1)
2.2 TEST I Test current output I (Sect. 7.1.2)
2.3 TEST F Test frequency output F (Sect. 7.1.3)
2.4
PROCESSOR
Submenu 1.1.0 Basis parameters
,
100%
see Fct. 3.1.1
required? See Fct. 3.1.2
Q
, see Fct. 3.1.3
R100%
see Fct. 3.2.1
see Fct. 3.2.2
See Fct. 3.2.3
See Fct. 3.2.4
See Fct. 3.2.5
Submenu 1.3.0 Current output I
see Fct. 3.3.6
output required? See Fct. 3.3.7
Cutoff “on” value SMU-I,
see Fct. 3.3.8
see Fct. 3.3.9
Frequency output F
Pulse rate for 100% flowrate or for
sound velocity, see Fct. 3.4.3 or
Pulse value for flowrate unit,
see Fct. 3.4.3
output required? See Fct. 3.4.3
see Fct. 3.4.7
see Fct. 3.4.8
Start with → key, duration approx. 30
Sec. Test interruption with ↵ key.
• 0 mA • 4 mA • 20 mA
• 2 mA • 10 mA • 22 mA
Displayed value present directly at
current output. Actual value again
present at output after pressing the ↵
key.
• 1 Hz• 100 Hz
• 10 Hz • 1000 Hz
Displayed value present directly at
current output. Actual value again
present at output after pressing the ↵
key.
Test microprocessor (Sect. 7.1.4)
Start with ↵ key, duration approx. 2
Sec. End of test: NO ERROR or
ERROR displayed.
12
Page 13
Fct. Text Description and settings
3.0 INSTALL Main menu 3.0 Installation
3.1.0
BASIS.PARAM.
3.1.1 FULL SCALE Full-scale range for flowrate Q
Submenu 3.1.0 Basisparameters
select from list
Unit:
100%
under Fct. 3.2.1
Value:
9.5*10-7 - 150.8 m3/Sec or
3.9 - 1,987,200 US Gal/min
After selecting unit, call numerical
value with ↵ key, 1st digit flashes.
3.1.2 REV. SCALE Different range for reverse flow
required? Setting NO or YES
3.1.3 REV. VALUE Full-scale range for reverse flow
(appears only if YES set
under Fct. 3.2.1)
Unit:
select from list
under Fct. 3.2.1
Value:
9.5*10-7 - 150.8 m3/Sec or
3.9 - 1,987,200 US Gal/min
Value must not be larger
than that of Fct. 3.1.1!
After selecting unit, call numerical
value with ↵ key, 1st digit flashes.
3.1.4 ZERO SET Zero setting
• FIXED.VALUE
• VALUE.MEASU.
(Carry out only at “zero” flow and
with completely filled measuring
tube).
1) Inquiry: CALIB. NO or YES
2) if YES: calibration (duration
approx. 20 Sec) with zero display
in PERCENT of Q
100%
3) Inquiry: STORE NO or YES.
3.1.5 METER SIZE Meter size
mm or inches
Unit:
Value:
25 - 1200 mm or
0.98 - 48 inches
After selecting unit, call numerical
value with ↵ key, 1st digit flashes.
3.1.6 GK VALUE Primary head constant GK
(see also primary head nameplate).
Range: 0.5 - 14
3.1.7 FLOW DIR Define direction of forward flow,
see Sect. 5.4.
Setting:
+ or −, acc. to direction of
arrow
on primary head.
3.1.8 MIN SP. S. Minimum sound velocity
Minimum value used for I
or F0%
0%
(when function SOUND.VELO.
selected in 3.3.1 or 3.4.1)
Value:
0 to 5000 m/s
3.1.9 MAX SP. S Maximum sound velocity
Maximum value used for I
or F
100%
100%
(when function SOUND.VELO.
selected in 3.3.1 or 3.4.1)
Value:
1 - 5000 m/s
13
Page 14
Fct. Text Description and settings
3.2.0 DISPLAY Submenu 3.2.0 Display
3.2.1 DISP. FLOW Unit for flowrate
3
• m
/Sec • Liter/Sec
• US Gal/Sec
• m3/min • Liter/min
• US Gal/min
• m3/hr • Liter/hr • US Gal/hr
• h Liter/hr or US.MGal/DAY
(factory set, can be changed
as required, see Fct. 3.6.6,
3.6.7+3.6.8 and Sect. 5.14)
• PERCENT
• NO DISPLAY
3.2.2 DISP. TOTAL Function of totalizer display
• + TOTAL. (forward totalizer)
• −- TOTAL. (reverse totalizer)
• +/− TOTAL. (forward and reverse
totalizers, alternating)
• SUM TOTAL.
