2.3Description of the Displays and User Interfaces 2-2
2.3.1Main unit display 2-2
2.3.2SL3000(XL) Sniffer line with probe display 2-4
2.3.3Built-in PRO-Check reference leak 2-5
3 Operation of the Protec P3000 3-1
3.1Start-Up 3-1
3.2Controls on the main display unit 3-2
3.3Controls on the probe display unit 3-5
3.4Performing measurements 3-5
3.4.1Standard Operation Mode 3-6
3.4.2I•Guide Operating Mode 3-9
3.4.2.1Starting the I•Guide Mode 3-9
3.4.2.2Selecting an I•Guide Program 3-10
kina26eIVZ.fmTechnical Handbook(1209)
3.4.2.3Using an I•Guide Program 3-10
3.4.3The Info Page 3-13
Content0-3
3.5Calibration and Self-Test 3-15
3.5.1Verifying a calibration (proof function) 3-15
3.5.2Internal cali bration 3-17
3.5.3External calibration 3-17
3.6Standby 3-20
3.7Shutdown 3-20
3.8Storage for fast availability as back-up unit 3-20
4 Equipment Settings 4-1
4.1Menu Structure 4-1
4.2The Service Menu 4-3
4.3Selecting gas equivalents and setting trigger values 4-3
4.3.1Setting gas parameters 4-3
4.3.2Selecting a set of gas parameters 4-7
4.4Settings Sub-menu 4-8
4.4.1Vacuum & Access 4-8
4.4.2Audio Functions 4-15
4.4.3Display Settings 4-17
4.4.4Setting-up / editing an I•Guide Program 4-19
4.4.5Miscellaneous Settings 4-21
4.5Interfaces 4-23
4.5.1Control location 4-23
4.5.2Recorder outputs 4-23
4.5.3RS232 Protocol 4-25
4.5.4Select PLC inputs 4-27
4.5.5Baud rate & end sign 4-27
4.5.6PRO-Check (only available in advanced mode) 4-27
4.6The Info Menu 4-28
4.7History & Maintenance 4-32
5 Protec P3000 Messages 5-1
5.1Error Messages and Warnings 5-1
6 Equipment Connections 6-1
6.1I/O Port (Control Inputs and Outputs) 6-1
6.1.1Ground connectors 6-2
6.1.224V Output 6-2
6.1.3PLC Inputs 6-2
6.1.4PLC Outputs 6-3
6.1.4.1Relay outputs 6-5
6.1.4.2Recorder Outputs 6-5
6.1.5How to perform a calibration? 6-6
6.2RS232 interface 6-7
7 Maintenance 7-1
7.1Maintenance schedule 7-1
7.2Exchanging the air filter 7-2
7.3Exchanging the external fuses 7-4
kina26eIVZ.fmTechnical Handbook(1209)
0-4Content
7.4Replacing filters in the sniffer line 7-4
7.4.1Replacing the felt discs of the capillary filter
(for SL3000 only) 7-5
7.4.2Replacing the felt discs when using the water
protection tip (for SL3000 only) 7-7
7.4.3Checking / replacing the sinter filter
(for SL3000 only) 7-7
7.5Replacing the filter pad of the sniffer tip
(for SL3000XL only) 7-9
7.6Switching the capillary filter
(for SL3000 sniffer line only) 7-11
7.6.1Switching from metal to plastic capillary filter 7-11
7.6.2Switching from plastic to metal capillary filter 7-12
7.7Replacing the gas reservoir of the
PRO-Check 7-13
8 The gas library 8-1
kina26eIVZ.fmTechnical Handbook(1209)
Content0-5
General Safety Precautions
Warning
Caution
Warning
Warning
Indicates procedures that must be strictly observed to prevent haza rd s to per sons.
Indicates procedures that must strictly be observed to prevent damage to or
destruction of the Protec P3000 leak detector.
NoticeIndicates special requirements the user must comply with.
