Druck PC6-PRO Operating Manual

PRESSURE
PC6
CALIBRATOR
operating
MENU
ENTER
manual
PRINT
LOG
ON/OFF
PC6
SI Pressure Instruments Ltd. Garretts Green Lane, Birmingham, B33 OYA. UK. Tel: +44 (0)121 784 6855 Fax: +44 (0)121 784 4795. sales@si-pressure.com
Part No: 31-0432 Issue 3. DIM.5221
www.si-pressure.com
CONTENTS
Welcome ( Indicator Note ) Safety Warning Quick Reference
Exit, Menu selection, Data Entry, Tare) ..................
Manufacturer's Specifications
Description 1
1.0
Introduction ...................................... 2
1.1
General ............................................. 2
1.2
Keypad ............................................. 2
1.3
LCD .................................................. 3
1.4
Connection Panel ............................. 3
1.4.1
1.4.2
1.4.3
1.4.4
1.4.5
PC6-PRO Arrangement Functions (menu 1) Functions (menu 2) Functions (menu 3)
Operating Instructions 9
1.5
Introduction .........................................10
1.6 Connections:
1.6.1
Pressure Input .................................... 10
1.6.2
EPM ................................................... 10
1.6.3
Voltage/Current Input ....................... 10
1.6.4
Switch Test.......................................... 11
1.6.5
RS232 ................................................ 11
1.6.6
Charger ............................................. 11
1.6.7
Voltage Output .................................. 11
1.7 Messages & Parameters:
....................................... ii
(Keys, Activating functions,
2 mm Sockets ............................. 3
Quick connect coupling ............. 3
RS232 ......................................... 3
Charger ....................................... 3
EPM ............................................. 3
....................................
.................................... 6
.................................... 8
................. i
iii
............... iv
............................... 4
5
1.7.1
Retained parameters on switch off ...... 12
1.7.2
Battery low state ................................ 12
1.7.3
Tare Value .......................................... 12
1.7.4
Maximum / Minimum ......................... 12
1.7.5
Void leak test ..................................... 12
1.7.6
Switch Off .......................................... 12
1.7.7
Auto Switch Off .................................. 12
1.7.8
Pressure Overload ............................ 12
1.7.9
Low Battery Detect ............................ 12
1.7.10
Pressure Switch Sate Change .......... 12
1.7.11
Printer Busy ........................................ 12
1.7.12
Outside User Set Alarm Value ........... 12
1.7.13
Display Hold ...................................... 13
1.7.14
Connection Fault ............................... 13
1.7.15
Short Detected .................................. 13
1.7.16
Voltage / Current Overscale .............. 13
1.7.17
RS232 Output Format ....................... 13
1.7.18
Alarm Flag ......................................... 13
1.7.19
Symbols ............................................. 13
How to Select Functions (Menu 1) 14
1.8
Units ................................................... 15
1.9
Max,Min,Switch,Temp.,
1.9.1
Tare .................................................... 15
1.9.2
V / mA ................................................. 15
2.0
Leak Test ............................................ 17
2.1
File (logged files) ................................ 18
Set-menu (menu 2) 20
2.2
Alarm Settings - High / Low ............... 21
2.3
V / mA ................................................ 22
2.4
V-Out, Display light, Port ................... 26
2.5
Resolution ........................................... 26
2.6
Event ...................................................26
2.7
Log ..................................................... 27
2.8
RS232 ................................................ 29
2.8.1
DPM .................................................... 30
Date&Time,% 15
2.9
Date & Time ....................................... 30
3.0
Language ............................................ 31
3.1
Zero .................................................... 31
3.2
Power_Off ........................................... 31
Cal-menu (menu 3) 32
3.2
Zero .................................................... 32
3.3
Span ................................................... 33
3.4
History (overload, zero/span) ............. 33
3.5
Language ....................... .................... 34
Flowcharts 35
Menu1,Menu2&Menu3 36 Menu 1
Menu 2
Menu 3
Operating Instructions for Hydraulic ............. 55
Test Pump HTP1
Operating Instructions for Pneumatic ............ 57
Test Pump TP1 Operating Instruction for Low Pressure Test …60 Pump LTP1
Unit, Leak ....................................... 37
V / mA ............................................ 38
File ................................................. 39
High & Low .................................... 40
V / mA (measure) ........................... 41
V-out, Display light ....... 47
Resolution, Event, RS232 .............. 47
Log ................................................. 48
Date & Time, Zero, EPM, Language 49
Zero ................................................ 50
Span ............................................... 51
History ............................................ 52
Language ........................................ 53
Pressure Generation 54
Index 63
..........................
