Anritsu MS2650, MS2660B, MS2660C Operation Manual

MS2650/MS2660B/C Series
Spectrum Analyzer
Operation Manual
(Detailed Operating Instructions
10th Edition
For safety and warning information, please read this
manual before attempting to use the equipment.
Additional safety and warning information is
provided within the MS2650/MS2660B/C Series Spectrum Analyzer Operation Manual Vol. 1 (Basic Operating Instructions). Please also refer to this document before using the equipment.
Keep this manual with the equipment.
)
ANRITSU CORPORATION
Document No.: M-W1251AE-10.0

Safety Symbols

To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the
following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings
of the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu
equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this
manual.
Symbols used in manual
DANGER
WARNING
CAUTION
This indicates a very dangerous procedure that could result in serious injury or death if not performed properly.
This indicates a hazardous procedure that could result in serious injury or death if not performed properly.
This indicates a hazardous procedure or danger that could result in light-to-severe injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manual
The following safety symbols are used inside or on the equipment near operation locations to provide information
about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols
and take the necessary precautions BEFORE using the equipment.
This indicates an obligatory safety precaution. The obligatory operation is
This indicates a warning or caution. The contents are indicated symbolically in or
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
This indicates a prohibited operation. The prohibited operation is indicated symbolically in or near the barred circle.
indicated symbolically in or near the circle.
near the triangle.
MS2650/MS2660B/C Series Spectrum Analyzer Operation Manual Vol. 2 (Detailed Operating Instructions)
14 March 1997 (First Edition) 7 March 2008 (10th Edition)
Copyright © 1997-2008, ANRITSU CORPORATION. All rights reserved. No part of this manual may be reproduced without the prior written permission of the publisher. The contents of this manual may be changed without prior notice. Printed in Japan
ii
Equipment Certificate
Anritsu Corporation certifies that this equipment was tested before shipment using calibrated measuring instruments with direct traceability to public testing organizations recognized by national research laboratories, including the National Institute of Advanced Industrial Science and Technology, and the National Institute of Information and Communications Technology, and was found to meet the published specifications.
Anritsu Warranty
Anritsu Corporation will repair this equipment free-of-charge if a malfunction occurs within one year after shipment due to a manufacturing fault, under the condition that this warranty is void when:
The fault is outside the scope of the warranty conditions described in
the operation manual.
The fault is due to mishandling, misuse, or unauthorized modification
or repair of the equipment by the customer.
The fault is due to severe usage clearly exceeding normal usage.
The fault is due to improper or insufficient maintenance by the
customer.
The fault is due to natural disaster including fire, flooding,
earthquake, etc.
The fault is due to use of non-specified peripheral equipment,
peripheral parts, consumables, etc.
The fault is due to use of a non-specified power supply or in a
non-specified installation location.
In addition, this warranty is valid only for the original equipment purchaser. It is not transferable if the equipment is resold.
Anritsu Corporation shall assume no liability for injury or financial loss of the customer due to the use of or a failure to be able to use this equipment.
Anritsu Corporation Contact
In the event that this equipment malfunctions, contact an Anritsu Service and Sales office. Contact information can be found on the last page of the printed version of this manual, and is available in a separate file on the CD version.
iii
r
r
Notes On Export Management
This product and its manuals may require an Export License/Approval by
the Government of the product's country of origin for re-export from you
country.
