Anritsu ME7838D4, ME7838G4 Quick Start Guide

Quick Start Guide
VectorStar
ME7838D4/G4 Multiport Broadband/Banded Millimeter-Wave System
High Performance Modular Broadband/Banded Millimeter-Wave Vector Network Analyzer (VNA) Multiport Measurement System
Anritsu Company 490 Jarvis Drive Morgan Hill, CA 95037-2809 USA
http://www.anritsu.com
Part Number: 10410-00770
Copyright 2021 Anritsu Company, USA. All Rights Reserved.
Published: February 2021
Revision: C

1. ME7838D4 and ME7838G4 Main Components

High Performance Modular Broadband/Banded Millimeter-Wave Vector Network Analyzer (VNA) Measurement System
This guide provides quick setup instructions for the ME7838D4 and ME7838G4 Multiport Broadband/Banded mm-Wave VNA System assemblies. For additional safety and compliance information, and for more details about the assembly, configuration, setup, and initial testing of the equipment, refer to the VectorStar™ ME7838x4 Series Multiport Broadband Vector Network Analyzers Installation Guide – 10410-00734.
This and all other documentation that supports the ME7838D4 and ME7838G4 is available on the VectorStar product web page:
http://www.anritsu.com/test-measurement/products/me7838a4
On this web page, you can select various tabs for more information about your instrument. Included is a Library tab which contains links to all the latest technical documentation related to this instrument.
1. ME7838D4 and ME7838G4 Main Components

ME7838D4 Broadband Systems

The ME7838D4 Multiport Broadband system consists of the following components:
MS4647A or MS4647B VNA with Option 007 (Receiver Offset) and Option 08x (Modular Broadband
Connection Capability)
MN4697C Multiport Test Set
3736B Broadband Test Set
3739C Broadband Test Set
Four MA25300A Millimeter-Wave Modules
Front and rear panel cables

ME7838G4 Broadband Systems

The ME7838G4 Multiport Broadband system consists of the following components:
MS4647B VNA with Option 007 (Receiver Offset) and Option 08x (Modular Broadband Connection
Capability)
MN4697C Multiport Test Set
3736B Broadband Test Set
3739C Broadband Test Set
Four MA25400A Millimeter-Wave Modules
Front and rear panel cables

ME7838D4 Banded Systems

The ME7838D4 Multiport Banded system consists of the following components:
MS464xA/B VNA with Option 08x
MN4694C Multiport Test Set
3736B Broadband Test Set
3739C Broadband Test Set
Four 3744A-EE, 3744A-EW, or four OML/VDI Millimeter-Wave Modules
Front and rear panel cables
QSG-2 PN: 10410-00770 Rev. C VectorStar ME7838D4/G4 QSG
1. ME7838D4 and ME7838G4 Main Components

ME7838G4 Banded Systems

The ME7838G4 Multiport Banded system consists of the following components:
MS464xB VNA with Option 08x
MN4694C Multiport Test Set
3736B Broadband Test Set
3739C Broadband Test Set
Four 3744A-EE, 3744A-EW, or four OML/VDI Millimeter-Wave Modules
A MS464xA/B VNA unit is heavy. To avoid personal injury, it must be lifted and maneuvered by at least two people during installation.
Caution
If mounting on a workbench surface, first position the 3739C Broadband Test Set with access to its front and rear panels. Stack the remaining test sets on top of one another, then finally the VNA.
If mounting into rack or console, make sure the Test Sets have been installed, and that the rack/console is carefully positioned on a flat and level surface. If equipped, make sure any casters are locked. Use two people to lift the VNA unit and two to guide it into its shelf rails.
The test loops on the front and rear panels of the VNA are delicate. Be careful not to bump or bend the test loops.
VectorStar ME7838D4/G4 QSG PN: 10410-00770 Rev. C QSG-3

