Zeiss CenterMax Operating Instructions

CenterMax

Coordinate measuring machine for series production
Operating Instructions

Read this first! • Please read these operating instructions, before starting up the

measuring machine.
We reserve the right to make changes in the version and scope of delivery of the basic CMM and its options, the program packages and their respective documents.
The passing on or duplication of this document as well as the use or disclosure of its contents are prohibited unless express permission is granted. Violations will be prosecuted.
All rights reserved, especially in connection with the granting of pat­ents or the registration of a utility model.
We reserve the right to revise this manual and perform technical mod­ifications of the CMM and its components.
All product names are registered trademarks or trademarks belonging to the respective proprietors.
Carl Zeiss CenterMax: Operating Instructions Industrial Metrology Revision status: 1.1 Business Group Date of issue: 01/02 D-73446 Oberkochen Order No.: 61211-1020202

Contents

Chapter 1 Introduction
Foreword
Information about these operation instructions . . . . . . . . . . . . . . . 1
Symbols / danger warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
For your orientation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
CE marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Safety of the machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Standards and regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Additional documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Separate documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Supplementary literature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Proper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Requirements for person responsible for CMM . . . . . . . . . . . . . . . 1-9
Definition of a specialist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
Safety symbols on the CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-10
Requirements for the safe use of the CMM . . . . . . . . . . . . . . . . 1-10
Safety devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-11
Chapter 2 Description of the CMM
Identification of a CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Design of the CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
61211-1020202 Operating Instructions Contents1
Components and their functions . . . . . . . . . . . . . . . . . . . . 2-4
Bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Measuring table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Coordinate axes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Control cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Operating elements of the control cabinet . . . . . . . . . . . . . . . . 2-8
Control console . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Operation of the joysticks. . . . . . . . . . . . . . . . . . . . . . . . . . . .2-10
Calibration standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Probe system types . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Components of a probe system . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Functions of the components . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Standard version of the probe system. . . . . . . . . . . . . . . . . . .2-15
VAST probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
limiting values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Temperature-sensing probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Probe components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19
Design and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19
Probe elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-21
Principles for assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-22
Screwing technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-23
Clamping technique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
Storage of probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-24
Probe rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Chapter 3 Technical data
Coordinate measuring machine . . . . . . . . . . . . . . . . . . . . . 3-2
General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
CMM parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
61211-1020202 Operating InstructionsContents 2
Connection data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Travel speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Environmental conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
VAST probe head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Chapter 4 Transport and installation
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
To be observed on delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Transport conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Selecting the installation site . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Prerequisites for installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Chapter 5 Preparations for start-up
Before you start! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Connecting the power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Connecting the compressed-air supply . . . . . . . . . . . . . . . . . . . . . 5-3
Requirements for compressed-air supply . . . . . . . . . . . . . . . . . 5-4
Compressed-air connection to the CMM . . . . . . . . . . . . . . . . . 5-4
Pressure gages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
Adjusting the pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Filter units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
Mounting the probe rack (option) . . . . . . . . . . . . . . . . . . . 5-8
Prerequisites for the probe rack . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Installation procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Visual check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Start-up preparation checklist . . . . . . . . . . . . . . . . . . . . . 5-15
61211-1020202 Operating Instructions Contents3
Chapter 6 Start-up
Clamping the workpiece on the measuring table . . . . . . 6-2
Safety measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Clamping the workpiece . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Workpiece positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Lowering the workpiece onto the measuring table . . . . . . . . . . . . 6-4
Switching the CMM on . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Safety measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Switch-on sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
Checking the protective circuit . . . . . . . . . . . . . . . . . . . . . . 6-8
Function of the protective circuit . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
Mounting / changing the probe head . . . . . . . . . . . . . . . 6-11
Probe assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Screwing technique or clamping technique? . . . . . . . . . . . . . . . .6-12
Safety measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Criteria and limiting values for assembling probes . . . . . . . . . . . 6-13
Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
Inserting / removing the probe . . . . . . . . . . . . . . . . . . . . 6-16
Safety measures and information . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Inserting the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
Removing the probe (manually) . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Start-up checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Chapter 7 Measuring operation
What you should know! . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Coordinate systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Probing directions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Types of measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Single point probing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Multipoint measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Scanning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
61211-1020202 Operating InstructionsContents 4
Collision protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Axis clamping with VAST probe system . . . . . . . . . . . . . . . . . . . . 7-6
All axes unclamped . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Clamped axes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Self-centering probing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
Temperature-sensing probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Notes on temperature measurement . . . . . . . . . . . . . . . . . . . .7-9
Possibilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Operating mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
Measuring software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Preparations for a measuring run . . . . . . . . . . . . . . . . . . 7-14
Prerequisites for a perfect measuring run . . . . . . . . . . . . . . . . . .7-14
Reference point travel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Reference point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-14
Carrying out a reference point travel . . . . . . . . . . . . . . . . . . .7-16
Probe calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
What you should know!. . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-17
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-17
Calibration procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
Causes of large deviations . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-21
Temperature compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-22
What is temperature compensation? . . . . . . . . . . . . . . . . . . . 7-22
Connecting the temperature sensors . . . . . . . . . . . . . . . . . . . 7-22
Probing the workpiece . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
How to ensure correct measurement . . . . . . . . . . . . . . . . . . . . . 7-24
Probing procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-24
Selecting the type of probing . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Probing sequence. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-24
Probing conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-24
Probing speed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-25
Aligning the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-26
Preventing measuring errors. . . . . . . . . . . . . . . . . . . . . . . . . . 7-26
Special features of VAST probe system . . . . . . . . . . . . . . . . . . . . 7-27
Setting the measuring force . . . . . . . . . . . . . . . . . . . . . . . . . . 7-27
Precision positioning during probing . . . . . . . . . . . . . . . . . . . 7-28
61211-1020202 Operating Instructions Contents5
Notes on scanning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-29
Automatic probe change . . . . . . . . . . . . . . . . . . . . . . . . . 7-30
Tips for effective operation . . . . . . . . . . . . . . . . . . . . . . . 7-31
Evaluating the measuring data . . . . . . . . . . . . . . . . . . . . 7-32
Limiting values for deviations . . . . . . . . . . . . . . . . . . . . . . . . . . .7-32
Causes of large deviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-32
Terminating a measuring run . . . . . . . . . . . . . . . . . . . . . . 7-33
Effect of system shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-33
Shutdown procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-33
Chapter 8 Errors and faults
Errors occurring prior to the measuring run . . . . . . . . . . . 8-2
Faults during the measuring run . . . . . . . . . . . . . . . . . . . . 8-3
Special measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-5
After collision . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Probe head outside the measuring volume . . . . . . . . . . . . . . . . . . 8-5
If nothing else helps, ... – hotline! . . . . . . . . . . . . . . . . . . . 8-9
Chapter 9 Maintenance and care
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Care measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
Inspection measures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
Checking the probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Cleaning/changing the filter mat . . . . . . . . . . . . . . . . . . . . . . . 9-9
Changing the compressed air filters . . . . . . . . . . . . . . . . . . . . 9-10
Index
Glossar
61211-1020202 Operating InstructionsContents 6
Appendix
Application examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
61211-1020202 Operating Instructions Contents7
61211-1020202 Operating InstructionsContents 8

Foreword

Information about these operation instructions

CenterMax coordinate measuring machine (CMM) is described in these operating instructions.
These operating instructions address operators and users of the coor­dinate measuring machine.
Short information about the contents:
Warranty and safety-relevant information in Chapter 1 “Intro-
duction” on page 1-1.
– What is what, e.g. quill, coordinate axes? What is required for
operating the CMM? The answers can be found in Chapter 2 “Description of the CMM” on page 2-1.
– Dimensions, connection data, travel speed, environmental condi-
tions of the CMM and probe system data can be found in
Chapter 3 “Technical data” on page 3-1.
– You want to install the CMM at another place. Or you want to
install an additional CMM. For more information, please refer to
Chapter 4 “Transport and installation” on page 4-1.
– Making connections, mounting of probe racks, general visual
check prior to start-up. For more information, please refer to
Chapter 5 “Preparations for start-up” on page 5-1.
– The workpiece and the CMM must be prepared before starting the
measuring run. To do so, please refer to Chapter 6 “Start-up” on page 6-1.
– Connections made, workpiece clamped on measuring table, probe
system prepared. The preparations for the measuring run can begin Chapter 7 “Measuring operation” on page 7-1.
– In case the CMM is not doing what you expect it to, please refer to
Chapter 8 “Errors and faults” on page 8-1.
– A CMM also requires care, especially for ensuring safe and correct
measurements. The corresponding measures are explained
Chapter 9 “Maintenance and care” on page 9-1.
Further chapters:
Detailed listing of all chapters: Contents.
Explanation of important terms: Glossar.
Examples of CMM applications: Appendix.
Foreword61211-1020202 Operating Instructions

Symbols / danger warnings

Three special symbols which always denote important information are used in this manual.
These symbols appear in the left-hand margin of the page and directly adjacent to the corresponding text.
!
!
NOTE
Danger!
Special caution is advisable here. The warning triangle and the adja­cent text indicate a possible danger of injury.
Failure to observe this warning may result in injury to the operator.
Important!
This symbol warns you of situations which might cause measuring errors, interfere with a measuring run or even lead to collisions result­ing in damage to the CMM and/or the workpiece.
This symbol indicates additional and helpful information.