WELCOME
This detailed operating manual will help you to become familiar
with the many features of the PC6-PRO Calibrator. The simple step by step
instructions will quickly guide you through the procedures for using the
calibrator and other items needed for accurately testing and calibrating
numerous types of instruments. Please t
the whole manual b
efore you begin to use the PC6-PRO.
Two versions of the PC6-PRO are available:-
PC6-PRO Calibrator:
All of the functions shown in this manual apply to this unit. The PC6-PRO Calibrator is supplied with a LTP1 Low pressure pneumatic pump for pressures up to 3 bar, TP1 40 bar or HTP1 hydraulic pump for pressures up to 1000 bar.
PC6-PRO Indicator:
This is a simplified version of the PC6-PRO Calibrator and does not include the Voltage/Current input/output, functions.
or EPM
ake time to carefully read
pneumatic pump for pressures up to
© SI Pressure Instruments Ltd.
iPage
SAFETY WARNING!
HIGH PRESSURE
Uncontrolled release of high pressure is hazardous to personnel and may cause damage to equipment. Before connection of any pressure component to the PC6­PRO ensure that the component(s) is/are isolated from the pressure supply and any internal pressure is released slowly.
:
RECHARGEABLE Ni-MH BATTERIES
Rechargeable Ni-MH batteries must be recycled or disposed of properly. May explode if damaged or disposed of in fire. DO NOT short-circuit. CAUTION: Use charger supplied by SI Pressure Instruments only.
Ni-MH
iiPage
QUICK REFERENCE
Keys
ON/OFF
MENU
Activating Functions
Select menu position cursor press the Enter key
Exit
To exit out of any menu select 'Exit' option, or press key.
Press to turn PC6-PRO on and off
Press to access the menus
Press to accept functions or settings
Press to move cursor up/left or increase values
Press to move cursor down/right or decrease values
Press to print/log/start event depending on previous setting
MENU
under function
MENU
Menu Selection
Pressure Display
1st Press for Menu 1
2nd Press for Menu 2
3rd Press back to Pressure Display
MENU
MENU
MENU
Data Entry
Press or to position cursor and press
to select required alphanumeric. Press
key to release cursor.
For next data entry, repeat the above procedures then press once all data is entered
Note: To quickly scroll through alphanumeric lists hold down the or key.
Tare
the pressure display mode. The unit will bleep after two seconds when the Max. and Min. values reset. After four seconds another bleep will sound as the displayed pressure now resets to zero. This is a temporary
function and will be cancelled when the unit is switched off.
MENU
Hold down the key for 4 seconds while in
iiiPage
MANUFACTURER'S SPECIFICATIONS
Model
PC6-PRO-0.1-C G PC6-PRO-1-C G or A PC6-PRO-2-C PC6-PRO-5-C G or A PC6-PRO-10-C PC6-PRO-20-C PC6-PRO-25-C G or A PC6-PRO-40-C G or A PC6-PRO-70-C SG PC6-PRO-100-C SG PC6-PRO-200-C SG PC6-PRO-350-C SG PC6-PRO-700-C SG PC6-PRO-1000-C SG
GorA
GorA GorA
G = Vented Gauge, A = Absolute, SG = Sealed Gauge ( For Indicators, substitute ‘C’ with an ‘I’ ). Alternative pressure ranges are available to special order
Overload Pressure Pressure Display Resolution EPM
Measurement
Voltage*
(Auto Range )
Current*
Sourcing
Voltage* Current*
Range
0 to 100mV
0to5V
0to50V
0to50mA
Range
0to10V
0to50mA
Pressure Range
bar psi
-0.1 to 0.1 1.5
-1 to 1 15
-1 to 2 30
-1 to 5 75
-1 to 10 150
-1 to 20 300
-1 to 25 360
-1 to 40 600 0 to 70 1000 0 to 100 1500 0 to 200 3,000 0 to 350 5,000 0 to 700 10,000 0 to 1000 15,000
10% FS 6 Digits External Pressure Modules are available in
all of the above ranges. Differential units are available as EPM’s. ( See Appendix for EPM details ).
