Anritsu MT8213E, MT8212E User Manual

User Guide

Cell Master

MT8212E and MT8213E
MT8212E 2 MHz to 4 GHz Cable and Antenna Analyzer 9 kHz to 4 GHz Spectrum Analyzer 10 MHz to 4 GHz Power Meter
MT8213E 2 MHz to 6 GHz Cable and Antenna Analyzer 9 kHz to 6 GHz Spectrum Analyzer 10 MHz to 6 GHz Power Meter
Anritsu Company 490 Jarvis Drive Morgan Hill, CA 95037-2809 USA
http://www.anritsu.com
Part Number: 10580-00250
Copyright 2020 Anritsu Company, USA. All Rights Reserved.
Revision: AA
Published: April 2020

Table of Contents

Chapter 1—General Information
1-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Additional Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Contacting Anritsu for Sales and Service . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-2 Instrument Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
1-3 Available Models. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-4 Available Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
1-5 Calibration and Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1-6 Instrument Care and Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . 1-6
Connector Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
ESD Caution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
RF Input Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1-7 Secure Environment Workplace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Cell Master Memory Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Erase All User Files in Internal Memory . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Recommended Usage in a Secure Environment . . . . . . . . . . . . . . . . 1-10
Chapter 2—Instrument Overview
2-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 Turning On the Cell Master. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-3 Instrument Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
LED Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Front Panel Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2-4 Interface Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Touch Screen Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Symbols and Icons on the Title Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Symbols and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Display Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Touch Screen Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Calibration Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2-5 Data Entry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Numeric Values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Selection Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Text Entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
2-6 Mode Selector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
2-7 Connector Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Cell Master UG PN: 10580-00250 Rev. AA Contents-1
Table of Contents (Continued)
2-8 Soft Carrying Case . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
2-9 Tilt Bail Stand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22
Chapter 3—Quick Start Guide
3-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 Measurement Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-3 Cable & Antenna Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Select the Measurement Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Set the Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Set the Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Turn on Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Single Limit Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
DTF Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Calibrate with OSL Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
3-4 Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Set Start and Stop Frequencies. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Enter the Center Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Select a Signal Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Set the Measurement Frequency Bandwidth. . . . . . . . . . . . . . . . . . . . . 3-8
Set the Amplitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Power Offset Set Up for Compensating External Loss . . . . . . . . . . . . . 3-9
Set the Span . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Single Limit Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Create a Limit Envelope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Setting Up Markers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Select a Measurement Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
External Power On. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Chapter 4— File Management
4-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4-2 Managing Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
File Types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Save Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Recall Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Copy Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Delete Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
4-3 File Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Contents-2 PN: 10580-00250 Rev. AA Cell Master UG
Table of Contents (Continued)
4-4 File Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Save Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
File Type Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Save Location Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
Save On... Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
Recall Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
Copy Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Delete Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-16
Chapter 5—System Operations
5-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5-2 System Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5-3 System Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
System Options Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
System Options 2/2 Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Power-On Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Display Settings Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Brightness Settings Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Reset Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
5-4 Preset Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
5-5 Self Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
5-6 Firmware Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Chapter 6—GPS (Option 31)
6-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-2 Activating the GPS Feature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6-3 Resetting GPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6-4 Saving and Recalling Traces with GPS Information . . . . . . . . . . . . . . . . . . 6-3
Saving Traces with GPS Information. . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Recalling GPS Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
6-5 GPS Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Chapter 7—Bias Tee (Option 10)
7-1 Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Chapter 8—Anritsu Tool Box
8-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-2 Software Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8-3 Anritsu Software Tool Box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-2
Cell Master UG PN: 10580-00250 Rev. AA Contents-3
Table of Contents (Continued)
8-4 Software Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Line Sweep Tools (LST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Master Software Tools (MST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
easyTest Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
easyMap Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Wireless Remote Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
Chapter 9—Web Remote Control
9-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
9-2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
LAN Connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1
Connection to a Wireless Router . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
9-3 Web Remote Control Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
User Login . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Home Page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Capture Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Capture Trace . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
File List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Device Management (not as Administrator) . . . . . . . . . . . . . . . . . . . . . 9-8
Device Management (Administrator) . . . . . . . . . . . . . . . . . . . . . . . . . 9-10
Logout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12
Appendix A—Measurement Guides
A-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Appendix B—LAN and DHCP
B-1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B-2 LAN Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
B-3 Ethernet Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Ethernet Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2
DHCP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4
B-4 ipconfig Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
B-5 Ping Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6
Appendix C—Glossary of Terms
C-1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .C-1
C-2 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1
Index
Contents-4 PN: 10580-00250 Rev. AA Cell Master UG

Chapter 1 — General Information

1-1 Introduction

The Cell Master User Guide is part of a set of manuals that describe all of the instrument functions and their use. This manual covers the instrument overview, system functions and other common features, along with a brief guide to basic measurement concepts and setups. Instrument operations are explained in various document types as listed below.

