Joop Geesinkweg 140
1114 AB Amsterdam, The Netherlands
Australia Sponsor
FUJIFILM SonoSite Australasia Pty Ltd
114 Old Pittwater Rd
Brookvale, New South Wales 2100, Australia
Caution:
Edge, SiteLink, SonoCalc, SonoHD2, SonoMB, SonoMBe, SonoSite, and the SonoSite logo are registered (in some jurisdictions) and
unregistered trademarks owned by FUJIFILM SonoSite, Inc.
DICOM is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital
communications of medical information.
Non-SonoSite product names may be trademarks or registered trademarks of their respective owners.
The SonoSite ultrasound system(s) referenced in this document may be covered by one or more of the following U.S. patents:
Patents: US 8,439,840; US 8,398,408; US 8,355,554; US 8,216,146; US 8,213,467; US 8,147,408; US 8,137,278; US 8,088,071; US 8,066,642;
US 8,052,606; US 7,819,807; US 7,804,970; US 7,740,586; US 7,686,766; US 7,604,596; US 7,591,786; US 7,588,541; US 7,534,211; US
7,449,640; US 7,169,108; US 6,962,566; US 6,648,826; US 6,575,908; US 6,569,101; US 6,471,651; US 6,416,475; US 6,383,139; US
6,364,839; US 6,203,498; US 6,135,961; US 5,893,363 ; US 5,817,024; US 5,782,769; US 5,722,412; US 8,805,047; US 8,527,033; US
8,858,436; US 8,861,822; US 8,956,296; AU: 730822; AU: 727381; CA 2,372,152; CA: 2,371,711; CN 98108973.9; CN 98106133.8; CN
97113678.5; DE 69831698.3; DE 69830539.6; DE 69730563.5; DE 602004027882.3; DE 602004023816.3; DE: 60034670.6; DE 60029777.2;
EP 1589878; EP 1552792; EP 1180971; EP 0875203; EP 0815793; EP 1180970; EP 0881492; ES 2229318; ES 159878; ES 1552792; ES
0881492; FR 158978; FR 1552792; FR 1180970; FR 0881492; FR 0875203; FR 0815793; GB 158978; GB 1552792; GB 1180971; GB 1180970;
GB 0881492; GB 0875203; GB 0815793; IT 1589878; IT 1552792; IT 0881492; IT 0815793; JP 4696150; KR 532359; KR 528102; NO 326814;
NO 326202 and pending.
Federal (United States) law restricts this device to sale by or on the order of a physician.
Before servicing the SonoSite Edge II Ultrasound System, please read this manual.
The ultrasound system has multiple configurations and feature sets. All are described in this service manual
but not every option may apply to your system. System features depend on your system configuration,
transducer, and exam type.
Refer to the SonoSite Edge II Ultrasound System User Guide for additional information regarding safety,
system controls, operation, capabilities, and specifications.
This chapter also defines labeling symbols, specifications, and standards.
Audience
The intended audience of this manual is properly trained field and in-house service personnel.
Contact Information
Questions and comments are encouraged. SonoSite is interested in your feedback regarding the service
manual. If you encounter difficulty with the system, use the information in this manual to help correct the
problem. If the problem is not covered here, contact SonoSite Technical Support as follows:
Technical Support (USA, Canada)1-877-657-8118
Technical Support fax:1-425-951-6700
Technical Support e-mail:ffss-service@fujifilm.com
SonoSite website:www.sonosite.com
International Technical Support:Contact your local representative or call (USA) +425-951-1330
Italy
Tel 0039 02 94753655
e-mail: ERAF-service@fujifilm.com
Asia/PacificTel: (+65) 63805581
e-mail: FFSS-APACME-service@fujifilm.com
Chapter 1: Introduction1
Conventions, symbols, and terms
LOT
REF
The user guide follows these conventions:
•A WARNIN G describes precautions necessary to prevent injury or loss of life.
•A Caution describes precautions necessary to protect the products.
•Numbered steps in procedures must be performed in order.
