Bio-Rad Bio-Beads S-X Media User Manual

Bio-Beads®S-X Beads
Gel Permeation Chromatography
Instruction Manual
Table of Contents
Page
Technical Description........................................ 1
Mechanism.......................................................... 2
Instructions for Use............................................ 3
Swelling the Beads.............................................. 3
Packing the Column........................................... 5
Collecting Fractions........................................... 8
Regeneration....................................................... 9
Applications........................................................10
Product Information..........................................21
Introduction
Bio-Beads S-X beads are a series of porous crosslinked polystyrene polymers used for gel permeation separations of lipophilic polymers and low molecular weight, hydrophobic materials in the presence of organic solvents. These non­aqueous spherical beads are used in much the same way aqueous gels are used, except that they are swollen with organic solvents during the separation.
Technical Description
Bio-Beads S-X beads are neutral, porous styrene divinylbenzene copolymer beads. The beads in the Bio­Beads S-X series have exclusion limits from 400 to 14,000 daltons. This range makes them particularly suitable for the fractionation and separation of low molecular weight organ­ic polymers and other hydrophobic substances. The amount of divinylbenzene crosslinkage determines the pore size, and hence the molecular weight exclusion limit of a particu­lar gel in this series. The beads are available with crosslink­ages from 1-12%. Pore dimensions and exclusion limits are also influenced by the eluant employed; maximal expansion
1
of the matrix is achieved with relatively nonpolar, aromatic solvents. The beads are typically used with benzene, toluene, xylene, carbon tetrachloride, and mixtures of sol­vents.
Mechanism
Gel filtration, also called gel permeation, is the mode of separation which occurs with Bio-Beads S-X beads. Large compounds, greater than the molecular exclusion limit, pass through the column unhindered, whereas small compounds, within the molecular weight operating range, will be retained in the column. The small compounds permeate the pores of the Bio-Beads S-X beads, and thus they take longer to pass through the column. This mechanism requires an eluant which is mobile, and, therefore, Bio-Beads S-X beads must always be used in a column mode.
2
Instructions for Use
Bio-Beads S-X beads are supplied dry, and must be swollen prior to packing into a chromatographic column. The general instructions are:
1. Swell the beads.
2. Assemble the column.
3. Pour the beads into the column.
4. Add the sample and proceed with the separation.
The instructions below describe swelling the beads and packing the column in details.
Swelling the Beads
Before use, swell the Bio-Beads S-X beads in an organ­ic solvent, such as:
Aromatics Methylene chloride Benzene Orthodichlorobenzene Carbon tetrachloride Perchloroethylene Dimethylformamide Tetrahydrofuran Ketones Trichlorobenzene
These organic solvents allow maximal swelling of the beads. If polar solvents, such as water or methanol, are used,
3
the Bio-Beads S-X beads will not swell, and the pore size will be minimal. The chosen solvent should be the one used for the separation, and the same as the solvent in which the sample is dissolved.
The solvents should be of highest quality available, and preferably redistilled if non-volatile matter is present in them. Some solvents, e.g. tetrahydrofuran, develop perox­ides on standing in contact with air. Solvents should gener­ally be degassed prior to use and protected from atmospheric contamination, to prevent later outgassing dur­ing the chromatographic run.
Bio-Beads S-X1 beads will swell considerably, and should be placed in at least six times the resin weight of sol­vent (w/w). The higher crosslinked resins will not swell as much, so they will not require as much elution solvent. The swelling should always be done in the presence of excess solvent to prevent the resin from drawing up all the solvent, and possibly not swelling fully. The higher crosslinked resins will require more time to become fully swollen.
Bio-Beads S-X12 and S-X8 beads may require swelling overnight, whereas the lower crosslinked resins will be fully swollen in a few hours. Complete swelling is necessary to prevent swelling after packing, which could break the col-
4
umn. If the amount of swelling is unknown, it can be checked by swelling a known weight of beads and measur­ing the volume.
After the beads are fully swollen, they are packed into a chromatographic column and washed with the solvent in which they were swollen. Normally, the sample is dissolved and the elution is performed with this same solvent, to pre­vent swelling or shrinking of the resin during the run. If the beads swell during the run, a glass column may break.