The INFICON Protec P3000 leak detector has been designed for safe and efficient
operation when used properly and in accordance with this Technica l Handb ook. It is
the responsibility of the user to carefully read and strictly observe all safety
precautions described in this chapter and throughout this Technical Handbook. The
Protec P3000 must only be operated in the proper condition and under th e conditions
described in this Technical Handbook. It must be operated and maintained by train ed
personal only. Consult local, state, and national agencies regarding specific
requirements and regulations. Adress any further safety, operation and / or
maintenance questions to our nearest office.
Failure to observe the following precautions could result in serious personal
injury:
Danger of explosion!
To use the Protec P3000 in explosion hazard areas could cause ignition of
flammable mixtures.
The Protec P3000 must only be operated outside of explosion hazard areas.
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Protec P3000 with the ground conductor unconnected is not
permissible.
kina26eIVZ.fmTechnical Handbook(1209)
0-6General Safety Precautions
Warning
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a
Warning
Warning
Warning
Warning
Warning
close distance as this may cause permanent damage to the eye.
Danger of electric shock.
Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not
be exeeded or reached to ground or ground equipment conductors.
According to the type of in- or outputs lower voltages had to be accepted. For this,
please refer to the information given in the responding chapters.
For all maintenance on the Protec P3000, the Protec P3000 must be disconnected
from power.
Before exchanging the air filter the Protec P3000 must be disconnected from
power.
Before exchanging the fuses the Protec P3000 must be disconnected from power.
kina26eIVZ.fmTechnical Handbook(1209)
General Safety Precautions0-7
Failure to observe the following precautions could result in damage to the
Caution
Caution
Caution
Caution
Caution
Caution
Caution
equipment:
The Protec P3000 must not be operated while standing in water or when exposed
to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Avoid contact of the Protec P3000 with bases, acids and solvents as well as
exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Protec P3000, a space of at least
20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear
must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying
the leak detector at the sides of the main unit must not be covered at any time as
these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity
of the Protec P3000.
Before connecting the Protec P3000 to the mains you must make sure that the
mains voltage rating of the Protec P3000 coincides with the locally available mains
voltage.
Do not suck in any liquids.
0-8General Safety Precautions
kina26eIVZ.fmTechnical Handbook(1209)
Caution
Permissible maximum input voltage PLC 28 V.
Caution
Caution
Caution
Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.
Maximum load rating relay outputs is 60 V DC / 25 V AC and 1 A per relay.
The air filter should be checked for contamination at least every 6 months and
should be definitely exchanged after 2 years.
kina26eIVZ.fmTechnical Handbook(1209)
General Safety Precautions0-9
0-10General Safety Precautions
kina26eIVZ.fmTechnical Handbook(1209)
1General Information
Caution
Caution
Caution
The Protec P3000 helium leak detector is supplied ready for operation. However, we
recommend that you carefully read the Technical Handbook to ensure optimum
operating conditions right from the start. This handbook contains important
information on functions, installation, start-up and operation of the Protec P3000.
If not stated otherwise this technical handbook applies to all three configurations of
the Protec P3000 (see Section 1.1.2). Sections specific to one configuration are
marked as "for … only". Sections marked as "for Protec P300 0XL only" always apply
to the Protec P3000XL with the SL3000XL sniffer line (capable of
1.1Introduction
1.1.1Intended Use
The Protec P3000 is a helium leak detector for sniffer applications. It may be used to
localise and quantify leaks in test samples if there is helium under an overpressure
within the test sample and when searching the test sample with a sniffer probe from
the outside (sniffer method). The use of this sniffer probe is mandatory for proper
operation and it is available as an accessory (Cat. No. 525-001 to 525-004).
HIGHFLOW mode).
The Protec P3000 must not be operated while standing in water or when exp osed
to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Avoid contact of the Protec P3000 with bases, acids and solvents as well as
exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-1
1.1.2Available Configurations
The Protec P3000 leak detector is available in four different configurations:
The Standard Pr otec
P3000
The standard Protec P3000 should be used for high sensitivity applications. It
requires the SL3000 sniffer line.
The Protec P3000,
RC version
The Protec P3000, RC version is the standard Protec P3000 but with an external
display unit. It requires the SL3000 sniffer line.
