Eppendorf Thermoblock for slides DC User Manual

Page 1
Exchangeable Thermoblock
for slides DC
Operating instructions
Page 2
Table of contents
1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5 Cleaning / Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
6Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
No part of this publication may be reproduced without the prior permission of the copyright owner.
Copyright
©
2005 Eppendorf AG, Hamburg
2
Page 3
1 General / 2 Safety instructions
1 General
Only experienced laboratory personnel are permitted to use the slides thermo­block. The exchangeable Thermoblock for slides DC is especially designed for processing either DNA or protein microarrays using the Eppendorf hybridization frames in research laboratories in the fields of bioscience, medicine and chemistry. The Thermoblock for slides DC can be used on both the Thermomixer comfort or the ThermoStat plus. It can be used for rotational movement up to a maximum of 1,400 rpm and at a maximum temperature of 70 °C.
2 Safety instructions
– When exchanging a heated block, please keep in mind that contact with the
surface of the opened exchangeable thermoblock, its underside or the thermal base on the basic device can result in burns! Exchange the thermoblock once it has cooled down.
– Make certain that the slides are inserted precisely into the recesses. The
Thermoblock for slides DC is compatible with slides with a maximum length of
76.70 mm. – Make sure that the block is clean (no glass particles) before placing slides. – Make sure that the pressure plate is correctly fixed. See mounting part of this
manual.
– The lid of the Thermoblock for slides DC must be tightly closed (lid parallel to
block) prior to using the mix function.
– Be sure to observe the maximum rotational speed of 1,400 rpm and the
maximum temperature setting of 70 °C. The ThermoStat plus basic device has no block recognition function. There is no automatic limitation of the
temperature control. – Do not use damaged thermoblocks. – Exchangeable thermoblocks may only be used in conjunction with the
ThermoStat plus and Thermomixer comfort devices. See their respective
operating manuals for more information. – This operating manual must also be handed over if you give the slides
thermoblock to someone else.
3
Page 4
3 Mounting
1. To mount the exchangeable Thermoblock for slides DC, open the lid by
simultaneously pressing the two pressure points on the fr ont edge of the lid. The pressure points are labeled with "push".
Figure 1: Drawing of the Thermoblock for slides DC
1 Pressure plate 2 Screws
3
3 Lid seal 4 Label
2 1
4
2. Alternately unscrew the 2 screws of the pressure plate until the pressure is
completely released (Figure 1).
3. Adjust the exchangeable Thermoblock for slides DC on the basic device in
such a way that the label can be seen from the front. Screw it tight using a Hexagon key.
4
Page 5
3 Mounting
4. Place the slides (up to 4) in the Thermoblock for slides DC. The slides have
to be inserted with the sticker facing the front part of the thermoblock (Figure 2).
Figure 2: Drawing of the opened Thermoblock for slides DC
1 Slide 2 Pressure plate location ridges 3 Holes to remove the slide 4 Front part of the thermoblock
2 1
3
4
2
5
Page 6
s
3 Mounting
5. Position the pressure plate on top of the Thermoblock for slides DC and
push it firmly into the location ridges. Hold both plate and thermoblock tightly and alternately fasten both screws
completely using the screwing tool (Figure 3).
Screwing tool
Pressure plate location ridge
Pressure plate location ridge
Figure 3: Schematic drawing of the Thermoblock for slides DC assembly and the
crewing tool.
6
Pressure plate
Thermoblock for slides DC
assembly
Page 7
3 Mounting / 4 Application
6. Close the lid and push down the 2 pressure points on the front edges of the
lid. It is absolutely necessary to ascertain that the lid has been correctly closed prior to starting the device.
In the case of the Thermomixer comfort, block identification occurs after the exchangeable Thermoblock for slides DC has been correctly positioned. This results in an automatic limitation of maximum possible rotational speed (1400 rpm) and temperature setting (70 °C).