(sum of + and −- totalizers)
• ALL TOTAL. (alternating,
sum, + and −-totalizers)
• TOTAL. OFF
(totalizer switched off)
3.2.3 UNIT TOTAL. Unit for totalizer display
3
• m
• Liter • US Gal
• h Liter or US.Mgal
(see Fct. 3.2.1 “hLiter/hr” and
“US.Mgal/DAY”)
3.2.4 DISP. SP. S Display of sound velocity (in m/s)
required?
Setting: NO or YES
3.2.5 CYCL. DISP. Cyclic display required?
3.2.6 ERROR
MSG.
Setting:
Which error messages to be
displayed?
NO or YES
• NO MESSAGES
(no error messages)
• US ERROR
(only ultrasonic errors)
• TOTAL.ERROR
(only errors of internal totalizer)
• ALL ERRORS (all errors)
14
Page 15
Fct. Text Description and settings
3.3.0 CUR.OUTP. I Submenu 3.3.0 Current output I
3.3.1 FUNCTION I Function, current output I
• OFF (switched off)
• F/R IND.
(F/R indication, e.g. for F)
• 1 DIR. (1 flow direction)
• I<I 0 PCT
(Forward / Reverse flow,
e.g. in 0 - 20 mA range:
F=10 - 20 mA and R=10 - 0 mA)
• 2 DIR. (Forward/Reverse flow,
F/R-measurement)
• SP. SOUND (sound velocity)
3.3.2 RANGE I Range for current output I, see
Sect. 5.7.2
• 0 - 20 mA
• 4 - 20 mA
• OTHER RANGE
(see Fct. 3.3.3, 3.3.4 + 3.3.5)
3.3.3 I 0 PCT. Current for 0% flow (I0%)
(appears only if OTHER RANGE
set under Fct. 3.3.2).
Value:
00 to 16 mA
3.3.4 I 100 PCT. Current for 100% flow (I
100%
)
of full-scale range (Fct. 3.1.1)
(appears only if OTHER RANGE
set under Fct. 3.3.2).
Value:
04 - 20 mA
(value must be at least 4 mA
greater than that of Fct. 3.3.4).
3.3.5 I MAX mA Current limitation (I
max
)
see Fct. 5.7.2
(appears only if OTHER RANGE is
set under Fct. 3.3.2)
Value:
0420 mA
(value must be at least 4 mA
greater than that of Fct. 3.3.4).
3.3.6 TIMECONST.I Time constant of current output I
Value:
0.04 - 3600 Sec
3.3.7 L.F.CUTOFF I Low flow cutoff (SMU) for current
output required?
Setting: NO or YES
3.3.8 L.F.CUTOFF
ON
Cutoff “on” value for SMU-I
(appears only if YES set
under Fct. 3.3.7)
Value:
01 - 19 PERCENT of Q
100%
(Fct. 3.1.1)
3.3.9 L.F.CUTOFF
OFF
Cutoff “off” value for SMU-I
(appears only if YES set
under Fct. 3.3.7)
Value:
02 - 20 PERCENT of Q
100%
(Fct. 3.1.1), value must be greater
than that of Fct. 3.3.8.
15
Page 16
Fct. Text Description and settings
3.4.0 FREQ.
OUTP.F
Submenu 3.4.0 Frequency output
F
3.4.1 FUNCTION F Function, frequency output F
• OFF (switched off)
• F/R IND.
(F/R indication, e.g. for I)
• 1 DIR. (1 flow direction)
• 2 DIR. (forward / reverse flow,
F/R measurement)
• SP. SOUND (sound velocity)
3.4.2 PULSOUTP Unit of frequency output F
• PULSRATE
(setting in pulses per unit time)
• PULSE/UNIT (setting in pulses
per unit volume)
3.4.3 PULSRATE Pulse rate for 100% flowrate or for
sound velocity, see Fct. 3.1.1 or
3.1.8 + 3.1.9 (appears only if
PULSRATE set under Fct. 3.4.2)
Value:
2.778*10
-3
- 1000 PuLSe/Sec (= Hz)
or 0.1667 - 60,000 PuLSe/min
or 10 - 3,600,000 PuLSe/hr
After selecting unit, call numerical
value with ↵ key, 1 digit flashes.
3.4.3 PULSE/UNIT Pulse value for flowrate unit
(appears only if PULSE/UNIT set
under Fct. 3.4.2)
Unit:
PuLSe per m3, Liter, US Gal or
unit of Fct. 3.6.6, 3.6.7+3.6.8
Value:
0.0001 to 9.9999*10
(no check, but Q
9
PuLSe
pulse value
100%
*
≤ 3.600.000 pulses/hr).
