INFICON P3000XL User Manual

TECHNICAL HANDBOOK
kina26e1-p (1209)
Catalog No. 520-001
520-002 520-003 520-004
520-103 520-104 520-105 520-106
from software version V 2.9
Protec P3000(XL)
Helium Sniffer Leak Detector
0-2
Content
Content 0-3 General Safety Precautions 0-6
1 General Information 1-1
1.1 Introduction 1-1
1.1.1 Intended Use 1-1
1.1.2 Available Configurations 1-2
1.1.3 Technical Data 1-2
1.2 Support by INFICON 1-3
1.2.1 Service Centers 1-5
1.3 Unpacking 1-7
1.3.1 Supplied Equipment 1-7
1.3.2 Accessories 1-8
1.4 Notes on How to Use This Handbook 1-9
1.4.1 Symbols of Vacuum Technology 1-9
1.4.2 Definition of Terms 1-10
1.5 Instrument Views of the Protec P3000 1-11
1.6 Installation 1-12
1.6.1 Set up 1-12
1.6.2 Mechanical Connections 1-13
1.6.3 Electrical Connections 1-16
1.6.4 RS232 Interface 1-17
1.6.5 I/O Port 1-17
2 How the Protec P3000 Works 2-1
2.1 Description of the Functions 2-1
2.2 Description of the Subassemblies 2-1
2.2.1 Backing Pump 2-1
2.2.2 Wise TechnologyTM Sensor 2-1
2.2.3 Valve Holder 2-2
2.2.4 Control Assembly 2-2
2.3 Description of the Displays and User Interfaces 2-2
2.3.1 Main unit display 2-2
2.3.2 SL3000(XL) Sniffer line with probe display 2-4
2.3.3 Built-in PRO-Check reference leak 2-5
3 Operation of the Protec P3000 3-1
3.1 Start-Up 3-1
3.2 Controls on the main display unit 3-2
3.3 Controls on the probe display unit 3-5
3.4 Performing measurements 3-5
3.4.1 Standard Operation Mode 3-6
3.4.2 I•Guide Operating Mode 3-9
3.4.2.1 Starting the I•Guide Mode 3-9
3.4.2.2 Selecting an I•Guide Program 3-10
3.4.2.3 Using an I•Guide Program 3-10
3.4.3 The Info Page 3-13
Content 0-3
3.5 Calibration and Self-Test 3-15
3.5.1 Verifying a calibration (proof function) 3-15
3.5.2 Internal cali bration 3-17
3.5.3 External calibration 3-17
3.6 Standby 3-20
3.7 Shutdown 3-20
3.8 Storage for fast availability as back-up unit 3-20
4 Equipment Settings 4-1
4.1 Menu Structure 4-1
4.2 The Service Menu 4-3
4.3 Selecting gas equivalents and setting trigger values 4-3
4.3.1 Setting gas parameters 4-3
4.3.2 Selecting a set of gas parameters 4-7
4.4 Settings Sub-menu 4-8
4.4.1 Vacuum & Access 4-8
4.4.2 Audio Functions 4-15
4.4.3 Display Settings 4-17
4.4.4 Setting-up / editing an I•Guide Program 4-19
4.4.5 Miscellaneous Settings 4-21
4.5 Interfaces 4-23
4.5.1 Control location 4-23
4.5.2 Recorder outputs 4-23
4.5.3 RS232 Protocol 4-25
4.5.4 Select PLC inputs 4-27
4.5.5 Baud rate & end sign 4-27
4.5.6 PRO-Check (only available in advanced mode) 4-27
4.6 The Info Menu 4-28
4.7 History & Maintenance 4-32
5 Protec P3000 Messages 5-1
5.1 Error Messages and Warnings 5-1
6 Equipment Connections 6-1
6.1 I/O Port (Control Inputs and Outputs) 6-1
6.1.1 Ground connectors 6-2
6.1.2 24V Output 6-2
6.1.3 PLC Inputs 6-2
6.1.4 PLC Outputs 6-3
6.1.4.1 Relay outputs 6-5
6.1.4.2 Recorder Outputs 6-5
6.1.5 How to perform a calibration? 6-6
6.2 RS232 interface 6-7
7 Maintenance 7-1
7.1 Maintenance schedule 7-1
7.2 Exchanging the air filter 7-2
7.3 Exchanging the external fuses 7-4
0-4 Content
7.4 Replacing filters in the sniffer line 7-4
7.4.1 Replacing the felt discs of the capillary filter (for SL3000 only) 7-5
7.4.2 Replacing the felt discs when using the water protection tip (for SL3000 only) 7-7
7.4.3 Checking / replacing the sinter filter (for SL3000 only) 7-7
7.5 Replacing the filter pad of the sniffer tip (for SL3000XL only) 7-9
7.6 Switching the capillary filter (for SL3000 sniffer line only) 7-11
7.6.1 Switching from metal to plastic capillary filter 7-11
7.6.2 Switching from plastic to metal capillary filter 7-12
7.7 Replacing the gas reservoir of the PRO-Check 7-13
8 The gas library 8-1
Content 0-5
General Safety Precautions
Warning
Caution
Warning
Warning
Indicates procedures that must be strictly observed to prevent haza rd s to per sons.
Indicates procedures that must strictly be observed to prevent damage to or destruction of the Protec P3000 leak detector.
Notice Indicates special requirements the user must comply with.
The INFICON Protec P3000 leak detector has been designed for safe and efficient operation when used properly and in accordance with this Technica l Handb ook. It is the responsibility of the user to carefully read and strictly observe all safety precautions described in this chapter and throughout this Technical Handbook. The Protec P3000 must only be operated in the proper condition and under th e conditions described in this Technical Handbook. It must be operated and maintained by train ed personal only. Consult local, state, and national agencies regarding specific requirements and regulations. Adress any further safety, operation and / or maintenance questions to our nearest office.
Failure to observe the following precautions could result in serious personal injury:
Danger of explosion! To use the Protec P3000 in explosion hazard areas could cause ignition of
flammable mixtures. The Protec P3000 must only be operated outside of explosion hazard areas.
Only 3-core mains cables having a protective ground conductor must be used. Operation of the Protec P3000 with the ground conductor unconnected is not permissible.
0-6 General Safety Precautions
Warning
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a
Warning
Warning
Warning
Warning
Warning
close distance as this may cause permanent damage to the eye.
Danger of electric shock. Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not be exeeded or reached to ground or ground equipment conductors. According to the type of in- or outputs lower voltages had to be accepted. For this, please refer to the information given in the responding chapters.
For all maintenance on the Protec P3000, the Protec P3000 must be disconnected from power.
Before exchanging the air filter the Protec P3000 must be disconnected from power.
Before exchanging the fuses the Protec P3000 must be disconnected from power.
General Safety Precautions 0-7
Failure to observe the following precautions could result in damage to the
Caution
Caution
Caution
Caution
Caution
Caution
Caution
equipment:
The Protec P3000 must not be operated while standing in water or when exposed to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Avoid contact of the Protec P3000 with bases, acids and solvents as well as exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Protec P3000, a space of at least 20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying the leak detector at the sides of the main unit must not be covered at any time as these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity of the Protec P3000.
Before connecting the Protec P3000 to the mains you must make sure that the mains voltage rating of the Protec P3000 coincides with the locally available mains voltage.
Do not suck in any liquids.
0-8 General Safety Precautions
Caution
Permissible maximum input voltage PLC 28 V.
Caution
Caution
Caution
Permissible max. voltage and current for open collector outputs are: 28 V; 50 mA.
Maximum load rating relay outputs is 60 V DC / 25 V AC and 1 A per relay.
The air filter should be checked for contamination at least every 6 months and should be definitely exchanged after 2 years.
General Safety Precautions 0-9
0-10 General Safety Precautions
1 General Information
Caution
Caution
Caution
The Protec P3000 helium leak detector is supplied ready for operation. However, we recommend that you carefully read the Technical Handbook to ensure optimum operating conditions right from the start. This handbook contains important information on functions, installation, start-up and operation of the Protec P3000.
If not stated otherwise this technical handbook applies to all three configurations of the Protec P3000 (see Section 1.1.2). Sections specific to one configuration are marked as "for … only". Sections marked as "for Protec P300 0XL only" always apply to the Protec P3000XL with the SL3000XL sniffer line (capable of
1.1 Introduction
1.1.1 Intended Use
The Protec P3000 is a helium leak detector for sniffer applications. It may be used to localise and quantify leaks in test samples if there is helium under an overpressure within the test sample and when searching the test sample with a sniffer probe from the outside (sniffer method). The use of this sniffer probe is mandatory for proper operation and it is available as an accessory (Cat. No. 525-001 to 525-004).
HIGH FLOW mode).
The Protec P3000 must not be operated while standing in water or when exp osed to drip water. The same applies to all other kinds of liquids.
This Protec P3000 should only be used in rooms.
Avoid contact of the Protec P3000 with bases, acids and solvents as well as exposure to extreme climatic conditions.
Ensure sufficient air cooling (see also Section 1.1.2)
General Information 1-1
1.1.2 Available Configurations
The Protec P3000 leak detector is available in four different configurations:
The Standard Pr otec P3000
The standard Protec P3000 should be used for high sensitivity applications. It requires the SL3000 sniffer line.
The Protec P3000, RC version
The Protec P3000, RC version is the standard Protec P3000 but with an external display unit. It requires the SL3000 sniffer line.
The Protec P3000XL
The Protec P3000XL is the leaks at a much higher distance from the potential leak at reduced sensitivity but can be switched back to normal flow at inc reased sensitivity. In order to use both flow modes (high and low) it requires the use of the SL 3000XL sniffer line. It may also be operated with the normal SL3000 sniffer line, in this case the high flow mode is disabled, however.
The Protec P3000XL, RC version
The Protec P3000XL, RC version is identical to the Protec P3000XL, but with an external display unit. It requires the SL3000XL sniffer line.
1.1.3 Technical Data
Physical Data
Lowest detectable leak rate
Measurement range
Helium sensor Wise Technology Sensor response time 450 ms
HIGH FLOW version of the Protec P3000. It is able to detect
For Protec P3000 1 x 10 For Protec P3000XL in For Protec P3000XL in
LOW FLOW mode 1 x 10 HIGH FLOW mode 1 x 10
-7
mbar l/s
-7
mbar l/s
-6
mbar l/s
For Protec P3000 5 decades For Protec P3000XL in
HIGH FLOW mode 4 decades
TM
Sensor
1-2 General Information
Electrical Data
Other data
Gas flow through the capillary
For Protec P3000 260 - 360 sccm* For Protec P3000XL in
HIGH FLOW mode 2660 - 3500 sccm*
Time until ready for operation approx. 3 min
* Measured at 1 atm (1013 mbar) at sea altitude. Actual flow may vary with increased altitude and low atmospheric pressure.
Mains voltages and mains frequencies 90 - 127 V, 50 / 60 Hz (fixed) 115 - 140 V, 60 Hz
187 - 265 V, 50 / 60 Hz Power consumption 200 VA Type of protection IP 20 Overvoltage category II Mains cord 2.5 m Noise level < 54 dBA
Dimensions (w x h x d) in mm 610 x 370 x 265 Weight 27 kg Permissible ambient temperature
10 °C to 45 °C (during operation)
Permissible storage temperature -40 °C to 60 °C Relative humidity max. 80% for temperature up to +31°C,
decreasing linearly to 50% at +40°C Contamination level II
(according to IEC 61010 / Part 1:
“normally only non-conductive pollution
may occur. Occasionally, however, a
temporary conductivity caused by
condensation can be tolerated.”) Max. altitude above sea level 2000m
1.2 Support by INFICON
INFICON Service If equipment is returned to INFICON or an authorised INFICON representative
indicate whether the equipment is free of substances damaging to health o r whether it is contaminated. If it is contaminated also indicate the nature of the hazard. INFICON must return any equipment without a Declaration of Contamination to the sender’s address. You will find an appropriate form below.
General Information 1-3
General
Declaration of Contamination
Legally binding declaration:
I/we hereby declare that the information on this form is complete and accurate and that I/we will assume any further costs that may arise. The contaminated product will be dispatched in accordance with the applicable regulations.
Organization/company
Address Post code, place
Phone Fax
Email
Name
Date and legally binding signature Company stamp
1) or not containing any amount of hazardous residues that exceed the permissible ex­ posure limits
Process related contamination of product:
toxic no 1) yes
caustic no 1) yes
biological hazard no
yes 2)
explosive no
yes  2)
radioactive no
yes 2)
other harmful substances n o1) yes
The service, repair, and/or disposal of vacuum equipment and components will only be carried out if a correctly completed declaration has been submitted. Non-completion will result in delay. This declaration may only be completed (in block letters) and signed by authorized and qualified staff.
Copies: Original for addressee - 1 copy for accompanying documents - 1 copy for file of sender
Harmful substances, gases and/or by-products
Please list all substances, gases, and by-products which the product may have come into contact with:
Trade/product name
Chemical name (or symbol)
Precautions associated with substance
Action if human contact
Description of product
Type Article Number Serial Number
Reason for return
Operating fluid(s) used (Must be drained before shipping.)
The product is free of any sub­stances which are damaging to health yes
This form can be downloaded from our website.
2) Products thus contam i­ nated will not be ac­ cepted without written evidence of decontam i­ nation!
We reserve the right to alter the design or any data given in this handbook. The illustrations are not binding.
Fig. 1-1 Declaration of Contamination
1-4 General Information
1.2.1 Service Centers
Algeria jhj@agramkow.dk Finland jhj@agramkow.dk Agramkow Sonderborg Belarus leakdetection.service@inficon.com France Christophe.Zaffanella@oerlikon.com INFICON GmbH Cologne Belgium leakdetection.service@inficon.com Germany leakdetection.service@inficon.com INFICON GmbH Cologne Brazil fernandoz@prestvacuo.com.br Hungary adam.lovics@kon-trade.hu PV Pest Vácuo Ltda. Santa de Parnaíba Bulgaria leakdetection.service@inficon.com India asdash@hotmail.com INFICON GmbH Cologne Canada reachus@vpcinc.ca Ireland reach.unitedkingdom@inficon.com Vacuum Products Canada Ltd. Ontario Central America infoqro@meisa.com Italy davide.giovanetti@inficon.com MEISA S.a. de C.V. Querètaro China reach.china@inficon.com Israel urimark@mark-tec.co.il INFICON LTD Hong Kong INFICON LTD Beijing INFICON LTD Guangzhou INFICON LTD Shanghai Czech Republic filip.lisec@inficon.com Korea reach.korea@inficon.com INFICON GmbH Pilsen Denmark jhj@agramkow.dk INFICON Ltd.
