Fisher Phero User manual

Page 1
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0
Instruction Manual
PHEROsub E Series
Horizontal Agarose Gel
Electrophoresis Systems
Catalogue Numbers
111100-001 111100-002 111100-003
PHERO-sub 0710-E 111210-001 PHERO-sub 1008-E
111300-001 111300-002 111300-003
PHERO-sub 1010-E 111400-001
111400-002 111400-003 111400-004
PHERO-sub 1515-E
111500-001 111500-002 111500-003
PHERO-sub 2020-E 111600-001
111600-002 111600-003 111600-004
PHERO-sub 2632-E
1
Page 2
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0
Contents
SAFETY PRECAUTION................................................................................................. 3
PACKING LISTS:........................................................................................................... 4
PHERO-sub 0710-E.................................................................................................. 4
PHERO-sub 1010-E.................................................................................................. 4
PHERO-sub 1515-E.................................................................................................. 5
PHERO-sub 2020-E.................................................................................................. 5
PHERO-sub 2632-E.................................................................................................. 6
PHERO-sub 1008-E.................................................................................................. 6
Usage Guidance and restrictions:.................................................................................. 7
Care and Maintenance:-................................................................................................ 8
Setting up the Horizontal Gel Tanks:-............................................................................ 9
Gel Preparation:-.......................................................................................................... 10
Gel Pouring:-................................................................................................................ 11
Using trays with Casting Dams (Note not PHERO-sub 2632-E):-............................11
Using The Flexicaster (Note not PHERO-sub 2632-E) :-......................................... 11
Using Traditional tape method:-............................................................................... 12
Casting in the Mini Fast Unit:-.................................................................................. 13
Running the Gel:-......................................................................................................... 14
Gel Staining and Viewing:-........................................................................................... 15
References................................................................................................................... 15
Combs – MC Denotes Multi Channel Pipette compatible............................................ 17
Warranty...................................................................................................................... 20
2
Page 3
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

SAFETY PRECAUTION

WHEN USED CORRECTLY, THESE UNITS POSE NO HEALTH RISK.
HOWEVER, THESE UNITS CAN DELIVER DANGEROUS LEVELS OF
ELECTRICITY AND ARE TO BE OPERATED ONLY BY QUALIFIED PERSONNEL
FOLLOWING THE GUIDELINES LAID OUT IN THIS INSTRUCTION MANUAL.
ANYONE INTENDING TO USE THIS EQUIPMENT SHOULD READ THE COMPLETE
MANUAL THOROUGHLY.
THE UNIT MUST NEVER BE USED WITHOUT THE SAFETY LID CORRECTLY IN
POSITION.
THE UNIT SHOULD NOT BE USED IF THERE IS ANY SIGN OF DAMAGE TO THE
EXTERNAL TANK OR LID.
THESE UNITS COMPLY WITH THE STATUTORY CE SAFETY DIRECTIVES:
73/23/EEC: LOW VOLTAGE DIRECTIVE: IEC 1010-1:1990 plus AMENDMENT
1:1992
EN 61010-1:1993/BS EN 61010-1:1993
3
Page 4
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

PACKING LISTS:

PHERO-sub 0710-E

111100-001, 111100-002, 111100-003
Units include tank, lid and electrodes and include the following accessories:-
Tray Tray Dams Combs Loading Guides Cables
111100-001
111130-001 7 x 7cm (W x L)
111140-002 Pack of 2
111121-004 1mm, 8 sample
111140-003 – Strips 111140-004 – Platform
111110-006
111100-002
111131-001 7 x 10cm (W x L)
111140-002 Pack of 2
2 x 111121-004 1mm, 8 sample
111140-003 – Strips 111130-004 – Platform
111110-006
111100-003
111130-001 111131-001
111140-002 Pack of 2
2 x 111121-004 1mm, 8 sample
111140-003 – Strips 111140-004 – Platform
111110-006

