Direct current and alternating currentGround connection
Protective ground connection
A terminal which must be connected
to ground prior to establishing any
other connections.
Symbols for certain types of information
Proline Promass H 100
A grounded terminal which, as far as
the operator is concerned, is
grounded via a grounding system.
Equipotential connection
A connection that has to be connected
to the plant grounding system: This
may be a potential equalization line
or a star grounding system depending
on national or company codes of
practice.
SymbolMeaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
Reference to page
Reference to graphic
Visual inspection
Symbols in graphics
SymbolMeaningSymbolMeaning
1, 2, 3,...Item numbers
A, B, C, ...ViewsA-A, B-B, C-C, ... Sections
Hazardous area
Series of steps
Safe area (non-hazardous area)
Flow direction
4Endress+Hauser
Page 5
Proline Promass H 100
12
3
Function and system design
Measuring principle
The measuring principle is based on the controlled generation of Coriolis forces. These forces are
always present in a system when both translational and rotational movements are superimposed.
Fc = 2 · ∆m (ν · ω)
Fc = Coriolis force
∆m = moving mass
ω = rotational velocity
ν = radial velocity in rotating or oscillating system
The amplitude of the Coriolis force depends on the moving mass ∆m, its velocity ν in the system and
thus on the mass flow. Instead of a constant rotational velocity ω, the sensor uses oscillation.
In the sensor, an oscillation is produced in the measuring tube. The Coriolis forces produced at the
measuring tube cause a phase shift in the tube oscillations (see illustration):
• If there is zero flow (i.e. when the fluid stands still), the oscillation measured at points A and B has
the same phase (no phase difference).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the
outlet (3).
Measuring system
A0016772
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the
tube oscillations at the inlet and outlet. System balance is created by exciting an eccentrically
arranged swinging mass to antiphase oscillation. The measuring principle operates independently of
temperature, pressure, viscosity, conductivity and flow profile.
Density measurement
The measuring tube is continuously excited at its resonance frequency. A change in the mass and
thus the density of the oscillating system (comprising measuring tube and fluid) results in a
corresponding, automatic adjustment in the oscillation frequency. Resonance frequency is thus a
function of medium density. The microprocessor utilizes this relationship to obtain a density signal.
Volume measurement
Together with the measured mass flow, this is used to calculate the volume flow.
Temperature measurement
The temperature of the measuring tube is determined in order to calculate the compensation factor
due to temperature effects. This signal corresponds to the process temperature and is also available
as an output signal.
The device consists of a transmitter and a sensor. If a device with Modbus RS485 intrinsically safe is
ordered, the Safety Barrier Promass 100 is part of the scope of supply and must be implemented to
operate the device.
The device is available as a compact version:
The transmitter and sensor form a mechanical unit.
• Additionally for device version with local display:
Via Web browser (e.g. Microsoft Internet Explorer)
• Also for device version with 4-20 mA HART, pulse/frequency/switch
output:
Via Web browser (e.g. Microsoft Internet Explorer)
• Also for device version with EtherNet/IP output:
– Via Web browser (e.g. Microsoft Internet Explorer)
– Via Add-on Profile Level 3 for automation system from Rockwell
Automation
– Via Electronic Data Sheet (EDS)
• Also for device version with PROFINET output:
– Via Web browser (e.g. Microsoft Internet Explorer)
– Via device master file (GSD)
Proline Promass H 100
Sensor
Promass H• Single bent tube
• Simultaneous measurement of flow, volume flow, density and
temperature (multivariable)
• Minimal pressure losses and chemical-resistant materials
• Dual-channel safety barrier for installation in non-hazardous locations
or zone 2/div. 2:
– Channel 1: DC 24 V power supply
– Channel 2: Modbus RS485
• In addition to current, voltage and power limitation, it offers galvanic
isolation of circuits for explosion protection.
• Easy top-hat rail mounting (DIN 35 mm) for installation in control
A0016763
cabinets
6Endress+Hauser
Page 7
Proline Promass H 100
23
7
8
6
9
10
11
4
1
5
Equipment architecture
A0016779
1Possibilities for integrating measuring devices into a system
1Automation system (e.g. PLC)
2EtherNet/IP
3PROFIBUS DP
4Modbus RS485
54-20 mA HART, pulse/frequency/switch output
6Safety Barrier Promass 100
7Modbus RS485 intrinsically safe
8Non-hazardous area
9Non-hazardous area and Zone 2/Div. 2
10 Intrinsically safe area and Zone 1/Div. 1
SafetyIT security
We only provide a warranty if the device is installed and used as described in the Operating
Instructions. The device is equipped with security mechanisms to protect it against any inadvertent
changes to the device settings.
IT security measures in line with operators' security standards and designed to provide additional
protection for the device and device data transfer must be implemented by the operators themselves.
Endress+Hauser7
Page 8
Input
Measured variableDirect measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring rangeMeasuring ranges for liquids
Proline Promass H 100
DNMeasuring range full scale values
[mm][in][kg/h][lb/min]
8³⁄₈0 to 20000 to 73.50
15½0 to 65000 to 238.9
2510 to 180000 to 661.5
401½0 to 450000 to 1654
5020 to 700000 to 2573
min(F)
to
max(F)
Measuring ranges for gases
Measuring ranges only valid for Promass H with tantalum 2.5W.
The full scale values depend on the density of the gas and can be calculated with the formula below:
max(G)
ρ
G
max(G)
max(F)
max(G)
=
<
· ρG : x
max(F)
Maximum full scale value for gas [kg/h]
Maximum full scale value for liquid [kg/h]
max(F)
[mm][in][kg/m3]
8³⁄₈60
15½80
25190
401½90
50290
can never be greater than
max(G)
Gas density in [kg/m³] at operating conditions
DNx
max(F)
To calculate the measuring range, use the Applicator sizing tool → 76
Recommended measuring range
"Flow limit" section → 52
Operable flow range
Over 1000 : 1.
Flow rates above the preset full scale value are not overridden by the electronics unit, with the result
that the totalizer values are registered correctly.
8Endress+Hauser
Page 9
Proline Promass H 100
Input signalExternal measured values
To increase the accuracy of certain measured variables or to calculate the corrected volume flow for
gases, the automation system can continuously write different measured values to the measuring
device:
• Operating pressure to increase accuracy (Endress+Hauser recommends the use of a pressure
measuring device for absolute pressure, e.g. Cerabar M or Cerabar S)
• Medium temperature to increase accuracy (e.g. iTEMP)
• Reference density for calculating the corrected volume flow for gases
Various pressure transmitters and temperature measuring devices can be ordered from Endress
+Hauser: see "Accessories" section → 77
It is recommended to read in external measured values to calculate the following measured variables:
• Mass flow
• Corrected volume flow
HART protocol
The measured values are written from the automation system to the measuring device via the HART
protocol. The pressure transmitter must support the following protocol-specific functions:
• HART protocol
• Burst mode
Digital communication
The measured values can be written from the automation system to the measuring via:
• PROFIBUS DP
• Modbus RS485
• EtherNet/IP
• PROFINET
Output
Output signalCurrent output
Current output4-20 mA HART (active)
Maximum output values• DC 24 V (no flow)
Load0 to 700 Ω
Resolution0.38 µA
DampingAdjustable: 0.07 to 999 s
Assignable measured
variables
Pulse/frequency/switch output
FunctionCan be set to pulse, frequency or switch output
VersionPassive, open collector
Maximum input values• DC 30 V
Voltage dropFor 25 mA: ≤ DC 2 V
Pulse output
• 22.5 mA
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
• 25 mA
Endress+Hauser9
Page 10
Pulse widthAdjustable: 0.05 to 2000 ms
Maximum pulse rate10000 Impulse/s
Pulse valueAdjustable
Assignable measured
variables
Frequency output
Output frequencyAdjustable: 0 to 10000 Hz
DampingAdjustable: 0 to 999 s
Pulse/pause ratio1:1
Assignable measured
variables
Switch output
Switching behaviorBinary, conductive or non-conductive
Switching delayAdjustable: 0 to 100 s
Number of switching
cycles
Assignable functions• Off
• Mass flow
• Volume flow
• Corrected volume flow
• Mass flow
• Volume flow
• Corrected volume flow
• Density
• Reference density
• Temperature
The range of options increases if the measuring device has one or more
application packages.
