•Accessories and Supplies List ....................................................................M-8
Index ................................................................................................ N-i
emographics.....................K-3
Humphrey Field Analyzer II-i series User Manual2660021145640 A
vi
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
(1) Introduction & Instrument Setup
Intended Use 1-1
Purpose of This User Manual 1- 4
Symbols and Labels 1- 5
Instrument Disposition 1- 8
Disposal 1- 8
Instrument Installation 1- 9
Tips to Avoid Damage 1- 9
Product Compliance1-1 1
Product Safety1-1 1
Electromagnetic Compatibility (EMC)1-12
Accessory Equipment1-16
About Visual Fields1-17
The Humphrey Advantage1- 20
1-1
System Components1- 23
Additional Components1- 25
System Assembly1- 31
You are about to use the most advanced aut
Analyzer IIII-
i
, including a brief discussion of visual fields and a summary of important instrument features.
After reading this chapter you will be familiar with:
• the importance of visual field testing
• general principles of perimetry
• unique features of the HFA II-
• installation and safety precautions
• connecting the printer and optional external devices
i
series (HFA™ II-
i
). This introductory section covers general information about the HFA
i
mated perimeter available, the Humphrey® Field
o
Intended Use
The Carl Zeiss Meditec, Inc. Humphrey Field Analyzer II-i is an automatic perimeter which is
intended to be used to measure the visual field of the eye.
Indications for Use
The Humphrey Field Analyzer II-i is an automated perimeter intended to identify visual field defects
for the purposes of screening, monitoring and assisting in the diagnosis and management of ocular
diseases such as glaucoma, and related neurological disorders.
Note: These perimetry results are an aid to interpre
is still the most important element in determining the clinical significance of the results, including
considering the limitations of the statistical package.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
tation, not a diagnosis. The doctor’s judgment
1-2
Introduction & Instrument Setup
Essential Performance
The Essential Performance of the instrument is to p
Patient Population
The Humphrey Field Analyzer IIof diagnostic evaluation of the eye. This includes (but is not limited to) patients with the following
disabilities or challenges:
•Wheelchair user
• Very low or not measurable visual acuity
• Postural problems
• Fixation problems
• Deafness
• Large body, but not those above 99th percentile based on anthropomorphic data
There is a general requirement that the patient be able to
in the chin and forehead rest of the instrument (with or without supplemental human or mechanical
support).
Part of the Body
The Humphrey Field Analyzer IIpatient's hand and fingers (or similar ability) are also required to press the Patient Response button.
i
may be used on all adults and children over the age of six in need
i
physically interacts with the patient’s forehead and chin. The
rovide accurate visual field measurements.
sit upright and be able to place their face
Application
The Humphrey Field Analyzer IIsites operate the instrument for 10 hours or less per day, indoors, within a medical office or hospital
setting. This setting shall have clean air free of soot, vapors from adhesives, grease, or volatile
organic chemicals. Other Operating Environment specifications are given in Appendix (A), "Product
Specifications,". Application related warnings are given in Chapter (1), "Introduction & Instrument
Setup," and elsewhere.
User Profile
We assume that users are clinicians with pr
ophthalmic equipment, and in diagnostic interpretation of the test results. Specific assumptions
regarding the profiles of individuals performing instrument operation or data interpretation are
given below. This manual contains information that will aid in the proper instrument operation and
interpretation of the resultant data.
Instrument Operation
Demographic
The user should be adult, and at least one of the following:
• Ophthalmologist
•Optometrist
•Nurse
i
is designed for continuous use, although it is expected that most
ofessional training or experience in the use of
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
• Certified Medical Technician
• Ophthalmic Photographer
• Non-certified Assistant
Occupational Skills
The user should be able to perform all of the following tasks:
• Power on the instrument
• Enter, find, and modify patient identifying data
• Clean surfaces that contact patient
• Position patient with the instrument, including moving the pat
height, and the patient’s chair
• Select and initiate a test
• Review and save a test or try again
• Generate an analysis report
• Review the analysis report for completeness
• Save, print, or export an analysis report
•Archive data
• Power off the instrument
1-3
ient, the instrument, the table
Data Interpretation
Demographic
The user should be one of the following:
• Ophthalmologist or other Medical Doctor
• Optometrist or equivalent
Occupational Skills
The user should have the following skills:
• See Instrument Operation above
• Ability to work with elderly patien
Job Requirements
The user should have training and certification in the analysis and tr
or other eye-related medical issues as required by governing bodies.
ts and those with disabilities
eatment of ophthalmic diseases
Humphrey Field Analyzer II-i series User Manual2660021145640 A
1-4
Introduction & Instrument Setup
Purpose of This User Manual
Carl Zeiss Meditec designed this User Manual to serve as a training, usage and reference guide.