The Protec P3000XL
The Protec P3000XL is the
leaks at a much higher distance from the potential leak at reduced sensitivity but can
be switched back to normal flow at inc reased sensitivity. In order to use both flow
modes (high and low) it requires the use of the SL 3000XL sniffer line. It may also be
operated with the normal SL3000 sniffer line, in this case the high flow mode is
disabled, however.
The Protec P3000XL,
RC version
The Protec P3000XL, RC version is identical to the Protec P3000XL, but with an
external display unit. It requires the SL3000XL sniffer line.
1.1.3Technical Data
Physical Data
Lowest detectable leak rate
Measurement range
Helium sensorWise Technology
Sensor response time450 ms
HIGHFLOW version of the Protec P3000. It is able to detect
For Protec P30001 x 10
For Protec P3000XL in
For Protec P3000XL in
LOWFLOW mode1 x 10
HIGHFLOW mode 1 x 10
-7
mbar l/s
-7
mbar l/s
-6
mbar l/s
For Protec P30005 decades
For Protec P3000XL in
HIGHFLOW mode 4 decades
TM
Sensor
1-2General Information
kina26e Chapter 1.fmTechnical Handboook(1209)
Electrical Data
Other data
Gas flow through the capillary
For Protec P3000260 - 360 sccm*
For Protec P3000XL in
HIGHFLOW mode 2660 - 3500 sccm*
Time until ready for operationapprox. 3 min
* Measured at 1 atm (1013 mbar) at sea altitude. Actual flow may vary with increased
altitude and low atmospheric pressure.
187 - 265 V, 50 / 60 Hz
Power consumption200 VA
Type of protectionIP 20
Overvoltage categoryII
Mains cord2.5 m
Noise level< 54 dBA
Dimensions (w x h x d) in mm610 x 370 x 265
Weight27 kg
Permissible ambient temperature
10 °C to 45 °C
(during operation)
Permissible storage temperature -40 °C to 60 °C
Relative humiditymax. 80% for temperature up to +31°C,
decreasing linearly to 50% at +40°C
Contamination levelII
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”)
Max. altitude above sea level2000m
1.2Support by INFICON
INFICON Service
If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health o r whether
it is contaminated. If it is contaminated also indicate the nature of the hazard.
INFICON must return any equipment without a Declaration of Contamination to the
sender’s address. You will find an appropriate form below.
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-3
General
Declaration of Contamination
Legally binding declaration:
I/we hereby declare that the information on this form is complete and accurate and that I/we will assume any further costs that may
arise. The contaminated product will be dispatched in accordance with the applicable regulations.
Organization/company
Address Post code, place
Phone Fax
Email
Name
Date and legally binding signature Company stamp
1) or not containing any amount
of hazardous residues that
exceed the permissible ex posure limits
Process related contamination of product:
toxic no 1)yes
caustic no 1) yes
biological hazard no
yes 2)
explosive no
yes 2)
radioactive no
yes 2)
other harmful substances n o1) yes
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has
been submitted. Non-completion will result in delay.
This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Copies:
Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which the product may have come into contact with:
Trade/product name
Chemical name
(or symbol)
Precautions associated
with substance
Action if human contact
Description of product
Type
Article Number
Serial Number
Reason for return
Operating fluid(s) used (Must be drained before shipping.)
The product is free of any substances which are damaging to
health yes
This form can be downloaded
from our website.
2) Products thus contam i nated will not be ac cepted without written
evidence of decontam i nation!
We reserve the right to alter the design or any data given in this handbook.
The illustrations are not binding.
Fig. 1-1 Declaration of Contamination
1-4General Information
kina26e Chapter 1.fmTechnical Handboook(1209)
1.2.1Service Centers
Algeriajhj@agramkow.dkFinlandjhj@agramkow.dk
Agramkow
Sonderborg
Belarusleakdetection.service@inficon.com FranceChristophe.Zaffanella@oerlikon.com
INFICON GmbH
Cologne
Belgiumleakdetection.service@inficon.com Germanyleakdetection.service@inficon.com
INFICON GmbH
Cologne
Brazilfernandoz@prestvacuo.com.brHungaryadam.lovics@kon-trade.hu
PV Pest Vácuo Ltda.