Agramkow Sonderborg Egypt jhj@agramkow.dk Latvia leakdetection.service@inficon.com Agramkow Sonderborg Estonia leakdetection.service@inficon.com Lithuania leakdetection.service@inficon.com INFICON GmbH Cologne
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +55 114 154 4888 Fax: +55 114 154 4888
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +905.672.7704 Fax: +905.672.2249
Phone: +52 44 22 25 42 80 Fax: +52 44 22 25 41 57
Phone: +852.2862.8863 Fax: +852.2865.6883 Phone: +86.10.6590.0164 Fax: +86.10.6590.0521 Phone: +86.20.8723.6889 Fax: +86.20.8723.6003 Phone: +86.21.6209.3094 Fax: +86.21.6295.2852
Phone +420 734 331 758 Fax: +420 604 203 037
Phone: +45 744 336 36 Fax: +45 744 336 46
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Agramkow Sonderborg
OLV France Orsay
INFICON GmbH Cologne
Kontrade Budaörs
Dashpute 400 064
INFICON Blackburn
INFICON GmbH Castelnuovo
Mark Technologies Ltd. Kiriat Ono Japan reach.japan@inficon.com
INFICON Co. Ltd. Yokohama
INFICON Ltd. Sungnam city
Suwon City INFICON Ltd. Cheonan City
INFICON GmbH Cologne
INFICON GmbH Cologne
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +33 476 351 584 Fax: +33 476 351 584
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +36 23 50 38 80 Fax: +36 23 50 38 96
Phone: +91 22 888 0324 Fax: +91 22 888 0324
Phone: +44 1254 678 250 Fax: +44 1254 698 577
Phone: +39 045 6 40 25 56 Fax: +39 045 6 40 24 21
Phone: +972 35 34 68 22 Fax: +972 35 34 25 89
Phone: +81.45.471.3396 Fax: +81.45.471.3387
Phone: +82 312 062 890 Fax: +82 312 063 058 Phone: +82 312 062 890 Fax: +82 312 063 058 Phone: +82 312 062 890 Fax: +82 312 063 058
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
General Information 1-5
Mexico infoqro@meisa.com Spain richard.cunill@leyboldoptics.com MEISA S.a. de C.V. Querètaro Netherlands leakdetection.service@inficon.com Sweden jhj@agramkow.dk INFICON GmbH Cologne Norway jhj@agramkow.dk Syria leakdetection.service@inficon.com Agramkow Sonderborg Poland kamola@vakpol.com Taiwan Susan.Chang@inficon.com VAK-POL & GAZ Sp. zo.o Pulawy Portugal leakdetection.service@inficon.com Tunisia leakdetection.service@inficon.com INFICON GmbH Cologne Republic of South Africa vacuquip@hotmail.com Turkey jhj@agramkow.dk Vacuquip Randburg Russia leakdetection.service@inficon.com Ukraine leakdetection.service@inficon.com INFICON GmbH Cologne Singapore reach.singapore@inficon.com United Kingdom reach.unitedkingdom@inficon.com INFICON PTE LTD. Singapur Slovakia filip.lisec@inficon.com United Arab Emirates leakdetection.service@inficon.com INFICON GmbH Pilsen Slovenia medivak@siol.net USA service.usa@inficon.com Medivac Ljubljani South America except Brazil infoqro@meisa.com Inficon Inc.
MEISA S.a. de C.V. Querètaro
Phone: +52 442 225 42 80 Fax: +52 442 225 41 57
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +48 60 23 15 212 Fax: +48 60 23 15 212
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +27 73 15 78 355 Agramkow
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +65.890.6250 Fax: +65.890.6266
Phone +420 734 331 758 Fax: +420 604 203 037
Phone: +386 15 63 91 50 Fax: +386 17 22 04 51
Phone: +52 44 22 12 36 15 Fax: +52 44 22 12 19 40
Leybold Optics Ibérica Barcelona
Agramkow Sonderborg
INFICON GmbH Cologne
INFICON Company Limited Chupei City, HsinChu Hsien
INFICON GmbH Cologne
Sonderborg
INFICON GmbH Cologne
INFICON Blackburn
INFICON GmbH Cologne
Inficon Inc. East Syracuse, NY
San Jose, CA Inficon Inc. Austin, TX
Phone: +34 93 66 60 778 Fax: +34 93 66 64 612
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +886.3.5525.828 Fax: +886.3.5525.829
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +45 741 236 36 Fax: +45 744 336 46
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +44 1254 678 250 Fax: +44 1254 698 577
Phone: +49 221 56788-112 Fax: +49 221 56788-9112
Phone: +1.315.434.1167 Fax: +1.315.434.2551 Phone: +1.408.361.1200 Fax: +1.408.362.1556 Phone: +1.512.448.0488 Fax: +1.512.448.0398
1-6 General Information
1.3 Unpacking
Unpack the Protec P3000 leak detector immediately after it has been received even if it is to be put into operation at some later date. Examine the shipping container for any external damage. Completely remove all packaging materials.
Notice Retain the shipping container and the packaging materials in the event of
possible complaints concerning any damages.
Check if the Protec P3000 leak detector is complete (see Section 1.3.1) and carefully subject it to a visual inspection. If any damage is discovered please immediately inform the forwarding agent and the insurers. If it is required to exchange the damaged part please contact our or ders department.
1.3.1 Supplied Equipment
The Protec P3000 leak detector is ready for operation. Before installation please read Section 1.5. Included with the leak detector are the following items:
Protec P3000 (main unit)
Mains cord, 3m long
Set of fuses (3 x 10 pcs.)
Spare air filter
8 mm hexagonal wrench
Documentation
Operating Instructions (kima26)Technical Handbook Protec P3000 (kina26e1)Spare Parts List Protec P3000 (kiua26d2)Interface Description Protec P3000 (kins26e1)
Notice The SL3000 and SL3000XL sniffer lines are available in different
configurations and need to be ordered separately in the desired length. The SL3000(XL) sniffer line is not part of the Protec P3000 shipment. (see Section 1.3.2 Accessories)
Notice The PRO-Check reference leak is an accessory (see Section 1.3.2
Accessories) and needs to be ordered seperately
Notice For the RC versions the display unit and the co nnecting cable ar e not part
of the standard Protec P3000 shipment and need to be ordered separate ly (see Section 1.3.2)
General Information 1-7
1.3.2 Accessories
For Protec P3000
SL3000 Sniffer line for Protec P3000 Cat. No. / Ref. No.
SL3000-3, 3 m length 525-001 SL3000-5, 5 m length 525-002 SL3000-10, 10 m length 525-003 SL3000-15, 15 m length 525-004
Sniffer tips for SL3000
ST 312, 120 mm, rigid 122 13 FT 312, 120 mm, flexible 122 14 FT 200, 200 mm, rigid 122 18 FT 250, 250 mm, flexible 122 66 ST 385, 385 mm, rigid 122 15 FT 385, 385 mm, flexible 122 16 FT 600, 600 mm, flexible 122 09 ST 500, 500 mm, rigid, 45° angled 122 75
Water protection tip for sniffer 122 46
For Protec P3000XL only
SL3000XL Sniffer line for Protec P3000XL
Sniffer tips for SL3000XL
For RC versions only
External display unit
Connecting cable for external display unit
Connecting cable for external display unit
SL3000XL-3, 3 m length 521-011 SL3000XL-5, 5 m length 521-012 SL3000XL-10, 10 m length 521-013 SL3000XL-15, 15 m length 521-014
ST 312XL, 120 mm, rigid 122 80 FT 312XL, 120 mm, flexible 122 81 ST 385XL, 385 mm, rigid 122 82 FT 385XL, 385 mm, flexible 122 83 FT 250XL, 250 mm, flexible 122 85
for benchtop use 551-100 for rack mounting 551-101
5m length 551-102
0.7m length 551-003
1-8 General Information
For all Protec P3000
Warning
Caution
configurations
Holder for SL3000(XL) 525-006 PRO-Check reference leak for helium 521-001 Calibrated sniffer leaks for helium
S-TL 4, approx.1 x 10 S-TL 5, range 10 S-TL 6, range 10
-4
mbar l/s 122 37
-5
mbar l/s 122 38
-6
mbar l/s 122 39
1.4 Notes on How to Use This Handbook
Important remarks concerning operational safety and protection are emphasised as follows:
Notice Indicates special requirements the user must comply with.
Indicates procedures that must be strictly observed to prevent hazards to persons.
Indicates procedures that must strictly be observed to prevent damage to or destruction of the Protec P3000 leak detector.
The references to diagrams, e.g. (2-1/6) consist of the Section No., Fig. No. and the Item No. in that order. For example: (2-1/6) means Section 2, Fig. 1 and Item No. 6 (here: mains switch).
1.4.1 Symbols of Vacuum Technology
In the following some important symbols of vacuum technology as used in this handbook are shown:
Diaphragm pump Vacuum gauge
General Information 1-9
1.4.2 Definition of Terms
Main menu
This menu is shown first after operating the Menu push-button.
Sub-menus
Comprise all menus which may be accessed from the main menu. Unauthorised changes to many of these sub-menus may be prevented by a password (see also Section 4.3.1).
Menu item
A single menu line.
Default condition
Status of the Protec P3000 when supplied from the factory.
Service menu
Comprises the menu lines in the “Service” sub-menu. The service menu is accessed by scrolling in the basic menu using the navigation push-b uttons (see also Section
3.2).
Autozero
Determination and compensation of the helium background. With this function, the internal ZERO level of the leak rate signal is determined in order to avoid a readout of the internal helium background and mistaking it as a properly measured value. If subsequently negative leak rates are obtained due to this correction , the stored offset values are changed so that ZERO will be the lowest value which can be obtained. In this way the values adapt automatically to a decaying background (adaptive background correction).
Internal background
The existing partial pressure in the measurement system. The level of the internal background is measured all the time and subtracted from the measured signal.
I•Guide Mode
In the I•Guide Mode different testing plans can be pre-programmed. During testing the operator is then constantly prompted for the next action and thus guid ed through the testing plan.
Unit under test
Test object that needs to be leak checked.
Display limit
Limits the measurement data displayed depending on the unit of measurement and the operator settings.
1-10 General Information
1.5 Instrument Views of the Protec P3000
1
2
3
4
5
Fig. 1-2 Instrument views of Protec P3000 Pos. Description Pos. Description
1 Main display 4 Lemo Connector for SL3000 sniffer line 2 Speaker 5 Handle fore carrying the Protec P3000 3 PRO-Check reference leak
General Information 1-11
1.6 Installation
Caution
Caution
1.6.1 Set up
How to remove the transportation lock: The transportation lock is located on the bottom side of the Protec P3000 and
consists of a yellow knurled screws. Please remove this screw before starting-up the leak detector. The Protec P3000 is supplied ready for operation. Initial start-up is described in Section 3.1.
Fig. 1-3 Removing the transportation lock before starting
Before installation remove the transportation lock.
In order to ensure adequate ventilation of the Protec P3000, a space of at least 20 cm 8) (8 in.) must be kept unobstructed to the sides. The clearance at the rear must be no less than 10 cm (4 in.). Moreover, the Protec P3000 handles for carrying the leak detector at the sides of the main unit must not be covered at any time as these acts as air inlet and outlet. Avoid the presence of heat sources in the vicinity of the Protec P3000.
1-12 General Information
1.6.2 Mechanical Connections
Caution
Twisting the sniffer line spoils the cable. Do not twist the sniffer line.
SL3000(XL) Sniffer line
In order to operate the Protec P3000 it is essential for the SL3000(XL) sniffer line to be connected. The connection for the sn iffe r line is loca te d at the fro nt of the Prot ec P3000 left of the PRO-Check reference leak. Insert th e plug into the opening with the red dot on the plug and the slot in the front cover aligned until the connector engage s. To disconnect the plug, retract the coupling and remove the probe’s line.
Water protection tip (optional)
If you intend to perform leak testing on parts that are not completely dry (e.g. due to condensation after performance testing), we strongly recommend to use a water protection tip.
To install the water protection tip,
1 screw off the metallic capillary filter at the very end of the sniffer tip and 2 install the water protection tip instead.
Notice Please do not forget to re-install the little rubber se al when switching to the
water protection tip.
Fig. 1-4 Installing water protection tip
General Information 1-13
PRO-Check Reference leak (optional)
Please insert the PRO-Check reference leak into the opening in the housing of the main unit. Make sure that the Sub-D plug is pr operly connected with the PRO-Ch eck leak.
Notice When properly inserted, the PRO-Check reference leak will still protrude by
On first usage of your PRO-Check you need to initialize the use of this reference leak in the Protec P3000 software.
Please perform the following steps:
1 Insert the PRO-Check into the appropriate opening of the Protec P3000 2 In the software menu go to HISTORY & MAINTENANCE / REPLACE PRO-CHECK.
approx. 10 mm.
Notice This menu item is only available when the Protec P3000 is set to
MODE (see Section 4.4.3, Keyword: User Mode)
ADVANCED
3 On the certificate, which is delivered with the PRO-Check, you will find a serial
number and a 12-digit-code. Enter the serial number in the first line of the open submenu and the 12-digit-code in the sec on d line an d press OK.
Notice The PRO-Check reference leak must be installed in the Protec P3000
when pressing
Fig. 1-5 Initializing the PRO-Check reference leak
OK.
Notice PRO-Check Warntime Expire Date (See Section 7.7.5).
For RC versions only
The RC version has no built-in display unit but a connectors plate is mounted instead . Please connect the external display unit with the 5m connecting table (Cat.-no. 551-
002)
1-14 General Information
Fig. 1-6 Protec P3000 with external display unit for: (a) benchtop use (left side), (b) rack mounting (rigth side)
Holder for SL3000 sniffer line (optional)
An optional holder for the SL3000 sniffer line is available as cat.-no. 525-006. The holder may be installed on the right or left sid e of the main unit (for right- or left handed operators) as shown in Fig. 1-8.
Fig. 1-7 Usage of sniffer line holder The installation is described in Fig. 1-8. There are two little slots on the front side at
the very top area of the blue square front of the main unit. Ho ld the holder horizontally and then insert the two little hooks of the holder into the two slots (either on the right or the left side). With the hooks still inserted, let the holder flap down. It will automatically attach to the metal front by the magnet on the backside of the holder. Now insert the sniffer probe grip into the opening of the holder and let it sink down until it rests in the holder.
General Information 1-15
Fig. 1-8 Installation of sniffer line holder
Caution
1
23 456
1.6.3 Electri cal Conn ectio ns
Fig. 1-9 Electrical connections
Pos. Description Pos. Description 1 Headphone port 4 Name plate 2 I/O Port 5 Power switch 3 RS232 interface 6 Power connector
Notice The local regulations for electrical connections must always be observed
(in Germany VDE 0100). The mains voltage rating for the Protec P3000 can be read off from the name plate left of the power switch. The mains voltage setting of the Protec P3000 is fixed and can not be changed. A separate fuse for each of the mains conductors has been inte g ra te d into the mains socket (Fig. 1-9/6).
The mains voltage is applied to the Protec P3000 via the detachable mains cable which is supplied with the leak detector. A main power socket is available for this purpose at the rear of the main unit.
Before connecting the Protec P3000 to the mains you must make sure that the mains voltage rating of the Protec P3000 coincides with the locally available mains voltage.
