Repairs/Modifications
Repairs on this device shall be performed only by DrägerService, a division
of North American Dräger (NAD), or its Authorized Service Centers.
Information about repairs can be obtained from Dräger or Authorized
Dealers. NAD or Dräger will not be responsible for injury to persons or
damage to property arising directly or indirectly out of unauthorized repairs
or modifications to this device. Furthermore, any unauthorized repairs or
modifications void any warranty extended by Dräger.
This document is provided for your information only. It will not be
exchanged or updated without request.
Trademarks
The Dräger name and logo and the name Babytherm® are registered
trademarks of Dräger Medizintechnik GmbH.
For correct and effective use of the product and in order to minimize
hazards it is mandatory to carefully read and to observe all portions of this
manual.
The design of the equipment, the accompanying literature, and the labeling
on the equipment take into consideration that the purchase and use of the
equipment are restricted to trained professionals, and that certain inherent
characteristics of the equipment are known to the trained operator.
Instructions, warnings, and caution statements are limited, therefore,
largely to the specifics of the Dräger design. This publication excludes
references to various hazards which are obvious to a medical professional
and operator of this equipment, to the consequences of product misuse, and
to potentially adverse effects in infants with abnormal conditions. Product
modification or misuse can be dangerous. Dräger disclaims all liability for
the consequences of product alterations or modifications, as well as for the
consequences which might result from the combination of this product with
other products whether supplied by Dräger or by other manufacturers if
such a combination is not endorsed by Dräger.
Introduction
Introduction
The operators of the heated bassinet system must recognize their
responsibility for choosing appropriate safety monitoring that supplies
adequate information on equipment performance and infant condition.
Infant safety may be achieved through a wide variety of different means
ranging from electronic surveillance of equipment performance and infant
condition to simple, direct observation of clinical signs. The responsibility
for the selection of the best level of infant monitoring lies solely with the
equipment user.
Limitation of Liability
Dräger's liability, whether arising out of or related to manufacture and sale
of the goods, their installation, demonstration, sales representation, use,
performance, or otherwise, including any liability based upon Dräger's
Product Warranty, is subject to and limited to the exclusive terms and
conditions as set forth, whether based upon breach of warranty or any other
cause of action whatsoever, regardless of any fault attributable to Dräger
and regardless of the form of action (including, without limitation, breach of
warranty, negligence, strict liability, or otherwise).
1
Page 8
Introduction
Dräger shall not be liable for, nor shall buyer be entitled to recover any
special incidental, or consequential damages or for any liability incurred by
buyer to any third party in any way arising out of or relating to the goods.
Limited Warranty
All Dräger products are guaranteed to be free of defects in material and
workmanship for a period of one year from date of delivery.
The following are exceptions to this warranty:
1. Defects caused by misuse, mishandling, tampering, or by modifications
THE STATED EXPRESSED WARRANTlES ARE IN
LlEU OF ALL OTHER WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING, WITHOUT LIMITATION,
WARRANTlES OF MERCHANTABILITY, FITNESS
FOR ANY PARTICULAR PURPOSE, OR
NONINFRINGEMENT.
not authorized by Dräger or its representatives are not covered.
2. All consumable and disposable products are guaranteed to be free from
defects upon shipment only.
Any product which proves to be defective in workmanship or material will
be replaced, credited, or repaired with Dräger holding the option. Dräger is
not responsible for deterioration, wear, or abuse. In any case, Dräger will
not be liable beyond the original selling price.
This warranty is rendered void and Dräger shall not be held liable for
condition as resultant from:
1. Damage to the unit is incurred as a result of mishandling.
2. The customer fails to maintain the unit in a proper manner.
3. The customer uses any parts, accessories or fittings not specified or sold
by Dräger.
4. Sale or service is not performed by an authorized service/dealer.
Application of this warranty is subject to the following conditions:
1. Dräger or its authorized representative must be promptly notified, in
writing, upon detection of the defective material or equipment.
2. Contact Dräger Customer Service at 1-800-4DRAGER for a Return
Authorization Number.
3. Defective material or equipment must be returned, shipping prepaid, to
Dräger or its authorized representative.
4. Examination by Dräger or its authorized representative must confirm
that the defect is covered by the terms of this warranty.
2
Page 9
Introduction
5. Notification in writing, of defective material or equipment must be
received by Dräger or its authorized representative no later than two (2)
weeks following expiration of this warranty.
The above is the sole warranty provided by Dräger. No other warranty
expressed or implied is intended. Representatives of Dräger are not
authorized to modify the terms of this warranty.
Dräger, Telford, PA
3
Page 10
Introduction
g
Definitions and Symbols
Warnings
Cautions
Notes
Terminology
A WARNING statement refers to conditions with a possibility of personal
injury or death if disregarded.
A CAUTION statement designates the possibility of damage to equipment
or improper operation if disregarded.
A NOTE provides additional information intended to clarify apparently
contradictory or confusing situations.
Inspectionexamination of actual condition
Servicemeasures to maintain specified condition
Repairmeasures to restore specified condition
Maintenanceinspection, service, and repair, where
necessary
Preventive
Maintenance
Maintenance measures at regular
intervals
Typing conventions in this manual:
LED messages are printed in bold type:
Symbols
Example:
CAUTION
!