Resolution
1uV
0.1mV 1mV
1uA
Resolution
0.2mV
1uA
Accuracy
0.025%FS
0.025%FS
0.025%FS
0.025%FS Accuracy
0.025%FS
0.025%FS
Accuracy (FS)
up to
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
0.025%
Input resistance
1MW 1M
W
1M
W
3
W
Max. Load
20mA
330 Ohm
* Temperature coefficient ± 0.005% of reading / °C w.r.t. 20°C
Pressure Units
bar, mbar, Mpa, kPa, p.s.i, kg/cm², atm, inH 0,
mH 0, mmH 0, inHg, mmHg, unit1, unit2 (user
22
defined, factory set only)
Overload Warning Temperature Display
Flashing display and audible tone at 110% of FS Temperature of pressure sensor module
(Accuracy: ±0.5°C)
Pressure Switch Test Operating Temperature Calibration Temperature Calibrated Temp. range Storage Temperature Temperature Effects RS232 Parameters
Status OPEN/CLOSED 0 to 50°C 20°C ± 2°C 0 to 50°C
-20°C to 70°C
0.001% of reading/°C between 10-40°C baud rate 1200, 2400, 4800, 9600, stop bits 1 or
2, status ON/OFF
Zero Reset Data Capacity
Manual keyboard operation 437 records into a maximum of 20 files, 32k
memory additional 585 records, ( 32k memory option available )
Display Humidity Resolution Select Power Supply
Backlit LCD, 16 character x 2 line alpha numeric 5 to 95% Relative Humidity non condensing Increase or decrease by a factor of 10 6V NI-MH rechargeable battery pack (see warning
page ii)/ mains operation via charger (supplied).
Voltage Output
Regulated D.C supply 10V ± 0.1% max current
30mA
Unregulated D.C supply 24V ± 10% max current
50mA
Battery Life Recharge Time Low Battery Dimensions Weight Microprocessor Recommended Recalibration Period
Fully charged: up to 10 hours 14-16 hours for full charge Continuous check, audio and visual warning 92x110x59mm 850 grams (Calibrator head only) H8/2138
1 Year (Max)
2
ivPage
MANUFACTURER'S SPECIFICATIONS (continued)
Electrical Connections
Ground Standard 2mm socket Voltage/Current Input/output Standard 2mm sockets Pressure Switch Input Standard 2mm sockets Battery 2 pole miniature round connector RS232 8 pole miniature round connector EPM (P2) 14 pole miniature round connector
Pressure Media
Pressure Connection
Software (optional)
System requirements: 486 with maths coprocessor.
Water, Mineral Oil, Non-corrosive gases (for other media, contact SI). 3/8"BSP male ‘quick-fit’ / 1/4”BSP or 1/4”NPT male
(Pneumatic).
3/8"BSP female ‘quick-fit’ / 1/4”BSP or 1/4”NPT male
(Hydraulic).
SiCal
The ‘ ’ software will present downloaded data in
various graphical forms such as gauges or a scrolling bar & remotely control the PC6-PRO via the RS232 cable linked to a PC. The downloaded data can be stored in a variety of file types compatible with most analysis database and word processing programs.
Cal. procedures can be uploaded to the PC6-PRO.
Windows 3.1 or higher is required.
vPage
DESCRIPTION
1Page
1.0 Introduction
The Pressure Calibrator is a microprocessor controlled, precision instrument powered by an internal 6V
PC6-PRO rechargeable battery pack. It is portable and capable of accurate pressure measurement, voltage or current measurements and sourcing when calibrating transducers, pressure switches, etc. This section describes the equipment from an external viewpoint, allowing the user to become familiar with the various controls and connections provided.
1.1 General
The PC6-PRO is housed in a three part anodised aluminium case with a liquid crystal display unit at the front and a six button membrane keyboard. The main body contains the circuitry, connections, internal transducer and battery pack.
1.2 Keypad
The Keypad is a non-tactile membrane keyboard with six buttons. The keyboard layout is shown below.
An audible tone accompanies
Note:
every key operation.
MENU
PRINT
LOG
ENTER
ZERO
ON/OFF
2Page
1.3 Liquid Crystal Display (LCD)
The LCD is a 16 character x 2 line alphanumeric LED backlit display which is capable of displaying special characters.