1-16 General Information
Warning
Only 3-core mains cables having a protective ground conductor must be used. Operation of the Protec P3000 with the ground conductor unconnected is not permissible.
1.6.4 RS232 Interface
The Protec P3000 is equipped with a RS232 interface which is located on the rear right side of the main unit. This interface is of the DCE type (Data Communications Equipment) and allows the connection of a PC for monitoring and data logging. The connection is provided through a commercially available Sub-D plug. For further information see “Interface Description Protec P3000” (kins26e1).
1.6.5 I/O Port
The I/O port allows communication with and control through external equipment via PLC and analog data. For details see Section 6.1.
Through this connection some functions of the Protec P3000 can be controlled externally or measurement data or the status of the Protec P3000 may be communicated to external equipment.
Through relay changeover contacts the trigger levels as well as the operating mode (Ready) of the Protec P3000 may be monitored.
General Information 1-17
2 How the Protec P3000 Works
2.1 Description of the Functions
The Protec P3000 takes in helium through the SL3000 sniffer line, detects the amount of helium by means of a helium sensitive sensor and converts the sensor signal into quantitative leak rates.
The Protec P3000 is composed of the following principal subassemblies:
A Wise Technology sensor as helium detector
A valve holder for controlling different operating stati
A diaphragm pump for pumping the gas to the sensor
An inlet system for the gas flow
The correspond ing electrical and electronic subassemblies for supplying power and for signal conditioning.
The detector operates under gross vacuum conditions, i. e. the operating pressure at the detector is several 100 mbar. T he under pressure is generated by a diaphragm pump. The pressure in front of the sensor is measured with a piezo-resistant pressure gauge and is about 250 mbar under normal operating conditions.
2.2 Description of the Subassemblies
2.2.1 Backing Pump
A diaphragm pump in the Protec P3000 serves as the backing pump. All data and further information on this pump are given in the Operating Instructions. The backing pump generates the flow of the gas through the SL3000 sniffer line.
2.2.2 Wise TechnologyTM Sensor
The helium detector (Wise TechnologyTM sensor) consists of a closed glass container with a measurement device for the precise determination of the pressure inside the glass housing and a mem brane chip wit h a large nu mber of small quartz windows. The membrane is permeable only for helium, all other components of air are retained by the membrane and from the inside of the glass housing. The Quartz membrane is heated so that the permeation for helium is sufficiently high and fast.
Inside of the glass housing the total pressure is measured precisely. As only helium can enter the glass housing the total pressure is equal to the partial pressure of helium. The determined total pressure inside the housing is proportional to the helium partial pressure outside the sensor.
2-1 How the Protec P3000 Works
2.2.3 Valve Holder
The valve holder carries the valves that control the gas flow to the Wise Technology sensor. These valves are used to select the sensitivity of the system, to activate a protection mode against high helium contamination and to set the system into
STANDBY mode. The Protec P3000 software continuously analyses the situtation and
sets the correct valve position via the control unit.
2.2.4 Control Assembly
The control assembly (microprocessor) is the central assembly of the Protec P3000’s electronics. All other subassemblies are controlled and monitored by this assembly. The microprocessor which is located here is thus continuously informed about the status of the entire Protec P3000 and can respond accordingly. In order to accept commands from the operator and to output measured values and messages, the control subassembly is linked to the display unit.
2.3 Description of the Displays and User Interfaces
2.3.1 Main unit display
This subassembly is used to communicate with the operator. It accepts commands from the 8 keys on both sides of the display and outputs measurement results and messages via the display.
Fig. 2-1 Main unit display Pos.Description Pos. Description
1 Menu buttons 1 to 4 2 Menu buttons 5 to 8
How the Protec P3000 Works 2-2
For RC versions only
The RC version has a connectors plate for the external display unit instead of the built-in main unit display. Two LED's left of the plug provide information about the status of the Protec P3000, even when the external display unit is disconnected:
The green LED will indicate that the Protec P3000 is in operation (switched on).
The red LED will be blinking in case of an error message, continuous red light
If no display unit is connected, error messages or warnings may be acknowledged by pressing both buttons of the SL3000 sniffer line simultaneously.
The external display unit also offers four buttons:
The START / STOP buttons have no function (the external displa y unit may also
The MENU button will open the software menu.
The ZERO button will set the current background reading to zero. (For details on
The green LED will show continuous green light if an external display unit is connected and will blink if no external display unit is detected.
indicates a warning.
be used with other INFICON leak detectors which need these buttons)
the ZERO function see Section 4.4.1)
Fig. 2-2 Conncetor plate with LEDs
2-3 How the Protec P3000 Works
2.3.2 SL3000(XL) Sniffer line with probe display
1
3
2
The probe handle also offers a small display for operating the Protec P3000 remotely without access to the main unit during normal leak detection operation.
1 Probe display 2 Button 3 ZERO button
Fig. 2-3 SL3000(XL)Sniffer line with probe display In addition to the display, the sniffer probe offers two press buttons. The left button
will function as the ZERO button in any operating mode. By pressing the left button the current background reading is set to ZERO. For details on the ZERO function see section 4.4.1)
The right probe button is used for different functions depending on the current operating mode the Protec P3000 is set to:
in standard operation mode:
no function for Protec P3000for Protec P3000XL: switching between
in I•Guide mode: – navigation through I•Guide program (for all configurations)
Fig. 2-4 Probe handle
HIGH FLOW and LOW FLOW mode
The probe handle also offers some LEDs in the flange of the sniffer tip in order to illuminate the location currently being leak tested.
How the Protec P3000 Works 2-4
Warning
Do not stare into the LEDs of the sniffer line intentionally for extended times or at a close distance as this may cause permanen t da ma g e to th e ey e.
The LEDs offer some bundled light. The intentional starring suspends the lid closing reflex and also the eyes do not move anymore which may le ad to overhea ting of the retina.
When looking into the LEDs „incidentally“ the eye is protected by the lid closing reflex. Also the permanent movement of the eyes prevents overheating and consequent damage of the retina.
2.3.3 Built-in PRO-Check reference leak
A built-in PRO-Check reference leak is available for all Protec P3000. The PRO­Check reference leak can be used for verifying the correct functioning of the Protec P3000 including the correct calibration and can also be used for re-calibrating the Protec P3000 if necessary.
The PRO-Check reference leak is inserted in the front of the h ousing. Insertion of the sniffer tip into the cone-shaped test leak opening is automatically detected via a light barrier.
Fig. 2-5 Built-in PRO-Check reference leak
2-5 How the Protec P3000 Works
Fig. 2-6 PRO-Check reference leak detached from the main unit for remote use
1
2
In cases where the main unit is placed in a difficult or inconvenient to access area the built-in PRO-Check reference leak can be removed from the main unit for easie r access and connected to the main unit via the Sub-D connector with a commonly available extension cord. The PRO-Check reference leak can then be placed in an area where it is convenient for the operator to be reached.
Fig. 2-7 Connections for remote use Pos. Description Pos. Description
1 Connector at built-in test leak 2 Connector at Protec P3000 housing
How the Protec P3000 Works 2-6
Notice The PRO-Check reference leak is not part of the Protec P3000 shipment
and needs to be ordered as a separate part no. (see section 1.3.2
Accessories).
Notice If you have not purchased the PRO-Check reference leak, warning 71 (“No
communication with test leak”) will be issued on first start-up. Please go to
SETTINGS / INTERFACES / PRO-CHECK and set the PRO-Check to “DISABLED
to prevent future warnings (see section 4.5.6)
Notice PRO-Check Warntime Expire Date (See Section 7.7.5).
2-7 How the Protec P3000 Works
3 Operation of the Protec P3000
3.1 Start-Up
Assemble the Protec P3000 (see Section 1.5). Connect the mains cord and the SL3000(XL) sniffer line, and then switch on the Protec P3000. The mains switch is located on the rear.
Fig. 3-1 Connection of the mains cord Pos. Description Pos. Description
1 Power Switch 2 Power cord connection
The Protec P3000 will automatically start-up without operator interaction. After power on the message "Wait for heater" will be displayed with the pump not being started yet. The foreline pressure as well as the flow through the sniffer line will be stated during this time.
After the pumps are switched on a self-test procedure will be performed during which all hardware will be checked. The message "wait for sensor current" will be displayed next. A bar graph will indicate the progress of this warm-up stage. The expected warm-up time left will be stated below the bar graph.
Notice The start-up procedure normally lasts 2 - 3 min. After extended times of
power off, however, the start-up of the Protec P3000 may take up to 20 min. max. To improve the start-up behaviour of the Protec P3000 after extended times ofpower off, see Section 3.8.
After switching on and completion of the run-up phase the Protec P3000 will be ready to make measurements. There is no separate start function. The SL3000(XL) sniffer lines are designed to maintain an inlet pressure low enough to make measurements.
Notice The Protec P3000 will only operate after having connected the sniffer line. A calibration in accordance with Section 3.5 is recommended not before 20 minutes
after having switched on the instrument (warm-up phase).
Operation of the Protec P3000 3-1
Notice If you have not purchased the PRO-Check reference leak, warning 71 (“No
communication with test leak”) will be issued on first start-up. Please go to Settings / Interfaces PRO-Check and set the PRO-Check to “disabled” to prevent future warnings (see section 4.5.6).
3.2 Controls on the main display unit
All set-up and control functions are integrated into the main disp lay unit via the menu structure. The functions of the 8 control keys are displayed on the LC display. Dur ing measurements the main interface is the probe handle display giving all necessary information for proper leak testing.
After start-up the Protec P3000 will automatically go into measurement mode.
Bar graph display
Fig. 3-2 Measurement screen Pos. Description Pos.Description
1 Audio volume buttons 4 Info button 2 Menu button 5 List of gas parame ter s bu tto n s 3 Calibration button
The currently detected leak rate will be display in a bar graph in logarithmic scale. The currently selected trigger level is indicated by a black line, the currently selected search level is indicated by a dotted line. If the search level is exceeded the shape of a bell is displayed at the top of the display, if the trigger is exceeded the bell starts to blink (to ring).
3-2 Operation of the Protec P3000
Gas type (refrigerant equivalent)
The currently selected gas type (helium or refrigerant equivalent) is indica ted on th e upper left side of the display.
For P3000XL only: If the Protec P3000XL is operated in inverted characters (on dark background) in the main display as well on the sniffer display.
Fig. 3-3 Measurement screen with leak
HIGH FLOW mode, the gas type is displayed in
Pos. Description 1 Selected gas type (refrigerant equivalent or He) 2 Indication of search level being exceeded 3 Indication of active warning 4 Audio volume level 5 Bar graph display of leak rate
Audio Volume Buttons
The two middle keys on the left side of the display allow to adjust the volume of the alarm sound at any time. When pressing any of the two buttons the currently selected volume will be demonstrated by the loudspeaker as well as by a bar graph in the status line. The selected value is also displayed as the first entry of the status line at the bottom of the display and only applies to the loud speaker in the main unit. For selecting different types of alarms see section 4.4.2.
Menu Button
The button on the bottom left side of the display will open the main menu at any time. The menu mode offers the user many possibilities of entering instrument settings and special functions.
CAL Button
With the button on the upper right side of the display an external calibration of the Protec P3000 can be started at any time. For details on how to perform an external
calibration see Section 3.5.3.
Operation of the Protec P3000 3-3
ZERO Button
When shortly pressing the zero button, the zero level will be updated. For details on the ZERO function see Section 4.4.1.
List of gas parameters button
Up to four different gas parameter sets can be stored in the Protec P3000. If more than one set of gas parameters are set-up, the list button on the right side of the display will appear in the menu screen. When pressing this button, a new set of gas parameters (e.g. different refrigerant equivalent, different trigger level, etc) can be selected. For details on how to set up the different gas parameters see section 4.3.1.
Info Button
Status Line
Lock Softkeys
When pressing the status of the Protec P3000 will be displayed. For details see section 3.4.3.
In the bottom line of the main display status information is indicated. First the currently selected volume for the audio alarm is stated. Next, a small black triangle with an exclamation mark may indicate an active warning.
The keys C be protected.
AL, ZERO und can be locked. So the settings of these functions can
i info button (bottom right side of the display) information on the
Abb. 3-4 Function key
3-4 Operation of the Protec P3000
3.3 Controls on the probe display unit
1
3
2
Warning
On the display of the probe handle similar information as on the main display is shown.
Description
Pos.
Bar graph indicating the leak rate
1
Absolute leak rate
2
Gas equivalent
3
Fig. 3-5 Sniffer display in standard operation mode The currently detected leak rate is indicated as a bar graph. In a second line the
numerical leak rate (in the same unit of measurement as on the main display) is shown. In the third line the type of gas equivalent is stated (e.g. He or R134a).
For Protec P3000XL only: If the Protec P3000XL is operated in displayed in inverted characters (on dark background).
HIGH FLOW mode, the gas equivalent is
The sniffer probe offers two press buttons. By pressing the left button the current background reading is set to ZERO. For details on the ZERO function please refer to
4.4.1. ZERO.
The right probe button is used for navigation when working in I•Guide mode (see Section 3.4.2) or for starting an internal calibration (see Section 3.5.2)
3.4 Performing measurements
The Protec P3000 offers two modes of operation:
The Standard Operation Mode (compatible to the Protec mode)
The I•Guide Operating Mode
Danger of electric shock. Don’t touch voltaged parts with the sniffer tip. Test samples need to be
disconnected from electricity before leak testing.
Standby Function
The Protec P3000 offers a contaminants into the sniffer probe during idle times, thus saving filter and sensor life time.
STANDBY function to prevent the needless intake of
Operation of the Protec P3000 3-5
The Protec P3000 may automatically go into standby after a preset time of
Caution
inoperation (see Chapter 4.3.1 for details on how to set-up this function) and operation will automatically resume if the probe is picked up again.
The Protec P3000 can also be set into probe button for 2 seconds. If the Protec P3000 has been set into manually operation will only resume by pressing any of the two probe buttons again.
Operating in moist environment
Do not suck in any liquids.
For operation in an environment where moisture (e.g. condensating water) may occur, a water-protection tip is available (Cat. no. 12246) which will protect the Protec P3000 against intake of liquids if necessary. For details on how to install the water potection tip see 1.6.2)
Protection mode and high helium background
STANDBY mode manually by pressing the left
STANDBY mode
The Protec P3000 is equipped with a protection mode against contamination with huge amounts of helium. This contamination mode will help the leak detector to clean up faster after detection of gross leaks. If the Protec P3000 switches to
GROSS MODE the item „GROSS“ is displayed and an
acustic signal is to hear. If a huge amount of helium is detected the Protec P3000 may show the message
HELIUM CONTAMINATED. The leak detector will come back to measurement mode
automatically after clean up and the displayed item „GROSS“ disappears.
