The information contained in this document is subject to change
without notice.
Agilent T echnologies makes no war ranty of any kind with r egard to this
material, including but not limited to, the implied warranties of
merchantability and fitness for a partic ular purpose. Agilent
Technologies shall not be liable for errors contained herein or for
incidental or consequential damages in connection with the furnishing,
performance, or use of this material.
Where to Find the Latest Information
Documentation is updated periodically. F or the late st information about
Agilent PSA or ESA spectrum analyzers, including firmware upgrades
and application information, please visit the following Internet URLs:
The analyzer can generate various messages that appear on the display during operation.
There are four types of messages.
This chapter descri b e s th e fo ll owing types of messag es di sp lay ed on the an a lyze r :
Instrument Messages
User Error
Messages
Informational
Messages
Status
Messages
Pop-up
Messages
Beginning on page 34, User Error Messages appear when an attempt has been
made to set a parameter incorrectly or an operation has failed (such as saving
a file). These messages are often generated during remote operation when an
invalid programming command has been entered. These messages appear in
the status line at the b otto m of t he d ispl ay. If yo u a re using the de fault d is pla y
colors, the message will appear in yellow. The message will remain until you
preset the analyzer, press
line.
Beginning on page 82, Informational Messages provide information that
requires intervention. These messages appear on the status line at the bottom
of the display. If you are using the default display colors, the message will
appear in green (ESA) or yellow (PSA). The message will remain until you
preset the analyzer, press
line. The information provided in brackets, for example <filename> or
<directory>, is a variable that represents a specific input provided previously.
Beginning on page 86, Status Messages appear on the right side of the
analyzer display and/or set status bits in the SCPI Status Register system.
These messages indicate a condition that may result in erroneous data being
displayed. Most messages will only be displayed until the error condition is
corrected. Multiple messages can be displayed and will be listed in the display
area. In each case the name of the corresponding status bit is indicated in
parenthesis. It will be noted if only a status bit is used (no message).
These messages indicate a condition that may require intervention. They
appear in the middle of the display in a framed box. The message will remain
until the appropriate intervention has taken place or the condition has been
corrected.
ESC, or another message is displayed on the status
ESC, or another message is displayed on the status
30Chapter 1
Instrument Messages
Instrument Messages Introduction
Error Queues
There are two types of error queues, front panel and remote interface. These two queues
are viewed and managed separately. Refer to Table 1-1. for more information on th e
characteri st i cs of the Error Que u e.
Front panel A summary of the last 11 error messages preceded by an error number
may be viewed in the Error Queue by pressing,
When a remo te interface in i t i a tes activity th a t g e n erates an erro r, the
messages are output to the remote bus. When output to the remote
interface, they are preceded by an error number.
Remote interface
(SCPI) When a user error condition occurs in the analyzer as a result of SCPI
(remote interface) activi ty, i t is reported to both the front panel display
error queue and the SCPI error queue. If it is a result of front panel
activity , it reports to the front panel display error queue, and depending on
the error, may also report to the SCPI error queue.
Table 1-1. Characteristics of the Error Queue
System then Show Errors.
Characteristic
Capacity
(number of errors)
Overflow Handling
Viewing Entries
Clearing the Queue
Front-Panel Displ ay
Error Queue
1130
Circular (rotating).
Drops oldest error as new error comes in.
Power up.
Send a *CLS command.
Read last item in the queue.
a. Error history incl udes the date and time an erro r last occu rred, the erro r number,
the error message, and the number of times the error occurred.
Instrument Messages
Chapter 131
Instrument Messages
Instrument Messages Introduction
Error Message Format
The system–defined error numbers are chosen on an enumerated ( “1 of N”) basis . The er ror
messages are listed in numerical orde r accord ing to the e rro r message number. Status and
Informational messages without numbers will be listed in alphabetical order following the
numerical listing.
In this chapter, an explanation i s in cl u ded with each error to furth e r cla ri fy its meaning.
The last error described in each class (for example, –400, –300, –200, –100) is a “generic”
error. There are also references to the IEEE Standard 488.2-1992, IEEE Standard Codes, Formats, Protocols and Common Commands for Use with ANSI /IEEE Std 488.1-1987.
New York, NY, 1992.
Error messages are displayed at the bottom of the screen in the status line. If you are
using the default display colors, the message will appear in yellow. The error number is
available through the remote interface and the show errors screen; it is not display ed in
the status line.
Figure 1-1. Error Message Example
Instrument Messages
32Chapter 1
Instrument Messages
Instrument Messages Introduction
Error Message Types
Events do not generate more than one type of error. F or example, an event that generates
a query error will not generate a device-specific, execution, or command error.
−499 to −400:
Query Errors
−399 to −300:
Device-Specific
Errors
−299 to −200:
Execution Errors
−199 to −100:
Command Errors
These errors indicate that the analyzer output queue control has detected a
problem with the message exchange protocol described in IEEE 488.2,
Chapter 6. Errors in this class set the query error bit (bit 2) in the event
status register (IEEE 488.2, section 11.5.1). These errors correspond to
message exchange protocol errors described in IEEE 488.2, 6.5. In this case:
• Either an attempt is being made to read data from the output queue when
no output is either present or pending, or
• Data in the output queue has been lost.
An error number in the range –399 to –300 indicates that the analyzer has
detected an error where some device operations did not properly complete,
possibly due to an abnormal hardware or firmware condition. This is not a
error in response to a SCPI query or command, or command execution. The
occurrence of any error in this class will cause the device-specific error bit
(bit 3) in the event status register to be set.
These errors indicate that an error has been detected during analyzer
execution.
These errors indicate that the analyzer parser detected an IEEE 488.2
syntax error. Errors in this class set the command error bit (bit 5) in the
event status register (IEEE 488.2, section 11.5.1). In this case:
• Either an IEEE 488.2 syntax error has been detected by the parser (a
control-to-device message was received that is in violation of the IEEE
488.2 standard. Possible violations include a data element which violates
device listening formats or whose type is unacceptable to the device.), or
Instrument Messages
64 to 799:
Device-Specific
Errors
Greater than
10000:
Measurement
Applications
Errors
• An unrecognized header was received. These include incorrect
device-specific headers and incorrect or unimplemented IEEE 488.2
common commands.
These errors indicate that a device operation did not properly complete,
possibly due to an abnormal hardware or firmware condition. These codes
are also used for self-test response errors. Errors in this class set the
device-specific error bit (bit 3) in the event status register (IEEE 488.2,
section 11.5.1).
The <error_message> string for a positive error is not part of the SCPI
standard. A positive error indicates that the analyzer detected an error
within the GPIB system, within the analyzer firmware or hardware, during
the transfer of block data, or during calibration.
These errors indicate that an error has been detected while executing
measurements requiring a personality option such as the GSM Measurement
Personality or those measurements found under the
key in SA mode.
Chapter 133
MEASURE front-panel
Instrument Messages
Instrument Messages
Instrument Message s
Error Messages
NOTEError numbers are displayed in the error queue, not on the display.
To s ee an error number, press System, Show Errors.
None Hardware Fail
ESA message NOT USED
A hardware failure has occurred. Get in touch with your service center.
0: No Error
0 No error
Instrument Messages
The queue is empty. Either every error in the queue has been read, or the
queue was cleare d b y po w e r-on or *CLS.
−499 to −400: Query Errors
NOTEError numbers are displayed in the error queue, not on the display.
To s ee an error number, press System, Show Errors.
–440 Query UNTERMINATED after indefinite response
Indicates that a query wa s received in the same program message after a
query requesting an indefinite response was executed (see IEEE 488.2,
6.3.7.5).
–430 Query DEADLOCKED
Indicates that a condition causing a DEADLOCKED query error occurred
(see IEEE 488.2, 6.3.1.7). For example, both the input buffer and the
output buffer are full and the analyzer cannot continue. The analyzer
automatically discards output to correct the deadlock.
–420 Query UNTERMINATED
Indicates that a condition causing an UNTERMINATED query error
occurred (see IEEE 488.2, 6.3.2.2). For example, the device was addressed
to talk and an i n co mplete prog ra m message was re c e ived.
–410 Query INTERRUPTED
Indicates that a condition causing an INTERRUPTED query error
34Chapter 1
Instrument Messages
Instrument Messages
occurred (see IEEE 488.2, 6.3.2.7). For example, a query was followed by
DAB or GET before a response was completely sent.
−400 Query Error
This is a generic query error for devices that cannot detect more specific
errors. The code indicates only that a query error as defined in IEEE
488.2, 11.5.1.1.7 and 6.3 has occurred.
−399 to −300: Device-Specific Errors
NOTEError numbers are displayed in the error queue, not on the display.
To s ee an error number, press System, Show Errors.
−340 Calibration failed
The instrument requires an
pressing
System, Alignments, Align All Now.
Align All Now. Restore the alignment by
−330 Self-test failed; EEPROM checksum for <card>
The card identification header for a hardware card is incorrect. If the card
is not properly identified, the instrument is likely to be non-functional.
Report this error to the nearest Ag ilent Technologies sales or service office .
−321 Out of memory
An internal operation needed more memory than was available. Report
this error to the nearest Agilent Technologies sales or service office.
−310 System error; Pretune DAC Calibration failed
System error; Pretune DAC Cal failed midpoint verification
System error
An internal system problem w a s detected. Report this error to the nearest
Agilent Technologies sales or service office.
−300 Device-specific error
This is a generic device-dependent error for devices that cannot detect
more specific errors. The code indicates only that a device-d ependen t er ror
as defined in IEEE 488.2, 11.5.1.1.6 has occurred. Repo rt this error to the
nearest Agilent Technologies sales or service office.
Instrument Messages
−299 to −200: Execution Errors
NOTEError numbers are displayed in the error queue, not on the display.
To s ee an error number, press System, Show Errors.
Chapter 135
Instrument Messages
Instrument Messages
−253 Corrupt media
A removable media was found to be bad or incorrectly formatted. Any
existing data on the media may have been lost.
−250 Mass storage error; EEPROM write timeout on EEPROM
Failure to initialize EEPROM. Report this error to the nearest Agi l ent
Technologies sales or service office.
−241 Hardware missing
Hardware missing; no such SIO address
Missing device hardware. For example, an option is not installed.
−230 Data corrupt or stale
Data corrupt or stale; EEPROM copy of <file>
Data corrupt or stale; NRAM copy of <file>
Possibly invalid data. A new measu r ement was started b u t n o t co m pleted.
Instrument Messages
−224 Illegal parameter value
You have sent a parameter for this command that is not allowed. See the
Function Reference.
−223 Too much data; not all points entered
Too much data; <description of the type of data exceeded>
Indicates that a legal program dat a element of block, expression, or st ring
type was received that cont a i n ed more data tha n th e d e v i ce co u ld handle
due to memory or related device-specific requirements.
−222 Data out of range
A legal program data element was par sed b ut could not be executed
because the interpreted value was outside the legal range defined by the
device (see IEEE 488.2 11.5.1.1.5). The displayed results may be clipped.
−221 Settings conflict;…
…Averaging unavailable when Sig ID is on
User in ext mix with Sig ID on and try to select Averaging
Mode.
…Cannot center unpreselected external mixer
Mixer Type is Unpreselected and user tries Pres elector
Centering.
…Command incompatible with band pair marker
Using remote commands , you have tried to adjust the start
or stop frequency of a span pair marker. You can adjust
only the center and span.
36Chapter 1
Instrument Messages
Instrument Messages
…External mixing unavailable when preamp on
Internal preamp is on and user tries to select external
mixing.
…FFT unavailable when Sig ID is on
User in ext mix with Sig ID on and try to select
Manual:FFT
…Settings Conflict, Preamp unavailable in
external mixing mode
User in ext mix and tries turning on internal preamp.
…Sig ID requires EXTERNAL mixing mode
Input Mixer set ti n g i s INT E RNAL and the use r tr i e s Si g
ID.
…Sig ID unavailable in Manual:FFT
User in ext mix, in FFT, user tries Sig ID.
…Sig ID unavailable when Averaging is on
User in ext mix, in Averaging Mode, user tries Sig ID .
…Sig ID unavailable when Signal Track is on
User in ext mix, in Signal Track Mode, user tries Sig ID.
…Signal ID unavailable with Presel Mixer
Instrument Messages
User in Presel mixer, and tries to turn on Sig ID.
…Signal Track unavailable when Sig ID is on
User in ext mix with Sig ID on and try to select Signal
Track.
Settings conflict; parameter currently disabled
This parameter is grayed out (unavailable) in the current context. Check
the individual parameter help/d ocumentation for more information.
−200 Execution Error
For devices that cannot detect more specific errors, this code indicates only
that an execution error as defined in IEEE 488.2, 11.5.1.1.5 has occurred.
−199 to −100: Command Errors
NOTEError numbers are displayed in the error queue, not on the display.
To s ee an error number, press System, Show Errors.
Chapter 137
Instrument Messages
Instrument Messages
–178 Expression data not allowed
A legal expression data was encountered, but was not allowed by the
device at this point in parsing.
–171 Invalid expression
The expression data element was invalid (see IEEE 488.2, 7.7.7.2). For
example, unmatched parentheses or an illegal character.
–170 Expression data error
This error, as well as errors –171 through –179, is generated when parsing
an expression data element. This particular error message is used if the
device cannot detect a more specific error.
–168 Block data not allowed
A legal block data element was encountered, but not allowed by the device
at this point in the parsing.
–161 Invalid block data
Instrument Messages
A block data element was expected, but w as invalid (s ee IEEE 488.2,
7.7.6.2). For example, an END message was received before the end length
was satisfied.
–160 Block data error
This error, as well as errors –161 through –169, is generated when parsing
a block data element. This particular error message is used if the device
cannot detect a more specific error.
–158 String data not allowed
A string data element was encount ered, but not allowed by the device at
this point in the parsing.
–151 Invalid string data
A string data element was expected, bu t was invalid (see IEEE 488.2,
7.7.5.2). For example, an END message was received before the terminal
quote character.
–150 String data error
This error, as well as errors –151 through –159, is generated when parsing
a string data element. This particul ar error message is used if the device
cannot detect a more specific error.
–148 Character data not allowed
A legal character data element was encountered where prohibited by the
device.
38Chapter 1
–144 Character data too long
The character data element contains mo re than twelv e characters (see
IEEE 488.2, 7.7.1.4).