1.4.2
either a 3/8"BSP (1/4”BSP / 1/4”NPT adaptors) male for low pressure systems (below 40 bar) or 3/8”BSP female (1/4”BSP / 1/4”NPT adaptors) for systems greater than 40 bar.
Pressure connections are made to
COUPLING
Pressure reading Unit of measurement
Special Symbols
Function Function Value
1.4 Connection Panel
The connection panel is located at the top of the case and provides input / output connections as detailed in Fig.2
1.4.1
the top of the case provide input or output current and voltage and switch status.
Voltage outputs of 10 and 24V are available for use with sensors, transducers, transmitters, etc.
Six standard 2mm electrical sockets mounted along
(See Page 4, Fig 2)
1.4.3
connector is used for communication with RS232 standard communication protocols/equipment. The user software ‘ ’ (optional) can be used to remotely
SiCal
monitor or download data from the PC6­PRO via this connector.
(Optional RS232 cable - Part No. 01-0625)
1.4.4
connector is used for the battery charger (provided).
1.4.5
connector is used to connect to the EPM (External Pressure Module). These optional modules cover pressure ranges from 0 to 1000 bar.
An 8 way circular miniature
The 2 way circular miniature
The 14 way circular miniature
RS232
CHARGER
EPM
3Page
ARRANGEMENT OF PC6-PRO
Front View: LCD & keypad
Top View: connection panel
Keypad
See Page 2
54-0158 ¼”BSP 54-0363 PC6-PRO- 40 bar or below
¼”NPT
Fig 1
Display.
MENU
PRINT
General Purpose Adaptors
LOG
ON/OFF
ENTER
ZERO
Quick Connect Coupling.
54-0101 3/8”BSP 54-0364
¼”NPT
PC6-PRO- Above 40 bar
BATTERY
CHARGER
+
-
EPM
CHARGER RS232
+10V +24V
(See Page 10
for V/mA
connections)
Fig 2
Battery Charger Input Socket
See Page 11
External Pressure Module Socket
See Page 10
RS232 Communication
See Page 11
Socket
EPM
GND
VmASWITCH
OUTPUTS
+24 V Volts output. +10 V Volts output. GND Common Ground Socket
mA Milliamp Input (controls input current in
V Volt Input (output in source mode). SWITCH Pressure Switch Input.
IN/OUT
source mode).
4Page
PRESSURE DISPLAY -
Menu 1
UNITS
bar mbar
Pa kPa
M MegaPascals kiloPascals
p.s.i kg/cm² atm nH O mH O mmH O inHg mmHg Unit1 (Factory Set Only) Unit2 (Factory Set Only) Exit
FUNCTIONS (MENU 1)
Allows the user to operate the PC6 for pressure measurements. (Display Functions)
Selection of pressure unit.
bar millibar
pounds force per square inch kilograms force per square centimetre atmospheres i inches of Water (at 4°C)
2 2
metres of water (at 4°C) millimetres of water (at 4°C) inches of Mercury (at 0°C) mm of Mercury (at 0°C) user defined unit second user defined unit Returns to menu 1
2
MAX MIN % TARE V/mA
Measure
Source
Exit SWITCH Leak
Start
Set-Time
Exit Tem p C Date & Time File
View
Export
Import
Delete
Exit Exit
Displays the maximum pressure measured since the last reset (50 samples per second). Displays the minimum pressure measured since the last reset (50 samples per second). Displays the pressure as a percentage between user pre-set high and low alarm values. Displays the pressure at the time of tare. Volts / milliamps functions
Actual Converted Exit
Actual Converted PtoV/PtomA Exit
Displays the following input measurements.
Displays actual input measurements. Displays converted input measurements. Returns to Pressure Display.
Displays the following outputs.
Displays actual output measurements. Displays converted output measurements. Displays Pressure to Volts / milliamps conversion. Returns to Pressure Display.
Returns to pressure display Indicates the status of the pressure switch under test (OPEN/CLOSED). Allows the PC6 to measure pressure changes over time.
Starts the leak test.
Sets the leak test time.
Returns to Pressure Display. Allows the user to monitor the temperature around the selected pressure transducer in pressure display mode. Allows the user to monitor the date & time of the on board real time clock in pressure display mode. Allows the user to view or manipulate a file previously stored in the filing system.
