Anritsu MS2711A User's Guide

MS2711A
Hand Held
Spectrum Analyzer
User's Guide
Hand-Held Spectrum Analyzer for Measuring,
Monitoring and Analyzing Signal Environments
color cover part #: 00986-00040
WARRANTY
-
-
LIMITATION OF WARRANTY
The foregoing warranty does not apply to Anritsu connectors that have failed due to normal wear. Also, the warranty does not apply to defects resulting from improper or inadequate maintenance by the Buyer, unauthorized modification or misuse, or op eration outside the environmental specifications of the product. No other warranty is expressed or implied, and the remedies provided herein are the Buyer's sole and exclusive remedies.
-
TRADEMARK ACKNOWLEDGMENTS
MS-DOS, Windows, Windows for Workgroups, Windows NT, and Windows 95 are registered trademarks of the Microsoft Corporation. Anritsu and Cable Mate are trademarks of Anritsu Company.
NOTICE
Anritsu Company has prepared this manual for use by Anritsu Company personnel and customers as a guide for the proper installation, operation and maintenance of Anritsu Company equipment and computer programs. The drawings, specifications, and information contained herein are the property of Anritsu Company, and any un­authorized use or disclosure of these drawings, specifications, and information is prohibited; they shall not be reproduced, copied, or used in whole or in part as the basis for manufacture or sale of the equipment or software programs without the prior written consent of Anritsu Company.

Table of Contents

Chapter 1 - General Information 1-1
Introduction..................................1-1
Description ..................................1-1
Standard Accessories.............................1-1
Options ....................................1-2
Printers ....................................1-2
Optional Accessories .............................1-2
Performance Specifications .........................1-4
Preventive Maintenance ...........................1-6
Calibration ..................................1-6
Annual Verification..............................1-6
Anritsu Service Centers ...........................1-6
Chapter 2 - Quick Start Guide 2-1
Introduction..................................2-1
Turning the MS2711A On for the First Time ...............2-1
Front Panel Overview ............................2-2
Test Panel Connectors ............................2-5
Making Spectrum Analyzer Measurements ................2-6
Required Equipment......................................2-6
Procedure ............................................2-6
Making aMeasurement ....................................2-6
Selecting theFrequency ....................................2-6
Selecting theSpan .......................................2-6
Selecting theAmplitude ....................................2-7
Selecting BandwidthParameters . . .............................2-7
Selecting SweepParameters . . . . . .............................2-7
Adjusting Markers.......................................2-7
Adjusting Limits ........................................2-8
Adjusting AttenuatorSettings. ................................2-8
Making a Basic Measurement ........................2-9
Example – Measuring a 900 MHz signal ..................2-9
Set thecenter frequency ....................................2-9
Set thefrequency span ....................................2-10
Set theamplitude .......................................2-11
Activate themarker......................................2-12
Save theDisplay........................................2-13
Recall theDisplay .......................................2-13
Printing....................................2-13
Printer SwitchSettings ....................................2-13
Printing aScreen .......................................2-14
Determining Remaining Battery Life ...................2-15
Symbols ...................................2-16
Self Test ...................................2-16
June 2002 10580-00048 Copyright 2000-2002 Anritsu Company Revision: D
i
Error Codes .................................2-17
Self TestErrors ........................................2-17
Range Errors .........................................2-17
Using the Soft Carrying Case........................2-18
Chapter 3 - Key Functions 3-1
Introduction..................................3-1
Hard Keys............................................3-1
Soft Keys ............................................3-4
Chapter 4 - Advanced Measurement Fundamentals 4-1
Introduction..................................4-1
Advanced Measurement Fundamentals...................4-1
Chapter 5 - Field Measurements 5-1
Introduction..................................5-1
Occupied Bandwidth .............................5-1
XdB DownMethod.......................................5-1
N% Method...........................................5-1
Occupied Bandwidth Measurement with the MS2711A ..........5-1
Required Equipment......................................5-1
Procedure ............................................5-1
Adjacent Channel Power Leakage .....................5-2
Adjacent Channel Power Measurement with the MS2711A........5-2
Required Equipment......................................5-2
Procedure ............................................5-2
Out-of-Band Spurious Emission Measurements ..............5-3
Out-of-band Spurious Measurement with the MS2711A .........5-3
Required Equipment......................................5-3
Procedure ............................................5-3
In-Band/Out-Of-Channel Measurements..................5-4
In-band Spurious Measurement with the MS2711A ............5-4
Required Equipment......................................5-4
Procedure ............................................5-4
Field Strength.................................5-5
Required Equipment......................................5-5
Procedure ............................................5-5
ANTENNA CALCULATIONS . . . .............................5-6
Making Power Measurements with the Power Meter ...........5-7
Displaying Powerin dBm and Watts .............................5-7
Displaying RelativePower. . . . . . .............................5-7
AM/FM Demodulation ............................5-8
ii
Chapter 6 - Software Tools Program 6-1
Description ..................................6-1
Requirements .................................6-1
Communication Port Setting.........................6-1
Changing COM Port Settings–Windows 95/98/NT/2000 .........6-1
Software Installation .............................6-2
Plot Capture..................................6-3
Capture multipleto database . ................................6-3
Capture multipletraces to PC screen.............................6-4
Single tracecapture ......................................6-4
Program Operation..............................6-4
Entering Antenna Factors ..........................6-5
Uploading Antenna Factors .........................6-5
Saving a Plot as a Windows Metafile ....................6-6
Saving Data to a Spreadsheet ........................6-6
Saving Data as a Data File ..........................6-6
Saving Data to a Database ..........................6-6
“Drag-n-Drop” ................................6-7
Printing ....................................6-7
iii
Chapter 1 General Information
MODE
FREQ / SPAN
AMPLITUDE
MS2711A
BW/SWEEP
1
TRACE MEAS
3
RECALL
SAVE SETUP
SETUP
5
LIMIT
MARKER
7
RECALL
SAVE
DISPLAY
DISPLAY
9
PRINT
ON
OFF
ESCAPE
CLEAR
2
4
6
ENTER
8
SINGLE
CONT
+
0
-
/
SYS
.
Figure 1-1. Anritsu MS2711A Hand Held Spectrum Analyzer™
1-0
Chapter 1 General Information
Introduction
This chapter provides a description, performance specifications, optional accessories, pre ventive maintenance, and calibration requirements for the Anritsu Hand Held Spectrum An alyzer model listed below. Throughout this manual, this instrument may be referred to as an Anritsu HHSA.
Model
MS2711A 100 kHz to 3000 MHz
Frequency Range
Description
The Anritsu HHSA is a synthesized-based hand held spectrum analyzer that provides quick and accurate signal results. Measurements can be easily made by using the main instrument functions: frequency, span, amplitude and bandwidth. Dedicated keys for common func­tions and a familiar calculator type keypad are available for fast data entry. Automatic time and date stamping of measurement data of up to 200 results is provided as is storage and re­call of up to 10 different measurement setups. A large, high-resolution liquid crystal display (LCD) provides easy viewing in a variety of lighting conditions. The Anritsu HHSA is ca­pable of up to two and one-half hours of continuous operation from a fully charged battery and can be operated from a 12.5 Vdc source (which will also simultaneously charge the bat­tery). Built-in energy conservation features can be used to extend battery life over an eight-hour workday.
The Anritsu HHSA is designed for monitoring, measuring, and analyzing signal environ­ments. Typical measurements include: in-band interference, transmit spectrum analysis, an tenna isolation and cell area interference. A full range of marker capabilities such as peak, center and delta functions are provided for faster, more comprehensive measurement of dis played signals. Limit lines are available for creating quick, simple pass/fail measurements. A menu option provides for an audible beep when the limit value is exceeded. To permit use in low-light environments, the LCD can be back lit using a front panel key.
-
-
1
-
-
Standard Accessories
Anritsu HHSA Software Tools, a PC based software program, provides an on-line database record for storing measurement data. Software Tools can also convert the Anritsu HHSA display to a Microsoft Windows 95/98ä graphic. Measurements stored in the Anritsu HHSA internal memory can be down-loaded to a PC using the included null-modem serial cable. Once stored, the graphic trace can then be displayed, scaled, or enhanced with mark ers and limit lines. Historical graphs can be overlaid with current data using the PC mouse in a drag-and-drop fashion. The underlying data can be extracted and used in spreadsheets or for other analytical tasks.
-
1-1
Chapter 1 General Information
The following standard accessories are supplied with the MS2711A:
Soft Carrying Case
·
AC-DC Adapter
·
Automotive Cigarette Lighter 12 Volt DC Adapter
·
CD ROM containing the Software Tools program.
·
Serial Interface Cable (Null Modem Type)
·
Rechargeable NiMH Battery
·
One year Warranty (includes battery, firmware, and software)
·
User's Guide
·
Options
Option 5 — Add RF Wattmeter Power Monitor (RF detector not included)
·
Printers
2000-766 HP DeskJet Printer, with Interface Cable, Black Print Cartridge,
·
and U.S. Power Cable
2000-753 Serial-to-Parallel Converter Cable
·
2000-661 Black Print Cartridge
·
·
2000-662 Rechargeable Battery for DeskJet Printer
·
2000-663 Power Cable (Europe) for DeskJet Printer
·
2000-664 Power Cable (Australia) for DeskJet Printer
·
2000-665 Power Cable (U.K.) for DeskJet Printer
·
2000-667 Power Cable (So. Africa) for DeskJet Printer
·
2000-1008 Sieko DPU-414-30BU Thermal Printer with Internal
Battery, Thermal Printer Paper, Serial Cable, Power Cable
·
2000-755 Five rolls of Thermal Paper
·
2000-1002 U.S. Adapter for Seiko DPU-414-30B
·
2000-1003 Euro Adapter for Seiko DPU-414-30B
·
2000-1004 Battery Pack Adapter for Seiko DPU-414-30B
·
2000-1012 Serial 9-pin male to 9-pin female cable for Seiko DPU-414-30B
·
2000-1046 Serial-to-parallel Converter Cable w/ DIP switch labeling, 36-pin
female Centronics to DB25 female
Optional Accessories
Optional Accessories are shown in Table 1-1
1-2
Chapter 1 General Information
Table 1-1. Optional Accessories
Part Number Description
5400-71N50 42N50A-30
15NN50-1.5A Test Port Cable Armored, 1.5 meter, N(m) to N(m), 3.5 GHz 15NN50-3.0A Test Port Cable Armored, 3.0 meter, N(m) to N(m), 3.5 GHz 15NN50-5.0A Test Port Cable Armored, 5.0 meter, N(m) to N(m), 3.5 GHz 15NNF50-1.5A Test Port Cable Armored, 1.5 meter, N(m) to N(f), 3.5 GHz 15NNF50-5.0A Test Port Cable Armored, 3.0 meter, N(m) to N(f), 3.5 GHz 15NNF50-3.0A Test Port Cable Armored, 5.0 meter, N(m) to N(f), 3.5 GHz 15ND50-1.5A Test Port Cable Armored, 1.5 meter, N(m) to 7/16 DIN(m), 3.5 GHz 15NDF50-1.5A Test Port Cable Armored, 1.5 meter, N(m) to 7/16 DIN(f), 3.5 GHz 800-109 Detector Extender Cable, 7.6m (25 ft.) 800-110 Detector Extender Cable, 15.2m (50 ft.) 800-111 Detector Extender Cable, 30.5m (100 ft.) 800-112 Detector Extender Cable, 61m (200 ft.) 510-90 Adapter 7-16 (f) to N(m), 3.5 GHz 510-91 Adapter 7-16 (f) to N(f), 3.5 GHz 510-92 Adapter 7-16 (m) to N(m), 3.5 GHz 510-96 Adapter 7-16 DIN (m) to 7/16 DIN (m), 3.5 GHz 510-97 Adapter 7-16 DIN (f) to 7/16 DIN (f), 3.5 GHz 1091-26 Adapter, DC to 18 GHz, 50 Ohm, N(m)-SMA(m) 1091-27 Adapter, DC to 18 GHz, 50 Ohm, N(m)-SMA(f) 1091-172 Adapter, DC to 1.3 GHz, 50 Ohm, N(m)-BNC(f) 34NN50A Precision Adapter, DC to 18 GHz, 50 Ohm, N(m)-N(m) 34NFNF50 Precision Adapter, DC to 18 GHz, 50 Ohm, N(f)-N(f) 48258 Spare Soft Carrying Case 40-115 Spare AC/DC Adapter 806-62 Spare Automotive Cigarette Lighter/12 Volt DC adapter 800-441 Spare Serial Interface Cable 760-215A Transit Case for Anritsu Hand Held Spectrum Analyzer 633-27 Rechargeable Battery, NiMH 2300-347 Anritsu Hand Held Spectrum Analyzer Software Tools 10580-00048 Anritsu HHSA User’s Guide, Model MS2711A 10580-00049
10580-00050 2000-1029 Battery Charger (NiMH), with Universal Power Supply 2000-1030 Portable Antenna, 50 Ohm, SMA (m), 1.71 to 1.88 GHz 2000-1031 Portable Antenna, 50 Ohm, SMA (m), 1.85 to 1.99 GHz 2000-1032 Portable Antenna, SMA (m), 2.4 to 2.5 GHz, 50 Ohm 2000-1034 Portable Antenna, 50 Ohm, SMA (f), 806-960 MHz 2000-1035
70-28
RF Detector, N(m), 50 ohm, 1 to 3000 MHz Attenuator, 30 dB, 50 Watt, bi-directional
Anritsu HHSA Programming Manual, Model MS2711A Anritsu HHSA Maintenance Manual, Model MS2711A
Portable Antenna, 50 Ohm, SMA (m), 902-960 MHz Headset
1-3
Chapter 1 General Information
Performance Specifications
Performance specifications are provided in Table 1-2. Unless otherwise noted, specified values are obtained after a five minute warmup period at a constant ambient temperature. The typical values are provided for reference, and are not guaranteed.
