Eppendorf 5430 R User Manual

Page 1
Register your instrument!
www.eppendorf.com/myeppendorf
Centrifuge 5430/5430 R
Operating manual
Page 2
Copyright
2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior
permission of the copyright owner.
Trademarks
Eppendorf
, the Eppendorf logo. CombiSlide®, and QuickLock® are registered trademarks of
Eppendorf AG, Hamburg, Germany.
Centriplus
Microtainer
Trademarks are not marked in all cases with ™ or
is a registered trademark of Millipore Corporation, Billerica, USA.
is a registered trademark of Becton Dickinson, Franklin Lakes, NJ, USA.
in this manual.
Operating manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Declarations and Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Menu structure of Centrifuge 5430/5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Rotor F-35-6-30: Adapters for round-bottom tubes and blood taking systems . . . . . . . . . . . . . . . . . . . . . . 86
5427 901.124-04/072013
Page 3
Table of contents
Centrifuge 5430/5430 R
English (EN)
Operating manu alCentrifuge 5430/5430 Rseeon p.Fig.Tab.p.

English (EN)Operating manual

Table of contents
1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Danger symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Danger levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Abbreviations used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Main illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Delivery package. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1 Centrifuge 5430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2 Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.4 Rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4.1 rcf display and calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3
3 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3 Application limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3.1 Declaration concerning the ATEX directive (94/9/EC) . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3.2 Maximum service life for accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4 Information on product liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5 Warnings for intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.1 Personal injury or damage to the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.2 Incorrect handling of the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.3 Incorrect handling of the rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5.4 Extreme strain on the centrifuging tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.5 Aerosol-tight centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6 Safety notes on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.1 Selecting the location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2 Preparing installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3 Installing the instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.1 Overview of operating controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2 Menu navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.3 Configure centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.1 Set menu language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.2 Setting the date and time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.4 Preparing for centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.1 Switching on the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.2 Inserting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.3 Automatic rotor detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.4 Manual rotor detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.5 Loading the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.4.6 Closing the centrifuge lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Page 4
Table of contents
Centrifuge 5430/5430 R
4
English (EN)
5.5 Cooling (only 5430 R) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.6 Centrifuging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.7 Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.8 User instructions on rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.5.1 Temperature adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.2 Temperature display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.3 Temperature monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.4 FastTemp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.5 FastTemp pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.5.6 Continuous cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.6.1 Centrifugation with time setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.6.2 Centrifuging in continuous operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.6.3 Short-spin centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.6.4 Removing the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.8.1 Rotor F-35-6-30: rotor removal tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.8.2 Rotor A-2-MTP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.8.3 Rotor FA-45-24-11-HS: using the special rotor key . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.8.4 QuickLock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6 Operating controls and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.1 Device menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2 Settings in the device menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.1 Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.2 Use program keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.3 Other menu items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.2.4 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.1 Prepare cleaning/disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.2 Cleaning/disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.2.1 Cleaning and disinfecting the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.2.2 Cleaning and disinfecting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.3 Additional service instructions for Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.4 Glass breakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.5 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.6 Decontamination before shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.1 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.2 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.3 Emergency release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9 Transport, storage and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.3 Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
10 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.1 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.3 Weight/dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10.4 Application parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Page 5
Table of contents
Centrifuge 5430/5430 R
English (EN)
11 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.1 Centrifuge 5430 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.2 Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.3 Rotors, rotor lids and seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
11.3.1 Rotors with QuickLock rotor lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
11.3.2 Rotors with rotor lid thread. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11.3.3 Rotors with rotor lid for attaching. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.4 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.4.1 Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.4.2 Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
11.4.3 Fuses for Centrifuge 5430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
5
Page 6
Table of contents
Centrifuge 5430/5430 R
6
English (EN)
Page 7
Operating instructions
Centrifuge 5430/5430 R
English (EN)

1 Operating instructions

1.1 Using this manual
Read this operating manual completely before using the device for the first time. Please also note the
operating instructions for the accessories, if applicable.
This operating manual is part of the product. Thus, it must always be easily accessible.
Enclose this operating manual when transferring the device to third parties.
If this manual is lost, please request another one. For the current version, please refer to our webpage
www.eppendorf.com
The Centrifuge 5430/5430 R is available in two versions: key pad or rotary knobs. This operating manual generally describes how to operate the keypad version but it also applies to the rotary knob version.
(international) or www.eppendorfna.com (North America).
1.2 Danger symbols and danger levels
The safety instructions in this manual appear with the following danger symbols and danger levels:
7
1.2.1 Danger symbols
Biohazard Explosion
Electric shock Crushing
Hazard point Material damage
1.2.2 Danger levels
DANGER Will lead to serious injuries or death.
WARNING May lead to serious injuries or death.
CAUTION May lead to light to moderate injuries.
NOTICE May lead to material damage.
1.3 Symbols used
Example Meaning
You are requested to perform an action.
1.
2.
• List.
Perform these actions in the sequence described.
Press this key to perform the described action.
Text Ter ms from the display of the device.
References useful information.
Page 8
Operating instructions
Centrifuge 5430/5430 R
8
English (EN)
1.4 Abbreviations used
MTP
Micro test plate
PCR
Polymerase chain reaction
PTFE
Polytetrafluorethylene
RZB/rcf
Relative centrifugal force – g-force in m/s
rpm
Revolutions per minute – in rpm
2
UV
Ultraviolet radiation
Page 9

2 Product description

1
2
3
4
5
678
1
2
3
4
5
1
9
6
8
7
2.1 Main illustration
Abb. 2-1: Front and rear view of Centrifuge 5430
Product description
Centrifuge 5430/5430 R
English (EN)
9
Fig. 2-1: Front and rear view of Centrifuge 5430
Abb. 2-2: Front and rear view of Centrifuge 5430 R
Fig. 2-2: Front and rear view of Centrifuge 5430 R
Page 10
10
Product description
Centrifuge 5430/5430 R English (EN)
1 Emergency lid release
On both sides of the device (see Emergency release on p. 57).
2 Centrifuge lid
3 Control panel
Keys and dials (dependent on the device version) for operating the centrifuge(see p. 28).
4Window
Visual control for rotor stop or option for speed check via stroboscope.
5Display
Depiction of the current centrifuging parameters and device settings (see p. 28).
6 Mains connection
Connection socket for the mains cable supplied. Only 5430: The fuse holder is located beneath (see Fuses on p. 53).
7 Mains switch
Switch for switching the device on and off. Switch position 0: The device is switched off. Switch position I: The device is switched on.
8ID plate
9USB port
Interface for error analysis and software updates by the Technical Service.
2.2 Delivery package
2.2.1 Centrifuge 5430
Quantity Order no.
(International)
1– See chapter Ordering Information for corresponding device
1 5301 850.249 022654403 4 A (230 V), 2 pieces oder 5427 850.341 022654381 8.0 AT UL (120 V/100 V), 2 pieces
1 5416 301.001 022634305 Standard 1– Mains power cable
1 5427 901.124 5427901124 Languages: EN, DE, FR, ES, IT, PT 1 5427 901.132 5427901132 Languages: DA, FI, EL, NL, SV (230 V devices only)
2.2.2 Centrifuge 5430 R
Quantity Order no.
(International)
1– See chapter Ordering Information for corresponding device
1 5416 301.001 022634305 Standard 1– Mains power cable
1 5427 901.124 5427901124 Languages: EN, DE, FR, ES, IT, PT 1 5427 901.132 5427901132 Languages: DA, FI, EL, NL, SV (230 V devices only)
Order no. (North America)
Order no. (North America)
Description
Centrifuge 5430
version, equipment and order number
Fuse
Rotor key
Operating manual Centrifuge 5430/5430 R
Description
Centrifuge 5430 R
version, equipment and order number
Rotor key
Operating manual Centrifuge 5430/5430 R
Page 11
Product description
Centrifuge 5430/5430 R
English (EN)
2.3 Features
The multifaceted Centrifuge 5430 has a capacity of 48 × 2.0 mL and reaches max. 30,130 × g/17,500 min-1. The versatility is reflected in the available rotor options. You can select from 12 different rotors to centrifuge the following tubes for your various applications:
• Tubes (0.2 to 5.0 mL)
•PCR strips
• Microtainers
•Spin columns
•Cryo tubes
• Conical tubes (15/50 mL)
•Microplates
•PCR plates
• Deepwell plates (max. height 29 mm)
• Slides (with CombiSlide adapter)
Five program keys for rapid loading and saving of parameters, as well as another 45 program places, a large display and menu-controlled operation all make it easier to use the centrifuge. The Centrifuge 5430 has been designed based on latest ergonomic studies. This facilitates an intuitive and easy operation.
11
The Centrifuge 5430 is available with two different operator panels: one easy to clean keypad or blue rotary knobs to quickly set the centrifugation parameter.
The Centrifuge 5430 R has an additional temperature control function for centrifugation between -11°C and +40°C. The Fast Temp function is used to start a temperature control run without samples in order to quickly bring the rotor chamber, and rotor, bucket and adapter, to the set temperature. This temperature control cycle can also be started automatically at specified times using the Fast Temp pro function.
Page 12
12
Product description
Centrifuge 5430/5430 R English (EN)
2.4 Rotors
The Centrifuge 5430/5430 R can be operated with the following rotors. Before using tubes, observe the manufacturer's recommended specifications on resistance to centrifugation (max. g-force).
Rotor FA-45-48-11
With aerosol-tight QuickLock rotor lid
Rotor F-45-48-11
With polypropylene rotor lid
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
48 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
Outer ring: 18,213 × g Inner ring: 16,048 × g/ 12,700 rpm
≤ 20 s/20 s ≤ 61 s/65 s (soft)
48 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
Outer ring: 18,213 × g Inner ring: 16,048 × g/ 12,700 rpm
≤ 20 s/20 s ≤ 61 s/65 s (soft)
Max. load per rotor
(1)
bore
(2)
3.75 g
3.75 g
Notes
• Aerosol-tight QuickLock rotor lid (aluminum).
(3)
Rotor FA-45-30-11
With aerosol-tight QuickLock rotor lid
Rotor F-45-30-11
With polypropylene rotor lid
30 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
30 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
20,817 × g/ 14,000 rpm
≤ 15 s/15 s ≤ 61 s/65 s (soft)
20,817 × g/ 14,000 rpm
≤ 15 s/15 s ≤ 61 s/65 s (soft)
3.75 g
3.75 g
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
•PTFE-coated (particularly resistant to chemicals), marked: coated
•Spin columns available, better with rotor FA-45-24-11-kit.
Page 13
Product description
Centrifuge 5430/5430 R
English (EN)
13
Rotor FA-45-24-11-Kit
With aerosol-tight QuickLock rotor lid
Rotor FA-45-16-17
With aerosol-tight QuickLock rotor lid
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
24 spin columns or 1.5/2.0 mL tubes. With adapters:
• 0.2 mL PCR tubes
19.090 × g/ 13,200 rpm
≤ 15 s/16 s ≤ 78 s/90 s (soft)
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
16 tubes 5.0 mL. 21,191 × g/
14,200 rpm
≤ 20 s/20 s ≤ 61 s/66 s (soft)
Max. load per rotor
(1)
bore
(2)
3.75 g
9.5 g
Notes
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
• Uniquely high edge, for all commercial spin columns.
Also observe the note on centrifugation with open tube lids (see
Fixed-angle rotors on p. 33)
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
Rotor S-24-11-AT
With aerosol-tight QuickLock rotor lid
24 tubes 1.5/2.0 mL. 16,049 × g/
12,700 rpm
This rotor exclusively is to be used with 1.5/2.0 mL reaction
≤ 13 s/16 s
≤ 61 s/66 s (soft) vessels. Spin Columns and the adapters including the corresponding 0.2 mL, 0.4 mL,
0.5 mL and 0.6 mL vessels must not be used in this rotor.
3.75 g
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
•The rotor must always be operated with a rotor lid.
Page 14
14
Product description
Centrifuge 5430/5430 R English (EN)
Rotor FA-45-24-11-HS
Rotor F-45-64-5-PCR
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
24 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
30,130 × g/ 17,500 rpm
≤ 21 s/16 s ≤ 61 s/65 s (soft)
• 0.5 mL tubes
• 0.6 mL Microtainer
64 PCR tubes (0.2 mL) or
13,543 × g/
11,800 rpm eight 5-tube or 8-tube PCR strips, each with the enclosed
≤ 12 s/15 s
≤ 62 s/65 s (soft) adapters.
Max. load per rotor
(1)
bore
(2)
3.75 g
3.4 g (without adapter)
Notes
• Aerosol-tight
(3)
rotor
lid (aluminum).
• Max. g-force/speed (30,130 × g/17,500 rpm) only with tubes approved by the man­ufacturer for this speed.
• PTFE-coated (particu­larly resistant to chemicals), marked:
coated
• Spin columns avail­able, better with rotor FA-45-24-11-kit.
•The rotor can only be tightened and loos­ened using the special rotor key for rotor FA-45-24-11-HS (see p. 42).
Rotor F-45-18-17-Cryo
18 Cryo tubes or 18 sealable centrifugation tubes, max. Ø: 16.9 mm. With supplied adapters: max. Ø: 13.4 mm, max. tube length: 50 mm.
8,324 × g/ 8,900 rpm
≤ 8 s/11 s ≤ 77 s/85 s (soft)
8.7 g • Setting the g-force/ speed in increments of 10 × g or 10 rpm.
Page 15
Product description
Fixed-angle
rotor
Aerosol-tight
version
Angle of
bores
Number of
bores
of
bores (mm)
AT
24 11
Aerosol-tight
version
Number of
bores
of
bores (mm)
Swing-bucket
rotor
Centrifuge 5430/5430 R
English (EN)
15
Rotor F-35-6-30
Rotor A-2-MTP
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
6 conical tubes 50 mL, with or without skirted bottom or 6 conical tubes 15 mL, with
7,745 × g/ 7,830 rpm
≤ 23 s/23 s ≤ 62 s/67 s (soft)
enclosed adapters or 6 Centriplus centrifuge filter units with adapters.
Two buckets to hold:
•Microplates
2,204 × g/ 4,680 rpm
• Cell culture plates
•PCR plates
• Deepwell plates (max.
≤ 18 s/21 s ≤ 63 s/67 s (soft)
height 29 mm)
• Slide (with CombiSlide adapter)
Max.
Notes load per rotor
(1)
bore
(2)
110 g • The rotor can only be
removed and inserted using the supplied removal tool.
•Centrifugation of round-bottom tubes and blood taking systems can be completed using additional adapters (see p. 86).
170 g (per bucket)
• PCR plates can only be centrifuged with appropriate adapters.
• Max. loading height: 29 mm.
Only 5430 R: More efficient cooling via centrifugation without wind shield upper shell (see
Swing-bucket rotor on p. 34).
(1) Maximum load per rotor bore for adapter + tube + contents.
(2) According to DIN 58 970 (device versions: 230 V, 120 V and 100 V, 50 to 60 Hz).
(3) Aerosol tightness tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK) (see certificates at the end of this operating manual).
For the rotors and rotor lids labeled coated, color fluctuations may occur as a result of the production process. These fluctuations have no effect on service life or resistance to chemicals.
Page 16
16
Product description
Centrifuge 5430/5430 R English (EN)
2.4.1 rcf display and calculation
Use the rpm/rcf key to switch the speed of centrifugation display from rpm to g-force (rcf). Ensure that the g-force displayed during switching is standardized to suit the rotor in question without an adapter. When adapters are used, the following maximum g-forces (rcf) can be achieved at maximum speed:
Rotor Adapter Max.centrifugation
radius r
Rotor FA-45-48-11/ Rotor F-45-48-11
Rotor FA-45-30-11/ Rotor F-45-30-11
Rotor FA-45-24-11-Kit
Rotor FA-45-16-17 For 5.0 mL micro test tubes 9.4 21,191
Rotor S-24-11-AT Without adapter 8.9 16,049
Rotor FA-45-24-11-HS
Rotor F-45-64-5-PCR
Rotor F-45-18-17-Cryo
Without adapter Outer ring: 10.1
Inner ring: 8.9
For 0.2 mL PCR tubes Outer ring: 8
Inner ring: 6.8
For 0.4 mL tubes Outer ring: 10.1
Inner ring: 8.9
For 0.5 mL tubes Outer ring: 9
Inner ring: 7.8
For 0.6 mL Microtainer Outer ring: 10.1
Inner ring: 8.9
Without adapter 9.5 20,871
For 0.2 mL PCR tubes 7.4 16.215
For 0.4 mL tubes 9.5 20,871
For 0.5 mL tubes 8.4 18,407
For 0.6 mL Microtainer 9.5 20,817
Without adapter 9.8 19,090
For 0.2 mL PCR tubes 7.7 15,000
For 0.4 mL tubes 9.8 19,090
For 0.5 mL tubes 8.7 16,950
For 0.6 mL Microtainer 9.8 19,090
Without adapter 8.8 30,130
For 0.2 mL PCR tubes 6.7 22,940
For 0.4 mL tubes 8.8 30,130
For 0.5 mL tubes 7.7 26,364
For 0.6 mL Microtainer 8.8 30,130
For PCR strips, inside 7.7 11,987
For PCR strips, outside 8.7 13,543
Without adapter 9.4 8,320
For Cryo tubes 9.0 7,970
max
[cm]
Max. g-force (rcf)
Outer ring: 18,210 Inner ring: 16,048
Outer ring: 14,425 Inner ring: 12,261
Outer ring: 18,210 Inner ring: 16,048
Outer ring: 16,229 Inner ring: 14,065
Outer ring: 18,210 Inner ring: 16,048
Page 17
Product description
Centrifuge 5430/5430 R
English (EN)
17
Rotor Adapter Max.centrifugation
radius r
max
[cm]
Max. g-force (rcf)
Rotor F-35-6-30* For 15 mL conical tubes 11.0 7,540
For 50 mL conical tubes 10.5 7,197
for Centriplus centrifuge filter units 11.1 7,567
A-2-MTP rotor Without adapter 9.0 2,204
for 384 well PCR plates 7.7 1,885
for 96 well PCR plates 7.3 1,788
CombiSlide adapter 7.7 1,885
*) Centrifugation of round-bottom tubes and blood taking systems can be done using additional adapters (see p. 86).
To determine the g-force (rcf) for a special adapter, you can perform a calculation according to DIN 58 970 with the following formula:
-5
rcf = 1.118 · 10
· n2 · r
max
n: speed in min-1 (rpm)
: max. centrifuging radius in cm
r
max
Example:
In Rotor FA-45-30-11, the 0.5 mL adapter has a maximum radius of 8.4 cm. At 7,000 rpm, a maximum g-force of 4,600 × g is achieved.
Page 18
18
Safety
Centrifuge 5430/5430 R English (EN)

