Agilent Technologies 490 User Manual

Agilent 490 Micro Gas Chromatograph
User Manual
Agilent Technologies
Notices
CAUTION
WARNING
© Agilent Technologies, Inc. 2017
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
Manual Part Number
G3581-90001
Edition
Sixth edition, November 2017
Printed in China
Agilent Technologies, Inc. 412 Ying Lun Road Waigoaqiao Freed Trade Zone Shanghai 200131 P.R.China
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
(June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
A CAUTION notice denotes a hazard. It calls attention to an oper­ating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14

Contents

1 Introduction
Safety Information 8
Important safety warnings 8 Hydrogen safety 8 Safety symbols 9 Safety and regulatory information 10 General safety precautions 10
Shipping Instructions 13
Cleaning 13
Instrument Disposal 13
2 Instrument Overview
Principle of Operation 16
Front View 17
Back View 18
Inside View 19
Carrier Gas Connection 21
Power 23
Power source 23 Power Requirements 23 Disposal 24 Specifications 24
Ambient Pressure 25
Ambient Temperature 25
Maximum Operation Altitude 25
Micro GC Cycle with Constant Pressure 26
Micro GC Cycle with Ramped Pressure 27
3 Installation and Use
Pre-Installation Requirements 30
490 Micro GC User Manual 3
Inspect the Shipping Packages 30
Unpack the Micro GC 31
Review the Packing List 32
490 Micro GC Installation 33
Step 1: Connect carrier gas 33 Step 2: Connect to calibration gas or checkout sample 33 Step 3: Install power supply 33 Step 4: Connect to computer or local network 34 Step 5: Install Chromatography Data System 34 Step 6: Assign IP address 34
Restore the Factory Default IP Address 38
Create the Test Method 40
Perform a Series of Runs 41
Shut Down Procedure 42
Long Storage Recovery Procedure 42
4 Sample Gas Handling
5 GC Channels
Using the external filter unit 46
Heated sample lines 47
How to connect your sample to the 490 Micro GC 48
Rear inlet (heated or unheated) 48 Internal inlet 49 Internal bracket for Genie filter 51
490-Micro GC Optional Pressure Regulators 53
G3581-S0003 53 G581-S0004 56
Manual Injection 59
Manual injection guidelines 59 Injection Procedure 60 Field upgrade kits 60 Manual injection flow diagrams 61
Carrier Gas 66
Micro Electronic Gas Control (EGC) 67
Inert Sample Path 67
Injector 67
4 490 Micro GC User Manual
Column 68
Molsieve 5Å columns 69 CP-Sil 5 CB columns 70 CP-Sil CB columns 71 PoraPlot 10m column 72 Hayesep A 40 cm heated column 73 COX and AL203/KCI columns 74 MES (NGA) and CP-WAX 52 CB columns 75 Column conditioning 76
Backflush Option 77
Tuning the backflush time (except on a HayeSep A channel) 79 Tuning the backflush time on a HayeSep A channel 80 To disable backflush 81
Backflush to Detector 82
CP-Sil 5 CB Backflush to detector 82 Al2O3 Backflush to detector 82 Tuning the backflush time 83 To disable backflush 85 Set invert signal time 85 Checkout information 86 C6+ Calorific value calculation 88
TCD Detector 88
6 Channel Exchange and Installation
Tools required 90
Replacement procedure for Micro GC channel 91
Replacement procedure for Micro GC channel with RTS option 99
Replacement procedure for Molsieve filters with the RTS option 103
Carrier gas Tube Stop Modification Kit 105
7 Communications
Access the Connection Ports 108
490 Chromatography Data Systems 110
Ethernet Networks 111
IP Addresses 112 Example network configurations 112
USB VICI Valve 116
Configure Multiple VICI Valves with OpenLab EZchrom 116
USB Wi-Fi 118
490 Micro GC User Manual 5
8 Errors
Frequently Asked Questions (FAQ) 121
Glossary of network terms 121
External Digital I/O 123
External Analog I/O 124
Error Handling 126
Error List 127
6 490 Micro GC User Manual
Agilent 490 Micro Gas Chromatograph User Manual
1 Introduction
Safety Information 8 Shipping Instructions 13 Cleaning 13 Instrument Disposal 13
This chapter provides important information about using the Agilent 490 Micro Gas Chromatograph (Micro GC) safely. To prevent any injury to you or any damage to the instrument it is essential that you read the information in this chapter.
Agilent Technologies
7
1 Introduction
WARNING
WARNING

