Input Impedance:≧1M// less than 100pF.
AC Conversion Type: Reading less Full Scale 15% add 4 dgt.
AC Conversions are ac-coupled, true rms responding, calibrated to the rms
value of a sine wave input. Accuracies are given for sine wave at full scale and
non-sine wave below half scale. For non-sine wave add the following Crest
Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
CF 2 @ 600V
1.5 @ 750V
(2) DC Volts: Auto-ranging
Range Resolution Accuracy
Over voltage
protection
Over voltage
protection
400.0V 100mV
±(0.7% reading + 2 digits) 1000V rms
1000V 1V
Input Impedance:≧1M
(3) Resistance Auto-ranging
Range Resolution Accuracy
Over voltage
protection
400.0 100m±(1.0% reading + 3 digits) 600V rms
Continuity: Built-in buzzer sound when resistance is less than 30 approximately.
4
(4) Frequency: Hz
Range Resolution Accuracy
20Hz ~ 400Hz 1Hz ±(0.1% reading + 2 digits)
Min. Input Frequency:20Hz
Sensitivity: 3A rms for ACA (A~)(>400Hz Unspecified)
Operating Temperature: 0℃ to 30℃(≦80%RH), 30℃ to 50℃(≦75%RH)
AC Conversion Type: Reading less Full Scale 15% add 4 dgt.
AC Conversions are ac-coupled, true rms responding, calibrated to the rms
value of a sine wave input. Accuracies are given for sine wave at full scale and
non-sine wave below half scale. For non-sine wave add the following Crest
Factor corrections:
For Crest Factor of 1.4 to 2.0, add 1.0% to accuracy.
For Crest Factor of 2.0 to 2.5, add 2.5% to accuracy.
For Crest Factor of 2.5 to 3.0, add 4.0% to accuracy.
CF 2 @ 600V
1.5 @ 750V
(6) DCA:
Range Resolution Accuracy
Over voltage
protection
0A ~ 200.0A ±(2.9% reading + 3 Amp)
200A ~ 400.0A
0.1A
±(1.9% reading + 2 Amp)
AC 1000A
for 1 min.
400A ~ 1000A 1A ±(2.9% reading + 5 Amp)
Overload
protection
1000A rms
Additional error according to remanence: 1% max. of current crest.
5
(7) Auto Power Off (APO)
The meter will automatically shut itself off after approximately 30 minutes after power on.
(8) Peak Hold: ±(3% reading +20 digits)
* >750V Unspecified.
* >800A Unspecified.
(9) Min/Max Hold: Add ± 15 dgt to accuracy for ACA.
Add ± 15 dgt and add remanence to accuracy for DCA.
* Automatically switch to the low resolution range at Peak Hold and Min/Max Hold.
6