PHERO-sub 1010-E

111130-001, 111130-002, 111130-003
Units include tank, lid and electrodes and include the following accessories:-
Tray Tray Dams Combs Loading Guides Cables
111130-001
111330-001 10 x 7cm (W x L)
111340-001 Pack of 2
2 x 111321-007 1mm, 16 sample
111340-003 – Strips 111340-004 – Platform
111310-006
111130-002
111331-001 10 x 10cm (W x L)
111340-001 Pack of 2
2 x 111321-007 1mm, 16 sample
111340-003 – Strips 111340-004 – Platform
111310-006
111130-003
111330-001 111331-001
111340-001 Pack of 2
2 x 111321-007 1mm, 16 sample
111340-003 – Strips 111340-004 – Platform
111310-006
4
Page 5
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

PHERO-sub 1515-E

111400-001, 111400-002, 111400-003, 111400-004
Units include tank, lid and electrodes and include the following accessories:-
Tray Tray Dams Combs Loading Guides Cables
111400-001
111430-001 15 x 7cm (W x L)
111440-002 Pack of 2
2 x 111421-007 1mm, 20 sample
111440-003 – Strips 111440-004 – Platform
111410-006
111400-002
111431-001 15 x 10cm (W x L)
111440-002 Pack of 2
2 x 111421-007 1mm, 20 sample
111440-003 – Strips 111440-004 – Platform
111410-006
111400-003
111432-001 15 x 15cm (W x L)
111440-002 Pack of 2
2 x 111421-007 1mm, 20 sample
111440-003 – Strips 111440-004 – Platform
111410-006
111400-004
111430-001 111431-001 111432-001
111440-002 Pack of 2
2 x 111421-007 1mm, 20 sample
111440-003 – Strips 111440-004 – Platform
111410-006

PHERO-sub 2020-E

111500-001, 111500-002, 111500-003
Units include tank, lid and electrodes and include the following accessories:-
Tray Tray Dams Combs Loading Guides Cables
111500-001
111530-001 20 x 10cm (W x L)
111540-002 Pack of 2
2 x 111521-006 1mm, 20 sample
111540-003 – Strips 111540-004 – Platform
111510-006
111500-002
111531-001 20 x 20cm (W x L)
111540-002 Pack of 2
2 x 111521-006 1mm, 20 sample
111540-003 – Strips 111540-004 – Platform
111510-006
111500-003
111530-001 111531-001
111540-002 Pack of 2
2 x 111521-006 1mm, 20 sample
111540-003 – Strips 111540-004 – Platform
111510-006
5
Page 6
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

PHERO-sub 2632-E

111600-001, 111600-002, 111600-003, 111600-004
Units include tank, lid and electrodes and include the following accessories:-
Tray Combs Loading Guides Cables
111600-001
111630-001 26 x 16cm (W x L)
6 x 111621-001 1mm, 28 sample
111640-003 – Strips 111640-004 – Platform
111610-006
111600-002
111631-001 26 x 24cm (W x L)
6 x 111621-001 1mm, 28 sample
111640-003 – Strips 111640-004 – Platform
111610-006
111600-003
111632-001 26 x 32cm (W x L)
6 x 111621-001 1mm, 28 sample
111640-003 – Strips 111640-004 – Platform
111610-006
111600-004
111630-001 111631-001 111632-001
6 x 111621-001 1mm, 28 sample
111640-003 – Strips 111640-004 – Platform
111610-006

PHERO-sub 1008-E

Unit includes tank, lid and electrodes and the following accessories:-
Casting Gates Combs Cables
111200-001
111240-002 Pack of 2
2 x 111221-003
1.5mm, 8 sample
111210-006
The packing lists should be referred to as soon as the units are received to
ensure that all components have been included. The unit should be checked for
damage when received.
Please contact your supplier if there are any problems or missing items.
6
Page 7
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Usage Guidance and restrictions:

• Maximum altitude 2,000m.
• Temperature range between 4°C and 65°C.
• Maximum relative humidity 80% for temperatures up to 31OC decreasing linearly to
50% relative humidity at 40OC.
• Not for outdoor Use.
This apparatus is rated POLLUTION DEGREE 2 in accordance with IEC 664.
POLLUTION DEGREE 2, states that: “Normally only non-conductive pollution occurs.
Occasionally, however, a temporary conductivity caused by condensation must be
expected”.
7
Page 8
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Care and Maintenance:-