Table 1-2. Performance Specifications (1 of 2)
Description Value
Frequency
Frequency Range 100 kHz to 3.0 GHz Frequency Reference
Aging Accuracy
Frequency Span 100 kHz to 3 GHz in 1, 2, 5 step selections in auto mode,
Sweep Time ³ 650 ms full span; 400 ms zero span SSB Phase Noise –75 dBc/Hz, 30 kHz offset@1GHz Spurious Responses
Input Related £-45 dBc
Spurious
Residual Responses £-95 dBm
±1 ppm/yr ±2 ppm
plus zero span
Resolution Bandwidth
Selections 10 kHz, 30 kHz, 100 kHz and 1 MHz Accuracy
Video Bandwidth
Selection 100 Hz to 300 kHz in 1-3 sequence
Amplitude
Measurement Range +20 dBm to –97 dBm Displayed Average Noise
Level: £ –97 dBm (full span) typical Dynamic Range > 65 dB Total Level Accuracy ±2 dB ³ 200 kHz, typical
Maximum Safe Input Level +20 dBm, maximum measurable safe input
Amplitude Units
Log Scale Mode dBm, dBV, dBmV, dBmV Display Range 2 to 15 dB/division, in 1 dB steps, 10 divisions displayed
Attenuator Range: 0 to 50 dB, selected manually or automatically
± 20% typical
±3 dB < 200 kHz, typical
+27 dBm, maximum damage +27 dBm, peak pulse power +50 Vdc
coupled to the reference level. Resolution in 10 dB steps.
1-4
Chapter 1 General Information
Table 1-2. Performance Specifications (2 of 2)
Description Value
Display
Type Monochrome LCD with back light capability Resolution 640 ´ 480 Marker Modes Standard
delta marker to peak marker to center
Memory
Trace Storage 200 traces Setup Storage 10 setups Displayed Traces 2 traces
Inputs
RF Input 50 W Connector Female, Type N Maximum Input Level + 20 dBm, + 50 Vdc RF Input VSWR 2.0:1
RS232 Interface
Type Null modem Baud Range 9600 to 56k baud
Printer Interface
Drivers Epson ESC/P
Epson ESC/P RAST Epson ESC/P2 HP PCL3
General Characteristics
Dimensions 10 ´ 7 ´ 2.25 inches
25.4 ´ 17.8 ´ 6.1 centimeters Weight 4 pounds (1.8 kg) including battery Power Requirements NIMH battery: 10.8 volts, 1800 mA maximum
External DC input: +11 to +15 Vdc, 1250 mA max.
Temperature Operating: 0 to +50° C, 85% or less humidity
Non-operating: –20 to +75° C Electromagnetic Compatibility Meets European community requirements for CE marking Safety Conforms to EN 61010-1 for Class 1 portable equipment
1-5
Chapter 1 General Information
Preventive Maintenance
Anritsu HHSA preventive maintenance consists of cleaning the unit and inspecting and cleaning the RF connector on the instrument and all accessories.
Clean the Anritsu HHSA with a soft, lint-free cloth dampened with water or water and a mild cleaning solution.
CAUTION: To avoid damaging the display or case, do not use solvents or abra sive cleaners.
Clean the RF connectors and center pins with a cotton swab dampened with denatured alco hol. Visually inspect the connectors. The fingers of the N (f) connectors and the pins of the N(m) connectors should be unbroken and uniform in appearance. If you are unsure whether the connectors are good, gauge the connectors to confirm that their dimensions are correct.
Visually inspect the test port cable(s). The test port cable should be uniform in appearance, not stretched, kinked, dented, or broken.
Calibration
The Anritsu HHSA loads factory calibration data during start-up, eliminating the need for daily calibration checks.
Annual Verification
Although the Anritsu HHSA does not require daily field calibration, Anritsu recommends an annual calibration and performance verification by local Anritsu service centers. Anritsu service centers are listed in Table 1-3 beginning on the following page.
The Anritsu HHSA itself is “self calibrating”, meaning that there are no field-adjustable components.
-
-
Anritsu Service Centers
Table 1-3 provides a listing of the Anritsu Service Centers.
1-6
Table 1-3. Anritsu Service Centers
Chapter 1 General Information
UNITED STATES
ANRITSU COMPANY 685 JarvisDrive Morgan Hill,CA 95037-2809 Telephone: (408)776-8300 FAX: 408-776-1744
ANRITSU COMPANY 10 NewMapleAve., Suite305 Pine Brook,NJ 07058 Telephone: 973-227-8999 FAX: 973-575-0092
ANRITSU COMPANY 1155E. Collins Blvd Richardson, TX75081 Telephone: 1-800-ANRITSU FAX: 972-671-1877
AUSTRALIA
ANRITSU PTY.LTD. Unit 3,170 Foster Road Mt Waverley,VIC 3149 Australia Telephone: 03-9558-8177 FAX: 03-9558-8255
BRAZIL
ANRITSU ELECTRONICALTDA. Praia deBotafogo 440. Sala 2401 CEP22250-040,Rio deJaneiro,RJ, Brasil Telephone: 021-527-6922 FAX: 021-53-71-456
CANADA
ANRITSU INSTRUMENTSLTD. 700 SilverSeven Road, Suite 120 Kanata, OntarioK2V 1C3 Telephone: (613)591-2003 FAX: (613) 591-1006
CHINA (SHANGHAI)
ANRITSU ELECTRONICSCO LTD 2F,Rm.B, 52Section Factory Bldg. NO 516Fu Te Road (N) Waigaoqiao Free Trade Zone Pudong, Shanghai200131 PR CHINA Telephone: 86-21-58680226 FAX: 86-21-58680588
FRANCE
ANRITSU S.A 9 Avenue du Quebec Zone deCourtaboeuf 91951 LesUlis Cedex Telephone: 016-09-21-550 FAX: 016-44-61-065
GERMANY
ANRITSU GmbH Grafenberger Allee54-56 D-40237 Dusseldorf Germany Telephone: 0211-968550 FAX: 0211-9685555
INDIA
MEERA AGENCIES (P) LTD A-23 HauzKhas New Delhi,India 110 016 Telephone: 011-685-3959 FAX: 011-686-6720
ISRAEL
TECH-CENT, LTD 4 RaulValenberg St. Tel-Aviv,Israel 69719 Telephone: 972-36-478563 FAX: 972-36-478334
ITALY
ANRITSU Sp.A Rome Office Via E. Vittorini, 129 00144 RomaEUR Telephone: (06)50-2299-711 FAX: 06-50-22-4252
JAPAN
ANRITSU CUSTOMER SERVICELTD. 1800 OnnaAtsugi—shi Kanagawa-Prf. 243Japan Telephone: 0462-96-6688 FAX: 0462-25-8379
KOREA
ANRITSU SERVICE CENTER 8F SanwonBldg. 1329-8 Seocho-Dong Seocho-Ku Seoul, Korea137-070 Telephone: 82-2-581-6603 FAX: 82-2-582-6603
SINGAPORE
ANRITSU (SINGAPORE)PTE LTD 10, HoeChiang Road #07-01/02 Keppel Towers Singapore 089315 Telephone:65-282-2400 FAX:65-282-2533
SOUTH AFRICA
ETESCSA 12 SurreySquare OfficePark 330 SurreyAvenue Ferndale, Randburt,2194 South Africa Telephone: 011-27-11-787-7200 Fax: 011-27-11-787-0446
SWEDEN
ANRITSU AB Botvid Center Fittja Backe13A 145 84 Stockholm, Sweden Telephone: (08)534-707-00 FAX: (08)534-707-30
TAIWAN
ANRITSU CO.,LTD. 6F, No.96, Section 3 Chien KuoN. Road Taipei, Taiwan, R.O.C. Telephone: (02)515-6050 FAX: (02) 509-5519
UNITED KINGDOM
ANRITSU LTD. 200 CapabilityGreen Luton, Bedfordshire LU1 3LU,England Telephone: 015-82-43-3200 FAX: 015-82-73-1303
1-7
Chapter 2 Quick Start Guide
MS2711A
MODE
FREQ / SPAN
AMPLITUDE
BW/SWEEP
Figure 2-1. MS2711A Power ON/OFF Key
1
TRACE MEAS
3
RECALL
SAVE SETUP
SETUP
5
LIMIT
MARKER
7
RECALL
SAVE
DISPLAY
DISPLAY
9
PRINT
ON OFF
ESCAPE
CLEAR
2
4
6
ENTER
8
SINGLE
CONT
+
0
-
/
SYS
.
ON/OFF Key
2-0
Chapter 2 Quick Start Guide
Introduction
This chapter provides a brief overview of the Anritsu MS2711A HHSA. The intent of this manual is to provide the user with a starting point for making basic measurements. For more detailed information, users may want to consult Chapter 3, Key Functions or Chapter 4, Advanced Measurement Functions.
Turning the MS2711A On for the First Time
The Anritsu MS2711A HHSA is a lightweight, hand held, battery operated unit designed specifically for field environments and applications requiring mobility. It is capable of up to
2.5 hours of continuous operation from a fully charged, field-replaceable battery. Built-in energy conservation features allow battery life to be extended over an eight-hour workday.
The Anritsu MS2711A HHSA can also be operated from a 12.5 Vdc source (which will also simultaneously charge the battery). This can be achieved with either the Anritsu AC-DC Adapter (P/N 40-115) or 12.5 Vdc Automotive Cigarette Lighter Adapter (P/N 806-62). Both items are included as standard accessories (see Chapter 1).
2
Press the ON/OFF front panel button (Figure 2-1, facing page).
The HHSA takes about five seconds to perform a series of self-diagnostic and adjustment routines. At completion, the screen displays the Anritsu logo, the model number, the version of firmware. It also prompts you to press ENTER to continue.
The Anritsu HHSA is now ready for operation. No additional keystrokes or in stallation are required. For information on making measurements with the Anritsu HHSA, refer to “Making a Basic Measurement,” on page 2-9. For ad vanced applications, refer to Chapter 4, “Advanced Measurement Fundamentals.”
-
-
2-1
Chapter 2 Quick Start Guide
Front Panel Overview
The Anritsu HHSA menu-driven user interface is easy to use and requires little training. Hard keys on the front panel are used to initiate function-specific menus. There are four function hard keys located below the display, Mode, Frequency/Span, Amplitude and Band width/Sweep.
-
There are seventeen keypad hard keys located to the right of the display. Twelve of the key pad hard keys are dual purpose, depending on the current mode of operation. The dual pur pose keys are labeled with one function in black, the other in blue.
There are also six soft keys that change function depending upon the current mode selec
­tion. The current soft key function is indicated in the active function block to the right of the display. The locations of the different keys are illustrated in Figure 2-2, below.
Soft Keysctive Function Block
MODE
FREQ / SPAN
AMPLITUDE
MS2711A
BW/SWEEP
TRACE MEAS
SAVE SETUP
LIMIT
SAVE DISPLAY
ON OFF
ESCAPE
CLEAR
2
1
3
4
RECALL SETUP
6
5
MARKER
ENTER
8
7
SINGLE
RECALL
CONT
DISPLAY
PRINT
+
0
-
/
SYS
.
9
Keypad Hard Keys
-
-
Function Hard Keys
Figure 2-2. MS2711A Soft Keys
Figures 2-3 and 2-4 illustrate the menu structures and soft key labels for each menu selec tion. Refer to Chapter 3, Key Functions, for more detailed hard and soft key descriptions.
2-2
-
MODE MENU SOFT KEYS
A
SPECTRUM
NALYZER
(DEFAULT)
FREQ/SPAN
CENTER
START
STOP
SPAN
CENTER SPAN START
GHz
EDIT
MHz
FULL
ZERO
STOP
kHz
Chapter 2 Quick Start Guide
Hz
SPAN
UP
1-2-5
SPAN
DOWN
1-2-5
BACK
AMPLITUDE
ATTEN
UNITS
BW/SWEEP
RBW
VBW
POWER MONITOR
DETECTION
POWER MONITOR
REF
LEVEL
AUTO
dBm
RBW
AUTO
AUTO
POSITIVE
PEAK
UNITS REL
SCALE
MANUAL
dBV
VBW
MANUAL
MANUAL
AVERAGE
ATTEN
dBmV dBuV
MAX
HOLD
NEGATIVE
PEAK
OFFSET ZERO
EDIT
EDIT
EDIT
UNITS
DETEC-
TION
BACK
BACK
BACK
BACK
Figure 2-3. Function Hard Keys Menu Structure
2-3
Chapter 2 Quick Start Guide
A
MODE MENU SOFT KEYS
SPECTRUM
NALYZER
(DEFAULT)
TRACE
TRACE
RESET
A
B
VIEW
A->B
B
CLEAR
A-B
-> A
RECALL
B
TRACE
->B
A+B
-> A
TRACE
B
BACK
%
TYPE
MAIN
BW
EDIT
2
2
AM/FM
DEMOD
VOLUME
CHANNEL
MULTIPLE
LOWER
LIMITS
SEGMENT
SEGMENT
M3
MEAS
FIELD
STRNGTH
OBW
AM/FM
DEMOD
ACP
LIMIT
SINGLE
LIMIT
MULTIPLE
UPPER LIMITS
MULTIPLE
LOWER
LIMITS
MARKER
FIELD
STRNGTH
ON/OFF
METHOD
ON/OFF
CENTER
FREQ
SINGLE
LIMIT
ON/OFF
SEGMENT
SEGMENT
M1
1
1
OBW
SELECT
ANTENNA
DEMOD
CHANNEL
MULTIPLE
UPPER
LIMITS
SEGMENT
SEGMENT
M2
dBc
ADJ
BW
BEEP
AT
LEVEL
3
3
CHANNEL
POWER
CHANNEL
SPACING
LIMIT BEEP
SEGMENT
4
SEGMENT
4
M4
ACP
MEASURE
MEASURE
SEGMENT
5
SEGMENT
5
ALL
OFF
BACK
BACK
BACK
BACK
BACK
BACK
BACK
ON/OFF
ON/OFF
OPTIONS
PRINTER
HOUR
SYS
OPTIONS
CLOCK
M1
M2 M3 M4
Figure 2-4. Keypad Hard Keys Menu Structure
2-4
EDIT
EDIT
CLOCK
CHANGE
DATE
FORMAT
MINUTE
DELTA
(Mx - M1)
SELF
TEST
MONTH
MARKER
TO
PEAK
MARKER
TO
PEAK
DAY
MARKER
FREQ TO
CENTER
MARKER
FREQ TO
CENTER
YEAR
BACK
BACK
BACK
Chapter 2 Quick Start Guide
Test Panel Connectors
The connectors and indicators located on the test panel are listed and described below.