–141 Invalid character data
Either the character data element contains an invalid character or the
particular element received is not valid for the header.
–140 Character data error
This error, as well as errors –141 through –149, is generated when parsing
a character data element. This particular error message is used if the
device cannot detect a more specific error.
–138 Suffix not allowed
A suffix was encountered aft er a numer ic element which does not allow
suffixes.
–134 Suffix too long
Instrument Messages
Instrument Messages
The suffix contained more than twelve characters (see IEEE 488.2,
7.7.3.4).
–131 Invalid suffix
Attempt to se n d a SC P I command wit h a su ff ix that does not ma tch th e
current units for the function.
The suffix does not follow the syntax described in IEEE 488.2, 7.7.3.2, or
the suffix is inappropriate for this device.
–130 Suffix error
This error, as well as errors –131 through –139, is generated when parsing
a suffix. This particular err or message is used if the devic e can not dete ct a
more specific error.
–128 Numeric data not allowed
A legal numeric data element was rec eived, but the device does not accept
one in this position for the header.
–124 Too many digits
The mantissa of a decimal-numeric data element contained more than 255
digits excluding leading zeros (see IEEE 488.2, 7.7.2.4.1).
Instrument Messages
–123 Exponent too large
The magnitude of an exponent was greater than 32000 (see IEEE 488.2,
7.7.2.4.1).
–121 Invalid character in number
An invalid character for the data type being parsed was encountered. For
example, an alpha in a decimal numeric or a “9” in octal data.
Chapter 139
Instrument Messages
Instrument Messages
−120 Numeric data error; VGR
Numeric data error
This error, and errors −121 through −129, is generated when parsing a
data element that appears to be numeric, including non-decimal numeric
types. This message is used if the device cannot detect a more specific
error.
–114 Header suffix out of range
The value of a header suffix attached to a program mnemonic makes the
header invalid.
–113 Undefined header
The header is syntactically correct, but it is undefined for this specific
device. For example, *XYZ is not defined for any device.
–112 Program mnemonic too long
The header contains more than twelve characters (see IEEE 488.2,
7.6.1.4.1).
Instrument Messages
–111 Header separator error
A character which is not a legal header separator was encountered while
parsing the header.
–110 Command header error
An error was detected in the header. This message is used whe n the device
cannot detect the more specific errors described for errors –111 through
–119.
–109 Missing parameter
Fewer parameters were received than required for the header. For
example, the *ESE common command re quires one pa ram eter, so receiving
*ESE is not allowed.
–108 Parameter not allowed
More parameters were received than e xpected for the head er. F or example ,
the *ESE common command only accepts one parameter, so receiving *ESE 0,1 is not allowed.
–105 GET not allowed
A Group Execute Trigger was received within a program message (see
IEEE 488.2, 7.7). Correct the GPIB controller program so that the GET does
not occur wi th i n a l i n e o f G P IB pr o g ra m code.
–104 Data type error
The parser recognized a data element that is not allowed . For example,
numeric or string data was expe cted, but block data was encountered.
40Chapter 1
–103 Invalid separator
The parser was expecting a separator and encountered an illegal
character. For example, the semicolon was omitted after a program
message unit.
–102 Syntax error
An unrecognized command or data type was encounter ed. F or examp le, a
string was received when the dev i ce does not accept strings .
–101 Invalid character
A syntactic command contains a character which is invalid for that type.
For example , a head er c ontaining an ampers and, suc h as “SETUP&”. This
error might be used in place of error numbers –114, –121, –141, and some
others.
−100 Command error
This is a generic syntax error for devices that cannot detect more specific
errors. The code in di c a te s o nl y t hat a command err o r a s de fi n e d in
IEEE 488.2, 11.5.1.1.4 has occurred.
Instrument Messages
Instrument Messages
64 to 799: Device-Specific Errors
64 System, Alignments, Align All Now, Needed
102 There is too m u ch 50 MHz energy a t th e R F p o rt f o r a li g n ments to run.
Reduce input power and run alignments again.
103 Sampling Oscillator Unlock; Failure acquiring SO frequency
lock
Report this error to the nearest Ag ilent Technologies sales or service office .
104 1st LO Unlock; Failure acquiring FracN LO frequency lock
1st LO Unlock; Failure acquiring single loop FracN LO lock
The first LO on has lost phase l ock. Report this error t o the nearest Agilent
Technologies sales or service office.
105 2nd LO Unlock
The second LO has lost phase lock. Report this error to the nearest Agilent
Technologies sales or service office.
106 4th LO Unlock
The fourth LO has lost phase lock. Report this error to the nearest Agilent
Technologies sales or service office.
Instrument Messages
107 Sample Clock Unlock
Report this error to the nearest Ag ilent Technologies sales or service office .
Chapter 141
Instrument Messages
Instrument Messages
108 Cal Oscillator Unlock
Report this error to the nearest Ag ilent Technologies sales or service office .
129 Meas Uncal
The measurement is uncalibrated, usually due to sweep ing a narrow RBW
filter too quickly. Check the sweep time, span, and bandwidth settings, or
press
Auto Couple.
143 Final IF Overload
Either increase input attenuation or de crease the input signal level.
163 Freq Count: Reduce Span/RBW ratio
The span is too wide for the current resolution bandwidth. Either reduce
the span or increase the RBW.
187 Directory already exists
190 Unable to save file, invalid path
191 File Operation Status; directory <name> created
Instrument Messages
File Operation Status; <name> directory deleted
File Operation Status; <name> file copied
File Operation Status; <name> file deleted
File Operation Status; <name> file loaded
File Operation Status; <name> file saved
File Operation Status; Volume <name> formatted
196 Can’t Auto-Couple Sweep Time in Zero Span
197 Marker Freq Count HW Failure
201 Option not installed
The desired operation cannot be performed because a required option is
not installed. For example, pressing
Source with no tracking generator
installed in the analyzer will generate this error.
202 No peak found
No signal peak was foun d th a t mee t s th e cr i te r ia u n de r
Search Param Criteria.
204 TG Frequency Limit
Peak Search,
The tracking generator has reached the limit of its allowable frequency
range.
205 Command not recognized
Indicates that the command sent from the remote interface was not
recognized. Check the programming gu ide for correct syntax.
42Chapter 1
206 Unable to initialize flatness data
A failure occurred in setting th e flatness data in the internal EEROM. Get
in touch with your local Agilent Technologies sales and service office.
207 Unable to store flatness data
A failure occurred in setting th e flatness data in the internal EEROM. Get
in touch with your local Agilent Technologies sales and service office.
208 Preselector is not available at this frequency.
Instrument Messages
Instrument Messages
Preselection of frequencies below
∼ 3 GHz is not supported. Preselector
centering can be performed only for frequencies above 3.025 GHz.
209 Preselector centering failed
An attempt to center the preselector failed . You tried to center the
preselector with the analyzer span too wide. Preselector centering can be
performed only for spans below 1.123 GHz.
211 RBW limited to 1 kHz when Span > 5 MHz
In spans greater than 5 MHz, narrow (digital) resolution bandwidths,
below 1 kHz, are not available.
213 Span limited to 5 MHz when RBW < 1 kHz
In narrow (digital) resolution bandwidths, below 1 kHz, spans greater
than 5 MHz are not available .
214 TG start freq is less than 9 kHz
Tracking generator uncalibrated below 9 kHz.
215 TG start freq is less than 1/2 res bw
Tracking generator unc alibrated at start frequenc ies below 1/ 2 the current
resolution bandwidth.
Instrument Messages
216 Invalid Baud Rate
Attempt to use invalid baud rate. Refer to the programming language
chapter of Agilent Tec hnologies ESA Series Spectrum Ana lyzer Programmer’s Guide for more information.
217 RS-232 Interface Error
An error occurred on the serial interface.
219 Command not valid in this model
Indicates that t h e command sent from the remote interface does not apply
to this analyzer model number. For example, attempting to center the
preselector in an analyzer without a preselector will generate this error.
220 Trace Menu is unavailable when Sig Ident is ON
Chapter 143
Instrument Messages
Instrument Messages
221 Invalid option, unable to uninstall package
Y ou have attempted to remove a personality that is not currently installed.
Verify command was entered correctly.
222 Command not valid when no measurement is active
Indicates that the command sent fro m the remote interface must be iss ued
while a measurement is running in the analyzer.
223 Trigger Offset unavailable in swept spans
Trigger Offset is only available in Zero Span. Refer to “Trig” in your
Agilent Spectrum Analyzer User’s Guide for a description of this function.
224 Option not licensed.
The selected option requires a license. Refer to the installation procedures
in the user’s guide available for this particular option.
226 Actual Trig Delay = Remote Trig Delay + Remote Trig Offset
228 Preselector centering failed - inadequate peak
Instrument Messages
You tried to center the preselector with a trace peak of insuffic ient level.
You may have pressed
Presel Center with th e m a rker placed be l o w th e
signal peak.
229 3
− 26.5 GHz preselector bypassed, can’t adjust
The microwave preselector has been bypassed. Attempts to press or send
the SCPI commands for
Presel Center or Presel Adjust will result in this
error.
230 Unpreselected external mixing - no preselector
While using external mixing with an unpreselected mixer, attempts to
press or send the SCPI commands for
Presel Center or Presel Adjust will
result in th i s error.
231 Preselector unavailable while in external mixing
The internal preselectors cannot be used while in external mixing.
232 External mixing is off, can’t adjust external preselector
The preselector is unavailable when external mixing is turned of f.
238 Trace smoothing: VBW filt or Average Detector; cannot use both
239 HW Diag is OFF
332 Average Type incompatible for scale.
Amplitude Scale command should be sent prior to the Average Typ e
command.
44Chapter 1
601 PSA only
612 ESA only File not found
The analyzer could not find the specified file.
601 Floppy disk full
The floppy disk is full. Clear some space by deleting unwanted files.
602 PSA only
605 ESA only Media is protected
A save was attempted to a write-protected device.
602 Floppy disk error
An unknown error has occurred while accessing the floppy disk.
603 PSA only
615 ESA only File is corrupt
Corrupted file
The file that you were trying to load is corrupt.
Instrument Messages
Instrument Messages
604 PSA only
607 ESA only File name error
File name error; Directory does not have a default file type
File name error; Directory does not support extenders
File name error; Empty filename
File name error; Illegal extender
File name error; Illegal filename character
File name error; Only one: is allowed
File name error; Only one extender character allowed
An invalid file name was specified. Use filenames with a maximum of 8
characters (letters and digit s only) , and us e a 3-character extension. File
names are not case-sensitive. This error also occurs if you try to delete a
nonexistent file.
615 PSA only
604 ESA only File already exists
You attempted to save to a file that already exis ts . Eithe r delete or renam e
the existing file or select a new name, then try again.
Instrument Messages
602 PSA only
605 ESA only Media is protected
A save was attempted to a write-protected device.
Chapter 145
Instrument Messages
Instrument Messages
605 Data corrupt or stale; EEPROM copy of <file>
WARNING: Contact the factory if you see this error.
This error indicates that a file st ored in EEPROM has beco me cor rupt and
the Hamming codes are unable to repair the data.
When the EEPROM file has been corrupted, the system will store default
values into both the Non-Volatile RAM and EEPROM copies of the file. If
this error occurs often, it may indicate that there is a problem with the
physical memory.
Data corrupt or stale; RAM copy of <file>
This error indicates that a file stored in Non-Volatile RAM has become
corrupt and the Hamming codes are unable to repair the data.
When a Non-Volatile file has this error, the system will first try to verify
the state of the data in EEPROM and if valid it will copy the data in
EEPROM into the Non-Volatile ram copy of the file. When this happens,
the system will remain calibrated. If this error occurs often, it may
indicate that there is a problem with the physical memory.
Instrument Messages
606 Data questionable; EEPROM copy of <file>
EEPROM error occurred. The EEPROM copy of an internal file is either
corrupt or otherwise unusable. The system automatically updates the
non-volatile RAM copy of the EEPROM copy using a default initialization.
The actual EEPROM file is l e ft as it is. Report this err o r to the nearest
Agilent Technologies sales or service office.
606 Media is corrupt
A save was attempted to a corrupt device.
604 PSA only
607 ESA only File name error
File name error; Directory does not have a default file type
File name error; Directory does not support extenders
File name error; Empty filename
File name error; Illegal extender
File name error; Illegal filename character
File name error; Only one: is allowed
File name error; Only one extender character allowed
An invalid file name was specified. Use filenames with a maximum of 8
characters (letters and digit s only) , and us e a 3-character extension. File
names are not case-sensitive. This error also occurs if you try to delete a
nonexistent file.
46Chapter 1
609 Restoration of NVRAM data
This error indicates that a file stored in Non-Volatile RAM has been
corrected based on the Hamming codes stored with the file. After the
correction, the file will co ntain t he same data as wh en the instr ument w a s
shut off. If this error continues to occur on succes sive power cycles, i t may
indicate that there is a problem with the physical memory.
627 PSA only
609 ESA only Media is not writable
A save was attempted to a read-only device or device that could not be
written to. Try a known-good disk.
ERROR/NUMBER “609 Restoration of NVRAM data” IS ALREADY
USED FOR HAPPENING ./storage/directory/NRAM_Directory.cc
HAP_nramDataCorrected
610 File access is denied
The file is protected or hidden and cannot be accessed.
611 File access IO busy;
duplicate :SERVice[:PRODuction]:CALibrate:BEGin
Instrument Messages
Instrument Messages
File access IO busy; :SERVice[:PRODuction]:CALibrate:END
without a BEGin
601 PSA only
612 ESA only File not found
The analyzer could not find the specified file.
613 Flash memory is full
The internal flash memory is full. Clear some space by deleting unwanted
files. If your analyzer has a serial number less than US41440000 or
MY41440000 and Option B72 has not been installed, you may also
increase the flash memory size by purchasing Option B72.
763 PSA only
614 ESA only Bad, missing, or unformatted disk
Bad or missing floppy disk
The floppy disk is not inserted properly, is not formatted, or the directory
cannot be read. Insert a known good disk and try again.
615 PSA only
604 ESA only File already exists
You attempted to save to a file that already exis ts . Eithe r delete or renam e
the existing file or select a new name, then try again.