3Safety

3.1 Intended use
The Centrifuge 5430/5430 R is intended exclusively for indoor use and for separating aqueous solutions and suspensions of various densities in approved test tubes.
3.2 User profile
This device may only be operated by trained specialist staff. They must have carefully read the operating manual and be familiar with the function of the device.
3.3 Application limits
3.3.1 Declaration concerning the ATEX directive (94/9/EC)
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive substances.Do not use this device to process any explosive or highly reactive substances.Do not use this device to process any substances which may generate an explosive
atmosphere.
Due to its design and the environmental conditions inside the device, the Centrifuge 5430/5430 R is not suitable for use in a potentially explosive atmosphere.
The device only must be used in a safe environment, such as the open environment of a ventilated laboratory or fume hood. The use of substances which could create a potentially explosive atmosphere is not permitted. The final decision on the risks associated with the use of these types of substances is the responsibility of the user.
3.3.2 Maximum service life for accessories
WARNING! Risk of injury from chemically or mechanically damaged accessories.
Even minor scratches and cracks can lead to serious internal material damage.
Protect all accessory parts from mechanical damage.Inspect the accessories for damage before each use. Replace any damaged accessories.Do not use rotors, rotor lids or buckets with signs of corrosion or mechanical damage (e.g.,
deformations).
Do not use any accessories whose maximum service life has been exceeded.When inserting the buckets and rotors, ensure that they do not become scratched.
CAUTION! Risk of injury due to chemically damaged rotor lids or caps.
Transparent rotor lids or caps made from PC, PP or PEI may lose their strength under the impact of organic solvents (e.g., phenol, chloroform).
If rotor lids or caps have come into contact with organic solvents, they should be cleaned
immediately.
Regularly check the rotor lids and caps for damages and cracks.Immediately replace rotor lids or caps that have cracks or milky stains.
Page 19
Safety
Centrifuge 5430/5430 R
English (EN)
The rotors listed below, and the corresponding buckets and rotor lids, have a maximum service life of the number of years or cycles listed in the table. Since 2012, Eppendorf has been stating the maximum service life of the rotors and accessories both in years and in the maximum number of cycles. The decisive factor for the service life is which case occurs first, generally this is the number of years in operation.
Rotor Maximum service from commissioning onward
FA-45-48-11 (With QuickLock)
F-45-48-11 (With rotor lid thread)
FA-45-30-11 (With QuickLock)
FA-45-24-11-Kit (With QuickLock)
FA-45-16-17 (With QuickLock)
S-24-11-AT (With QuickLock)
A-2-MTP With the corresponding bucket and upper shell of the wind shield
QuickLock rotor lid - 3 years
Rotor lid and caps made of polycarbonate (PC), polypropylene (PP) or polyetherimide (PEI)
Plastic adapters - 1 year
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 7 years
-3 years
19
For all other rotors and rotor lids of this centrifuge there is no service life limit as long as the following requirements are met:
•Proper use,
• Recommended maintenance
• Undamaged condition
The date of manufacture is stamped on the rotors in the format 03/10 (= March 2010) or on the inside of the
plastic rotor lids in the form of a clock .
To ensure aerosol tightness, the following applies:
• Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.
• Replace the seal of QuickLock rotor lids after 50 autoclaving cycles.
Page 20
20
Safety
Centrifuge 5430/5430 R English (EN)
3.4 Information on product liability
In the following cases, the designated protection of the device may be compromised. Liability for any resulting property damage or personal injury is then transferred to the operator:
• The device is not used in accordance with the operating manual.
• The device is used outside of its intended use.
• The device is used with accessories or consumables which are not recommended by Eppendorf.
• The device is maintained or repaired by people not authorized by Eppendorf.
• The user makes unauthorized changes to the device.
3.5 Warnings for intended use
Read the operating manual and observe the following general safety instructions before using the Centrifuge 5430/5430 R.
3.5.1 Personal injury or damage to the equipment
WARNING! Electric shock due to damage to device or mains cable.
Only switch on the device if the device and mains cable are undamaged.Only use devices that have been properly installed or repaired.In case of danger, disconnect the device from the mains supply by pulling the power plug
from the device or the mains socket or, by using the isolating device intended for this purpose (e.g. emergency stop switch in the laboratory).
WARNING! Lethal voltages inside the device.
Ensure that the housing is always closed and undamaged so that no parts inside the device
can be contacted by accident.
Do not remove the housing of the device.Do not allow any liquids to penetrate the inside of the housing.Do not allow the device to be opened by anyone except service personnel who have been
specifically authorized by Eppendorf.
WARNING! Risk from incorrect supply voltage
Only connect the device to voltage sources which correspond to the electrical
requirements on the name plate.
Only use sockets with a protective earth (PE) conductor and suitable power cable.
WARNING! Damage to health due to infectious liquids and pathogenic germs.
When handling infectious liquids and pathogenic germs, observe the national regulations,
the biological security level of your laboratory, the material safety data sheets, and the manufacturer's application notes.
Use aerosol tight sealing systems for the centrifugation of these substances.When working with pathogenic germs belonging to a higher risk group, more than one
aerosol-tight bioseal must be used.
Wear personal protective equipment.For full instructions regarding the handling of germs or biological material of risk group II
or higher, please refer to the "Laboratory Biosafety Manual" (Source: World Health Organization, current edition of the Laboratory Biosafety Manual).
Page 21
Safety
Centrifuge 5430/5430 R
English (EN)
WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
When opening or closing the centrifuge lid, do not reach between the lid and device or into
the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or accept any liability for damage resulting from the use of incorrect or non-recommended accessories and spare parts, or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
NOTICE! Damage to device due to spilled liquids.
1. Switch off the device.
2. Disconnect the device from the power supply.
3. Carefully clean the device and the accessories in accordance with the cleaning and disinfection instructions in the operating manual.
4. If a different cleaning and disinfecting method is to be used, contact Eppendorf AG to ensure that the intended method will not damage the device.
21
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a warmer environment.
After installing the device, wait for at least 4 h. Only then connect the device to the mains
power supply.
3.5.2 Incorrect handling of the centrifuge
NOTICE! Damage from knocking against or moving the device during operation.
If the rotor bangs against the rotor chamber wall, it will cause considerable damage to the device and rotor.
Do not move or knock against the device during operation.
Page 22
22
Safety
Centrifuge 5430/5430 R English (EN)
3.5.3 Incorrect handling of the rotors
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with rotor and rotor lid firmly tightened.If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Immediately press the start/stop key to stop centrifuging.
CAUTION! Risk of injury due to asymmetric loading of a rotor.
Load rotors symmetrically with identical tubes or plates and buckets.Only load adapters with suitable tubes or plates.Always use tubes or plates of the same type (weight, material/density and volume).Check that loading is symmetrical by balancing the adapters and tubes or plates used with
scales. The device automatically detects imbalances during operation and stops the run immediately with an error message and a signal tone. Check the loading, balance the tubes and re-start the centrifugation.
CAUTION! Risk of injury from overloaded rotor.
The Centrifuge 5430/5430 R is designed for the centrifugation of material with a max. density of 1.2 g/mL at maximum speed and filling volume and/or load.
Do not exceed the maximum load of the rotor.
NOTICE! Damage to rotors from aggressive chemicals.
Rotors are high-quality components which withstand extreme stresses. This stability can be impaired by aggressive chemicals.
Avoid using aggressive chemicals, including strong and weak alkalis, strong acids,
solutions with mercury, copper and other heavy metal ions, halogenated hydrocarbons,
concentrated saline solutions and phenol.
If the rotor is contaminated by aggressive chemicals, clean it immediately using a neutral
cleaning agent. This applies to the rotor bores, in particular.
Due to the manufacturing process, color variations may occur on rotors marked "coated".
These color variations do not effect service life or resistance to chemicals.
NOTICE! If handled incorrectly, the rotor can fall over.
The rotor can fall over if the buckets are used as a handle.
Remove the buckets before inserting and/or removing a swing-bucket rotor.Always use both hands to carry the rotor cross.
CAUTION! Risk of injury due to chemically damaged rotor lids or caps.
Transparent rotor lids or caps made from PC, PP or PEI may lose their strength under the impact of organic solvents (e.g., phenol, chloroform).
If rotor lids or caps have come into contact with organic solvents, they should be cleaned
immediately.
Regularly check the rotor lids and caps for damages and cracks.Immediately replace rotor lids or caps that have cracks or milky stains.
Page 23
3.5.4 Extreme strain on the centrifuging tubes
CAUTION! Risk of injury from overloaded tubes.
Note the loading limits specified by the tube manufacturer.Only use tubes which are approved by the manufacturer for the required rcf.
NOTICE! Risk from damaged tubes.
Damaged tubes must not be used, as this could cause further damage to the device and the accessories and loss of the samples.
Before use, visually check all of the tubes for damage.
NOTICE! Risk from open tube lids.
Open tube lids can brake off during centrifugation and damage the rotor and the centrifuge.
Carefully seal all tube lids before centrifuging.
Exception: Note the information on the centrifugation of spin columns in the rotor FA-45-24-11-Kit (see p. 34).
Safety
Centrifuge 5430/5430 R
English (EN)
23
NOTICE! Hazard to plastic tubes from organic solvents.
The density of plastic tubes is reduced when organic solvents (e.g., phenol, chloroform) are used, i.e. the tubes could become damaged.
Observe the manufacturer's specifications for chemical resistance of the tubes.
NOTICE! Sample tubes heat up.
In uncooled centrifuges, the temperature in the rotor chamber, rotor and sample can increase to above 40 °C, based on the run time, g-force (rcf)/speed and ambient temperature.
Note that this can reduce the centrifugation resistance of the sample tubes.Please note the temperature resistance of the samples.
Page 24
24
Safety
Centrifuge 5430/5430 R English (EN)
3.5.5 Aerosol-tight centrifugation
WARNING! Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this purpose are used. The designation of aerosol-tight fixed-angle rotors always starts with FA.The aerosol-tight rotors and rotor lids of this centrifuge are additionally marked with a red ring on the rotor and a red rotor lid screw. Aerosol-tight swing-bucket rotors are marked with AT (aerosol-tight).
For aerosol-tight centrifugation, always simultaneously use rotors and rotor lids which are
marked as aerosol-tight in the centrifuge intended for the corresponding purpose. The
details specifying in which centrifuge you may use the aerosol-tight rotors and rotor lids
can be found on the rotor and, beginning from production date of October 2003, on the
upper side of the rotor lid.
Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor
lid.
Only use aerosol-tight buckets with the corresponding caps.
WARNING! Damage to health as a result of limited aerosol tightness and incorrect usage.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive solvents can impair the aerosol-tightness of the rotors and rotor lids.
Check the integrity of the seals of the aerosol-tight rotor lids or caps before each use.Only use aerosol-tight rotor lids or caps if the seals are undamaged and clean.Lightly grease the threads of the rotor lid screw with pivot grease after every proper
autoclaving (121 °C, 20 min.) (int. order no. Int. 5810 350.050, North America
022634330).
Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.For QuickLock rotor lids, the seal must be replaced after 50 autoclaving cycles.Never store aerosol-tight rotors or buckets closed.
3.6 Safety notes on the device
Symbol Meaning Location
Hazard point. 5430: Rear of the device
CAUTION
Always tighten the rotor with the enclosed rotor key.
CAUTION
Close all tubes and use a rotor lid.
5430 R: Right side of the device
Upper side of device, under the centrifuge lid.
Upper side of device, under the centrifuge lid.
Page 25

