Eppendorf SOP User Manual

rating Procedure for Pipettes
www.eppendorf.com/myeppendorf
P
N)Operating Procedure for Pipettes
Eppendorf SOP
Standard Operating Procedure for Pipettes
Copyright
©
2013 Eppendorf AG, Hamburg. No part of this publication may be
reproduced without the prior permission of the copyright owner.
®
eppendorf Eppendorf Multipette Biomaster
, the Eppendorf logo, Eppendorf Reference®, Eppendorf Research®,
®
, Eppendorf Xplorer®, Varipette®, Varispenser®,
®
, Combitips® and PICASO® are registered trademarks of
Eppendorf AG.
AESOP13640-08/052013
Table of contents
Eppendorf SOP
English (EN)
Table of contents
1 Test and pipette conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1.1 Balances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1.1.1 Balance type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
1.1.2 Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.1.3 Minimum requirements for balances . . . . . . . . . . . . . . . . . .6
1.2 Measuring station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2.1 Evaporation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2.2 Measuring station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2.3 Test room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2.4 Temperature differences. . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2.5 Test liquid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
1.2.6 Operating Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.1 Scope of testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.1.1 Adjusting adjustable volume pipettes . . . . . . . . . . . . . . . . . . 8
2.1.2 Multi-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.1.3 Multipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.1.4 Bottle-top dispenser and Top Buret . . . . . . . . . . . . . . . . . . .9
2.2 Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.1 Work method. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2.2 Removing the test volume from the reservoir. . . . . . . . . . . . 9
2.2.3 Dispensing the test volume into the weighing vessel . . . . . 10
3
3 Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
3.1 Calculating the systematic error . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.2 Calculating the random error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
4 Sterilization and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
4.1 Sterilization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.1.1 Reference and Biomaster . . . . . . . . . . . . . . . . . . . . . . . . . .12
4.1.2 Research plus and Reference 2. . . . . . . . . . . . . . . . . . . . . . 12
4.1.3 Research and Research pro . . . . . . . . . . . . . . . . . . . . . . . .13
4.1.4 Top Buret . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
4.1.5 Varipette and Multipette . . . . . . . . . . . . . . . . . . . . . . . . . . .13
4.1.6 Varispenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
4.1.7 Xplorer and Xplorer plus . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Table of contents
4
Eppendorf SOP English (EN)
5 Leakage check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5.1 Pipette leakage check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
6.1 Potential error causes and corrections . . . . . . . . . . . . . . . . . . . . . . . 15
7 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
7.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
7.2 Adjusting adjustable-volume pipettes. . . . . . . . . . . . . . . . . . . . . . . .17
7.2.1 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
7.2.2 Biomaster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
7.2.3 Reference variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
7.2.4 Reference 2 adjustable-volume. . . . . . . . . . . . . . . . . . . . . .20
7.2.5 Research variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
7.2.6 Research plusadjustable volume - adjustment to
environmental parameters. . . . . . . . . . .. . . . . . . . . . .. . . . . . 25
7.2.7 Research plus adjustable volume - change to the factory
adjustment.. . . . . . . . .. . . . .. . . . . . . . . . .. . . . . . . . . . . . . . . .26
7.2.8 Research Multi-channel . . . . . . . . . . . . . . . . . . . . . . . . . . .31
7.2.9 Research plus Multi-channel . . . . . . . . . . . . . . . . . . . . . . .31
7.2.10 Varispenser plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
7.2.11 Xplorer and Xplorer plus . . . . . . . . . . . . . . . . . . . . . . . . . .32
7.3 Adjusting fixed-volume pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . .33
7.3.1 Reference fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
7.3.2 Reference2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
7.3.3 Research fixed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
7.3.4 Research plus fixed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
7.4 Physical influences of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
8 Factor Z for distilled water . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
8.1 Factor Z overview table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
Table of contents
Eppendorf SOP
English (EN)
9 Technical specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.1 Fixed-volume pipettes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9.1.1 Reference fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9.1.2 Reference 2 fixed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
9.1.3 Research fixed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
9.1.4 Research plus fixed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
9.2 Adjusting adjustable volume pipettes. . . . . . . . . . . . . . . . . . . . . . . .41
9.2.1 Reference variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
9.2.2 Reference 2 adjustable-volume. . . . . . . . . . . . . . . . . . . . . .42
9.2.3 Research adjustable-volume . . . . . . . . . . . . . . . . . . . . . . . .44
9.2.4 Research pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
9.2.5 Research plus adjustable-volume . . . . . . . . . . . . . . . . . . . .46
9.2.6 Biomaster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9.2.7 Varipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
9.2.8 Xplorer and Xplorer plus . . . . . . . . . . . . . . . . . . . . . . . . . .48
9.3 Multi-channel pipettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.3.1 Research . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
9.3.2 Research pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
9.3.3 Research plus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
9.3.4 Reference2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
9.3.5 Xplorer and Xplorer plus . . . . . . . . . . . . . . . . . . . . . . . . . .53
9.4 Multipette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
9.4.1 MultipettePlus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
9.4.2 MultipetteM4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
9.4.3 Multipette stream / Multipette Xstream . . . . . . . . . . . . . . .56
9.5 Varispenser / Top Buret . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9.5.1 Varispenser and Varispenser plus. . . . . . . . . . . . . . . . . . . .57
9.5.2 Top Buret . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57
9.6 Error limits in accordance with EN ISO 8655. . . . . . . . . . . . . . . . . . 57
9.6.1 Air-cushion pipettes, fixed and adjustable volume. . . . . . .58
9.6.2 Positive displacement pipettes (Biomaster) . . . . . . . . . . . .58
9.6.3 Dispenser (Multipette) . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
9.6.4 Single-stroke dispenser (Varispenser) . . . . . . . . . . . . . . . .60
9.6.5 Piston stroke burets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
5
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
Test and pipette conditions
6
Eppendorf SOP English (EN)

