Endress+Hauser Gammapilot M FMG60 Operating Instructions Manual

Page 1
Products Solutions Services
BA00236F/00/EN/14.14 71253004
Valid for software version: V01.03.06
Operating Instructions
Gammapilot M FMG60
AUTHORIZED DISTRIBUTOR: InstrumentsAndControl.com Houston, Texas USA sales@InstrumentsAndControl.com 281-609-7170
Page 2
Endress+Hauser
Short instructions
L00-FMG60xxx-05-00-00-en-050
Contents of the Operating Instructions
These Operating Instructions describe how to install and commission the radiometric compact transmitter Gammapilot M (communication version 4-20 mA with HART). All the functions that are required for standard measuring tasks have been included. Also, the Gammapilot M provides additional functions for optimizing the measuring point and for converting the measured value. These functions are not included in these Operating Instructions.
You can find an overview of all the device functions in the Appendix.
A detailed description of all the instrument functions is given in Operating Instructions BA00287F/00/EN "Gammapilot M - Description of Device Functions" which can be found on the CD-ROM supplied.
E
+
-
+
E
+
-
-
……
……
100 (HART) 2457 (PA,FF)
333 (HART) 33333 (PA,FF)
……
E
E
E
E
52025626
KA242F/00/a2/02.07
*00
*0
*2
*C
*4
*5
*6
*A
*A0
*01
*02 *03 *04
*06 *09
*0A
*0B
*A1
*A4
*3
*9 *92
*A3
*C0
*05
*1 *10 *11 *12 *13
*14
*16*15 *11
*19
*17
*18*15 *11
*1B*11 *1C *1D *1E *1F
*1A
*15
*19
++
*07 *08
slave, end slave
safety settings
system parameters
linearisation
gammagraphy
output (HART, FF) profibus param.(PA)
diagnostics
present error
previous error
unlock parameter
temp. compens.
display language
reset
tag marking
density/concentration
density calibr.
stop/edit
start
avg. pulse rate
density value
calibr. point
absorp. coeff.
ref. pulse rate
calibr. point
calibration
next point
no
yes
Contrast: or
*indicates the meas.mode: L: Level S: Level limit D: Density C: Concentration
calibration backgr.
calib
start
avg. pulse rate
bgr. pulse rate
stop/edit
calibr. point
value full
full calibr.
stop/edit
calibration
start
avg. pulse rate
next point
value empty
empty calibr.
stop/edit
calibration
start
avg. pulse rate
no
yes
level/
limit
density unit
pipe diam. unit
pipe diam.
output damping
min. density
max. density
measured value
Group selection
basic setup
todays date
beam type isotope operating
mode
meas. mode
Gammapilot M - Quick Setup
Page 3
Gammapilot M FMG60 HART
Endress+Hauser 3
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Installation, commissioning, operation . . . . . . . . . 4
1.3 Hazardous area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Radiation protection . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5 Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2 Identification . . . . . . . . . . . . . . . . . . . . . . 8
2.1 Parts of the Gammapilot M . . . . . . . . . . . . . . . . . . . 8
2.2 Nameplates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.4 Supplied documentation . . . . . . . . . . . . . . . . . . . . . 9
2.5 Certificates and approvals . . . . . . . . . . . . . . . . . . 10
2.6 Registered trademarks . . . . . . . . . . . . . . . . . . . . . 10
3 Installation . . . . . . . . . . . . . . . . . . . . . . . 11
3.1 Incoming acceptance, transport, storage . . . . . . 11
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . 12
3.3 Water cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.4 Installation check . . . . . . . . . . . . . . . . . . . . . . . . . 19
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.1 Terminal compartments . . . . . . . . . . . . . . . . . . . . 20
4.2 Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4.3 Terminal assignment . . . . . . . . . . . . . . . . . . . . . . 21
4.4 Fieldbus connectors . . . . . . . . . . . . . . . . . . . . . . . 23
4.5 Potential equalization . . . . . . . . . . . . . . . . . . . . . . 24
4.6 Wiring in terminal compartment 1 . . . . . . . . . . 25
4.7 Wiring in terminal compartment 2 . . . . . . . . . . 26
4.8 Connecting the
remote display and operation FHX40 . . . . . . . . . 27
4.9 Wiring in cascading mode . . . . . . . . . . . . . . . . . . 28
4.10 Wiring example for
level limit detection 200/400 mm . . . . . . . . . . . 29
4.11 Measuring solids flow . . . . . . . . . . . . . . . . . . . . . . 30
4.12 Post-connection check . . . . . . . . . . . . . . . . . . . . . 30
5 Operation. . . . . . . . . . . . . . . . . . . . . . . . . 31
5.1 Overview over the operating options . . . . . . . . . 31
5.2 Display operation . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.3 Alternative operation options . . . . . . . . . . . . . . . 36
5.4 Lock/unlock configuration . . . . . . . . . . . . . . . . . . 37
5.5 Reset to the default configuration . . . . . . . . . . . 37
6 Commissioning. . . . . . . . . . . . . . . . . . . . 38
6.1 Calibration: overview . . . . . . . . . . . . . . . . . . . . . . 38
6.2 Switching on the device . . . . . . . . . . . . . . . . . . . . 39
6.3 Basic setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
6.4 Calibration for
level measurement and limit detection . . . . . . . 46
6.5 Function group "Safety settings" (*2) . . . . . . . . . 55
6.6 SIL locking (for level limit detection
200/400 mm PVT scintillator) . . . . . . . . . . . . . . . 56
6.7 Calibration for density and concentration
measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.8 Density measurement/
temperature-compensated . . . . . . . . . . . . . . . . . . 69
6.9 Gammagraphy detection . . . . . . . . . . . . . . . . . . . . 69
7 Maintenance and repairs. . . . . . . . . . . 70
7.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7.2 Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7.3 Repairs to Ex- or SIL-approved devices . . . . . . . . 70
7.4 Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
7.5 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.6 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
7.7 Contact addresses of Endress+Hauser . . . . . . . . . 71
8 Accessories. . . . . . . . . . . . . . . . . . . . . . . 72
8.1 Commubox FXA195 HART . . . . . . . . . . . . . . . . . . 72
8.2 Commubox FXA291 . . . . . . . . . . . . . . . . . . . . . . . . 72
8.3 ToF adapter FXA291 . . . . . . . . . . . . . . . . . . . . . . . 72
8.4 Field Xpert SFX100 . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.5 Remote display FHX40 . . . . . . . . . . . . . . . . . . . . . . 73
8.6 Mounting device FHG60 (for level measurement and level limit measurement) . . . 75
8.7 Clamping device for
density measurement FHG61 . . . . . . . . . . . . . . . . 77
8.8 Measurement section for
density measurement FHG62 . . . . . . . . . . . . . . . . 77
9 Troubleshooting . . . . . . . . . . . . . . . . . . 78
9.1 System error messages . . . . . . . . . . . . . . . . . . . . . 78
9.2 Possible calibration errors . . . . . . . . . . . . . . . . . . . 81
9.3 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . 82
10 Technical Data. . . . . . . . . . . . . . . . . . . . 83
10.1 Additional technical data . . . . . . . . . . . . . . . . . . . . 83
10.2 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
11 Appendix . . . . . . . . . . . . . . . . . . . . . . . . 88
11.1 Operating menu for level measurements . . . . . . 88
11.2 Operating menu for level limit detection . . . . . . . 90
11.3 Operating menu for density and concentration
measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Page 4
Safety instructions Gammapilot M FMG60 HART
4 Endress+Hauser
1 Safety instructions
1.1 Designated use
The Gammapilot M is a compact transmitter for non-contact level, level limit, density and concentration measurement. The measuring range of a single Gammapilot M extends up to 2 m (6.6 ft). Large measuring ranges of any size can be implemented, however, by cascading several Gammapilot M devices. When used for level limit detection, the Gammapilot M is certified according to IEC 61508 for safety-related operation up to SIL 2/3.
1.2 Installation, commissioning, operation
The Gammapilot M is fail-safe and is constructed to the state-of-the-art. It meets the appropriate standards and EC directives. However, if you use it improperly or other than for its designated use, it may pose application-specific hazards, e.g. product overflow due to incorrect installation or configuration. Installation, electrical connection, startup, operation and maintenance of the measuring device must therefore be carried out exclusively by trained specialists authorized by the system operator. Technical personnel must have read and understood these Operating Instructions and must adhere to them. You may only undertake modifications or repair work to the device when it is expressly permitted by the Operating Instructions.
1.3 Hazardous area
Measuring systems for use in hazardous environments are accompanied by separate "Ex documentation", which is an integral part of these Operating Instructions. Strict compliance with the installation instructions and ratings as stated in this supplementary documentation is mandatory.
• Ensure that all personnel are suitably qualified.
• Observe the specifications in the certificate as well as national and local standards and regulations.
CAUTION
!
Detector or cooling jacket can be damaged if the cooling water freezes.
Empty cooling jacket or protect against freezing.
WARNING
!
Depending on the certificate version observe the associated Safety Instructions (ä 83).
WARNING
!
The three screws, connecting the pipe housing to the compartment housing, must not be opened.
A0018068
Page 5
Gammapilot M FMG60 HART Safety instructions
Endress+Hauser 5
1.4 Radiation protection
The Gammapilot M is used in conjunction with a radioactive source, contained in a source container. When handling radioactive sources, the following instructions have to be observed:
1.4.1 Basic regulations on radiation protection
WARNING
!
When handling radioactive sources, all unnecessary radiation exposure should be avoided. All unavoidable radiation exposure should be kept as low as possible. Three measures are used for this:
A0016373
A Screening BTime CDistance
Shielding
Ensure the best possible shielding between the radiation source and yourself as well as all other individuals. Effective shielding is provided by source containers (FQG60, FQG61/ FQG62, FQG63, QG2000) and all high-density materials (lead, iron, concrete).
CAUTION
!
When working with source containers, all the instructions for mounting and usage outlined in the following documents must be observed:
Time
Stay as short as possible in the area exposed to radiation.
Distance
Keep as far away as possible from the radiation source. The local radiation intensity decreases as the square-root of the distance from the radiation source.
A
B
C
Source Container Document
FQG60 TI00445F/00/EN
FQG61, FQG62 TI00435F/00/EN
FQG63 TI00446F/00/EN
QG2000 TI00346F/00/EN
BA00223F/00/EN
Page 6
Safety instructions Gammapilot M FMG60 HART
6 Endress+Hauser
1.5 Symbols
1.5.1 Safety symbols
1.5.2 Electrical symbols
1.5.3 Tool symbols
Symbol Meaning
A0011189-DE
DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury.
A0011190-DE
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
A0011191-DE
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minor or medium injury.
A0011192-DE
NOTICE!
This symbol contains information on procedures and other facts which do not result in personal injury.
Symbol Meaning
A0018338
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system..
A0018339
Protective ground connection
A terminal which must be connected to ground prior to establishing any other connections.
Symbol Meaning
A0011221
Allen key
DANGER
WARNING
CAUTION
NOTICE
Page 7
Gammapilot M FMG60 HART Safety instructions
Endress+Hauser 7
1.5.4 Symbols for certain types of information
1.5.5 Symbols in graphics
Symbol Meaning
A0011182
Allowed
Indicates procedures, processes or actions that are allowed.
A0011183
Preferred
Indicates procedures, processes or actions thar are preferred.
A0011184
Forbidden
Indicates procedures, processes or actions that are forbidden.
A0011193
Tip
Indicates additional information.
A0015484
Reference to page
Refers to the corresponding page number.
, , ... Series of steps
Symbol Meaning
1, 2, 3, 4, ... Item numbers
, , ... Series of steps
A, B, C, D, ... Views
A0011187
Hazardous area
Indicates a hazardous area.
A0011188
Safe area (non-hazardous area)
Indicates a non-hazardous location.
-
.
Page 8
Identification Gammapilot M FMG60 HART
8 Endress+Hauser
2 Identification
2.1 Parts of the Gammapilot M
1. Terminal head
2. Mounting collar
3. Detector pipe
4. Measuring range marks
5. Terminal compartment 2
6. Terminal compartment 1
7. Supplementary nameplate
8. Centering knob
9. Instrument nameplate
10. Cooling water connections
11. Marking of the radiation
window (only for versions with collimator)
12. Potential equalization
terminal of the water cooling jacket
A0018069
A Gammapilot M without water cooling jacket B Gammapilot M with water cooling jacket C Gammapilot M with collimator
1
2
3
6
5
8
4
4
9
7
4
4
10
10
AB/C
12
11
C
Page 9
Gammapilot M FMG60 HART Identification
Endress+Hauser 9
2.2 Nameplates
2.3 Scope of delivery
• Ordered version of the device (including Operating Instructions)
• Endress+Hauser operating program (on the CD-ROM supplied)
• Accessories as ordered
2.4 Supplied documentation
2.4.1 Operating Instructions (BA00236F/00/EN)
Describe how to install and commission the Gammapilot M (communication version 4-20 mA HART). Those functions of the operating menu are included, which are required for standard measuring tasks. Any additional functions are contained in the "Description of Device Functions" (BA00287F/00/EN).
2.4.2 Description of Device Functions (BA00287F/00/EN)
Contains a detailed description of all the functions of the Gammapilot M and applies for all communication versions. This document can be found as a PDF file on the CD-ROM supplied and can be downloaded from the Internet under "www.de.endress.com" ( download).
2.4.3 Safety instructions
Additional safety instructions (XA, ZE, ZD) are supplied with certified device versions. Refer to the nameplate for the names of the safety instructions that apply to your device variant. An overview of the certificates and approvals can be found in TI00363F/00/EN, chapter "Certificates and approvals".
Device nameplate
A0018070
1 Order code (as defined by the Ordering information) 2 Serial number 3 Measuring range 4Power supply 5Output signal 6 Max. ambient temperature
Supplementary nameplate (Examples)
A0018070
7 No. of certificate 8Equipment group and category
9 Type of protection 10 Reference to additional safety-relevant information 11 Reference to additional certificates (e.g. WHG, SIL) 12 Specification of required temperature resistance of
the connected cables (only for instrument versions with water cooling jacket)
Ser.-No.:
Order Code:
IP 66/67
Gammapilot M
Made in Germany, D-79689 Maulburg
L=
mm
Dat. xx.xx
250001828- -
TYPE 4X / 6 Encl.
Ta
-40°C
°C
1 2
3
4
5
6
000000- -
KEMA 04 ATEX
11
10
7
8
9
000000- -
11
10
T Ta + 20 K
T Ta + 20 K
0044
Page 10
Identification Gammapilot M FMG60 HART
10 Endress+Hauser
2.5 Certificates and approvals
CE mark, declaration of conformity
The device is designed to meet state-of-the-art safety requirements, has been tested and left the factory in a condition in which it is safe to operate. The device takes into account applicable standards and regulations which are listed in the EC declaration of conformity and thus meets the legal requirements of the EC Directives. Endress+Hauser confirms the successful testing of the instrument by affixing to it the CE mark.
2.6 Registered trademarks
HART
®
Registered trademark of the HART Communication Foundation, Austin, USA
ToF
®
Registered trademark of Endress+Hauser GmbH+Co. KG, Maulburg, Germany
Page 11
Gammapilot M FMG60 HART Installation
Endress+Hauser 11
3Installation
3.1 Incoming acceptance, transport, storage
3.1.1 Incoming acceptance
Check the packing and contents for any signs of damage. Check the shipment, make sure nothing is missing and that the scope of supply matches your order.
3.1.2 Transport
CAUTION
!
Follow the safety instructions and transport conditions for devices of more than 18 kg (39.69 lbs).
