Bio-Rad Bio-Plex Assay Builder User Manual

Page 1
Bio-Rad Laboratories, Inc.
2000 Alfred Nobel Dr. Hercules, CA 94547 USA 1-800-424-6723 (in the US)
Bio-Plex Phospho-Histone H3
TM
Bio Plex
-
Phoshoprotein Detection
Instruction Manual
Now Includes Protocols for
Lysate Preparation!
4110018 Rev C 0406
For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723)
Page 2
Table of Contents
Section 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Section 2 Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Section 3 Required and Recommended Materials . . . . .3
Section 4 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Section 5 Lysate Preparation . . . . . . . . . . . . . . . . . . . . . 6
Section 6 Assay Instructions . . . . . . . . . . . . . . . . . . . . . 9
Section 7 Data Acquisition . . . . . . . . . . . . . . . . . . . . . . 15
Section 8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . 20
Section 9 Safety Considerations . . . . . . . . . . . . . . . . . .23
Section 10 References . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Page 3
Section 1 Introduction
Bio-Plex phosphoprotein assays and Bio-Plex total target assays are bead-based multiplex assays (Luminex xMAP technology) that detect the phosphorylation of proteins in lysates derived from cell culture or tissue samples. These 96-well plate-format assays allow profiling of the specific phosphorylation state of up to 100 different proteins using as few as two wells and as little as 25 µl of lysate per well. The Bio-Plex total target assay reports the abundance of the target protein in one well, while the Bio-Plex phosphoprotein assay reports the level of phosphorylation of that protein in a separate well. These instructions apply to both assays. For a current list of all Bio-Plex phosphoproteins and total target assays, visit www.bio-rad.com/products/phosphoproteins.
Available as Singleplex or Premixed Multiplex Assays
Bio-Plex phosphoprotein and total target assays are available as singleplex or premixed multiplex assays. Singleplex assays are designed to be flexible. They can be used individually to test for a single phosphoprotein at a time, or they can be combined to create a multiplex assay to test for a specific set of phosphoproteins in a single sample. Premixed multiplex assays are the more convenient format for repeat testing of a specific set of phosphoproteins. Both the coupled beads and the detection antibodies are premixed and quality tested at Bio-Rad. These assays are only available through Bio-Rad’s online x-Plex assay service (
For research use only. Not for diagnostic procedures.
www.bio-rad.com/bio-plex/x-plex/).
1
Page 4
Section 2
metI noitamrofnIgniredrO
syassAnietorpohpsohPxelP-oiB
tadnuofebnacsyassafotsiltnerruC
/snietorpohpsohp/stcudorp/moc.dar-oib.www
syassAtegraTlatoTxelP-oiB
tadnuofebnacsyassafotsiltnerruC
/snietorpohpsohp/stcudorp/moc.dar-oib.www
*tiKsisyLlleCxelP-oiB
)yassAlleW-69enO(110403-171#golatacdaR-oiB
)syassAlleW-69neT(210403-171#golatacdaR-oiB
nietorpohpsohPxelP-oiB
*tiKtnegaeRnoitceteD
)yassAlleW-69enO(400403-171#golatacdaR-oiB
)syassAlleW-69neT(500403-171#golatacdaR-oiB
slairetaMrehtO
dellitsidelirets,spittepipdnastepiP
tnebrosba,liofmunimula,retaw
egufirtnecorcimlm5.1,slewotrepap
sebuterutluclm51,sebut
Principle
The principle of these 96-well plate-format, bead-based assays is similar to a capture sandwich immunoassay. An antibody directed against the desired target protein is covalently coupled to internally dyed beads. The coupled beads are allowed to react with a lysate sample containing target protein. After a series of washes to remove unbound protein, a biotinylated detection antibody specific for a different epitope is added to the reaction. The result is the formation of a sandwich of antibodies around the target protein. Streptavidin-phycoerythrin (streptavidin-PE) is then added to bind to the biotinylated detection antibodies on the bead surface.