Before re-exporting the product or manuals, please contact us to confirm
whether they are export-controlled items or not.
When you dispose of export-controlled items, the products/manuals need
to be broken/shredded so as not to be unlawfully used for military purpose.
Disposal Procedure
The product that you have purchased contains a rechargeable battery.
The battery is recyclable. At the end of its useful life, under various
state and local laws, it may be illegal to dispose of this battery into the
municipal waste stream. Check with your local solid waste officials fo
details in your area for recycling options or proper disposal.
iv
Front Panel Power Switch
To prevent malfunction caused by accidental touching, the front power switch of this equipment turns on the power if it is pressed continuously for about one second in the standby state. If the switch is pressed continuously for one second in the power-on state, the equipment enters the standby state.
In the power-on state, if the power plug is removed from the outlet, then reinserted into it, the power will not be turned on. Also, if the lines is disconnected due to momentary power supply interruption or power failure, the power will not be turned on (enters the standby state) even if the line is recovered.
This is because this equipment enters the standby state and prevents incorrect data from being acquired when the line has to be disconnected and reconnected.
For example, if the sweep time is 1,000 seconds and data acquisition requires a long time, momentary power supply interruption (power failure) might occur during measurement and the line could be recovered automatically to power-on. In such a case, the equipment may mistake incorrect data for correct data without recognizing the momentary power supply interruption.
If this equipment enters the standby state due to momentary power supply interruption or power failure, check the state of the measuring system and press the front power switch to restore power to this equipment.
Further, if this equipment is built into a system and the system power has to be disconnected then reconnected, the power for this equipment must also be restored by pressing the front power switch.
Consequently, if this equipment is built into remote monitoring systems that use MODEMs, the standby function of this equipment must be modified.
v
ABOUT DETECTION MODE
This instrument is a spectrum analyzer which uses a digital storage system. The spectrum analyzer makes level measurements in frequency steps obtained by dividing the frequency span by the number of measurement data points (501). This method of measurement cannot detect the signal peak level if the spectrum of a received signal is narrower than these frequency steps. To resolve this problem, this instrument usually operates in positive peak detection mode and normal detection mode. In the positive peak detection mode, the highest level within the frequency range between the sample points can be held and traced. In the normal detection mode, both the positive peak and the negative peak can be traced. Positive peak detection mode should be used for almost all measurements including normal signal level measurement, pulsed noise analysis, and others. It is impossible to measure the signal level accurately in sample detection mode or in negative peak detection mode. Use of sample detection mode is restricted to random noise measurement, occupied frequency bandwidth measurement for analog communication systems, and adjacent-channel leakage power measurement, etc.
Measurement Item
• Normal signal.................................................................................................. POS PEAK
• Random noise.................................................................................................. SAMPLE
• Pulsed noise..................................................................................................... NORMAL (POSI-NEG)
• Occupied frequency bandwidth, adjacent-channel leakage power.......... SAMPLE
(for analog communication systems)
• Occupied frequency bandwidth, adjacent-channel leakage power.......... POS PEAK or SAMPLE
(for digital communication systems)
When a detection mode is specified as one of the measurement methods, make the measurement in the specified detection mode.
vi