•Items in bulleted lists do not require performance in sequence.
Labeling symbols
The following symbols are used on the products, packaging, and containers.
Table 1: Labeling Symbols
SymbolDefinition
Alternating Current (AC)
Class 1 device indicating manufacturer’s declaration of conformance with
Annex VII of 93/42/EEC
Class 1 device requiring verification by the Notified Body of sterilization or
measurement features, or to a Class IIa, IIb, or III device requiring
verification or auditing by the Notified Body to applicable Annex(es) of
93/42/EEC
Attention, see the user guide
Follow instructions for use.
Device complies with relevant Australian regulations for electronic devices.
Batch code, date code, or lot code type of control number
Biological risk
Device complies with relevant Brazilian regulations for electro-medical
devices.
Canadian Standards Association. The “C” and “US” indicators next to this
mark signify that the product has been evaluated to the applicable CSA
and ANSI/UL Standards, for use in Canada and the US, respectively.
Catalog number
Collect separately from other household waste (see European Commission
Directive 93/86/EEC). Refer to local regulations for disposal.
2Chapter 1: Introduction
Table 1: Labeling Symbols (Continued)
STERILE R
STERILE EO
SymbolDefinition
Corrugated recycle
Dangerous voltage
Date of manufacture
Manufacturer
Direct Current (DC)
Do not get wet.
Do not stack over 2 high.
Do not stack over 5 high.
Do not stack over 10 high.
Electrostatic sensitive devices
Device complies with relevant FCC regulations for electronic devices.
Fragile
GELGel
Sterilized using irradiation
Sterilized using ethylene oxide
Hot
Chapter 1: Introduction3
Table 1: Labeling Symbols (Continued)
SN
SymbolDefinition
Device emits a static (DC) magnetic field.
Non-ionizing radiation
Paper recycle
Serial number type of control number
Temperature limitation
Atmospheric pressure limitation
Humidity limitation
Submersible. Protected against the effects of temporary immersion.
Water-Tight Equipment. Protected against the effects of extended
immersion.
Handle transducer with care.
Follow manufacturer’s instructions for disinfecting time.
Disinfect transducer.
Type BF patient applied part
(B = body, F = floating applied part)
Defibrillator proof type CF patient applied part
Underwriter’s Laboratories labeling
Pollution Control Logo. (Applies to all parts/products listed in the China
RoHS disclosure table. May not appear on the exterior of some
parts/products because of space limitations.)
China Compulsory Certificate mark (“CCC Mark”). A compulsory safety
mark for compliance to Chinese national standards for many products sold
in the People’s Republic of China.
4Chapter 1: Introduction
Table 1: Labeling Symbols (Continued)
SymbolDefinition
WARNING:
Connect Only
Accessories and
Peripherals
Recommended
by FujiFilm
SonoSite
WARNING: Connect Only Accessories and Peripherals Recommended by
FujiFilm SonoSite
Chapter 1: Introduction5
6Chapter 1: Introduction
Chapter 2: Specifications
This chapter contains information regarding system specifications and accessory compatibility. The
information applies to the ultrasound system, transducers, accessories, and peripherals.
Specifications
Dimensions
System
•Length: 13 in. (33 cm)
•Width: 12.4 in. (31.5 cm)
•Height: 2.5 in. (6.3 cm)
Display
•Length: 9.7 in. (24.6 cm)
•Height: 7.3 in. (18.5 cm)
•Diagonal: 12.1 in. (30.7 cm)
Environmental limits
Note: The temperature, pressure, and humidity limits apply only to the ultrasound system, transducers,
and battery.
Operating (system, battery, and transducer)
10–40°C (50–104°F), 15–95% R.H.
700 to 1060hPa (0.7 to 1.05 ATM)
Mode of Operation:
Continuous 35°C or below
Non-Continuous above 35°C (30 minutes on /30 minutes off)
Shipping and storage (system and transducer)
-35–65°C (-31–149°F), 15–95% R.H.