4 Installation

4.1 Selecting the location
NOTICE! If an error occurs, the objects in the immediate proximity of the device will be damaged.
In accordance with recommendations in EN 61010-2-020, leave a safety clearance of
30 cm around the device during operation.
Please remove all materials and objects from this area.
NOTICE! Damage from overheating.
Do not install the device near heat sources (e.g. heating, drying cabinet).Do not expose the device to direct sunlight.Ensure unobstructed air circulation. Maintain a clearance of at least 30 cm around all
ventilation grilles.
Installation
Centrifuge 5430/5430 R
English (EN)
25
Select the location for the device according to the following criteria:
• Suitable power connection as per the name plate (230 V/120 V/100 V).
• Stable, horizontal and resonance-free lab bench.
• A well ventilated environment which is protected from direct sunlight to prevent the device from heating up more.
4.2 Preparing installation
Prerequisites
The weight of the centrifuge is 29 kg (5430) or 56 kg (5430 R). A second person is needed to unpack and position the device.
Retain the packaging material and the transport protection device for subsequent transport or storage. See also the instructions relating to transport (see p. 58).
Perform the following steps in the sequence described.
Centrifuge 5430 Centrifuge 5430 R
1. Open the packaging board.
2. Remove the covering cardboard.
3. Remove the accessories.
4. Grip from the strap retainers and have two persons lift the centrifuge out of the box.
5. Pull off the strap retainers, do not cut.
6. Remove the front and rear transport securing devices from the centrifuge.
7. Remove the plastic sleeve.
8. Carefully lift the centrifuge on one side and pull off the transport securing device of the motor on the underside of the centrifuge.
1. Open the packaging board.
2. Remove the accessories.
3. Lift and remove the front and rear transport securing devices.
4. Grip from the fabric straps and have two people lift the centrifuge out of the box.
5. Pull off the textile straps, do not cut.
6. Remove the plastic sleeve.
Page 26
26
Installation
Centrifuge 5430/5430 R English (EN)
4.3 Installing the instrument
Prerequisites
The device is on a suitable lab bench.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a warmer environment.
After installing the device, wait for at least 4 h. Only then connect the device to the mains
power supply.
NOTICE! Centrifuge 5430 R: compressor damage after improper transport.
Only switch on the centrifuge 4 h after installation.
Perform the following steps in the sequence described.
1. Let the device warm up to ambient temperature for at least 4 hours to avoid damaging electronic assemblies as a result of condensation and avoid damaging the compressor (only 5430 R).
2. Check that the mains voltage and frequency match the requirements on the device type plate.
3. Connect the centrifuge to the mains and switch it on using the mains/power switch on the rear of the device (Centrifuge 5430) or at the right side of the device (Centrifuge 5430 R).
• The standby key lights green.
•The display is active.
Only 5430: Lid opens automatically
4. Only 5430: Remove the transport securing device of the motor shaft.
5. If the scope of delivery includes a rotor, dismantle and remove it using the supplied rotor key.
6. Only 5430: Remove the transport securing device of the air guide ring.
Page 27
Centrifuge 5430/5430 R
7. Use the details included in the scope of delivery to check that the delivery is complete.
8. Check all parts for any transport damage. Contact your dealer if any damage is found.
Retain the packaging material and the transport protection device for subsequent transport or storage. See also the instructions relating to transport (see p. 58).
Installation
27
English (EN)
Page 28
28
1
2
3
4
5
6
7891011
13 12
Operation
Centrifuge 5430/5430 R English (EN)