When the beads are used for the first time, low molecu­lar weight polystyrene trapped inside the pores of the beads will tend to cause slow column equilibration. Although swelling the beads in one of the solvents listed above will remove some of the low molecular weight polystyrene, sev­eral column volumes of the running buffer may be necessary to reach baseline equilibration.
Packing the Column
Metal columns are often used in gel permeation chromatography, though glass columns offer the important advantages of visibility of packing and therefore are better suited.
5
Put together a clean column assembly. Place a small amount of elution solvent in the column to prevent bubble formation at the base of the poured column packing. Prepare a solvent reservoir, which should be placed at an elevation higher than the top of the column. The solvent reservoir may be connected to the column by tubing. Any type of connec­tion must be air-tight and clean.
The Bio-Beads S-X beads should be placed in an approximately 50% slurry of beads and elution solvent. A good practice at this point is to degas this mixture by soni­cating it under vacuum. Safety precautions should be observed. Double check that the stopcock below the column is closed. Swirl the slurry to get a homogeneous mixture, then slowly and consistently pour the slurry into the column. It is best to pour a long narrow column in sections, in order to obtain a uniform and reproducible column. A glass rod may be used to facilitate pouring the mixture down the side of the column. This will also help eliminate the formation of bubbles. When it is clear that a few centimeters of packed bed have settled, open the stopcock below the column and allow the mobile phase to begin flowing. Begin with a slow flow rate and gradually increase to 10-15% above the final operating flow rate, if possible. The purpose of this elevated flow rate is to pack the column under more pressure than
6
will be used during the separation. (This is not necessary for Bio-Beads S-X1 beads.) Maintain at least a few centimeters of liquid above the resin bed.
Never allow the packed beads to become dry, because this will cause air pockets and channeling within the bed, resulting in poor efficiency and low resolution. If a small column does go dry, add extra solvent, cap both ends, invert several times until the resin is fully slurried, allow the resin bed to settle, and begin flow as normal. An alternative technique, for larger columns, is to backwash the resin. This will resuspend the beads and allow repacking. This technique should only be used if the packed bed fills less than half the column. Simply connect a piece of tubing to the column outlet, and, at a very slow flow rate, flow the elu­tion solvent into the column. This will resuspend the packing material. Care must be taken that enough space is available for the resuspended material, as well as the additional solvent. When the material is fully suspended, and the air pockets have risen above the packing material, stop the flow. Allow the mate­rial to settle.
Note: Bio-Beads S-X beads will float in high density solvents, e.g. chloroform. First swell the media in tetrahydrofuran and pack the column as previously described. Insert an adjustable flow adaptor into the column to constrain the beads in a packed
7
position. Then, displace the tetrahydrofuran with 3 bed volumes of chloroform.
Flow Rate
Bio-Beads S-X beads have the capacity for different flow rates, depending on the crosslinkage. The 1% and 2% crosslinked resins (Bio-Beads S-X1 beads) are very soft when fully swollen and should only be used in gravity flow proce­dures. Bio-Beads S-X3 beads can withstand 5 ml/min with a backpressure of 300 psi. Bio-Beads S-X8 and S-X12 beads can withstand up to 5,000 psi backpressure.
Collecting Fractions
Separated compounds are often collected for further analy­sis. Fractions can be constantly collected, for example in 2 ml increments, or they can be collected specifically when the com­pound of interest is eluting. In order to know when the com­pounds are eluting from the column, pass the eluant through a detector. Ultraviolet (UV) detectors are commonly used, though a preparative cell may be necessary so that the flow is not restricted too much. Use the information from the UV chro-
8
matogram to establish the time periods to use when collecting a specific fraction.
Regeneration
Regeneration of Bio-Beads S-X beads may be necessary if compounds have become trapped within the pores of the resin, for example if a series of eluants has been used. Bio-Beads S-X beads are hydrophobic, and can also absorb compounds. To wash the resin, swell it to its maximum with a solvent such as methylene chloride, toluene, or tetrahydrofuran.
After long periods of storage, small loose polystyrene which had been trapped inside the beads will leach out. This polystryrene (2-10 chain length) has a very high extinction coef­ficient, so even trace amounts will have a high UV absorption. To eliminate this problem, wash with several bed volumes of solvent.
