Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of
shipment.
Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batteries,
diskettes. and documentation.
During the Wmanty period, we will, at
To exercise this warranty, write or call your local Keithley representative, or contact Keithley headquarters in Cleveland, Ohio, You will
be given prompt assistance and return instructions. Send the product, transportation prepaid, to the indicated service facility. Repairs
will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days.
OUT
option, either repair or replace any product that proves to be defective.
LIMITATION OF WARRANTY
This warranty does not apply to defects resulting from product modification without Keithley’s express written consent, or misuse of
any product or part. This warranty also does not apply to fuses, software, non-rechargeable batteries, damage from battery leakage, or
problems arising from normal wear or failure to follow instructions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR USE. THE REMEDIES PROVIDED HEREIN ARE
BUYER’S SOLE AND EXCLUSIVE REMEDIES.
NEITHER KEITHLEY INSTRUMENTS, INC. NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF THE USE OF ITS INSTRUMENTS AND
SOFTWARE EVEN IF KEITHLEY INSTRUMENTS, INC., HAS BEEN ADVISED IN ADVANCE OF THE POSSIBILITY OF
SUCH DAMAGES. SUCH EXCLUDED DAMAGES SHALL INCLUDE, BUT ARE NOT LIMITED TO: COSTS OF REMOVAL
AND INSTALLATION, LOSSES SUSTAINED AS THE RESULT OF INJURY TO ANY PERSON, OR DAMAGE TO PROPERTY.
current Mode "S Voltage Made Waveforms
Top Cover Assembly
Bortom Cover Assembly.
............................
........................
........................
............................
............................
............................
..........................
...........................
..........................
.............................
...........................
Switch 5101.
.........................
..........................
..........................
........................
Check
.........................
.......................
.....................
...................
..............
..............
...............
Page
I
1
3
3
4
6
6
6
6
6
7
7
8
13
13
14
14
18
18
18
18
18
21
21
0676
I
iii
4
CAUTION
MODEL 225
CAUTION
IF THE DECADE CURRENT DIALS ARE
SET
TO O-O-O and if the Output Selector switch is not at STANDBY, a
COMPLIANCE VOLTAGE up to 100 volts may be present at the oucpuc terminals. For r&x"rn operator
safety the Model 225 Current Source should always be set to STA!IDBY mode when no current output is
needed.
dial seteings and/or Voltage Compliance.
The Output Selector switch (S105) should also be set to
This will ensure that no transient currente will be
STANDBY
prior to changing current
generated.
WARNING
This insrrument is no? approved for use in "hazardous locations" as defined in the National
Electrical Code, Classes I, II, and III.
Class I: Those locations in which flameble gases or vapors are or may be present in the air in
quantifies sufficient to produce explosive or ignitable mixcures.
Class II: Those locations that are hazardous because of the presence of combustible dust.
Class III: Those locations that are hazardous because of the presence of easily ignitable fibers
or flyings.
5
0676
MODEL 225
OUTPUT:
RESOLUTION: Three significant figuresfrom lo-’
ACCURACY: z 0.5% of reading. 2 0.05% of full
STABILITY: -0.02% of reading, _cO.O05% of
NOISE: Less than 0.01% rms of full rangeabove
SPECIFICATIONS
D.C. Current: lo-’ ampere full range to lo-’
ampere in seven decade ranges, 3.digit in.
line readout.
Voltage: 100 voltsmaximum. Compliancelimit
continuously variable from 10 to 100 volts.
Polarity: Positive or negative.
Floating: c 500 volts maximum off chassis
ground, less than 5 ppm of full range change
in output current per volt.
to 10-l ampere. “Trim” potentiometer permits
0.02% of full range or better resolution.
ra”ge.
full range on the 10-l to 1O-6 ampere range,
(e-0.1% of reading 20.02% of range on the
10-T ampere range) the first hour or in sub.
sequent E-hour periods after a l.hour warm-
up and at reasonably constant ambient tern.
perature.
5 HZ.
SPECIFICATIONS
1
LOADREGULATION: t0.005% offullrangefrom
no load to full load on the lo-’ to 10-6 am.
pere range, 50.05% on the lo-’ ampere
range (with FILTER “OFF”).