Santa de Parnaíba
Bulgarialeakdetection.service@inficon.com Indiaasdash@hotmail.com
INFICON GmbH
Cologne
Canadareachus@vpcinc.caIrelandreach.unitedkingdom@inficon.com
Vacuum Products Canada Ltd.
Ontario
Central Americainfoqro@meisa.comItalydavide.giovanetti@inficon.com
MEISA S.a. de C.V.
Querètaro
Chinareach.china@inficon.comIsraelurimark@mark-tec.co.il
INFICON LTD
Hong Kong
INFICON LTD
Beijing
INFICON LTD
Guangzhou
INFICON LTD
Shanghai
Czech Republicfilip.lisec@inficon.comKoreareach.korea@inficon.com
INFICON GmbH
Pilsen
Denmarkjhj@agramkow.dkINFICON Ltd.
Unpack the Protec P3000 leak detector immediately after it has been received even
if it is to be put into operation at some later date. Examine the shipping container for
any external damage. Completely remove all packaging materials.
NoticeRetain the shipping container and the packaging materials in the event of
possible complaints concerning any damages.
Check if the Protec P3000 leak detector is complete (see Section 1.3.1) and carefully
subject it to a visual inspection. If any damage is discovered please immediately
inform the forwarding agent and the insurers. If it is required to exchange the
damaged part please contact our or ders department.
1.3.1Supplied Equipment
The Protec P3000 leak detector is ready for operation. Before installation please
read Section 1.5. Included with the leak detector are the following items:
NoticeThe SL3000 and SL3000XL sniffer lines are available in different
configurations and need to be ordered separately in the desired length. The
SL3000(XL) sniffer line is not part of the Protec P3000 shipment. (see
Section 1.3.2 Accessories)
NoticeThe PRO-Check reference leak is an accessory (see Section 1.3.2
Accessories) and needs to be ordered seperately
NoticeFor the RC versions the display unit and the co nnecting cable ar e not part
of the standard Protec P3000 shipment and need to be ordered separate ly
(see Section 1.3.2)
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-7
1.3.2Accessories
For Protec P3000
SL3000 Sniffer line for Protec P3000Cat. No. / Ref. No.
SL3000-3, 3 m length525-001
SL3000-5, 5 m length525-002
SL3000-10, 10 m length525-003
SL3000-15, 15 m length525-004
Sniffer tips for SL3000
ST 312, 120 mm, rigid122 13
FT 312, 120 mm, flexible122 14
FT 200, 200 mm, rigid122 18
FT 250, 250 mm, flexible122 66
ST 385, 385 mm, rigid122 15
FT 385, 385 mm, flexible122 16
FT 600, 600 mm, flexible122 09
ST 500, 500 mm, rigid, 45° angled122 75
Water protection tip for sniffer122 46
For Protec P3000XL only
SL3000XL Sniffer line for Protec P3000XL
Sniffer tips for SL3000XL
For RC versions only
External display unit
Connecting cable for external display unit
Connecting cable for external display unit
SL3000XL-3, 3 m length521-011
SL3000XL-5, 5 m length521-012
SL3000XL-10, 10 m length521-013
SL3000XL-15, 15 m length521-014
ST 312XL, 120 mm, rigid122 80
FT 312XL, 120 mm, flexible122 81
ST 385XL, 385 mm, rigid122 82
FT 385XL, 385 mm, flexible122 83
FT 250XL, 250 mm, flexible122 85
for benchtop use551-100
for rack mounting551-101
5m length551-102
0.7m length551-003
kina26e Chapter 1.fmTechnical Handboook(1209)
1-8General Information
For all Protec P3000
Warning
Caution
configurations
Holder for SL3000(XL)525-006
PRO-Check reference leak for helium521-001
Calibrated sniffer leaks for helium
S-TL 4, approx.1 x 10
S-TL 5, range 10
S-TL 6, range 10
-4
mbar l/s122 37
-5
mbar l/s122 38
-6
mbar l/s122 39
1.4Notes on How to Use This Handbook
Important remarks concerning operational safety and protection are emphasised as
follows:
NoticeIndicates special requirements the user must comply with.
