eppendorf is a registered trademark of Eppendorf AG, Hamburg, Germany.
Microtainer is a registered trademark of Becton Dickinson, Franklin Lakes, NJ, USA.
Trademarks are not marked in all cases with ™ or
®
in this manual.
Your Online Product Registration
Register your Eppendorf instruments online and get rewarded!
Learn more: www.eppendorf.com/myeppendorf
Please read this operating manual completely before using the device for the first time.
Please view this operating manual as part of the product and keep it somewhere easily
accessible.
When passing the device on to third parties, be sure to include this operating manual.
If this manual is lost, please request another one. The current version can be found on our
website www.eppendorf.com.
The Centrifuge 5424 / 5424 R is available in two versions: keypad or rotary knobs. This
operating manual generally describes how to operate the keypad version. However, it also
applies to the rotary knob version.
BiohazardExplosion
EN
Operating manual
Electric shockCrushing
Hazard pointMaterial damage
1.2.2Degrees of danger
The degree of danger is a part of a safety note and distinguishes the possible results of
non-observance from each other.
DANGERWill lead to severe injuries or death.
WARNINGMay lead to severe injuries or death.
CAUTIONMay lead to light to moderate injuries.
NOTICEMay lead to material damage.
1.3Abbreviations used
PCRPolymerase chain reaction
PTFEPolytetrafluorethylen
rcfRelative centrifugal force
rpmRevolutions per
UVUltraviolet radiation
minute
7
1
3
4
2
1
5
6
7
2
8
9
1
3
4
10
11
2
1
5
7
6
9
8
2
EN
Operating manual
Centrifuge 5424 / 5424 R — Operating manual
2 Product description
2Product description
2.1Main illustration
The depiction of the Centrifuge 5424 / 5424 R can also be found on the front fold-out page (see Fig. 1 and Fig. 2).
Abb. 1:Front and rear view of the Ce ntrifuge 5424
Abb. 2:Front and rear view of the Ce ntrifuge 5424 R
Fig. 2:Front and rear view of the Centrifuge 5424 R
1 Centrifuge lid2 Monitoring glass
Visual control for rotor stop or option for speed control
via stroboscope.
3Display
Depiction of the current centrifuging parameters and
device settings (see Fig. 4 on p. 19).
4 Control panel
Keys and dials (dependent on the device version) for
operating the centrifuge (see Fig. 3 on p. 18).
5 Emergency lid release
6ID plate
(see Emergency release on p. 32)
7 Mains switch
Switch for switching the device on and off.
Switch position 0: The device is switched off.
Switch position I: The device is switched on.
8 Mains connection
Connection socket for the mains cable supplied.
9 Fuse holder10 USB port (only 5424 R)
Interface for error analysis and software updates by
the Technical Service.
11 Condensation water tray (only 5424 R)
8
Centrifuge 5424 / 5424 R — Operating manual
2 Product description
2.2Delivery package
EN
2.2.1Centrifuge 5424
Quantity Order No.
(International)
1--
1
or
15416 301.001022634305
15703 350.102022639609
1--Power cable
1
1
2.2.2Centrifuge 5424 R
Quantity Order No.
1--
1
or
15416 301.001022634305
1--Power cable
15404 850.0855404850085Tray for condensation water
1
1
5424 852.122
5424 852.130
5404 900.023
5404 900.031
(International)
5424 852.122
5424 852.130
5404 900.023
5404 900.031
Order No.
(North America)
950004266
950004240
5404900023
5404900031
Order No.