Unlimited
• On
• Diagnostic behavior
• Limit value
– Mass flow
– Volume flow
– Corrected volume flow
– Density
– Reference density
– Temperature
– Totalizer 1-3
• Flow direction monitoring
• Status
– Partially filled pipe detection
– Low flow cut off
The range of options increases if the measuring device has one or more
application packages.
Proline Promass H 100
PROFIBUS DP
Signal encodingNRZ code
Data transfer9.6 kBaud…12 MBaud
Modbus RS485
Physical interfaceIn accordance with EIA/TIA-485-A standard
Terminating resistor• For device version used in non-hazardous areas or Zone 2/Div. 2: integrated
and can be activated via DIP switches on the transmitter electronics module
• For device version used in intrinsically safe areas: integrated and can be
activated via DIP switches on the Safety Barrier Promass 100
10Endress+Hauser
Page 11
Proline Promass H 100
EtherNet/IP
StandardsIn accordance with IEEE 802.3
PROFINET
StandardsIn accordance with IEEE 802.3
Signal on alarm
Depending on the interface, failure information is displayed as follows:
Current output
4-20 mA
Failure modeChoose from:
• 4 to 20 mA in accordance with NAMUR recommendation NE 43
• 4 to 20 mA in accordance with US
• Min. value: 3.59 mA
• Max. value: 22.5 mA
• Freely definable value between: 3.59 to 22.5 mA
• Actual value
• Last valid value
HART
Device diagnosticsDevice condition can be read out via HART Command 48
Pulse/frequency/switch output
Pulse output
Failure modeChoose from:
• Actual value
• No pulses
Frequency output
Failure modeChoose from:
• Actual value
• 0 Hz
• Defined value: 0 to 12500 Hz
Switch output
Failure modeChoose from:
• Current status
• Open
• Closed
PROFIBUS DP
Status and alarm
messages
Diagnostics in accordance with PROFIBUS PA Profile 3.02
Modbus RS485
Failure modeChoose from:
• NaN value instead of current value
• Last valid value
Endress+Hauser11
Page 12
Proline Promass H 100
EtherNet/IP
Device diagnosticsDevice condition can be read out in Input Assembly
PROFINET
Device diagnosticsIn accordance with "Application Layer protocol for decentral device periphery and
distributed automation", version 2.3
Local display
Plain text displayWith information on cause and remedial measures
BacklightRed backlighting indicates a device error.
Status signal as per NAMUR recommendation NE 107
Operating tool
• Via digital communication:
– HART protocol
– PROFIBUS DP
– Modbus RS485
– EtherNet/IP
– PROFINET
• Via service interface
• Via Web server
Ex connection data
Plain text displayWith information on cause and remedial measures
Additional information on remote operation → 67
Web browser
Plain text displayWith information on cause and remedial measures
Light emitting diodes (LED)
Status informationStatus indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
• EtherNet/IP network available
• EtherNet/IP connection established
• PROFINET network available
• PROFINET connection established
• PROFINET blinking feature
These values only apply for the following device version:
Order code for "Output", option M "Modbus RS485", for use in intrinsically safe areas
12Endress+Hauser
Page 13
Proline Promass H 100
Safety Barrier Promass 100
Safety-related values
Terminal numbers
Supply voltageSignal transmission
2 (L-)1 (L+)26 (A)27 (B)
U
nom
U
= AC 260 V
max
= DC 24 V
U
nom
U
= AC 260 V
max
= DC 5 V
Intrinsically safe values
Terminal numbers
Supply voltageSignal transmission
20 (L-)10 (L+)62 (A)72 (B)
Uo = 16.24 V
Io = 623 mA
Po = 2.45 W
With IIB
1)
: Lo = 92.8 µH, Co = 0.433 μF, Lo/Ro = 14.6 μH/Ω
1)
: Lo= 372 µH, Co = 2.57 μF, Lo/Ro = 58.3 μH/Ω
With IIC
For an overview and for information on the interdependencies between the gas group - sensor - nominal
diameter, see the "Safety Instructions" (XA) document for the measuring device
Low flow cut off
Galvanic isolation
1)The gas group depends on the sensor and nominal diameter.
Transmitter
Intrinsically safe values
Order code for
"Approval"
• Option BM: ATEX II2G + IECEx Z1 Ex ia, II2D Ex tb
• System units
– Mass flow
– Mass
– Volume flow
– Volume
– Corrected volume flow
– Corrected volume
– Density
– Reference density
– Temperature
– Pressure
• Concentration application package
– Coefficients A0 to A4
– Coefficients B1 to B3
Order code for "Output":
Option M: Modbus RS485, for use in intrinsically safe areas (connection via Safety Barrier Promass 100)
Endress+Hauser27
Page 28
Proline Promass H 100
L
L
1
2
+
_
1
2
EtherNet/IP connection version
Order code for "Output", option N
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code for
"Housing"
Options
A, B
Options
A, B, C
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, stainless
• Option C: ultra-compact, stainless
6EtherNet/IP terminal assignment
1Power supply: DC 24 V
2EtherNet/IP
Connection methods available
Output
Device plugs
→ 30
Device plugs
→ 30
Power
supply
Terminals• Option L: plug M12x1 + thread NPT ½"
Device plugs
→ 30
Possible options for order code
"Electrical connection"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Option Q: 2 x plug M12x1
A0017054
Order code for
"Output"
Option NDC 24 VEtherNet/IP
Order code for "Output":
Option N: EtherNet/IP
Terminal number
Power supplyOutput
2 (L-)1 (L+)Device plug M12x1
28Endress+Hauser
Page 29
Proline Promass H 100
L
L
1
2
+
_
1
2
PROFINET connection version
Order code for "Output", option R
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code for
"Housing"
Options
A, B
Options
A, B, C
Order code for "Housing":
• Option A: compact, coated aluminum
• Option B: compact, stainless
• Option C: ultra-compact, stainless
7PROFINET terminal assignment
1Power supply: DC 24 V
2PROFINET
Connection methods available
Output
Device plugs
→ 30
Device plugs
→ 30
Power
supply
Terminals• Option L: plug M12x1 + thread NPT ½"
Device plugs
→ 30
Possible options for order code
"Electrical connection"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Option Q: 2 x plug M12x1
A0017054
Order code for "Output":
Option R: PROFINET
Terminal number
Order code for
"Output"
Option RDC 24 VPROFINET
Power supplyOutput
2 (L-)1 (L+)Device plug M12x1
Endress+Hauser29
Page 30
Safety Barrier Promass 100
A
Safe area
1
L+2L
27
26
A B
Power
supply
24V
DC
Modbus
RS485
10207262
L+LA B
Power
supply
Modbus
RS485
Hazardous area
Safety Barrier
Promass 100
Power
Communication
Lift panel for
bus termination
1
2
A
1
2
4
3
5
8Safety Barrier Promass 100 with terminals
1Non-hazardous area and Zone 2/Div. 2
2Intrinsically safe area
Proline Promass H 100
A0016922
Pin assignment, device plug
Order codes for the M12x1 connectors, see the "Order code for electrical connection" column:
• 4-20 mA HART, pulse/frequency/switch output → 24
• PROFIBUS DP→ 25
• Modbus RS485 → 26
• EtherNet/IP → 28
• PROFINET→ 29
Supply voltage
For all connection versions except MODBUS RS485 intrinsically safe (device side)
Device plug MODBUS RS485 intrinsically safe with supply voltage → 31
PinAssignment
1L+DC 24 V
2Not assigned
3Not assigned
4L-DC 24 V
A0016809
5Grounding/shielding
CodingPlug/socket
APlug
The following is recommended as a socket:
• Binder, series 763, part no. 79 3440 35 05
• Alternatively: Phoenix part no. 1669767 SAC-5P-M12MS
– With the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
– With the order code for "Output", option N: EtherNet/IP
• When using the device in a hazardous location: Use a suitably certified socket.
30Endress+Hauser
Page 31
Proline Promass H 100
3
2
4
1
5
3
2
4
1
5
1
2
4
3
5
4-20 mA HART with pulse/frequency/switch output
Device plug for signal transmission (device side)
PinAssignment
1+4-20 mA HART (active)
2-4-20 mA HART (active)
3+Pulse/frequency/switch output (passive)
4-Pulse/frequency/switch output (passive)
A0016810
• Recommended plug: Binder, series 763, part no. 79 3439 12 05
• When using the device in a hazardous location, use a suitably certified plug.
5Grounding/shielding
CodingPlug/socket
ASocket
PROFIBUS DP
For use in the non-hazardous area and Zone 2/Div. 2.