Instrument Messages
Chapter 147
Instrument Messages
Instrument Messages
603 PSA only
615 ESA only File is corrupt
Corrupted file
The file that you were trying to load is corrupt.
617 Wrong density floppy inserted
The floppy disk has the wrong densit y. It sho u ld be 1.44 MB.
618 Illegal write access of Flash memory
Attempt to write to an unavailable area of internal flash memory.
619 Can’t Auto-Couple RBW in Zero Span
You sent a remote command to set the RBW into auto while in zero span.
(Remote interface only.)
620 Can’t Auto-Couple Sweep Time in Zero Span
You sent a remote command to set the sweep time to auto while in zero
span. (Remote interface only.)
Instrument Messages
622 External reference missing or out of range
The external frequency reference signal is missing, has too low an
amplitude, or does not match the fr equency value that you previously
entered into instrument memory.
623 Printer not responding
Printer not responding; ioctl ERROR
Check the printer. It may not be connected properly or turned on.
625 Printer out of paper
Load paper in pr i n te r.
626 Print failed; Can’t open print file desc
Print failed; DisplayExport illegal print language
Print failed; DisplayExport insufficient RAM space
Print failed; DisplayExport undetermined error
Print failed; Error transferring image to printer
Print failed; Printer ioctl failed
Print failed; Unknown printer response
627 PSA only
609 ESA only Media is not writable
A save was attempted to a read-only device or device that could not be
written to. Try a known-good disk.
48Chapter 1
Instrument Messages
Instrument Messages
651 Connect RF OUT to INPUT
Attempt to align the tracking generator without its output connected.
Connect the tracking generator RF OUT to the analyzer
INPUT.
652 Connect Amptd Ref Output to Input
For Agilent Technologies E4402B, E4403B, E4404B, E4405B, E4407B, and
E4408B only: you must connect the
INPUT with the appropriate cable.
AMPTD REF OUTPUT to the analyzer
653 Auto Align not available when using Calibration Defaults
The Auto Alignment system cannot be used until an Align Now All is
executed by pressing
System, Alignments, Align Now, All. On all Agilent
Technologies ESA spectrum analyzer models except Agilent Technologies
E4401B and E4411B, you must connect the
AMPTD REF OUT to the INPUT
with the appropriate cable to perform this alignment. For Agilent
Technologies E4401B and E4411B only: di sconnect any signals from the
INPUT prior to performing this procedure.
701 Invalid printer response
In attempting to identify the printer an invalid response was receiv ed.
Check that you are using a supported printer. Be sure you are using the
proper cable and that it is securely fastened.
702 Unsupported printer
A printer which is recognized, but known to be unsupported was
identified. This printer cannot be used with the analyzer. For example, a
printer only supported by Microsoft Wi ndows will generate this error.
703 Unknown printer
In attempting to identify the pri nter, a valid response w as re ceived but t he
printer is not known to the analyzer. Use the
Print Setup to configure the printer.
704 Printer interface error
An error occurred while trying to p rint. Mak e sure the pri nter is t urned on
and properly connected.
705 Printer Type is None
The current printer typ e is set to
Change the type in the
Print Setup menu and try again.
None, so no print opera tions ar e possible.
727 In <filename>: [DATA] header missing
Instrument Messages
Custom printer menu under
This message indicates that the d ata section of a fi le did not begin with t he
token [
DATA].
Chapter 149
Instrument Messages
Instrument Messages
728 In <filename>, line <nnn>: separator missing
The [
HEADER] section of a file contains entries requiring an equal (=) s ign,
such as <keyword> = <value >. This mess age appears if the equal sign does
not appear on th e li n e.
729 In <filename>: error reading file
Appears when loading data from a limit line or cor rections disk file and a
failure to the file occurs.
730 In <filename>, line <numeric_value>: line too long
When loading data from a limit line or corrections disk file, this message
will appear if the length of any line in the file exceeds 255 characters.
731 In <command>: bad data count (<numeric_value>): expected
multiple of <numeric_value>
This message indicates that the da ta sent to a correc tions or limit table vi a
DAT A or MERGE commands does not have the expected length for the
the
table. F or example , this message would appear if an attempt were made to
merge 7 numeric values into a limit table , since each logical e ntry require s
3 values (frequency, amplitude, and connected).
Instrument Messages
732 In <filename>, line <numeric_value>: error parsing tokens
This message may appear when loading data from a limit line or
corrections disk file . It indica tes a problem in the attempt to break a string
of text into toke n s. Ther e m ay b e too fe w to ke n s i n th e st ri n g. In oth e r
words, the file content must match the expected format. This typically
happens when there are too few numeric values in the [
DATA] section of a
limit or corre c ti o n s fi l e.
733 In <filename>, line <numeric_value>: <xxx> is not numeric
This message may appear when loading data from a limit line or
corrections disk file. It indicates that a non-numeric token <xxx> was
found where a numeric token was expected. In other words, the file
content must match the expe ct e d fo r m a t .
734 Interpolation error: cannot compute log of
<negative_frequency_value>
Occurs when the frequency interpolation of a limit line is set to log and the
start frequency of the instrument is negative. The
<negative_frequency_value> is limited to
– 80 MHz, so it may not match the frequency that caused the error.
735 In <filename>: bad amplitude unit <unit>
This message indicates that unit <unit> is not recognized or supported.
50Chapter 1
736 Too many data values at <freq_or_time_value>
This message may appear when data is sent to a corrections or limit table
using the
DATA or MERGE commands. These tables limit the number of
amplitudes associated with a frequency or time to 2 or less. This message
will appear if an attempt is made to attach 3 or more values to a frequency
or time.
751 Instrument state set to initial values
Instrument state may be corrupt, state has been reset to
initial values
While trying to load a trace or state , the state information was found to be
in error. This may be because the state had been stored on a later revision
of analyzer firmware. A default set of state variables was loaded instead.
There is nothing wrong with the analyzer.
An error in the internal instrume nt state has b een detec ted. The st ate has
been reset to a default value.
752 Unable to load state from file
Instrument Messages
Instrument Messages
Attempt to load a state from a file failed.
753 Unable to save state to file
Attempt to save a state to a file failed. See the associated error messages
for the cause (press
System, Show Errors).
For PSA: If this error occur s af te r a mo de switch that take s lo n g e r th e n
normal, you may need to free up some instrument memory. When you
switch to a different measurement mode, the analyzer must temporarily
store instrument state data. If you have less then 500 kB of available
memory, the switching process can take over 1 minute and result in this
error. Use the File key menu to delete any unnecessary files from
instrument memory.
755 Unable to load state from register
Attempt to load a state from an internal state register failed.
756 Unable to save state to register
Attempt to save a state to an in te r nal state register failed. Se e th e
associated error messages for the cause (press
System, Show Errors).
757 Unable to load user state, factory preset was done
A user preset failed, so the factory preset values were used. Save a valid
state into user preset and try again.
Instrument Messages
758 Unable to save user state
Attempt to save a user preset state failed. See the associated error
messages fo r th e ca u se (press
Chapter 151
System, Show Err ors).
Instrument Messages
Instrument Messages
759 Unable to load state saved from firmware Rev A.xx.xx; Restore
Sys\nDefaults
Unable to load state
A saved state fil e fr om a newer firmware revision was a tt e m pted to be
loaded into an older instrument.
760 ESA only Unable to query state
Query of state over the remote interface w a s unsuccessful.
761 ESA only Unable to set state
Attempt to set the state over the remote interface was unsuccessful.
762 ESA only Incorrect filename, allowable extensions are .trc or .csv
Attempt to save a trace to a file with an incorrect extension.
762 ESA only Unable to load file
A failure occurred while loading a file; the file was not loaded.
Instrument Messages
763 PSA only
614 ESA only Bad, missing, or unformatted disk
Bad or missing floppy disk
The floppy disk is not inserted properly, is not formatted, or the directory
cannot be read. Insert a known good disk and try again.
763 ESA only Incorrect filename, allowable extensions are .gif or .wmf
Attempt to save a screen image to a file with an incorrect extension.
764 Unable to save file
Attempt to sav e a fi l e f a i le d ; th e file was not saved . Se e th e associated
error messa g e s fo r th e cause (pres s
System, Show Err ors).
765 Unable to load file
Attempt to load a file failed; the file was not loaded.
766 Unable to format drive
767 Failed to Initialize ISTATE regions. Fatal LDS error
Attempt to initialize the instrument state has failed. Cycle instrument
power. If this fails to correct the problem, contact your nearest Agilent
Technologies service center.
768 Unable to load user state, required instrument mode is not
installed
Attempt to load a state failed, because the state was saved with a
measurement personality that is not currently loaded. Load the
appropriate personality and try again.
52Chapter 1
Instrument Messages
Instrument Messages
769 ESA only Invalid instrument mode
You have attempted to switch to an instrument mode that is currently not
installed. Confirm that the mode name (for INST:SEL) or number (for
INST:NSEL) was entered correctly and that the requested personality is
actually installed in the instrument.
770 Incorrect filename, allowable extensions are .gif or .wmf
You sent a remote command to save a screen file bu t did not specif y a valid
extension.
Incorrect filename, allowable extension is .sta
You sent a remote command to save a state file but did not specify a valid
extension.
Incorrect filename, allowable extensions are .trc or .csv
You sent a remote command to save a trace file but did not specify a valid
extension.
770 Instrument mode requested is not supported
Instrument mode specified with: INST command is not valid. Refer to
Chapter 5, “Instrument Subsystem” of Agilent Technologies ESA Series Spectrum Analyzer Programmer’s Guide for more information.
771 Store Ref trace before turning on Normalize
A reference trace must be available for the Normalize function to be
activated. Refer to “View/Trace” in the Agilent ESA Spectrum Analyzer
User’s Guide where the
Normalize key function is explained in detail.
772 Cannot load a directory
Cannot load a directory, please choose a file
You have selected a directory instead of a file when attempting to pe rform
the Load function under the
File front-panel key.
780 No Peak Found
781 Video Trigger cannot be active with Average Detector
You sent a remote command to do one of the following:
• Tur n on video trigger while the Average Detector or a Marker Function
is active.
• Tur n on the A verage Detector or a Marker Function while V ideo Trigger
is active.
Instrument Messages
782 Video Trigger cannot be used with Mkr Func due to Average
Detector
783 Allowable span for current center frequency exceeded
784 Allowable CF for current span exceeded
Chapter 153
Instrument Messages
Instrument Messages
787 WBIF Self Test Failed
Variable resistor 122 performs a loop-bac k self -test on power up . It does an
ADC image calibration, then stores a sine wave in RAM and plays it out
the DAC. It then measures the carrier power, image rejection and
signal-to-noise ratio. If this test fails, the wideband IF pc assemblies are
either incorrectly installed or have a hardware failure. Call your Agilent
Service Representative.
There is diagnostic information available for the CE. The self test can be
run with DIAG:WBIF:TEST? This comm a n d re t u rns 4 values: pa ss ed,
carrier power, image rejection and signal-to-noise ratio.
First look at the carrier power. If it is below –50, the LVDS cable is
probably missing or faulty. If the carrier power is above 0 dBm, image
rejection less than 40 but signal-to-noise ratio is more than 50, there is
probably a bad ADC.
788 WBIF ADC Image Calibration Failed
The Opt 122 card uses two ADCs that must be kept in perfect
synchronization to avoid unw anted images appearing in the spec trum. The
instrument will occasionally perform an ADC Image Calibration (formerly
known as Step Cal) to re-synchronize the ADCs . This calibration is only
known to fail if there is a hardware problem.
Instrument Messages
789 WBIF Firmware Reset
Opt 122 has a local microprocessor controlling the hardware trigger. If
this microp r ocessor reset s, it wi ll a u to m a tically resta rt and begin taki n g
data again, posting this notice to the us er. While this should never happen,
it is a self-recovering error that requires no action on the part of the user.
If, however, it starts to occur frequently, contact your Agilent servic e
representative to find out if there is a firmware upgrade to solve this
problem.
791 Acquiring Data...
795 No trigger before timeout; auto-triggered
797 Awaiting Trigger, no AUTO Trig
801 Memory limit caused Data Acquisition to be truncated
805 doing Alignment...
Greater than 10000: Measurement Applications Errors
NOTEError numbers are displayed in the error queue, not on the display.
To s ee an error number, press System, Show Errors.
54Chapter 1
Instrument Messages
Instrument Messages
10008 to 10517: Power Suite Error Messages An error number in this ra nge indic ates
the instrument has detected an error relating to the Power Suite functionality.
10008 Preferred resolution bandwidth not available.
The calculated required resolution bandwidth for this measurem ent is not
available.
10010 One or more harmonics past freq limit: number decreased.
Highest harmonic was past the frequency limit of the analyzer, so the
number of measured harmonics was decreased.
10011 First harmonic is past analyzer frequency limit.
First harmonic was past the frequency limit, so no harmonics can
be measured. The analyzer will simply measure the fundamental
. If
you have the Agilent E4407B analyzer, Option AYZ allows the use of
HP/Agilent 11970 Series, and HP/Agilent 11974 exte rnal mixers to extend
the frequency range to 110 GHz. Operation to 325 GHz is also possible
with non-HP/Agilent mixers.
10012 No Fundamental > 0 Hz found in given span.
There were no frequencies greater than 0 Hz in the starting
span, so the centre frequency at measurement entry will be used
as the fundamental
.
10013 No Fundamental > -50 dBm found in given span.
There was no fundamental f ound, so the cent re freq uency at measuremen t
entry will be used as the fundamental.
10020 Signal Tracking is not available when Noise Correction is on.
If signal tracking i s turned on while noi se correctio n is on , it wi ll be tu rned
off. To use the signal tracking function, first turn nois e correction off.
10138 Valid GSM burst not found in frame (Burst Type)
No active GSM b u rsts that match th e selected burst type have been
detected in the RF input sig n a l . Search was perfo rmed over a comp l e te
GSM frame.
10139 Valid GSM burst not found in specified timeslot (Burst Type)
No active GSM b u rsts that match th e selected burst type have been
detected in the RF input signal. Search was performed over the specified
timeslot.
Instrument Messages
10140 Valid GSM burst not found in frame (Ref Burst)
No active GSM b u rsts that match th e selected burst type have been
detected in the RF input signal. Search was performed over the specified
complete G S M frame using the Ref Burst type setting.