1 Test and pipette conditions

These test and pipetting conditions are valid for all Eppendorf pipettes and dispensers. Follow EN ISO 8655 "Volume-measuring equipment with pistons".
For the calibration or adjustment of pipettes, the scales and measuring station should fulfill the following requirements:

1.1 Balances

1.1.1 Balance type

Use semi-microbalances and microbalances to calibrate pipettes.
Some companies offer balances that are specially designed to meet the requirements of pipette calibration, e.g. Sartorius and Mettler.

1.1.2 Accuracy

When selecting the balances, ensure that they are suitable for the accuracy of the pipette. This helps to ensure that deviations within a measuring series are recorded exactly for the assessment of the systematic and random measurement deviations in accordance with EN ISO 8655.
For a pipette volume of less than or equal to 10 µl, balances with a 6-digit display must be used. With larger volumes, balances with a 5-digit display are sufficient.

1.1.3 Minimum requirements for balances

Selected volume* of the pipette to be tested
1 µL to 10 µL 0.001 >10 µL to 100 µL 0.01 >100 µL to 1 000 µL 0.1 >1 mL to 10 mL 0.1 >10 mL to 200 mL 1
* For practical reasons, the nominal volume can be used for the selection of the balances.
Display resolution (mg)
Test and pipette conditions
Eppendorf SOP
English (EN)

1.2 Measuring station

1.2.1 Evaporation

Take evaporation protection into account while carrying out the measurement.
With volumes < 50 µl in particular, errors caused by the evaporation of the test liquid must be taken into consideration.
This can be ensured by using a liquid trap or other equipment to prevent evaporation.

1.2.2 Measuring station

For electronic measurement data processing, we recommend calibration software and accessories for your balance (PICASO, order no. 3113 004.001).
(see also EN ISO 8655, Part 6)

1.2.3 Test room

The tests should be carried out in a draft-free room under constant climatic conditions.
The test room should have a constant temperature between 15 °C and 30 °C and a constant relative humidity above 50 %.
7

1.2.4 Temperature differences

Before the test, the device to be tested and test liquid must have stood in the test room for a sufficient amount of time, at least 2 hours, in order to reach equilibrium with the storage conditions.
Direct sunlight and other influences which could affect the temperature should be avoided at all costs.

1.2.5 Test liquid

Distilled or deionized water of "Quality 3 in accordance with ISO 3696", degassed or at equilibrium with air. The water must be at room temperature.

1.2.6 Operating Manual

Observe the operating instructions for your pipette.
Calibration
8
Eppendorf SOP English (EN)

2 Calibration

2.1 Scope of testing

Note concerning the nominal volume:
The nominal volume of an adjustable-volume pipette is the largest volume to be set by the user and specified by the manufacturer.
When using the Combitip with the Multipette, the nominal volume is the largest possible dispensing volume of the Combitip. This means that for the Multipette plus and Multipette M4, the nominal volume is 1/5 of the filling volume of the Combitip advanced. For the Multipette stream / Xstream, the nominal volume is the filling volume of the Combitip advanced.