Device plug for signal transmission (device side)
PinAssignment
1Not assigned
2APROFIBUS DP
3Not assigned
4BPROFIBUS DP
A0016811
• Recommended plug: Binder, series 763, part no. 79 4449 20 05
• When using the device in a hazardous location, use a suitably certified plug.
5Grounding/shielding
CodingPlug/socket
BSocket
MODBUS RS485
Device plug for signal transmission with supply voltage (device side), MODBUS RS485 (intrinsically
safe)
PinAssignment
1L+Supply voltage, intrinsically safe
A0016809
2A
3B
4L-Supply voltage, intrinsically safe
5Grounding/shielding
CodingPlug/socket
APlug
Modbus RS485 intrinsically safe
• Recommended socket: Binder, series 763, part no. 79 3439 12 05
• When using the device in a hazardous location: Use a suitably certified socket.
Endress+Hauser31
Page 32
Proline Promass H 100
3
2
4
1
5
3
2
4
1
3
2
4
1
Device plug for signal transmission (device side), MODBUS RS485 (not intrinsically safe)
For use in the non-hazardous area and Zone 2/Div. 2.
PinAssignment
1Not assigned
2AModbus RS485
3Not assigned
4BModbus RS485
A0016811
• Recommended plug: Binder, series 763, part no. 79 4449 20 05
• When using the device in a hazardous location, use a suitably certified plug.
EtherNet/IP
Device plug for signal transmission (device side)
5Grounding/shielding
CodingPlug/socket
BSocket
PinAssignment
1+Tx
2+Rx
3-Tx
4-Rx
A0016812
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
CodingPlug/socket
DSocket
PROFINET
Device plug for signal transmission (device side)
PinAssignment
1+TD +
2+RD +
3-TD –
4-RD –
A0016812
CodingPlug/socket
DSocket
Recommended plug:
• Binder, series 763, part no. 99 3729 810 04
• Phoenix, part no. 1543223 SACC-M12MSD-4Q
• When using the device in a hazardous location, use a suitably certified plug.
Supply voltage
The power unit must be tested to ensure it meets safety requirements (e.g. PELV, SELV).
32Endress+Hauser
Page 33
Proline Promass H 100
Transmitter
For device version with communication type:
• HART, PROFIBUS DP, EtherNet/IP: DC 20 to 30 V
• Modbus RS485, device version:
– For use in the non-hazardous area and Zone 2/Div. 2: DC 20 to 30 V
– For use in the intrinsically safe area: power supply via Safety Barrier Promass 100
Safety Barrier Promass 100
DC 20 to 30 V
Power consumptionTransmitter
Order code for "Output"
Option B: 4-20 mA HART with pulse/frequency/switch output3.5 W
Option L: PROFIBUS DP3.5 W
Option M Modbus RS485, for use in non-hazardous areas and Zone 2/
Div. 2
Option M: Modbus RS485, for use in intrinsically safe areas2.45 W
Option N: EtherNet/IP3.5 W
Option R: PROFINET3.5 W
Safety Barrier Promass 100
Order code for "Output"
Option M: Modbus RS485, for use in intrinsically safe areas4.8 W
Option M Modbus RS485, for use in non-hazardous
areas and Zone 2/Div. 2
Option M: Modbus RS485, for use in intrinsically safe
areas
Option N: EtherNet/IP145 mA18 A (< 0.125 ms)
Option R: PROFINET145 mA18 A (< 0.125 ms)
Maximum
Power consumption
Maximum
Power consumption
Maximum
Current consumption
90 mA10 A (< 0.8 ms)
145 mA16 A (< 0.4 ms)
switch-on current
3.5 W
Maximum
Safety Barrier Promass 100
Power supply failure
Order code for "Output"
Option M: Modbus RS485, for use in intrinsically safe
areas
• Totalizers stop at the last value measured.
Maximum
Current consumption
230 mA10 A (< 0.8 ms)
Maximum
switch-on current
• Depending on the device version, the configuration is retained in the device memory or in the
plug-in memory (HistoROM DAT).
• Configuration is retained in the plug-in memory (HistoROM DAT).
• Error messages (incl. total operated hours) are stored.
Endress+Hauser33
Page 34
Electrical connectionConnecting the transmitter
121234
ABC
4
4...20 mA
+
–
5
2
1
3
6
+
_
AHousing version: compact, aluminum coated
BHousing version: compact, stainless
1Cable entry or device plug for signal transmission
2Cable entry or device plug for supply voltage
CHousing version: ultra-compact, stainless
3Device plug for signal transmission
4Device plug for supply voltage
Proline Promass H 100
A0016924
• Terminal assignment → 23
• Pin assignment, device plug → 30
In the case of device versions with a connector, the transmitter housing does not need to be
opened to connect the signal cable or power supply cable.
Connection examples
Current output 4-20 mA HART
9Connection example for 4-20 mA HART current output (active)
1Automation system with current input (e.g. PLC)
2Cable shield, observe cable specifications
3Connection for HART operating devices
4Resistor for HART communication (≥ 250 Ω): observe maximum load
5Analog display unit: observe maximum load
6Transmitter
A0016800
34Endress+Hauser
Page 35
Proline Promass H 100
1
+
_
12345
2
+
–
+
–
3
1
+
_
+
_
2
+
_
3
Pulse/frequency output
A0016801
10Connection example for pulse/frequency output (passive)
1Automation system with pulse/frequency input (e.g. PLC)
2Power supply
3Transmitter: observe input values → 9
Switch output
11Connection example for switch output (passive)
1Automation system with switch input (e.g. PLC)
2Power supply
3Transmitter: observe input values
A0016802
Endress+Hauser35
Page 36
Proline Promass H 100
2
1
A
B
A
B
A
B
3
3
2
1
A
B
A
B
A
B
3
4
4
PROFIBUS DP
A0021429
12Connection example for PROFIBUS DP, non-hazardous area and Zone 2/Div. 2
1Control system (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3Transmitter
If baud rates > 1.5 MBaud an EMC cable entry must be used and the cable shield must continue
as far as the terminal wherever possible.
Modbus RS485
Modbus RS485, non-hazardous area and Zone 2/Div. 2
A0016803
13Connection example for Modbus RS485, non-hazardous area and Zone 2/Div. 2
1Control system (e.g. PLC)
2Cable shield: the cable shield must be grounded at both ends to comply with EMC requirements; observe cable
specifications
3Distribution box
4Transmitter
36Endress+Hauser
Page 37
Proline Promass H 100
21
A
B
A
B
3
L+
L-
A
B
L+
L-
L- L+
A
B
567
8
4
124
5
5
3
Modbus RS485 intrinsically safe
A0016804
14Connection example for Modbus RS485 intrinsically safe
1Control system (e.g. PLC)
2Cable shield, observe cable specifications
3Safety Barrier Promass 100
4Observe cable specifications
5Non-hazardous area
6Non-hazardous area and Zone 2/Div. 2
7Intrinsically safe area
8Transmitter
17Connection example for HART input (burst mode) via current output (active)
1Cable shield, observe cable specifications
2Resistor for HART communication (≥ 250 Ω): observe maximum load
3Connection for HART operating devices
4Analog display unit
5Transmitter
6Sensor for external measured variable
A0019828
38Endress+Hauser
Page 39
Proline Promass H 100
3
4...20 mA
4
521
+
-
6
4...20 mA
+
7
3
4
52
+
-
-
+
A0019830
18Connection example for HART input (master mode) via current output (active)
1Automation system with current input (e.g. PLC).
Prerequisite: automation system with HART version 6, HART commands 113 and 114 can be processed.
2Cable shield, observe cable specifications
3Resistor for HART communication (≥ 250 Ω): observe maximum load
4Connection for HART operating devices
5Analog display unit
6Transmitter
7Sensor for external measured variable
Potential equalizationRequirements
No special measures for potential equalization are required.
Please consider the following to ensure correct measurement:
• Same electrical potential for the fluid and sensor
• Company-internal grounding concepts
For devices intended for use in hazardous locations, please observe the guidelines in the Ex
documentation (XA).
Terminals
Transmitter
Spring terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Safety Barrier Promass 100
Plug-in screw terminals for wire cross-sections0.5 to 2.5 mm2 (20 to 14 AWG)
Cable entries
• Cable gland: M20 × 1.5 with cable 6 to 12 mm (0.24 to 0.47 in)
• Thread for cable entry:
– NPT ½"
– G ½"
– M20
Cable specificationPermitted temperature range
• –40 °C (–40 °F) to +80 °C (+176 °F)
• Minimum requirement: cable temperature range ≥ ambient temperature +20 K
Power supply cable
Standard installation cable is sufficient.
