INFICON P3000XL User Manual

TECHNICAL HANDBOOK
kina26e1-p (1209)
Catalog No. 520-001
520-002 520-003 520-004
520-103 520-104 520-105 520-106
from software version V 2.9
Protec P3000(XL)
Helium Sniffer Leak Detector
0-2
Content
Content 0-3 General Safety Precautions 0-6
1 General Information 1-1
1.1 Introduction 1-1
1.1.1 Intended Use 1-1
1.1.2 Available Configurations 1-2
1.1.3 Technical Data 1-2
1.2 Support by INFICON 1-3
1.2.1 Service Centers 1-5
1.3 Unpacking 1-7
1.3.1 Supplied Equipment 1-7
1.3.2 Accessories 1-8
1.4 Notes on How to Use This Handbook 1-9
1.4.1 Symbols of Vacuum Technology 1-9
1.4.2 Definition of Terms 1-10
1.5 Instrument Views of the Protec P3000 1-11
1.6 Installation 1-12
1.6.1 Set up 1-12
1.6.2 Mechanical Connections 1-13
1.6.3 Electrical Connections 1-16
1.6.4 RS232 Interface 1-17
1.6.5 I/O Port 1-17
2 How the Protec P3000 Works 2-1
2.1 Description of the Functions 2-1
2.2 Description of the Subassemblies 2-1
2.2.1 Backing Pump 2-1
2.2.2 Wise TechnologyTM Sensor 2-1
2.2.3 Valve Holder 2-2
2.2.4 Control Assembly 2-2
2.3 Description of the Displays and User Interfaces 2-2
2.3.1 Main unit display 2-2
2.3.2 SL3000(XL) Sniffer line with probe display 2-4
2.3.3 Built-in PRO-Check reference leak 2-5
3 Operation of the Protec P3000 3-1
3.1 Start-Up 3-1
3.2 Controls on the main display unit 3-2
3.3 Controls on the probe display unit 3-5
3.4 Performing measurements 3-5
3.4.1 Standard Operation Mode 3-6
3.4.2 I•Guide Operating Mode 3-9
3.4.2.1 Starting the I•Guide Mode 3-9
3.4.2.2 Selecting an I•Guide Program 3-10
3.4.2.3 Using an I•Guide Program 3-10
3.4.3 The Info Page 3-13
Content 0-3
3.5 Calibration and Self-Test 3-15
3.5.1 Verifying a calibration (proof function) 3-15
3.5.2 Internal cali bration 3-17
3.5.3 External calibration 3-17
3.6 Standby 3-20
3.7 Shutdown 3-20
3.8 Storage for fast availability as back-up unit 3-20
4 Equipment Settings 4-1
4.1 Menu Structure 4-1
4.2 The Service Menu 4-3
4.3 Selecting gas equivalents and setting trigger values 4-3
4.3.1 Setting gas parameters 4-3
4.3.2 Selecting a set of gas parameters 4-7
4.4 Settings Sub-menu 4-8
4.4.1 Vacuum & Access 4-8
4.4.2 Audio Functions 4-15
4.4.3 Display Settings 4-17
4.4.4 Setting-up / editing an I•Guide Program 4-19
4.4.5 Miscellaneous Settings 4-21
4.5 Interfaces 4-23
4.5.1 Control location 4-23
4.5.2 Recorder outputs 4-23
4.5.3 RS232 Protocol 4-25
4.5.4 Select PLC inputs 4-27
4.5.5 Baud rate & end sign 4-27
4.5.6 PRO-Check (only available in advanced mode) 4-27
4.6 The Info Menu 4-28
4.7 History & Maintenance 4-32
5 Protec P3000 Messages 5-1
5.1 Error Messages and Warnings 5-1
6 Equipment Connections 6-1
6.1 I/O Port (Control Inputs and Outputs) 6-1
6.1.1 Ground connectors 6-2
6.1.2 24V Output 6-2
6.1.3 PLC Inputs 6-2
6.1.4 PLC Outputs 6-3
6.1.4.1 Relay outputs 6-5
6.1.4.2 Recorder Outputs 6-5
6.1.5 How to perform a calibration? 6-6
6.2 RS232 interface 6-7
7 Maintenance 7-1
7.1 Maintenance schedule 7-1
7.2 Exchanging the air filter 7-2
7.3 Exchanging the external fuses 7-4
0-4 Content
7.4 Replacing filters in the sniffer line 7-4
7.4.1 Replacing the felt discs of the capillary filter (for SL3000 only) 7-5
7.4.2 Replacing the felt discs when using the water protection tip (for SL3000 only) 7-7
7.4.3 Checking / replacing the sinter filter (for SL3000 only) 7-7
7.5 Replacing the filter pad of the sniffer tip (for SL3000XL only) 7-9
7.6 Switching the capillary filter (for SL3000 sniffer line only) 7-11
7.6.1 Switching from metal to plastic capillary filter 7-11
7.6.2 Switching from plastic to metal capillary filter 7-12
7.7 Replacing the gas reservoir of the PRO-Check 7-13
8 The gas library 8-1
Content 0-5
General Safety Precautions
Warning
Caution
Warning
Warning
Indicates procedures that must be strictly observed to prevent haza rd s to per sons.