500 to 1060hPa (0.5 to 1.05 ATM)
Shipping and storage (battery)
-20–60°C (-4–140°F), 15–95% R.H. (For storage longer than 30 days, store at or below room temperature.)
500 to 1060hPa (0.5 to 1.05 ATM)
Electrical specifications
Power Supply Input: 100-240 VAC, 50/60 Hz, 2.0 A Max @ 100 VAC
Power Supply Output #1: 15 VDC, 5.0 A Max
Power Supply Output #2: 12 VDC, 2.3 A Max
Combined output not exceeding 75 watts.
Chapter 2: Specifications7
Battery specifications
The battery is comprised of six lithium-ion cells plus electronics, a temperature sensor, and battery contacts.
Run time is up to two hours, depending on imaging mode and display brightness.
Compatible accessories and peripherals
SonoSite has tested the SonoSite Edge II ultrasound system with the following accessories and peripherals
and has demonstrated compliance to the requirements of IEC60601-1-2:2007.
You may use these SonoSite accessories and third-party peripherals with the SonoSite Edge II ultrasound
system.
WARNING:
WARNING:
Accessories and peripherals compatible with SonoSite Edge II ultrasound system
DescriptionPart NumberMaximum Cable Length
C8x transducerP080106.0 ft/1.8 m
C11x transducerP076786.5 ft/2.0 m
C35x transducerP219816.0 ft/1.8 m
rC60xi transducerP210705.5 ft/1.7 m
rC60xi transducer armoredP216365.5 ft/1.7 m
rP19x transducerP210156.0 ft/1.8 m
rP19x transducer armoredP215566.0 ft/1.8 m
HFL38xi transducerP203115.5 ft/ 1.7 m
HFL38xi transducer armoredP203775.5 ft/1.7 m
Use of the accessories with medical systems other than the SonoSite
Edge II ultrasound system may result in increased emissions or decreased
immunity of the medical system.
Use of accessories other than those specified may result in increased
emissions or decreased immunity of the ultrasound system.
HFL50x transducerP076936.0 ft/1.8 m
HSL25xP206797.5 ft/2.3 m
ICTx transducerP076906.0 ft/1.8 m
L25x transducerP076917.5 ft/2.3 m
L25x transducer armoredP229507.5 ft/2.3 m
L38xi transducerP127426.0 ft/1.8 m
L38xi transducer armoredP196266.0 ft/1.8 m
L52x transducer (Vet)V000337.5 ft/2.3 m
L52x transducer armored (Vet)V209627.5 ft/2.3 m
P10x transducerP076966.5 ft/2.0 m
TEExi transducerP198257.5 ft/2.3 m
8Chapter 2: Specifications
Accessories and peripherals compatible with SonoSite Edge II ultrasound system (Continued)
Bar code scanner P141664.8 ft/1.5 m
Kit, PowerPackP13559—
Black & white printerP13745—
Hybrid black & white printerP20006
Black & white printer power
—3.3 ft/1 m
cable
Black & white printer USB
—10.8 ft/3.3 m
cable
Color printerP13983—
Color printer power cable—3.3 ft/1 m
Color printer video cable—6.0 ft/ 1.8 m
ECG lead wiresP1420224 in/ 0.6 m
ECG moduleP085015.8 ft/1.8 m
Edge Mini-DockP15078—
Edge StandP15800—
Dual USB Footswitch P146899.8 ft/3.0 m
Power cord (system)P00848 (USA)10 ft/3 m
Power Supply/Battery ChargerP098236.8 ft/ 2 m
PowerParkP12822—
Storage bin kit, Universal/Edge
P22048
stands
Triple Transducer ConnectP16535—
USB wireless adapterP17725—
EU compliant wireless adapterP22625
Edge Battery PackP15051—
Wireless mounting kitP17724
Chapter 2: Specifications9
10Chapter 2: Specifications
Chapter 3: Safety
This chapter contains electrical and clinical safety information required by regulatory agencies. The
information applies to the ultrasound system, transducers, accessories, and peripherals.