For your orientation

The following typographical conventions are used in this manual:
Example Description
not Words are emphasized by placing them in
italics.
– Italics are often used to set apart an inter-
mediate heading, e.g. Type of measurement:
Protective circuit Specially emphasized words are printed in
bold type.
no grease film Words specially emphasized in warnings are
also printed in bold type.
Blue bold type is sometimes used to mark a listing, e.g.
ST-ATAC: ... VAST: ...
"Measurement on
page ...
Page ...
Foreword 61211-1020202 Operating Instructions
Reference to a text passage where you can find more information.
Example Description
1 Remove probe.
2 Uncover adapter
plate receptacle.
Place the work­piece ...
Switch off the drives so that ...
1 Turning lever for ... 2 Metal plate with ...
A Adapter B Probe head
Instructions which must be followed in chronological order.
Instructions
Instructions in warnings, e.g. danger warn­ing.
Description of item numbers in a drawing
Description of item numbers in a drawing
Foreword61211-1020202 Operating Instructions
Foreword 61211-1020202 Operating Instructions
Chapter

1 Introduction

.................................................................................................................................
This chapter informs you about the applications of the CMM and the prerequisites for safe use.
This chapter contains:
General information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Additional documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Proper use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
1-161211-1020202 CenterMax Operating Instructions
Introduction

General information

Scope of delivery

The standard version of the CenterMax coordiante measuring machine is delivered with the following components:
Coordinate measuring machine (CMM)
Control
Control panel standard version
Computer with peripheral units
Software e.g. UMESS-UX
Documents
Probe system
Master probe
Reference sphere – for calibration of probes
Temperature sensor – for measuring the workpiece temperature
Temperature probe.

CE marking

The CE marking confirms the system's conformity with the basic requirements of the following CE directives:
Machinery directive 98/37/EC
Low voltage directive 73/ 23/EEC
EMC directive 89/336/EEC.
The fact that the CMM meets the above requirements is indicated by the CE marking on its type plate.
1-2
61211-1020202 CenterMax Operating Instructions
General information

Safety of the machine

Approvals for the CMM:
Europe GS marking of the professional associations of
precision engineering and electrical engineering.

Standards and regulations

The CMM is designed, manufactured and tested according to the fol­lowing standards and regulations:
Regulations – BGV A1 (VBG 1): Accident prevention regulations –
General regulations
– BGV A2 (VBG 4): Accident prevention regulations –
Electrical systems and equipment
Standards – DIN/VDE 0100:
Regulations for electric power installations with nominal voltages of up to 1000V
– EN 61010 part 1:
Measuring machines, control units, controllers and laboratory equipment
– EN 60204 part 1:
Safety of machines – electrical equipment of machines
– EN 292, part 1:
Safety of machines – basic terminology
– EN 292, part 2:
Safety of machines – technical directives and specifications
– EN 61326:
EMC – interference immunity; interference emission, class A
– CSA C22.2 No. 1010.1:
Safety requirements for Electrical Equipment for measurement, control and laboratory use.
– UL 3101-1:
Standard for Electrical Equipment for Laboratory use.
1-3CenterMax Operating Instructions61211-1020202
Introduction

Additional documents

Separate documents

Components for which separate documents are available are referred to in these operating instructions.
Separate documents are available for the following components:
Standard The following documents are delivered with each CMM:
General safety instructions
Control console standard.
Option The following document is delivered if the CMM is equipped with the
corresponding option.
– Loading device.

Supplementary literature

"Einfach messen – und was Sie dazu wissen sollten". The primer of measurement technology.
Carl Zeiss, Unternehmensbereich Industrielle Messtechnik, order no.: 612302-9002.
1-4
61211-1020202 CenterMax Operating Instructions

Warranty

Warranty
Notes
– The specifications and statements in the German operating
instructions are prevalent and binding for translations into other languages.
– All rights pertaining to changes in the coordinate measuring
machine and its options, the software packages and the pertaining documents reserved.
– All rights reserved, especially in cases of granting a patent or regis-
tering a utility model.
The warranty does not cover the following:
incidentals
wearing parts.
Exclusion of warranty
The manufacturer cannot be held liable.....
for actions contrary to the instructions in this manual,
if the coordinate measuring machine version delivered by us was
changed,
– if maintenance work is not carried out by personnel specially
trained at Zeiss,
– if measures for care are not taken by the operator or user accord-
ing to the specified measures,
– if original spare parts are not used for maintenance and repair
work,
– if the necessary maintenance work and measures for care are not
carried out according to the Customer Documentation.
All information regarding maintenance work and measures for care, incidentals and wearing parts is contained in the Customer Documentation manual.
1-5CenterMax Operating Instructions61211-1020202
Introduction

Proper use

Warranty
The coordinate measuring machine (CMM) may be used only for its intended purpose. The user is liable for any damage incurred during improper use of the CMM.
CMM
The CMM can be used to determine the geometric sizes of work­pieces. The workpieces may be made of metal or plastic.
Movable axes The CMM can be moved in all axes and probes the workpiece in the x,
y and z coordinate directions. A rotary axis is available when using a rotary table.
Dimensions, position and form
!
Geometric sizes are for example width, length, height as well as the diameter and depth of bores.
The position of bores is calculated from the measuring data. Further­more, the form of workpieces, e.g. the form of a turbine bucket can be determined by means of a special software.
Danger!
The CMM must not be used for determining geometric sizes of work­pieces. The CMM must not be used for any purposes other than those listed below.
Only workpieces and objects required for measurement may be placed on the measuring table.
Functions of the CMM:
– Workpiece probing in six directions (±x, ±y, ±z).
This is ensured by means of three moving machine axes. Probing is executed by a probing system. For measurement, the workpieces must be fastened on the measuring table.
– The measuring table has the following functions:
1-6
Support of one or more workpieces
Support of the calibration tools
Storage of a probe rack (option)
Storage of a rotary table (option) – the rotary table is either inte-
grated in the measuring table or attached on top of it.
– Fastening of the workpieces and installation of the probe racks,
calibration tools and the rotary table via the threaded holes pro­vided in the measuring table.
61211-1020202 CenterMax Operating Instructions
Proper use
Improper use
The CMM must not be used for purposes other than the proper use.
Examples for use contrary to the proper use:
The machine axes must not be used to move objects.
The z axis of the CMM must not be used to lift objects.
The CMM must not be used to measure living objects or parts of
the body.

Probe system

The probe system is a hightech product which may only be used only for its intended purpose.
Probing CMM probe systems are designed to determine the coordinates of a
workpiece. This is usually achieved via a probing, during which the workpiece is probed by a probe element (stylus tip). In some cases optical measuring techniques are used.
Linear scales The probe system comprises several components, such as the probe
head and linear scales. The linear scales are integrated in the machine axes of the CMM and protected by covers. These covers must be attached when the CMM is in operation to prevent it from being used improperly.
Probe
The probe consists of several components: adapter plate, probe styli and probe joints. A probe may be equipped with one or more styli.
The probe element is located at the end of the stylus.
Functions of the probe:
– The adapter plate holds and positions the probe exactly in the
probe head.
– Workpiece probing is carried out by the probe element.
Probe head
The probe head is inserted in the quill. The probe used for probing is inserted below the probe head. Handle the probe head and probe very carefully.
Functions of the probe head:
Holding the probe.
Exact positioning of the probe.
1-7CenterMax Operating Instructions61211-1020202
Introduction
– Detection of probe deflection and transmission of the signal to the
computer. The computer calculates the coordinates of the probed point.
Improper use
The probe head and probe must not be used for purposes other than the proper use.
Examples:
The probe head must not be used as a support.
The probe must not be used as lever arm, e.g. to loosen a ring bolt.
The probe must not be used as striking tool.
1-8
61211-1020202 CenterMax Operating Instructions

Safety

Safety

Requirements for person responsible for CMM

The person responsible for the CMM must ensure that:
The installation site requirements are met.
The operator has been properly instructed regarding handling and
operation of the CMM.
– The operator has received the operating instructions for working
with the CMM. The operating instructions must be suitable for all local and operating conditions and available in the language spo­ken by the operating personnel.
The operating instructions must always be available within easy reach of the CMM.
The operator must know and observe the safety instructions.
Maintenance work and work on electrical equipment must be car-
ried out only by competent specialists.

