3B Scientific Magnetic Field Sensor User Manual

3B SCIENTIFIC
Magnetic Field Sensor ±2000 mT 1009941
04/12 Hh
®
PHYSICS
1. Description
Sensor box with attached sensor probe, 140 mm long, 7 mm wide and 1.9 mm thick, for
measuring magnetic flux density B of a mag-
netic field acting outside the box in a tangential plane, primarily for use with the basic Hall­effect apparatus (1009934).
At the tip of the probe there is a Hall sensor with an active area of about 0.044 mm², which produces a steady output signal that is rati­ometric (i.e., proportional to the operating volt­age).
The Hall sensor has an extended temperature range of -20°C to +180°C for measuring the flux density of heated germanium semiconductor crystals (probes).
There are two latching range buttons for 0.2 T and 2 T plus an additional Tare button for cali­bration. The current setting of the measure­ment range is shown visually by a light-emitting diode on the left of the relevant button.
The stand clamping rod can be adjusted to hold the sensor in the desired position and
orientation in the magnetic field that is being measured.
The box is also suitable for use with the connec­tor box (115 V, 50/60 Hz) (1009954) or connector box (230 V, 50/60 Hz) (1009955). See the rele­vant technical data.
2. Safety instructions
The magnetic field sensor is not suitable for applications where safety is essential.
Only use the magnetic field sensor for
educational purposes!
3. Equipment supplied
1 sensor box with permanently attached probe 1 miniDIN 8-pin connecting cable, 60 cm 1 clamping rod, length 120 mm 1 instruction sheet
1
4. Technical data
Measurement ranges: 200 mT, 2000 T Configuration: Tangential Sensor type: Linear Hall-effect
sensor
Position of Hall sensor: 135 mm, with refer-
ence to the front of the sensor box casing
Tare range: 200 mT range: +/- 100 mT 2 T range: +/- 1 T
Non-linearity: Max. ±1,5 % of the
total measurement range
Temperature dependence:
() ( )
,KBTB
⎢ ⎣
T
⎛ ⎜
K
= 3000008801300
When used with a connector box 200 mT range: Conversion factor:
125 mT/V, 1.60 V at 200 mT
2 T range: Conversion factor:
1250 mT/V, 1.60 V at 2000 mT
5. Operation
Note:
To avoid permanent damage to the Hall
sensor in the sensor probe tip, do not sub­ject it to any mechanical pressure!
Do not bend the sensor probe tip!
Hold the sensor box by hand in the mag-
netic field to be measured, or use the clamping rod to position it as required in the experimental setup.
Set the tangential orientation of the sensor
element as required and measure the magnetic field.
Insert the sensor probe vertically through
the positioning hole marked "MFS" on the top of the basic Hall-effect apparatus till it comes to rest. The centre of the active sur­face on the sensor will then be within the uniform magnetic field of the electro­magnet and right next to the semiconduc­tor crystal.
Read off the value of the magnetic flux
density from the 3B NETlog
The sensor box is automatically detected by
the 3B NETlog
TM
unit.
Any change in the measuring range is auto-
matically transmitted to the 3B NETlog
TM
display.
TM
unit.
5.1 Zero calibration for the sensor box
When a measuring range has been se-
lected, hold down the Tare button for about 1 s. Zero calibration is carried out auto­matically.
The zero display follows on the 3B NETlog
TM
display row corresponding to the selected sen­sor input.
Under certain circumstances it may be
necessary to carry out zero calibration again between measurements.
Zero calibration should not be carried out
inside the pole pieces of a transformer. The poles may possess some remanence which would need to be taken into account.
6. Experimental applications
Magnetic fields of permanent magnets and coils Hysteresis of transformers Remanence Saturation effects in ferrite cores
7. Sample experiment
Experiment to measure magnetic flux den­sity involving the Hall effect in semiconduc­tors
Equipment needed:
TM
1 3B NETlog
(115 V, 50/60 Hz) 1000539
1 Transformer with Rectifier (115 V, 50/60 Hz)
1003315
1 DC Power Supply 20 V, 5 A (115 V, 50/60 Hz)
1003311
or
1 3B NETlog
TM
(230 V, 50/60 Hz) 1000540
1 Transformer with Rectifier (230 V, 50/60 Hz)
1003316
1 DC Power Supply 20 V, 5 A (230 V, 50/60 Hz)
1003312 1 Hall Effect Basic Apparatus 1009934 1 P-Doped Ge on Circuit Board 1009810
or 1 N-Doped Ge. on Circuit Board 1009760
1 Magnetic Field Sensor ±2000 mT 1009941 1 U Core 1000979 2 Coils D with 600 Taps 1000988 1 Pair of Pole Shoes 1009935 1 Set of 15 Safety Experiment Leads 1002843
Set up the experiment as in Fig. 1.
Switch the transformer and rectifier to the
12-V setting to supply the basic Hall-effect apparatus with 12 V AC.
2
Loading...
+ 2 hidden pages