Signal cable
Current output
For 4-20 mA HART: Shielded cable recommended. Observe grounding concept of the plant.
Pulse/frequency/switch output
Standard installation cable is sufficient.
Endress+Hauser39
Page 40
Proline Promass H 100
PROFIBUS DP
The IEC 61158 standard specifies two types of cable (A and B) for the bus line which can be used for
every transmission rate. Cable type A is recommended.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance<30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
Modbus RS485
The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used
for every transmission rate. Cable type A is recommended.
>0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
Cable typeA
Characteristic impedance135 to 165 Ω at a measuring frequency of 3 to 20 MHz
Cable capacitance<30 pF/m
Wire cross-section
Cable typeTwisted pairs
Loop resistance≤110 Ω/km
Signal dampingMax. 9 dB over the entire length of the cable cross-section
ShieldCopper braided shielding or braided shielding with foil shield. When grounding
>0.34 mm2 (22 AWG)
the cable shield, observe the grounding concept of the plant.
EtherNet/IP
The standard ANSI/TIA/EIA-568-B.2 Annex specifies CAT 5 as the minimum category for a cable
used for EtherNet/IP. CAT 5e and CAT 6 are recommended.
For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of ODVA Organization
PROFINET
Standard IEC 61156-6 specifies CAT 5 as the minimum category for a cable used for PROFINET. CAT
5e and CAT 6 are recommended.
For more information on planning and installing PROFINET networks, see: "PROFINET Cabling
and Interconnection Technology", Guideline for PROFINET
Connecting cable between Safety Barrier Promass 100 and measuring device
Cable typeShielded twisted-pair cable with 2x2 wires. When grounding the cable shield,
observe the grounding concept of the plant.
Maximum cable resistance 2.5 Ω, one side
Comply with the maximum cable resistance specifications to ensure the operational reliability
of the measuring device.
40Endress+Hauser
Page 41
Proline Promass H 100
The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
hazardous areas .
Wire cross-sectionMaximum cable length
[mm2][AWG][m][ft]
0.52070230
0.7518100328
1.017100328
1.516200656
2.514300984
Performance characteristics
Reference operating
conditions
Maximum measured error
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at 2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the Applicator sizing tool → 76
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Design fundamentals → 44
Mass flow and volume flow (liquids)
±0.10 % o.r.
Mass flow (gases)
±0.50 % o.r. (tantalum 2.5W)
Density (liquids)
Under
reference operating conditions
[g/cm³][lbs/in³][g/cm³][lbs/in³][g/cm³][lbs/in³]
±0.0005±0.00097±0.02±0.039±0.002±0.0039
Standard density
calibration
1)
Wide-range
density specification
2) 3)
1)Valid over the entire temperature and density range
2)Valid range for special density calibration: 0 to 2 g/cm³, +5 to +80 °C (+41 to +176 °F)
3)Order code for "Application package", option EF "Special density and concentration "
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)
Endress+Hauser41
Page 42
Proline Promass H 100
Zero point stability
DNZero point stability
[mm][in][kg/h][lb/min]
8³⁄₈0.400.015
15½0.650.024
2511.800.066
401½9.000.331
50214.000.514
Flow values
Flow values as turndown parameter depending on nominal diameter.
In the case of analog outputs, the output accuracy must also be considered for the measured
error; in contrast, this need not be considered in the case of fieldbus outputs (e.g. Modbus
RS485, EtherNet/IP).
The outputs have the following base accuracy specifications.
Current output
AccuracyMax. ±5 µA
Pulse/frequency output
o.r. = of reading
AccuracyMax. ±50 ppm o.r. (across the entire ambient temperature range)
42Endress+Hauser
Page 43
Proline Promass H 100
Repeatability
Response time
Influence of ambient
temperature
o.r. = of reading; 1 g/cm3 = 1 kg/l; T = medium temperature
Base repeatability
Mass flow and volume flow (liquids)
±0.05 % o.r.
Mass flow (gases)
±0.25 % o.r. (tantalum 2.5W)
Design fundamentals → 44
Density (liquids)
±0.00025 g/cm
3
Temperature
±0.25 °C ± 0.0025 · T °C (±0.45 °F ± 0.0015 · (T–32) °F)
The response time depends on the configuration (damping).
Current output
o.r. = of reading
Temperature coefficientMax. ±0.005% o.r./°C
Pulse/frequency output
Influence of medium
temperature
Temperature coefficientNo additional effect. Included in accuracy.
Mass flow and volume flow
When there is a difference between the temperature for zero point adjustment and the process
temperature, the typical measured error of the sensor is ±0.0002 % of the full scale value/°C
(±0.0001 % of the full scale value/°F).
Density
When there is a difference between the density calibration temperature and the process
temperature, the typical measured error of the sensor is ±0.0001 g/cm3 /°C (±0.00005 g/cm3 /°F).
Field density calibration is possible.
Wide-range density specification (special density calibration)
If the process temperature is outside the valid range (→ 41) the measured error is
±0.0001 g/cm3 /°C (±0.00005 g/cm3 /°F)
Endress+Hauser43
Page 44
[kg/m ]
3
[°C]
-40
0
-80
40 80 120 160 200 240 280 320 360 400
[°F]
-50
0
50
100150
200
0
2
4
6
8
10
14
12
16
1
2
18
20
1Field density calibration, for example at +20 °C (+68 °F)
ZeroPoint
BaseAccu
⋅ 100
³
± BaseAccu
ZeroPoint
BaseAccu
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
±
2Special density calibration
Proline Promass H 100
A0016615
Influence of medium
pressure
Design fundamentals
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
The table below shows the effect on accuracy of mass flow due to a difference between calibration
pressure and process pressure.
o.r. = of reading
DNPromass H zirconium 702/R 60702Promass H tantalum 2.5W
BaseAccu = base accuracy in % o.r., BaseRepeat = base repeatability in % o.r.
MeasValue = measured value; ZeroPoint = zero point stability
Calculation of the maximum measured error as a function of the flow rate
Flow rateMaximum measured error in % o.r.
A0021332
A0021333
44Endress+Hauser
A0021339
A0021334
Page 45
Proline Promass H 100
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
³
½ ⋅ ZeroPoint
BaseRepeat
⋅ 100
<
ZeroPoint
MeasValue
⋅ 100
½ ⋅±
E [%]
Q [%]
0510 15 20 25 30 35 40 45 5060 6555
0
0.5
1.0
1.5
2.0
2.5
70 75 80 85 90 95 100
Calculation of the maximum repeatability as a function of the flow rate
Flow rateMaximum repeatability in % o.r.
A0021335
A0021336
Example for max. measured error
EError: Maximum measured error as % o.r. (example)
QFlow rate as %
Installation
A0021340
A0021337
A0016709
Mounting location
No special measures such as supports are necessary. External forces are absorbed by the construction
of the device.
A0023344
To prevent measuring errors arising from accumulation of gas bubbles in the measuring tube, avoid
the following mounting locations in the pipe:
• Highest point of a pipeline.
• Directly upstream of a free pipe outlet in a down pipe.
Installation in down pipes
However, the following installation suggestion allows for installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter
prevent the sensor running empty while measurement is in progress.
Endress+Hauser45
Page 46
1
2
3
4
5
19Installation in a down pipe (e.g. for batching applications)
1Supply tank
2Sensor
3Orifice plate, pipe restriction
4Valve
5Batching tank
Proline Promass H 100
A0015596
Orientation
DNØ orifice plate, pipe restriction
[mm][in][mm][in]
8³⁄₈60.24
15½100.40
251140.55
401½220.87
502281.10
The direction of the arrow on the sensor nameplate helps you to install the sensor according to the
flow direction (direction of medium flow through the piping).
OrientationRecommendation
AVertical orientation
A0015591
BHorizontal orientation, transmitter
head up
CHorizontal orientation, transmitter
head down
DHorizontal orientation, transmitter
head at side
→ 20, 47
A0015589
→ 20, 47
A0015590
A0015592
1)
Exceptions:
2)
Exceptions:
1)Applications with low process temperatures may decrease the ambient temperature. To maintain the
minimum ambient temperature for the transmitter, this orientation is recommended.
2)Applications with high process temperatures may increase the ambient temperature. To maintain the
maximum ambient temperature for the transmitter, this orientation is recommended.
