Including information on: M1754A Signal
Averaged ECG (SAECG), M1755A A.02.00, and
A.03.XX Software, M1756A FAX Modem, and
M1790A Vectorcardiography (VCG)
HEWLETT
PACKARD
HP Part No. M1700-91909
Printed in USA September 1995
Edition 7
E0995
Notice
The information in this document may change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this
material, including, but not limited to, the implied warranties of
merchantability and fitness for a particular purpose. Hewlett-Packard
shall not be liable for errors herein or for incidental or consequential
damages in connection with the furnishing, performance or use of this
material.
This document contains or refers to proprietary information which
is protected by copyright. All rights are reserved. Copying or other
reproduction of this document without prior written permission of
Hewlett-Packard Company is prohibited.
Responsibility of the
Manufacturer
Warranty
Hewlett-Packad only considers itself responsible for any effects
on safety, reliability and performance of the equipment if
all
the
following are true:
n
Assembly operations, extensions, re-adjustments, modifications or
repairs are done by persons authorized by Hewlett-Packard.
m
The electrical installation of the relevant room complies with the
IEC reqirements (IEC regulated countries only).
n
The instrument is used according to the instructions for use
presented in this manual.
Hewlett-Packard warrants this medical product against defects in
materials and workmanship for a period of:
1. One year on instrument.
2. Lifetime on m1700A and M1701A patient module (remote front
end). Lifetime is defined as the end of the product's support life.
This is a minimum of five years after the cardiograph is removed
from Hewlett-Packard’s price list.
3. One year on M1702A patient module (rcmote front end).
4. All other accessories including cables, battery, and consumables
are under standard, 90-day warranty.
If Hewlett-Packard receives notice of such defects during the warranty
period, Hewlett-Packard shall, at its option, either repair or replace
hardware products which prove to be defective.
Hewlett-Packard software and firmware products that are designated
by Hewlett-Packard for use with a hardware product, when properly
installed on that hardware product, are warranted not to fail to
execute their programming instructions due to defects during the
warranty period. Hewlett-Packard shall repair or replace software
media and firmware that do not execute their programming
instructions due to such defects. Hewlett-Packard does not warrant
that the operation of the software, firmware, or hardware shall be
uninterrupted or error free.
If Hewlett-Packard is unable, within a reasonable time, to repair or
replace any product to a condition as warranted, Buyer shall be
entitled to a refund of the purchase upon return of the product to
Hewlett-Packard.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from any
of the following:
1. Improper or inadequate maintenance by Buyer.
2. Buyer-supplied software or interfacing.
3. Unauthorizd modification or misuse.
4. Operation outside of the environmental specifications for the
product.
5. Improper site preparation and maintenance.
THE WARRANTY SET FORTH ABOVE IS EXCLUSIVE
AND NO OTHER WARRANTY, WHETHER WRITTEN OR
ORAL, IS EXPRESSED OR IMPLIED. HEWLETT-PACKARD
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR
Alternating current.
Equipotential (this is on the ground
if
The niaxin1unl weight that the cart can hold.
Please see chapter 4,
“Pcrfomance Verification ant1 Maintenance,”
lug).
for safety requirements that apply to the cardiograph.
Warning statements describe conditions or actions that can result in
personal injury or loss of life.
Ca,ution
tlaitia.gc
Notes couta.in zdditiollal infuonl;I.tioIr
sta?.Lci~rc~~~Ls
Lo
LIIC 0qrtip111enL or sol’t~v;\w.
tlcstrriI)(~
wi~tlitions or actions Ll1a.t can
on
c:artliogra.ph usage.
result in
Sof tkey
a
r~~prwriits Llic> Lrrnpor~lry
Ll~r display.
rcplTsc~llLs
liC’.\iS
011 LllP
I‘I~OIIL
IiCy
lal,l~cls Iliat appear on
pwl.
V
Preface
This nlanual contains service information for the Hewlett-Packard
M1700A PageWriter XLi, M1701A PageWriter XLs, and M1702A
PageWriter XLe cardiographs. The informat,ion and procedures in
this nlanual apply to all models unless otherwise specified.
This manual is orgmizetl as follows:
Chapter l-Introduction. Contains a general description of
the cardiographs, lists of technical specifications, and lists of
accessories ant1 options.