Additional Documentation

Document Part Number Description (Required Option)
10100-00065 Important Product Information, Compliance, and Safety Notices
10580-00215
10580-00349
10580-00234
ODTF-1 Optical Distance-to-Fault User Guide ODTF-1 Optical Distance-to-Fault Module
Spectrum Analyzer Measurement Guide Spectrum Analyzer PIM Hunting Interference Analyzer (Option 25) Channel Scanner (Option 27) C/W Signal Generator (Option 28) Gated Sweep (Option 90) Coverage Mapping (Option 431) EMF Measurements (Option 444)
3GPP Signal Analyzer Measurement Guide GSM/GPRS/EDGE Measurements (Option 880) * GSM/EDGE RF Measurements (Option 40) * GSM/EDGE Demodulation (Option 41) W-CDMA/HSPA+ Measurements (Option 881) * W-CDMA/HSPA+ RF Measurements (Option 44) * W-CDMA/HSPA+ Demodulation (Option 65) * W-CDMA/HSPA+ Over-the-Air Measurements (Option 35) TD-SCDMA/HSPA+ Measurements (Option 882) * TD-SCDMA/HSPA+ Measurements (Option 60) * TD-SCDMA/HSPA+ Demodulation (Option 61) * TD-SCDMA/HSPA+ Over-the-Air Measurements (Option 38) LTE/LTE-A FDD/TDD Measurements (Option 883) LTE and TD-LTE Coverage Mapping (included) TD-LTE/LTE OTA EMF Measurements (Option 444) * TD-LTE/LTE-A RF Measurements (Option 551) * TD-LTE/LTE-A Modulation Measurements (Option 552) * TD-LTE/LTE-A Over-the-Air Measurements (Option 556) LTE 256QAM Demodulation Measurements (Option 886) NB-IoT Measurements (Option 887)
Cell Master UG PN: 10580-00250 Rev. AA 1-1
1-1 Introduction General Information
Document Part Number Description (Required Option)
3GPP2 Signal Analyzer Measurement Guide CDMA/EV-DO Measurements (Option 884) * CDMA RF Measurements (Option 42)
10580-00235
* CDMA Demodulation (Option 43) * CDMA Over-the-Air Measurements (Option 33) * EV-DO RF Measurements (Option 62) * EV-DO Demodulation (Option 63) * EV-DO Over-the-Air Measurements (Option 34)
WiMAX Signal Analyzer Measurement Guide WiMAX Fixed/Mobile Measurements (Option 885) * Fixed WiMAX RF Measurements (Option 46)
10580-00236
* Fixed WiMAX Demodulation (Option 47) * Mobile WiMAX RF Measurements (Option 66) * Mobile WiMAX Demodulation (Option 67) * Mobile WiMAX Over-the-Air Measurements (Option 37)
Digital Video Broadcast Signal Analyzer Measurement Guide ISDB-T Digital Video Measurements (Option 30) ISDB-T SFN Measurements (Option 32)
10580-00237
ISDB-T BER Measurements (Option 79) DVB-T/H Digital Video Measurements (Option 64) DVB-T/H SFN Measurements (Option 78) DVB-T/H BER Measurements (Option 57)
Backhaul Analyzer Measurement Guide
10580-00238
* T1 Analyzer (Option 51) * E1 Analyzer (Option 52) * T3/T1 Analyzer (Option 53)
Power Meter Measurement Guide
10580-00240
High-Accuracy Power Meter (Option 19) Power Meter (Option 29)
10580-00241 Cable and Antenna Analyzer
10580-00242
2-Port Transmission Measurement Guide (Option 21) Bias-Tee (Option 10)
10580-00256 SCPI Programming Manual
10580-00280
PIM Master User Guide PIM Analyzer (supported on legacy instruments only)
CPRI LTE RF Analyzer Measurement Guide
10580-00415
CPRI LTE RF Measurements (Option 752; requires Option 759) * Option 751 (obsolete) is functionally identical to the combined Options 752 and 759
10580-00434
OBSAI LTE RF Analyzer Measurement Guide OBSAI LTE RF Measurements (Option 753; requires Option 759)
10580-00455 EMF Measurement Guide (Option 444)
11410-00485
* Indicates the option is obsolete.
Cell Master Technical Data Sheet Performance Specifications, Instrument Options and Accessories
1-2 PN: 10580-00250 Rev. AA Cell Master UG

General Information 1-2 Instrument Description

Read the Handheld Instruments Product Information, Compliance, and Safety Guide (PN: 10100-00065) for important safety, legal, and regulatory notices before operating the equipment. For additional information and literature covering your product, visit the product page of your instrument and select the Library tab:
http://www.anritsu.com/en-US/test-measurement/products/mt8212e
http://www.anritsu.com/en-US/test-measurement/products/mt8213e