ATTENTION
SEt
:
:
Refer to accompanying documents before
operatin
Consulter l es documents ci -joints avant de faire
fonctionner l’appareil.
Denotes IEC601-1 type of applied part.
Protective earth symbol.
equipment.
4
Page 11
General WARNINGS and CAUTIONS
Warning: Strictly follow this Operator's Instruction Manual. Any use of
the product requires full understanding and strict
observation of all portions of these Instructions. The
equipment is only to be used for the purpose specified under
"Intended Use." Observe all WARNINGS and CAUTIONS as
rendered throughout this manual and on labels on the
equipment.
Warning: This device is to be used only in rooms with line power
installations complying with national safety standards for
hospital infant rooms. To maintain grounding integrity,
connect only to a "hospital grade" receptacle. Always
disconnect power before servicing.
Warning: Do not use in the presence of flammable anesthetics. Turn off
and disconnect oxygen supply when performing cleaning and
service. Keep matches, lighted cigarettes, and all other
sources of ignition out of the room in which the Babytherm is
located. Small quantities of flammable agents, such as ethyls
and alcohol, can cause a fire in connection with oxygen left in
the Babytherm in the presence of ignition sources.
Introduction
Warning: Do not use cellular phones or similar wireless communication
equipment within 33 feet (10 m) of the Babytherm 8000 LDR.
Wireless phones may cause malfunction in electromedical
equipment.
Caution: Restriction of Distribution
: Federal Law restricts this device
to sale by or on the order of a physician.
Caution: Maintenance
your local DrägerService
: In case of malfunction of this device, contact
SM
technician or our Factory
Authorized Technical Service Center. The device must be
inspected by competent and factory authorized technical
service representatives at regular intervals. A record must be
kept on this preventive maintenance. Maintenance or repair
of the Babytherm 8000 LDR shall be performed only by
Dräger authorized technical service representatives.
Warning: Never operate the heated bassinet if it has suffered physical
damage or does not seem to operate properly. In this case,
always refer servicing to properly trained and factory
authorized service personnel.
5
Page 12
Introduction
Precautions During Preparation
Warning: Use of gel mattress when mattress heater is switched off will
result in infant becoming hypothermic. Use foam rubber
mattress when mattress heater is not in use for extended
periods (> 15 min.).
Caution: Never route infant circuits or lines over top of panels – they
may get pinched when folding up panels or when installing
cover.
Precautions During Operation
Warning: To prevent infant fall, do not leave infant unattended when
any panels are down. All panels must be erect and secure
during transport to prevent infant fall.
Warning: When folding up surrounding panels, take care not to pinch
infant body parts, circuits, or other items, such as sheets.
Warning: Temperature Monitoring
skin condition at regular intervals and adjust temperature
setting to individual infant needs. Infant temperature must
be monitored particularly carefully during phototherapy or
while a radiant warmer is used in conjunction with the
Babytherm heated bassinet.
Warning: Mattress Heater
mattress must be placed directly on the heated tray surface.
Always take into account warm-up time of the mattress
(approximately 1 hour) before placing an infant in the
bassinet.
Warning: To minimize infant overheating, always consider effect of
additional heat sources, such as phototherapy lamps or
radiant warmers. They increase the heat directed to the
infant. Avoid direct sunlight.
Warning: Do not obstruct conductive heat transfer by placing
additional blankets or other insulating covers underneath the
infant.
Warning: Set temperatures in the lower extended temperature range
only on the order of a physician. Check the infant's
temperature and skin condition at regular intervals.
: When mattress heating is used, the gel
: Check the infant’s temperature and
Warning: Check the infant's temperature and skin condition at regular
intervals.
6
Page 13
Introduction
Warning: Always use a proper fitting thin sheet (Part No. 2M21272).
Do not fold the sheet under the gel mattress more than 3". To
prevent infant fall in Trendelenberg or reverse Trendelenberg
position always use proper fitting mattress sheet.
Warning: When used with an infant radiant warmer, check its
operator’s manual for directions of the correct heat level and
distance to the infant.
Warning: Use of radiant warmers may increase insensible water loss.
Measures to maintain desired fluid levels should be
considered.
Warning: Use of Radiant Warmer
recommended. Dräger cannot warrant or endorse the safe
performance of third party radiant warmers for use with the
Babytherm 8000 LDR. Specifically, combinations with
radiant warmers that are not intended for use with heated
bassinets bear risks of excessive overall heat transfer to the
infant.
Caution: Remove from mattress all objects with sharp edges. Do not
fold gel mattress. The gel mattress may become damaged.
Precautions During Care
Warning: Always follow accepted hospital procedures for handling
equipment contaminated with body fluids.
Warning: Always disconnect AC line cord before cleaning and
disinfecting.
Caution: Do not launder mattress in an automatic washing machine.
Do not autoclave mattress.
Precautions During Maintenance
Warning: To minimize the risk of infection, clean and disinfect bassinet
and accessories before any maintenance according to
established hospital procedures – this applies also when
returning units or parts for repair.
: Use of the Dräger RH600 warmer is
Warning: Maintenance work on the Babytherm 8000 LDR may be
performed by trained and factory authorized staff only.