Notice The Protec P3000 should not be switched off or set to
HELIUM CONTAMINATED is displayed as this will trap the increased levels of
STANDBY mode while
helium inside the sensor and it cannot be pumped away for clean up of the Protec P3000. Switching off the Protec P3000 while contaminated with large concentrations of helium will in fact lead to significantly increased start-up times afterwards. If this has happened unintentionally, just switch on the Protec P3000 and let it sit in "Warm up" until it reaches measurement mode.
Notice In
HIGH FLOW mode leak rates in the range < 1x10
-5
mbar l/s are reduced
during a time of 30 Sec. to the current background value.
For details on how to set up the
CONTAMINATION LIMIT please refer to Section 4.4.1.
3.4.1 Standard Operation Mode
Provided the Protec P3000 has been set-up to meet the requirements of the particular application and it has been calibrated (see Section 3.5), a measurement is performed as follows:
3-6 Operation of the Protec P3000
First briefly operate the ZERO button on the sniffer probe. This will ensure that the Protec P3000 eliminates all interfering influences which may affect the ZERO level (i.e. the detection limit of 1x10
-7
mbar l/s). Next hold the tip of the sniffer as close as possible to the suspected leak, if required the tip may even touch the te st object. If a welded seam or alike needs to be tested, the tip should be moved at a velocity of no more than 13 cm/s (5 inch per second) along the welded seam. The distance between tip and test sample should be as small as possible.
If a leak is detected the bar will grow. The Protec P3000 continuously compares the measured leak rates with the programmed trigger levels. If the trigger is exceeded the background colour of the probe display will change from green to red. At the same time an alarm sound will be released by the speaker in the probe handle and the probe handle will start to slightly vibrate. As an additional indication of exceeding the trigger value the three white LEDs in the flange of the sniffer tip will start flashing.
Fig. 3-6 Sniffer display when detecting a leak
As soon as an acoustic alarm sounds, the tip should be removed from the spot being tested. After displaying a constant leak rate the ZERO button should be pressed again to repeat the test. Thus, a measurement error can be prevented and the lea k can be located.
For Protec P3000XL only:
The Protec P3000XL allows the use of the HIGH FLOW mode (requires use of SL3000XL sniffer line). When set to HIGH FLOW mode leaks may be detected at a larger distance from the potential leak. The high flow mode is indicated by the gas type being displayed in inverted colors (o n the pr obe display as well as on the m ain display). When testing for joints, the sniffer tip should be no further than 10 mm (0.4 in) away from the joint. If testing a weld ed seam (or alike), the sniffer tip should not be moved faster than 4 cm/s (1.6 in/s) at a maximum distance of 10 mm (0.4 in) from
the welded seam.
Operation of the Protec P3000 3-7
If a leak is detected (and accessibility allows so) switch the Protec P3000XL to LOW FLOW mode (by pressing the right probe button) for easier localization of the leak. The gas type will now be displayed not-inverted. Now search for the leak again by bringing the sniffer tip as close as po ssible to the potentially leaking area. Localize the leak by moving the sniffer tip back and forth across the suspected ar ea. The leak will occur where the leak rate signal shows the maximum leak rate.
3-8 Operation of the Protec P3000
3.4.2 I•Guide Operating Mode
The I•Guide operating mode has been introduced to suppor t the operator in applyin g proper sniffer leak detection technique.
The I•Guide operator guiding mode allows to store pre-programmed parameters for different units under test. The number of locations that need to be tested per specimen, the time each location needs to be tested for as well as the time required to move to the next location may be programmed. In add ition, a ma ximum allowab le global leak rate for the total unit under test is stor ed. In the I•Guide mode up to 10 pre-programmed testing procedures can be stored.
Notice If the number of locations to be tested is set to zero, the Protec P3000 will
operate in a continious mode without checking for a global leak rate but still issue the timer signal for proper testing.
3.4.2.1 Starting the I•Guide Mode
To start the I•Guide Mode go to the main menu and choose SET UP IGUIDE. In the opening menu go to the top line item and change the setting to side push button and press enabled I•Guide program in the list. A message screen will pop-up notifying the operator that I•Guide has been activated (the gas selected in the I·Guide program you choose).
For how to set-up and I•Guide program please refer to Section 4.4.4. To switch back to the Standard Operation Mode select “
ON with the left hand
OK. The Protec P3000 will automatically choose the first
OFF” and press “OK”.
Fig. 3-7 Switching to I•Guide Mode
Operation of the Protec P3000 3-9
3.4.2.2 Selecting an I•Guide Program
To open the SELECT I•GUIDE menu press the PROGRAM LIST button on the right side of the display.
Fig. 3-8 Measuring Screen in I•G uid e M od e Pos. Description
1 Program list In the opening S
and press ok. The new program is now loaded.
Fig. 3-9 Selecting an I•Guide program
3.4.2.3 Using an I•Guide Program
In the measuring screen of the I•Guide operating mode the selected program, the gas type stored in the program as well as the summarized global leak rate will be displayed.
ELECT I•GUIDE menu highlight the program number you intend to use
For Protec P3000XL only:
If the Protec P3000XL is operated in inverted characters (on dark background) in the main display as well as on th e sniffer probe display.
3-10 Operation of the Protec P3000
HIGH FLOW mode, the gas type is displayed in
For all Protec P3000
1
2
3
4
5
3
7
6
configurations
In the I•Guide message line the Protec P3000 will prompt the user for action. First it will ask to move to the first testing position. Also on the probe display the message “okay pos. 1?” will be shown. Please confirm with the right probe button that the sniffer tip has reached the right location.
Fig. 3-10 I•Guide screens during measurement Pos.Description
1 Selected program 2 Gas type stored in the selected program 3 I•Guide message 4 Measuring time 5 Summarized global leak rate per unit under test 6 ZERO button 7 I•Guide button
After the first location has been confirmed the message “leak check point1” will be indicated on the main unit display and the elap sing measu rement time ( stored in the testing program) will be shown in the lower part of the menu page. Please make sure to hold the sniffer tip in the right testing location during the total measuring time. During this time a ticking sound will be issued by the main unit and a beep will indicate that the measuring time has elapsed and the sniffer tip can be moved again.
After the measurement time has elapsed the message “Move to point 2” will be displayed on the main unit. The Probe display will read “tip to pos. 2”. Please move the sniffer tip to the next testing location and if the wait time indicated in the display has elapsed you may start the next measurement. If the operator tries to start the next measurement before the wait time has elapsed, the message “please wait” will be displayed in the message line until a next measurement is allowed. Please confirm that the sniffer tip has been positioned properly by pressing the right probe button so that the next measurement can be started.
Operation of the Protec P3000 3-11
Fig. 3-11 I•Guide screens requesting the next location
After checking the pre-programmed nu mber of location s the result of the testin g for the total unit under test will be displayed as shown in the following screen shot. The selected testing program as well as the gas type stored in the program will be stated once more followed by the summarized global leak rate. If the global leak rate is lower than the global leak rate trigger the message “Global leak check okay!” will be displayed followed by the testing results for each tested location.
Notice For each location where no leak is detected, the currently selected lower
diplay limit is added to the global leak rate as this is the maximum leak rate which still may occur for each position (worst case approach).
Fig. 3-12 Result of I•Guide program: unit under test passed Pos. Description Pos. Description
1 Program na me 3 Results of each location tested 2 Gas type
If the summarized global leak rate exceeds the global trigger the message “Global trigger exceeded!” will be displayed.
3-12 Operation of the Protec P3000
Fig. 3-13 Result of I•Guide program: unit under test failed By pressing the right probe button the next testing cycle may be started.
Notice The I•Guide mode can also be used as a timer signal only. If the number of
Notice The I•Guide mode can also be used to summarize leak rates on demand.
3.4.3 The Info Page
By pressing the info button on the main display a general info page will open. Information on the software version used, date & time information, the currently set audio volume and the minimum audio volume are stated.
points is set to 0, the Protec P3000 will prompt for the next location to be tested continuously without using the global leak rate function.
If the number of points is set to 99, a results screen with summarized global leak rate will be displayed after the right button has been pressed for 2 s conintously (or after the 98th point automatically).
Fig. 3-14 Info page without errors or warnings If an active warning exists this will be displayed in the info page instead of the gas
info lines.
Operation of the Protec P3000 3-13
Fig. 3-15 Info page with active warning When operating in I•Guide Mode, the info page will state information about the
currently selected program: the selected program name, the gas used for this program, the number of points to be checked for this program, the selected measuring and wait time as well as the global trigger.
3-14 Operation of the Protec P3000
3.5 Calibration and Self-Test
The Protec P3000 can be calibrated internally with the built-in PRO-Check re ference leak or externally with an external calibrated leak (Cat. No. 122 37 - 122 39).
Notice If a calibration is started during the first 20 minutes after power on, a
warning will be issued. Do not calibrate the Protec P3000 during the first 20 minutes after start-up.
Also a verification of the calibration may lead to wrong results in the first 20 minutes after start-up.
Only confirm and continue with the calibration if the real warm-up has been longer than 20 min. (e.g. after a quick restart of the Protec P3000.)
The built-in PRO-Check reference leak can be used for a self-test of the Protec P3000 as well as for internal calibration.
Notice The PRO-Check reference leak is a temperature compensated leak. It
must only be used when electrically connected to the main unit, either when inserted into its port or when connected to the main unit via a Sub-D extension cord. The leak rate printed onto the body of the PRO-Check is only valid at 20°C (68F) and will vary greatly with temperature and time.
To compensate for this the PRO-Check reference leak is equippend with a temperature sensor and a compensation curve is stored in the software which automatically compensates the test leak rate for changes in temperature when connected to the main unit.
A calibration or verification with the PRO-Check reference leak not being connected to the main unit will cause a wrong calibration of the Protec P3000 and / or will lead to wrong testing results.
3.5.1 Verifying a calibration (proof function)
Notice A verification can only be performed while the unit is in one of the two
measurement modes: standard operation mode or I•Guide operation mode. A verification will not be started if the main menu is opened.
If the sniffer tip is inserted into the opening of the PRO-Check reference leak a verification of the calibration (proof function) will be started automatically. While holding the sniffer tip in the test leak opening, the unit will check the re ading from the PRO-Check. Afterwards the operator will be requested to remove the sniffer tip from the leak opening.
Notice Any time during the verification procedure an internal calibration may be
started by pressing either the right sniffer probe button or the CAL button on the main display.
The results of the verification will be displayed in a summarizing screen. If the verification process was successful, the message "sensitivity check OK" will be displayed.
Operation of the Protec P3000 3-15
Fig. 3-16 Results of proof function for Protec P3000 If the value measured for the PRO-Check is out of range, the message "recalibration
required!" will be displayed in inverted colors. For returning to the measuring mode, please press the right probe button or press
OK” on the main display.
For Protec P3000XL only
For the Protec P3000XL both calibration factors for
HIGH FLOW and LOW FLOW mode
will be verified (see fig. 3-16). However, only the information for the selected flow mode will be relevant for passing or failing the verification. The results for the non­selected mode will be displayed for reference at the bottom of the results screen.
Fig. 3-17 Results of proof function for Protec P3000XL
3-16 Operation of the Protec P3000
3.5.2 Internal calibration
Notice A calibration can only be performed while the unit is in one of the two
measurement modes: standard operation mode or I•Guide operation mode. A calibration will not be started if the main menu is opened.
If the sniffer tip is inserted into the opening of the PRO-Check reference leak with the right probe button pressed while in measurement mode a calibration will be started automatically. While holding the sniffer tip in the test leak opening , the Protec P3000 will measure the test leak. Afterwards the operator will be requested to remove the sniffer tip from the leak opening.
After completion of the internal calibration a screen summarizing the results of the calibration will be displayed. The old calibration factor as well as the new calibration factor will be displayed. If warnings are active and have been acknowledged during the calibration process, the message is endorsed with the information "with active warnings".
Fig. 3-18 Results of internal calibration To avoid unintended overwriting of a previous external (more accurate) calibration,
the operator needs to “Accept new values (anyway)?”.
3.5.3 External calibration
For external calibration it is recommended to use leak rates > 5x10-6 mbar l/s for Protec P3000 and leak rates of > 5x10 P3000XL at
HIGH FLOW (3000 sccm).
If the reference leak rate is < 5x10 calibrates in
LOW FLOW mode.
Notice If significantly increased backgrounds are prevalent in your production
environment, larger leak rates for the calibration leak may be required.
The external calibration is a semi-automatic process during which the user will have to follow some instructions. The calibration process may be starte d via the “ button from the measuring mode at any time (except when the menu is open or the function has been locked). A running calibration process may be cancelled by operating the “
ESC” button.
-5
mbar l/s when calibrating the Protec
-5
mbar l/s the Protec P3000XL automatically
CAL” -
Operation of the Protec P3000 3-17
After pressing the CAL button, please check whether the leak rate equals the leak rate of the external leak you plan to use. If the leak rate is different press
EDIT LEAK RATE
and enter the correct leak rate value. Press “START” afterwards to begin with the calibration process.
Fig. 3-19 Setting-up the leak rate of the external leak Please hold the sniffer tip to the outlet of the external calibrated leak. Hold the sniffer
tip steady and very close to the opening, however, do not clog the opening with the sniffer tip. Some air also needs to enter the sniffer tip in addition to the helium from the external calibrated leak. If the leak rate signal indicated by the bar graph is stable press “
OK”. Keep holding the sniffer tip steady in front of the opening while the Protec
P3000 reads the leak rate of the calibrated leak. During this time the text “Please wait…” will be displayed.
Fig. 3-20 Calibration screen during external calibration When the analysis of the calibrated leak signal is completed a message “sniff air!”
will be displayed. Remove the sniffer tip from the opening of the calibrated leak and hold the sniffer tip into the air, as far as possible from any sources of helium. Wait until the bar graph shows a stable signal again and press “ok” once more.
3-18 Operation of the Protec P3000
Fig. 3-21 Request to sniff air during external calibration A message “Please wait…” will be displayed until the calibration is finished.
After completion of the external calibration a screen summarizing the results of the calibration will be displayed. The old calibration factor as well as the new calibration factor will be displayed. If warnings are active and have been acknowledged during the calibration process, the message is endorsed with the information "with active warnings".
Fig. 3-22 Results of external calibration
Operation of the Protec P3000 3-19
3.6 Standby
If not used, the Protec P3000 will go into a STANDBY mode automatically after a predetermined time in order to avoid wear of parts during times not used. In this
STANDBY mode the flow through the sniffer line will be shut down to conserve filters
in the sniffer line as well as to save sensor life time. If the operator picks up the sniffer line, a movement sensor will detect this action and
will set the Protec P3000 back into normal operation. The Protec P3000 will be ready for measurement again after approx. 5 seconds.
For details on how to set the time until the Protec P3000 goes into please refer to section 4.3.1 (
VACUUM & ACCESS, STANDBY DELAY)
3.7 Shutdown
To shut down the Protec P3000, set the ON / OFF main switch (Fig. 3-1/6) to the „0“ position regardless of the operating mode of the Protec P3000. Nothing else is required. The parameters entered are saved by Protec P3000. After switching on, the Protec P3000 will revert to the same status it was in when it was switched off before.