Chapter 2-Installation aud Configuration. Snnlmarizes the
cardiograph insta.lla.tion ant1 explains how to Configure the
cardiograph for specific customer requireinents.
Chapter 3-Operation. Suniniarizes how to operate tlie
cardiograph.
Chapter 4-Performance Verification aud Maintenance. Explains
how to CheCli the ca.idiograph’s perforlna.nce using built-in
self-tests, a.ntl lists lnai
nlciln.ncr
procctlurcs a.ntl sa.fety requirements
that apply to llir (~r7rtliogi~aph.
Chapter 5-Theory of Operation.
the cardiogra~pli Works and
sul~assrinl~lies.
tlcscri1)c.i; the opcrn.tion of the major
Provides a.11 overview of how
Chapter 6-Troubleshooting. (:ollta7dns proc-ctlures ant1 err01
codes to aid t;lle srrvicc person in ioca.lizin g faults to ;L repla.ceable
subnsseinbly.
Chapter 7--Removal and Replacement. (~ollt.ains procedures
for
rciuovitlg
snl~assrnil~lics.
Chapter 8-Parts List.
replnwnl)lr
Appendix A-Connector Pin Assignments.
the
signals assig1ird
alit1 wplncing
parts,
antI provitlrs n5srtnl)l.~ tlm.\\:it~gs.
Lo
c:l(.I~ of 1.110 ci~i~tliograpll’s major
Lists
part IIUIII~~I.S for llrc cwliograph’s
Itlc~ltifies and defines
tlio sul~a.ss;e~nl~ly iliLrrColineclions.
Appendix B-M1754A Signal Averaged ECG System. Enables
the ca.rtliograph to
clclwl,
tlisplny, mtl record lii~h-fi,equency,
low-miplitutlr E,(:C; sign&.
Appendix C-M1756A Direct Digital ECG Fax. Pcmits the
PageWriter XLi to rcceivc or lmisniil from fa.x inachnines.
Appendix D-Vectrocardiography M1790A.
Preventing the VCG Report from Printing
The VCG Report
VCG Report Formats
Formats
..................
Manual Report Formats
Troubleshooting
Parts List
...................
...............
.............
...........
................
......
.......
....
C-1
C-2
C-3
C-3
C-4
C-4
C-5
C-5
C-6
C-7
C-8
D-l
D-2
D-2
D-3
D-3
D-4
D-4
D-4
D-5
D-5
D-6
D-7
D-7
D-10
D-10
D-12
Index
Contents-7
Figures
1-1. M1700A PageWriter XLi Cardiograph with Preview
Display.
2-1. Installing the Hayes® Modem.
2-2. Connecting the Fax Modem Cables.
2-3. Connecting the Printer Cable.
2-4. Connecting the Patient Module.
2-5. The Line Module and AC Switch.
2-6. Installing the Battery. .............
2-7. Fitting the Battery Holder Tab Into the Battery Slot.
2-8. Loading the Paper. ..............
3-1. The Keyboard and Front Panel.
3-2. An Auto 3x4 ECG (3x4).
3-3. An Auto 3x4 ECG With Three Rhythm Strips (3x4,
3R). ...................
3-4. A Manual 6-Lead ECG.
3-5. An SAECG Report. .............
4-1. 12-Lead ECG from ECG Simulator.
4-2. Cleaning the Digital Array Printhead.
5-1. CPU Initialization Flow Chart.
5-2. Simplified System Block Diagram.
5-3. Patient Module Block Diagram.
5-4. CPU Assembly Block Diagram.
5-5. System Memory Block Diagram.
5-6. I/O Control Block Diagram.
5-7. Serial Communications Circuitry.
5-8. Digital Signal Processor Block Diagram.
5-9. The Disk Drive and its Control Circuit.
6-1. Sample Error Log Printout.
6-2. Troubleshooting Flowchart.
6-3. Voltage Test Locations.
7-1. Opening the Battery Compartment.
7-2. Fitting the Battery Holder Tab Into the Battery Slot.
7-3. Removing the Keyboard.
7-4. User Interface Screw Locations.
7-5. Removing the Printhead Cover.
7-6. Printer Drive and Keyboard Ground Plane Screw
Locations. ................
