Keithley 2260B-80-13, 2260B-250-4, 2260B-250-9, 2260B-30-108, 2260B-30-72 Adjustment Manual

...
www.tek.com/keithley
Series 2260B
Multi-Range Programmable DC Power Supplies
Verification and Adjustment Manual
077104802 / November 2016
*P077104802*
077104802
A Greater Measure of Confidence
Series 2260B
Programmable Power Supplies
Verification and Adjustment Manual
© 2016, Keithley Instruments
Cleveland, Ohio, U.S.A.
All rights reserved.
Any unauthorized reproduction, photocopy, or use of the information herein, in whole or in part,
without the prior written approval of Keithley Instruments is strictly prohibited.
All Keithley Instruments product names are trademarks or registered trademarks of Keithley Instruments. Other brand names are trademarks or registered trademarks of their respective
holders.
Microsoft, Visual C++, Excel, and Windows are either registered trademarks or trademarks of
Microsoft Corporation in the United States and/or other countries.
Document number: 077104802 / November 2016

Safety precaut ions

The following safety precauti o ns shou ld be observ ed befor e usin g th is produ ct and any associa t ed inst rum enta tion. Although some instruments and accessories would normally be used with nonhazardous voltages, there are situations where hazardous conditions may be present.
This product is intended for use by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read and follow all installation, operation, and maintenance information carefully before using the product. Refer to the user documentation for complete product specifications.
If the product is used in a manner not specified, the protection provided by the product warranty may be impaired. The types of product users are:
Responsible body is the individual or group responsible for the use and maintenance of equipment, for ensuring that the equipment is operated within its specifications and operating limits, and for ensuring that operators are adequately trained.
Operators use the product for its intended function. They must be trained in electrical safety procedures and proper use of the instrument. They must be protected from electric shock and contact with hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it operating properly, for example, setting the line voltage or replacing consumable materials. Maintenance procedures are described in the user documentation. The procedures explicitly state if the operator may perform them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe installations, and repair products. Only properly trained service personnel may perform installation and service procedures.
Keithley Instruments products are designed for use with electrical signals that are measurement, control, and data I/O connections, with low transient overvoltages, and must not be directly connected to mains voltage or to voltage sources with high transient overvoltages. Measurement Category II (as referenced in IEC 60664) connections require protection for high transient overvoltages often associated with local AC mains connections. Certain Keithley Instruments measuring instruments may be connected to mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and instrument labels, do not connect any instrument to mains. Exercise extreme caution when a shock hazard is present. Lethal voltage may be present on cable connector jacks or test
fixtures. The American National Standards Institute (ANSI) states that a shock hazard exists when voltage levels greater than 30 V RMS, 42.4 V peak, or 60 VDC are present. A good safety practice is to expect that hazardous voltage is present in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times. The responsible body must ensure that operators are prevented access and/or insulated from every connection point. In some cases, connections must be exposed to potential human contact. Product operators in these circumstances must be trained to protect themselves from the risk of electric shock. If the circuit is capable of operating at or above 1000 V, no conductive part of the circuit may be exposed.
Do not connect switching cards directly to unlimited power circuits. They are intended to be used with impedance-limited sources. NEVER connect switching cards directly to AC mains. When connecting sources to switching cards, install protective devices to limit fault current and voltage to the card.
Before operating an instrument, ensure that the line cord is connected to a properly-grounded power receptacle. Inspect the connecting cables, test leads, and jumpers for possible wear, cracks, or breaks before each use.
When installing equipment where access to the main power cord is restricted, such as rack mounting, a separate main input power disconnect device must be provided in close proximity to the equipment and within easy reach of the operator.
For maximum safety, do not touch the product, test cables, or any other instruments while power is applied to the circuit under test. ALWAYS remove power from the entire test system and discharge any capacitors before: co nnecting or disconnecting cables or jumpers, installing or removing switching cards, or making internal changes, such as installing or removing jumpers.
Do not touch any object that could provide a current path to the common side of the circuit under test or power line (earth) ground. Always make measurements with dry hands while standing on a dry, insulated surface capable of withstanding the voltage being measured.
For safety, instruments and accessories must be used in accordance with the operating instructions. If the instruments or accessories are used in a manner not specified in the operating instructions, the protection provided by the equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories. Maximum signal levels are defined in the specifications and operating information and shown on the instrument panels, test fixture panels, and switching cards.
When fuses are used in a product, replace with the same type and rating for continued protection against fire hazard. Chassis connections must only be used as shield connections for measuring circuits, NOT as protective earth (safety ground)
connections. If you are using a test fixture, keep the lid closed while power is applied to the device under test. Safe operation requires the use
of a lid interlock.
screw is present, connect it to protective earth (safety ground) using the wire recommended in the user documentation.
If a
The
symbol on an instrument means caution, risk of danger. The user must refer to the operating instructions located in the
user documentation in all cases where the symbol is mark ed on the instr u ment .
The
symbol on an instrument means caution, risk of electric shock. Use standard safety precautions to avoid personal
contact with these voltages.
The
The
If this
symbol on an instrument shows that the surface may be hot. Avoid personal contact to prevent burns.
symbol indicates a connection terminal to the equipment frame.
symbol is on a product, it indicates that mercury is present in the display lamp. Please note that the lamp must be
properly disposed of according to federal, state, and local laws. The WARNING heading in the user documentation explains dangers that might result in personal injury or death. Always read
the associated information very carefully before performing the indicated procedure. The CAUTION heading in the user documentation explains h az ards that coul d dama ge the instrument. Such damage may
invalidate the warranty. Instrumentation and accessories shall not be conne cted to human s. Before performing any maintenance, disconnect the line cord and all test cables. To maintain protection from electric shock and fire, replacement components in mains circuits — including the power
transformer, test leads, and input jacks — must be purchased from Keithley Instruments. Standard fuses with applicable national safety approvals may be used if the rating and type are the same. The detachable mains power cord provided with the instrument may only be replaced with a similarly rated power cord. Other components that are not safety-related may be purchased from other suppliers as long as they are equivalent to the original component (note that selected parts should be purchased only through Keithley Instruments to maintain accuracy and functionality of the product). If you are unsure about the applicability of a replacement component, call a Keithley Instruments office for information.
Unless otherwise noted in product-specific literat ure , Kei thle y Instruments instruments are designed to operate indoors only, in the following environment: Altitude at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F); and pollution degree 1 or 2.
To clean an instrument, use a damp cloth or mild, water-based cleaner. Clean the exterior of the instru ment only . Do not apply cleaner directly to the instrument or allow liquids to enter or spill on the instrument. Products that consist of a circuit board with no case or chassis (e.g., a data acquisition board for installation into a computer) should never require cleaning if handled according to instructions. If the board becomes contaminated and operation is affected, the board should be returned to the factory for proper cleaning/servicing.
Safety precaution revision as of March 2016.