(North America)
950004266
950004240
5404900023
5404900031
Description
Centrifuge 5424
See chapter Ordering Information for corresponding device version,
equipment and order number
Fuses
3.15 AT for 230 V, 2 pieces
6.3 AT for 120 V/100 V, 2 pieces
Rotor key
Standard
Captain Eppi rotor key holder
1 piece
Operating manual Centrifuge 5424/5424 R
Languages: EN, DE, FR, ES, IT, PT
Languages: DA, FI, EL, NL, SV (230 V devices only)
Description
Centrifuge 5424 R
See chapter Ordering Information for corresponding device version,
equipment and order number
Fuses
3.15 AT for 230 V, 2 pieces
6.3 AT for 120 V/100 V, 2 pieces
Rotor key
Standard
Operating manual Centrifuge 5424/5424 R
Languages: EN, DE, FR, ES, IT, PT
Languages: DA, FI, EL, NL, SV (230 V devices only)
Operating manual
2.3Features
The versatile Centrifuge 5424 / 5424 R has a capacity of 24 x 2 mL and reaches a maximum of
21,130 x g / 15,000 rpm (5424: 120 V, 5424 R: 230 V, 120 V, 100 V) or 20,238 x g / 14,680 rpm
(5424: 230 V, 100 V). You can select between four different rotors to centrifuge the following
tubes for your various applications:
• Micro test tubes (0.2 to 2.0 mL)
• PCR strips
• Microtainers (0.6 mL)
• Spin columns (1.5/2.0 mL)
The Centrifuge 5424 R has an additional temperature control function for centrifugation between
-10°C and +40°C. The fast temp function can be used to start a temperature control run without
samples to adjust the rotor chamber quickly to the set target temperature.
9
EN
Centrifuge 5424 / 5424 R — Operating manual
2 Product description
2.4Rotors
You can operate the Centrifuge 5424 / 5424 R with the following rotors. Before use of sample
tubes, please note the manufacturer's specifications with regard to centrifugation resistance
(max. rcf).
Operating manual
Rotor
FA-45-24-11
Rotor
FA-45-24-11-
Special
Rotor
F-45-18-11-
Kit
Rotor
F-45-32-5-
PCR
Max. capacityMax. g-force (rcf) /
speed (rpm)
without adapter
(5424: 120 V,
5424 R)
Max. load per rotor bore
24 micro test tubes of 1.5/2.0 mL
each or spin columns.
With adapters:
• 0.2 mL PCR tubes
21,130 x g /
15,000 rpm
3.75 g
• 0.4 mL/0.5 mL micro test
tubes
• 0.6 mL Microtainers
24 micro test tubes of 1.5/2.0 mL
each or spin columns.
With adapters:
• 0.2 mL PCR tubes
21,130 x g /
15,000 rpm
3.75 g
• 0.4 mL/0.5 mL micro test
tubes
• 0.6 mL Microtainers
18 spin columns or 1.5/2.0 mL
micro test tubes.
With adapters:
18,111 x g /
15,000 rpm
3.75 g
• 0.2 mL PCR tubes
• 0.4 mL/0.5 mL micro test
tubes
• 0.6 mL Microtainers
Four PCR strips of 5/8 or
32 PCR tubes of 0.2 mL each.
18,615 x g /
15,000 rpm
3.5 g
Max. g-force (rcf) or
speed (rpm)
without adapter
(5424: 230 V/100 V)
(1)
20,238 x g /
14,860 rpm
20,238 x g /
14,860 rpm
17,347 x g /
14,860 rpm
17,829 x g /
14,860 rpm
Notes
• Aerosol-tight
lid (aluminum).
(2)
rotor
• Spin columns
possible, better with
rotor F-45-18-11-Kit.
• Aerosol-tight
lid (aluminum).
(2)
rotor
• PTFE-coated
(particularly resistant
to chemicals),
marked: coated
• Spin columns
possible, better with
rotor F-45-18-11-Kit.
• Particularly high edge
for all commercially
available spin
columns. See the note
about centrifugation
with open tube lids in
this regard(see Load the rotor on p. 21).
10
(1) Maximum load per rotor bore for adapter + tube + content.
(2) Aerosol tightness tested and certified by the Centre for Emergency Preparedness and
Response, Health Protection Agency, Porton Down (UK) (see certificates at the end of this
operating manual).
For the rotors and rotor lids labeled coated, color fluctuations may occur as a result of the
production process. These fluctuations have no effect on service life or resistance to chemicals.