Chapter 155
Instrument Messages
Instrument Messages
10141 Sync word not found in frame (Burst Type)
One or more active GSM bursts that match the selected burst type have
been detected in the RF input signal, but none contain the selected
Training Sequence Code (TSC). The search was performed over the
complete G SM frame.
10142 Sync word not found in specified timeslot (Burst Type)
One or more active GSM bursts that match the selected burst type have
been detected in the RF input signal, but none contain the selected
Training Sequence Code (TSC). The search was only performed over the
specified timeslot setting.
10143 Sync word not found in frame (Ref Burst)
One or more active GSM bursts that match the selected burst type have
been detected in the RF input signal, but none contain the selected
Training Sequence Code (TSC). The search was only performed using the
Reference Burst type and the Ref TSC settings over the complete GSM
frame.
The RF Communications Hardware (Option B7E) and Digital Signal Fast
ADC (Option B7D) cards required to perform the demodulation are not
present in the analyzer.
10147 Opt B7D bootrom requires upgrade.
The (Option B7D) bootrom revision is not supported by the currently
loaded personality version. Refer to the user’s guide for the personality in
use.
10148 Opt BAH DSP algorithm files not installed. Meas unavailable.
Contact your Service Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are not present in the analyzer.
10148 B7D DSP Code Fail. Reload Opt BAH. Contact your Service
Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are not present in the IDSP: drive.
The Digital Signal Processing alg orithm code fi le re vision is no t su pporte d
by the currently loaded personality version. Refer to the user’s guide for
the person a lity in use for mo re i n formation on in s tallation/u p g r ade.
The Digital Signal Processor algorithm coefficient file revision is not
supported by the currently loaded personality version. Refer to the user ’s
guide for the personality in use.
The Digital Signal Processor algorithm files required to perform the
demodulation are corrupt and cannot be loaded properly. Reinstall
measurement personality.
10151 B7D DSP Code Fail. Reload Opt BAH. Contact your Service
Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are present in the IDSP: drive and cannot be loaded
properly into DSP RAM. Reinstall measurement personality.
10152 Lost trigger, aborting measurement.
The selected trigger source w as present at the start of the measurement,
but timed out before the measurement completed.
10153 DSP algorithm timeout, aborting measurement
The Digital Signal Processor demodulation algorithm timed-out for an
unknown reason. This message normally indicates a problem with the
modulated signal.
10154 Measurement not defined for Out of Band.
You have changed to an out-of-band frequency range. The band
measurement only operates in the selected band.
10155 No Fast ADC hardware installed. Meas unavailable.
The analyzer cannot use sweep times of less than 5 msec when Option B7D
or Option AYX is not installed. Therefore, the measurement will not be
executed.
10156 No Fast ADC hardware installed.
The analyzer cannot use sweep times of less than 5 msec when Option
B7D or Option AYX is not installed resulting in all the radio st andards not
being supported.
Instrument Messages
10156 The calibration data is invalid, and has been cleared.
10157 Tracking Generator hardware is not present. Meas unavailable.
The measurement requires a built-in tracking generator.
Chapter 157
Instrument Messages
Instrument Messages
10158 This operation requires a measurement to be active.
10159 Entire trace is below threshold level
The measurement cannot operate properly because the trace has
completely fallen below the threshold level. Change the threshold level to
view trace.
10160 Valid burst not found
Cannot find a valid burst, either because it is not fully on the display, or
the level is too low to be detected properly.
10160 Upper Custom Mask is Invalid!
The user-spec ifi ed upp er custo m mask cannot be re solve d into a limit line .
The format may be bad.
10161 No Fast ADC hardware installed.
Cannot use sweeptimes less than 5 ms, due to the lack of a fadc, therefore
the meas will be severely restricted.
Instrument Messages
10161 Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be resolved into a limit line.
The format may be bad.
10162 Resolution BW <300 kHz.
This error message is a warning that the resolution bandwidth has been
set below 300 kHz. The test results will not meet GSM specifications.
10163 Cannot find the Power vs Time Limits File.
The limit line definition f ile for the GSM standards has been deleted. This
message is displayed while the
Measure key is grayed out. Reinstall the
GSM measurement personality.
10164 Band Measurement not defined for Out of Band.
You are attempting to monitor the band but have set the frequency outside
the band. Reset the band for the particular standard for which you are
testing or use the channel setting whic h does not re quire a fr equenc y to be
Meas Setup, Method (Channel)).
set. (
10165 Carrier Not Present.
10166 Cannot update the list of cable types.
The cable file may have been moved or deleted accident ally. Reinst all the
measurement personality. This message is applicable to the GSM (Option
BAH) and Distance to Fault (Option 226) measurement personality
options.
58Chapter 1
Instrument Messages
Instrument Messages
10166 Cannot locate the file that contains the list of cable types.
The cable file may have been moved or deleted accidentally. Reinstall the
measurement personality. This message is applicable to the GSM (Option
BAH) and Distance to Fault (Option 226) measurement personality
options.
10167 The file containing the list of cable types has been updated.
10168 Cannot update the list of cable types on drive C:
The file update failed.
10170 The Cable Fault Measurement is active. Mode Setup is disabled.
Mode setup is not available in the cable fault utility.
10172 Sweep Time too fast (<2 sec)
The sweep time must be set to 2 seconds or longer fo r the results to be
valid.
10173 Carrier Not Present.
10177 There are no spurs to inspect.
You have attempted to switch the
Inspect Spur menu key to the On position
after the measurement has finished, but found no spurs.
10178 Carrier Not Present.
10179 Carrier Present. Test Stopped!
A carrier was found in the transm i t band. Either disable the carrier or
insert a bandpass filter for th e re c eive bandwid th .
10180 Gate option not installed. Results may not be accurate.
This measurement method requires the use of the time-gate (op tion 1D6)
in order to gate the spectrum during the 50-90% part of the burst. If the
gate option is not installed, the measurement will still run although this
warning will be displayed.
10181 Upper Custom Mask is Invalid!
The user-spec ifi ed upp er custo m mask cannot be re solve d into a limit line .
The format may be bad.
10182 Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be resolved into a limit line.
The format may be bad.
Instrument Messages
10183 Resolution BW
The test will not meet EDGE specifications bec ause the res olution BW has
fallen below 300 kHz.
Chapter 159
< 300kHz.
Instrument Messages
Instrument Messages
10184 Cannot find the Power vs Time Limits File.
Cannot find the limit line definition file for the EDGE standards. This
message is displayed while the
Measure key is grayed out.
10185 Gate option not installed. Results may not be accurate.
10186 Measurement does not support the current radio standard.
The measurement you have chosen is currently greyed out. Sel ect a radio
standard which is supported or configure this measurement manually
Mode Setup, Radio Std. None).
(
10186 EDGE EVM only supports EDGE TCH burst type (Burst Type)
10187 Radio standard is not supported by the current measurement.
The standard you have chosen is currently greyed out. Select a radio
standard which is supported by the current m easurement or configure t his
measurement manually (
Mode Setup, Radio Std. None).
10187 EDGE ORFS only supports EDGE TCH burst type (Burst Type)
Instrument Messages
10188 EDGE PVT only supports EDGE TCH burst type (Burst Type)
10189 EDGE PVT only supports EDGE TCH burst type (Ref Burst)
10190 Select a valid Radio Standard/Measurement combination to
enable.
The Enable All Measurements menu key was pressed when it was
disabled. To enable select a radio standard that is supported by
the current measurement or a measurement that supports the
current radio standard.
10190 EDGE ORFS only supports EDGE TCH burst type (Ref Burst)
10191 Mask unavailable for current Span - increase to display mask.
The current span setting is either narrower than the mask width
or so wide that there are too few display points to allow the
mask to be drawn. Increase or decrease the span to display the
mask.
10191 EDGE EVM only supports EDGE TCH burst type (Ref Burst)
10192 Sweep Point/Span ratio below minimum - results may be
inaccurate.
The sweep point to span ratio is below the minimum required to
ensure the bucket ratio is large enough to test DVB-T masks
10218 Hardkeys are disabled.
Some of the forms (for example Receiver Spurious in GSM) do not allow
you to close the form without either formally accepting or canc elling the
form settings. For this reason, all of the hardkeys are disabled until you
terminat e th e fo rm.
60Chapter 1
Instrument Messages
Instrument Messages
10219 Awaiting trigger
The measurement requires a trigger to be present. If the trigger does not
occur or is delayed, this message will be displayed. Check your trigger
settings.
10227 Measurement suspended until carrier is turned off.
The receive channel power and the receive spur measurements are
specified with the attenuation se t to 0 dB. To prevent overload, the
frequency spectrum of interest is moni tored for signal levels which exceed
a specified threshold before setting the attenuator to 0 dB. If a carrier is
found, this message is displayed and the completion of the measurement
will not occur until the carrier is removed. The carrier check may be
turned off using the pro perties form under the fr ont-pane l
Mode Setup key.
You m ay al so ch a nge the signal threshold which de te rmines a carri e r on
the properties form.
10228 Cannot correlate to input signal.
This error is normally ge nerated because of one of the follo wing reasons : 1.
There is no carrier signal.
2. Walsh channels other than the pil ot are active.
3. There is some other modulation problem that will prevent the
measurement from being made.
This problem must be corrected before the measurement can continue.
10229 The regression portion failed.
This message occurs when (Option B7D) is not functioning properly.
Demodulation measurements (modulation accuracy and code domain)
might fail as a result of this error.
10230 DSP timed out, resetting DSP.
Digital Signal Processor was unable to finish the selected measurement
within the given period of time. Restart the me asurement.
10231 Measurement failed for unknown reasons.
Check instrument settings and re start measurement.
10232 RF Signal not found.
This message is generated if there is no signal at the center frequency that
is greater than 10 dB above the displayed average noise level.
10233 Level is low, results may degrade.
The signal being measured is of low power. The results may not be as
accurate as they would be if the signal level was higher.
10237 RF Board LO Unlocked. Contact service center.
This message occurs if the local oscillator on the (Option B7E) is in an
unlocked state. This indicates broken hardw are.
Instrument Messages
Chapter 161
Instrument Messages
Instrument Messages
10238 RF Board SR Osc Unlocked. Contact service center.
This message occurs if the sample rate (SR) oscillator on Option B7E is in
an unlocked state. This indicates broken hardware.
10239 Opt Freq Ref setting does not match external reference.
This message is generated if Source is set to External on the properties
form under the front-panel
Mode Setup key and the frequency on the s ame
form is set to a frequency that does not match the frequency of the signal
being used as the external reference.
10240 RF Board RF Osc Unlocked. Contact service center.
This message occurs if the reference osc illator on the (Opti on B7E) is in an
unlocked state. This indicates broken hardware.
10241 RF Board could not detect any bursts in signal.
This message is generated when the trigger is set to RF Burst and (Option
B7E) cannot detect a burst.
10245 Error reading file: SPCLIMIT.CSV. Cannot use custom limits.
Instrument Messages
The file could be missing or corrupt . Create a new limits f ile . Alternati vely,
the actual limits defined in the file might not allow the measure ment to be
executed. Redefine the limits or use the default limits. Restart the
The file is missing or corrupt. Please reinstall the cdmaOne perso n ality.
10247 Error reading file: CDPDMDA. Please reinstall cdmaOne.
The file is missing or corrupt. Please reinstall the cdmaOne perso n ality.
10248 Error reading file: CDPPMCO. Please reinstall cdmaOne.
The file is missing or corrupt. Please reinstall the cdmaOne perso n ality.
10249 Error reading file: CDPPMDA. Please reinstall cdmaOne.
The file is missing or corrupt. Please reinstall the cdmaOne perso n ality.
10250 Error reading file: RHODMDA. Please reinstall cdmaOne.
The file is missing or corrupt. Please reinstall the cdmaOne perso n ality.
10251 Error reading file: RHOPMCO. Please reinstall cdmaOne.
The file is missing or corrupt. Please reinstall the cdmaOne perso n ality.
10256 Error reading file: OOBSTAB.CSV. Use Edit Table
This error is generated when you try to load a table (using the
key on page 2 of the edit table form menu) before a table has been saved.
You must first save a table using the
table using the
62Chapter 1
Load Table key.
Save Table key before trying to load a
| Save Table.
Load Table
Instrument Messages
Instrument Messages
10259 Table could not be saved.
This message occurs if the C: drive is full or corrupt. Check the amount of
space left on the drive.
10260 Table could not be loaded.
When trying to load a table, the previous table has been somehow
corrupted. Use the
Save Table key to save a valid table. Then edit the valid
table, save it, and try to load it again.
10264 Emission bandwidth not found. Consider increasing span.
This error is normally generated when attempti ng occupied bandwidth
measurements. The “X dB” value you entered (
Meas Setup, X dB) to
calculate the emission bandwidth is the difference between the highest
point on the trace and the point “X dB” down on either side of the
maximum. If the actual difference is less than the value entered, the
emission bandwidth cannot be computed. Some responses to this situation
are as follows:
1. Connect a signal to the input. (If there is no signal present, the
difference between the trace minimum and maximum will generally be
less than “X dB”.)
2. Increase the span. (If the signal is wide, the shoulders of the signal
might not be present on the screen, and again, the diff erence between
the trace minimum and maximum will be less than “X dB”.)
3. Center the signal. (There must be a point on the trace that is “X dB”
down from the maximum on both sides of that maximum.)
10286 Burst not found.
Either the signal being analyzed has insufficient power, the rising or
falling edges cannot be detected, or the burst is less than
126 microseconds.
10287 Valid Bluetooth burst not found. (Check Packet Type)
The burst that has been found does not correspond to the currently
selected Bl u et o oth packet type (t h e b u rs t l ength may be too s h o rt ).
10288 Option B7D or AYX FADC hardware required. Meas unavailable.
The measurement you are attempting requires either the DSP and Fast
ADC (Option B7D) or the Fast ADC (Option AYX) card to perform the
demodulation, but neither are present in the analyzer.
You have selected (by remote SCPI command) preamble sync with no
Bluetooth FM demodulation card present (Option 106) .
10321 Start Marker must be at least 1% < Stop Marker.
You have attempted to input a start marker value that will result in the
difference between the start and s top markers being less than 1%.