Safety Information

Important safety warnings

There are several important safety notices that you should always keep in mind when using the Micro GC.
When handling or using chemicals for preparation or use within the Micro GC, all applicable local and national laboratory safety
When
practices must be followed. This includes, but is not limited to, correct use of Personal Protective Equipment, correct use of storage vials, and correct handling of chemicals, as defined in the laboratory’s internal safety analysis and standard operating procedures. Failure to adhere to laboratory safety practices could lead to injury or death.

Hydrogen safety

Hydrogen is a commonly used GC carrier gas. When mixed with air, hydrogen can form explosive mixtures and has other dangerous characteristics.
When using hydrogen (H2) as the carrier gas, be aware that hydrogen gas can create a fire or explosion hazard. Be sure that the supply is turned off until all connections are made.
Hydrogen is flammable. Leaks, when confined in an enclosed space, may create a fire or explosion hazard. In any application using hydrogen, leak test all connections, lines, and valves before operating the instrument. Always turn off the hydrogen supply at its source before working on the instrument.
Hydrogen is combustible over a wide range of concentrations. At atmospheric pressure, hydrogen is combustible at concentrations from 4 % to 74.2 % by volume.
Hydrogen has the highest burning velocity of any gas.
Hydrogen has a very low ignition energy.
Hydrogen that is allowed to expand rapidly from high pressure into
the atmosphere can self-ignite.
Hydrogen burns with a nonluminous flame which can be invisible under bright light.
8 490 Micro GC User Manual

Safety symbols

Introduction 1
Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
See accompanying instructions for more information.
Indicates a hot surface.
Indicates hazardous voltages.
Indicates earth (ground) terminal.
Indicates potential explosion hazard.
Indicates electrostatic discharge hazard.
Indicates a hazard. See the Agilent 490 GC user documentation for the item labeled.
Indicates that you must not discard this electrical/electronic product in domestic household waste
490 Micro GC User Manual 9
1 Introduction

Safety and regulatory information

This instrument and its accompanying documentation comply with the CE specifications and the safety requirements for electrical equipment for measurement, control, and laboratory use (CEI/IEC 1010-1)
This device has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
CSAUS and FCC-b.
C
NOTICE This instrument has been tested per applicable
requirements of EMC Directive as required to carry the European Union CE Mark. As such, this equipment may be susceptible to radiation/interference levels or frequencies, which are not within the tested limits.