12.5-15VDC (1100 mA)
When using the AC-DC Adapter, always use a three-wire power cable con nected to a three-wire power line outlet. If power is supplied without grounding the equipment in this manner, there is a risk of receiving a severe or fatal electric shock.
Battery Charging
External Power
Headphone Jack
Serial Interface
RF In 50-Ohm RF input for spectrum analysis and stimulus-response measurements.
RF Det RF detector input for the Power Monitor (Option 5).
DC input to power the unit or for battery charging. Input is 12.5 to 15 Vdc @ 1100 mA.
WARNING
-
Indicator light to show that the battery is being charged. The indicator light auto matically shuts off when the battery is fully charged.
Indicator light to show that the Anritsu is being powered by the external charg ing unit.
Allows connection of audio headphones for monitoring AM/FM demodulation.
RS232 DB9 interface to a COM port on a personal computer (for use with the Anritsu Software Tools program). Also provides a serial interface to a printer.
-
-
Battery Compartment
Contains the NiMH battery, part number 633-27. To open, turn the fastener ¼-turn counter-clockwise, pull up and remove the battery compartment door. Remove the battery by pulling straight up on the battery lanyard. Replacement is the opposite of removal.
2-5
Chapter 2 Quick Start Guide
Making Spectrum Analyzer Measurements
Required Equipment
MS2711A Hand Held Spectrum Analyzer
·
Test Port Extension Cable, Anritsu 15NNF50-1.5A
·
Procedure
Step 1. Press the ON/OFF key.
Step 2. Press the MODE key and use the Up/Down arrow key to select Spectrum Ana
lyzer mode. Press ENTER to set the mode.
Making a Measurement
Step 1. Step 2. Locate and display the signal(s) of interest by selecting the desired frequency,
Connect the input cable to the RF In test port.
span, and amplitude value.
Selecting the Frequency
Step 1. Step 2.
Press the FREQ/SPAN key to display the Frequency menu. To enter a center frequency, select the CENTER soft key and enter the desired
center frequency on the keypad. Select the GHz, MHz, kHz,orHz softkey to accept the center frequency input.
or
To set a specific frequency band, select the START soft key and enter the de­sired start frequency on the keypad. Select the GHz, MHz, kHz,orHz softkey to accept the start frequency input. Then select the STOP soft key and enter the desired stop frequency on the keypad. Select the GHz, MHz, kHz,orHz softkey to accept the stop frequency input.
Selecting the Span
-
Step 1. Step 2. Step 3.
NOTE: To quickly move the span value up or down, select the SPAN UP 1-2-5 or SPAN DOWN 1-2-5 soft keys. These keys facilitate a zoom-in, zoom-out in a 1-2-5 sequence.
Press the FREQ/SPAN key to display the Frequency menu. Press the SPAN soft key to display the Span menu. For a full span, select the FULL soft key.
or
For no span, select the ZERO soft key.
or
To set a specific span, select the EDIT soft key and enter the desired span on the keypad. Select the GHz, MHz, kHz,orHz soft key to accept the span.
2-6
Selecting the Amplitude
Chapter 2 Quick Start Guide
Step 1. Step 2.
Step 3.
Step 4.
Step 5.
NOTE: Select AUTO coupling to help insure that harmonics and spurs are not introduced into the measurements.
Press the AMPLITUDE key. Press the REF LEVEL soft key and use the up/down arrow key or directly enter
the desired reference level from the keypad. Press ENTER to set the amplitude level.
Press the SCALE soft key and use the up/down arrow key or directly enter the desired scale from the keypad. Press ENTER to set the scale.
Press the UNITS soft key and select the desired units from the soft keys pre sented. Press ENTER to set the units selection.
Press the ATTEN soft key to set the attenuation. Select either the AUTO or MANUAL soft key. AUTO automatically couples the attenuator setting to the reference level. MANUAL allows editing of the attenuator setting. Press the EDIT soft key and use the up/down arrow key to select the attenuator level. Press ENTER to set the attenuator level.
Selecting Bandwidth Parameters
Step 1. Step 2.
Press the BW/SWEEP key to display the bandwidth menu. To select the resolution bandwidth, press the RBW soft key.
-
Step 3.
Step 4. Step 5.
Select AUTO for automatic resolution bandwidth selection, or select EDIT and use the Up/Down arrow key to select the resolution bandwidth. Press ENTER to set the resolution bandwidth.
To select the video bandwidth, press the VBW soft key. Select AUTO for automatic video bandwidth selection, or select EDIT and use
the Up/Down arrow key to select the video bandwidth. Press ENTER to set the video bandwidth.
Selecting Sweep Parameters
Step 1.
Step 2.
To toggle maximum hold ON or OFF, press the MAX HOLD soft key. Maxi mum hold displays the maximum response of the input signal.
To set the detection method, press the DETECTION soft key and select either POSITIVE PEAK, AVERAGE, or NEGATIVE PEAK detection.
Adjusting Markers
Step 1. Step 2. Step 3.
Press the MARKER key to call up the Markers menu. Press the M1 soft key to select the M1 marker function. Press the EDIT soft key and enter an appropriate value using the keypad or
Up/Down arrow key. Select the GHz, MHz, kHz,orHz softkey to accept the marker frequency input. Pressing the ON/OFF soft key activates or deactivates the M1 marker function.
-
2-7
Chapter 2 Quick Start Guide
Step 4. Step 5. Repeat the steps for markers M2, M3, and M4.
Press the BACK soft key to return to the Markers Menu.
Adjusting Limits
Step 1. Step 2. Enter the desired numerical value using the keypad or Up/Down arrow key. Step 3.
Press the LIMIT key.
Press ENTER when the data entry is complete.
Adjusting Attenuator Settings
Step 1. Step 2. Step 3.
Step 4.
NOTE: AUTO coupling helps insure that harmonics and spurs are not intro­duced into the measurements.
Press the AMPLITUDE key. Press the ATTEN soft key. Select AUTO to automatically couple the attenuator setting to the reference
level. Select EDIT to adjust the attenuation setting. Use the Up/Down arrow key to se
lect the attenuation setting and press ENTER to set.
-
2-8
Chapter 2 Quick Start Guide
Making a Basic Measurement
Making a basic measurement with the Anritsu HHSA is similar to conventional spectrum analyzers. Users need simply to power-on and tune the Anritsu Hand Held Spectrum Ana lyzer to locate and display a signal on the screen of the hand held spectrum analyzer. Once a signal is displayed the user can measure the signal input in four simple steps to determine the frequency and amplitude of the signal.
These steps include:
Setting the center frequency.
r
Setting the frequency span.
r
Setting the amplitude.
r
Activating the marker.
r
-
NOTE: Frequency, span and amplitude are the fundamental functions for spec trum analyzer measurements. However, by using marker functions, you are able to easily read out frequencies and amplitudes on the spectrum analyzer trace. This lets you make relative measurements, automatically locate the sig­nal of the highest amplitude on a trace, and tune the spectrum analyzer to track a signal automatically. For more information, please refer to Chapter 4, “Ad­vanced Measurement Fundamentals.”
Example – Measuring a 900 MHz signal
Step1.
Step2. Connect a signal generator to the input of the Anritsu HHSA and configure it to
Set the center frequency
Step1.
Step2.
NOTE: To set the center frequency to 900 MHz, use the numerical keypad to the right of the display. The data keys allow you to select the exact numeric value of the active function, which in this example is the center frequency. When activating the center frequency function, the hand held spectrum ana lyzer is set to the center-frequency span mode.
Press the ON/OFF key, then the ENTER key when prompted.
provide a -10dBm, 900 MHz signal.
Press the FREQ/SPAN key. Press the CENTER frequency soft key.
-
-
Step3.
Enter 900 on the keypad and select the MHz soft key . Observe that the signal resembles that shown in Figure 2-5.
2-9
Chapter 2 Quick Start Guide
Figure 2-5. Signal at 900 MHz
Set the frequency span
Step 1. Press the FREQ/SPAN key. Step 2.
The Span menu is now displayed in the active function block. When activating the span function, the hand held spectrum analyzer is set to the center-fre­quency span mode.
Choices within the span menu include EDIT, ZERO, FULL SPAN, SPAN UP 1-2-5, and SPAN DOWN 1-2-5. The span can be set to 0 Hz using either the data keys or activating ZERO SPAN in the span menu. To reduce the fre­quency span ¾ for example, to 20 MHz ¾ either press 20 on the keypad and press the MHz soft key, or use the down arrow key to “step down” to the value. (Like data keys, step keys can also be used to change the numeric value of the active function.) A span of 20 MHz is shown in Figure 2-6.
Press SPAN soft key.
Figure 2-6. 20 MHz Frequency Span
2-10
Set the amplitude
Chapter 2 Quick Start Guide
Generally, placing the signal peak at the reference level provides the best measurement ac curacy. The following steps will adjust the signal peak to the reference level (Figure 2-7).
Step 1. Press the AMPLITUDE key. Step 2.
Step 3. Step 4. Press the +/– key and 10 on the keypad and press ENTER to set the reference
Press the ATTEN soft key and the AUTO soft key to select automatic attenua tion.
Press BACK soft key and then the REF LEVEL soft key.
level to –10 dBm. Observe that the display resembles that shown in Figure 2-7.
-
-
Figure 2-7 Reference Level set at –10 dBm
Step 5.
Step 6.
Press the SCALE soft key and use the key pad or up/down arrow key to select a scale of 10 dB/. Press the ENTER key to set the scale.
Press the UNITS soft key and select dBm as the amplitude units.
2-11
Chapter 2 Quick Start Guide
Activate the marker
The marker reads both the frequency and the amplitude, and it displays these values in the message area at the bottom of the display. In this case, the marker will read 900 MHz and –10.00 dBm, as shown in Figure 2-8.
Step 1. Press the MARKER key. Step 2. Step 3. Step 4.
Step 5. Read both frequency and the amplitude values identified by the selected marker.
Press the M1 soft key . Press the ON/OFF soft key to activate the selected marker. Press the MARKER TO PEAK soft key to set marker M1 to the highest point
on the trace.
These values are displayed in the message area at the bottom of the display (Figure 2-8).
Figure 2-8. Activated Normal Marker
NOTE: Marker’s can be placed the peak of the signal by pressing PEAK, or by using the up/down arrow key to move the marker manually. When using more than one marker, it may be desirable to use the marker delta function, defined as delta (M1–Mx) in the active function box. The marker delta function reads the difference in amplitude and frequency between two signals and displays the values in the area below the display.
2-12
Save the Display
Chapter 2 Quick Start Guide
Step 1. Step 2. To enter a name for the display, press the soft key letter group that contains the
Step 3.
Press the SAVE DISPLAY key.
desired letter, then select the soft key for that letter. Continue to select letters up to a maximum of 16 characters.
Press ENTER to set the saved display name.
Recall the Display
Step 1. Step 2. Select the desired display using the Up/Down arrow key. Step 3.
Press the RECALL DISPLAY key.
Press ENTER to recall the selected display.
Printing
Printing is accomplished using the Seiko DPU-414 thermal printer, the Hewlett Packard DeskJet 340 ink jet printer, or any printer listed in theMS2711A printer selection table . Figure 2-9 shows a setup diagram for two types of printers. Refer to the printer manual for setup details.
Printer Switch Settings
Set the switches on the serial-to-parallel interface cable to the HP Deskjet 340 ink jet printer or other parallel interface printer as follows:
SW1 SW2 SW3 SW4 SW5 SW6 SW7 SW8
OFF ON OFF OFF OFF OFF ON OFF
Set the switches, SW1, SW2, and SW3, on the Seiko DPU-414 thermal printer as follows:
Switch
SW1 OFF ON ON ON ON OFF ON ON SW2 ON ON ON ON ON ON ON OFF SW3 ON ON ON OFF OFF ON ON ON
1 2 3 4 5 6 7 8
2-13
Chapter 2 Quick Start Guide
Printing a Screen
Step 1. Connect the printer as shown in Figure 2-9.