46Endress+Hauser
Page 47
Proline Promass H 100
12
> 22.5
(> 0.89)
> 22.5
(> 0.89)
If a sensor is installed horizontally with a curved measuring tube, match the position of the sensor to
the fluid properties.
A0014057
20Orientation of sensor with curved measuring tube
1Avoid this orientation for fluids with entrained solids: Risk of solids accumulating.
2Avoid this orientation for outgassing fluids: Risk of gas accumulating.
Inlet and outlet runs
Special mounting
instructions
Mounting Safety Barrier
Promass 100
No special precautions need to be taken for fittings which create turbulence, such as valves, elbows
or T-pieces, as long as no cavitation occurs → 53.
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology. Calibration
takes place under reference conditions → 41. Therefore, a zero point adjustment in the field is
generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates
• Under extreme process or operating conditions (e.g. very high process temperatures or very highviscosity fluids).
21Minimum distance between additional Safety Barrier Promass 100 or other modules.Engineering unit mm
(in)
Ambient temperature range
Environment
Measuring device Non-Ex–40 to +60 °C (–40 to +140 °F)
Ex na, NI version –40 to +60 °C (–40 to +140 °F)
Endress+Hauser47
A0016894
Page 48
Proline Promass H 100
Ex ia, IS version• –40 to +60 °C (–40 to +140 °F)
• –50 to +60 °C (–58 to +140 °F) (order code for "Test, certificate",
option JM))
Readability of the local display–20 to +60 °C (–4 to +140 °F)
The readability of the display may be impaired at temperatures outside
the temperature range.
Safety Barrier Promass 100–40 to +60 °C (–40 to +140 °F)
If operating outdoors:
‣
Avoid direct sunlight, particularly in warm climatic regions.
Weather protection covers can be ordered from Endress+Hauser: see "Accessories" section
Temperature tables
In the following tables, the following interdependencies between the maximum medium
temperature Tm for T6 to T1 and the maximum ambient temperature Ta apply when operating the
device in hazardous areas.
Ex ia, CCSAUS IS
SI units
Order code for "Housing"T
Option A "Compact coated alu"
Option B "Compact, stainless"
Option C "Ultra-compact, stainless"
1)The following applies for specified sensors with a maximum medium temperature Tm = 205 °C: Tm = 170 °C
2)The following applies for specified sensors with a maximum medium temperature Tm = 205 °C: Tm = 205 °C
Determining the temperature class and surface temperature with the temperature table
• In the case of gas: Determine the temperature class as a function of the ambient temperature T
a
and the medium temperature Tm.
• In the case of dust: Determine the maximum surface temperature as a function of the maximum
ambient temperature Ta and the maximum medium temperature Tm.
Example
• Measured maximum ambient temperature: Tma = 47 °C
• Measured maximum medium temperature: Tmm = 108 °C
A0019758
22Procedure for determining the maximum surface temperature
1.Select device (optional).
Endress+Hauser49
Page 50
Proline Promass H 100
2.In the column for the maximum ambient temperature Ta select the temperature that is
immediately greater than or equal to the measured maximum ambient temperature Tma that is
present.
Ta = 50 °C.
The row showing the maximum medium temperature is determined.
3.Select the maximum medium temperature Tm of this row, which is larger or equal to the
measured maximum medium temperature Tmm.
The column with the temperature class for gas is determined: 108 °C ≤ 120°C → T4.
4.The maximum temperature of the temperature class determined corresponds to the maximum
surface temperature for dust: T4 = 135 °C
Storage temperature
Climate class
Degree of protection
Vibration resistance
Shock resistance
–40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F) (standard version)
–50 to +80 °C (–58 to +176 °F) (Order code for "Test, certificate", option JM)
DIN EN 60068-2-38 (test Z/AD)
Transmitter and sensor
• As standard: IP66/67, type 4X enclosure
• With the order code for "Sensor options", option CM: IP69K can also be ordered
• When housing is open: IP20, type 1 enclosure
• Display module: IP20, type 1 enclosure
Safety Barrier Promass 100
IP20
Compact version
• Vibration, sinusoidal according to IEC 60068-2-6
– 2 to 8.4 Hz, 3.5 mm peak
– 8.4 to 2000 Hz, 1 g peak
• Vibration broad-band random, according to IEC 60068-2-64
– 10 to 200 Hz, 0.003 g2/Hz
– 200 to 2000 Hz, 0.001 g2/Hz
– Total: 1.54 g rms
Compact version
Shock, half-sine according to IEC 60068-2-27
6 ms 30 g
Shock resistance
Interior cleaning
Electromagnetic
compatibility (EMC)
Compact version
Rough handling shocks according to IEC 60068-2-31
• Cleaning in place (CIP)
• Sterilization in place (SIP)
Options
Oil- and grease-free version for wetted parts, without inspection certificate
Order code for "Service", option HA
• Depends on the communication protocol:
– HART, PROFIBUS DP, Modbus RS485, EtherNet/IP:
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
– PROFINET: as per IEC/EN 61326
• Complies with emission limits for industry as per EN 55011 (Class A)
• Device version with PROFIBUS DP: Complies with emission limits for industry as per EN 50170
Volume 2, IEC 61784
The following applies for PROFIBUS DP: If baud rates > 1.5 MBaud, an EMC cable entry must be
used and the cable shield must continue as far as the terminal wherever possible.
For details, refer to the Declaration of Conformity.
50Endress+Hauser
Page 51
Proline Promass H 100
PN40
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
200
400
600
0
800
[bar][psi]
300[°F]350250200150100500-50
[°C]200150100500-50
400
0
10
20
30
40
50
[bar][psi]
200
400
600
0
Class 300
Class 150
Process
Medium temperature range
Density
Pressure-temperature
ratings
Sensor
• Zirconium 702/R 60702: –50 to +205 °C (–58 to +401 °F)
• Tantalum 2.5W: –50 to +150 °C (–58 to +302 °F)
Seals
No internal seals
0 to 5000 kg/m3 (0 to 312 lb/cf)
The following pressure-temperature ratings refer to the entire device and not just the process
connection.
Flange connection according to EN 1092-1 (DIN 2501)
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TJ
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TJ
The material load curves for the temperature range +150 to +205 °C (+302 to +401 °F) apply only to
the order code for "Measuring tube material", option TJ
Secondary containment
pressure rating
The sensor housing is filled with dry nitrogen and protects the electronics and mechanics inside.
The following secondary containment pressure rating is only valid for a fully welded sensor housing
and/or a device equipped with closed purge connections (never opened/as delivered).
DNSecondary containment
pressure rating
(designed with a safety factor
≥ 4)
[mm][in][bar][psi][bar][psi]
8³⁄₈253621702465
15½253621602320
251253621301885
401½16232851200
50216232851200
Secondary containment burst pressure
If there is a risk of measuring tube failure due to process characteristics, e.g. with corrosive
fluids, we recommend the use of sensors whose secondary containment is equipped with special
pressure monitoring connections (order code for "Sensor option", option CH "Purge connection").
With the help of these connections, the fluid collected in the secondary containment can be
bled off in the event of tube failure. This is especially important in high-pressure gas
applications. These connections can also be used for gas purging (gas detection).
Do not open the purge connections unless the containment can be filled immediately with a
dry, inert gas. Use only low gauge pressure to purge. Maximum pressure: 5 bar (72.5 psi).
If a device fitted with purge connections is connected to the purge system, the maximum
nominal pressure is determined by the purge system itself or by the device, depending on which
component has the lower nominal pressure.
Dimensions:
Flow limit
Select the nominal diameter by optimizing between the required flow range and permissible pressure
loss.
For an overview of the full scale values for the measuring range, see the "Measuring range"
section
• The minimum recommended full scale value is approx. 1/20 of the maximum full scale value
• In most applications, 20 to 50 % of the maximum full scale value can be considered ideal
• A low full scale value must be selected for abrasive media (such as liquids with entrained solids):
52Endress+Hauser
flow velocity < 1 m/s (< 3 ft/s).
• For gas measurement the following rules apply:
– The flow velocity in the measuring tubes should not exceed half the sound velocity (0.5 Mach)
– The maximum mass flow depends on the density of the gas: formula
Page 53
Proline Promass H 100
t
a
0
20
60
40
80
[in]
1
2
3
0
t
100
4
[mm]
T
40(104)
T
60(140)
80100120140
200300400
160180200
[°F]
T
m
[°C]
Pressure loss
System pressure
Thermal insulation
To calculate the pressure loss, use the Applicator sizing tool → 76
It is important that cavitation does not occur, or that gases entrained in the liquids do not outgas.