General safety precautions

Follow the following safety practices to ensure safe equipment operation:
Perform periodic leak checks on all supply lines and
pneumatic plumbing.
Do not allow gas lines to become kinked or punctured. Place
lines away from foot traffic and extreme heat or cold.
Store organic solvents in fireproof, vented and clearly
labeled cabinets so they are easily identified as either toxic, or flammable, or both types of materials.
Do not accumulate waste solvents. Dispose of such materials
through a regulated disposal program and not through municipal sewage lines.
10 490 Micro GC User Manual
Introduction 1
WARNING
WARNING
This instrument is designed for chromatographic analysis of appropriately prepared samples. It must be operated using appropriate gases or solvents and within specified maximum ranges for pressure, flows, and temperatures as described in this manual. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
It is the responsibility of the customer to inform Agilent customer support representatives if the instrument has been used for the analysis of hazardous samples, prior to any instrument service being performed or when an instrument is being returned for repair.
Avoid exposure to potentially dangerous voltages.
Disconnect the instrument from all power sources before removing protective panels.
When it is necessary to use a non-original power cord and
plug, make sure the replacement cord adheres to the color coding and polarity described in the manual and all local building safety codes.
Replace faulty or frayed power cords immediately with the
same type and rating.
Place this instrument in a location with sufficient ventilation
to remove gases and vapors. Make sure there is enough space around the instrument for it to cool off sufficiently.
Before plugging the instrument in or turning the power on,
always make sure that the voltage and fuses are set appropriately for your local power source.
Do not turn on the instrument if there is a possibility of any
kind of electrical damage. Instead, disconnect the power cord and contact your local Agilent sales office.
The supplied power cord must be inserted into a power
outlet with a protective ground connection. When using an extension cord, make sure that the cord is also properly grounded.
Do not change any external or internal grounding
connections, as this could endanger you or damage the instrument.
490 Micro GC User Manual 11
1 Introduction
The instrument is properly grounded when shipped. You do
not need to make any changes to the electrical connections or to the instrument chassis to ensure safe operation.
When working with this instrument, follow the regulations
for Good Laboratory Practices (GLP). Take care to wear safety glasses and appropriate clothing.
Do not place containers with flammable liquids on this
instrument. Spilling liquid over hot parts may cause fire.
This instrument may use flammable or explosive gases, such
as hydrogen gas under pressure. Before operating the instrument be sure to be familiar with and to follow accurately the operation procedures prescribed for those gases.
Never try to repair or replace any component that is not
described in this manual without the assistance of an Agilent service engineer. Unauthorized repairs or modifications will result in rejection of warranty claims.
Always disconnect the AC power cord before attempting any
type of maintenance.
Use proper tools when working on the instrument to prevent
danger to you or damage to the instrument.
Do not attempt to replace any battery or fuse in this
instrument other than as specified in the manual.
Damage can result if the instrument is stored under
unfavorable conditions for prolonged periods. (For example, damage will occur if stored while subject to heat, water, or other conditions exceeding the allowable operating conditions).
Do not shut off column flow when the oven temperature is
high, since this may damage the column.
This unit has been designed and tested in accordance with
recognized safety standards and designed for use indoors.
If the instrument is used in a manner not specified by the
manufacturer, the protection provided by the instrument may be impaired.
Substituting parts or performing any unauthorized
modification to the instrument may result in a safety hazard.
Changes or modifications not expressly approved by the
responsible party for compliance could void the user's authority to operate the equipment.
12 490 Micro GC User Manual

Shipping Instructions

Cleaning

Introduction 1
If your Micro GC must be shipped for any reason, it is very important to follow these additional shipping preparation instructions:
Place all the vent caps on the back of the Micro GC (see
Figure 3 on page 18).
Always include the power supply.
Include, if used, the inlet filter(s).
To clean the surface of the Micro GC:
1 Switch the Micro GC off. 2 Remove the power cable.

Instrument Disposal

3 Put protection plugs on the sample and carrier gas inlets. 4 Put protection plugs on the column vents. 5 Use a soft brush (not hard or abrasive) to carefully brush
away all dust and dirt.
6 Use a soft, clean cloth dampened with mild detergent to
clean the outside of the instrument.
Never clean the inside of the instrument.
Never use alcohol or thinners to clean the instrument;
these chemicals can damage the case.
Be careful not to get water on the electronic components.
Do not use compressed air to clean the instrument.
When the Micro GC or its parts have reached the end of their useful life, dispose of them in accordance with the environmental regulations that are applicable in your country.
490 Micro GC User Manual 13
1 Introduction
14 490 Micro GC User Manual
Agilent 490 Micro Gas Chromatograph User Manual
2 Instrument Overview
Principle of Operation 16 Front View 17 Back View 18 Inside View 19 Carrier Gas Connection 21 Power 23 Ambient Pressure 25 Ambient Temperature 25 Maximum Operation Altitude 25 Micro GC Cycle with Constant Pressure 26 Micro GC Cycle with Ramped Pressure 27
There are several versions of the Agilent 490 Micro GC. All of them use GC channels, each of which consists of an Electronic Gas Control (EGC) injector, column, and detector.
The Micro GC is a self-contained package with all of the normal GC components. It is available as a dual channel cabinet version (one or two GC channels) or a quad channel cabinet version (up to four GC channels). A computer with a chromatography data system (CDS) is needed to complete the system.
This chapter provides a brief overview of the 490 Micro GC.
Agilent Technologies
15
2 Instrument Overview