SEIKO
PRINTER
SERIAL CABLE
2000-756
HP 340
DESKJET
SERIAL-TO-PARALLEL
INTERFACE CABLE
2000-753
Figure 2-9. Printer Setup
Step 2. Obtain the desired measurement display Step 3. Step 4.
Press the SYS key and the OPTIONS soft key. Press the PRINTER soft key and select from the displayed menu of supported
printers..
Step 5. Step 6.
Press the PRINT key. Press ENTER.
2-14
Chapter 2 Quick Start Guide
Determining Remaining Battery Life
When the AC-DC adapter is disengaged from the Site Master, a battery indicator symbol is continuously displayed at the top-left corner of the display (Figure 2-10). A totally black bar indicates a fully charged battery.
Battery Monitor
Figure 2-10. Battery Monitor
During operation, the battery condition can be viewed by pressing the SYS key and select­ing the SELF TEST soft key. The battery condition will be displayed as a percentage of charge remaining.
SELF TEST
VOLTAGE........ .BATTERY 11.1V
TEMPERATURE.... 24°C
BATTERY CAL......PASSED
MEMORY..........PASSED
LO................PASSED
BATTERY CHARGE = 84%
PRESS ENTER TO CONTINUE
Figure 2-11. Self Test Battery Condition Display
2-15
Chapter 2 Quick Start Guide
Symbols
Table 2-1 provides a listing of the symbols used as condition indicators on the LCD display.
Table 2-1. LCD Icon Symbols
Icon Symbol
Anritsu HHSA is in power conservation mode. To resume sweeping,
HOLD
T
press the ENTER key. After 10 minutes without a key press, the Anritsu HHSA will automatically enter power conservation
Lock fail indication. Check battery. (If the Anritsu HHSA fails to lock with a fully charged battery, call your Anritsu Service Center.)
Indicates the remaining charge on the battery. The inner white rectangle grows longer as the battery charge depletes.
Self Test
At turn-on, the Anritsu HHSA runs through a series of quick checks to ensure the system is functioning properly. Note that the battery voltage and temperature are displayed in the lower left corner below the self test message. If the battery is low, or if the ambient temper­ature is not within the specified operational range, Self Test will fail. If Self Test fails AND the battery is fully charged AND the Anritsu HHSA is within the specified operating range, call your Anritsu Service Center.
2-16
Error Codes
Self Test Errors
A listing of Self Test Error messages is given in Table 2-2.
Table 2-2. Self Test Error Messages
Chapter 2 Quick Start Guide
Error Message
BATTERY LOW
EXTERNAL POWER LOW
PLL FAILED
EEPROM R/W FAILED
OUT OF TEMP. RANGE
NOTE: A listing of current Anritsu service centers is givenin Table 1-3 , page 1-7.
Battery voltage is less than 10 volts. Charge battery. If condition per sists, call your Anritsu Service Center.
External supply voltage is less than 10 volts. Call your Anritsu Ser vice Center
Phase-locked loops failed to lock. Charge battery. If condition per sists with a fully charged battery, call your Anritsu Service Center
Non-volatile memory system has failed. Call your Anritsu Service Center.
Ambient temperature is not within the specified operating range. If the temperature is within the specified operating range and the con­dition persists, call your Anritsu Service Center.
Description
Range Errors
A listing of Range Error messages is given in Table 2-3.
Table 2-3. Range Error Messages
Error Message
Description
-
-
-
NO STORED SWEEP AT THIS LOCATION
USE OPTIONS MENU TO SELECT A PRINTER
CANNOT ZERO NO DETECTOR INSTALLED
CANNOT ZERO INPUT SIGNAL TOO HIGH
POWER MONITOR OPTION NOT INSTALLED
Attempting to recall a display from a location that has not been previ ously written to. That is, the location does not contain stored sweep.
Attempting to print a display with no printer selected. Select a printer, then retry.
Attempting to perform a Power Monitor zero adjust function with no RF detector connected to the Anritsu HHSA.
Attempting to perform a Power Monitor zero adjust function with an input of greater than –20 dBm.
Attempting to enter Power Monitor mode with no option 5 installed.
2-17
-
Chapter 2 Quick Start Guide
Using the Soft Carrying Case
The soft carrying case has been designed such that the strap can be unsnapped to allow the case to be easily oriented horizontally; thus allowing the Anritsu controls to be more easily accessed (Figure 2-12).
Figure 2-12. Using the Soft Carrying Case
2-18
Chapter 3 Key Functions
Introduction
This chapter provides a description of the Anritsu MS2711A Hand Held Spectrum Analyzer keypad controls. There are two kinds of keys available on the MS2711A front panel: Hard Keys and Soft Keys.
Hard Keys
Hard keys are those keys on the front panel that are labeled in black or blue and perform specific functions as explained below. There are four function hard keys, located below the display, and seventeen keypad hard keys located to the right of the display. Twelve of the keypad hard keys are dual purpose, depending on the current mode of operation.
Function Hard Keys
MODE Sets the Anritsu HHSA to a specific mode of operation. Use the Up/Down
arrow key to select either spectrum analyzer mode or power monitor mode (if Option 5 is installed). Press the ENTER key to implement the selec­tion.
3
FREQ/SPAN Activates a menu of frequency and span related functions for selection us-
ing the soft keys. Available choices from this menu include CENTER, SPAN, START frequency and STOP frequency. Soft key selection of any of these functions brings up a set of choices or selections specific to that procedure.
AMPLITUDE
BW/SWEEP Activates a menu of bandwidth and sweep related functions including
Activates a menu of amplitude related functions including REF LEVEL, SCALE, ATTEN, and UNITS. Soft key selection of any of these func
tions brings up a set of choices or selections specific to that procedure.
RBW, VBW, MAX HOLD, and DETECTION. Soft key selection of any of these functions brings up a set of choices or selections specific to that procedure.
-
3-1
Chapter 3 Key Functions
Keypad Hard Keys
The following keypad hard key functions are printed in black on the keypad keys.
0-9 These keys are used to enter numerical data as required to setup or per
form measurements.
+/–
·
ESCAPE
CLEAR
UP/DOWN
ARROWS
ENTER Implements the current action or parameter selection.
ON
OFF
SYS
The plus/minus key is used to enter positive or negative values as required to setup or perform measurements.
The decimal point is used to enter decimal values as required to setup or perform measurements.
Exits the present operation or clears the display. If a parameter is being edited, pressing this key will clear the value currently being entered and restore the last valid entry. Pressing this key again will close the parame ter. During normal sweeping, pressing this key will move up one menu level.
Increments or decrements a parameter value. The specific parameter value affected typically appears in the message area of the LCD.
Turns the Anritsu HHSA on or off. When turned on, the system state at the last turn-off is restored. If the ESCAPE/CLEAR key is held down while the ON/OFF key is pressed, the factory preset state will be restored.
Allows selection of system setup parameters. Choices are OPTIONS, CLOCK, and SELF TEST.
-
-
3
The following keypad hard key functions are printed in blue on the keypad keys.
Turns the liquid crystal display (LCD) back-lighting ON or OFF. (Leaving back lighting off conserves battery power.)
LCD Contrast adjust. Use the Up/Down arrow key and ENTER to adjust the display contrast to suit ambient conditions.
TRACE Activates a menu that allows comparison of a current runtime trace
(Trace A) to a saved trace (Trace B).
MEAS Activates a menu of measurement related functions including field
strength, occupied bandwidth (OBW), and AM/FM Demodulation.
SAVE SETUP
RECALL SETUP
Saves the current system setup to 1 of 10 internal non-volatile memory lo cations. When the key is pressed, a Save Setup selection box appears on the display. Use the Up/Down arrow key to select a setup location and press the ENTER key to implement.
Recalls a previously saved setup from memory location 1 through 10. When the key is pressed, a Recall Trace selection box appears on the dis play. Select a setup using the Up/Down arrow key and press the ENTER key to implement. Setup location 0 recalls the factory preset state.
3-2
-
-
Chapter 3 Key Functions
LIMIT Activates a menu of Limit functions including single limit, multiple upper
limits, multiple lower limits, and limit beep.
MARKER Activates a menu of marker functions for markers 1 through 4. Each
marker, when selected, activates marker to peak and marker frequency to center functions. Markers 2 through 4 also include delta marker functions referenced to M1.
SAVE DISPLAY
RECALL DISPLAY
SINGLE
CONT
PRINT Prints the current display to the selected printer.
Saves the displayed trace to non-volatile memory. When the key is pressed, TRACE NAME: appears on the display. Select the alphanumeric characters for that trace name and press ENTER key to implement.
Recalls a previously saved trace from memory. When the key is pressed, a Recall Trace selection box appears on the display. Select a trace using the Up/Down arrow key and press the ENTER key to implement.
Toggles the units sweep between continuous sweep mode and single sweep mode. The default is continuous sweep. When single sweep is selected, the Anritsu HHSA will sweep once and hold until activated. The currently selected sweep mode appears on the LCD. Single sweep mode can be used to conserve battery power.
3-3
Chapter 3 Key Functions
Soft Keys
Soft keys are the six unlabeled keys (see Figure 2-2, page 2-2) that perform different func tions depending on the currently selected soft key menu. The available selections for the softkeys are displayed in the active function block.
The following paragraphs relate the menus displayed in the active function block when the various function hard keys or keypad hard keys are selected.
Frequency Menu
Pressing the FREQ/SPAN key activates a menu of frequency related choices in the active function block. Use the corresponding softkey to select the desired function, explained below. Use the keypad or Up/Down arrow key to change or enter values shown in the mes sage area. Complete the entry by pressing the appropriate softkey frequency designation.
CENTER Activates the center frequency function and sets the Anritsu HHSA to the
center frequency. A specific center frequency can be entered by using the keypad or Up/Down arrow key. Select the GHz, MHz, kHz,orHz softkey to accept the center frequency input.
START Activates the start frequency and sets the Anritsu HHSA in the Start/Stop
mode. Select the desired start frequency and complete the entry by press­ing the GHz, MHz, kHz,orHz softkey to accept the start frequency in- put.
STOP Activates the stop frequency and sets the Anritsu HHSA in the Start/Stop
mode. Select the desired stop frequency and complete the entry by press­ing the GHz, MHz, kHz,orHz softkey to accept the stop frequency input.
-
-
Span Menu
Pressing the SPAN soft key activates a menu of span related choices in the active function block. Use the corresponding softkey to select the desired function, explained below. Use the keypad or Up/Down arrow key to change or enter values shown in the message area.
EDIT Allows direct entry of a specific frequency span. Use the keypad or
Up/Down arrow key to select the span of choice and complete the entry by pressing the GHz, MHz, kHz,orHz softkey.
FULL Sets the Anritsu HHSA to its maximum frequency span, 2.999 GHz.
ZERO Sets the span to 0 Hz.
SPAN
UP
1-2-5
SPAN
DOWN
1-2-5
BACK Returns to the previous menu level.
Activates the span function so that the span may be increased quickly in a 1-2-5 sequence.
Activates the span function so that the span may be reduced quickly in a 1-2-5 sequence.
3-4
Amplitude Menu
Chapter 3 Key Functions
Pressing AMPLITUDE activates a menu of amplitude related functions. Use the corre sponding softkey to select the desired amplitude function.
REF LEVEL Activates the amplitude reference level function.
SCALE Activates the scale function in a 2 through 15 dB logarithmic amplitude
scale.
ATTEN Sets the Anritsu HHSA input attenuator so that it is either coupled auto
matically to the reference level (AUTO) or manually adjustable (MAN UAL).
UNITS
Bandwidth Menu
Pressing BW/SWEEP activates a menu of bandwidth related functions. Use the corre­sponding softkey to select the desired bandwidth function.
RBW Sets the resolution bandwidth so that it is either coupled automatically to
Choose from the menu of amplitude related units. Selection of dBm sets absolute decibels relative to 1 mW as the amplitude unit. Selection of dBV, dBmV or dBmV sets absolute decibels relative to 1 volt, 1 millivolt, or 1 microvolt respectively as the amplitude unit.
the span (AUTO) or manually adjustable (MANUAL).
-
-
-
VBW Sets the video bandwidth so that it is either coupled automatically to the
span (AUTO) or manually adjustable (MANUAL).
Sweep Menu
Pressing BW/SWEEP activates a menu of sweep related functions. Use the corresponding softkey to select the desired sweep function.
MAX HOLD Displays and holds the maximum responses of the input signal.
DETECTION
BACK Returns to the previous menu level.
Trace Menu
Pressing the TRACE key activates a menu of trace related functions. Use the corresponding softkey to select the desired trace function.
NOTE: For this function, Trace A is always the runtime trace, and Trace B is al ways a saved trace. Refer to page 3-9 for information on saving and recalling traces.
Accesses a menu of detector modes including POSITIVE PEAK detec tion, AVERAGE detection and NEGATIVE PEAK detection.
-
-
3-5
Chapter 3 Key Functions
RESET
A
A ® B Stores the current runtime trace into the Trace B buffer.
A–B
®
A
A+B
®
A
TRACE
B
Measurement Menu
Pressing MEAS on the data keypad activates a menu of measurement related functions. Use the corresponding softkey to select the measurement function.
Clears the current runtime trace.
Performs a subtraction trace math operation.
Performs an addition trace math operation.
Accesses a menu of saved trace options. VIEW B - Provides a trace overlay of the saved trace with the current runtime trace.
CLEAR B - Turns off trace overlay. RECALL TRACE ® B - Recalls the saved trace B buffer. BACK - Returns to the previous menu.