Electrical safety
This system meets EN60601-1, Class I/internally-powered equipment requirements and Type BF and Type
CF isolated patient-applied parts safety requirements.
This system complies with the applicable medical equipment requirements published in the Canadian
Standards Association (CSA), European Norm Harmonized Standards, and Underwriters Laboratories (UL)
safety standards. See “Standards” on page 21.
For maximum safety observe the following warnings and cautions.
WARNING:
WARNING:
WARNING:
To avoid the risk of injury, do not operate the system in the presence of
flammable gasses or anesthetics. Explosion can result.
To avoid the risk of electrical shock or injury, do not open the system
enclosures. All internal adjustments and replacements, except battery
replacement, must be made by a qualified technician.
To avoid the risk of electrical shock:
• This equipment must be connected only to a supply mains with protective
earth.
• Use only properly grounded equipment. Shock hazards exist if the power
supply is not properly grounded. Grounding reliability can be achieved only
when equipment is connected to a receptacle marked “Hospital Only” or
“Hospital Grade” or equivalent. The grounding wire must not be removed or
defeated.
• When using the system in an environment where the integrity of the protective
earth conductor arrangement is in doubt, operate the system on battery
power only and disconnect the power supply.
• Do not let the bar code scanner or external mouse touch the patient.
• Do not touch any of the following:
• The power supply and the patient at the same time
• The ungrounded signal input/output connectors on the back of the
ultrasound system
• The system battery contacts (inside the battery compartment)
• The system transducer connector when the transducer or Triple
Transducer Connect (TTC) is disconnected
• The system transducer connector on the TTC if no transducers are
connected
Chapter 3: Safety11
• Do not connect the system power supply or docking system to a multiple
portable socket outlet (MPSO) or extension cord.
• Before using the transducer, inspect the transducer face, housing, and cable.
Do not use the transducer if the transducer or cable is damaged.
• Always disconnect the power supply from the system before cleaning the
system.
• Do not use any transducer that has been immersed beyond the specified
cleaning or disinfection level. See Chapter 7, “Maintenance”.
• Use only accessories and peripherals recommended by SonoSite, including
the power supply. Connection of accessories and peripherals not
recommended by SonoSite could result in electrical shock. Contact SonoSite
or your local representative for a list of accessories and peripherals available
from or recommended by SonoSite.
WARNING:
WARNING:
WARNING:
Caution:
To avoid the risk of electrical shock and fire hazard:
• Inspect the power supply, AC power cords, cables, and plugs on a regular
basis. Ensure that they are not damaged.
• The power cord set that connects the power supply of the ultrasound system
or the stand to mains power must only be used with the power supply or
docking system, and cannot be used to connect other devices to mains
power.
To prevent injury to the operator/bystander, the transducer must be removed
from patient contact before the application of a high-voltage defibrillation pulse.
To avoid possible electrical shock or electromagnetic interference, verify proper
operation and compliance with relevant safety standards for all equipment
before clinical use. Connecting additional equipment to the ultrasound system
constitutes configuring a medical system. SonoSite recommends verifying that
the system, all combinations of equipment, and accessories connected to the
ultrasound system comply with JACHO installation requirements and/or safety
standards such as AAMI-ES1, NFPA 99 OR IEC Standard 60601-1-1 and
electromagnetic compatibility standard IEC 60601-1-2 (Electromagnetic
compatibility), and are certified according to IEC Standard 60950 (Information
Technology Equipment (ITE)).
Do not use the system if an error message appears on the LCD display. Note
the error code and call SonoSite Technical Support for further assistance.
Caution:
To avoid increasing the system and transducer connector temperature, do not
block the airflow to the ventilation holes on the side of the system.
12Chapter 3: Safety
Electrical safety classification
Class I equipmentThe ultrasound system is classified as Class I equipment
when powered from the external power supply or mounted
on the stand because the external power supply is a Class
1 protectively earthed power supply.
The stand has no protective earth. Ground bond testing is
not applicable to the ultrasound system or the stand.