Cleaning Horizontal Units
Units are best cleaned using warm water and a mild detergent. Water at
temperatures above 600 C can cause damage to the unit and components.
The tank should be thoroughly rinsed with warm water or distilled water to prevent
build up of salts but care should be taken not to damage the enclosed electrode and
vigorous cleaning is not necessary or advised.
Air drying is preferably before use.
The units should only be cleaned with the following:-
Warm water with a mild concentration of soap or other mild detergent.
Compatible detergents include dishwashing liquid, Hexane and Aliphatic hydrocarbons
The units should not be left to in detergents for more than 30 minutes.
The units should never come into contact with the following cleaning agents,
these will cause irreversible and accumulative damage:-
Acetone, Phenol, Chloroform, Carbon tetrachloride, Methanol, Ethanol, Isopropyl
alcohol
Alkalis.
Rnase Decontamination
This can be performed using the following protocol:-
Clean the units with a mild detergent as described above.
Wash with 3% hydrogen peroxide (H2O2) for 10 minutes.
Rinsed with 0.1% DEPC- (diethyl pyrocarbonate) treated distilled water,
Caution: DEPC is a suspected carcinogen. Always take the necessary precautions
when using.
RNaseZAP™ (Ambion) can also be used. Please consult the instructions for use with
acrylic gel tanks.
8
Page 9
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Setting up the Horizontal Gel Tanks:-

Instructions for fitting Electrode Cables.
1. Note the position of the lid on the unit. This shows the correct polarity and the
correct orientation of the cables, black is negative and red positive.
2. Remove the lid from the unit. Note if the lid is not removed, fitting the cables
may result in un-tightening of the gold plug and damage to the electrode.
3. Screw the cables into the tapped holes as fully as possible so that there is no
gap between the lid and the leading edge of the cable fitting.
4. Refit the lid.
Instructions for fitting Loading Guides (Not FMMS10).
These can be fitted to enhance visibility of the wells if desired. They can be
fitted to the white vinyl platform sheet or to the unit itself.
1. Seat the tray in the unit and note the position of the comb grooves. The
samples run black to red but the trays can be used frontward or backwards so
ensure that the comb grooves closest to the black electrode are marked.
2. Remove the tray.
3. Peel the back off of the loading guide and carefully apply the loading guide
directly to the gel platform.
The unit is now ready to be used.
9
Page 10
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Gel Preparation:-

1. Table 1 below shows the volume of agarose solution required to make the desired
agarose gel for each unit tray size. For a standard 0.7% agarose gel, add 0.7
grammes of agarose to 100 ml of 1x TAE or TBE solution. The same 1 x solution
should be used in the tank buffer solution.
Table 1.
PHERO-sub 0710-E PHERO-sub 1010-E PHERO-sub 1515-E PHERO-sub 2020-E Tray Gel
volume for
a 0.5mm
thick gel
Tray Gel
volume for
a 0.5mm
thick gel
Tray Gel
volume for
a 0.5mm
thick gel
Tray Gel
volume for
a 0.5mm
thick gel
7 x 7cm 25ml 10 x 7cm 35ml 15 x 7cm 52.5ml 20 x 10cm 100ml 7 x10cm 35ml 10 x 10cm 50ml 15 x 10cm 75ml 20 x 20cm 200ml
15 x 15cm 112.5ml
PHERO-sub 2632-E PHERO-sub 1008-E Tray Gel volume for
a 0.5mm thick
gel
Gel volume for a
0.5mm thick gel
26 x 16cm 208ml 40ml 26 x 24cm 312ml 26 x 32cm 416ml
2. Add the agarose powder to a conical flask.
3. Add the appropriate amount of 1x TAE or TBE solution from the table above. To
prevent evaporation during the dissolving steps below, the conical flask should be
covered with parafilm.
4. Dissolve the agarose powder by heating the agarose either on a magnetic hot plate
with stirring bar or in a microwave oven. If using the microwave method, the
microwave should be set at around a 400 watt or medium setting and the flask swirled
every minute. The solution should be heated until all crystals are dissolved. This is
best viewed against a light background. Crystals appear as translucent crystals. These
will interfere with sample migration if not completely dissolved.
The gel must be cooled to between 50°C and 60°C degrees before pouring.
10
Page 11
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Gel Pouring:-

The multiSUB range of units allow three different methods of gel casting:-
Casting Dams
Flexicaster
Traditional Tape

