EFD 750 INSTALLATION GUIDE

Introduction
The 750 Series fluid dispense valves are simple to use and will operate many millions of cycles without maintenance. The 750 Series incorporates a variety of compact, precise, adjustable diaphragm valves for dispensing low to medium viscosity fluids. These include the 750V, 750V-SS, 751V, 752V, 752V-SS, 752V-UH, 752V-UHSS, 752V-DVD and 752V-UHDVD models.
All 750 Series valves use the same actuating assembly. But fluid body styles and diaphragms differ according to the model. The 752V Series fluid bodys are con­structed with the outlet at the end of the fluid body. The 750V and 751V fluid bodys are constructed with the outlet on the side of the fluid body.
Each valve is shipped with a dispensing tip adapter, fluid inlet fitting and 5-foot actuating air hose installed.
750 Series Dispense Valve
INSTALLATION GUIDE
USA: 800-556-3484 or +1-401-434-1680 Europe: 0800 585733 or +44 (0) 1582 666334 Asia: +86 (21) 5854 2345 technical@efd-inc.com www.efd-inc.com
Electronic pdf files of EFD®manuals are also available at www.efd-inc.com/manuals.html.
®
A NORDSON COMPANY
Stroke control knob
Stroke reference ring
Fluid supply line
Inlet fitting
Control air hose
Fluid body
Dispensing tip
Installation
Set desired deposit size by adjusting valve open time. Refer to valve controller operating manual.
11
1. Mount valve with an EFD universal
valve mount (#7002VM) or other fixture.
2. Connect fluid supply line to valve. If
3/8” OD tubing is used, change to fitting #7610BP supplied.
3. Connect the fluid supply line to reservoir.
The reservoir can accept either OD or 1/4” OD or 3/8” OD tubing using #62518PT supplied.
4. Connect valve control air hose to
ValveMate
controller (or other pneu­matic switch) used to control valve open time.
5. Choose a dispensing tip—small tips
(20 gauge) for low-viscosity fluids and larger (14 gauge) for higher viscosities.
6. Fill reservoir by pouring fluid directly
into tank liner or manufacturer’s bottle placed inside reservoir. Secure cover prior to setting pressure.
7. Set reservoir pressure to low for thin
fluids and higher for thick fluids.
8. Set the diaphragm stroke starting with
no more than 1/2 turn open.
*
9. Place a cup under the dispensing tip
and actuate the valve until fluid lines, valve and dispensing tip are free of air.
10. Set desired flow rate by adjusting
fluid reservoir pressure or changing dispensing tip.
*
Do not overtighten the stroke adjustment knob or open it more than two full turns. If open more than two turns, pressurized liquid could force open the diaphragm seal, resulting in continuous liquid flow.
9
Open
Close
8
Air Output
ut
Exhaust
r
Prior to installing this valve, please read the associated reservoir and valve controller operating instructions to become familiar with the operation of all components of the dispensing system.
1
7
5
2
6
4
10
3
To exhaust
To pressurize
A NORDSON COMPANY
®
Steady
Test
Run
Setup Program
5
20
0
Time Set
Clear
Fast Slow
POWER
RUN
SETUP
CYCLE
Fast Slow
Time Set
Pressure
STOP
Time
Purge
Loading...
+ 2 hidden pages