ABOUT THIS MANUAL

(1) Composition of MS2650/MS2660B/C series spectrum analyzer Operation Manuals
The MS2650/MS2660B/C Spectrum Analyzer operation manuals of the standard type are composed of the
following three documents. Use them properly according to the usage purpose.
Basic operating instruction part
Vol. 1
Detailed operating instruction part
Composition
of
Operation Manuals
Vol. 2
Programming part
Vol. 3
(Remote control part)
(PTA control part)
Basic operating instruction part: Basic Operating Instructions: Provides information on the MS2650/
MS2660B/C series outline, preparation before use, panel description,
basic operation, soft-key menu and performance tests.
Detailed operating instruction part: Detailed Operating Instructions: Provides information on the detailed
panel operating instructions on the spectrum analyzer that expand
on the basic operation and soft-key menu in the Basic Operating
Instruction Part.
Programming part: Composed of the Remote Control Part and PTA Control Part. The
Remote Control Part provides information on RS-232C remote
control GPIB remote control and sample programs, while the PTA
Control Part describes about PTA operation and PTL commands.
I

TABLE OF CONTENTS

For Safety ................................................................................................................. iii
ABOUT THIS MANUAL.................................................................................................... I
SECTION 1 BASIC OPERATION PROCEDURE ........................................................ 1-1
Signal Display .............................................................................................................................. 1-3
Marker Operation ......................................................................................................................... 1-6
"Measure" Function Check .......................................................................................................... 1-8
Screen Hard Copy ........................................................................................................................ 1-9
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY .......................................... 2-1
Setting Observation Frequency .................................................................................................... 2-3
Setting Level Range ..................................................................................................................... 2-8
Offsetting Reference Level ........................................................................................................ 2-13
SECTION 3 MARKER FUNCTIONS ......................................................................... 3-1
Changing Zone Marker Position and Width ................................................................................ 3-4
Marker Mode ............................................................................................................................... 3-7
Display Line ............................................................................................................................... 3-10
Multimarker ............................................................................................................................... 3-11
Marker Search ............................................................................................................................ 3-16
Setting Parameters Using Marker Values .................................................................................. 3-22
SECTION 4 SIGNAL SEARCH FUNCTION ............................................................. 4-1
Detecting Peaks ............................................................................................................................ 4-3
Moving the Measurement Point ................................................................................................... 4-5
II
SECTION 5 SELECTING THE DISPLAY METHOD ................................................. 5-1
Display Mode ............................................................................................................................... 5-3
Storage Mode ............................................................................................................................. 5-15
Detection Mode .......................................................................................................................... 5-22
Time Domain ............................................................................................................................. 5-26
SECTION 6 SELECTING THE SWEEP METHOD ................................................... 6-1
Sweep Mode ................................................................................................................................ 6-3
Trigger Mode ............................................................................................................................... 6-5
Zone Sweep and Signal Tracking .............................................................................................. 6-12
Time Gate Function ................................................................................................................... 6-14
Domain Sweep ........................................................................................................................... 6-22
SECTION 7 COUPLED FUNCTION ......................................................................... 7-1
From Auto to Manual Operation ................................................................................................. 7-4
SECTION 8 AUTOMATIC CALIBRATION AND
LEVEL CORRECTION FUNCTIONS .................................................... 8-1
Automatic Calibration Function CAL ........................................................................................ 8-3
Preselector tuning ........................................................................................................................ 8-6
Measurement System Level Correction ....................................................................................... 8-8
SECTION 9 SYSTEM SETTING AND PRESET FUNCTION ................................... 9-1
Coupled Function Common/Independent Setting Mode ............................................................. 9-4
SECTION 10 SAVE/RECALL FUNCTION ................................................................ 10-1
Internal Register ......................................................................................................................... 10-4
Memory Card File Management .............................................................................................. 10-11
SECTION 11 COPY/SOUND MONITOR/TV IMAGE MONITOR .............................. 11-1
Direct Plotting ............................................................................................................................ 11-3
Saving Screen Image Data to Memory Card ............................................................................. 11-9
III
Displaying a Title ..................................................................................................................... 11-11
SOUND Monitor ...................................................................................................................... 11-12
TV Image Monitor ................................................................................................................... 11-14
SECTION 12 PTA/DEFINE FUNCTIONS ................................................................. 12-1
PTA Program Editing and Loading ........................................................................................... 12-3
User-Definition Function ........................................................................................................... 12-6
SECTION 13 MEASUREMENT ................................................................................ 13-1
Measure Measurement Function ................................................................................................ 13-3
Measurement Examples ............................................................................................................. 13-8
SECTION 14 TRACKING GENERATOR .................................................................. 14-1
Tracking Generator Menus ........................................................................................................ 14-3
Normalize/Instant-Normalize Functions .................................................................................... 14-4
Transmission Characteristics Measurement .............................................................................. 14-6
Reflection Characteristics Measurement ................................................................................. 14-13
Notes on Active Device Measurement .................................................................................... 14-17
SECTION 15 EMC, Measurement of Field Strength ................................................. 15-1
Menu of EMC ............................................................................................................................ 15-3
Field Strength Measurement ...................................................................................................... 15-4
User Antenna Factor Setting, Save/Load to /from a Memory Card .......................................... 15-7
Caution: When Performing Field Strength Automatic Measurement ..................................... 15-13
EMC Measurement .................................................................................................................. 15-15
APPENDIX A SOFT-KEY MENU ............................................................................... A-1
Soft-key Menu List ..................................................................................................................... A-4
Menu Tree ................................................................................................................................... A-6
APPENDIX B KEYWORDS INDEX........................................................................... B-1
IV
.