Table of Contents

Introduction ............................................................................................................... 1-1
Welcome .............................................................................................................................. 1-1
Extended warranty ............................................................................................................... 1-1
Contact information .............................................................................................................. 1-1
Series 2260B available models ............................................................................................ 1-2
Verification preparation ............................................................................................ 2-1
Verification inform ation ......................................................................................................... 2-1
Verification equipment.......................................................................................................... 2-2
Constant voltage verification ................................................................................... 3-1
Constant voltage (CV) verification tests ............................................................................... 3-1
Voltage programming and measurement accuracy verification ................................................. 3-1
Constant voltage load regulation verification ............................................................................. 3-2
Constant voltage line regulation verification .............................................................................. 3-3
Constant voltage ripple and noise verification ........................................................................... 3-5
Constant current verification ................................................................................... 4-1
Constant current (CC) verification tests ............................................................................... 4-1
Current programming and measurement accuracy verification ................................................. 4-1
Constant current load regulation verification ............................................................................. 4-3
Constant current line regulation verification .............................................................................. 4-4
Constant current noise verification ............................................................................................ 4-6
Verification test record forms .................................................................................. 5-1
30 Volt verification test record .............................................................................................. 5-2
80 Volt verification test record .............................................................................................. 5-4
250 Volt verification record .................................................................................................. 5-6
800 Volt verification record .................................................................................................. 5-8
2260B Adjustment equipment .................................................................................. 6-1
Adjustment equipment ......................................................................................................... 6-1
Analog interface adjustment .................................................................................... 7-1
Remove top cover ................................................................................................................ 7-1
Confirm location of adjustment points .................................................................................. 7-2
Analog interface adjustment procedure ............................................................................... 7-3
Table of Contents
Verification and Adjustment Manual
Series 2260B Programmab le Power Supplies
Constant voltage adjustment ................................................................................... 8-1
Connect device equipment .................................................................................................. 8-1
Constant voltage adjustment procedure .............................................................................. 8-2
Constant current adjustment ................................................................................... 9-1
Connect device equipment .................................................................................................. 9-1
Constant current adjustment procedure ............................................................................... 9-2
Adjustment code glossary ..................................................................................... 10-1
Adjustment codes ............................................................................................................... 10-2
Index ............................................................................................................................. 1
Series 2260B available models ................................................ 1-1

Welcome

Thank you for choosing a Keithley Instruments product. The Series 2260B Programmable DC Power Supplies is designed for use in the laboratory and for test applications. It also has excellent regulation and low output voltage ripple. The digital displays provide accurate readings of voltage and current and also provide for easy, precise setting of output values using digital entry of current and voltage values. Output voltage can be set from the front panel, using a remote analog voltage or resistance, or over any of the digital interfaces: LAN, USB, GPIB, or RS-485. Voltage and current analog outputs are also available for remote monitoring and analog control.
Section 1

Introduction

In this section:
Welcome .................................................................................. 1-1
Extended warranty ................................................................... 1-1
Contact information .................................................................. 1-1

Extended warranty

Additional years of warranty coverage are available on many products. These valuable contracts protect you from unbudgeted service expenses and provide additional years of protection at a fraction of the price of a repair. Extended warranties are available on new and existing products. Contact your local Keithley Instruments representative for details.

Contact information

If you have any questions after you revie w the infor mation in this documentation, please contact your local Keithley Instruments office, sales partner, or distributor. You can also call the corporate headquarters of Keithley Instruments (toll-free inside the U.S. and Canada only) at 1-800-935-5595, or from outside the U.S. at +1-440-248-0400. For worldwide contact numbers, visit the
Instruments website (http://www.tek.com/keithley).
Keithley
Section
ent Manual
2260B-30-36
0 V to 30 V
0 A to 36 A
360 W
2260B-30-72
0 V to 30 V
0 A to 72 A
720 W
2260B-30-108
0 V to 30 V
0 A to 108 A
1080 W
2260B-80-13
0 V to 80 V
0 A to 13.5 A
360 W
2260B-80-27
0 V to 80 V
0 A to 27 A
720 W
2260B-80-40
0 V to 80 V
0 A to 40.5 A
1080 W
2260B-250-4
0 V to 250 V
0 A to 4.5 A
360 W
2260B-250-13
0 V to 250 V
0 A to 13.5 A
1080 W
2260B-800-1
0 V to 800 V
0 A to 1.44 A
360 W
2260B-800-2
0 V to 800 V
0 A to 2.88 A
720 W
2260B-800-4
0 V to 800 V
0 A to 4.32 A
1080 W
1: Introduction Series 2260B Programmab le Power Supplies Verification and Adjustm

Series 2260B available models

The 2260B series consists of 12 models, divided into three different model types covering four power capacities: 360 watt models, 720 watt models, and 1080 watt models.
Model number Output voltage Output current Power
2260B-250-9 0 V to 250 V 0 A to 9 A 720 W
1-2 077104802 / November 2016
Verification equipment .............................................................. 2-1
In this section:
Verification information ............................................................. 2-1

Verification information

In order to ensure that the power supply is working properly and performance accuracy, we recommend that you use the equipment listed for verifying your equipment. Also, note the following conditions and parameters that you should follow when performing verification procedures.
It is recommended that you verify instrument specifications when:
Section 2

Verification preparation

Using the power supply in a new environment.
Removing the cover for any reason.
The following are the environmental conditions that are required when performing verification and adjustment:
Indoor location only; no direct sunlight; dust free area
Relative humidity <80%
Temperature +18 to 28 °C (64 to 82 °F)
Warm-up time ≥30 minutes
If the verification fails:
You will need to perform the adjustment procedure.
If the adjustment does not accomplish the desired result you will have to send the instrument to
your local Keithley Instruments office, sales partner, or distributor for repair (see
Information (on page 1-1).
Contact
Section
Verification and Adjustment Manual
Precision current shunt
3 A (0.1 Ω) 0.02% TC = 10 ppm / ºC
GW Instek PCS-1000 or
30 A (0.01 Ω) 0.02% TC = 10 ppm / ºC
300 A (0.001 Ω) 0.02% TC = 10 ppm / ºC
Electronic loads
60 V, 240 A minimum with transient capability and a slew rate of 1 A/μs or better.
Various manufacturers of
500 V, 60 A minimum with transient capability
and a slew rate of 0.4 A/μs or better.
and a slew rate of 0.2 A/μs or better.
AC power supply Adjustable to highest rated input voltage range
Ametek 3001i or equivalent Power: 3000 VA
Oscilloscope
Sensitivity: 1 mV
Tektronix DPO4014B or
Bandwidth limit: 20 MHz
Probe: 1:1 with JEITA RC-9131B
Voltage resolution: 0.1 mV
Accuracy: <0.01% mV of reading
2: Verification preparation Series 2260B Programmab le Power Supplies

Verification equipment

Here is a list of the recommended equipment used for verifying the specifications of the programmable DC power supplies:
Type Specifications Recommended instrument
1000 V, 12 A minimum with transient capability
Digital multimeter
equivalent
programmable DC power supply for electronic loads
equivalent
Keithley DMM 7510 or equivalent
Note that the following sections will indicate how to set up the equipment based on the type of verification that you are performing.
2-2 077104802 / November 2016
Constant voltage (CV) verification tests ................................... 3-1

Constant voltage verific at io n

In this section:

Constant voltage (CV) verification tests

There are four verification tests in this series. These tests verify display panel accuracy, verify programmed voltage, measure output voltage changes based on load or no load, measure output voltage based on AC line voltage changes, and measure ripple and noise.