Contacting Anritsu for Sales and Service

To contact Anritsu, visit the following URL and select the services in your region:
www.anritsu.com/contact-us.
1-2 Instrument Description
The Cell Master MT821xE is a handheld multi-function Base Station Analyzer designed to make Cable and Antenna Analysis, Spectrum Analysis and Power Meter measurements in the field. In addition, the Cell Master can be equipped with 2-port Transmission Measurement capability, Interference Analyzer, Channel Scanner, Coverage Mapping, CW Signal Generator, GSM/EDGE Analyzer, W-CDMA/HSDPA Analyzer, TD-SCDMA Analyzer, CDMA Analyzer, EVDO Analyzer, Fixed and Mobile WiMAX Analyzer, PIM Analyzer, T1/T3, E1 Analyzer, LTE Analyzer, and TD-LTE Analyzer, thus eliminating the need to carry multiple instruments to the field.
The cable & antenna analyzer includes Return Loss, Cable Loss, VSWR, Distance-to-fault
-Return Loss, Distance-to-Fault SWR, 1-port phase and smith chart measurements. The 2-port transmission measurement option includes two power levels and access to a built-in 32V bias tee.
The bright 8.4" TFT color display provides easy viewing in a variety of lighting conditions. The Cell Master MT821xE is equipped with a Li-Ion battery delivering more than three hours of battery life.
The combination of a touch screen and keypad enables users to navigate menus with the touch screen and enter numbers with the keypad.
Time and date stamping of measurement data is automatic. The internal memory provides for the storage and recall of up to 1000 measurement setups and up to 1000 traces using Master Software Tools (MST). Measurements and setups can be stored in a USB flash drive or transferred to a PC by using the included USB cable.
Not all after-market USB drives are compatible with the Cell Master. Many drives come with a second partition that contains proprietary firmware. This partition must
Note
Cell Master UG PN: 10580-00250 Rev. AA 1-3
be removed. Only one partition is allowed. Refer to the individual manufacturer for instructions on how to remove it. You may also try reformatting a drive that contains a single partition using FAT32 format.

1-3 Available Models General Information

1-3 Available Models
Table 1-1 lists the Cell Master models described in this User Guide.
Tab le 1 -1. Cell Master Model
Model Frequency Range
MT8212E Cable & Antenna Analyzer, 2 MHz to 4 GHz
Spectrum Analyzer, 9 kHz to 4 GHz Power Meter, 10 MHz to 4 GHz
MT8213E Cable & Antenna Analyzer, 2 MHz to 6 GHz
Spectrum Analyzer, 9 kHz to 6 GHz Power Meter, 10 MHz to 6 GHz

1-4 Available Options

Available options for the Cell Master are shown in Table 1-2.
Tab le 1 - 2 . Available Options (Sheet 1 of 2)
MT8212E MT8213E Description
MT8212E-0010 MT8213E-0010 Bias-Tee
MT8212E-0019 MT8213E-0019
MT8212E-0021 MT8213E-0021 2-Port Transmission Measurement
MT8212E-0025 MT8213E-0025 Interference Analyzer (Option 31 recommended)
MT8212E-0027 MT8213E-0027 Channel Scanner
MT8212E-0028 MT8213E-0028
MT8212E-0030 MT8213E-0030 ISDB-T Digital Video Measurements
MT8212E-0031 MT8213E-0031 GPS Receiver (requires GPS Antenna)
MT8212E-0032 MT8213E-0032 ISDB-T SFN Measurements
MT8212E-0057 MT8213E-0057
MT8212E-0064 MT8213E-0064 DVB-T/H Digital Video Measurements
MT8212E-0078 MT8213E-0078 DVB-T/H SFN Measurements
MT8212E-0079 MT8213E-0079
MT8212E-0090 MT8213E-0090 Gated Sweep
MT8212E-0098 MT8213E-0098
MT8212E-0099 MT8213E-0099
MT8212E-0413 MT8213E-0413 Ethernet Connectivity
High-Accuracy Power Meter (requires External Power Sensor)
C/W Signal Generator (requires CW Signal Generator Kit, P/N 69793)
DVB-T/H BER Measurements (requires Option 64; cannot be ordered with Option 759)
ISDB-T BER Measurements (requires Option 30; cannot be ordered with Option 759)
Standard Calibration to ISO17025 and ANSI/NCSL Z540-1. Includes calibration certificate.
Premium Calibration to ISO17025 and ANSI/NCSL Z540-1. Includes calibration certificate, test report, and uncertainty data.
1-4 PN: 10580-00250 Rev. AA Cell Master UG

General Information 1-5 Calibration and Verification

Table 1 - 2. Available Options (Sheet 2 of 2)
MT8212E MT8213E Description
MT8212E-0431 MT8213E-0431 Coverage Mapping (requires Option 31)
MT8212E-0444 MT8213E-0444 EMF Measurements (requires Anritsu Isotropic Antenna)
MT8212E-0752 MT8213E-0752 CPRI LTE RF Measurements (requires Option 759)
MT8212E-0753 MT8213E-0753 OBSAI LTE RF Measurements (requires Option 759)
MT8212E-0759 MT8213E-0759
MT8212E-0880 MT8213E-0880
MT8212E-0881 MT8213E-0881
MT8212E-0882 MT8213E-0882
MT8212E-0883 MT8213E-0883
MT8212E-0884 MT8213E-0884
MT8212E-0885 MT8213E-0885
MT8212E-0886 MT8213E-0886
MT8212E-0887 MT8213E-0887 NB-IoT Measurements
RF over Fiber Hardware (requires Option 752 or 753; cannot be ordered with Option 57 or 79)
GSM/GPRS/EDGE Measurements (combines functionality of Options 40 and 41; cannot be ordered with Option 40 or
41)
W-CDMA/HSPA+ Measurements (Option 31 recommended)
TD-SCDMA/HSPA+ Measurements (requires Option 31 for full functionality)
LTE/LTE-A FDD/TDD Measurements (requires Option 31 for full functionality)
CDMA/EV-DO Measurements (requires Option 31 for full functionality)
WiMAX Fixed/Mobile Measurements (requires Option 31 for full functionality)
LTE 256QAM Demodulation Measurements (requires Option 883 or Option 552*)
* Indicates the option is obsolete.
1-5 Calibration and Verification
Anritsu recommends annual calibration and performance verification by local Anritsu service centers. The Cable and Antenna Analyzer mode requires calibration standards for OPEN, SHORT, and LOAD (OSL) or InstaCal module, which are sold separately.
Note
The Cell Master is self-calibrating and there are no field-adjustable components. The OSL calibration components are crucial to the integrity of the calibration. As a result, they must be verified periodically to ensure performance conformity. This is especially important if the OSL calibration components have been accidentally dropped or over-torqued.
Contact information for Anritsu Service Centers is available at: www.anritsu.com/contact-us.
Cell Master UG PN: 10580-00250 Rev. AA 1-5
Anritsu recommends allowing the instrument to warm up to typical operation temperature (approximately 15 minutes) before calibrating.