3.8 Storage for fast availability as back-up unit
STANDBY mode
Due to accumulation of Helium (from air) in the sensor unit during storage the start­up time for the Protec P3000 will be approx. 1.5 min per day of storage with a maximum start-up time of approx. 1 hour.
If you want to use the Protec P3000 as a back-up unit to ensure uninterrupted availability of your production line, the Protec P3000 should be stored as follows.
1 Shut down the Protec P3000 but keep it connected to a power outlet. 2 Put a clock timer between the Protec P3000 and the power outlet. 3 Program the clock timer to switch on the Protec P3000 for 1 hour twice per week (at
least every 4 days).
Notice The clock timer should be battery buffered so that in case of a loss of power
in the factory, the clock timer does not loose its pre-programmed window for power on.
With this approach the max. start-up time will be approx. 7 min.
3-20 Operation of the Protec P3000
Operation of the Protec P3000 3-21
4 Equipment Settings
4.1 Menu Structure
By pressing the menu button the main menu page opens. The following options will be displayed:
Fig. 4-1 Main menu screen In the subsequent sub-menus the following frequently used functions will occur:
↑ ↓ The “up” and “down” keys allow scrolling through listed menu items. The
currently selected menu item is highlighted by inverted colours. The “up” and “down” keys are also used for changing numerical values to a new setting.
? Help function – by pressing this key additional information on how to use
this page can be displayed.
Esc Going back to the pre viou s pa ge while om ittin g any cha n ge s. Back Going back to the next higher level in the menu structure. OK Accepting a selected item from a list or accepting newly entered values.
Exit all menus and return to main screen. If the menu button is presse d later on again, the software will jump to the menu page that has been exited by the button. By pressing the back key the display will go back one level and finally all the way back to the measurement display page.
The menu structure of the main menu is illustrated on the following page for quick reference.
4-1 Equipment Settings
Main menu Standby/Start Library Gas
Service User Gas Gas Trigger Gas 1 Name (edit) Helium
Gas 2 Additional He Gas 3 M ode Gas 4 Trigger&U nit
Display limit Search level Equivalent correction (He)
Settings Vakuum & access Zero time
Contamination limit Flow limits (only P3000) limit LOW FLOW
limit HIGH FLOW
Flow settings (only P3000XL) Additional low flow error
Flow mode Li mi ts l ow flow e rr or
Sn i ffe r bu tton Standby delay CALibration Change menu-PIN
Audio Beep sound
Au dio intern Audio Sn i ffer Al arm typ e Vo lume
Setup I-Guide I-Guide ON/OFF
M e a s . 1. .. 10 ( e d i t ) N a m e
Gas
No. of p oints
Measuring time
Wait time
Global trigger
Misc ellaneo us Language
Time & d ate PRO-Check Warntime Expiry Date Pressure u n it Leak rate filter Al arm delay Sn i ffe r li g h t
Display Contrast
Pe ak h old Flow Displa y (1) User mode Lock Softkey s
Interfaces Control location
Recorde r out p u ts Select PLC inputs RS232 protocol Baud rate & End sign PRO-Check
e
View error list Calibration history Sn i ffe r ti p fi l te r Maintenance intervall Maintenance history Confirm maintenance (1) (1) only in user mode "Advanced mode" Replace PRO-Check (1)
Info General data
Sensor Global Sensor Errors PRO-Check data Sniffer line data I/O-P o rt data Analog data
History&M aintenanc
Equipment Settings 4-2
4.2 The Service Menu
The service menu is password protected. When entering the service menu a PIN is required to be entered. The service menu should only be used by trained service personnel. The PIN for entering the service menu will be distributed during service training. Please refer to the service documentation for the submenus and functions of the service menu.
4.3 Selecting gas equivalents and setting trigger values
The Protec P3000 can detect leak rates as helium leak rates or in refrigerant equivalent leak rates. Up to four different sets of parameters can be stored by the Protec P3000. Only one type of leak rate, however, will be displayed in the measurement screen optional with the equiva lent Helium leak rate in addition. In the gas / trigger menu the chosen gas equivalen ts are displayed with their curr ent trigger value.
With default settings only helium is set up in the first set of parameters. The remaining three sets are blank ("<------>"). Only sets that have been edited will be listed in the list of gas parameters that can be loaded directly from the measuring screen through the list button.
4.3.1 Setting gas parameters
When pressing one of the GAS1 to GAS4 buttons an info page regarding the chosen set of parameters will open. Parameter sets that are not in use are marked blank („<----->)“. Changes to the current setting can be made by scrolling to the line item to be changed by use of the the
EDIT key.
Fig. 4-2 Editing Gas 1 In the opening info page the name of the gas equivalent (or helium), the
measurement mode and the currently selected trigger value will be displayed. This info is followed by the value for the lower display limit. In addition, the correction factor, the helium concentration used, the fill pressure for helium and the fill pressure for refrigerant later on are listed.
UP and DOWN keys and selecting the line item by pressing
4-3 Equipment Settings
Name (Gas equivalent)
Additional Helium
For displaying the leak rate in refrigerant equivalent leak rates go to the menu and go to the
SELECTION
submenu choose the refrigerant equivalent from the GAS LIBRARY. Gas
NAME line item. Press the EDIT button and in the opening GAS
EDIT GAS sub
parameters can be fully deleted by setting the gas name back to blank („<----->“). Notice If more than one set of parameter is set up (i.e. maximum two gases are
not named „<----->“) a list button will appear on the right side of the measurement screen. When pressing the list button a short-cut to selecting the different parameter sets will be provided.
This submenu only is available when operating in gas equivalents. In this case the Protec P3000 may display the Helium leak rate in addition to the refrigerant equivalent leak rate in the main measurement screen. This feature may be selected or not selected.
Default: not selected
Fig. 4-3 Selecting additional Helium
Notice The option “Additional He” is only available if at least one parameter set
(gas 1 … gas 4) is configured to helium. The additional helium will be displayed in the unit of measurement of the
lowest gas number set to helium.
Equipment Settings 4-4
Editing gas equivalent parameters
Fig. 4-4 Selecting gases from the library
Back in the edit gas submenu scroll to the
EDIT.
HELIUM PERCENTAGE line item and press
Fig. 4-5 Setting a refrigerant equivalent correction factor In the opening submenu the helium fill pressure, the fill pressure for the gas
equivalent (refrigerant) later on and the concentration of helium used can be entered. In the lower right corner the
CORRECTION FACTOR (from helium to refrigerant) will be
displayed. If a set of parameters is entered that exceeds the boundaries of the Protec P3000 the correction factor will be displayed in inverted colours. In this case adjust the parameters until the correction factor is displayed in normal colours again . Press OK if all parameters are as desired.
4-5 Equipment Settings
Notice This menu should also be used when working with diluted Helium but still
displaying the leak rate as a Helium leak rate.
Notice This menu should be used at any time when the Helium fill pressure in pre-
testing differs from the refrigerant fill pressure later on to display the correct leak rate value, even if the leak rate is still displayed as a Helium leak rate.
Fig. 4-6 Editing parameters for refrigerant equivalent leak rates In the EDIT GAS submenu you can now enter the TRIGGER value in the refrigerant
equivalent value. A set of parameters (refrigerant type, fill pressure, helium concentration) can also be
saved as user gas. In the measurement screen the gas type will now be displayed as refrigerant
equivalent leak rate (e.g. R134a <- He).
Trigger and unit of measurement
In the the corresponding line item. You can change the trigger level setting by the use of the left You can leave the submenu without making any changes by using the
EDIT GAS X info page the TRIGGER & UNIT sub-menu page will open by selecting
UP and DOWN keys. The correct settings must be confirmed by the OK button.
ESC key.
The leak rate unit can be changed by the use of the right Available leak rate units include mbar l/s, Pam
3
/s, Torr l/s, atm cc/s, g/a, oz/yr,
UP and DOWN keys,
lb/yr, ppm. Default value: 2 x 10
-5
mbar l/s
Fig. 4-7 Setting-up trigger level and unit of measurement Pos. Description Pos. Description
1 Increasing the trigger value 3 Changing the unit of measurement
2 Decreasing the trigger value 4 Changing the unit of measurement
Equipment Settings 4-6
Display limit
Fig. 4-8 Editing the lower display limit The lower display limit for each gas can be edited. To change the lower display limit
go to the “
DISPLAY LIMIT” line item in the “EDIT GAS” submenu and press “EDIT”. The
lower display limit can be increased in factors of the lowest display limit. In the line below the factor the absolute lower display limit in the currently chosen unit of measurement is stated. Available factors are 1, 2, 5, 10, 20, 50, 100.
For the Protec P3000XL the lowest display limit of 1x10
-6
mbar l/s equals factor 1. The equivalent absolute lower display limits at other factors are the lower display limit in
LOW FLOW mode mulitplied by a factor of 10.
Notice For Protec P3000XL only:
When switching from display limit factor will remain unchanged, i.e. if the lower display limit factor is set to 2, the lower display limit will be adopted from 2x10
-6
1x10
mbar l/s automatically when switching from LOW FLOW to HIGH FLOW.
Default value: 2
Search level
An additional warning level can be entered as percentage of the trigger level. A warning signal will be issued for leaks that are below the trigger level but above the search level in order not to miss small leaks. The search level may be set to values between 5 and 100%. Default value: 90%
4.3.2 Selecting a set of gas parameters
A set of gas parameters can be selected by setting the Mode in the Gas Edit menu to “enabled”. Enabling a set of parameters will automatically disable all other three sets of parameters so that only one set of parameters is used at the time.
LOW FLOW to HIGH FLOW (or vice versa) the lower
-7
mbar l/s to
4-7 Equipment Settings
4.4 Settings Sub-menu
In the settings sub-menu several settings which are important for the leak testing procedure can be edited.
4.4.1 Vacuum & Access
In the Vacuum Access sub-menu setting monitoring functions like the ZERO function, the flow limits and the contamination limit setting can be edited.
Zero time
The
ZERO function is automatically updated when the backgr ound decre ases so that
negative values will never be displayed, thus avoiding unintended missing of leaks. The
ZERO TIME is the time the leak rate has to be negative in order for the Protec
P3000 to update the background level. The
1.0 and 9.9 s. Default value: 5.0 s
Contamination Limit
The Protec P3000 offers a protection mode against high helium concentrations (contamination with helium).
This feature should be used to prevent the sensor from accelerated aging when seeing large helium concentrations very frequently. The set to levels of accumulated amount of helium sniffed over the last couple of seconds. If this accumulated amount of helium exceeds the preset level ( message „contaminated“ will be displayed until the amount of helium in the system has decreased again sufficiently. However, the lower the the smaller the largest possibly detected leak will be.
ZERO TIME can be set to values between
CONTAMINATION LIMIT can be
LOW, NORMAL AND HIGH. The contamination limit function will track the
LOW, NORMAL, HIGH) the
CONTAMINATION LIMIT set ting
Notice You should start with the
CONTAMINATION LIMIT set to LOW. If you feel that
you get the „contaminated“ message too often, you may increase the
CONTAMINATION LIMIT by one level at a time (to get the best usage out of
your Wise Technology Sensor).
Default: N
ORMAL
The lower the contamination limit is set, the longer the life time of the Wise Technology sensor of the Protec P3000.
Equipment Settings 4-8
Fig. 4-9 Setting up the contamination limit
Notice The Protec P3000 should not be switched off or set to STANDBY mode while
Flow limits (for Protec P3000 only)
In this submenu the minimum and maximum gas flow through the capillary is entered. If the actual gas flow drops below this limit (e.g. if the capillary becomes partially blocked) or exceeds this limit (e.g. if the sniffer line is damaged so that ambient air is sucked in) a n a l arm is issued. The closer the to the actual gas flow the more sensitive the Protec P3000 will respond to beginning clogging of the filters and the sniffer lin e itself. The closer the to the actual gas flow the more sensitive the Protec P3000 will respond to damages to the sniffer line.
Notice The flow through the sniffer line depends on the atmospheric pressure. If
The
LOWER FLOW LIMIT can be set to values between 150 and 240 sccm. Default
value is 180. The default value of 350.
HELIUM CONTAMINATED as this will trap the increased levels of helium inside
the sensor and it cannot be pumped away for cle an up of the Protec P3000. Switching of the Protec P3000 while contaminated with large concentrations of helium will in fact lead to significantly increased start-up times afterwards. If this has happened by mistake, just switch on the Protec P3000 and let it sit in "Warm up" until it reaches measurement mode.
LOWER FLOW LIMIT is set
UPPER FLOW LIMIT is set
operated at high altitude the flow through the sniffer line may drop significantly (approx. 20% per 1000 m altitude). In this case, please adjust the flow limits accordingly.
UPPER FLOW LIMIT can be varied between 300 and 650 with a factory
4-9 Equipment Settings
Fig. 4-10 Changing the flow limits of the SL3000 sniffer line
Flow settings (For Protec P3000XL only)
The Protec P3000XL can be used in
LOW FLOW mode as well as in HIGH FLOW mode.
In the flow settings submenu all settings related to both flow modes can be adjusted. When entering the
FLOW SETTINGS submenu four further submenus will be displayed:
Fig. 4-11 Setting up flow settings
Flow limits
LOW FLOW
In this submenu the UPPER FLOW LIMIT and LOWER FLOW LIMIT for the LOW FLOW mode may be set-up. The
LOWER FLOW LIMIT can be set to values between 150 and 240
sccm. Default value: 180 The
UPPER FLOW LIMIT can be varied be tween 300 and 650 sccm. Defaul t value: 395
Equipment Settings 4-10
Flow limits HIGH FLOW
Fig. 4-12 Variation of flow limits In this submenu the UPPER FLOW LIMIT and LOWER FLOW LIMIT for the HIGH FLOW mode
may be set-up. The
LOWER FLOW LIMIT can be set to values between 1500 and 2400
sccm. Default value: 1800 The
UPPER FLOW LIMIT can be varied between 3000 and 3990 sccm.
Default value: 3950
Notice To be able to measure in H
is equipped with two capillaries (~300sccm and ~2700 sccm). In L mode only the smaller capillary is used, in H
IGH FLOW and LOW FLOW mode, the SL3000XL
OW FLOW
IGH FLOW mode both are used,
but only the gas from the smaller one is directed to the Wise Technology sensor.
To monitor the correct flow through the smaller capillary in (if blocked, no leaks are detected any more), the flow limits in H mode are converted into corresponding Technology sensor. If these
PRESSURE LIMITS are exceeded or undercut,
PRESSURE LIMITS at the Wise
HIGH FLOW mode
IGH FLOW
this is an indication of the smaller capillary’s flow being out of the normal range and a warning message (W41 or W42) is issued.
Additional low flow error (only available in ADVANCED mode)
Notice This submenu is only available when in
ADVANCED mode and if a SL3000XL
sniffer line is connected to a Protec P3000XL main unit.
