Druck ADTS 505 Data sheet

GE
Measurement & Control
ADTS 505
Druck Air Data Test Set
GE is the foremost supplier of Air Data Test Sets and Systems, with over 25 years of experience in the design and manufacture of advanced pressure measuring instruments and sensors.
The ADTS 505 is the latest in a series of reliable, compact and high accuracy Air Data Test Sets.
GE’s continuous research and development, customer feedback and experience gained from manufacturing thousands of automatic pressure controllers. This has enabled performance, ease of maintenance, and operational simplicity to be optimized.
Features
High accuracy—RVSM compliant
Fully automatic control
Rugged ightline construction
Wide-ranging aircraft compatibility
Integral pneumatic supplies
Protection for aircraft instruments
Standard remote hand terminal
High Accuracy
Fully Automatic Control
Compliant with RVSM (Reduced Vertical Separation
Minima) requirements, the ADTS 505 is a precision twin-channel Ps and Pt pressure control test set for
the accurate calibration/verication of civil aircraft
pitot statics.
It utilizes advanced micro-machined, piezo­resistive and resonant silicon sensors, developed
by GE specically for high accuracy measurement
applications. This technology provides excellent thermal and long term stability.
Suitable for use with a wide range of xed or rotary wing civil aircraft, the ADTS 505 enables vital ight
instrumentation such as Altimeters, Airspeed
Indicators, Rate of Climb Indicators, Mach Meters
and Air Data Computers to be quickly and accurately
tested on the ightline.
User Friendly
The sophisticated, yet user-friendly display is fully programmable for a range of test requirements. It
can be congured, with limits and preferred units of
measurement, for individual aircraft types. The level of information displayed can be determined by the operator to suit the particular task at hand.
The ADTS 505 has fully automatic control by means of the menu-driven high contrast, electroluminescent display and tactile membrane keypad. The desired altitude, airspeed or rate of climb can be input, and the ADTS 505 will automatically generate and continuously maintain the desired value.
A choice of units is provided for Altitude, Airspeed,
Rate of Climb, Engine Pressure Ratio and other
related test functions. Control and measurement is available for these parameters as well as Static, Pitot and Calibrated Airspeed.
Testing With Condence
Protection for the aircraft instruments under test includes a power failure protection feature, which, in the event of an interruption to the electrical supply, locks the pneumatic system and allows safe restoration to ambient ground conditions.
The ADTS 505 will operate with leaking systems, which in practice can often occur. The required aim value is maintained, enabling the test to be completed.
Advanced Functions
In addition to the standard pitot-static functions of the ADTS 505, there are many additional advanced features which include automatic go-to-ground, leak test mode and device under test protection limits amongst others. This enables the operator to
undertake ightline testing accurately and eciently,
without compromising safety standards.
Cost Eective
Highly portable, quick and easy to use, the ADTS 505 is also designed for low maintenance, with a recommended 12 month recalibration interval. Operational productivity is optimized and aircraft downtime is minimized.
Easy Flightline Operation
The self-contained ADTS 505 simplies accurate and RVSM compliant pitot-static testing for a wide
range of civil aircraft. Featuring a highly compact and lightweight yet rugged construction, it enables single-handed portability for convenient use on the
ightline.
Integral Pneumatic Supplies
Matched pressure and vacuum pumps are included within the ABS case, which also provides storage for the pneumatic test hoses supplied with AN4 aircraft connections. The ADTS 505 is conveniently powered from any aircraft or national single phase supply, with no user switching required.
Weighing just 15 kg (33 lb), the compact ADTS 505 oers a convenient yet high performance solution
for responding quickly to pitot static and pressure
related Air Data Test requirements on the ightline.
The ADTS 505 is housed in a rugged weatherproof case, complete with internal pneumatic pumps for pressure and vacuum generation. Automatic control is implemented by simple menu instruction and keypress at the front panel.
ADTS 505 Specication
Parameter Operating range Resolution Accuracy Repeatability
Altitude -609 to 18288 m
Static Sensor 35(3) to 1355 mbar abs (1 to 40 inHg) 0.01 mbar (0.0003 inHg) ±0.1 mbar
Airspeed 20 to 650 knots
Airspeed (Qc) 0 to 2500 mbar di (1 to 74 inHg) 0.01 mbar (0.0003 inHg) ±0.1% reading
Rate of Climb 0 to 1829 m/min(5) (0 to 6000 ft/min) 0.30 m/min (1 ft/min) ±2% of value ±0.5%
Mach 0.6 to 2.8 0.001 Better than 0.005 0.001 rising to 0.005
Engine Pressure Ratio (EPR) 0.1 to 10 0.001 Better than 0.005
(1)
(-2,000 to 60,000 ft) 0.30 m (1 ft) 0.91 m (3 ft) at sea level
(4)
0.1 kts ±0.5 kts at 50 kts ±0.4 kts
2.13 m at 9144 m
8.83 m at 18288 m
(±0.003 inHg)
±0.125 mbar
(2) ,
(2)
(7 ft at 30,000 ft)
(2)
(29 ft at 60,000 ft)
±0.30 m (±1 ft) ±0.60 m (±2 ft) ±1.82 m (±7 ft)
±0.05 mbar
(±0.0015 inHg)
0.05 mbar rising to
0.17 mbar
(measurement)
1. 105,000 ft (32004 m) available (measure mode).
