Anritsu ME7838D4, ME7838G4 Quick Start Guide

Quick Start Guide

VectorStarME7838D4/G4

Multiport Broadband/Banded

Millimeter-Wave System

High Performance Modular Broadband/Banded Millimeter-Wave Vector Network Analyzer (VNA) Multiport Measurement System

ME7838D4, 70 kHz to 145 (150) GHz

ME7838G4, 70 kHz to 220 (226) GHz

Anritsu Company

Part Number: 10410-00770

490 Jarvis Drive

Revision: C

Morgan Hill, CA 95037-2809

Published: February 2021

USA

Copyright 2021 Anritsu Company, USA. All Rights Reserved.

http://www.anritsu.com

 

1. ME7838D4 and ME7838G4 Main Components

High Performance Modular Broadband/Banded Millimeter-Wave Vector Network Analyzer (VNA) Measurement System

This guide provides quick setup instructions for the ME7838D4 and ME7838G4 Multiport Broadband/Banded mm-Wave VNA System assemblies. For additional safety and compliance information, and for more details about the assembly, configuration, setup, and initial testing of the equipment, refer to the VectorStar™ ME7838x4 Series Multiport Broadband Vector Network Analyzers Installation Guide – 10410-00734.

This and all other documentation that supports the ME7838D4 and ME7838G4 is available on the VectorStar product web page:

http://www.anritsu.com/test-measurement/products/me7838a4

On this web page, you can select various tabs for more information about your instrument. Included is a Library tab which contains links to all the latest technical documentation related to this instrument.

1.ME7838D4 and ME7838G4 Main Components

ME7838D4 Broadband Systems

The ME7838D4 Multiport Broadband system consists of the following components:

MS4647A or MS4647B VNA with Option 007 (Receiver Offset) and Option 08x (Modular Broadband Connection Capability)

MN4697C Multiport Test Set

3736B Broadband Test Set

3739C Broadband Test Set

Four MA25300A Millimeter-Wave Modules

Front and rear panel cables

ME7838G4 Broadband Systems

The ME7838G4 Multiport Broadband system consists of the following components:

MS4647B VNA with Option 007 (Receiver Offset) and Option 08x (Modular Broadband Connection Capability)

MN4697C Multiport Test Set

3736B Broadband Test Set

3739C Broadband Test Set

Four MA25400A Millimeter-Wave Modules

Front and rear panel cables

ME7838D4 Banded Systems

The ME7838D4 Multiport Banded system consists of the following components:

MS464xA/B VNA with Option 08x

MN4694C Multiport Test Set

3736B Broadband Test Set

3739C Broadband Test Set

Four 3744A-EE, 3744A-EW, or four OML/VDI Millimeter-Wave Modules

Front and rear panel cables

QSG-2

PN: 10410-00770 Rev. C

VectorStar ME7838D4/G4 QSG

1. ME7838D4 and ME7838G4 Main Components

ME7838G4 Banded Systems

The ME7838G4 Multiport Banded system consists of the following components:

MS464xB VNA with Option 08x

MN4694C Multiport Test Set

3736B Broadband Test Set

3739C Broadband Test Set

Four 3744A-EE, 3744A-EW, or four OML/VDI Millimeter-Wave Modules

 

A MS464xA/B VNA unit is heavy. To avoid personal injury, it must be lifted and

 

maneuvered by at least two people during installation.

Caution

If mounting on a workbench surface, first position the 3739C Broadband Test Set

 

with access to its front and rear panels. Stack the remaining test sets on top of one

 

another, then finally the VNA.

 

If mounting into rack or console, make sure the Test Sets have been installed, and

 

that the rack/console is carefully positioned on a flat and level surface. If equipped,

 

make sure any casters are locked. Use two people to lift the VNA unit and two to

 

guide it into its shelf rails.

 

The test loops on the front and rear panels of the VNA are delicate. Be careful not to

 

bump or bend the test loops.

 

 

VectorStar ME7838D4/G4 QSG

PN: 10410-00770 Rev. C

QSG-3

2. Rear Panel Connections Between System Components (MS464xA VNA)

2.Rear Panel Connections Between System Components (MS464xA VNA)

In this section, connect the cables between the VNA and the Test Sets as shown in Figure 1 and Table 1.

 

7

 

7

7

1

2

 

 

7

 

 

 

 

 

 

 

19

 

 

 

 

 

 

MS464xA

 

 

 

 

 

 

VNA

 

 

 

 

 

 

8

 

 

 

 

 

 

9

 

12

14

 

16

18

 

11

13

 

15

17

 

 

 

 

 

 

20

 

 

 

 

MN469xC

 

 

 

 

Test Set

 

 

 

 

9

8

7

7

7

7

2

 

 

 

21

 

 

 

 

3736B

1

 

 

 

Test Set

 

 

 

 

4

6

 

 

 

3

 

 

 

6

6

 

 

5

 

 

 

 

IF Outputs

6

 

CAUTION

a1

b

a2

b2

Do not operate with

 

 

power cord ungrounded.