In the case of the ThermoStat plus there is no block identification function for technical reasons. Please set for max. temperature of 70 °C.
7. After the incubation step, open the Thermoblock for slides DC lid and
unscrew the pressure plate (caution: hot plate) by alternating between both screws.
4 Application
An Eppendorf hybridization frame must be positioned above the microarray on the slide surface in order to carry out the reaction. In the case of using DualChip™ microarrays from Eppendorf, hybridization frames are already placed on the slides. Four slide positions are provided on the exchangeable Thermoblock for slides DC.
7
Page 8
5 Cleaning / Service
–To clean the Thermoblock for slides DC, switch off the power supply and
unplug the device. – Ascertain that the surface of the opened Thermoblock for slides DC and its
underside as well as the thermal base on the basic device have cooled down
completely to minimize the risk of burns. – Unscrew the block from the basic device. – Clean the exchangeable Thermoblock for slides DC immediately if sample
liquid finds its way into the bores. A mild soap solution or lab cleaner can be
used for cleaning. The device should be protected from organic solutions and
aggressive chemicals. For disinfection purposes, please use alcohol-based
infectants. – The sealant material in the lid can be removed and cleaned or replaced.
5.1 Replacing the snap seal
If sample liquid finds its way into the two snap seals of the block, remove them from the block and wash them separately in a soap solution or in an ultrasonic bath or replace them. The removal of the locking devices from the exchangeable Thermoblock for slides DC is carried out by pressing both of the side catches beneath the block together and simultaneously pressing them out of the block (see Figure 4). Be aware of the direction when reinserting the locking systems into the block: the directional marking on the right side must point forward while that on the left side points to the rear.
Figure 4: Changing the snap seal
1. Press the side catches together simultaneously
2. Push out the lock
3 Directional marking
2.
3
1.
8
Page 9
5 Cleaning / Service
5.2 Replacing the lid
The lid, including the seal, can also be replaced when damaged: To do this, loosen and remove the 2 slotted head screws at the side of the lid with a slotted head screwdriver (N° 2). Replace the lid and refasten it by retightening the 2 slotted head screws (Figure 5).
Figure 5: Replacing the lid
9
Page 10
6 Technical data
Slide positions: max. of 4 positions Slide dimensions: Length 74.5 – 76.7 mm
Width 24.7 – 26.3 mm
Height 0.9 – 1.2 mm Rotational speed: max. 1400 rpm Temperature control: to a maximum of 70 °C possible
– T
Temperature homogeneity: T
T
max max
2 K * (at 70 °C)
min
– T
0,5 K * (at 45 °C)
min
Heating-up speed of the thermoblock: > 5 K / min * Chemical resistance: The use of strong acids, bases or
acetone is not recommended.
This could result in damage
to the surface. Cooling-off period from 70 °C to 10 K above the ambient temperature: 20 min
Data determined with a Thermomixer comfort.