5 Operation

5.1 Overview of operating controls
The Centrifuge 5430/5430 R is available in two versions: keypad or rotary knobs. This operating manual generally describes how to operate the keypad version. However, it also applies to the rotary knob version.
Before using the Centrifuge 5430/5430 R for the first time, familiarize yourself with the operating controls and the display.
Abb. 5-1: Control panel and display of Centrifuge 5430/5430 R (keypad version).
Fig. 5-1: Control panel and display of Centrifuge 5430/5430 R (keypad version).
1 Select program
Press briefly: load the stored centrifuging
parameters. Press and hold (> 2 sec): save the current centrifuging parameters (see p. 47).
2 Short Spin centrifugation
(see p. 40)
3Release lid
4 Start and stop centrifugation
5 Activate/deactivate standby mode
Key lights green: centrifuge is ready for operation. Key lights red: standby mode active (see p. 41).
6Display
8 Adjust the temperature (only 5430 R)
9 Adjust the centrifuging duration
Dependent on device version, designed as key or rotary knob.
10 Start the temperature control cycle Fast Temp
(only 5430 R)
(see p. 36)
11 Switch the displayed centrifuging speed (rpm/
rcf)
12 Call and select the menu parameters
(see p. 47)
13 Navigating the menu
(see p. 29)
7 Set speed of centrifugation
Dependent on device version, designed as key or rotary knob.
Page 29
Abb. 5-2: Display of the Ce ntrifuge 5430/5430 R
1
2 3 4 5 7
6
8
10
9
111213
Fig. 5-2: Display of the Centrifuge 5430/5430 R
Operation
Centrifuge 5430/5430 R
English (EN)
29
1 Program number (if enabled)
2 Status of the function At set rpm
: Start of operation when reaching 95% of the
preset g-force (rcf) or speed.
: Start of run time immediately.
3 Status of the key lock
: Centrifuging parameters cannot be modified
unintentionally.
: No key lock.
4 Status of the loudspeaker
: Switched on.
: Switched off.
5 Status of the centrifuge
: Centrifuge lid unlocked.
: Centrifuge lid locked.
(flashing): Centrifuging in progress.
6 Temperature control cycle programming (only
5430 R)
: Fast Temp pro is enabled, i.e., start time and temperature of the temperature control run have been programmed (see p. 37).
8Soft ramp
: Rotor accelerates and brakes slowly.
No symbol: Rotor accelerates and brakes rapidly.
9Standard display
10 Extended display (if enabled)
11 g-force/speed
12 Temperature (only 5430 R)
13 Centrifuging duration
7Time
Please also read the precise description of the individual functions (see p. 47).
5.2 Menu navigation
The menu consists of two levels. To change settings, generally proceed as follows.
1. Open menu.
Page 30
30
Operation
Centrifuge 5430/5430 R English (EN)
2. Select the desired menu item.
3. Confirm selection.
4. Select the setting of the parameters in question.
5.
6.
Confirm changed setting. A tick appears in front of the confirmed setting.
Keep pressing the key until you reach the desired menu level or exit the menu. Some menus can only be exited by selecting and confirming the menu item Back /
Zurück / Retour / Atrás.
5.3 Configure centrifuge
5.3.1 Set menu language
Proceed as follows to set menu language.
1. Open the menu.
2. Select Settings.
3. Confirm your selection.
4. Select Language.
5. Confirm your selection.
6. Select English, Deutsch, Francais or Espanol.
7.
8. Press key several times to exit the menu.
Confirm your selection. A tick appears in front of the selected language. The setting takes effect immediately.
5.3.2 Setting the date and time
Proceed as follows to set date and time.
1. Open the menu.
Page 31
2. Select Settings.
3. Confirm your selection.
4. Select Date/time.
5. Confirm your selection.
6. Set date.
7. Confirm setting.
Operation
Centrifuge 5430/5430 R
English (EN)
31
8. Set time format (12 h/24 h).
9. Confirm setting.
10. Set time.
11. Confirm setting.
12. Press key several times to exit the menu.
There is no automatic switch between summer time and winter time.
Page 32
32
Operation
Centrifuge 5430/5430 R English (EN)
5.4 Preparing for centrifugation
5.4.1 Switching on the centrifuge
1. Switch on the centrifuge using the mains power switch or the standby key.
Only 5430: After switching on at the mains power switch, the centrifuge lid opens automatically.
2. Open the closed centrifuge lid by pressing the open key.
The parameter settings of the last run are displayed.
5.4.2 Inserting the rotor
Swing-bucket rotors: remove the buckets before inserting and/or removing the rotor. Use
both hands to pick up the rotor cross.
Rotor F-35-6-30: only use the rotor removal tool supplied to insert or remove the rotor.
1. Fit the rotor vertically on the motor shaft.
2. Insert the supplied rotor key into the rotor nut.
Rotor FA-45-24-11-HS: Use the special rotor key.
3. Turn rotor key clockwise until the rotor nut is firmly tightened.
5.4.3 Automatic rotor detection
The centrifuge has automatic rotor detection. It detects a newly inserted rotor during centrifugation and displays its name for approx. 2 seconds. The set g-force (rcf)/speed (rpm) is automatically limited to the maximum permissible value of the rotor, if necessary.
If you start centrifuging immediately after a rotor change, the centrifuge has not carried out an automatic rotor detection yet. The speed set for the previous rotor may exceed the maximum permitted speed for the new rotor. In this case the centrifuge stops after the automatic rotor detection and displays the error message Note C. The new maximum permitted speed appears in the display. You can then restart the centrifuging with this setting or adjust the speed as necessary.
Always check the set g-force (rcf)/speed (rpm) after a rotor change and adjust it if necessary.
5.4.4 Manual rotor detection
CAUTION! Risk of injury when turning the rotor manually.
When turning a swing-bucket rotor, pay special attention to ensure that your fingers do not
get jammed or get caught on the swinging buckets.
In order to trigger the rotor detection manually, turn the rotor counterclockwise by hand.
• The name of the rotor appears in the display.
• The g-force (rcf)/speed (rpm) automatically is limited to the maximum value of the rotor.
Page 33
5.4.5 Loading the rotor
CAUTION! Risk of injury due to asymmetric loading of a rotor.
Load rotors symmetrically with identical tubes or plates and buckets.Only load adapters with suitable tubes or plates.Always use tubes or plates of the same type (weight, material/density and volume).Check that loading is symmetrical by balancing the adapters and tubes or plates used with
scales. The device automatically detects imbalances during operation and stops the run immediately with an error message and a signal tone. Check the loading, balance the tubes and re-start the centrifugation.
CAUTION! Risk from damaged or overloaded tubes.
When loading the rotor, observe the safety precautions on dangers as a result of
overloaded or damaged tubes (see Warnings for intended use on p. 20).
Operation
Centrifuge 5430/5430 R
English (EN)
33
The device automatically detects imbalances during operation and stops the run immediately with an error message and a signal tone.
Check the load, balance the tubes and restart the run.
5.4.5.1 Fixed-angle rotors
Rotor lid
• Fixed-angle rotors may only be operated with the appropriate rotor lid in each case. This is clearly shown by the identical rotor name labeling on the rotor and on the rotor lid.
• To carry out an aerosol-tight centrifugation, an aerosol-tight rotor (label: red ring) and the corresponding aerosol-tight rotor lid (label: aerosol-tight and red lid screw) must be used.
To load the rotor, proceed as follows:
1. Check the maximum load (adapter, tube and contents) per rotor bore.
The information about this can be found on the rotor and in this operating manual (see Rotors on p. 12).
2. Load rotors and adapters only with the tubes intended for them.
3. Insert tubes opposite each other in pairs into the rotor bores. To ensure symmetric loading, tubes that are arranged opposite each other must be of the same type and contain the same filling quantity.
To minimize weight differences between filled sample tubes, we recommend taring with a scale. This will reduce wear on the drive and reduce running noise.
4. Attach and tighten rotor lid.
Page 34
34
Operation
Centrifuge 5430/5430 R English (EN)
Spin columns
For centrifuging spin columns in the rotor FA-45-24-11-Kit, you can leave the tube lids open. However, this can only be done using the tubes provided by kit manufacturers for this purpose. For reliable centrifugation, you must lean the open tube lids against the edge of the rotor. Ensure that this does not involve the lids projecting beyond the edge of the rotor, and then put on the associated rotor lid.
5.4.5.2 Swing-bucket rotor
Prerequisites
• A combination of rotor, carrier and adapter, approved by Eppendorf.
• Matching and tested tubes and plates.
NOTICE! Filling the plates too high can cause overflowing.
During the run the meniscuses in the tubes along the edges of the plates are at an angle. This is due to the centrifugal forces and cannot be avoided.
Fill the wells of the plates to a maximum of 2/3 of the max. capacity.
To load the rotor, proceed as follows:
1. Check the carrier grooves for cleanliness and grease lightly with pivot grease (order no. Int.: 5810 350.050 / North America: 022634330).
Dirty grooves and pivots prevent carriers from swinging out evenly.
2. Hang the buckets into the rotor.
All rotor positions must be loaded with carriers.
3. Check that all carriers are hanging properly and can swing freely.
4. Carry out a manual loading and swing test the first time a tube or plate type is used.
5. Check the maximum load per carrier (adapter, tube or plate and contents) and the loading height.
The information about this can be found on the rotor and in this operating manual (see Rotors on p. 12).
Page 35
Operation
Centrifuge 5430/5430 R
English (EN)
6. Load the buckets symmetrically.
Abb. 5-3: Incomplete, but symmetric loading of the buckets. The pegs of each bucket must be uniformly loaded .
Fig. 5-3: Incomplete, but symmetric loading of the buckets. The pegs of each bucket must be uniformly
loaded.
Abb. 5-4: Symmetrical loa ding of the plates.
35
Fig. 5-4: Symmetrical loading of the plates.
The plate arrangement shown on the right-hand side is incorrect, as the buckets will not swing properly.
The plates have some play in the buckets.
7. Check the loading of the bucket.
Carry out a brief centrifugation test at low speed (e.g., 1000 rpm) when you use a tube or
plate type for the first time.
Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield in order to guarantee precise and quick temperature control of samples. Note that the centrifugation noise will increase slightly in this case. This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor lid.
5.4.6 Closing the centrifuge lid
WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
When opening or closing the centrifuge lid, do not reach between the lid and device or into
the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
1. Check the correct attachment of the rotor and rotor lid.
2. Push down the centrifuge lid until the lid latch engages and the lid is automatically closed.
The centrifuge will close automatically.
The open key lights up blue. The symbol appears in the display.
Page 36
36
Operation
Centrifuge 5430/5430 R English (EN)
5.5 Cooling (only 5430 R)
5.5.1 Temperature adjustment
Set the temperature using the temp arrow keys between -11°C and +40°C.
You can also modify the temperature during centrifugation.
At higher ambient temperature a brief fan noise is possible until the desired temperature has been reached. This indicates a heavy cooling performance.
5.5.2 Temperature display
If the rotor is stopped: Target temperature During centrifugation: Actual temperature
The set temperature appears in the extended display.
5.5.3 Temperature monitoring
After the target temperature has been reached, the centrifuge reacts to temperature deviations during centrifugation as follows:
Deviation from set value Action
± 3 °C Temperatures on the display are flashing.
± 5 °C Periodic warning tone and display Error 18. Centrifugation is stopped
automatically.
5.5.4 FastTemp
This function can be used to start a temperature control run directly without samples with a rotor and temperature-specific speed in order to quickly adjust the rotor chamber, including the rotor, buckets and adapters, to the previously set nominal temperature.
The FastTemp pro function for programming the temperature control run with defined start times is described in the next section.
Prerequisites
• The centrifuge is switched on.
• The rotor and rotor lid are properly attached.
• The centrifuge lid is closed.
• Temperature and g-force (rcf)/speed (rpm) for the centrifugation are set (see Centrifuging on p. 39).
1. Press the fast temp key.
The display shows Fast Temp, the remaining duration of the temperature control run as well as the current temperature and g-force (rcf)/speed (rpm).
The temperature control run automatically ends when the set temperature is reached. A periodic signal tone sounds.
2. Press the start/stop key to terminate the temperature control run early.
After the set temperature has been reached and the temperature control run is complete, the centrifuge keeps the rotor chamber with the centrifuge lid closed at the set target temperature if the temperature is below the ambient temperature. However, independent of the target temperature, 4 °C must be met via this continuous cooling in order to prevent the rotor chamber from freezing.
Page 37
Operation
Centrifuge 5430/5430 R
English (EN)
The centrifuge stops the cycle automatically if the rotor or the buckets have reached the set temperature. Therefore, there may be a delay of approx. 30 min between the display of the set temperature and the automatic end of the temperature control run.
Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield in order to guarantee precise and quick temperature control of samples. Note that the centrifugation noise will increase slightly in this case. This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor lid.
5.5.5 FastTemp pro
You can have the previously described temperature control run FastTemp (see p. 36) start automatically at a specified time. Two options are available:
Once The temperature control cycle is started once at the set time.
Repeatedly The temperature control cycle is started at the set time on the next specified weekday.
This is repeated for an unlimited period of time with each weekday specified.
37
5.5.5.1 Programming the start time
1. Select Fast Temp pro in the device menu.
2. Select Once or Repeatedly.
This selection only appears as long as the FastTemp pro function has not already been activated. Otherwise it is only possible to edit or delete the programmed start time.
3. Only with Repeatedly: Activate/deactivate weekdays with menu/enter, select Continue and confirm with menu/enter.
4. Enter date and time for the one-time or repeated start of the temperature control run as well as the set temperature and confirm with menu/enter.
An overview of the current settings is displayed.
5. Edit the settings again or save.
6. Exit the menu.
FastTemp pro is now activated. In the display the symbol appears as long as an automatic start
of a temperature control run is still outstanding. In the standby mode Fast Temp pro is displayed.
• The temperature control run FastTemp (see p. 36) starts automatically at the set time.
• After a one-off programmed temperature control run, the following symbol is extinguished . With
several programmed temperature control runs, the FastTemp pro remains active until you deactivate it. To do this, select Fast Temp pro in the device menu and delete the settings.
5.5.5.2 Preparing the centrifuge
Ensure that the centrifuge is switched on or in the standby mode during the start time set and the rotor
and rotor lid are properly attached and the centrifuge lid is closed.
5.5.5.3 Automatic start of the temperature control cycle
1. If the centrifuge is in standby mode, it switches to the operating mode 1 min before the set start time.
2. At the start time the temperature control run FastTemp (see FastTemp on p. 36) begins. Fast Temp pro
appears in the display.
Automatically starting the temperature control cycle is not possible during centrifugation.
Page 38
38
Operation
Centrifuge 5430/5430 R English (EN)
5.5.6 Continuous cooling
If the rotor stops, the rotor chamber will be maintained at the target temperature if the following requirements have been met:
• The centrifuge is switched on.
• The centrifuge lid is closed.
• The target temperature is lower than the ambient temperature.
• The centrifuge is not in standby mode.
During continuous cooling the following applies:
• The set and actual temperature are displayed alternately.
• Irrespective of the set temperature, the temperature does not go below 4 °C to prevent the rotor chamber from freezing and from increased condensation in the device.
• The temperature adjustment is slower because the rotor does not rotate during this process.
To end continuous cooling, open the centrifuge lid or press the standby key.
If the centrifuge is not used for more than 8 hours, the continuous cooling is switched off automatically. The device then switches to standby mode. This protects against ice formation in the rotor chamber and increased condensation in the device. With FastTemp you can quickly reach the desired temperature again (see p. 36).
You can change the continuous cooling to endless operation at your own risk. To do so, in the device menu under Continuous cooling enable the (see p. 48) item.
Page 39
Operation
Centrifuge 5430/5430 R
English (EN)
5.6 Centrifuging
CAUTION! Risk from incorrectly-loaded rotors and damaged/overloaded tubes!
Before commencing centrifugation, follow the safety instructions relating to risks from
asymmetrically loaded and/or overloaded rotors and from overloaded, damaged and/or open tubes (see Warnings for intended use on p. 20).
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with rotor and rotor lid firmly tightened.If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Immediately press the start/stop key to stop centrifuging.
Each of the centrifuging variants described here must be preceded by the preparation described above (see Preparing for centrifugation on p. 32).
39
Only 5430 R: Please also note the instructions on cooling (see Cooling (only 5430 R) on p. 36).
5.6.1 Centrifugation with time setting
Perform the following steps in the sequence described.
1. Use the time arrow keys to set run time.
2. Only 5430 R: Use the temp arrow keys to adjust the temperature.
3. Use the speed arrow keys to the g-force (rcf)/speed.
4. Press start/stop to start centrifuging.
During centrifugation:
blinks in the display as long as the rotor is running.
• The remaining run time will be displayed in minutes. The last minute is counted down in seconds.
Only 5430 R: The current temperature will be displayed.
• The current g-force (rcf) or rotor speed is displayed.
• The shortcut keys, the , open and short keys, and all menu items which directly affect
centrifugation, are blocked during centrifugation.
End of centrifugation
• The centrifuge automatically stops after the set time has elapsed. The elapsed centrifugation will be
shown in a blinking display during the braking process. A signal tone sounds when the rotor is at a standstill.
Only 5430: The centrifuge lid opens automatically. The display shows the symbol .
Only 5430 R: The centrifuge lid remains closed to maintain the sample temperature. It can be
opened lid by pressing the blinking open key.
5. Remove the material for centrifuging.
• During the run you can modify the total run time, the temperature (only 5430 R), the speed and the rpm/rcf indication. The new parameters are adopted immediately. Please note that the shortest new total run time that can be set is the elapsed time plus 2 minuts.
• You can also terminate the centrifugation before the set run time has elapsed by pressing the start/stop key.
Page 40
40
Operation
Centrifuge 5430/5430 R English (EN)
5.6.2 Centrifuging in continuous operation
Perform the following steps in the sequence described.
1. Use the time arrow keys to set the continuous run.
The continuous operation function can be set above 99:59 h or below 30 seconds. The timer shows to indicate continuous operation.
2. Only 5430 R: Use the temp arrow keys to adjust the temperature.
3. Use the speed arrow keys to the g-force (rcf)/speed.
4. Press start/stop to start centrifuging.
blinks in the display as long as the rotor is running. Time is counted upwards, first in 30-second increments and then in minute increments from ten minutes.
5. Press start/stop to end centrifuging after the desired time period.
• The centrifugation time will be shown in a blinking display during the braking process.
• A signal tone sounds when the rotor is at a standstill.
Only 5430: The centrifuge lid opens automatically. The display shows the symbol .
Only 5430 R: The centrifuge lid remains closed to maintain the sample temperature. It can be opened lid by pressing the blinking open key.
6. Remove the material for centrifuging.
5.6.3 Short-spin centrifugation
You can carry out a short-spin cycle with the currently set or with the maximum g-force (rcf)/speed of the rotor used. This is set in the device menu (see Other menu items on p. 48) before executing the following steps in the sequence specified:
1. For short-spin cycle with the current g-force (rcf)/speed, set this directly with the arrow keys speed.
2. Only 5430 R: Use the temp arrow keys to adjust the temperature.
3. Start the short run: Hold the short key pressed down.
blinks in the display as long as the rotor is running.
• The time is counted upwards in seconds.
• During short run centrifuging all other keys are blocked. However, short run centrifuging is interrupted if another key is pressed simultaneously.
4. End short spin run: Release the short key.
• The centrifugation time will be shown in a blinking display during the braking process.
• A signal tone sounds when the rotor is at a standstill.
Only 5430: The centrifuge lid opens automatically. The display shows the symbol .
Only 5430 R: The centrifuge lid remains closed to maintain the sample temperature. It can be opened lid by pressing the blinking open key.
5. Remove the material for centrifuging.
During the braking process, you can restart the centrifugation up to two times by pressing the short key again.
The soft ramp does not work during short-spin centrifugation.
Page 41
Operation
Centrifuge 5430/5430 R
English (EN)
5.6.4 Removing the rotor
Swing-bucket rotors: remove the buckets before inserting and/or removing the rotor. Use
both hands to pick up the rotor cross.
Rotor F-35-6-30: only use the rotor removal tool supplied to insert or remove the rotor.
1. Turn the rotor nut counterclockwise using the supplied rotor key.
Rotor FA-45-24-11-HS: use special rotor key.
2. Remove the rotor vertically in an upward motion.
3. Only 5430 R: Switch off the centrifuge after use and empty the condensation water tray (remove it
from the left side of the device). Leave centrifuge lid fully opened and protect it against closing.
5.7 Standby mode
The centrifuge automatically changes from the ready state to the standby mode if the following prerequisites are met:
• The centrifuge was not used for the time set in the device menu (1 to 60 min)(see Settings on p. 49)
Only 5430 R: The centrifuge lid is open.
41
In the Standby mode, the following applies :
• The standby key lights red.
Only 5430 R: The rotor chamber is not cooled (see Continuous cooling on p. 38).
In the Ready state, the following applies:
• The centrifugation parameters are displayed.
• The standby key lights green.
Only 5430 R: The rotor chamber is cooled when the centrifuge lid is closed (see Continuous cooling on
p. 38).
You can switch between standby mode and ready state at any time when centrifugation is not performed by pressing the standby key.
Page 42
42
1. 2.
1. 2.
Operation
Centrifuge 5430/5430 R English (EN)
5.8 User instructions on rotors
5.8.1 Rotor F-35-6-30: rotor removal tool
5.8.1.1 Inserting the rotor
Prerequisites
The rotor nut is loose.
Use the rotor removal tool to insert the rotor in the centrifuge and remove it again.
1. Position the rotor removal tool with the narrow side on the rotor thread.
2. Tighten the rotor removal tool clockwise with approx. 3 revolutions.
3. Grip the rotor removal tool and insert the rotor.
4. Loosen and remove the rotor removal tool by turning it counterclockwise.
5. Inserting the rotor: Tighten the rotor using the enclosed rotor key (see Inserting the rotor on p. 32).
5.8.1.2 Loosening the rotor lid
Use the rotor removal tool to loosen a firmly tightened rotor lid screw.
1. Position the rotor removal tool with the wide side on the rotor lid screw.
2. Loosen the rotor lid screw by turning the rotor removal tool counterclockwise.
Page 43
Operation
Centrifuge 5430/5430 R
English (EN)
5.8.2 Rotor A-2-MTP
5.8.2.1 Inserting the rotor
NOTICE! If handled incorrectly, the rotor can fall over.
The rotor can fall over if the buckets are used as a handle.
Remove the buckets before inserting and/or removing a swing-bucket rotor.Always use both hands to carry the rotor cross.
5.8.2.2 Position and remove the upper shell of the wind shield
The upper shell of the wind shield is used for noise reduction.
1. Before it is used for the first time, remove the attachment that is mounted on the knob of the upper shell of the wind shield .
2. Position the upper shell of the wind shield and rotate a max. 1/4 revolution until it is lowered onto the rotor hub.
The lock indicator must protrude so far over the knob that the red marking is clearly visible:
43
The loose positioning of the upper shell is intentional and aids in self-centering.
3. Pull on the knob of the upper shell of the wind shield in order to lift it.
Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield in order to guarantee precise and quck temperature control of samples. Note that the centrifugation noise will increase slightly in this case. This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor lid.
5.8.3 Rotor FA-45-24-11-HS: using the special rotor key
5.8.3.1 Tightening the rotor
1. Insert the rotor key for rotor FA-45-24-11-HS in the rotor nut.
2. Turn the rotor key clockwise until it spins through ("click").
The rotor is correctly tightened.
5.8.3.2 Loosening the rotor
Turn the rotor nut and rotor key for Rotor FA-45-24-11-HS counterclockwise.
The rotor key for Rotor FA-45-24-11-HS can only be used to tighten and loosen this rotor. For the other rotors described in this operating manual, use the rotor key that is delivered with the Centrifuge 5430/5430 R.
Page 44
44
Operation
Centrifuge 5430/5430 R English (EN)
5.8.4 QuickLock
Aerosol-tight rotors FA-45-48-11, FA-45-30-11, FA-45-24-11-Kit, FA-45-16-17 and S-24-11-AT have a quick lock for the rotor lid (QuickLock).
Rotors FA-45-30-11 and FA-45-24-11-Kit have been redesigned and are now only available with a quick lock (QuickLock) instead of a rotor lid thread.
Closing the rotor lid
1. Check the correct positioning of the external sealing ring in the groove.
2. Place the rotor lid on the rotor in a vertical motion.
3. To lock the rotor, turn the red rotor lid screw clockwise as far as it will go, and after an audible "click" is heard.
The rotor is correctly locked after the audible "click" is heard!
If the locking system is stiff, lightly lubricate the pins in the rotor lid screw with pivot grease.
Page 45
Operating controls and function
Centrifuge 5430/5430 R
English (EN)