Indicates procedures that must strictly be observed to prevent damage to or destruction of the Protec P3000 leak detector.
Notice Indicates special requirements the user must comply with.
The INFICON Protec P3000 leak detector has been designed for safe and efficient operation when used properly and in accordance with this Technica l Handb ook. It is the responsibility of the user to carefully read and strictly observe all safety precautions described in this chapter and throughout this Technical Handbook. The Protec P3000 must only be operated in the proper condition and under th e conditions described in this Technical Handbook. It must be operated and maintained by train ed personal only. Consult local, state, and national agencies regarding specific requirements and regulations. Adress any further safety, operation and / or maintenance questions to our nearest office.
Failure to observe the following precautions could result in serious personal injury:
Danger of explosion! To use the Protec P3000 in explosion hazard areas could cause ignition of
flammable mixtures. The Protec P3000 must only be operated outside of explosion hazard areas.
Only 3-core mains cables having a protective ground conductor must be used. Operation of the Protec P3000 with the ground conductor unconnected is not permissible.
0-6 General Safety Precautions
Warning
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a
Warning
Warning
Warning
Warning
Warning
close distance as this may cause permanent damage to the eye.
Danger of electric shock. Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not be exeeded or reached to ground or ground equipment conductors. According to the type of in- or outputs lower voltages had to be accepted. For this, please refer to the information given in the responding chapters.
For all maintenance on the Protec P3000, the Protec P3000 must be disconnected from power.
Before exchanging the air filter the Protec P3000 must be disconnected from power.
Before exchanging the fuses the Protec P3000 must be disconnected from power.
General Safety Precautions 0-7
Failure to observe the following precautions could result in damage to the
Caution
Caution
Caution
Caution
Caution
Caution
Caution
equipment:
The Protec P3000 must not be operated while standing in water or when exposed to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Avoid contact of the Protec P3000 with bases, acids and solvents as well as exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Protec P3000, a space of at least 20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying the leak detector at the sides of the main unit must not be covered at any time as these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity of the Protec P3000.
Before connecting the Protec P3000 to the mains you must make sure that the mains voltage rating of the Protec P3000 coincides with the locally available mains voltage.
Do not suck in any liquids.
0-8 General Safety Precautions
Caution
Permissible maximum input voltage PLC 28 V.
Caution
Caution
Caution
Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.
Maximum load rating relay outputs is 60 V DC / 25 V AC and 1 A per relay.
The air filter should be checked for contamination at least every 6 months and should be definitely exchanged after 2 years.
General Safety Precautions 0-9
0-10 General Safety Precautions
1 General Information
Caution
Caution
Caution
The Protec P3000 helium leak detector is supplied ready for operation. However, we recommend that you carefully read the Technical Handbook to ensure optimum operating conditions right from the start. This handbook contains important information on functions, installation, start-up and operation of the Protec P3000.
If not stated otherwise this technical handbook applies to all three configurations of the Protec P3000 (see Section 1.1.2). Sections specific to one configuration are marked as "for … only". Sections marked as "for Protec P300 0XL only" always apply to the Protec P3000XL with the SL3000XL sniffer line (capable of
1.1 Introduction
1.1.1 Intended Use
The Protec P3000 is a helium leak detector for sniffer applications. It may be used to localise and quantify leaks in test samples if there is helium under an overpressure within the test sample and when searching the test sample with a sniffer probe from the outside (sniffer method). The use of this sniffer probe is mandatory for proper operation and it is available as an accessory (Cat. No. 525-001 to 525-004).