This is prevented by means of a sufficiently high system pressure.
For this reason, the following mounting locations are recommended:
• At the lowest point in a vertical pipe
• Downstream from pumps (no danger of vacuum)
A0015594
In the case of some fluids, it is important that the heat radiated from the sensor to the transmitter is
kept to a minimum. A wide range of materials can be used for the required insulation.
Endress+Hauser53
aMinimum distance to insulation
tmaximum Insulation thickness
The minimum distance between the transmitter housing and the insulation is 10 mm (0.39 in) so
that the transmitter head remains completely exposed.
26Maximum recommended insulation thickness depending on the temperature of the medium and the
ambient temperature
tInsulation thickness
T
m
T
40(104)
T
60(140)
Medium temperature
Maximum recommended insulation thickness at an ambient temperature of Ta = 40 °C (104 °F)
Maximum recommended insulation thickness at an ambient temperature of Ta = 60 °C (140 °F)
A0019919
A0023177
Page 54
Proline Promass H 100
NOTICE
Danger of overheating with insulation
Ensure that the temperature at the lower end of the transmitter housing does not exceed
‣
80 °C (176 °F)
NOTICE
The insulation can also be thicker than the maximum recommended insulation thickness.
Prerequisite:
Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣
Ensure that a sufficiently large area of the housing support remains exposed. The uncovered part
‣
serves as a radiator and protects the electronics from overheating and excessive cooling.
Heating
Vibrations
Some fluids require suitable measures to avoid loss of heat at the sensor.
Heating options
• Electrical heating, e.g. with electric band heaters
• Via pipes carrying hot water or steam
• Via heating jackets
Heating jackets for the sensors can be ordered as accessories from Endress+Hauser → 75.
NOTICE
Danger of overheating when heating
Ensure that the temperature at the lower end of the transmitter housing does not exceed
‣
80 °C (176 °F).
Ensure that convection takes place on a sufficiently large scale at the transmitter neck.
‣
Ensure that a sufficiently large area of the housing support remains exposed. The uncovered part
‣
serves as a radiator and protects the electronics from overheating and excessive cooling.
The high oscillation frequency of the measuring tubes ensures that the correct operation of the
measuring system is not influenced by plant vibrations.
54Endress+Hauser
Page 55
Proline Promass H 100
A
DC
B
E
H
F
L
G
K
Mechanical construction
Dimensions in SI unitsCompact version
Order code for "Housing", option A "Compact coated alu"
A0019436
DN
[mm]A[mm]B[mm]C[mm]D[mm]
8136147.593.5542613698.51108
15136147.593.55426136912.0108
25136147.593.55426138217.6121
40136147.593.55428545825.5173
50136147.593.55429653740.5241
1)If using a display, order code for "Display; Operation", option B: values + 28 mm
2)Dependent on the specific process connection
1)
E
[mm]
1)
F
[mm]G[mm]H[mm]L[mm]K[mm]
2)
2)
2)
2)
2)
92
92
92
132
167
Endress+Hauser55
Page 56
Order code for "Housing", option B "Compact hygienic, stainless"
DC
B
A
E
H
F
L
G
K
Proline Promass H 100
A0019437
DN
[mm]A[mm]B[mm]C[mm]D[mm]
1)
E
[mm]
8133.5136.87858.82573658.51108
15133.5136.87858.825736512.0108
25133.5136.87858.825737817.6121
40133.5136.87858.828145425.5173
50133.5136.87858.829253340.5241
1)
F
[mm]G[mm]H[mm]L[mm]K[mm]
2)
2)
2)
2)
2)
92
92
92
132
167
1)If using a display, order code for "Display; Operation", option B: values + 14 mm
2)Dependent on the specific process connection
56Endress+Hauser
Page 57
Proline Promass H 100
ADC
B
E
H
F
G
K
L
C
D
A
B
E
L
+1.5 (+0.06)
−2.0 (–0.08)
±3.5 (±0.14)
DN 125 (5")≥DN 100 (4")≤
Order code for "Housing", option C "Ultra-compact hygienic, stainless"
A0019438
DN
[mm]A[mm]B[mm]C[mm]D[mm]
1)
E
[mm]
8114.9123.667.755.92563648.51108
15114.9123.667.755.925636412.0108
25114.9123.667.755.925637717.6121
40114.9123.667.755.928045325.5173
50114.9123.667.755.929153240.5241
1)
F
[mm]G[mm]H[mm]L[mm]K[mm]
2)
2)
2)
2)
2)
92
92
92
132
167
1)If using a display, order code for "Display; Operation", option B: values + 14 mm
2)Dependent on the specific process connection
Flange connections
Fixed flange EN 1092-1, ASME B16.5, JIS B2220
27Engineering unit mm (in)
A0015621
Endress+Hauser57
Page 58
Proline Promass H 100
Flange according to EN 1092-1 (DIN 2501): PN 40
1.4404 (316/316L)
Order code for "Process connection", option D2W
DN
[mm]
1)
8
1595654 × Ø142017.3440
25115854 × Ø1419.028.5580
401501104 × Ø1821.543.1794
501651254 × Ø1823.554.51071
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 µm
1)DN 8 with DN 15 flanges as standard
Flange according to ASME B16.5: Cl 150
1.4404 (316/316L)
Order code for "Process connection", option AAW
DN
[mm]
1)
8
159060.34 × Ø15.712.815.7440
2511079.44 × Ø15.715.126.7580
4012598.44 × Ø15.717.540.9794
50150120.74 × Ø19.123.652.61071
Surface roughness (flange): Ra 3.2 to 6.3 µm
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
95654 × Ø142017.3336
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
9060.34 × Ø15.712.815.7336
L
[mm]
L
[mm]
1)DN 8 with DN 15 flanges as standard
Flange according to ASME B16.5: Cl 300
1.4404 (316/316L)
Order code for "Process connection", option ABW
DN
[mm]
1)
8
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
9566.74 × Ø15.714.215.7336
159566.74 × Ø15.714.215.7440
2512588.94 × Ø19.117.526.7580
40155114.34 × Ø22.320.640.9794
50165127.08 × Ø19.123.652.61071
Surface roughness (flange): Ra 3.2 to 6.3 µm
1)DN 8 with DN 15 flanges as standard
Flange JIS B2220: 20K
1.4404 (316/316L)
Order code for "Process connection", option NEW
DN
[mm]
1)
8
A
[mm]
B
[mm]
C
[mm]
D
[mm]
E
[mm]
95704 × Ø151415336
1595704 × Ø151415440
25125904 × Ø191625580
L
[mm]
L
[mm]
58Endress+Hauser
Page 59
Proline Promass H 100
B
D
A
C
H
A
G
L
Flange JIS B2220: 20K
1.4404 (316/316L)
Order code for "Process connection", option NEW
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [kg].
Weight in US units
All values (weight) refer to devices with EN/DIN PN 40 flanges. Weight information in [lbs].
DN
[mm]
810
1511
2517
4034
5067
DN
[in]
3/822
½24
137
1½75
2148
Weight [kg]
Weight [lbs]
64Endress+Hauser
Page 65
Proline Promass H 100
1
2
3
Safety Barrier Promass 100
49 g (1.73 ounce)
MaterialsTransmitter housing
• Order code for "Housing", option A "Compact, aluminum coated":
Aluminum, AlSi10Mg, coated
• Order code for "Housing", option B "Compact, stainless":
Stainless steel 1.4301 (304)
• Order code for "Housing", option C "Ultra-compact, stainless":
Stainless steel 1.4301 (304)
• Window material for optional local display (→ 67):
– For order code for "Housing", option A: glass
– For order code for "Housing", option B and C: plastic
Cable entries/cable glands
29Possible cable entries/cable glands
1Cable entry in transmitter housing with internal thread M20 x 1.5
2Cable gland M20 x 1.5
3Adapter for cable entry with internal thread G ½" or NPT ½"
Order code for "Housing", option A "Compact, coated aluminum"
The various cable entries are suitable for hazardous and non-hazardous areas.
Cable entry/cable glandMaterial
Cable gland M20 × 1.5Nickel-plated brass
Adapter for cable entry with internal thread G ½"
Adapter for cable entry with internal thread NPT ½"
Order code for "Housing", option B "Compact, stainless"
The various cable entries are suitable for hazardous and non-hazardous areas.