Definition of a specialist

A specialist is a person who due to his
training
experience
instruction
and knowledge of
relevant standards
legal regulations
accident prevention regulations
and operating conditions
has been authorized by the safety officer responsible for the CMM to perform the currently required actions and therefore is in a position to recognize and avoid potential dangers.
1-9CenterMax Operating Instructions61211-1020202
Introduction

Safety symbols on the CMM

The following symbols are attached to the coordinate measuring machine (CMM) and the control cabinet.
Danger symbol for electric voltage: This symbol indicates a personal health
hazard of persons. The symbol is attached to the front and rear
side above or below the door latch.
!
!
Danger!
Any work on the control cabinet must be carried out only by an elec­trical engineer.
Before opening the control cabinet, disconnect the control cabinet from the power supply:
• Switch off the main switch – position "0“ – and secure it so that it cannot be restarted.
In case of plug connection: Disconnect the power plug from the socket.
Danger!
Components in the control cabinet are still live, e.g. line filter, even after switching off the main switch.
Danger symbol: This symbol indicates a danger for persons and
the CMM and its components. This symbol is attached to the CMM, the con-
trol panel or the control cabinet.
1-10

Requirements for the safe use of the CMM

In order to ensure safe operation of the coordinate measuring machine (CMM), certain requirements must be met.
Observe the following:
Always comply with the generally recognized accident prevention regulations and safety instructions.
Operate the CMM only with the protective devices provided for that purpose.
Do not remove any covers, protective equipment or warning
signs.
61211-1020202 CenterMax Operating Instructions
Safety
Only use power cables and connectors which are in perfect condi­tion.
Insert the power connectors only in sockets equipped with a per­fect protective ground connection.
Do not exceed the maximum permissible workpiece weight.
No warranty claims The manufacturer of the machine cannot be held liable for any dam-
age caused by unauthorized interventions in the measuring system. In case of unauthorized manipulations, all warranty claims against the manufacturer and supplier as well as the validity of the EC declaration of conformity are null and void.
Documents Correct operation of the CMM is required for a safe measuring run. It
is assumed that you are familiar with the documents included in the scope of delivery.
Ensure that the operating instructions for the CMM are always available within easy reach of the CMM.
NOTE
Where necessary, further safety instructions are mentioned in these operating instructions. Furthermore, a separate document regarding "Safety" is also available.
Please read the notes in the document entitled "Safe operation of bridge CMMs"

Safety devices

The coordinate measuring machine (CMM) is equipped with safety devices. Some devices are standard devices, others are optional.
Standard The standard safety devices include:
The EMERGENCY STOP button on the control panel
The quickstop button on the control cabinet
Thrust force limiting (hardware) and drive monitoring (software)
Drop protection of the quill
Collision protection for the probe head and the probe; effective in
manual mode (v
= 70mm/s).
max
Option Optional safety devices are:
Light barrier
Footswitch mats.
1-11CenterMax Operating Instructions61211-1020202
Introduction
1-12
61211-1020202 CenterMax Operating Instructions
Chapter

2 Description of the CMM

.................................................................................................................................
This chapter provides a general overview of the CMM and its compo­nents.
This chapter contains:
Identification of a CMM. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Design of the CMM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Components and their functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Probe rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2-161211-1020202 CenterMax Operating Instructions
Description of the CMM

Identification of a CMM

The CenterMax is identified by the size of its working range. This range is specified by its length in the x, y and z axes.
Identification of CenterMax:
Identification: 11/12/7
Ref. no. Working range
11 x: 1100 mm
12 y: 1200 mm
7z:700 mm
NOTE
Type pl a t e The identification of your CMM is given on the type plate. The type
The working range does not correspond to the measuring range. Please refer to the technical data for more information on the measur­ing range.
plate is located on the front side of the measuring table.
12 43
1 Serial number 2 Year of manufacture 3 Number of the CMM type 4 Series and identification of the CMM type.
2-2
61211-1020202 CenterMax Operating Instructions

Design of the CMM

A CMM of the series CenterMax is illustrated in the following.
Design of the CMM
1
6
1 Quill cover 2 X-bridge with concertina cover 3 Drive-side and guideway-side 4 Probe head 5 Measuring table 6 Guideway-side
2
3
4
5
2-3CenterMax Operating Instructions61211-1020202
Description of the CMM

Components and their functions

Bridge

The CenterMax is a bridge coordinate measuring machine. The bridge is composed of a crossbeam and two guideways. The quill is guided on the crossbeam of the x-bridge.
4
1 Crossbeam 2 Drive side and guideway side 3Quill 4Guideway.
3
1
2

Measuring table

The workpieces to be measured are placed on the measuring table.
Threaded holes Threaded holes are provided in the measuring table. These are used to
fasten workpieces and calibration tools on the measuring table. The threaded holes are distributed on the measuring table. For more infor­mation, please refer to the document "Installation Instructions“.
2-4
61211-1020202 CenterMax Operating Instructions
Components and their functions
Measuring volume
The measuring volume is the space in which workpieces can be probed.
1
1 Measuring volume

Coordinate axes

Travelling in the x, y and z axes is possible with the coordinate mea­suring machine.
Travel directions The drawing shows the possible travel directions of the bridge.
2-5CenterMax Operating Instructions61211-1020202
Description of the CMM
1
x direction
Crossbeam The x-bridge crossbeam is mounted on the columns and supports the
quill. The quill is guided on the crossbeam and can be moved back and forth. Thus, probing in the x direction is possible.
y direction
Columns The drive and guideway side columns support the cross beam and the
quill and the frictionless air bearings enable movement of the com­plete bridge in the y direction.
z direction
Quill The quill is guided along the crossbeam. It can be moved in the
vertical direction and allows probing in the z direction. The probe sys­tem is located at the bottom of the quill.

Control

Control cabinet
32 bit / C99 The control for the CenterMax is located in a separate control cabinet.
The CenterMax is equipped with a 32 bit control. The control has the designation C99.
IP54 The housing of the control cabinet has protection class IP54.
Front side of the control cabinet:
2
3
4
2-6
1 Operating elements 2 Type plate for control cabinet 3Door lock 4 Warning symbol for electric voltage: The control cabinet must be opened
only by an electrical engineer.
61211-1020202 CenterMax Operating Instructions
Components and their functions
1
Rear side of the control cabinet:
2
3
4
5
6 7
8
9 10
1Fan 2Air filter 3 Warning label: Replace or clean soiled air filter! 4Door lock 5 Warning symbol for electric voltage: The control cabinet must be opened
only by an electrical engineer.
6 Plate with technical data 7 Attention symbol: Observe the information on the plate! 8 Connector for peripheral devices 9 Power supply of the control cabinet 10 Cables leading to the CMM.
2-7CenterMax Operating Instructions61211-1020202
Description of the CMM
Operating elements of the control cabinet
The following operating elements and indicator lamps are located on the control cabinet:
187 6 4 2
53
1 Main switch for power supply 2 Rotary switch for the control 3 Indicator lamp for control; green 4 Rotary switch for drives 5 Indicator lamp for drives; green 6 Key-operated switch for operating modes 7 Working-hour meter 8 Push-and-turn switch for quick stop; for cases of emergency
The functions of the operating elements are explained in the follow­ing.
Power supply Main switch (1): The main switch is used to switch the power supply
of the coordinate measuring machine on and off.
a Direction of rotation for switching on and off; 0=On, 1= Off
a
b
c
b Detent opening; in position 0 the switch can be protected against unauthorized switch-on.
c Position 0; power supply is switched off.
Control
Rotary switch (2) and indicator lamp (3): The rotary switch is used to switch the control on. If it is in the ON position, the control is switched on. The indicator lamp above the switch lights up.
Drives Rotary switch (4) and indicator lamp (5): The rotary switch is used to
switch the drives on.
Turn the switch clockwise. The switch does not remain in the new position. It returns to its initial position.
2-8
61211-1020202 CenterMax Operating Instructions
Components and their functions
When the drives are switched on, the indicator lamp above the switch lights up.
Operating modes Key-operated switch (6): The key-operated switch is used to deter-
mine the operating mode in which the measurements are to be car­ried out.
– MAN: The measurement is carried out manually. The workpiece
is probed by means of the joysticks on the control console. The maximum speed is limited Travel speed on page 3-3.
OFF: No travel movement possible.
AUTO: The measurement is carried out automatically, e.g. by
means of a CNC program. Before setting this operating mode, you should test the CNC program in the MAN operating mode.
Working-hour meter The working-hour meter (7) counts the number of hours during
which the drives remain activated.
Quickstop button Push-and-turn switch (8): The quickstop button can be pressed to
stop all travel movements in cases of emergency. It engages when pressed. All drives are switched off.
The quickstop button must be released before switching on the drives. Release the button by turning it counterclockwise. Then, the drives can be switched on again by means of the rotary switch
page 2-8.