*
10
Page 11
7 Ordering information
Product Description Order no. for
Thermoblock for slides DC 022670786 Lid with seal
Snap seal
Products
Thermomixer comfort / R 115 V 220 V ThermosStat plus (without block) Silverquant
®
adapter
North America
0022670727 0022670743
022670107 022670158 022670204 956015073
Order no. International
5368 000.100 5368 851.002
5368 852.009
5355 000.011
5352 000.010 5368 000.037
11
Page 12
Eppendorf Offices
ASEAN
Eppendorf AG Regional Office in Malaysia Tel. +60 3 8023 2769 Fax +60 3 8023 3720 E-Mail: eppendorf@eppendorf.com.my Internet: www.eppendorf.com.my
AUSTRALIA / NEW ZEALAND
Eppendorf South Pacific Pty. Ltd. Tel. +61 2 9889 5000 Fax +61 2 9889 5111 E-mail: Info@eppendorf.com.au Internet: www.eppendorf.com.au
AUSTRIA
Eppendorf AG c/o Schott Austria Tel. +43 1 29017560 Fax +43 1 290175620 E-Mail: gilch.p@eppendorf.de Internet: www.eppendorf.com
BRAZIL
Eppendorf do Brasil Ltda. Tel. +55 11 30 95 93 44 Fax +55 11 30 95 93 40 E-Mail: eppendorf@eppendorf.com.br Internet: www.eppendorf.com.br
CANADA
Brinkmann Instruments, Ltd. Tel. +1 905 826 5525 Fax +1 905 826 5424 E-Mail: canada@brinkmann.com Internet: www.brinkmann.com
CHINA
Eppendorf AG Tel. +86 21 68760880 Fax +86 21 50815371 E-Mail: market.info@eppendorf.cn Internet: www.eppendorf.cn
FRANCE
EPPENDORF FRANCE S.A.R.L. Tel. +33 1 30 15 67 40 Fax +33 1 30 15 67 45 E-Mail: eppendorf@eppendorf.fr Internet: www.eppendorf.fr
GERMANY
Eppendorf Vertrieb Deutschland GmbH Tel. +49 02232 418-0 Fax +49 02232 418-155 E-Mail: vertrieb@eppendorf.de Internet: www.eppendorf.de
INDIA
Eppendorf India Limited Tel. +91 44 52111314 Fax +91 44 52187405 E-Mail: info@eppendorf.co.in Internet: www.eppendorf.co.in
ITALY
Eppendorf s.r.l. Tel. +390 2 55 404 1 Fax +390 2 58 013 438 E-Mail: eppendorf@eppendorf.it Internet: www.eppendorf.it
Page 13
Eppendorf Offices
JAPAN
Eppendorf Japan Co. Ltd. Tel. +81 3 5825 2363 Fax +81 3 5825 2365 E-Mail: info@eppendorf.jp Internet: www.eppendorf.jp
NORDIC
Eppendorf Nordic Aps Tel. +45 70 22 2970 Fax +45 45 76 7370 E-Mail: nordic@eppendorf.dk Internet: www.eppendorf.dk
SPAIN
Eppendorf Ibérica S.L. Tel. +34 91 651 76 94 Fax +34 91 651 81 44 E-Mail: iberica@eppendorf.de Internet: www.eppendorf.es
SWITZERLAND
Vaudaux-Eppendorf AG Tel. +41 61 482 1414 Fax +41 61 482 1419 E-Mail: vaudaux@vaudaux.ch Internet: www.eppendorf.com
UNITED KINGDOM
Eppendorf UK Limited Tel. +44 1223 200 440 Fax +44 1223 200 441 E-Mail: sales@eppendorf.co.uk Internet: www.eppendorf.co.uk
USA
Eppendorf North America Tel. +1 516 334 7500 Fax +1 516 334 7506 E-Mail: info@eppendorf.com Internet: www.eppendorfna.com
OTHER COUNTRIES
see: www.eppendorf.com/worldwide
Page 14
is a registered trademark.
®
In touch with life
Your local distributor: www.eppendorf.com/worldwide
Eppendorf AG · 22331 Hamburg · Germany · Tel. +49 40 5 38 01-0 · Fax +49 40 5 38 01-556
E-Mail: eppendorf@eppendorf.com
Eppendorf North America, Inc. · One Cantiague Road, P.O. Box 1019
Westbury, N.Y. 11590-0207 USA · Tel. +1 516 3 34 75 00 · Free phone +1 800 6 45 30 50
Fax +1 516 3 34 75 06 · E-Mail: info@eppendorf.com
Application Support
Europe, International: Tel. +49 180 3 66 67 89 · E-Mail: support@eppendorf.com
North America: Tel. +1 800 6 45 30 50 ext. 2258 · E-Mail: support_NA@eppendorf.com
Asia, Pacific: Tel. +60 3 80 23 27 69 · E-Mail: support_Asia@eppendorf.com
eppendorf
B 5368 900.100-01/0805 · Printed in Germany
Loading...