Note:AC powered peripherals that may be used with the system
are Class I and are individually protectively earthed. Ground bond
testing may be conducted on each AC powered peripheral.
Internally powered
equipment
Type BF applied partsUltrasound transducers
Type CF applied partsECG module/ECG leads
IPX-7 (watertight
equipment)
IPX-8 (watertight
equipment)
Non AP/APGUltrasound system power supply, docking system, and
Ultrasound system not connected to the power supply
(battery only)
Ultrasound transducers
Footswitch
peripherals. Equipment is not suitable for use in the
presence of flammable anaesthetics.
Equipment safety
To protect your ultrasound system, transducers, and accessories, follow these precautions.
Caution:
Caution:
Excessive bending or twisting of cables can cause a failure or intermittent
operation.
Improper cleaning or disinfecting of any part of the system can cause
permanent damage. For cleaning and disinfecting instructions, see Chapter 7,
“Maintenance”.
Caution:
Caution:
Caution:
Caution:
Chapter 3: Safety13
Do not submerge the transducer connector in solution. The cable is not
liquid-tight beyond the transducer connector/cable interface.
Do not use solvents such as thinner or benzene, or abrasive cleaners on any
part of the system.
Remove the battery from the system if the system is not likely to be used for a
month or more.
Do not spill liquid on the system.
Battery safety
To prevent the battery from bursting, igniting, or emitting fumes and causing personal injury or equipment
damage, observe the following precautions.
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
The battery has a safety device. Do not disassemble or alter the battery.
Charge the batteries only when the ambient temperature is between 0° and
40°C (32° and 104°F).
Do not short-circuit the battery by directly connecting the positive and negative
terminals with metal objects.
Do not touch battery contacts.
Do not heat the battery or discard it in a fire.
Do not expose the battery to temperatures over 60°C (140°F). Keep it away
from fire and other heat sources.
Do not charge the battery near a heat source, such as a fire or heater.
Do not leave the battery in direct sunlight.
Do not pierce the battery with a sharp object, hit it, or step on it.
Do not use a damaged battery.
Do not solder a battery.
The polarity of the battery terminals isfixed and cannot be switched or
reversed. Do not force the battery into the system.
Do not connect the battery to an electrical power outlet.
Do not continue recharging the battery if it does not recharge after two
successive six hour charging cycles.
WARNING:
WARNING:
WARNING:
Caution:
Do not ship a damaged battery without instructions from SonoSite Technical
Support. (See “Technical Support (USA, Canada)” on page 1.)
If the battery leaks or emits an odor, remove it from all possible flammable
sources.
Periodically check to make sure that the battery charges fully. If the battery fails
to charge fully, replace it.
To avoid the battery becoming damaged and causing equipment damage,
observe the following precautions:
• Do not immerse the battery in water or allow it to get wet.
• Do not put the battery into a microwave oven or pressurized container.
• If the battery emits an odor or heat, is deformed or discolored, or in any way
appears abnormal during use, recharging or storage, immediately remove it
and stop using it. If you have any questions about the battery, consult SonoSite
or your local representative.
• Store the battery between -20°C (-4°F) and 60°C (140°F).
• Use only SonoSite batteries.
• Do not use or charge the battery with non-SonoSite equipment. Only charge
the battery with the system.
14Chapter 3: Safety
Clinical safety
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
WARNING:
Non-medical (commercial) grade peripheral monitors have not been verified or
validated by SonoSite as being suitable for diagnosis.
To avoid the risk of a burn hazard, do not use the transducer with high
frequency surgical equipment. Such a hazard may occur in the event of a
defect in the high frequency surgical neutral electrode connection.
Do not use the system if it exhibits erratic or inconsistent behavior.
Discontinuities in the scanning sequence are indicative of a hardware failure
that must be corrected before use.
Some transducer sheaths contain natural rubber latex and talc, which can
cause allergic reactions in some individuals. Refer to 21 CFR 801.437, User
labeling for devices that contain natural rubber.
Perform ultrasound procedures prudently. Use the ALARA (as low as
reasonably achievable) principle and follow the prudent use information
concerning MI and TI.