Control console

The CenterMax is equipped with the standard control console on delivery. However, it can also be delivered with the Dynalog control console on request. Each of these control consoles is described in a separate manual.
2-9CenterMax Operating Instructions61211-1020202
Description of the CMM
Standard control console
EMERGENCY STOP but­ton
1
2
1 Control knob for speed adjustment 2 Joystick for the z axis and the rotary table (option); the joystick is
equipped with a push button for setting intermediate positions
3 EMERGENCY STOP button 4Keypad 5 Joystick for the x and y axis; the joystick is equipped with a push button
for setting intermediate positions.
The drives are switched off by pressing the EMERGENCY STOP but­ton. Press the button to switch off the drives. The button engages. No further travel movements are possible.
3
4
5
The button must be unlocked in order to switch the drives on again.
Unlock the button by turning it. Then the button is pushed up. The button is unlocked.
Operation of the joysticks
The joysticks are required for travel control. This applies to manual and automatic probing. Prior to automatic probing, manual probing is required to program the automatic measuring run.
Right joystick
x and y direction The right joystick is used to move the probe in the x and y direction.
The bridge moves in the y direction, the quill in the x direction.
2-10
61211-1020202 CenterMax Operating Instructions
Components and their functions
Deflect the joystick in the direction required to move the probe toward the workpiece. The right joystick is used to execute the fol­lowing actions:
NOTE
Axis / Direction
x direction -x Push to the left.
y direction -y Push to the rear.
You may reverse the travel direction of the joysticks. This is advisable when operating the control console and performing probing from behind the CMM.
Action Right joystick
-x +x
The quill moves to the left.
+x Push to the right.
The quill moves to the right.
-y
The bridge moves away from the opera­tor.
+y Pull to the front.
The bridge moves towards the operator.
+y
z direction and direc­tion of rotation of the rotary table (option)
You can reverse the travel direction by pressing a button on the con­trol console Operating instructions for the control console.
Left joystick
The left joystick is used to move the quill in the z direction.
Direction Action Left joystick
z direction -z Push to the rear.
The quill moves downwards.
+z Pull to the front.
The quill moves upwards.
-z
+z
2-11CenterMax Operating Instructions61211-1020202
Description of the CMM
Direction Action Left joystick
NOTE
Direction of rotation of the rotary table (option)
If the joysticks are not activated within a certain period of time, they will be blocked. The blocking must be cancelled before it is possible to carry out travel movements by means of the joystick Operating instructions for the control panel.
Push to the left. The rotary table turns
clockwise.
Push to the right. The rotary table turns
counterclockwise.

Calibration standard

A high-precision sphere is used as the calibration standard. The sphere is made of a ceramic material and has a diameter of 30mm.
The calibration standard shown is included in the scope of delivery. The calibration standard has a magnetic stand.
NOTE
1
2
3
1 Calibration sphere 2 Rotary switch for activating the magnet 3Magnetic stand
A metal plate is required for mounting the calibration standard on the measuring table. This plate is delivered with the calibration standard. For more information see Mounting the calibration standard on the measuring table on page 7-18.
2-12
61211-1020202 CenterMax Operating Instructions
Components and their functions

Options

The following special equipment is available for CenterMax:
Loading device
Illuminating device.
2-13CenterMax Operating Instructions61211-1020202
Description of the CMM

Probe system

Probe system types

The CenterMax is equipped with the VAST probe system.
VAST
Application Universal
Measuring principle measuring
Further information page 2-15

Components of a probe system

The probe system comprises:
VAST
Probe head ×
Adapter plate ×
Probe stylus ×
Functions of the components
Probing The probe is used for probing the workpiece. One or more probes are
mounted on the adapter plate. The adapter plate and the probe form a single unit.
Holding + securing The adapter plate is used to hold one or more probes and fasten
them to the probe head. Proper fastening is ensured by means of a three-point bearing. The adapter plate is held by a magnet integrated in the adapter plate receptacle of the probe head.
A pin located on the adapter plate receptacle is mated to a recess on the edge of the adapter plate. The adapter plate must be inserted in the adapter plate receptacle so that the pin fits into this recess.
2-14
61211-1020202 CenterMax Operating Instructions
Probe system
Registration + transmis­sion
The probe head holds the adapter plate and the mounted probes. Furthermore, the probe head registers the deflection of the adapter plate and the probe during probing. A signal is sent to the computer after each deflection.
Standard version of the probe system
The probe system is delivered with different components. The stan­dard equipment comprises:
VAST
Probe head VAST
Master probe 1
Adapter plate 1
Probe rack
one rail
Adapter platea for
eightfold
2
probe rack
a
The adapter plate is identical with the adapter plate for the master probe.
b
Number of holders
b
NOTE
The probe system can be extended.

VAST probe system

Application
Universal The VAST probe system can be used universally. Measurements per-
formed with this system provide information on the dimensions, position and form of a workpiece.
Special applications:
– In case of adverse ambient conditions: e.g. vibrations (floor vibra-
tions and sound).
– For measuring many measuring points – multipoint measurement,
scanning.
If high accuracy is required.
If long, heavy probes are used.
Special probes In addition to conventional probes, a temperature-sensing probe may
be used. The temperature-sensing probe is used to measure the work­piece temperature.
2-15CenterMax Operating Instructions61211-1020202
Description of the CMM
0
Components
A
1 2
B
600g
450mm
A VAST probe head B Probe – adapter plate and probe configuration
a
1 Adapter plate receptacle 2 Pin for positioning the probe.
limiting values
The limiting values for weight, length and torque of a probe must be considered when assembling or mounting the probe.
Probe weight
The adapter plate receptacle may only be loaded with a maximum weight of 600g.
Probe length
The length of the probe head including extension must not exceed 450mm.
2-16
61211-1020202 CenterMax Operating Instructions
Torque
The maximum torque of the probe is 0.1Nm.
Probe system
!
NOTE
Important!
The weight of all components and the length of the probe styli, exten­sions and joining elements must be considered when assembling probes.
Observe also the weights indicated for the individual probes, join­ing elements and the adapter plate when assembling probes.
Observe the lengths indicated for the probes, extensions and other joining elements when assembling probes.

Temperature-sensing probe

Application
The temperature-sensing probe is used to measure the the tempera­ture of the workpiece surface and the surrounding air. Both measure­ments can also be carried out via a CNC program. Thus, temperature monitoring is possible throughout the entire measurement.
The VAST probe system is required for the operation of a tempera­ture-sensing probe.
For more information see Notes on temperature measurement on page 7-9 and Possibilities on page 7-10.
2-17CenterMax Operating Instructions61211-1020202
Description of the CMM
Design
A
1
2
B
A Adapter plate B RST temperature-sensing probe C Temperature sensor
1 Connector 2 Knurled ring with internal thread.
C
2-18
61211-1020202 CenterMax Operating Instructions

Probe

Probe

Probe components

The probe consists of
adapter plate
one or more styli
joining elements.
Probe configuration A distinction is made between probe configuration and probe combi-
nation. The probe configuration may consist of one or more styli. Sev­eral styli may form a probe combination.
NOTE
1
2
1 Probe combinations 2 Probe configuration
It is basically possible to create a probe configuration with several probe combinations. However, certain assembly criteria must be observed in order to ensure exact measurement Criteria and limit- ing values for assembling probes on page 6-13.

Design and function

Adapter plate
VAST The VAST adapter plate has the following functions; it must ...
hold the probes
fit into the adapter plate receptacle of the probe head
enable probe change via a probe rack.
2-19CenterMax Operating Instructions61211-1020202
Description of the CMM
A
The adapter plate is inserted in the probe head and held by magnetic force. The groove in the adapter plate allows correct positioning. The pin in the adapter plate receptacle of the probe head must engage in this groove.
The VAST adapter plate is equipped with a probe mounting cube with five threaded holes used for mounting probe components. The size of the connection thread is M5.
1
2
3
A VAST adapter plate
1 Anchor plate 2 Groove for pin in the adapter plate receptacle of the probe head; the pin
allows correct positioning of the adapter plate.
3 M5 threaded hole for probe components, e.g. styli.
Joining elements
The probe kits for the probe systems contain probes or styli of differ­ent sizes and components used to create probe configurations, e.g. extensions.
Probes
Shaft + probe element A probe or stylus consists of a shaft and a probe element or stylus tip.
Shafts differ with regard to size and material. Furthermore, probe ele­ments differ with regard to form page 2-21.
2-20
61211-1020202 CenterMax Operating Instructions
Probe

Probe elements

Probe elements may have various forms. Examples of probe elements:
12 3 4
1Sphere 2Cylinder 3Cone 4Disc
Application of the probe elements
Sphere The sphere is the standard probe element suitable for most measuring
jobs.
Cylinder Cylinders are preferably used for probing thin sheet-metal parts and
narrow workpiece edges.
Cone Cones are required for two applications: For ...
self-centering probing
2-21CenterMax Operating Instructions61211-1020202
Description of the CMM
High-precision probing of specified positions.
Disc A disc is used for probing deep grooves and large holes. For large
holes, a disc offers the advantage of shorter travel paths as compared with the sphere.
2-22
!