SECTION 1 BASIC OPERATION PROCEDURE

SECTION 1
BASIC OPERATION PROCEDURE
TABLE OF CONTENTS
Signal Display
Turn the power on
Execute automatic calibration
Set the signal to the center of the screen
Enlarge and display the signal
Marker Operation
Check of the zone marker function
The "marker CF" function check
"Measure" Function Check
Screen Hard Copy
.....................................................................................................................................
.......................................................................................................................
................................................................................................................................
..................................................................................................................
...............................................................................................................................
.....................................................................................................
...................................................................................
....................................................................................................
1-3
1-3
1-4
1-4
1-5
1-6
1-8
1-9
1-1
SECTION 1 BASIC OPERATION PROCEDURE
1-2
SECTION 1 BASIC OPERATION PROCEDURE
SECTION 1 BASIC OPERATION PROCEDURE
The basic operation procedure of this equipment is
explained here. The operations are listed on the right.
Also, the explanation will advance assuming that a
500 MHz signal is applied to the input connector.
Please read this manual while operating this
equipment.
( : Panel key, : Soft key)
<Actual operations>

Signal Display

Turn the power on

Press the standby button on the rear panel, then
press the power switch (0) on the front panel. In
this case, continue pressing the power switch for
one second or more.
(I) Signal display
1) Turn the power on,
2) execute automatic calibration,
3) set the signal to the center of the screen, and
4) enlarge and display the signal.
(II) Marker operation
Check of the zone marker function.
The "marker CF" function check.
(III) "Measure" function check
(IV) Screen hard copy
Press Preset key.
Press Preset All key in the menu.
Fig. 1-1
The power is turned on/off only when the power
switch is pressed for one second or more. This
prevents the power from being turned on/off easily
by mistake.
When panel key (hard key) is pressed, the related
soft key menu is displayed.
Partial resettings are enabled. This resetting
includes only the display-related resetting or the
resetting of special modes such as zone sweep.
1-3
SECTION 1 BASIC OPERATION PROCEDURE

Execute automatic calibration

Wait after switching on the power supply of the
machine (warm up period) till the internal temperature
becomes stable. This period is approximately 10
minutes.
After warm up, execute automatic calibration.
CAL
Press Shift key then 0 key.
Select All Cal from the menu displayed on the display.
Fig. 1-2

Set the signal to the center of the screen

Automatic Calibration is carried out by using an
internal source without need for any external cable
connection.
See "Detailed Operation Instructions" for detail
information about contents of calibration.
1-4
Press Frequency key.
Fig. 1-3
When pressing Frequency, Span, Amplitude or
Coupled Function key(s) which is used frequently,
Center Frequency, Span, Reference Level, RBW
or VBW function is selected and numeric value
for the function can be entered into Entry area.
This reduces key operation times.
This display section is called Entry area. Selecting
the menu displays the current set value of the
parameter. The set value can be changed by
entering data in Entry area.
Press Menu On/Off key
Fig. 1-4
Press Menu On/Off key to return to previous screen.
Use the ten-key pad (numeric keys) to enter 500 MHz.
SECTION 1 BASIC OPERATION PROCEDURE
The display of the soft key menu can be switched
on/off using Menu On/Off key. When the menu
disappears, the scale is enlarged. Also, when the
menu is displayed, the scale is reduced.
Fig. 1-5

Enlarge and display the signal

Press Span key , then press the V down key several times to enlarge the signal display.
The following three methods to input numeric
values to parameters are provided: direct input by
the ten-key pad (numeric keys), up/down keys,
and rotary knob.
Fig. 1-6
1-5
SECTION 1 BASIC OPERATION PROCEDURE