FIELD STRENGTH
OBW
AM/FM
DEMOD
Accesses a menu of field strength measurement options.
ON/OFF - Turns field strength measurements on or off. SELECT ANTENNA - Selects an antenna profile to be used for field
strength measurements. BACK - Returns to the previous menu.
Activates the occupied bandwidth menu. Select either % or dBc method of occupied bandwidth measurement.
Selecting % allows entry of the desired % of occupied bandwidth to be measured.
Selecting dBc allows entry of the desired power level (dBc) to be mea sured.
Accesses a menu of AM/FM demodulation options.
ON/OFF - Turns AM/FM demodulation on or off. DEMOD TYPE - Selects the type of demodulation to perform:
FM - WIDE BAND FM - NARROW BAND
AM VOLUME - Adjust the demodulation speaker volume from 0 to 100. BACK - Returns to the previous menu.
-
3-6
Chapter 3 Key Functions
CHANNEL POWER
ACP Accesses a menu of Adjacent Channel Power ratio measurement options:
Save Setup Menu
Pressing SAVE SETUP on the data keypad saves the current system setup to 1 of 10 inter nal non-volatile memory locations. When the key is pressed, a Save Setup selection box ap pears on the display. Use the Up/Down arrow key to select a setup and press the ENTER key to implement.
Recall Setup Menu
Activates Channel Power measurement. Channel power is measured in dBm. Channel Power density is measured in dBm/Hz.
CENTER FREQ - Activates the center frequency function and sets the Anritsu HHSA to the center frequency. A specific center frequency can be entered using the keypad or Up/Down arrow key. Select the GHz, MHz,
kHz,orHz softkey to accept the center frequency input. MAIN CHANNEL BW - Sets the bandwidth of the main channel. ADJACENT CHANNEL BW - Sets the bandwidth of the adjacent chan
nel.
CHANNEL SPACING - Sets the channel spacing. MEASURE - Begins the ACP measurement. BACK - Returns to the previous menu.
-
-
-
Pressing RECALL SETUP on the data keypad recalls a previously saved setup from mem­ory location 1 through 10. When the key is pressed, a Recall Trace selection box appears on the display. Select a setup using the Up/Down arrow key and press the ENTER key to im­plement. Setup location 0 recalls the factory preset state.
Limit Menu
Pressing LIMIT on the data keypad activates a menu of limit related functions. Use the cor­responding softkey to select the desired limit function. Then use the Up/Down arrow key to change its value, which is displayed in the message area at the bottom of the display.
SINGLE
LIMIT
MULTIPLE
UPPER
LIMITS
Sets a single limit value in dBm. Menu choices are:
ON/OFF
EDIT
BEEP AT LEVEL
BACK
Sets multiple user defined upper limits, and can be used to create an upper limit mask for quick pass/fail measurements. Menu choices are:
SEGMENT 1
SEGMENT 2
SEGMENT 3
SEGMENT 4
SEGMENT 5
BACK
3-7
Chapter 3 Key Functions
MULTIPLE
LOWER
LIMITS
LIMIT
BEEP
Marker Menu
Pressing MARKER on the data keypad activates a menu for the four different markers. Use the corresponding softkey to select the desired marker. Then use the marker second level menu to turn the markers on or off, and to edit marker parameters and values.
M1 Selects the M1 marker parameter, displays both amplitude and frequency,
M2 Selects the M2 marker parameter, displays both amplitude and frequency,
Set multiple user defined lower limits, and can be used to create a lower limit mask for quick pass/fail measurements. Menu choices are:
SEGMENT 1
SEGMENT 2
SEGMENT 3
SEGMENT 4
SEGMENT 5
BACK
Turns the audible limit beep indicator on or off.
and opens the Markers second level menu.
and opens the Markers second level menu.
M3 Selects the M3 marker parameter, displays both amplitude and frequency,
and opens the Markers second level menu.
M4 Selects the M4 marker parameter, displays both amplitude and frequency,
and opens the Markers second level menu.
ALL OFF Turns off all markers.
Marker Menu Second Level Menu
The markers second level menu to turn the markers on or off, and to edit marker parameters and values.
ON/OFF Turns the selected marker on or off.
EDIT Opens the selected marker parameter for data entry. Enter the desired
marker frequency using the keypad or Up/Down arrow key. Select the GHz, MHz, kHz,orHz softkey to accept the marker frequency input.
DELTA Displays delta frequency and amplitude for the selected marker with re
spect to the M1 marker.
MARKER
TO
PEAK
Places the selected marker at the highest point on a trace.
-
3-8
MARKER
FREQ TO
CENTER
Places the selected marker equal to the center frequency
Chapter 3 Key Functions
Save Display Menu
Pressing the SAVE DISPLAY key saves the displayed trace to non-volatile memory. When the key is pressed, “Trace Name:” appears on the display. Select the alphanumeric charac ters for that trace name and press ENTER key to implement.
Recall Display Menu
Pressing the RECALL DISPLAY key recalls a previously saved trace from memory. When the key is pressed, a Recall Trace selection box appears on the display. Select a trace using the Up/Down arrow key and press the ENTER key to implement.
Single Continuous Menu
Toggles the units sweep from continuous sweep mode to single sweep mode, as indicated to the left side of the display. The default is continuous sweep. When single sweep is acti vated, the Anritsu HHSA will sweep once and hold until activated. Single sweep mode can be used to conserve battery power.
Print Menu
Selecting the PRINT key prints the current display to the selected printer.
-
-
System Menu
Pressing SYS on the data keypad activates a menu of system related functions. Use the cor­responding softkey to select the system function.
OPTIONS Displays a second level menu of options functions (see below).
CLOCK Displays a second level menu of clock functions (see below).
SELF TEST Starts an instrument self test featuring battery calibration, memory, and
local oscillator testing.
Options Menu Second Level
Provides access to a second level menu of options related functions. Use the corresponding softkey to select the desired function.
PRINTER Displays a menu of supported printers. Use the Up/Down arrow key and
ENTER key to select the connected printer.
CHANGE
DATE
FORMAT
Clock Menu Second Level
Pressing this soft key changes the format of the date as displayed in the message area. Choices are DD/MM/YYYY, YYYY/MM/DD, MM/DD/YY, DD/MM/YY, or YY/MM/DD.
Provides access to a second level menu of clock related functions. Use the corresponding softkey to select the desired function.
HOUR The hour is displayed in the message area. Enter the desired hour (0-23)
using the keypad and press ENTER to accept.
3-9
Chapter 3 Key Functions
MINUTE The minute is displayed in the message area. Enter the desired minute
(0-59) using the keypad and press ENTER to accept.
MONTH The month is displayed in the message area. Enter the desired month
(1-12) using the keypad and press ENTER to accept.
DAY The day as displayed in the message area. Enter the desired day (1-31) us
ing the keypad and press ENTER to accept.
YEAR The year as displayed in the message area. Enter the desired year
(1997-2038) using the keypad and press ENTER to accept.
BACK Returns user to the top-level systems menu.
Power Monitor Menu
When the mode of operation is changed to Power Monitor, the power monitor menu provides for setting power monitor parameters.
UNITS Toggles between dBm and Watts.
REL Turns relative mode OFF, if currently ON. If relative mode is currently
OFF, turns it ON and causes the power level to be measured and saved as the base level. Subsequent measurements are then displayed relative to this saved value. With units of dBm, relative mode displays dBr; with units of Watts, relative mode displays % (percent).
OFFSET Turns offset OFF, if currently ON. IF Offset is currently OFF, turns it ON
and opens the Offset parameter for data entry using the keypad. Press ENTER when data entry is complete. Offset is the attenuation (in dB) in­serted in the line between the DUT and the RF detector. The attenuation is added to the measured input level prior to display.
-
ZERO Turns Zero OFF, if currently ON. IF Zero is currently OFF, turns it ON
and initiates collection of a series of power level samples, which are aver aged and saved. This saved value is then subtracted from subsequent mea surements prior to display.
3-10
-
-
Chapter 4 Advanced Measurement Fundamentals
Introduction
This chapter demonstrates spectrum analyzer measurement techniques using the Anritsu HHSA. Part I will focus on advanced measurement fundamentals. Part II will focus on mea surement techniques with examples of typical applications. Each application focuses on dif ferent features of the Anritsu HHSA. Measurement applications and procedures covered in this section include:
Resolving Closely Spaced Signals
r
Harmonic Distortion
r
AM/FM Modulation
r
AM/FM Demodulation
r
Segmented Limit Lines
r
Trace Overlay
r
-
-
Advanced Measurement Fundamentals
Advanced measurements require the use of additional spectrum analyzer functions beyond frequency, span, amplitude and marker functions. In particular, this section will focus on resolution bandwidth, video bandwidth, sweep, and attenuator functions.
Effect of Resolution Bandwidth
Resolution Bandwidth is determined by the intermediate frequency (IF) filter bandwidth. The spectrum analyzer traces the shape of its IF filter as it tunes past a signal. If more than one IF is used in a spectrum analyzer, the narrowest one dominates and is considered the resolution bandwidth.
The choice of resolution bandwidth depends on several factors. Filters take time to settle. That is, when a signal first appears at the input of the filter, it will take a while before the signal appears at the output. Additionally, the output of the filter will take some time to set tle to the correct value, so that it can be measured. The narrower the filter bandwidth (reso lution bandwidth) the longer the settling time.
The choice of resolution bandwidth will depend on the signal being measured. If two sig nals are to be measured individually, then a narrow bandwidth is required. If a wider band width is used, then the energy of both signals will be included in the measurement. Thus, the wider bandwidth does not have the ability to look at frequencies selectively but instead must measure across their entire frequency range at all times.
4
-
-
-
-
4-1
Chapter 4 Advanced Measurement Fundamentals
Therefore, a broadband measurement would include the fundamental frequency, harmonics, spurious responses, and noise in the measurement. On the other hand a narrow-band mea surement will filter out all but the desired frequency components, resulting in a measure ment that includes only the fundamental. There are advantages to each. The ultimate decision will be dependent on the type of measurement required by the user.
There is always some amount of noise present in a measurement. Noise is often broadband in nature; that is, is exists at a broad range of frequencies in the frequency domain. If the noise is included in the measurement, the measured value will be in error (too large) de pending on the noise level. With a wide bandwidth more noise is included in the measure ment. With a narrow bandwidth, very little noise enters the resolution bandwidth filter, and the measurement is more accurate. If the resolution bandwidth is narrower, the noise floor will drop on the display of the spectrum analyzer. This is because the IF filter of the ana lyzer has been made narrower in bandwidth, which lets in less noise. As the measured noise level drops, smaller signals that were previously obscured by the noise can now be mea sured.
-
-
-
-
-
-
As a general rule of thumb, most field spectrum analyzer measurements are made at a reso lution bandwidth of 30 kHz.
Effect of Video Bandwidth
Spectrum analyzers typically use another type of filtering after the detector called VIDEO FILTERING. This filter also affects the noise on the display but in a different manner than the resolution bandwidth. In video filtering, the average level of the noise remains the same but the variation in the noise is reduced. Hence, the effect of video filtering is a “smooth­ing” of the signal noise. The resultant effect on the analyzer’s display is that the noise floor compresses into a thinner trace, while the position of the trace remains the same. Thus, changing the video bandwidth (VBW) does not improve sensitivity; however, it does im­prove discernability and repeatability when making low-level measurements.
As a general rule of thumb, most field spectrum analyzer measurements are made at a video bandwidth that is a factor of 10 to 100 less than the resolution bandwidth. Thus, for a reso lution bandwidth of 30 kHz, the typical video bandwidth setting options are either 3 kHz or 300 Hz.
Sweep Limitations
The user normally has control over the SWEEP TIME (the elapsed time of each sweep, sometimes called SCAN TIME), the frequency range over which the analyzer sweeps, and the resolution bandwidth. The analyzer cannot be swept arbitrarily fast while maintaining its specified accuracy, but will have a sweep rate limitation depending on the resolution bandwidth selected. The sweep rate is not usually chosen by the user but is determined by the frequency range swept divided by the sweep time.
-
-
The limitation on sweep rate comes from the settling or response time of the resolution bandwidth filter. If an analyzer is swept very quickly, the filter does not have time to re spond, and the measurement is inaccurate. Under such conditions, the analyzer display tends to have a “smeared” look to it, with the spectral lines being wider than normal and shifted to the right. (Fortunately, the Anritsu Hand Held Spectrum Analyzer has mecha nisms designed into it that unburden the user from having to calculate the sweep rate.)
4-2
-
-
Chapter 4 Advanced Measurement Fundamentals
Attenuator Functions
Attenuation adjusts the hand held spectrum analyzer input attenuator. In AUTO mode, the input attenuator is coupled to the reference level. In manual (MANUAL) mode, the input at tenuation can be adjusted by using the Up/down arrow key. The attenuator range is 0 to 50 dB, in 10 dB steps.
-
Attenuation is normally a coupled function and is automatically adjusted when the refer ence level changes. The reference level will not change however, when the attenuator changes. The attenuator should be adjusted so that the maximum signal amplitude at the in put mixer is –30 dBm or less. For example, if the reference level is + 20 dBm, the attenua tion is 50 dB for an input signal of –30 dBm at the mixer (20 – 50 = –30). This prevents signal compression.
NOTE: It is best to begin all measurements in AUTO attenuation mode.