Data from the reaction are then acquired using the Bio-Plex suspension array system (or Luminex 100 system), a dual-laser, flow-based microplate reader system. The contents of the well are drawn up into the reader. The lasers and associated optics detect the internal fluorescence of the individual dyed beads as well as the fluorescent signal on the bead surface. This identifies each assay and reports the level of target protein in the well. Intensity of fluorescence detected on the beads indicate the relative quantity of targeted proteins. A high-speed digital processor efficiently manages the data output, which is further analyzed and presented as fluorescence intensity on Bio-Plex Manager accompanying software package. If specific wells are identified for comparison, the ratio of fluorescence intensity between those wells is automatically calculated.
TM
Software, the
Section 3 Required and Recommended Materials
Required Materials
The following are required for phosphoprotein detection: 1) Bio-Plex phosphoprotein or total target assays to test for specific target proteins in lysate samples, 2) a Bio-Plex cell lysis kit to optimally lyse cell culture or tissue samples, and 3) a Bio-Plex phosphoprotein detection reagent kit in order to prepare the assays and acquire data on the instrument.
1
2
3
Bio-Plex phosphoprotein and total target assays feature CST antibodies, exclusively developed for Bio-Rad.
2
*Buffers contained in this kit have not been optimized for the use with Bio-Plex cytokine assays.
3
Page 5
Recommended Materials
metI noitamrofnIgniredrO
metsySyarrAnoisnepsuSxelP-oiB
)metsySxenimuLro(
)FTHsedulcni(502000-171#golatacdaR-oiB
tiKnoitadilaVxelP-oiB 100302-171#golatacdaR-oiB
tiKnoitarbilaCxelP-oiB 060302-171#golatacdaR-oiB
CD ItiKyassAnietorP 1110-005#golatacdaR-oiB
rekahSetalPretitorciM
4rofrekahs4-STMrelttuhcS-AKI
5264ledoMeniL-baLrosetalporcim
elbapac,tnelaviuqero(rekahSetalP
)mpr001,1–003fo
0008023#golatacAKI
sutarappAmuucaVetalPretliF
muucavneercSitluMeropilliM ™muruAdaR-oiBrodlofinam
dlofinammuucav
TNATROPMI etalpretliffoesuehT:
enoehtnahtrehtosdlofinam
dehsinimidnitluseryamdeificeps
rof6noitcesees;ecnamrofrepyassa
yassasihtotcificepssnoitcurtsni
R0690MVAM#golataceropilliM
0746-237#golatacdaR-oiB
rexetroV
rexetrov-inimdnarbRWV
121-61885#golatacRWV
riovreseRtnegaeR
tnegaerlm05ratsoC.cnI,gninroC
0784riovreser
2784-422#golatacdaR-oiB
ediroulFlynofluslyhtemlynehP
)FSMP(
6267P#golatacamgiS
edixofluSlyhtemiD )OSMD( 0562D#golatacamgiS
rednirGeussiT
setnoK
2000-000588TK#golatacRWV
tnenopmoC egarotS
syassategratlatotdnanietorpohpsohpxelP-oiB
sdaebdelpuoC )x05(
tonoD.kradehtniCº4taerotS
.ezeerf
seidobitnanoitceteD )x52( .Cº4taerotS
setasyL
)lm/gµ002=noitartnecnocnietorp(
nezorfebnacsetasyL.Cº02–taerotS
.semit5otpudewahtdna
tiksisylllecxelP-oiB
reffubhsawlleC .Cº4taerotS
reffubsisyllleC .Cº4taerotS
1rotcaf,reffubsisyllleC )x052(
nezorfebnacsrotcaF.Cº02–taerotS
.semit5otpudewahtdna
2rotcaf,reffubsisyllleC )x005(
nezorfebnacsrotcaF.Cº02–taerotS
.semit5otpudewahtdna
tiktnegaernoitcetednietorpohpsohpxelP-oiB
reffubyassA .Cº4taerotS
reffubhsaW .Cº4taerotS
tneulidydobitnanoitceteD .Cº4taerotS
reffubnoisnepsuseR .Cº4taerotS
EP-nidivatpertS )x001( .kradehtniCº4taerotS
For optimal results, the use of the items listed below is recommended.