Voltage programming and measurement accuracy verification

Section 3
This test verifies that the voltage programming and measurement functions are within specifications.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 1: 2260B Voltage adjustment devices
Section
Verification and Adjustment Manual
3: Constant voltage verification Series 2260B Programmab le Power Supplies
Before proceeding, make sure that the 2260B instrument is off.
To perform voltage programming and measurement accuracy:
1. Connect the AC power supply to the 2260B.
2. Connect the DMM on the PCS-1000 directly across the +S and -S terminals to an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. Press the Set button on your 2260B and program the output voltage to 0.1 and the output current to its maximum value with the load off (see Series 2260B available models (on page 1-1
)).
5. On the 2260B press the Output button. Note that CV is displayed on the 2260B front panel and the output current reading on the PCS-1000 should be approximately zero.
6. Record the DMM output voltage reading and the voltage indicated on the 2260B. The readings should be within the limits specified in the test record form for the appropriate model (see
Verification test record forms (on page 5-1
) for the appropriate model, under Voltage
Programming and Measurement).
7. Program the output voltage to its maximum rated value (see Series 2260B available models (on page 1-1)).
8. Record the DMM output voltage reading and the voltage indicated on the 2260B.
The readings should be within the limits specified in the test record form for the appropriate model (see Verification test record forms (on page 5-1
) for the appropriate model, under Voltage
Programming and Measurement).

Constant voltage load regulation verification

This test measures the change in output voltage resulting from a change from full load to no load.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 2: 2260B Constant voltage load regulation
Verifying voltage programming and measurement accuracy is complete.
3-2 077104802 / November 2016
Series 2260B
Constant voltage verification
Programmable Power Supplies Verification and Adjustment Manual Section 3:
Before proceeding, make sure that the 2260B instrument is off.
To perform CV load regulation verification:
1. Connect the AC power supply to the 2260B.
2. Connect the DMM on the PCS-1000 directly across the +S and -S terminals to an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. Press the Set button on your 2260B and program the output current to its maximum value and the output voltage to the maximum rated value (see Series 2260B available models (on page
1-
1)).
5. On the 2260B press the Output button.
6. On the electronic DC load instrument, set the output load to 1 A. Note that CV is displayed on the 2260B front panel. If CV is not displayed, adjust the load so that the output current drops and CV is displayed.
7. Record the DMM output voltage reading.
8. Open the load and record the DMM voltage reading.
The difference between the DMM readings in steps 7 and 8 is the load effect, which should not exceed the value listed in the test record form (see Verification test record forms (on page 5-1
) for the
appropriate model, under CV load regulation). Verifying CV load regulation is complete.

Constant voltage line regulation verification

This test measures the change in output voltage resulting from a change in AC line voltage from the minimum to maximum value within the line voltage specifications.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 3: 2260B Constant voltage line regulation
077104802 / November 2016 3-3
Section
Verification and Adjustment Manual
3: Constant voltage verification Series 2260B Progr am mab le Power Supplies
Before proceeding, make sure that the 2260B instrument is off.
To perform CV line regulation verification:
1. Connect the AC power supply to the 2260B.
2. Connect the DMM on the PCS-1000 directly across the +S and -S terminals to an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. Set the variable AC voltage to nominal line voltage.
5. Press the Set button on the 2260B and program the output current to its maximum value and the output voltage to its maximum rated value (see Series 2260B available models (on page 1-1
)).
6. On the 2260B press the Output button.
7. On the electronic DC load instrument, set the electronic output load to 1 A. Note that CV is displayed on the 2260B front panel. If CV is not displayed, adjust the load so that the output current drops and CV is displayed.
8. Adjust the AC power source to the low-line voltage (85 VAC for 100/120 nominal line; 170 VAC for 200/240 nominal line).
9. Record the DMM output voltage reading.
10. Adjust the AC power source to the high-line voltage (132 VAC for 100/120 nominal line; 265 VAC for 200/240 nominal line).
11. Record the DMM output voltage reading.
The difference between the DMM reading in steps 9 and 11 is the source effect, which should not exceed the value listed in the test record form (see Verification test record forms (on page 5-1
) for the
appropriate model, under CV line regulation). Verifying CV line regulation is complete.
3-4 077104802 / November 2016
Series 2260B
Constant voltage verification
Programmable Power Supplies Verification and Adjustment Manual Section 3:

Constant voltage ripple and noise verification

This test measures the DC output voltage with the (10:1) 50 Ω probe.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 4: 2260B Constant voltage ripple noise verification
077104802 / November 2016 3-5
Section
ual
3: Constant voltage verification Series 2260B Programmab le Power Supplies Verification and Adjustment Man
Before proceeding, make sure that the 2260B instrument is off.
To perform CV ripple and noise verification:
1. Connect the AC power supply to the 2260B.
2. Connect the DMM on the PCS-1000 directly across the +S and -S terminals to an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. On the oscilloscope, set the AC coupling to 20 MHz bandwidth, the scale to 10 mV Sampling mode to average, and the measurement to peak-to-peak.
5. Press the Set button on your 2260B and program the power supply to output current to its maximum value and the output voltage to its maximum rated value (see
Series 2260B available
models (on page 1-1)).
6. On the 2260B press the Output button. Let the oscilloscope run for 20 seconds in order to generate enough measurement points. The result should not exceed the peak-to-peak limits in the test record form.
7. Use the DMM to measure the RMS noise voltage using AC voltage (see next figure).
The result should not exceed the RMS limits in the test record form (see Verification test record forms (on page 5-1) for the appropriate model, under CV ripple and noise - RMS).
Verifying CV ripple and noise is complete.
3-6 077104802 / November 2016
Constant current verification tests ............................................ 4-1

Constant current verification

In this section:

Constant current (CC) verification tests

There are four verification tests in this series. These tests verify display panel accuracy, verify programmed voltage, measure output voltage changes based on load or no load, measure output voltage based AC line voltage changes, and measure output noise.