3.1.3 Storage
Pack the measuring instrument so that it is protected against impacts for storage and transport. The original packing material provides the optimum protection for this. The permissible storage temperature is:
• -40 to +50 °C (-40 to +122 °F) for devices with PVT scintillator
• -40 to +60 °C (-40 to +140 °F) for devices with NaI crystal
Page 12
Installation Gammapilot M FMG60 HART
12 Endress+Hauser
3.2 Installation conditions
3.2.1 Dimensions, weight
Gammapilot M (without water cooling jacket)
A0018072
Gammapilot M with water cooling jacket or collimator
A0018073
Type
Measuring
length A
[mm (in)]
Without water cooling
jacket
With water cooling jacket
Overall
length
B [mm (in)]
Weight
[kg (lbs)]
1)
1) The given weight data are for 316L version. The weight of the aluminium version is reduced by 5,3 kg (11.69 lbs).
Overall
length
B [mm (in)]
Weight
without water
[kg (lbs)]
)
Weight
including
water
[kg (lbs)]
)
NaI 50 (1.97) 621 (24.4) 14 (30.87) 631 (24.8) 18 (39.69) 20 (44.10)
NaI with collimator 50 (1.97) 663 (26.1) 35 (77.18)  
PVT 200 (7.87) 780 (30.7) 15 (33.08) 790 (31.1) 20 (44.10) 24 (52.92)
PVT 400 (15.7) 980 (38.6) 16 (35.28) 990 (39) 23 (50.72) 29 (63.95)
PVT 800 (31.5) 1380 (54.3) 20 (44.10) 1390 (54.7) 31 (68.36) 40 (88.20)
PVT 1200 (47.5) 1780 (70.1) 24 (52.92) 1790 (70.5) 37 (81.59) 50 (110.25)
PVT 1600 (63) 2180 (85.8) 28 (61.74) 2190 (86.2) 45 (99.23) 61 (134.51)
PVT 2000 (7.87) 2580 (102) 31 (68.36) 2590 (102) 51 (112.46) 72 (158.76)
160x190
(6.3x7.48)
ø120
(4.72)
ø80
(3.15)
537 (22.6)
mm (in)
A
B
160x190
(6.3x7.48)
537 (22.6)
A
B
164 (6.46)
mm (in)
12 (0.47)
ø140
(5.51)
Page 13
Gammapilot M FMG60 HART Installation
Endress+Hauser 13
3.2.2 Installation conditions for level measurement
Conditions
• For level measurements the Gammapilot M is mounted vertically; if possible the detector head should point downwards.
• The exit angle of the source container must be exactly aligned to the measuring range of the Gammapilot M. Observe the measuring range marks of the Gammapilot M.
• In cascading mode no gap should occur between the measuring ranges of the different Gammapilot M.
• The source container and the Gammapilot M must be mounted as close to the vessel as possible. Any access to the beam must be blocked so that no persons or part of their body (hand, arm, head) may come into the area of the beam.
• In order to enlarge the lifetime, the Gammapilot M should be protected against direct sun. If necessary, a protective cover should be applied.
• The mounting device FHG60 (ä 72, "Accessories") or an equivalent mounting device should be used for fastening the Gammapilot M. The mounting device must be installed in a way such that it can withstand the weight of the Gammapilot M
1)
under all operating conditions (e.g. vibrations).
NOTICE
The Gammapilot M should be given additional support to prevent damage to the connecting cable or to the unit if it falls off.
Examples
A0018074
A Vertical cylinder; the Gammapilot M is mounted vertically with the detector head pointing downwards;
the gamma ray is aligned to the measuring range. B Cascading of multiple Gammapilot M; there is no gap between the measuring ranges C Wrong: Gammapilot M mounted inside the tank insulation D Conical tank outlet (here with sun protection cover) E Horizontal cylinder (here with sun protection cover) F Right: Tank insulation removed for Gammapilot M
1Support
1) The weights of the various versions of the Gammapilot M are summarized in the section "Dimensions, weight".
A
D
B C
EF
1
Page 14
Installation Gammapilot M FMG60 HART
14 Endress+Hauser
3.2.3 Installation conditions for level limit detection
Conditions
• For level limit detection, the Gammapilot M should be mounted horizontally at the height of the desired level limit.
• The exit angle of the source container must be exactly aligned to the measuring range of the Gammapilot M. Observe the measuring range marks of the Gammapilot M.
• The source container and the Gammapilot M must be mounted as close to the vessel as possible. Any access to the beam must be blocked so that no persons or part of their body (hand, arm, head) may come into the area of the beam.
• In order to enlarge the lifetime, the Gammapilot M should be protected against direct sun. If necessary, a protective cover should be applied.
• The mounting device FHG60 (ä 72, "Accessories") or an equivalent mounting device should be used for fastening the Gammapilot M. The mounting device itself must be fitted in a way such that it can withstand the weight of the Gammapilot M
2)
under all operating conditions to be expected.
More information with regard to the safety-related deployment of Gammapilot M can
be found in Functional Safety Manual SD00230F/00/EN and SD00324F/00/EN.
Examples
A0018075
A Maximum fail-safe mode B Minimum point level detection
2) The weights of the various versions of the Gammapilot M are summarized in the section "Dimensions, weight".
A
B
Page 15
Gammapilot M FMG60 HART Installation
Endress+Hauser 15
3.2.4 Installation conditions for density and concentration measurement
Conditions
• If possible, density and concentration should be measured at vertical pipes with a feed
direction from bottom to top.
• If only horizontal pipes are accessible, the path of the ray should also be arranged
horizontally to reduce the influence of air bubbles and sediments.
• The Endress+Hauser clamping device (ä 72, "Accessories") or an equivalent clamping
device should be used for fastening the radiation source container and the Gammapilot M to the measuring tube.The clamping device itself must be installed in a way such that it can withstand the weight of the source container
3)
and the Gammapilot M4) under all
operating conditions.
• The sample point may not be further than 20 m (66 ft) from the measuring point.
• The distance of the density measurement to pipe bends is 3 x pipe diameter and 10 x
pipe diameter for pumps.
Configuration of the measuring system
The configuration of the source container and the Gammapilot M depends on the pipe diameter (or the length of the irradiated measuring path respectively) and the measuring range. These two parameters determine the measuring effect (relative change of the pulse rate). The measuring effect increases with the length of the radiation path through the medium. Therefore, diagonal irradiation or the use of a measurement section is necessary for small pipe diameters.
For the configuration of the measuring system please contact your Endress+Hauser sales organization or use the "Applicator"
 5)
configuration software.
A0018076
A Vertical beam (90°) B Diagonal beam (30°) C Measurement section 1Sample point
NOTICE
General
To increase the accuracy for density measurements, the use of a collimator is
recommended. The collimator screens the detector against environmental radiation.
When planning, the total weight of the measuring system must be taken into
consideration.
The Gammapilot M should be given additional support to secure it against falling or
prevent damage to the connecting cable.
A clamping device and a measurement section are available as accessories (ä 72,
"Accessories").
3) The weights of the source containers are specified in TI00445F/00/EN (FQG60), TI00435F/00/EN (FQG61, FQG62), TI00446F/00/EN (FQG63) or TI00346F/00/EN (QG2000).
4) The weights of the various versions of the Gammapilot M are summarized in the section "Dimensions, weight".
5) The "Applicator"

is available from your Endress+Hauser sales organization.
ABC
1
Page 16
Installation Gammapilot M FMG60 HART
16 Endress+Hauser
3.2.5 Empty pipe detection
A0018077
1 Gammapilot M 2 Monitoring detector FTG20 or FMG60 3SPS
Mounting the FTG20 or FMG60 on the FMG60 for empty pipe detection
If the pipe becomes empty as a result of operational processes, the radiation on the detector side can hit dangerous levels.
• In such instances, the irradiation channel must be closed immediately for reasons of radiation protection.
• A high local dose rate also causes the detector unit (scintillator and photomultiplier) to age quickly.
The best way of avoiding such a situation is to mount a second radiometric measuring system that monitors the radiation intensity. If high radiation levels occur, an alarm is output and/or the source container is automatically switched off through pneumatic action for example.
1
2
3
Page 17
Gammapilot M FMG60 HART Installation
Endress+Hauser 17
3.3 Water cooling
For the versions of the Gammapilot with water cooling jacket, the following applies:
• Material: 316L
• Water connection: 2 x G 1/4"A, DIN ISO 228
• Inlet temperature: max. 40 °C (104 °F)
• Outlet temperature: max. 50 °C (122 °F) (temperature monitoring recommended)
• Water pressure: 4 to 6 bar (60 to 90 psi)
A0018078
A T <75 °C (167 °F) B T <120 °C (248 °F)
CAUTION
!
Detector or cooling jacket can be damaged if the cooling water freezes.
Empty cooling jacket or protect against freezing.
WARNING
!
Pressurized water cooling system!
Do not open the cylinder screws (see figure below) when pressurized.
A0023205
1 Cylinder screws
AB
G1/4"A
G1/4"A
*
1
CAUTION
!
To consider when using the water cooling jacket
It is recommended to ground the water
cooling jacket separately at the provided earth terminal (see picture above)
The ambient temperature of the
compartment housing must not exceed 75 °C (167 °F). This is also valid, if water cooling is applied.
The three screws, connecting the pipe
housing to the compartment housing, must not be opened.
A0018068
Page 18
Installation Gammapilot M FMG60 HART
18 Endress+Hauser
3.3.1 Mounting versions
A0018079
A Recommended mounting position for level measurement: compartment housing at the bottom B In exceptional cases (e.g. shortage of space) the compartment housing may be located at the top C Mounting position for point level detection and density measurement
CAUTION
!
The water inlet must always be at the bottom to ensure that the water cooling jacket is completely filled.
A
B C
Page 19
Gammapilot M FMG60 HART Installation
Endress+Hauser 19
3.3.2 Required flow rate
The required flow rate depends on
• the ambient temperature at the water cooling jacket
• the inlet temperature
• the measuring range of the Gammapilot M
Typical values are given in the following tables:
Ambient temperature T
A
=75°C (167°F)
Ambient temperature T
A
= 100 °C (212 °F)
Ambient temperature T
A
= 120 °C (248 °F)
3.4 Installation check
After installing the device, carry out the following checks:
• Is the device damaged (visual inspection)?
• Does the device correspond to the measuring point specifications for ambient temperature, measuring range etc.?
• If available: Are the measuring point number and labeling correct (visual inspection)?
• Is the measuring device sufficiently protected against direct sunlight?
• Are the cable glands tightened correctly?
Inlet temperature
°C (°F)
Measuring range in mm (in)
50 (1.97) 200 (7.87) 400 (15.7) 800 (31.5) 1200 (47.2)1600 (63) 2000 (78.7)
20 (68) 30 l/h 30 l/h 30 l/h 41 l/h 55 l/h 70 l/h 84 l/h
25 (77) 30 l/h 30 l/h 30 l/h 45 l/h 61 l/h 77 l/h 93 l/h
30 (86) 30 l/h 30 l/h 33 l/h 50 l/h 68 l/h 86 l/h 104 l/h
35 (95) 30 l/h 30 l/h 38 l/h 59 l/h 80 l/h 101 l/h 122 l/h
40 (104) 30 l/h 30 l/h 47 l/h 72 l/h 98 l/h 124 l/h 149 l/h
Inlet temperature
°C (°F)
Measuring range in mm (in)
50 (1.97) 200 (7.87) 400 (15.7) 800 (31.5) 1200 (47.2)1600 (63) 2000 (78.7)
20 (68) 30 l/h 30 l/h 38 l/h 59 l/h 80 l/h 101 l/h 122 l/h
25 (77) 30 l/h 30 l/h 42 l/h 64 l/h 87 l/h 110 l/h 133 l/h
30 (86) 30 l/h 30 l/h 47 l/h 73 l/h 98 l/h 124 l/h 150 l/h
35 (95) 30 l/h 30 l/h 54 l/h 84 l/h 113 l/h 143 l/h 173 l/h
40 (104) 33 l/h 33 l/h 66 l/h 101 l/h 137 l/h 173 l/h 210 l/h
Inlet temperature
°C (°F)
Measuring range in mm (in)
50 (1.97) 200 (7.87) 400 (15.7) 800 (31.5) 1200 (47.2) 1600 (63) 2000 (78.7)
20 (68) 30 l/h 30 l/h 45 l/h 70 l/h 94 l/h 119 l/h 144 l/h
25 (77) 30 l/h 30 l/h 50 l/h 77 l/h 104 l/h 131 l/h 158 l/h
30 (86) 30 l/h 30 l/h 55 l/h 85 l/h 115 l/h 146 l/h 176 l/h
35 (95) 32 l/h 32 l/h 64 l/h 98 l/h 133 l/h 168 l/h 203 l/h
40 (104) 38 l/h 38 l/h 75 l/h 116 l/h 157 l/h 199 l/h 240 l/h
Page 20
Wiring Gammapilot M FMG60 HART
20 Endress+Hauser
4 Wiring
4.1 Terminal compartments
4.2 Cable entries
NOTICE
Cable entries
On delivery, not more than one cable gland is present for each of the terminal
compartments. If further cable glands are required (e.g. for cascading mode), they must be supplied by the user.
Connecting cables should be routed away from the housing from below to prevent
moisture from penetrating the connection compartment. Otherwise, a drain loop should be provided or the Gammapilot M should be fitted with a weather protection cover.
The Gammapilot M has got two terminal compartments:
• Terminal compartment 1, for – Power supply – Signal output (depending on the instrument
version)
• Terminal compartment 2, for – Signal output (depending on the instrument
version) – PT-100 input (4-wire) – Pulse input for cascading mode – Pulse output for cascading mode – Display and operating module FHX40
(or VU331)
NOTICE
Depending on the instrument version, the signal output is located in the terminal compartment 1 or 2.
Maximum cable length:
• For cascade, 20 m (66 ft) each
• For PT-100 2 m (6.6 ft) (temperature should be measured as close as possible to density measurement)
A0018082
1
2
The number and type of cable entries depend on the instrument version ordered. The following types may occur:
• Gland M20x1.5 tightening diameter (clamping range):
7.0 to 10.5 mm
• Cable entry M20x1.5
• Cable entry G1/2
• Cable entry NPT1/2
• M12 connector (see "Fieldbus connectors")
• 7/8" connector (see "Fieldbus connectors")
In addition, Gammapilot M has a socket to connect the separate display and operating unit FHX40. The housing of the Gammapilot M does not have to be opened to connect the FHX40.
A0018083
1 Cable entries for terminal compartment 2 2 Socket for FHX40 3 Cable entries for terminal compartment 1
1 2
3
Page 21
Gammapilot M FMG60 HART Wiring
Endress+Hauser 21
4.3 Terminal assignment
Terminal compartment 1
A0018084
1 90 to 253VAC, 18 to 36 VDC
Terminal compartment 2
A0018085
1
14
15
2
(3)
L1L+N
L-
+-
(4)
1
0
9
3
++
+
5
7
10
4
--
-
6
11
12
13
SIM WP
PT100
IN CASCADE OUT
OFF
ON
8
)
Page 22
Wiring Gammapilot M FMG60 HART
22 Endress+Hauser
Terminal(s) Meaning
0 Grounding of the cable screen
1)
1) Rated cross section > 1 mm2 (17 AWG)
1, 2 Power supply
2)
2) Rated cross section max. 2.5 mm2 (14 AWG)
Compartment 2: 3, 4 Compartment 1: (3)
1
,(4)
1
Signal output, depending on communication version:
• 4-20mA with HART
•PROFIBUS PA
• Foundation Fieldbus
(Depending on the device version ordered, the signal output is in connection compartment 1 or 2, see below)
NOTICE
For the versions of the Gammapilot M with fieldbus plug connectors (M12 or 7/8"), the signal output is wired in compartment 2 on delivery and connected to the fieldbus plug connector (see below, section "Fieldbus connectors"). In this case, the housing needs not to be opened for connecting the signal line.