Principles for assembly

There are two ways to assemble a probe:
Screwing technique
Clamping technique
Preference should be given to the screwing technique.
Probes may be assembled individually. Probe kits with different com­ponents are available.
Important!
The limiting values for probes must be observed when assembling probes.
61211-1020202 CenterMax Operating Instructions
Probe
Screwing technique
The probe components have threads. In this way the components can be screwed together. Depending on the probe system, different thread sizes are used: M2, M3 and M5.
– M5: VAST
Probe kits
There are different probe kits, e.g. for small and large workpieces, for each probe system.
Probe kit Order no. Use
MPH, small 600891-9021 for VAST; small work-
pieces
MPH, large 600891-9031 for VAST; large work-
pieces
Technical data of probe components
The probe components differ with regard to material and geometry, e.g. length and diameter of the probe element. The probe weight var­ies accordingly.
The following table shows the technical data of some available probes. This data shows the influence of the material and the geo­metric sizes (e.g. length of the probe) on the probe weight. A table with the technical data of other probe components is included with the probe kits.
Extract from the technical data of probes
dL DMLd
[mm] [g]
Hard metal
3 2511152 4
3 33,5 11 23.5 2 4
5 5011403.59
8 63.5 11 50.5 6 25
Weight
S
8 114.5 19 92.5 6 59
9 64.5 19 42.5 6 38
10 65.5 11 52.5 6 27
Ceramic material
5 5311433.55
a
2-23CenterMax Operating Instructions61211-1020202
Description of the CMM
Extract from the technical data of probes
Advantages and disad­vantages of the clamp­ing technique
dL DMLd
8 63.5 11 50.5 6 8
8 114.5 19 101.5 6 38
9 64,5 19 42.5 6 34
a
The probe shaft is made of ceramic material.
10 65.5 11 52.5 6 9
[mm] [g]
Weight
S
Clamping technique
Laterally oriented styli are inserted in grooved discs and clamped by the vertically oriented stylus. The end stylus is screwed into the probe extension or directly into the adapter plate. In this way the laterally oriented styli are clamped.
The probe components for the clamping technique and assembly instructions are included in the probe kit.
Advantages:
high flexibility
low weight.
Disadvantages:
– high measuring inaccuracy caused by poorly definable clamping
joints.
– with VAST: loss of rigidity at each joint.

Storage of probes

Once the probe has been assembled, it can be used as long as required. It does not need to be disassembled. Careful handling is necessary. This applies particularly to the adapter plate and probe ele­ments.
Probe protection The probe must be protected against the following influences:
Force of impact and other external mechanical infuences
Dust and dirt.
A probe rack is recommended for storage if you measure every day with the CMM. When no measurements are carried out with the CMM over a longer period, the probes should be stored in a place protected against environmental influences:
2-24
61211-1020202 CenterMax Operating Instructions
Separate packing materials
e.g. box made of cardboard, styrofoam, wood
When using a metal box, wrap the probe in a cloth.
This is necessary to protect the adapter plate and probe ele­ments against being scratched.
Drawer
Wrap the probe in a cloth or lay it on a soft support.
This is necessary to protect the adapter plate and probe ele­ments against being scratched.
Special probe cabinet
The probes are inserted in holders. The probe cabinet is available from ZEISS.
Probe
2-25CenterMax Operating Instructions61211-1020202
Description of the CMM

Probe rack

Probe system order numbers
Components of the probe rack
Probe racks supplied for retrofitting purposes are delivered unassem­bled.
Scope of delivery:
Profile rail for fastening the holders.
Holders (storage positions):
8 holders belong to a probe rack. The holders are preassembled.
– Screws and mounting aids.
Two probe racks are available for CenterMax. One rack is included in the standard version and the other rack can be ordered:
Order no.
1038-094 Included in the CenterMax standard
version
1038-093 Option
NOTE
2-26
The probe rack with the order no. 1038-093 is mounted at the drive end. However, the measuring range in the x direction is reduced.
61211-1020202 CenterMax Operating Instructions
Chapter

3 Technical data

.................................................................................................................................
The following text provides technical data on the CMM and the probe system.
CMM: Data referring to dimensions, measuring range, weight, noise level, parameters, environmental conditions, requirements for the electric power supply and compressed-air supply.
Probe system: Data referring to several probe systems, e.g. limiting values and environmental conditions.
This chapter contains:
Coordinate measuring machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Probe system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3-161211-1020202 CenterMax Operating Instructions
Technical data

Coordinate measuring machine

NOTE
The table below lists a selection of technical data. For more data, please refer to the document "Installation Instructions".

General data

CenterMax 11/12 /7
Dimensions
Width [mm] 2090
Length [mm] 2126
Height
Measuring range
x [mm] 1100
y [mm] 1200
z [mm] 700
Weight
CMM [kg] 6000
a
[mm] 3000
Workpiece [kg] 1000
Noise level of the CMM [dBA] <70
a
An additional distance is required for mounting work: min. 200mm.

CMM parameters

Category Characteristic
value
Overvoltage category III
Pollution degree 2
Protection class 1
3-2
61211-1020202 CenterMax Operating Instructions
Coordinate measuring machine

Connection data

Electric data
Line voltage 100, 110, 115, 120, 125, 230,
240 VAC (±10%)
Type of current 1/N/PE
Frequency 50 to 60Hz (±3.5%)
Power consumption max. 2500VA
Fuse
for 230V or 240V C 25A
for 100V or 125V 25A
Compressed air
Certain requirements have to be met before connecting the com­pressed air in order not to impair the function of the CMM.
Requirements
Pressure 6 - 10bar
Degree of purity 99%; without solid particles, gas-
eous proportions of oil and water
Air consumption max. 20l/min with 5bar

Travel speed

Creep speed: SLOW function
0 to 5 mm/s
Setup mode: MAN operating mode
0 to 70mm/s
Series measuring operation: AUTO operating mode
Axis Vector
max. 300mm/s max. 425mm/s (or 520mm/s)
3-3CenterMax Operating Instructions61211-1020202
Technical data

Environmental conditions

The following conditions must be fulfilled to guarantee perfect opera­tion of the coordinate measuring machine.
Ambient temperature for storage
Ambient temperature for operational readiness
Ambient temperature for measuring operation
Relative humidity 40% to 70%
+5°C to +40°C
1535°C
1535°C
3-4
61211-1020202 CenterMax Operating Instructions

Probe system

Probe system

VAST probe head

CMM dimensions
Length 91mm
Width 91mm
Height 200mm
Probing directions 6; ± x, ±y, ± z
Measuring force 0.05 to 1N / continuous
Diameter of adapter plate 69mm
Probe head deflection max. ±5mm
Probe weight, max. 600g / incl. adapter plate
Probe length, max. 450 mm / stylus + extension
Repeatability following probe change
<1µm / with 200mm probe length
Environmental conditions
Ambient temperature for opera­tional readiness
+5°C to +40°C
3-5CenterMax Operating Instructions61211-1020202
Technical data
3-6
61211-1020202 CenterMax Operating Instructions
Chapter

4 Transport and installation

.................................................................................................................................
Do you want to install the CMM at another location or install an addi­tional CMM? In either case, this chapter provides the basic informa­tion you will need on the required condition of the installation site and transport considerations.
This chapter contains:
Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
4-161211-1020202 CenterMax Operating Instructions
Transport and installation

Transport

Separate document
NOTE
Requirements regard­ing the fork-lift truck
You received the document "Installation Instructions prior to the delivery of the coordinate measuring machine. This document informs you about all measures to be taken for transport and installation. In the following, we will refer to some of the most important points con­tained in this document.

To be observed on delivery

The coordinate measuring machine is delivered disassembled on transport pallets. The individual parts are wrapped in insulated pack­aging. For overseas transport, the individual parts are packed in ship­ping containers.
A fork-lift truck with sufficient carrying force has to be used for trans­port. The fork-lift truck must be suitable for the weight of the individ­ual transport containers. In order to avoid tilting of the load, the forks must be completely underneath the transport pallets or shipping con­tainers and the fork width must be variable. You will find the weight and dimensions of the individual parts in the document "Installation Instructions“.
!
Ambient temperature +5°C to +40°C
Prerequisites for trans­port
Important!
The packing material or the transport containers must not be dam­aged. The packing material may only be removed at the installation site by a ZEISS service engineer.
The transport pallets and shipping containers must be stored in a cov­ered area until the machine is installed. The ambient temperature in the storage area must be within +5°C to +40°C.