Chemical Resistance
Bio-Beads S-X beads are highly chemically resistant, though explosive mixtures may form in the presence of strong oxidizing reagents such as chromic acid, nitric acid, and hydro­gen peroxide. The beads may be autoclaved.
9
Applications
Harmon reviewed the many different types of gels that are available to the chromatographer, and reported that the relative­ly soft gels, such as the lower Bio-Beads (particularly S-X1 ) products swell appreciably in many solvents.1Remarkable sep­arations can therefore be achieved at low flow rates.
Many different types of compounds have been separated on Bio-Beads S-X beads. The beads have been used for analysis and quantification of pesticides Beads S-X beads are the basis of the official EPA procedure for the measurement of organic priority pollutants in sludge.8The beads are useful for separation of polycyclic aromatic com-
9-11
pounds,
assessment of tissue reaction to biomaterial, fractionation of halogenated environmental contaminants. Other applications include separation of tall oil components,
19, 20
lipids,
alkalines,21fatty acids,22a variety of hydrocarbons, and polystyrenes.24These beads can also be used to determine polymer molecular weights and molecular weight distribu-
24-28
tions.
They have been used for the separation of low molec­ular weight trimethylsilylated silicic acid,29and for the isolation of low molecular weight polar organics in fatty tissues for sub­sequent GLC-MA analysis.30They have also been used for the
2-5
10
and rodenticides.
6,7
12, 13
Bio-
and
14-17
18 23
analysis of fish lipid extracts,31and for fractionation of food grade poly (vinyl chloride) resin.
32, 33
A variety of lipophilic polymer substances has been suc-
cessfully purified on beds of Bio-Beads S-X beads (Figure
11
5
3
2
1
Relative solute concentration
4
7
6
1).
Relative effluent volume
Fig. 1. Separation of triglycerides and hydrocarbons on Bio-Beads S-X1 and S-X2 beads in benzene (two beds in series). 1 = tristearin; 2 = trimyristein; 3 = trilaurin; 4 = tri-
caprylin; 5 = tricaproin; 6 = hexadecane; 7 = undecane.
12
21
Bio-Beads S-X2 and S-X8 beads have been used in multi­column system for separating various components of tall oil, wood resin, and gum resin by gel permeation chromatography. Figure 2 shows the fractionation of tall oil with Bio-Beads S-X beads. All of the injected acid sample was fully recovered, and no delay in elution time was noticed.
Resin acid
Fatty acid dimer
Fatty acid
Resin acid dimer
Recorder response
90 100 110 120 130 ml
Elution volume
0.2 0.3 0.4 0.5 0.6 0.7
Fig. 2. Gel permeation chromatogram of tall oil.
13
18
Ault and Spurgeon used Bio-Beads S-X3 beads for specific separation of chlorinated pesticides form animal fat.
Aitzetmuller has used Bio-Beads S-X beads in benzene to fractionate a 1 g sample of polymeric trioleins. The large pore size of the gel effectively resolved dimeric triglycerides (m.w. ca 1,800) from monomeric triglycerides.
14
35
34
References
Reference Application
1. Harmon, D. J., Sep. Sci., 5, 403 (1970). Bead swelling in
2. Stalling, D. L., et al., J.A.O.A.C., 55, 32 (1972). Pesticides
3. Johnson, L. D., et al., J.A.O.A.C., 59, 174 Pesticides
(1976).
4. Steinwandter, H., Fresenius, Z. Anal. Chem., Pesticides
313, 536 (1982).
5. Blaha, J. J. and Jackson, P. J., J.A.O.A.C., Pesticides
68 (6), 1095 (1985).
6. Hunter, K., J. Chromatog., 299, 405 (1984). Rodenticides
7. Hunter, K., J. Chromatog., 321, 255 (1985). Rodenticides
8. Haile, C. L. and Lopez-Avila, V., U.S. Envi- Sludge ronmental Protection Agency, Project Summary No. 600/S4-84-001, March, 1984.
9. Friley, B. K., Phelps, J. B. and Kincaid, J. R., Polycyclic aromatics J. Chromatog., 258, 310 (1983).
10. Chamberlain, W. J., Snook, M. E., Baker, J. L. Polycyclic aromatics
and Chortyk, O. T., Anal. Chim. Acta, 222, 235 (1979).