While we offer training in the use of the HFA IIinterpretation. This manual does not attempt to do so.
To fully appreciate the capabilities of the HFA IIrecommend that you rely on this User Manual as your training and reference guide. It has been
designed to make learning easy. The concise step-by-step instructions and accompanying
illustrations help you get started quickly and with more confidence.
We think you will enjoy working with the HFA IIlearn and easy to operate. For optimum results:
• Read your User Manual in the order written.
• Read it while sitting at the instrument.
• Practice using the HFA II-
i
by first testing staff members, before using it with patients.
i
, we do not offer instruction in diagnostic
i
and to develop good testing techniques, we
i
. The friendly touch control makes it inviting to
Note: HFA IIaltering the button locations, functions, and text from the prior software version. This manual
depicts the screens in the prior software style. Regardless of style, text and buttons on screen
images depicted in this manual are identical to those in the Ver. 5.x software screen images.
Model Differentiation
This guide contains instructions for Models
information is relevant to all models, some information applies only to particular models. When a
feature or function applies only to specific models, this guide specifies the model number(s), often in
parentheses, in a prominent location. An example of this is found in the discussion of “SWAP
(Blue-Yellow) Testing (Models 745i and 750i),” on page 1- 2 1. Conversely, model numbers are not
specified when information is standard or optional on all models. Y
your instrument on the rear panel of the HFA IIbutton located in the upper, left-hand corner of the screen (see “The Information Button,” on
page 2-3). Refer to Appendix (B), "Product Features,", if you ar
capabilities of your instrument.
Text Conventions
The terms “select,” “choose,” “touch,” and “press” a
initiate an operator action using the touch screen, external keyboard, glidepad, trackball, or mouse.
The terms “hard disk” and “hard drive” are used interchangeably, in reference to the data storage
device standard on all HFA II-
i
series system software Ver. 5.x has updated the style of the user interface without
720i, 740i, 745i and 750i. Although much of the
ou can find the model number of
i
or you may access this information via the “i”
e unsure about the particular
re used interchangeably. Each term means to
i
models.
UPPER CASE LETTERS are reserved for references to specific command buttons found on the touch
een. The exceptions to this are messages on test printouts, the words STATPAC, SITA™, SWAP,
scr
HFA II-
i
, and headings.
Italicized words
figures, pictures, tables, and special notes in this manual.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
are used to identify the icon buttons on the right border of the screen, the titles of
Introduction & Instrument Setup
Bold words are used to highlight warnings and section headings.
1-5
This manual means “left-click” when it says, “click,” except
Chains of menu items are indicated with the use of the “>”
“
File>Exit
Access Menu Options
To access the options offered through each menu, click on the menu headings. Then click on an
op
Electronic User Manual Access
The HFA IIDocumentation CD included in the instrument accessory kit. If you do not have Adobe Reader
installed, go to www.adobe.com to download and install the free Adobe Reader.
Additional References
The User Manual cannot possibly cover every situatio
especially interpretation questions. Your HFA IIprovides an overview of visual field results.
R. Anderson and Vincent Michael Patella (Mosby, Inc., St. Louis), is recommended for in-depth
information and analysis of visual fields.
” directs you to select Exit in the File menu.
tion to select it. Click outside all menu options to make the options disappear.
• Some menus are fields tagged with a down-arrow (
options, click on the down-arrow.
• Grayed-out menu options or
i
User Manual in Acrobat PDF format for use on a computer is on the HFA II-i User
buttons are not available.
i
comes with a copy of
Automated Static Perimetry, Second Edition,
where “right-click” is specified.
symbol between items. For example,
drop-down lists). To access these menu
n you may encounter with the HFA II-i,
Essential Perimetry
, which
by Douglas
Symbols and Labels
WARNING
CAUTION
Must follow Instruction for Use
Type B applied parts
Alternating Current
Fuse
Humphrey Field Analyzer II-i series User Manual2660021145640 A
1-6
Introduction & Instrument Setup
CAUTION: Hot Surface
Manufacturer
Date of Manufacture
Authorized European Community Representative
Serial number
Catalog number / part number
Model number
Patent
System Software USB Flash Drive
Calibration Software USB Flash Drive
GPA Sample Data USB Flash Drive
European Conformity
Disposal of the Product within the EU. Do not dispose via domestic waste disposal system or
communal waste disposal facility.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
Power On
Power Off
Projector Lamp
Brightness
Air Intake Filter
Serial RS-232
Communication Port
VGA Video Port
Printer
Keyboard/Mouse
USB Port
Network Port
Patient Response
Button
Port
DO NOT USE
Additional symbols appearing on the HFA II-i:
1-7
Figure 1.1 : Additional HFA II-i Symbol Definitions
Humphrey Field Analyzer II-i series User Manual2660021145640 A
1-8
Introduction & Instrument Setup
Protective Packing Symbols
The protective packing symbols on the shipping carton specify the handling requirements and the
tran
sport and storage conditions for the HFA IIsymbols in the event that your HFA IIuse.