Principle of Operation

The 490 Micro GC can be equipped with one to four independent column channels. Each column channel is a complete, miniaturized GC with electronic carrier gas control, micro-machined injector, narrow-bore analytical column and micro thermal conductivity detector (µ TCD), Figure 1.
Figure 1 490 Micro GC setup
The 490 Micro GC analytical channels can optionally be equipped with a back flush option. The advantages include the protection of the stationary column phase against moisture and carbon dioxide. Next to that, it results in shorter analysis times as late elution compounds, which are not of interest, do not enter the analytical column.
16 490 Micro GC User Manual

Front View

Ready LED
LED OFF: System not ready LED ON: System is ready
Run LED
LED OFF: No run LED blinking: Run in progress
Error LED
LED OFF: No error LED blinking: Error present
See “Error List” on page 127
Power LED
LED OFF: No power LED ON: Power OK LED blinking: Voltage < 10 Volt
Sample 1 and Sample 2
Sample gas inlet connector (for unheated front inlets)
See “Manual Injection” on page 59
Power On/Off Switch
Switch the Micro GC ON or OFF
Instrument Overview 2
Figure 2 Front view of the 490 Micro GC
490 Micro GC User Manual 17
2 Instrument Overview
Power connector
Power connector (male) See “Power” on page 23
Vents
It is possible to connect long vent lines to these fittings in order to safely guide hazardous fumes to a fume hood or other appropriate vent.
Carrier gas input
Carrier gas input connector See “Carrier Gas Connection” on page 21

Back View

Figure 3 Back view of the 490 Micro GC
18 490 Micro GC User Manual

Inside View

Assign IP address switch
See “Ethernet Networks” on page 111.
LAN indicators
Red LED: Transmit data Green LED: Receive data
Ethernet (LAN) connector
Ethernet RJ45 connector.
See “Ethernet
Networks” on page 111.
COM 1
RS-232 communication interface
Digital I/O
Digital input and output signals, such as start_stop, ready_out, and start_in.
See “External Digital
I/O” on page 123.
COM 3 and COM 4
RS-485 (4-wire) communication interface.
See Table 1 on page 20.
Analog I/O
External analog I/O signals. See “External Analog
I/O” on page 124.
COM 2
RS-232 (2-wire) communication interface. See “490 Chromatography
Data Systems” on
page 110.
USB
communication interface. See “USB VICI Valve” on
page 116 and “USB
Wi-Fi” on page 118
SD Card Slot
No function supported.
Instrument Overview 2
Open the right side cover and the cable connectors will be visible. See Figure 4.
Figure 4 Cable connectors (main board G3581-65000 shown)
490 Micro GC User Manual 19
2 Instrument Overview
The Micro GC provides communications ports as shown in
Table 1, depending on the model.
Table 1 Micro GC communication ports
Port Connection 490 Micro GC 490-Mobile Micro GC 490-PRO Micro GC
LAN Ethernet Interface with PC Interface with PC Interface with PC
COM 1 RS232 Not available Not available Valco stream selector;
Serial MODBUS
COM 2 RS232 Valco stream selector
Field case
LCD
COM 3 RS485
RS232 RS422
COM 4 RS485
RS232 RS422
Not available Not available Not available
Not available Not available Not available
Valco stream selector Field case
LCD
Not available Not available Not available
Not available Not available Not available
Valco stream selector; Serial MODBUS
LCD
Serial MODBUS Not available Not available
Serial MODBUS Not available Not available
*
*
;
*
*
Analog I/O Analog I/O Analog I/O Analog I/O
Digital I/O Digital I/O;
ready in - ready out; start in - start out; extension boards
USB VICI Valves, WIFI interface VICI Valves, WIFI interface,
Digital I/O; ready in - ready out; start in - start out; extension boards
USB Storage
Digital I/O; ready in - ready out; start in - start out; extension boards
VICI Valves, WIFI interface, USB Storage
* Optional PRO license required † Optional accessory ‡ Extension boards not included
20 490 Micro GC User Manual