Allows the user to see logged file.
Outputs a logged file
Imports a file ( refer to ‘ ’ Instructions for exporting files to PC6 PRO)
Deletes a logged file.
Returns to Pressure Display. Returns to Pressure Display mode.
SiCal
(SHORTCUT: To abort any menu level, simply press the MENU key).
5Page
SETTING MENU Menu 2
High Low V/mA
Continued on
Page 7
- To set user specific functions. (Set Functions)
Allows the user to set a High Pressure Alarm Limit.
Measure
Zero FS Text Root Exit
Source
Range
Convert
PtoV/mA
Setup
Exit
(by using the increase/decrease facility above which an alarm sounds.) Allows the user to set a Low Pressure Alarm Limit. (by using the increase/decrease facility below which an alarm sounds.)
Voltage/ current input settings for Converted Measurement.
Voltage/ current output settings.
Zero FS Exit
Zero FS Text Exit
Zero FS Exit
Status Root Step Ramp
Status Rate Dwell_time Exit
Exit
Returns to Pressure Display.
Set Zero. Set Full scale value. User defined measurement text. Switch measurement root function ‘on’ or ‘off’. Reverts back to Pressure Display.
Sets the following values:-
Sets Zero value Sets Full Scale value Returns to Source sub-menu.
Sets the following converted values:-
Sets converted Zero values. Sets converted Full Scale values. User defined converted text. Returns to Source sub-menu.
Sets Pressure to V / mA conversion factors.
Set zero values. Set Full Scale values. Returns to Source sub-menu.
Sets the parameters for voltage/current sourcing.
Switches output on or off. Switch sourcing root function ‘on’ or ‘off’. Set step value (%FS) Settings for Ramp mode
Returns to Source sub-menu.
Switches the Power Supply outputs +10V & +24V ON or OFFVout
Switches ON or OFF the LED back light.Display Light
Switches ramp mode on or off. Sets the ramp rate (V or mA/Sec). Sets the time for which the output remains at the same level. Returns to Source sub-menu.
6Page
Resolution Event
Log
RS2332
Date/Time
Language Zero
Port Power_Off
Exit
Key Timed Exit
Log View Delete Status Export Import Exit
Status Baud-rate Stop-bits Exit
Set View Exit
Key Timed Exit
FUNCTIONS (MENU 2) Continued.
Selects low, normal or high resolution. Sets up the event status for RS232 output and for logging functions.
Sets key driven output or logging. Sets timed output defined by the user. Reverts back to Pressure Display.
Allows the user to setup, amend and view files for logging.
To select a log file and start logging. To look at a file. Deletes a file. To enable or disable logging. To export logged files. To import files from a PC. (see ‘SiCal’ software instructions for exporting files to the PC6) Reverts back to Pressure Display.
Settings for RS232 communications.
To enable RS232 communication. To select the baud rate. To select the number of stop bits. Reverts back to Pressure Display.
Sets up Date & Time parameters
Sets the date & time. To look at date & time.
Reverts back to Pressure Display. Allows user to change language. Will reset pressure reading to Zero. ( This will remove Zero offset permanently / NOT available for absolute units ). Selects current pressure module (P1, P2, DPM) Allows the user to set Power_Off time or to switch the function on or off.
Switches the auto-power function off.
Allows the user to set the auto-power off time in minutes.
Reverts back to Pressure Display. Returns to Pressure Display
Page 7
CAL MENU
Menu 3
ZERO
SPAN
HISTORY
Language EXIT
FUNCTIONS (MENU 3)
Allows user to set zero. ( for absolute units, see page 32 )
CONTINUE EXIT
Allows user to set span.
CONTINUE EXIT
Allows user to see overload and zero-span history.
OVERLOAD ZERO-SPAN PC6 S/No Displays the unit’s serial number.
EXIT
Allows user to download different languages. Returns to Pressure Display.
Starts zero calibration procedure. Returns to Cal Menu.
Starts span calibration procedure. Returns to Cal Menu.
Displays period of time over which overload occurred. Displays date last calibrated, last span set, range and overload value.
Displays the serial number of the currently selected pressure module.PM S/No Returns to Cal Menu.