Current programming and measurement accuracy verification

Section 4
This test verifies that the current programming and measurement functions are within specifications.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 5: 2260B Current analog adjustment devices
Section
Verification and Adjustment Manual
4: Constant current verification Series 2260B Programmab le Power Supplies
Before proceeding, make sure that the 2260B instrument is off.
To perform current programming and measurement accuracy:
1. Connect the AC power supply.
2. Connect the current input on the PCS-1000 across the +S and -S terminals through an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. Press the Set button on the 2260B, program the output voltage to 15 volts and set the output current to zero (see Series 2260B available models (on page 1-1
), if needed).
5. Set the electronic load to a short and on the 2260B press the Output button. Note that CC is displayed on the 2260B front panel.
6. Record the precision current shunt (PCS-1000) input current readings and the current indicated on the 2260B. The readings should be within the limits specified in the test record form for the appropriate model (see Verification test record forms (on page 5-1
) for the appropriate model,
under Current Programming and Measurement, if needed).
7. Program the output current to its maximum rated value (see Series 2260B available models (on page 1-1), if needed).
8. Set the electronic load to a short.
9. Record the precision current shunt (PCS-1000) input current readings and the current indicated on the 2260B.
The readings should be within the limits specified in the test record form for the appropriate model (see Verification test record forms (on page 5-1
) for the appropriate model, under Current
Programming and Measurement, if needed). Verifying Current Programming and Measurement Accuracy is complete.
4-2 077104802 / November 2016
Series 2260B
Constant current verification
Programmable Power Supplies Verification and Adjustment Manual Section 4:

Constant current load re gulation verification

This test measures the change in output current resulting from a change from full scale to short circuit.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 6: 2260B Constant current load regulation
Before proceeding, make sure that the 2260B instrument is off.
To perform CC load regulation verification:
1. Connect the AC power supply.
2. Connect the current input on the PCS-1000 across the +S and -S terminals through an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. With the electronic load in CR mode, press the Set button and program the 2260B output voltage to 15 volts.
5. Press the Output button on your 2260B.
6. Using the electronic load, set the output current to the maximum rated output. Note that CC is displayed on the 2260B front panel. If CC is not displayed, adjust the load so that the voltage drops and CC is displayed.
7. Record the PCS-1000 input current reading.
8. Short the electronic load and record the indicated current reading on the PCS-1000.
The difference in the current readings in steps 7 and 8 is the load effect, which should not exceed the value listed in the test record form for the appropriate model (see Verification test record forms
(on
page 5-1) for the appropriate model, under CC load regulation, if needed). Verifying CC load regulation is complete.
077104802 / November 2016 4-3
Section
Verification and Adjustment Manual
4: Constant current verification Series 2260B Programmab le Power Supplies

Constant current line re gulation verification

This test measures the change in output current that results from a change in AC line voltage from the minimum to maximum value within the line voltage specifications.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 7: 2260B Constant current line regul ation
4-4 077104802 / November 2016
Series 2260B
Constant current verification
Programmable Power Supplies Verification and Adjustment Manual Section 4:
Before proceeding, make sure that the 2260B instrument is off.
To perform CC line regulation verification:
1. Connect the AC power supply.
2. Connect the PCS-1000 current input across the +S and -S terminals through an electronic DC load, as shown in the previous figure.
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. Set the variable AC voltage to nominal line voltage.
5. On your 2260B press the Set button and program the output current to its maximum rated value (see Series 2260B available models (on page 1-1
), if needed).
6. Press the Output button on your 2260B.
7. With the electronic load in CR mode short the load and program the 2260B output voltage to 15 volts. Note that CC is displayed on the 2260B front panel. If CC is not displayed, adjust the load so that the voltage drops and CC is displayed.
8. Adjust the AC power source to the low-line voltage (85 VAC for 100/120 nominal line; 170 VAC for 200/240 nominal line).
9. Record the PCS-1000 input current reading.
10. Adjust the AC power source to the high-line voltage (132 VAC for 100/120 nominal line; 265 VAC for 200/240 nominal line).
11. Record the PCS-1000 input current reading.
The difference between the PCS-1000 reading in steps 9 and 11 is the source effect, which should not exceed the value listed in the test record form (see Verification test record forms (on page 5-1
) for
the appropriate model, under CV line regulation, if needed). Verifying CC line regulation is complete.
077104802 / November 2016 4-5
Section
Verification and Adjustment Manual
4: Constant current verification Series 2260B Programmab le Power Supplies

Constant current noise ver ification

This test measures the AC output voltage with the PCS-1000 and DMM.
Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this verification test while instruments are powered on. Turn off the 2260B instrument from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Figure 8: 2260B Constant current noise verification
Before proceeding, make sure that the 2260B instrument is off.
To perform CC noise verification:
1. Connect the AC power supply.
2. Connect the PCS-1000 current input across the +S and -S terminals through an electronic DC load (see previous figure).
3. Turn on the 2260B and, if necessary, all of the other instruments.
4. Press the Set button on your 2260B and program the power supply output current to its maximum value and the output voltage to 15 volts (see Series 2260B available models (on page 1-1
), if
needed).
5. On the 2260B press the Output button.
6. Use the DMM to measure the RMS noise voltage using AC voltage (see previous figure).
The result should not exceed the RMS limits in the test record form (see Verification test record forms (on page 5-1) for the appropriate model, under CV ripple and noise - RMS, if needed).
Verifying CC ripple and noise is complete.
4-6 077104802 / November 2016
800 Volt verification record ....................................................... 5-8
Section 5

Verification test record forms

In this section:
30 Volt verification test record .................................................. 5-1
80 Volt verification test record .................................................. 5-4
250 Volt verification record ....................................................... 5-6
Section
Verification and Adjustment Manual
Model  2260B-30-36
2260B-30-72
2260B-30-108
Serial number:
Verified by
Name:
Company/contact:
Year:
Month:
Day:
Environment
Temperature:
Humidity:
Minimum voltage out
All
-10.1 mV
+10.1 mV
Measurement accuracy
All
-20.1 mV
+20.1 mV
Rated voltage out
All
29.960 V
30.040 V
Measurement accuracy
All
29.95 V
30.5 V
All
-20 mV
+20 mV
All
-18 mV
+18 mV
5: Verification test record forms Series 2260B Programmab le Power Supplies

