Bio-Rad Bio-Scale Mini CHT 40 µm Cartridges User Manual

Page 1
Bio-Scale™Mini CHT
Ceramic Hydroxyapatite Cartridges, 5 ml
Instruction Manual
Catalog # 732-4322
Page 2
2
Page 3
5.2 Wash Alternatives ..................20
5.3 Autoclaving ............................20
5.4 Storage ..................................20
Section 6 ..Technical Support.............................21
Section 7 ..Ordering Information.........................22
Section 8 ..References .......................................24
Page 4
Page 5
The Bio-Scale Mini CHT cartridges are packed with Type I or Type II ceramic hydroxyapatite supports. These supports are based on hydroxyapatite, a form of calcium phosphate used in chromatographic separations of biomolecules. CHT ceramic hydroxyapatite is a spherical, macroporous form of hydroxyapatite. Unlike most other chromotagraphic absorbents, CHT is both the ligand and the support matrix. CHT Type I has a higher protein binding capacity and greater capacity for acidic proteins. CHT Type II has a lower protein binding capacity but provides better resolution of nucleic acids and certain other proteins. The Type II material also has a very low affinity for albumin and is especially suitable for the purification of many species types and classes of immunoglobulins. Applications of hydroxyapatite chromatography include the purification of different subclasses of monoclonal and polyclonal antibodies, antibodies that differ in light chain composition, antibody fragments, isozymes, supercoiled DNA from linear duplexes, and single-stranded from double
2
Page 6
Table 2. Product Description
TTyyppee II TTyyppee IIII
FFuunnccttiioonnaall ggrroouuppss
Ca
2+
, PO4, OH Ca2+, PO4, OH
OObbsseerrvveedd ddyynnaammiicc bbiinnddiinngg ccaappaacciittyy
llyyssoozzyymmee ((LLyyss))
25 mg Lys/g CHT 12.5 mg Lys/g CHT
NNoommiinnaall ppoorree ddiiaammeetteerr
600–800 Å 800–1,000 Å
MMaaxxiimmuumm bbaacckkpprreessssuurree
100 bar (1,500 psi) 100 bar (1,500 psi)
N
Noommiinnaall mmeeaann ppaarrttiiccllee ssiizzee
20 ± 2, 40 ± 4, and 80 ± 8 µm
BBuullkk ddeennssiittyy
0.63 g/ml 0.63 g/ml
OObbsseerrvveedd ddyynnaammiicc bbiinnddiinngg ccaappaacciittyy
IIggGG
25–60 mg IgG/ml CHT* 15–25 mg IgG/ml CHT*
TTyyppiiccaall lliinneeaarr fflloow
w rraattee rraannggee
50–1,000 cm/hr
ppHH ssttaabbiilliittyy
6.5–14
BBaassee ssttaabbiilliittyy
at least 21 months in 1 N NaOH
RReeggeenneerraattiioonn
500 mM sodium phosphate, pH 7
1,000 mM trisodium phosphate, pH 11–12
AAuuttooccllaavvaabbiilliittyy ((bbuullkk))
121°C, 20 min in phosphate buffer, pH 7
SSaanniittiizzaattiioonn
1–2 N NaOH
RRe
eccoommmmeennddeedd ccoolluummnn ssttoorraaggee
0.1 N NaOH
SShheellff lliiffee ((ddrryy,, uunnuusseedd mmaatteerriiaall))
85 months stored dry, sealed, and at room temperature
* 40 µm particles, 300 cm/hr, 5 mM sodium phosphate, pH 6.5
4
Page 7
(Use orange lock rings and medium size barb fittings with 1.6 mm tubing.)
Fig. 1. Biologic LP setup.
6
Platen pressure
screw
See detail
23
6
1
9
45
.
78
0
C
rm
la
A
Lock-ring
Tubing
Luer fittimg
Page 8
3. Hold the cartridge vertically with outlet in the downward direction. Connect the inlet of the cartridge to the male luer fitting on the MV-6 sample inject valve (Figure 2). If not using the MV-6 sample inject valve, connect a barb to male luer fitting on the 1.6 mm ID tubing, then connect to the top of the female luer on the Bio-Scale mini cartridge. For optimum performance, a cartridge should be mounted vertically with the arrow on the cartridge pointing downward.