 

 

Ext

 

 

 

 

22

 

Alc

 

 

 

 

 

Out

 

External I/O

WARNING

85-240 VAC

3739C

 

 

able parts inside.

 

 

 

No operator service-

~ LINE INPUT

 

 

 

 

47-63 Hz

 

 

 

 

 

Refer servicing to

150 VA Max

 

Test Set

Ext

 

 

qualified personnel.

Fuse 250V 2A T

 

 

 

Module Interface

 

 

Analog

 

 

Cable Length

 

 

 

In

 

 

 

CAUTION

 

 

For continued fire protection Standard Extended replace only with specified

type and rated fuse.

Figure 1 ME7838D4 – Rear Panel Cable Connections (MS464xA VNA)

QSG-4

PN: 10410-00770 Rev. C

VectorStar ME7838D4/G4 QSG

2. Rear Panel Connections Between System Components (MS464xA VNA)

Table 1.

ME7838D4 Multiport Rear Panel Cable Connections (1 of 2)

 

 

 

 

 

 

Index

 

Part Number

Cable Description

Connection From

Connection To

 

 

 

 

 

 

1

 

3-806-225

BNC (male-male), 24 in

MS464xA port labeled:

3736B port labeled:

 

Ext Analog Out

EXT ANALOG IN

2

 

3-806-225

BNC (male-male), 24 in

MS464xA port labeled:

3736B port labeled:

 

Ext ALC

EXT ALC OUT

 

 

 

 

3736B port labeled:

3739C port labeled:

3

 

3-806-225

BNC (male-male), 24 in

(To Test Set) EXT ANALOG

 

EXT ANALOG IN

 

 

 

 

OUT

 

 

 

 

 

 

 

 

 

 

 

4

 

3-806-225

BNC (male-male), 24 in

3736B port labeled:

3739C port labeled:

 

(To Test Set) EXT ALC IN

EXT ALC OUT

5

 

 

DB-25

3736B port labeled:

3739C port labeled:

 

 

TO TEST SET EXTERNAL I/O

EXTERNAL I/O

 

 

 

 

 

 

a

SMA male-male flexible:

3736B ports labeled:

3739C ports labeled:

 

 

3-73598-1

3-72243-1

(IF INPUTS)

(IF OUTPUTS)

 

 

(5 cable

6

 

bundle)

3-72243-2

a1

a1

 

 

 

 

 

3-72243-3

b1

b1

 

 

 

a2

a2

 

 

 

3-72243-4

 

 

 

b2

b2

 

 

 

 

 

 

 

 

 

SMA male-male flexible:

MS464xA ports labeled:

3736B ports labeled:

 

 

 

3-72243-21

(Inputs/Outputs)

(IF OUTPUTS)

7

 

3-80734a

3-72243-22

a1 IF

a1 IF

 

 

(5 cable

3-72243-23

b1 IF

b1 IF

 

 

a2 IF

a2 IF

 

 

bundle)

3-72243-24

 

 

b2 IF

b2 IF

 

 

 

 

 

 

 

 

 

8

 

 

DB-25

MS464xA port labeled:

3736B port labeled:

 

 

External I/O

TO VNA EXTERNAL I/O

 

 

 

 

 

 

 

 

 

 

9

 

2100-1-R

GPIB, 1 m

MS464xA port labeled:

MN469xC port labeled:

 

Dedicated GPIB

IEEE 488.2 GPIB

 

 

 

 

 

 

 

 

 

 

11

 

3-62112-81a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

b2 loop out

TO VNA b2 OUTPUT

 

 

 

semi-rigid

12

 

3-62112-80a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

b2 loop in

TO VNA b2 INPUT

 

 

 

semi-rigid

13

 

3-62112-81a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

P2 source loop out

TO VNA Port 2 Src OUTPUT

 

 

 

semi-rigid

14

 

3-62112-80a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

P2 source loop in

TO VNA Port 2 Src INPUT

 

 

 

semi-rigid

15

 

3-62112-81a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

b1 loop out

TO VNA b1 OUTPUT

 

 

 

semi-rigid

16

 

3-62112-80a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

b1 loop in

TO VNA b1 INPUT

 

 

 

semi-rigid

17

 

3-62112-81a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

P1 source loop out

TO VNA Port 1 Src OUTPUT

 

 

 

semi-rigid

VectorStar ME7838D4/G4 QSG

PN: 10410-00770 Rev. C

QSG-5

2. Rear Panel Connections Between System Components (MS464xA VNA)

Table 1.