2. Rear Panel Connections Between System Components (MS464xA VNA)

WARNING
No operator service­able parts inside. Refer servicing to qualified personnel.
Ext Alc Out
Ext Analog In
a
1
b a
2
b
2
IF Outputs
External I/O
~ LINE INPUT
47-63 Hz 85-240 VAC 150 VA Max
Fuse 250V 2A T
CAUTION
Do not operate with power cord ungrounded.
CAUTION
For continued fire protection replace only with specified type and rated fuse.
Cable Length
Standard
Extended
Module Interface
MN469xC
Test Set
3736B
Test Set
3739C
Test Set
MS464xA
VNA
1
1
2
2
3
4
11 151617
18121314
19
20
21
22
5
6
6
6
6
7 7 7 7
7 7 7 7
8
8
9
9
2. Rear Panel Connections Between System Components (MS464xA VNA)
In this section, connect the cables between the VNA and the Test Sets as shown in Figure 1 and Table 1.
Figure 1 ME7838D4 – Rear Panel Cable Connections (MS464xA VNA)
QSG-4 PN: 10410-00770 Rev. C VectorStar ME7838D4/G4 QSG
2. Rear Panel Connections Between System Components (MS464xA VNA)
Table 1. ME7838D4 Multiport Rear Panel Cable Connections (1 of 2)
Index Part Number Cable Description Connection From Connection To
1 3-806-225 BNC (male-male), 24 in
2 3-806-225 BNC (male-male), 24 in
3 3-806-225 BNC (male-male), 24 in
4 3-806-225 BNC (male-male), 24 in
5
3-73598-1
(5 cable
6
bundle)
DB-25
SMA male-male flexible:
a
3-72243-1
3-72243-2
3-72243-3
3-72243-4
SMA male-male flexible:
3-72243-21
7
3-80734
(5 cable
bundle)
a
3-72243-22
3-72243-23
3-72243-24
8DB-25
9 2100-1-R GPIB, 1 m
SMA male-male
11 3-62112-81
12 3-62112-80
13 3-62112-81
14 3-62112-80
15 3-62112-81
16 3-62112-80
17 3-62112-81
a
semi-rigid
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
MS464xA port labeled:
Ext Analog Out
MS464xA port labeled:
Ext ALC
3736B port labeled:
(To Test Set) EXT ANALOG OUT
3736B port labeled:
(To Test Set) EXT ALC IN
3736B port labeled:
TO TEST SET EXTERNAL I/O
3736B ports labeled:
(IF INPUTS)
a1 b1 a2 b2
MS464xA ports labeled:
(Inputs/Outputs)
a1 IF b1 IF a2 IF b2 IF
MS464xA port labeled:
External I/O
MS464xA port labeled:
Dedicated GPIB
MS464xA port labeled:
b2 loop out
MS464xA port labeled:
b2 loop in
MS464xA port labeled:
P2 source loop out
MS464xA port labeled:
P2 source loop in
MS464xA port labeled:
b1 loop out
MS464xA port labeled:
b1 loop in
MS464xA port labeled:
P1 source loop out
3736B port labeled:
EXT ANALOG IN
3736B port labeled:
EXT ALC OUT
3739C port labeled:
EXT ANALOG IN
3739C port labeled:
EXT ALC OUT
3739C port labeled:
EXTERNAL I/O
3739C ports labeled:
(IF OUTPUTS)
a1 b1 a2 b2
3736B ports labeled:
(IF OUTPUTS)
a1 IF b1 IF a2 IF b2 IF
3736B port labeled:
TO VNA EXTERNAL I/O
MN469xC port labeled:
IEEE 488.2 GPIB
MN469xC port labeled:
TO VNA b2 OUTPUT
MN469xC port labeled:
TO VNA b2 INPUT
MN469xC port labeled:
TO VNA Port 2 Src OUTPUT
MN469xC port labeled:
TO VNA Port 2 Src INPUT
MN469xC port labeled:
TO VNA b1 OUTPUT
MN469xC port labeled:
TO VNA b1 INPUT
MN469xC port labeled:
TO VNA Port 1 Src OUTPUT
VectorStar ME7838D4/G4 QSG PN: 10410-00770 Rev. C QSG-5
2. Rear Panel Connections Between System Components (MS464xA VNA)
Tab le 1 . ME7838D4 Multiport Rear Panel Cable Connections (2 of 2)
Index Part Number Cable Description Connection From Connection To
18 3-62112-80
SMA male-male
a
semi-rigid
MS464xA port labeled:
P1 source loop in
MN469xC port labeled:
TO VNA Port 1 Src INPUT
19 MS464xA VNA
20 MN469xC Test Set
21 3736B Test Set
22 3739C Test Set
a. Tighten each cable in this group using an 8 mm (5/16 in) torque end wrench set to 0.9 N·m (8 lbf·in).
QSG-6 PN: 10410-00770 Rev. C VectorStar ME7838D4/G4 QSG
3. Rear Panel Connections Between Components (MS464xB VNA)
WARNING
No operator service­able parts inside. Refer servicing to qualified personnel.
Ext Alc Out
Ext Analog In
a
1
b a
2
b
2
IF Outputs
External I/O
~ LINE INPUT
47-63 Hz 85-240 VAC 150 VA Max
Fuse 250V 2A T
CAUTION
Do not operate with power cord ungrounded.
CAUTION
For continued fire protection replace only with specified type and rated fuse.
Cable Length
Standard
Extended
Module Interface
MN469xC
Test Set
3736B
Test Set
3739C
Test Set
MS464xB
VNA
3
11 151617
18
121314
19
20
21
22
5
6
6
6
6
7 7 7 7
7 7 7 7
8
8
9
9
Ext In ALC 1
and
Ext In ALC 2
(With Option 031)
Ext ALC 1
(With
Option 031)
1
1
2
2
2
2
4
Ext ALC 2
(With
Option 031)
In this section, connect the cables between the VNA and the Test Sets as shown in Figure 2 or Figure 3, and
Table 2.

