Buchi Rotavapor R-300 Data Sheet

Rotavapor® R-300
Technical data sheet
The Rotavapor® R-300 meets the highest expectations in convenience and versatility for rotary evaporation. Its modular design allows for easy extension of the R-300 to a fully integrated sys­tem where a central interface regulates each component.
Overview of Rotavapor® R-300 stand-alone and systems
System
Rotavapor® R-300
Rotavapor® R-300 / R-300 Hand Lift** 1 1
Heating Bath B-301 / B-305** * 1
Interface I-300 / I-300 Pro (incl. VacuBox)** * 1
Vacuum Pump V-300 - 1
Recirculating Chiller F-305 / F-308** - *
* optional, according to order code ** according to order code
“Rotavapor® Dynamic”
Rotavapor® R-300
The Rotavapor® R-300 represents high reliability and convenience in laboratory rotary evaporation applications. The Rotavapor® R-300 is available with either an electronic or manual lift, and can optionally include the Hea­ting Bath B-301 or B-305 and, a glass assembly of your choice and the Interface I-300 or I-300 Pro.
All accessories and connections are included for out of the box usage.
2
Order code
Choose the configuration according to your needs:
Rotavapor® R-300
1 3 01 R 0
Lift System
1 Manual lift 2 Electronic lift
Heating Bath
0 Without 1 B-301: 20 – 95 °C (water), max. flask size: 1 L 5 B-305: 20 – 220 °C, max. flask size: 5 L
Standard Joint
0 Without (w/o evaporating flask, w/o vapor duct) 1 SJ 29/32 (incl. evaporating flask, vapor duct) 2 SJ 24/40 (incl. evaporating flask, vapor duct)
Glass Assembly
N Without (w/o receiving flask)
1)
V Vertical (incl. receiving flask) A Diagonal (incl. receiving flask) C Cold trap (incl. receiving flask) S Vertical reflux (incl. receiving flask)
Protective Coating
0 Without 1 Safety coating (P+G)
1)
2)
Interface
0 Without 1 Without, with Woulff bottle 2 I-300, with Woulff bottle (incl. VacuBox) 3 I-300 Pro, with Woulff bottle (incl. VacuBox) 4 I-300, with valve unit (incl. VacuBox) 5 I-300 Pro, with valve unit (incl. VacuBox)
Voltage
1 220 – 240 V 2 100 – 120 V
1)
If glass assembly E, CR or BY is required, for Glass Assembly select “N”, for Standard Joint select either “1” or “2” and for Protective
Coating select “0” (vapor duct and evaporating flask are included). Glass assembly E, CR or BY need to be ordered separately.
2)
”Safety coating (P+G)” applies to condenser and receiving flask
3
Scope of delivery of Rotavapor® R-300
All configurations are delivered ready to use and are complete of:
Components Rotavapor® R-300
Rotavapor® R-300 / R-300 Hand Lift** 1
Heating Bath B-301 / B-305** *
Glass assembly** *
Vapor duct (with Combi-Clip) **
Vacuum seal (WD26) 1
Evaporating flask (1 liter) **
Receiving flask (1 liter) **
Flyer “List of solvents” 1
Set of required cooling hoses **
Power cable (1 for heating bath, 1 for Rotavapor®) **
Interface I-300 / I-300 Pro** *
VacuBox *
* optional, according to order code ** according to order code
4
System “Rotavapor® Dynamic”
The rotary evaporator R-300 system meets the highest expectations in convenience and versatility. It is bundled with the Vacuum Pump V-300, controlled by the Interface I-300 / I-300 Pro and can include the Recirculating Chiller F-305 / F-308. The interface is the central control unit for all process parameters. Rotation speed, heating-, cooling-, vapor-temperature and pressure are perfectly synchronized to optimize the distillation process while increasing efficiency and convenience. All the “Rotavapor® Dynamic” systems are P+G coated. All accessories and connections are included for out of the box usage.