Warning: Never operate the heated bassinet if it has suffered physical
damage or does not seem to operate properly. In this case,
always refer servicing to properly trained and factory
authorized service personnel.
7
Page 14
Introduction
Intended Use
Babytherm 8000 LDR is a heated bassinet intended as a therapeutic device
providing warmth, with or without the use of a canopy, for premature and
full-term neonates and infants up to 16 lbs.
8
Page 15
Mattress Heating
The Babytherm 8000 LDR features a mattress heating system that uses
conductive heat to warm up infants and to keep them warm.
The system consists of a gel mattress, a heating plate and an electronic
control and monitoring unit. The gel is enclosed by a soft, skin compatible
film.
The mattress molds itself to the body contours of the infant lying on it,
resulting in a large contact surface, a property that helps promote thermal
heat transfer to the infant and minimizes pressure points. A cold gel
mattress, on the other hand, dissipates infant thermal energy. The gel
material has an excellent capacity for storing heat. When the power supply
is disconnected (e. g. during transport within the hospital), the infant is
kept warm for approximately 15 minutes if sufficiently insulated (blanket,
cover).
Description
Description
Features
The heating plate consists of an aluminum plate with a heater element
attached to its under side. It distributes heat evenly over the entire bed
area.
The control and monitoring unit is located under the heating plate and faces
the side most frequently accessed by caregivers.
The micro-processor controlled system continuously monitors the status of
the electronic temperature control and alarm system, and alerts the user if
any faults are detected.
With its ergonomic design including:
•foldable side panels for increased access to the infant
•tilting bed for Trendelenberg and reverse Trendelenberg position
•pop-up side table – integrated hand rail
•cable management system
•storage space in the cabinet
•5 inch diameter castors, two with locks
•optional canopy
it serves an important function for newborn babies from the first minute
after birth until discharge to home.
9
Page 16
Description
10
Page 17
Storage of Supplies
Always observe the stated weight limits when placing items on the shelf.
Installing Side Panels
The bassinet is surrounded by two side panels and two end panels, which
can be opened independently of each other. End panels have cutouts with
U-grommets for routing circuits and infant cables.
Preparation
Preparation
SIDE
PANEL
HINGE
DI0001
Figure 1. Installation of Side Panel
1. Insert lower pins on panel hinges into the left- and right-hand slots in
bassinet base.
2. Position panel semi-upright.
3. Press panel down firmly, until pins lock into position at the bottom of
slots.
•Tilt panel fully upright and let it drop into locking position.
4. Install all four panels in the same fashion.
5. Verify the panels are seated correctly.
11
Page 18
Preparation
Opening Side
Panels
To open Babytherm side panels:
1. Lift panel all the way up, until upper pins on hinge brackets are out of
their guide slots.
2. Fold down.
Installing Canopy (
OPTIONAL
Place canopy over side and end panels so that the flanges on the canopy
interlock with the end panel cutouts.
Selecting a Mattress
1. Select mattress type:
•Operation with mattress heating:
•Operation without mattress heating:
2. Place the mattress in bassinet and cover it with a thin sheet.
Warning: Always use a proper fitting thin sheet (Part No. 2M21272)
)
Gel mattress (Part No. 2M20827) [
Foam rubber mattress (Part No. 2M21012) [
[
PROVIDED
]. Do not fold the sheet under the gel mattress
more than 3". To prevent infant fall in Trendelenberg or
reverse Trendelenberg position always use proper fitting
mattress sheet.
PROVIDED
].
OPTIONAL
].
Tilting the Bed
Warning: Do not obstruct conductive heat transfer by placing
additional blankets or other insulating covers placed
underneath the infant.
Warning: Use of gel mattress when mattress heater is switched off will
result in infant becoming hypothermic. Use foam rubber
mattress when mattress heater is not in use for extended
periods (> 15 min.).
The bed can be locked in three positions: head up, level, or feet up.
1. Pull handle of locking mechanism up.
2. Tilt the bed as desired.
Tilt angles:
•Trendelenberg position: 10°
•Reverse Trendelenberg position: 10°
3. Release locking handle, the bed locks into position.
12
Page 19
Tests of Readiness for Operation
1. Verify that supplied voltage agrees with the values indicated on the
rating plate next to the power cable.
2. Inspect the power plug and power cord for signs of wear.
3. Inspect gel mattress and infant area for signs of damage.
Preparation
Preuse
Checkout
Procedure
Prior to each use:
1. Disinfect the Babytherm per institutional procedures.
2. Lock the side and end panels securely into place.
3. Verify that all panels are free of cracks and sharp edges.
4. Verify that the correct mattress is in the bassinet.
•Operation with mattress heater
•Operation without mattress heater
⇒
gel mattress,
⇒
foam rubber mattress.
5. Check the bed tilt positions and lock securely into required angle.
6. Verify that all patient lines and circuits are securely and properly
installed.
Caution: Never route infant circuits or lines over top of panels – they
may get pinched when folding up panels or when installing
cover.
