Figure 1
➁ Rotate fluid
chamber to desired
position.
➀ Set stroke
adjustment two
full turns out.
Note: Do not rotate fluid chamber more than one
complete turn.
Install an appropriate size dispensing tip on
the output tip adapter. The adapter accepts
EFD SafetyLok™tips. Use small diameter
(20 to 25 gage) for low viscosity fluids, and
larger tips (14 to 18 gage) for higher viscosity fluids.
Fill the fluid reservoir (refer to reservoir user's
guide).
After filling, check to be certain the reservoir
cover is secured and that all connections
are tight.
Set the reservoir pressure. For low viscosity
fluids, start with a pressure of approximately
5 psi (0.3 bar). For higher viscosity, use 20 to
40 psi (1.4 to 2.8 bar). Increase or decrease
as necessary.
Note: If solvents or other watery fluids are to be
dispensed in very small amounts, specify 0 to 15
psi (0 to 1.0 bar) precision regulator to control
reservoir pressure.
Set the needle stroke. Start at one full turn
open.
Set the valve control air pressure to 70 psi
(4.8 bar). Refer to the VALVEMATE™7000
manual.
Open the valve with an air pulse long
enough to fill the valve and start fluid flow.
Then test the dispensed amount with a
nominal time setting.
Increase or decrease reservoir pressure,
needle stroke and valve open time to arrive
at the desired deposit size.
Setup
Mount the 740V valve with an EFD universal
valve mount (#7002VM) or other appropriate
fixture, using the 1/4-28 UNF tapped hole on
the air cylinder body.
The 740V Series valves are shipped with the
fluid inlet fitting installed for use with 1/4" OD
x 1/8" ID flexible fluid hose. An optional fluid
inlet fitting for 3/8" OD tube is enclosed.
Connect the control air line to the valve controller or other pneumatic switch that is used
to control operating air to the valve. (Refer to
EFD VALVEMATE™7000 operating manual.)
Connect the fluid feed hose to the fluid inlet
fitting and fluid reservoir.
The valve fluid chamber with inlet fitting can
be rotated to prevent possible interference
with machine parts (See Figure 1).
To reposition the fluid chamber, first turn the
needle stroke control out two full turns ➀ ,
then rotate the chamber ➁ to the desired
position.
7 / Setup Instructions