After selecting unit, call numerical
value with ↵ key, 1 digit flashes.
3.4.4 PULSWIDTH Pulse width for frequenc ≤ 10 Hz
• 30 mSec • 200 mSec
• 50 mSec • 500 mSec
• 100 mSec
3.4.5 TIMECONST.F Time constant
of frequency output F
• 40 mSec
• SAME AS I
(time constant for F same as for I,
see Fct. 3.3.6)
3.4.6 L.F.CUTOFF F Low-flow cutoff (SMU) for frequency
output required?
Setting:
NO or YES
3.4.7 CUTOFF ON Cutoff “on” value SMU-F
(appears only if YES set
under Fct. 3.4.6)
Value:
01 - 19 PERCENT of Q
100%
(Fct. 3.1.1)
3.4.8 CUTOFF
OFF
Cuttoff “off” value SMU-F
(appears only if YES set
under Fct. 3.4.6)
Value:
02 - 20 PERCENT of Q
100%
(Fct. 3.1.1),, value must be greater
than of Fct. 3.3.8
16
Page 17
Fct. Text Description and settings
3.5.0 USER DATA Submenu 3.5.0 User data
3.5.1 LANGUAGE Language for display texts, see
Sect. 5.11
• GB/USA (English)
• D (german)
• F (French)
• N (Dutch)
3.5.2 ENTRY.CODE.1 Entry code 1 for setting level
required?
See Sect. 5.12
• NO = Entry with key only
• YES = Entry with key and
9-keystroke code
Setting of the code under Fct. 3.5.3
3.5.3 CODE 1 Set Code 1, see Sect. 5.12
(appears only if YES set under Fct.
3.6.2)
• Factory setting:
→, →, →, ↵, ↵. ↵, ↑, ↑, ↑
• Different code required:
Press any 9-keystroke combination
and then press the same
combination again.
Each keystroke acknowledged by
“
”.
*
WRONG CODE appears if 1
nd
entries are not equal.
2
st
and
Press ↵ + → keys
and repeat entries.
3.5.4 LOCATION
Tag name setting
(measurement point no.)
max. 10 digits. Required only
for flowmeters of “HHC” design
(operator control via Hand-Held
Communicator MIC 500, connected
to current output).
Factory setting:
ALTOMETER
Characters assignable to each
place: A..Z / a..z / 0..9 / + / -./
underscore character =
blank character.
3.5.5 OUTP. HOLD Hold values of outputs
during settings?
Setting:
NO or YES
3.5.6 UNIT TEXT Text for user-defined unit
Factory setting:
hLiter/hr or US.MGal/DAY
Characters assignable
to each place:
A..Z / a..z / 0..9 / + / −-./
underscore
character = blank character.
Fraction bar ”/” in 7
th
place is
unalternable.
3.5.7 FACT. QUANT Conversion factor for quantity FM
F
= quantity per 1 m3!
M
Factory setting:
1.00000 E1 (for hecto Liter)
or 2.64172 E-4 (for US M.gallons)
Value setting:
0.00001*10
9
−
- 9.99999*10+9
3.5.8 FACT. TIME Conversion factor for time FT
F
in seconds!
T
Factory setting:
3.60000 E3 (for hour) or
8.64000 E4 (for day)
Value setting:
0.00001*10
9
−
- 9.99999*10+9
17
Page 18
Fct. Text Description and settings
3.5.9 TOTAL.
RESET
3.5.10 ENABL.
RESET
3.5.11 PLAUSIB
ERR.
3.5.12 WEIGHT
P.OK
3.5.13 N.ER.PLAU
SIB.
Totalizer reset (+ and −- totalizer
together)
Inquiry: NO or YES
Enable totalizer reset
for RESET/QUIT menu.
Inquiry: NO or YES
Error limit in % of measured value for
plausibility statement. Measured
values that are outside the specified
band are not processed. Every
measured value outside the specified
band will increase an internal counter
by “1”, until a maximum counter
value (see Fct. 3.6.13) has been
reached. The corresponding
measurement channel will then be
made inactive and an indication will
be visible on the display.
Value setting: 1 - 99 PERCENT
Factory setting: 20 PERCENT
Weight factor for correct
measurements. The internal
plausibility counter is increased by
the number programmed, when the
measured value is correct. The
higher the number the faster an
inactive channel will become active
again.
Value setting: 1 - 50
Factory setting: 4
Limit value for the counter of
incorrectly measurements (see Fct.