10322 Stop Marker must be at least 1% > Start Marker.
Instrument Messages
You have attempted to input a stop marker value that will result in the
difference between the start and s top markers being less than 1%.
10323 Unable to Calculate Result using Current Setup.
Y ou have changed the setup parameters such that the marker lines used to
measure the power cannot be displayed therefore accurate measurements
cannot be made.
10340 ’10101010’ pattern not detected - results may be inaccurate.
This message is displayed if the measurement cannot detect the
‘10101010’ pattern in the payload. The measurement will continue and
carry out the calculations on the payload data supplied, but may not be
correct.
10350 Payload data pattern ‘10101010’ not present.
This message is displayed when the “Payload Data” parameter is set to
Auto and the measurement has not detected either of the required
patterns in the payload.
10351 Required payload data pattern ‘10101010’ not present.
This message is displayed after successfully measuring and holding the
‘11110000’ pattern if the measurement is restarted and the ‘10101010’ data
pattern is not detected.
10352 Required payload data pattern ‘11110000’ not present.
This message is displayed after successfully measuring and holding the
‘10101010’ pattern if the measurement is restarted and the ‘11110000’ data
pattern is not detected.
64Chapter 1
10353 There is no valid result to hold.
Instrument Messages
Instrument Messages
You have attempted to hold either
∆ f1 or ∆f2 before it has been measured.
10360 Can’t compute result - not enough transitions.
This message is displayed when the measurement cannot f ind either a 111,
000, 101 or 010 pattern and is there fore unable to calc ulate the low or high
frequency deviations.
10389 Valid EDGE burst not found in frame (Burst Type)
No active EDGE bursts that match the select ed burst type have been
detected in the RF input signal. Search was performed over the complete
EDGE frame.
10390 Valid EDGE burst not found in specified timeslot (Burst Type)
No active EDGE bursts that match the select ed burst type have been
detected in the RF input signal. Search was performed over the specified
timeslot.
10391 Valid EDGE burst not found in frame (Ref Burst)
No active EDGE bursts that match the select ed burst type have been
detected in the RF input signal. Search was performed over the specified
complete ED G E fr a m e us i n g th e Re f B u rs t setting.
10392 B7D DSP Code Fail. Reload Opt. 202. Contact your Service
Center
The DSP algorithm files required to perform the demodulation are not
present in the instrument ISDP: drive.
10393 Opt 252 DSP algorithm code file requires upgrade.
10394 Opt 252 DSP algorithm coef. file requires upgrade.
10395 B7D DSP Code Fail. Reload Opt. 202. Contact your Service
Center.
The DSP algorithm files required to perfor m the demodul ation are present
in the instrument ISDP: drive, but cannot be loaded pr operly into DSP
RAM.
10396 EDGE Upgrade (252) unlicensed. Please contact your Agilent
Sales Rep.
10400 Valid signal not found.
This error is normally generated because of one of the following reasons:
1. There is no carrier signal.
2. The carrier signal power has changed during the measurement, causing
ADC to over/under range.
3. There is some other modulation problem that will prevent the
measurement from being made . This prob lem must be corre cted before the
measurem e nt can contin ue.
Instrument Messages
Chapter 165
Instrument Messages
Instrument Messages
10401 Valid burst not found in frame (Burst Type).
No active bursts that match the Burst Type have been detected in the RF
input signal. The search was performed over the complete frame.
10402 Valid burst not found in specified timeslot (Burst Type).
No active bursts that match the Burst Type have been detected in the RF
input signal. The search was only performed over the specified timeslot
setting.
10403 Sync word not found in frame (TSC).
One or more active EDGE bursts that match the Normal Burst Type have
been detected in th e RF Inp u t si g n a l , b u t n one co n tain the selected
Training Sequence Code (TSC). The search was performed over the
complete E DG E fr ame.
10404 Sync word not found in specified timeslot (TSC).
One or more active EDGE bursts that match the Normal Burst Type have
been detected in th e RF Inp u t si g n a l , b u t n one co n tain the selected
Training Sequence Code (TSC). The search was only performed over the
specified timeslot setting.
Instrument Messages
10409 Wideband calibration not valid until Align Now, All performed.
Wideband calibration corrections cannot be applied to current
demodulation results. Results will not meet specified demodulation
performance levels until an Align Now, All has been performed.
Wideband calibration corrections cannot be applied to the current
demodulation results. Results will not meet specified demodulation
performance levels until
Wideband Cal has been selected and successf ully
performed.
10411 Measurement uncalibrated, symbol rate exceeds +/-10% nominal.
Wideband calibration corrections cannot be applied to current
demodulation results, due to the
Symbol Rate setting exceeding +/-10% of
the nominal value. Results may not meet specified demodulation
performance levels.
10412 Cal Data corrupt. Wideband cal failed. Using previous data.
Wideband calibration procedure failed for an unknown reason. New
calibratio n da ta was not saved. Previo u s w i de b a n d ca l i b ra tion data is
being used, if available. However, it is not acceptable. Before re-attempting
Wideband Cal, en sure connect i o n between the
INPUT 50 Ω is in place. If the error persists , g et in touch with your service
AMPTD REF OUT and the
center.
66Chapter 1
Instrument Messages
Instrument Messages
10413 B7D DSP Code Install Fail Opt.229. Contact your Service
Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are not present in the analyzer. Reinstall the measurement
personality and contact your service center.
10414 Opt 229 DSP algorithm code file requires upgrade.
The Digital Signal Processing alg orithm code fi le re vision is no t su pporte d
by the currently loaded personality version. Refer to the user’s guide for
the person a lity in use for mo re i n formation on in s tallation/u p g r ade.
10415 Opt 229 DSP algorithm coef. file requires upgrade.
The Digital Signal Processor algorithm coefficient file revision is not
supported by the currently loaded personality version. Refer to the user ’s
guide for the personality in use for more information on
installation/upgrade.
10416 B7D DSP Code Reload Fail Opt.229. Contact your Service Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are corrupt and cannot be loaded properly. Reinstall the
measurement personality and contact your service center.
10417 This operation requires a measurement to be active.
The analyzer cannot perform this operation, because it requires a
measurement to be running and no measurement is running.
10418 Demod Format setting does not allow this value for
Points/Symbol.
Invalid Points/Symbol selection attempted for the given Demod Format
setting.
10419 Carrier Not Present.
A carrier signal/burst is expected at t he analyzer input . This signal cannot
be found; however, the measurement will still run. Check input signal
connection.
10420 Valid burst not found.
This error only occurs when
RF Amptd is selected for Burst Sync. It
indicates that a valid burst envelope which meets the specified Burst
Search Threshold w as not found withi n the spe cified Burs t Searc h Length.
Note that the demodulation measurement will proceed by attempting to
demodulate the signal without Burs t Sync alignm ent. P o ssible caus es may
be:
Instrument Messages
1. Carrier signal is not actually bursted.
2. Burst Search Threshold and/or Burst Search Length may need to be
adjusted.
Chapter 167
Instrument Messages
Instrument Messages
10421 Cannot lock to carrier.
This error only occurs when attempting to demodulate OFFSET QPSK
signals. It indicates that the demodulation algorithm is unable to lock to
your signal. Possible causes may be:
1. There is no carrier signal.
2. Carrier sign a l is pre se n t, but Demod set ti n gs do not match the
modulation for m a t of the O F FSET QPSK signal be i ng me a s ured.
3. OFFSET QPSK signal is too noisy to achieve carrier lock.
10425 B7D DSP Code Install Fail Opt.231. Contact your Service
Center.
The Digital Signal Processor algorithm files required to
for the 89600 VSA SW are not present in the
analyzer. Reinstall the measurement personality and
contact your service center.
The Digital Signal Processing algorithm code file
revision is not supported by the currently loaded
personality version. Refer to the user’s guide for the
personality in use for more information on
installation/upgrade.
The Digital Signal Processor algorithm coefficient file
revision is not supported by the currently loaded
personality version. Refer to the user’s guide for the
personality in use for more information on
installation/upgrade.
10428 B7D DSP Code Reload Fail. Opt.231. Contact your Service
Center.
The Digital Signal Processor algorithm files required for
the 89600 VSA SW are corrupt and cannot be
loaded properly. Reinstall the measurement personality
and contact your service center.
68Chapter 1
10450 Avg Mode changed to Repeat for Full Meas Type.
Meas Type has been changed to Full. Exponential Av erage Mode is not
available for Full Meas Type therefore Average Mode has been changed to
Repeat.
10451 Avg Mode changed to Exp for Examine Meas Type.
Meas Type has been changed to Examine. Repeat Average Mode is not
available for Examine Meas Type therefore Average Mode has been
changed to Expo n e n ti a l .
10452 Meas Type changed to Examine for Exp Avg Mode.
Average Mode has been changed to Exponential. Full Meas Type is not
available for Exponential Average Mode therefore Meas Type has been
changed to Exa m i n e.
10453 Meas Type changed to Full for Repeat Avg Mode.
Average Mode has been changed to Repeat. Examine Meas Type is not
available for Repeat A verage Mode therefore Me as Type has been changed
to Full.
Instrument Messages
Instrument Messages
10454 Valid burst not found.
Cannot find a valid burst, either because it is not fully on the display, or
the level is too low to be detected properly.
Using the current instrument settings, the measurement cannot locate
four signals with enough energy or resolution to perform an accurate
measurement.
10510 Signals to not fit expected Intermod (TOI) pattern.
Based on the current analyzer settings, the measurement cannot find two
signals of high enough amplitude that can be resolved in such a manner
that the third order products would lie within the measured span.
10511 100 spurs found. Additional spurs ignored.
There are too many spurs for the table (the limit is 100), and any
additional spurs that are found will be ignored
10512 No spurs have been found.
You have started a measurement in examine meas type, in single or
continual sweep mode, or full meas type in single sweep mode, but no
spurs were found.
Instrument Messages
10513 No ranges are defined.
There are no active ranges in the range table. The user will need to
activate at least one range.
Chapter 169
Instrument Messages
Instrument Messages
10514 Avg Mode changed to Repeat for Full Meas Type
Meas Type has been changed to Full. Exponential Av erage Mode is not
available for Full Meas Type therefore Average Mode has been changed to
Repeat.
10515 Avg Mode changed to Exp for Examine Meas Type
Meas Type has been changed to Examine. Repeat Average Mode is not
available for Examine Meas Type therefore Average Mode has been
changed to Expo n e n ti a l .
10516 Meas Type changed to Examine for Exp Avg Mode
Average Mode has been changed to Exponential. Full Meas Type is not
available for Exponential Average Mode therefore Meas Type has been
changed to Exa m i n e.
10517 Meas Type changed to Full for Repeat Avg Mode
Average Mode has been changed to Repeat. Examine Meas Type is not
available for Repeat A verage Mode therefore Me as Type has been changed
to Full.
Instrument Messages
10524 to 10561: Phase Noise Error Messages (Option 226) An error number in this
range indicates the instrument has detected an error relating to the Phase Noise
personality.
10524 This measurement does not support the *.CSV file format.
You cannot load or save base instrument traces , as this is not suppo rted by
the Log Plot measurement.
10525 Invalid format: CSV files are not supported by this
measurement.
You cannot load or save base instrument traces , as this is not suppo rted by
the Log Plot measurement.
10525 Use View/Trace menu when loading or saving logarithmic traces.
You cannot load CSV format trace files or s ave t rac es in t he CS V format i n
the Log Plot measurement of the Phase Noise Measurement Personality.
10526 Cannot display Ref Trace because it has no data.
Reference trace cannot be displayed , as there is currently no dat a assigned
to it - use the functions under the St ore Ref Trace menu, or load a trace to
assign some data.
10527 Invalid Marker Trace. Marker moved to Raw Trace.
Cannot place markers on the reference trace, because the reference trace
is currently turned off or has no data.
70Chapter 1
Instrument Messages
Instrument Messages
10529 Failed to Load trace. Bad file format.
The load trace operation could not be completed, as the input file was not
in the expected format. You can only load traces that were previously
saved using the ’Save Trace’ feature.
10530 Trace file contains no compatible traces.
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace fo rmat that is incompatible with
the version you are running. Please check you are running the most up to
date version of the personality.
10531 Trace file was created by incompatible version of Opt. 226
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace fo rmat that is incompatible with
the version you are running. Please check you are running the most up to
date version of the personality.
10532 Cannot open trace file for writing. Save Failed.
Cannot write the trace file to the destination filename. This could be
because the disk is full, or possibly due to a filename error. If using a
floppy disk, check there is a formatted disk in the drive.
10533 Cannot save Ref Trace because it contains no data.
Check the Ref Trace is turned on, and contains so me valid trace data.
10534 Trace file saved successfully.
The trace saving operation was successful.
10535 Carrier Not Present
No Carrier > -50 dBm found at the analyzer in put within the search span.
Solution: Modify center frequency to be closer to actual carrier, or
alternatively apply carrier of sufficient amplitude at the current center
frequency.
10535 Cancellation trace is not set to Reference or has no data.
When performing phase noise cancellation, you need to supply a reference
trace that w il l be used to cance l o u t th e background noise of the a n a lyzer.
The reference trace must be in Refe rence (View) mode, and selected by the
Ref Trace parameter under the Cancellation menu.
10536 Cancellation disabled while measuring DANL Floor.
Phase Noise cancellation does not make sense when measuring the DANL
Floor, so for this reason it has been disabled.
Instrument Messages
Chapter 171
Instrument Messages
Instrument Messages
10537 Cancellation trace has different X-Scale to Smoothed trace.
When performing phase noise cancellation, you need to supply a reference
trace that w il l be used to cance l o u t th e background noise of the a n a lyzer.
The reference trace must have been measured over the same range of
offsets, and on the sa me ins trument as the current measurement being
performed.
10560 Carrier Not Present. Verify frequency and amplitude settings.
No Carrier > –50 dBm found at the analyzer input within the search span.
Solution: Modify center frequency to be closer to actual carrier, or
alternatively apply carrier of sufficient amplitude at the current center
frequency.
10561 Signal Tracking disabled when measuring DANL Floor (Removal).
Measuring the DANL Floor with DANL Method set to Removal requires
that the user remove the signal and attach a load to the analyzer RF
Input. Signal Tracking requires that the carrier be measured many times
per measurement, so the two are inherently incompatible.