7. Verify that the storage doors and the pop-up table are working properly.
8. Verify that the two locking casters function properly.
9. Lay gel mattress flat on the heating plate.
10. Connect plug to properly grounded, hospital grade power outlet.
11. Press button.
Check
a. all LEDs illuminate for approximately 2 seconds
b. numeric displays show
88.8
c.an audible alarm is generated
d. displays and LEDs extinguish
e. the audible alarm is silent
f.after 2 seconds, actual temperature is displayed.
Note: This function check is possible during operation without affecting
any of the settings.
12. Contact your DrägerService authorized service/dealer in case of failure.
13
Page 20
Preparation
General Precautions During Infant Care
Warning: To prevent infant fall, do not leave infant unattended when
any panels are down. All panels must be erect and secure
during transport to prevent infant fall.
•Check that all 4 panels are closed and locked securely into
place.
Warning: To allow for adequate air flow while canopy is in use, do not
obstruct openings around side panels of the bassinet.
14
Page 21
y
The heating system provided with Babytherm 8000 LDR is a conductive
mattress heater. Please refer to “Operation with Concurrent Use of a
Radiant Warmer” on page 31 for information about use of Babytherm 8000
LDR in the presence of radiant warmers.
Warnings and Cautions
Heat Therapy
Heat Therap
Heater Safety
Precautions
Mattress
Repair
Warning: Temperature Monitoring
skin condition at regular intervals and adjust temperature
setting to individual infant needs. Infant temperature must
be monitored particularly carefully during phototherapy or
while a radiant warmer is used in conjunction with the
Babytherm heated bassinet.
Should the gel mattress become punctured or ruptured, use the Dräger
mattress repair kit (Part No. 2M21324) to close the opening. In an
emergency clear adhesive tape may be used to temporarily fix the mattress.
Caution: Remove from mattress all objects with sharp edges. Do not
fold gel mattress. The gel mattress may become damaged.
Warning: Mattress Heater
mattress must be placed directly on the heated tray surface.
Always take into account warm-up time of the mattress
(approximately 1 hour) before placing an infant in the
bassinet.
Warning: To minimize infant overheating, always consider effect of
additional heat sources, such as phototherapy lamps or
radiant warmers. They increase the heat directed to the
infant. Avoid direct sunlight.
: When mattress heating is used, the gel
: Check the infant’s temperature and
Warning: This device is to be used only in rooms with line power
installations complying with national safety standards for
hospital infant rooms. To maintain grounding integrity,
connect only to a "hospital grade" receptacle. Always
disconnect supply before servicing.
15
Page 22
Heat Therapy
Mattress Heater Control Panel
678910
°C
>37 °C
<35 °C
°C
12345
Legend
(1)
Button for extending range of settings to
temperatures from 30° C to 38.5° C
(2)
Display of mattress set temperature
• to the left of display: button for
decreasing set temperature
• to the right of display: button for
increasing set temperature
(3)
Alarm silence button for silencing
intermittent audible alarm for 15 minutes;
• Yellow LED in button illuminates while
intermittent audible alarm is silenced
40 °C
±1 °C
Inop
Babytherm 8000
Check
(6)
Display of actual (measured)
mattress temperature
(7)
Red LED ; illuminates in case
of power failure
(8)
Red LED 40° C; flashes when
measured mattress temperature
exceeds 40° C
!
DI0002
16
(4)
Button for checking function of LEDs,
(9)
displays and audible alarm
(5)
Power switch
(10)
Figure 2. Babytherm Mattress Heater Control Panel
Red LED ±1° C; flashes when
measured mattress temperature
deviates more than ±1° C from
set value
Red LED Inop; illuminates in
case of unit malfunction
Page 23
Mattress
Heater
Alarms
Heat Therapy
±1° C alarm•Red LED ±1° C (9) flashes
•actual value display flashes
•intermittent audible alarm.
40° C alarm•Red LED 40° C (8) flashes
•actual value display flashes
•continuous audible alarm.
Power failure•Red LED ; (7) illuminates
•continuous audible alarm.
Inoperable
device
•Red LED Inop (10) illuminates
•Err appears on actual value display alternately with
measured value
•Error code is flashing in set value display,
continuous audible alarm.
Sensor alarm•3 flashing dashes >>– – –<< in the middle of the
actual value display (6)
Capacitor
discharge
alarm
•Acc appears on actual value
•display (6) alternately with measured value.
•Intermittent audible alarm.
Note:Due to periods of non-use, the capacitor powering the power failure
alarm may become discharged. By turning the unit on, the backup
capacitor will be charged automatically and the alarm will silence as
soon as the capacitor is charged. Press alarm silence button to
silence the Acc alarm until the capacitor is charged.
17
Page 24
Heat Therapy
Control
Systems
Self-Test
°C
40 °C
±1 °C
Inop
Babytherm 8000
>37 °C
<35 °C
DI0003
Figure 3. Babytherm Power On and Self-Test
Check
Switching heater unit on activates the Babytherm systems self-test:
1. Press power switch until it engages = ON.
a. A microprocessor self-test now checks proper function of the unit
including all temperature sensors, power relays, AC and DC supply
voltages and microprocessor watchdogs.
•all LEDs
•the numeric display shows
illuminate
88.8
•an audible alarm is generated.
b. After approximately 2 seconds, the unit shows the default set
temperature of 37.0° C (display blinking).
!