Choose the configuration according to your needs:
11 S
3 0R 0
Lift System
1 Manual lift 2 Electronic
Heating Bath
1 B-301: 20 – 95 °C (water), max. flask size: 1 L 5 B-305: 20 – 220 °C, max. flask size: 5 L
Standard Joint
1 SJ 29/32 (incl. evaporating flask, vapor duct) 2 SJ 24/40 (incl. evaporating flask, vapor duct)
Glass Assembly
V Vertical (condenser and receiving flask are P+G coated) C Cold trap (condenser and receiving flask are P+G-LT coated)
3)
Interface
S I-300, with Woulff bottle (incl. VacuBox) P I-300 Pro, with Woulff bottle (incl. VacuBox)
Recirculating Chiller
0 Without 1 F-305: cooling capacity 550 W at 15 °C (-10 – 25 °C) 2 F-308: cooling capacity 900 W at 15 °C (-10 – 25 °C)
Voltage
1 220 – 240 V 2 100 – 120 V
3)
When ordering the glass assembly C, a recirculating chiller is not needed
5
Scope of delivery of “System Rotavapor® Dynamic”
All configurations are delivered ready to use and are complete of:
Components “System Rotavapor® Dynamic”
Rotavapor® R-300 / R-300 Hand Lift** 1
Heating Bath B-301 / B-305** 1
Glass assembly 1
Vapor duct (with Combi-Clip) 1
Vacuum seal (WD26) 1
Evaporating flask (1 liter) 1
Receiving flask (1 liter) 1
Flyer “List of solvents” 1
Set of required cooling and vacuum hoses 1
Set of required power cables 1
Interface I-300 / I-300 Pro** 1
VacuBox 1
Vacuum Pump V-300 1
Recirculating Chiller F-305 / F-308** *
* optional, according to order code ** according to order code
6
Technical data
Rotavapor® R-300
Dimension (WxHxD)
Instrument
Shipping box
Weight 13.5 kg (electronic lift)
Connection voltage 100 – 240 VAC
Power consumption 10 0 W
Frequency 50/60 Hz
Protection class IP21
Immersion angle adjustable 40 °
Stroke distance 220 mm
End stop position adjustment range 170 mm (electronic lift)
400 x 615 x 320 mm
700 x 605 x 590 mm
13.0 kg (manual lift)
100 mm (manual lift)
Rotation speed 10 – 280 rpm
Maximum flask load 3 kg
Approvals CE / CSA
Heating Bath B-300 Base
Dimension (WxHxD) 183 x 82 x 286 mm
Weight 0.8 kg
Connection voltage 100 – 120 V / 220 – 240
Frequency 50 – 60 Hz
Power consumption (with bath) 150 0 W
Protection class IP21
Approvals CE / CSA
Heating Bath B-301 Bath Heating Bath B-305 Bath
Dimension (WxHxD) 218 x 192 x 205 mm 307 x 202 x 275 mm
Weight 2.0 kg 4.2 kg
Operation voltage (according to order code)
Frequency 50 – 60 Hz 50 – 60 Hz
100 – 120 V
220 – 240 V
100 – 120 V
220 – 240 V
7
Heating Bath B-301 Bath Heating Bath B-305 Bath
Power consumption 1250 W 150 0 W
Heating Power 110 0 W 1300 W
Protection class IP21 IP21
Controlled temperature
ambient to 95 °C ambient to 220 °C
range
Maximum flask size 1000 mL 5000 mL
Adjustment accuracy ± 1 °C ± 1 °C
Temperature deviation at 60 °C ± 1 °C
at 95 °C ± 1 °C
at 60 °C ± 1 °C
at 95 °C ± 1 °C
at 180 °C ± 3 °C
at 220 °C ± 4 °C
Heating media water water / oil
Approvals CE / CSA CE / CSA
Heating bath performance
Heating-up time of Heating Bath B-301 and B-305
Water Oil (Ucon HTF 14)
50 °C 95 °C 180 °C 220 °C
B-301 4 min 13 min --- ---
B-305 8 min 22 min 27 min 37 min
Water
100
50
0
0
Temperature (°C)
∆T
70 °C (water)
: B-301 -> 13 min; B-305 -> 22 min
5 10 15 20 25
Time (min)
B-301 B-305
Special features
Rotavapor® R-300
End stop positioner Adjustable via button within a range of 170 mm (electronic) / 100 mm (manual)
Prevents glass breakage and damages to heating bath pan
Patented multifunctional Combi-Clip
Tool for evaporating flask fixation and removal, as well as vapor duct release