3.6.11). When ‘0’ is set, the
plausibility function will become
inactive.
Value setting: 0 - 10,000
Factory setting: 0
18
Page 19
7.2 Error messages
Fct. Text Description and settings
4.0 PARAM.
ERROR
4.1.0 FLOW
VELOC.
4.1.1 FULL SCALE Full-scale range for flowrate Q
4.1.2 METER SIZE Meter size, see Fct. 3.1.5
4.2.0 F/R FLOW FULL-SCALE RANGE(S) for
4.2.1 FULL SCALE Full-scale range for forward
4.2.2 REV. SCALE Different range for reverse flow
4.2.3 REV. VALUE Full-scale range for reverse flow
4.3.0 I RANGE CURRENT OUTPUT I RANGE
4.3.1 I 0 PCT Current for 0% flow (I0%), see Fct.
4.3.2 I 100 PCT Current for 100% flow (I
4.4.0 I MAXIMUM CURRENT LIMITATION
4.4.1 I 100 PCT Current for 100% flow (I
4.4.2 I MAX mA Setting of max. output current
4.5.0 LFC. I RANG. LOW-FLOW CUTOFF RANGE I
4.5.1 L.F. CUTOFF I Low-flow cutoff (SMU) for current
4.5.2 CUTOFF ON Cutoff “on” value SMU-I, see Fct.
4.5.3 CUTOFF
OFF
4.6.0 LFC. F
RANG.
4.6.1 L.F. CUTOFF F Low-flow cutoff (SMU) for
4.6.2 CUTOFF ON Cutoff “on” value SMU-F,
4.6.3 CUTOFF
OFF
Main menu 4.0 Parameter error
FLOW VELOCITY “v” incorrect:
Ensure condition
0.5 m/s ≤ v ≤ 60 m/s
or 1.5 ft/s ≤ v ≤ 180 ft/s is met!
see Fct. 3.1.1
100%
forward/reverse flow incorrect:
Ensure condition F ≥ R is met!
LOW-FLOW CUTOFF RANGE F
incorrect:
Ensure condition cutoff “off”.
Cutoff “on” ≥ 1% is met!
frequency output required?
See Fct. 3.4.6
see Fct. 3.4.7
Cutoff “off” value SMU-F,
see Fct. 3.4.8
,
19
Page 20
Fct. Text Description and settings
4.7.0 F > 1 kHz OUTPUT FREQUENCY too high:
must be less than 1 kHz!
4.7.1 FULL SCALE Full-scale range for flowrate Q
100%
see Fct. 3.1.1
4.7.2 PULSOUTP. Unit of frequency output F,
see Fct. 3.4.2
4.7.3 PULSRATE
or
PULSE/UNIT
Pulse rate for 100% flowrate or
for sound velocity,
see Fct. 3.4.3 or
Pulse value for flowrate unit,
see Fct. 3.4.3
4.8.0 F <> PULSW. FREQUENCY/PULSE WIDTH
ASSIGNMENT is incorrect
4.8.1 PULSOUTP. Unit of frequency output F,
see Fct. 3.4.2
4.8.2 PULSRATE
or
PULSE/UNIT
Pulse rate for 100% flowrate or
for sound velocity, see Fct 3.4.3
or Pulse value for flowrate unit,
see Fct. 3.4.3
4.8.3 PULSWIDTH Pulse width for frequencies
≤ 10 Hz, see Fct. 3.4.4
4.9.0 PULS/T.TIME Incorrect ASSIGNMENT of UNIT
for F and SOUND VELOCITY
4.9.1 FUNCTION F Function of frequency output F,
see Fct. 3.4.1
4.9.2 PULSOUTP. Unit of frequency output F,
see Fct. 3.4.2
4.10.0 LFC.
I/T.TIME
LOW-FLOW CUTOFF I incorrect:
Ensure low-flow cutoff is “off”
when function of current output
is sound velocity.
4.10.1 FUNCTION I Function of current output I,
see Fct. 3.3.1
4.10.2 L.F.CUTOFF I Low-flow cutoff (SMU) for current
output required? See Fct. 3.3.7
4.11.0 LFC.
F/T.TIME
LOW-FLOW CUTOFF F incorrect:
Ensure low-flow cutoff is “off”
when function of frequency
output is sound velocity.
4.11.1 FUNCTION F Function of frequency output F,
see Fct. 3.4.1
4.11.2 L.F.CUTOFF F Low-flow cutoff (SMU) for
frequency output required?