i
must be stored for a period of time, prior to its set up and
Handling RequirementsTransport and Storage Conditions
i
as it is shipped from the factory. Note these
Fragile, Handle with Care
Keep Dry
This end upAtmospheric Pressure Limits (500 hPa
Relative Humidity (10% to 100%, including
condensation
Tem
per atur e (-40 to +70 deg. C)
to 1060 hPa)
Instrument Disposition
When it comes time to upgrade the HFA, please contact Carl Zeiss Meditec to inquire about trade-in
or upgrade values we may offer. Should you not wish to trade in the instrument, please see the
Disposal section below.
Disposal
CAUTION: This product contains electronic components. At the end of its lifetime, the
product should be disposed of in accordance with the relevant national regulations.
Disposal of the Product within the European Union (EU)
In accordance with applicable EU guidelines at the time at
market, the product specified on the consignment note is not to be disposed of via the domestic
waste disposal system or communal waste disposal facilities.
For further information on disposal of this product, please cont
manufacturer or its legal successor company. Please read the latest internet information provided by
the manufacturer.
Where the product or its components are resold, the seller must inform the buyer that the product
must be disposed of in accor
Humphrey Field Analyzer II-i series User Manual2660021145640 A
dance with the currently applicable national regulations.
which the pr
oduct was brought onto the
act your local dealer or the
Introduction & Instrument Setup
User Changes to Software or Hardware
The HFA II-i is a medical device. The software and hardware have been designed in accordance with
U.S., European and other international medical device standards designed to protect clinicians,
users and patients from potential harm caused by mechanical, diagnostic or therapeutic failures.
1-9
WARNING: Unauthorized modification of HF
peripherals) can jeopardize the safety of operators and patients, the performance of
the instrument, and the integrity of patient data; it also voids the instrument
warranty.
Approved Software
Use of software supplied or approved by Carl Zeiss Meditec for the HFA IIcurrent list of approved software call Carl Zeiss Meditec Customer Care: In the U.S., call
800-341-6968. Outside the U.S., contact your local Carl Zeiss Meditec distributor.
Note: Carl Zeiss Meditec does not provide technical support for the use of unappr
software.
A II-i software or hardware (including
i
is authorized. For the
oved third party
Instrument Installation
Only an authorized Carl Zeiss Meditec service representative should install the
with the buyer, Carl Zeiss Meditec schedules one free on-site installation appointment after
instrument delivery. System installation and operator training require approximately one-half
business day.
Care in Handling
Use extreme care when handling and transporting the HFA IIcontains fragile optics that have been precisely aligned at the factory.
i
shipping boxes. The instrument
HFA II-i. In consultation
Installation Requirements
• The HFA IIconfigure your HFA II-
• An isolation transformer is required when connecting peripher
Device approved (i.e., printer, USB drive, etc.) within 1.5 meters (4.9 feet) away from the
patient, such that the patient cannot touch a peripheral device with any part of his or her body
while being examined.
i
should operate on a dedicated power outlet. Based on your specification, we
i
at the factory to use either 100V, 115V, or 230V line voltage.
al devices that are not Medical
Tips to Avoid Damage
Note: Users are not authorized to dismantle or modify the HFA II-i hardware. To transport the
instrument outside the office, you must consult with a Carl Zeiss Meditec service technician. Failure
i
to do so voids all warranties offered with the HFA II-
• Only Carl Z
In the case of malfunction, error messages or operational problems, call Carl Zeiss Meditec
Customer Care: In the U.S., call 800-341-6968. Outside the U.S., contact your local Carl Zeiss
Meditec distributor.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
eiss Meditec authorized technicians should disassemble or service this instrument.
.
1-10
Introduction & Instrument Setup
• This instrument has no special measures to protect against harmful ingress of water or other
liquids (classified IPXO—ordinary equipment). Do not place containers of liquid on or near the
instrument, nor use aerosols on or near it.
• In case of emergency related to the instrument, unplug the power cor
call for service immediately.
• With the exception of the main power fuses and k
in the instrument. For the replacement of any component, accessory, or peripheral, except
fuses or the keyboard, call Carl Zeiss Meditec Customer Care: In the U.S., call 800-341-6968.