Instrument Messages
10601 to 10700: GSM and EDGE Erro r Messages (Options BAH and 202) An error
number in this range indicates the instrument has detected an error relating to the GSM
or EDGE personality.
10602 Sync word not found
An attempt to position a GSM trace into the mask, when not enough data
was present. Try using the
Restart key to clear the problem. This can be
caused by a bad GSM burst, or the RF Sync Delay set too far.
10604 GSM burst out of limits
The GSM signal did not fit into the mask in the Power vs. Time
A GSM-like burst was acquired, but its timing is not valid. Ensure the
correct
Burst Type has been selected.
10614 Valid GSM burst not found
In a GSM measurement, data was acquired but a GSM burst was not
found, with the timeslot mode disabled.
10616 Cannot synchronize frame trigger
Cannot synchronize the frame trigger to the even second clock.
72Chapter 1
Instrument Messages
Instrument Messages
10618 Dynamic range not optimum
10620 Cannot synchronize to RF amplitude (burst error)
10622 GSM RF sync delay is out of range
Change RF Sync Delay.
10624 Sync word not found
In a GSM measurement using demodulation, the training sequence code
(sync word) could not be found.
10626 Signal too noisy
In a GSM measurement, indicates that a burst could not be found in a
signal that appears noisy.
10628 Sync is RF Ampl (not Training Seq). Bits not accurate
10630 Marker X value not adjusted when right of Rise&Fall mark
10632 Incorrect trigger holdoff
10634 Break freq > FFT filter edge
10636 SCPI marker query not available in GSM Rise&Fall
10638 GSM Pwr Meas requires trig delay <
−50 µs. Delay set to −50 µs
10640 Carrier frequency outside device’s transmit band
The entered channel/carrier frequen cy is not within the range of your
current mode setup select ion of standard and device.
10642 ADC overload -- Carrier not at expected frequency
The carrier frequency of the signal may not match the instruments
channel frequency setting.
10644 Requested timeslot number not present
The selected timeslot is not on. (Timeslot is referenced to the trigger
point.)
10646 Tx Band Spur measurement not defined for mobiles
Only base station testing is available.
10648 Carrier power too low for optimum dynamic range
For better dynamic range, transmit band spur measurements require
>10 dBm signal power at the RF input port.
Instrument Messages
10650 Unexpected carrier frequency (BMT only)
The transmit band spur measurement only all ows bottom (B), middle (M),
and top (T) channel frequencies for eac h sup ported frequency band. The
carrier frequency must be se t t o the bot tom, middl e or top f reque ncy of t he
current frequency band.
Chapter 173
Instrument Messages
Instrument Messages
10652 EVM Measurement only supports EDGE TCH burst type
10654 Unable to demodulate signal
10655 Input overload
Excessive input power has been detected which will cause the ADC to clip
the signal. Red u ce the signal level, change the att en u a t or setting (unde r
Input menu), or press Restart if the RF Input Range is Auto.
10656 Tx Band Spur measurement does not support this frequency band
The transmit band spur measurement does not support all of the
commercially available freque ncy bands . You need to change your s election
under
Mode Setup, Radio, Band to one of the supported bands.
10657 Pretrigger too long
10658 Valid GSM burst not found for selected timeslot
In a GSM measurement, data was acquired but a GSM burst was not
found, with the timeslot mode enabled.
Instrument Messages
10701 to 10800: W-CDMA Error Messages (Option BAF) An error number in this
range indicates the instrument has detected an error relating to the W-CDMA personality.
10702 Signal too noisy
10704 Input power too low
The instrument only identifies a channel as active if it meets the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be changed using the Active Set Threshold
function in the Meas Setup menu.
10768 Cannot correlate to input signal
Cannot correlate to the input signal and no active channel is found. (from
the composite EVM measurement) An active channel must meet the
default threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be chang ed using the active set threshold
function in the Meas Setup menu.
10772 Burst not found.
Either the signal being analyzed has insufficient power, the rising or
falling edges cannot be detected, or the burst is less than
126 microseconds.
10776 Cannot sync DPCCH pilot.
This error message appears when an in Slot power measurement
(Chip Power) is selected and the DPCCH sync is not done
correctly
74Chapter 1
.
Instrument Messages
Instrument Messages
10778 Valid PRACH burst not found.
This error message appears when an PRACH power measurement
(
Waveform/Chip Power) is selected and a valid PRACH burst is not
found.
10801 to 10900: cdma2000 Error Messages (Option B78) An error number in this
range indicates the instr ument has det ect ed an error re lating t o the cdma2000 per sona lity.
10824 Signal too noisy
10826 Input power too low
The instrument only identifies a channel as active if it meets the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be changed using the Active Set Threshold
function in the Meas Setup menu.
10868 Can not get long code phase (RS-232)
For MS (mobile station) measurements, the long code phase information
could not be obtained f rom the signal a t the RS-2 32 port (fr om code domain
power measurement or composite EVM measurement).
10872 Cannot correlate to input signal
Cannot correlate to the input signal and no active channel is found. (from
composite EVM measurement) An active channel must meet the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be chang ed using the active set threshold
function in the Meas Setup menu.
10901 to 10925: NADC Error Messages (Option BAE) An error number in this range
indicates the instrument has detected an error relating to the NADC personality.
10902 Sync word not found
In an EVM measurement, the sync word is not found and the
synchronization cannot be established when
Burst Sync menu.
Sync Word is selected in the
10904 Valid NADC burst not found
A valid NADC burst is not found when the
Device is MS.
10906 Signal too noisy
The valid EVM measurement cannot be performed, because the input
signal is too noisy.
Instrument Messages
10908 Burst Delay out of limit for EVM (2 ms)
In an EVM measurement, the
Chapter 175
Burst Delay value must be less than 2 ms.
Instrument Messages
Instrument Messages
10926 to 10950: PDC Error Messages (Option BAE) An error num ber in this range
indicates th e in strument has detected an error relating to the PDC pers o n al i ty.
10926 Sync word not found
In an EVM measurement, the sync word is not found and the
synchronization cannot be established when
Burst Sync menu.
Sync Word is selected in the
10928 Valid PDC burst not found
A valid PDC burst is not found when the Device is MS.
10930 Signal too noisy
The valid EVM measurement cannot be performed, because the input
signal is too noisy.
10932 Burst Delay out of limit for EVM (2 ms)
Instrument Messages
In an EVM measurement, the
Burst Delay value must be less than 2 ms.
10951 to 10975: cdmaOne Error Messages (Option BAC) An error number in this
range indica t e s th e instrument has detected an e rr o r re l a ting to the cdma O n e pe rsonality.
10952 Signal exceeds maximum allowable power
10954 Input overload
Excessive input power has been detected which will cause the ADC to clip
the signal. Reduce the si gnal leve l, c hange t he attenu ator /max to tal power
setting (under
Input menu), or press Restart if the RF Input Range is Auto.
10956 Channel center frequency outside device’s transmit band.
10960 No power at carrier frequency
No power was detected as a CW or a modulated signal.
10962 Cannot correlate to input signal
A correlation failure with the pilot CDMA channel occu rred during
synchronous demodulation.
10976 to 11000: Phase Noise Error Messages (Option 226) An error number in this
range indicates the instrument has detected an error relating to the Phase Noise
personality.
10976 CMD ERR, <name of offending command>
When a command or command argument is not recogn ized by the p arser a
command err o r i s gen erated.
76Chapter 1
10977 PARAMETER ERROR, <name of offending parameter>
When the “GRAT” or “ANNOT” 8566 series commands are used with
arguments other than ON, OFF, 1 or 0.
10978 INVALID ITEM, <offending item>
The invalid item message appears when a value that is not part of the
commands list of enumerated arguments is entered.
10979 INVALID SAVE REG, <invalid register>
When the RCLS or RC commands are used to recall a state register that
wasn’t previously saved with the SAVES or SV commands.
10980 INVALID TRACE, <trace>
When the argument to a t race command is not p art of t he pre defined trace
enumeration (TRA, TRB or TRC).
10981 CMD NOT SUPPORTED, <command>
Command not supported error s are generated when a valid 8566 c ommand
that is not supported by Chameleon is executed.
Instrument Messages
Instrument Messages
10982 LOG FILE FULL, CLEAR FOR FURTHER ERROR LOGGING
10983 LOG FILE CLEARED
12000 to 12999: Noise Figure Error Messages (Option 219) An error number in this
range indicates the instrument has detected an error relating to the Noise Figure
personality.
12116 Two entries already exist at frequency.
When specifying limit line point, if there are more than two poi nts at the
same frequency, the new limit point is discarded and the existing limit is
used.
12219 Command not valid in this mode.
Indicates that the command s ent from the Remote inte rface does not appl y
for this DLP.
12303 User cal invalidated
The existing user cal has been invalidated because of one of the following
reasons:
• Meas mode changed
• Freq outside cal range
• Fixed LO changed
• S ideband changed
Instrument Messages
Chapter 177
Instrument Messages
Instrument Messages
12304 Alignment Failed
Calibration failed because an error was encountered at some point during
the calibration (for example gain < 0)
12305 Mode Setup Error
A mode setup error occurred because of one of the following:
• S ystem input frequency out of range
• External LO frequency out of range
• Stop freq must be less than fixed LO freq
• Start freq must be greater than start IF freq
• LO – Stop freq must be >= min system input freq
• Start freq must be greater than fixed LO freq
• Stop IF freq must be less than fixed LO freq
• Start – LO freq must be >= min system input freq
• Stop freq must be less than stop RF freq
• Start freq must be greater than start RF freq
12307 Required Input attenuation not calibrated
Instrument Messages
Corrected measurements have been requested and the required RF
front-end setting of x dB has not been calibrated.
12308 Invalid frequency list for measurement mode
A frequency within the frequency list cannot be used to make a
measurement in the current mode.
12309 No entries in frequency list
A measurement was attempted wit h List frequency mode or a SCPI query
of the frequency list table was made and the frequency list table is empty.
12310 No entries in ENR table
A measurement was attempted o r a SCPI query of an ENR table w as made
and there were no entries in the relevant ENR table (Common, Meas or
Cal).
12311 No entries in Limit Line table
An attempt is made to either display or test against a limit line table
which has no entries.
12312 RF re-range required: Meas Restarted.
During a continuous measurement, a change of RF front-end attenuation
was required. To do this the measurement needs to be restarted.
12313 IF over range req. RF re-range: Meas Restarted.
During a continuous measurement, an IF section over range condition
occurred, requiring a change of RF front-end attenuation. To do this, the
measurem e n t n eeds to be resta r ted.
78Chapter 1
12314 No entries in loss table
A measurement is attempted or a SCPI query of a before or after los s table
is made and there are no entries in the relevant loss table
12316 T hot must be greater than T cold
A spot T hot has been specified which is not greater than T cold.
12603 Illegal MSDOS name given
An invalid filename has been specified. Use filenames with a maxi mum of
8 characters (letters and digits only) and use a 3 character extens ion. Note
that lowercase and uppercase are perce ived as the same.
12604 File already exists
Attempt to store a file that already exists. Delete or rename the old file
and try again
12605 Media is protected
A store was attempted to a write-protected device.
Instrument Messages
Instrument Messages
12606 Media is not writable
A store was attempted to a read-only device.
12612 File does not exist
The file you were trying to recall could not be found.
12762 Incorrect filename, allowable extension CSV
Attempt to load/store table data with an incorrect extension.
12768 Failed to load ENR data
A problem occurred when trying to load or store table data.
12769 Failed to store ENR data
A problem occurred when trying to load or store table data.
12770 Incorrect filename, allowable extension ENR
Attempt to load/store table data with an incorrect extension.
12771 Failed to load Frequency List
A problem occurred when trying to load or store table data.
12772 Failed to store Frequency List
Instrument Messages
Attempt to load/store table data with an incorrect extension.
12773 Incorrect filename, allowable extension LST
Attempt to load/store table data with an incorrect extension.
12774 Failed to load Limit line
Chapter 179
Instrument Messages
Instrument Messages
Attempt to load/store table data with an incorrect extension.
12775 Failed to store Limit line
Attempt to load/store table data with an incorrect extension.
Attempt to load/store table data with an incorrect extension.
12778 Failed to store Trace
A problem occurred when trying to load or store table data.
12779 Failed to load Loss data
A problem occurred when trying to load or store table data.
12780 Failed to store Loss data
A problem occurred when trying to load or store table data.
12781 Incorrect filename, allowable extension LOS
Attempt to load/store table data with an incorrect extension.
Instrument Messages
12995
Noise Figure DLP requires Noise Source Control Hardware
You have attempted to initiate the Noise Figure mode with the Noise Source
Control hardware installed in your instrument. You must install the Noise
Source Control hardware before running the Noise Figure mode.
13001 to 13100: 1xEV -DO Error Messages (Option 204) An error numb e r i n th i s
range indicates the instrument has detected an error relating to the 1xEV personality.
13024 Signal too noisy
13026 Input power too low
The instrument only identifies a channel as active if it meets the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be changed using the Active Set Threshold
function in the Meas Setup menu.
13070 Cannot correlate to input signal
Cannot correlate to the input signal and no active channel is found. (from
composite EVM measurement) An active channel must meet the defaul t
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be chang ed using the active set threshold
function in the Meas Setup menu.
13074 Valid 1xEV burst not found
In a 1xEV measurement, data was acquired but a 1xEV burst was not
found, with the timeslot mode disabled.
80Chapter 1
Instrument Messages
Instrument Messages
13101 to 13200: Flexible Digital Modulation Analysis Error Messages (Option
241) An error number in this range indicates the instrument has detected an error
relating to the Flexible Digital Modulation Analysis perso nality.
13102 Valid signal not found
The input signal cannot be demodulated becaus e of inappropriate
parameter settings or incorrect signal
13104 Valid burst not found
The burst signal cannot be detected because of inap propriate parameter
settings or incorrect signal.
13106 Sync word not found
The sync word cannot be detected because of inappropriate parameter
settings or incorrect signal.
13108 Points/Symbol changed to an even number because of demod format
Offset QPSK allows only even values for points/symbol. The user changed
the demod format to OQPSK while the points/symbol was set to an odd
value. Therefore, the points/symbol was re-set to an even number.
13110 Sync Pattern includes non-hex characters
The sync pattern should only consist of hex c haracters. Invali d characters
are converted to 0.