Transport conditions

The transport route from the storage area to the installation site must be selected carefully. It might be necessary to prepare it in order for it to meet the requirements.
Observe the following regarding transport:
– Weight: Is the bearing capacity of the floor along the routes suit-
able for the weight of the transport containers and the fork-lift truck?
4-2
61211-1020202 CenterMax Operating Instructions
Transport
Dimensions: Are the height and width of the doorways and
routes sufficient for the transport containers and the fork-lift truck?
Fork-lift truck: Does the fork-lift truck comply with the
requirements? page 4-2.
4-3CenterMax Operating Instructions61211-1020202
Transport and installation

Installation

Separate document
NOTE
You received the document "Installation Instructions prior to the delivery of the coordinate measuring machine. This document informs you about all measures to be taken for transport and installation. In the following, we will refer to some of the most important points con­tained in this document.

Selecting the installation site

The place of installation must meet the following requirements. The following questions must be clarified:
Is the room height sufficient?
Is enough space available?
Is the floor suitable for the weight of the CMM and the fork-lift
truck?
– How strong are the vibrations at the installation site?
Disturbing vibrations may for example be caused by machines in adjacent rooms or buildings.
– Does the installation site comply with all of the environmental con-
ditions required for perfect operation of the CMM?
– Is the installation site free from coarse dirt particles, such as casting
sand and metal chips?
– Have the requirements for the compressed air supply been met?
For more information, please refer to the document "Installation Instructions“.
Carrying out vibration analysis
A vibration analysis must be carried out in any case. The result of the analysis must be made available to ZEISS for planning purposes. An additional foundation may be necessary.
ZEISS can carry out a vibration analysis for you. You can order it under the order no. 6000 08-0276.
Dial +49.(0)7364.20.3471 if you want ZEISS to carry out a vibra­tion analysis.
4-4
61211-1020202 CenterMax Operating Instructions
Installation

Prerequisites for installation

The following preparations must be made prior to installation of the CMM by ZEISS service engineers:
Provision of steel plates
Installation of power supply
Installation of compressed-air supply
Adaptation of CMM to room temperature
You will find the data required for preparation in the document "Installation Instructions“.
Adaptation to room temperature
During start-up of the CMM, the temperature of all CMM parts must be stabilized at a minimum temperature of 15°C.
At least two days in advance
Store all parts of the CMM in a place having an ambient tempera­ture of at least 15°C at least two days prior to installation.
4-5CenterMax Operating Instructions61211-1020202
Transport and installation
4-6
61211-1020202 CenterMax Operating Instructions
Chapter

5 Preparations for start-up

.................................................................................................................................
The preparations include many activities not performed daily or weekly. One exception to this is the visual check. The visual check should be part of your routine daily measuring operation.
It is possible that the probe rack must be removed, replaced or modi­fied. In this case, please refer to the corresponding information in this chapter.
This chapter contains:
Before you start! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Mounting the probe rack (option) . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Visual check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Start-up preparation checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
5-161211-1020202 CenterMax Operating Instructions
Preparations for start-up

Before you start!

Initial start-up by a ser­vice engineer
Initial start-up is carried out by a ZEISS service engineer. However, you must familiarize yourself with the preparations required for start-up and know as well as observe the corresponding safety instructions.
Please read the information provided in the document "General safety instructions“.
5-2
61211-1020202 CenterMax Operating Instructions

Connections

Connections

Connecting the power supply

The electrical power supply must be located close to the installation site to enable electrical connection of the CMM. This must be done before installing the CMM.
Observe the electrical data provided in the chapter "Technical datapage 3-3.
NOTE
By ZEISS service engi­neers
For more information on the electric power supply, please refer to the document entitled "Installation Instructions“.
The CMM is fitted with a power plug.
Connect the power plug to the electric outlet provided.
Connection of the power supply is part of the CMM installation pro­cedure. It may be carried out only by ZEISS service engineers.
The CMM power supply is provided by the control cabinet. A permanent connection must be provided for the control cabinet power supply.
Connect the control cabinet power cord to the power supply line.
Protect this connection with a cutout switch.
1
2
1 Power supply (permanent connection) 2 Control cabinet.

Connecting the compressed-air supply

The building compressed-air supply must be close to the CMM in order to be able to connect it to the CMM supply line. This must be done before installing the CMM.
5-3CenterMax Operating Instructions61211-1020202
Preparations for start-up
Observe the requirements regarding the compressed-air supply in the chapter "Technical datapage 3-3.
NOTE
For more information on the connection of the compressed-air supply, please refer to the document "Installation Instructions“.
Requirements for compressed-air supply
Pressure in the supply line:
Required minimum pressure: 6bar
Permissible maximum pressure: 10bar.
Quality of the compressed air:
Degree of purity 99%
Free from solid particles
Free from water in liquid or gaseous form
Free from oil in liquid or gaseous form.
Compressed-air connection to the CMM
The compressed-air connection is located on the right-hand side of the CMM behind the doors.
1
2 3
1 CMM (top plan view) 2 Compressed-air connection of the CMM on the right-hand side
page 5-5
3 Compressed-air supply at the installation site.
The connection for the dampers of the CenterMax is located behind the left door. The connections for the axes and the special equipment are located behind the middle door.
5-4
61211-1020202 CenterMax Operating Instructions
Connections
Pressure gages
There are five pressure gages for five devices. These devices and their required pressures are listed below:
231 45 678
!
1 Compressed-air connection to the CMM 2 Air filters; coarse and fine 3 to 7: Compressed-air gages and connectors for five
different devices
8 Pressure controller
Indicated pressures:
Device No.
y drive 3 2bar
Weight balancer 4 5bar
Option: 5 5bar
Option: e.g. loading device 6 5 bar
Air bearings in the CMM 7 5bar
a
The values set by the ZEISS service engineer may slightly deviate from the indicated values.
Important!
The valves of pressure gages which are not connected must be closed. The connection shank should be covered by a cap.
Required pressure
a
5-5CenterMax Operating Instructions61211-1020202
Preparations for start-up
Compressed-air valve with pressure gage
1321
1 Control knob for pressure adjustment – locked; only for service engineers 2 Air pressure gage 3 Connection of the compressed-air line.
!
Adjusting the pressure
Important!
The pressure is adjusted individually by a ZEISS service engineer during the initial start-up. After this, the pressure must not be readjusted.
If the required pressure is not indicated, proceed as follows:
Identify the cause of the decrease or increase of pressure “Faults during the measuring run on page 8-3.
Call a ZEISS service engineer if you cannot find or eliminate the cause.
Filter units
Several filter combinations are offered for the CMM. The CMM fea­tures a combination of a fine filter and an extra fine filter as standard equipment. If the required compressed air quality cannot be main­tained, additional filters can be installed later. The possible versions are specified below::
Combination
5-6
1 Fine and extra fine
filter
61211-1020202 CenterMax Operating Instructions
Standard
Combination
Connections
2 Combination 1 +
membrane dryer for filtering out moisture
3 Combination 2 +
combination 1
Optional
Optional
5-7CenterMax Operating Instructions61211-1020202
Preparations for start-up

Mounting the probe rack (option)

The probe rack is mounted by ZEISS service engineers during the ini­tial start-up. If the probe rack is to be installed in another location or a second probe rack is to be installed, you have to observe certain points.

Prerequisites for the probe rack

What are the requirements for the probe rack?
The requirements must be known before installing a probe rack.
– When using probes with long horizontal styli, it may prove impossi-
ble to use all holders.
Possibly two probe racks
Clarify questions in advance
The requirements of the measuring job influence the number of num­ber of holders which can be used and the height of the probe rack. It may be necessary to install two probe racks.
The following questions must be clarified before installing the probe rack:
– Number of holders used for the measurement?
Space required by the widest probe?
– Number of probes required for workpiece measurement?
How to install the probe rack
The profile rail is attached to the guideway side. Two threaded holes are provided for this purpose. Threaded bolts are then screwed into the threaded holes and secured with a lock nut Installing a probe rack on page 5-12.