SonoSite does not currently recommend a specific brand of acoustic standoff. If
an acoustic standoff is used, it must have a minimum attenuation
of .3dB/cm/MHz.
Some SonoSite transducers are approved for intraoperative applications if a
market-cleared sheath is used.
To avoid injury and reduce risk of infection to the patient, observe the following:
• Follow Universal Precautions when inserting and maintaining a medical
device for interventional and intraoperative procedures.
• Appropriate training in interventional and intraoperative procedures as
dictated by current relevant medical practices as well as in proper operation
of the ultrasound system and transducer is required. During vascular access,
the potential exists for serious complications including without limitation the
following: pneumothorax, arterial puncture, guidewire misplacement, and
risks normally associated with local or general anesthesia, surgery, and
post-operative recovery.
WARNING:
To avoid device damage or patient injury, do not use the P10x or P21x needle
guide bracket on patients with pacemakers or medical electronic implants. The
needle guide bracket for the P10x and P21x transducers contains a magnet
that is used to ensure the bracket is correctly oriented on the transducer. The
magnetic field in direct proximity to the pacemaker or medical electronic
implant may have an adverse effect.
Hazardous materials
WARNING:
Chapter 3: Safety15
Products and accessories may contain hazardous materials. Ensure that
products and accessories are disposed of in an environmentally responsible
manner and meet federal and local regulations for disposing hazardous
materials.
Electromagnetic compatibility
The ultrasound system has been tested and found to comply with the electromagnetic compatibility (EMC)
limits for medical devices to IEC 60601-1-2:2001. These limits are designed to provide reasonable
protection against harmful interference in a typical medical installation.
WARNING:
Caution:
The Sonosite Edge II ultrasound system should not be used adjacent to or
stacked with other equipment. If such use occurs, verify that the SonoSite Edge
II ultrasound system operates normally in that configuration.
Medical electrical equipment requires special precautions regarding EMC and
must be installed and operated according to these instructions. Portable and
mobile RF communications equipment can affect the ultrasound system.
Electromagnetic interference (EMI) from other equipment or interference
sources could result in performance disruption of the ultrasound system.
Evidence of disruption may include image degradation or distortion, erratic
readings, equipment ceasing to operate, or other incorrect functioning. If this
occurs, survey the site to determine the source of disruption, and take the
following actions to eliminate the source(s).
• Turn equipment in the vicinity off and on to isolate disruptive equipment.
• Relocate or re-orient interfering equipment.
• Increase distance between interfering equipment and your ultrasound
system.
• Manage use of frequencies close to ultrasound system frequencies.
• Remove devices that are highly susceptible to EMI.
• Lower power from internal sources within facility control (such as paging
systems).
• Label devices susceptible to EMI.
• Educate clinical staff to recognize potential EMI-related problems.
• Eliminate or reduce EMI with technical solutions (such as shielding).
• Restrict use of personal communicators (cell phones, computers) in areas
with devices susceptible to EMI.
• Share relevant EMI information with others, particularly when evaluating new
equipment purchases which may generate EMI.
• Purchase medical devices that comply with IEC 60601-1-2 EMC Standards.
Caution:
To avoid the risk of increased electromagnetic emissions or decreased
immunity, use only accessories and peripherals recommended by SonoSite.
Connection of accessories and peripherals not recommended by SonoSite to
the ultrasound system may result in malfunction of the ultrasound system or
other medical electrical devices in the area. Contact SonoSite or your local
representative for a list of accessories and peripherals available from or
recommended by SonoSite. See the SonoSite accessories user guide.
Electrostatic discharge
Caution:
16Chapter 3: Safety
Electrostatic discharge (ESD), or static shock, is a naturally occurring
phenomenon. ESD is common in conditions of low humidity, which can be
caused by heating or air conditioning. ESD is a discharge of the electrical
energy from a charged body to a lesser or non-charged body. The degree of
discharge can be significant enough to cause damage to a transducer or an
ultrasound system. The following precautions can help reduce ESD: anti-static
spray on carpets, anti-static spray on linoleum, and anti-static mats.