POWER
SWITCH
18
40 °C
±1 °C
°C
>37 °C
<35 °C
Figure 4. Setting Babytherm Temperature Control
Inop
Babytherm 8000
Check
c.The display alternates between SEt = "Set value" and measured
temperature.
2. Press down [
R]
or up [P] button to accept the default value as set value.
Display now shows the measured mattress temperature continuously.
!
Page 25
Power Failure
ging
Alarm Backup
Rechar
Heat Therapy
40 °C
±1 °C
°C
Inop
Babytherm 8000
Failure of
Self-Test
>37 °C
<35 °C
Check
!
DI0005
Figure 5. Insufficient Charge for Power Failure Alarm
The display may alternate between
measured temperature with a simultaneous intermittent audible alarm.
This means that the power failure alarm backup capacitor is not sufficiently
charged.
1. Continue operation. After approximately 30 minutes, the backup
capacitor will be fully charged. The display will then show the measured
temperature continuously without an audible alarm.
To silence the audible alarm:
2. Press alarm silence button , its yellow LED will illuminate.
An intermittent audible alarm can be silenced for 15 minutes.
°C
= "rechargeable backup" and
Acc
40 °C
±1 °C
Inop
Babytherm 8000
>37 °C
<35 °C
Check
!
Figure 6. Control Panel Shows Failure of Self-Test
The display may alternate between
Inop LED illuminates. The unit is inoperative: see “Troubleshooting” on
page 36 or consult qualified service personnel.
and measured temperature, and red
Err
19
Page 26
Heat Therapy
Control
Systems
Operator Test
Control Panel
Checkout
40 °C
±1 °C
°C
Inop
Babytherm 8000
>37 °C
<35 °C
Figure 7. Operator Checkout of Control Systems
Check
In order to check LEDs, displays and audible alarm:
1. Press button.
Check
a. all LEDs illuminate for approximately 2 seconds
b. numeric displays show 88.8
c.an audible alarm is generated
d. displays and LEDs extinguish
e. the audible alarm is silent
f.after 2 seconds, actual temperature is displayed.
Note: This function check is possible during operation without affecting
any of the settings.
2. Perform control systems check at least once daily.
!
20
Page 27
Power Failure
Alarm
Checkout
°C
40 °C
±1 °C
Inop
Heat Therapy
Babytherm 8000
Figure 8. Operator Checkout of Power Failure Alarm
To check power failure alarm:
1. Remove power plug from outlet.
2. Reconnect power plug.
Heater Operation
1. Set temperature to 38° C during warm-up. Allow approximately 1 hour
>37 °C
<35 °C
Check
!
DI0008
Red LED illuminates and continuous audible alarm is generated.
Red LED extinguishes. Audible alarm is silenced. Unit continues to
operate with temperature values set prior to power failure.
for thorough warm-up. During warm-up and in standby mode, cover the
gel mattress with a thick blanket to minimize heat loss to the
environment.
2. Remove the thick blanket and place the infant on the gel mattress (now
only covered with a thin sheet).
3. Set mattress temperature according to infant’s needs and ambient
conditions.
4. Monitor baby’s temperature and adjust mattress temperature as
needed. Allow time for the heated gel mattress to react to any changes to
the temperature set point.
Caution: Always use a thin sheet to cover the mattress.
Warning: Do not obstruct conductive heat transfer by placing
additional blankets or other insulating covers underneath the
infant.
21
Page 28
Heat Therapy
Operation without Concurrent Use of a Radiant Warmer
Figure 9. Babytherm Operation without Radiant Warmer or Canopy
When no radiant warmer is used:
DI0009
1. Cover infant with a blanket.
Warning: Use of gel mattress when mattress heater is switched off will
result in infant becoming hypothermic. Use foam rubber
mattress when mattress heater is not in use for extended
periods (> 15 min.).
22
Page 29
Oxygen Therapy with Canopy
Heat Therapy
Operator
Precautions
An oxygen analyzer should be used when oxygen is administered. The
following table illustrates approximate oxygen concentrations with the
indicated flows.
FlowApproximate
Concentration
5 L/min O
2
30% O
10 L/min O240% O
15 L/min O250% O
Adjusting Mattress Temperature
°C
>37 °C
<35 °C
2
2
2
40 °C
±1 °C
Inop
Check
Babytherm 8000
!
DI0010
Figure 10. Adjusting Temperature Within Normal Range
Temperature can be varied in steps of 0.1 degree Celsius.
Setting the temperature in the normal range (35° C to 37° C):
1. Press and hold the down [R]or up [P]button until desired temperature
is displayed.
>37 °C
<35 °C
40 °C
±1 °C
°C
Inop
Check
Babytherm 8000
!
DI0011
Figure 11. Adjusting Temperature in Extended Range
23
Page 30
Heat Therapy
Setting temperature in extended range:
•extended lower range: 30° C to 35° C,
•extended upper range: 37° C to 38.5° C.
2. Press the extended range button.
The yellow LED in the button is illuminated.
3. Within 1 minute, input a temperature above 37° C or below 35° C.
•Hold the down [R]or up [P]button until desired temperature is
showing on the display.
If there is a delay between pressing the extended range button and
entering the temperature value, the LED in the extended range button
will extinguish and the extended temperature ranges will be locked
again.