See Fct. 3.4.6
4.12.0 V.S.
min>max
MAX. SOUND VELOCITY MUST
BE LARGER THAN MIN. SOUND
VELOCITY.
4.12.1 MIN SP. S Minimum sound velocity,
sound velocity for I
or F0%
0%
4.12.2 MAX SP. S Maximum sound velocity,
sound velocity for I
100%
or F
100%
,
20
Page 21
8. Addition for the GFM 700 F-EEx ATEX and GFM 700 HT EEx-ATEX
8.1 Safety instructions
This product is designed for use in accordance with EN IEC 61010-1 for Installation
Category 2 and Pollution Degree 2. Hazardous voltages are present within this product
during normal operation. The product is designed for Protection Class I and should
never be operated without protective earthing. The product shall also never be
operated with covers removed unless equivalent protection of the operator and its
environment from accidental contact with hazardous internal voltages is provided.
Always follow basic and local safety precautions when using this product to reduce
risk of injury from electrical shock, spread of fire or other dangerous situations.
WARNING !
No changes may be made to the devices. Unauthorized changes might affect the explosion
safety of the devices.
These additional instructions are an extension to the standard Installation and Operating
Instructions and only applies for the EEx version of the GFM 700 F gas flowmeter. All technical
information described in the standard Installation and Operating Instructions are applicable,
when not specifically excluded or replaced by the instructions in these additional instructions.
Be sure to follow these instructions !
8.2 SYSTEM COMPONENTS
8.2.1 General information
The Altosonic gas flowmeter system GFM 700 F-EEx consists of the ultrasonic gas flow sensor GFS 700 F-EEx
or GFS 700 F/HT-EEx (high-temperature version) in combination with the ultrasonic gas flow converter GFC 700
F-EEx. Both the sensor and the converter can be installed in a hazardous location of Zone 1 or 2. The schematic
block diagram of the gas flowmeter system is given in Figure 1 (see Section 4 Connection diagram for details).
Figure 1:
Schematic block diagram of the GFM 700 F-EEx gas flowmeter.
GFS 700 F-EEx
GFS 700 F/HT-EEx
gas flow sensor
Important
The prescriptions and regulations as well as the electrical data described in the EG certificate of
conformity must be obeyed.
Beside the instructions for electrical installations in non-hazardous locations according to the applicable
National standard (equivalent to IEC 364, i.e. VDE 0100) especially the regulations in EN 60079-14
“Electrical installations in hazardous locations” or equivalent national standard (e.g. DIN VDE 0165) must
be followed.
Installation, establishment, utilization and maintenance are only allowed to be executed by personnel
with an education in explosion safety.
Hazardous locations of Zone 1 or 2
or
MR04 connecting cable
Equipotential bonding
GFC 700 F-EEx
gas flow converter
Signal
outputs
Power
supply
21
Page 22
8.2.2 Gas flow sensor
There are two types of gas flow sensors, namely the GFS 700 F-EEx (standard version) and the GFS 700 F/HTEEx (high-temperature version). The GFS 700 F/…-EEx gas flow sensors are in conformance with the European
Directive 94/9 EG (ATEX 100a) and approved by KEMA according to the European Standards EN 50xxx under
the following number.
KEMA 00 ATEX 2119 X
The GFS 700 F/…-EEx gas flow sensors have type of protection flameproof enclosure "d" according to EN 50018
(measuring sensors) and increased safety "e" according to EN 50019 (junction box). Both types of gas flow
sensors are designed for ambient temperatures (i.e. Ta) in the range of -40°C to +60°C. The maximum
permissible temperature of the gas(es) to be measured is restricted by the temperature class of the explosive
atmosphere that (possibly) surrounds the gas flow sensor, see therefore Table 1 and 2 below.
Temperature class
T6 80°C
T5 95°C
T4 130°C
T3 180°C
Table 1: Temperature classification of GFS 700 F-EEx (standard).
Table 2: Temperature classification of GFS 700 F-EEx (standard).
The GFS 700 F-EEx gas flow sensor is marked with the code II 2G EEx de IIC T6…T3, the high-temperature
version GFS 700 F/HT-EEx bears code II 2G EEx de IIC T6…T1. See also the respective Certificate of
Conformity in section 5.1 of these additional instructions.
Maximum gas temperature at Ta ≤ 60°C
Maximum gas temperature at Ta ≤ 60°C
8.2.3 Gas flow converter
The GFC 700 F-EEx gas flow converter also conforms to the European Directive 94/9 EG (ATEX 100a) and is
approved by KEMA in accordance with the European Standards of the EN 50xxx series. The GFC 700 F-EEx has
the following approval number.
