Drager
“MEDICAL
OPERATION
Minimum
Outlet
Dew
Power
Current
Circuit
Weight:
Noise
Compressor
*
PRECAUTIONS
AIR
AND
Continuous
Pressure:
Point
Depression:
Requirement:
Leakage:
Breaker:
126
Ibs.
Level:
52
Safety
COMPRESSOR
SERVICE
SPECIFICATIONS:
37-49
15
Amps
dBa + 2
Valve:
MANUAL
Output:
psi
3° C below
115V,
60Hz,
normal
less
than
dBa
85
psig t 5
60
Ipm © 37
room
10
Amps
polarity
85
microamps
ambient
open
psig
psi
ground
Before
with
all
use
the
service
WARNING:
-
WARNING:
-
WARNING:
WARNING:
-
CAUTION:
|
CAUTION:
operating
the
precautions,
compressor
manual
To
personnel.
Do
anesthetics.
The
outlet
Never
are
It
the
moisture
(portable
emptied
The
lubricated.
or
servicing
if
to
perform
avoid
not
compressor
not
is
recommended
inlet
compressor
notes
repair
scheduled
shock
use
the
or
an
adaptor.
put
oxygen
approved
of
all
in
case
or
daily.
the
air
and
is
needed.
hazard,
compressor
is
into
for
blenders,
the
piped)
is
compressor,
warnings.
Replace
maintenance
refer
grounded.
the
compressor.
oxygen
that
you
respirators,
drying
fails.
The
of
the
oil-less
read
and
understand
Drager
all
in
the
DO
use.
install a transparent
system
water
is
not
broken
and
repairs.
servicing
presence
NOT
plug
The
etc.
to
of
the
trap
should
type
and
responsible
or
worn
to
qualified
of
it
electrical
protect
compressed
MUST
parts
flammable
into
an
ungrounded
components
water
them
be
checked
NOT
be
all
the
accompanying
for
damages
immediately.
trap
at
from
air
source
and
caused
Follow
manuals.
by
misuse.
the
instructions
Be
familiar
Do
ОКАСЕК
JUNE
not
in
0-18
this
ОЛ
199!
|
TABLE
OF
CONTENTS
ROUTINE
WEEKLY
QUARTERLY
YEARLY
BETWEEN
TROUBLE
A.
B.
C.
D.
E.
F.
-
SERVICE
A.
B.
EXTERNAL
A.
B.
REPLACEMENT
WIRING
SCHEMATIC
PNEUMATIC
MAINTENANCE
8000-9000
SHOOTING
PRESSURE
GAUGE
OVERHEATING
NOISE/EXCESSIVE
WATER
HIGH
LOW
TO
ADJUSTING
AIR
AIR
INSTALLING
PROBLEMS
PRESSURE
PRESSURE
If
the
If
the
CHECK
RESERVOIR
OUTLET
PARTS
compressor
compressor
FOR A FAULTY
THE
MODIFICATIONS
OPTIONAL
LIST
SCHEMATIC
HOURS
NEEDLE
VIBRATION
is
is
PRESSURE
KIT
EXTERNAL
STICKS
not
running
running
SOLENOID
SWITCH
ROUTINE
OR
VIBRATES
AIR
RESERVOIR
MAINTENANCE
KIT
NN
Mn
OO
Y“
Y“
Y
“Y
00
© © © ©
©
O
Note:
WEEKLY
Remove
and
the
reinstall.
QUARTERLY
Wash/replace
1.
_Unthread
door.
Locate
Unscrew
Snap
elements
well,
and
4.
Reassemble
After
performing
specifications
foam
inlet
Replacement
the
air
pump
the 8 screws
the
muffler
intake
the
cap
out.
allow
to
the
inside
by
filter
part
intake
in
filter
off
of
the
Wash
dry
and
of
the
reversing
any
maintenance
on
page
on
the
front
#13989.
filter
from
the
the
left
side
assy.
from
intake
them
in
warm
reinstall.
filter
with a clean
steps 1 through
1,
use
of
the
elements:
rear
of
the
the
muffler.
filter
and
soapy
While
procedure,
an
RT-200
compressor.
door
and
compartment.
pull
the
water.
the
cap
is
dry
cloth.
3.
or
equivalent
remove
filter
Rinse
off,
verify
Wash
wipe
the
it
it
performance
measuring
in
warm
“FILTER,
FIGURE
device.
soapy
COMPLETE
968-402
i
of
the
water,
INTAKE
unit
rinse
—
to
the
well,
allow
“
FILTER,
901-426
to
FILTER,
901-425
FOAM
dry
FELT