Using trays with Casting Dams (Note not PHERO-sub 2632-E):-

1. Fit the casting dams over each end of the tray and place onto a level surface.
The dams should be fitted so that there is no gap between the sides of the tray
and the groove in the dams. This will ensure that there is no possibility of gel
leakage.
2. Place the comb(s) in the grooves. Each tray has more than one comb grove so
that multiple combs can be used. Using multiple combs increases sample
number available per gel but decreases run length and care must be taken to
ensure that samples from the first wells do not migrate into the lanes of the
second comb wells.
3. Pour in the agarose carefully so as not to generate bubbles. Any bubbles that
do occur can be smoothed to the edge of the gel and dispersed using a pipette
tip.
4. Allow the agarose to set, ensuring that the gel remains undisturbed.
5. Carefully remove the gel casting gates and comb and transfer the gel including
tray to the main tank.

Using The Flexicaster (Note not PHERO-sub 2632-E) :-

1. Level the Flexicaster base by adjusting the feet so that the bubble is exactly
central.
2. Insert the desired length tray into the Flexicaster such that one end of the tray is
pushed up and seals against the silicone mat of the permanent end of the
Flexicaster.
3. Position the movable end of the Flexicaster so that the silicone mat is pushed
against the other end of the tray.
11
Page 12
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0
4. Turn the cam so that the silicone mat tightly seals against the side of the tray.
Pour in the agarose carefully so as not to generate bubbles. Any bubbles that
do occur can be smoothed to the edge of the gel and dispersed using a pipette
tip.
5. Allow the agarose to set, ensuring that the gel remains undisturbed.
6. Carefully remove the gel casting gates and comb and transfer the gel including
tray to the main tank.

Using Traditional tape method:-

1. Autoclave or plastic backed general tape should be used. A length 5cm longer
than the width of each end of the tray should be cut. One length should be
placed over one end of the tray and stuck m1cm in from the tray edge. This
should then be folded and the edges sealed securely. Repeat for the other end
and place onto a level surface for gel pouring.
2. Place the comb(s) in the grooves. Each tray has more than one comb grove so
that multiple combs can be used. Using multiple combs increases sample
number available per gel but decreases run length and care must be taken to
ensure that samples from the first wells do not migrate into the lanes of the
second comb wells.
3. Pour in the agarose carefully so as not to generate bubbles. Any bubbles that
do occur can be smoothed to the edge of the gel and dispersed using a pipette
tip.
4. Allow the agarose to set, ensuring that the gel remains undisturbed.
5. Carefully remove the gel casting gates and comb and transfer the gel including
tray to the main tank.
12
Page 13
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Casting in the Mini Fast Unit:-

1. Fit the casting gates into the grooves in the unit which are 10mm from the
platinum wire. Ensure that these are seated as deeply as possible. This will
ensure that a good seal is formed and that there is no possibility of gel leakage.
A small amount of Vaseline or sealing grease on the bottom and side edges of
the gates will give further leakage protection.
2. Place the comb(s) in the grooves. Each tray has more than one comb grove so
that multiple combs can be used. Using multiple combs increases sample
number available per gel but decreases run length and care must be taken to
ensure that samples from the first wells do not migrate into the lanes of the
second comb wells.
3. Pour in the agarose carefully so as not to generate bubbles. Any bubbles that
do occur can be smoothed to the edge of the gel and dispersed using a pipette
tip.
4. Allow the agarose to set, ensuring that the gel remains undisturbed.
5. Carefully remove the gel casting gates and comb.
13
Page 14
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Running the Gel:-

1. Mix the sample to be loaded with sample buffer – see solutions for common sample
buffers. Usually 3ul of sample buffer is adequate but less may be used with sample
volumes of less than 10ul.
2. Fill the unit with buffer until the gel is just flooded with buffer. This will give the
fastest resolution times. For enhanced quality of resolution of sample, fill the unit to
5mm above the gel.
3. Load the samples into the wells using pipettes. Multi-channel pipettes can be used
for
loading samples with MC compatible combs, see listing in accessories for identification
of these.
4. Carefully place the lid on the tank and connect to a power supply.
5. Typically gels are run at between 90 and 150 volts. However, maximum voltages
are indicated on the serial badge of each unit. It should be noted that higher voltages
generally give faster but poorer quality sample resolution.
14
Page 15
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Gel Staining and Viewing:-