LINE REGULATION: k 0.005% Of full range for
10% change in line voltage.
OVERLOAD PROTECTION: Voltage limited tocorn.
pliance voltage setting of polarity selected.
Automatic recovery from overload.
MODULATION: Transformer input permitsmodw
lation of current from 50 Hz to 500 Hr.
Input 2 approximately 500 ohms.
FILTER: For operation with inductive loads up to
100 millihenries having greater than 10.6L/R
ratio. Limits shunt output impedance to 1
microfarad shunted bygreaterthan 1010ohms.
ENVIRONMENT: 50% relative humidity limit at
25’C.
CONNECTORS: Output (front and rear panels):
BNC
POWER: 105.125, 210.250 volts (switch se.
lected), 50.60 Hz; 25 watts.
DIMENSIONS, WEIGHT: 5%” high x 8%” wide x
10” deep; net weight, 8 pounds.
0571
225 CURRENT SOURCE
.IWE
1
6
GENERAL DESCRIPTION
MODEL 225
SECTION 1.
1-I. GENERAL.
full scale ranges frarn 10-7 to 10-l
1-2. FEATURES.
The Model 225 is a current source with
ampere.
GENERAL DESCRIPTION
c. Accuracy. The three decade dials adjust the out-
put current to +.05% of full range and +.5% of reading.
d. Polarity. The OUTPUT SELECTOR Switch provides
polarity selection without changing connections c" the
a. Resolution.
The thpe decade dials plus a "Trim"
load.
potentiometer provides a resolution of .02% of full
range.
e. Floating Capability. The instrument LOW can be
floated up t" 2500 volts above chassis.
b. Compliance Voltage. A front panel Compliance
Cantral provides a" adjustable maximum compliance voltsee from 10 to 100 volts.
f. Modulation. The ""tput current can be modulated
over a frequency range from 50 t" 500 Hz.
TABLE l-l.
Front Panel Controls
CQ"trOl Punctianal Description
Power Switch (S202) Controls line power to instrument. Decimal point serves
as a pilot lamp.
OUTPUT SELECTOR Switch (SlO5)
VOLTAGE COMPLIANCE Control
Selects positive or negative polarity or standby pasitio".
Sets the maximum compliance voltage which can be developed
(R117> at the OUT terminal.
Paragraph
2-4, a
2-4, b
2-4, c
LIMIT Indicator (DSlOl)
When lighted, the instrument is in Voltage mode.
Range Switch (SlOl) Selects the full scale range. The decimal point is auto-
matically positioned by the Range switch.
Selectar Switches (S102,
Selects the magnitude af current in 3 decade positions.
s103, SlO4)
FINE Control (R13S) Adjusts current.
FILTER Switch (5106)
When set to "ON", the switch connects a 1 p capacitor
at the OUT terminal.
Pravides line power to instrument.
Sets instrument for 117 or 234 V line operatia".
Type 3AG. slow slow.
Output high cannectio".
Output law connection.
Ground (chassis) connection.
Functional Description Paragraph
117 v: 1,4A; 234 Y: l/S‘4
2-2, e
2-4, d
2-4, e
2-4, f
2-4, s
2-1, a
2-1, a
2-1, a
2.1, a
!4ODUlATE Receptacle (5101) Connection to modulation transformer.
Shorting Link
7
Provides cbnnection between LOW and GROUND.
7
2-1, b
0571
MODEL 225
GENERAL DESCRIPTION
0571
3
8
OPERATION
MODEL 225
SECTION 2.
2-l.
CONNECTIONS.
8. OUtPUt conneerors.
Front Panel.
1.
panel are BNC type receptacles (Keithley
cs-15).
The OVP receptacle (5103) provides conr,ection to the Model 225 Output High. (The center
terminal is Output High while the outer shield is
connected to chassis ground.) Connections to the
output load can be made using the OUT receptacle
when the Model 225 is not floating. (For a more
complete discussion of floating operation, refer to
paragraph 2-2 f.) The LOW receptacle (JlO5) provides
a connection to the Model 225 Output Low.
center terminal is Output Low while the outer shield
is connected to chassis ground.)