• Flanges according to EN 1092-1 (DIN 2501) / according to ASME B16.5 / according to JIS B2220
List of all available process connections → 66
Seals
Welded process connections without internal seals
Safety Barrier Promass 100
Housing: Polyamide
Process connections
Surface roughness
Operating concept
Fixed flange connections:
– EN 1092-1 (DIN 2501) flange
– EN 1092-1 (DIN 2512N) flange
– ASME B16.5 flange
– JIS B2220 flange
For information on the different materials used in the process connections → 65
All data relate to parts in contact with fluid.
Not polished
Operability
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Quick and safe commissioning
• Individual menus for applications
• Menu guidance with brief explanations of the individual parameter functions
Reliable operation
• Operation in the following languages:
– Via "FieldCare" operating tool:
English, German, French, Spanish, Italian, Chinese, Japanese
– Via integrated Web browser (only available for device versions with HART, PROFIBUS DP,
PROFINET and EtherNet/IP):
English, German, French, Spanish, Italian, Dutch, Portuguese, Polish, Russian, Turkish, Chinese,
Japanese, Bahasa (Indonesian), Vietnamese, Czech, Swedish, Korean
• Uniform operating philosophy applied to operating tools and Web browser
• If replacing the electronic module, transfer the device configuration via the plug-in memory
(HistoROM DAT) which contains the process and measuring device data and the event logbook. No
need to reconfigure.
For devices with Modbus RS485, the data recovery function is implemented without the plug-in
memory (HistoROM DAT).
• Troubleshooting measures can be called up via the operating tools and Web browser
• Diverse simulation options
• Status indicated by several light emitting diodes (LEDs) on the electronic module in the housing
compartment
Local display
A local display is only available for device versions with the following communication protocols:
HART, PROFIBUS-DP, PROFINET, EtherNet/IP
The local display is only available with the following device order code:
Order code for "Display; Operation", option B: 4-line; lit, via communication
Display element
• 4-line liquid crystal display with 16 characters per line.
• White background lighting; switches to red in event of device errors.
• Format for displaying measured variables and status variables can be individually configured.
• Permitted ambient temperature for the display: –20 to +60 °C (–4 to +140 °F). The readability of
the display may be impaired at temperatures outside the temperature range.
Remote operationVia HART protocol
This communication interface is available in device versions with a HART output.
30Options for remote operation via HART protocol
1Control system (e.g. PLC)
2Field Communicator 475
3Computer with operating tool (e.g. FieldCare, AMS Device Manager, SIMATIC PDM)
4Commubox FXA195 (USB)
5Field Xpert SFX350 or SFX370
6VIATOR Bluetooth modem with connecting cable
7Transmitter
Via PROFIBUS DP network
This communication interface is available in device versions with PROFIBUS DP.
A0016948
Endress+Hauser67
Page 68
12
3
444
31Options for remote operation via PROFIBUS DP network
4
213
555
1Automation system
2Computer with PROFIBUS network card
3PROFIBUS DP network
4Measuring device
Proline Promass H 100
A0020903
Via Ethernet-based fieldbus
This communication interface is available in device versions with EtherNet/IP.
32Options for remote operation via Ethernet-based fieldbus
1Control system, e.g. "RSLogix" (Rockwell Automation)
2Workstation for measuring device operation: with Add-on Profile Level 3 for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
4Ethernet switch
5Measuring device
Via PROFINET network
This communication interface is available in device versions with PROFINET.
A0016961
68Endress+Hauser
Page 69
Proline Promass H 100
3
12
444
1
2
3
A0026545
33Options for remote operation via PROFINET network
1Automation system, e.g. Simatic S7 (Siemens)
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Switch, e.g. Scalance X204 (Siemens)
4Measuring device
Service interfaceVia service interface (CDI-RJ45)
This communication interface is present in the following device version:
• Order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
• Order code for "Output", option L: PROFIBUS DP
• Order code for "Output", option N: EtherNet/IP
• Order code for "Output", option R: PROFINET
HART
A0016926
34Connection for the order code for "Output", option B: 4-20 mA HART, pulse/frequency/switch output
1Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
Endress+Hauser69
Page 70
Proline Promass H 100
1
2
3
1
2
3
PROFIBUS DP
A0021270
35Connection for order code for "Output", option L: PROFIBUS DP
1Service interface (CDI -RJ45) of the measuring device with access to the integrated Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
EtherNet/IP
A0016940
36Connection for order code for "Output", option N: EtherNet/IP
1Service interface (CDI -RJ45) and EtherNet/IP interface of the measuring device with access to the integrated
Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
70Endress+Hauser
Page 71
Proline Promass H 100
1
2
3
1
2
3
PROFINET
A0016940
37Connection for order code for "Output", option R: PROFINET
1Service interface (CDI -RJ45) and PROFINET interface of the measuring device with access to the integrated
Web server
2Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or with
"FieldCare" operating tool with COM DTM "CDI Communication TCP/IP"
3Standard Ethernet connecting cable with RJ45 plug
Via service interface (CDI)
This communication interface is present in the following device version:
Order code for "Output", option M: Modbus RS485
Modbus RS485
1Service interface (CDI) of the measuring device
2Commubox FXA291
3Computer with "FieldCare" operating tool with COM DTM "CDI Communication FXA291"
A0016925
Certificates and approvals
CE mark
C-Tick symbol
Endress+Hauser71
The measuring system is in conformity with the statutory requirements of the applicable EC
Directives. These are listed in the corresponding EC Declaration of Conformity along with the
standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
The measuring system meets the EMC requirements of the "Australian Communications and Media
Authority (ACMA)".
Page 72
Proline Promass H 100
Ex approval
The measuring device is certified for use in hazardous areas and the relevant safety instructions are
provided in the separate "Safety Instructions" (XA) document. Reference is made to this document on
the nameplate.
The separate Ex documentation (XA) containing all the relevant explosion protection data is
available from your Endress+Hauser sales center.
ATEX/IECEx
Currently, the following versions for use in hazardous areas are available:
Ex ia
Category (ATEX)Type of protection
II2GEx ia IIC T6…T1 Gb
II2GEx ia IIC T6…T1 Gb or Ex ia IIB T6…T1 Gb
II1/2G, II2DEx ia IIC T6…T1 Ga/Gb or Ex ia IIB T6…T1 Ga/Gb
Ex tb IIIC Txx °C Db
II2G, II2DEx ia IIC T6…T1 Gb or Ex ia IIB T6…T1 Gb
Ex tb IIIC Txx °C Db
Ex nA
Category (ATEX)Type of protection
II3GEx nA IIC T6…T1 Gc or Ex nA IIC T5-T1 Gc
CSA
C
US
Currently, the following versions for use in hazardous areas are available:
IS (Ex i)
• Class I Division 1 Groups ABCD
• Class II Division 1 Groups EFG and Class III
NI (Ex nA)
Class I Division 2 Groups ABCD
HART certificationHART interface
The measuring device is certified and registered by the FieldComm Group. The measuring system
meets all the requirements of the following specifications:
• Certified according to HART 7
• The device can also be operated with certified devices of other manufacturers (interoperability)
Certification PROFIBUSPROFIBUS interface
The measuring device is certified and registered by the PROFIBUS User Organization (PNO). The
measuring system meets all the requirements of the following specifications:
• Certified in accordance with PROFIBUS PA Profile 3.02
• The device can also be operated with certified devices of other manufacturers (interoperability)
Certification PROFINETPROFINET interface
The measuring device is certified and registered by the PNO (PROFIBUS User Organization ). The
measuring system meets all the requirements of the following specifications:
• Certified according to:
– Test specification for PROFINET devices
– PROFINET Security Level 1 – Net load test
• The device can also be operated with certified devices of other manufacturers (interoperability)
72Endress+Hauser
Page 73
Proline Promass H 100
EtherNet/IP certification
Modbus RS485 certification
Pressure Equipment
Directive
The measuring device is certified and registered by the ODVA (Open Device Vendor Association).
The measuring system meets all the requirements of the following specifications:
• Certified in accordance with the ODVA Conformance Test
• EtherNet/IP Performance Test
• EtherNet/IP PlugFest compliance
• The device can also be operated with certified devices of other manufacturers (interoperability)
The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed
all the test procedures carried out and is certified by the “MODBUS/TCP Conformance Test
Laboratory” of the University of Michigan.
The devices can be ordered with or without a PED approval. If a device with a PED approval is
required, this must be explicitly stated in the order. For devices with nominal diameters less than or
equal to DN 25 (1"), this is neither possible nor necessary.