6 Operating controls and function

6.1 Device menu
Most menu levels contain the additional menu item Back / Zurück / Retour / Atrás. Detailed information on the menu can be found in the following chapters.
Tab. 6-1: Menu structure of the Centrifuge 5430/5430 R in four different menu languages.
English Deutsch Français Español Display
Programs
• Load program
•Save program
• Delete program
Soft ramp
•On
•Off
Key lock
•On
•Off
Programme
•Programm laden
•Programm speichern
•Programm löschen
Softrampe
•An
•Aus
Tastensperre
•An
•Aus
Programmes
•Charger prog.
• Enregistrer prog.
• Supprimer prog.
Rampe douce
•Marche
•Arrêt
Verrouilla. de touches
•Marche
•Arrêt
Programas
•Cargar programa
•Guardar programa
• Borrar programa
Rampa suave
• Encendido
•Apagado
Bloqueo del teclado
• Encendido
•Apagado
45
At set rpm
•On
•Off
Short spin
• Maximum speed
•Current speed
Continuous cooling (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•One-time-use
• Repeated use
Settings
•Display –S
andard display
t
– Extended display
•Alarm –On –Off
At set rpm
•An
•Aus
Short Spin
• Maximaler Wert
• Aktueller Wert
Dauerkühlung (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Einmal
•Mehrmals
Einstellungen
•Anzeige –Standardanzeige – Erweiterte
Anzeige
•Lautsprecher –An –Aus
At set rpm
•Marche
•Arrêt
Short Spin
• Vitesse max
• Vitesse actuelle
Refrigération continue (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Une fois
•Plusieurs fois
Réglages
•Affichage –Affichage
standard
–Affichage large
•Signal sonore –Marche –Arrêt
At set rpm
• Encendido
•Apagado
Short Spin
• Velocidad máximo
• Velocidad actual
Refrigeración continua (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Una vez
• Varias veces
Ajustes
•Indicador –Indicador
estándar
–Indicador
extendido
•Altavoz – Encendido –Apagado
Page 46
46
Operating controls and function
Centrifuge 5430/5430 R English (EN)
English Deutsch Français Español Display
•Volume –Cancel –Save –Default
• Date/Time • Datum/Uhrzeit • Date/Heure • Fecha/Hora
• Contrast –Cancel –Save –Default
• Language –English –Deutsch –Français –Español
•Standby –On –Off –Set time
•Lid release
(5430) –Automatic –Manual
• Lautstärke –Abbrechen –Speichern –Lieferzustand
•Kontrast –Abbrechen –Speichern –Lieferzustand
•Sprache –English –Deutsch –Français –Español
•Standby –An –Aus – Zeit einstellen
•Deckelöffnung
(5430) –Automatisch –Manuell
•Volume – Annuler – Enregistrer –Réglage usine
• Contraste – Annuler – Enregistrer –Réglage usine
•Langue –English –Deutsch –Français –Español
• Mise en veille –Marche –Arrêt –Réglage du temps
• Ouverture
couvercle (5430) –Automatique –Manuelle
•Volumen –Cancelar –Guardar – Est. de Fábrica
• Contraste –Cancelar –Guardar – Est. de Fábrica
•Idioma –English –Deutsch –Français –Español
• Standby – Encendido –Apagado –Establecer tiempo
• Apertura de la tapa
(5430) –Automática –Manual
5430: only for Centrifuge 5430 5430 R: only for Centrifuge 5430 R
Page 47
Operating controls and function
Centrifuge 5430/5430 R
English (EN)
6.2 Settings in the device menu
6.2.1 Programs
The Centrifuge 5430/5430 R has more than 50 programmable memory locations.
Load program Load the selected program. This appears in the display with number and name
and can be started immediately using the <>start/stop key. When selecting a program with a too high g-force (rcf)/speed for the rotor used, it flashes and a safety message appears.
Save program Save the set centrifugation parameters (centrifugation duration, temperature
(only 5430 R), g-force (rcf)/speed, soft ramp and At set rpm) under the selected number. In addition, you can assign a program name with a maximum of 20 characters. The centrifuging duration, the temperature (only 5430 R) and the g-force/rotational speed can still be changed in this menu using the corresponding arrow buttons time, temp (only 5430 R) and speed.
Delete program Delete the selected program.
Program numbers 1 to 5 cannot be deleted.
47
These functions are only available with the centrifuge at a standstill.
If the program memory is empty the menu item Delete program is exited automatically after the last program has been deleted. You will furthermore be unable to call up this menu item if the program memory is empty.
Program numbers which are already occupied can be overwritten.
6.2.2 Use program keys
You can also save and load Programs 1 to 5 directly by pressing the program keys:
6.2.2.1 Load program
Press the desired program key briefly.
The pressed program key illuminates in blue, the parameters are displayed.
By pressing again, you can exit the selected program again. The blue light of the key will then go out. The parameters of the most recent centrifugation are then displayed again.
6.2.2.2 Save program
1. Set the centrifugation parameters (centrifugation duration, temperature (only 5430 R), g-force (rcf)/ speed, soft ramp and At set rpm).
2. Press the desired program key for at least 2 seconds.
A signal tone sounds and the program key you pressed lights up.
The centrifuging parameters are saved under the appropriate program number (1-5).
Page 48
48
Operating controls and function
Centrifuge 5430/5430 R English (EN)
6.2.3 Other menu items
Menu item/meaning Setting Function Display
Soft ramp
Reduce speed of acceleration and braking ramp. Not used for Short Spin centrifuging.
Key lock
Set the current centrifugation parameters permanently to prevent the time, temperature (only 5430 R), g-force (rcf) or speed, soft ramp and At set rpm from being unintentionally modified.
At set rpm
Set start of centrifuging run time.
Short Spin
Before the start of a short run (see Short-spin centrifugation on p. 40) it is possible to switch between the maximum and currently set g-force (rcf) or speed. Soft ramp is not used for Short Spin centrifuging.
Continuous cooling (only 5430 R) Time limitation of continuous cooling (see p. 38) . Continuous cooling is only activated when the rotor is stopped and the centrifuge lid is closed.
Fast Temp pro (only 5430 R) Time and temperature programming for the automatic temperature control run. The selection Once/Repeatedly only appears if no Fast Temp pro has been set (see FastTemp pro on p. 37).
on Rotor accelerates and brakes slowly.
off Rotor accelerates and brakes rapidly.
on Set the centrifugation parameters
permanently.
off Release the permanent settings.
on The set time is counted down only
once 95% of the specified g-force (rcf) or speed has been reached.
off The set time is counted down
immediately.
Maximum value
Current value
8 h Preset value.
Endless operation of continuous
Once Set the date and time for the start of a
Repeatedly Set the week days and the common
Short-spin run at maximum g-force (rcf) or speed of the rotor used.
Short run at set g-force (rcf)/speed.
cooling. Caution! Icing possible! Set at own risk!
temperature control run.
start time for several repeated temperature control runs. Fast Temp pro is active from the date set.
Page 49
Operating controls and function
Centrifuge 5430/5430 R
English (EN)
6.2.4 Settings
Menu item/meaning Setting Function Display
Display
Select standard display or extended display of centrifuging parameters.
Loudspeaker
Switch loudspeaker on and off. In the event of error messages, a signal tone sounds even if the loudspeaker is switched off.
Volume
Adjust the speaker volume using the menu
arrow keys and in 5 stages. The signal tone for error messages is always issued at least at medium volume.
Date/time
Set date and time. The system does not switch automatically between summer and winter time.
Contrast
Adjust the display contrast using the menu
arrow keys and .
Language Set menu language (English, Deutsch,
Standby
Switch standby mode on and off. If the centrifuge is not used during the set time period and no continuous cooling takes place (only 5430 R), it switches to the standby mode (see p. 41).
Lid release (only 5430) Automatic Lid opens automatically at the end of
Standard display
Extended display
On Switch on loudspeaker.
Off Switch off loudspeaker.
Cancel Exit menu item without saving.
Save Save volume just set.
Default Restore default volume.
Cancel Exit menu item without saving.
Save Save the contrast just set.
Default Restore default contrast.
On Switch on standby mode.
Off Switch off standby mode.
Set time Using the arrow keys, set the time
Manual Lid remains closed at the end of
If the centrifuge is at rest the target values are displayed, and during centrifugation the actual values of run time, temperature (only 5430 R) and the g-force (rcf)/speed.
In addition to the standard display, specified values are always shown at the bottom of the display.
In the date display, set year (YYYY), month (MM) and day (DD). In the time display, set hours (hh) and minutes (mm). Before setting the clock time the time format is selected (12 h / 24 h).
Français or Español) . (see Set menu language on p. 30).
after which the centrifuge should automatically change to the standby mode (1 to 60 min).
centrifuging when the rotor stops.
centrifuging when the rotor stops and can be opened using the now flashing key open.
49
Page 50
50
Maintenance
Centrifuge 5430/5430 R English (EN)

