GE
Sensing
General
Pressure Measurement
The DPI 150 is available with the following selection of
internal pressure sensor ranges. It is also available
without internal pressure range for use instead with
IDOS UPM and IDOS UPMP external sensors.
Standard Pressure Ranges
• 10 inH2O (25 mbar)
• 1 and 3 psig (70 and 200 mbar)
• 5, 10, 15, 30, 50, 100, 150, 300, 500, 1000, 1500, 2000,
3000 psig (350, 700 mbar, 1, 2, 3.5, 7, 10, 20, 35, 70,
100, 135, 200 bar)
• All versions available with negative gauge calibration
(specify option D)
• For absolute pressure ranges specify option E,
Barometric Reference
• Absolute pressure ranges as above, plus atmospheric
pressure
Over Range
1.1 x FS pressure range
Maximum Working Pressure
• 1.2 x FS pressure range for 10 to 3000 psi
(700 mbar to 200 bar)
• 2 x FS pressure range for 10 inH
2
O to 5 psi
(25 to 350 mbar)
Pressure Media
Pressure range 30 psi (2 bar) and below:
• Silicon, pyrex, titanium and structural adhesive
• Media must be non-conductive
Pressure range 50 to 3000 psi (3.5 to 200 bar):
• 316 stainless steel and Hastelloy C276
• Reference port: dry, non-corrosive gas only
Display
High-contrast panel backlit LCD
Readout
±9999999 maximum, updated two times per second
Pressure Units
24 units plus two user-defined and airspeed in mph,
km/h and kts
Languages
English, Chinese, French, German, Italian, Japanese,
Portuguese and Spanish
Process Features
Hold, Maximum/minimum value, tare and
programmable filter
Performance
Precision
• Precision 0.01% FS from 15 to 3000 psi
(1 to 200 bar)*
• Precision 0.03% FS below 15 psi (1 bar)*
• Precision includes non-linearity, hysteresis,
repeatability and temperature effect between
64.4°F and 82.4°F (18°C and 28°C), for both absolute
and gauge pressures
• Calibration Standard (Deadweight Tester) accuracy
0.005% of reading
*Precision assumes regular zeroing.
Negative Gauge Precision
Maximum error at any negative pressure value is equal
to maximum error at the equivalent positive pressure
value.
Measurement Stability
• 0.01% of reading per year from 15 to 3000 psi
(1 to 200 bar).
• 0.02% of reading per year below 15 psi (1 bar)
Barometric Reference Precision
Precision for the optional barometric reference:
0.002 psi (0.15 mbar). Includes non-linearity, hysteresis,
repeatability and temperature effects between
41°F and 122°F (5°C and 50°C). Long term stability
0.002 psi (0.15 mbar) per year.
Electrical
Communications
• RS232 interface supplied as standard. (SCPI protocol)
• IEEE-488 optional
Power Supply
11V to 26 VAC or VDC, 10VA, via 0.08 in (2.1 mm) jack,
supplied with AC/DC power adaptor 90 to 264 VAC,
45 to 65Hz.
Environmental
Temperature
• Operating 41°F to 122°F (5°C to 50°C)
• Calibrated 71°F (23°C)
• Storage -4°F to 140°F (-20°C to 60°C)
Humidity
Compliant with Def. Stan. 66-31 8.6 Cat. 3
DPI 150
Specifications