Centrifuge 5424 / 5424 R — Operating manual
2 Product description
2.4.1Rcf display and calculation
Use the rpm/rcf key to switch the display of centrifugation speed between speed (rpm) and
g-force (rcf). Ensure that the g-force displayed upon switching is standardized to suit the rotor
FA-45-24-11 without an adapter. When other rotors and adapters are used, you can achieve the
following maximum g-forces (rcf) at the maximum speed (see p. 10):
EN
Operating manual
RotorAdaptersMax. centrifugation
radius r
FA-45-24-11,
FA-45-24-11-Special
F-45-18-11-Kitwithout adapter7.218,11117,347
FA-45-32-5-PCRwithout adapter7.418,61517,829
without adapter8.421,13020,238
0.2 mL6.315,84815,179
0.4 mL8.421,13020,238
0.5 mL7.318,36317,558
0.6 mL8.421,13020,238
0.2 mL5.112,82912,288
0.4 mL7.218,11117,347
0.5 mL6.115,34514,697
0.6 mL7.218,11117,347
To determine the g-force (rcf) for a specific adapter, you can calculate per DIN 58 970 using the
following formula:
rcf = 1.118 · 10-5 · n2 · r
n: revolutions per minute (rpm)
r
: max. centrifuging radius in cm
max
Example
In rotor FA-45-24-11, the 0.5 mL adapter has a maximum radius of 7.3 cm. At 7,000 rpm a
maximum g-force of 4,000 x g is reached.
The Centrifuge 5424 / 5424 R is intended exclusively for indoor use and for separating aqueous
solutions and suspensions of various densities in approved test tubes.
Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable for any
damage resulting from the use of non-recommended accessories and spare parts or from the
improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
This device may only be operated by correspondingly trained specialist staff. This staff must have
carefully read the operating manual and be familiar with the function of the device.
3.3Application limits
3.3.1Declaration concerning the ATEX directive (94/9/EC)
Danger of explosion!
Do not operate the device in areas where work is being carried out with explosive substances.
Do not use this device to process any explosive, radioactive or highly reactive substances.
DANGER!
3.3.2Maximum service life for accessories
Do not use this device to process any substances which could create an explosive
atmosphere.
Due to its design and the environmental conditions on the inside of the device, the Centrifuge
5424 / 5424 R is not suitable for use in a potentially explosive atmosphere.
The device must therefore only be used in a safe environment, such as in the open environment
of a ventilated laboratory or an extractor hood. The use of substances which may contribute to a
potentially explosive atmosphere is not permitted. The final decision with regard to the risks
connected with the use of such substances is the responsibility of the user.
Risk of injury from chemically or mechanically damaged accessories.
Even small scratches and cracks can lead to serious internal material damage.
Protect all parts from mechanical damage.
CAUTION!
Check accessories regularly.
Do not use rotors or buckets with signs of corrosion or mechanical damage (e.g.
deformations).
Do not use accessories whose maximum service life has been exceeded.
12
AccessoryMaximum service life from first
commissioning
Transparent polypropylene rotor lids3 years
Plastic adapters1 year
Centrifuge 5424 / 5424 R — Operating manual
3 Safety
For the other rotors and rotor lids of this centrifuge (see Rotors on p. 10) there is no limit for their
service life, as long as the following conditions are met: proper use, recommended maintenance
and undamaged condition. Lids of aerosol-tight rotors must be replaced after 50 autoclave cycles
to ensure aerosol tightness.
The date of manufacture is stamped on the rotors in the format 03/07 (= March 2007) or on the
inside of the plastic rotor lids in the form of a clock . This is for information only and does not
have any reference to the service life.
3.4Information on product liability
In the following cases, the protection provided by the device may be impaired. The liability for the
function of the device passes to the operator if:
• The device is not used in accordance with the operating manual.
• The device is used outside of the range of application described in the preceding chapters.
• The device is used with accessories or consumables (e.g., tubes) which are not
recommended by Eppendorf.
• The device is maintained or repaired by persons not authorized by Eppendorf.
• The owner has made unauthorized modifications to the device.
EN
Operating manual
3.5Warnings for intended use
Read the operating manual first and observe the following general safety instructions before
using the Centrifuge 5424 / 5424 R.
3.5.1Personal injury or damage to the equipment
Risk of electrical shock from damage to the device or mains cable.
The device may only be switched on if the device and mains cable are undamaged.
Only use devices which have been professionally installed or repaired.
WARNING!
Lethal voltages inside the device.