2.
asac Panel.
panel is also a BNC type. Connections can be made
to either the front panel receptacle s the rear
panel receptacle with satisfactory results. The
rear panel LOW terminal (3104) is a binding post
(Keithley part no. BP-1lG) which is connected to the
center terminal of 5105 (on the front panel). The
rear panel GROUND terminal (3201) is a binding past
(Keithley part’no. BP-118) which is connected to
chassis ground.
LOW to chassis ground connections.
b. Modulate Connector. The NOD”UTE receptacle
0101) is a BNC type connector (Keithley part no. CS-
15). This connector can be used for modulating the
output current using an ac modulation signal. The
center terminal is connected to a transformer-coupled
modulation circuit while the outer shield is connected
to chassis ground.
c. Cables. Connections should be made using coaxial cables which are low-noise types with graphite
coating between dielectric and shield braid. For
custom length cables, Keithley Part No. SC-9 Low Noise
Coaxial Cable should be used. A mating BNC connector
is available by ordering Keithley part no. CS-44.
The connectors on the front
part
no.
(The
The OUT receptacle on the rear
The shorting link (supplied) permits
OPERATION
2-2.
MEASUREMENT CONSIDERATIONS.
a. Accuracy. The accuracy is specified in terms
of the front panel dial setting or reading as well as
the full range selected. (The FINE current control
must be set
accuracies.) Since the total accuracy of the Model
225
is the s of the readina accuracy and the full
a accuracy, the user should select the lowest
full range for best possible accuracy. Table 2-l.
shows the total accuracy for typical dial settings.
(The accuracy on the lOOkA range is +.05% of 1000 or
o.s”A plus 2.5%
Output
lhrL-elLt setting
1 x lo-6A
1 x lo-6A O-1-0 100 $ + .055 NA
1 x
104A
1 x lo-6A
When the OUTPUT SELECTOR is set to + or - and VOLTAGE When the OUTPUT SELECTOR is set to + or - and VOLTAGE
COMPLIANCE is set to 100, up to 100 volts will be COMPLIANCE is set to 100, up to 100 volts will be
present at the output terminals. present at the output terminals.
b. Stability. The stability is specified in terms
of the front panel dial setting or reading as well as
the full range selected.
specification is the e of the readinR stability and
the full range stability, the user should select the
lowest full range for best possible stability.
c. Line Regulation.
as +..005% of full range for a corresponding 10% change
in line voltage.
d. Load Rezulation.
ified a8 2.005% of full range far a change from no
load to full load.
tion corresponds to an output compliance voltage change
from 0 volts to 100 volts.
to
“0” position to obtain the stated
of 1000 or %A
for a coca1 of 5.5nA.)
TABLE 2-l.
Accuracy
Decade Dial
For Typical Dial Settings
RANGE
Setting
Total Accuracy
(FINE c0ntr01
set to “0”)
o-o-1 1000 &k4 + .505 VA
1-o-o
10-O-O 1000 UA + .0055 p.4
10 @
+ .OlO *
CAUTION CAUTION
Since the m stability
The line regulation is stated
The load regulation is spec-
The no load to full load varia-
FIGURE 4. Co”*ectm”s.
e. Output Impedance.
The effective output impedance of the Model 225 is a function of the load regulation specification and the high to low insulation
resistance. The output impedance for each Range setting is given in Table 2-2. The insulation resistance is 2 x 1012 ohms with the FILTER Switch OFF, but
is 1 x lOlo
ohms with the’FILTER Switch ON. There-
fore the maximum output impedance is 2 x lOl2 on 100
nA and 1OOOti ranges with the FILTER Switch OFF. The
effective output impedance R. ~a” be decermtned by the
following equations.
0676
MODEL 225
TABLE 2-2.
Output Impedance
Range setting
100 m.4
10 mA
1000 liA
100 ,,A
O”tput Impedance*
2 x 10; n
2 x log n
2 x lOlOG
2 x 1OlI R
10 UA
1000 nA
100 nA
2 x l0l2 “,
“, : :;12 n
*With Filter switch set to off. I
For the 100 mA range,
R,, = 103n and % regulation = 0.005%
since I regulation = 100 x
RL
qq
I 1
Then % -
The Model 225 will deliver the dialed current for
a load resistance from 0 ohms to WK. The val’ue
of w is determined by the dialed current I and
the compliance voltage setting VC where Rw.K =
WI.