The Multi Sub trays and the Mini Fast unit allow staining to be performed without
removing the gel from the tray (unit for Fast Mini) if this is preferred.
1. Transfer the gel to a vessel containing the appropriate volume of 0.5 µg/ml ethidium
bromide stain for 15–30 minutes, see solutions for stock stain concentration and adjust
to the volume used accordingly. The entire gel should be covered.
NOTE:- Ethidium bromide is a suspected carcinogen and the necessary safety
precautions should be undertaken.
2. De-stain the gel for 10–30 minutes in distilled water again ensuring the gel is
completely immersed.
3. Rinse the gel twice for a couple of seconds with distilled water.
4. Transfer the gel to a UV Transilluminator.
5. The samples will often appear as brighter, clearer bands when photographed or
viewed using a gel documentation system. However if the gel bands are too faint
then the staining procedure should be adjusted so that there is less de-staining. If
there is too much background then the staining procedure should be adjusted so
that there is more de-staining.

References

1. Sambrook, Fritsch, and Maniatis, Molecular Cloning A Laboratory Manual,
Second Edition,
Cold Spring Harbor Laboratory Press, 1989.
2. Current Protocols in Molecular Biology, Greene Publishing Associates and
Wiley-Interscience,1989.
15
Page 16
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0
Solutions:-
1x TAE 40 mM tris (pH 7.6), 20 mM acetic acid, 1 mM EDTA.
50x (1L) dissolve in 750 ml distilled water: 242 g tris base (FW = 121)
57.1 ml glacial acetic acid 100 ml 0.5 M EDTA (pH 8.0). Fill to 1 litre with distilled water.
1x TBE 89 mM tris (pH 7.6), 89 mM boric acid, 2 mM EDTA 10x (1L) dissolve in 750 ml distilled water: 108 g tris base (FW = 121) 55 g boric acid (FW = 61.8) 40 ml 0.5 M EDTA (pH 8.0) Fill to 1 litre with distilled water.
Sample Loading Dye
10x sample buffer stock consists of 50% glycerol, 0.25% bromophenol blue, and
0.25% xylene cyanole FF in 1x TAE buffer. Only 1–10 ml of the 10x loading dye should be prepared.
Ethidium Bromide Solution
Add 10 mg of Ethidium Bromide to 1 ml distilled water.
16
Page 17
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Combs – MC Denotes Multi Channel Pipette compatible.