Ensure that the housing is always closed and undamaged so that no parts inside the device
can be contacted by accident.
WARNING!
WARNING!
Do not remove the housing of the device.
Do not allow any liquids to enter the inside of the housing.
Do not allow the device to be opened by anyone except service personnel who have been
specifically authorized by Eppendorf.
Danger from using an incorrect power supply.
Only connect the device to voltage supplies which correspond with the electrical requirements
on the name plate.
13
EN
Operating manual
3 Safety
WARNING!
WARNING!
Centrifuge 5424 / 5424 R — Operating manual
Damage to health when handling infectious liquids and pathogenic germs.
Observe the national regulations for handling these substances, the biological security level
of your laboratory, the material safety data sheets and the manufacturer's application notes.
Use suitable aerosol-tight closure systems for the centrifugation of these substances.
When working with pathogenic germs belonging to a higher risk group, more than one
aerosol-tight bioseal must be provided for.
Wear personal protective equipment (PPE).
Follow the instructions regarding hygiene, cleaning and decontamination.
For complete instructions regarding the handling of germs or biological material of risk group
II or higher, please refer to the "Laboratory Biosafety Manual" (Source: World Health
Organization, current edition of the Laboratory Biosafety Manual).
Centrifuge lid can crush. Keep hands clear.
When opening or closing the device lid, do not reach between the lid and device or into the
latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
Damage to device by spilling liquids in the rotor or rotor chamber
1. Switch the device off.
2. Disconnect the device from the power supply.
NOTICE!
NOTICE!
3. Clean the device and the accessories carefully in accordance with the cleaning and
disinfection instructions in the operating manual.
4. If a different cleaning and disinfecting method is to be used, contact Eppendorf to ensure that
the intended method will not damage the device.
Damage to electronic components through formation of condensation.
After the device has been moved from a cool to a warmer environment, condensation can form
inside the device.
Wait at least three hours (5424) or four hours (5424 R) before connecting the device to the
power supply.
Only 5424: Alternatively: let the device run for half an hour before transporting it.
3.5.2Incorrect handling of the centrifuge
Damage from knocking against or moving the device during operation.
A rotor banging against the rotor chamber wall can cause considerable damage to the device and
rotor.
NOTICE!
Do not move or knock against the device during operation.
3.5.3Incorrect handling of the rotors
Risk of injury from improperly attached rotors and rotor lids.
Centrifuge only with the rotor and rotor lid firmly tightened.
If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
CAUTION!
14
properly secured. Stop centrifugation immediately by pressing the start/stop key.
3 Safety
CAUTION!
CAUTION!
NOTICE!
Centrifuge 5424 / 5424 R — Operating manual
Risk of injury from asymmetric loading of rotors.
Load rotors symmetrically with identical tubes.
Only load adapters with suitable tubes.
Always use tubes of the same type (weight, material/density and volume).
Check symmetric loading by balancing the adapters and tubes used with scales.
Risk of injury from overloaded rotor.
The Centrifuge 5424 / 5424 R is designed for the centrifugation of material with a max. density of
1.2 g/mL at maximum speed and volume and/or load.
Observe the information on each rotor relating to maximum load (adapter, tube and contents)
per rotor bore and make sure it is not exceeded.
Damage to rotors from aggressive chemicals.
Rotors are high-quality components which withstand extreme stresses. This stability can be
impaired by aggressive chemicals.
Avoid the use of aggressive chemicals, including strong and weak alkali, strong acids,
solutions with mercury, copper and other heavy metal ions, halogenated hydrocarbons,
concentrated saline solutions and phenol.
If the rotor is contaminated by aggressive chemicals, clean it immediately using a neutral
cleaning agent. This applies to the rotor bores in particular.
EN
Operating manual
3.5.4Extreme strain on the centrifuging tubes
Risk of injury from overloaded tubes.
Note the loading limits specified by the tube manufacturer.
Only use tubes which are approved by the manufacturer for the required rcf.
CAUTION!
Risk from damaged tubes.
Damaged tubes must not be used, as this could cause further damage to the device and the
accessories and loss of the samples.
NOTICE!
NOTICE!
NOTICE!
Before use, carry out a visual check of all tubes for any damage.