5, 6 Pulse input (for cascading mode; is used for master and slave)
7, 8 Pulse output (for cascading mode; is used for slave and end slave)
9, 10, 11, 12 PT-100 input (4-wire)
13 Plug for display and operating module VU331 (normally in FHX40);
is wired on delivery and connected to the FHX40 plug
14 Protective earth
)
15 Protective earth or grounding of the cable screen
)
The cables used at terminals 14 or 15 must at least have the same cross section as the cables at terminals 1 and 2.
Page 23
Gammapilot M FMG60 HART Wiring
Endress+Hauser 23
4.4 Fieldbus connectors
For the versions with a connector M12 or 7/8", the housing does not have to be opened for connecting the signal line.
4.4.1 Pin assignment for M12 connector
4.4.2 Pin assignment for 7/8" connector
Feature 30 of the ordering information: Power supply wiring/output wiring
Terminal compartment for
Supply voltage Signal output
ANon-Ex; Non-Ex 1 2
A0018082
BEx e; Ex ia 1 2
CEx e; Ex e 1 1
D Ex d (XP); Ex d (XP) 1 1
E Ex d (XP); Ex ia (IS) 1 2
F Dust-Ex; Dust-Ex 1 1
G Ex e, Dust-Ex; Ex e, Dust-Ex 1 1
H Ex d, Dust-Ex; Ex d, Dust-Ex 1 1
J Ex e, Dust-Ex; Ex ia, Dust-Ex 1 2
K Ex d, Dust-Ex; Ex ia, Dust-Ex 1 2
L Dust-Ex; Ex ia 1 2
A0011175
PIN Meaning
1Signal +
2Not assigned
3Signal
4Earth
A0011176
PIN Meaning
1Signal
2Signal +
3Shield
4Not assigned
1
2
21
3
4
2
1
4
3
Page 24
Wiring Gammapilot M FMG60 HART
24 Endress+Hauser
4.5 Potential equalization
Before wiring, connect the potential
matching line to the external ground terminal of the transmitter. If a water cooling jacket is present, it must be separately connected to the potential matching line.
CAUTION
!
In Ex applications, the instrument must only be grounded on the sensor side. Further safety instructions are given in the separate documentation for applications in explosion hazardous areas.
NOTICE
For optimum electromagnetic compatibility the potential matching line should be as short as possible and at least
2.5 mm
2
(14 AWG) in cross-section.
A0018086
*
Page 25
Gammapilot M FMG60 HART Wiring
Endress+Hauser 25
4.6 Wiring in terminal compartment 1
CAUTION
!
Before connection please note the following:
When using the instrument in hazardous areas, make sure to comply with national
standards and the specifications in the Safety Instructions (XAs). Make sure you use the specified cable gland.
The supply voltage must comply to the data on the nameplate.Switch off power supply before connecting the instrument.Connect potential matching line to the outer transmitter ground terminal and to the
ground terminal of the water cooling jacket (if present) before connecting up the instrument (see "Potential equalization", ä 24).
Connect protective earth to the protective earth terminal (see "Terminal compartments",
ä 20).
According to IEC/EN 61010 a suitable power switch has to be provided for the
instrument.
The cable isolations must comply with the supply voltage and the overvoltage category.The temperature resistance of the connecting cable must comply with the ambient
temperature.
The procedure
Using a 3 mm Allen key, loosen the
cover clamp for the connection compartment cover.
Unscrew the cover of the terminal
compartment.
Push the power cable and (if required)
the signal cable) through the appropriate cable glands or cable threads.
Wire up according to the terminal assignment diagram.
Tighten the cable glands or threads.
Screw the cover securely back onto the terminal compartment..
Adjust the cover clamp so it is set over
the cover and tighten.
A0019826
1
L1 L+
N L-
+
-
Page 26
Wiring Gammapilot M FMG60 HART
26 Endress+Hauser
4.7 Wiring in terminal compartment 2
CAUTION
!
Before connection please note the following:
Connect potential matching line to transmitter ground terminal and to the ground
terminal of the water cooling jacket (if present) before connecting up the instrument ("Potential equalization", ä 24).
The cable isolations must comply with the supply voltage and the overvoltage category.The temperature resistance of the connecting cable must comply with the ambient
temperature.
The procedure
Unscrew the cover of the terminal
compartment.
Push the following cables through the
appropriate cable glands or threads:
- signal cable (if the signal output is located in terminal compartment 2)
- PT-100 cable (if present)
- cascading cables (input and/or output, if required)
Wire up according to the terminal assignment diagram.
Tighten the cable glands or threads.
Screw the cover securely back onto the terminal compartment.
A0018927
+
-
PT100
IN CASCADE OUT
SIM WP
OFF ON
3
4
5
6
7
8
9
10
11
12
+ - - +
0
13
2
Page 27
Gammapilot M FMG60 HART Wiring
Endress+Hauser 27
4.8 Connecting the remote display and operation FHX40
For some Dust-Ex versions of the Gammapilot M, the FHX40 connector is protected by a metal sleeve.
Loosen and remove the sleeve with an Allen wrench.
Connect the display and operating unit FHX40
Attach the sleeve and fasten the Allen screw.
A0018090
The remote display and operating unit FHX40 is available as accessory. It is connected to the FHX40-connector of the Gammapilot M via the supplied cable. To do this, the housing of the Gammapilot M needs not to be opened.
A0018089
1 Gammapilot M FMG60 2 Cable of the display and operating unit FHX40
2
1
Page 28
Wiring Gammapilot M FMG60 HART
28 Endress+Hauser
4.9 Wiring in cascading mode
A0018091
A Power supply (90 to 253VAC or 18 to 36VDC ) BJunction box M Master SSlave EEnd-Slave
NOTICE
For the positioning of the power switch according to IEC/EN 61010, there are two options:
At the side of the power supply (one switch for all transmitters)At the side of the transmitters (an individual switch for every transmitter)
WARNING
!
For applications in hazardous areas it is not allowed to interconnect the HART signal circuits for multidrop operation.
L1
L+
N
L-
+
-
L1
L+
N
L-
+
-
L1
L+
N
L-
+
-
L1/L+
N/L-
AC/DC
A
B
M
S
E
+
-
PT100
IN CASCADE OUT
SIM WP
OFF
ON
3
4
9
10
11
12
+ - - +
13
+
-
PT100
IN CASCADE OUT
SIM WP
OFF
ON
3
4
9
10
11
12
+ - - +
13
+
-
PT100
IN CASCADE OUT
SIM WP
OFF
ON
3
4
9
10
11
12
+ - - +
13
+ -
+ - - +
- +
Page 29
Gammapilot M FMG60 HART Wiring
Endress+Hauser 29
4.10 Wiring example for level limit detection 200/400 mm
The output signal is linear between the free and covered calibration (e.g. 4 to 20 mA) and can be evaluated in the control system. If a relay output is needed, the following Endress+Hauser process transmitters can be used:
• RTA421: for non-Ex applications, without WHG, without SIL
• RMA42: for Ex-applications; with WHG and SIL certificates
A0018092
A Wiring with RTA421 switching unit 1 Gammapilot M B Wiring with process control system 2 4 to 20 mA C Wiring with RMA42 switching unit 3 RTA421 D When installing in hazardous areas, please 4 SPS
observe the corresponding safety instructions 5 RMA42
4.10.1 Ex applications in connection with RMA42
Observe the following Safety Instructions:
• XA00303F/00/A3: ATEX II 2 (1) G for Gammapilot M
• XA00304F/00/A3: ATEX II 2 (1) D for Gammapilot M
• XA00095R/09/A3: ATEX II (1) G [Ex ia] IIC, ATEX II (1) D [Ex ia] IIIC for RMA42
4.10.2
SIL applications for Gammapilot M FMG60 in conjunction with RMA42 (for point level detection 200/400 mm PVT scintillator)
• The Gammapilot M meets SIL2/3 as per IEC 61508, see: – Functional safety manual SD00230F/00/EN (Maximum point level detection) – Functional safety manual SD00324F/00/EN (Minimum point level detection)
• The RMA42 meets SIL2 as per IEC 61508:2010 (Edition 2.0), see functional safety manual SD00025R/09/EN.
4
D
5
-
.
X8
X4
-
+
1
1
1
A
B
C
3
2
2
2
.
.
16
11
+
-
+
+
+
-
-
-
Page 30
Wiring Gammapilot M FMG60 HART
30 Endress+Hauser
4.11 Measuring solids flow
In conjunction with a density measuring device, such as Endress+Hauser's "Gammapilot M", Promag 55S also determines the rate of solids with regard to the mass, volume or percentage content. The following order information is required for this purpose for Promag 55S: Order option for software function "Solids flow" (F-CHIP) and order option for a current input.
A0018093
Solids flow measurement (m) with the aid of a density and flow measuring device. If the density of the solids (
S
) and the density of
the transporting liquid (
C
) are also known, the solids flow rate can be calculated.
1 Flow measuring device (Promag 55S)
Volume flow (V). The solids density (
S
) and the density of the transporting liquid
(
C
) also have to be entered in the transmitter
2 Density measuring device (e.g. "Gammapilot M")
Total density
M
(transporting liquid and solids)
4.12 Post-connection check
After wiring the device, carry out the following checks:
• Is the protective earth connected?
• Is the Potential Equalization Line connected?
• Are the terminals correctly assigned?
• Are the cable glands and dummy plugs tight?
• Are the fieldbus connectors and the FHX40 connector fixed securely?
• Are the lids screwed tightly onto the terminal compartments?
• For dust ignition-proof devices: Is the protective sleeve for the FHX40 socket correctly attached?
• Is the cover of the terminal compartment 1 secured by the cover clamp?
WARNING
!
The Gammapilot M may only be operated, if the cover of the terminal compartment 1 is tightly closed.
r
C
r
S
1
2
r
M
m
V
Page 31
Gammapilot M FMG60 HART Operation
Endress+Hauser 31
5Operation
5.1 Overview over the operating options
5.1.1 4 to 20mA with HART protocol
A0018095
1 PLC (programmable logic controller) 5 Gammapilot M 2 Computer with operating tool (e.g. FieldCare) 6 Field Xpert SFX100
3 Commubox FXA291 with ToF Adapter FXA291 7 VIATOR Bluetooth modem with connection cable 4 FHX40 8 Commubox FXA195 (USB)
If the HART communication resistor is not built into the supply unit, it is necessary to insert a communication resistor of 250 into the 2-wire line.
Operation via the service interface
• With the display and operating unit FHX40
• With a personal computer, Commubox FXA291 with ToF Adapter FXA291 (USB) and the operating program "FieldCare". FieldCare is a graphical operating software for devices from Endress+Hauser. It assists with commissioning, securing data, signal analysis and documentation of the measuring point.
Operation via HART
• With Fiel Xpert SFX100
• With the Commubox FXA195 and the operating program "FieldCare"
4
5
8
22
3
7
6
1
Page 32
Operation Gammapilot M FMG60 HART
32 Endress+Hauser
5.2 Display operation
5.2.1 Display and operating elements
The LCD module VU331 for displaying and operating is inside the remote display and operating unit FHX40. The measured value can be read off through the FHX40 sight glass. In order to operate the instrument, the FHX40 must be opened by removing the four screws.
A0018096
1 Gammapilot M 2 FHX40 3Operating module VU331
Display and operating module VU331
A0018097
1 Operating keys 4 Function name 2 Bargraph 5 Parameter Identification number 3Symbols
Display symbols
The following table describes the symbols that appear on the liquid crystal display:
1
2
3
ENDRESS + HAUSER
E
+
1
2
3
4
5
Symbol Meaning
ALARM_SYMBOL
This alarm symbol appears when the instrument is in an alarm state. If the symbol flashes, this indicates a warning.
LOCK_SYMBOL
This lock symbol appears when the instrument is locked, i.e. if no input is possible.
COM_SYMBOL
This communication symbol appears when data transmission via HART, PROFIBUS PA or FOUNDATION Fieldbus, for example, is in progress.
SIMULATION_SWITCH_ENABLE
This communication symbol appears when simulation in FOUNDATION Fieldbus is enabled via the DIP switch.
Page 33
Gammapilot M FMG60 HART Operation
Endress+Hauser 33
Function of the keys
Key(s) Meaning
O
or
V
Navigate upwards in the selection list. Edit numeric value within a function.
S
or
W
Navigate downwards in the selection list. Edit numeric value within a function.
X
or
Z
Navigate to the left within a function group.
F
Navigate to the right within a function group, confirmation.
O
and
F
or
S
and
F
Contrast settings of the LCD.
O
and S and
F
Hardware lock / unlock After a hardware lock, an operation of the instrument via display or communication is not possible! The hardware can only be unlocked via the display. An unlock parameter must be entered to do so.
Page 34
Operation Gammapilot M FMG60 HART
34 Endress+Hauser
5.2.2 The operating menu
Function code
The functions of the Gammapilot M are arranged in an operating menu. To ensure easy orientation within the menu, a unique position code is indicated on the display for each function. This code consists of one alphabetic and two numeric characters.
A0019876-EN
1Measuring mode 2Function group 3Function
• The alphabetic character specifies the current measuring mode of the Gammapilot M: – L: level – S: limit (switch) – D: density – C: concentration – *: no measuring mode selected yet
• The first numeric character identifies the function group: – basic setup *0calibration *1Safety settings *2 – ...
• The second numeric character numbers the individual functions within the function group:
basic setup *0
today's date *01beam type *02isotope *03operating mode *04 – ...
Hereafter, the position is always given in brackets after the function name. "*" (not yet selected) is always indicated as the measurement method, e.g. "present date" (*01).
1 2
3
Page 35
Gammapilot M FMG60 HART Operation
Endress+Hauser 35
Operation using the onsite display VU331
Selection and configuration in Operation menu:
1. Change from Measured Value Display to Group Selection by pressing F.
2. Press S or O to select the required Function Group and confirm by pressing F. The active selection is marked by a
in front of the menu text.
3. Activate Edit mode with O or S.
Selection menus
a. Select the required parameter in the function selected with S or O.
b. F confirms selection;
appears in front of the selected parameter.
c. F confirms the edited value; system quits edit mode.
d. Simultaneous pressing of O and S interrupts selection; system quits edit mode.
Typing in numerals and text
a. Press O or S to edit the first character of the numeral / text.
b. F positions the cursor at the next character; continue with a. until you have
completed your input.
c. If a symbol appears at the cursor, press F to accept the value entered; system quits
edit mode.
d. If a symbol appears at the cursor, press F to return to the previous character (e.g.
for correction of entries).
e. Simultaneous pressing of O and S interrupts selection; system quits edit mode.
4. Press F to select the next function.
5. Press O and S simultaneously once; return to previous function. Press O and S simultaneously twice; return to Group Selection.
6. Press O and S simultaneously to return to Measured value display.
Basic setup Todays date Isotop Return to
group selection
Calibration 3 137 Cs
Safety setting 60Co
Temp. Comp. no comp.
Linearisation
...
>3 s
ESC ESC
S S
2 x
ESC
S
ESC
S
F
ENDRESS + HAUSER
E
+
S
F
F
S
Page 36
Operation Gammapilot M FMG60 HART
36 Endress+Hauser
5.3 Alternative operation options
Apart from local operation, you can also parameterise the measuring instrument and view measured values by means of a HART protocol. There are two options available for operation:
• Operation via the universal handheld operating unit Field Xpert SFX100
• Operation via the Personal Computer (PC) using the operating program FieldCare
NOTICE
The device can also be operated locally using the keys. If operation is prevented by the keys being locked locally, parameter entry via communication is not possible either.
5.3.1 Operation via Field Xpert SFX100
Compact, flexible and robust industry handheld terminal for remote parametrization and measured value inspection via the HART current output or FOUNDATION Fieldbus. For details refer to BA00060S/04/EN.
5.3.2 FieldCare operating program
FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you can configure all Endress+Hauser devices as well as devices from other manufacturers that support the FDT standard. Hardware and software requirements you can find on the internet: www.endress.com
select your country search: FieldCare FieldCare Technical Data.