4
Section 4 Storage
Store the individual components as specified. Note that lysates are shipped and stored separately from the coupled beads and detection antibodies. Factors are shipped and stored separately from the cell wash and cell lysis buffers.
5
Page 6
Section 5
erutluC
lessev
muidemerutluC
emulov
noitulosgnisyL
emulov
setoN
etalpllew-69 llew/lµ001 llew/lµ57 otsllecworG
ecneulfnoc%58–08
gnivaeldnemmoceR ytpmesllewlanretxe
tceffeegdeoteud
hsiderutlucmc01 lm01 lm3–2 otsllecworG
ecneulfnoc%09–08
Lysate Preparation
3. Prepare an adequate volume of lysing solution (refer to the table on the left). For 10 ml of lysing solution, add 40 µl of 20 µl of and set aside on ice. Then add 40 µl of 500 mM PMSF.
factor 2
to 9.9 ml of
cell lysis buffer
factor 1
. Vortex gently to mix
and
This section provides instructions for preparing lysates derived from cell culture and tissure samples. For optimal recovery and sensitivity with the phospho-Histone H3 assay, refer to suggested protocol for lysate preparation.
1. Rinse the samples with
Adherent Cells — Stop the treatment reaction by aspirating the culture medium and quickly rinsing the cells with ice-cold cell wash buffer. The volume of cell wash buffer required is the same as the volume of aspirated cell culture medium. Keep the cells on ice.
Suspension Cells — Stop the treatment reaction by adding ice­cold wash buffer to the cells. The volume of cell wash buffer required is twice that of the culture medium. Centrifuge the cells at 1,000 rpm for 5 min at 4ºC. Aspirate the supernatant.
Tissue Samples — Rinse the tissue sample with cell wash buffer once. Cut the tissue into 3 x 3 mm pieces and transfer them to a 2 ml tissue grinder.
2. Prepare 500 mM PMSF by dissolving 0.436 g PMSF in 5 ml DMSO. Store as 0.5 ml aliquots at –20ºC. Aliquots can be frozen and thawed up to 5 times.
Lysing Solution Volume Guide
cell wash buffer
as follows:
4. Lyse the samples:
Adherent and Suspension Cells
a) Immediately add the lysing solution to the cells. The amount of
lysing solution needed depends on the cell concentration in the culture vessel (see table on the left).
b) Agitate the cells as follows:
Culture Plate — For suspension cells, place the plate on ice and pipet the contents of the wells up and down 5 times. For adherent cells, scrape the cells with a cell scraper. For both, agitate the plate on a microplate shaker at 300 rpm for 20 min at 4ºC.
Other Culture Vessel — Transfer the cell lysate to a centrifuge tube and rotate for 20 min at 4ºC.
HINT: Freeze-thawing the lysate once using dry ice or a –20ºC freezer may increase the extent of the lysis. Alternatively, briefly sonicate (eg., with a Sonifier 450 as follows: Duty cycle = 40, Output = 1, Pulse sonicating = 18 times).
c) Centrifuge the samples at 4,500 g for 20 min at 4ºC.
Tissue Samples
a) Immediately add 500 µl of lysing solution to the tissue grinder
and grind the tissue sample on ice using about 20 strokes.
b) Transfer the ground tissue to a clean microcentrifuge tube
and freeze the sample at –70ºC.
c) Thaw the samples, then sonicate on ice as suggested above.
6
d) Centrifuge the samples at 4,500 g for 4 min.
5. Collect the supernatant without disturbing the pellet.
6. Determine the lysate protein concentration. The protein concentration should be 200–900 µg/ml. It may be necessary to test-lyse your samples with different volumes of lysing solution to obtain the specified protein concentration range.
7
Page 7
7. Add an equal volume of
8. If the lysate is not tested immediately, store at –20ºC. The lysate is stable for up to 5 freeze-thaw cycles.
Suggested protocol for lysate preparation of Histone H3 assay:
1. Follow steps 1–3 above
4. Lyse the samples:
a) Immediately add the lysing solution to the cells. The amount of
lysing solution needed depends on the cell concentration in the culture vessel (see table on the left).
b) Briefly sonicate (e.g., with a Sonifer 450 as follows:
Duty cycle = 40, Output = 1, Pulse sonicating = two 10 min pulses with a 1 min break in between).
c) Agitate the cells. Transfer the cell lysate to a centrifuge tube and
rotate for 20 min at 4
d) Centrifuge the samples at 4,500 g for 20 min at 4ºC.