Carrier Gas Connection

CAUTION
Instrument Overview 2
The carrier gas line is connected to the Micro GC at the back panel Carrier 1 or Carrier 2 port.
Do not use any kind of plastic tubing since air will diffuse through the tubing, which may cause noisy baselines and decreased sensitivity. The metal tubing must be clean for GC use. Buy either flamed or chromato­graphically clean tubing.
Specifications for the carrier gas:
Pressure: 550 kPa ± 10 kPA (80 psi ± 1.5 psi) Purity: 99.999 % minimum Dry and free of particles: Gas Clean filters recommended
Gas Clean filters are recommended to remove any traces of moisture and oxygen. For low-level analysis, consider using a better grade of carrier gas.
Gas Clean filters are filled with nitrogen. If you are not using nitrogen as the carrier gas, flush filters and gas lines after installation of a new filter.
The type of analysis you want to perform dictates the type of carrier gas to use. The difference between the relative thermal conductivity of the carrier gas and the sample components should be as high as possible. See Table 2 for several relative thermal conductivities.
Table 2 Relative thermal conductivities
Relative thermal
Carrier gas
Hydrogen 47.1 Ethane 5.8
Helium 37.6 Propane 4.8
Methane 8.9 Argon 4.6
Oxygen 6.8 Carbon dioxide 4.4
Nitrogen 6.6 Butane 4.3
Carbon monoxide 6.4
conductivities
Carrier gas
Relative thermal conductivities
490 Micro GC User Manual 21
2 Instrument Overview
WARNING
WARNING
Your Micro GC is configured for a specific carrier gas, either He and H2 or N2 and Ar. Make certain that any carrier gas selection in your Agilent data system corresponds to the carrier gas physically connected to your Micro GC. Use only the carrier gas corresponding to this configuration. If you change the carrier gas type plumbed to the Micro GC, you must change the corresponding carrier gas type in the data system.
Hydrogen is flammable. If you are using hydrogen as a carrier gas, pay particular attention to possible leaks at connections inside and outside of the Micro GC (use an electronic leak tester).
22 490 Micro GC User Manual

Power

CAUTION
Instrument Overview 2

Power source

90 to 264 Vac, frequency between 47 to 63 Hz.
The room power outlet circuit must be exclusively reserved
for the instrument(s).
The network should be properly grounded.
Installation Category (overvoltage category): II

Power Requirements

The Micro GC requires 12 V Vdc, 150 W.
The Gasifier requires 12 V Vdc, 150 W.
Only use the power supply provided with your Micro GC.
This Power Supply, see Figure 5, is tailored to meet the power needs of your Micro GC. See Table 3 on page 24 for specifications.
Figure 5 Model GST220A12-AG1 (P/N G3581-60080)
490 Micro GC User Manual 23
2 Instrument Overview

Disposal

Specifications

Disposal of the Power Supply must be carried out in accordance with all environmental regulations applicable in your country.
Table 3 Power supply specifications
Feature Model: GST220A12-AG1
Input voltage 85 Vac to 264 Vac
Input frequency 47-63 Hz
Inrush current 120A/230VAC
Output voltage 12.0 Vdc
Voltage adjust ± 5 %
Output power 180 W
Over voltage protection 105 %-135 % rated output voltage
Ripple and noise 80mV Vp-p
Operating temperature -30 °C to +70 °C
Storage temperature -40 °C to +85 °C
Humidity 20 % to 90 % non condensing
Safety standard UL60950-1, TUV EN60950-1, BSMI CNS14336, CSA
C22.2, CCCGB4943, PSE J60950-1 Approved
RFI/EMC standard In compliance with CISPR22 (EN55022) Class B and
FCC Part 15/CISPR 22 class B, CNS13438 class B, GB9254, EN61000-3-2, EN61000-3-3, EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11 (light industry level, criteria A)
Dimensions 210 × 85 × 46 mm (L×W×H)
Weight 1.1 kg approximately
24 490 Micro GC User Manual

Ambient Pressure

The Micro GC automatically shuts down if the ambient pressure is greater than 120 kPa.