FieldCare supports the following functions:
• Configuration of transmitters in online operation
• Loading and saving device data (upload/download)
• Documentation of the measuring point
Connection options:
• HART via Commubox FXA195 and the USB port on a computer
• Commubox FXA291 with ToF Adapter FXA291 via service interface
Page 37
Gammapilot M FMG60 HART Operation
Endress+Hauser 37
5.4 Lock/unlock configuration
5.4.1 Software security locking
Enter a number 100 into the "unlock parameter" (*A4) function in the "diagnostics" (*A) function group. The symbol appears on the display. Inputs are no longer possible.
If you try to change a parameter, the device jumps to the "unlock parameter" (*A4) function. Enter "100". Now change the parameters.
5.4.2 Hardware security locking
Press S, O and F simultaneously. Inputs are no longer possible. If you try to change a parameter, the following appears:
Press S, O and F simultaneously. The "unlock parameter" (*A4) function appears. Enter "100". Now change the parameters.
NOTICE
A hardware locking can only be unlocked again via the display by pressing the O, S and F keys at the same time again. It is not possible to unlock the hardware by
communication.
5.5 Reset to the default configuration
It is advisable to reset the customer parameters if you want to use a device with an unknown history. Effects of resetting:
• All customer parameters are reset to their default values.
• Linearization is switched to "linear", but the table values are kept. The table can be
switched back on in the "linearization" (*4) function group in the "linearization" (*40/ *46) function.
In order to carry out the reset, enter "333" in the "reset" (*A3) function in the "diagnostics" (*A) function group.
CAUTION
!
A reset may lead to impairment of the measurement. As a rule, a basic calibration is required after a reset. All the calibration data are deleted after a reset. A complete calibration is needed to put the measurement back into operation.
NOTICE
The default values of each parameter are shown in bold in the menu overview in the appendix.
On-site display
unlock parameter 0A4
Hardware locked
Page 38
Commissioning Gammapilot M FMG60 HART
38 Endress+Hauser
6 Commissioning
NOTICE
This chapter describes the commissioning of the Gammapilot M using the operating module VU331 (which is located in the remote display and operating unit FHX40). Commissioning using the "FieldCare" or the "Field Xpert SFX100" is similar. For further instructions to the operating program "FieldCare" see BA00027S/04/EN or the BA00060S/04/EN supplied together with the Field Xpert SFX100.
6.1 Calibration: overview
A detailed description of the functions used can be found in the following sections: ä 40, "Basic setup" ä 46, "Calibration for level measurement and limit detection" ä 61, "Calibration for density and concentration measurements" ä 69, "Density measurement/temperature-compensated"
Power up instrument
see ä 39, "Switching on the device"
Basic setup
see ä 40, "Basic setup"
Calibration for level measure­ment and level limit detection
Calibration for density and con­centration measurements
Calibration for density measure­ments/ temperature compensa­ted
see ä 46, "Calibration for level measurement and limit detection"
see ä 61, "Calibration for den­sity and concentration measure­ments"
see ä 69, "Density measure­ment/temperature-compensated"
Additional settings
see "Description of Instrument Functions" BA00287F/00/EN
Page 39
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 39
6.2 Switching on the device
NOTICE
Error messages A165 "electronics defect" and A635 "present date not defined"
The Gammapilot M contains 2 real-time clocks for the decay compensation, which are permanently compared to each other for safety reasons. In order to bridge voltage interruptions, the clocks are buffered with a capacitor. To ensure that the clocks work correctly and retain the date in the event of a voltage interruption, this capacitor must have a minimum charge. If the A165 "Electronics defect" or A635 "Present date not defined" error message appears after switching on the Gammapilot M, then the capacitor may possibly not yet be charged sufficiently. In this case, the Gammapilot M must be operated at the operating voltage for at least 20 to 30 minutes, in order to charge the capacitor. After this, the date must be entered correctly. If the error message still persists subsequent to this, it can be deleted by switching the Gammapilot M off and on.
After switching on the supply voltage, the instrument is first initialized. Due to internal memory tests, this takes approx. 2 minutes.
On-site display Meaning
FMG60
Then, the following appear for approximately five seconds:
•Device type
•Software version
• Type of the communication signal
V01.03.06 HART
On first power up you are requested to select the language for the display texts.
Language 092
Select the language with the O and S keys. Confirm your choice by pressing F twice.
Englisch
Français
Español
After that the measured value display appears. Now you can perform the basic setup and the calibration. Press F to switch to the group selection.
Group selection
Press F again to enter the first function of the "basic setup" function group.
Basic setup
Calibration
Safety settings
Page 40
Commissioning Gammapilot M FMG60 HART
40 Endress+Hauser
6.3 Basic setup
6.3.1 "Present date" (*01)
6.3.2 "Beam type" (*02)
6.3.3 "Isotope" (*03)
On-site display Meaning
Present date *01 Date and time of the basic setup are specified in this function.
Each of these values must be confirmed by "E" after it has been entered.
17.11.04 10:30
dd.mm.yy hh:mm
On-site display Meaning
Beam type *02
This function is used to specify whether the radiation source used emits radiation continuously or whether it is modulated (for grammagraphy suppression).
• Standard/continuous (permanent, continuous radiation)
• Modulated (modulated radiation source)
Standard/cont.
Modulated
On-site display Meaning
Isotope *03
This function is used to specify which isotope is used for the measurement. The Gammapilot M needs this information for the decay compensation.
137 Cs
60 Co
no compens.
Page 41
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 41
6.3.4 "Operating mode" (*04)
NOTICE
The selection can be performed only once and the function is automatically locked after that. It can only be unlocked again by a reset of the Gammapilot M ("Reset" (*A3) function).
A0018107
A One Gammapilot M is sufficient for measuring ranges up to 2 m (6.6 ft ); For larger measuring ranges as many
Gammapilot M as required can be connected (cascading mode). By Software settings they are defined as B Master CSlave(s) or DEnd-Slave 1 4 to 20 mA HART; PROFIBUS PA; FOUNDATION Fieldbus
Options/display:
Stand alone: This option is selected if the Gammapilot M is used as a single instrument.
Master: This option is selected if the Gammapilot is located at the beginning of a
cascading chain. It receives pulses from a connected slave, adds its own pulses and calculates the measuring value from this total.
Slave: This option is selected if the Gammapilot M is located in the middle of a cascading
chain. It receives the pulses from an additional connected slave or end-slave, adds its own pulses and transmits this total to the next device (master or slave). After selecting this option, the basic setup is finished. When cascading several transmitters the further calibration is performed on the master only.
End slave: This option is selected if the Gammapilot M is located at the end of a cascading
chain. It does not receive pulses from another device but transmits its own pulses to the next device (master or slave). After selecting this option, the basic setup is finished. When cascading several transmitters the further calibration is performed on the master only.
Not defined: Is displayed if no operating mode has been selected yet. In order to continue
the basic setup, a selection is necessary.
NOTICE
If a "Slave" or an "End-slave" are connected to the "FieldCare", the pulse rate of this device is displayed in the header instead of the measured value.
On-site display Meaning
Operating mode *04
This function is used to specify in which operating mode the Gammapilot M will be used.
Stand alone
Master
Slave
A
B
1
C
D
Page 42
Commissioning Gammapilot M FMG60 HART
42 Endress+Hauser
6.3.5 "Measuring mode" (*05)
Further options:
• Level measurement (continuous)
• Level limit detection
• Density measurement (with temperature compensation if required)
• Concentration measurement (density measurement followed by linearization)
NOTICE
The selection can be performed only once and the function is automatically locked after that. It can only be unlocked again by a reset of the Gammapilot M ("Reset" (*A3) function).
A0018108
A Level measurement (continuous)
Density B Level limit detection E Concentration C Density measurement (with temperature compensation if required) D Concentration measurement (density measurement followed by linearization)
On-site display Meaning
Meas. mode *05
This function is used to select the desired measuring mode.
Level
Limit
Density
A
B
C
D
r
E
Page 43
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 43
6.3.6 "Density unit" (*06)
Further options:
•g/cm
3
•g/l
• lb/gal; [1g/cm
3
= 8,345 lb/gal]
•lb/ft
3
; [1g/cm3 = 62,428 lb/ft3]
• 1°Brix = [270 (1 - 1/x)]
• °Baumé; [1°Baumé = 144.3 (1 - 1/x)]
• °API; [1°API = 131.5 (1.076/x - 1)]
• °Twaddell; [1°Twaddell = 200 (x-1)]
"x" refers to the density in g/cm
3
. The formula indicates how many degrees this density
corresponds to.
6.3.7 "Min. density" (*07)
6.3.8 "Max. density" (*08)
6.3.9 "Pipe diameter unit" (*09)
On-site display Meaning
Density unit *06
This function is needed for density and concentration measure­ments only. It is used to select the density unit.
g/cm3
g/l
lb/gal
On-site display Meaning
Min. density *07
This function is needed for density and concentration measure­ments only. It is used to specify the lower limit of the density range. The output current for this density is 4 mA.
0,9500 g/cm³
On-site display Meaning
Max. density *08
This function is needed for density and concentration measure­ments only. It is used to specify the upper limit of the density range. The output current for this density is 20 mA.
1,2500 g/cm³
On-site display Meaning
Pipe diam. unit *09
This function is needed for density and concentration measure­ments only. It is used to select the unit for the pipe diameter.
1 inch = 25,4 mm
mm
inch
Page 44
Commissioning Gammapilot M FMG60 HART
44 Endress+Hauser
6.3.10 "Pipe diameter" (*0A)
A0018109
Always specify the complete irradiated measuring path L in the "pipe diameter" (*0A) function. Depending on the installation, this value may be larger than the actual pipe diameter.
On-site display Meaning
Pipe diam. *0A
This function is needed for density and concentration measure­ments only. It is used to specify the irradiated measuring path L. With standard installation, this value is identical to the inner pipe diameter D
I
. For other installations (in order to enlarge the irradiated measuring path) it may be larger (see figure). The pipe walls are not to be considered a part of the measuring path.
200 mm
1
1
1
L
L
L
a
D
I
a
L
30°
2.0 x D
I
Page 45
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 45
6.3.11 "Output damping" (*0B)
A0018110
1 Level change (or density change) 2Measured value
Range of values
1 to 999 s
Default
The default depends on the selected "measuring mode" (*05):
•Level: 6 s
• Limit: 6 s
• Density: 60 s
• Concentration: 60 s
Selecting the output damping
The best value of the output damping depends on the process conditions. By enlarging the output damping, the measured value becomes considerably steadier but also slower. In order to dampen the influence of strongly fluctuating surfaces or stirrers, it is advisable to enlarge the output damping. On the other hand, if rapid changes of the measured value have to be detected accurately, the output damping may not be selected to large.
On-site display Meaning
Output damping *0B
This function is used to specify the output damping (in seconds) by which changes of the measured value are attenua­ted. After a surge in the level or density it takes 5 x until the new measured value is reached.
60 s
5 x t
t
L
1
2
Page 46
Commissioning Gammapilot M FMG60 HART
46 Endress+Hauser
6.4 Calibration for level measurement and limit detection
6.4.1 Basic principles
The calibration points for the measurement are entered in the "calibration" (*1) function group. Each calibration point consists of a level and the associated pulse rate.
Calibration points for level measurement
A0018111
A Background calibration B Full calibration C Empty calibration
Background calibration
Refers to the following situation:
• The radiation is switched off.
• Within the measuring range, the vessel is filled as far as possible (ideally: 100%).
The background calibration is necessary, in order to register the natural background radiation at the mounting position of the Gammapilot M. The pulse rate of this background radiation is automatically subtracted from any other measured pulse rate. That means: only the part of the pulse rate which originates from the applied radiation source is taken into account and is displayed. As opposed to the radiation of the applied source, the background radiation remains nearly constant during the complete measurement. Therefore, it is not submitted to the automatic decay compensation of the Gammapilot M.
Full calibration
Refers to the following situation:
• The radiation is switched on.
• Within the measuring range, the vessel is filled as far as possible (ideally: 100%, minimum 60%).
If the vessel cannot be filled to at least 60% during the calibration, the full calibration can alternatively performed with the radiation being switched off, which is a way of simulating a filling of 100%. In this case, the full calibration is identical to the background calibration. As the pulse rate of the background radiation is automatically subtracted, the displayed pulse rate is about 0 cps.
NOTICE
This type of simulated calibration is not possible with self-radiating media. In this case it is always necessary to perform the background and full calibration with the vessel filled to 100%.
C
B
A
OFF
ON ON
Page 47
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 47
Empty calibration
Refers to the following situation:
• The radiation is switched on.
• Within the measuring range, the vessel is emptied as far as possible (ideally: 0%, maximum 40%).
Calibration points for limit detection
A0018112
A Background calibration B Covered calibration C Free calibration
Background calibration
Refers to the following situation:
• The radiation is switched off.
• If possible, the radiation path is completely covered.
The background calibration is necessary, in order to register the natural background radiation at the mounting position of the Gammapilot M. The pulse rate of this background radiation is automatically subtracted from any other measured pulse rate. That means: only the part of the pulse rate which originates from the applied radiation source is taken into account and is displayed. As opposed to the radiation of the applied source, the background radiation remains nearly constant during the complete measurement. Therefore, it is not submitted to the automatic decay compensation of the Gammapilot M.
Covered calibration
Refers to the following situation:
• The radiation is switched on.
• If possible, the radiation path is completely covered.
If the radiation path cannot be completely covered during the calibration, the covered calibration can alternatively be performed with the radiation being switched off, which is a way of simulating complete covering. In this case, the covered calibration is identical to the background calibration. As the pulse rate of the background radiation is automatically subtracted, the displayed pulse rate is about 0 c/s.
NOTICE
This type of simulated calibration is not possible with self-radiating media. In this case it is always necessary to perform the background calibration and the covered calibration with the radiation path completely covered.
Free calibration
Refers to the following situation:
• The radiation is switched on.
• The radiation path is completely free.
AB
C
OFF ON ON
Page 48
Commissioning Gammapilot M FMG60 HART
48 Endress+Hauser
Methods for entering the calibration points
Automatic calibration
For an automatic calibration, the vessel is filled to the required value. For the background calibration the radiation remains switched off, for the other calibration points the radiation is switched on. The Gammapilot M automatically records the pulse rate. The associated level is entered by the user.
Manual calibration
If during the commissioning of the Gammapilot M one or more calibration points cannot be realized (e.g. because the vessel cannot be sufficiently filled or emptied), the calibration point must be entered manually. That is, not only the level but also the associated pulse rate must be entered by the user. For details concerning the calculation of the count rate please refer to your Endress+Hauser sales organization.
NOTICE
Calibration date and calibration
When calibrating manually, the calibration date is not set automatically. Instead, it must
be entered manually into the "calibration date" (*C7) function.
A manually entered calibration point should be replaced by an automatic calibration as
soon as the associated level occurs during the operation of the plant. This recalibration is advisable because calibration points entered automatically result in more precise measurement results than calculated ones.
Page 49
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 49
6.4.2 Background calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the background calibration is entered. The individual functions are explained in the sections below.
"Background calibration" (*10)
Options:
stop/edit This option must be selected if – No background calibration is to be performed but the pulse rate of an existing
background calibration is to be displayed instead. – A manual background calibration is to be performed. After selecting this option, the Gammapilot M changes to the "bgr. pulse rate" (*12) function, where the existing pulse rate is displayed and can be changed if required.
start This option is used to start an automatic background calibration. The Gammapilot M changes to the "avg. pulse rate" (*11) function.
*10 Backgr. calibr. 0: stop/edit 1: start
*11 avg. pulse rate
*10 bgr. pulse rate
-1: not calibrated
On-site display Meaning
Backgr. cal. *10
This function is used to start the background calibration.
stop/edit
start
E
E
0
Page 50
Commissioning Gammapilot M FMG60 HART
50 Endress+Hauser
"Avg. pulse rate" (*11)
A0018118
If the value is sufficiently stable, the function can be left by pressing "E". Thereafter, the Gammapilot M changes to the "backgr. calib." (*10) function. Select "stop/
edit" to stop the averaging procedure. The value is then automatically transmitted to the "bgr. pulse rate" (*12) function.