KEMA 00 ATEX 2118 X
The GFC 700 F-EEx ultrasonic gas flow converter has type of protection flameproof enclosure "d" to EN 50018
(electronics compartment and optional for terminal compartment) and/or increased safety "e" according to EN
50019 (standard for terminal compartment). The flow converter is suitable for ambient temperatures in the range
of -40 to +60°C and is marked with temperature classification T6, because a maximum surface temperature of
80°C (85°C minus 5 K safety margin) is not exceeded.
The GFC 700 F-EEx is marked with code II 2G EEx d IIC T6 for the terminal compartment in type of protection "d"
or with code II 2G EEx de IIC T6 (terminal compartment in "e"). The Certificate of Conformity is listed in section
5.2 of these additional instructions.
22
Page 23
8.3 ELECTRICAL CONNECTION
8.3.1 General
The GFC 700 F-EEx gas flow converter is always connected to one of two in Section 1.2 (see previous page)
described GFS 700 F/…-EEx ultrasonic gas flow sensors via the factory installed MR04 type connecting cable.
Both apparatus must also be incorporated into the equipotential bonding system via respectively the external M5
clamp terminal that is pressed into the connecting flange at the bottom of the signal converter housing (of GFC
700 F-EEx) and the M8 terminal on the warning plate of the GFS 700 F/…-EEx gas flow sensor.
8.3.2 Gas flow converter
The MR04 connecting cable runs into the flameproof ("EEx d") electronics compartment through an "EEx d"
approved cable gland. The cable is factory installed and tightly clamped by this gland. The gland may under no conditions be (re)assembled by the customer, because it could damage the flameproof properties of the
electronics compartment.
The field cables that enter the terminal compartment of the GFC 700 F-EEx flow converter (power supply, current
output and binary outputs) are non-intrinsically safe. For connection of measuring devices to the output
terminals, the wiring requirements for type of protection of the compartment (standard: increased safety "e",
option: flameproof enclosure "d") must be according to the international or national standard (e.g. DIN VDE 0165,
section 5.6) involved.
For flameproof conduit systems, the terminal compartment must have type of protection flameproof enclosure "d"
according to EN 50018. The conduits must be sealed by "EEx d" approved (within the ATEX 100a directive)
sealing devices (i.e. stopping box) directly at the conduit entrances of the as flameproof enclosure performed
terminal compartment.
The arrangement of the terminals in the terminal compartment is shown by Figure 2 below.
Figure 2:
Arrangement of terminals.
Pulse Status
output output
B1 B- B2 I+ I L N
L
BINARY CURRENT
OUTPUTS OUTPUT
L
Protective conductor terminal
NOTE:
The status output is not supported by the current software yet, but will be supported by future software
versions. The current and pulse output are galvanically insulated from each other as well as from all other in-
/output circuits
.
23
Page 24
I
≤ 150 mA
B1B- B2
U
000
ext
≤ 32V
24 V
DC
AC
Ri ≤ 680
Ω
I+I
+
mA
_
Electronic or
electro-mechanical
totalizer
Figure 3: Passive pulse output. Figure 4: Active current output.
NOTE:
The status output (terminals B1 and B-) can only be configured as passive output, the current output
(terminals I+ and I) can only be configured as active output.
See page 6 of the standard Installation and Operating Instructions for the electrical data of the available power
supplies (e.g. mains voltages, etc.).
8.3.3 Replacement of electronics unit or power fuse(s)
IMPORTANT !
The following instructions must be followed carefully if the GFC 700 F-EEx gas flow converter housing has to
be opened respectively closed !
Before opening:
♦ Make absolutely sure that there is no explosion hazard !
♦ Gas-free certificate !
♦ Make sure that all connecting cables are safely isolated from the power supply !
♦ Allow the prescribed waiting time to elapse before opening the housing:
20 minutes for temperature class T6
11 minutes for temperature class T5
When the instructions above are strictly followed, the cover (includes a glass window) of the electronics
compartment may be removed. First unscrew the recessed head screw of the interlocking device by a hollowhead screw wrench size 3, until the cover can rotate freely. Unscrew the cover with the special plastic wrench
(black) that is supplied with the apparatus.
After opening:
♦ The copper grounding strip at the back of the electronics unit must be securely screwed to the housing (back-
end of electronics compartment) by screw C (see Figure 5). The electronics unit is mounted to the back-end of
electronics compartment by two screws B. Before screws B and C can be accessed, the display unit must be
removed via screws A.
♦ Before the cover is screwed back into the housing, the screw-thread must be clean and well-greased with
an acid and resin-free grease, e.g. silicone grease.