1-6 Instrument Care and Preventive Maintenance General Information

1-6 Instrument Care and Preventive Maintenance
Instrument care and preventive maintenance consist of cleaning the unit and inspecting and cleaning the RF connectors and all accessories. Clean the instrument with a soft, lint-free cloth dampened with water or water and a mild cleaning solution.
Caution To avoid damaging the display or case, do not use solvents or abrasive cleaners.

Connector Care

Clean the RF connectors and center pins with a cotton swab dampened with denatured alcohol. Visually inspect the connectors. The fingers of the N(f) connectors and the pins of the N(m) connectors should be unbroken and uniform in appearance. If you are unsure whether the connectors are undamaged, gauge the connectors to confirm that the dimensions are correct. Visually inspect the test port cable(s). The test port cable should be uniform in appearance, and not stretched, kinked, dented, or broken.
To prevent damage to your instrument, do not use pliers or a plain wrench to tighten the Type-N connectors. The recommended torque is 12 lbf · in to 15 lbf· in (1.36 N· m to 1.70 N·m). Inadequate torque settings can affect measurement accuracy. Over-tightening connectors can damage the cable, the connector, the instrument, or all of these items.
Visually inspect connectors for general wear, cleanliness, and for damage such as bent pins or connector rings. Repair or replace damaged connectors immediately. Dirty connectors can limit the accuracy of your measurements. Damaged connectors can harm the instrument. Connection of cables carrying an electrostatic potential, excess power, or excess voltage can damage the connector, the instrument, or both.
Connecting Procedure
1. Carefully align the connectors. The male connector center pin must slip concentrically into the contact fingers of the female connector.
2. Push connectors straight together. Do not twist or screw them together. A slight resistance can usually be felt as the center conductors mate.
3. To tighten, turn the connector nut, not the connector body. Major damage can occur to the center conductor and to the outer conductor if the connector body is twisted.
4. If you use a torque wrench, initially tighten by hand so that approximately 1/8 turn or 45 degrees of rotation remains for the final tightening with the torque wrench.
Relieve any side pressure on the connection (such as from long or heavy cables) in order to assure consistent torque. Use an open-end wrench to keep the connector body from turning while tightening with the torque wrench.
Do not over-torque the connector.
Disconnecting Procedure
1. If a wrench is needed, use an open-end wrench to keep the connector body from turning while loosening with a second wrench.
2. Complete the disconnection by hand, turning only the connector nut.
3. Pull the connectors straight apart without twisting or bending.
1-6 PN: 10580-00250 Rev. AA Cell Master UG
General Information 1-6 Instrument Care and Preventive Maintenance

ESD Caution

The Cell Master, like other high performance instruments, is susceptible to electrostatic discharge (ESD) damage. Coaxial cables and antennas often build up a static charge, which (if allowed to discharge by connecting directly to the instrument without discharging the static charge) may damage the Cell Master input circuitry. Instrument operators must be aware of the potential for ESD damage and take all necessary precautions.
Operators should exercise practices outlined within industry standards such as JEDEC-625 (EIA-625), MIL-HDBK-263, and MIL-STD-1686, which pertain to ESD and ESDS devices, equipment, and practices. Because these apply to the Cell Master, it is recommended that any static charges that may be present be dissipated before connecting coaxial cables or antennas to the instrument. This may be as simple as temporarily attaching a short or load device to the cable or antenna prior to attaching to the Cell Master. It is important to remember that the operator may also carry a static charge that can cause damage. Following the practices outlined in the above standards will ensure a safe environment for both personnel and equipment.

RF Input Warning

The Anritsu Cell Master is a sensitive measuring instrument designed to measure low power levels. Avoid damaging this sensitive circuitry by observing the maximum input levels printed on the instrument connector labeling and specified in the product technical data sheet.
Typical maximum input is +33 dBm (±50 VDC) and could be less if additional features, such as a preamplifier, are in use. Be sure to review the product technical data sheet or Anritsu website for recommended components and accessories that can help you protect your instrument. These include a variety of adapters, attenuators, filters, and RF detection accessories.