NOTICE
Bgr. Pulse rate
The maximum integration time is 1000 s. After this time, the value is automatically
transmitted to the "bgr. pulse rate" (*1B) function.
The integration is not terminated by pressing "E" in the "avg. pulse rate" (*11) function.
It is continued until the selection of "stop/edit" in the "backgr. calib." (*10) function. This may result in a slight deviation between the last displayed average pulse rate and the final "bgr. pulse rate" (*12).
"Background pulse rate" (*12)
On-site display Meaning
Avg. pulse rate *11
The average pulse rate is displayed in this function (after selec­tion of "start" in the previous function). Initially, this value fluc­tuates (because of the decay statistics), but due to the integra­tion it reaches an average value in the course of time. The longer the averaging is performed the lower are the remaining fluctuations.
186 cps
c/s
t
"E"
On-site display Meaning
Backgr. pulse rate *12
The pulse rate of the background calibration is displayed in this function. By pressing "E" the displayed value can be confirmed and the background calibration completed. "-1" indicates, that no background calibration is present yet. In this case there are two options:
• Either return to the "background calibration" (*10) func­tion and restart the background calibration
• Or enter a known or calculated pulse rate (manual calibra­tion). Thereafter, the Gammapilot M changes to the "calibr. point" (*13) or (*1A) function.
186 cps
Page 51
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 51
6.4.3 Full and empty calibration or covered and free calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the full and empty calibration (for level measurements) or the covered and free calibration (for level limit detection) are entered. The individual functions are explained in the sections below. The functions are only accessible after the background calibration has been performed.
NOTICE
The "value full" (*14) and "value empty" (*17) functions only appear if the "level" option was selected in the "measurement method" function (*05).
"Calibration point" (*13)
*14 value full [%] target value at calibr. point
*15 calibration 0: stop/ edit 1: start
*11 avg. pulse rate
*16 full calibr. [c/s]
-1: not calibra­ted
Group selection
013 calibr. point 0: full/ covered 1: empty/free
*19 next point 0: no 1: yes
*17 value empty [%] target value at calibr. point
*15 calibration 0: stop/ edit 1: start
*11 avg. pulse rate
*18 empty calibr. [c/s]
-1: not calibra­ted
On-site display Meaning
Calibr. point *13
This function is used to select which calibration point ("full/ covered" or "empty/free") will be entered.
full/covered
empty/free
1
E
0
1
E
0
Page 52
Commissioning Gammapilot M FMG60 HART
52 Endress+Hauser
"Value full" (*14) / "Value empty" (*17)
Range of values
"Calibration" (*15)
Options:
stop/edit This option must be selected if – the calibration point is not to be entered (e.g. because it has already been entered). The
pulse rate of the calibration point is then displayed in the following function, "full calibr." (*16) or "empty calibr." (*18). If required, this value can be changed.
– the calibration point is to be entered manually. This can be done in the following
function, "full calibr." (*16) or "empty calibr." (*18).
start This option is used to start the automatic entering of the calibration point. The Gammapilot M then changes to the "avg. pulse rate" (*11) function.
On-site display Meaning
value full *14
These functions are needed for level measurements only. They are used to specify the level at which the full or empty calibration are performed.
100%
value empty *17
0%
optimum value minimum value maximum value
Value full (*14) 100% 60% 100%
Value empty (*17) 0% 0% 40%
On-site display Meaning
calibration *15
This function is used to start the automatic entering of the selected calibration point.
stop/edit
start
Page 53
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 53
"Avg. pulse rate" (*11)
A0018118
Initially, the pulse rate strongly fluctuates. In the course of time an average value is reached.
If the value is sufficiently stable, the function can be left by pressing "E". Thereafter, the Gammapilot M changes to the "calibration" (*15) function. Select "stop/
edit" to stop the averaging procedure. The value is then automatically transmitted to the "full calibr." (*16) or "empty calibr." (*18) function respectively.
NOTICE
Avg. Pulse rate
The maximum integration time is 1000 s. The value is then transmitted automatically to
the "full calibr." (*16) or "empty calibr." (*18) function.
The integration is not terminated by pressing "E" in the "avg. pulse rate" (*11) function.
It is continued until the selection of "stop/edit" in the "calibration" (*15) function. This may result in a slight deviation between the last displayed average pulse rate and the final "full calibr." (*16) or "empty calibr. (*18).
"Full calibration" (*16) / "Empty calibration" (*18)
On-site display Meaning
avg. pulse rate *11
The average pulse rate is displayed in this function (after selection of "start" in the previous function). Initially, this value fluctuates (because of the decay statistics), but due to the integration it reaches an average value in the course of time. The longer the averaging is performed the lower are the remaining fluctuations.
2548 cps
c/s
t
"E"
On-site display Meaning
full calibr. *16 The pulse rate of the respective calibration points is displayed
in these functions. The displayed value must be confirmed by pressing "E". "-1" indicates, that no background calibration is present yet. In this case there are two options:
• either return to the "calibration" (*15) function and restart the calibration
• or enter a knwon or calculated pulse rate (manual calibration)
33 cps
empty calibr. *18
2548 cps
Page 54
Commissioning Gammapilot M FMG60 HART
54 Endress+Hauser
"Next point" (*19)
Options:
no This option must be selected after both calibration points have been entered. After this selection the Gammapilot M returns to the group selection and the calibration is completed.
yes This option must be selected, if only one calibration point has been entered yet. After this selection the Gammapilot M returns to the "calibr. point" (*13) function and the next point can be entered.
6.4.4 Additional settings
After the calibration has been finished, the Gammapilot M outputs the measuring value via the current output and via the HART signal. The complete measuring range (0 to 100%) is mapped to the range (4 to 20 mA) of the output current. Many additional functions are available for optimizing the measuring point. They can be configured as required. A detailed description of all instrument functions is given in Operating Instructions BA00287F/00/EN, "Gammapilot M - Description of Device Functions". This document can be found on the supplied CD-ROM.
6.4.5 Configuration of the contactor for level limit detection
For safety-related applications, the threshold values in the Safety Manual
SD00230F/00/EN and SD00324F/00/EN must be observed.
The calculation of the switching signal from the continuous signal is not performed within the Gammapilot M but in the connected evaluation unit or process transmitter. For further information refer to the Operating Instructions of the respective instrument. If using the Endress+Hauser process transmitter RTA421 or RMA42 the following settings are recommended:
For maximum fail-safe mode
• Switch threshold (SETP) = 75%
• Hysteresis (HYST) = 50%
On-site display Meaning
next point *19
This function is used to specify, if a further calibration point is to be entered or not.
no
yes
Page 55
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 55
6.5 Function group "Safety settings" (*2)
6.5.1 "Output on alarm" (*20)
6.5.2 "Output on alarm" (*21)
On-site display
group selection *2Þ
safety settings
temp. compensation
linearization
On-site display Meaning
outp. on alarm *20
This function determines what value the output of the Gamma­pilot M assumes in the event of an alarm condition.
MIN -10% 3.6 mA
MAX 110% 22 mA
Hold
(*20)
Output on alarm
4 to 20 mA with HART
PROFIBUS PA
FOUNDATION Fieldbus
MIN 3.6 mA -99999
MAX 22 mA +99999
Hold The last measured value is held
User-specific (can only be selected for HART devices)
As defined in "Output on alarm"
(*21)
Not possible
On-site display Meaning
outp. on alarm *21
This function is used to specify what user-specific value the current output should assume in the event of an alarm condi­tion. The value is entered in mA. This function is only available for HART devices. It is only active if the "user-specific" option has been selected in the "output on alarm" (*20) function. Range of values: 3.6 to 22 mA
22.00 mA
Page 56
Commissioning Gammapilot M FMG60 HART
56 Endress+Hauser
6.6 SIL locking (for level limit detection 200/400 mm PVT
scintillator)
SIL locking ("Security locking" (022) function) is in the "Safety settings function group" (S2). It can only be accessed in the "stand alone" operating mode in conjunction with the "level limit" measurement method (see also "Requirements for locking"). As soon as SIL locking or unlocking is started, communication via the display or via FieldCare is significantly slower. This is due to internal readback and parameter validation. However, this only applies during the locking or unlocking phase and does not affect the measurement itself. In the event of locking, all the parameters are locked apart from the manufacturer reset code. The parameters can only be viewed. Only the manufacturer reset code can be modified. Locking starts by entering a four-digit password (1000 to 9999). Then there follows a sequence of prompts for the most important parameter s which all have to be confi rmed. T he locking is concluded by confirming the password. The device is locked as soon as the password is confirmed. The password is no longer visible. If a parameter or the password is displayed incorrectly and thus the password or a parameter is not confirmed, the locking procedure is aborted. The FMG60 is then in an unlocked state, as it was before the locking procedure began.
Prerequisites for locking
The following parameters must be configured for locking to be possible:
1. Operating mode = stand alone
2. Measurement mode = limit
3. Communication = HART Ex i or HART Ex e/d
4. Type of scintillator = PVT
5. Detector length = 200 or 400
6. SW version = 01.02.00 or 01.02.02
7. Radiation source = Cs or Co
Check whether the calibration values of the readback path are in the valid range.
6.6.1 List of the parameters to be confirmed
The following parameters can be modified by the user and thus they have to be confirmed. The detector length must be confirmed since it cannot be defined with regard to the safety function in the final check and is only in the service segment following any repair work.
1. Date
2. Beam type (standard or modulated
6)
)
3. Source type (Cs or Co)
4. Output damping
5. Calibration date
6. Background pulse rate cps
7. Free calibration cps
8. Covered calibration cps
9. Gammagraphy hold time (can only be configured for standard beam type) or 10 for
modulated source
10. Output current 3.0 mA
11. Detector length
6) Modulated just for max. level limit detection.
Page 57
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 57
6.6.2 Function "security locking" (*22) (SIL unlocking)
The FMG60 can be unlocked in SIL2/3 mode by entering the password. If the password is entered correctly, the FMG60 is unlocked. If an incorrect password is entered, the FGM60 goes back to the group selection. The device cannot be unlocked by switching the power ON and then OFF.
Further options:
• unlocked
• locked
• device locked
• device unlocked
6.6.3 Forgotten your password?
The password cannot be viewed when the device is locked. For this reason, the password can only be deleted by a manufacturer reset. At the same time, all the parameters are set to default values and the calibration data are deleted. The device then has an error current.
6.6.4 Function "password" (*23) (security password)
The password is always a four-digit number in the range from 1000 to 9999. Other values are not valid. After locking, 0000 is displayed. The password itself is not.
On-site display
security locking *22
unlocked
locked
device locked
On-site display
password *23
Page 58
Commissioning Gammapilot M FMG60 HART
58 Endress+Hauser
6.6.5 Function "confirm Iout" (*24) (output current during locking)
The output current is shifted to <3.6, typically 2.4 mA, by means of the 2nd switch-off path by selecting "Device locked" so the user can clearly see that the FMG60 has actually been locked. The user must explicitly confirm this current value. The FMG60 is only set to the "device locked" state and the current output is only released again once the system has successfully run through the entire locking sequence. If the FMG60 is switched off and then on again during the locking process, the FMG60 resumes normal, unlocked operation. If a parameter is not confirmed, the FMG60 remains in the "device locked" state. The device can be switched to "unlocked" during the locking process; it then operates in the normal measuring mode. The "device locked" state can also be disabled by means of a total reset (7864) which also deletes all the calibration parameters, however. The correct locking state can be determined using the "partial stroke test".
6.6.6 Function "confirm sequence" (*25) (display check)
To verify that numbers are shown properly on the display, the sequence of numbers > 0123456789 .- < is the first element to be confirmed on the display. The user must confirm the numbers are displayed correctly. If an error occurs in how the numbers are displayed, the user must abort the locking action.
6.6.7 Function "confirm backgr." (*26)
On-site display
confirm Iout *24
not valid
valid
On-site display
confirm sequence *24
>0123456789 .-<
not valid
valid
On-site display Meaning
confirm backgr. *26
Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the data you entered. Select "not valid" if you want to abort SIL locking.
not valid
valid
Page 59
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 59
6.6.8 Function "confirm cal." (*27)
Further options:
• Not valid
•Valid
6.6.9 Function "confirm source" (*28)
Further options:
• Not valid
•Valid
6.6.10 Function "confirm time" (*29) (output damping)
Further options:
• Not valid
•Valid
6.6.11 Function "confirm date" (*2A)
Further options:
• Not valid
•Valid
On-site display Meaning
confirm cal. *27
Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the data you entered. Select "not valid" if you want to abort SIL locking.
not valid
On-site display Meaning
confirm source *28
Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the data you entered. Select "not valid" if you want to abort SIL locking.
not valid
On-site display Meaning
confirm time *29 Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the
data you entered. Select "not valid" if you want to abort SIL locking.
not valid
On-site display Meaning
confirm date *2A Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the
data you entered. Select "not valid" if you want to abort SIL locking.
not valid
Page 60
Commissioning Gammapilot M FMG60 HART
60 Endress+Hauser
6.6.12 Function "confirm length" (*2B)
6.6.13 Function "confirm password" (*2C)
6.6.14 Function "password" (*2D) (unlock password)
The password is always a four-digit number in the range from 1000 to 9999. Other values are not valid. To unlock the device, enter the the four-digit number password.
On-site display Meaning
confirm length *2B
Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the data you entered. Select "not valid" if you want to abort SIL locking.
not valid
valid
On-site display Meaning
confirm password *2C
Select "valid" if the data displayed (see _ _ _ _ _ _ ) match the data you entered. Select "not valid" if you want to abort SIL locking.
not valid
valid
On-site display
password *2D
Page 61
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 61
6.7 Calibration for density and concentration measurements
6.7.1 Basic principles
The calibration points for the measurement are entered in the "calibration" (*1) function group. Each calibration point consists of a density value and the associated pulse rate.
Calibration points for density and concentration measurements
Function of the calibration points
For density and concentration measurements the Gammapilot M needs (apart from the length of the irradiated measuring path) the following two parameters:
• The absorption coefficient of the material measured
• The reference pulse rate I
0
7)
.
It calculates these parameters automatically from the following calibration points:
• Background calibration (calibration with the radiation switched off)
• Up to nine calibration points for samples of various known densities.
NOTICE
With self-radiating media it is always necessary to perform the background calibration with a filled pipe. A simulated calibration with an empty pipe is not possible in this case.
A0018128
0 Background calibration 1-9 Calibration points for various densities
Two-point calibration
The recommended comparisen procedure for high exactness standards about the whole measuring range is the two-point calibration. First the background calibration occurs. The two calibration points will be adapt. They should lie differ considerably. After input of both calibration points, the Gammapilot M calculates the parameters I
0
und .
One-point calibration
If a two-point calibration is not possible, a one-point calibration can be carried out. That means, that apart from the background calibration only one further calibration point is used. This calibration point should be located as near as possible to the operating point. Densities in the proximity of this operating point are measured fairly precisely, whereas the precision may decrease with increasing distance to the operating point. In one-point calibration, the Gammapilot M only calculates the reference pulse rate I
0
.
For the absorption coefficient it uses the standard value = 7.7 mm
2
/g in this instance.
7) I0 is the pulse rate for the tube being empty. The value is significantly higher than any real pulse rate occurring during the measurement.
OFF ON
0
1
9
Page 62
Commissioning Gammapilot M FMG60 HART
62 Endress+Hauser
Multiple-point calibration
The multiple-point calibration is recommended particularly for measurements in a large density area or for especially exact measurements. Up to 9 calibration points can be used about the whole measuring range. The calibration points should be located as far from each other as possible and should be uniformly distributed over the measuring range. After the calibration points have been entered, the Gammapilot M automatically calculates the parameters I
0
and . Multiple-point calibration is especially advisable for measurement in a
wide range of densities or for especially precise measurements.