PHERO-sub 0710-E PHERO-sub 1010-E PHERO-sub 1515-E PHERO-sub 2020-E
No. of Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number
0.75mm Thick Combs
1 111120-001 1 111320-001 1 111420-001 1 111520-001 2 111120-002 2 111320-002 2 111420-002 2 111520-002 4 111120-003 4 111320-003 4 111420-003 4 111520-003 8 111120-004 8 111320-004 10 111420-004 10 111520-004 8 MC 111120-005 10 MC 111320-005 10 MC 111420-005 16 111520-005 10 111120-006 12 111320-006 12 111420-006 20 MC 111520-006 12 111120-007 16 111320-007 14 MC 111420-007 25 111520-007 16 111120-008 20 MC 111320-008 16 111420-008 30 111520-008
18 MC 111420-009 36 111520-009 20 111420-010 40 MC 111520-010 28 MC 111420-011 30 111420-012
PHERO-sub 0710-E PHERO-sub 1010-E PHERO-sub 1515-E PHERO-sub 2020-E
No. of Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number
1mm Thick Combs
1 111121-001 1 111321-001 1 111421-001 1 111521-001 2 111121-002 2 111321-002 2 111421-002 2 111521-002 4 111121-003 4 111321-003 4 111421-003 4 111521-003 8 111121-004 8 111321-004 10 111421-004 10 111521-004 8 MC 111121-005 10 MC 111321-005 10 MC 111421-005 16 111521-005 10 111121-006 12 111321-006 12 111421-006 20 MC 111521-006 12 111121-007 16 111321-007 14 MC 111421-007 25 111521-007 16 111121-008 20 MC 111321-008 16 111421-008 30 111521-008
18 MC 111421-009 36 111521-009 20 111421-010 40 MC 111521-010 28 MC 111421-011 30 111421-012
17
Page 18
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0
PHERO-sub 0710-E PHERO-sub 1010-E PHERO-sub 1515-E PHERO-sub 2020-E
No. of Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number
1.5mm Thick Combs
1 111122-001 1 111322-001 1 111422-001 1 111522-001 2 111122-002 2 111322-002 2 111422-002 2 111522-002 4 111122-003 4 111322-003 4 111422-003 4 111522-003 8 111122-004 8 111322-004 10 111422-004 10 111522-004 8 MC 111122-005 10 MC 111322-005 10 MC 111422-005 16 111522-005 10 111122-006 12 111322-006 12 111422-006 20 MC 111522-006 12 111122-007 16 111322-007 14 MC 111422-007 25 111522-007 16 111122-008 20 MC 111322-008 16 111422-008 30 111522-008
28 MC 111422-009 36 111522-009 20 111422-010 40 MC 111522-010 28 111422-011 30 111422-012
PHERO-sub 0710-E PHERO-sub 1010-E PHERO-sub 1515-E PHERO-sub 2020-E
No. of Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number No. of
Samples
Part Number
2mm Thick Combs
1 111123-001 1 111323-001 1 111423-001 1 111523-001 2 111123-002 2 111323-002 2 111423-002 2 111523-002 4 111123-003 4 111323-003 4 111423-003 4 111523-003 8 111123-004 8 111323-004 10 111423-004 10 111523-004 8 MC 111123-005 10 MC 111323-005 10 MC 111423-005 16 111523-005 10 111123-006 12 111323-006 12 111423-006 20 MC 111523-006 12 111123-007 16 111323-007 14 MC 111423-007 25 111523-007 16 111123-008 20 MC 111323-008 16 111423-008 30 111523-008
18 MC 111423-009 36 111523-009 20 111423-010 40 MC 111523-010 28 MC 111423-011 30 111423-012
18
Page 19
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0
PHERO-sub 2632-E Combs:
Description Part Number
Comb 28 sample MC, 0.75mm thick 111620-001 Comb 56 sample MC, 0.75mm thick 111620-002 Comb 28 sample MC, 1mm thick 111621-001 Comb 56 sample MC, 1mm thick 111621-002 Comb 28 sample MC, 1.5mm thick 111622-001 Comb 56 sample MC, 1.5mm thick 111622-002 Comb 28 sample MC, 2mm thick 111623-001 Comb 56 sample MC, 2mm thick 111623-002
PHERO-sub 1008-E Combs:-
Description Part Number
Comb 1 sample , 1 dye , 1mm thick 111220-001 Comb 4 sample, 1mm thick 111220-002 Comb 8 sample, 1mm thick 111220-003 Comb 12 sample, 1mm thick 111220-004 Comb 16 sample, 1mm thick 111220-005 Comb 20 sample MC, 1mm thick 111220-006 Comb 1 sample , 1 dye , 1.5mm thick 111221-001 Comb 4 sample, 1.5mm thick 111221-002 Comb 8 sample, 1.5mm thick 111221-003 Comb 12 sample, 1.5mm thick 111221-004 Comb 16 sample, 1.5mm thick 111221-005 Comb 20 sample MC, 1.5mm thick 111221-006
19
Page 20
Manual
PHERO-sub E series
BIOTEC-FISCHER GmbH Daimlerstrasse 6 35447 Reiskirchen / Germany Fon: +49-6408-6072 Fax: +49-6408-64165 e-mail: info@biotec-fischer.de Revision: 1.0

Warranty

The multiSUB Horizontal Electrophoresis units have a warranty against manufacturing
and material faults of twelve months from date of customer receipt.
If any defects occur during this warranty period, your supplier will repair or replace the
defective parts free of charge.
This warranty does not cover defects occurring by accident or misuse or defects
caused by improper operation.
Units where repair or modification has been performed by anyone other than your
supplier or an appointed distributor or representative are no longer under warranty
from the time the unit was modified.
Units which have accessories or repaired parts not supplied by your supplier or it’s
associated distributors have invalidated warranty.
Your supplier cannot repair or replace free of charge units where improper solutions or
chemicals have been used. For a list of these please see the Care and Maintenance
subsection.
If a problem does occur then please contact your supplier.
20