Allows one handed flask removal
8
Rotavapor® R-300
Immersion angle adjustable
Safety stop (electronic lift only)
Adjustable within 7-stages
enables the use of different flask sizes, adjustable for individual application
Automatic operation stop by power failure
Lifts flask automatically out of heating bath
2 lift speeds Moves quickly until 2 cm above close to depth stop, it than moves
slowly to depth stop
Allows a smooth operation
Earthquake safety Lug on backside for bath and Rotavapor
®
Enables fixation on lab bench
Lift height indication Position is digitally indicated on display of heating bath
Allows reproducibility of parameter for future applications
Color code Green cables indicate BUCHI communication cable, marked with
“COM”
Allows a simple understanding of set up and easy play & plug
Large top hole V-condenser
Top hole with screw cap (SVL22)
Allows easy cleaning of condenser
Ergonomic handle Electronic and manual version available
Allows to conveniently change the position of lift
Display Rotation speed, heating temperature, lift position (set and actual
value) displayed on Heating Bath B-300 Base
Allows user to monitor all parameters at a glance
Heating Bath B-301 / B-305
Universal use (B-305 only)
Useable for multiple applications up to 220 °C
Allows the use of water and oil as a heating media
Over heat protection Mechanical and electronic over-temperature control
Cuts off power when actual exceeds set temperature and by rapid tempera­ture increase
Carry handles Two integrated laterally ergonomically designed carry handles
Allows easy carrying for re-filling or emptying the heating bath, without the risk of burn
Bath guidance Integrated in Heating Bath B-300 Base
Allows an easy positioning of bath along horizontal axe, for individual application
Temperature limit fixation (B-305 only)
Desired maximum temperature can be fixed at 95 °C, 180 °C or 220 °C
Prevents accidental change of set temperature when water is used as heating media
Key-button locking functions
To lock the set temperature of the heating bath
Prevents accidental change of set conditions during process
9
Heating Bath B-301 / B-305
Infrared communication Transition of heating bath data to interface
Allows immaculate, fast communication
Kettle connection Cordless bath
Allows easy emptying and refilling of bath without unplugging
Heating Bath B-300 Base extendable
Functional with two different heating baths
Allows the use of B-301 or B-305, adequate to application
LCD-display Digital, displays all operating parameters
Indicates set and actual values of heating bath temperature, rotation speed and lift position
Optional accessories
Protection shield Special shaped, robust shield for user protection
enables handling on evaporating flask while being protected
Top cover for heating bath
To cover B-305 when not in use to safe energy and for less evapo­ration of heating medium, no drip off, convenient handle prevents burning skin
Complementary products
Pump V-300 Chiller F-305 Chiller F-308
R-300 autonomous manually controlled manually controlled
R-300 with I-300 / I-300 Pro and VacuBox*
speed controlled controlled controlled
*To connect a Vacuum Pump V-300 and a Recirculating Chiller F-305 or F-308 to an Interface I-300 or Inter­face I-300 Pro a VacuBox is needed. A VacuBox is included when ordering a Rotavapor® R-300 with an Interface I-300 or Interface I-300 compatibility pack.