Recalibration
The Gammapilot M provides a further calibration point ("10") for recalibration. This point can be entered, if the measuring conditions have changed, e.g. by deposit in the measuring tube. After entering of the recalibration point, I
0
is recalculated according to the current measuring conditions. The absorption coefficient is kept unchanged from the original calibration.
Methods for entering the calibration points
Automatic calibration
For an automatic calibration, the desired calibration point is realized at the measuring tube, i.e. the measuring tube is filled with a medium of the desired density. For the background calibration the radiation remains switched off, for the other calibration points the radiation is switched on. The Gammapilot M automatically records the pulse rate. The associated density is determined in the laboratory and entered by the user.
Manual calibration
In order to achieve a high measuring accuracy, it is advisable to determine the pulse rates for a couple of samples of the same density and to calculate the average density and average pulse rate for these samples. These values can then be entered manually into the Gammapilot M. If possible, this procedure should be repeated at a further density. The both density values should be as far from each other as possible.
NOTICE
When calibrating manually, the calibration date is not set automatically. Instead, it must be entered manually into the "calibration date" (*C7) function.
Page 63
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 63
6.7.2 Background calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the background calibration is entered. The individual functions are explained in the sections below.
"Background calibration" (*10)
Options:
stop/edit This option must be selected if – no background calibration is to be performed but the pulse rate of an existing
background calibration is to be displayed instead. – a manual background calibration is to be performed. After selecting this option, the Gammapilot M changes to the "bgr. pulse rate" (*12) function, where the existing pulse rate is displayed and can be changed if required.
start This option is used to start an automatic background calibration. The Gammapilot M changes to the "avg. pulse rate" (*11) function.
*10 backgr. calib. 0: stop/edit 1: start
*11 avg. pulse rate
*12 bgr. pulse rate
-1: not calibrated
On-site display Meaning
backgr. cal. *10
This function is used to start the background calibration.
stop/edit
start
E
1
E
0
Page 64
Commissioning Gammapilot M FMG60 HART
64 Endress+Hauser
"Avg. pulse rate" (*11)
A0018118
If the value is sufficiently stable, the function can be left by pressing "E". Thereafter, the Gammapilot M changes to the "backgr. calib." (*10) function. Select "stop/
edit" to stop the averaging procedure. The value is then automatically transmitted to the "bgr. pulse rate" (*12) function.
NOTICE
Bgr. Pulse rate
The maximum integration time is 1000 s. After this time, the value is automatically
transmitted to the "bgr. pulse rate" (*1B) function.
The integration is not terminated by pressing "E" in the "avg. pulse rate" (*11) function.
It is continued until the selection of "stop/edit" in the "backgr. calib." (*10) function. This may result in a slight deviation between the last displayed average pulse rate and the final "bgr. pulse rate" (*12).
"Background pulse rate" (*12)
On-site display Meaning
avg. pulse rate *11
The average pulse rate is displayed in this function (after selec­tion of "start" in the previous function). Initially, this value fluctuates (because of the decay statistics), but due to the integration it reaches an average value in the course of time. The longer the averaging is performed the lower are the remaining fluctuations.
186 cps
c/s
t
"E"
On-site display Meaning
backgr. pul. rate *12
The pulse rate of the background calibration is displayed in this function. By pressing "E" the displayed value can be confirmed and the background calibration completed. "-1" indicates, that no background calibration is present yet. In this case there are two options:
• either return to the "background calibration" (*10) function and restart the background calibration
• or enter a known or calculated pulse rate (manual calibration). Thereafter, the Gammapilot M changes to the "calibr. point" (*13) or (*1A) function.
186 cps
Page 65
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 65
6.7.3 Calibration points
Excerpt from the operating menu
The following excerpt from the operating menu shows, how the density calibration points are entered. The individual functions are explained in the sections below. The functions are only accessible after the background calibration has been performed.
"Calibration point" (*1A)
Further options:
• "1" to "9" : Calibration points for various densities
• "10": recalibration point After entering of the recalibration point, I
0
is recalculated according to the current measuring conditions. The absorption coefficient is kept unchanged from the original calibration. The calibration point "10" can be entered if the measuring conditions have changed, e.g. due to buildup in the measuring tube.
*1A calibr. point
1...9 point 10 for recalibration
*15 calibration 0: stop/edit 1: start
*11 avg. pulse rate
*1B density calib. (c/s)
*1C density value
Group selection
*1D calibr. point 0: not used 1: used 2: clear
*1E absorp. coeff.
*1F ref. pulse rate
*19 next point 0: no 1: yes
On-site display Meaning
calibr. point *1A
This function is used to select, which calibration point will be entered.
1
2
3
E
0
1
Page 66
Commissioning Gammapilot M FMG60 HART
66 Endress+Hauser
"Calibration" (*15)
Options:
stop/edit This option must be selected if – the calibration point is not to be entered (e.g. because it has already been entered). The
pulse rate of the calibration point is then displayed in the following function "density calib." (*1B). If required, this value can be changed.
– the calibration is to be entered manually. For this purpose, the Gammapilot M changes
to the "density calib." (*1B) function.
start This option is used to start the automatic entering of the calibration point. The Gammapilot M then changes to the "avg. pulse rate" (*11) function.
"Avg. pulse rate" (*11)
A0018118
If the value is sufficiently stable, the function can be left by pressing "E". Thereafter, the Gammapilot M changes to the "calibration" (*15) function. Select "stop/
edit" to stop the averaging procedure. The value is then automatically transmitted to the "density calibr." (*1B) function.
NOTICE
Density calibration
The maximum integration time is 1000 s. After this time, the value is automatically
transmitted to the "density calibration" (*1B) function.
During the integration a sample of the measured material must be taken. Its density must
be determined in the laboratory).
The integration is not terminated by pressing "E" in the "avg. pulse rate" (*11) function.
It is continued until the selection of "stop/edit" in the "calibration" (*15) function. This may result in a slight deviation between the last displayed average pulse rate and the final "density calibration" (*1B).
On-site display Meaning
calibration *15
This function is used to start the automatic entering of the selected calibration point.
stop/edit
start
On-site display Meaning
avg. pulse rate *11 The average pulse rate is displayed in this function (after selec-
tion of "start" in the previous function). Initially, this value fluc­tuates (because of the decay statistics), but in the course of time it reaches an average value. The longer the averaging is performed the lower are the remaining flucuations.
1983 cps
c/s
t
"E"
Page 67
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 67
"Density calibration" (*1B)
"Density value" (*1C)
NOTICE
When entering the value, temperature influences have to be taken into account. The density entered must refer to the temperature at which the pulse rate has been determined. If the density and the pulse rate have been determined at different temperatures, the density value must be corrected accordingly.
"Calibration point" (*1D)
Options:
not used The calibration point is not used. However, it can be reactivated at a later point of time.
used The calibration point is used.
clear The calibration point is deleted. It cannot be reactivated at a later point of time.
On-site display Meaning
density calibr. *1B
The pulse rate of the respective calibration point is displayed in this function. The display value must be confirmed by pressing "E". "-1" indicates, that no pulse rate is present yet. In this case there are two options:
• either return to the "calibration" (*15) function and restart the calibration
• or enter a known or calculated pulse rate (manual calibra­tion)
1983 cps
On-site display Meaning
density value *1C
This function is used to enter the density of the calibration point. The value must be determined from the sample in a laboratory measurement.
0.9963 g/cm3
On-site display Meaning
calibr. point *1D
This function is used to specify, if the current calibration piont is to be used.
not used
used
clear
Page 68
Commissioning Gammapilot M FMG60 HART
68 Endress+Hauser
"Absorption coefficient" (*1E)
NOTICE
If only one calibration point is currently active, the absorption coefficient is not calculated. The last valid value is used instead. At the first commissioning or after a reset, the default value, =7.70mm
2
/g, is used. The value can be changed by the user.
"Reference pulse rate" (*1F)
NOTICE
I0 is the pulse rate for the tube being empty (theoretical reference value). Generally, the value is significantly higher than any real pulse rate occurring during the measurement.
"Next point" (*19)
Options:
no This option must be selected if no further calibration point is to be entered or changed. After this selection the Gammapilot M returns to the group selection and the calibration is completed.
yes This option must be selected, if a further calibration point is to be entered or changed. The Gammapilot M returns to the "calibr. point" (*1A) function and the next point can be entered or changed.
On-site display Meaning
absorp. coeff. *1E
This function displays the absorption coefficient which results from the currently active calibration points. The displayed value should be used for plausibility checking.
7.70 mm2/g
On-site display Meaning
ref. pulse rate *1F This function displays the reference pulse rate IO, which results
from the currently active calibration points. The value cannot be edited.
31687 cps
On-site display Meaning
next point *19
This function is used to specify, if a further calibration point is to be entered or not.
no
yes
Page 69
Gammapilot M FMG60 HART Commissioning
Endress+Hauser 69
6.7.4 Linearization (for concentration measurements)
If the concentration is to be measured in a unit other than the "density unit" (*06), a linearization must be performed after the calibration. This can be done in the "Linearization" (*4) function group. The functions of this group and the linearization procedure are described in Operating Instructions BA00287F/00/EN, "Gammapilot M ­Description of Device Functions". This document can be found on the supplied CD-ROM.
6.7.5 Additional settings
After the calibration has been finished, the Gammapilot M outputs the measuring value via the current output and via the HART signal. The complete measuring range [min. density (*07) to max. density (*08)] is mapped to the current range 4 to 20 mA. Many additional functions are available for optimizing the measuring point. They can be configured as required. A detailed description of all instrument functions is given in Operating Instructions BA00287F/00/EN, "Gammapilot M - Description of Device Functions". This document can be found on the supplied CD-ROM.
6.8 Density measurement/temperature-compensated
Perform density measurement as described in the section "Calibration for density and concentration measurements" and then perform temperature calibration (see "Temperature compensation" section in BA00287F/00/EN "Gammapilot M - Description of Device Functions").
6.9 Gammagraphy detection
See "Gammagraphy" section in BA00287F/00/EN "Gammapilot M - Description of Device Functions".
Page 70
Maintenance and repairs Gammapilot M FMG60 HART
70 Endress+Hauser
7 Maintenance and repairs
7.1 Exterior cleaning
When cleaning the exterior, always use cleaning agents that do not attack the surface of the housing and the seals.
7.2 Repair
The Endress+Hauser repair concept assumes that repairs to the Gammapilot M are performed by the Endress+Hauser Service in Endress+Hauser workshops. For more information contact the Service Department at Endress+Hauser.
7.3 Repairs to Ex- or SIL-approved devices
When carrying out repairs to Ex- or SIL-approved devices, please note the following:
• Repairs to Ex- or SIL-approved devices may only be carried out by Endress+Hauser Service in Endress+Hauser workshops.
• Comply with the prevailing standards, national Ex-area regulations, safety instructions (XA) and certificates.
• Only use original spare parts from Endress+Hauser.
• A certified device may only be converted into a different certified device version by Endress+Hauser Service in Endress+Hauser workshops.
• Document all repair work and conversions.
7.4 Replacement
CAUTION
!
An uploading/download procedure is not permitted for safety-related use of the device.
After a complete device or electronic module has been replaced, the parameters can be downloaded into the device again via the communication interface. As a prerequisite, the data have to have been uploaded to the PC beforehand using the "FieldCare".
Level measurement and level limit detection
Measurement can be continued without a new setup. However, the calibration values should be checked as soon as possible, since the mounting position may have slightly changed.
Density and concentration measurement
After the replacement a new setup and calibration must be performed.
Page 71
Gammapilot M FMG60 HART Maintenance and repairs
Endress+Hauser 71
7.5 Return
7.5.1 Returning devices
The measuring device must be returned if repairs of a factory calibration are required, or if the wrong measuring device has been ordered or delivered. According to legal regulations, Endress+Hauser, an ISO-certified company, is required to follow certain procedures when handling returned products that are in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and conditions on the Endress+Hauser website at www.services.endress.com/return-material
7.6 Disposal
When disposing, separate and recycle the device components based on the materials.
7.7 Contact addresses of Endress+Hauser
Contact addresses can be found on our homepage: www.endress.com/worldwide. If you have any questions, please contact your Endress+Hauser subsidiary.
Page 72
Accessories Gammapilot M FMG60 HART
72 Endress+Hauser
8 Accessories
8.1 Commubox FXA195 HART
For intrinsically safe HART communication with FieldCare via the USB port. For details, see TI00404F/00/EN.
8.2 Commubox FXA291
The Commubox FXA291 connects Endress+Hauser field devices with a CDI interface (Endress+Hauser Common Data Interface) and the USB port of a computer or laptop. For details, see TI00405C/07/EN.
For Gammapilot M, you also need the "ToF adapter FXA291" accessory.
8.3 ToF adapter FXA291
The ToF adapter FXA291 connects the Commubox FXA291 with the Gammapilot M by means of the USB port of a computer or laptop. For details, see KA00271F/00/A2.
8.4 Field Xpert SFX100
Compact, flexible and robus industry handheld terminal for remote parametrization and measured value inspection via the HART current output or FOUNDATION Fieldbus. For details refer to Operating Instructions BA00060S/04/EN.
Page 73
Gammapilot M FMG60 HART Accessories
Endress+Hauser 73
8.5 Remote display FHX40
8.5.1 Dimensions
A0018138
A Wall-mounting (without mounting bracket) 1 Gammapilot M 3 Cable B Pipe-mounting (mounting bracket and plate supplied optionally) 2 Separat housing FHX40 4 Pipe
8.5.2 Ordering information
Use the cables provided for the corresponding communication version of the device to connect the remote display FHX40.
82 (3.23)
mm (in)
6.3 (0.25)
106 (4.17)
122 (4.8)
120 (4.72)
max. 80 (3.15) min. 30 (1.18)
96 (3.78)
160 (6.3)
8.5 (0.33)
118 (4.65)
180 (7.09)
1
2
3
h
b
t
AB
4
88(3.46)
010 Approval
A Non-hazardous area 2 ATEX II 2G Ex ia IIC T6 3 ATEX II 2D Ex ia IIIC T80°C H ATEX II 3G Ex ic IIC T6, T5 Gc (in preparation) G IECEx zone1 Ex ia IIC T6/T5 S FM IS Cl.I Div.1 Gr.A-D, zone 0 U CSA IS Cl.I Div.1 Gr.A-D, zone 0 N CSA General Purpose K TIIS Ex ia IIC T6 C NEPSI Ex ia IIC T6/T5 Gb Y Special version, TSP-no. to be spec.
020 Cable
1 20m/65ft (> HART) 5 20m/65ft (> PROFIBUS PA/FOUNDATION Fieldbus) 9 Special version, TSP-no. to be spec.
030 Additional Option
A Basic version B Mounting bracket, pipe 1"/ 2" Y Special version, TSP-no. to be spec.