Installation procedure

The holders are preassembled on the profile rail. The profile rail can be attached directly to the guideway side.
5-8
!
Important!
Exact probe rack alignment is necessary to enable trouble-free probe change. The deviation must not exceed 0.1mm over the total length of the profile rail.
Align the probe rack so that the deviation of the profile rail does not exceed 0.1mm.
61211-1020202 CenterMax Operating Instructions
Mounting the probe rack (option)
Attaching the holders to the profile rail
1
2
3
1 Holder (storage position) 2 Profile rail 3 Guideway for holder – at the top or at the bottom.
The holders are fastened to the profile rail using Allen screws. Two screws are required for VAST holders.
5-9CenterMax Operating Instructions61211-1020202
Preparations for start-up
The holders are generally preassembled. The procedure for moving, removing or adding a holder is described in the following.
A Cap B Holder for probe system VAST;
a: top plan view; b: front view
1 Mounting aid 2Screw; M6×16 3 Cover of a holder 4 Brush for cleaning the adapter plate
Remove the cap (A).
Move the cover of the holder in the direction of the arrow (3) and
loosen the two hexagon-socket screws (2) slightly; do not remove them completely from the mounting aid.
Slide the holder on the profile rail. The mounting aid must be guided in the profile rail.
Proceed in the same way for sliding further holders on the profile rail.
The permissible number depends on the probe system and the pro­file rail length.
Reattach the cap to the profile rail.
5-10
61211-1020202 CenterMax Operating Instructions
Mounting the probe rack (option)
Distribute the holders evenly on the profile rail and tighten the screws (2).
5-11CenterMax Operating Instructions61211-1020202
Preparations for start-up
Installing a probe rack
In the case of the CenterMax, two different positions on the guide­way side can be used for installing the profile rail. If you only have one profile rail, you must mount it at the lower position first. This is neces­sary for reasons of safety.
A
1
B
2 3
A CenterMax: top plan view. B Guide side
1 Holders (storage positions) 2 Mounting option for upper position 3 Mounting the profile rail in the lower position
5-12
61211-1020202 CenterMax Operating Instructions
Mounting the probe rack (option)
1
2
3
NOTE
1 Threaded holes provided on the guide side 2 Threaded bolts 3 Fastening of the profile rail
Depending on whether the profile rail is to be mounted at the top or at the bottom, threaded bolts of varying length are required.
If you mount the profile rail at the top, you have to use long threaded bolts.
If you mount the profile rail at the bottom, you have to use short threaded bolts.
Three threaded bolts are required per profile rail. Two threaded bolts are screwed into the threaded holes at the front and one threaded bolt is screwed into the threaded hole at the rear.
If you require more holders than can be mounted on two profile rails, it is possible to mount a third profile rail on the drive side. However, this will decrease the effective measuring range.
5-13CenterMax Operating Instructions61211-1020202
Preparations for start-up

Visual check

NOTE
The visual check must be part of the routine work to be carried out during the daily measuring operation.
Housing covers
The CMM may only be operated if all CMM housing covers are mounted.
Mount all housing covers before taking further measures.
Cable for power supply
The cable must be in perfect condition. It must not be ...
...bent or
...damaged.
Make sure that the cable is in perfect condition.
Ensure that no heavy objects are lying on the cable.
Compressed-air line
The line and connections must be in perfect condition. The line must not be ...
...bent or
...damaged.
Make sure that the line is in perfect condition.
Make sure that ...
no heavy objects are lying on the line.
the line is connected correctly to the supply line and the CMM.
Cables
Cable duct The cables from the control cabinet to the CMM are installed in a
cable duct.
Make sure that the cable duct is not damaged.
5-14
61211-1020202 CenterMax Operating Instructions

Start-up preparation checklist

Start-up preparation checklist
Housing covers
Are all CMM housing covers mounted?
Power supply
Is the power supply connected?
Are the cables in perfect condition?
Compressed-air supply
Is the compressed-air supply connected?
Does the compressed-air supply have sufficient pressure?
Is the required pressure indicated?
Probe rack (optional)
Have all requirements regarding the probe rack been met?
Has the probe rack been installed correctly?
Is probe change possible without collision?
5-15CenterMax Operating Instructions61211-1020202
Preparations for start-up
5-16
61211-1020202 CenterMax Operating Instructions
Chapter

6 Start-up

.................................................................................................................................
This chapter describes the work to be carried out regularly.
You want to clamp a new workpiece on the measuring table or pre­pare the probe system for the measuring run? Or you want to switch the CMM on again after it has been put out of operation? The follow­ing pages provide the corresponding information.
Furthermore, the procedure for checking the protective circuit for the drives is described.
This chapter contains:
Clamping the workpiece on the measuring table . . . . . . . . . . . . . . . 2
Switching the CMM on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Checking the protective circuit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Mounting / changing the probe head. . . . . . . . . . . . . . . . . . . . . . . 11
Probe assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Inserting / removing the probe. . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Start-up checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
6-161211-1020202 CenterMax Operating Instructions
Start-up

Clamping the workpiece on the measuring table

Safety measures

!
!
!
Danger!
You are exposed to the danger area of the CMM when working on the measuring table. Certain safety measures have to be observed to avoid injuries.
Switch off the drives to prevent unintentional travel movements.
Activate the "crane interlock when using lifting equipment to
place the workpiece on the measuring table. For more information see ➤ Operating instructions for the measur-
ing software.
Danger!
There must not be any grease film on the measuring table. Work- pieces may slide and fall off of the measuring table, thus endangering personnel. Make sure that the measuring table is free of lubricants before placing a workpiece on top of it.
Remove any greasy or oily substances from the measuring table.
Important!
The workpieces must have full contact with the measuring table. Any hard particles lying underneath the workpiece may lead to inaccurate measurements. Furthermore, the surface of the measuring table and the workpiece may be damaged.
6-2
Clean the measuring table before placing a workpiece on the mea­suring table.
61211-1020202 CenterMax Operating Instructions
Clamping the workpiece on the measuring table
!
!
Important!
Certain safety measures have to be observed to avoid damage to the CMM and the measuring table.
Observe the permissible workpiece weight page 3-2.
Lower the workpiece onto the measuring table carefully. Lower
heavy workpieces at a maximum speed of 2mm/s. The CMM may be damaged if the workpiece is set down hard.
Make sure that the workpiece or lifting device does not strike the bridge, quill, or drive side of the CMM.
Even light jolts may affect the functioning of the CMM.
Place the workpiece upright on the measuring table.
Do not slide the workpiece on the measuring table.
The surface of the table could be scratched when sliding the work­piece, thus impairing the measuring accuracy of the CMM.

Clamping the workpiece

Important!
The maximum permissible torque for the threaded holes in the mea­suring table is 40Nm.
Tighten the screws to a maximum torque of 40Nm.
NOTE
Metal, plastic, wood Chucks are generally made of metallic materials. Chucks made of
Workpieces made of soft materials may be deformed by chucks, thus falsifying the results of the measurement.
If necessary, use special chucks to avoid deforming of the work­pieces.
Chucks
Chucks are screwed onto the measuring table using the threaded (M12) holes.
plastic or wood may also be used if necessary.

Workpiece positioning

Position the workpiece so that all required measurements can be car­ried out without changing the workpiece position.
6-3CenterMax Operating Instructions61211-1020202
Start-up
Measuring operation without collision
When using a probe rack, position the workpiece and probe rack so that measurements and probe change can be carried out without col­lision.
Arrange the workpiece and probe rack so that no collisions can occur during the measuring run.
For more information on installing a probe rack,, please refer to .

Lowering the workpiece onto the measuring table

Workpieces can be lowered onto the measuring table either manually or by means of a lifting equipment.
Crane lock Activate the "crane interlock when using lifting equipment to lower
workpieces onto the measuring table. The crane lock is an option. This option requires additional hardware and software. For more information, please refer to the operating instructions for the measur­ing software.
Heavy workpieces
Heavy workpieces have a hard impact if lowered into the measuring table too quickly.
Lower heavy workpieces at a maximum speed of 2mm/s.
Elastic supports First place the workpiece on elastic supports.
Then replace the elastic elements with rigid supports.
NOTE
Metal blocks clamped by clamping devices can be used as rigid sup­ports. Screw fasteners can be used alternatively. These fasteners can be screwed into the threaded holes in the measuring table.
6-4
61211-1020202 CenterMax Operating Instructions

Switching the CMM on

Safety measures

Switching the CMM on
!
Danger!
A visual check must be carried out before switching the CMM on. The CMM may be switched on only if it is in perfect condition.
Make sure that the CMM is in perfect condition “Visual check on page 5-14.