Marker Operation

Here, checks that the signal frequency and level are displayed in a marker display area. The zone marker
automatically fetches the highest level signal within the zone and displays the frequency and level.
Fig. 1-7
To check Marker CF function, shift the signal from the center intentionally.
Press Frequency key and More key in order, and then Scroll key two times.
The soft key menu marked by an asterisk on the
upper right indicates that the menu can further be
opened by pressing the key. Adversely, the soft
key menu not marked indicates that the menu
cannot be opened any more, so to speak, the end
of menu opening.
The following items can easily be checked by the
soft key menu tab: How many pages of the soft
key menu being displayed currently are there?,
Fig. 1-8
and what page is displayed now?
To turn over the page, press More key.
1-6
Press Peak Search key.
Fig. 1-9
The marker fetches the signal.
Press More to open 2nd page, and press Marker
key.
SECTION 1 BASIC OPERATION PROCEDURE
*Advanced operation memo: It is convenient that
the page can also be turned over by repeatedly
pressing the panel key. This method is used when
key (s), such as Measure key, has a number of
pages. Besides, the Freq/Ampl and Marker-related
keys do not turn over the page by repeatedly
pressing the panel key. For these keys, because
the first page is important specially, it should always
be displayed when the panel key is pressed.
When the soft key menu is pressed, a menu of
function related to the menu is further displayed.
In this case, as shown in the figure on the left, the
thick line (the line on the preceding page) is
displayed at the left of the soft key menu. This
indicates that a new menu is overlapped with the
preceding page.
Fig. 1-10
Press marker CF key.
The page opened by pressing the soft key can
return to the preceding page by the Return key.
Besides, it can be checked that which soft key
menu was pressed previously to open the current
menu, as the menu title is displayed on the upper
row of the soft key.
Fig. 1-11
Here, return to the screen of Fig. 1-8 and ensure that the screen changes to that of Fig. 1-11 only by pressing
the CF key.
1-7
SECTION 1 BASIC OPERATION PROCEDURE

"Measure" Function Check

Press Preset key and Preset All key in order.
Press Peak Search key.
If the zero beat signal level (local feed through) is larger than the signal level and the marker fetches the zero
beat level, press "Next peak" key and put the marker on the signal.
Fig. 1-12
Press the Measure key and Frequency Count key to set the function of high accuracy frequency measurement
of the marker points.
Then, press the Count On key and start measurement.
Fig. 1-13
The frequency of marker points is displayed at the top left of the screen.
Incidentally, the internal counter correctly operates even at the full span condition, so an operation to reduce
frequency span otherwise required is not necessary in this model.
The soft-key menu display can be switched On/
Off by the Menu On/Off key.
However, keys that condition setting is not
possible unless a menu is On unconditionally
make the soft-key menu display On when pressing
a panel key.
From the screen after executing measurement, press
another panel key and change parameters, and then,
pressing again the Measure key will automatically
return to the menu of this screen and not to page 1
of the menu (page learning function).
It is a useful function when repeating measurement.
1-8

Screen Hard Copy

The screen can be hard-copied with the VP-600 printer (Epson) via an RS232C interface, and the procedures
are described below:
1) As illustrated below, connect the RS-232C connector and printer with an attached RS-232C cable.
2) Press the Copy key, and the currently displayed screen is hard-copied.
If the printed copy is improper, check if the RS-232C interface is correctly set in the following sequence.
3) Press the Shift key and then the Interface key.
4) Press the Connect to Controller key several times to get None on the display, and press the Connect to
Prt/Plt key several times and get RS-232C on the display.
Now the printer can be operated with RS-232C.
5) Press the RS232C Setup key and set so that (or check if) the setting of RS-232C interface is the same
between the main body and printer.
(For the setting/checking of the RS-232C interface on the printer side, refer to the instruction manual of
the printer.)
SECTION 1 BASIC OPERATION PROCEDURE
6) Press the Shift key and then the Copy Cont key.
7) Press the Printer/Plotter key and select Printer.
8) Press the Printer Setup key, and then press the VP-600 key.
9) Press the Magnify key several times and make the display 1×1.
10) Press the Copy key, and the currently displayed screen is hard-copied.
Rear panel
OPT01
OPT15
Phone
IF Out(455kHz/10.695MHz)
OPT07
Composite Out
10MHz STD
Ref In Buff Out
0dBm > 2Vp-p
OPT06
Trig/Gate In
(± 10V)
Sweep Status(Z)Video(Y)Sweep(X)
7.5
Buff Out
OPT15
TTL0-0.5V0-10V
Freq Adj.
WARNING
CAUTION
Line Input 320VA Max
50/60/400Hz 100−120V 5A 200−240V 5A
RS−232C
Off
On
RS-232C cable
Printer
Fig. 1-14
1-9
SECTION 1 BASIC OPERATION PROCEDURE
1-10
.

SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY

SECTION 2
FREQUENCY/AMPLITUDE DATA ENTRY
This section describes the data entry function related to frequency and amplitude in the Freq/Ampl section on the front panel.
TABLE OF CONTENTS
Setting Observation Frequency
Center-Span Mode
Start-Stop Mode
Setting Step Size with Step Keys
Setting Frequency Scroll Step Size
Fixing the frequency bands
Setting Full Scan
Setting Zero Span
Setting Level Range
Setting Log/Linear Scale
Selecting Reference Level Units
Setting Reference Level
Setting Reference Level Step Size
Offsetting Reference Level
Setting Attenuator
Preamp Setting
Setting 50 Ω→75 Impedance Transformer
Setting Level Frequency Correction Coefficient
......................................................................................................................
..........................................................................................................................
.........................................................................................................................
.......................................................................................................................
............................................................................................................................
.......................................................................................................................
...........................................................................................................................
...........................................................................................................
................................................................................................
.........................................................................................................
.............................................................................................................
.................................................................................................
..............................................................................................................
..................................................................................................................
............................................................................................
.............................................................................................
.............................................................................
..........................................................................
2-3
2-4
2-5
2-6
2-6
2-7
2-8
2-8
2-9
2-10
2-11
2-12
2-13
2-14
2-16
2-16
2-16
2-17
2-1
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY
2-2
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY
Trig/Gate
MS2661B
Spectrum Analyzer 9kHz–3GHz
RF Input 50
+30dBm
±50V DC Max
+30dBm(PRE AMP On)
Local
Remote
Stby
On
Preset
Multi Mkr
Marker
Frequency
Span
Amplitude
Continuous
Single
Save
Recall
Measure
TG
A/B,A/BG
A,B
A/Time
Time
TV Monitor
CF RLV
Peak
Search
User
Marker
TG Output 50
9kHz–3GHz
RVS PWR
+20dBm
0V DC Max
EntryMarkerFreq/Ampl
dBm dB
V sec
mV msec
µV µsec
0
Cal
.
– / +
BS
CE
8 9 GHz7
PTA
EMC
HoldDefine
.
Hz
EnterInterface Mem Card
2 3 kHz1
Sound System Title
5 6 MHz4
Copy
Copy Cont
Shift
Peak
Menu On/Off
More
F 1
F 2
F 3
F 4
F 5
F 6
RBW VBW Sweep Time Atten
Display
Coupled Finction
Memory Card
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY

Setting Observation Frequency

The observation frequency of the spectrum analyzer is set in the following two modes:
Center-Span
Start-Stop
The frequency setting upper and lower limits are For the MS2651B/2661B/2661C, 0 to 3 GHz
For the MS2653B/2663B/2663C, 0 to 8.1 GHz, respectively.
The Frequency key is used as the header key for setting the frequency, and the Span key is used
as the header key for setting the frequency span.
Frequency
Span
2-3
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY

Center-Span Mode

(1) Setting center frequency
Frequency
(2) Setting frequency span
Span
TEN KEY
TEN KEY
Key
キー
キー
(1/2/5 sequence)
Key
(Use the CF Step Size key to set the step size.)
2-4
Note: When frequency span is 200 kHz or less, warming up might be necessary until the observation
frequency becomes stable after turning on the power.
When "Warm Up" message is being indicated at the top right corner of the indicator, please wait
for approximately 3 minutes, and start the measurements after the message disappears.