2.1.1 Adjusting adjustable volume pipettes

For variable pipettes, check 3 different volumes with 10 measured values each:
• the nominal volume,
• approx. 50 % of the nominal volume,
• 10 % of the nominal volume.

2.1.2 Multi-channel pipettes

For Multi-channel pipettes check each channel separately with 3 different volumes, each with 10 measured values:
• the nominal volume,
• approx. 50 % of the nominal volume,
• 10 % of the nominal volume.

2.1.3 Multipette

With the
• Multipette,
• Multipetteplus,
• Multipette stream,
• MultipetteXstream
• MultipetteM4
check the nominal volume with 10 measured values with the Eppendorf Combitip used.
Calibration
Eppendorf SOP
English (EN)

2.1.4 Bottle-top dispenser and Top Buret

For bottle-top dispensers and the Top Buret, check the nominal volume with 10 measured values.

2.2 Measurement

2.2.1 Work method

1. Place the selected pipette tip on the tip cone of the pipette or on the corresponding Combitip in the Multipette.
2. Implement the following settings:
• Adjustable-volume piston-stroke pipettes: the smallest volume to be tested
• Multipettes: the nominal volume
• Bottle-top dispenser and Top Buret: the nominal volume
3. For all Multipettes with fully drawn up Combitip, always discard the first dispensing step.
4. Fill test liquid up to a height of min. 3 mm in the weighing vessel.
5. For piston-stroke pipettes fill the pipette tip 5x with test liquid and empty it (pre-wet), in order to create a moisture balance in the dead air volume.
6. Replace the single-use tip.
7. Pre-wet the tip 1x.
9

2.2.2 Removing the test volume from the reservoir

1. Hold the pipette vertically.
2. Dip the pipette tip into the test liquid by a few millimeters.
3. Draw in the volume to be tested slowly and evenly. The waiting time of 1 to 3 seconds, for Research 1 to 10 mL 5 seconds must be observed. (the waiting time corresponds to the size of the tip, see operating instructions.)
4. Pull the pipette tip slowly out of the liquid, wiping it on the vessel wall.
Calibration
10
Eppendorf SOP English (EN)

2.2.3 Dispensing the test volume into the weighing vessel

1. Rest the filled tip up against the wall of the weighing vessel at an angle.
2. Dispense the test liquid slowly until the first stop (measuring stroke).
3. Press the control button to the second stop (blow-out) and dispense the remaining liquid in the tip (does not apply for dispensers and burets).
4. Hold down the control button and pull the tip up the vessel wall.
5. Let the control button slide back into position.
6. Determine the weight.
7. Complete all measurements of a measuring series as described and calculate the systematic and random error (see p. 11).
8. In case of adjustable-volume pipettes, determine the measurement with the nominal volume, 50 % and 10 % of the nominal volume. Always begin the test with 10 % of the nominal volume.
Evaluation
Eppendorf SOP
English (EN)

3 Evaluation

3.1 Calculating the systematic error

Mean value of the dispensed volume:
n = Number of measured values
To convert the measured values into volume values, use correction factor Z for the dependency of the test liquid on temperature and air pressure for each individual value (see Factor Z for distilled water on p. 36).
11
Systematic error e
Systematic error e
in micro liter:
s
in percent:
s

3.2 Calculating the random error

Random error as repeat standard s:
Random error as coefficient of variation CV:
The specifications of the tested pipette can be found in the relevant operating instructions or technical specifications (see Technical specifications on p. 37).
Sterilization and cleaning
12
Eppendorf SOP English (EN)

4 Sterilization and cleaning

Further notes on cleaning and sterilization can found in the relevant operating manual.

4.1 Sterilization

Parameter for autoclaving
• 121 °C
• 20 minutes
• 1 bar overpressure

4.1.1 Reference and Biomaster

The Reference pipettes and the Biomaster are fully autoclavable.
1. Prior to autoclaving, unscrew the upper and lower part of the pipette apart by approx. one turn to enable the vapor to enter more easily.
2. After autoclaving, allow the pipette to cool down to room temperature and to dry completely and then screw together.