Compatibility of older generation’s peripherals
Pump Pump V-300
V-700 /
V-710
Pump V-10 0
Chiller F-100
R-300 autonomous autonomous autonomous autonomous manually
R-300 with I-300 / I-300 Pro
R-300 with I-300 / I-300 Pro and LegacyBox
speed controlled
speed controlled
(no need of
LegacyBox)
regulated* regulated* autonomous ---
speed
regulated regulated regulated
controlled
R-300 with I-100 regulated regulated regulated autonomous regulated
Chiller F-105 / F-108
controlled
10
Pump V-300
Pump
V-700 /
V-710
Pump V-10 0
Chiller F-100
Chiller F-105 / F-108
R-300 with V-850 / V-850
To connect the Vacuum Pump V-100 or V-710 and the Recirculating Chiller F-100 or F-105 to the Interface I-300 or Interface I-300 Pro a Vacubox and a LegacyBox is needed. A VacuBox is included when ordering a Rotavapor R-300 with Interface I-300 or an Interface I-300 compatibility pack.
∙ Communication cable 11060649 is additionally needed
Explanation of terms
Autonomous: the peripheral (chiller / pump) can be turned On and Off by its own power
Regulated:
Speed controlled:
Manually controlled: the parameters can be adjusted via the peripherals control buttons
regulated regulated regulated autonomous regulated
switch only
the peripheral (pump and chiller) is automatically turned On and Off via inter­face -> ON/Off communication
the peripheral (pump) is turned On and Off as well as speed controlled via the inter face
Controlled: the temperature of the chiller can be also set via interface (bi-directional)
Exemplary solutions
The following items are listed on two solutions, shown in the “Laboratory Rotary Evaporation Solutions” brochure.
Solution “Rotavapor® Dynamic Pro” Order number
Rotavapor® R-300, condenser V, SJ 29/32, P+G safety coating, Heating Bath B-305, Interface I-300 Pro, VacuBox, Vacuum Pump V-300, Recirculating Chiller F-308 (220 – 240 V version)
Foam sensor 110 61167
AutoDest sensor 110 5 9 22 5
Level sensor 11060954
Solution “Rotavapor® Dynamic” Order number
Rotavapor® R-300, condenser V, SJ 29/32, P+G safety coating, Heating Bath B-301, Interface I-300, VacuBox, Vacuum Pump V-300 (220 – 240 V version)
11S R 3 0 02 51V P 21
11S R 3 0 0211V S 01
Protection shield 110 6140 2
11
Accessories – Rotavapor® glass assemblies
The following items have to be purchased in addition to a system that is ordered without glass.
All glass assemblies include a 1 liter receiving flask (corresponding to order code - either without coating, with P+G or P+G-LT), required hoses and a ball joint clamp. Evaporating flask, vacuum seal, vapor duct (WD26) and condenser holder are not included. All condensers allow feeding of the evaporating flask via stopcock.
Cooling Abbr./ Name Type / Application
surface
area (max.) Order code
Ver tical Vertical condenser
∙ For standard applications, the most common
condenser ∙ Connection for vapor temperature sensor ∙ Automatic distillation with Interface I-300 / I-300
Pro possible ∙ Use of foam sensor possible
Diagonal Diagonal condenser
∙ For standard applications where height is limited
Cold trap Dry ice condenser
∙ For distillation of solvents with low boiling points ∙ Maximum condensation of vapors ∙ No cooling water needed, but i.e. dry ice or ice ∙ Use of foam sensor possible
Reflux Vertical reflux condenser with shut-off valve
∙ For combining reflux reaction and distillation ∙ Connection for vapor temperature sensor ∙ Automatic distillation with Interface I-300 / I-300
Pro possible ∙ Use of foam sensor possible
150 0 c m
150 0 c m
500 cm
150 0 c m
2
110 6 2 4 3 2
110 6 2 4 3 3 P+G
2
04 8168
04 8169 P+G
2
040640
04 0 6 42 P+G
2
048290
048291 P+G-LT
Cold trap reflux
Dry ice condenser with shut-off valve
∙ For combining reflux reaction and distillation ∙ For distillation of solvents with low boiling points ∙ Maximum condensation of vapors ∙ Connection for vapor temperature sensor ∙ No cooling water needed, but i.e. dry ice or ice ∙ Use of foam sensor possible
500 cm
2
048292
048293 P+G-LT
12
Abbr./ Name Type / Application
Cooling surface area (max.) Order code
Expansion Descending condenser with expansion vessel
∙ Ideal for distillations exhibiting foaming and
“bumping” tendencies/products ∙ Connection for vapor temperature sensor ∙ Automatic distillation with Interface I-300 / I-300
Pro possible
Double jacket
Vertical reflux condenser with double jacket for inten­sive condensation and with shut-off valve
∙ Vertical intensive condenser with double jack-
et and distributor for particularly efficient
condensation ∙ For combining reflux reaction and distillation ∙ For particularly efficient and intensive condensation ∙ Connection for vapor temperature sensor ∙ With an additional SJ 29/32 joint on top for flexible
expansion ∙ Use of foam sensor possible
150 0 c m
1300 cm
2
110 61112
110 61113 P+ G
2
048176
048297 P+G
For more information concerning the glass assemblies please check the “Laboratory Evaporation Glassware” brochure.