13112 Sync Word Length changed to new upp er limit
The upper limit of Sync word bits is 128. If the code word length (sync
word length) exceeds this value, then the sync word length is cut off. This
warning may happen only when the demod format is changed .
13114 Sync offset value not valid
The sync offset value is too far off from a value that would be reasonable to
measure valid portion of data. So the measurement gives up trying to
apply the offset value.
13116 Search Length is shortened because of high IF Band Width
The upper limit of the search length is 50 ms when the IF bandwidth is
> 6.7 MHz, and it is 100 ms when IF BW is < 6.7 MHz.
Instrument Messages
Instrument Messages Without Numbers
Error Messages Without Numbers
Error Attempt to change signal data failed.
EMC message, NOT USED
Chapter 181
Instrument Messages
Instrument Messages
Error Awaiting valid signal.
This indicates that software enhanced triggering is ON and 5 successive
failures have been detected within a demod measurement
Error Signal frequency malformed. Signal not added.
Error Unable to retrieve signal (remote).
EMC message NOT USED
Error Unable to uninstall personality, file not deletable.
This message occurs when you try to delete a personality which has been
marked as non-deletable. The personality is marked non-de letable at the
factory. Get in touch with your nearest service center for further
assistance.
Informational Messages
These messages simply provide information; you are not required to do anything.
Information in brackets (such as <directoryname>), is a previously-provided input.
Instrument Messages
Informational messages appear at the bottom of the screen in the status line.
Informational <directoryname> directory deleted
The directory indicated has been successfully deleted.
Informational <directoryname> directory deleted
The directory indicated has been successfully deleted.
Informational <name1> <directoryname1> directory renamed to <name2>
<directoryname2>
Directory name1 has been successfully renamed to directory name2.
Informational <filename> file copied
The filename indicated has been successfully copied.
Informational <filename> file deleted
The filename indicated has been successfully deleted.
Informational <filename> file loaded
The filename indicated has been successfully loaded.
Informational <filename> file saved
The filename indicated has been successfully saved.
Informational <filename> too many data entries
This message may appear when loading data from a limit line or ampcor
disk file. The [DATA] section of such a file can contain at mo st 200 lines of
data. This message is displayed if that limit is exceed ed.
82Chapter 1
Instrument Messages
Instrument Messages
Informational <name1> <filename1> file renamed to <name2> <filename2>
File name1 has been successfully renamed to file name2.
Informational Atten auto set to 15 dB
Indicates that an input signal has been detected which is of sufficient level
to damage the input circuitry and the input attenuator has been
automatically set to 15 dB. If the signal level is reduced, the attenuator
will stay at 15 dB. This overload protection occurs at an input power level
of 13 dBm (68 dBmV for Option 1DP)
± 7 dB when the input attenuation is
auto coupled and set to <15 dB. To return to the original measurement
setup, reduce the input signal level and press
Attenuation (Auto).
Amplitude. Then press
Overload protection is only available in the Agilent Technologies E4401B
and E4411B.
Informational Auto ranging...
Displayed during autoranging.
Informational B7D and/or B7E not found. Code Domain not available.
Digital Signal Processing and Fast Analog to Digital Converter (Option
B7D) and/or RF Communications Hardware (Option B7E) are not install ed
options on your analyzer. Code domain is therefore not available .
Informational B7D and/or B7E not found. Mod Acc not available.
Digital Signal Processing and Fast Analog to Digital Converter (B7D)
and/or RF Communications Hardware (B7E) are not installed options on
your analyzer. Modulation accuracy is therefore not available.
Informational Carrier Not Present.
A carrier signal/burst is expected at t he analyzer input . This signal cannot
be found; however, the measurement will still run. Check input signal
connection.
Informational Channel frequency outside device’s transmit band.
Reset channel number or frequency.
Informational Default spur table values loaded.
No spur table has been previously saved when the Out-of-Band Spurious
measurement begins. Press
ranges of interest and press
Meas Setup, Edit Table to enter the frequency
Save Table to save that information. This
saved table will be loaded the next time the measurement is run.
Instrument Messages
Chapter 183
Instrument Messages
Instrument Messages
Informational Device = Mobile. Code Domain not available.
Code Domain measurement is grayed out when the device is set to mobile
under the
Mode Setup front-panel key. Code Domain measurement is only
accessible w h en th e device is set to ba se and RF Commu n i cations
Hardware (Option B7E) or Enhanced Memory Upgrade (Option B72) are
installed.
Informational Directory already exists
Each directory and file must have a unique name. The directory name you
have entered is currently being used on the selected drive. You may either
enter a new name or rename the directory currently existent.
Informational Device = Mobile. Mod Acc not available.
Modulation accuracy measurement is grayed out when the device is set to
mobile under
Mode Setup, front-panel key. Modulation accuracy is only
accessible w h en th e device is set to ba se and RF Commu n i cations
Hardware (Option B7E) or Enhanced Memory Upgrade (Option B72) are
installed.
Instrument Messages
Informational Directory already exists
Each directory and file must have a unique name. The directory name you
have entered is currently being used on the selected drive. You may either
enter a new name or rename the directory currently existent. Refer to “File
Menu Functions” in the Agilent ESA Spectrum Analyzer User’s Guide.
Informational ENR table will be extrapolated
One or more calibration or measurement frequency po ints exceeds the
currently loaded Cal or Meas ENR Table frequency ranges. The
corresponding ENR tables lowest frequency ENR value will be re-used for
frequencies less than the table range, and the highest freque ncy ENR
value will be re-used for frequencies greater than the table range.
Informational Entire trace is below the threshold level.
The measurement cannot operate properly because the trace has fallen
completely below the threshold level. Change the threshold level to view
signal.
Informational Invalid format: CSV files are not supported by this
measurement.
When utilizing the Phase Noise Personality: Option 226, the *.CSV file
format is not available.
Informational Invalid EDGE burst timing.
The burst has manually been positioned off the display.
84Chapter 1
Instrument Messages
Instrument Messages
Informational Measurement halted. Press a measurement key to continue.
This error occurs after you choose
Cancel to refrain from setting the
attenuator to 0 dB during the Receive Channel Power and Receive Spur
measurements.
Informational No Fast ADC hardware installed. Meas unavailable.
The analyzer cannot use sweep times of less than 5 milliseconds when
Option B7D or Option AYX is not installed. Therefore the measurement
will not be executed.
Informational Not enough frequency range to measure harmonics for channel.
Selected harmonics are above the frequency range of the analyzer.
THIS MESSAGE IS NO LONGER USED FOR FW A.08.00 AND ABOVE.
Informational One or more harmonics beyond freq limit: number decreased.
The highest harmonic was beyond the frequency limit of the analyzer, so
the number of measured harmonics was decreased.
Informational Option activated
This message is displayed after entering the selected option’s License Key.
Informational Please set RF input range (INPUT menu) to manual first.
In order to manually set the reference level and/or the attenuat i on under
the
Amplitude front-panel key, the RF Input Range menu under the Input
front-panel key must be set to
Man (manual).
Informational Shutdown in process.
Instrument Messages
The analyzer is responding to the
Standby key selection, and is executing
the shutdown procedure.
Informational Table loaded successfully.
When the
Load Table key was pressed on the second page of the edit table
form while in the out-of-band spurious measur ement, the f ile was present.
The information has been loaded into the measurement where it may be
edited again. This message will also appear when the out-of-band spurious
measurement is opened if a spur table has been previously saved.
Informational Table saved successfully.
This message appears aft e r yo u press the
Save Table key on the second
page of the edit table for the out-of-band spurious measurement. It
indicates that the current spur table has been written successfully to disk
and is available to be loaded by means of the
Load Table key.
Informational The calibration data is invalid, and has been cleared.
A parameter has changed that affects calibration. Therefore the
calibration data has been reset, and for best results recalibration is
recommended.
Chapter 185
Instrument Messages
Instrument Messages
Informational The file containing the list of cable types has been updated.
The file update was successful.
Informational This measurement does not support the *.CSV file format.
When utilizing the Phase Noise Personality: Option 226, *.CSV (comma
separated values) file format is no t available.
Informational This operation requires a measurement to be active.
The analyzer cannot perform this operation, because it requires a
measurement to be running and no measurement is running.
Informational Tracking Peak Needed.
This message is displayed when there has been a change in Resolution
Bandwidth, Span, or Alignment since the previous Tracking Peak.
The message does not apply to the E4401B or E4411B Spectrum Analyzer.
Informational Use trigger delay to align bit p0 center on marker.
Tells you where to manually align P0.
Instrument Messages
Informational WARNING: You are about to delete all of the contents of
directory xxxxxx “x:\xxxxx\”. Press Delete Now again to
proceed or any other key to abort.
If you select a directory or subdirectory to delete, this popup message is
displayed when you press
Delete Now. (xxxxxx “x:\xxxxx\” in the message
is the full path and directory name).
Informational
Waiting for end of data acquisition to access floppy disk
Informational Volume <name> formatted
The indicated disk has been successfully formatted.
Status Messages
These messages indicate conditi ons that c an caus e the dis pl ay o f incorr ect data. The name
of the corresponding status bi t a ppears in parenthesis. Some messages disp lay only the
status bit (as noted).
Status messages appear in the upper right portion of the screen grid.
Status * (Invalid Data)
Data on the screen may not match the sc reen annotation. For example,
while analyzer settings are changing or when any trace is in view mode.
Status 50 MHz Osc Unlevel (50 MHz Osc Unleveled)
The internal 50 MHz amplitude reference source has become unleveled.
This condition must be corrected befo re a valid alignment can be
performed.
86Chapter 1
Status (ADC Align Failure) status bit only, no message
The alignment routine was unable to align the analog-to-digital converter
(ADC).
Status Align Now All Needed (Align Needed)
Instrument Messages
Instrument Messages
The instrument requires complete alignment. Press
Align Now, All. On all Agilent Technologies ESA spectrum analyzer models
except Agilent Technologies E4401B and E4411B, you must connect the
AMPTD REF OUT to the INPUT with the a p p ropriate cab l e to perform th is
alignment. For Agilent Technologies E4401B and E4411B only: disconnect
any signals from the
message recurs, load defaults (
perform
Alignment Now, All.
INPUT prior to performing this procedure. If this
System, Alignments, Load Defaults) and then
Status Align Now RF Needed (Align Now RF Needed)
The instrument requires RF alignment. Press
, RF (EXT Cable). On all Agilent Technologies ESA spectrum analyzer
Now
System, Alignments, Align
models except Agilent Technologies E4401B and E4411B, you must
connect the
AMPTD REF OUT to the INPUT with the appropriate cable to
perform this alignment. For Agilent Technologies E4401B and E4411B
only: disconnect any signals from the
INPUT prior to pe rforming thi s
procedure.
Status Align RF Skipped (Align RF Skipped)
The RF alignment has been skipped because a 50 MHz signal w as dete cted
at the INPUT; al ignment wil l r esume when the 50 MHz signal is re moved.
The alignment will not work when there is too much input power at
50 MHz. The instrument may not continue to measure properly. To remove
the message, remove the 50 MHz input signal, then perform an
RF. Press System, Alignments, Align Now, RF. On all Agilent Technologies
ESA spectrum analyzer models except E4401B and E4411B , you must
connect the
AMPTD REF OUT to the INPUT with the appropriate cable to
perform this alignment. For Agilent Technologies E4401B and E4411B
only: disconnect any signals from the
INPUT prior to pe rforming thi s
procedure.
System, Alignments,
Instrument Messages
Align Now,
If this message occurs and you are going to make a measurement near
50 MHz, select
System, Alignments, Auto Align, and All but RF.
Status Align RF Skipped, Align signal amplitude too low
The RF alignment has been skipped because the calibration output power
is too low.
Chapter 187
Instrument Messages
Instrument Messages
Status DC Coupled
Indicates the input of the analyze r i s DC coupled (
). This setting is necessary when measuring frequencies below 100
(DC)
Input/Output, Coupling
kHz on E4402B with Option UKB, E4404B, and E4405B analyzers. For
E4407B analyzers with Option UKB, you must set the coupling to DC
when measuring below 10 MHz. Take care to limit the input level to 0 Vdc
and +30 dBm whenever you are in DC coupled mode.
Status Demod ON: reduce span for audible detection
When the Demod function is active and the speaker is turned on, the ratio
of the resolution bandwidth to span must be gre ater than 0.002 to properly
demodulate and listen to the resulting aud io signal. You must decrease the
span to continue the measurement
Status Ext Ref (no corresponding status bit)
The instrument’s frequency refer ence has been set to
This message will appear even if an external reference is not connected to
the instrument. The external reference frequency must also be entered.
Indicates that the frequency reference is being supplied by an external
10 MHz source.
Status External LO frequency out of range
One or more external LO frequencies are out of range. Check that the LO
frequency limits are set correctly and check the entered measurement
frequencies and measurement mode .
External Reference.
Instrument Messages
Status Final IF Overload (IF/ADC Overrange)
The final IF section has been overloaded. Measurement results may be
invalid. Either increase the input attenuation or decrease the input signal
level.
Status First IF Overload (IF/ADC Overrange)
The first IF section has been overloaded. Measurement results may be
invalid. Either increase the input attenuation or decrease the input level.
Status Flat corr off (no corresponding status bit)
Indicates that the flatness corrections have been turned off.
Status (FM Demod Align Failure) status bit only, no message
A failure has occurred during the FM Demod alignment. Measurement
results may be invalid.
Status Freq corr off (no corresponding status bit)
Indicates that the frequency corrections have been manually disabled.
Press
System, Alignments, Freq Correct, (On) to restore.
88Chapter 1
Status Freq Count: Reduce Span/RBW ratio
The span is too wide for the current resolution bandwidth. Either reduce
the span or increase the RBW.
Status Freq outside cal range
The existing user cal has been invalidated because the current
measurement frequencies lie partially or wholly outside the range of
frequencies used for user-cal.
Status Frequency Reference Error (Freq Ref Unlocked)
The frequency reference lock loop has unlocked or is near an unlock
condition . T h is message can oc cur when either the intern a l o r external
reference is selected. If exter nal reference is select ed, assure the frequenc y
of the external reference source exactly matches the user entered value
displayed on the reference menu key.
The frequency reference has been tuned too far off of 10 MHz. This
condition may be corrected by cycling power on the analyzer.