HIGH FLOW mode).
The Protec P3000 must not be operated while standing in water or when exp osed to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Avoid contact of the Protec P3000 with bases, acids and solvents as well as exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
General Information 1-1
1.1.2 Available Configurations
The Protec P3000 leak detector is available in four different configurations:
The Standard Pr otec P3000
The standard Protec P3000 should be used for high sensitivity applications. It requires the SL3000 sniffer line.
The Protec P3000, RC version
The Protec P3000, RC version is the standard Protec P3000 but with an external display unit. It requires the SL3000 sniffer line.
The Protec P3000XL
The Protec P3000XL is the leaks at a much higher distance from the potential leak at reduced sensitivity but can be switched back to normal flow at inc reased sensitivity. In order to use both flow modes (high and low) it requires the use of the SL 3000XL sniffer line. It may also be operated with the normal SL3000 sniffer line, in this case the high flow mode is disabled, however.
The Protec P3000XL, RC version
The Protec P3000XL, RC version is identical to the Protec P3000XL, but with an external display unit. It requires the SL3000XL sniffer line.
1.1.3 Technical Data
Physical Data
Lowest detectable leak rate
Measurement range
Helium sensor Wise Technology Sensor response time 450 ms
HIGH FLOW version of the Protec P3000. It is able to detect
For Protec P3000 1 x 10 For Protec P3000XL in For Protec P3000XL in
LOW FLOW mode 1 x 10 HIGH FLOW mode 1 x 10
-7
mbar l/s
-7
mbar l/s
-6
mbar l/s
For Protec P3000 5 decades For Protec P3000XL in
HIGH FLOW mode 4 decades
TM
Sensor
1-2 General Information
Electrical Data
Other data
Gas flow through the capillary
For Protec P3000 260 - 360 sccm* For Protec P3000XL in
HIGH FLOW mode 2660 - 3500 sccm*
Time until ready for operation approx. 3 min
* Measured at 1 atm (1013 mbar) at sea altitude. Actual flow may vary with increased altitude and low atmospheric pressure.
Mains voltages and mains frequencies 90 - 127 V, 50 / 60 Hz (fixed) 115 - 140 V, 60 Hz
187 - 265 V, 50 / 60 Hz Power consumption 200 VA Type of protection IP 20 Overvoltage category II Mains cord 2.5 m Noise level < 54 dBA
Dimensions (w x h x d) in mm 610 x 370 x 265 Weight 27 kg Permissible ambient temperature
10 °C to 45 °C (during operation)
Permissible storage temperature -40 °C to 60 °C Relative humidity max. 80% for temperature up to +31°C,
decreasing linearly to 50% at +40°C Contamination level II
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”) Max. altitude above sea level 2000m
1.2 Support by INFICON
INFICON Service If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health o r whether it is contaminated. If it is contaminated also indicate the nature of the hazard. INFICON must return any equipment without a Declaration of Contamination to the sender’s address. You will find an appropriate form below.
General Information 1-3
General
Declaration of Contamination
Legally binding declaration:
I/we hereby declare that the information on this form is complete and accurate and that I/we will assume any further costs that may arise. The contaminated product will be dispatched in accordance with the applicable regulations.
Organization/company
Address Post code, place
Phone Fax
Email
Name
Date and legally binding signature Company stamp
1) or not containing any amount of hazardous residues that exceed the permissible ex­ posure limits
Process related contamination of product:
toxic no 1) yes
caustic no 1) yes
biological hazard no
yes 2)
explosive no
yes  2)
radioactive no
yes 2)
other harmful substances n o1) yes
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has been submitted. Non-completion will result in delay. This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Copies: Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which the product may have come into contact with:
Trade/product name
Chemical name (or symbol)
Precautions associated with substance
Action if human contact
Description of product
Type Article Number Serial Number
Reason for return
Operating fluid(s) used (Must be drained before shipping.)
The product is free of any sub­stances which are damaging to health yes
This form can be downloaded from our website.
2) Products thus contam i­ nated will not be ac­ cepted without written evidence of decontam i­ nation!