Separation distance
PPP
Recommended separation distances between portable and mobile RF communications equipment
and the SonoSite Edge II ultrasound system
The SonoSite Edge II ultrasound system is intended for use in an electromagnetic
environment in which radiated radio frequency (RF) disturbances are controlled. The
customer or the user of the SonoSite Edge II ultrasound system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the SonoSite Edge II ultrasound
system as recommended below, according to the maximum output power of the
communications equipment.
Rated
maximum
output power
of transmitter
Watts
0.010.120.120.23
0.10.380.380.73
11.2 1.2 2.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d=1.2
80 MHz to 800 MHz
d=1.2
800 MHz to 2.5 GHz
d=2.3
Chapter 3: Safety17
Guidance and manufacturer’s declaration
WARNING:
The SonoSite Edge II wireless adapter contains an IEEE 802.11 transmitter that utilizes the ISM frequency
band from 2.412 to 2.4835 GHz and implements two methods of transmission:
•IEEE 802.11b with Complementary Code Keying (CCK), Differential Quaternary Phase Shift Keying
(DQPSK), and Differential Binary Phase Shift Keying (DBPSK) at 16 dB
•IEEE 802.11g with Orthogonal Frequency Division Multiplexing (OFDM) at 13 dBm
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The SonoSite Edge II ultrasound system is intended for use in the electromagnetic
environment specified below. The customer or the user of the SonoSite Edge II ultrasound
system should assure that it is used in such an environment.
Other equipment, even equipment that complies with CISPR emission
requirements, can interfere with the SonoSite Edge II ultrasound system.
Group 1The SonoSite Edge II ultrasound system uses RF
energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
ClSPR 11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The SonoSite Edge II ultrasound system is intended for use in the electromagnetic
environment specified below. The customer or the user of the SonoSite Edge II ultrasound
system should assure that it is used in such an environment.
Immunity
Test
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
IEC 60601 Test
Level
±6.0KV contact
±8.0KV air
Class AThe SonoSite Edge II ultrasound system is
suitable for use in all establishments other than
domestic and those directly connected to the
Class A
Complies
public low-voltage power supply network which
supplies buildings used for domestic purposes.±
Compliance
Level
±6.0KV contact
±8.0KV air
Electromagnetic
Environment
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
18Chapter 3: Safety
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (Continued)
P
The SonoSite Edge II ultrasound system is intended for use in the electromagnetic
environment specified below. The customer or the user of the SonoSite Edge II ultrasound
system should assure that it is used in such an environment.
Immunity
Test
Electrical fast
Transient burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
IEC 60601 Test
Level
±2KV for power
supply lines
±1KV for
input/output lines
±1KV line(s) to
line(s)
±2KV line(s) to earth
>5% U
T
(>95% dip in UT) for
0.5 cycle
40% U
T
(60% dip in UT) for 5
cycles
70% U
T
(30% dip in UT) for
25 cycles
>5% U
T
(>95% dip in UT) for
5s
Compliance
Level
±2KV for power
supply lines
±1KV for
input/output lines
±1KV line(s) to
line(s)
±2KV line(s) to
earth
>5% U
T
(>95% dip in UT)
for 0.5 cycle
40% U
T
(60% dip in UT)
for 5 cycles
70% U
T
(30% dip in UT)
for 25 cycles
>5% U
T
(>95% dip in UT)
for 5s
Electromagnetic
Environment
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the SonoSite Edge II
ultrasound system requires
continued operation during
power mains interruptions, it
is recommended that the
SonoSite Edge II ultrasound
system be powered from an
uninterruptible power supply
or a battery.
Power
Frequency
Magnetic Field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
3 A/m3 A/mPower frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
3Vrms
150 kHz to 80 MHz
3 VrmsPortable and mobile RF
communications equipment
should be used no closer to
any part of the SonoSite
Edge II ultrasound system
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance
d = 1.2
Chapter 3: Safety19
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity (Continued)
P
P
The SonoSite Edge II ultrasound system is intended for use in the electromagnetic
environment specified below. The customer or the user of the SonoSite Edge II ultrasound
system should assure that it is used in such an environment.