Note: A change of more than 1° C in the settings will trip the
temperature deviation alarm. Press the alarm silence button
to silence the audible alarm. The display will flash until the
temperature is within 1° C of the set temperature.
°C
°C
Figure 12. Control Panel Showing Temperature in Extended Range
When the temperature is outside a range from 5° C to 45° C:
•three dashes in the bottom of the display =
the temperature is below 5° C
•three dashes in the top of the display =
the temperature is above 45° C.
Note: This may occur, e. g., when the Babytherm 8000 LDR is stored in
ambient conditions beyond the range of 5° to 45° C.
DI0012
24
Page 31
Heater
Alarms
Heat Therapy
40 °C
±1 °C
°C
Inop
Babytherm 8000
>37 °C
<35 °C
Check
!
DI0013
Figure 13. Control Panel Showing Set Temperature Alarm Condition
Deviation from set temperature:
In case of deviations greater than ±1° C between actual and set mattress
temperature:
•red LED ±1° C flashes
•an intermittent audible alarm is generated
•the display flashes.
The intermittent audible alarm can be silenced for approximately
15 minutes:
1. Press the alarm silence button , its yellow LED will illuminate.
•The red LED ±1° C remains illuminated.
When the actual mattress temperature has returned to a value within ±1° C
of the set temperature:
•All visual and audible alarms are reset.
After the mattress heater is initially switched on:
•audible alarm is silenced for 1 hour (warm-up)
•red LED ±1° C is illuminated
•yellow LED in the
button is illuminated.
25
Page 32
Heat Therapy
Power-Down
Procedure
40 °C
±1 °C
°C
>37 °C
<35 °C
DI0014
Figure 14. Control Panel Showing High Temperature Alarm Condition
Inop
Check
Babytherm 8000
40° C alarm:
In case of high temperature:
•red LED 40° C flashes
•a continuous audible alarm sounds.
•the display flashes.
Note: The continuous alarm cannot be manually silenced. The alarm
resets automatically when the temperature decreases below
.
40° C
This procedure applies only for equipment processing between uses.
1. Press the power switch on the front panel.
!
26
All displays will be extinguished.
2. Disconnect power plug from outlet.
3. Store the power cord using the cable holder on the unit.
Page 33
Using the Canopy
Extreme care must me taken when using the canopy accessory (Part No.
2M21030) together with an infant radiant warmer. The radiant energy from
the overhead radiant warmer will be absorbed by the canopy. The infant will
not receive direct radiant energy from the radiant warmer, but a much
lower indirect radiation from the warm canopy.
The use of the canopy may be desired with stable infants. When warming a
cool infant, the use of the canopy will result in a much longer heat-up time.
Heat Therapy
Warning: To prevent possible injury to operator or patient, do not use
radiant warmer while canopy is in use.
•The maximum measured CO2 concentration with the canopy in place
is 0.1%, when measured in accordance with IEC 601-2-19,
subclause 105.
•Always use a foam rubber mattress (Part No. 2M21012).
•Do not use a gel mattress without mattress heating, or when mattress
heating is switched off.
Warning: Temperature Monitoring
skin condition at regular intervals and adjust temperature
setting to individual infant needs. Infant temperature must
be monitored particularly carefully during phototherapy or
while a radiant warmer is used in conjunction with the
Babytherm 8000 LDR heated bassinet.
: Check the infant’s temperature and
27
Page 34
Heat Therapy
36.3°C
38.0°C
Figure 15. Babytherm Operation with Radiant Warmer and Mattress Heating
Using radiant warmers and mattress heating with a gel mattress:
To add warmth from below, in addition to a radiant warmer.
1. Set mattress heater temperature to, e. g., 38° C.
2. Set skin temperature at the radiant warmer to the desired skin
temperature of the infant.
Warning: Use of gel mattress when mattress heater is switched off will
result in infant becoming hypothermic. Use foam rubber
mattress when mattress heater is not in use for extended
periods (> 15 min.).
DI0015
28
Page 35
36.3°C
Heat Therapy
37.0°C
Figure 16. Babytherm Operation for Neutral Balance
To achieve a neutral balance:
1. Set mattress heater temperature to, e.g., 37° C.
2. Set skin temperature at the radiant warmer to the desired skin
temperature of the infant.
DI0016
29
Page 36
Heat Therapy
36.3°C
36.0°C
DI0017
Figure 17. Babytherm Operation to Cool Infant
To cool the infant from below:
1. Set mattress heater temperature to, e.g, 36° C.
2. Set skin temperature at the radiant warmer at the desired skin
temperature of the infant.
Note:Always consider the effect of the two independent heat
sources: The overhead radiant warmer and the heated gel
mattress.
Warming and cooling down an infant can be achieved with different settings
of both warmers. Always choose the appropriate settings with regards to the
infant’s needs and the ambient conditions.
An overhead radiant warmer will supply a large amount of energy quickly.
The radiant warmer energy can be controlled by the infant’s skin
temperature via a skin temperature sensor.
30
Page 37
The heated gel mattress provides additional heat (or cooling) depending on
the settings relative to the infant’s temperature. The heated gel mattress
will provide gentle, conductive heat at lower level with slower increase (or
decrease) compared to an infant radiant warmer.