Outside the U.S., contact your local Carl Zeiss Meditec distributor.
• Although this instrument is designed for con
not in use for an extended period.
• This instrument operates according to specifications under standar
lighting conditions, without exposure to any direct sunlight.
eyboard, there are no user-replaceable parts
tinuous operation, it should be turned off when
HFA II-i Embedded License
Each HFA II-i is issued with an embedded VxWorks® operating system license.
d from the wall outlet and
d indoor office (fluorescent)
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
0297
Product Compliance
Complies with 93/42/EEC Medical Device Directive.
Complies with US and Canadian medical electrical system safety requirements.
Product Safety
This instrument is classified as follows:
• Class I Equipment
•
Type B – Degree of protection against electric shock of applied part (chinand forehead rests,
and patient button).
• Ordinary Equipment (IPX0)–
• Continuous Oper
General Safety Requirements
• Although the HFA II-i is designed for continuous operation, it should be turned off when not
used for an extended period of time. The HFA II-
• Use the instrument cover to protect the HFA II-i at all times when it is not in use.
• Do NOT place the cover over the instrument when the HFA IIairflow can cause overheating and damage to sensitive components.
• Do NOT connect or disconnect cables while power is on.
• Do NOT place any objects on top of the instrument.
• Do NOT place any container holding liquid near the instrument.
• Use only a stand or table recommended by Carl Zeiss Meditec.
– Protection against electrical shock.
Degree of protection against ingress of liquids (none).
ation – Mode of operation.
i
should be used in a cool, dry dust-free setting.
i
is turned on, as loss of proper
1-11
Warnings and Cautions
w
WARNING: Do NOT block the ventilation openings. These allo
heat generated during operation. A buildup of heat due to ventilation opening
blockage can cause failures which may result in a fire hazard.
WARNING: To prevent electric shock, the instrument must be plugged into an earthed
ground outlet. Do not remove or disable the ground pin. Only an authorized Carl
Zeiss Meditec service representative may install the instrument.
WARNING: Do not use the printer or the instrument or the optional power table with
an extension cord or a power strip (multiple portable socket outlet). For additional
safety, do not plug the printer and the instrument (or the optional power table) into
the same wall outlet. Failure to observe this warning could result in electrical shock to
the patient and/or examiner.
WARNING: Do not open the instrument covers. Opening the instrument covers could
expose you to electrical and optical hazards.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
for the release of
1-12
Appliance Coupler
(input power plug)
HFA Instrument
Introduction & Instrument Setup
WARNING: If the instrument is externally connected to non-medical peripheral
devices (i.e. printer, storage devices, etc.), the complete system must comply with the
system requirements in standard IEC 60601-1. This standard requires the usage of an
Isolation Transformer to power the non-medical peripheral device(s) if located within
1.5 m from the patient. If the peripheral device is located outside the patient
environment (beyond 1.5 m) and is connected to the HFA, a separation device must
be used or there shall be no metal to metal connection between the non-medical
peripheral device and the HFA.
The person or the responsible organization connecting additional devices or
reconfiguring the system must evaluate the complete system to ensure compliance to
the applicable IEC 60601-1 requirements. The instrument operator must not attempt
to touch the patient and the peripheral device simultaneously.
WARNING: Do not reconfigure system components on the table, nor add non-system
devices or components to the table, nor replace original system components with
substitutes not approved by Carl Zeiss Meditec. Such actions could result in failure of
the table height adjustment mechanism, instability of the table, tipping and damage
to the instrument, and injury to operator and patient.
CAUTION: The appliance coupler (shown on the left) is the main disconnect device of
the instrument. Position the instrument in such a way to have easy access to
disconnect the appliance coupler in case of an emergency.
CAUTION: In case of an emergency, disconnect the appliance coupler.
WARNING: This instrument may cause ignition of flammable gases or vapors. Do NOT
use in the presence of flammable anesthetics such as nitrous oxide, or in the presence
of pure oxygen.
WARNING: Avoid tipping. Do not use the instrument on an uneven or sloped surface.
Do not roll the table in deep pile carpet or over objects on the floor such as power
cords. Failure to observe these precautions could result in tipping of the instrument
and/or table and resulting injury to operator or patient and damage to the
instrument.
Electromagnetic Compatibility (EMC)
Note: The HFA II-i needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided herein.
Note: Portable and mobile RF communications equipment can affect medical electrical equipment.
WARNING: The use of accessories, transducers and cables other than those specified
may result in increased emissions or decreased immunity of the equipment.