11. Snook, M. E., Chamberlain, W. J., Severson, Polycyclic aromatics
R. F. and Chortyk, O. T., Anal. Chem., 47, 1155 (1975).
15
various solvents
Reference Application
12. Hood, C. I., Schoen, F. J., Coleman, S. E. and Tissue reaction
Mickley, L. D., J. Biomed. Materials Res., 18, 1031 (1984).
13. Schoen, F., et al., J. Biomed Materials Res., 20, Tissue reaction
709 (1986).
14. Stalling, D. L., Smith, L. M. and Petty, J. D., Halogenated ASTM Special Publication No. 686, page 302 environmental (1979). contaminants
15. Musial, C. and Uthe, J. F., J.A.O.A.C., 69 (3), Halogenated 462 (1986). environmental
16. Ault, J. A., Schofield, C. M., Johnson, L. D. Halogenated
and Waltz, R. H., J. Ag. and Food Chem., 27, environmental 825 (1979). contaminants
17. LeBel, G. and Williams, D. T., J.A.O.A.C., Halogenated 69 (3), 451 (1986). environmental
18. Chang, T-L., Anal. Chem., 40 (6), 989 (1968). Tall oil
19. Tipton, C. L. Paulis, J. W. and Pierson, M. D., Lipids J. Chromatog., 14, 486 (1964).
20. Hirsch, J., Colloq. Int. Centre Nat. Res. Sci., Lipids 99, 11 (1960).
21. Chang, T-L., Anal. Chim. Acta, 39, 519 (1967). Alkalines
22. Chang, T-L., Anal. Chim. Acta, 42, 51 (1968). Fatty acids
16
contaminants
contaminants
Reference Application
23. Henfrickson, J. G., J. Chromatog., 32, 543 Hydrocarbons (1968).
24. Cantow, M. J. R., et al., J. Polymer Sci., 5, 987 MW determination (1967)
25. Coll, H., Separ. Sci., 5, 273 (1970). MW determination
26. Pickett, H. E., et al., J. Applied Polymer Sci., MW determination 10, 917 (1966).
27. Cantow, M. J. R., et al., J. Polymer Sci., MW determination (Part C), 16, 13 (1967).
28. Cantow, M. J. R., et al., J. Polymer Sci., MW determination (Part A-1), 5, 1391 (1967).
29. Shimono, T., Toshiyuki, I. and Tarutani, T., Trimethysilylated J. Chromatog., 179, 323 (1979). silicic acid
30. Kuehl, D. H. and Leonard, E. N., Anal Chem., Polar organics 50, 182 (1978). in fatty tissues
31. Burns, B. G., et al., J.A.O.A.C., 64, 282 (1981). Fish lipid extracts
32. Gilbert, J., Shepherd, M. J. and Wallwork, Food grade PVC M. A., J. Chromatog., 320, 361 (1985). fractionation
33. Waliszewski, S. M. and Szymczynski, G. A., Food grade PVC J. Chromatog., 321, 480 (1985). fractionation
34. Ault, J. A. and Spurgeon, T. E., J.A.O.A.C., Chlorinated pesticides 67 (2), 284 (1984). from animal fat
35. Aitzetmuller, K., J. Chromatog., 71, 355 Triglycerides
(1972).
17
Product Information
Catalog Product Mesh Exclusion Operating ml/g Number Description Size Limit Range Benzene
152-2150 Bio-Beads S-X1 200-400 14,000 600-14,000 7.5
152-2151 Bio-Beads S-X1 200-400 14,000 600-14,000 7.5
152-2750 Bio-Beads S-X3 200-400 2,000 up to 2,000 4.75
152-3350 Bio-Beads S-X8 200-400 1,000 up to 1,000 3.1
152-3650 Bio-Beads S-X12200-400 400 up to 400 2.5
Beads, 100 g
Beads, 1 kg
Beads, 100 g
Beads, 100 g
Beads, 100 g
M.W. M.W. Bed Vol.
18
Swollen
Bio-Rad Laboratories, 2000 Alfred Nobel Dr., Hercules, CA 94547
LIT263 Rev C
Loading...