Indicates procedures that must be strictly observed to prevent hazards to persons.
Indicates procedures that must strictly be observed to prevent damage to or
destruction of the Protec P3000 leak detector.
The references to diagrams, e.g. (2-1/6) consist of the Section No., Fig. No. and the
Item No. in that order. For example: (2-1/6) means Section 2, Fig. 1 and Item No. 6
(here: mains switch).
1.4.1Symbols of Vacuum Technology
In the following some important symbols of vacuum technology as used in this
handbook are shown:
Diaphragm pumpVacuum gauge
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-9
1.4.2Definition of Terms
Main menu
This menu is shown first after operating the Menu push-button.
Sub-menus
Comprise all menus which may be accessed from the main menu. Unauthorised
changes to many of these sub-menus may be prevented by a password (see also
Section 4.3.1).
Menu item
A single menu line.
Default condition
Status of the Protec P3000 when supplied from the factory.
Service menu
Comprises the menu lines in the “Service” sub-menu. The service menu is accessed
by scrolling in the basic menu using the navigation push-b uttons (see also Section
3.2).
Autozero
Determination and compensation of the helium background. With this function, the
internal ZERO level of the leak rate signal is determined in order to avoid a readout
of the internal helium background and mistaking it as a properly measured value. If
subsequently negative leak rates are obtained due to this correction , the stored offset
values are changed so that ZERO will be the lowest value which can be obtained. In
this way the values adapt automatically to a decaying background (adaptive
background correction).
Internal background
The existing partial pressure in the measurement system. The level of the internal
background is measured all the time and subtracted from the measured signal.
I•Guide Mode
In the I•Guide Mode different testing plans can be pre-programmed. During testing
the operator is then constantly prompted for the next action and thus guid ed through
the testing plan.
Unit under test
Test object that needs to be leak checked.
Display limit
Limits the measurement data displayed depending on the unit of measurement and
the operator settings.
1-10General Information
kina26e Chapter 1.fmTechnical Handboook(1209)
1.5Instrument Views of the Protec P3000
1
2
3
4
5
Fig. 1-2 Instrument views of Protec P3000
Pos. DescriptionPos. Description
1Main display4Lemo Connector for SL3000 sniffer line
2Speaker5Handle fore carrying the Protec P3000
3PRO-Check reference leak
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-11
1.6Installation
Caution
Caution
1.6.1Set up
How to remove the transportation lock:
The transportation lock is located on the bottom side of the Protec P3000 and
consists of a yellow knurled screws. Please remove this screw before starting-up the
leak detector. The Protec P3000 is supplied ready for operation. Initial start-up is
described in Section 3.1.
Fig. 1-3 Removing the transportation lock before starting
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Protec P3000, a space of at least
20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear
must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying
the leak detector at the sides of the main unit must not be covered at any time as
these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity
of the Protec P3000.
1-12General Information
kina26e Chapter 1.fmTechnical Handboook(1209)
1.6.2Mechanical Connections
Caution
Twisting the sniffer line spoils the cable.
Do not twist the sniffer line.
SL3000(XL) Sniffer line
In order to operate the Protec P3000 it is essential for the SL3000(XL) sniffer line to
be connected. The connection for the sn iffe r line is loca te d at the fro nt of the Prot ec
P3000 left of the PRO-Check reference leak. Insert th e plug into the opening with the
red dot on the plug and the slot in the front cover aligned until the connector engage s.
To disconnect the plug, retract the coupling and remove the probe’s line.
Water protection tip
(optional)
If you intend to perform leak testing on parts that are not completely dry (e.g. due to
condensation after performance testing), we strongly recommend to use a water
protection tip.
To install the water protection tip,
1 screw off the metallic capillary filter at the very end of the sniffer tip and
2 install the water protection tip instead.