.................
.........
.......
.........
........
.......
........
...........
............
.......
......
.........
.......
........
........
........
...........
........
.....
.....
.........
............
...
......
..
..........
..........
............
......
...........
........
........
1-2
2-2
2-3
2-4
2-7
2-8
2-11
2-12
2-15
3-2
3-6
3-7
3-8
3-9
4-8
4-11
5-3
5-5
5-7
5-9
5-11
5-13
5-14
5-15
5-15
5-17
5-23
5-24
5-25
5-27
6-3
G-G
6-24
7-3
7-4
7-5
7-6
7-7
7-8
Contents-8
7-7. Removing the Printhead Cover.
........
7-8. Printer Drive and Keyboard Ground Plane Screw
Locations.
7-9. The Printer Drive Gears.
................
...........
7-10. Opening and Removing the Top Cover Assembly. .
7-11. Removing the Preview Screen Assembly.
7-12. Removing the Preview Screen Assembly.
.....
.....
7-13. Opening and Removing the Top Cover Assembly. .
7-14. Opening and Removing the Top Cover Assembly. .
7-15. Removing the Printhead Cover and Printhead.
7-16. Removing the Printhead Cover and Printhead.
7-17. Printhead Resistance Selector.
.........
..
..
7-18. Preview Logic Board Removal (Serial Number Prefix
3208A or Later).
.............
7-19. Preview Logic Board Assembly Screw Locations. .
7-20. Screw and Connector Locations on the CPU
Assembly.
7-21. Disk Drive Mounting Screws.
8-1. Patient Module Diagram.
................
.........
...........
8-2. Main Assembly Exploded Diagram (Serial Number
Prefix 3208A or Later).
..........
8-3. Main Assembly Exploded Diagram (Serial Number
8-11. Cabling for Hayes Modem and Direct Connect (Quick
Connect) Transmission
8-12. Cabling for LaserJet Printing
B-1. Connecting the Patient Module.
B-2. An SAECG Report.
B-3. A Template Screen.
..............
B-4. SAECG Patient Module Block Diagram.
C-1. Cabling for FAX Modem Transmission Only
...........
.........
........
.............
....
...
C-2. Cabling for FAX Modem and Direct Connect (Quick
Connect) Transmission
D-1. The VCG Report
...............
D-2. VCG1: Right Sagittal, 0° to 360°.
D-3. VCG2: Left Sagittal, 0° to 360°.
D-4. VCG3: Right Sagittal, -180° to +180°.
D-5. VCG4: Left Sagittal, -180° to +180°.
...........
.......
........
.....
......
7-9
7-10
7-11
7-13
7-14
7-16
7-18
7-19
7-21
7-22
7-24
7-26
7-28
7-30
7-32
8-2
8-4
8-9
8-15
8-17
8-19
8-21
8-23
8-24
8-25
8-26
8-27
B-2
B-6
B-8
B-14
C-9
C-10
D-6
D-7
D-8
D-8
D-9
Contents-9
Tables
1-1. Physical Specifications
1-2. Patient Module Specifications
1-3. SAECG Patient Module Specifications
1-4. ECG to Paper Specifications
1-5. Power Supply and Battery Specifications
1-6. Environmental Specifications
1-7. Miscellaneous Specifications
1-8. Country Option Configurations
2-1. Line Voltage Settings .............
2-2. Line Fuse Ratings for Voltage Settings
2-3. Global Configuration Parameters
2-4. Global Configuration Parameters (continued)
2-5. Global Configuration Parameters (continued)
2-6. Site Transmission Configuration
2-7. Site Receive Configuration
2-8. LaserJet Printer Configuration
6-1. Power-on Errors ...............
6-2. Power Supply and Battery
6-3. Patient Module and Patient Data Cable
6-4. CPU Assembly (CPU, FGA, ROM, RAM)
6-5. Printer ...................
6-6. Disk Drive (M1700A and M1701A only)
6-7. Preview Display and Logic (M1700A)
6-8. Keyboard and Keyboard Display (LCD)
6-9. Transmit and Modem .............
7-1. Printhead Resistance Codes
8-1. Patient Module Parts List
8-2. Main Assembly Parts List (Serial Number Prefix
3208A or Later) ..............