Features of the glass assemblies
Top in l e t
For easy cleaning of the condenser
1
Condenser holder cpl. (048180) is recommended
Characteristics V C A CR S E BY
Top inlet
Inner glass tube
Inner glass tube
For reduced chimney effect upon vacuum applications
Stopcock (available for R-215, R-210, R-3; see FAQ)
Used for aeration and feeding (by using the feeding hose)
Inner glass plate
Prevention of solvent back flow
Connector for sensor
For measurement of vapor temperature during distillation
(SVL22;
extra large)
Flask feeding via stop­cock possible
Inner glass plate
13
Connector for vapor temperature sensor
Condenser holder
1 1 1 1 1
1
Reflux possibility (part
included)
Condensate trap
Rotavapor® accessories
Adapter B-300. For Heating Bath B-305
Required when Heating Bath B-305 is used with a Rotavapor® R-210, R-215 or R II. Heating Bath B-301 does not fit.
Vapor temperature sensor. Incl. cap nut. seal GL14
Measures the vapor temperature inside the system. Meant to be used with the Interface I-300/I-300 Pro.
Lid. For Heating Bath B-305
To safe energy and to minimize loss of water when heating bath is not in use.
Set earthqake fastener. Lug for backside
To fasten Rotavapor® on lab bench.
Order number
110 61317
11060707
110 5 9 5 0 0
110 6 2 3 8 6
Extraction unit Soxhlet. 200 mL, incl. extraction part, reduction part
For Soxhlet extraction applications, Meant to be used with glass assembly S.
Extraction unit Soxhlet. 500 mL, incl. extraction part, reduction part
For Soxhlet extraction applications. Meant to be used with glass assembly S.
Extraction thimbles. set. 25 pcs, 200 mL, cellulose
For Soxhlet extraction unit.
Flange screwed connection. For glass assembly V/C/A
Content: Flange nut, pressure spring
Level sensor. Incl. O-ring, cable, rubber band
Prevents an overflow in the receiving flask of condenser/secondary condenser. Placed on receiving flask. Meant to be used with a Interface I-300/I-300 Pro.
01174 4
01174 5
01810 6
048237
11060954
Holder. For extraction unit, steel, 750 mm, Ø12 mm 0119 0 4
Communication cable. Mini-DIN to RJ45, 1.5 m
Connection between Vacuum Controller V-850/V-855 and Vacuum Pump V-300 or between Vacuum Cont­roller V-850/V-855 and Recirculating Chiller F-3xx.
Condensate trap. Grey, TPE, for condenser V, hose connector Ø8 mm
Collects and drains condensate which may accumulate at the condenser.
Condensate trap. Green, TPE, for condenser V, hose connector Ø8 mm
Collects and drains condensate which may accumulate at the condenser
11060649
110 6 2 9 5 5
110 619 8 5
Condenser holder. For glass assembly V/C/S/CR/BY
Content: Holding rod, rubber band, cross sleve
048180
14
Order number
Cooling water valve. 24 VAC
Valve opens cooling water feed during distillation. Meant to be used with a vacuum controller/interface.