5. Collect the supernatant without disturbing the pellet.
6. Determine the lysate protein concentration. The protein concentration should be 200–900 µg/ml. It may be necessary to test-lyse your samples with different volumes of lysing solution to obtain the specified protein concentration range.
7. Add equal volume of assay buffer to the lysate.
8. Freeze (overnight) at -20ºC and thaw before testing.
assay buffer
.
ºC.
to the lysate.
Section 6 Assay Instructions
The following instructions apply to Bio-Plex phosphoprotein and total target singleplex, custom-premixed, and mutliplex assays. Do not mix phosphoprotein assays with its corresponding total target assays (e.g. phospho-Akt and total Akt).
Plan Experiment
1. Assign which wells of a 96-well plate will be used for each lysate (see the example below). Keep in mind that the instrument reads wells down the plate and not across. Consider assigning the wells vertically. A pullout worksheet has been provided in this manual that may be used as a reference during the different assay steps.
Example Plate
2. Determine the total number of wells that will be used in the assay. Include a 25% excess (or add 2 wells for every 8 wells used) to ensure that enough diluted coupled beads, detection antibodies, and streptavidin-PE are prepared. Record these numbers on the worksheet since they will be referenced throughout the assay.
8
9
Page 8
yassA emuloV
xelpelgniS xelpelgnishcaemorflµ1
xelptilumdeximerP
xelpitlumdeximerphcaemorflµ1
dnaxelpelgniS
xelpitlumdeximerp
dnaxelpelgnishcaemorflµ1
xelpitlumdeximerphcaemorflµ1
tamroFyassA deximerPdnaxelpelgniS xelpitluM
forebmuN
sllew
:yassarofdesU 81 sllew
:ssecxe%52edulcnI
81 =%52.1xsllew 32
=sllewforebmunlatoT sllew32
:yassarofdesU 42 sllew
:ssecxe%52edulcnI
42 =%52.1xsllew 03
=sllewforebmunlatoT sllew03
foemuloV
sdaebdelpuoc
)x05(
:syassadeximerP
=xelp-3 32
)83p,a-BkI,2KRE(
:syassaxelpelgniS
03=2KRE
a-BkI
= 03
03=83p
=latoT 09
foemuloV
sdaebdelpuoc
)x1(
sllew32 =lµ05x 0511 sllew03 =lµ05x 0051
hsawfoemuloV
reffub
0511 32 = 7211 0051 09 = 0141
Thaw Lysates
1. Retrieve the experiment lysates and lysates shipped with the assays from –20ºC storage. These lysates were prepared using the protocol in section 5 and contain 50% assay buffer.
Prepare Coupled Beads
Protect the beads from light by covering the tubes with aluminum foil. Keep all tubes on ice until ready to use. Coupled beads must be mixed manually prior to use when combining singleplex or premixed assays.
NOTE: Refer to the table provided with the lysate packaging to identify which lysates shipped with the assays (visit www rad.com/products/phosphoproteins/ to download the PDF). Select the treated and untreated lysates from the table, which is used to determine the assay performance of each Bio-Plex phosphoprotein and total target assay. These should not be considered as references. Instead, activation signals and ratios should be based on experimental control lysates. For determining total protein
DC
concentrations, consider Bio-Rad’s
protein assay kit (Bio-Rad
catalog #500-0112).
2. Thaw the lysates at room temperature and then place them on ice.
3. If necessary, it is possible to further dilute the lysates. Lysing solution freshly prepared (as specified in section 5) and assay buffer are required. Use a 1:1 mixture of lysing solution and
assay buffer
further dilute the lysate.