Battery Replacement

The battery can be replaced without the use of tools. The battery compartment door is located on the lower left side of the instrument (when you are facing the measurement display). To remove the battery:
1. Slide the catch toward the bottom of the instrument.
2. Pull the top of the door away from the unit.
3. Lift out the battery door.
4. Pull straight out on the lanyard to remove the battery pack from the instrument.
When inserting the battery, the battery label should face the back of the instrument
Note
Cell Master UG PN: 10580-00250 Rev. AA 1-7
and the end where the battery contacts are located slides in first. If the battery door does not latch closed, the battery may be inserted incorrectly.
1-6 Instrument Care and Preventive Maintenance General Information
Figure 1-1. Battery Compartment
The battery that is supplied with the Cell Master may need charging before use. The battery can be charged while it is installed in the Cell Master by using the AC-DC Adapter, or outside the Cell Master with the optional Dual Battery Charger. Refer to “Symbols and Indicators”
on page 2-9 for a description of battery symbols.
Use only Anritsu Company approved batteries, adapters, and chargers with this
Note
instrument. Anritsu recommends removing the battery for long-term storage of the instrument.
When using the Automotive Power Adapter, always verify that the supply is rated
Caution
for a minimum of 60 Watts @ 12 VDC, and that the socket is clear of any dirt or debris. If the adapter plug becomes hot to the touch during operation, discontinue use immediately.
1-8 PN: 10580-00250 Rev. AA Cell Master UG

General Information 1-7 Secure Environment Workplace

1-7 Secure Environment Workplace
This section details the types of memory in the Cell Master, how to delete stored user files in internal memory, and recommended usage in a secure environment workplace.

Cell Master Memory Types

The instrument contains non-volatile disk-on-a-chip memory, EEPROM, and volatile DRAM memory. The instrument does not have a hard disk drive or any other type of volatile or non-volatile memory.
Disk-On-A-Chip (DOC): DOC is used for storage of instrument firmware, factory calibration information, user measurements, setups, and .jpg screen images. User information stored on the DOC is erased by the master reset process described below.
EEPROM: This memory stores the model number, serial number, and calibration data for the instrument. Also stored here are the user-set operating parameters such as frequency range. During the master reset process, all operating parameters stored in the EEPROM are set to standard factory default values.
RAM Memory: This is volatile memory used to store parameters needed for the normal operation of the instrument along with current measurements. This memory is reset whenever the instrument is restarted.
External USB Flash Drive (not included with the instrument): This memory may be selected as the destination for saved measurements and setups for the instrument. The user can also copy the contents of the internal disk-on-chip memory to the external flash memory for storage or data transfer. The external Flash USB can be reformatted or sanitized using software on a PC.
Refer to the Chapter 4, “File Management” for additional information on saving and copying files to the USB flash drive.

Erase All User Files in Internal Memory

Perform a Master Reset:
1. Turn the instrument on.
2. Press the Shift button, then the System (8) button.
3. Press the System Options submenu key.
4. Press the Reset key, then the Master Reset key.
A message box will be displayed to warn the user that all settings will be returned to factory default values and all user files will be deleted. This is a standard file deletion and does not involve overwriting existing information.
5. Press the ENTER button to complete the master reset.
6. The instrument will reboot and the reset is complete.
Cell Master UG PN: 10580-00250 Rev. AA 1-9
1-7 Secure Environment Workplace General Information

Recommended Usage in a Secure Environment

Set the instrument to save files to the external USB Flash drive:
1. Attach the external Flash drive and turn the instrument on.
2. Press the Shift button, then the File (7) button.
3. Press the Save submenu key.
4. Press the Change Save Location submenu key, then select the USB drive with the
rotary knob, Up/Down arrow keys, or the touch screen.
5. Press the Set Location submenu key.
The external USB drive is now the default location for saving files.
Not all USB drives are compatible with the instrument. Many drives come with a second partition that contains proprietary firmware. This partition must be
Note
removed. Only one partition is allowed. Refer to the individual manufacturer for instructions on how to remove it. Some drives can be made to work by reformatting them using the FAT32 format.
1-10 PN: 10580-00250 Rev. AA Cell Master UG

Chapter 2 — Instrument Overview

2-1 Introduction

This chapter provides an overview of the Anritsu Cell Master. It describes the instrument front panel, touch screen display, and the connector panel. For detailed information on the instrument’s test and measurement functions, refer to one of the Anritsu user documents listed in Appendix A, “Measurement Guides”.

2-2 Turning On the Cell Master

The Anritsu Cell Master is capable of approximately three hours of continuous operation from a fully charged, field-replaceable battery (refer to “Battery Replacement” on page 1-7). The Cell Master can also be operated from a 12 VDC source (which will simultaneously charge the battery). This can be achieved with either the Anritsu AC-DC Adapter or the Automotive Power Adapter. Both items are included as standard accessories with the Cell Master. Refer to the instrument Technical Data Sheet.
When using the Automotive Power Adapter, always verify that the supply is rated
Caution
for a minimum of 60 Watts @ 12 VDC, and that the socket is clear of any dirt or debris. If the adapter plug becomes hot to the touch during operation, discontinue use immediately.
To turn on the Cell Master, press the green On/Off button on the front panel (see Figure 2-1
on page 2-2).
The Cell Master takes approximately 60 seconds to complete power-up and to load the application software. At the completion of this process, the instrument is ready for use.
Note
Keep the fan inlet and exhaust port clear of obstructions at all times for proper ventilation and cooling of the instrument.
Cell Master UG PN: 10580-00250 Rev. AA 2-1