ME7838D4 Multiport Rear Panel Cable Connections (2 of 2)

 

 

 

 

 

 

Index

 

Part Number

Cable Description

Connection From

Connection To

 

 

 

 

 

 

18

 

3-62112-80a

SMA male-male

MS464xA port labeled:

MN469xC port labeled:

 

P1 source loop in

TO VNA Port 1 Src INPUT

 

 

 

semi-rigid

19

 

MS464xA VNA

 

 

 

 

 

 

 

 

20

 

MN469xC Test Set

 

 

 

 

 

 

 

 

21

 

3736B Test Set

 

 

 

 

 

 

 

 

 

22

 

3739C Test Set

 

 

 

 

 

 

 

 

 

a. Tighten each cable in this group using an 8 mm (5/16 in) torque end wrench set to 0.9 N·m (8 lbf·in).

QSG-6

PN: 10410-00770 Rev. C

VectorStar ME7838D4/G4 QSG

Anritsu ME7838D4, ME7838G4 Quick Start Guide

3.Rear Panel Connections Between Components (MS464xB VNA)

3.Rear Panel Connections Between Components (MS464xB VNA)

In this section, connect the cables between the VNA and the Test Sets as shown in Figure 2 or Figure 3, and Table 2.

Cable Connections (Systems with Option 031 Dual Source Architecture)

 

2

 

 

 

 

19

Ext In ALC 1

 

 

 

 

MS464xB

and

 

 

 

 

 

VNA

Ext In ALC 2

2

 

 

 

 

 

(With Option 031)

 

 

 

 

 

 

1

 

 

 

 

 

 

 

12

14

 

16

18

 

11

13

 

15

17

 

 

 

 

 

 

20

 

 

 

 

 

MN469xC

 

 

 

 

 

Test Set

Ext ALC 2

 

 

 

 

9

 

 

 

 

 

(With

 

 

 

 

 

Option 031)

 

 

 

 

 

 

8

7

7

7

7

2

 

 

 

 

21

 

 

 

 

 

3736B

1

 

 

 

 

Test Set

 

 

 

 

 

4

3

6

 

 

 

 

 

 

 

 

5

6

6

 

 

 

 

IF Outputs

6

 

CAUTION

 

a1

b

a2

b2

 

Do not operate with

 

 

 

 

power cord ungrounded.

 

 

2

Ext

 

 

 

 

 

22

 

Alc

 

 

 

 

 

 

Out

 

External I/O

 

WARNING

~ LINE INPUT

3739C

Ext ALC

 

 

 

able parts inside.

85-240 VAC

 

 

 

 

 

 

No operator service-

47-63 Hz

 

 

(With

Ext

 

 

 

Refer servicing to

150 VA Max

 

Test Set

 

 

Module Interface

Fuse 250V 2A T

 

 

 

 

 

 

qualified personnel.

 

 

Option 031)

Analog

 

 

Cable Length

 

 

 

In

 

 

 

 

CAUTION

 

 

 

 

 

 

Standard

Extended

For continued fire protection

 

 

 

 

 

 

replace only with specified

 

 

 

 

 

 

 

 

type and rated fuse.

 

 

Figure 2 ME7838D4/G4 Multiport BB/mmW VNA System – Rear Panel Cables (With Option 031)

VectorStar ME7838D4/G4 QSG

PN: 10410-00770 Rev. C

QSG-7

3. Rear Panel Connections Between Components (MS464xB VNA)

Cable Connections (Systems without Option 031)

 

 

 

 

 

19

 

 

 

 

 

MS464xB

 

 

 

 

 

VNA

2

 

 

 

 

 

1

 

 

 

 

 

 

12

14

 

16

18

11

13

 

15

17

 

 

 

 

 

 

20

 

 

 

 

 

MN469xC

 

 

 

 

 

Test Set

 

 

 

 

 

9

 

8

7

7

7

7

2

 

 

 

 

21

 

 

 

 

 

3736B

1

 

 

 

 

Test Set

 

 

 

 

 

4

3

6

 

 

 

 

5

6

6

 

 

 

 

IF Outputs

6

 

CAUTION

a1

b

a2

b2

 

Do not operate with

 

 

 

power cord ungrounded.

 

 

Ext

 

 

 

 

 

22

 

Alc

 

 

 

 

 

 

Out

 

External I/O

 

WARNING

85-240 VAC

3739C

 

 

 

able parts inside.