8Page
OPERATING
INSTRUCTIONS
9Page
1.5 Introduction
The PC6-RPO enables the user to carry out checks on pressure transducers, transmitters, pressure switches, pressure gauges and process indicators quickly and accurately. It is intended for use as a portable , on-site calibrator, used in close proximity with the equipment under test and without the need to remove the unit from its parent system. This section describes the connections to be made when checking switches, transmitters, etc. It also lists menu options, selection procedures and purpose of options available.
1.6 Connections
1.6.1 Pressure Input The PC6-PRO measures pressure
directly via the connector found underneath the case (i.e. the connection between the PC6-PRO and a hand held pressure test system). Alternatively, connect the item under test to the PC6-PRO with the adaptor and seals provided.
1.6.2 EPM (External Pressure Module)
Different pressure ranges can be read by attaching an External Pressure Module to the 14-way socket.
(See Fig.3 opposite)
Warning: The PC6-PRO must be switched off prior to
connecting or disconnecting EPM .
1.6.3 Voltage / Current Input / Output Voltage and Current
readings can be taken by connecting the outputs from the item under test to the 'V' or 'mA' and 'GND' input sockets on the connection panel using the connection leads provided.
Warning: When measuring voltage or current, make sure 'GND' is connected (where applicable, see fig. 3a)
EPM
Fig.3
CHARGER RS232
+10V +24V
OUTPUTS
EPM
GND
VmASWITCH
IN/OUT
Fig.3a
+10V +24V
OUTPUTS
GND
IN/OUT
V mA SWITCH
Input
+V
Two Wi re
Transmitter
4to20mA
Three Wire Transmitter
0to5V
-V
Output
0V
10Page
1.6.4 Switch Test The switching pressure of a pressure switch can be monitored by connecting the contacts of the pressure switch between the INPUT 'GND' and 'SWITCH' sockets on the connection panel and pressurizing the pressure switch.
1.6.5 RS232
If the RS232 facility is to be used then connections to the 8­way socket must be made to the specifications given in the table below. Alternatively an RS232 communication cable can be purchased from SI suitable for most RS232 applications as shown below.
To RS232 Standard 9-way socket on PC6 'D'type female
Cable Part No: 01-0625
connector
RS232 Connections Table
PIN No.
1 2 3 4 5 6 7 8
Note:
INPUT/OUTPUT
Output Output
Input
Output
Input
Output
Common
Input Pin 6 and pin 8 should not be used.
Data Terminal Ready (DTR)
Transmit Data (TxD) Receive Data (RxD)
Request To Send (RTS)
Clear To Send (CTS)
Do not use
Do not use
SIGNAL
Ground
STATE
Held High
Active High
RS232 Output
Connect the PC6-RPO to a printer or computer using a suitable cable (not supplied) as detailed above and set the parameters as detailed in Para. 2.6 & 2.8. The output can now be obtained as ‘single shot' or ‘timed' as follows:
Single Shot
when RS232 status & Log status are OFF. The output is obtained each time the key is pressed (provided any previous output
- This output is always available in Display mode
PRINT
LOG
is complete).
Warning:
The PC6-PRO must be switched off prior to
connecting or disconnecting the RS232 cable.
Timed
status is ON & Log status is OFF. The output cycle is initiated by pressing the key. Pressing the key cancels the output.
- This output is available in Display mode when RS232
PRINT
LOG
MENU
Once initiated, the output is repeated at the selected time interval (see ‘Event' function, page 26 ) until the key is pressed.
1.6.6 DPM
(DIgital Pressure Module) An alternative to the EPM,
MENU
the DPM communicates via the RS232 socket and is available in
-1 to 1000bar/15,000psi pressure ranges with accuracies up to
0.025%
1.6.7 Battery Charger
The internal battery pack may be recharged (when indicated by the battery low message) by plugging the charger (supplied) into the 'CHARGER' socket.
CAUTION:
When recharging internal batteries, only the charging
unit supplied with the PC6-PRO should be used.
1.6.8 Voltage Output
A voltage of +10V or +24V may be sourced from the PC6-PRO by connecting to the 2mm sockets. A typical application for this voltage output would be to power a pressure transmitter. ( see Fig. 3. page 10 ) To switch on voltage output function:­a) Press the key twice to activate 'Set' menu. b) Select 'V-out' and press the key. c) Select 'ON' and press the key.