8-3. Main Assembly Parts List (Serial Number Prefix
Prior to 3208A) ..............
8-4. Fastener List .................
8-5. Printer Parts List ..............
8-6. Preview Display Assembly Parts List
8-7. Cart Parts List ................
8-8. Cart Fastener List ..............
8-9. M1705B Cart Parts List
8-10. Data Communications Package Parts List
8-11. Direct Connect Cable Parts List
8-12. Hayes Modem Cable Parts List
8-13. Hayes Modem and Direct Connect Cables Parts List
8-14. LaserJet Cable Parts List
............
.........
.....
..........
....
..........
..........
.........
.....
........
. .
. .
........
...........
.........
...........
.....
....
.....
......
.....
..........
...........
......
............
....
........
.........
...........
1-5
1-5
1-6
1-6
1-7
1-8
1-8
1-14
2-9
2-9
2-20
2-22
2-23
2-29
2-30
2-31
6-8
6-9
6-12
6-14
6-15
6-17
6-19
6-20
6-21
7-23
8-3
8-5
8-10
8-14
8-16
8-18
8-20
8-20
8-22
8-23
8-24
8-25
8-26
8-27
Contents-10
A-1. Keyboard and Keyboard Display (LCD) Connector
J1....................
A-2. Printhead Logic Connector J2
A-3. Printer Motor and Sensor Connector J3
A-4. Battery Connector J4
B-4. Patient Module and Patient Data Cable
B-5. SAECG Parts List
C-1. Transmit and Fax
...............
..............
C-2. Fax Modem Parts List
C-3. Fax Modem Cable Parts List
...........
.....
............
..........
C-4. Fax Modem and Quick Connect Cable Parts List .
D-1. AHA Lead Set and Patient Module Configurations
D-2. IEC Lead Set and Patient Module
D-3. Global Configuration Fields
D-4. VCG Report Formats
D-5. Troubleshooting
D-6. VCG Parts List
.............
...............
...............
..........
.......
A-2
A-3
A-4
A-5
A-5
A-6
A-7
A-8
A-9
A-10
A-11
A-12
A-12
B-4
B-5
B-15
B-15
B-16
C-7
C-8
C-9
C-10
D-2
D-2
D-4
D-7
D-10
D-12
Contents-11
Introduction
1
This chapter introduces you to the M1700A PageWriter XLi,
M1701A PageWriter XLs, and M1702A PageWriter XLe cardiographs
and lists their technical specifications.
Warning
Note
These cardiographs are only to be serviced by qualified personnel.
Safe and effective use of medical instrumentation requires periodic
inspection and preventive maintenance. Perform the preventive
maintenance procedures in Chapter 4 of this manual at required
intervals to ensure satisfactory instrument performance.
The cardiographs use a thermal printhead to record waveforms and
label the ECG report. The paper supplied with the cardiographs is a
special thermal paper designed specifically to work with the thermal
printhead and the photo detector used to advance the paper.
The cardiograph’s clisk drive (M1700A & M1701A only) is a critical
part. Disks of substandard quality can damage the clrive or destroy
patient clata.
Hewlett-Packarcl guarantees the performance of the cardiographs only
when used with paper and disks that meet or exceed Hewlett-Packard
specifications.
PageWriter XL
Cardiographs
The M1700A PageWriter XLi is Hewlett-Packard’s full-featured,
interpretive cardiograph. The M1701A PageWriter XLs is the
standard, non-interpretive cardiograph. The M1702A PageWriter
XLe is the basic, non-interpretive cardiograph. Each cardiograph
uses a patient module, which is a remote “front end” that acquires
and digitizes clata. The M1700A and M1701A patient modules
feature a start button ant1 an LCD display that indicates signal
quality. The M1702A patient module does not feature a display
or start button. The cardiograph main unit has the user controls,
the disk drive (M1700A and M1701A only), the printer, and all the
processing circuitry. All models use the same enclosure, except when
the h41700A is equipped with preview display. The models fit onto
the M1705A cart clesignecl for these carcliographs. Figure l-1 shows
the M1700A with preview display mounted
on
the cart.
Introduction l-l
Figure l-l. M1700A PageWriter XLi Cardiograph with Preview Display.