4.1.2 Research plus and Reference 2

The Research plus pipettes and Reference 2 pipettes are fully autoclavable.
1. You can put the Research plus and Reference 2 into the autoclave as a complete unit or with the lower part removed. Do not disassemble the lower part.
2. For 2.5 mL, 5 mL and 10 mL pipettes: remove the old protection filter. Add a new protection filter and install it after autoclaving. Only autoclave the protection filter once. After autoclaving:
3. Cool the pipette down to room temperature and leave to dry.
4. For 2.5 mL, 5 mL / 10 mL pipettes: the protection filter swells during autoclaving. Slightly squeeze the protection filter when installing it into the cone tip.
Sterilization and cleaning
Eppendorf SOP
English (EN)

4.1.3 Research and Research pro

With the Research and Research pro pipettes, the lower section is autoclavable.
With single-channel models prior to autoclaving
1. Pull off the ejector sleeve with depressed ejector.
2. Unscrew the lower part of the pipette.
With the Multi-channel versions, you can autoclave the entire lower section.
After autoclaving, assemble the parts only after room temperature has been reached and the parts have been dried.

4.1.4 Top Buret

The Top Buret can not be autoclaved.

4.1.5 Varipette and Multipette

The
• Varipette,
• Multipette,
• Multipetteplus,
• Multipette stream
• MultipetteXstream
• MultipetteM4
13
are not autoclavable.

4.1.6 Varispenser

The Varispenser and Varispenser plus are only autoclavable when fully assembled.
With theVarispenser plus
1. Set the discharge valve toggle to the dispensing position (–>).
2. Release the volume setting knob, move to the middle position and leave released.
3. Place the dispenser on a cloth and autoclave, avoiding contact with hot surfaces.
4. Only use the dispenser again when it has cooled down to room temperature.
Sterilization and cleaning
14
Eppendorf SOP English (EN)

4.1.7 Xplorer and Xplorer plus

The lower part of theXplorer and Xplorer plus pipettes can be autoclaved.
1. Pull off the ejector sleeve with depressed ejector.
2. On the lower part of the pipette, push the ring labeled PUSH UP TO RELEASE upward until the lower part is released.
After autoclaving, assemble the parts only after room temperature has been reached and the parts have been dried.

4.2 Cleaning

All areas of the underside of the pipette can be cleaned in a soap solution or isopropanol 60 % provided that no differing instructions are given in the operating manual.
1. Clean the parts in soap solution or isopropanol.
2. Rinse the parts in distilled water.
3. Allow the parts to dry completely and then assemble.
4. Lightly lubricate the piston of the pipette (Eppendorf special grease
Leakage check
Eppendorf SOP
English (EN)

5 Leakage check

5.1 Pipette leakage check

In order to carry out a leakage check on a pipette, follow the procedure outlined below:
1. For adjustable-volume pipettes: set the nominal volume.
2. For volumes <20 µl pre-wet the tip several times.
3. Hold the pipette vertically with a full tip for approx. 30 sec. Do not touch the pipette tip.
4. Observe the meniscus of the liquid on the tip opening. If there is a leak in the pipette, a drip will form on the tip opening.

6 Troubleshooting

6.1 Potential error causes and corrections

Observe the error descriptions in the operating instructions.
Error Cause Elimination
Droplets on the inner wall of the pipette tip.
Pipette is dripping, pipetted volume is incorrect.
Uneven wetting of the plastic wall.
• Tip loose.
• Incorrect pipette tip.
Pipette is leaking because:
•The piston is contaminated.
• The piston is damaged.
• The seal is damaged.
Attach a new pipette tip.
Press the tip on firmly. Use an original Eppendorf
tip.
Clean the piston and
lubricate slightly.
Replace the piston and
piston sealing and lubricate slightly.
Replace the seal.
15
Troubleshooting
16
Eppendorf SOP English (EN)
Error Cause Elimination
Liquid cannot be aspirated correctly or dripping occurs.
• Leakage in the Combitip advanced.
• Combitip advanced has heated up.
Replace the Combitip
advanced with a new Combitip advanced.
Ensure that the
temperature is uniform as the liquid expands when heated.
The control button is jammed, runs jerky.
•The piston is contaminated.
• The seal is contaminated.
Clean the piston and
lubricate slightly.
Disassemble the pipette,
clean all seals, replace if necessary.
• The piston is damaged.
Replace the piston and
piston seal and lubricate slightly.
• Penetration of solvent vapors
Unscrew the lower part of
the pipette and aerate the lower part.
Clean the piston and
lubricate slightly.
Adjustment
Eppendorf SOP
English (EN)