4. Connect the cartridge outlet to the 1.6 mm ID tubing leading to the BioLogic LP optics module or Econo UV monitor. It is recommended to use the shortest length (approximately 10 cm) of
1.6 mm ID tubing. Connect a barb to female luer to the 1.6 mm ID tubing, and then connect to the bottom of the male luer on the Bio-Scale Mini cartridge.
8
Page 9
Section 3 Connecting to Other Liquid Chromatography Systems
The Bio-Scale Mini cartridges can be connected to any liquid chromatography system, provided that the maximum pressure limit (3 bar, 45 psi, or 300 KPa) of the cartridges is not exceeded. It is recommended that the system pressure limit be set according to the cartridge pressure limit. Pressures in excess of 3.4 bar are usually caused by restrictions in tubing or detector cells downstream from the cartridge. Bio-Rad offers two fitting kits for easy connection of a Bio-Scale Mini cartridge to a BioLogic DuoFlow, HPLC- or FPLC-type system.
10
Page 10
3.3 FPLC Systems
The luer to M6 adaptor fittings kit (catalog number 732-0111) provides fittings necessary to connect the cartridge to the M6 fittings found on FPLC or related systems.
Alternatively, connection can be made by using the following Upchurch Scientific
®
Quick Connect Luer Adaptors: two Upchurch P-621 adaptor, 1/4–28, to metric adaptors, one Upchurch P-619 adaptor, 1/4–28, to male luer, and one Upchurch P-628 adaptor, 1/4–28, to female luer. Assemble the luers to the 1/4–28 metric adaptors. Attach the adaptor with the male luer to the column inlet line of the FPLC system and the one with the female luer to the FPLC column outlet. To prevent tubing or cartridge failure, do not exceed the maximum recommended flow rate.
12
Page 11
5. Equilibrate the cartridge with running buffer and
operatate the cartridge according to the CHT instruction manual or according to your protocol.
4.1 Sample Preparation
Proper pH and ionic strength are necessary for consistent and reproducible results. Sample can be exchanged into the starting buffer or diluted to the starting buffer's concentration. This can be achieved by diluting the sample to the ionic strength of the starting buffer, dialyzing against the starting buffer, or exchanging it into the starting buffer. Buffer exchange can be accomplished using the Bio-Scale Mini P6 cartridge, Bio-Spin
®
6 or Bio-Spin 30 columns, Econo-Pac 10DG desalting columns, or Bio-Gel
®
P-6DG gel filtration gel. The choice of product will depend on sample volume. All samples should be filtered through a 0.45 µm filter prior to cartridge application.
14
Page 12
calcium. Common buffers for hydroxyapatite chromatography are listed in Table 4.
An appropriate starting point for purifying samples is a linear gradient from 5–500 mM sodium phosphate pH 6.8, spanning 10 to 20 column volumes at 2.0 ml/min for the 5 ml cartridge. For separations of monoclonal antibodies from aggregates, an alternative purification protocol is eluting with a linear gradient from 0-2 M sodium chloride in 5-20 mM sodium phosphate, pH 6.5. The separation can be optimized by changing the gradient profile. At the end of each run the cartridge can be regenerated with 500 mM potassium or sodium phosphate buffer, neutral pH; or 400 mM trisodium phosphate, pH 11–12; followed by starting buffer. Return to the desired flow rate and proceed with the next separation.