2-3 Instrument Front Panel Instrument Overview

1
2
3
4
5
6
7
8
9
10
2-3 Instrument Front Panel
The Cell Master uses a touch screen and keypad for data input. See Figure 2-1. The five bottom menu keys and up to eight submenu keys on the right side of the display are touch screen keys. The menu and submenu keys vary depending on the current measurement mode, installed options, and instrument function selected.
Figure 2-1. Cell Master Front Panel and Interface Screen
1. Fan Inlet Port
2. Touch Screen Submenu Keys
3. Menu Key
4. Shift Key
5. Numeric Keypad and Shift Menu Keys (0 to 9, printed in blue above each key)
6. Fan Inlet Port
7. On/Off Button
8. Touch Screen Main Menu Keys
9. Calibration Status and Type
10. Fan Exhaust Port
2-2 PN: 10580-00250 Rev. AA Cell Master UG
Instrument Overview 2-3 Instrument Front Panel

LED Indicators

Power LED
The Power LED is located on the left of the On/Off button. The LED is solid green when the instrument is on, and blinks slowly when the unit is off but is connected to an external power source.
Charge LED
The Charge LED is located on the right of the On/Off button. The LED blinks slowly when the battery is charging and is solid green when the battery is fully charged.

Front Panel Keys

The numeric keypad, rotary knob, and the four arrow keys can all be used to change the value of the currently selected parameter.
Numeric Keypad
Keys 0 through 9 are used for numeric input, with an alternate function printed in blue above each of the keys. Press the Shift key, then a numeric key, to access the instrument menu or function indicated by the key label.
Some of the alternate functions associated with the numeric keypad are not available in all measurement modes. Refer to the Measurement Guides listed in Appendix A.
Shift Key
Pressing the Shift key followed by a number key executes the function that is indicated in blue above the number key. When the Shift key is active, its icon is displayed at the far right of the title bar, above the sweep window.
Figure 2-2. Shift Key Icon in Title Bar
Esc Key
Press this key to cancel the parameter change being made and exit the current menu function, if applicable.
Arrow Keys
The four arrow keys are used to scroll through a list and highlight the item you wish to select, or to change the value of the currently selected parameter. The arrow keys can also be used to move markers. The rotary knob performs similar functions.
Enter Key
Press this key to apply a parameter value or instrument setting you have entered, or to select a highlighted item from a list.
Cell Master UG PN: 10580-00250 Rev. AA 2-3
2-3 Instrument Front Panel Instrument Overview
Rotary Knob
Turn the rotary knob to change numerical values, scroll through selectable items in a list, or to move markers. Values or items may be within a dialog box or an edit window.
Menu Key
The Menu key displays a grid of shortcut icons for installed measurement modes and user-selected menus and setup files. See Figure 2-3. Press one of the icons in the top rows to switch to the corresponding application. These icons are system-generated and cannot be moved or deleted.
An alternative to the Menu key is to press Shift, then the Mode (9) key to display the Mode Selector list box. Refer to “Mode Selector” on page 2-16.
Figure 2-3. Menu Key Screen
Shortcut icons displayed in the top rows of the Menu screen vary with the
Note
instrument model, firmware version, and installed options. Help for the Menu screen is available by pressing the icon in the lower right corner of the display.
2-4 PN: 10580-00250 Rev. AA Cell Master UG
Instrument Overview 2-3 Instrument Front Panel
User-Created Shortcuts
To create a shortcut to any submenu key or main menu key, press and hold the key until a grid appears, showing the open locations on the Menu screen where you can place the new shortcut. Press one of the cells in the displayed grid to place the shortcut. See Figure 2-4.
Figure 2-4. Placement Grid for User-Created Shortcuts
To create a shortcut to a setup file (.stp), press Recall under the File menu, then press and hold the desired file name until a grid is displayed. Select the display location of the new shortcut as described above.
User-defined shortcuts remain on the Menu screen until deleted. To delete or move a shortcut icon, press the Menu key, then press and hold the shortcut until the Customize Button dialog appears. See Figure 2-5. Press the appropriate button to delete or move the shortcut icon.
Figure 2-5. Customize Button Dialog
Press Esc to close the dialog without deleting or moving the shortcut. Also use Esc to exit the Menu screen.
Cell Master UG PN: 10580-00250 Rev. AA 2-5