Start-Stop Mode

(1) Start frequency
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY
Frequency
(2) Stop frequency
Frequency
Notes: Because the and keys are the step keys for the center frequency, the start and stop
Start Freq
Stop Freq
TEN KEY
TEN KEY
frequencies are also changed.
The stop frequency may also vary depending on the values of the frequency span setting resolution
and start frequency.
2-5
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY

Setting Step Size with Step Keys

To use the keys to change the step size of the center frequency, register the step size as follows:
Frequency CF Step Size

Setting Frequency Scroll Step Size

Frequency
Scroll Step Size
Copy Cont
Stby
Spectrum Analyzer
MS2661B
Preset
Remote
Local
Copy
On
9kHz–3GHz
Memory Card
TEN KEY
Menu On/Off
TG Output 50
9kHz–3GHz
F 1
F 2
F 3
F 4
F 5
F 6
More
Frequency
Span
Amplitude
Peak
CF RLV
Save
Recall
Measure
Display
A/B,A/BG
A/Time
A,B
Time
Coupled Finction
RBW VBW Sweep Time Atten
RVS PWR
+20dBm
0V DC Max
1 div
2 div
EntryMarkerFreq/Ampl
PTA
8 9 GHz7
EMC
5 6 MHz4
Sound System Title
2 3 kHz1
Cal 0
.
.
CE
BS
HoldDefine
EnterInterface Mem Card
– / +
Hz
RF Input 50
+30dBm
±50V DC Max
+30dBm(PRE AMP On)
Shift
dBm dB
V sec
mV msec
µV µsec
Multi Mkr
Marker
Marker
Peak Search
Continuous
Single
TV Monitor
Trig/Gate
User
TG
Select a step size. The selected
scroll step size is highlighted. 5 div
10 div
2-6
2 div
Tr-A
return
Span
Return to the top menu.
Scroll
The waveform scrolls 2 divs to the right.
Span
Scroll
The waveform scrolls 2 divs to the left.

Fixing the frequency bands

This function is an MS2653B/2663B/2663C dedicated function. It cannot be used in the MS2651B/2661B/
2661C.
In the MS2653B/2663B/2663C, the 0 to 8.1 GHz frequency range consists of the following three bands:
Band 0 .................. .0 to 3.2 GHz
Band 1-................ 2.92 to 6.5 GHz
Band 1+............... 6.4 to 8.1 GHz
In the initial state, the Auto Band mode that is operated by selecting the optional frequency band is selected
according to the range of frequencies to be observed.
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY
Frequency band configuration
Frequency range to be observed
0 2.92 3.2 6.4 6.5 8.1
Band 0 Band 1
­Band 1
(GHz)
+
Selection of frequency bands according to range of frequencies to be observed in Auto Band mode
Perform the following to set the frequency bands, for example, when the frequency bands are switched:
Span
Band
Auto Band
Manual Band 0
Manual Band 1
Manual Band 1
-
Select any of these buttons.
+
return
2-7
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY

Setting Full Scan

In the normal operating state, pressing
analyzer to be swept over the full span. However, this setting also initializes the parameters except the frequency
range.
To set the full span and leave the other parameters unchanged, perform the following key operations.
Span
For the MS2651B/2661B/2661C, 0 to 3 GHz
For the MS2653B/2663B/2663C Auto band, 0 to 8.1 GHz
For Band 0, 0 to 3.2 GHz
For Band 1-, 2.92 to 6.5 GHz
For Band 1+, 6.4 to 8.1 GHz

Setting Zero Span

The Spectrum Analyzer can operate as a selective level meter in which the horizontal axis is graduated as a
time axis by setting the frequency span to 0 Hz. The rising and falling edges of burst waves can also be
observed and measured.
Preset
the key allows the entire frequency range of the spectrum
Full Span
2-8
Performing any of the following key operations allows the spectrum analyzer to operate in the zero panel (time
domain) mode.
Hz
Time
Span
Span
0
Zero Span
For further details on the zero span (time domain) mode, see SECTION 5, "SELECTING THE DISPLAY
METHOD."
In the frequency and time domains, the RBW, VBW, Sweep time and other coupling functions time can be set
to different values. For further details, see SECTION 9, "SYSTEM SETTING AND PRESET FUNCTION."