10
.bio-
1. Vortex the coupled beads (50x) at medium speed for 5 sec.
2. Prepare a sufficient volume of coupled beads (1x) using
wash buffer
. When preparing a multiplex assay, use equal volumes of each bead (see sample below). Each well requires 1 µl of coupled beads (50x) for each target adjusted to a final volume of 50 µl (refer to the table below). These calculations can be done on the worksheet.
Volume of Coupled Beads (50x) in Each Well
to
Example Coupled Bead Calculations
11
Page 9
yassA emuloV
xelpelgniS xelpelgnishcaemorflµ1
xelptilumdeximerP
xelpitlumdeximerphcaemorflµ1
dnaxelpelgniS
xelpitlumdeximerp
dnaxelpelgnishcaemorflµ1
xelpitlumdeximerphcaemorflµ1
tamroFyassA deximerPdnaxelpelgniS xelpitluM
forebmuN
sllew
:yassarofdesU 81 sllew
:ssecxe%52edulcnI
81 =%52.1xsllew 32
=sllewforebmunlatoT sllew32
:yassarofdesU 42 sllew
:ssecxe%52edulcnI
42 =%52.1xsllew 03
=sllewforebmunlatoT sllew03
foemuloV
noitceted
)x52(seidobitna
:syassadeximerP
=xelp-3 32
)83p,a-BkI,2KRE(
:syassaxelpelgniS
03=2KRE
a-BkI
= 03
03=83p
=latoT 09
foemuloV
noitceted
)x1(seidobitna
sllew32 =lµ52x 575 sllew03 =lµ52x 057
foemuloV
noitceted
ydobitna
tneulid
575 32 = 255 057 09 = 066
Calibrate Vacuum Apparatus
mreT snoitceriDdeliateD
hsaW
folµ001ddA reffubhsaw anoetalpretlifehtecalP.llewhcaeot
muucavybreffubehtevomerdnasutarappamuucavdetarbilac
.lewotrepapnaelcahtiwetalpretlifehtfomottobehttolB.noitartlif
.deificepssataepeR
esniR
folµ001ddA reffubnoisnepsuser etalpretlifehtecalP.llewhcaeot
muucavybreffubehtevomerdnasutarappamuucavdetarbilacano
.lewotrepapnaelcahtiwetalpretlifehtfomottobehttolB.noitartlif
.deificepssataepeR
etabucnI
retlifehtecalP.epatgnilaesfoteehswenahtiwetalpretlifehtrevoC
.liofmunimulahtiwrevocnehtdnarekahsetalporcimanoetalp
,ces03rofmpr001,1taerutarepmetmoortaetalpretlifehtekahS
.emitnoitabucnideificepsehtrofmpr003taneht
muucaV
retlif-
evomerdnasutarappamuucavdetarbilacanoetalpretlifehtecalP
htiwetalpretlifehtfomottobehttolB.noitartlifmuucavybreffubeht
.lewotrepapnaelca
The vacuum apparatus must be calibrated at the beginning of the assay to ensure an optimal bead yield. For more detailed instructions, refer to the Bio-Plex suspension array system hardware instruction manual.
1. Prewet all the wells of a 96-well filter plate with 100 µl of wash buffer.
2. Place the filter plate on the vacuum apparatus and turn on the vacuum to the maximum level.
3. Press on the filter plate and note the time required to remove the buffer from the wells by vacuum filtration. The evacuation time should be 2–5 sec.
If the evacuation time is <2 sec, the pressure is too high. Open the vacuum control valve slightly and repeat steps 1–3.
If the evacuation time is >5 sec, the pressure is too low. Close the vacuum control valve slightly and repeat steps 1–3.
Assay Key
The following terms are repeated throughout the assay procedure. Refer to these detailed instructions when
wash, rinse, incubate, and
vacuum-filter are shown in bold.
Assay Procedure
Bring all buffers to room temperature. Avoid bubbles when pipetting.
Wash the desired number of wells in a 96-well filter plate. If fewer
1. than 96 wells are required, cover the unused wells with sealing tape for later use.
2. Vortex the coupled beads (1x) for 5 sec at medium speed. Add 50 µl to each well and immediately
3.
Wash twice.
vacuum-filter.