Ambient Temperature

The Micro GC automatically shuts down if the ambient temperature exceeds 65 °C.

Maximum Operation Altitude

Certified up to 2000 meters above sea level.
Instrument Overview 2
490 Micro GC User Manual 25
2 Instrument Overview
Pressure
Time
Initial pressure
Run started
Equilibration time
Sample time
Pressurization time
Download method
Inject time Run time
Inject Start data
acquuisition
Run finished
After the run sampling is begun. This means that the sample is (mostly) sucked into the sample loop.
Pressurization delay (120 mS), used to pressurize the sample to the same pressure as the column head pressure.
During injection time the sample is transported to the column by the carrier gas.
Before a run is started all used method parameters must be downloaded via the data system.
Fixed equilibration delay of 40 ms. This time allows the sample to settle in the sample loop.
The run can be started when the Micro GC is ready (Ready LED is lit). The data system will wait until data is received from the Micro GC.
The real analysis (run) and data acquisition starts.
After the specified run time has elapsed, the run is finished.

Micro GC Cycle with Constant Pressure

The timing diagram below provides an overview of the constant pressure cycle of the Micro GC.
This description is only for one channel. In most cases a dual-channel system is used. When a dual-channel system is used, the sequence is the same, but the timing settings can differ. If the sample time on channel A and channel B are different, the longest time is used for both channels. Also the run time can be specified per channel; the data acquisition stops per channel as soon as the run time has elapsed. The total analysis time depends on the longest run time.
26 490 Micro GC User Manual

Micro GC Cycle with Ramped Pressure

Pressure
Time
Initial pressure
Run started
Equilibration time
Sample time
Pressurization time
Download method
Inject time
Run time
Inject Start data
acquuisition
Run finished
Final pressure
Not ready
Pressure release
Pressure rise
Ready
Ready
Initial time
Rise time Final time
Relieve time
Stabilization time
The remaining final time depends on the total run time, the duration of the initial time and the pressure rise. This means that it is possible that the final time is zero. Another situation is that the final pressure is limited because of these settings. The software will check all parameter values and change them into realistic values.
Note: During the run time, there can be only one pressure ramp to a higher pressure.
The pressure rise is started, the duration is depending on two (2) parameters:
• Pressure rise
• Final pressure
Stabilization time for the pressure after it has been returned to the initial pressure. Fixed at 500 mS.
Relieve time, the time needed to decrease the column head pressure from the final pressure to the initial pressure.
As soon as the final pressure is reached, the rise stops and the final time begins. The pressure remains the same.
During initial time the column head pressure remains the same.
The timing diagram below provides an overview of the ramped (programmed) pressure cycle of the Micro GC. The timing before the injection is identical to the constant pressure cycle.
Instrument Overview 2
490 Micro GC User Manual 27
2 Instrument Overview
28 490 Micro GC User Manual
Agilent 490 Micro Gas Chromatograph User Manual
3 Installation and Use
Pre-Installation Requirements 30 Inspect the Shipping Packages 30 Unpack the Micro GC 31 Review the Packing List 32 490 Micro GC Installation 33 Restore the Factory Default IP Address 38 Create the Test Method 40 Perform a Series of Runs 41 Shut Down Procedure 42 Long Storage Recovery Procedure 42
This chapter describes how to install and use the instrument. For an initial installation, an example of a typical packing list is also included. The actual packing list and included parts depend on the options ordered.
Agilent Technologies
29
3 Installation and Use

Pre-Installation Requirements

Prepare the installation site as described in the Site Preparation Guide (G3581-90002), including the recommended Gas Clean filters.

Inspect the Shipping Packages

The Micro GC will arrive in one large box and one or more smaller cartons. Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to your local Agilent office.
30 490 Micro GC User Manual
Loading...
+ 104 hidden pages