♦ Screw the cover as tight as possible into the housing by hand, until it cannot be opened by hand anymore.
Screw the recessed head screw of the interlocking device tight.
24
Page 25
A
B
Copper earth strip
Display unit
Flat cable of
display unit
C
A
Figure 5: Display unit (left) / Electronics unit after removal of display unit (right).
B
8.3.4 Replacement of electronics unit
See the standard Installation and Operating Instructions for detailed information about resetting and
reprogramming the new electronics unit after replacement. Important customer specific data (like the value of the
internal totalizer) should be noted before replacing the electronics unit !
Before commencing work, note the instructions in Section 2.3 ("Before opening").
Then continue as follows:
1. Remove the cover of the electronics compartment.
2. Unscrew the two screws A of the display unit and turn display unit carefully aside or remove the unit
completely by taking out the flat cable connector.
3. Unscrew the two mounting screws B of the electronics unit as well as screw C, which fixes the copper earth
strip at the back of the housing. A screwdriver with a long shaft (200 mm) can best be used to unscrew C
(e.g. screwdriver type Philips No. 2).
4. Pull the electronics unit carefully out of the converter housing, till the SMB connectors of the coaxial cables
can be unplugged easily. Then remove the complete electronics unit.
5. Check the new electronics unit if the voltage setting (only applicable for AC supplies) and power fuse rating
are correct. Change the voltage setting (see section 2.3.3) or exchange the fuse if necessary.
6. Carefully insert the electronics unit till the numbered SMB connectors can be connected to the corresponding
numbered SMB receptacles on the electronics unit. Then mount the unit completely into the housing and fix
the screws. First C, then B and finally screw the display unit on the electronics via screws A, after the flat
cable connector is connected.
7. Screw the cover of the electronics compartment back into the housing.
Note the instructions of section 2.3 ("After opening") during reassembling.
IMPORTANT !
Carefully keep the coaxial cables to the side of the housing, while inserting or removing the electronics unit
into respectively from the converter housing. This is to prevent damaging of the coaxial cables !
25
Page 26
8.3.5 Replacement of power fuse(s)
a) AC versions 115/230 VAC and 100/200 VAC
Before commencing work, note the instructions in Section 2.3 ("Before opening").
Then continue as follows:
1. Remove the cover of the electronics compartment.
2. Unscrew the two screws A of the display unit and turn the display unit carefully aside.
3. The fuse-holder in which the power fuse is mounted is accessible now to exchange the defect power fuse F1
for one with the same rating. The rating depends on the voltage of the mains (T200mA for 100/115VAC and
T125mA for 200/230 VAC setting). See also the yellow sticker on the mains transformer as shown in Figure 6
below.
4. Reassemble in reverse order (points 2 and 1).
Note the instructions of section 2.3 ("After opening") during reassembling.
Mains transformer
AC
115/230 V
version
Sticker with fuse rating
Mains fuse F1 (in fuse-holder)
Voltage selector
Indication of
voltage selector
(black dot = notch)
SIDE OF DISPLAY UNIT
Figure 6: Power supply 115/230 VAC.
b) 24 VAC/DC version
Before commencing work, note the instructions in Section 2.3 ("Before opening").
Then continue as follows:
1. Remove the cover of the electronics compartment.
2. Unscrew the two screws A of the display unit and disconnect the display unit via the flat cable connector (see
right picture of Figure 5 on the previous page).
3. Unscrew the copper earth strip (screw C) with the and the mounting screws B of the electronics unit. Pull the
unit out until the SMB connectors of the coaxial cables can be disconnected from the electronics. Then take
out the complete electronics unit. Be careful with the coaxial cables, so that they do not damage while
removing the electronics unit from the flow converter housing. Continued on the next page !
26
Page 27
4. The power fuses F1 and F2 (see Figure 7 below) can be replaced now. The 24 VAC/DC power supply uses
two sub-miniature fuses of T1.25A in accordance with IEC 127-3.
5. Reassemble in reverse order (points 3 through 1).
Note the instructions of section 2.3 ("After opening") during reassembling.
Location of power fuses
(see detail in Figure 8)
Figure 7: GFC 700 electronics unit with 24 VAC/DC power supply.
F2
F1
Figure 8: Location of fuses F1, F2 on 24 VAC/DC unit.
27
Page 28
8.3.6 Changing power supply voltage (not for 24 VAC/DC version)
Before commencing work, note the instructions in Section 2.3 ("Before opening").
Then continue as follows:
1. Remove the cover of the electronics compartment.
2. Unscrew the two screws A of the display unit and turn display unit carefully aside or remove the unit
completely by taking out the flat cable connector.
