Eppendorf Serological Pipets User Manual

Instructions for use Eppendorf Serological Pipetsseeon p.Fig.Tab.p.
10 mL in 1/10 TD 20° C
1
2
3
4
5
6
7
8
9
10
10 mL in 1/10 TD 20° C
5678
1
2
4
3
9 10 11
English (EN)
Instructions for use
Eppendorf Serological Pipets English (EN)

1 Operating instructions

1.1 Using this manual

Before using the consumables for the first time, read these instructions for use and the operating manual of the device that you use the consumables with.

2 Product description

2.1 Main illustration

1Outlet opening
2Graduation
3Zero point
4 Negative graduation
5Nominal volume
7Pipette type
8 Test temperature
9 Color code
10 Filter
11 Aspiration area
6 Increment

2.2 Overview of Eppendorf Serological Pipets with color codes

Nominal volume 1 mL 2 mL 5 mL 10 mL 25 mL 50 mL
Color code
Accuracy [± mL] 0.02 0.04 0.1 0.2 0.5 1
Graduation [mL] 0.01 0.02 0.1 0.1 0.2 0.5
Negative graduation [mL] -0.2 -0.8 -3 -3 -9 -6
Length [mm] 277 277 338 338 343 343
Diameter at the aspiration area [mm]
4.6 6.4 8.2
Max. diameter [mm] 9.9 14.6 18.6
7.8 8.3 8.3
Instructions for use
Eppendorf Serological Pipets
2
English (EN)

2.3 Features

Eppendorf Serological Pipets are single-use items for aspiring and dispensing liquids using manual and electronic pipette controllers (e.g., Eppendorf Easypet 3 and Eppendorf Pipet Helper). The pipettes are calibrated to blow-out. The various pipette sizes are marked with a color code.

2.4 Materials

NOTICE! Aggressive substances may damage the Eppendorf Serological Pipets and accessories.
Check the resistance to chemicals before using organic solvents or aggressive chemicals.
Material
Pipette Polystyrene (PS)
Filter Polyethylene (PE)/polyester (PES)

2.5 Resistance to chemicals

■■■ Resistant
The chemical can be used.
■■ Limited resistance and/or suitable for limited use
The chemical can be used for a limited period of time. Dispense immediately after liquid aspiration. Extended contact may damage the Eppendorf Serological Pipets.
Unstable
Material and prinitng are unstable. The chemical cannot be used.
Solution Concentration in % Resistance
Ammonia solution 25 ■■
Acetic acid 12
Caustic soda 20
Nitric acid 65
Sulfuric acid 16
Acetone
Acetonitrile
Chloroform
Dimethyl sulfoxide 100 ■■■
Ethanol 70 ■■
Isoamyl alcohol ■■
Isopropanol 70 ■■