We reserve the right to alter the design or any data given in this handbook. The illustrations are not binding.
Fig. 1-1 Declaration of Contamination
1-4 General Information
1.2.1 Service Centers
Algeria jhj@agramkow.dk Finland jhj@agramkow.dk Agramkow Sonderborg Belarus leakdetection.service@inficon.com France Christophe.Zaffanella@oerlikon.com INFICON GmbH Cologne Belgium leakdetection.service@inficon.com Germany leakdetection.service@inficon.com INFICON GmbH Cologne Brazil fernandoz@prestvacuo.com.br Hungary adam.lovics@kon-trade.hu PV Pest Vácuo Ltda. Santa de Parnaíba Bulgaria leakdetection.service@inficon.com India asdash@hotmail.com INFICON GmbH Cologne Canada reachus@vpcinc.ca Ireland reach.unitedkingdom@inficon.com Vacuum Products Canada Ltd. Ontario Central America infoqro@meisa.com Italy davide.giovanetti@inficon.com MEISA S.a. de C.V. Querètaro China reach.china@inficon.com Israel urimark@mark-tec.co.il INFICON LTD Hong Kong INFICON LTD Beijing INFICON LTD Guangzhou INFICON LTD Shanghai Czech Republic filip.lisec@inficon.com Korea reach.korea@inficon.com INFICON GmbH Pilsen Denmark jhj@agramkow.dk INFICON Ltd.
Agramkow Sonderborg Egypt jhj@agramkow.dk Latvia leakdetection.service@inficon.com Agramkow Sonderborg Estonia leakdetection.service@inficon.com Lithuania leakdetection.service@inficon.com INFICON GmbH Cologne
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +55 114 154 4888 Fax: +55 114 154 4888
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +905.672.7704 Fax: +905.672.2249
Phone: +52 44 22 25 42 80 Fax: +52 44 22 25 41 57
Phone: +852.2862.8863 Fax: +852.2865.6883 Phone: +86.10.6590.0164 Fax: +86.10.6590.0521 Phone: +86.20.8723.6889 Fax: +86.20.8723.6003 Phone: +86.21.6209.3094 Fax: +86.21.6295.2852
Phone +420 734 331 758 Fax: +420 604 203 037
Phone: +45 744 336 36 Fax: +45 744 336 46
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Agramkow Sonderborg
OLV France Orsay
INFICON GmbH Cologne
Kontrade Budaörs
Dashpute 400 064
INFICON Blackburn
INFICON GmbH Castelnuovo
Mark Technologies Ltd. Kiriat Ono Japan reach.japan@inficon.com
INFICON Co. Ltd. Yokohama
INFICON Ltd. Sungnam city
Suwon City INFICON Ltd. Cheonan City
INFICON GmbH Cologne
INFICON GmbH Cologne
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +33 476 351 584 Fax: +33 476 351 584
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +36 23 50 38 80 Fax: +36 23 50 38 96
Phone: +91 22 888 0324 Fax: +91 22 888 0324
Phone: +44 1254 678 250 Fax: +44 1254 698 577
Phone: +39 045 6 40 25 56 Fax: +39 045 6 40 24 21
Phone: +972 35 34 68 22 Fax: +972 35 34 25 89
Phone: +81.45.471.3396 Fax: +81.45.471.3387
Phone: +82 312 062 890 Fax: +82 312 063 058 Phone: +82 312 062 890 Fax: +82 312 063 058 Phone: +82 312 062 890 Fax: +82 312 063 058
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
General Information 1-5
Mexico infoqro@meisa.com Spain richard.cunill@leyboldoptics.com MEISA S.a. de C.V. Querètaro Netherlands leakdetection.service@inficon.com Sweden jhj@agramkow.dk INFICON GmbH Cologne Norway jhj@agramkow.dk Syria leakdetection.service@inficon.com Agramkow Sonderborg Poland kamola@vakpol.com Taiwan Susan.Chang@inficon.com VAK-POL & GAZ Sp. zo.o Pulawy Portugal leakdetection.service@inficon.com Tunisia leakdetection.service@inficon.com INFICON GmbH Cologne Republic of South Africa vacuquip@hotmail.com Turkey jhj@agramkow.dk Vacuquip Randburg Russia leakdetection.service@inficon.com Ukraine leakdetection.service@inficon.com INFICON GmbH Cologne Singapore reach.singapore@inficon.com United Kingdom reach.unitedkingdom@inficon.com INFICON PTE LTD. Singapur Slovakia filip.lisec@inficon.com United Arab Emirates leakdetection.service@inficon.com INFICON GmbH Pilsen Slovenia medivak@siol.net USA service.usa@inficon.com Medivac Ljubljani South America except Brazil infoqro@meisa.com Inficon Inc.