Risk from open tube lids.
Open tube lids can break off during centrifugation and cause damage to the rotor as well as to
the centrifuge.
Carefully close all tube lids before centrifuging.
Exception: Note the information on the centrifugation of spin columns in the rotor F-45-18-11-Kit
(see Load the rotor on p. 21).
Hazard to plastic tubes from organic solvents.
When using organic solvents (e.g. phenol, chloroform) the density of plastic tubes is reduced,
i.e., the tubes could get damaged.
Follow the manufacturer's information about the chemical resistance of tubes.
15
EN
Centrifuge 5424 / 5424 R — Operating manual
3 Safety
Sample tubes heat up.
In uncooled centrifuges the temperature in the rotor chamber, rotor and sample can rise to above
40°C dependent on the run time, g-force (rcf) or speed (rpm) and ambient temperature.
NOTICE!
3.5.5Aerosol-tight centrifugation
Note that this can reduce the centrifugation resistance of the sample tubes.
Please note the temperature resistance of the samples.
Operating manual
Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this purpose
are used. These are always indicated by the prefix FA.
WARNING!
WARNING!
3.6Safety instructions on the device
DepictionMeaningLocation
Always use rotors and rotor lids marked aerosol-tight together for aerosol-tight centrifugation.
Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor lid.
Risk to health due to limited aerosol tightness in the event of incorrect use.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive solvents
can impair the aerosol-tightness of the rotors and rotor lid.
Regularly check that the seals of aerosol-tight rotor lids are undamaged.
Only use aerosol-tight rotor lids with undamaged and clean gaskets.
Thinly brush the threads of the rotor lid screw with pivot grease (order no. Int.: 5810 350.050 /
North America: 022634330) after every proper autoclaving (121°C, 20 min.). Do not apply
the pivot grease to the gaskets.
Replace aerosol-tight rotor lids after 50 autoclave cycles.
Aerosol-tight rotors should never be stored with rotor lids screwed on.
WARNING
General hazard point. Follow the operating
manual.
Right side of the device
16
CAUTION
Always tighten up the rotor using the rotor key
supplied.
If a fault occurs, objects in the immediate vicinity of the devices could get damaged.
In accordance with the recommendations of EN 61010-2-020, leave a safety distance of 30
cm clear around the device during operation.
NOTICE!
Damage from overheating.
Do not place the device close to sources of heat (e.g., radiator, drying cabinet).
Do not expose the device to direct sunlight.
NOTICE!
Ensure free circulation of air by maintaining a distance of at least 30 cm on all sides of the
device from adjacent devices or the wall and keep the underside of the device clear.
Make sure that the vents in the device are always free of obstruction.
Select the location for the device according to the following criteria:
• Suitable power connection as per the name plate (230 V/120 V/100 V).
• Stable, horizontal and resonance-free lab bench. Weight of the device: 13.4 kg (5424) or
21 kg (5424 R).
• A well ventilated environment which is protected from direct sunlight to prevent the device
from heating up more.
4.2Installing the instrument
Centrifuge 5424 R: Compressor damage after improper transport.
Only switch on the centrifuge 4 hours after installation.
Operating manual
NOTICE!
Hint!
Perform the following steps in the sequence described.
1. Place the device on a suitable lab bench.
2. Allow the device to warm up for at least 3 hours (5424) or 4 hours (5424 R) to the ambient
temperature to prevent damage to the electronic components from condensation and damage
to the compressor (only5424 R).
3. Check that the mains voltage and frequency match the requirements on the device type plate.
4. Connect the centrifuge to the mains and switch it on using the mains power switch on the
right side of the device.
• The display is active.
• Lid opens automatically
5. Use the details included in the scope of delivery to check that the delivery is complete.
6. Check all parts for any transport damage. Contact your dealer if any damage is found.
7. Only 5424 R: Insert the condensation water tray at the left side of the device into the holder
provided (see Fig. 2 on p. 8).
Retain the packaging material and the transport protection device for subsequent transport or
storage. See also the instructions relating to transport (see p. 33).