If the load resistance exceeds RMAx then
the Mcdel 225 will automatically switch into Volt-
age Limit mode indicated by the LIMIT lamp (DSlOl)..
f. Floatiw Operation. The instrument can be floated
“I) to +500 volts off chassis nround with less than 5
ppm of-full range change in &put current per volt
off ground.
on the rear panel m,,st be m.
For floating operation, the shorting link
In this case, two
coaxial cables should be used to connect to output
high and low.
The outer shell of the front panel O!Jf
and LOW receptacles are always at chassis ground for
safety when flokting the instrument.
A
filter capaci-
tor C217 (,05 ~0 is connected between the LOW and
GROUND binding posts to minimize line frequency pickup.
The LOW to GROUND isolation is approximately lo7 ohms.
A
typical example of floating operation is shown in
Fig”re 9. In this example, the Model 225 can be used
with a voltage supply such 8s Keithley Model 240A
to
extend the maximum compliance voltage to +600 volts.
The Model 225 can be adjusted for a compliance voltage
of +500 2100 volts de. The maximum current is 10 milliamperes (for the Model 24OA).
c. contra1 setti”*s.
The output current is selected
by three decade switches and a seven position Range
Switch. Refer to the discussion of the fronr panel
concrok in paragraph 2-4 before power is turned on.
2-4. FRONT PANEL CONTROLS.
a. Power Switch (S2022. This switch controls the
line-power to the instrument.
b. OUTPUT SELECTOR Switch (SlO5). This switch has
four positions designated “-“,
and “+“.
polarity.
The + and - positions set the output current
The two “STANDBY” positions divert ehe out-
“STANDBY”,
“STANDBY”,
put current as shown in Figure 5. The output load is
not disconnected at any position of the switch. (The
shunting effect should be considered, since a small
voltage can be developed across the lo-milliahm shunt
and therefore across the output terminal.)
c. VOLTAGE COMPLIANCE Control (R1171. This control
is a continuously variable potentiometer which can be
see for a compliance voltage from lo~volts to 100
volts.
d. Range Switch (SlOl). This switch selects the
full scale ranpe from 10-7 BmDePe to 10-l amoere.
The dial is deiignated in re&s of engineering Airs;
that is, nA, IJA, and mA. The full range is determined
by the position of the lighted decimal point end the
designated engineering unit.
e. Selector Switches (SIOZ, SlO3. S104L. These
switches select the magnitude of ‘the o”tput current
in three decades. The position
of
the dials determines
the significant digits from O-O-O to V-9-9.
f. FINB Control (Rl38)-. This control is the inner
dial of a dual concentric control. The fully clock-
wise position “10” represents 1 digit added t., the
right-hand dial (S104). The fully counter clockwise
position is the “0” or off position. When the FINE
Control is
set
to “0”. the output current is determined
only by the calibrated dials within rated specifica-
tion. This control cm be used to “trim” the output
current to within +.02% of full range, although the
trim dial is adjustable t., within 2.005% of full range.
NOTE
For example, if the three decade dials were set
for
9-9-9 on the 1000 !,A range, then the FINE
control could be set to 1500 clockwise to obtain
a current .Y”tp”t of approximately 999.5 +4.
2-3. PRELIMINARY PROCEDURES.
a. Power.
Check the 117-234V Switch for proper linevoltage setting. Check the fuse for proper rating.
Set the Power Switch to off before connecting the power
card.
b. Connections.
The output load should be cwnected
as described in paragraph 2-l.
0972B
Since the stabilitv specification is +.02% (or better)
of reading, the useebie resolution w&d be ~i.2 ,,A.
If the FINE control was set to “O”, then the output
would be 999.0 ti 2.5 ,JA.
g. FILTER Switch (S106). This switch should be
used whenever the Model 225 must deliver current to
an inductive load.
10
5
OPERATION MODEL 225
OPERATING CHARACTERISTICS. The Model 225 Current
2-5.