NoticePlease do not forget to re-install the little rubber se al when switching to the
water protection tip.
Fig. 1-4 Installing water protection tip
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-13
PRO-Check Reference
leak (optional)
Please insert the PRO-Check reference leak into the opening in the housing of the
main unit. Make sure that the Sub-D plug is pr operly connected with the PRO-Ch eck
leak.
NoticeWhen properly inserted, the PRO-Check reference leak will still protrude by
On first usage of your PRO-Check you need to initialize the use of this reference leak
in the Protec P3000 software.
Please perform the following steps:
1 Insert the PRO-Check into the appropriate opening of the Protec P3000
2 In the software menu go to HISTORY & MAINTENANCE / REPLACEPRO-CHECK.
approx. 10 mm.
NoticeThis menu item is only available when the Protec P3000 is set to
MODE (see Section 4.4.3, Keyword: User Mode)
ADVANCED
3 On the certificate, which is delivered with the PRO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open
submenu and the 12-digit-code in the sec on d line an d press OK.
NoticeThe PRO-Check reference leak must be installed in the Protec P3000
when pressing
Fig. 1-5 Initializing the PRO-Check reference leak
OK.
NoticePRO-Check Warntime Expire Date (See Section 7.7.5).
For RC versions only
The RC version has no built-in display unit but a connectors plate is mounted instead .
Please connect the external display unit with the 5m connecting table (Cat.-no. 551-
002)
1-14General Information
kina26e Chapter 1.fmTechnical Handboook(1209)
Fig. 1-6 Protec P3000 with external display unit for: (a) benchtop use (left side), (b)
rack mounting (rigth side)
Holder for SL3000
sniffer line (optional)
An optional holder for the SL3000 sniffer line is available as cat.-no. 525-006. The
holder may be installed on the right or left sid e of the main unit (for right- or left
handed operators) as shown in Fig. 1-8.
Fig. 1-7 Usage of sniffer line holder
The installation is described in Fig. 1-8. There are two little slots on the front side at
the very top area of the blue square front of the main unit. Ho ld the holder horizontally
and then insert the two little hooks of the holder into the two slots (either on the right
or the left side). With the hooks still inserted, let the holder flap down. It will
automatically attach to the metal front by the magnet on the backside of the holder.
Now insert the sniffer probe grip into the opening of the holder and let it sink down
until it rests in the holder.
NoticeThe local regulations for electrical connections must always be observed
(in Germany VDE 0100). The mains voltage rating for the Protec P3000
can be read off from the name plate left of the power switch. The mains
voltage setting of the Protec P3000 is fixed and can not be changed. A
separate fuse for each of the mains conductors has been inte g ra te d into
the mains socket (Fig. 1-9/6).
The mains voltage is applied to the Protec P3000 via the detachable mains cable
which is supplied with the leak detector. A main power socket is available for this
purpose at the rear of the main unit.
Before connecting the Protec P3000 to the mains you must make sure that the
mains voltage rating of the Protec P3000 coincides with the locally available mains
voltage.
kina26e Chapter 1.fmTechnical Handboook(1209)
1-16General Information
Warning
Only 3-core mains cables having a protective ground conductor must be used.
Operation of the Protec P3000 with the ground conductor unconnected is not
permissible.
1.6.4RS232 Interface
The Protec P3000 is equipped with a RS232 interface which is located on the rear
right side of the main unit. This interface is of the DCE type (Data Communications
Equipment) and allows the connection of a PC for monitoring and data logging. The
connection is provided through a commercially available Sub-D plug. For further
information see “Interface Description Protec P3000” (kins26e1).
1.6.5I/O Port
The I/O port allows communication with and control through external equipment via
PLC and analog data. For details see Section 6.1.
Through this connection some functions of the Protec P3000 can be controlled
externally or measurement data or the status of the Protec P3000 may be
communicated to external equipment.