4. Vortex the thawed lysates gently for 3 sec. Add 50 µl of lysate to each well, changing the pipet tip after every volume transfer.
Incubate for 15–18 hr (or overnight).
Volume of Detection Antibodies (25x) in Each Well
Example Detection Antibody Calculations
12
13
Page 10
5. The next day, prepare a sufficient volume of detection antibodies (1x)
yassA tegraTlatoTdnanietorpohpsohP
sllewforebmunlatoT =32+03 35
)x001(EP-ASfoemuloV =lµ5.0x35 5.62
)x1(EP-ASfoemuloV =lµ05x35 0562
reffubhsawfoemuloV =lµ5.62–lµ0562 5.3262
using
detection antibody diluent
page). Each well requires 1µl of detection antibodies (25x) for each target adjusted to a final volume of 25 µl (refer to the table on the previous page).
6. After the incubation, slowly remove and discard the sealing tape,
7.
8. Vortex the detection antibodies gently and add 25 µl to each well,
9. After the incubation, slowly remove and discard the sealing tape,
10.
11. Keep the plate in the dark and prepare a sufficient volume of
Example Streptavidin-PE Calculations
vacuum-filter.
then
Wash 3 times.
changing the pipet tip after every volume transfer.
then
vacuum-filter.
Wash 3 times.
streptavidin-PE (1x) using well requires 0.5 µl of streptavidin-PE (100x) adjusted to a final volume of 50 µl. Store in the dark after preparation.
(see the example on the previous
Incubate for 30 min.
wash buffer
(see example below). Each
Section 7 Data Acquisition
Recommendations for acquiring data using the Bio-Plex suspension array system are listed below. Alternatively, refer to the Bio-Plex Manager software user guide or the instructions provided with the Luminex 100 instrument.
Prepare System
1. Empty the waste bottle and fill the sheath fluid bottle before starting. This will prevent fluidic system backup and potential data loss.
2. Turn on the reader and microplate platform (and HTF if present). Allow the system to warm up for 30 min.
3. Select Startup and follow the instructions to prepare the reader to acquire data. If the system is idle for 4 hr the lasers will automatically turn off and a 30 min warm-up period will again be required prior to acquiring data. Select Warm up and wait for the optics to reach operational temperature.
Calibrate With High RP1 Target Value
Calibrate using Bio-Plex calibration beads and target values. Daily calibration is recommended before acquiring data.
1. Select Calibrate and confirm that the default values for CAL1 and CAL2 are the same as the values on the Bio-Plex calibration bead labels. Use the Bio-Plex High RP1 target value for CAL2 calibration for Bio-Plex phosphoprotein and total target assays.
12. Vortex the diluted streptavidin-PE vigorously and add 50 µl to each
13. After the incubation, slowly remove and discard the sealing tape,
14.
15. Add 125 µl of
14
Incubate for 10 min.
well.
then
vacuum-filter.
Rinse 3 times.
If the data is not acquired immediately, the assay may be stored in the dark at 4ºC for up to 24 hrs.
resuspension buffer
to each well. Incubate for 30 sec.
NOTE: When acquiring data for Bio-Plex phosphoprotein or total target assays with a Luminex 100, Luminex Data Collector software, and Luminex calibration beads, it is necessary to convert the Luminex CAL2 calibration bead RP1 target value using the following equation:
Bio-Plex High RP1 target value = (Luminex RP1 target value) x 4.55
Add the new target value to the Luminex software by selecting Calibrate, then New under the Reporter Channel in the Start Calibration dialog. Enter the new target value and save it as a new lot. Then calibrate using the new RP1 target value.
15
Page 11
2. Select OK and follow the instructions for CAL1 and CAL 2 calibration.
Prepare Protocol
1. Open a new protocol by selecting File, then New from the main menu. Locate the steps at the left of the protocol menu.
2. Select Step 1 (Describe Protocol) and enter information about the assay.
3. Select Step 2 (Select Analytes) and select the panel for Phosphoproteins or Total Targets. Choose the target proteins for all the assays on the plate. If both phosphoprotein and total target assays are run on the same plate, two separate protocols must be entered.