 

 

 

 

No operator service-

~ LINE INPUT

 

 

 

 

 

47-63 Hz

 

 

 

 

 

 

Refer servicing to

150 VA Max

 

Test Set

Ext

 

 

 

qualified personnel.

Fuse 250V 2A T

 

 

 

Module Interface

 

 

Analog

 

 

Cable Length

 

 

 

In

 

 

 

 

CAUTION

 

 

 

 

 

Standard

Extended

For continued fire protection

 

 

 

 

 

replace only with specified

 

 

 

 

 

 

 

type and rated fuse.

 

 

Figure 3 ME7838D4/G4 Multiport BB/mmW VNA System – Rear Panel Cables (No Option 031)

QSG-8

PN: 10410-00770 Rev. C

VectorStar ME7838D4/G4 QSG

3. Rear Panel Connections Between Components (MS464xB VNA)

Table 2.

ME7838D4/G4 Multiport Rear Panel Cable Connections (MS464xB) (1 of 2)

 

 

 

 

 

Index

 

Part Number

Cable Description

Connection From

Connection To

 

 

 

 

 

 

1

 

3-806-225

BNC (male-male), 24 in

MS464xB port labeled:

3736B port labeled:

 

Ext Analog Out

EXT ANALOG IN

 

 

 

 

VNA port labeled:

3736B port labeled:

 

 

3-806-225

BNC (male-male), 24 in

Ext In ALC 1

 

 

EXT ALC OUT

 

 

 

 

(No Option 031) (Figure 3)

 

 

 

 

 

 

 

 

 

VNA port labeled:

3739C port labeled:

2

 

3-806-225

BNC (male-male), 24 in

Ext In ALC 1

 

EXT ALC OUT

 

 

 

 

(With Option 031) (Figure 2)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

VNA port labeled:

3736B port labeled:

 

 

3-806-225

BNC (male-male), 24 in

Ext In ALC 2

 

 

EXT ALC OUT

 

 

 

 

(With Option 031) (Figure 2)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

3736B port labeled:

3739C port labeled:

3

 

3-806-225

BNC (male-male), 24 in

(To Test Set) EXT ANALOG

 

EXT ANALOG IN

 

 

 

 

OUT

 

 

 

 

 

 

 

 

 

3736B port labeled:

 

4

 

3-806-225

BNC (male-male), 24 in

(To Test Set) EXT ALC IN

3739C port labeled:

 

 

 

 

(Figure 3)

EXT ALC OUT

 

 

 

 

(Not used with Option 031)

 

5

 

 

DB-25

3736B port labeled:

3739C port labeled:

 

 

TO TEST SET EXTERNAL I/O

EXTERNAL I/O

 

 

 

 

 

 

 

 

 

 

 

 

3-73598-1a

SMA male-male flexible:

3736B ports labeled:

3739C ports labeled:

 

 

(5 cable

3-72243-1

(IF INPUTS)

(IF OUTPUTS)

6

 

bundle)

3-72243-2

a1

a1

 

 

 

 

 

3-72243-3

b1

b1

 

 

 

a2

a2

 

 

 

3-72243-4

 

 

 

b2

b2

 

 

 

SMA male-male flexible:

MS464xB ports labeled:

3736B ports labeled:

 

 

 

3-72243-21

(Inputs/Outputs)

(IF OUTPUTS)

7

 

3-80734a

3-72243-22

a1 IF

a1 IF

 

 

(5 cable

3-72243-23

b1 IF

b1 IF

 

 

a2 IF

a2 IF

 

 

bundle)

3-72243-24

 

 

b2 IF

b2 IF

 

 

 

 

 

 

 

 

 

8

 

 

DB-25

MS464xB port labeled:

3736B port labeled:

 

 

External I/O

TO VNA EXTERNAL I/O

 

 

 

 

 

 

 

 

 

 

9

 

2100-1-R

GPIB, 1 m

MS464xB port labeled:

MN469xC port labeled:

 

Dedicated GPIB

IEEE 488.2 GPIB

 

 

 

 

 

 

 

 

 

 

11

 

3-62112-81a

SMA male-male

MS464xB port labeled:

MN469xC port labeled:

 

b2 loop out

TO VNA b2 OUTPUT

 

 

 

semi-rigid

12

 

3-62112-80a

SMA male-male

MS464xB port labeled:

MN469xC port labeled:

 

b2 loop in

TO VNA b2 INPUT

 

 

 

semi-rigid

13

 

3-62112-81a

SMA male-male

MS464xB port labeled:

MN469xC port labeled:

 

P2 source loop out

TO VNA Port 2 Src OUTPUT

 

 

 

semi-rigid

VectorStar ME7838D4/G4 QSG

PN: 10410-00770 Rev. C

QSG-9