Example 1: Resolving Closely Spaced Signals
Signal resolution is determined by the intermediate frequency (IF) filter bandwidth. The hand held spectrum analyzer, as do conventional spectrum analyzers, traces the shape of its IF filter as it tunes past a signal. Thus, if two equal-amplitude signals are close enough in frequency, the filter shapes for the two signals can fall on top of one another and appear as a single response. Conversely, if two signals are not equal in amplitude but are still close to­gether, the smaller signal can be hidden under the response of the larger one.
Measurement of Two Signals Having Equal Amplitudes
To resolve two signals of equal amplitude, the resolution bandwidth must be less than or equal to the frequency separation of the two signals. For example, to resolve two signals of equal amplitude with a frequency separation of 30 kHz, a resolution bandwidth of 30 kHz or less should be used. However, most spectrum analyzer IF filter bandwidths are not exact, varying by as much as ± 20%. This should be taken into consideration whenever testing for closely spaced signals. Thus, you may want to select the resolution bandwidth within the lower end of the specification (i.e., –20%, or 24 kHz) to ensure accurate measurements on two signals spaced to within 30 kHz. In this case the next smallest resolution bandwidth would be 10 kHz.
-
-
-
Example
Connect two signal sources to the spectrum analyzer input and set the frequency of one source to 900.0 MHz and the other source to 900.030 MHz. Set both sources to the same amplitude, preferably – 20 dBm. On the MS2711A HHSA:
Step 1. Set the span to 100 kHz. Step 2. Set the resolution bandwidth to 30 kHz and the video bandwidth to 1 kHz.
The two signals should be easily observable on the spectrum analyzer display, spaced 30 kHz apart.
Step 3. Change the resolution bandwidth to a wider value and note that you can no lon
ger see that there are two signals present.
4-3
-
Chapter 4 Advanced Measurement Fundamentals
Example
Keeping the same setup as the previous example, change the two source frequencies to
900.0 MHz and 900.020 MHz. Set both sources to the same amplitude, preferably –20 dBm.
Step 1. Set the span to 100 kHz Step 2. Set the resolution bandwidth to 100 kHz, video bandwidth to 1 kHz. Step 3. Change the resolution bandwidth from 100 kHz to 30 kHz then to 10 kHz to de
tect the presence of two signals.
Remember, the resolution bandwidth must be equal to or less than the frequency separation of the signal. Therefore, a 10 kHz resolution bandwidth must be used. The next larger filter, 30 kHz, would exceed the 20 kHz separation and thus would not resolve the two signals. Also, keep in mind that noise side-bands (phase noise) can also affect resolution.
Measurement of Two Signals Having Unequal Amplitudes
Typically, in real world environments, closely spaced signals do not have equal amplitudes. Often, the difference between closely spaced signals can be as much as –90 dBm. To re solve two signals of unequal amplitude, the resolution bandwidth must be less than or equal to the frequency separation of the two signals (the same as resolving two equal amplitude signals). However, in this case the largest resolution bandwidth that will resolve the two un­equal signals is determined primarily by the shape factor of the IF filter, rather than by the 3 dB bandwidth. Shape factor is defined as the ratio of the 60 dB bandwidth to the 3 dB band­width of the IF filter.
Therefore, to resolve two signals of unequal amplitude, the half-bandwidth of a filter at the point equal to the amplitude separation of the two signals must be less than the frequency separation of the two signals. However, if you do not know the specific shape factor of the IF Filter, not all is lost. Simply perform this measurement as if the signals had equal ampli­tudes but pay close attention to potential signals having unequal power levels that are closely spaced. This will take some adjusting among the various resolution and video band width and span functions.
-
-
-
4-4
Chapter 4 Advanced Measurement Fundamentals
Example
Connect two signal sources to the spectrum analyzer input. Set the frequency of one source to 1900.0 MHz and the other source to 1900.100 MHz. Set one source to the + 20 dBm, and the other to – 15 dBm.
Step 1. Set the span to 2 MHz Step 2. Set the resolution bandwidth to 30 kHz and the video bandwidth to 300 Hz.
The two signals should be easily observable on the spectrum analyzer, each with different amplitudes and spaced 100 kHz apart
Step 3. Change the frequency in the second source from 1900.100 MHz to 1900.050
MHz gradually by turning the knob on the signal source and observe the effect.
Step 4. Change the resolution bandwidth to 10 kHz.
The two signals should still be observable with the 10 kHz resolution band width, but may be difficult to detect with the 30 kHz resolution bandwidth. Nar rowing the span may help in detecting the differences in these two signals.
NOTE: Spectrum analyzer sweep time is inversely proportional to the square of the resolution bandwidth. So, if the resolution bandwidth is reduced by a factor of ten, the sweep time is increased by a factor of 100. For fastest measurement times, use the widest resolution bandwidth that still permits resolution of all de­sired signals.
Example 2: Harmonic Distortion
Most transmitting devices and signal sources contain harmonics. Measuring the harmonic content of such sources is frequently required. In fact, measuring harmonic distortion is one of the most common uses of spectrum analyzers.
The following harmonic distortion measurement applies an important group of spectrum analyzer operating skills: setting the frequency span, using start and stop frequencies; set ting the video bandwidth; and making relative measurements using two markers. It also demonstrates setting a signal to center frequency using a marker, and setting the frequency step size to the value for the center frequency.
-
-
-
4-5
Chapter 4 Advanced Measurement Fundamentals
Example
Step 1. Connect a signal source to the Anritsu HHSA spectrum analyzer. Step 2. Set the input frequency to 10 MHz and the signal level to + 20 dBm. Step 3. Set the START frequency at 1 Mhz Step 4. Set the STOP frequency to 100 MHz Step 5. Set the resolution bandwidth to improve visibility by smoothing the noise:
a. Press BW/SWEEP b. Press RBW, then MANUAL and EDIT.
Step 6. Use the step down (down arrow key) to select the desired resolution bandwidth,
100 kHz in this case.
Step 7. Press ENTER once the desired resolution bandwidth has been selected. Step 8. Set the video bandwidth to improve visibility by smoothing the noise:
a. Press BW/SWEEP b. Press VBW, then MANUAL and EDIT.
Step 9. Use the step down (down arrow key) to select the desired video bandwidth, 3
kHz in this case.
Step 10. Press ENTER once the desired video bandwidth has been selected. Step 11. Press AMPLITUDE and then select REF LEVEL Step 12. Key in the desired REF LEVEL (+20) on the data keypad and press ENTER. Step 13. Press MARKER and select M1 Step 14. Activate M1 ON and press PEAK Step 15. Press MARKER and select M2 Step 16. Activate M2 ON and use Up/Down key to place M2 to the center of the first har
monic
Step 17. Select DELTA (M2-M1) marker function. This will provide frequency and am
plitude information with respect to marker M1.
Step 18. Press MARKER and select M3 Step 19. Activate M3 ON and use Up/Down key to place M3 to the center of the second
harmonic
Step 20. Select DELTA (M3-M1) marker function. This will provide frequency and am
plitude information with respect to marker M1.
-
-
-
4-6
Example 3: AM/FM Modulation
Chapter 4 Advanced Measurement Fundamentals
Modulation is the process of translating some low frequency or baseband signal (voice, mu sic, or data) to a higher frequency. In the modulation process, some characteristic of a car rier signal (usually amplitude or frequency) is changed in direct proportion to the instantaneous amplitude of the baseband signal.
The following procedure describes how to measure signals with AM and FM types of mod ulation on them. It shows how to tune the signal on the spectrum analyzer display.
Amplitude Modulation
Step 1. Connect the signal source to the spectrum analyzer input. Step 2. Set the source to a carrier frequency of 100 MHz with an amplitude modulation
of about 35 kHz
Step 3. Set the spectrum analyzer center frequency to 100 MHz. Set the span to 500
kHz.
Step 4. To determine the frequency of the carrier, press MARKER, activate M1 and
press PEAK, ENTER.
Additional modulation information can be easily determined from the carrier signal and sideband. For example, the difference between the carrier frequency and the sideband fre­quency can be found by pressing MARKER, activating M2 and pressing DELTA, ENTER.
The markers read the frequency difference between the two signals, which is equal to the modulating frequency. The marker also reads the difference in amplitude. This difference in amplitude between the two signals can be used to determine the percentage of modula­tion.
-
-
-
NOTE: Unequal amplitudes of the lower and upper sidebands indicate inciden­tal FM on the input signal. Incidental FM can reduce the accuracy of percent age-of-modulation measurements.
Frequency Modulation
This section contains general information about frequency modulation, as well as a proce dure for calculating FM deviation using a spectrum analyzer.
For sinusoidal modulation where either the modulation frequency or the FM deviation can be varied, the spectrum analyzer can be used to accurately set up a modulation index corre sponding to a Bessel null. The following example illustrates how to verify the FM deviation accuracy of a signal generator with FM capability. We will use a carrier frequency of 100 MHz and test for a FM deviation accuracy at a 25 kHz rate using the modulation index for the first carrier null (2.401).
Step 1. Connect the signal source to the spectrum analyzer input. Set the source to 100
MHz.
Step 2. Change the signal source settings as follows:
a. Set the carrier frequency to 100 MHz b. Set Span to 100 kHz
-
4-7
-
-
Chapter 4 Advanced Measurement Fundamentals
c. Set resolution bandwidth to 10 kHz d. Set video bandwidth to 1 kHz
NOTE: Incidental AM from a source signal can cause the frequency null to shift, resulting in errors to the procedure above. Incidental AM is very low for most RF signal generators but can be significant in microwave signal generators. Non-symmetrical side lobes indicate the presence of incidental AM.
Example 4: AM/FM Demodulation
Amplitude modulation (AM) and frequency modulation (FM) are common modulation techniques used to broadcast information. In the United States and Canada, the AM broad cast band is 535 kHz to 1605 kHz, while the FM broadcast band covers 88 MHz to 108 MHz.
Demodulation of AM and FM signals can be done with any spectrum analyzer with zero span capability and with a wide enough bandwidth to fully encompass the modulated sig nal, provided that the spectrum analyzer has a speaker or headset jack and AM/FM demod ulator. (The Anritsu HHSA comes standard with built-in AM/FM demodulator, internal speaker and headset jack).
-
-
-
Example
Step 1. Provide a signal to the spectrum analyzer by connecting an antenna to the input
port of the spectrum analyzer.
Step 2. Tune to the FM band by pressing the MEASURE key. Step 3. Press DEMOD TYPE Step 4. Select FM WIDEBAND by using the appropriate softkey Step 5. Activate FM demodulation ON by using the appropriate softkey. Step 6. Select the appropriate resolution bandwidth, video bandwidth and reference
level to enhance the selection of desired signal. (For wide-band FM, the resolu tion bandwidth should be 1 MHz. Reference level should be roughly –30 dBm)
Step 7. Position the marker on the signal of interest. If the signal of interest is the high
est in amplitude, press MARKER TO PEAK directly.
Step 8. To obtain continuous demodulation of the signal, set the span to zero span. Step 9. Press FREQ/SPAN Step 10. Select SPAN by using the appropriate softkey Step 11. Select ZERO by using the appropriate softkey.
-
-
Setting the hand held spectrum analyzer to zero span will ensure continuous demodulation. Otherwise, demodulation will occur only during the spectrum analyzer sweep.
4-8
Chapter 4 Advanced Measurement Fundamentals
Example 5: Field Strength Measurements
All antennas have losses or gain that can cause errors in measurements. The MS2711A can correct for antenna loss or gain errors using Field Strength Measurements.
The antenna factors must be uploaded to the MS2711A using the Anritsu Software Tools provided with the unit. These antenna factors can then be used to correct for the measure ment error.
-
Step 1. Enter the antenna factor information for the specific antenna into the antenna ed
itor of the SW Tools.
Step 2. Upload the antenna factors to the MS2711A. Step 3. Step 4. Step 5.
The MS2711A will now automatically scale the screen by the antenna factors entered.
Example 6: Creating a Spectral Mask
Quick go/no-go measurements can be performed by establishing test limits. When using test limits, the user is able to quickly identify signals exceeding established limits as failing. To aid users in establishing limits, the Anritsu HHSA features both single limit and multi­ple limit functions.
Example
Step 1. Connect a signal source to the Anritsu HHSA spectrum analyzer. Step 2. Set the input frequency to 900 MHz and the signal level to - 20 dBm.
Press the MEAS function key on the MS2711A. Select the FIELD STRENGTH soft key from the measurement menu. Press the SELECT ANTENNA soft key and use the Up/Down arrow key to se
lect the desired antenna factor file. Press the ENTER key to select.
-
-
Step 3. Set the center frequency at 900 MHz, resolution bandwidth to 30 kHz, video
bandwidth to 300 Hz, and the span to 20 MHz.
Step 4. Press LIMIT and select MULTIPLE UPPER LIMITS using the appropriate
softkey.
Step 5. Select SEGMENT 1 using the appropriate softkey, and enter the start frequency,
start limit, end frequency, and end limit as prompted in the message area. (Hint: span is ten divisions wide; therefore divide span by 10 to determine span per division and desired starting point). Enter start frequency = 890 MHz, start limit = -40, end frequency = 898 MHz, end limit = -40.
Step 6. Select SEGMENT 2 using the appropriate softkey, and enter the start frequency,
start limit, end frequency, and end limit as prompted in the message area. (Hint: start frequency and start limit should be equal to end frequency and end limit of Segment 1).
4-9
Chapter 4 Advanced Measurement Fundamentals
Enter start frequency = 898 MHz, start limit = -40, end frequency = 899 MHz, end limit = 0.