WARNING: The HFA II-i should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the equipment or system should
be observed to verify normal operation in the configuration in which it will be used.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
Guidance and manufacturer’s declaration - electromagnetic emissions
The HFA II-i is intended for use in the electromagnetic environment specified below. The customer or user of the
HFA II-i should ensure that it is used in such an environment
The HFA II-i uses R
its internal function. Therefore, its
RF emissions are very low and are
not likely to cause any interference
in nearby electronic equipment.
The HFA II-i is suit
establishments, other than domestic
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
F energy only for
able for
use in all
Humphrey Field Analyzer II-i series User Manual2660021145640 A
1-14
Introduction & Instrument Setup
Guidance and manufacturer’s declaration - electromagnetic immunity
The HFA II-i is intended for use in the electromagnetic environment specified below. The customer or user of the
HFA II-i should ensure that it is used in such an environment
Immunity TestIEC 60601 test levelCompliance levelElectromagnetic environment - guidance
Electrostatic
Discharg
e (ESD)
± 6 kV contact
± 8 kV air
IEC 61000-4-2
Electrical fast
ient/burst
trans
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
Surge IEC 61000-4-5± 1 kV differential
mode
V common
± 2 k
mode
Voltage dips, short
interruptions, and
oltage variations on
v
power supply input
lines. IEC 61000-4-11
Power Frequency
Hz)
0/60
(5
<5% U
40% U
70% U
<5% U
(>95% dip in
T
) for 0.5 cycle
U
T
(60% dip in
T
) for 5 cycles
U
T
(30% dip in
T
) for 25 cycles
U
T
(95% dip in
T
U
) for 5 sec.
T
3 A/m3 A/mPower frequency magnetic fields should
magnetic field
IEC 61000-4-8
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
V for
± 1 k
input/output lines
± 1 kV differential
mode
V common
± 2 k
mode
<5% U
40% U
70% U
<5% U
(>95% dip in
T
) for 0.5 cycle
U
T
(60% dip in
T
) for 5 cycles
U
T
(30% dip in
T
) for 25 cycles
U
T
(95% dip in
T
U
) for 5 sec.
T
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
cal commercial or hospital
typi
environment.
Mains power quality should be that of a
cal commercial or hospital
typi
environment.
Mains power quality should be that of a
typical com
mercial or hospital
environment. If the user of the HFA II-i
requires continued operation during
power mains interruptions, it is
recommended that the HFA II-i be
powered from an uninterruptible source.
be at levels char
acteristic of a typical
location in a typical commercial or
hospital envrionment.
Note: U
is the a.c. mains voltage prior to application of the test level.
T
Humphrey Field Analyzer II-i series User Manual2660021145640 A
d1.17 P=
d1.17 P=
d2.33 P=
Introduction & Instrument Setup
Guidance and manufacturer’s declaration - electromagnetic immunity
The HFA II-i is intended for use in the electromagnetic environment specified below. The customer or user of the
HFA II-i should ensure that it is used in such an environment
Immunity TestIEC 60601 test levelCompliance levelElectromagnetic environment - guidance
Portable and mobile RF communications
equipmen
any part of the HFA II-i, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
t should be used no closer to
1-15
Conducted RF
IEC 6
1000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 8 0 M H z
3 V/m
80 MHz to 2,5 GHz
3 V
3 V/m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
where P is the maximum output power
rat
ing of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF transmitters,
y
as determined b
a
survey,
should be less than the
an electromagnetic site
compliance level in each frequency
b
range.
Interference may occur in the
vicinity of equipment marked with the
following symbol:
Note 1: At 80 MHz and 800 MHz, the higher frequency applies.
Note 2: These guidelines may not apply in all situations. Electromag
netic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM broadcast, cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the HFA II-i is used exceeds the applicable RF
compliance level above, the HFA II-i should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the HFA II-i.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
P
P
P
1-16
Introduction & Instrument Setup
Recommended separation distances between portable and mobile RF communications equipment and the
-i
HFA II
The HFA II-i is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the HFA II-i can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the HFA II-i
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmit
For transmitters rated at a maximum output power not listed above, the r
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency applies.