Through relay changeover contacts the trigger levels as well as the operating mode
(Ready) of the Protec P3000 may be monitored.
kina26e Chapter 1.fmTechnical Handboook(1209)
General Information1-17
2How the Protec P3000 Works
2.1Description of the Functions
The Protec P3000 takes in helium through the SL3000 sniffer line, detects the
amount of helium by means of a helium sensitive sensor and converts the sensor
signal into quantitative leak rates.
The Protec P3000 is composed of the following principal subassemblies:
•A Wise Technology sensor as helium detector
•A valve holder for controlling different operating stati
•A diaphragm pump for pumping the gas to the sensor
•An inlet system for the gas flow
•The correspond ing electrical and electronic subassemblies for supplying power
and for signal conditioning.
The detector operates under gross vacuum conditions, i. e. the operating pressure
at the detector is several 100 mbar. T he under pressure is generated by a diaphragm
pump. The pressure in front of the sensor is measured with a piezo-resistant
pressure gauge and is about 250 mbar under normal operating conditions.
2.2Description of the Subassemblies
2.2.1Backing Pump
A diaphragm pump in the Protec P3000 serves as the backing pump. All data and
further information on this pump are given in the Operating Instructions. The backing
pump generates the flow of the gas through the SL3000 sniffer line.
2.2.2Wise TechnologyTM Sensor
The helium detector (Wise TechnologyTM sensor) consists of a closed glass
container with a measurement device for the precise determination of the pressure
inside the glass housing and a mem brane chip wit h a large nu mber of small quartz
windows. The membrane is permeable only for helium, all other components of air
are retained by the membrane and from the inside of the glass housing. The Quartz
membrane is heated so that the permeation for helium is sufficiently high and fast.
Inside of the glass housing the total pressure is measured precisely. As only helium
can enter the glass housing the total pressure is equal to the partial pressure of
helium. The determined total pressure inside the housing is proportional to the
helium partial pressure outside the sensor.
2-1How the Protec P3000 Works
kina26e Chapter 2.fmTechnical Handbook(1209)
2.2.3Valve Holder
The valve holder carries the valves that control the gas flow to the Wise Technology
sensor. These valves are used to select the sensitivity of the system, to activate a
protection mode against high helium contamination and to set the system into
STANDBY mode. The Protec P3000 software continuously analyses the situtation and
sets the correct valve position via the control unit.
2.2.4Control Assembly
The control assembly (microprocessor) is the central assembly of the Protec P3000’s
electronics. All other subassemblies are controlled and monitored by this assembly.
The microprocessor which is located here is thus continuously informed about the
status of the entire Protec P3000 and can respond accordingly. In order to accept
commands from the operator and to output measured values and messages, the
control subassembly is linked to the display unit.
2.3Description of the Displays and User
Interfaces
2.3.1Main unit display
This subassembly is used to communicate with the operator. It accepts commands
from the 8 keys on both sides of the display and outputs measurement results and
messages via the display.
Fig. 2-1 Main unit display
Pos.DescriptionPos. Description
1Menu buttons 1 to 42Menu buttons 5 to 8
kina26e Chapter 2.fmTechnical Handbook(1209)
How the Protec P3000 Works2-2
For RC versions only
The RC version has a connectors plate for the external display unit instead of the
built-in main unit display. Two LED's left of the plug provide information about the
status of the Protec P3000, even when the external display unit is disconnected:
•The green LED will indicate that the Protec P3000 is in operation (switched on).
•The red LED will be blinking in case of an error message, continuous red light
If no display unit is connected, error messages or warnings may be acknowledged
by pressing both buttons of the SL3000 sniffer line simultaneously.
The external display unit also offers four buttons:
•The START / STOP buttons have no function (the external displa y unit may also
•The MENU button will open the software menu.
•The ZERO button will set the current background reading to zero. (For details on
The green LED will show continuous green light if an external display unit is
connected and will blink if no external display unit is detected.
indicates a warning.
be used with other INFICON leak detectors which need these buttons)
the ZERO function see Section 4.4.1)
Fig. 2-2 Conncetor plate with LEDs
2-3How the Protec P3000 Works
kina26e Chapter 2.fmTechnical Handbook(1209)
Loading...
+ 102 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.