17
EN
Operating manual
Centrifuge 5424 / 5424 R — Operating manual
5 Operation
5Operation
5.1Overview of operating controls
Abb. 3:Control panel of the Centr ifuge 5424 / 5424 R
The Centrifuge 5424 / 5424 R is available in two versions: keypad or rotary knobs. This operating
manual generally describes how to operate the keypad version. However, it also applies to the
rotary knob version.
Before using the Centrifuge 5424 / 5424 R for the first time, familiarize yourself with the operating
controls and the display.
The depiction of the operating controls and the display can also be found on the front fold-out
page (see Fig. 3 and Fig. 4).
Fig. 3:Control panel of the Centrifuge 5424 / 5424 R
1 Call and select the menu parameters
(see Device menu on p. 20)
3 Display4 Short spin centrifugation
5 Release lid6 Start and stop centrifugation
7 Set the speed of centrifugation
Dependent on device version designed as a key or
dial.
9 Adjust the centrifuging duration
Dependent on device version designed as a key or
dial.
11 Select the menu item
(see Menu navigation on p. 21)
Please also read the precise description of the individual menu functions (see p. 20).
2 Switch the displayed centrifuging speed (rpm/rcf)
(see Rcf display and calculation on p. 11)
(see Short spin centrifugation on p. 26)
8 Adjust the temperature (only 5424 R)
10 Start the temperature control run fast temp
(only 5424 R)
18
5 Operation
Centrifuge 5424 / 5424 R — Operating manual
Abb. 4:Display of the Centrifug e 5424 / 5424 R
Fig. 4:Display of the Centrifuge 5424 R. The display of the Centrifuge 5424 is similar.
1 Centrifuging duration2 Status of the key lock (LOCK)
: Centrifuging parameters cannot be modified
unintentionally.
: No key lock.
3 Status of the function ATSET
: Start of operation when reaching 95% of the
preset g-force (rcf) or speed (rpm).
: Start of operation immediately.
5 Soft ramp
: Rotor accelerates and brakes slowly.
No symbol: Rotor accelerates and brakes rapidly.
4 Temperature (only 5424 R)
6 g-force (rcf) or speed (rpm)
(see Rcf display and calculation on p. 11)
EN
Operating manual
7 Status of the centrifuge
: Centrifuge lid unlocked.
: Centrifuge lid locked.
(Flashing): centrifuging in progress.
5 Soft ramp
: Rotor accelerates and brakes slowly.
No symbol: Rotor accelerates and brakes rapidly.
9 Status of the loudspeaker
: Switched on
: Switched off
8 Status of the centrifugation speed display
rcf: g-force (relative centrifugal force)
rpm: revolutions per minute
6 g-force (rcf) or speed (rpm)
(see Rcf display and calculation on p. 11)
19
EN
Centrifuge 5424 / 5424 R — Operating manual
5 Operation
5.2Device menu
Tab. 1:Menu structure of the Centrifuge 5424 / 5424 R. All menu levels contain the additional menu item Back.
Operating manual
Level 1 (M I)Level 2
(M 2)
SOFT
Soft ramp: Reduce speed of acceleration and
braking ramp. Not used for short spin
centrifugation.
LOCK
Key lock: Set the current centrifugation
parameters permanently to prevent the time,
temperature (only 5424 R) g-force (rcf) or
speed (rpm) from being unintentionally
modified.
ATSET
Set start of centrifuging run time.
SHORT
Before the start of a short run (see Short spin
centrifugation on p. 26) it is possible to switch
between the maximum and currently set
g-force (rcf) or speed (rpm). The SOFT function
is not used for short spin centrifugation.
OnRotor accelerates and brakes slowly.
OffRotor accelerates and brakes rapidly.
OnSet the centrifugation parameters permanently.
Off
OnThe set time is counted down only once 95% of
OffThe set time is counted down immediately.
MAXShort spin run at maximum g-force (rcf) or
SetShort run at set g-force (rcf) or speed (rpm).
FunctionDisplay
When you select the keys time, temp (only
5424 R) and speed, the display shows SAFE.
the specified g-force (rcf) or speed (rpm) has been
reached.
speed (rpm) of the rotor used.
TEMP (only 5424 R)
Set the time limitation of continuous cooling
(see p. 24).