Step 7. Select SEGMENT 3 using the appropriate softkey, and enter the start frequency,
start limit, end frequency, and end limit as prompted in the message area. (Hint: start frequency and start limit should be equal to end frequency and end limit of Segment 2). Enter start frequency = 899 MHz, start limit = 0, end frequency = 901 MHz, end limit = 0.
Step 8. Select SEGMENT 4 using the appropriate softkey, and enter the start frequency,
start limit, end frequency, and end limit as prompted in the message area. (Hint: start frequency and start limit should be equal to end frequency and end limit of Segment 3). Enter start frequency = 901 MHz, start limit = 0, end frequency = 902 MHz, end limit = -40.
Step 9. Select SEGMENT 5 using the appropriate softkey, and enter the start frequency,
start limit, end frequency, and end limit as prompted in the message area. (Hint: start frequency and start limit should be equal to end frequency and end limit of Segment 4). Enter start frequency = 902 MHz, start limit = -40, end frequency = 910 MHz, end limit = -40.
Example 7: Trace Overlay
The MS2711A HHSA can be used to compare frequency spectrums.
Example:
Step 1. Connect a signal source to the Anritsu HHSA spectrum analyzer. Step 2. Set the input frequency to 900 MHz and the signal level to –20 dBm. Step 3. Set the center frequency at 900 MHz, resolution bandwidth to 30 kHz, video
bandwidth to 300 Hz, and the span to 20 MHz.
Step 4. Step 5.
Step 6. Set the input frequency to 901 MHz and the signal level to –20 dB. Step 7. Step 8.
Press the TRACE key. Select the A ® B soft key to save the current runtime trace to the Trace B
buffer.
Press the TRACE B soft key to go to the Trace B menu. Select the VIEW B soft key to view the traces simultaneously.
4-10
Chapter 4 Advanced Measurement Fundamentals
Example 8: Power Monitor Measurements
A RF Wattmeter/Power meter option is also available on the Anritsu HHSA. Power meter measurement is accomplished using a broadband (1 MHz to 3000 MHz) RF detector. Mea sured power can be displayed in dBm or Watts.
Example
To use the power monitor, use must first set the Anritsu HHSA to the power monitor mode and zero the power monitor.
Step 1. Press the MODE key Step 2. With no applied power to the DUT, press ZERO soft key from the power menu.
Wait a few seconds while the Anritsu HHSA accumulates samples of the quies cent power level. When complete, ZERO ADJ: ON is displayed in the message area.
Step 3. Insert an attenuator between the DUT and the RF detector, sufficient to insure
that the power to the Anritsu HHSA is no greater than 20 dBm.
Step 4. Press the OFFSET soft key. Step 5. Enter the attenuation in dB using the keypad. Press ENTER to complete the en-
try. The message area will show OFFSET in ON along with the entered value in dB.
-
-
Step 6. Press the UNITS soft key to display power in Watts. Step 7. With the desired base power level input to the Anritsu HHSA, press the REL
soft key. The message area will show REL: ON and the power reading will indi­cate 100%.
Step 8. Press the UNITS soft key to display power in dBm. Since REL is ON, the power
reading will be in dBr (relative to base power).
4-11/4-12
Chapter 5 Field Measurements
Introduction
This chapter provides examples of various field measurements:
Occupied Bandwidth
r
Power Monitor
r
Adjacent Channel Power
r
Out-of-Band Spurious Emissions
r
In-Band Spurious Emissions
r
Field Strength
r
AM/FM Demodulation
r
Occupied Bandwidth
A common measurement performed on radio transmitters is that of occupied bandwidth (OBW). This measurement calculates the bandwidth containing the total integrated power occupied in a given signal bandwidth. There are two different methods of calculation de pending on the technique to modulate the carrier.
XdB Down Method
The occupied frequency bandwidth is defined as the bandwidth between the upper and lower frequency points at which the signal level is XdB below the peak carrier level.
N% Method
The occupied frequency bandwidth is calculated as the bandwidth containing N% of the power transmitted where N can be between 1% and 99%.
Occupied Bandwidth Measurement with the MS2711A
-
To make an occupied bandwidth measurement with the MS2711A hand held spectrum ana lyzer, perform the steps below.
Required Equipment
r
MS2711A Hand Held Spectrum Analyzer
r
30 dB, 50 watt, Bi-Directional, DC – 18 GHz, N(m) – N(f), Attenuator
r
Test Port extension cable, Anritsu 15NNF50 – 1.5A
Procedure
Step 1. Using the test port extension cable and 30 dB, 50 watt, Bi-directional attenuator,
connect the MS2711A to appropriate transmit test port.
Step 2. Step 3.
Press the ON/OFF key on the MS2711A. Press the MODE key.
-
5
5-1
Chapter 5 Field Measurements
Step 4. Use the Up/Down arrow key to scroll to Spectrum Analyzer mode and press
ENTER.
Step 5. Enter the center frequency of interest. Step 6. Select the appropriate span, amplitude level, resolution and video bandwidths. Step 7. Step 8.
Step 9. Select measurement method (dB Down or % of Power) in the active function
Step 10. Adjust the dBc or % by pressing the appropriate key. Step 11.
Press the MEAS key. Once in the Measurement menu, press the OBW soft key in the active function
block.
block by pressing the METHOD soft key.
Press MEASURE to initiate the measurement.
Adjacent Channel Power Leakage
Another common transmitter measurement is that of adjacent channel leakage power. This is defined as the ratio of the amount of leakage power in an adjacent channel to the total transmitted power. Therefore, this test measures the amount of integrated power within ad jacent channel bandwidths. The resulting signal level is determined by this amount of isola­tion between channels. In order to calculate the upper and lower adjacent channel values, the spectrum analyzer needs five parameters to be specified:
Channels separation
r
r
Measurement channel bandwidth
r
Adjacent channel bandwidth (if different from the measurement channel band­width), and the center frequency of the reference channel
-
r
Power level of the primary channel as well as the lower and upper adjacent channels.
r
The adjacent channel power leakage measurement is applicable to both modu lated and unmodulated signals and provides a means of assessing the transmit ters selectivity.
Adjacent Channel Power Measurement with the MS2711A
To make a channel power measurement with the MS2711A hand held spectrum analyzer, simply execute the following steps.
Required Equipment
r
MS2711A Hand Held Spectrum Analyzer
r
30 dB, 50 watt, Bi-Directional, DC – 18 GHz, N(m) – N(f), Attenuator
r
Test Port extension cable, Anritsu 15NNF50 – 1.5A
Procedure
Step 1. Using the test port extension cable and 30 dB, 50 watt, Bi-directional attenuator,
connect the MS2711A to appropriate transmit test port.
Step 2.
Press the ON/OFF key on the MS2711A.
-
-
5-2
Chapter 5 Field Measurements
Step 3. Step 4. Use the Up/Down arrow key to scroll to Spectrum Analyzer mode and press
Step 5. Enter the center frequency for the channel of interest. Step 6. Set the span wide enough to include the primary channel bandwidth and upper
Step 7. Select the appropriate amplitude level, resolution and video bandwidths. Step 8. Use markers to read and compare to the specified power level for adjacent chan
Press the MODE key.
ENTER.
and lower channel bandwidths.
nel power compliance.
Out-of-Band Spurious Emission Measurements
Out-of-band spurious measurements are made on signals outside the system band. These signals, which can interfere with other communication systems, can be categorized into har monics and unknown spurious emissions. Real time monitoring of spurious emissions from a transmitter can uncover unwanted signals before they interfere with other users of the ra dio spectrum, rendering the transmitting system non-compliant. In order to determine com­pliance with the allowable level of spurious emissions, the spectrum analyzer needs three parameters to be specified:
-
-
-
Measurement channel bandwidth (in band, Figure 5-1)
r
r
Measurement channel bandwidth (out of band, Figure 5-1)
r
Allowable level of spurious emissions
Out-of-band Spurious Measurement with the MS2711A
To make an out-of-band spurious measurement with the MS2711A hand held spectrum ana lyzer, simply execute the following steps.
Required Equipment
r
MS2711A Hand Held Spectrum Analyzer
r
30 dB, 50 watt, Bi-Directional, DC – 18 GHz, N(m) – N(f), Attenuator
r
Test Port extension cable, Anritsu 15NNF50 – 1.5A
Procedure
Step 1. Using the test port extension cable and 30 dB, 50 watt, Bi-directional attenuator,
connect the MS2711A to appropriate transmit test port.
Step 2. Step 3. Step 4. Use the Up/Down arrow key to scroll to Spectrum Analyzer mode and press
Press the ON/OFF key on the MS2711A. Press the MODE key.
ENTER.
-
Step 5. Enter the center frequency for the channel of interest.
5-3
Chapter 5 Field Measurements
Step 6. Set the span wide enough to include the primary channel bandwidth and upper
and lower channel bandwidths.
Step 7. Select the appropriate amplitude level, resolution and video bandwidths. Step 8. Use limit line and markers to read and compare observed signals to the specified
allowable level of out-of-band spurious emissions for the corresponding chan nel transmit frequency.
In-Band/Out-Of-Channel Measurements
The in-band/out-of-channel measurements are those measurements that measure distortion and interference within the system band, but outside of the transmitting channel. These measurements include (1) in-band spurious emissions and (2) adjacent channel power ratio (also called spectral regrowth). There are stringent regulatory controls on the amount of in terference that a transmitter can cause to neighboring channels. The standards specify the amount of interference allowed by the system. In order to determine compliance with the allowable level of spurious emissions, the spectrum analyzer needs two parameters to be specified:
-
-
Out-of-band In band Out of band
Figure 5-1. Example of In- and Out-of Bandwidth Signals
r
Measurement channel bandwidth
r
Allowable level of spurious emissions
In-band Spurious Measurement with the MS2711A
To make an out-of-band spurious measurement with the MS2711A hand held spectrum ana lyzer, simply execute the following steps.
Required Equipment
r
MS2711A Hand Held Spectrum Analyzer
r
30 dB, 50 watt, Bi-Directional, DC – 18 GHz, N(m) – N(f), Attenuator
r
Test Port extension cable, Anritsu 15NNF50 – 1.5A
Procedure
Step 1. Using the test port extension cable and 30 dB, 50 watt, Bi-directional attenuator,
connect the MS2711A to appropriate receive test port.
-
5-4
Step 2.
Press the ON/OFF key on the MS2711A.
Chapter 5 Field Measurements
Step 3. Step 4. Use the Up/Down arrow key to scroll to Spectrum Analyzer mode and press
Step 5. Enter the center frequency for the channel of interest. Step 6. Set the span wide enough to include the primary channel bandwidth and upper
Step 7. Select the appropriate amplitude level, resolution and video bandwidths.
Press the MODE key.
ENTER.
and lower channel bandwidths.
Use limit line and markers to read and compare observed signals to the speci fied allowable level of in-band spurious emissions for the corresponding chan nel transmit/receive frequency.
Signal resolution is determined by the intermediate frequency (IF) filter band width. The hand held spectrum analyzer, as does conventional spectrum analyz ers, traces the shape of its IF filter as it tunes past a signal. Thus, if two equal-amplitude signals are close enough in frequency, the filter shapes for the two signals can fall on top of one another and appear as a single response. Con versely, if two signals are not equal in amplitude but are still close together, the smaller signal can be hidden under the response of the larger one.
Field Strength
-
-
-
-
-
The procedure below details how to make field strength measurements.
Required Equipment
r
MS2711A Hand Held Spectrum Analyzer
r
30 dB, 50 watt, Bi-Directional, DC – 18 GHz, N(m) – N(f), Attenuator
r
Test Port extension cable, Anritsu 15NNF50 – 1.5A
Procedure
Step 1. Use the antenna editor feature of the Software Tools to define an antenna (see
page 6-5).
Step 2. Upload the antenna information to the MS2711A HHSA (see page 6-5). Step 3. Press the MEAS key. Step 4. Select the FIELD STRENGTH soft key. Step 5. Press the SELECT ANTENNA soft key. Step 6. Use the Up/Down arrow key to choose the desired antenna and select by press
ing the ENTER key.
Step 7. To change the unit of measurement, press the AMPLITUDE hard key, then
press the UNITS soft key and select dB, dBV, dBmV or dBmV using the desig nated soft key.
-
-
The MS2711A HHSA will automatically scale the display by the antenna factors selected.
5-5
Chapter 5 Field Measurements
ANTENNA CALCULATIONS
The following is a list of various antenna calculations should you find it necessary to con vert from one to another:
Conversion of signal levels from mW to mV in a 50-ohm system:
2
V
P
=
R
where: P = power in Watts
V = voltage level in Volts R = resistance in Ohms
-3
For power in milliwatts (10
V
dB V
()( )
Power density to field strength. An alternate measure of field strength to electric field is power density:
Pd
=
where: E = field strength in V/m
P
=+107
dBmm
2
E
120p
P = Power density in W/m
W), and voltage in microvolts 10-6V:
2
-
Power density at a point:
PtG
P
d
In the far field, where electric and magnetic fields are related by the impedance of free space:
P
d
Pt= power transmitted in Watts G r = distance from the antenna in meters
t
2
r=4
p
= power density in W/m
= gain of transmitting antenna
t
2
5-6
Chapter 5 Field Measurements
Making Power Measurements with the Power Meter
Power measurement is accomplished using a broadband (1 MHz to 3000 MHz) RF detector, P/N 5400-71N50. The power monitor displays the measured power in dBm or Watts.
Entering Power Monitor Mode
Step 1.
Press the MODE key.
Zeroing the Power Monitor
Step 2.