MENU
ENTER
ZERO
ENTER
ZERO
11Page
1.7 Messages & Parameters
The PC6-PRO registers pressure input at switch on after the initial start­up message. The pressure reading is displayed on the top line of the LCD (see Page 3). Pressure values are monitored, alarm values are compared, alarm warnings and other information messages displayed until the unit is switched off.
Memory Errors:
Any read / write errors to memory detected at switch on will be reported as: Memory Error ‘Error No.’
Error Numbers:-
- Onboard eeprom; - IPM; - 2nd Onboard eeprom.
089
Errors with Pressure Modules will be indicated by the message "No P1 (P2 or DPM) proceeded by "PM EEPROM ERROR" at power on.
When switched on, the unit registers the same state as when last
1.7.1
switched off. The display mode (Operating/Set), units of measurement, function and parameter settings, alarm limits and RS232 output control, EPM selection, logged parameters are all retained at switch off, with the exception of the following conditions:
A battery low state occurring prior to the switch off. When switched
1.7.2
on, the unit registers the latest retained set of parameters.
Any TARE value set during use is not retained and therefore
1.7.3
should be set, if required, each time the unit is switched on.
Maximum and Minimum values are not retained. Current values
1.7.4
reflect pressure monitored from the start of the pressure input.
If the unit is switched off during a LEAK test then the test will be
1.7.5
void. When switched on again the unit will start up in LEAK test mode.
1.7.6 Switch off
When the key is pressed the 'switch off' message is displayed for two seconds accompanied by an audible bleep (Bleep rate 1 Hz). Normal operation may be resumed by pressing any key during switch off.
ON/OFF
1.7.7 Auto Switch off
The PC6-PRO will automatically turn itself of
after a set period (See Page 31 for instructions)
1.7.8 Pressure Overload
(Bleep rate 4 Hz) ‘OVERLOAD' occurs when
input pressure exceeds overload pressure level.
1.7.9 Low Battery Detect
(Bleep rate 1 Hz). ‘Low Battery' occurs when internal battery pack drops below 5.5V. Message flashes for 4 seconds, after a further 30 seconds, low battery detection is re-enabled.
1.7.10 Pressure Switch State Change (Bleep rate 2 Hz) Occurs in pressure mode with the pressure switch state display being
selected. When the connected pressure switch changes state, the pressure value (upper line) is frozen and the pressure switch state (lower line) display is flashed at 1 Hz unless a log file is open. Normal display update resumes when the key is pressed.
1.7.11 Printer Busy
ENTER
ZERO
(Bleep rate 1 Hz)
If the receiving device is not ready for input (e.g. device not connected or handshake LOW), 10 seconds after starting an output cycle (single shot or timed output) the message 'Printer Busy' is flashed on the display, together with an audible warning. If the device does not accept input within the next 5 seconds then the output cycle is cancelled and output is stopped until the key is pressed again.
1.7.12 Outside User Set Alarm Values
PRINT
LOG
(Bleep rate 2 Hz)
Occurs when measured pressure is lower than the user set Low alarm value or greater than the user set High alarm value and the
message is flashed. Indication ceases when pressure is equal to
limits’
’outside
either alarm value or between the values.
1.7.13 Display Hold
Occurs in pressure mode. If the key is pressed once (less than 1
(Bleep rate 2 Hz)
ENTER
ZERO
second) the pressure value (upper line) is frozen and the lower line is flashed with the ‘Display Hold’ message at 1 Hz. Normal display update resumes when any top row key is pressed.
Note: The display contents can be printed out during display hold by pressing the key.
PRINT
LOG
12Page
1.7.14 Connection Fault:
In voltage / current source mode, if the PC6-PRO is unable to control the voltage/current output (e.g. because there is no loop back in the current mode), the message “Connection Fault” will flash on the lower line accompanied by a bleep at 1 Hz. The message will continue until the fault is rectified or the user exits from source mode.
1.7.15 Short Detected:
If the 24V output is ON and it is shorted to 0V, the message “*Short detected*” will be displayed on the lower line accompanied by a bleep at 1 Hz. This message will be cleared when any key is pressed. In this condition the 24V supply will be switched off for safety.