MEISA S.a. de C.V. Querètaro
Phone: +52 442 225 42 80 Fax: +52 442 225 41 57
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +48 60 23 15 212 Fax: +48 60 23 15 212
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +27 73 15 78 355 Agramkow
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +65.890.6250 Fax: +65.890.6266
Phone +420 734 331 758 Fax: +420 604 203 037
Phone: +386 15 63 91 50 Fax: +386 17 22 04 51
Phone: +52 44 22 12 36 15 Fax: +52 44 22 12 19 40
Leybold Optics Ibérica Barcelona
Agramkow Sonderborg
INFICON GmbH Cologne
INFICON Company Limited Chupei City, HsinChu Hsien
INFICON GmbH Cologne
Sonderborg
INFICON GmbH Cologne
INFICON Blackburn
INFICON GmbH Cologne
Inficon Inc. East Syracuse, NY
San Jose, CA Inficon Inc. Austin, TX
Phone: +34 93 66 60 778 Fax: +34 93 66 64 612
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +886.3.5525.828 Fax: +886.3.5525.829
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +44 1254 678 250 Fax: +44 1254 698 577
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +1.315.434.1167 Fax: +1.315.434.2551 Phone: +1.408.361.1200 Fax: +1.408.362.1556 Phone: +1.512.448.0488 Fax: +1.512.448.0398
1-6 General Information
1.3 Unpacking
Unpack the Protec P3000 leak detector immediately after it has been received even if it is to be put into operation at some later date. Examine the shipping container for any external damage. Completely remove all packaging materials.
Notice Retain the shipping container and the packaging materials in the event of
possible complaints concerning any damages.
Check if the Protec P3000 leak detector is complete (see Section 1.3.1) and carefully subject it to a visual inspection. If any damage is discovered please immediately inform the forwarding agent and the insurers. If it is required to exchange the damaged part please contact our or ders department.
1.3.1 Supplied Equipment
The Protec P3000 leak detector is ready for operation. Before installation please read Section 1.5. Included with the leak detector are the following items:
Protec P3000 (main unit)
Mains cord, 3m long
Set of fuses (3 x 10 pcs.)
Spare air filter
8 mm hexagonal wrench
Documentation
Operating Instructions (kima26)Technical Handbook Protec P3000 (kina26e1)Spare Parts List Protec P3000 (kiua26d2)Interface Description Protec P3000 (kins26e1)
Notice The SL3000 and SL3000XL sniffer lines are available in different
configurations and need to be ordered separately in the desired length. The SL3000(XL) sniffer line is not part of the Protec P3000 shipment. (see Section 1.3.2 Accessories)
Notice The PRO-Check reference leak is an accessory (see Section 1.3.2
Accessories) and needs to be ordered seperately
Notice For the RC versions the display unit and the co nnecting cable ar e not part
of the standard Protec P3000 shipment and need to be ordered separate ly (see Section 1.3.2)
General Information 1-7
1.3.2 Accessories
For Protec P3000
SL3000 Sniffer line for Protec P3000 Cat. No. / Ref. No.