In this submenu an
ENABLED an additional flow error (E55 / E56) will be issued in addition to the standard
low flow warnings. Only if the
LIMITS LOW FLOW ERROR will be available.
If a
LOW FLOW ERROR occurs, the Protec P3000XL will stop measuring. Only when the
cause of the low flow problem has been resolved and the flow is above the
ERROR again, the Protec P3000XL will go back into measurement mode.
FLOW
ADDITIONAL LOW FLOW ERROR may be ENABLED or DISABLED. If
ADDITIONAL LOW FLOW ERROR is ENABLED, the button
LIMITS LOW
4-11 Equipment Settings
The
LIMITS for the LOW FLOW ERROR may be set -up in the LIMITS LOW FLOW ERROR
submenu (see below). Default setting: disabled
Flow mode (only available when SL3000XL sniffer line is connected)
In this submenu the Protec P3000XL can be switched between
HIGH FLOW mode.
LOW FLOW mode and
Default setting:
HIGH FLOW Mode
Limits low flow error
This submenu is only available if the In this submenu the limits for the
actual flow through the sniffer line drops below the
ADDITIONAL LOW FLOW ERROR is set to ENABLED.
ADDITIONAL LOW FLOW ERROR may be set-up. If the
LIMITS LOW FLOW ERROR, the
Protec P3000XL will stop measuring. Only when the cause of the low flow problem has been resolved and the flow is above the
LIMITS LOW FLOW ERROR again, the
Protec P3000XL will go back into measurement mode.
Fig. 4-13 Additional limits for low flow error
Two different values for the HIGH FLOW and LOW FLOW mode may be entered. The value for the HIGH FLOW mode will also be converted into a corresponding LOWER PRESSURE LIMIT. If the corresponding LOWER PRESSURE LIMIT is undercut, this is an indication of the smaller capillary's flow being blocked in SL3000XL sniffer line and also E55 / E56 will be issued and the Protec P3000XL will stop measuring.
Error limits can be set to values between 150 and 240 sccm for LOW FLOW mode and to values between 1500 and 2400 sccm for HIGH FLOW mode.
Default value: 150 sccm for LOW FLOW mode
1500 sccm for HIGH FLOW mode
Sniffer button
In this submenu the use of the right sniffer button for switching between
LOW FLOW
and HIGH FLOW mode during leak testing (except when in I·Guide mode) may be
ENABLED or DISABLED. If the use of the SNIFFER BUTTON is DISABLED, the Protec
P3000XL may only be switched from through the
FLOW MODE submenu or via RS232.
HIGH FLOW to LOW FLOW mode (or vice versa)
Default setting: enabled
Equipment Settings 4-12
Standby Delay
Cal
The Protec P3000 offers a times when the Protec P3000 is not used. During the
STANDBY mode to preserve filters and sensor life time in
STANDBY mode the flow through
the sniffer line will be stopped, thus no dirt is pumped through the filters and the sensor is not subjected to the helium background presented in the testing environment.
The time until the Protec P3000 goes into
STANDBY DELAY time. The STANDBY delay time can be set to values between 10 s and
1 h. If the probe is not moved for the set switch to
STANDBY mode automatically. If the probe is moved during this time, the
STANDBY mode is determined by the
STANDBY delay time the Protec P3000 will
counter will be set back to zero and start again. If the probe is picked up when the Protec P3000 is already in
STANDBY mode the
Protec P3000 will go back into operation automatically and will be ready for measurement again in approx. 5 sec.
Default value: 10 min.
Notice It is recommended to keep the
STANDBY delay time as short as possible to
minimize wear of the systems without causing too much interruption of your operations.
In the CALibration submenu an internal calibration can be function is set to
DISABLED, only a proof function with the built-in PRO-Check
ENABLED or DISABLED. If the
reference leak will be possible. Default value:
ENABLED
Change menu PIN
Any changes to the current settings may be password protected. The menu PIN can be changed in the “change menu PIN” submenu.
Fig. 4-14 Setting-up a Menu-PIN for paramete r pr ot ec tion If a new PIN is entered the Protec P3000 software will ask to repeat the new PIN to
prevent entering errors. Afterwards the menu PIN will be changed to the new value.
4-13 Equipment Settings
Notice The operator needs to go back up through the main menu for this change
to become effective immediately. Otherwise, the new PIN will be activated after 5 minutes.
A menu PIN of “0000” means no password protection, all menus can be freely accessed.
Default setting: 0000
Equipment Settings 4-14
4.4.2 Audio Functions
In the Audio sub-menu all settings regarding the different audio alarms of the Protec P3000 can be edited.
Beep sound
Beep sounds are sounds that are issued as confirmation of certain functions, e.g. completion of a calibration process. This acknowledgement sounds can be enabled or disabled in the beep sound submenu.
Default value: enabled.
Audio intern
The internal loudspeaker of the main unit can be switched on and off. This will not effect the headphones outlet.
Default value: enabled.
Audio sniffer
Alarm type
The loudspeaker of the probe handle can be set to issue an alarm when the
TRIGGER
level is exceeded, when the SEARCH level is exceeded or can be disabled completely. Default value:
TRIGGER.
Different types of audio alarm of the main un it speaker can be selected. Optio ns are
SETPOINT, PINPOINT, and TRIGGER ALARM.
SETPOINT The frequency of the tone will change when the trigger level is
exceeded. Below the trigger level the loudspeaker will remain off.
TRIGGER ALARM
An acoustic alarm with a continuous low tone will sound as soon as the search level is exceeded. An acoustic alarm which consists of two tones sounds as soon as the trigger level is exceeded. Below the search level the loudspeaker will remain off.
If the Protec P3000 is set to TRIGGER ALARM as the alarm type an additional button will be shown on the lower right side called TONE 1,2,3. With this button different alarm sounds can be selected for the trigger alarm to clearly distinguish between nearby other leak detectors.
PINPOINT The external loudspeaker is always on. Within the leak rate range of
+/- one decade of the trigger level the frequency will change depending on the leak rate. Outside this range the tone will be constantly low or high. This setting is recommended if also leaks which are still below the trigger level need to be pinpointed or if the expected leak rates are within the range of the trigger level.
4-15 Equipment Settings
Default value: Trigger Alarm
Volume
The
MINIMUM VOLUME and the currently selected VOLUME for the loudspeaker in the
main unit can be edited. The
MINIMUM VOLUME is the lowest level the volume can be
set to from the main display and should prevent unintentional turning off of the main unit audio alarms. Both values will also be valid for the headphones outlet. Both volumes (minimum and current) can be chosen between 0 and 15.
Default value for minimum volume: 2 Default value for the current volume: 2.
Equipment Settings 4-16
4.4.3 Display Settings
In the DISPLAY SETTINGS submenu the contrast of the display can be changed and the peak hold function can be set-up.
Contrast
In the
CONTRAST submenu the contrast of the main unit display can be adjusted.
Values between 0 and 99 can be entered. Also the display can be switched to inverted colours.
Default value: not inverted, Level 30
Notice If by accident the display contrast has been set too high or too low so that
it cannot be read off anymore, this can be fixed as follows: Shut off the Protec P3000 and switch it back on. During run-up press the third button from the top on both sides of the display simultaneously until the contrast is back to normal. This setting is not saved to the EEPROM but needs to be confirmed manually through the contrast menu. Otherwise the Protec P3000 will go back to the out of range setting after the next start-up.
Peak hold
Fig. 4-15 Setting-up the contrast of the main dis play
The
PEAK HOLD function can be enabled or disabled in this submenu. If the PEAK HOLD
function is enable the maximum leak rate will be displayed for a preset time in addition to the currently me asured value in the main display. Th e time setting will have no effect if this function is disabled.
Default: value disabled, 5 s.
4-17 Equipment Settings
Fig. 4-16 Setting-up the peak hold function
Flow display (only available in ADVANCED mode)
In the
FLOW DISPLAY submenu the continuous display of the flow through the sniffer
line during measurement s can be indicated in the as integer number.
Notice After certain events like acknowledging warnings or when switching the
Default setting: disabled
User mode
In this submenu either the
NORMAL mode only the most relevant information for normal operation will be
displayed whereas in the will become available.
ENABLED or DISABLED. If ENABLED, the flow will be
STANDARD OPERATION MODE only underneath the leak rate bar graph
flow mode, the flow monitoring may be disabled for a short period of time while the flow stabilizes to a final value again and no new warnings will be issued. During this time a message ”Flow control inactive” will be displayed instead of the flow value.
NORMAL or the ADVANCED mode can be selected. In the
ADVANCED mode additional information for advanced users
Equipment Settings 4-18
4.4.4 Setting-up / editing an I•Guide Program
When selecting the SETUP PROGRAM button a submenu page will open showing the list of 10 pre-programmable I•Guide programs and their current status (enabled / disabled). If a program is enabled it will be offered in the list of programs to select from when operating in the I•Guide mode. To change the status of a program use the
ON / OFF button on the left side of the display.
Fig. 4-17 Setting-up the I•Guide Mode To edit the parameters of a program select the program by scrolling to the
corresponding menu line and open the selected program by pressing the “E button.
DIT
4-19 Equipment Settings
Fig. 4-18 Editing an I•Guide program In the opening
EDIT PROGRAM sub-menu the name of the program, the gas type
selected for this program, the number of locat ions to be leak tested, the minimum measuring time for each location as well as the wait time between two leak tests allowed for moving the sniffer to the next location is stated. In addition, the summarized global trigger value corresponding to the allowable leak rate for the complete unit under test is stored.
Each parameter can be edited by scrolling to the appropriate line item and selecting the highlighted item by pressing the
EDIT button.
Name
Gas type
Each program can be named with a free text, descriptive name up to 6 characters. The name can be edited by highlighting the name line and pressing
EDIT. An
alphanumerical entry screen will open. Press ok after entering the desired name description.
No. of points
For editing the gas equivalent go to the “ equivalents that are listed in the “
GAS” line item and press “EDIT”. Only gas
GAS/TRIGGER” submenu (Gas1 to Gas4) are
available for selection. The gas equivalent is sta ted in plain text in pa renthesis. Th e list of maximum four available gas equivalent can be scrolled with the up and down buttons.
Notice The I•Guide Mode refers to the gas equivalent selected in the “
TRIGGER” submenu. If you change a gas equivalent in the “GAS” submenu
GAS/
the gas equivalent selected for the I•Guide program will automatically change as well.
The number of potential leak locations to be checked can be varied between 0 and 99 and can be edited with the up and down keys.
Default value: 4
Notice The I•Guide mode can also be used as a timer signal only. If the number of
points is set to 0, the Protec P3000 will prompt for the next location to be tested continuously without using the global leak rate function.
Notice The I•Guide mode can also be used to summarize leak rates on demand.
If the number of points is set to 99, a results screen with summarized global leak rate will be displayed after the right button has been pressed for 2 s continiously (or after the 98th point automatically).
Measuring time
The
MEASURING TIME during which the sniffer tip needs to be hold to the correct
location can be programmed to values between 0.7 and 25.0 s. The measuring time should not be less than the response time of the Protec P3000. We recommend to use the following minimum measuring times:
Sniffer line length min. measuring time 3m 0.7 s 5m 0.9 s 10m 1.4 s 15m 3.0 s
Please use the
Default value: 1.0 s
UP and DOWN buttons for editing this parameter.
Equipment Settings 4-20
Wait time
The
WAIT TIME between two leak checks allowed to move the sniffer to the next
potentially leaking location can be varied between 0.1 s an d 25.0s in step s of 0.1s . Please use the
Default value: 3.0 s
Global trigger
In the
GLOBAL TRIGGER submenu the maximum allowable leak rate for the complete
unit under test can be edite d. The same ranges as for a single trigger value are allowed. The unit of measurement for the global trigger will be the same as selected for this gas type in the edit gas submenue.
Default value: 2 x 10
4.4.5 Miscellaneous Settings
Language
Different languages can be chosen in which the software can be operated. The current selection includes English, German, Spanisch, Italian, French, Portugese, Japanese (Katakana) and Chinese (Mandarine).
Default value: English
UP and DOWN buttons for editing this parameter.
-3
mbar l/s (or the equivalent in any other unit of me asurement)
Date & Time
Sniffer light
Pressure unit
Notice If the Protec P3000 has been set to a language th at you do not understand,
turn off the power and press the second button from the top of the display on both sides simultaneously during run-up. This will reset the Protec P3000 to English as the menu language. This setting is not saved automatically. Please go to the
LANGUAGE submenu afterwards and select
the appropriate language.
In this submenu the internal clock of the Protec P3000 can be set. On the first page the date needs to be entered in the format DD.MM.YYYY. The lower right button (→) opens the second page on which the time in the format HH.MM is entered.
The intensity of the sniffer light can be varied to the desired level. The intensity can be selected between 1 and 6. Also the light can be switched on and off via the D
ISABLE / ENABLE buttons. Any new setting must be confirmed with the OK button.
Default value: on, level 4.
4-21 Equipment Settings
The pressure unit of all displayed values of the Protec P3000 can be set to Pa, Torr , atm or mbar.
Default value: mbar
Leak rate filter
Alarm delay
For all normal operation the I•Filter should be selected. I•Filter is an intelligent filtering algorithm that yields the best results with respect to noise level and stability of leak rate values specifically developed for use in the Protec P3000 leak detector.
Only in cases where the older Protec model is replaced by an Protec P3000 and the leak detector is used in a fixed testing system it may be required to stick with the “
FIXED” filter that was used in the Protec.
Default: I•Filter
Under conditions with very unstable background conditions it may be favourable to not issue an alarm immediately but only release an audio alarm if the trigger leve l has been exceeded for an extended time. This alarm delay can be entered in 1/10 of a second. Values between 0 and 9.9 sec will be accepted. Default value: 0.0 sec (i.e. disabled).
Equipment Settings 4-22
4.5 Interfaces
In the INTERFACES submenu the CONTROL LOCATION, the RECORDER OUTPUTS. the PLC inputs, the RS232 PROTOCOL and the BAUD RATE & END SIGN can be edited.
4.5.1 Control location
The CONTROL LOCATION can be “LOCAL”, “RS232” or “LOCAL AND RS232”. I n LOCAL mode the RS232 interface can only be used to read data, but not for control purposes. In RS232 mode, the Zero and Cal functions are controled solely via the interface. In interface but also by input from the main display.
Default setting: local and RS232
Notice Even with the control location set to RS232 some parameter may still be
4.5.2 Recorder outputs
The RECORDER OUTPUT SCALE is linear and logarithmic. The linear output signal is provided through channel 1 (pin 1 of the I/O port) and the logarithmic output is provided through channel 2 (Pin 14 of the I/O port) (see Section 6.1).
In logarithmic output scale the voltage range is 0 - 10 V. Beginning at 1V, each leak rate decade is spread over 2 volts, i.e.:
1…3V: 1 3…5V: 2 5…7V: 3 7…9V: 4
Thus a leak rate range of 4 decades will be output by way of an analog signal. The four decades are set through the value and the unit of measurement for the trigger level. The software of the Protec P3000 will set the output voltage in such a manner that the trigger level itself will always be set within the second decade (i.e. within the range of 3 to 5 Volts).