Immunity
Test
Radiated RF
IEC 61000-4-3
Radiated RF
IEC 61000-4-3
(continued)
IEC 60601 Test
Level
3Vim
80 MHz to 2.5 GHz
Compliance
Level
3 V/m
Electromagnetic
Environment
d = 1.2
80 MHz to 800 MHz
d = 2.3
800 MHz to 2,5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic Site
a
survey
, should be less than
the compliance level in each
b
frequency range
.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Note:UT is the AC mains voltage prior to application of the test level.
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the SonoSite ultrasound system is
used exceeds the applicable RF compliance level above, the SonoSite ultrasound system should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the SonoSite ultrasound system.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
FCC Caution: Changes or modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
•This device may not cause harmful interference.
•This device must accept any interference received, including interference that may cause undesired
operation.
20Chapter 3: Safety
Immunity testing requirements
The SonoSite Edge II ultrasound system complies with the essential performance requirements specified in
IEC 60601-1-2 and IEC 60601-2-37. Results of immunity testing show that the SonoSite Edge II ultrasound
system meets these requirements and is free from the following:
•Noise on a waveform or artifacts or distortion in an image or error of a displayed numerical value that
cannot be attributed to a physiological effect and that may alter the diagnosis
•Display of incorrect numerical values associated with the diagnosis to be performed
•Display of incorrect safety related indications
•Production of unintended or excessive ultrasound output
•Production of unintended or excessive transducer assembly surface temperature
•Production of unintended or uncontrolled motion of transducer assemblies intended for intra-corporeal
use
Standards
Electrical safety standards
AAMI ES60601-1:2005 + C1(2009) + A2(2010) Medical electrical equipment, Part 1: General requirements
for basic safety and essential performance (3
rd
edition plus Corrigendum 1 and Amendment A2)
CSA C22.2 No. 60601-1-08 + TC 2(2011) Medical electrical equipment – Part 1: General requirements for
basic safety and essential performance (3
IEC 60601-1:2005 + C1(2006) + C2(2007) Medical electrical equipment, Part 1: General requirements for
basic safety and essential performance (3
CSA C22.2 60601-2-37:08 Medical electrical equipment - Part 2-37: Particular requirements for the basic
safety and essential performance of ultrasound medical diagnostic and monitoring equipment
IEC 60601-2-37 (ed .2.0) Medical electrical equipment - Part 2-37: Particular requirements for the basic
safety and essential performance of ultrasound medical diagnostic and monitoring equipment
CSA C22.2 60601-6-07 Medical Electrical Equipment part 1-6: General requirements for basic safety and
essential performance – Collateral Standard: Usability
IEC 60601-1-6:2010 Medical Electrical Equipment part 1-6: General requirements for basic safety and
essential performance – Collateral Standard: Usability.
rd
edition plus Corrigendum 2)
rd
edition plus Corrigendum 1 and Corrigendum 2)
EMC standards classification
CISPR 11:2009, Industrial, Scientific, and Medical (ISM) Radio-Frequency Equipment Electromagnetic
Disturbance Characteristics—Limits and Methods of Measurement.
IEC 60601-1-2:2007, Medical Electrical Equipment—General Requirements for Basic Safety and Essential
Performance-Collateral Standard. Electromagnetic Compatibility Requirements and Tests.
The Classification for the ultrasound system, docking system, accessories, and peripherals when
configured together: Group 1, Class A.
Acoustic standards
NEMA UD 2-2004, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.
NEMA UD 3-2004, Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on
Diagnostic Ultrasound Equipment, American Institute of Ultrasound in Medicine.
Biocompatibility standards
AAMI/ANSI/ISO 10993-1, Biological evaluation of medical devices—Part 1: Evaluation and testing (2009).
Chapter 3: Safety21
Loading...
+ 67 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.