SL3000-3, 3 m length 525-001 SL3000-5, 5 m length 525-002 SL3000-10, 10 m length 525-003 SL3000-15, 15 m length 525-004
Sniffer tips for SL3000
ST 312, 120 mm, rigid 122 13 FT 312, 120 mm, flexible 122 14 FT 200, 200 mm, rigid 122 18 FT 250, 250 mm, flexible 122 66 ST 385, 385 mm, rigid 122 15 FT 385, 385 mm, flexible 122 16 FT 600, 600 mm, flexible 122 09 ST 500, 500 mm, rigid, 45° angled 122 75
Water protection tip for sniffer 122 46
For Protec P3000XL only
SL3000XL Sniffer line for Protec P3000XL
Sniffer tips for SL3000XL
For RC versions only
External display unit
Connecting cable for external display unit
Connecting cable for external display unit
SL3000XL-3, 3 m length 521-011 SL3000XL-5, 5 m length 521-012 SL3000XL-10, 10 m length 521-013 SL3000XL-15, 15 m length 521-014
ST 312XL, 120 mm, rigid 122 80 FT 312XL, 120 mm, flexible 122 81 ST 385XL, 385 mm, rigid 122 82 FT 385XL, 385 mm, flexible 122 83 FT 250XL, 250 mm, flexible 122 85
for benchtop use 551-100 for rack mounting 551-101
5m length 551-102
0.7m length 551-003
1-8 General Information
For all Protec P3000
Warning
Caution
configurations
Holder for SL3000(XL) 525-006 PRO-Check reference leak for helium 521-001 Calibrated sniffer leaks for helium
S-TL 4, approx.1 x 10 S-TL 5, range 10 S-TL 6, range 10
-4
mbar l/s 122 37
-5
mbar l/s 122 38
-6
mbar l/s 122 39
1.4 Notes on How to Use This Handbook
Important remarks concerning operational safety and protection are emphasised as follows:
Notice Indicates special requirements the user must comply with.
Indicates procedures that must be strictly observed to prevent hazards to persons.
Indicates procedures that must strictly be observed to prevent damage to or destruction of the Protec P3000 leak detector.
The references to diagrams, e.g. (2-1/6) consist of the Section No., Fig. No. and the Item No. in that order. For example: (2-1/6) means Section 2, Fig. 1 and Item No. 6 (here: mains switch).
1.4.1 Symbols of Vacuum Technology
In the following some important symbols of vacuum technology as used in this handbook are shown:
Diaphragm pump Vacuum gauge
General Information 1-9
1.4.2 Definition of Terms
Main menu
This menu is shown first after operating the Menu push-button.
Sub-menus
Comprise all menus which may be accessed from the main menu. Unauthorised changes to many of these sub-menus may be prevented by a password (see also Section 4.3.1).
Menu item
A single menu line.
Default condition
Status of the Protec P3000 when supplied from the factory.
Service menu
Comprises the menu lines in the “Service” sub-menu. The service menu is accessed by scrolling in the basic menu using the navigation push-b uttons (see also Section
3.2).
Autozero
Determination and compensation of the helium background. With this function, the internal ZERO level of the leak rate signal is determined in order to avoid a readout of the internal helium background and mistaking it as a properly measured value. If subsequently negative leak rates are obtained due to this correction , the stored offset values are changed so that ZERO will be the lowest value which can be obtained. In this way the values adapt automatically to a decaying background (adaptive background correction).
Internal background
The existing partial pressure in the measurement system. The level of the internal background is measured all the time and subtracted from the measured signal.
I•Guide Mode
In the I•Guide Mode different testing plans can be pre-programmed. During testing the operator is then constantly prompted for the next action and thus guid ed through the testing plan.
Unit under test
Test object that needs to be leak checked.
Display limit
Limits the measurement data displayed depending on the unit of measurement and the operator settings.
1-10 General Information
1.5 Instrument Views of the Protec P3000
1
2
3
4
5
Fig. 1-2 Instrument views of Protec P3000 Pos. Description Pos. Description
1 Main display 4 Lemo Connector for SL3000 sniffer line 2 Speaker 5 Handle fore carrying the Protec P3000 3 PRO-Check reference leak
General Information 1-11
1.6 Installation
Caution
Caution
1.6.1 Set up
How to remove the transportation lock: The transportation lock is located on the bottom side of the Protec P3000 and
consists of a yellow knurled screws. Please remove this screw before starting-up the leak detector. The Protec P3000 is supplied ready for operation. Initial start-up is described in Section 3.1.
Fig. 1-3 Removing the transportation lock before starting
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Protec P3000, a space of at least 20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying the leak detector at the sides of the main unit must not be covered at any time as these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity of the Protec P3000.
1-12 General Information
1.6.2 Mechanical Connections
Caution
Twisting the sniffer line spoils the cable. Do not twist the sniffer line.