With no power applied to the DUT, press the ZERO soft key from the Power menu. Wait a few seconds while the Anritsu HHSA accumulates samples of the quiescent power level. When complete, ZERO ADJ: ON is displayed in the mes sage area.
Measuring High Input Power Levels
Step 3. Insert an attenuator between the DUT and the RF detector, sufficient to insure
that the input power to the Anritsu is no greater the 20 dBm.
Step 4. Step 5.
Press the OFFSET soft key. Enter the attenuation in dB using the keypad. Press ENTER to complete the en-
try. The message area will show OFFSET is ON along with the entered value in dB.
Displaying Power in dBm and Watts
Step 6.
Press the UNITS soft key to display power in Watts.
Displaying Relative Power
-
Step 7.
Step 8.
With the desired base power level input to the Anritsu HHSA, press the REL soft key. The message area will show REL: ON and the power reading will indi cate 100%.
Press the UNITS soft key to display power in dBm. Since REL is ON, the power reading will be in dBr, relative to the base power level.
5-7
-
Chapter 5 Field Measurements
AM/FM Demodulation
Amplitude modulation (AM) and frequency modulation (FM) are common modulation techniques used to broadcast information. In the United States and Canada, the AM broad cast band is 535 kHz to 1605 kHz, while the FM broadcats band covers 88 MHz to 108 MHz.
Demodulation of AM and FM signals can be accomplished with the MS2711A hand held spectrum analyzer using zero span and a wide enough bandwidth to fully encompass the modulated signal. An alternative, and possibly easier, method of examining a demodulated signal with the MS2711A is to use the built-in AM/FM demodulator and speaker or op tional external headset.
The following example illustrates the use of the MS2711A built-in AM/FM demodulator and speaker. The example includes tunning the hand held spectrum analyzer to the band of interest, activating and moving a marker to the signal that is to be demodulated, selecting the appropriate zero span, and demodulating and listening to the signal of interest with the built-in demodulator and speaker. To find the signal of interest for demodulation:
Step 1. Connect an antenna to the input port of the MS2711A.
-
-
Step 2. Step 3. Step 4.
Step 5. Step 6. Step 7.
Step 8.
Step 9. Step 10.
Press the FREQ/SPAN key Press the START soft key and enter 88 MHz Press the STOP soft key and enter 108 MHz
To demodulat an FM signal, you must first locate the signal you are trying to de­modulate. Using a marker will facilitate this process:
Press the MARKER key. Select the M1 soft key and then the ON/OFF soft key to activate marker 1. Select the EDIT soft key and set marker M1 to the signal of interest (if the signal
of interest is the highest in amplitude, press the MARKER TO PEAK key).
The marker frequency should correspond to the desired signal of interest, and now becomes the center frequency which must be entered. To improve chances of receiving the desired signal, the attenuation setting should be set to zero and the resolution bandwidth should bet wide enough to encompass the entire de modulated signal, i.e., 1 MHz.
Press the MEAS key and select AM/FM DEMOD to activate the demodulation function.
Press the ON/OFF soft key to turn AM/FM demodulation on. Press the DEMOD TYPE soft key and select WIDE BAND FM.
-
5-8
Step 11.
Step 12. Step 13.
Press the VOLUME soft key, use the Up/down arrow key to adjust the volume setting and press ENTER
Press the FREQ/SPAN key and the SPAN soft key. Select the ZERO soft key to select zero span.
Chapter 6 Software Tools Program
Description
The Anritsu Software Tools program provides the means for transferring the measured trace, along with any applied markers and/or limit line, to the screen of a personal computer (PC) running Windows 95/98/NT/2000.
Requirements
The Anritsu Software Tools program will run on any computer that runs Windows 95/98/NT/2000. Typically, this means having a PC with the following characteristics:
Pentium or better microprocessor running enhanced mode (100 MHz or better,
·
recommended). 32 MBytes of memory, minimum.
·
Hard Disk Drive, with approximately 15 MBytes of available space.
·
Communication Port Setting
The Anritsu Software Tools communicates with the Anritsu through a standard COM port on the PC. It is important that your Windows COM port settings conform to the actual hardware settings. Since various add-in devices such as sound cards, modems, and network cards use IRQ (interrupts), it is possible that your computer has non-standard COM port settings. Please consult your computer vendor for COM port address and IRQ information.
Changing COM Port Settings–Windows 95/98/NT/2000
Refer to Figure 5-1 while performing the following procedure.
Step 1. Step 2. Step 3. Step 4. Step 5.
NOTE: If Windows doesn’t show any available COM Ports, consult your computer manufacturer.
Open the Windows Control Panel. Double click on the SYSTEM icon. The System Properties window appears. Select Device Manager. The Device List appears. Double click on the item Ports (COM & LPT) in the device list. Double click on the Communications Port you want to set. The
Communications Port Properties window appears.
6
Step 6.
Choose Port Settings, then change to the following settings if necessary.
Baud Rate: 9600
6-1
Chapter 6 Software Tools Program
Figure 6-1. Windows 95 COM Port Setting Dialog Boxes
Data Bits: 8
Parity: None Stop Bits: 1 Flow Control: None
Step 7.
Step 8.
NOTE: If you changed the COM Port Base Address and/or IRQ, you will need to restart Windows.
Choose Resource and verify the COM Port Base Address and IRQ. Click OK when done.
Click OK again to close the System Properties window.
Software Installation
The Anritsu Software Tools program is a conventional Microsoft Windows program. Installation is similar to all other such programs. For users new to Windows, a detailed procedure is given below.
Step 1. Insert the Anritsu Anritsu Software Tools CD For Windows disk into the CD
drive.
Step 2. Select Run under the Start menu. Step 3. Type D: \SETUP.EXE, when the dialog box appears (Assumes CD ROM is
drive D:).
Step 4.
Press the Enter key to select the default directory, C:\PROGRAM FILES\SITEMASTER SOFTWARE TOOLS, and begin the Setup routine.
6-2
Step 5. When the Setup program prompts for the Program Group, select “Site Master
Software Tools” and select CONTINUE.
Chapter 6 Software Tools Program
NOTE: The Setup routine will create a new Program Manager Group named
“Site Master Software Tools”.
Step 6. Step 7. Double-click on the “Anritsu Software Tools ” icon to open the Software Tools
Step 8.
When the installation is complete, click OK.
program. Click on Settings, in the top menu bar, and select Communications. Select the
appropriate COM port number for the serial interface cable (null modem type).
Plot Capture
Plots (traces) can be captured either individually from the Anritsu HHSA display or in multiples from one or more stored-display locations. Both methods are described below.
The recommended method is the Capture Multiple Traces option in the Capture pull-down menu. Using this method, you can download to the Software Tools program in one operation all of the data residing in the up-to-70 Stored Display memory locations. The downloaded traces can be stored in a database or appear in cascade on the PC screen.
NOTE: Trace scale can be captured as per Anritsu HHSA or as autoscale by the capture program. To select per Anritsu HHSA or autoscaling, click on Settings and Default Plot Settings from the top menu bar and pull-down menu.
Capture multiple to database
Step 1. Connect the supplied cable. Step 2. Open the “Anritsu Software Tools” group, in the Windows Program Manager.
(Select the Windows pull-down menu from the Program Manager menu bar, and select Anritsu Software Tools.)
Step 3. Step 4.
Step 5. Follow the database instructions to download the plot(s) to either a new database
Step 6. Enter the number(s) of the stored-display memory location(s) (1 to 50) from
Step 7. Observe that the “Acquiring Control” box appears on the screen, then disappears
Double-click on the “Anritsu Software Tools” icon to open the program. Click on Capture, in the top menu bar, and select Capture Multiple to
Database from the drop-down menu.
or an existing database.
which you wish to store to the database, and click “OK”.
as traces are automatically acquired. The “Database” box appears when the plot(s) has been completely captured into the database.
6-3
Chapter 6 Software Tools Program
Capture multiple traces to PC screen
Step1. Perform steps 1, 2, and 3 of the capture-multiple-to-database procedure. Step2.
Step3. Enter the number (or numbers) of the stored-display memory location(s) (1 to
Step4. Select “OK.” Step5. Observe that the “Acquiring Control” box appears on the screen, then disappears
Click on “Capture Multiple Traces” icon or click on Capture, in the top menu bar, and select Capture Multiple Traces from the drop-down menu.
50) from which you wish to display traces in Anritsu Software Tools or select from the drop-down list.
as the traces are automatically acquired.
Single trace capture
Step1. Perform steps 1, 2, and 3 of the capture-multiple-to-database procedure. Step2.
Step3. Observe that a “Waiting for Data” box appears on the screen. Step4. On the Anritsu HHSA, Step5. Capture a new trace and store it to a memory location. Step6. Recall the stored plot to the screen. Step7.
Click on “Start a Plot Capture” icon or click on Capture, in the top menu bar, and select Start Capture from the drop-down menu.
Press the ENTER key (on the Anritsu HHSA) to start the screen capture.
NOTE: The ENTER key on the Anritsu HHSA must be pressed while the PC screen shows “Waiting for Data” for the transfer of information to be complete.
Step 8. On the PC, observe that the “Waiting for Data” box disappears and the trace
capture process begins. Within a short time, the Anritsu HHSA trace will appear.
Program Operation
The captured trace on the PC can be scaled and have its limit line, markers, and properties changed. (Select Plot Properties under the View menu to make these changes.) The operation of the various menus that allow these operations to be accomplished is straightforward. To read about the operation of the menus, refer to the on-line help screens. They can be accessed from the Help menu, in the top menu bar.
6-4
Chapter 6 Software Tools Program
Entering Antenna Factors
The antenna factor is a parameter of an antenna that is used in the calculation of field strength during radiated emissions measurement. It relates the voltage output of the measurement antenna to the value of the incident field producing that voltage. The units are volts output per volt/meter incident field or reciprocal meters. Antennas used for radiated emissions testing are individually calibrated (the antenna factor is directly measured) at all appropriate distances. The calibrations produce values that are defined as the “equivalent free space antenna factor.” The calibration procedure corrects for the presence of the reflection of the antenna in the ground plane, giving the value that would be measured if the antenna were in “free space.”
Step 1. Once the Site Master Software Tools has been opened, select Antenna Editor
from the Tools menu on the tool bar. A pop-up box will appear on the screen of the PC.
Step 2. Click on “Edit Antenna” to enter an antenna name, description, frequencies, and
antenna factors. Enter the frequencies in ascending order, starting with lowest frequency first. A maximum of 60 antenna factors may be entered. Use the arrow keys on the keyboard to move between entries.
NOTES: Only one input of frequency and antenna factor is allowed per row. Multiple antenna factors for a single antenna must be entered individually. For example, an antenna having an antenna factor of 5 from 2.0 to 2.25 GHz and an antenna factor of 4 from 2.25 to 2.5 GHz should be entered as follows:
Frequency (MHz) Antenna Factor
2000 5 2251 4 2500 4
If necessary, an antenna factor of zero (0) may be entered.
Step 3. Select Save from the File menu to save the antenna factors to the hard disk.
Uploading Antenna Factors
To upload antenna information from Software Tools to the MS2711A:
Step 1. Connect the RS232 cable between the PC and the MS2711A. Step 2. Click on the “Upload” button on the tool bar (or select Upload from the Tools
menu). It is important to note that the MS2711A should sweep very quickly during the data transfer, at least at 5 seconds, otherwise the Software Tools program will default to a “time-out” condition (to improve chances of successful upload, increase RBW and VBW settings to the maximum position).
Step 3. To determine if the antenna information has been successfully uploaded to the
MS2711A, press the measure function key (#4 key) and activate the field strength measurement by pressing the ON/OFF soft key.
6-5
Chapter 6 Software Tools Program
Saving a Plot as a Windows Metafile
Plots can be saved as Windows Metafiles (.WMF). The metafile may be imported into other graphic programs, but cannot be reloaded into the Anritsu Software Tools program.
To save a plot as a Windows Metafile, click on File, in the top menu bar, and select Save as Metafile from the drop down menu.
Saving Data to a Spreadsheet
The data points from a plot can be exported to a spreadsheet via the clipboard. To transfer data to a Windows spreadsheet program:
Step 1.
Step 2. Capture or load the desired plot. Step 3.
Step 4. Open the spreadsheet program and place the cursor where the first data point
Step 5.
Select Settings and Clipboard Format from the top menu bar and pull-down menu and choose Formatted Text.
Copy the data to the clipboard by selecting the “Copy to Clipboard” icon or Copy from the Edit menu.
should appear. Select Paste from the spreadsheet program’s Edit menu.
Saving Data as a Data File
The data points from a plot can be saved as a data file (.dat).
Step 1. Step 2. Step 3. Enter a file name and location. Step 4.
Select the File menu. Select Save.
Click OK.
Saving Data to a Database
Plots can be saved to a database. Comments can be added to the plot data saved. Queries of the database provide a means of comparing plots in the database. Refer to the on-line help screens for operating instructions.
To save a plot to a database, click on File, in the top menu bar, and select Save Plot to Database from the drop-down menu.
6-6
Chapter 6 Software Tools Program
“Drag-n-Drop”
Anritsu Software Tools is Windows based. Graphs can “Drag-n-Drop” onto each other. Anritsu Software Tools allows quick comparison of “before” and “after” measurements.
Recent data is compared to a historical PC database record, which is usually recorded during site installation/commissioning.
Printing
Captured traces may be printed from a PC using Anritsu Software Tools. Once a captured trace has been downloaded choose Print under the File menu for printing options. The printer setup can be altered, plots can be scaled, and multiple plots can be printed from the Print dialog box.
6-7/6-8
Loading...