Note 2: These guidelines may not apply in all situations. Electr
and reflection from structures, objects and people.
ter
150 kHz to 80 MHz80 MHz to 800 MHz800 MHz to 2,5 GHz
W
0,010.1170.1170.233
0,10.3700.3700.737
11.1701.1702.330
103.7003.7007. 368
10011.70011.70023.300
Separation distance according to frequency of transmitter
m
d = 1.17
d = 1.17
e
commended separation distance d in
omag
netic propagation is affected by absorption
d = 2.33
Accessory Equipment
WARNING: Accessory equipment connected to the analog and digital interfaces must
be certified according to the respective IEC standards (e.g., IEC 60950 for data
processing equipment and IEC 60601-1 for medical equipment). Furthermore, all
configurations shall continue to comply with the applicable system requirements of
IEC 60601-1 standard. Any person who connects additional equipment to the signal
input part or signal output part configures a medical system, and is therefore
responsible for ensuring that the complete system complies with the applicable
system requirement of IEC 60601-1 standard. If in doubt, consult the technical service
department or your local representative.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
About Visual Fields
When asked to assess one’s own vision, the average person often will confidently reply, “I see
20/20”, “20/100” or whatever the result of their visual acuity test. Fortunately, doctors appreciate
the complexities involved in evaluating visual function and rely on an extensive and varied battery of
diagnostic tests and instruments as part of the ocular examination. Without question, one of the
most essential tools in the modern ophthalmic office is the automated perimeter, used to evaluate
the visual field.
1-17
The purpose of visual field testing, or perimet
• diagnosing ocular diseases, especiall
• evaluating neurological diseases
• monitoring the progress of ocular and neurological diseases
Visual field testing can lead to early detection and tr
visual fields play a major role in identifying visual field defects and evaluating the efficacy of the
therapy used to control the disease process.
What Visual Field Tests Measure
When evaluating visual performance, clinicians ar
resolution and contrast sensitivity. Resolution is the ability to identify discrete forms (letters,
numbers, symbols), and is commonly measured with the visual acuity test. Resolution rapidly
diminishes with increasing distance from the fovea and is, therefore, a poor indicator of overall
visual performance.
A better means of evaluating visual function—especially
is contrast sensitivity testing. Contrast sensitivity is the ability to detect a stimulus (spot of light or
other target) against a darker or brighter background. Humphrey perimetry may be thought of as
contrast sensitivity testing applied throughout the peripheral visual field.
In perimetry, the term “threshold” is used to describe
threshold represents the point at which a stimulus is seen 50% of the time and missed 50% of the
time. The assumption is that all stimuli brighter than the threshold value will be seen and all stimuli
dimmer will be missed. Reviewing the threshold value at each point tested in the visual field is an
important part of the diagnostic process.
ry, is to provide information critical to:
y glaucoma
eatment of disease. In the case of glaucoma,
e primarily interested in two retinal functions:
those areas less sensitive than the fovea—
a very specific level of stimulus detection. The
Visual field tests can yield information that is gener
exacting and quantitative, as with threshold tests. In deciding which test type is most appropriate
for a patient, the practitioner is influenced by many factors, including the patient’s presenting
complaint, family history, age, degree of cooperation, and time available to run the test.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
al in nature, as with screening tests, or more
1-18
60º
90º
60º
70º
Superior
Inferior
Nasal
Tem po ra l
Blind Spot
Fovea
Introduction & Instrument Setup
Normal Versus Pathologic Fields
The visual field normally extends more than 90
70
º inferiorly. That means a person can potentially perceive stimuli within this range while staring at
º temporally, 60º nasally and superiorly, and about
a fixed point.
Figure 1.2 The Boundaries of the Normal Visual Field
A more comprehensive understanding of the normal field takes into account that visual sensitivity is
not constant (or equal) throughout the range. As previously stated, vision is most acute at the fovea
and decreases toward the periphery of the retina. It is easy to see why the visual field is often
expressed as a “hill of vision in a sea of darkness”.
Figure 1.3 The Normal Hill of Vision
Several factors affect the normal hill of vision, causing variations in its overall height and shape.
Among them are a patient’s age, ambient light, stimulus size, and stimulus duration. In general,
deviations from the normal hill of vision are viewed as visual field defects and caused by some
pathological change.
A visual field defect, or scotoma, is categorized as either re
area that has depressed vision or less than normal sensitivity; an absolute defect is an area where
the perception of light is absent. The point at which the optic nerve enters the retina is referred to as
lative or absolute. A relative defect is an
the blind spot, and is an example of an absolute scotoma.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Introduction & Instrument Setup
Some defect patterns are characteristic of certain diseases, a fact which makes visual field testing a
valuable part of the diagnostic process. Furthermore, by having patients repeat the same tests at
later dates, practitioners gain insight into the progression of the disease and the effectiveness of
treatment.
Methods of Testing the Visual Field
1-19
Over the years, visual field testing devices have varied in size, complexity
The fundamental premise has remained the same, however; patients must respond when they see a
stimulus.
Static threshold testing evaluates retinal function. The te
being used. In static testing, predefined test locations in the visual field are probed. Through a series
of stimulus presentations of varying brightness intensities, the threshold value is determined for
each test point. When evaluating static test results, clinicians are looking at the topography or
contour of the hill of vision, and whether depressions are evident.