1.7.16 Voltage / Current Overscale:
In Voltage measure mode, if a voltage in excess of 55V is connected to the PC6-PRO the message “*OVERSCALE*” will be displayed on the lower line, until the voltage falls within the input limits. In Current measure mode, a current in excess of 55mA enters the mA terminal, the PC6-PRO will display “*OVERSCALE*” on the upper line and the lower line will scroll the message “Check connections: Press MENU to continue”. Pressing the key will remove the message
MENU
provided the current measured is within range. This message may also appear in the source mode under connection fault conditions. Source mode will change to measure mode for safety. If the current overscale message occurs when the PC6-PRO is in timed RS232 output mode, the unit will exit from the timed output mode. If there is an excess voltage at the mA input when the PC6 is placed into current measurement mode, the message “Connection Fault” will be scrolled on the lower line. Once the fault has been rectified, press the
MENU
key to return to measuring mode.
1.7.18 Alarm Flags
A single character is output to indicate the various
pressure alarm conditions as given below:
* overload pressure level exceeded
(pressure warnings, 1.7.8 & 1.7.12) > pressure exceeds user set high limit < pressure below user set low limit “ ” a space if no alarms apply. c connection fault s short detected
Note: If two (or more) conditions occur which require a message display, the highest priority condition in the above list will be displayed. The sequence of messages given above set the priority except 'Switch Off' which comes last.
1.7.19 Display Symbols*
Log Status ON
Pressure Port 1 selected
Log Status ON File Open
Pressure Port 2 (EPM) selected
Top line symbols
Timed Log / output
Nudge Mode (UP)
Keyed Log / output
Nudge Mode (DOWN)
Bottom line symbols
* If more than one symbol is applicable, they are displayed alternatively.
or
or
Rs232 Output Status ON
Source Step Mode
Source Mode Ramp
DPM Status ON
Measure Mode & P-V / P-mA
Root Function
1.7.17 RS232 Output Format
LF Time & date ( 1100) +LF
23:13:30 08 display top line (16 characters)+LF display lower line (16 characters)+LF alarm flag (1 character)+LF
The output format is as follows:
13Page
HOW TO SELECT
FUNCTIONS
(MENU 1)
14Page
1.8 Units
To change the units of pressure, proceed as follows:
• Press the key once, then press key to choose
MENU
the units sub-menu
• Select kPa by pressing the key and then press key.
The display will indicate the pressure in kPa and the pre-selected 2nd line function.
1.9 Max., Min., Switch, Temp C, Date & Time, %
To display any one of these functions, proceed as follows:
• Press the key once and then scroll to the required function using the or key and then press key to select it.
The display will show the pressure on the top line and the 2nd line will show the function that has now been selected.
MENU
ENTER
ZERO
ENTER
ZERO
°
ENTER
ZERO
To measure input voltage or current select ‘Measure’ and press the key.
ENTER
ZERO
Two options are now available; ‘Actual’ and ‘Convert’.
• ‘Actual’ is selected by pressing the key.
ENTER
ZERO
An example display below shows the Measure Mode symbol on the second line.
ENTER
• Select ‘Convert’ using the key and press the
ZERO
key. To setup the conversion parameters, i.e. zero, full scale or text; see 2.3 on page 22.
1.9.1 Tare
The Tare value can be displayed on the bottom line by selecting ‘Tare’ from the menu using the key and pressing the key.
To activate function: hold down the key for 4 seconds
:
Tare
ENTER
ZERO
ENTER
ZERO
while in the pressure display mode. The unit will bleep after two seconds when the Max. and Min. values reset. At four seconds another bleep will sound as the displayed pressure now resets to zero. This is a temporary function and will be cancelled when the unit is switched off.
1.9.2 V, mA (procedures are the same unless otherwise stated)
Two main options are available for ‘V’ and ‘mA’:-
• Select ‘Measure’ for V and mA input.
• Select ‘Source’ for V and mA output.
To output voltage or current:-
• Select ‘Source’ using the key and press the
ENTER
ZERO
key.
Three options become available; ‘Actual’, ‘Convert’ and ‘P V’ / mA.
• ‘Actual’ is selected by pressing the key.
ENTER
ZERO
An example display below shows the Source Mode symbol on the second line.
• Select ‘Converted’ using the key and then pressing
ENTER
ZERO
the key. To set the converted parameters refer to section 2.3.1 on page 23.
15Page
Loading...
+ 51 hidden pages