The PageWriter XL family of cardiographs can print ECGs in Auto
and Manual formats for different clinical tests or applications. Both
cardiographs print all traces with a continuous feed, high-resolution,
digital array printer. This produces electrocardiograms on standard
page lengths that require
no
cutting or mounting. Records are clearly
annotated with ECG data and patient information. The M1700A
also labels records with automatic ECG measurements and ECG
interpretation for both adult and pediatric patients.
The Ml700 ant1 M1701A cardiographs use flexible disks for storing
ECGs. Exact ECG copies can be made at the bedside or Iater. Each
ECG is automatically logged and can be automatically stored by the
cardiograph. A list of all stored ECGs can be printed, facilitating
administrative reporting and billing. Typically, 50-150 ECGS can
be stored per flexible disk, although the actual quantity depends on
usage.
From a service standpoint, the PageWriter XL cardiographs are
similar except for five major hardware differences:
H The h41700A preview display allows you to view the ECG prior to
recording.
1-2 Introduction
n
The M1700A CPU assembly has extra nlenlory to accommodate
the measurement and interpretation software, but uses the same
ROM as the M1701A.
n
Each PageWriter has a unique CPU asscnibly.
n
The M1700A and M1701A load software from, and store ECGs to,
the disk drive. The M1702A software is in ROM, and does not
store ECGs.
n
The M1700A and M1701A patient modules feature a display that
indicates signal quality and a start button. The M1702A patient
module does not offer these features.
All other model differences are found in the cardiograph’s software.
The M1700A and M1701A software is installed in the cardiograph
from flexible disk, while the M1702A software is installed from ROM.
After installation, the software is stored in dynamic RAM (DRAM).
Re-installation is required only when a user desires a different
configuration, or if DRAM loses its contents due to a low battery.
The PageWriter XL family of cardiographs is designed for long-term
reliability. The modular clesign makes extensive use of VLSI and
gate array technology, resulting in a minimum number of major
subassemblies. The modular approach means less down-time for the
user, since replacing subassemblies allows quicker field repairs. The extended self-test efficiently identifies faulty subassemblies, further
speeding the repair process.
Note
:’
t?
M1705A Cart
As of February 1992, all new PageWriter XL cardiographs are
manufactured to a revised mechanical design specification. All units
having a serial number prefix of 3208A or later conform to this
revised specification. This manual describes all specific differences.
The following paragraphs describe some of the significant options
available for the PageWriter XL cardiographs.
The M1705A Cart provides mobility for the PageWriter XL family
of cardiographs. The large wheels make the cart easy to move, yet
steady when in use. Slots in the cardiograph’s feet and a thumbscrew
secure the cardiograph to the cart rails, The tray directly below the
cardiograph provides storage for acquired ECGs, the user’s guide,
and several flexible disks. The tray also has molded fittings that
accommodate the modem in the M170GA/B Telecommunications
Package. Storage for the power cable is built in, and other built-in
cable retainers hold the patient clata cable out of the way. Two
compartmented trays provide storage for the patient module,
spare patient electrodes and leaclwires, additional thermal paper,
consumable supplies, and miscellaneous items. An optional paper
management tray is also available for the M1705A.
Introduction 1-3
M1706A/B
Telecommunications
Package
With the addition of the M1706A or M1706B Telecommunications
Package, the M1700A can transmit ECGs over telephone lines to
another site. The receiver at the other site can be a Hewlett-Packard
ECG Management system or another M1700A. This feature is not
available for the M1701A.
The M1706A or M1706B Telecommunications Package consists
of a Hayes@ modem, a bracket that attaches the modem to the
cart, a power cable that connects to the modem power outlet on
the cardiograph, and data cables that connect the modem to the
cardiograph and the telephone system. Local laws, regulations, and
phone system standards determine the specific type of modem and
accessories included and the types of cables included in the M1706A
or M1706B Telecommunications Package.
M1755A A.03.XX
A.02.XX Software
A.01 .Xx Software
Inquiries
Software
This software is shipped with PageWriter XLi on a standard basis.
This revision provides the PageWriter XLi with severity-only
analysis, LaserJet printing option, and extended configurable
transmission phone numbers. This revision includes all A.02.XX
enhancements and also provides fax transmission, SAECG, and VCG
capabilities. The software is shipped with all new PageWriter XLi
cardiographs in late 1992.