The heated gel mattress will significantly reduce the time to heat up an
infant compared to an infant radiant warmer alone.
Operation with Concurrent Use of a Radiant Warmer
Heat Therapy
Warning: Use of Radiant Warmer
Dräger RH600 warmer. Dräger cannot warrant or endorse
the safe performance of third party radiant warmers for use
with the Babytherm 8000 LDR. Specifically, combinations
with radiant warmers that are not intended for use with
heated bassinets bear risks of excessive overall heat transfer
to the infant.
Warning: Temperature Monitoring
skin condition at regular intervals and adjust temperature
setting to individual infant needs. Infant temperature must
be monitored particularly carefully during phototherapy or
while a radiant warmer is used in conjunction with the
Babytherm 8000 LDR heated bassinet.
Warning: To minimize infant overheating, always consider effect of
additional heat sources, such as phototherapy lamps or
radiant warmers. They increase the heat directed to the
infant. Avoid direct sunlight.
Warning: When used with an infant radiant warmer check its operators
manual for directions of the correct heat level and distance to
the infant.
Warning: Use of radiant warmers may increase insensible water loss.
Measures to maintain desired fluid levels should be
considered.
: Dräger recommends the use of the
: Check the infant’s temperature and
Warning: To prevent possible injury to operator or patient, do not use
radiant warmer while canopy is in use.
31
Page 38
Heat Therapy
32
Page 39
g
g
Care
Cleaning/Disinfecting/Sterilizin
Cleaning/Disinfecting/Sterilizin
The Babytherm 8000 LDR heated bassinet must be thoroughly cleaned and
disinfected after each infant. Perform any disinfection procedures according
to established hospital procedures as well as to the following additional
instruction.
Warning: Always follow accepted hospital procedures for handling
equipment contaminated with body fluids.
Before any cleaning procedures, the bassinet must first be partially
disassembled. For the cleaning and disinfecting of accessories, see their
respective Operating Instructions.
Disassembly
1. Switch off the Babytherm warming bed.
2. Disconnect power cord from AC receptacle.
Warning: Always disconnect AC line cord before cleaning and
disinfecting.
3. Remove any ancillary infant care equipment from unit.
4. Remove mattress from bassinet.
5. Store the mattress flat; do not fold or roll.
Caution: Remove all sharp objects from mattress. Do not fold gel
mattress. The gel mattress may become damaged.
Cleaning and Disinfection Procedures
To prevent any damage, we recommend using only detergents and
disinfectants compatible with the materials used in the bassinet and its
components and accessories. Use aldehydes, or quarternary ammonium
compounds according to manufacturer’s label and established hospital
disinfection procedures.
Always follow the instruction labels specifically with respect to prescribed
concentrations and the necessary exposure times.
Warning: To minimize any risk of infection, clean and disinfect bassinet
and accessories before any maintenance according to
established hospital procedures – this also applies when
returning units or parts for repair.
33
Page 40
Cleaning/Disinfecting/Sterilizin
g
Caution: Do not use alcohol on the side and end panels; it will damage
materials.
Caution: Do not launder mattress in an automatic washing machine.
Do not autoclave mattress.
1. Follow manufacturer’s instructions for use of the cleaning/disinfection
detergent for effective wipe-disinfection.
2. Wipe-disinfect the following surfaces:
•side and end panels, inside and outside
•resting surface
•acrylic canopy
•gel mattress (foam rubber mattress)
•wooden bassinet, including attachments.
3. Remove visible impurities with a cloth and detergent.
4. Wipe with a damp cloth.
5. Dry with a clean cloth.
Before Next Use
1. Reassemble components.
2. Operate unit for a few hours without an infant, in order to eliminate any
3. Check the wood surface for sharp nicks. Any sharp nicks on the surface
Before an infant is placed in the Babytherm 8000 LDR:
4. Set up accessories for therapy as needed.
5. Check readiness for operation. See “Tests of Readiness for Operation” on
6. Verify that mattress is sufficiently preheated.
disinfectant residues from prior disinfection (at mattress temperature
setting of 37° C).
of the wood bassinet need to be repaired by authorized service
personnel.
page 13.
34
Page 41
Inspection and Maintenance
Inspection and Maintenance
Precautions
Repair and
Replacement
of Parts
Preventive
Maintenance
Intervals
Warning: To minimize any risk of infection, clean and disinfect bassinet
and accessories before any maintenance according to
established hospital procedures – this applies also when
returning units or parts for repair.
Warning: Always disconnect power before servicing.
Warning: Never operate the heated bassinet if it has been damaged or
does not seem to operate properly. In this case always refer
servicing to properly trained and factory authorized service
personnel.
Gel mattress For damage to the gel mattress use the
Dräger repair kit (Part No. 2M21324).
GrommetsReplace broken or lost grommets.
The Accreditation Manual for hospitals requires each piece of equipment to
be tested prior to initial use and at least annually thereafter. To comply with
this standard, it is recommended that you participate in our Periodic
Manufacturer’s Service Program. This service can be performed by
authorized DrägerService technicians.