30 Volt verification test record

Print this page and record your verification results that can be used for reference in the future. Make sure to keep this record with your power supply.
Constant voltage tests:
Voltage programming and measurement
Model Minimum
specifications
CV load regulation Model Minimum
specifications
Results Maximum specifications
Results Maximum specifications
CV line regulation Model Minimum
CV ripple and noise Model Minimum
Peak-to-peak
RMS
5-2 077104802 / November 2016
Results Maximum specifications
specifications
Results Maximum specifications
specifications
30-36 N/A 60 mV 30-72 N/A 80 mV 30-108 N/A 100 mV 30-36 N/A 7 mV 30-72 N/A 11 mV 30-108 N/A 14 mV
Series 2260B
Verification test record forms
30-36
-41 mA
+41 mA
30-72
-77 mA
+77mA
30-108
-113 mA
+113 mA
30-36
-41 mA
+41 mA
30-72
-77 mA
+77mA
30-108
-113 mA
+113 mA
Programmable Power Supplies Verification and Adjustment Manual Section 5:
Constant current tests:
Current programming and measurement
Minimum current out 30-36 -30 mA +30 mA
Measurement accuracy 30-36 -40 mA +40 mA
Model Minimum
Results Maximum specifications
specifications
30-72 -60 mA +60 mA 30-108 -100 mA +100 mA
30-72 -70 mA +70 mA 30-108 -100 mA +100 mA
Rated current out
30-36 35.934 A 36.066 A 30-72 71.898 A 72.102 A 30-108 107.862 A 108.138 A
Measurement accuracy 30-36 35.924 A 36.076 A
30-72 71.858 A 72.142 A 30-108 107.79 A 108.21 A
CC load regulation Model Minimum
Results Maximum specifications
specifications
CC line regulation Model Minimum
specifications
Results Maximum specifications
077104802 / November 2016 5-3
Section
Verification and Adjustment Manual
Model  2260B-80-13
2260B-80-27
2260B-80-40
Serial number:
Verified by
Name:
Company/contact:
Year:
Month:
Day:
Environment
Temperature:
Humidity:
Minimum voltage out
All
-10.1 mV
+10.1 mV
Measurement accuracy
All
-20.1 mV
+20.1 mV
All
-45 mV
+45 mV
All
-43 mV
+43 mV
5: Verification test record forms Series 2260B Programmab le Power Supplies

80 Volt verification test record

Print this page and record your verification results that can be used for reference in the future. Make sure to keep this record with your power supply.
Constant voltage tests:
Voltage programming and measurement
Rated voltage out All 79.91 V 80.090 V Measurement accuracy All 79.9 V 80.1 V
Model Minimum
specifications
Results Maximum specifications
CV load regulation Model Minimum
specifications
Results Maximum specifications
CV line regulation Model Minimum
specifications
Results Maximum specifications
CV ripple and noise Model Minimum
specifications
Peak-to-peak 80-13 N/A 60 mV
80-27 N/A 80 mV 80-40 N/A 100 mV
RMS 80-13 N/A 7 mV
80-27 N/A 11 mV 80-40 N/A 14 mV
Results Maximum specifications
5-4 077104802 / November 2016
Series 2260B
Verification test record forms
80-13
-18 mA
+18 mA
80-27
-32 mA
+32 mA
80-40
-45 mA
+45 mA
80-13
-18 mA
+18 mA
80-27
-32 mA
+32 mA
80-40
-45 mA
+45 mA
Programmable Power Supplies Verification and Adjustment Manual Section 5:
Constant current tests:
Current programming and measurement
Minimum current out 80-13 -10 mA +10 mA
Measurement accuracy 80-13 -20 mA +20 mA
Rated current out 80-13 12.87 A 13.13 A
Measurement accuracy 80-13 12.967 A 13.033 A
Model Minimum
Results Maximum specifications
specifications
80-27 -30 mA +30 mA 80-40 -40 mA +40 mA
80-27 -40 mA +40 mA 80-40 -50 mA +50 mA
80-27 26.943 A 27.057 A 80-40 39.92 A 40.080 A
80-27 26.933 A 27.067 A 80-40 39.91 A 30.09 A
CC load regulation Model Minimum
specifications
CC line regulation Model Minimum
specifications
Results Maximum specifications
Results Maximum specifications
077104802 / November 2016 5-5
Section
Verification and Adjustment Manual
Model  2260B-250-4
2260B-250-9
2260B-250-13
Serial number:
Verified by
Name:
Company/contact:
Year:
Month:
Day:
Environment
Temperature:
Humidity:
Minimum voltage out
All
-200 mV
+200 mV
Measurement accuracy
All
-200 mV
+200 mV
All
-130 mV
+130 mV
All
-128 mV
+128 mV
5: Verification test record forms Series 2260B Programmable Pow er Supplies

250 Volt verification record

Print this page and record your verification results that can be used for reference in the future. Make sure to keep this record with your power supply.
Constant voltage tests:
Voltage programming and measurement
Rated voltage out All 249.47 V 250.53 V Measurement accuracy All 249.47 V 250.53 V
Model Minimum
specifications
Results Maximum specifications
CV load regulation Model Minimum
specifications
Results Maximum specifications
CV line regulation Model Minimum
specifications
Results Maximum specifications
CV ripple and noise Model Minimum
specifications
Peak-to-peak 250-4 N/A 80 mV
250-9 N/A 100 mV 250-13 N/A 120 mV
RMS 250-4 N/A 15 mV
250-9 N/A 15 mV 250-13 N/A 15 mV
Results Maximum specifications
5-6 077104802 / November 2016
Series 2260B
Verification test record forms
250-4
-9.000 mA
+9.000 mA
250-9
-14.000 mA
+14.000 mA
250-13
-18.00 mA
+18.00 mA
250-4
-9.5 mA
+9.5 mA
250-9
-14 mA
+14 mA
250-13
-18.5 mA
+18.5 mA
Programmable Power Supplies Verification and Adjustment Manual Section 5:
Constant current tests:
Current programming and measurement
Minimum current out 250-4 -5 mA +5 mA
Measurement accuracy 250-4 -5 mA +5 mA
Rated current out 250-4 3.991 A 4.001 A
Measurement accuracy 250-4 3.991 mA 4.001 mA
Model Minimum
Results Maximum specifications
specifications
250-9 -10 mA +10 mA 250-13 -15 mA +15 mA
250-9 -10 mA +10 mA 250-13 -20 mA +20 mA
250-9 8.9810 A 9.0190 A 250-13 12.972 A 13.028 A
250-9 8.981 A 9.019 A 250-13 12.967 A 13.033 A
CC load regulation Model Minimum
specifications
CC line regulation Model Minimum
specifications
Results Maximum specifications
Results Maximum specifications
077104802 / November 2016 5-7
Section
l
Model  2260B-800-1
2260B-800-2
2260B-800-4
Serial number:
Verified by
Name:
Company/contact:
Year:
Month:
Day:
Environment
Temperature:
Humidity:
Minimum voltage out
All
-400 mV
+400 mV
Measurement accuracy
All
-400 mV
+400 mV
All
-405 mV
+405 mV
All
-403 mV
+403 mV
5: Verification test record forms Series 2260B Programmab le Power Supplies Verification and Adjustment Manua