This software revision provides the PageWriter XLi with enhanced
preview displays, interactive query, and FAA transmission. The
software also gives the XLi fax transmission using the M175GA
modem and SAECG capabilities. This revision is only available for
support purposes.
The original software for PageWriter XLi provides all basic features
with preview display. This revision is only available for support
purposes.
Refer any questions or comments regarding these instruments
to the nearest Hewlett-Packard Sales/Service Office or to one of
Hewlett-Packard’s Service Dispatch Centers. They can provide
you with additional information about service training, special
applications, and schematics. Always identify the instrument
by model number and serial number in all correspondence.
Hewlett-Packard sales and service offices are listed at the end of this
manual. Toll-free numbers for Service Dispatch Centers are listecl in
Chapter 8, “Parts List.”
1-4 Introduction
Specification Data
The following tables list the technical specifications for the
cardiographs and the mobile cart. Specifications are the same for all
models except as noted.
Table 1-1. Physical Specifications
Parameter
Dimensions (1 x w x h)
patient module
cardiograph
cart
Weight
patient module
cardiograph
cart
Connector life
patient module
RS-232
Chemical resistance, cleaners
Parameter Specification
Resolution 5 /iv after processing
4 in x 5 in x 1.5 in (10.11 cm x 12.7
15.5 in x 22.25 in x 5.5 in (39.37 cm
33.37 in x 17.12 in x 36.0 in (84.76 cm x 43.48 cm x 91.44 cm)
15 oz. (425.25 g) or less
46 lbs. (20.87 kg) or less (includes battery and preview display)
38 lbs. (17.24 kg) or less
350 cycles (insertions)
350 cycles (insertions)
Withstands the following: isopropyl alcohol (except. leadwires and patient
data cable), mild soap and water, chlorine bleach and water (30 ml/l of
water).
Table 1-2. Patient Module Specifications
SDecification
cm x 3.81 cm)
x
56.5 cm x 14 cm)
I
Input impedance
Gain accuracy
Input bias 40 nA or less for electrode wire input, less than 500 nA for Right Leg
Patient isolation, risk currents
Defibrillator protection
Crosstalk rejection 50 dB or greater with 10 mV p-p on other channels
Calibration signal 1 mV f2% after processing
DC offset tolerance
25 MSt @ 10 Hz, typically 1OOMQ @ DC
f3% of
110 dB or greater with AAMI load Common mode rejection
10 pA RMS or less @ 240 V AC 50 Hz, or 120 V AC 60 Hz
System recovery 5 seconds after 400 joule discharge, no damage
4 kHz Sample rate
*300 mV
input
signal +lO ILV
Introduction l-5
I
Table 1-3. SAECG Patient Module Specifications
Parameter
Resolution
Input impedance
Gain accuracy
Input bias
Common mode rejection
Patient isolation risk currents
I,--
Crosstalk rejection
I
Calibration signal
Sample rate
DC offset tolerance
Specification
1.22 /IV after processing in high-resolution mode
5 /IV after processing in standard mode
25 MR @ 10 Hz, typically 1OOMfi @ DC
&2% of input signal
] 40 nA or less for electrode wire input, less than 500 nA for Right Leg
120 dB or greater with AAMI test load
10 ItA RMS
System recovery 5 seconds after 400 joule discharge, no damage
50 dB or greater with 10 mV p-p on other channels in standarcl mode
I
1 mV f2% after processing
8 kHz
I
355 mV
Table 1-4. ECG to Paper Specifications
or less @ 240 V AC 50 Hz,
or 120 V AC 60 Hz
I
I
Paranwter
Frequency response
Auto ECG
Manual ECG
SAECG
power line
ECG resolution on paper
voltage axis
time axis
Swxificatiolk
3 dB down at (user selectable): 0.05-150 Hz, 0.15-150 Hz, 0.5-150 Hz,
0.05-100 Hz, 0.15-100 IIz, or 0.5-100 Ilz, 0.15-40 Hz, 0.5-40 Hz
3 dB down at (user selectable): 0.05-40 Hz, 0.05-150 Hz, 0.5-40 Hz, 0.5-100
Hz, 0.05-100 Hz
3 dB down at (user selectable): 25-300 IIz, 40-300 Hz, SO-300 Hz
115 dB rejection at 50 or 60 11s with less than 300 mV DC offset
200 dots/in (8 dok/mm)
500 dots/in (20 dots/mm) at 25 mm/s and slower
250 dots/in (10 dots/mm) at 50 mm/s
1-6 Introduction
Table 1-5. Power Supply and Battery Specifications
Parameter
AC line frequency
100 V setting
120 V setting
220 V setting
240 V setting
Battery voltage
50 and 60 Hz
90-110 V AC
108-132V AC
198-253 V AC
216-269 V AC
18 V DC, 6.5 Ahr
Specification
Battery recharge time (from 70% down)
to 90% capacity
to full capacity
8 hours typical
16 hours typical
Battery capacity, M1700A with preview
without storage
with storage
continuous rhythm (Manual mode)
continuous idle
RS-232 bit transmission rates 300, 1200, 4800, 9600, 19200 baud
Printhead life Typically 100,000 pages
I
1-8 Introduction
Options and
Accessories
PageWriter XLi, XLs
and XLe Cardiographs
These tables list the options and accessories available for PageWriter
XL cardiographs.