LOCAL AND RS232 mode the Protec P3000 can be controlled via the
able to be changed through the main units software menues. Please lock these by activating the menu PIN to prevent unitended changes.
st
decade
nd
decade
rd
decade
th
decade
4-23 Equipment Settings
Examples:
Trigger level Output voltage (1 to 9V) will correspond to
3 g/a 0.1 to 1000 g/a
0.2 oz/yr 0.01 to 100 oz/yr
-4
5x10
mbar l/s 1x10
The following table has been provided to clarify the way in which the output voltage represents the leak rate.
-5
to 1x10-1 mbar l/s
Converting an output
leak rate 10
Trigger onentexp
U3
2
------------ -+
=
10
1–
535, 3–
2
---------------------+
1 5oz, yr=
10
0–
56, 3–
2
----------------- -+
20g a=
10
0–
64,3
2
----------------+
50g a=
10
4–
28, 3–
2
----------------- -+
8105–mbarl s=
10
4–
655, 3–
2
---------------------+
15mbarl s,=
voltage to a leak rate
If the recorder outputs a certain voltage and you want to convert this to a leak rate, the following formula applies:
(2 is used since one decade is s pread over 2 Volts. 3 is used because the trigger level will always be within the range of 3 to 5 V).
Example calculations:
Trigger
Trigger exponent
Output voltage
Leak rate
0.1 oz/yr -1 5.35V
3 g/a 0 5.60V
6.4V
5 x 10-4 mbar l/s
-4 2.8V
6.556V
Equipment Settings 4-24
Converting a trigger leak rate to a trigger output voltage
If you want to know the corresponding voltage for a specific trigger leak rate, the following formula applies:
U = 3 + 2·log (mantissa of scientific notation of trigger value) (The factor of 2 is used since one decade is spread over 2 Volts. The summand of 3
is used because the trigger level will always be within the range of 3 to 5 V).
Example
Trigger leak rate Scientific notation Mantissa Trigger voltage 2*10 5 g/a 5*10
0.4 oz/yr 4*10
Recorder Output
In linear output scale different settings can be chosen in the submenu
OUTPUT
The recorder output can be set so that the max. 10 V corresponds to either 1 x 10 mbar l/s, 1 x 10-3 mbar l/s, 1 x 10-2 mbar l/s, 1 x 10-4 Torr l/s, 1 x 10-3 Torr l/s, 1x10 corresponds to 10 x trigger level, i. e. 1 V equals the selected trigger level.
In the modes "E U = 10 V will be output through channels 1 and 2.
Default setting:
4.5.3 RS232 Protocol
The RS232 protocol can be set to “ASCII”, “DIAGNOSTICS”, “PRINTER AUTO” or “PRINTER
MANUAL
widely used for measuring equipment. For details on this protocol please refer to the Interface Description (kins26e1).
-5
mbarl/s 2*10
-5
0
-1
2 3+2·log(2) = 3.6 V 5 3+2·log(5) = 4.4 V 4 3+2·log(-1) = 4.2 V
RECORDER
. The voltage range spans from 0 V to 10 V.
-2
Torr l/s. Alternatively, the recorder otuput can be set so that 10 V
RROR”, "NOT READY TO MEASURE” and "STANDBY” a voltage of
AUTO
”. The ASCII protocol is a protocol similar to the SCPI, a standard protocol
-4
Diagnostics
4-25 Equipment Settings
Default setting: ASCII
The
DIAGNOSTICS protocol is a binary protocol used by INFICON for analysis in quality
control and trouble shooting during manufacturing and service.
Printer manual
Printer auto
The
PRINTER MANUAL protocol allows to send leak rates on pressing the right sniffer
probe button.
Notice When in any of the two printer modes, no ASCII commands must be sent
to the Protec P3000 as this will interrupt the printer mode.
Format
Date Time Gas equivalent Leak rate Unit
Example
03.05.2005 12:31 He 6x10
-4
mbar l/s
When set to I·Guide mode the PRINTER MANUAL mode is without function.
In
PRINTER AUTO the leak rate will be sent any time the set trigger level is exceeded.
At the time the signal drops below the trigger level the maximum detected leak rate will be sent.
Format
Date Time Gas equivalent Leak rate Unit
Example
03.05.2005 12:31 R134a 2.34 g/a
When set to I·Guide mode the leak rate results will be sent after each test point together with the test point number. After the test cycle has been completed, the summarized global leak rate will be sent.
Example:
Point no. Date Time Leak rate Unit Point 01 8.3 E-6 mbar l/s Point 02 2.0 E-6 mbar l/s Point 03 8.4 E-6 mbar l/s Point 04 1.2 E-6 mbar l/s Global 13.04.2007 11:57:03 1.8 E-6 mbar l/s
Equipment Settings 4-26
4.5.4 Select PLC inputs
In the SELECT PLC INPUTS submenu the user can select which pin on the I/O port (suitable for PLC input) represents which command. The default setting is as follows:
Pin Default Command Pin Default Command 7 Standby 13 Cal 8 Clear Error 20 Zero 9 Cal Abort 25 Not used
To change these settings select the appropriate pin with the UP and DOWN arrows on the left side of the display and afterwards select the desired command from the list of commands with the UP and DOWN buttons on the right side of the display. Press OK to save your settings.
Fig. 4-19 Customizing PLC inputs
4.5.5 Baud rate & end sign
The baud rate can be set to values between 1200 and 19200. Default setting: 9 600. The end sign can be set to Carriage Return (CR), Line Feed (LF) or to CR+LF. Default setting: CR+LF.
4.5.6 PRO-Check (only available in ADVANCED mode)
In this submenu the PRO-Check reference leak can be disabled completely so that for Protec P3000 leak detectors without a PRO-Check onboard the leak detector does not check for the electrical connection with the reference leak anymore and no error message is issued.
Default value: Enable
4-27 Equipment Settings
4.6 The Info Menu
The INFO MENU lists all internal data that may be useful for any trouble-shooting of the Protec P3000. The button “ one page whereas the lower right button allows to go to the next page. The page number is displayed in the upper right corner.
Menu item Format Description
Page 1: General Data
Foreline pressure mbar Flow sccm Flow through sniffer line Time since power on Min Operation hours h Serial number 11-digit number Wise Serial number Software version x.xx.xx
INFO MENU consists of 10 pages. On each page the upper left
BACK” allows to leave the info menu, the lower left button allows to go back
Circuit temperature °C Temperature of main board Test leak temperature °C
1 = Protec P3000 2 = Protec P3000 with
Sensor block Protec P3000 1 / 2 / 3
Page 2: Sensor Global
I akt. A B-Filter A I-Filter A Q-Filter A Leak rate mbar l/s helium leak rate PWM-Membrane Integer Heater Voltage V High Voltage V
extended measurement range 3 = Protec P3000XL
Wise-State Integer
Equipment Settings 4-28
Menu item Format Description
Measure Background Press button
If pressing the MEASURE BACKGROUND button the Protec P3000 will go into a special measurement mode and determine the background concentration of Helium currently prevalent in the environment. This process may take a couple of seconds. This feature is not a continuous signal display but should be used for trouble-shooting purposes only.
Page 3: Sensor Errors
Sensor Error Hex Code Sensor / Valve Warning Hex Code Adjust Error Hex Code
Page 4: PRO-Check data
Gas Helium Gas type of internal leak Leak rate nom. / at T mbar l/s / mbar l/s Version / Checksum Hex Code Serial no. Serial no reservoir Manufacturing date DD.MM.YYYY Expiry Date DD.MM.YYYY Gain / Offset Test leak temperature °C / °F State
Page 5: Sniffer data
Type
Software version x.x Length 3m / 5m / 10m / 15m Serial no. 9000 xxx xxxx Switch left on / off
SL3000 / SL3000XL / system
4-29 Equipment Settings
Switch right on / off Color Green / red Bar graph
Menu item Format Description
Acceleration x / y x / x
Flow at calibration sccm / sccm
Pressure at calibration mbar / mbar
2 values for Protec P3000XK
2 values for Protec P3000XL
Page 6: I/O Port data
Here the current PIN assignment is displaye d . T he de fa ult con fig ur a tion is sh ow n in the picture.
Page 7: Analog data
AIN3 Sniffer length V AIN4 +5V II Leak V AIN5 +24V III ext V AIN6 +5V I Sniffer V AIN8 –15V MC50 V AIN9 +15V MC50 V AIN10 +24V MC50 V AIN11 +24V I QMS V AIN12 +24V II V
Page 8: Analog data
AIN0 V AIN0 offset V Foreline pressure mbar AIN1 V
Equipment Settings 4-30
Menu item Format Description
Flow sccm AIN2 V
Page 9: Info valves
V1: bypass open / close V2: measure open / closed V3: purge inlet open / closed V4: purge sniffer open / closed V5: sniffer open / closed V6 / V7: Air pressure mbar Normal mode mbar Gross mode mbar Zero pressure mbar
Page 10: RS232 Info
Protec P3000 sniffer ASCII string
Sniffer Protec P3000 ASCII string
Host Protec P3000 ASCII string
Protec P3000 Host ASCII string
HIGH FLOW open / closed / not installed for P3000XL only
Command sent from main unit to sniffer
Command sent from sniffer to main unit
Command sent from host to Protec P3000
Command sent from Protec P3000 to host
4-31 Equipment Settings
4.7 History & Maintenance
In HISTORY & MAINTENANCE historical data that has been collected during the operation of the Protec P3000 can be viewed as well as information about the maintenance status of the leak detector.
View Error List
In the
ERROR LIST all errors (and warnings) tha t have occurred d uring the operat ion
of the Protec P3000 are summarized. In the list the occurred is stated followed by error or warning nu mbe r (E xx being errors and W xx being warnings) with a short description of the error or warning. The full error message or warning may be displayed when scrolling to the appropriate line item and pressing the magnifying glass-button.
DATE and TIME when the error
Calibration history
Fig. 4-20 Example of an Protec P3000 error list
In this listing all calibrations that were conducted during the operation of the Protec P3000 are collected. The
DATE and TIME when they were performed are stated,
followed by the type of calibration (internal / external) and the calibration factor.
Fig. 4-21 Example of a Protec P3000 Calibration history
Equipment Settings 4-32
Sniffer tip filter
The maintenance interval for the sniffer tip filter depends on the operating environment and may be set to values between 10 and 999 hours. If set to reminder message for the sniffer tip filter will be disabled.
Default value: 100 hours For details on how to perform the maintenance tasks see Chapter 7, Maintenance. When confirming that the filter has been r eplaced (without changing the interval time)
press okay only and the counter will restart.
the
Fig. 4-22 Setting the maintenance interval for the sniffer tip filter and confirming its replacement
Maintenance intervall
In the maintenance interval sub-menu the number of operating hours for the main unit in total, as well as the time until the next maintenance is due for the diaphram pump, the main unit air filter and the optimization for the Wise Technology sensor.
When the number of operating hours have expired a warning will be issued as a reminder to perform the appropriate task. The reminders are issued for the diaphragm pump and the air filter of the main unit. Please also refer to Section 5.1 (Error Messages and Warnings)!
4-33 Equipment Settings
Fig. 4-23 Example of a maintenance inter va l menu page
Maintenance history
In the maintenance history list all maintenance tasks performed are listed. The date and the time the maintenance tasks were performed, the number of operation hours the Protec P3000 had worked at the time when the maintenance task was performed and the type of maintenance that was conducted are listed. Details about each maintenance task may be viewed when highlighting the appropriate line item and pressing the
Fig. 4-24 Example of maintenance history list
Confirm maintenance (only in ADVANCED mode)
VIEW button.
In this submenu the replacement of the main unit air filter can be confirmed and the date and the number of operation hours at which the maintenance task was performed is saved in the software.
Notice Maintenance of the diaphragm pump can only be con firm e d in th e
(password protected) service menu by trained personnell.
Fig. 4-25 Confirming maintenance task For details on how to perform the maintenance tasks see Chapter 6, Maintenance.
Equipment Settings 4-34
Replace PRO-Check (only available if
ADVANCED mode
ENABLED and in
When replacing the gas reservoir of the PRO-Check reference leak, the new serial number of the replacement gas reservoir as well as a code containing the calibration data of the new reservoir needs to be entered in this sub menu. For details on ho w to replace the gas reservoir of the PRO-Check reference leak please refer to Section
7.7.
Fig. 4-26 Activating new PRO-Check gas reservoir
Notice PRO-Check Warntime Expire Date (See Section 7.7.5).
4-35 Equipment Settings
5 Protec P3000 Messages
During leak detection operations the LCD display will indicate information which supports the operator in running the Protec P3000. Besid es measurement data, also current equipment conditions, operating hints as well as warnings and error messages can be displayed.
5.1 Error Messages and Warnings
The Protec P3000 is equipped with comprehensive self-diagnostic functions. When a faulty condition is sensed by the control board, this condition is indicated to the operator via the LCD display as far as possible.
Errors
Errors are events which fo rce an inter ruption of t he measurem ents in pro gress and which the Protec P3000 is not capable of rectifying on its own. Errors are indicated in plain text together with an error number.
The Protec P3000 remains in the error status. After the fault cause has been removed the error can be acknowledged by pressing the P3000 then returns to the measurement mode.
RESTART button. The Protec
Warnings
Warnings are issued if an abnormal condition is detected that may cause increased inaccuracies of measurements but will not interrupt measuring completely.
In the following table all possible error messages possible reasons for the problem and the recommendations on how to fix the proplem are li sted. If no recommendation is given (or the recommendation does not fix the problem), please call your nearest INFICON service representative (see Section 1.2.1).
Notice In case of enquiries please keep the serial number and the software
version number of the Protec P3000 at hand.
The following abbrevation is used in the listing + error messages below: MC50 CPU-board
Protec P3000 Messages 5-1
Error
Caution
no.
E1 24V of the MC50 too low Fuse F1 on the motherboa rd has
Plain text message Possible Reaso n Recommendation
blown
Replace fuse F1 on motherboard*
E2 24V of sensor heating too low Fuse F2 on the motherboard has
blown E3 24V II too low Fuse F3 on the motherboard blown Replace fuse F3 on motherboard* W4 24V of the OPTION socket too
low
W5 5V of the sniffer is too low Fuse F5 on the motherboard has
W6 5V of the internal Leak too low PRO-Check electronics defective Replace PRO-Check reference leak or
E7 -15V of the MC50 is too low Motherboard defective Call nearest INFICON service
E8 15V of the MC50 is too low Motherboard defective Call nearest INFICON service
E9 High Voltage error Problem in sensor electronics Call nearest INFICON service
W11 Wise current unstable Current from Wise Technology
E12 Wise Sensor not ignited Current from Wise Technology
E13 Problem in Heater Control Heater control for Wise
W14 Dis charge gone out Sensitivity problem of Wise
W15 Replace sniffer tip filter Filter in the sniffer tip is clogged Replace the filter in the sniffer tip!