800 Volt verification record

Print this page and record your verification results that can be used for reference in the future. Make sure to keep this record with your power supply.
Constant voltage tests:
Voltage programming and measurement
Rated voltage out All 798.8 V 801.2 V Measurement accuracy All 798.8 V 801.2 V
Model Minimum
specifications
Results Maximum specifications
CV load regulation Model Minimum
specifications
Results Maximum specifications
CV line regulation Model Minimum
specifications
Results Maximum specifications
CV ripple and noise Model Minimum
specifications
Peak-to-peak 800-1 N/A 150 mV
800-2 N/A 200 mV 800-4 N/A 200 mV
RMS 800-1 N/A 30 mV
800-2 N/A 30 mV 800-4 N/A 30 mV
Results Maximum specifications
5-8 077104802 / November 2016
Series 2260B
Verification test record forms
800-1
-6.00 mA
+6.00 mA
800-2
-7.00 mA
+7.00 mA
800-4
-9.00 mA
+9.00 mA
800-1
-6.00 mA
+6.00 mA
800-2
-7.00 mA
+7.00 mA
800-4
-9.00 mA
+9.00 mA
Programmable Power Supplies Verification and Adjustment Manual Section 5:
Constant current tests:
Current programming and measurement
Minimum current out 800-1 -2 mA +2 m A
Measurement accuracy 800-1 -2 mA +2 mA
Rated current out 800-1 0.997 A 1.003 A
Measurement accuracy 800-1 0.997 A 1.003 A
Model Minimum
Results Maximum specifications
specifications
800-2 -4 mA +4 m A 800-4 -6 mA +6 m A
800-2 -4 mA +4 m A 800-4 -6 mA +6 mA
800-2 1.994 A 2.006 A 800-4 3.936 A 4.064 A
800-2 1.994 A 2.006 A 800-4 3.936 A 4.064 A
CC load regulation Model Minimum
specifications
CC line regulation Model Minimum
specifications
Results Maximum specifications
Results Maximum specifications
077104802 / November 2016 5-9
Adjustment equipment ............................................................. 6-1
3 A (0.1 Ω) 0.01% TC = 3 ppm / °C
30 A (0.01 Ω) 0.01% TC = 10 ppm / °C
300 A (0.001 Ω) 0.02% TC = 10 ppm / °C
Electronic loads
80 V, 280 A minimum with transient capability and a slew rate of 2.8 A/μs or better.
Various manufacturers of
500 V, 40 A minimum with transient capability and a slew rate of 0.4 A/μs or better.
850 V, 40 A minimum with transient capability and a slew rate of 0.4 A/μs or better.
DC power supply Voltage resolution: 1 mV
Keithley programmable supply 2220-30-01
Accuracy: <0.05% mV of reading
Digital multimeter Voltage resolution: 0.1 mV
Keithley DMM 7510 or Accuracy: <0.01% mV of reading
In this section:

Adjustment equipment

Here is a list of the recommended equipment used for calibrating the programmable DC power supplies:
Type Specifications Recommended instrument
Precision current shunt
Section 6

2260B Adjustment equipment

GW Instek PCS-1000 or equivalent
programmable DC electronic loads
dual-channel DC power
equivalent
Analog interface adjustment procedure .................................... 7-3
In this section:
Remove top cover .................................................................... 7-1
Confirm location of adjustment points ...................................... 7-2

Remove top cover

Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this adjustment while instruments are powered on. Turn off all of the instruments from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Section 7

Analog interface ad justment

Section
Verification and Adjustment Manual
7: Analog interface adjustment Series 2260B Programmable Power Supplies
Before you can make adjustments on the instrument, you will need to remove the 2260B top cover. On the 360 watt models (2260B-30-36, 2260B-80-1 3, 2260 B-250-4, and 2260B-800-01), note that
there are nine screws that you will need to remove in order to remove the top cover (see next figure).
The next figure only shows seven screws. On the 720 watt models (2260B-30-72, 2260B-80-2 7, 2260 B-250-9, 2260B-800-2), there are ten
screws that you will need to remove. On the 1080 watt models (2260B-30-108, 2260B-80-40, 2260B-250-13, 2260B-800-4), there are
eleven screws that you will need to remove.
Figure 9: 2260B Top cover

Confirm location of adjustment points

The location of the adjustment points on the 2260B circuit board (VR400 and VR401) are shown in the next figure:
Figure 10: 2260B circuit board
7-2 077104802 / November 2016
Series 2260B
Analog interface adjustment
Programmable Power Supplies Verification and Adjustment Manual Section 7:
Make sure that you connect the following devices as indicated in the next figure. Also, make sure that you set the Model 2220-30-1 DC power supply, channel one and channel two to 10.00 V.
Figure 11: Analog adjustment de vices

Analog interface adjustment procedure

To start the adjustment procedure:
1. Turn on the programmable DC power supply (Series 2260B) by pressing the Function key.
The Function key lights up, the display indicates "F - 01." Additionally, note that the "x" in the diagram below indicates that the value is not fixed.
2. Rotate the Voltage knob until the display indicates "F - 00".
077104802 / November 2016 7-3
3. Press the Current knob to select the numeric indicator.
Section
Verification and Adjustment Manual
2260B-30-36
3036
2260B-30-72
3072
2260B-30-108
3010
2260B-80-13
8013
2260B-80-27
8027
2260B-80-40
8040
2260B-250-4
2545
2260B-250-9
2509
2260B-250-13
2515
2260B-800-1
8014
2260B-800-2
8028
2260B-800-4
8043
7: Analog interface adjustment Series 2260B Programmable Pow er Supplie s
4. Use the Current knob to select the appropriate password for the model that you are calibrating. Here's the list of models and passwords:
Model number Password
5. Press the Current knob to enter the password (note that the highlighted numeral is the one that you are changing).
6. Once the number needed is highlighted, press the Current knob to move to the next numeric position.
7. Press the Voltage knob to enter the password.
The display will indicate "C000".
8. Turn on the external power supply (the 2220-30-01) and set the output to 10.000 V ±0.001 V, for both channel 1 and 2.
9. Press the Voltage knob to prepare to execute the external output voltage adjustment (C010).
Make sure that the DC electronic load instrument is turned off.
10. Press the Voltage knob (C011 is displayed); the 2260B instrument Output button illuminates and CV is displayed.
7-4 077104802 / November 2016
Series 2260B
og interface adjustment
2260B-30-36
29.994
30.006
2260B-30-72
29.994
30.006
2260B-30-108
29.994
30.006
2260B-80-27
79.984
80.016
2260B-80-40
79.984
80.016
2260B-250-4
249.950
250.050
2260B-250-9
249.950
250.050
2260B-250-13
249.950
250.050
2260B-800-1
799.840
800.160
2260B-800-2
799.840
800.160
2260B-800-4
799.840
800.160
Programmable Power Supplies Verification and Adjustment Manual Section 7: Anal
Make sure that the reading on the DMM is within the specified voltage range. If not, adjust the voltage resistor (VR401) on the 2260B instrument (see Confirm location of adjustment points (on page 7-2)).
Here's the list of models with adjustment ranges for point VR401:
Model number minimum volts maximum volts
2260B-80-13 79.984 80.016
11. Press the Voltage knob to return to external voltage adjustment (C010).
12. Rotate the Voltage knob until it indicates C020, in order to prepare to execute the external output current adjustment.
Make sure that the load key and short key on the DC electronic load instrument are turned on.
13. Press the Voltag e knob (C0 21 is disp layed); the 2260B Output button illuminates and CC is displayed.
077104802 / November 2016 7-5
Section
Verification and Adjustment Manual
2260B-30-36
35.9928
36.0072
2260B-30-72
71.9856
72.0144
2260B-30-108
107.9784
108.0216
2260B-80-13
13.4973
13.5027
2260B-80-27
26.9946
27.0054
2260B-80-40
40.4919
40.5081
2260B-250-4
4.4991
4.5009
2260B-250-9
8.9982
9.0018
2260B-250-13
13.4973
13.5027
2260B-800-1
1.4397
1.4403
2260B-800-2
2.8794
2.8806
2260B-800-4
4.3191
4.3209
7: Analog interface adjustment Series 2260B Programmable Power Supplies
Make sure that the reading on the PCS-1000 current meter is within the specified current range. If not, adjust the current resistor (VR400) on the 2260B instrument (see Confirm location of adjustment
points (on page 7-2)).
Here's the list of models with adjustment ranges for point VR401:
Model number minimum current maximum current
14. Press the Voltage knob to return to the external current adjustment (C020).
15. Rotate the Voltage knob to exit the external current adjustment display (C030 is displayed).
16. Press the Voltage knob to return to the main adjustment display.
17. Rotate the Current knob to exit adjustment mode (C000:2).
7-6 077104802 / November 2016
Series 2260B
Analog interface adjustment
Programmable Power Supplies Verification and Adjustment Manual Section 7:
18. Press the Voltage knob to exit adjustment mode.
The "x" in the diagram above indicates that the value is not fixed.
19. Turn off the 2600B instrument and put the top cover back on.
Analog interface adjustment is complete.
077104802 / November 2016 7-7
Constant voltage adjustment procedure ................................... 8-2
In this section:
Connect device equipment ....................................................... 8-1