Cable Connections (Systems with Option 031 Dual Source Architecture)

3. Rear Panel Connections Between Components (MS464xB VNA)

Figure 2 ME7838D4/G4 Multiport BB/mmW VNA System – Rear Panel Cables (With Option 031)
VectorStar ME7838D4/G4 QSG PN: 10410-00770 Rev. C QSG-7
3. Rear Panel Connections Between Components (MS464xB VNA)
WARNING
No operator service­able parts inside. Refer servicing to qualified personnel.
Ext Alc Out
Ext Analog In
a
1
b a
2
b
2
IF Outputs
External I/O
~ LINE INPUT
47-63 Hz 85-240 VAC 150 VA Max
Fuse 250V 2A T
CAUTION
Do not operate with power cord ungrounded.
CAUTION
For continued fire protection replace only with specified type and rated fuse.
Cable Length
Standard
Extended
Module Interface
MN469xC
Test Set
3736B
Test Set
3739C
Test Set
MS464xB
VNA
3
11 151617
18
121314
19
20
21
22
5
6
6
6
6
7 7 7 7
7 7 7 7
8
8
9
9
1
1
2
2
4

Cable Connections (Systems without Option 031)

Figure 3 ME7838D4/G4 Multiport BB/mmW VNA System – Rear Panel Cables (No Option 031)
QSG-8 PN: 10410-00770 Rev. C VectorStar ME7838D4/G4 QSG
3. Rear Panel Connections Between Components (MS464xB VNA)
Table 2. ME7838D4/G4 Multiport Rear Panel Cable Connections (MS464xB) (1 of 2)
Index Part Number Cable Description Connection From Connection To
1
3-806-225 BNC (male-male), 24 in
3-806-225 BNC (male-male), 24 in
2
3-806-225 BNC (male-male), 24 in
3-806-225 BNC (male-male), 24 in
3
4
5
3-806-225 BNC (male-male), 24 in
3-806-225 BNC (male-male), 24 in
3-73598-1
(5 cable
6
7
bundle)
3-80734
(5 cable
bundle)
8
9
2100-1-R GPIB, 1 m
11 3-62112-81
12 3-62112-80
13 3-62112-81
DB-25
SMA male-male flexible:
a
3-72243-1
3-72243-2
3-72243-3
3-72243-4
SMA male-male flexible:
3-72243-21
3-72243-22
a
3-72243-23
3-72243-24
DB-25
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
SMA male-male
a
semi-rigid
MS464xB port labeled:
Ext Analog Out
VNA port labeled:
Ext In ALC 1
(No Option 031) (Figure 3)
VNA port labeled:
Ext In ALC 1
(With Option 031) (Figure 2)
VNA port labeled:
Ext In ALC 2
(With Option 031) (Figure 2)
3736B port labeled:
(To Test Set) EXT ANALOG OUT
3736B port labeled:
(To Test Set) EXT ALC IN
(Figure 3)
(Not used with Option 031)
3736B port labeled:
TO TEST SET EXTERNAL I/O
3736B ports labeled:
(IF INPUTS)
a1 b1 a2 b2
MS464xB ports labeled:
(Inputs/Outputs)
a1 IF b1 IF a2 IF b2 IF
MS464xB port labeled:
External I/O
MS464xB port labeled:
Dedicated GPIB
MS464xB port labeled:
b2 loop out
MS464xB port labeled:
b2 loop in
MS464xB port labeled:
P2 source loop out
3736B port labeled:
EXT ANALOG IN
3736B port labeled:
EXT ALC OUT
3739C port labeled:
EXT ALC OUT
3736B port labeled:
EXT ALC OUT
3739C port labeled:
EXT ANALOG IN
3739C port labeled:
EXT ALC OUT
3739C port labeled:
EXTERNAL I/O
3739C ports labeled:
(IF OUTPUTS)
a1 b1 a2 b2
3736B ports labeled:
(IF OUTPUTS)
a1 IF b1 IF a2 IF b2 IF
3736B port labeled:
TO VNA EXTERNAL I/O
MN469xC port labeled:
IEEE 488.2 GPIB
MN469xC port labeled:
TO VNA b2 OUTPUT
MN469xC port labeled:
TO VNA b2 INPUT
MN469xC port labeled:
TO VNA Port 2 Src OUTPUT
VectorStar ME7838D4/G4 QSG PN: 10410-00770 Rev. C QSG-9