Druck DPI 142 Data sheet

Precision Barometric Indicator
Ranges up to 50 psi absolute
Precision 0.01% FS
Long term stability 0.01% FS/yr
Altitude display in feet and meters
Programmable analog output
RS232 and IEEE 488 communications
The DPI 142 is a high accuracy barometric indicator with digital communications suitable for integration into any datalogging or integrated information system. It provides highly accurate
readings using the resonant pressure sensor to better than 0.01% FS over 10°C to 40°C, with excellent long term stability. It can be used as a transfer calibration standard in laboratories or as a company reference standard.
The DPI 142 features easy to use software menus accessed via the front panel push buttons. The instrument also includes functional screens to measure barometric pressure, display a barograph (1 day, 2 day or 5 day trend), read altitude (feet/meters) and leak test pressure or altitude.
Digital communications allows easy data transfer of instrument readings and the analog output option enables the user to set a selectable output to be proportional to readings of pressure or altitude.
DPI 142
GE Infrastructure
Sensing
T 203 746 0400 • F 203 746 2494 www.gesensing.com
All specifications are subject to change for product improvement without notice. GE®is a registered trademark of General Electric Co. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
© 2004. All rights reserved. PDSA177 08/04
SPECIFICATIONS
Standard Specification
Standard Pressure Ranges
10.9 - 16.7 psia (750 - 1150mbar absolute) (barometric only).
0.5 - 19 psia (35 - 1310mbar absolute).
0.5 - 38 psia (35 - 2620mbar absolute).
0.5 - 50 psia (35 - 3500mbar absolute).
Over Range
1.1 x FS pressure range.
Maximum Working Pressure
58 psia absolute.
Pressure Media
Non-corrosive dry gases only.
Display
Panel
High contrast, back-lit LCD.
Readout
±9999999 maximum, updated 2 times per second.
Pressure Units
24 units plus two user-defined and altitude in feet (ft) or meters (m).
Languages
English, Chinese, French, German, Italian, Japanese, Portuguese, Spanish.
Process Features
Hold, Maximum/minimum value, Tare and programmable filter.
Performance
Precision
Precision 0.01% FS includes non-linearity, hysteresis, repeatability and temperature effects over 10° to 40°C.
Measurement Stability
Better than 100 ppm (0.01% FS) per year.
Electrical
Communications
RS232 serial interface supplied as standard (SCPI Protocol). IEEE-488 optional.
Power Supply
11V to 26V AC or DC, 10VA, via 0.083 inch (2.1mm) Jack, supplied with AC/DC power adaptor 90 to 264 VAC, 45 to 65 Hz.
Environmental
Temperature
Operating: 5° to 50°C Calibrated: 23°C Storage: -20° to 60°C
Humidity
Compliant with Def Stan. 66-31 8.6 cat. 3
Vibration
Compliant with Def Stan. 66-31 8.4 cat. 3
Shock
Mechanical shock conforms to EN61010
Conformity
Electrical and mechanical safety: EN61010 EMC Emission: EN61326-1 EMC Immunity: EN61326-1 Certification: CE marked
Physical
Weight
2.2 pounds (1kg) nominal.
Dimensions
7.3” W x 7.7” D x 3” H (185mm W x 195mm D x 75mm H)
Pressure Connection
1/8 NPT female
Options
(A) Analog Output
0 - 10V, 0 - 5V, -5V to 5V, 0/4 - 20mA outputs selectable. Accuracy 0.05% FS, update rate 2 readings per second. Programmable between minimum and full scale pressure for proportional output against pressure.
(B) IEEE 488 digital communications
Full computer control is available via a databus using the SCPI protocol. IEEE parallel D connector is provided on the rear panel.
(C) Panel mounting kit
Two sided plates and front panel cutout enable easy mounting to racks and panels.
Supplied as Standard
User handbook, calibration certificate and AC/DC power adaptor.
Calibration Standards
Instruments manufactured are calibrated against precision pressure calibration equipment which is traceable to International Standards.
Related Products
We manufacture a wide range of pressure sensors, indicators, calibrators, controllers, Air Data Test Systems and deadweight testers. The range of portable calibrators also covers temperature and electrical parameters.
Please refer to gesensing.com for further information and datasheets.
Ordering Information
Please state the following:
(1) DPI 142 (2) Pressure range, absolute (3) Options required.
Continuing development sometimes necessitates specification changes without notice.
Loading...