Connect device equi pme nt

Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this adjustment while instruments are powered on. Turn off all of the instruments from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Section 8

Constant voltage adjustment

Make sure that you connect the following devices as indicated in the next figure.
Figure 12: 2260B Constant voltage adjustment devices
Section
Verification and Adjustment Manual
2260B-30-36
3036
2260B-30-72
3072
2260B-30-108
3010
2260B-80-13
8013
2260B-80-27
8027
2260B-80-40
8040
2260B-250-4
2545
2260B-250-9
2509
2260B-250-13
2515
2260B-800-1
8014
2260B-800-2
8028
2260B-800-4
8043
8: Constant voltage adjustment Series 2260B Programmable Power Supplies

Constant voltage (CV) adjustment procedure

To start the adjustment procedure:
1. Turn on the programmable DC power supply (Series 2260B) by pressing the Function key.
The function key lights up, the display indicates "F - 01." Additionally, note that the "x" in the diagram below indicates that the value is not fixed.
2. Rotate the voltage knob until it indicates "F - 00".
3. Use the Current knob to select the appropriate password for the model that you are calibrating. Here's the list of models and passwords:
Model number Password
4. Press the voltage knob to enter the enter the password.
The display will indicate "C000".
8-2 077104802 / November 2016
Series 2260
Constant voltage adjustment
B Programmable Power Supplies Verification and Adjustment Manual Section 8:
5. Use the current knob to enter "constant voltage" adjustment mode. To enter adjustment mode, rotate the current knob until "1" is displayed on the bottom display (C000:1).
6. Press the voltage knob to enter voltage adjustment (C110).
7. Press the voltage knob two times to enter the voltage offset adjustment (C111).
8. Press the voltage knob to set the offset voltage value (C112).
The "x" in the diagram above indicates that the value is not fixed.
9. Rotate the current knob to set the offset value on the 2260B display so that it is the same as the reading on the PCS-1000 voltmeter.
10. Press the voltage knob to enter the output voltage middle adjustment (C113).
11. Press the voltage knob to set the middle voltage value.
077104802 / November 2016 8-3
Section
Verification and Adjustment Manual
8: Constant voltage adjustment Series 2260B Programmable Power Supplies
The "x" in the diagram above indicates that the value is not fixed.
12. Rotate the current knob to set the middle voltage value on the 2260B display so that it is the same as the reading on the PCS-1000 voltmeter.
13. Press the voltage knob to enter the output voltage full-scale adjustment (C115).
14. Press the voltage knob to set the full-scale voltage value (C116).
The "x" in the diagram above indicates that the value is not fixed.
15. Rotate the current knob to set the full-scale voltage value on the 2260B display so that it is the same as the reading on the PCS-1000 voltmeter.
16. Press the voltage knob to enter the OVP (overvoltage protection) adjustment (C117).
17. Press the voltage knob to start the automatic OVP adjustment (C118).
8-4 077104802 / November 2016
Series 2260B
Constant voltage adjustment
Programmable Power Supplies Verification and Adjustment Manual Section 8:
The display will indicate "OK" when the adjustment is complete for approximately five seconds. Additionally, the output turns on and the CV indicator is lit. Also, the instrument automatically returns to C110 after the automatic OVP adjustment ends.
18. Rotate the voltage knob to go to save mode (C130).
19. Rotate the current knob until "1" is shown on the bottom display (C130:1)
20. Press the voltage knob to save the adjustment results.
The display will indicate "OK" and the instrument will automatically return to "C000" when the adjustment value is saved.
21. Rotate the current knob to go to "exit adjustment" mode (C000:2).
077104802 / November 2016 8-5
Section
Verification and Adjustment Manual
8: Constant voltage adjustment Series 2260B Programmable Power Supplies
22. Press the voltage knob to exit adjustment mode.
The "x" in the diagram above indicates that the value is not fixed. Constant voltage adjustment is complete.
8-6 077104802 / November 2016
Constant current adjustment procedure ................................... 9-2
In this section:
Connect device equipment ....................................................... 9-1

Connect device equi pme nt

Hazardous voltages may be present on all output and guard terminals. To prevent electrical shock that could cause injury or death, never make or break connections to any of the instruments used for this adjustment while instruments are powered on. Turn off all of the instruments from the front panel or disconnect the main power cord from the rear of the instrument before handling cables. Putting the equipment into an output-off state does not guarantee that the outputs are powered off if a hardware or software fault occurs. Precautions must be taken to prevent a shock hazard by surrounding the test device and any unprotected leads (wiring) with double insulation for up to 800 volts depending on the power supply that you have, Category I.
Section 9

Constant current adjustment

Make sure that you connect the following devices as indicated in the next figure.
Figure 13: Constant current adjustment devices
Section
Verification and Adjustment Manual
2260B-30-36
3036
2260B-30-72
3072
2260B-30-108
3010
2260B-80-13
8013
2260B-80-27
8027
2260B-80-40
8040
2260B-250-4
2545
2260B-250-9
2509
2260B-250-13
2515
2260B-800-1
8014
2260B-800-2
8028
2260B-800-4
8043
9: Constant current adjustment Series 2260B Programmable Power Supplie s