Setting Level Range

The table below shows the types of level display modes and the ranges of the reference level (top graticule of
the amplitude scale) for the different modes. When a preamp (Option) is used, the entire reference level range
is shifted 20 dB.
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY
50 (standard) 75 (with Opt.22)
MS2651B/2661B/2661C only
Reference Level rangeUnitsDisplay mode
–100 to +25 dBm
+8.8 to +133.8 dBµV
–51.2 to 73.8 dBmV
2.74 µV to 4.87 V
+14.8 to +139.8 dBµV
100 fW to 316 mW
– – – – – –
274 µV to 4.87 V
Log scale
Linear scale
dBm
dBµV
dBmV
V
dBµV (emf)
W
dBµV/m
V
–100 to +30 dBm
+7 to +137 dBµV
–53 to 77 dBmV
2.24 µV to 7.07 V
+13 to +143 dBµV
100 fW to 1.0 W
– – – – – –
224 µV to 7.07 V
dBm: dBm unit system where 1 mW/50 Ω or 75 Ω is defined as 0 dBm.
dBµV: dBµV unit system where 1µV is defined as 0 dBµV, and the terminal voltage display is
terminated into 50 or 75 Ω.
dBmV: dBmV unit system where 1 mV is defined as 0 dBmV, and the terminal voltage display is
terminated into 50 or 75 Ω.
dBµV (emf): dBµV (emf) unit system based on the open-voltage display, and dBµV +6 dB is fed as the
output value.
The Amplitude key is used as the header key for setting the amplitude level.
Amplitude
Preset
Remote
Local
Copy Cont
Copy
Stby
Spectrum Analyzer
MS2661B
9kHz–3GHz
On
Memory Card
Menu On/Off
TG Output 50
9kHz–3GHz
F 1
F 2
F 3
F 4
F 5
F 6
More
Frequency
Span
Amplitude
Peak
CF RLV
Save
Recall
Measure
Display
A/B,A/BG
A/Time
A,B
Time
Coupled Finction
RBW VBW Sweep Time Atten
RVS PWR +20dBm
0V DC Max
Multi Mkr
Marker
Marker
Peak Search
Continuous
Single
TV Monitor
Trig/Gate
User
TG
PTA
8 9 GHz7
EMC
5 6 MHz4
Sound System Title
2 3 kHz1
Cal 0
.
.
EntryMarkerFreq/Ampl
CE
BS
HoldDefine
EnterInterface Mem Card
– / +
Hz
RF Input 50
+30dBm
±50V DC Max
+30dBm(PRE AMP On)
Shift
dBm dB
V sec
mV msec
µV µsec
2-9
SECTION 2 FREQUENCY/AMPLITUDE DATA ENTRY

Setting Log/Linear Scale

To set the amplitude scale to log scale or linear scale, perform the following key operations.
(1) Setting log scale
Amplitude
(2) Setting linear scale
Amplitude
Log Scale 10 dB/div
The log scale is set when
this key is pressed.
Linear Scale 10%/div
The linear scale is set when
this key is pressed.
Select one of the scales. 5 dB/div
2 dB/div
1 dB/div
Return
Select one of the scales. 5%/div
2%/div
1%/div
Return
2-10
The reference level remains constant, independent of switching between log and linear scales.
When the reference level is set to less than –60 dBm in the log scale mode, the reference level of the linear
scale is switched to 224 µV/50 or 274 µV/75 Ω.
Loading...
+ 290 hidden pages