995 Marking
1 Tagging (TAG), see additional spec.
Page 74
Accessories Gammapilot M FMG60 HART
74 Endress+Hauser
8.5.3 Technical data (cable and housing)
8.5.4 Materials
L00-FMxxxxxx-00-00-06-de-003
Max. cable length 20 m (66 ft) (fixed length including the cast-on plugs)
Temperature range Temperature class T5: -40 to +75 °C (-40 to +167 °F)
Temperature class T6: -40 to +60 °C (-40 to +140 °F)
Degree of protection IP65/67 (housing); IP68 (cable) acc. to IEC 60529
Materials Housing: AlSi12; cable glands: nickle plated brass
Dimensions [mm (in) 122x150x80 (4.72x5.91x3.15) / HxWxD
Position Part Material
1
Housing/Cover
AlSi12, Screw: V2A
Ground terminal CuZn nickel-plated, Screw: V2A
2Display Glass
3 Cable gland CuZn nickel-plated
4Cable PVC
5 Mounting bracket 316 Ti (1.4571) or 316 L (1.4435) or 316 (1.4401)
6Nut V4A
7 Plate Screw set (M5) 316 Ti (1.4571)
Spring washer: 301 (1.4310) or V2A Screw: V4A, Nut: V4A
Made
in Ger
Made in Ger
manmany
Maulb
y Maulb
urg
4
1
2
3
6
7
5
3
Page 75
Gammapilot M FMG60 HART Accessories
Endress+Hauser 75
8.6 Mounting device FHG60 (for level measurement and level limit measurement)
8.6.1 Dimensions
A0018139
1 Spacing as large as possible 2 Retainers (number and size depending on selected application
)
; Allen screws to ISO 4762 are supplied)
3 Mounting clamps (number depending on selected application
)
)
4 with preferred mounting method "housing head down" - bracket (for "level" application only
8)
)
5 with alternative mounting method "housing head up" - bracket (for "level" application only
)
)
Size of the retainers (depending on the selected application):
CAUTION
!
Max. torque for the screws of the retainers: 6 Nm (4.42 lbf ft).
M6x20-A2
M6x80-A2
2
3
B
40 (1.57)
70 (2.76)
70 (2.76)
190 (7.48)
88 (3.46)
116 (4.57)
mm (in)
#
max. 6 Nm (4.42 lbf ft)
1
4 5
88 (3.46)
35 (1.38)
A
øD
øC
63 (2.48)
35 (1.38)
110 (4.33)
230 (9.06)
8) See "Use" and "Ordering information for complete mounting set".
Mounting position at the FMG60
A [mm (in)] B [mm (in)] øC [mm (in)] øD [mm (in)] Mounting
Scintillator pipe dimensions 196 (7.72) 126 (4.96) 80 (3.15)
40 to 65 (1.57 to 2.56)
(a)
Electronic pipe dimensions 210 (8.27) 150 (5.91) 102 (4.02) (b)
Water cooling jacket dimensions
230 (9.06) 200 (7.87) 140 (5.51) (c)
Page 76
Accessories Gammapilot M FMG60 HART
76 Endress+Hauser
8.6.2 Application hints
Allowed
Preferred
A0018141
A Level measurement, FMG60 without water jacket B Level measurement, FMG60 with water jacket C Level limit measurement, FMG60 without water jacket D Level limit measurement, FMG60 with water jacket E Such horizontal mounting not permitted 1 Retainer for pipe
80 mm (3.15 in)
2 Retainer for water cooling jacket ,
140 mm (5.51 in)
3 Retainer for pipe
102 mm (4.72 in)
4Bracket
CAUTION
!
Consider while mounting the device
The mounting device must be fitted in a way such that it can withstand the weight of the
Gammapilot M under all operating conditions to be expected.
For a measuring length of 1600 mm (63 in) or more, two retainers (vertical) or three
retainers (horizontal) must be used.
For vertical mounting usage of the bracket or a support unit fitted by the customer is
mandatory. Otherwise a sufficient stability and support of the Gammapilot M is not ensured.
For stability reasons the mounting version with the compartment housing pointing
upwards should only be used in exceptional cases (e.g. lack of space).
To prevent damage of the detector tube, the maximum torque of the retainer screws is
6Nm (4.42lbfft).
8.6.3 Ordering information
Detailed ordering information is available as follows:
• In the Product Configurator on the Endress+Hauser website: www.endress.com Select country Instruments Select device Accessories for FMG60
• From your Endress+Hauser sales center: www.endress.com/worldwide
1 2
2
3
22
1
11
2
2
1
44 44
1
1
A
B C
D
E
Product Configurator - the tool for individual product configuration
• Configuration data updated on a daily basis
• Depending on the device: Direct input of data specific to measuring point, such as measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic generation of order code with breakdown in PDF or Excel output format
• Possibility to order directly from the Endress+Hauser online shop
Page 77
Gammapilot M FMG60 HART Accessories
Endress+Hauser 77
8.7 Clamping device for density measurement FHG61
For details, see SD01221F/00/EN.
8.8 Measurement section for density measurement FHG62
For drawings and description, see SD00540F/00/EN.
Page 78
Troubleshooting Gammapilot M FMG60 HART
78 Endress+Hauser
9Troubleshooting
9.1 System error messages
9.1.1 Error signal
Errors occurring during commissioning or operation are signaled in the following way:
• Error symbol, error code and error description on the display and operating module
• Current output, configurable ("output on alarm" (*20) function): – MAX, 110%, 22 mA – MIN, -10%, 3.6 mA – HOLD (the last value is held) – User-specific value
9.1.2 Last error
The last error is displayed in the "diagnostics" (*A) function group in the "previous error" (*A1) function. The displayed item can be deleted in the "delete last error" function (*A2).
9.1.3 Types of error
Type of error Symbol Meaning
Alarm (A)
continuous
The output signal assumes a value which can be set using the "output on alarm" (*10) function:
• MAX: 110%, 22 mA
•MIN: -10%, 3.8mA
• Hold: last value is on hold
• User-specific value
An error message is displayed.
Warning (W)
flashing
The device continues measurement. An error message is displayed (alternating with the measured value).
Page 79
Gammapilot M FMG60 HART Troubleshooting
Endress+Hauser 79
9.1.4 Error codes
Code Description Remedy
A102 checksum error Call Endress+Hauser Service
W103 initializing Wait for completion of the initializing procedure
A106 downloading Wait for completion of the download
A110 checksum error Call Endress+Hauser Service
A111 electronics defect Switch instrument off/on;
If error still is present: call Endress+Hauser service or exchange transmitter
A113 electronics defect Switch instrument off/on;
If error still is present: call Endress+Hauser service or exchange transmitter
A114 electronics defect Switch instrument off/on;
If error still is present: call Endress+Hauser service or exchange transmitter
A116 download error Repeat download
A121 electronics defect Switch instrument off/on;
If error still is present: call Endress+Hauser service or exchange transmitter
W153 initializing Wait for completion of the initializing procedure
A160 checksum error Call Endress+Hauser Service
A165 electronics defect • Switch instrument off/on;
If error still is present: call Endress+Hauser service or exchange transmitter
•See Note "Error messages A165 "electronics defect" and A635 "present date not defined" on ä 39.
A291 slave error Check basic setup and connection of the slave transmitter
A503 wrong sensor type Call Endress+Hauser Service
W513 calibration integration running Wait, until a stable pulse rate has been reached; then,
terminate integration (by pressing "E" in the "average pulse rate" (*11) function)
W514 PT-100 calibration Wait for the calibration to complete.
if error still persists: call Endress+Hauser Service
A531 sensor electronics defect Switch instrument off/on;
If error still is present: call Endress+Hauser service or exchange transmitter
A532 sensor voltage error Call Endress+Hauser Service
A533 wrong sensor software version Call Endress+Hauser Service
A535 sensor regulation error Call Endress+Hauser Service
W536 high voltage near limit Call Endress+Hauser Service
A538 sensor communication error Call Endress+Hauser Service
A602 linearization table not plausible Check monotony of the linearization table; if necessary, adjust
table ("linearization" (*4) function group)
A612 linearization table not defined Enter or complete linearization table
( "linearization" (*4) function group) The linearization table must contain the end points 0% = 1000 cps (standardized) and 100% = 0 cps (standardized). When entering via FieldCare: Select the right type of table (lin. tab. "level" or "concentration")
W621 simulation on Switch off linearization
("output" (*6) function group "simulation" (*65) function)
W640 SIL lock device SIL_locking not completed
Page 80
Troubleshooting Gammapilot M FMG60 HART
80 Endress+Hauser
W642 I_back calibration running Calibration of the current readback path active
A631 background not calibrated Perform background calibration
("calibration" (*1) function group)
A632 full/covered not calibrated Perform full/covered calibration
("calibration" (*1) function group)
A633 empty/free not calibrated Perform empty/free calibration
("calibration" (*1) function group)
A634 density not calibrated • Check: Has at least one calibration point been entered and
activated? If not: Enter and activate calibration point(s). ("calibration" (*1) function group)
• Check: Is the "reference pulse rate" (*1F) larger than 2
32
? If yes: re-perform density calibration ("calibration" (*1)) function group
A635 present date not defined • Enter present date
("basic setup" (*0) function group "today's date" (*01) function)
•See Note "Error messages A165 "electronics defect" and A635 "present date not defined" on ä 39.
A636 calibration date not plausible Check calibration date and enter it again
("system parameters" (*C) function group "calibration date" (*C7) function)
A637 operating mode not defined Enter operating mode
("basic setup" (*0) function group "operating mode" (*04) function)
A638 measurement mode not defined Enter measurement mode
("basic setup" (*0) function group "measurement mode" (*05) function)
A639 temperature compensation not
complete
Enter at least two "temperature - density" value pairs ("temperature compensation" (*3) function group)
W662 high sensor temperature (warning) Install water cooling jacket or thermal shielding
A663 sensor temperature too high (alarm) Install water cooling jacket or thermal shielding
A664 temperature measurement error Check correct functioning and connection of the PT-100 sensor
W681 current out of range
(3.8 to 20.5 mA)
Check calibration and linearization
A692 gammagraphy detected (alarm) • Check, if interference radiation is present or the "hold time"
(*54) is too short.
• If no interference radiation is present: decrease gammagraphy sensitivity ("gammagraphy" (*5) function group "sensitivity" (*52) function)
W693 gammagraphy detected (warning) Wait for end of the gammagraphy measurement
W695 measurement counter overflow The local dose rate is too high (where applicable, reduce with
blind flange).
Code Description Remedy
Page 81
Gammapilot M FMG60 HART Troubleshooting
Endress+Hauser 81
9.2 Possible calibration errors
Error Possible cause and remedy
Pulse rate too low at empty vessel
• Radiation source switched off
Switch on source at the source container
• Incorrect alignment of radiation beam
Realign beam
• Buildup in the vessel
Clean vessel or Recalibrate (if buildup is stable)
• Fittings in the vessel have not been considered in the activity calculation
Recalculate activity and change source accordingly, if required
• Pressure in the vessel has not been considered in the activity calculation
Recalculate activity and change source accordingly, if required
• No radiation source in the source container
Load source capsule in the container
• Source too weak
Use source with higher activity
Pulse rate to high at empty vessel
• Activity too high
Attenuate radiation, e.g. by mounting a steel plate in front of the source
container; or exchange source
•External radiation source (e.g. by gammagraphy)
Shield off if possible; repeat calibration without external radiation source
Pulse rate too high at full vessel
•External radiation source (e.g. by gammagraphy)
Shield off if possible; repeat calibration without external radiation source
Page 82
Troubleshooting Gammapilot M FMG60 HART
82 Endress+Hauser
9.3 Software history
Date Software version Software modifications Documentation
Since
09.2004
01.01.02 Original software. BA236F/00/en/08.04 52023878 BA287F/00/en/08.04 52023818
Since
11.2005
01.01.04 Bug fix.
Concentration mode corrected. Density measurement recalibration corrected.
Since
08.2006
01.01.06 Bug fix.
Corrections for high and low pulse rate.
Since
04.2007
01.02.00
01.02.02
Software extended to include "SIL locking" function. BA236F/00/en/03.07
71041168 BA287F/00/en/04.07 71041170
BA236F/00/en/06.07 71041168 BA287F/00/en/06.07 71041170
Since
08.2008
01.03.00 Automatic restart after error A165 caused by
empty pipe at density measurements (pulse rate > 160000 c/s)
NOTICE
For instruments with SIL or WHG approvals, software version 01.02.02 remains valid.
BA236F/00/en/09.08 71082936 BA287F/00/en/06.07 71041170
Since
02.2009
01.03.02 New filtering function for Gamma Modulator
FHG65 implemented
BA236F/00/en/03.09 71091966 BA287F/00/en/06.07 71041170
Since
10.2010
01.03.06 Improved EMC stability for extreme interferences. BA236F/00/en/10.09 71104595 BA287F/00/en/06.07 71041170
Page 83
Gammapilot M FMG60 HART Technical Data
Endress+Hauser 83
10 Technical Data
10.1 Additional technical data
For the technical data, please refer to the Technical Information TI00363F/00/EN.
10.2 Documentation
This supplementary documentation can be found on our product pages at www.endress.com.
• Technical Information (TI00363F/00/EN)
• Operating Instructions "Description of Device Functions" (BA00287F/00/EN)
• Safety Manual "Functional Safety Manual": –
For max. point level detection
(SD00230F/00/EN)
– For min. point level detection (SD00324F/00/EN)
10.2.1 Clamping device for level and point level detection FHG60
For details, see SD01202F/00/EN.
10.2.2 Clamping device for density measurement FHG61
For details, see SD01221F/00/EN.
10.2.3 Measurement section for density measurement FHG62
For drawings and description, see SD00540F/00/EN.
10.2.4 Certificates and approvals
Safety Manual (SIL 2/3)
SIL 2/3 in accordance with IEC 61508 see
SD00230F/00/EN "Functional Safety Manual" (for max. point level detection)
• SD00324F/00/EN "Functional Safety Manual" (for min. point level detection)
Ex approval
The available certificates are listed in the ordering information. Observe the related safety instructions (XA) and Control Drawings (ZD).
Page 84
Technical Data Gammapilot M FMG60 HART
84 Endress+Hauser
Certificates
General
Approval number: NEPSI GYJ101145
Approval number: IECEx BKI 05.0001
Allocation to the option, see the Product Configurator on the Endress+Hauser website: www.endress.com Select country Instruments Select device Product page function: Configure this product
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
A Non-hazardous area A 1, 2, 3
F Non-hazardous area, WHG A 1
N CSA General Purpose A 1, 2, 3
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
CEx de [ia] IIC T6
C 1 XA00536F
C 2, 3 XA00537F
B 1 XA00536F
B 2, 3 XA00537F
D Ex d [ia] IIC T6
D 1 XA00536F
D 2, 3 XA00537F
E 1 XA00536F
E 2, 3 XA00537F
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
G Ex de [ia Ga] IIC T6 Gb
B 1 XA00449F
B 2, 3 XA00450F
C 1 XA00449F
C 2, 3 XA00451F
H Ex d [ia Ga] IIC T6 Gb
E 1 XA00449F
E 2, 3 XA00450F
D 1 XA00449F
D 2, 3 XA00451F
Page 85
Gammapilot M FMG60 HART Technical Data
Endress+Hauser 85
Approval number: KEMA 04 ATEX 1153
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
1 II 2(1) G Ex de [ia Ga] IIC T6 Gb
B 1 XA00303F
B 2, 3 XA00332F
C 1 XA00303F
C 2, 3 XA00334F
2 II 2(1) G Ex de [ia Ga] IIC T6 Gb, WHG
B 1 XA00303F
C 1 XA00303F
3 II 2(1) G Ex d [ia Ga] IIC T6 Gb
E 1 XA00303F
E 2, 3 XA00332F
D 1 XA00303F
D 2, 3 XA00334F
4 II 2(1) G Ex d [ia Ga] IIC T6 Gb, WHG
E 1 XA00303F
D 1 XA00303F
5 II 2(1) D Ex tb [ia Da] IIIC T80°C Db
F 1 XA00304F
F 2, 3 XA00335F
L 1 XA00304F
L 2, 3 XA00333F
6
II 2(1) G Ex de [ia Ga] IIC T6 Gb II 2(1) D Ex tb [ia Da] IIIC T80°C Db
J1
XA00303F XA00304F
J2, 3
XA00332F XA00333F
G1
XA00303F XA00304F
G2, 3
XA00334F XA00335F
7
II 2(1) Ex de [ia Ga] IIC T6 Gb, WHG II 2(1) D Ex tb [ia Da] IIIC T80°C Db, WHG
J1
XA00303F XA00304F
G1
XA00303F XA00304F
8
II 2(1) G Ex d [ia Ga] IIC T6 Gb II 2(1) D Ex tb [ia Da] IIIC T80°C Db
K1
XA00303F XA00304F
K2, 3
XA00332F XA00304F
H1
XA00303F XA00304F
H 2, 3 XA00334F
XA00335F
M
II 2(1) G Ex d [ia Ga] IIC T6 Gb, WHG II 2(1) D Ex tb [ia Da] IIIC T80°C Db, WHG
K 1 XA00303F
XA00304F
H 1 XA00303F
XA00304F
Page 86
Technical Data Gammapilot M FMG60 HART
86 Endress+Hauser
Approval number: ID 3022785
Approval number: CSA 1653884
Approval number: TC17525, TC19557 (NaJ-Scintillator)
Approval number: TC17524, TC19556 (PVT-Scintillator)
CE mark
The measuring system meets the legal requirements of the EC-guidelines. Endress+Hauser confirms the device passing the required tests by attaching the CE-mark.