3. Unscrew the two mounting screws B of the electronics unit as well as screw C, which fixes the copper earth
strip at the back of the housing. A screwdriver with a long shaft (200 mm) can best be used to unscrew C
(e.g. screwdriver type Philips No. 2).
4. Pull the electronics unit carefully out of the converter housing, till the SMB connectors of the coaxial cables
can be unplugged easily. Then remove the complete electronics unit.
5. The voltage setting of the power supply can be changed by turning the dummy dual-in-line block (i.e. voltage selector, see Figure 6 on page 6) over 180° in its socket. The position of the notch on the dummy dual-in-line
block indicates the voltage setting. Also see the sticker that is mounted on the mains transformer (see Figure
6).
6. Carefully insert the electronics unit back into the housing until the numbered SMB connectors can be
connected to the corresponding numbered SMB receptacles on the electronics unit. Then mount the unit
completely into the housing and fix the screws. First C, then B and finally screw the display unit on the
electronics via screws A, after the flat cable connector is connected.
7. Screw the cover of the electronics compartment back into the housing.
Note the instructions of section 2.3 ("After opening") during reassembling.
IMPORTANT !
Carefully keep the coaxial cables to the side of the housing, while inserting or removing the electronics unit
into respectively from the converter housing. This is to prevent damaging of the coaxial cables !
8.3.7 Gas flow sensor
The gas flow sensor is available in two designs, the GFS 700 F-EEx (standard) for gas temperatures up to a
maximum of 180°C and the GFS 700 F/HT-EEx (high-temperature) for higher gas temperatures. Both types are
connected to a GFC 700 F-EEx gas flow converter via a MR04 connecting cable (i.e. four RG 179 B/U coaxial
cables with additional insulation). Each coaxial cable is marked by a number "1" to "4". The core of each cable is
connected to a "EEx e" approved terminal in the terminal box of the flow sensor. The core of the cable with
number "1" is connected to the terminal marked with number "21", the screen is connected to the adjoining
terminal with number "1". Core number "2" is connected to terminal number "22", etcetera. An equipotential
bonding conductor must interconnect the flow sensor to the gas flow converter (see section 2.5 below).
8.3.8 Equipotential bonding system
The equipotential bonding conductor consists of a insulated copper wire with a minimum cross-sectional area of 4
2
mm
(AWG 10). One side of the wire is connected to the external M5 clamp terminal on the connecting flange at
the bottom of the converter housing, the other end is screwed to the M8 bolt/nut terminal on the warning plate of
the gas flow sensor. This end of the cable is therefor provided with a eye cable tag.
8.3.9 Maximum length of connecting cable
The maximum length of the connecting MR04 cable between the GFC 700 F-EEx gas flow converter and the GFS
700 F/…-EEx gas flow sensor is standard limited at 10 m (30 ft) due to measurement technical reasons. Longer
cables on request, but must be evaluated first.
28
Page 29
8.4 CONNECTING CABLES
NOTE:
The below described cables are shown in the connection diagram on the following page.
Cable A:
Quad coaxial cable
Type MR04 (to be supplied by Krohne Altometer)
Technical data:
Test voltage ≥ 500 V
Diameter of strand (core and screen) ≥ 0.1 mm
Distributed capacitance (core/screen) 67 pF/m
Distributed inductance (core/screen) 0.4 µH/m
Cable B:
Power supply cable
Cable parameters must be in accordance with the regulations in the EN 60079-14 "Electrical installations in
hazardous locations" or an equivalent national standard (e.g. DIN VDE 0165).
Rated voltage ≥ 500 V
Examples: H07..-., H05..-.
Cable C:
Signal cable for current output and binary outputs (pulse and status output).
Cable parameters must be in accordance with the regulations in the EN 60079-14 "Electrical installations in
hazardous locations" or an equivalent national standard (e.g. DIN VDE 0165).
Bonding conductor:
Minimum cross-sectional area 4 mm
2
(AWG 10)
29
Page 30
8.5 CONNECTION DIAGRAM
"EEx d" cable gland
"EEx e" junction box
(incl. terminals)
S1 = gas sensor 1
S2 = gas sensor 2
S3 = gas sensor 3
S4 = gas sensor 4
Gas sensors have type
of protection flameproof
enclosure "EEx d"
30
Page 31
9 ATEX approvals
9.1 GFC 700 F - EEx Signal converter
31
Page 32
32
Page 33
33
Page 34
GFS 700 F - EEx Primary head (flowmeter)
9.2
34
Page 35
35
Page 36
36
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.