Stopcock. For condenser C/CR, glass, SJ18.8/38
For aeration of system. Mounted on vacuum connection at cold trap outer part.
Stopcock. Incl. 3-way valve
For feeding for solvents. Content: Introduction hose 300mm, backfeed hose 600mm, cap nut GL10.
Stopcock. Professional, glass, SJ18.8/38
For aeration of system. Less cross-contamination compared to standard-stopcock (040627).
Stopcock. PTFE, SJ18.8/38
For aeration of system. For applications when grease should be avoided. Used instead of standard-stop­cock (040627).
Stopcock. Standard, glass, SJ18.8/38
For aeration of system.
Legacybox. Incl. accessory kit
Interface between Interface I-300/I-300 Pro (with VacuBox) to peripherals of older generations (Vacuum Pump V-7xx, Recirculating Chiller F-1xx).
Flask holder. EPDM, slip free
Holder for round-bottom flasks (50-5000 mL).
031356
040628
110 5 8 814
000637
023896
040627
110 6116 6
048 618
Flask holders. set. EPDM, slip free
Holder for round-bottom flasks (50-5000 mL).
Tubing, FEP. Ø6/8 mm, transparent, per m
Use: Vacuum, cooling media
Tubing. Natural rubber, Ø6/16 mm, red, per m
Use: Vacuum
Tubing. Nyflex, PVC-P, Ø8/14 mm, transparent, per m
Use: Vacuum, cooling media, feeding (industrial Rotavapor)
Tubing. PTFE, Ø4.7/5.5 mm, transparent, 330 mm
Use: To introduce solvent into evaporating flask during distillation.
Tubing. PTFE, Ø4.7/5.5 mm, transparent, 460 mm
Use: To introduce solvent into evaporating flask during distillation.
Tubing. Silicone rubber, Ø6/9 mm, transparent, per m
Use: Cooling media
Protection shield. Operator protection for heating bath
For Heating Bath B-301 and B-305.
Heating bath balls. 450 pcs, PP, Ø10 mm
To reduce energy consumption of heating bath and for less evaporation of the heating medium. For temp­eratures up to 100 °C.
11059916
027900
01762 2
0 0 4113
000646
000643
00413 3
110 6140 2
036405
Vacuum valve. Magnetic valve, 24V/4W, Mini-DIN, 1.5 m
Flow valve without flask, meant to be used with a centralized vacuum source or an unregulated vacuum pump. Meant to be used with the Interface I-300/I-300 Pro.
Vacuum connection. Incl. drain valve, hose barb
Connection piece for aeration of the system, placed between condenser and receiving flask.
11060706
001006
15
Order number
Water regulation nozzle. Flow Regulator, incl. hose clamp, sieve
Used when tap water is used to generate vacuum. Reduces water consumption.
Water jet pump. Plastic
Used when tap water is used to generate vacuum.
Rotavapor® accessories – Heating Baths
Heating bath. Heating Bath B-301, not incl. Base B-300, 110V
For up to 1 Liter evaporating flasks. For temperatures up to 95 °C. Meant to be used with water as heating media.
Heating bath. Heating Bath B-301, not incl. Base B-300, 230V
For up to 1 Liter evaporating flasks. For temperatures up to 95 °C. Meant to be used with water as heating media.
Heating bath. Heating Bath B-301, incl. Base B-300, 110V
For up to 1 Liter evaporating flasks. For temperatures up to 95 °C. Meant to be used with water as heating media.
Heating bath. Heating Bath B-301, incl. Base B-300, 230V
For up to 1 Liter evaporating flasks. For temperatures up to 95 °C. Meant to be used with water as heating media.
0116 0 6
002913
Order number
11B 3 0 1 0 0 2
11B301001
11B 3 0 11 0 2
11B 3 0 11 0 1
Heating bath. Heating Bath B-305, not incl. Base B-300, 110V
For up to 5 Liter evaporating flasks. For temperatures up to 220 °C. Meant to be used with water or oil as heating media.