Status (IF Align Failure) status bit only, no message
Instrument Messages
Instrument Messages
A failure has occurred during the IF alignment. Measurement results ma y
be invalid.
Status IF Gain fixed
The autoranging function of the analyzer has been turned off (
More, More, IF Gain (Fixed)). This setting is useful when measuring signals
that require fast measurement time, narrow resolution bandwidths (< 1
kHz), and < 70 dB of display range. For more information on this setting,
refer to IF Gain key description in the ESA Spectrum Analyzer User’s Guide.
Status IF Overload(IF/ADC Over Range)
The IF section has been overloaded. Measurement results may be invalid.
Status Input is internal (no corresponding status bit)
This message applies to the Agilent Technologies E4401B and E4411B only.
Indicates the
50 MHz Amptd Ref selection is On. W ith the 50 MHz amplitude
reference on, the input is routed throug h an internal signal path.
Status LO – Stop freq must be >= min system input freq
The current measurement mode requires that the stop RF (input to DUT)
frequency must be more than the minimum system input frequency away
from the fixed LO frequency.
Amplitude,
Instrument Messages
Status (LO Align Failure) status bit only, no message
A failure has occurred during the alignment of the local oscillator (LO).
Measurement results may be invalid.
Chapter 189
Instrument Messages
Instrument Messages
Status LO Out Unlevel (LO Out Unleveled)
Indicates the output of the local osci llator (LO) has become unleveled . This
condition must be corrected to make valid measurements.
Status LO Unlevel (LO Unleveled)
Indicates the internal circuitry of the local osc illator (LO) has become
unleveled. This condition must be corrected to make valid measurements.
Status LO Unlock (Synth Unlocked)
Indicates the phase locked circuitry of the local oscillator (LO) has become
unlocked. This condition must be corrected to make valid measurements.
Status Log Corr Off (no corresponding status bit)
The log amplifier corrections have been turned off.
Status Marker Count:Widen Res BW
The ratio of the resolution bandwidth to span must be greater than 0.002
for the marker count function to work properly. Increase the resolution
bandwidth or decrease the span to continue the measurement.
Instrument Messages
Status Meas Uncal (Oversweep)
The measurement is uncalibrated, usually due to sweep ing a narrow RBW
filter too quickly. Check the sweep time, span, and bandwidth settings, or
press
Auto Couple.
Status Measurement outside Internal Preamp range
The Internal Preamp is currently t urned on, but the measurement is being
performed outside the range of the preamp.
Status Measurement partially outside Internal Preamp range
The Internal Preamp is currently turned on, and the measurement range
is being performed partly within and partly outside this range.
Recommendation is that the user turns preamp off to ensure consistent
results across the entire measurement.
Status Overload: Reduce Signal and press <ESC> (Input Overload
Tripped)
This message applies to the Agilent Technologies E4401B and E4411B only.
A signal has been applied to the input connector that caused the overload
protection circuitry to e ngage. The input signal must be reduced. After th e
signal is reduced, press
ESC to reset the overload detector so that you can
continue using the analyzer.
CAUTIONExposing the analyzer to high le vels of inp ut power over a prolonged period of
time can damage the internal circuitry.
90Chapter 1
Status Peaking Signal (no corresponding status bit)
The instrument is executing a tracking generator peak.
Status Preferred resolution bandwidth not available.
The calculated required resolution bandwidth for this measurem ent is not
available.
Status (RF Align Failure) status bit only, no message
A failure has occurred during the alignment of the RF section.
Measurement results may be invalid.
Status Signal Ident On, Amptd Uncal (Signal Ident On)
Indicates that the amplitude measurement could be uncalibrated because
the signal id en tification fea t u re is on.
Status Signal level is low.
Indicates the signal can be correlat ed, however the level is below that
specified to ensure accurate measurement results.
Instrument Messages
Instrument Messages
Status Source LO Unlevel (Source LO Unleveled)
The internal circuitry of the local oscillator (LO) in the tracking generator
has become unleveled. This condition must be corrected to make valid
measurements.
Status Source LO Unlock (Source Synth Unlocked)
The phase-locked circuitry of the local oscillator (LO) in the tracking
generator has become unlocked. This condition must be corrected to make
valid measurements.
Status Source Unlevel (Source Unleveled)
Indicates the source power is set higher or lower than the analyzer can
provide, the frequency span extends beyond the sp ecified frequency range
of the tracking ge nerator, or the calib rati on data f or t he sourc e is incorr ect .
Status Start freq must be greater than fixed LO freq
The current measurement mode requires t hat the star t frequen cy must be
greater tha n th e fixed LO frequency.
Status Start freq must be greater than start RF freq
The current measurement mode re quires that the sta rt IF (output to DUT)
frequency must be greater than the start RF (input to DUT) frequency.
Instrument Messages
Status Start – LO freq must be >= min system input freq
The current measurement mode require s th at the start RF ( inpu t to DUT)
frequency must be more than the minimum system input frequency away
from the fixed LO frequency.
Chapter 191
Instrument Messages
Instrument Messages
Status Stop freq must be less than fixed LO freq
The current measurement mode requires that the stop frequency must be
less than the fixed LO frequency.
Status Stop freq must be less than stop RF freq
The current measurement mode requires that the stop IF (output to DUT)
frequency must be less than the stop RF (input to DUT) frequency.
Status Stop IF freq must be less than fixed LO freq
The current measurement mode requires that the stop IF (output to DUT)
frequency must be less than the fixed LO frequency.
Status System, Alignments, Align All Now Needed (Align Needed)
Instrument Messages
This message occurs only when
The instrument requires an
pressing
System, Alignments, Align All Now.
Auto Align is set to Alert.
Align All Now. Restore the alignment by
Status System input frequency out of range
One or more system input frequencies are out of range . If using a
frequency list, check that all entries are valid for current measurement
mode.
Status System Alignments, Align Now, All Required
Internal alignment correction data has been lost. Press
Alignments, Align Now, All to clear this message from the display.
Status (TG Align Failure) status bit only, no message
A failure has occurred during the tracking generator (TG) alignment.
Status Video shift off (no corresponding status bit)
Indicates the video shift has been manually disabled; this will impair
readings.
Annunciator Bar Messages
Annunciator Align
System,
This message appears in the left side of the annunciator bar in red . When
this message appears it indicates that th ere is one or more condition that
has affected the alignment. For the possible conditions and suggested
resolutions, refer to Align Now All Needed (Align Needed) on page 87,
Align Now RF Needed (Align Now RF Needed) on page 87, Align RF
Skipped (Align RF Skipped) on page 87, and System Alignments, Align
Now, All Required on page 92.
92Chapter 1
Annunciator Ext Ref
This message appears in the right side of the annunciator bar in green.
When this message appears it indic ates that there is one or more condi tion
that affects the reference source. For the possible conditions and suggested
resolutions, refer to Ext Ref on page 88.
Annunciator FreqRefUnlock
This message appears in the left side of the annunciator bar in red . When
this message appears it indicates that th ere is one or more condition that
results in a frequency reference error. For the possible co nditions and
suggested resolu tions, refer to Frequency Reference Error (Freq Ref
Unlocked) on page 89.
Annunciator Overload
This message appears in the left side of the annunciator bar in red . When
this message appears it indicates that th ere is one or more condition that
causes a system overload. For the possible conditions and suggested
resolutions, refer to IF Overload on page 89 and the Status overload on
page 90.
Instrument Messages
Instrument Messages
Annunciator Uncal
This message appears in the left side of the annunciator bar in red . When
this message appears it indicates that th ere is one or more condition that
results in uncalibrated results. For the possible conditions and suggested
resolutions, refer to Meas Uncal (Oversweep) on page 90.
Annunciator Unlevel
This message appears in the left side of the annunciator bar in red . When
this message appears it indicates that th ere is one or more condition that
affects the leveling of an inte rnal signal, such as the local osci llator (LO) or
the tracking generator (source). For the possible conditions and sugges ted
resolutions, refer to 50 MHz Osc Unlevel (50 MHz Osc Unleveled) on page
86, LO Out Unlevel (LO Out Unleveled) on page 90, Source LO Unlevel
(Source LO Unleveled) on page 91, and Source Unlevel (Source Unleveled)
on page 91.
Annunciator Unlock
This message appears in the left side of the annunciator bar in red . When
this message appears it indicates that th ere is one or more condition that
affects locking one or more of the local oscillators (LOs). For the possible
conditions and suggested re solutions , re fer to LO Unlock (Synth Unlocked)
on page 90.
Instrument Messages
Chapter 193
Instrument Messages
Instrument Messages
Instrument Messages
94Chapter 1
2Functional Tests
Functional Tests
95
Functional Tests
What You Will Find in This Chapter
What You Will Find in This Chapter
This chapter describes the functional tests and provides information on how to perform
them.
What Are the Functional Tests?
Functional tests are tests of various instrument parameters that give a high degree of
confidence that the analyzer is operating correctly. They are recommended as a chec k of
analyzer operation for incoming inspection or after a repair. Measurement uncertainty
analysis is not available for functional tests, and the analyzer is checked against limits
that are wider than the published specificat ions. The functional tests are designed to test
an analyzer operating within the temperatur e range def ined by the analyzer specifica tions
using a minimum set of test equipment. If a test does not pass, performance tests must be
run to confir m a p ro b l e m ex i sts.
Functional Test Versus Performance Verification
Functional tests use a minimum set of test equipment to check a much smaller range of
parameters (and a limited number of data points for each paramet er) than do pe rformance
verification tests. Functional tests use limits that are wider than the published
specifications; measurement uncertainty analysis is not available for functional tests.
NOTEIf a functional test does not pass , you must run p erformance ve rification te sts
to confirm that a proble m exists.
Performance verification tests span a wide range of instrument parameters and provide
the highest level of confidence that the instrument conforms to published specifications.
These tests can be time consuming and require extensive test equipment.
This chapter includes the following:
• Bef ore Performing a Functional Test on page 97 (what to do first).
• Test Equi pment on page 98 (a list of the equipment required for all of the tests).
• PSA Funct ional Tests (PSA only):
— Amplitude Accuracy at 50 MHz on page 102
— Displayed Average Noise Level (DANL) on page 107
— Frequency Readout Accuracy on page 109
— Frequency Response (Flatness) on page 111
— Amplitude Linearity on page 115
— Second Harmonic Distortion (SHD) on page 118
Functional Tests
• ESA Funct ional Tests (ESA only):
— Displayed Average Noise Level on page 122
96Chapter 2
Functional Tests
What You Will Find in This Chapter
— Frequency Readout Accuracy on page 139
— Marker Count Accuracy on page 141
— Frequency Response (Flatness) on page 142
— Reference Level Accuracy on page 146
— Resolution Bandwidth Switching Uncertainty on page 151
— Scale Fidelity on page 154
— Second Harmonic Spurious Responses on page 157
— Tracking Generator Level Flatness: Models E4401B and E4411B, Options 1DN and
• One or more tables in which to record the measurement results
Before Performing a Functional Test
1. Ensure that you have the proper test equipment.
2. Switch on the unit under test (UUT) and let it warm up (in accordance with warm-up
requirements in the instrument specifications).
3. Allow sufficient warm-up time for the required test equipment (refer to individ ual
instrument documentation for warm-up specifications).
4. Ensure that the analyzer’s frequency reference is set to Internal:
a. Press the
b. If the
is underlined.
5. Following instrument warm-up, perform the auto align routine:
Press
System, Alignments, Align All Now.
NOTEFunctional test accuracy depends on the pr ecision of the test equi pment use d.
System, Reference keys.
Freq Ref softkey does not have Int underlined, press the Freq Ref s oftkey until it
Ensure that all of the test equipment is calibrat ed before running a functional
test.
Functional Tests
Chapter 297
Functional Tests
Test Equipment
TEST EQUIPMENT
The table below summarizes the test equipment needed to perfo rm all of the funct ional
tests. Alternate equipment model numbers are given in case the recommended equipment
is not available.
If neither the recommended nor the alternative test equipment are available, subs titute
equipment that meets or exceeds the critical specifications listed.
Analyzer
ItemCritical Specifications
Adapters
PSA ESA3.5 mm (f) to 3.5mm (f)
(connector saver for
83630B)
PSA ESA3.5 mm (f) to N (f)1250-1745
PSA 2.4 mm (f) to 3.5 mm (m)E4446A, E4448A only
PSA BNC (f) to SMA (m)1250-1200
PSA BNC (m) to SMA (f)1250-1700
ESABNC (m) to SMA (f)1250-2015
PSA ESAType N (f) to BNC (m)1250-1534
PSA Type N (f) to N (f)1250-14721250-0777
PSA ESAType N (m) to 3.5 mm (f)1250-1744
PSA ESAType N (m) to 3.5 mm (m)1250-1743
PSA Type N (m) to BNC (f)1250-1476
Recommended
Agilent Model
5061-5311
11901D
Alternate
Agilent
Model
Functional Tests
ESAType N (m) to BNC (f)1250-0780
Attenuators
PSA 10 dB Step AttenuatorRange: 0 to 60 dB355D
PSA 20 dB Fixed AttenuatorAccuracy: < 0.5 dB8491A8491B
Cables
PSA ESAAPC 3.5 mm (1 meter)8120-4921
ESAAPC 3.5 mm E4407B and E4408B only11500E
ESAAPC 3.5 mm11500D
ESAType-N, 152-cm (60-in)11500D
ESAAPC 3.5 mm E4407B and E4408B only11500E
98Chapter 2
Functional Tests
Test Equipment
Analyzer
ItemCritical Specifications
PSA ESACable, BNC (2 required)120 cm (48 in.) BNC cable10503A
Signal Source
PSA Synthesized SweeperFrequency: 10 MHz to 26.5 GHz
Harmonic level:< −40 dBc
Amplitude range: 10 to −20 dBm
Frequency Accuracy: 0.02%
PSA Synthesized SweeperFrequency:10MHz to 50GHz
Harmonic level:< −40 dBc
Amplitude range: 10 to −20 dBm
Frequency Accuracy: 0.02%
E4446A, E4448A only
ESASynthesized SweeperFrequency: 10 MHz to 26.5 GHz
Harmonic level:< −40 dBc
Amplitude range: 10 to −20 dBm
Frequency Accuracy: 0.02%