SL3000(XL) Sniffer line
In order to operate the Protec P3000 it is essential for the SL3000(XL) sniffer line to be connected. The connection for the sn iffe r line is loca te d at the fro nt of the Prot ec P3000 left of the PRO-Check reference leak. Insert th e plug into the opening with the red dot on the plug and the slot in the front cover aligned until the connector engage s. To disconnect the plug, retract the coupling and remove the probe’s line.
Water protection tip (optional)
If you intend to perform leak testing on parts that are not completely dry (e.g. due to condensation after performance testing), we strongly recommend to use a water protection tip.
To install the water protection tip,
1 screw off the metallic capillary filter at the very end of the sniffer tip and 2 install the water protection tip instead.
Notice Please do not forget to re-install the little rubber se al when switching to the
water protection tip.
Fig. 1-4 Installing water protection tip
General Information 1-13
PRO-Check Reference leak (optional)
Please insert the PRO-Check reference leak into the opening in the housing of the main unit. Make sure that the Sub-D plug is pr operly connected with the PRO-Ch eck leak.
Notice When properly inserted, the PRO-Check reference leak will still protrude by
On first usage of your PRO-Check you need to initialize the use of this reference leak in the Protec P3000 software.
Please perform the following steps:
1 Insert the PRO-Check into the appropriate opening of the Protec P3000 2 In the software menu go to HISTORY & MAINTENANCE / REPLACE PRO-CHECK.
approx. 10 mm.
Notice This menu item is only available when the Protec P3000 is set to
MODE (see Section 4.4.3, Keyword: User Mode)
ADVANCED
3 On the certificate, which is delivered with the PRO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open submenu and the 12-digit-code in the sec on d line an d press OK.
Notice The PRO-Check reference leak must be installed in the Protec P3000
when pressing
Fig. 1-5 Initializing the PRO-Check reference leak
OK.
Notice PRO-Check Warntime Expire Date (See Section 7.7.5).
For RC versions only
The RC version has no built-in display unit but a connectors plate is mounted instead . Please connect the external display unit with the 5m connecting table (Cat.-no. 551-
002)
1-14 General Information
Fig. 1-6 Protec P3000 with external display unit for: (a) benchtop use (left side), (b) rack mounting (rigth side)
Holder for SL3000 sniffer line (optional)
An optional holder for the SL3000 sniffer line is available as cat.-no. 525-006. The holder may be installed on the right or left sid e of the main unit (for right- or left handed operators) as shown in Fig. 1-8.
Fig. 1-7 Usage of sniffer line holder The installation is described in Fig. 1-8. There are two little slots on the front side at
the very top area of the blue square front of the main unit. Ho ld the holder horizontally and then insert the two little hooks of the holder into the two slots (either on the right or the left side). With the hooks still inserted, let the holder flap down. It will automatically attach to the metal front by the magnet on the backside of the holder. Now insert the sniffer probe grip into the opening of the holder and let it sink down until it rests in the holder.
General Information 1-15
Fig. 1-8 Installation of sniffer line holder
Caution
1
23 456
1.6.3 Electri cal Conn ectio ns
Fig. 1-9 Electrical connections
Pos. Description Pos. Description 1 Headphone port 4 Name plate 2 I/O Port 5 Power switch 3 RS232 interface 6 Power connector
Notice The local regulations for electrical connections must always be observed
(in Germany VDE 0100). The mains voltage rating for the Protec P3000 can be read off from the name plate left of the power switch. The mains voltage setting of the Protec P3000 is fixed and can not be changed. A separate fuse for each of the mains conductors has been inte g ra te d into the mains socket (Fig. 1-9/6).
The mains voltage is applied to the Protec P3000 via the detachable mains cable which is supplied with the leak detector. A main power socket is available for this purpose at the rear of the main unit.
Before connecting the Protec P3000 to the mains you must make sure that the mains voltage rating of the Protec P3000 coincides with the locally available mains voltage.
1-16 General Information
Warning
Only 3-core mains cables having a protective ground conductor must be used. Operation of the Protec P3000 with the ground conductor unconnected is not permissible.
1.6.4 RS232 Interface
The Protec P3000 is equipped with a RS232 interface which is located on the rear right side of the main unit. This interface is of the DCE type (Data Communications Equipment) and allows the connection of a PC for monitoring and data logging. The connection is provided through a commercially available Sub-D plug. For further information see “Interface Description Protec P3000” (kins26e1).