35
Page 42
Troubleshootin
g
g
g
g
g
g
g
g
ging
g
g
Troubleshootin
Mattress Heater
Problem/Message Likely Causes Remedy
Red LED ±1° C flashes, actual value
display flashes, intermittent audible alarm
Red LED 40° C flashes, actual value
display flashes, continuous audible alarm.
Red LED illuminated, continuous
audible alarm.
Red LED Inop illuminates ,
actual value display alternately with
measured value.
Error code is flashin
continuous audible alarm.
3 flashin
of the actual value display.
Acc
alternately with measured value.
Intermittent audible alarm. See “Power
Failure Alarm Backup Rechar
pa
dashes >> - - - << in the middle
appears on actual value display
e 19.
Err
appears on
in set value display,
” on
Measured mattress temperature
deviates from set t emperature by
more than ±1° C.
Mattress temperature >40° C Durin
No electrical powerCheck power connection .
Device malfunction
Failed temperature sensors
Backup capacitor for power
failure al arm is not char
sufficiently.
ed
operation with radiant
Durin
warmer: reduce radiant warmer
output.
operation with radiant
warmer: reduce radiant warmer
output.
Call DrägerServiceSM.
Call Drä
Call Drä
Operate device for 30 minutes,
Press button [Alarm Silence],
audible alarm will be silenced for
15 minutes.
erServiceSM.
erServiceSM.
36
Page 43
Operator Precautions
g
g
g
g
g
g
g
g
g
g
g
g
g
g
g
OPERATOR PRECAUTIONSPRÉCAUTIONS À PRENDRE
Precautions
Precautions
Warning Fire Hazard
1.Do not switch off heater while infant is
lying on the gel mattress as this may
lead to accidental hypothermia.
2.Use foam mattress when heater is to
remain switched off.
3.When used with a radiant warmer,
check its operator’s manual for
directions on correct heat level and
distance to infant.
4.Check infant's temperature and skin
condition at re
temperature settin
5.An oxy
when oxy
Flowapproximate
5 L/min O
10 L/min O2 40% O
15 L/min O
WARNING - FIRE HAZARD!ATTENTION - RISQUE D’INCENDIE!
30% O
2
2
ular intervals and adjust
to infant's needs.
en analyzer should be used
en is administered.
concentration
2
2
50% O
2
1.Risque d'hypothermie. Ne pas arrêter le
chauffa
matelas-
2.Utiliser un matelas ca outchouc-mousse
lors d'un arrêt de longue durée du
chauffa
3.Lors de l'utilisation de chauff a
radiant, utiliser la notice d'utilisation
pour le reglage approprié de
température et dista nce e n fonct ion des
besoins du patient.
4.Surveillier ré
et l'état de la peau du nourisson et
régler le chauffage en fonction des
besoins de patient.
5.Lors de l'administration en oxy
utiliser un analyseur d'oxy
Débit recommandépour concentration
5 L/min O
10 L/min O
15 L/min O
e lors de l'utilisation d'un
el.
e.
ulièrement la températu re
approximative
2
2
2
30% O
40% O
50% O
e
ène.
2
2
2
ène,
Keep matches, li
other sources of i
Babytherm 8000 LDR is located.
hted cigarettes and all
nition out of room where
Garder toute source d'alluma
(ex: allumettes, ci
flammes) hors de la chambre des lits
chauffés stationnés.
arettes, étincelles et
e
37
Page 44
Caution
g
g
g
g
g
g
g
g
g
Precautions
CAUTION!ATTENTION!
Danger
Warning
To maintain
to a properly wired "hospital
receptacle.
To reduce risk of electric shock, do not
remove controller cover.
Federal (USA) law restricts this device to
sale by or on the order of a physician.
DANGER!DANGER!
Risk of explosion if used in the presence of
flammable anesthetics.
WARNING!ATTENTION!
1.To allow for adequate air flow while
canopy is in use, do not obstruct
openin
bassinet.
2.T o pre vent poss ible in jury to op erator or
patient, do not use radiant warmer while
canopy is in use.
3.Disconnect power before servicin
4.To reduce risk of electric shock, do not
remove cover.
5.Repairs on this equipment to be
performed only by Drä
its authorized servic e centers.
rounding integrity , connect onl y
rade"
s around side panels of the
.
erService
SM
or
Pour assurer une mise à la terre approprié
n'utiliser que les prises murales conforme
aux exi
Pour éviter tous risques d'électrocution ne
pas enelever le ou les panneaux de
protections.
Risque d'explosion à proximité d'a
anesthésiques inflammable.
1.Ne pas obstruer les trous d’aération
2.Pour éviter tous risques d’incident de
3.Débrancher l'appareil avant toute
4.Pour éviter tous risques d’électrocution
5.Les réparations et révisions de
ences de l'hôpital.
lorsque le couvercle est placé.
l’utilisateur ou du patient, ne pas mettre
le chauffa
le couvercle est placé.
intervention.
ne pas enlever le ou les panneaux de
protections.
l'appareil doivent être confiées
uniquement au Drä
personnel autorisé par le
DrägerService.
ents
e radiant en marche lorsque
erServiceSM ou
38
Page 45
Ambient conditions
Technical Data
Technical Data
Operation
Storage
Electrical
Temperature . . . . . . . . . . . . . . . . . . . . 20° C to 30° C (68° F to 86° F)