4. Select Step 3 (Format Plate) and click on the Plate Formatting tab. Click on and drag the cursor over all the wells that contain lysates.
Plate Formatting Example
5. Then select the Plate Groupings tab to display the plate grouping tools.
a) Select Group and drag the cursor across all the wells to
define each assay group (exclude the ones that contain lysates provided with the assays).
b) The first well of the group is automatically assigned as the
Reference well and the remaining wells in the group are Member wells. To change the Reference well, select Reference and click on the new Reference well.
c) Select Member/Reference or Reference/Member from the Ratio
pull-down list. When selecting the more common Member/Reference option, the ratio of fluorescence intensity of each well will be calculated against the fluorescence intensity of the Reference well. This value will be reported as a ratio in the results file.
Plate Grouping Example
16
6. Select Step 6 (Enter Sample Info) and enter sample information. This is the location where the wells are identified as containing either Bio-Plex phosphoprotein or total target assays (see the example on the right).
17
Page 12
Sample Information Example
Acquire Data
1. Shake the assay plate at 1,100 rpm for 30 sec immediately before acquiring data. Failure to do so will result in increased data acquisition time due to bead settling.
2. Check that the filter plate is flat. While pressing on one end of the plate, observe the distance that the opposite end of the plate is raised off a flat surface. If the distance is >1 mm, transfer all contents to a flat-bottom 96-well plate or another filter plate.
3. Visually inspect the plate and ensure that the assay wells are filled with buffer prior to placing the plate in the Bio-Plex microplate platform.
6. If acquiring data from more than one plate, empty the waste bottle and refill the sheath bottle after each plate. Select Wash Between Plates and follow the instructions for fluidics maintenance. Then repeat the Prepare Protocol and Acquire Data steps.
7. When data acquisition is complete, select Shut Down and follow the instructions.
Reacquire Data
It is possible to acquire data from a well or plate a second time using the Rerun/Recovery mode located below Start in Step 7 (Run Protocol).
1. Check the wells where data will be acquired a second time. Any previous data will be overwritten.
2. Remove the buffer by vacuum filtration and add 125 µl of resuspension buffer to each well. Cover the filter plate with a new sheet of sealing tape.
3. Repeat
Acquire Data steps 1–6 to acquire data a second time.
The data acquired should be similar to the data acquired initially; however, the data acquisition time will be extended since fewer beads are present in each well.
4. Slowly remove the sealing tape and any plate cover before placing the plate in the reader.
5. Select Step 7 (Run Protocol):
a) Specify data acquisition for
25 beads per region.
b) In Advanced Settings, confirm that the default DD gate values
are set to 4335 (low) and 10000 (high).
NOTE: When using a Luminex 100 instrument, set the gates according to the Luminex procedure located in the manual.
c) Select Start and save the .rbx file. Then follow the instructions for
data acquisition.
18
19
Page 13
Section 8 Troubleshooting
This troubleshooting section addresses problems that may be encountered with Bio-Plex phosphoprotein or total target assays. If the problems listed below are encountered, review the possible causes and solutions provided. This will assist in resolving problems directly related to the assay. Use the Bio-Plex validation kit to validate all the key functions of the array reader and assist in determining whether or not the array reader is functioning properly.