Constant current (CC) adjustment procedur e

To start the adjustment procedure:
1. Turn on the programmable DC power supply (Series 2260B) by pressing the Function key.
The function key lights up, the display indicates "F - 01." Additionally, note that the "x" in the diagram below indicates that the value is not fixed.
2. Rotate the voltage knob until it indicates "F - 00".
3. Use the Current knob to select the appropriate password for the model that you are calibrating. Here's the list of models and passwords:
Model number Password
4. Press the voltage knob to enter analog interface adjustment.
The display will indicate "C000" when completed.
9-2 077104802 / November 2016
Series 2260B
Constant current adjustmen t
Programmable Power Supplies Verification and Adjustment Manual Section 9:
5. Use the current knob to enter "constant voltage" adjustment mode. To enter the adjustment mode, rotate the current knob until "1" is displayed on the bottom display (C000:1).
6. Press the voltage knob to enter voltage adjustment (C110).
7. Rotate the voltage knob to go to current adjustment (C120).
8. Press the voltage knob to enter the output current offset adjustment (C121).
9. Press the voltage knob to set the offset value.
The "x" in the diagram above indicates that the value is not fixed.
10. Rotate the current knob to set the offset value on the 2260B display so that it is the same as the reading on the PCS-1000 current meter.
11. Press the voltage knob to enter the output current half-scale adjustment (C123).
077104802 / November 2016 9-3
Section
Verification and Adjustment Manual
9: Constant current adjustment Series 2260B Programmable Power Supplies
12. Press the voltage knob to set the half-scale current value (C124).
The "x" in the diagram above indicates that the value is not fixed.
13. Rotate the current knob to set the half-scale current value on the 2260B display so that it is the same as the reading on the PCS-1000 current meter.
14. Press the voltage knob to enter the output current full-scale adjustment (C125).
15. Press the voltage knob to set the full-scale value (C126).
The "x" in the diagram above indicates that the value is not fixed.
1. Rotate the current knob to set the full-scale value on the 2260B display so that it is the same as the reading on the PCS-1000 current meter.
2. Press the voltage knob to enter the OCP (overcurrent protection) adjustment (C127).
3. Press the voltage knob to start the automatic OCP adjustment (C128).
9-4 077104802 / November 2016
Series 2260B
adjustment
Programmable Power Supplies Verification and Adjustment Manual Section 9: Constant current
The display will indicate "OK" when the adjustment is complete. Additionally, the output turns on and the CC indicator is lit. Also, the instrument automatically returns to C120 after the automatic OCP adjustment ends.
4. Rotate the voltage knob to go to save mode (C130).
5. Rotate the current knob until "1" is shown on the bottom display (C130:1)
6. Press the voltage knob to save the adjustment results.
The display will indicate "OK" and the instrument will automatically return to "C000" when the adjustment value is saved.
7. Rotate the current knob to go to "exit adjustment" mode (C000:2).
077104802 / November 2016 9-5
8. Press the voltage knob to exit adjustment mode.
Section
Verification and Adjustment Manual
9: Constant current adjustment Series 2260B Programmable Power Supplie s
The "x" in the diagram above indicates that the value is not fixed. Constant current adjustment is complete.
9-6 077104802 / November 2016
Adjustment codes ................................................................... 10-2
Section 10

Adjustment code glossary

In this section:
Section
Verification and Adjustment Manual
F-01
X
Power applied; Function key lights up
F-00
0000
Enter adjustment mode
C000
0
Password entered; instrument ready for adjustment
C000
1
Enter constant voltage adjustment mode
C000
2
Exit adjustment mode
C010
- - - -
Standby to execute external voltage adjustment
C011
- - - -
Output button illuminates and CV is displayed
C020
- - - -
Standby to execute external current adjustment
C021
- - - -
Output button illuminates and CC is displayed
C030
- - - -
Exit external current adjustment
C110
- - - -
Enter voltage adjustment mode
C111
- - - -
Standby adjustment offset voltage value
C112
X
Adjustment offset voltage value
C113
- - - -
Standby adjustment half-scale voltage value
C114
X
Adjustment full-scale voltage value
C115
- - - -
Standby adjustment full-scale voltage value
C116
X
Adjustment full-scale voltage value
C117
- - - -
Standby OVP adjustment
C118
ok
OVP adjustment
C120
- - - -
Enter current adjustment mode
C121
- - - -
Standby adjustment half-scale current value
C122
X
Adjustment offset current value
C123
- - - -
Standby adjustment half-scale current value
C124
X
Adjustment half-scale curren t v alue
C125
- - - -
Adjustment full-scale current value
C126
X
Set the full-scale value
C127
- - - -
Standby OCP adjustment
C128
ok
OCP adjustment
C130
0
Save mode
C130
1
Save adjustment results
C130
ok
Indicates adjustment results saved
X X Exit adjustment mode
10: Adjustment code glossary Series 2260B Programmab le Power Supplies

Adjustment code s

Voltage display Current display Description
The "x" in the previous table indicates that the value is not fixed.
10-2 077104802 / November 2016

Index

2
2260B Adjustment equipment • 6-1 250 Volt verification record • 5-6
3
30 Volt verification test record • 5-2
8
80 Volt verification test record • 5-4 800 Volt verification record • 5-8
A
Adjustment code glossary • 10-1 Adjustment codes • 10-2 Adjustment equipment • 6-1 Analog interface adjustment • 7-1 Analog interface adjustment procedure • 7-3
C
Confirm location of adjustment points • 7-2, 7-5, 7-6 Connect device equipment • 8-1, 9-1 Constant current adjustment • 9-1 Constant current adjustment procedure • 9-2 Constant current line regulatio n verif icat ion • 4-4 Constant current load regulatio n verif ication • 4-3 Constant current noise verification • 4-6 Constant current verification • 4-1 Constant current verification tests • 4-1 Constant voltage (CV) verification tests • 3-1 Constant voltage adjustment • 8-1 Constant voltage adjustment procedure • 8-2 Constant voltage line regulation verification • 3-3 Constant voltage load regulation verification • 3-2 Constant voltage ripple and noise verification • 3-5 Constant voltage verification • 3-1 Contact information • 1-1, 2-1 Current programming and measurement accuracy
verification • 4-1
R
Remove top cover • 7-1
S
Series 2260B available models • 1-2, 3-2, 3-3, 3-4,
3-6, 4-2, 4-5, 4-6
V
Verification equipment • 2-2 Verification information • 2-1 Verification preparation • 2-1 Verification test record forms • 3-2, 3-3, 3-4, 3-6, 4-2,
4-3, 4-5, 4-6, 5-1
Voltage programming and measurement accuracy
verification • 3-1
W
Welcome • 1-1
E
Extended warranty • 1-1
I
Introduction • 1-1
All Keithley trademarks and trade names are the property of Keithley Instruments.
All other trademarks and trade names are the property of their respective companies.
Specifications are subject to change without notice.
Corporate Headquarters • 28775 Aurora Road • Cleveland, Ohio 44139 • 440-248-0400 • Fax: 440-248-6168 • 1-800-935-5595 • www.tek.com/keithley
Keithley Instruments
A Greater Measure of Confidence
12/15
Loading...