Source has a bipolar output, a constant cnreent up to
the maximum compliance setting, and automatic crossover to constant voltage mode.
a. Bipolar Output.
accept current and as such it has a bipolar output.
1. Current Source. The instrument can be used as
a current source with a resistive load as shown in
Figure 5.
current set by the front panel controls. Tbe voltage developed across the load resistance
IO x RL.
2.
a current sink as shown in Figure 6. In this case.
the polarity of the current is such that cw’eent
is flowing into the Model 225. The output voltage
“0 = E - IORL, where IC is selected by the front
panel dials, E is the external voltage source, and
RL is the load resistance.
b. Constant Current Mode. Tbe Model 225 will de-
liver the dialed current for resistive, capacitive,
and inductive loads unless the voltage required at the
output terminals exceeds the compliance voltage.
1. Resistive Load.
deliver current to a resistive load from 0 ohms to
w where k = VC/IO and VC is the compliance
voltage
In this case, the
Current Sink.
setting.
The
Model 225 can deliver 0’
Model 225 delivers B
RL is V, =
The instrument can be used as
The instrument can be used to
-‘Iff!?++iJ
___----------
FIGURE 5.
I
Resistive Load.
RI
I
FIGURE 6.
Current Sink.
2.
Capacitfve Load. The instrument can also be
used to charge a capacitive load as shown in Figure
7.
The capacitor will be charged to the compliance
voltage set on the front panel.
3.
Inductive Load. The instrument can
deliver current to an inductive load. When the ratio
of load inductance
than 10-6, the FILTER Switch m”st be set to “ON” for
stability.
citor across the output as shown in Figure 8.
:I
I
I
I
I
The FILTER Switch connects a 1 pF cap*-
L to load
FIGURE 8.
resistance R is greeter
Inductive Load.
be
used
to
FIGURC 7.
FIGURE 9. FIGURE 9.
Capacitive Load.
Floating Supply. Floating Supply.
6
0972
11
MODEL 225
OPERATION
C. constant “Oltsge Mode.
jwted for a compliance voltaza from 10 to 100 volts.
The voltage limiiing will o&r if the load resistance
is such that IORL - "C.
marically switch into a constant voltage mode (the
LIMIT lamp will be lighted) such that the output volt-
age will not exceed the compliance setting. If the
Mcdel 225 is used as a current sink, a,, output voltage
of 130 volts could be developed regardless of the con-
pliance setting. The voltage limiting characteristic
can be shown as in Figure 10.
2-6.
M!JDULATION. For convenience in making dynamic
measurements, the Model 225 has a MODULATE input (JlOl)
so that the current can be modulated in the 50 to 500
HZ range.
a. Circuirry. The Model 225 uses a transformer
coupled input with 500 ohms impedance.
b. Characteristics. The peak-to-peak modulation
can be adjusted up to a maximum of 40% of full range
from 50 CD 100 Hz decreasing to about 8% of full range
at 500 Hz (with a 10 volt rms sine wave input). The
modulation circuit is not calibrated so that the user
m"st check with a knwn load resistance. Within a
selected w the modulation is independent of decade
dial setting.
can affect the modulated current signal so that the
modulation calibration should take into account this
CapC.CitaWX.
C. "Olcage Complia*ce.
compliance can cause clipping of the modulated cur-
rent as shovn in Figure 11.
The capacitance of the output cables
The Model 225 can be ed-
The instrument "ill auto-
The serring of the voltage
t
1.3
,
I
F1mP.E 10.
RL-+
Volt Limit Characteristic
I
FIGURE 11.
Modulation.
I
I
I
NOTE
When the 225 output is modulated the average
value or dc level is determined by the dial set-
rings. The outpllt is essentially a sine-wave at
the input drive frequency. The peak-to-peak
modulation amplitude depends on both the magni-
tude and frequency of the drive signal so that
the exact modulation waveform will have to be
checked tith an actual output load connected.
Clipping of the output signal may occur depending on the compliance voltage setting. The
modulation feature is especially useful when
making dynamic impedance measurements on semiconductor devices, etc.
0972
12
7
Loading...
+ 26 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.