4.3 Scaling Up the Separation
For quick scale-up, two or three cartridges of the same type can be connected in series. Backpressure will increase with cartridges in series, so care should be taken to maintain pressures £45 psi. CHT ceramic
16
Page 13
CCHHTT SSuuiittaabbiilliittyy
ppHH BBuuffffeerr 1100 ccyycclleess 5500 ccyycclleess
7.0 MES + 5 mM PO
4
+ n/a
7.0 Acetate + 5 mM PO
4
+ n/a
7.0 Imidazole + 5 mM PO
4
++
7.0 Glycine + 5 mM PO
4
+ n/a
7.0 Arginine + 5 mM PO
4
+ n/a
7.0 HEPES + 5 mM PO
4
+ n/a
7.0 Tris + 5 mM PO
4
+ n/a
7.5 Phosphate (5 mM) + n/a
7.5 MES + 5 mM PO
4
+ n/a
7.5 Imidazole + 5 mM PO
4
++
7.5 Acetate + 5 mM PO
4
+ n/a
7.5 HEPES + 2 mM PO
4
+ n/a
7.5 HEPES + 5 mM PO
4
+ n/a
7.5 Tris + 2 mM PO
4
+ n/a
7.5 Tris + 5 mM PO
4
++
8.5 Tris + 5 mM PO
4
+ n/a
* All experiments performed in small scale columns. Each cycle used approximately 35 column
volumes of buffer to simulate an equilibration and long gradient, as well as five column volumes of 1 N NaOH to simulate regeneration.
18
Page 14
5.2 Wash Alternatives
Perform wash alternatives with any of the following alternative buffers in place of step 1. Continue with step 2 listed in Section 5.1.
1–2 M KCI or NaCI
6 M urea
8 M guanidine-HCI All the wash alternative buffers should contain 5 mM phosphate at neutral pH.
5.3 Autoclaving
Bio-Scale Mini cartridges are not autoclavable.
5.4 Storage
After the cartridges are washed with deionized water, Bio-Scale Mini CHT cartridges should be purged and stored with 0.1 N NaOH. Higher concentrations of NaOH may be used if desired. Used CHT, after being regenerated and sanitized, can be stored in up to 1.0 N NaOH at room temperature.
20
Page 15
Section 7 Ordering Information
Prepacked Bio-Scale Mini Cartridges*
Description 5 x 1 ml 1 x 5 ml 5 x 5 ml
CHT Ceramic Hydroxyapatite, Type I, 40 µm 732-4322 732-4324 CHT Ceramic Hydroxyapatite, Type II, 40 µm 732-4332 732-4334 UNOsphere
Q Support 732-4100 731-4102 731-4104 UNOsphere S Support 732-4110 731-4112 731-4114 Macro-Prep
®
High Q Support 732-4120 732-4122 732-4124 Macro-Prep High S Support 732-4130 732-4132 732-4134 Macro-Prep DEAE Support 732-4140 732-4142 732-4144 Bio-Gel P-6 Support 732-4502 732-4504 Affi-Prep
®
Protein A Support 732-4600 732-4602
Profinity
IMAC Support 732-4610 732-4612 732-4614
Affi-Gel
®
DEAE Blue Support 732-4632 732-4634
Affi-Gel Blue Support 732-4642 732-4644
* For the most up-to-date list of cartridge offerings, please visit
us online at www.bio-rad.com/cartridges/
** Larger package sizes of media are available for process scale
chromatography. Please contact your local Bio-Rad representative.
22
Page 16
Section 8 References
P. Gagnon et al., A Ceramic Hydroxyapatite-Based Purification Platform, BioProcess International 4 (2), 50–60 (2006)
T. Ogawa et al., Effect of pH on Gradient Elution of Proteins on Two Types of Macro-Prep Ceramic Hydroxapatite, Prep Tech ’95, Industrial Separation Science Conference, East Rutherford, NJ (1995)
S.R. Shepard et al., Discoloration of ceramic hydroxyapatite used for protein chromatography, J Chromatography A 891, 93–98 (2000)
E. Dolinski et al., Purification of a Fusion Protein by Ceramic Hydroxyapatite Chromatography, Second International Conferences on Hydroxyapatite and Related Products, San Francisco, Ca (2001)
M. Gorbunoff et al., The Interaction of Proteins with Hydroxyapatite I: Role of Protein Charge and Structure, Analytical Biochemistry136, 425–432 (1984)
24
Page 17
Loading...