Fuse F4 on the motherboard has
blown
blown
sensor did not stabilize within
20 min. after start-up!
Protec P3000 possibly not being
used for serveral days, sensor
current should stabilize with more
run-time.
sensor too low for more than 10
min after power on
Technology sensor defective.
Technology sensor
or
No ambient helium present (e.g.
sniffer purged with nitrogen)
Warntime every 2h.
Replace fuse F2 on motherboard*
Replace fuse F4 on motherboard*
Replace fuse F5 on motherboard*
call nearest INFICON service representative!
representative!
representative!
representative! Restart the Protec P3000.
If problem persists, call nearest INFICON service representative!
Restart the Protec P3000, if problem persists, call nearest INFICON service representative!
Call nearest INFICON service representative!
Make proof or calibration.
Restart with sniffer line connected to fresh air
If problem persists, call nearest INFICON service repesentative!
W17 Fore pump service interval
expired!
W18 Air filter service interval
expired!
E19 No communication with Wise
ADC
5-2 Protec P3000 Messages
> 10,000 operating hours since last
fore pump service
> 10,000 operating hours since last
main air filter service
Wise Technology sensor defective
or
CPU-board defective
Replace membrane of diaphragm pump!*
Clean or replace main air filter! See Section 7.2
Call nearest INFICON service representative!
Error
Plain text message Possible Reason Recommendation
no.
E20 Temperature at electronic unit
Ambient temperature too high Cool down environment, place Protec
is too high (>60°)
Ventilation failure Check if fans on both side of main unit
Air filter dirty Clean or replace main air filter
W24 24V for the ext. control unit too
low
For RC versions only: Fuse on the RC-driver board blown
E25 Remove sniffer from CAL port Sniffer inserted in calibration port
during start-up or sniffer needs to be removed during internal calibration process
Light barrier of PRO-Check dirty Blow out with fresh air and / or clean
E26 Background too high or Wise
sensor defective
Environment contaminated with helium or Wise Technology sensor defective!
W28 Real time clock reset! Please
CPU-board has been replaced Please enter date and time!
enter date and time!
Battery on CPU-board faulty Replace CPU-board*
P3000 in cooler area.
are running (check for air flow through inlets on both side of main unit housing)
See Section 7.2 Replace fuse on RC-driver board!*
Remove sniffer from calibration port of PRO-Check!
with cotton swab! Offer fresh air to sniffer line.
If problem persists call nearest INFICON service representative
See Section 4.4.5
E32 Wise Technology current too
high
Wise Technology sensor current exceeds threshold value
W34 Flow has changed! Flow has changed by more than
30% since last calibration (warning will disappear if change drops below 20% again).
W35 Flow through sniffer to o low Actual flow is lower than lower flow
limits (In
LOW FLOW mode)
Filter in sniffer line blocked
Capillary blocked
Filter in main unit blocked Lower flow limit setting too high
Do not switch off Protec P3000(XL)!!!
Supply fresh air through sniffer line quickly, restart the Protec P3000 and wait for Protec P3000 to recover! If problem persists, call nearest INFICON service repesentative!
Recalibrate the Protec P3000 (See Section 3.5) or replace sniffer line filters!
Replace sniffer tip filter See Section 7.4
Replace filters of sniffer line and recalibrate! (felt filter, capillary filter and / or filter pad in SL3000XL sniffer line and recalibrate) See Section 7.4 or Replace probe cable*
Replace internal filter* Decrease lower flow limit
See Section 4.4.1
Protec P3000 Messages 5-3
Error
Plain text message Possible Reaso n Recommendation
no.
E37 Flow through sniffer too high Actual flow is higher than upper
flow limits (
FLOW mode) or actual pressure is
LOW FLOW and HIGH
higher than upper pressure limit
HIGH FLOW mode only):
(
Capillary broken or leaky
Replace sniffer line and recalibrate or Replace probe cable*
Increase upper flow limit
Upper flow limit setting too low
See Section 4.4.1
E38 Pump error Fore pump defective Call nearest INFICON service
repesentative!
W39 Valve block error! Cannot recognize valve block
version during self-test W40 Ratio of
failed
HIGH FLOW to LOW FLOW
For Protec P3000XL only in HIGH
FLOW mode:
Flow in HIGH FLOW mode less than
5 times flow in LOW FLOW mode
(during start-up or calibration only)
Filter in sniffer line blocked
Internal filter in main unit blocked
Leak in
LOW FLOW capillary
Internal leak W41
HIGH FLOW too low For Protec P3000XL only in HIGH
FLOW mode:
Actual flow lower than lower flow
limit (for high flow mode) or actual
pressure at Wise Technology
sensor below corresponding
pressure limit
Filter in sniffer line blocked
Internal filter in main unit blocked
Leak in
LOW FLOW capillary
Internal Leak
Call nearest INFICON service representative!
Replace sniffer tip filter and recalibrate(See Section 7.4)
Replace internal filter* and recalibrate Replace sniffer line and recalibrate
If problem persists, call nearest INFICON service representative
Replace sniffer tip filter (See Section
7.4)
Replace internal filter* Replace sniffer line!
Or: Replace probe cable!* See Section 7
5-4 Protec P3000 Messages
Error
Plain text message Possible Reason Recommendation
no.
W42 Pressure at calibration differs
between
HIGH and LOW FLOW
For Protec P3000XL only in
FLOW mode:
HIGH
Pressure at Wise Technology sensor differs from LOW FLOW mode to high flow mode (during start-up or calibration only)
Filter in sniffer line blocked
Internal filter in main unit blocked
Capillary blocked
W53 Flow at calibration out of limits! Only for Protec P3000 or Protec
P3000XL in LOW FLOW mode: Flow at calibration higher than upper flow limit or lower than lower flow limit or limit low flow error (if enabled, whichever comes first)
Filter in sniffer line blocked!
Internal filter in main unit blocked
Leak in (low flow) capillary
Wrong flow limit or limit low flow error settings (for
W54 Flow at calibration out of limits! Only for Protec P3000XL in
FLOW mode:
LOW FLOW mode)
HIGH
Flow at calibration higher than upper flow limit or lower than lower flow limit or limit low flow error (if enabled, whichever comes first)
Filter in sniffer line blocked
Internal filter in main unit blocked
Leak in probe cable Wrong flow limit or limit low flow
error settings (for
HIGH FLOW mode)
Replace sniffer tip filter and recalibrate (See Section 7.4)
Replace internal filter* and recalibrate Replace sniffer line and recalibrate!
Or: Replace probe cable!* and recalibrate
See Section 7
Replace sniffer tip filter and recalibrate! (see section 7.4)
Replace internal filter* and recalibrate! Replace sniffer line (or: replace probe
cable*) and recalibrate! Check flow limit settings and limit low
flow error (for LOW FLOW mode)! See section 4.4.1
Replace sniffer tip filter and recalibrate! (see section 7.4)
Replace internal filter* and recalibrate! Replace sniffer line (or: replace probe
cable*) and recalibrate! Check flow limit settings and limit low
flow error (for HIGH FLOW mode)! See section 4.4.1
Protec P3000 Messages 5-5
Error
Plain text message Possible Reaso n Recommendation
no.
E55 Flow through sniffer below error
limit
E56
HIGH FLOW below error limit Only for Protec P3000XL in HIGH
W59 EEPROM parameter queue
overflow!
Only for Protec P3000 or Protec
P3000XL in LOW FLOW mode:
Flow through sniffer line below limit
low flow error for LOW FLOW mode
Filter in sniffer line blocked!
Internal filter in main unit blocked
Limit low flow error for LOW FLOW
mode too high
FLOW mode:
Flow through sniffer line below limit
low flow error for HIGH FLOW mode
or pressure at Wise Technology
sensor lower than corresponding
lower pressure limit
Filter in sniffer line blocked!
Internal filter in main unit blocked
Limit low flow error for HIGH FLOW
mode too high
May occur if software “update” to
older version is performed
Replace sniffer tip filter and recalibrate! (see section 7.4)
Replace internal filter* and recalibrate! Check limit low flow error (for LOW FLOW
mode)! See section 4.4.1
Replace sniffer tip filter and recalibrate! (see section 7.4)
Replace internal filter* and recalibrate! Check limit low flow error (for HIGH FLOW
mode)! See section 4.4.1 Restart the Protec P3000, if problem
persists, call nearest INFICON service representative!
W60 All EEPROM parameter lost!
Please check your settings!
New EEPROM has been installed,
EEPROM on motherboard is virgin
If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty W61 EEPROM parameter initialized! Software update performed and
new parameters have been
introduced
Newly introduced parameters are
listed below warning
If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty W62 EEPROM parameter lost! Parameter has been modified
during software update and reset
to default value
Affected parameters are listed
below warning
If message comes up repeatedly
during start-up, EEPROM on
motherboard is faulty W64 There are outstanding
warnings!
Acknowledged but still valid
warnings will be repeated every 2
hrs or on each new power on
All settings in software menu are reset to default! Please enter your settings again!
Replace EEPROM*
Acknowledge warning
Replace EEPROM*
Check setting of modified parameters in corresponding software menu and set to desired value!
Replace EEPROM*
Please double-check the warnings!
5-6 Protec P3000 Messages
Error
Plain text message Possible Reason Recommendation
no.
W65 Wrong date set! Wrong date in the Protec.
Wrong Hex code entered to PRO-Check
W66 New PRO-Check! New PRO-Check reference leak
installed in main unit
W67 PRO-Check will be expired on
DD.MM.YYYY
Warning for pending expiration will be issued adjusting 14, 30, 60, or 90 days before actual expiration.
W68 PRO-Check expired! PRO-Check has been used for
more than 1 year or has been manufactured more than 2 years ago.
W70 All EEPROM parameter of
PRO-Check lost!
W71 No communication with PRO-
Check!
EEPROM in PRO-Check is empty or faulty
No or defective electrical connection between PRO-Check and main unit
PRO-Check not installed in main unit
No PRO-Check available Disable PRO-Check in software menu
Check the date in the Protec. Check the Hex code entered to
PRO-Check Please enter serial number and code.
See Section 4.7, Replace PRO-Check Please order new PRO-Check gas
reservoir! (cat. no. 521-010) See Section 7.7.5
Please replace PRO-Check gas reservoir!
See Section 7.7
Replace PRO-Check!
Check connection of PRO-Check with main unit If problem persists, call nearest INFICON service repesentative!
Install PRO-Check!
(see Section 4.5.6)
W72 No communication with sniffer! No or defective electrical
connection between sniffer line and main unit
Check connection of sniffer line with main unit (disconnect and reconnect, if possible, try different sniffer line) If problem persists, call nearest INFICON service
W78 Signal differ ence between
calibrated leak and air too small!
Calibrated leak too small / empty during calibration or proof
Background too high during calibration
Not enough time expired for air signal to stabilized (acknowledged
Check leak rate of test leak or use leak with higher leak rate
Check helium background See Section 4.6, Info page 2.
Repeat calibration, allow sufficient time for air signal to stabilize!
too early)
W81 Calibration factor too low! Calibration factor determined to be
< 0.1 during calibration Leak rate of test leak is incorrect
Check correct setting of test leak value! (especially during external calibration)
W82 Calibration factor too high! Calibration factor during calibration
determined to be > 10 Leak rate of test leak is incorrect
Check correct setting of test leak value! (especially during external calibration)
Test leak not sniffed properly / not
long enough
Repeat calibration, sniff test leak
properly and for sufficient time
Protec P3000 Messages 5-7
Error no.
W86 Interna l calibration not possible Not in measure mode when
W87 Not supported gas in PRO-
Plain text message Possible Reaso n Recommendation
calibration is attempted Occurs only during internal
Check
calibration or proof function:
Wait until Protec P3000 enters measurement mode!
ECO-Check installed (accidentially)
EEPROM in PRO-Check not programmed
W88 PRO-Check defective Occurs only during internal
calibration or proof function: Temperature sensor defective
W89 Overrange! Protec P3000 is contaminated with
helium
Test leak value too high during external calibration!
W90 Calibration conditions not
maintained
Sniffer removed during internal calibration, internal calibration aborted
Take out ECO-Check, install PRO­Check!
Replace PRO-Check reference leak!
Replace PRO-Check reference leak!
Do not switch off Protec P3000(XL)!!! Keep Protec P3000 running while providing fresh air to sniffer line until warning is cleared.
If warning occurs frequently, increase contamination limit!
Check helium background See Section 4.6, Info page 2.
Use smaller test leak for external calibration!
Repeat calibration acknowledge
* To be performed only by authorized INFICON service personnel.
5-8 Protec P3000 Messages
6 Equipment Connections
Warning
1 2 3 4 5 6 7
8 9 10 11 12 13
14 15 16
17 18
19 20 21 22 23 24 25
Log 1.000V
Leck
Fehler
Leck
NO
Lin 0.100V
Ready
Kalibrierung stabil
24V Output
Standby
Zero
Nicht belegt
Kalibrierung abbrechen
bereit
Fehler rücksetzen
NO
Kalibrierung
C
NC
C
NC
The Protec P3000 is equipped with three electrical control connectors. The electr ical connections (head phone, I/O port and RS232) are located on the rear of the main unit directly next to the socket for the mains cable.
6.1 I/O Port (Control Inputs and Outputs)
For all contacts of the I/O Port a maximum voltage of 60 V DC or 25 V AC must not be exeeded or reached to ground or ground equipment conductors. According to the type of in- or outputs lower voltages had to be accepted. For this, please refer to the information given in the responding chapters.
Through this connection some functions of the Protec P3000 can be controlled externally or measurement data or the of the Protec P3000 status may be communicated to external equipment.
Through relay change over contacts the trigger levels as well as th e operating mod e (Ready) of the Protec P3000 may be monitored.
Fig. 6-1 Default Pin Assignment
Equipment Connections 6-1
6.1.1 Ground connectors
Caution
Pin 2, 3, 18, 19 and 21 are ground connectors.
6.1.2 24V Output
Pin 6 is a +24 V common output for supplying the PLC inputs and outputs, internally protected with fuse F4.
6.1.3 PLC Inputs
These inputs can be used to control the Protec P3000 via a programmable logic control (PLC).
Permissible maximum input voltage 28 V.
Technical data
24V nominal input Low level: 0 …7 V High level: 13 … 28 V
Pin assignment (default)
Pin Default Command
7 Standby 8 Clear Error 9 Calibration Abort 13 Calibration 20 Zero 25 Not used
All PLC inputs can be defined by the user from a list of commands. The commands can be selected in the details on this submenu please refer to Section 4.4.4. The currently selected commands can be viewed on INFO page #7.
SETTINGS / INTERFACES / SELECT PLC INPUTS submenu. For
Standby
Change from Change from
6-2 Equipment Connections
LOW to HIGH to activate STANDBY function. HIGH to LOW to deactivate STANDBY function (wake up).
Loading...