Country Options
Each country option includes appropriate power cord, voltage,
printer, patient module, and language. See Table 1-8, which shows
the configuration of each country option.
ABA
ABB
ABC
ABD
ABE
ABF
ABG
ABH
ABK
ABM
ABS
ABU
ABZ
AB4
ACF
ACQ
AKM
AKV
North American English Version
European English Version
French Canadian Version
German Version
Spanish Version
French Version
Australian Version
Dutch Version
Intercon English Version
Latin American Version
Swedish Version
U .I<. Version
Italian Version
Singapore Version
Japanese Version
South African Version
Chinese Version
South American Version
Delete Options
DOlDO2
DO3
DO4
DO5
DO6 _
Cable Options
Cable options are applicable to the M1700A only.
JO1
JO2
Delete preview display (M1700A only)
Delete cart (M1700A, M1701A only; the M1702A is
shipped without a cart)
Delete standard AHA patient module
Delete standard IEC patient module
Delete interpretation from the M1700A
Delete interpretation and preview from the M1700A
Direct transmission cable (24542M)
ECG Workstation adqter (92224F)
Introduction 1-9
Supplies/Accessories Options
A04
K01
Add paper tray to M1705A cart (M1705-89500)
Extra leads kit (Frank, pediatric, research leads)
(M1717B)
K02
K03
K04
K05
K06
g-meter patient data cable (M1720A)
Extra paper (8.5~11, 1 case/1600 sheets) (M1707A)
Extra A4 paper (1 case/1600 sheets) (M1709A)
Dust cover (M1700-87600)
Extra disposable electrodes (6 boxes each of 13943B,
lOOO/box; 1 set 13946A electrode adapters included)
Includes Ml754-17700 A.O1.OO SAECC software set, SAECG patient
module, and operator’s b
The M1754A option requires the M1755A (A.Oa.XX, A.03.OX or
higher) software. The Ml 754A option will not function with A.03.1X
software.
U.S. Modem (Hayes@ 288)
France Modem (Hayes@ 288)
Netherlands Modem (Hayes@ 288)
Quick Connect cables for Hayes@ modem
(M1706-61636 and R/11706-61637)
Upgrade XLe to XLs
Gde. Country option must also be ordered.
M1755A PageWriter XLi
Upgrade Kit
Note .I
t?
1-12 Introduction
Documentation Options
OBP
OBQ
Upgrade to version A.03.XX operating software. Includes two system
disks and two blank disks. Specify country option.
This software revision is the standard software that ships with all
.
XLi cardiographs as of late 1992.
SAECG User Training Video VHS/NTSC
(M1754-S9504)
SAECG User Training Video VHS/PAL
(M1754-S9505)
M1756A PageWriter XLi
Direct Digital ECG Fax
Modems
Includes M1756-07300 fax modem bracket and M1756-61600 fax
modem cable.
Note .I
4?
The M1756A option requires the M1755A (A.02.XX or higher)
software. ECG Fax capability is also available using the M1706B
modems with A.03.X3 or higher software.