Possible Causes Possible Solutions
Filter Plate Leakage
Vacuum setting too high
Low Signal (Good Signal From Lysates Provided with the Assays but Weak or No Signal From Experiment Lysates)
Protein concentration in lysate too low or too high
Low Signal (Weak or No Signal From Lysates Provided with the Assays and Experiment Lysates)
Detection antibody and/or streptavidin-PE diluted incorrectly
This could result in tearing of the filter. Confirm that the vacuum pressure is set as specified in the vacuum calibration procedure section. Also refer to the Vacuum Manifold Setup in the Bio-Plex suspension array system hardware instruction manual. Use the recommended filter plate vacuum apparatus
Verify the protein concentration in the cell lysate samples. Adjust the amount of lysing solution used in the lysate preparation to achieve an optimal protein concentration of 200–900 µg/ml prior to adding an equal part of assay buffer
Check the calculations and be careful to add the correct volumes for dilution
Possible Causes Possible Solutions
Expired beads, detection antibody,
Use new or unexpired components
and/or streptavidin-PE used
Incorrect incubation temperature used during incubation steps
Incubation time insufficient
Incubations should be at room temperature (20–22ºC)
Adhere to the recommended incubation times
Low Bead Count
Cell debris in lysate not cleared
Remove the cellular debris by centrifugation at 4,500 g for 20 min at 4ºC. Avoid disturbing the pellet while collecting the supernatant
Resuspension buffer not used after streptavidin-PE incubation
Rinse the beads 3 times with resuspension buffer after the streptavidin-PE incubation step
Vacuum setting too high
This results in bead loss. Calibrate the vacuum apparatus as specified
Filter plate not shaken enough before each incubation step and prior to data acquisition
Reader clogged
Shake the filter plate at 1,100 rpm for 30 sec before each incubation and immediately before acquiring data
Refer to the troubleshooting guide in the Bio-Plex hardware instruction manual
High Coefficient of Variation (CV)
Plate sealer reused
This could result in contamination. Use a new sheet of sealing tape for each incubation
Buffer not completely filtered from wells
Be sure that the wells are filtered completely and that no residual volume remains
20
21
Page 14
Possible Causes Possible Solutions
Contamination with wash buffer during wash steps
Microplate shaker set to an incorrect speed
Cellular debris not cleared from lysate
During the wash steps, do not splash wash buffer from one well to another. Filter the wells completely so that no residual volume remains. Also, be sure that the microplate shaker setting is not too high. Reduce the microplate shaker speed to minimize splashing
Check the microplate shaker speed and use the recommended setting. Setting the speed too high may cause splashing and contamination. Use the recommended plate shaker
Be sure to remove cellular debris by centrifugation as directed. Avoid disturbing the pellet while collecting the supernatant
Section 9 Safety Considerations
Eye protection and gloves are recommended while using this product. Consult the MSDS for additional information.
Human Source Material. Treat As Potentially Infectious.
The lysates provided with Bio-Plex phosphoprotein and total target assay contain components of human origin. The components are known to contain an agent that requires handling at Biosafety Level 2 containment [US Government Publication: Biosafety in Microbiological and Biomedical Laboratories (CDC, 1999)]. These agents have been associated with human disease. These components have not been screened for hepatitis B, human immunodeficiency viruses, or other adventitious agents. Handle Bio-Plex phosphoprotein positive and negative controls as potentially biohazardous material under at least Biosafety Level 2 containment.
Bead resuspension buffer not used after streptavidin-PE incubation
High Background Signal (From Both Lysates Provided with the Assays and Experiment Lysates)
Vacuum pressure too low, resulting in residue in wells
Wash steps performed incorrectly
Streptavidin-PE incubation step too long
22
Wash 3 times with bead resuspension buffer after streptavidin-PE incubation as described in the assay instructions
Use recommended filter plate vacuum apparatus with proper vacuum pressure setting
Perform washes as described in the assay instructions
Check suggested incubation times for appropriate steps of the assay. Follow the suggested time for incubation
23
Page 15
Section 10 References
Fulton R, McDade R, Smith P, Kienker L, and Kettman J Jr, Advanced multiplexed analysis with the FlowMetrix system, Clin Chem 43, 1749–1756 (1997)
Chang L, and Karin M, Mammalian MAP kinase signalling cascades, Nature 410, 37–40 (2001)
For a complete list of publications using Bio-Plex Phosphoprotein Detection Assays, refer to bulletin 5394.
xMAP is a trademark of Luminex Corp.
By purchasing this kit, which contains fluorescent labeled microsphere beads authorized by Luminex, you, the customer, acquire the right under Luminex's patent rights* to use this kit or any portion of this kit, including without limitation the microsphere beads contained herein, only with Luminex’s laser-based fluorescent analytical test instrumentation known under the name of Luminex 100, for example as marketed by Bio-Rad Laboratories, Inc. in the Bio-Plex system.
*Including, but not limited to US patent 5,981,180; 6,046,807; 6,057,107.
24
Loading...