1.6.5 I/O Port
The I/O port allows communication with and control through external equipment via PLC and analog data. For details see Section 6.1.
Through this connection some functions of the Protec P3000 can be controlled externally or measurement data or the status of the Protec P3000 may be communicated to external equipment.
Through relay changeover contacts the trigger levels as well as the operating mode (Ready) of the Protec P3000 may be monitored.
General Information 1-17
2 How the Protec P3000 Works
2.1 Description of the Functions
The Protec P3000 takes in helium through the SL3000 sniffer line, detects the amount of helium by means of a helium sensitive sensor and converts the sensor signal into quantitative leak rates.
The Protec P3000 is composed of the following principal subassemblies:
A Wise Technology sensor as helium detector
A valve holder for controlling different operating stati
A diaphragm pump for pumping the gas to the sensor
An inlet system for the gas flow
The correspond ing electrical and electronic subassemblies for supplying power and for signal conditioning.
The detector operates under gross vacuum conditions, i. e. the operating pressure at the detector is several 100 mbar. T he under pressure is generated by a diaphragm pump. The pressure in front of the sensor is measured with a piezo-resistant pressure gauge and is about 250 mbar under normal operating conditions.
2.2 Description of the Subassemblies
2.2.1 Backing Pump
A diaphragm pump in the Protec P3000 serves as the backing pump. All data and further information on this pump are given in the Operating Instructions. The backing pump generates the flow of the gas through the SL3000 sniffer line.
2.2.2 Wise TechnologyTM Sensor
The helium detector (Wise TechnologyTM sensor) consists of a closed glass container with a measurement device for the precise determination of the pressure inside the glass housing and a mem brane chip wit h a large nu mber of small quartz windows. The membrane is permeable only for helium, all other components of air are retained by the membrane and from the inside of the glass housing. The Quartz membrane is heated so that the permeation for helium is sufficiently high and fast.
Inside of the glass housing the total pressure is measured precisely. As only helium can enter the glass housing the total pressure is equal to the partial pressure of helium. The determined total pressure inside the housing is proportional to the helium partial pressure outside the sensor.
2-1 How the Protec P3000 Works
2.2.3 Valve Holder
The valve holder carries the valves that control the gas flow to the Wise Technology sensor. These valves are used to select the sensitivity of the system, to activate a protection mode against high helium contamination and to set the system into
STANDBY mode. The Protec P3000 software continuously analyses the situtation and
sets the correct valve position via the control unit.
2.2.4 Control Assembly
The control assembly (microprocessor) is the central assembly of the Protec P3000’s electronics. All other subassemblies are controlled and monitored by this assembly. The microprocessor which is located here is thus continuously informed about the status of the entire Protec P3000 and can respond accordingly. In order to accept commands from the operator and to output measured values and messages, the control subassembly is linked to the display unit.
2.3 Description of the Displays and User Interfaces
2.3.1 Main unit display
This subassembly is used to communicate with the operator. It accepts commands from the 8 keys on both sides of the display and outputs measurement results and messages via the display.
Fig. 2-1 Main unit display Pos.Description Pos. Description
1 Menu buttons 1 to 4 2 Menu buttons 5 to 8
How the Protec P3000 Works 2-2
For RC versions only
The RC version has a connectors plate for the external display unit instead of the built-in main unit display. Two LED's left of the plug provide information about the status of the Protec P3000, even when the external display unit is disconnected:
The green LED will indicate that the Protec P3000 is in operation (switched on).
The red LED will be blinking in case of an error message, continuous red light
If no display unit is connected, error messages or warnings may be acknowledged by pressing both buttons of the SL3000 sniffer line simultaneously.
The external display unit also offers four buttons:
The START / STOP buttons have no function (the external displa y unit may also
The MENU button will open the software menu.
The ZERO button will set the current background reading to zero. (For details on
The green LED will show continuous green light if an external display unit is connected and will blink if no external display unit is detected.
indicates a warning.
be used with other INFICON leak detectors which need these buttons)
the ZERO function see Section 4.4.1)
Fig. 2-2 Conncetor plate with LEDs
2-3 How the Protec P3000 Works
Loading...
+ 102 hidden pages