GOST
Approval for GOST available.
Overfill protection
WHG for level limit detection.
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
S
FM Cl. 1 Gp. A-D, Cl. II Gp. E-G, Cl. III, Cl. I Zone 1 Ex d [ia] IIC t6
D 1 XA01100F
D 2, 3 XA01108F
E 1 XA01102F
E 2, 3 XA01109F
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
P
CSA Cl. I Gp. A-D, Cl. II Gp. E-G, Cl. III, Cl. I Zone 1 Ex d [ia] IIC T6
D 1 XA01099F
D 2, 3 XA01110F
E 1 XA01101F
E 2, 3 XA01111F
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
K TIIS Ex d [ia] IIC T6 D 1 BA00236F
Approval Feature 010
Type of protection Power supply wiring/
Output wiring Option 030
Output (Communication) Option 040
Approval number
K TIIS Ex d [ia] IIC T6 D 1 BA00236F
Page 87
Gammapilot M FMG60 HART Technical Data
Endress+Hauser 87
Page 88
Appendix Gammapilot M FMG60 HART
88 Endress+Hauser
11 Appendix
11.1 Operating menu for level measurements
A0018143-EN
*0 *01
L50
L60
LA0
LC0
*04
L40
L61
L92
LA1
*03
L41
L42
L62
L93
LA2
LC2
*02
L63
LA3
L68
L95
LA4
LC4
L4
L5
L6
L9
LA
LC
L51 L52
L53
L1 L10 L11
E
L12
L13
L14
L54
*05
W
L17
L2 L20
L21
ESC
basic setup todays date
dd.mm.yy hh:mm
gammagr . detect
on
off
commun. address
present error
tag no.
operating mode
linearisation
linear
standard clear table
no. of preambles
language
previous error
isotope
137Cs
60 Co no compens.
table no
low output limit
on
off
back to home
specify time
Default: 100 s
clear last error
keep
erase
protocol + sw-no.
beam type
standard / cont.
modulated
input level
curr. output mode
standard
fixed current
current turn down
reset
for reset code see manual
Note! The Default values of the parameters are typed in bold.
4 mAvalue
no. of decimalas
x x.x
x.xxx
x.xx
unlock parameter
for unlock code see manual
serial no.
linearisation
gammagraphy
output
display
diagnostics
system parameters
L42
define table table on
span time
minimal time from min. to max. value
sensitivity
output gammagr .
slave end slave not defined
calibrationbackgr . calib.
start
stop/edit
avg. pulse rate bgr. pulse rate
-1: no
t calibrated
calibr . point
full/covered
empty/free
value full
target value at calibration point
hold time
master
stand alone
meas. mode
level (see separate menus for other selections)
*05
value empty
target value at calibration point
MIN, -10%, 3.6 mA MAX, 110%, 22 mA hold
safety settings output on alarm
MAX 110% 22 mA
hold MIN - 10% 3,6 mA
user specific
output on alarm
specify value
Page 89
Gammapilot M FMG60 HART Appendix
Endress+Hauser 89
A0018149-EN
L65
L96
L64
LA5
LC6
L66
L97
LA6
LC7
L67
L0B
L69
L15 L11
E
L16
L15 L11
E
L18
L19
L43L11
E
L44
L4B
L56L55
.
.
.
..
temperature unit
calibration date
°C
°F
dd.mm.yy hh:mm
pres. avg. pulsr avg. raw pulsr
decay compensated w/o background rate
at integration time
not compensated
sep. character
. period
, comma
display test
off
on
simulation
sim. off
sim pulserate sim. level sim. current
fixed cur . value
simulation value
output current
20 mA value
gammagr . count
keep erase
gammagr . count
avg. pulse rate normalized pulser
-1: not calibrated
next point
yes no
calibration
start
stop/edit
Return to Group selection
calibration
start
stop/edit
avg. pulse rateempty calibr
-1: not calibrated
next point
yes no
calibration
start
stop/edit
avg. pulse rate full calibr
-1: not calibrated
output damping
1 … 999 s
Default: 6 s
Page 90
Appendix Gammapilot M FMG60 HART
90 Endress+Hauser
11.2 Operating menu for level limit detection
A0018150-EN
S2 S20
S21
S22
S23 S24 S25
S2D
*0 *01
S50
S60
SA0
SC0
*04
S61
S92
SA1
*03
S62
S93
SA2
SC2
*02
S63
SA3
S68
S95
SA4
SC4
S5
S6
S9
SA
SC
S51 S52
. S53
S1 S10 S11
E
S12
S13
S54
*05
unlock parameter
for unlock code see manual
serial no.
reset
for reset code see manual
clear last error
keep
erase
protocol + sw-no.tag no.
Note! The Default values of the parameters are typed in bold.
system parameters
no. of decimalas
x x.x
x.xxx
x.xx
back to home
specify time
Default: 100 s
4 mA value
hold time
curr. output mode
standard
fixed current
current turn down
output gammagr
MIN, -10%, 3.6 mA MAX, 110%, 22 mA
hold
of
low output limit
on
f
sensitivity
gammagr . detect
on
off
commun. address
present error
no. of preambles
language
previous error
span time
minimal time from min. to max. value
gammagraphy
output
display
diagnostics
password confirm loutconfirm sequence
validvalid
password
not valid not valid
. calib. . pulse rate
. point
calibrationbackgr
start
stop/edit
avg. pulse rate bgr
-1: not calibrated
calibr
full/covered
empty/free
safe
ty settings output on alarm
MAX (22 mA)
hold MIN (<= 3.6 mA)
user specific
output on alarm
lock device
safety locking
unlocked
locked
unlock device
(see separate menus
meas. mode
limit
for other
selections)
operating mode
slave end slave not defined
master
stand alone
isotope
137Cs
60 Co no compens.
beam type
standard / cont.
modulated
basic setup todays date
dd.mm.yy hh:mm
Page 91
Gammapilot M FMG60 HART Appendix
Endress+Hauser 91
A0018151-EN
S27 S28 S29 S2A S2B S2CS26
.
S65
S96
S64
SA5
SC6
S66
S97
SA6
SC7
S67
S0B
S69
S15 S11
E
.
S16
S15 S11
E
. S18
S19
S56S55
simulation value
output current
simulation
sim. off
sim. pulserate sim. level sim. current
display test
off
on
avg. raw pulsr
calibration date
dd.mm.yy hh:mm
at integration time not compensated
.
sep. character
fixed cur. value
. period
, comma
pres. avg. pulsr
temperatiure unit
20 mA value
°C
°F
decay compensated
w/o background rate
..
gammagr. count
keep
erase
gammagr. count
conf. backgr
valid
not valid
conf. calibr
valid
not valid
conf. source
valid
not valid
conf. time
valid
not valid
conf. date conf. length
valid valid
valid
not valid not valid
not valid
conf. passwd
next point
yes no
full calibr
-1: not calibrated
empty calibr
-1: not calibrated
avg.
pulse rate
avg. pulse rate
calibration
start
stop/edit
calibration
start
stop/edit
stop/edit
output damping
1 999 s
Default: 6 s
Return to Group selection
Page 92
Appendix Gammapilot M FMG60 HART
92 Endress+Hauser
11.3 Operating menu for density and concentration measurements
A0018152-EN
ESC
*2
*0
*20
*01
*50
*60
*A0
*C0
*21
*04
*47
*61
*92
*A1
*03
*48
*49
*62
*93
*A2
*C2
*02
*63
*A3
*05
*68
*95
*A4
*C4
*4
*5
*6
*9
*A
*C
*51 *52
. *53
*1 *10 *11
E
*12
*1A
*15 *11
E
*54
*30 *31 *32 *33 *34
*3
*45
*46
ESC
Note! The Default values of the parameters are typed in bold.
present error
tag no.
language
previous error
back to home
specify time
Default: 100 s
clear last error
keep
erase
protocol + sw-no.
reset
for reset code see manual
no. of decimalas
x x.x
x.xxx
x.xx
unloc
k parameter
for unlock code see manual
serial no.
display
diagnostics
system parameters
of
. detectgammagr
on
off
commun. address no. of preambleslow output limit
on
f
curr. output mode
standard
fixed c
urrent
current turn down
4 mA value
gammagraphy
output
span time
minimal time from min. to max. value
sensitivity
output gammagr hold time
MIN, -10%, 3.6 mA MAX, 110%, 22 mA
hold
wad
linearisation
clear table
table no
input density
linearisation
edit table table on
unit selection
customer unit
g/cm3 g/l
lb/gal lb/ft3
°Brix °Baume °API °T
customer unit
% % mass %vol
sol/vol g dm/l
density unit
f..temp. compens. select temper temperature density linear coef
temp. compensation
reference lowest highest
clear
not used
used
operating mode
slave end slave
not defined
. point
calibration
start
stop/edit
avg. pulse rate
master
stand alone
meas. mode
density/concentration
(see separate menus for other selections)
output on alar
m
bgr
safety settings
basic setup
MAX 110% 22 mA
hold MIN - 10% 3,6 mA
user specific
todays date
dd.mm.yy hh:mm
output on alarm
specify value
isotope
137Cs
60 Co no compens.
beam type
standard / cont.
modul
ated
calibrationbackgr . calib.
start
stop/edit
avg. pulse rate . pulse rate
-1: not calibrated
calibr
1…10
Page 93
Gammapilot M FMG60 HART Appendix
Endress+Hauser 93
A0018153-EN
*65
*96
*64
*A5
*C6
*66
*97
*A6
*C7
*67
*09
*69
*1B
*19
*4A
*4B
*56*55
*06
*07
*1C *1D
*1E *1F
*35 *36
*08
*A7 *A8
*C8
*0A *0B
fixed cur
20 mA
.
..
simulation
sim. off
sim. pulserate. sim. density/concentr sim. current
sep. character
. value
. period
, comma
pres. avg. pulsr
temperatiure unit
simulation value
display test
off
on
avg. raw pulsr
calibration date
output current
value
gammagr . count
keep erase
gammagr . count
°C
°F
dd.mm.yy hh:mm
medium temp. density value
recalibration date
dd.mm.yy hh:mm
decay compensated w/o background rate
at integration time
not compensated
.
f.
y
Return to Group selection
input concentr
next point
yes no
square coefnext point
es
no
next point
yes no
1…
pipe diam. outp. damp.
999 s
Default: 60 s
ref. pulse ratef.calibrabsorp. coef
pipe diam. unit
density calibr
-1: not calibrated
density unit
g/cm3
g/l …
min. density
density value
max. density
without temp. compensation
. point
not used used clear
Page 94
Gammapilot M FMG60 HART
94 Endress+Hauser
Index
Symbols
*01 - Present date. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
*03 - Isotope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
*04 - Operating mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
*05 - Measuring mode . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
*06 - Density unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
*07 - Min. density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
*08 - Max. density . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
*09 - Pipe diameter unit. . . . . . . . . . . . . . . . . . . . . . . . . . 43
*0A - Pipe diameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
*0B - Output damping . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
*10 - Background calibration. . . . . . . . . . . . . . . . . . . 49, 63
*11 - Average pulse rate (background) . . . . . . . . . . 50, 64
*11 - Average pulse rate (density) . . . . . . . . . . . . . . . . . 66
*11 - Average pulse rate (level/limit) . . . . . . . . . . . . . . 53
*12 - Background pulse rate . . . . . . . . . . . . . . . . . . . 50, 64
*13 - Calibration point (level/limit) . . . . . . . . . . . . . . . . 51
*14 - Value full . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
*15 - Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
*15 - Calibration (density) . . . . . . . . . . . . . . . . . . . . . . . . 66
*16 - Full calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
*17 - Value empty. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
*18 - Empty calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 53
*19 - Next point (density) . . . . . . . . . . . . . . . . . . . . . . . . 68
*19 - Next point (level/limit). . . . . . . . . . . . . . . . . . . . . . 54
*1A - Calibration point . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
*1B - Density calibration . . . . . . . . . . . . . . . . . . . . . . . . . 67
*1C - Density value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
*1D - Calibration point . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
*1E - Absorption coefficient . . . . . . . . . . . . . . . . . . . . . . 68
*1F - Reference pulse rate . . . . . . . . . . . . . . . . . . . . . . . . 68
*20 - Output on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
*21 - Output on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Numerics
022 - Security locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
023 - Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
024 - Confirm Iout. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
025 - Confirm sequence . . . . . . . . . . . . . . . . . . . . . . . . . . 58
026 - Confirm backgr.. . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
027 - Confirm cal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
028 - Confirm source . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
029 - Confirm time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
02A - Confirm date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
02B - Confirm length . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
02C - Confirm password . . . . . . . . . . . . . . . . . . . . . . . . . . 60
02D - Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
A
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Automatic calibration . . . . . . . . . . . . . . . . . . . . . . . . . 48, 62
B
Background calibration. . . . . . . . . . . . . . . . . . . . . . . . 46–47
Basic setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
C
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Calibration errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
Calibration points . . . . . . . . . . . . . . . . . . . . . . . . . 46–47, 61
Cascading mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Centering knob . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Clamping device for density measurements . . . . . . 77, 83
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Commubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Concentration measurement. . . . . 15–16, 55–56, 61, 69
Cooling water connections. . . . . . . . . . . . . . . . . . . . . . . . . . 8
Covered calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
D
Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . 10
Density measurement. . . . . . . . . . . 15–16, 55–56, 61, 69
Detector pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Display operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32, 35
Display symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
E
Einpunkt-Abgleich . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Empty calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Error codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
F
FHX40. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27, 73
Fieldbus connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Free calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Full calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
Function code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
H
Hardware security locking . . . . . . . . . . . . . . . . . . . . . . . . . 37
HART protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Hazardous area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
I
Incoming acceptance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
K
Key assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Klemmvorrichtung für Dichtemessung . . . . . . . . . . . . . . 83
L
Level limit contactor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Level limit detection . . . . . . . . . . . . . . . . . . . . . . . . . . 14, 46
Level measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . 13, 46
Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Page 95
Gammapilot M FMG60 HART
Endress+Hauser 95
M
Manual calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 48, 62
Measurement section . . . . . . . . . . . . . . . . . . . . . . . . . 77, 83
Measuring range marks . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34, 88
Mounting collar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
mounting device FHG60 . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Multiple-point calibration . . . . . . . . . . . . . . . . . . . . . . . . 62
N
Nameplates. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8–9
O
One-point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Onsite display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Operating options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
P
PML connector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
R
Radiation protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Recalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
S
Safety instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Security locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
service interface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Service interface FXA291 . . . . . . . . . . . . . . . . . . . . . . . . . 72
Software history. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Software security locking . . . . . . . . . . . . . . . . . . . . . . . . . 37
Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
T
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Terminal compartments . . . . . . . . . . . . . . . . . . . . . . . . 8, 20
Terminal head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
ToF Tool - FieldTool Package . . . . . . . . . . . . . . . . . . . . . . 36
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Types of error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
U
Unlock parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
V
VU331. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
W
Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Water cooling. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Page 96
www.addresses.endress.com
71253004
Loading...