• With the PED/G1/x (x = category) marking on the sensor nameplate, Endress+Hauser confirms
compliance with the "Essential Safety Requirements" specified in Annex I of the Pressure
Equipment Directive 97/23/EC.
• Devices bearing this marking (PED) are suitable for the following types of medium:
– Media in Group 1 and 2 with a vapor pressure greater than, or smaller and equal to0.5 bar
(7.3 psi)
– Unstable gases
• Devices not bearing this marking (PED) are designed and manufactured according to good
engineering practice. They meet the requirements of Art.3 Section 3 of the Pressure Equipment
Directive 97/23/EC. The range of application is indicated in tables 6 to 9 in Annex II of the
Pressure Equipment Directive.
Other standards and
guidelines
• EN 60529
Degrees of protection provided by enclosures (IP code)
• IEC/EN 60068-2-6
Environmental influences: Test procedure - Test Fc: vibrate (sinusoidal).
• IEC/EN 60068-2-31
Environmental influences: Test procedure - Test Ec: shocks due to rough handling, primarily for
devices.
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use general requirements
• IEC/EN 61326
Emission in accordance with Class A requirements. Electromagnetic compatibility (EMC
requirements).
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment
• NAMUR NE 32
Data retention in the event of a power failure in field and control instruments with
microprocessors
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with
analog output signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics
• NAMUR NE 80
The application of the pressure equipment directive to process control devices
• NAMUR NE 105
Specifications for integrating fieldbus devices in engineering tools for field devices
• NAMUR NE 107
Self-monitoring and diagnosis of field devices
• NAMUR NE 131
Requirements for field devices for standard applications
• NAMUR NE 132
Coriolis mass meter
Endress+Hauser73
Page 74
Proline Promass H 100
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com → Select your
country → Products → Select measuring technology, software or components → Select the product
(picklists: measurement method, product family etc.) → Device support (right-hand column):
Configure the selected product → The Product Configurator for the selected product opens.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Application packages
Many different application packages are available to enhance the functionality of the device. Such
packages might be needed to address safety aspects or specific application requirements.
The application packages can be ordered with the device or subsequently from Endress+Hauser.
Detailed information on the order code in question is available from your local Endress+Hauser sales
center or on the product page of the Endress+Hauser website: www.endress.com.
Detailed information on the application packages:
• Special Documentation for the device
• Special Documentation for the device
Heartbeat Technology
PackageDescription
Heartbeat Verification
+Monitoring
Heartbeat Monitoring
Continuously supplies data, which are characteristic of the measuring principle, to
an external condition monitoring system for the purpose of preventive
maintenance or process analysis. These data enable the operator to:
• Draw conclusions - using these data and other information - about the impact
process influences (such as corrosion, abrasion, buildup etc.) have on the
measuring performance over time.
• Schedule servicing in time.
• Monitor the process or product quality, e.g. gas pockets.
Heartbeat Verification
Meets the requirement for traceable verification to DIN ISO 9001:2008 Chapter
7.6 a) "Control of monitoring and measuring equipment".
• Functional testing in the installed state without interrupting the process.
• Traceable verification results on request, including a report.
• Simple testing process via local operation or other operating interfaces.
• Clear measuring point assessment (pass/fail) with high test coverage within the
framework of manufacturer specifications.
• Extension of calibration intervals according to operator's risk assessment.
74Endress+Hauser
Page 75
Proline Promass H 100
Concentration
PackageDescription
Concentration
measurement and special
density
Calculation and outputting of fluid concentrations
Many applications use density as a key measured value for monitoring quality or
controlling processes. The device measures the density of the fluid as standard and
makes this value available to the control system.
The "Special Density" application package offers high-precision density
measurement over a wide density and temperature range particularly for
applications subject to varying process conditions.
With the help of the "Concentration Measurement" application package, the
measured density is used to calculate other process parameters:
• Temperature-compensated density (reference density).
• Percentage mass of the individual substances in a two-phase fluid.
(Concentration in %).
• Fluid concentration is output with special units (°Brix, °Baumé, °API, etc.) for
standard applications.
The measured values are output via the digital and analog outputs of the device.
Accessories
Various accessories, which can be ordered with the device or subsequently from Endress+Hauser, are
available for the device. Detailed information on the order code in question is available from your
local Endress+Hauser sales center or on the product page of the Endress+Hauser website:
www.endress.com.
Device-specific accessoriesFor the sensor
AccessoriesDescription
Heating jacketIs used to stabilize the temperature of the fluids in the sensor.
Communication-specific
accessories
AccessoriesDescription
Commubox FXA195
HART
Commubox FXA291Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser
HART Loop Converter
HMX50
Wireless HART adapter
SWA70
Water, water vapor and other non-corrosive liquids are permitted for use as fluids.
If using oil as a heating medium, please consult with Endress+Hauser.
For details, see Operating Instructions BA00099D
For intrinsically safe HART communication with FieldCare via the USB interface.
For details, see "Technical Information" TI00404F
Common Data Interface) and the USB port of a computer or laptop.
For details, see the "Technical Information" document TI405C/07
Is used to evaluate and convert dynamic HART process variables to analog current
signals or limit values.
For details, see "Technical Information" TI00429F and Operating Instructions
BA00371F
Is used for the wireless connection of field devices.
The WirelessHART adapter can be easily integrated into field devices and existing
infrastructures, offers data protection and transmission safety and can be operated
in parallel with other wireless networks with minimum cabling complexity.
For details, see Operating Instructions BA00061S
Fieldgate FXA320Gateway for the remote monitoring of connected 4-20 mA measuring devices via a
Web browser.
For details, see "Technical Information" TI00025S and Operating Instructions
BA00053S
Endress+Hauser75
Page 76
Proline Promass H 100
Fieldgate FXA520Gateway for the remote diagnostics and remote configuration of connected HART
measuring devices via a Web browser.
For details, see "Technical Information" TI00025S and Operating Instructions
BA00051S
Field Xpert SFX350Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices in the non-Ex area.
For details, see Operating Instructions BA01202S
Field Xpert SFX370Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
Fieldbus devices in the non-Ex area and the Ex area.
For details, see Operating Instructions BA01202S
Service-specific accessories
AccessoriesDescription
ApplicatorSoftware for selecting and sizing Endress+Hauser measuring devices:
• Calculation of all the necessary data for identifying the optimum flowmeter: e.g.
nominal diameter, pressure loss, accuracy or process connections.
• Graphic illustration of the calculation results
Administration, documentation and access to all project-related data and
parameters throughout the entire life cycle of a project.
Applicator is available:
• Via the Internet: https://wapps.endress.com/applicator
• On CD-ROM for local PC installation.
W@MLife cycle management for your plant
W@M supports you with a wide range of software applications over the entire
process: from planning and procurement, to the installation, commissioning and
operation of the measuring devices. All the relevant device information, such as
the device status, spare parts and device-specific documentation, is available for
every device over the entire life cycle.
The application already contains the data of your Endress+Hauser device. Endress
+Hauser also takes care of maintaining and updating the data records.
W@M is available:
• Via the Internet: www.endress.com/lifecyclemanagement
• On CD-ROM for local PC installation.
FieldCareFDT-based plant asset management tool from Endress+Hauser.
It can configure all smart field units in your system and helps you manage them. By
using the status information, it is also a simple but effective way of checking their
status and condition.
For details, see Operating Instructions BA00027S and BA00059S
DeviceCareTool for connecting and configuring Endress+Hauser field devices.
For details, see Innovation brochure IN01047S
Commubox FXA291Connects Endress+Hauser field devices with a CDI interface (= Endress+Hauser
Common Data Interface) and the USB port of a computer or laptop.
For details, see "Technical Information" TI00405C
76Endress+Hauser
Page 77
Proline Promass H 100
System components
AccessoriesDescription
Memograph M graphic
display recorder
The Memograph M graphic display recorder provides information on all relevant
measured variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on a SD card or USB stick.
iTEMPThe temperature transmitters can be used in all applications and are suitable for
the measurement of gases, steam and liquids. They can be used to read in the fluid
temperature.
Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer : Enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.
Standard documentationBrief Operating Instructions
Brief Operating Instructions containing the most important information for standard
commissioning are supplied with the device.
For details, see "Technical Information" TI00133R and Operating Instructions
BA00247R
For details, see "Fields of Activity", FA00006T
Supplementary devicedependent documentation
Operating Instructions
Documentation code
Measuring device
Promass H 100BA01189DBA01250DBA01177DBA01184DBA01428D