7 Maintenance

7.1 Prepare cleaning/disinfection
Clean all accessible surfaces of the device and the accessories at least weekly and when contaminated.
Clean the rotor regularly. This way the rotor is protected and the durability is prolonged.
Furthermore, observe the notes on decontamination (see Decontamination before shipment on p. 53)
when the device is sent to the authorized Technical Service for repairs.
The procedure described in the following chapter applies to the cleaning as well as to the disinfection or decontamination. The table below describes the steps required on top of this:
Cleaning Disinfecting/decontamination
1. Use a mild cleaning fluid to clean the accessible surfaces of the device and the accessories.
2. Carry out the cleaning as described in the following chapter.
1. Choose the disinfection method which corresponds to the legal regulations and guidelines in place for your range of application. For example, use alcohol (ethanol, isopropanol) or alcohol-based disinfectants.
2. Carry out the disinfection or decontamination as described in the following chapter.
3. Then clean the device and the accessories.
If you have any further questions regarding the cleaning and disinfection or decontamination or regarding the cleaning fluid to be used, contact the Eppendorf AG Application Support. The contact details are provided on the back of this manual.
7.2 Cleaning/disinfection
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.Do not spray clean/spray disinfect the housing.Only plug the device back in if it is completely dry, both inside and outside.
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, clean it immediately using a
mild cleaning agent.
NOTICE! Corrosion due to aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.Do not incubate the accessories in aggressive cleaning agents or disinfectants for longer
periods.
Page 51
Maintenance
Centrifuge 5430/5430 R
English (EN)
NOTICE! Damage from UV and other high-energy radiation.
Do not use UV, beta, gamma, or any other high-energy radiation for disinfecting.Avoid storage in areas with strong UV radiation
Autoclaving
All rotors, rotor lids and adapters, except the A-2-MTP rotor, can be autoclaved (121 °C, 20 min). Replace the lids of the aerosol-tight rotors after a maximum of 50 autoclaving cycles. Only QuickLock rotor lid: Only the seal needs to be replaced after a maximum of 50 autoclaving cycle.
Swing-bucket rotor
• Before cleaning the rotor, remove old pivot grease from grooves and pivots.
• Make sure that the grooves and pivots are clean. Dirty grooves and pivots prevent carriers from swinging out evenly.
• After cleaning, lubricate the pivots of the rotor and the grooves of the buckets with pivot grease (order no. Int.: 5810 350.050/North America: 022634330) so that the carriers can move freely in a swinging manner.
51
7.2.1 Cleaning and disinfecting the device
1. Open the lid. Switch off the device with the mains/power switch. Disconnect the power plug from the power supply.
2. Loosen the rotor nut by turning the rotor key counterclockwise.
3. Remove the rotor.
4. Clean and disinfect all accessible surfaces of the device, including the power cable, using a damp cloth and the recommended cleaning agents.
5. Thoroughly clean the rubber seals of the rotor chamber with water.
6. Rub the dry rubber seal with glycerine or talcum powder to prevent it from becoming brittle. Other components of the device, such as the lid latch, lid springs, motor shaft and rotor cone, must not be lubricated.
7. Clean the motor shaft with a soft, dry and lint-free cloth. Do not lubricate the motor shaft.
8. Check the motor shaft for damage.
9. Inspect the device for corrosion and damage.
10. Leave the centrifuge lid open when the device is not being used.
11. Only connect the device to the power supply if it is fully dry inside and out.
7.2.2 Cleaning and disinfecting the rotor
1. Inspect the rotor and accessories for damage and corrosion. Do not use any damaged rotors or accessories.
2. Clean and disinfect the rotors and accessories with the recommended cleaning agents.
3. Use a bottle brush to clean and disinfect the rotor bores.
Page 52
52
Maintenance
Centrifuge 5430/5430 R English (EN)
4. Clean and disinfect the rotor lids. ONLY QuickLock: It is imperative that the sealing ring be removed during this process so that the groove below it can be thoroughly cleaned.
5. Rinse the rotors and accessories thoroughly with distilled water. Rinse the rotor bores of fixed-angle rotors particularly thoroughly.
Do not immerse the rotor in liquid as liquid can get trapped inside the cavities.
6. Place rotors and accessories on a cloth to dry. Place fixed-angle rotors with the rotor bores facing downwards to allow the bores to also dry.
7. Correctly reinsert the rotor lid sealing ring in the clean and dry groove.
8. Clean the rotor cone with a soft, dry and lint-free cloth. Do not lubricate the rotor cone.
9. Inspect the rotor cone for damage.
10. Place the dry rotor onto the motor shaft.
11. Tighten the rotor nut firmly by turning it clockwise with the rotor key.
12. Load the fixed-angle rotor with the cleaned adapters or the swing-bucket rotor with the cleaned buckets and adapters, if necessary.
13. Leave the rotor lid open when the rotor is not being used.
7.3 Additional service instructions for Centrifuge 5430 R
Empty and clean the condensation water tray regularly and especially after liquid spillage in the rotor
chamber. Pull out the tray for condensation water from the left side of the centrifuge.
Clean the condensation water drain on a regular basis, e.g., using a bottle brush.
Regularly free the rotor chamber ice formations via thawing, by leaving the centrifuge lid open or
carrying out a short temperature control run at approx. 30 °C.
Leave the centrifuge lid open when not in use for a long period.
Wipe up condensate in the rotor chamber using a soft, absorbent cloth.
Remove dust deposits from the ventilation slits of the centrifuge using a brush or swab at the latest
every six months. First switch off the device and remove the power plug.
Regularly check the gas spring of the centrifuge lid for proper functioning.
A defective gas spring is an insufficient support for the centrifuge lid and could cause injury if the centrifuge lid falls down. We recommend that the gas spring be replaced by a service technician every 2 years.
Page 53
Maintenance
Centrifuge 5430/5430 R
English (EN)
7.4 Glass breakage
When using glass tubes there is a risk of glass breakage in the rotor chamber. The resulting glass splinters are swirled around in the rotor chamber during centrifugation and have a sandblasting effect on the rotor and accessories. The smallest glass particles become lodged in the rubber parts (e.g., the motor guide, the rotor chamber gasket, and the rubber mats of adapters).
NOTICE! Glass breakage in the rotor chamber
Glass tubes in the rotor chamber may break if the g-force is too high. Broken glass can damage the rotor, accessories and samples.
Please note the manufacturer's information on the recommended centrifugation
parameters (load and speed).
Effects of glass breakage in the rotor chamber:
• Fine black metal abrasion in the rotor chamber (in metal rotor chambers)
• The surfaces of the rotor chamber and accessories are scratched.
• The chemical resistance of the rotor chamber is reduced.
• Contamination of samples
• Wear on rubber parts
53
How to proceed in case of glass breakage
1. Remove all splinters and glass powder from the rotor chamber and accessories.
2. Thoroughly clean the rotor and rotor chamber. Thoroughly clean the bores of the fixed-angle rotors, in particular.
3. If required, replace rubber mats and adapters to prevent any further damage.
4. Regularly check the rotor bores for deposits and damage.
7.5 Fuses
The fuse holder is located under the mains power socket.
1. Pull out the power plug.
2. Remove the fuse holder towards the rear. The two fuses can now be reached. The fuses can be replaced.
Instead of removable fuses, the Centrifuge 5430 R features a thermal overcurrent protective switch. The mains power switch jumps to the '0' switch setting if the overcurrent protective switch is triggered.
1. Use the mains power switch to switch the device back on after more than 20 s.
2. If the mains power switch returns to the '0' switch setting, contact Technical Support.
7.6 Decontamination before shipment
If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer for disposal please note the following:
WARNING! Risk to health from contaminated device
1. Follow the instructions in the decontamination certificate. It is available as a PDF file on our webpage (www.eppendorf.com/decontamination
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
).
Page 54
54
Trou bleshooting
Centrifuge 5430/5430 R English (EN)

8 Troubleshooting

If you cannot remedy an error with the recommended measures, please contact your local Eppendorf partner. The contact address can be found online at: www.eppendorf.com
.
8.1 General errors
Symptom/message Cause Remedy
No display. No power connection.
No display. Power failure.
Lid of the device cannot be opened.
Lid of the device cannot be opened.
Device cannot be started.
Device shakes when it starts up.
Centrifuge brakes during a short run centrifugation, although the short key is pressed.
Tempera ture display flashes. (only for 5430 R)
Rotor is still running.
Power failure. 1. Check the mains fuse of the device
Lid of the device is not closed.
Rotor loaded unsymmetrically. 1. Stop the device and load
The short key was released briefly more than twice (protective function for the drive).
Temperature deviation from the set value: ±3 °C.
Check mains/power connection.
Check the mains fuse of the device
(see Fuses on p. 53).
Check the mains fuse of the
laboratory.
Wait for rotor to stop.
(see Fuses on p. 53).
2. Check the mains fuse of the laboratory.
3. Activate the emergency lid release (see p. 57).
Close device lid.
symmetrically.
2. Restart device.
Press the short key continuously
during a short run centrifugation.
Check the settings.Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.
Page 55
Troubleshooting
Centrifuge 5430/5430 R
English (EN)
8.2 Error messages
If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. Press open key to clear the error message.
3. If necessary, repeat centrifugation.
Some errors can have various causes. The actual cause is described in the message in the device display.
Symptom/message Cause Remedy
Note A Lid latch
Note B Imbalance
Note C Rotor detection
Centrifuge lid could not be locked.
Rotor loaded unsymmetrically.
Set g-force/speed too high, e.g. after a rotor change (see Automatic rotor detection on p. 32).
Try again to close centrifuge lid.
Load the rotor symmetrically and
balance it.
1. Check the g-force/speed.
2. Repeat the run.
55
Symptom/message Cause Remedy
Error 1 Rotor detection
Error 2 Electronics fault
Error 3 Speed control
Error 3 Speed control
Error 5 Lid latch
Error 6 Drive fault
Error 6 Drive fault
Error 7 Speed control
Error 8 Speed control
Error 9 to Error 14 Electronics fault.
Rotor not detected.
Check rotor.If this error message appears again,
test with a different rotor.
Electronics fault.
Error in speed measuring system.
Error in speed measuring system.
Prohibited opening of lid or lid switch is defective during a run.
Drive fault.
Switch centrifuge off and back on
again after > 20 s.
Insert and tighten rotor.
Wait for displayed time.
1. Wait for rotor to stop.
2. Open and close again the lid of the device.
3. Repeat the run.
Repeat the run.If this error message appears again,
switch centrifuge off and back on again after > 20 s.
Drive overheated.
Major deviation in the speed control. 1. Wait for rotor to stop.
•Drive fault.
• Rotor loose.
•Incorrect rotor.
Allow the drive to cool down for at
least 15 min.
2. Tighten rotor.
1. Wait for rotor to stop.
2. Tighten rotor.
3. Repeat the run.
Switch centrifuge off and back on
again after > 20 s.
Page 56
56
Trou bleshooting
Centrifuge 5430/5430 R English (EN)
Symptom/message Cause Remedy
Error 16 to Error 17 Electronics fault
Error 18 Room Temp. of rotor chamber
(only 5430 R)
Error 18 Room Temp. of rotor chamber
(only 5430 R)
Error 22 Electronics fault
(only 5430 R)
Error 25 Mains/power failure
Error 26 Electronics fault
(only 5430 R)
Error 27 Electronics fault
(only 5430 R)
Error 28 Electronics fault
Error 30 Lid latch
Error 30 Lid latch
Error 30 Lid latch
Electronics fault. Switch centrifuge off and back on
again after > 20 s.
Temperature deviation from set value in the rotor chamber: ΔT > 16 °C.
Temperature deviation from set value in the rotor chamber: ΔT > 50 °C.
Electronics fault.
Mains/power failure during a run.
Electronics fault.
Electronics fault.
Electronics fault.
Centrifuge lid could not be locked.
Centrifuge lid could not be released.
Centrifuge lid has not been opened wide enough.
Allow the device to cool down and
repeat cycle.
Allow the device to cool down and
repeat cycle.
Switch centrifuge off and back on
again after > 20 s.
Check the power supply.
Switch centrifuge off and back on
again after > 20 s.
Switch centrifuge off and back on
again after > 20 s.
Press the open key.
Try again to close centrifuge lid.
Switch the device off and back on
again.
If the error occurs again:
1. Switch off the device.
2. Activate the emergency lid release (see Emergency release on p. 57).
Open the centrifuge lid wider by
hand.
Page 57
Troubleshooting
Centrifuge 5430/5430 R
English (EN)
8.3 Emergency release
If the centrifuge lid cannot be opened, you can activate the emergency release manually.
WARNING! Risk of injury from rotating rotor.
If the emergency release of the lid is operated, the rotor may continue rotating for several minutes.
Wait until the rotor stop before operating the emergency release.Check the monitoring glass in the centrifuge lid.
Use the rotor key delivered with the Centrifuge 5430 for the emergency release. The rotor key for the rotor FA-45-24-11-HS is not suitable for this purpose.
1. Pull the mains/power plug.
2. Carry out the following steps for the emergency release on both the left side and right side of the centrifuge (see Fig. 1 and Fig. 2).
3. Only 5430: Remove the plastic cover for the emergency release.
4. Insert the centrifuge rotor key in the rear hexagonal opening until a noticeable resistance is felt.
5. Slightly press and turn the rotor key counterclockwise five to ten revolutions, as depicted on the openings of the emergency release.
This will release the centrifuge lid.
6. Open the centrifuge lid.
7. Remove the rotor key and put the plastic covers back on (Centrifuge 5430).
57
Page 58
58
Transport, storage and disposal
Centrifuge 5430/5430 R English (EN)

