SPECIFICATIONS
Standard Specification
Standard Pressure Ranges
10.9 - 16.7 psia (750 - 1150mbar absolute) (barometric
only).
0.5 - 19 psia (35 - 1310mbar absolute).
0.5 - 38 psia (35 - 2620mbar absolute).
0.5 - 50 psia (35 - 3500mbar absolute).
Over Range
1.1 x FS pressure range.
Maximum Working Pressure
58 psia absolute.
Pressure Media
Non-corrosive dry gases only.
Display
Panel
High contrast, back-lit LCD.
Readout
±9999999 maximum, updated 2 times per second.
Pressure Units
24 units plus two user-defined and altitude in feet (ft) or
meters (m).
Languages
English, Chinese, French, German, Italian, Japanese,
Portuguese, Spanish.
Process Features
Hold, Maximum/minimum value, Tare and programmable
filter.
Performance
Precision
Precision 0.01% FS includes non-linearity, hysteresis,
repeatability and
temperature effects over 10° to 40°C.
Measurement Stability
Better than 100 ppm (0.01% FS) per year.
Electrical
Communications
RS232 serial interface supplied as standard (SCPI
Protocol). IEEE-488 optional.
Power Supply
11V to 26V AC or DC, 10VA, via 0.083 inch (2.1mm) Jack,
supplied with AC/DC power adaptor 90 to 264 VAC, 45 to
65 Hz.
Environmental
Temperature
Operating: 5° to 50°C
Calibrated: 23°C
Storage: -20° to 60°C
Humidity
Compliant with Def Stan. 66-31 8.6 cat. 3
Vibration
Compliant with Def Stan. 66-31 8.4 cat. 3
Shock
Mechanical shock conforms to EN61010
Conformity
Electrical and mechanical safety: EN61010
EMC Emission: EN61326-1
EMC Immunity: EN61326-1
Certification: CE marked
Physical
Weight
2.2 pounds (1kg) nominal.
Dimensions
7.3” W x 7.7” D x 3” H (185mm W x 195mm D x 75mm H)
Pressure Connection
1/8 NPT female
Options
(A) Analog Output
0 - 10V, 0 - 5V, -5V to 5V, 0/4 - 20mA outputs selectable.
Accuracy 0.05% FS, update rate 2 readings per second.
Programmable between minimum and full scale pressure
for proportional output against pressure.
(B) IEEE 488 digital communications
Full computer control is available via a databus using the
SCPI protocol. IEEE parallel D connector is provided on the
rear panel.
(C) Panel mounting kit
Two sided plates and front panel cutout enable easy
mounting to racks and panels.
Supplied as Standard
User handbook, calibration certificate and AC/DC power
adaptor.
Calibration Standards
Instruments manufactured are calibrated against
precision pressure calibration equipment which is
traceable to International Standards.
Related Products
We manufacture a wide range of pressure sensors,
indicators, calibrators, controllers, Air Data Test Systems
and deadweight testers. The range of portable
calibrators also covers temperature and electrical
parameters.
Please refer to gesensing.com for further information and
datasheets.
Ordering Information
Please state the following:
(1) DPI 142
(2) Pressure range, absolute
(3) Options required.
Continuing development sometimes necessitates