2. Accuracy at ambient 5°C to 35°C (41°F to 95°F) for 0°C to 50°C (32°F to 122°F) x 1.5
3. 35 mbar (1 inHg) lowest calibration point. (Will measure below this value.)
4. Limits settable to prevent excessive Mach. (Civil limit Mach 1).
5. To 9144 m (30,000 ft) into 4 litres. Higher altitudes at lower rates available.
Scaling Factors
• Altitude: ft, meters
• Airspeed: knots, km/hr, mph
• Rate of Climb: ft/min, m/min, m/s, hm/s
• Others: mbar, inHg, inH2O, mmHg, kPa, hPa, psi
Rate Control/Indication
• Roc: Rate of Climb
• Rt Ps: Rate of Static
• Rt Pt: Rate of Pitot
• Rt Qc: Rate of Pt_Ps
• Rt CAS: Rate of calibrated airspeed
• Rt EPR: Rate of engine pressure ratio
Overpressure
Negligible calibration change with up to 1.25 x full scale
(FS) overload applied.
Calibration Stability
Better than 0.005% (Ps) and 0.08% (Qc) FS per annum.
Recalibration
Simple keypad instruction. 12 month interval suggested. Use of primary standard pressure reference is recommended.
Display
Large area, 1/4 VGA, high contrast, electro-luminescent display. 96 mm x 73 mm (3.7 in x 2.8 in), 320 x 240 pixels.
Response
Two readings-per-second display value update.
Power Supplies
Auto-selection between: 100/120/230 Vac, 50/60 Hz, Power - 200 VA 115 Vac, 400 Hz, Power - 200 VA
Power Failure Protection
System locks, and a manual let-down feature is provided.
Self Test
Integral test routines and reporting for both electrical and pneumatic systems.
Temperature Range
Calibrated: 5°C to 35°C (41°F to 95°F)
Operating: 0°C to 50°C (32°F to 122°F)
Storage: -20°C to 70°C (-4°F to 158°F)
Sealing
Weatherproof in operating mode (lid removed).
Humidity
0 to 95% non-condensing. “Tropicalized” specication.
Shock/Vibration
Designed to meet MIL-T-28800 Class 2.
Conformity
EN61010, EN61326. CE marked.
Physical
Weight 15 kg (33 lb) nominal
Dimensions (including lid): 265 mm x 520 mm x 355 mm (10.4 in x 20.5 in x 14 in)
Case
ABS moulded case with removable lid and storage for accessories.
Pneumatic Connections
AN4 for Ps and Pt, both color-coded.
Supplied with mating (approximately) 3 m (10 ft) long
exible hoses with AN4 ttings at one end.
Pneumatic Supplies
Integral pressure and vacuum pumps capable of
generating the following supply rates:
1829 m/min (6000 ft/min) into a 4 litre volume at 9144 m (30,000 ft)
300 knots/min into a 2 litre volume at 650 knots
Water/moisture content is vented automatically.
Standard Accessories
AC power lead—5 m (16 ft) length approximately. Ps and Pt hoses—3 m (10 ft) lengths approximately. Operator’s manual and calibration certicate also supplied as
standard.
Standard Remote Hand Terminal
Electroluminescent display and membrane keypad, operates exactly as front panel display interface. Includes 18 m connecting cable.
Calibration Standards
Instruments manufactured by GE are calibrated against precision calibration equipment traceable to international standards.
Options
Option C - Alternative Power Lead: C-1: UK 13 Amp Plug - 5m/16ft Lead C-2: USA 5-15p Plug - 5m/16ft Lead C-3: European CEE7/7 Plug - 5m/16ft Lead
Warranty Terms
New product is supplied with an initial warranty of 18
months. For peace of mind ,particularly if nal installation
is months away from your product purchase, extend coverage on your equipment beyond the initial period up to a total 4 year term.
• Improved cost predictability
• Increased assurance
Multi-year calibration and repair services agreements
Multi-year service agreements increase cost predictability
by providing xed rates for extended periods.
Ordering Information
Please state the following:
1. ADTS 505
2. Any options or special requirements
3. Supporting Services (order as separate items)
Supporting Services
GE provides services to enhance, support and complement the Aviation GSE range. Our highly trained
sta can support you, no matter where you are in the
world. Further details can be found in www.ge-mcs.com/en/services-and-support.html
Nationally accredited calibration
New product is supplied with factory calibration
certicates with measurements traceable back to
international standards. For applications where initial
nationally accredited calibration certicates are required
or periodic re-calibration is desired, GE can provide the solution.
Maintenance
Should your equipment need maintenance our global repair facilities are happy to serve. Work is conducted by trained approved technicians, using controlled original equipment parts and procedures so restoring the product to design condition.
Related Supporting Products
LSU105 Line Switching Unit - A manually operated port switching unit used to supply or isolate Ps and Pt pressures on multiple aircraft ports simultaneously.
www.ge-mcs.com
920-184E
© 2013 General Electric Company. All Rights Reserved. Specications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not aliated with GE.