9 Transport, storage and disposal

9.1 Transport
CAUTION! Risk of injury due to lifting and carrying heavy loads
The device is heavy. Lifting and carrying the device can lead to back injuries.
Transport and lift the device with an adequate number of helpers only.Use a transport aid to transport the device.
Remove the rotor from the centrifuge before transport.
Use the original packaging for transport.
Air temperature Relative humidity Atmospheric pressure
General transport -25 °C – 60 °C 10 % – 75 % 30 kPa – 106 kPa
Air freight -20 °C – 55 °C 10 % – 75 % 30 kPa – 106 kPa
9.2 Storage
Air temperature Relative humidity Atmospheric pressure
In transport packaging -25 °C – 55 °C 10 % – 75 % 70 kPa – 106 kPa
Without transport packaging
-5 °C – 45 °C 10 % – 75 % 70 kPa – 106 kPa
9.3 Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European Community:
Within the European Community, the disposal of electrical devices is regulated by national regulations based on EU Directive 2002/96/EC pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the business-to-business sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. To document this, they have been marked with the following identification:
Because disposal regulations may differ from one country to another within the EU, please contact your supplier if necessary.
Page 59

10 Technical data

10.1 Power supply
Centrifuge 5430
Mains connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz 100 V, 50 to 60 Hz
Current consumption 3 A (230 V)
6 A (120 V) 7 A (100 V)
Power consumption Maximum 475 W
EMC: interference emission (radio interference)
EMC: noise immunity EN 61326
Overvoltage category II
Protection class 1
Fuses 4.0 AT (230 V)
EN 61326 - category B
8.0 AT (120 V/100 V)
Technical data
Centrifuge 5430/5430 R
English (EN)
59
Centrifuge 5430 R
Mains connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz 100 V, 50 to 60 Hz
Current consumption 6 A (230 V)
12 A (120 V/ 100 V)
Power consumption Maximum 1050 W
EMC: Interference emission (radio interference)
EMC: Noise immunity EN 61326
Overvoltage category II
Protection class 1
Fuses Thermal overcurrent protective switch 7 A (230 V)
EN 61326 – Class B (230 V/120 V) EN 61326 – Class A (100 V)
Thermal overcurrent protective switch 15 A (120 V/100 V)
10.2 Ambient conditions
Ambience For indoor use only.
Ambient temperature Centrifuge 5430: 4 to 40°C
Centrifuge 5430 R: 10 to 40°C
Maximum relative humidity 75 %, non-condensing humidity
Atmospheric pressure Use up to an altitude of 2000 m above MSL.
Degree of pollution 2
Page 60
60
Technical data
Centrifuge 5430/5430 R English (EN)
10.3 Weight/dimensions
Centrifuge 5430
Dimensions Width: 335 mm (11.2 in.)
Depth: 415 mm (16.3 in.) Height: 250 mm (9.84 in.)
Weight without rotor 29 kg (63.9 lbs.)
Noise level < 60 dB(A) *
Centrifuge 5430 R
Dimensions Width: 380 mm (15.0 in.)
Depth: 640 mm (25.2 in.) Height: 296 mm (11.7 in.)
Weight without rotor 56 kg (123.5 lbs.)
Noise level < 60 dB(A) *
*) The noise level was measured frontally in a sound measuring room with accuracy class 1, 1 m from the device and at lab bench height.
10.4 Application parameters
Run time 30 s to 99:59 h, infinity ( )
Adjustable to 10 min in 0.5 min increments, then increments of 1 min
Temperature of 5430 R -11°C to 40°C
Relative centrifugal force (or rcf) 1 to 30.130 × g
Adjustable in increments of 10 × g up to 3,000 × g, then in increments of 100 × g
Speed 100 to 17,500 rpm
Adjustable in increments of 10 rpm up to 5,000 rpm, then in increments of 100 rpm
Maximum load 48 tubes with 2.0 mL each or 6 conical tubes with 50 mL each
Maximum kinetic energy 10,000 nm
Compulsory test log book No
Permitted density of the material for centrifuging at max. g-force/rpm and max. load
1.2 g/mL
Page 61
Technical data
Centrifuge 5430/5430 R
English (EN)
61
Rotor Lowest achievable temperature
-11°C set 23°C ambient temperature 60 min run time
FA-45-48-11 < 0 °C 12 700 rpm
F-45-48-11 < 0 °C 12 700 rpm
FA-45-30-11 < 0 °C 14 000 rpm
F-45-30-11 < 0 °C 14 000 rpm
FA-45-24-11-HS < 5 °C ? rpm
FA-45-24-11-Kit < 0 °C 13 200 rpm
F-45-64-5-PCR < 0 °C 11 800 rpm
F-45-18-17-Cryo < 0 °C 8 900 rpm
FA-45-16-17 < 0 °C 14 200 rpm
F-35-6-30 < 0 °C 7 8300 rpm
A-2-MTP < 0 °C 4 680 rpm
S-24-11-AT < 0 °C 12 700 rpm
Acceleration times and braking times of Centrifuge 5430/5430 R
Rotor Acceleration time/Deceleration time Without soft ramp
Acceleration time/Deceleration time With soft ramp
230 V 120 V 100 V
FA-45-48-11 F-45-48-11
FA-45-30-11 F-45-30-11
FA-45-24-11-Kit ≤ 15 / 16 s
FA-45-16-17 ≤ 20 / 20 s
S-24-11-AT ≤ 13 / 16 s
FA-45-24-11-HS ≤ 21 / 16 s
F-45-64-5-PCR ≤ 12 / 15 s
F-45-18-17-Cryo ≤ 8 / 11 s
F-35-6-30 ≤ 23 / 23 s
A-2-MTP ≤ 18 / 21 s
≤ 20 / 20 s ≤ 61 / 65 s (SOFT)
≤ 15 / 15 s ≤ 61 / 65 s (SOFT)
≤ 78 / 90 s (SOFT)
≤ 61 / 66 s (SOFT)
≤ 61 / 66 s (SOFT)
≤ 60 / 65 s (SOFT)
≤ 62 / 65 s (SOFT)
≤ 77 / 85 s (SOFT)
≤ 62 / 67 s (SOFT)
≤ 63 / 67 s (SOFT)
≤ 20 / 20 s ≤ 61 / 65 s (SOFT)
≤ 15 / 15 s ≤ 61 / 65 s (SOFT)
≤ 15 / 16 s ≤ 78 / 90 s (SOFT)
≤ 20 / 20 s ≤ 61 / 66 s (SOFT)
≤ 13 / 16 s ≤ 61 / 66 s (SOFT)
≤ 21 / 16 s ≤ 60 / 65 s (SOFT)
≤ 12 / 15 s ≤ 62 / 65 s (SOFT)
≤ 8 / 11 s ≤ 77 / 85 s (SOFT)
≤ 23 / 23 s ≤ 62 / 67 s (SOFT)
≤ 18 / 21 s ≤ 63 / 67 s (SOFT)
Speed to safely maintain 4°C sample temperature
4°C set 23 °C ambient temperature
≤ 27 / 20 s ≤ 61 / 65 s (SOFT)
≤ 20 / 15 s ≤ 61 / 65 s (SOFT)
≤ 20 / 16 s ≤ 78 / 90 s (SOFT)
≤ 30 / 20 s ≤ 61 / 66 s (SOFT)
≤ 16 / 16 s ≤ 61 / 66 s (SOFT)
≤ 30 / 16 s ≤ 60 / 65 s (SOFT)
≤ 15 / 15 s ≤ 62 / 65 s (SOFT)
≤ 8 / 11 s ≤ 77 / 85 s (SOFT)
≤ 27 / 27 s ≤ 62 / 67 s (SOFT)
≤ 18 / 21 s ≤ 63 / 67 s (SOFT)
Page 62
62
Ordering Information
Centrifuge 5430/5430 R English (EN)