In a second type of retinal evaluation, called kinetic testing, a
is moved into the visual field from a non-seeing area, until it is detected by the patient. Typically, the
stimulus is brought toward the center from several directions and the operator marks the location at
which the patient first detects the stimulus (threshold point).
Kinetic test results can only be reliably related to specific parts of the visual field if points are joined
to form an isopter
used during one kinetic test, and for each different target, a different isopter is mapped. When
reviewing several isopters, the clinician is visualizing different tiers in the hill of vision.
Patient Fixation and Test Reliability
In order for any visual field test to be useful clinically
factor affecting reliability is the steadiness of patient fixation. Unless the eye being tested fixates
accurately on the target while responding to stimuli, the results are unreliable.
, or ring of equal contrast sensitivity. Targets of varying size and brightness are
rm “static” refers to a stationary stimulus
light stimulus of fixed characteristics
, it must yield reliable results. One important
, and testing methodology.
Other factors adversely affecting reliability are:
• patient fatigue and anxiety
• poor test instructions
• patient discomfort
• improper near vision correction for central testing
Benefits of Computerized Perimetry
Certainly the advancements in microprocessor technology within the last 20 years have
profound effect on perimetry. Perimeters have evolved into a more precise measuring tool yielding
highly repeatable results.
These changes are better appreciated by examining t
both patient and professional:
• Reproducible testing conditions
• Data storage capability; results can be compared over time and
software
Humphrey Field Analyzer II-i series User Manual2660021145640 A
he benefits computerized perimeters bring to
analyzed using expert-system
had a
1-20
Introduction & Instrument Setup
• More sensitive testing; advances in algorithm development has made static perimetry superior
to the kinetic method for identifying defects
• Ease of operation; menu-driven software makes au
The Humphrey Advantage
Over 25 years of advancements in research, design and development are reflected in the Humphrey
Field Analyzer IIfrom around the world who generously have shared their best ideas with Carl Zeiss Meditec. With
tens of thousands of Humphrey Field Analyzers in use worldwide, Carl Zeiss Meditec took on the
challenge of improving the testing experience for the patient, the operator, and the practitioner.
Here are some of the features which differentiate the HFA IIavailable.
Ergonomic Design
The HFA II-i relieves many physical discomforts associated with visual field testing. The chin rest and
bowl shape allow patients to assume a more natural and re
The power table improves patient comfort by permitting the HFA IIof the patient adjusting to the instrument. This especially is important for wheelchair bound patients.
i
. Equally important, the latest models represent improvements suggested by users
tomated perimeters easy to learn and use
i
from other autoperimeters that are
laxed sitting position when taking tests.
i
to adjust to the patient instead
The patient response button is easy to operate, especially for patients who have limited use of their
hands; for instance, patients with arthritis. The uniquely-shaped button may be placed on a knee,
lap or the arm of a chair for better leverage. The cord angles away from the patient for greater
comfort. The response button beeps each time it is pressed to give immediate feedback to the
patient and to the operator.
Easy Operation
Sophisticated instrumentation need not be complicated. The HFA IIintended to make the instrument easier to use:
• Touch screen design speeds data input.
• Menu and icon commands simplify operation.
• On-screen video eye monitor is standard
• Confirmation screens reduce unintentional data loss.
• A keyboard, glidepad, trackball or mouse can be connected to the HFA IIinput devices.
• An optional keyboard with built-in glidepad is available for HFA models that were not delivered
with one.
Rapid Testing
Carl Zeiss Meditec’s SITA (Swedish Interactive Thre
precise visual field measurements with unprecedented speed. SITA is a rapid, reliable,
state-of-the-art autoperimetric technology that is available only with the Humphrey Field Analyzer.
With the SITA strategies, users can obtain visual field information in half the time it takes using
conventional testing algorithms without compromising accuracy. For further explanation of the SITA
testing strategy, refer to Appendix (K).
on all models.
sholding Algorithm) testing strategies allow
i
offers a number of features
i
as optional data
Humphrey Field Analyzer II-i series User Manual2660021145640 A
Sophisticated Data
Introduction & Instrument Setup
1-21
The Humphrey Field Analyzer’s statistical software, ST
visual field test results. With STATPAC, you can analyze test results at the time of examination, store
test results and analyze them at your convenience, or recall previously stored tests to analyze for
comparative purposes.
STATPAC includes several exclusive features
• Using results from a single test, STATPAC can poin
not be evident until subsequent tests were done.
• STATPAC can identify areas that look suspicious
normals data.
• Using results from a series of tests, STATPAC provides a
of changes in the patient’s visual field over time.
• The Glaucoma Hemifield Test (GHT) compares points in the
provide a plain language analysis of test results.
