March 2014
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Introduction
The GPIB to USB adapter is used to connect a GPIB controller to the USB B receptacle on a 2260B power supply.
The GPIB to USB adapter is currently only supported on this instrument.
Features:
•
Enables GPIB control of the instrument via the USB B receptacle.
•
No power adapter, all power requirements from USB.
•
The GPIB primary address can be assigned via the 2260B power supply.
Accessories:
•
User manual, USB type A-B cable.
Item Description
GPIB port 24 pin female GPIB port.
USB type A receptacle connects to the 2260B type B receptacle.
Ready indicator
The ready indicator lights when the adapter is ready to use.
Connection procedure
This section describes how to set the GPIB to USB adapter for use with the power supply.
1. Connect a GPIB cable from a GPIB controller to the GPIB port on the adapter.
2. Connect the USB cable from the rear panel device port (type B port) on the 2260B to the USB port (type A
port) on the GPIB to USB adapter.
¨
2260B-GPIB-USB GPIB to USB Adapter
Operation Instructions
USB type A receptacle
Figure 1

2260B-GPIB-USB GPIB to USB Adapter
2 March 2014
3. The Ready Indicator should be lit after the USB cable is connected.
Configure GPIB address from the 2260B
1. Connect a GPIB cable from a GPIB controller to the GPIB port on the adapter.
2. The display will show F-01 on the top and the configuration setting for F-01 on the bottom.
3. Rotate the voltage knob to change the F setting to F-22 (USB Mode).
4. Rotate the current knob to set the F-22 setting to 1 (Mode: GPIB-USB adapter).
5. Press the Voltage knob to save the configuration setting. ConF will be displayed when the settings are
successfully saved.
6. Rotate the voltage knob to change the F setting to F-23 (GPIB address).

2260B-GPIB-USB GPIB to USB Adapter
March 2014 3
7. Rotate the current knob to set the F-23 setting between 0 and 30. (GPIB address: 0~30)
8. Press the Voltage knob to save the GPIB address. ConF will be displayed when the GPIB address is
successfully saved.
9. Press the Function key again to exit the configuration settings. The function key light will turn off.
Trouble shooting
If the Ready indicator will not come on:
•
Check that the cables are correctly inserted.
•
Ensure the USB device powering the adapter is turned on and is functioning correctly.
Verify the connection
To check that the GPIB connection is working, National Instruments Measurement & Automation Explorer (MAX)
can be used. The following function check is based on version 5.0.0f1.
For further information about National Instruments, refer to the NI website at www.ni.com.
1. Complete the setup procedure described previously to configure the 2260B for GPIB.
2. Start the Measurement and Automation Explorer (MAX) program. Using Windows, press;
Start>All Programs>National Instruments>Measurement & Automation
NOTE
•
Only the 2260B can be used to change the GPIB address from the default GPIB address of 1.
•
GPIB secondary addresses are not supported.