2-4 Interface Screen Instrument Overview

1
2
3
4
5 6
7
8
9
10
11
1213
2-4 Interface Screen
Figure 2-6 and Figure 2-7 illustrate the Cell Master interface screen in Cable & Antenna
mode and Spectrum Analyzer mode, with touch screen menu keys, title bar, and measurement settings and results around the graph area, or sweep window.
Figure 2-6. Cable & Antenna Analyzer Return Loss Measurement Display
1. Measurement Settings Summary
2. Calibration Status, Type
3. Frequency Standard
4. Date and Time
5. GPS Icon
6. GPS Location
7. Save Icon
8. Save Screen Icon
9. Battery Charge Indicator
10. Trace Measurement Title
11. Submenu Touch Screen Keys
12. Main Menu Touch Screen Keys
13. Marker Table
2-6 PN: 10580-00250 Rev. AA Cell Master UG
Instrument Overview 2-4 Interface Screen
1
2
1112
3
4 5
6
7
8
9
10
Figure 2-7. Spectrum Analyzer Display
1. Measurement Settings Summary (Touch Screen Shortcuts)
2. Frequency Standard
3. Date and Time
4. GPS Icon
5. GPS Location
6. Save Icon
7. Save Screen Icon
8. Battery Charge Indicator
9. Trace Measurement Title
10. Submenu Touch Screen Keys
11. Main Menu Touch Screen Keys
12. Marker Table
Note
Many of the measurement settings displayed to the left and the top of the sweep window are actually touch screen shortcuts that you can press to bring up the corresponding menu.
Cell Master UG PN: 10580-00250 Rev. AA 2-7
2-4 Interface Screen Instrument Overview

Touch Screen Keys

Main Menu Touch Screen Keys
There are five main menu keys horizontally arranged along the bottom of the interface screen. These keys give access to the instrument’s test and measurement functions, which are documented in the applicable Measurement Guide. Refer to Appendix A, “Measurement
Guides”.
Different operation or measurement modes may display different main menu keys. To switch to another mode, press the Menu key, or press Shift followed by the Mode (9) key.
Note
Submenu Touch Screen Keys
Most of the instrument setup, control, and measurement functions are performed through the use of the submenu keys along the right side of the display. The key labels change as measurement settings and instrument setup parameters change. The current submenu title is displayed at the top of the submenu key block, which consists of up to eight touch screen keys. See Figure 2-6 on page 2-6.
The instrument model and installed options determine what measurement modes are available. Refer to Table 1-1 and Table1-2 onpage1-4.

Symbols and Icons on the Title Bar

Figure 2-8. Title Bar with Icons
The instrument title bar displays the system date and time. When GPS is on and is tracking satellites, its icon is followed by latitude and longitude coordinates.
Press the storage icon to open the touch screen keyboard for saving measurements, setups, limit lines, or screen display JPEG files. This shortcut to the “Save Menu” on page 4-11 is equivalent to pressing Shift and File (7), then Save. Refer to “Save Dialog Box” on page 4-3.
Press the camera icon to save a JPEG image of the current screen display.
The battery symbol indicates the charge remaining in the battery. The colored section inside the symbol changes size and color with the charge level. The Battery Charge LED (adjacent to the On/Off button) flashes when the battery is charging, and remains on steady when the battery is fully charged.
Use only Anritsu-approved batteries, adapters, and chargers with this instrument.
Caution
The Shift icon is displayed after the Shift key is pressed, and it remains displayed until another button is pressed.
2-8 PN: 10580-00250 Rev. AA Cell Master UG
Anritsu Company recommends removing the battery for long-term storage of the instrument.
Instrument Overview 2-4 Interface Screen
Shift

Symbols and Indicators

The following symbols, icons, and indicators convey the instrument status or condition on the display. The colors shown here are in the standard or default display mode.
Table 2 - 1. Symbols and Icons
Symbol Description
Green: Battery is 30% to 100% charged.
Yellow: Battery is 10% to 30% charged.
Red: Battery is 0% to 10% charged.
Green with Black Plug body: Battery is fully charged and
external power is applied.
Lightning Bolt: Battery is being charged (any color symbol).
Red Plug body: External power is applied, and no battery is
installed, or battery has lost communications with the instrument.
Storage Icon: Tap the floppy disk icon to display the Save screen and menu.
Camera Icon: Saves a JPEG image of the current screen display.
Shift Key Icon: This icon is displayed between the battery symbol
and the submenu keys after the Shift key has been pressed, and until another key is pressed.
GPS Icon: This icon is displayed right after the date and time when GPS is available. Refer to Chapter 6, “GPS (Option 31) ” for details.
Power Button with Power LED and Charge LED: This is a physical button with LED indicators. It is located near the numeric keypad.
Cell Master UG PN: 10580-00250 Rev. AA 2-9
2-4 Interface Screen Instrument Overview
Black on White Night Vision
Color on WhiteHigh Contrast

Display Settings

The Display Settings submenu lets you adjust the screen brightness level and control the auto-dimming function. Refer to “Brightness Settings Menu” on page 5-8.
You can also turn off the display entirely, as described in “Display Settings Menu”
on page 5-7. To turn the display back on, press any key (except the Power button) three times
in rapid succession.
In addition to the default display colors shown in Figure 2-6 on page 2-6, you can select different color schemes to suit the ambient lighting conditions. See Figure 2-9. Some color settings may not be available in all measurement modes.
Black on White — used for printing and viewing in broad daylight conditions
Night Vision — optimized for nighttime viewing
High Contrast — used in challenging viewing conditions
Color on White — used for printing and viewing in broad daylight conditions
Figure 2-9. Cell Master Display Color Settings
2-10 PN: 10580-00250 Rev. AA Cell Master UG
Instrument Overview 2-4 Interface Screen