11 Ordering Information

11.1 Centrifuge 5430
Order no. (International)
5427 000.011 022620533 230 V / 50 – 60 Hz – 022620584 120 V / 50 – 60 Hz, with US-plug
5427 000.216 022620525 230 V / 50 – 60 Hz – 022620509 120 V / 50 – 60 Hz, with US-plug
022620541 120 V / 50 – 60 Hz, with US-plug
022620568 120 V / 50 – 60 Hz, with US-plug
5427 000.615 022620540 230 V / 50 – 60 Hz – 022620596 120 V / 50 – 60 Hz, with US-plug
5427 000.410 - 230 V / 50 – 60 Hz – 022620511 120 V / 50 – 60 Hz, with US-plug
022620557 120 V / 50 – 60 Hz, with US-plug
022620572 120 V / 50 – 60 Hz, with US-plug
Order no. (North America)
Description
Centrifuge 5430, Keypad
without rotor
Centrifuge 5430, Keypad
with rotor FA-45-30-11 incl. rotor lid
Centrifuge 5430, Keypad
with rotor FA-45-24-11-Kit incl. rotor lid
Centrifuge 5430, Keypad
with rotor A-2-MTP incl. wind shield upper shell
Centrifuge 5430, Knobs
without rotor
Centrifuge 5430, Knobs
with rotor FA-45-30-11 incl. rotor lid
Centrifuge 5430, Knobs
with rotor FA-45-24-11-Kit incl. rotor lid
Centrifuge 5430, Knobs
with rotor A-2-MTP incl. wind shield upper shell
11.2 Centrifuge 5430 R
Order no. (International)
5428 000.210 022620678 230 V / 50 – 60 Hz – 022620667 120 V / 50 – 60 Hz, with US-plug
5428 000.015 022620612 230 V / 50 – 60 Hz – 022620601 120 V / 50 – 60 Hz, with US-plug
Order no. (North America)
Description
Centrifuge 5430 R, Keypad
without rotor
Centrifuge 5430 R, Keypad
with rotor FA-45-30-11 incl. rotor lid
Page 63
Ordering Information
Centrifuge 5430/5430 R
English (EN)
63
Order no. (International)
022620645 120 V / 50 – 60 Hz, with US-plug
5428 000.619 022620690 230 V / 50 – 60 Hz – 022620689 120 V / 50 – 60 Hz, with US-plug
5428 000.414 022620634 230 V / 50 – 60 Hz – 022620623 120 V / 50 – 60 Hz, with US-plug
022620656 120 V / 50 – 60 Hz, with US-plug
Order no. (North America)
Description
Centrifuge 5430 R, Keypad
with rotor A-2-MTP incl. wind shield upper shell
Centrifuge 5430 R, Knobs
without rotor
Centrifuge 5430 R, Knobs
with rotor FA-45-30-11 incl. rotor lid
Centrifuge 5430 R, Knobs
with rotor A-2-MTP incl. wind shield upper shell
11.3 Rotors, rotor lids and seals
11.3.1 Rotors with QuickLock rotor lid
Rotor FA-45-48-11
Order no. (International)
5427 754.008 5427754008 aerosol-tight, angle 45°, 48 places, max. tube diameter 11 mm,
5427 762.000 5427762000 aerosol-tight, aluminum
5409 718.002 5409718002 set of 5 pieces
Order no. (North America)
Description
Fixed-angle rotor FA-45-48-11
incl. rotor lid (aluminum)
Rotor lid for FA-45-48-11
Seal for rotor lid
FA-45-48-11 (5427 R/5430/5430 R)
Rotor FA-45-30-11
Order no. (International)
5427 753.001 5427753001 aerosol-tight, angle 45°, 30 places, max. tube diameter 11 mm,
5427 761.004 5427761004 aerosol-tight, aluminum
5820 762.004 5820762004 Set of 5 pieces
Order no. (North America)
Description
Fixed-angle FA-45-30-11
incl. rotor lid (aluminum)
Rotor lid for FA-45-30-11
Seal for rotor lid
FA-45-30-11 (5427 R/5430/5430 R)
Page 64
64
Ordering Information
Centrifuge 5430/5430 R English (EN)
Rotor FA-45-24-11 Kit
Order no. (International)
5427 752.005 5427752005 aerosol-tight, angle 45°, 24 places, max. tube diameter 11 mm,
5427 760.008 5427760008 aerosol-tight, aluminum
5820 767.006 5820767006 Set of 5 pieces
Rotor FA-45-16-17
Order no. (International)
5427 750.002 5427750002 aerosol-tight, angle 45°, 16 places, max. tube diameter 17 mm,
5427 751.009 5427751009 aerosol-tight, aluminum
5409 717.006 5409717006 Set of 5 pieces
Order no. (North America)
Order no. (North America)
Description
Fixed-angle rotor FA-45-24-11-Kit
incl. rotor lid (aluminum)
Rotor lid for FA-45-24-11-Kit
Seal for rotor lid
FA-45-24-11-Kit (5427 R/5430/5430 R), FA-45-48-11, FA-45-20-17 (5804/5804 R/5810/5810 R)
Description
Fixed-angle rotor FA-45-16-17
incl. rotor lid (aluminum)
Rotor lid for FA-45-16-17
Seal for rotor lid
FA-45-24-11 (5427 R), FA-45-16-17 (5430/5430 R)
Rotor S-24-11-AT
Order no. (International)
5427 757.007 5427757007 aerosol-tight, steel, angle 90°, 24 places, max. tube diameter
5427 758.003 5427758003 aerosol-tight, aluminum
5409 719.009 5409719009 set of 5 pieces
5409 721.003 5409721003 set of 2 pieces
Order no. (North America)
Description
Swing-bucket rotor S-24-11-AT
11 mm, incl. rotor lid (aluminum)
Rotor lid for S-24-11-AT
Seal for rotor lid
S-24-11-AT (5427 R/5430/5430 R)
Tube holder for S-24-11-AT
for 4 × 1,5 mL/2,0 mL Eppendorf tubes
Page 65
11.3.2 Rotors with rotor lid thread
Rotor F-45-48-11
Ordering Information
Centrifuge 5430/5430 R
English (EN)
65
Order no. (International)
5409 712.004 5409712004 aluminum, angle 45°, 48 places, max. tube diameter 11 mm,
5409 713.000 5409713000 Polypropylene
Rotor FA-45-30-11
Order no. (International)
5427 719.008 022654063 aerosol-tight, PTFE-coated, aluminum
Rotor F-45-30-11
Order no. (International)
5427 712.003 022654004 PTFE-coated, angle 45°, 30 places, max. tube diameter
5427 718.001 022654021 Polypropylene
Order no. (North America)
Order no. (North America)
Order no. (North America)
Description
Rotor F-45-48-11
incl. rotor lid (polypropylene)
Rotor lid for F-45-48-11
Description
Rotor lid
for F-45-30-11
Description
Rotor F-45-30-11
11 mm, incl. rotor lid (polypropylene)
Rotor lid
for F-45-30-11
Rotor FA-45-24-11-HS
Order no. (International)
5427 710.000 022654080 aerosol-tight, PTFE-coated, angle 45°, 24 places, max. tube
5427 711.007 022654101 aerosol-tight, PTFE-coated, aluminum
Order no. (North America)
Description
Rotor FA-45-24-11-HS
diameter 11 mm, incl. rotor lid (aluminum), incl. rotor key
Rotor lid
for FA-45-24-11-HS
Page 66
66
Ordering Information
Centrifuge 5430/5430 R English (EN)
Rotor FA-45-24-11-Kit
Order no. (International)
5427 704.000 022654144 aerosol-tight, aluminum
Rotor F-45-64-5-PCR
Order no. (International)
5427 714.006 022654209 angle 45°, 64 places, max. tube diameter 5 mm, incl. rotor lid
5427 720.006 022654225 aluminum
Rotor F-45-18-17-Cryo
Order no. (International)
5427 705.007 022654161 angle 45°, 18 places, max. tube diameter 17 mm, incl. rotor lid
5427 707.000 022654187 Polypropylene
Order no. (North America)
Order no. (North America)
Order no. (North America)
Description
Rotor lid
for FA-45-24-11-Kit
Description
Rotor F-45-64-5-PCR
(aluminum) and adapters
Rotor lid
for F-45-64-5-PCR
Description
Rotor F-45-18-17-Cryo
(polyproplene) and adapters
Rotor lid
for F-45-18-17-Cryo
Rotor F-35-6-30
Order no. (International)
5427 716.009 022654306 angle 35°, 6 places, max. tube diameter 30 mm, incl. rotor lid
5427 739.009 5427739009 angle 35°, 6 places, max. tube diameter 30 mm, incl. rotor lid
5427 715.002 022654322 aluminum
Order no. (North America)
Description
Rotor F-35-6-30
(aluminum) and adapters for 15/50 mL conical tubes
Rotor lid
for F-35-6-30
Page 67
11.3.3 Rotors with rotor lid for attaching
A-2-MTP rotor
Ordering Information
Centrifuge 5430/5430 R
English (EN)
67
Order no. (International)
5427 700.005 022634403 with 2 buckets and windshield upper shell
5427 722.009 022634420 Set of 2
5427 725.008 022654446 aluminum
The Aerosol tightness is tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK).
Order no. (North America)
Description
Rotor A-2-MTP
Spare MTP buckets for A-2-MTP
Wind shield upper shell for A-2-MTP
11.4 Accessories
11.4.1 Adapter
Order no. (International)
5425 715.005 022636260 for 1 PCR tube (0.2 mL, max. Ø 6 mm), set of 6 5425 717.008 022636243 for 1 sample tube (0.4 mL, max. Ø 6 mm), set of 6 5425 716.001 022636227 for 1 sample tube (0.5 mL, max. Ø 6 mm) or 1 Microtainer
5427 717.005 022654241 for PCR strips, set of 4 pieces
5702 752.002 022639498 for cryo tubes (max. Ø 13 mm) and sealable centrifuge tubes
5427 708.006 5427708006 for 1.5 mL HPLC vials, 18 pieces
5427 740.007 5427740007 13 × 65-89, set of 2 5427 741.003 5427741003 13 × 90-110, set of 2 5427 746.005 5427746005 for Eppendorf Tubes 5.0 mL, set of 2 5427 726.004 022654365 for 15 mL conical tubes, set of 2 5427 732.004 022654512 for 7 - 15 mL round-bottom tubes and blood collection tubes,
5427 735.003 022654538 for 9 - 15 mL round-bottom tubes and blood collection tubes,
5427 742.000 5427742000 13 × 65-89, set of 2
Order no. (North America)
Description
Adapter
used in FA-45-48-11, F-45-48-11, FA-45-30-11, F-45-30-11, FA-45-24-11-HS and FA-45-24-11-Kit
(0.6 mL, max. Ø 8 mm), set of 6
Adapter
used in F-45-64-5-PCR
Adapter
used in F-45-18-17-Cryo
(max. Ø 12.2 mm), max. length 50 mm, set of 6
Adapter
used in F-35-6-30, small tube bore
set of 2
set of 2
Adapter
used in F-35-6-30, large tube bore
Page 68
68
Ordering Information
Centrifuge 5430/5430 R English (EN)
Order no. (International)
5427 743.006 5427743006 13 × 90-110, set of 2 5427 747.001 5427747001 for Eppendorf Tubes 5.0 mL, set of 2 5427 727.000 022654349 for 50 mL conical tubes, set of 2 5427 723.005 022654331 for Centriplus centrifugal filter units, set of 6 5427 734.007 022654524 for 7 - 15 mL round-bottom tubes and blood collection tubes,
5427 738.002 022654545 for 9 - 15 mL round-bottom tubes and blood collection tubes,
5427 736.000 022654556 for 20 - 30 mL round-bottom tubes, set of 2 5427 737.006 022654567 for 50 mL round-bottom tubes, set of 2
5825 711.009 022638947 for 96-well PCR plates, set of 2 5825 713.001 022638955 for 384-well PCR plates, set of 2 5825 706.005 022638963 CombiSlide Adapter, set of 2
Order no. (North America)
Description
set of 2
set of 2
Adapter
used in A-2-MTP
11.4.2 Other accessories
Order no. (International)
5416 301.001 022634305 Standard 5427 730.001 5427730001 for rotor FA-45-24-11-HS
5427 728.007 5427728007
5810 350.050 022634330 Tube 20 mL
5428 850.418 022680452
Order no. (North America)
Description
Rotor key
Rotor removal tool for Rotor F-35-6-30
Pivot grease
Tray for condensation water
11.4.3 Fuses for Centrifuge 5430
Order no. (International)
5301 850.249 022654403 4 A (230 V), 2 pieces 5427 850.341 022654381 8.0 AT UL (120 V/100 V), 2 pieces
Order no. (North America)
Description
Fuse
Page 69
Centrifuge 5430/5430 Rseeon p.Fig.Tab.p.
Operating manual

Declarations and Certificates

Declarations and Certificates
Centrifuge 5430/5430 R
69
Page 70
Page 71
Page 72
Page 73
Page 74
Page 75
Page 76
Page 77
Page 78
Page 79
Page 80
Page 81
Page 82
Page 83
Declarations and Certificates
Centrifuge 5430/5430 R
83
Page 84
84
Declarations and Certificates
Centrifuge 5430/5430 R
Page 85

Annex

Centrifuge 5430/5430 R
Annex

Menu structure of Centrifuge 5430/5430 R

English Deutsch Français Español Display
Programs
• Load program
•Save program
• Delete program
Soft ramp
•On
•Off
Key lock
•On
•Off
Programme
•Programm laden
•Programm speichern
•Programm löschen
Softrampe
•An
•Aus
Tastensperre
•An
•Aus
Programmes
•Charger prog.
• Enregistrer prog.
• Supprimer prog.
Rampe douce
•Marche
•Arrêt
Verrouilla. de touches
•Marche
•Arrêt
Programas
•Cargar programa
•Guardar programa
• Borrar programa
Rampa suave
• Encendido
•Apagado
Bloqueo del teclado
• Encendido
•Apagado
85
At set rpm
•On
•Off
Short spin
• Maximum speed
•Current speed
Continuous cooling (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•One-time-use
• Repeated use
Settings
•D
i
splay –Standard display – Extended display
•Alarm –On –Off
At set rpm
•An
•Aus
Short Spin
• Maximaler Wert
• Aktueller Wert
Dauerkühlung (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Einmal
•Mehrmals
Einstellungen
•Anzeige –Standardanzeige – Erweiterte
Anzeige
•Lautsprecher –An –Aus
At set rpm
•Marche
•Arrêt
Short Spin
• Vitesse max
• Vitesse actuelle
Refrigération continue (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Une fois
•Plusieurs fois
Réglages
•Affichage –Affichage
standard
–Affichage large
•Signal sonore –Marche –Arrêt
At set rpm
• Encendido
•Apagado
Short Spin
• Velocidad máximo
• Velocidad actual
Refrigeración continua (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Una vez
• Varias veces
Ajustes
•Indicador –Indicador
estándar
–Indicador
extendido
•Altavoz – Encendido –Apagado
•Volume –Cancel –Save –Default
• Date/Time • Datum/Uhrzeit • Date/Heure • Fecha/Hora
• Lautstärke –Abbrechen –Speichern –Lieferzustand
•Volume – Annuler – Enregistrer –Réglage usine
•Volumen –Cancelar –Guardar – Est. de Fábrica
Page 86
86
Annex
Centrifuge 5430/5430 R
English Deutsch Français Español Display
• Contrast –Cancel –Save –Default
• Language –English –Deutsch –Français –Español
•Standby –On –Off –Set time
•Lid release
(5430) –Automatic –Manual
•Kontrast –Abbrechen –Speichern –Lieferzustand
•Sprache –English –Deutsch –Français –Español
•Standby –An –Aus – Zeit einstellen
•Deckelöffnung
(5430) –Automatisch –Manuell
• Contraste – Annuler – Enregistrer –Réglage usine
•Langue –English –Deutsch –Français –Español
• Mise en veille –Marche –Arrêt –Réglage du temps
• Ouverture
couvercle (5430) –Automatique –Manuelle
• Contraste –Cancelar –Guardar – Est. de Fábrica
•Idioma –English –Deutsch –Français –Español
• Standby – Encendido –Apagado –Establecer tiempo
• Apertura de la tapa
(5430) –Automática –Manual

Rotor F-35-6-30: Adapters for round-bottom tubes and blood taking systems

Adapter
(1)
Bottom shape
Tube dimensions: Ø × length (min to max)
(2)
Rotor bore
r
max
Max. g-force (rcf)
Ordering No. (International)
Ordering No. (North America)
2.6 – 7 mL round 13 × (65 to 89) mm small 9.4 cm 6,443 × g 5427 740.007 5427740007
2.6 – 7 mL round 13 × (65 to 89) mm large 8.9 cm 6,100 × g 5427 742.000 5427742000
2.6 – 7 mL round 13 × (90 to 110) mm small 10.9 cm 7,471 × g 5427 741.003 5427741003
2.6 – 7 mL round 13 × (90 to 110) mm large 10.4 cm 7,129 × g 5427 743.006 5427743006
5 mL conical only 5 mL tubes small 9.1 cm 6,237 × g 5427 746.005 5427746005
5 mL conical only 5 mL tubes large 10.4 cm 7,129 × g 5427 747.001 5427747001
7 – 15 mL round 16 × (74 to 103) mm small 10.2 cm 7,005 × g 5427 732.004 022654512
7 – 15 mL round 16 × (85 to 115) mm large 10.2 cm 7,005 × g 5427 734.007 022654524
9 – 15 mL round 16.8 × (84 to 125) mmsmall 11.3 cm 7,745 × g 5427 735.003 022654538
9 – 15 mL round 16.8 × (84 to 125) mmlarge 10.8 cm 7,403 × g 5427 738.002 022654545
20 – 30 mL round 26 × (97 to 110) mm large 10.4 cm 7,087 × g 5427 736.000 022654556
50 mL round 29 × (100 to 125)
large 11.1 cm 7,581 × g 5427 737.006 022654567
mm
1) One tube per adapter, 6 per rotor. For the diameters 13, 16, and 16.8 mm, 12 tubes of the same type can be simultaneously centrifuged in 6 of each small and large adapters.
2) Min. tube length below cap rim to max. tube length incl. cap.
Page 87
Page 88
Evaluate your manual
Give us your feedback. www.eppendorf.com/manualfeedback
Your local distributor: www.eppendorf.com/contact
Eppendorf AG · 22331 Hamburg · Germany eppendorf@eppendorf.com · www.eppendorf.com
Loading...