Heating bath. Heating Bath B-305, not incl. Base B-300, 230V
For up to 5 Liter evaporating flasks. For temperatures up to 220 °C. Meant to be used with water or oil as heating media.
Heating bath. Heating Bath B-305, incl. Base B-300, 110V
For up to 5 Liter evaporating flasks. For temperatures up to 220 °C. Meant to be used with water or oil as heating media.
Heating bath. Heating Bath B-305, incl. Base B-300, 230V
For up to 5 Liter evaporating flasks. For temperatures up to 220 °C. Meant to be used with water or oil as heating media.
11B 3 0 5 0 0 2
11B 3 0 5 0 0 1
11B 3 0 510 2
11B 3 0 510 1
Rotavapor® wear parts
Order number
Seal. For cap nut GL14 to FEP, EPDM 038225
Seals. set. 10 pcs, for hose barbs GL14, EPDM, black 040029
Seals. set. 10 pcs, for hose barbs GL14, FPM, green 040040
Seals. set. 20 pcs, for hose barbs GL14, silicone, red 040023
O-ring. FKM, 40.9/2.6 mm, for reflux insert 048078
O-ring. FPM, for screw cap GL10, Ø3.00/2.70mm 023900
Hose barb. Bent, GL14, incl. silicone seal 018 916
Hose barbs. set. 2 pcs, bent (1), straight (1), GL14, silicone seal
Content: Hose barbs, cap nuts, seals
04193 9
16
Order number
Hose barbs. set. 2 pcs, bent (1), straight (1), GL14 silicone seal
Content: Hose barbs, cap nut, screw caps seals
Hose barbs. set. 3 pcs, bent, GL14, silicone seal
Content: Hose barbs, seals
Hose barbs. set. 4 pcs, bent GL14, silicone seal
Content: Hose barbs, cap nuts, seals
Hose barbs. set. 4 pcs, bent, GL14, EPDM seal
Content: Hose barbs, cap nuts, seals
Hose barbs. set. 4 pcs, bent, GL14, FEP seal
Content: Hose barbs, cap nuts, seals
Hose barbs. set. 4 pcs, straight, GL14, EPDM seal
Content: Hose barbs, cap nuts, seals
Hose barbs. set. 4 pcs, straight, GL14, FPM seal
Content: Hose barbs, cap nuts, seals
Hose barbs. set. 4 pcs, straight, GL14, silicone seal
Content: Hose barbs, cap nuts, seals
Hose barbs. set. 6 pcs, bent (4), straight (2), GL14, silicone seal
Content: Hose barbs, cap nuts, seals
110 619 2 1
0419 87
037287
043129
040295
043128
040296
037642
038000
Screw caps. set. 5 pcs, GL14 040624
Cap nut. Screw cap with hole GL10 023875
Cap nuts. set. 10 pcs, screw cap with hole GL14, seal EPDM
Content: Hose barbs, cap nuts, seals
041999
Cap nuts. set. 10 pcs, screw cap with hole, GL14 04195 6
Vacuum gasket. WD26, PTFE, FDA-compliant 048021
Rotavapor® glass accessories – Vapor ducts
Vapor ducts compatible with WD26
SJ 29/32 24/40 29/42 24/29 34/35
For glass assembly V, C, S, E, CR, BY
For glass assembly A
For glass assembly C, V (analytical)
For glass assembly C and V (KD22)
110 6 2 18 6 110 6 2 18 7 110 6 2 4 6 4 11062909
110 6 2 26 7 110 6 22 6 8 1106 22 6 9 048167
110 6 2 4 6 5 110 6 24 6 6 110 6 24 67
11062910
Additional glassware
Additional glassware like evaporating flasks, drying flasks, receiving flasks, beaker flasks, Reitmeyer adapters (bump traps) and distillation spiders can be found in the “Laboratory Evaporation Glassware”-brochure.
www.buchi.com/laboratory-evaporation Quality in your hands
R-300 D ata Sheet en 1506 A / Techni cal data are subj ect to cha nge wit hout not ice /
Quali ty Systems ISO 9001
Loading...