Touch Screen Calibration

Calibration optimizes the response of touch input. It is recommended if the instrument does not respond as expected when you press the touch screen.
The Calibrate Touch Screen submenu key is in the “System Menu” on page 5-3. When pressed, a message box is displayed with instructions for calibration. See Figure 2-10. Press Enter to begin calibration, or press Esc to cancel. Alternatively, you can press 1 while the message box is displayed to use the arrow keys for navigation. Refer to “Disable Touch Screen to Use
Arrow Navigation”.
Figure 2-10. Touch Screen Calibration Message Box
During calibration, press the crosshairs in sequence as they appear on the screen. The process takes less than one minute.
Calibrate Touch Screen Shortcut
Another way to access touch screen calibration is to press Shift, then 0 (zero). This displays the touch screen calibration message box shown in Figure 2-10. The shortcut can be used if your touch inputs do not correspond to the appropriate locations on the screen to such an extent that you cannot access the Calibrate Touch Screen submenu key.
Disable Touch Screen to Use Arrow Navigation
If the touch screen is not functioning, you can use Arrow Navigation to simulate pressing the touch screen main menu keys and submenu keys. From the touch screen calibration message box (see Figure 2-10), press 1 to display the arrow navigation message box, illustrated in
Figure 2-11. Press 1 again to enter the arrow navigation mode, or press Esc to cancel.
Figure 2-11. Arrow Navigation Message Box
In arrow navigation mode, a red border highlights the currently selected key (see Figure 2-12
on page 2-12). Use the arrow keys to move the red selection box, then press the Menu key to
activate the highlighted key. Only the main menu keys and submenu keys can be activated using Arrow Navigation. This feature does not move the red selection box into other areas of the touch screen, like the display title bar, sweep window, or the instrument settings area on the left.
Cell Master UG PN: 10580-00250 Rev. AA 2-11
2-4 Interface Screen Instrument Overview
To save a measurement in arrow navigation mode, press Shift then File (7). Use the arrow keys to move the red selection box to the Save Measurement As submenu key. This submenu key must be used because the arrow navigation mode cannot be used to change data in popup windows in the measurement display. File names are determined by the current setting of the Save Measurement As submenu key. Refer to “Save Measurement As” on page 4-10.
You can save a JPEG image of the current display screen by pressing Shift, followed by Decimal, then +/–. The JEPG image shows screen data, but does not contain the additional measurement information that accompanies a saved measurement in a *.spa measurement file.
To return to normal touch entry mode, reboot the instrument (turn power Off, then On). If your touch screen has been damaged, refer to “Contacting Anritsu for Sales and Service”
on page 1-3.
Figure 2-12. Arrow Navigation Example
2-12 PN: 10580-00250 Rev. AA Cell Master UG
Instrument Overview 2-4 Interface Screen

Calibration Symbols

The current calibration status and type are displayed in the upper-left of the screen when in Cable & Antenna Analyzer mode. See Figure 2-6 on page 2-6. The five status messages are described next.
Cal Status: ON, Flex
The Cell Master has been calibrated with discrete Open, Short, and Load components. This is a FlexCal calibration indicating it is possible to change the frequency range after calibration.
Cal Status: ON, Standard
The Cell Master has been calibrated with discrete Open, Short, and Load components. This is a Standard calibration indicating it is not possible to change the frequency range after calibration without performing another calibration.
Cal Status: ON, Flex, Insta
The Cell Master has been calibrated with the InstaCal module. This is a FlexCal calibration indicating it is possible to change the frequency range after calibration.
Cal Status: ON, Standard, Insta
The Cell Master has been calibrated with the InstaCal module. The Cell Master has been calibrated with discrete Open, Short, and Load components. This is a Standard calibration indicating it is not possible to change the frequency range after calibration without performing another calibration.
Cal Status Off:
The Cell Master has not been calibrated.
For calibration procedures, refer to the Cable & Antenna Measurement Guide (PN: 10580-00241) listed in Appendix A.
Cell Master UG PN: 10580-00250 Rev. AA 2-13

2-5 Data Entry Instrument Overview

2-5 Data Entry
User input can be in the form of numeric values for instrument or measurement settings, selected values from a preset list, or alphanumeric text when entering file names, for example. To view or change a parameter value, access the appropriate submenu by pressing one of the five main menu keys along the bottom of the interface screen, then navigating through the touch screen submenus.
Other instrument functions are accessible from the numeric keypad, when used in combination with the Shift key. Refer to “Numeric Keypad” on page 2-3. Some of the parameter settings displayed on the left of the interface screen or under the title bar are touch screen shortcuts to related submenus.
Depending on the measurement mode, refer to the associated Measurement Guide listed in
Appendix A for a description of available menus and submenus.

Numeric Values

To modify a numeric parameter setting that is displayed on a submenu touch screen key, press the key to make it active. The display color of the currently set value changes to red. Use the rotary knob, arrow keys, or the numeric keypad to change the value. When entering a value from the keypad, the submenu typically shows the selectable units. Press the desired unit or press Enter to complete the entry.

Selection Lists

Some parameters and instrument functions are selectable from a pop-up list. These list boxes display the available selections, and value limits if applicable. Use the arrow keys or the rotary knob to scroll through the list and highlight the desired entry. You can also use the touch screen to make your selection.
To apply the selection, press Enter. To exit the selection list without making a change, press the Escape (Esc) key.
2-14 PN: 10580-00250 Rev. AA Cell Master UG
Loading...
+ 96 hidden pages