7Adjustment

7.1 General information

Prior to delivery, all pipettes are adjusted with distilled or de-ionized, degassed water under the conditions described in Chapters 1 to 3 and according to ISO 3696 (see Calibration on p. 8).
If you have doubts regarding the accuracy of the volume entered, please check the following points first:
• Is there a leak in the device (see Leakage check on p. 15)? Exception: Biomaster
• Does the temperature of the pipetted liquid correspond to: – the temperature of the device? – the ambient air temperature?
• Is the set volume correct?
• Has the liquid density and air pressure been taken into consideration?
• Does the density of the pipetted liquid deviate from the double-distilled, degassed water?
• Was the work completed correctly, as described in the "Calibration" chapter (see p. 8) and "Evaluation" chapter (see p. 11)?
• Were original Eppendorf pipette tips used?
Volume errors can also occur when pipetting liquids with high vapor pressure, where the density or viscosity deviates considerably from the values of water.
Adjustments should only be carried out on the device after these conditions have been taken into consideration.
17

7.2 Adjusting adjustable-volume pipettes

Observe the operating instructions for your pipette. There you will find detailed descriptions on adjustment.
The adjustment of the following adjustable-volume pipettes is a zero point offset:
• Research
• Research plus
• Reference
•Biomaster
• Reference2
Adjustment
18
Eppendorf SOP English (EN)

7.2.1 Procedure

Requirements
Device, original Eppendorf tip, test liquid and ambient air must have the same temperature (15 – 30°C) at a constant temperature of ±0.5°C during the test (according to EN ISO 8655, Part 6).
If your Research, Research plus or Reference has an adjustment, remove it before carrying out the adjustment. After adjusting, lock the adjustment opening with a new adjustment seal. Adjustment seals can be reordered (see operating manual).
1. Set the pipette to the smallest volume to be tested.
2. Use a matching original Eppendorf tip for single-channel pipettes. For Multi-channel pipettes, use any channel.
3. Pipette the set volume 10 times.
4. Conduct a weighing after every volume dispensing.
5. Calculate the mean value after 10 measurements (see Evaluation on p. 11). The calculated mean value of these weighings (observe conversion factor Z = mg to µL) yields the actual volume.
6. For adjusting, introduce the suitable tool into the opening intended for this purpose, or place it on the designated location, and adjust the actual volume (see the following figures or operating manual).
7. Check the set volume by measuring again. If the target volume does not match the result of the measurement, repeat steps 2 to 6.
8. After the adjustment, check the measured values for accuracy at 50% of the nominal volume and check the nominal volume for accuracy.

7.2.2 Biomaster

1
0
2
0
Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)
1. Use the pipette wrench to adjust the volume display of the pipette, with the piston stroke unchanged, to the actual volume value of the measurement.
2. Pull off the pipette wrench.
3. Set the pipette to the target volume in the usual manner.
4. Mark the adjustment on the pipette.

7.2.3 Reference variable

Adjustment
Eppendorf SOP
English (EN)
19
Auxiliary equipment
• Supplied pipette wrench (order no. 4910 092.001)
• Supplied red CAL seal
1. Insert side B of the pipette wrench into the adjustment opening in the lid.
0
6
2
6
2. Using the pipette wrench, set the digital display of the pipette, with unaltered piston stroke, to the actual volume value of the measurement.
3. Pull off the pipette wrench.
4. Set the pipette to the target volume in the usual manner.
5. After the adjustment has been successfully completed, lock the adjustment opening with a red CAL seal.
Adjustment
Reference 2
ADJ
Reference 2
Reference 2
20
Eppendorf SOP English (EN)

7.2.4 Reference 2 adjustable-volume

7.2.4.1 User adjustment
Changing the adjustment will change the volume by a specific value. Strictly speaking, the change only applies to the testing volume.
Example:
You readjust a 10 – 100 µL pipette, with a volume setting of 100 µL, by 1 µL (1 µL  1%). If the volume setting is 10 µL, the pipette is also adjusted by 1 µL ( 10%).
Auxiliary aids from the delivery package
•Pin
• Red plastic adjustment seal (ADJ)
1. Insert the pin in the middle of the gray adjustment seal (ADJ).
2. Remove the adjustment seal.
3. Insert the adjustment tool.
4. Turn the adjustment tool until the adjustment display shows the desired value.
5. Read the set value distortion-free using the aligning aid in the viewing window.
6. Carry out weighings to check the accuracy and precision.
1000 µL
ADJ
+-
11
Loading...
+ 44 hidden pages