• The HFA IIanalysis. These include databases for SITA and
databases for Full Threshold and FastPac™ test results.
• Another database consisting of stab
Analysis (GPA™) for following change in the progress of the disease. Refer to Chapter (8),
"Guided Progression Analysis (GPA)," for further details.
i
provides separate, clinically validated, age-normative databases for STATPAC
to help you identify visual field change:
le glaucoma patients is used with the Guided Progression
ATPAC, provides immediate expert analysis of
t out suspicious areas that otherwise might
but which, in fact, compare favorably with
highly sensitive and informative analysis
superior and inferior hemifields to
SITA-SWAP™, in addition to the original
SWAP (Blue-Yellow) Testing (M
Blue-Yellow perimetry is also known as Short-Wavelength Automated P
with Models 745
early detection of glaucomatous changes, according to published longitudinal studies.
Blue-Yellow perimetry differs from standar
chosen wavelength of blue light is used as the stimulus, and a specific color and brightness of
yellow light is used for the background illumination. The ability to use SITA for SWAP testing greatly
reduces the time involved. See Chapter (9), "Short-Wavelength Automated Perimetry (SWAP)," for
more information on SWAP and SITA-SWAP.
Automatic Fixation Monitoring
The HFA IItarget during testing. All models are equipped with a video eye monitor which presents a view of
the patient’s eye on-screen so that users can ensure proper patient fixation. Every HFA IIstandard Heijl-Krakau blind spot monitoring.
Models 740
real-time image analysis to verify the patient is looking at the fixation target and not looking
around. The Gaze Tracking device is unaffected by the patient’s head position. A continuous record
of fixation is available on the test screen for monitoring during the test. The Gaze Track graph is
included on the printout to provide a permanent record of the patient’s fixation.
i
and 750
i
employs several methods for ensuring that patients maintain proper fixation of the
i
, 745i and 750i also offer Gaze Tracking: a patented, high precision system which uses
i
). It has performed better than standard computerized perimetry for the
odels 745i and 750i)
erimetry, or SWAP, (available
d static white-on-white perimetry only in that a carefully
i
also offers
Humphrey Field Analyzer II-i series User Manual2660021145640 A
1-22
Introduction & Instrument Setup
For patients who require a trial lens, the Model 750i uses Head Tracking and Vertex Monitoring to
help ensure that the patient’s eye is both centered b
from the lens. These features help to eliminate the trial lens as a possible sour
field results.
Data Protection Features
ehind the lens and is held at the proper distance
ce of unreliable visual
Visual field results need to be saved a
of data storage methods to file the results. Five USB ports are provided for connection to USB
devices, two of which may be used simultaneously with USB storage devices. Floppy disk data
storage is available with an optional USB floppy disk drive. Using HFA-NET Pro, you can either
archive or back up vital test results to a network file server in your office. There are a number of
additional data protection features that work internally to safeguard your data from loss or serious
damage. This manual describes in great detail the procedures for creating extra copies of your data.
Refer to Chapter (11), "Database Management," for further information on database security. Refer
to Chapter (14), "Networking," for further information about connecting your HFA IIoffice computer network.
Risks of Internet Connectivity
CAUTION: When connected to the Internet, th
serious security risks, including viruses and worms that could disable your system or
adversely affect its performance. Internet connectivity enables third party software
drivers and updates to be downloaded to your system, either automatically or
intentionally. Installation of any unapproved software, including drivers, could
degrade the performance of the instrument and/or lead to corrupted diagnostics or
therapeutic information and may void the instrument warranty.
Networking Features
nd protected for future use. The HFA II-i offers you a number
i
into your
e HFA instrument may be vulnerable to
HFA II-
i
series perimeters offer many useful networking capabilities for patient data, test data, and
image files. These include:
• export patient data, test results, and image files (PDF and
• synchronize databases on two or more HFA II-
• back up patient data and test results to a
• restore patient data and test results from a
• import work lists from non-DICOM and DICOM EMR/PMS systems
• import patient information and exam data from a DICOM system (DICOM Gateway 2.0 only)
• export patient data, image files, and test r
• export raw exam data and reports to a DICOM system (DICOM Gateway 2.0 only)
Information on the Internet
New information about your HFA IIaddress is: www.meditec.zeiss.com/hfa.
Humphrey Field Analyzer II-i series User Manual2660021145640 A
i
may be found on the Carl Zeiss Meditec web site. The internet
i
perimeters via archiving and retrieval
file server for safe external storage
network file server to an HFA II-
esults to non-DICOM and DICOM EMR/PMS systems
TIFF format) to a file server
i
Loading...
+ 550 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.