ALARMOnSwitch on loudspeaker.
VOLVOL1
SLEEP
Standby mode
LID (only 5424)
Lid release, from software version 2.2.
8 hPreset value.
ooEndless operation of continuous cooling.
Icing possible! Note that this may reduce the
service life of the compressor.
OffSwitch off loudspeaker.
Adjust the speaker volume in 5 steps. The loud-
...
VOL5
OnIf the centrifuge has not been used for 15 min and
OffStandby mode deactivated.
AUTOThe centrifuge lid opens automatically after
OffThe centrifuge lid must be manually opened with
speaker must be switched on for the adjustment to
be audible.
the standby mode has been switched on, it
switches to standby mode. The EP logo then
appears in the display. When a button or knob is
used or the centrifuge lid is closed, the centrifuge
is reactivated. It is then ready for operation.
centrifugation.
the open key after centrifugation.
20
5 Operation
5.3Menu navigation
To change settings in the device menu, proceed as follows:
1.Open the menu.
2.Select the desired menu item.
3.Confirm your selection.
4.Select the setting of the parameters in question.
Centrifuge 5424 / 5424 R — Operating manual
EN
Operating manual
5.
6.Exit the menu.
To exit the second menu level without changing a parameter, select the BACK menu item and
confirm with menu/enter.
Hint!
Confirm the changed setting.
The BACK menu item belonging to the first menu level appears.
5.4Preparing for centrifugation
5.4.1Switch on centrifuge
1. Switch the centrifuge on, using the mains switch.
After switching on at the mains power switch, the centrifuge lid opens automatically.
2. Open the closed centrifuge lid by pressing the open key.
The parameter settings of the last run are displayed.
5.4.2Inserting the rotor
1. Place rotor vertically onto the motor shaft.
2. Insert the rotor key supplied into the rotor nut.
3. Turn rotor key clockwise until the rotor nut is firmly tightened.
5.4.3Load the rotor
CAUTION!
Risk of injury from asymmetric loading of rotors.
Load rotors symmetrically with identical tubes.
Only load adapters with suitable tubes.
Always use tubes of the same type (weight, material/density and volume).
Check symmetric loading by balancing the adapters and tubes used with scales.
21
EN
Operating manual
5 Operation
CAUTION!
Hint!
Centrifuge 5424 / 5424 R — Operating manual
Risk from damaged or overloaded tubes!
When loading the rotor, note the safety instructions with regard to hazards from overloaded or
damaged tubes(see Warnings for intended use on p. 13)
Rotor lid!
• Fixed-angle rotors may only be operated with the appropriate rotor lid in each case. This is
clearly shown by the identical rotor name labeling on the rotor and on the rotor lid.
• To carry out an aerosol-tight centrifugation, an aerosol-tight rotor (marking: red ring) and the
corresponding aerosol-tight rotor lid (marking: aerosol-tight and red lid screw) must be
used.
To load the rotor, proceed as follows:
1. Check the maximum load (adapter, tube and content) per rotor bore.
The information about this can be found on every rotor and in this operating manual
(see Rotors on p. 10).
2. Load rotors and adapters only with the tubes intended for them.
3. Insert tubes opposite each other in pairs into the rotor bores. For symmetrical loading, tubes
that are opposite each other must be of the same type and contain the same filling quantity.
In order to minimize weight differences between filled sample tubes, we recommend taring
with a scale. This will reduce wear on the drive and cut running noise.
4. Attach and tighten rotor lid.
With the rotors FA-45-24-11 and FA-45-24-11-Special centrifugation is also possible without a rotor lid.
Hint!
Please also note:
• The tube lids must be closed securely.
• The rotors are not aerosol-tight without rotor lid.
• The centrifugation is slightly louder.
• Spin columns must always be centrifuged with a rotor lid.
Spin columns
Hint!
For centrifuging spin columns in the rotor F-45-18-11-Kit, you can leave the tube lids open.
However, this is possible only using the tubes provided for this purpose by kit manufacturers. For
reliable centrifugation, you must lean the open tube lids against the edge of the rotor. Ensure that
this does not involve the lids projecting beyond the edge of the rotor, and then put on the
associated rotor lid.
22
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.