Switch-on sequence

Proceed as follows when switching the control on:
1 Connect power supply; switch power supply on by means of the
main switch
2 Switch the control on.
3 Switch the drives on.
4 Select the operating mode
5 Switch the computer on and start the measuring software.
6-5CenterMax Operating Instructions61211-1020202
Start-up
For more information, please see the following pages.
SCHNELLSTOP
quickstop mode - switch
1 Main switch for power supply 2 Rotary switch for the control 3 Indicator lamp for control; green 4 Rotary switch for drives 5 Indicator lamp for drives; green 6 Key-operated switch for operating modes
BETRIEBSARTEN
0125853
h
N
MA
0
AUTO
ANTRIEBE
drives
53
NETZ/STEUERUNG
START start
mains / control
AUS off
EIN on
HAUPTSCHALTER
main - switch
1642
NOTE
For more information on the operating elements, please refer to the chapter Operating elements of the control cabinet on page 2-8.
Connecting the power supply
Turn the main switch (1) clockwise to position "1“. The power supply for all electronic components of the CMM is
switched on.
Switching the control on
Turn the rotary switch (2) to the "ON" position. The indicator lamp above the switch lights up. The control is
switched on.
Wait for 30 s • Wait for approx. 30s before switching the drives on.
This is necessary since the internal computer requires a certain amount of time for booting. The booting process must be termi­nated before the drives are switched on. The LEDs on the control panel flash during the booting process.
6-6
61211-1020202 CenterMax Operating Instructions
Switching the CMM on
Switching the drives on
Turn the rotary switch (4) clockwise approx. 90° and hold it there until the indicator lamp above it lights up.
As soon as the indicator lamp above the switch lights up, the drives are switched on. The rotary switch does not engage. snaps back to its home position as soon as it is released.
Selecting the operating mode
It is also possible to set the operating mode later.
Turn the key-operated switch (6) to the required position.
MAN: Manual mode – Set-up mode. The travel speed is limited in
all three axes. You must set this operating mode for …:
– Joystick-controlled probing
!
!
NOTE
Programming CNC runs
Testing CNC programs.
AUTO: Automatic mode – series measurement mode. Travel at
maximum speed in all axes. You must set this operating mode for …:
– Workpiece measurement by means of a CNC program.
Danger!
When starting a CNC program, make sure that no one is in the dan­ger area of the CMM. The operating mode setting (MAN or AUTO) is irrelevant.
Check to make sure that no one is in the danger area of the CMM.
Important!
The AUTO mode may only be set if the CNC program was tested in the manual mode beforehand. This prevents collisions at high travel speeds in the automatic mode.
When starting a CNC program, the key should be removed and stored in a safe place to prevent the setting from being changed accidentally.
Switching the computer on
Switch the computer on and other peripheral devices, e.g. the printer.
Boot the computer; start the operating system.
Start the measuring software.
6-7CenterMax Operating Instructions61211-1020202
Start-up

Checking the protective circuit

The CMM is equipped with a protective circuit which monitors drive­forces during bridge positioning movements in the x and y axes. The shearing force monitoring is effective in all operating modes.

Function of the protective circuit

Drives must cut out automatically.
!
If the traveling motion is obstructed by a collision in the x or y axis and the shearing force therefore exceeds a value of approx. 150N, the drives must cut off automatically.
Danger!
In order to ensure the safety of the CMM during measurement, the function of the protective circuit must be checked regularly.
Check the protective circuit on a monthly basis.

Check

When performing the check, you must proceed as follows:
x axis
Move the crossbeam forward just far enough so that you can grasp the quill housing with one hand.
6-8
61211-1020202 CenterMax Operating Instructions
Checking the protective circuit
Move the probe (head) in the x axis at constant speed.
Try to stop the travel motion (v
the quill housing at the marked position (F).
This increases the shearing force of the drives. Once the shearing force has exceeded a certain value, all drives will be deactivated.
v
x
vx: Speed of quill in x axis
F: Force exerted against the quill housing by pressing it with one hand.
F
) by pressing one hand against
x
!
Check by a ZEISS service engineer
!
Reactivating the drives If you can stop the travel motions, you must reactivate the drives
Important!
You may press only against the quill housing.
Do not press against the quill or against the probe head. The probe head and quill could otherwise be damaged.
Under normal circumstances, the force of your hand should suffice to stop probe movement. If the movement can not be stopped in this way, the protective circuit must be checked immediately by a ZEISS service engineer.
Call your service engineer and arrange for him to check the protec­tive circuit.
Danger!
If the protective circuit is not functioning properly, the CMM must be taken out of operation until the protective circuit has been checked by a ZEISS service engineer.
Operational controls.
6-9CenterMax Operating Instructions61211-1020202
Start-up
y axis
Move the control panel to the drive side for y-axis travel. The drive side is located on the right side of the CMM.
!
v
y
vy: Quill speed in y axis
F: Force which must be exerted manually against the crossbeam on the drive side.
Perform crossbeam travel in the y axis at constant speed.
While the crossbeam is traveling, try to stop its movement by
pressing against the crossbeam with one hand.
Important!
You may press against the crossbeam only on the drive side.
F
6-10
Do not press against the probe head and quill or against the cross- beam on the guideway side.
The probe head, quill and crossbeam could be damaged.
61211-1020202 CenterMax Operating Instructions

Mounting / changing the probe head

Mounting / changing the probe head
!
Important!
With the CenterMax, the probe head must not be changed by the operator. The probe head may be changed only by a ZEISS service engineer.
6-11CenterMax Operating Instructions61211-1020202
Start-up

Probe assembly

Screwing technique or clamping technique?

Although the probe can be assembled by clamping its parts together, it is preferable to use the screwing technique.
Advantages of screwing technique:
Easy assembly
Complex probe configurations possible
Stable design.
The following example describes how to assemble probes using the screwing technique.
!
!
!
For further information please refer to the chapter “Storage of probes on page 2-24.

Safety measures

Important!
Probes must not be assembled on the probe head. The probe head could thus be damaged.
Do not attach any probe elements to the probe while it is inserted in the adapter plate receptacle.
Important!
Assemble the probe configuration carefully. Do not exert any force when screwing the components together.
Screw the components together manually. Assemble them fully.
Use the pin included in the probe kit to tighten the components.
Important!
Observe the limiting values regarding weight and length when assem­bling the probe. The limiting values depend on the probe system.
6-12
Observe the limiting values Limiting values on page 6-14.
61211-1020202 CenterMax Operating Instructions
Probe assembly

Criteria and limiting values for assembling probes

Criteria
An unfavorable probe assembly may impair the measuring accuracy. Therefore, it is necessary to observe the following criteria for assem­bly.
Only use probe elements with an M5thread.
Stable design with as few components as possible.
Symmetrical design with respect to weight; the center of gravity
must be located in the physical center; use counterweights if nec­essary.
Probe extensions may be used as counterweights.
NOTE
Check the balance by holding the adapter plate between two fingers and letting it swing back and forth until it comes to a stop. The probe must point in a vertical downward direction.
Do not assemble probes with more than one branching.
If possible, never use joints when assembling oblique probe config-
urations. It is preferable to use components shaped according to your corre-
sponding requirements. You can also make such components your­self, provided that you observe the permissible limiting values.
Use only components that are in perfect condition. Requirements for probe components:
6-13CenterMax Operating Instructions61211-1020202
Start-up
perfect condition
clean
grease-free
Observe the limiting values; probe length and probe weight.
Limiting values
The maximum weight and length must not be exceeded in order to ensure error-free measurement by the probe system.
The limiting values depend on the probe system used.
Maximum values VAST
Probe weight – including adapter plate
G
Probe length – including extension:
L
600g
450mm

Example

Connecting parts are usually required to assemble the required probe configuration. It is also possible to screw the probe directly into the adapter plate page 6-15.
6-14
61211-1020202 CenterMax Operating Instructions
Probe assembly
Screwing technique
Example of probe assembly via the screwing technique using the VAST adapter plate:
AB
A Probe stylus in adapter plate B Extension in adapter plate
6-15CenterMax Operating Instructions61211-1020202
Start-up

Inserting / removing the probe

Safety measures and information

The probe is held in place by an electromagnet or solenoid. If the measuring software initiates insertion or change of the probe, the magnetic force is reduced. The full magnetic force becomes effective after 20 seconds.
The measuring software is required for inserting and changing a probe.
Please also read the corresponding chapters in the operating instructions for the measuring software.
!
!
Important!
Probes must not be assembled on the probe head. The probe head could be damaged.
Insert only finished probes equipped with all necessary probe ele­ments.
Do not attach any probe elements to the probe as long as it is inserted in the adapter plate receptacle.
Important!
A VAST probe automatically falls off of the adapter plate receptacle after a specified period of time.
Hold the probe immediately after activating change in order to avoid damage to the probe, the workpiece or the measuring table.
The period after which the probe falls off is set in the measuring soft­ware.

Inserting the probe

Operating instructions for the measuring soft­ware
6-16
The procedure for inserting a probe is basically the same for all probe systems. This procedure must be initiated by the measuring software.
Please read the corresponding chapters in the operating instruc­tions for the measuring software.
1 Initiate the procedure by means of the measuring software.
Then, insert the probe within 20 seconds. If the probe is not inserted within this time, you have to repeat this procedure.
61211-1020202 CenterMax Operating Instructions
Inserting / removing the probe
2 Insert the probe in the adapter plate receptacle.
The pin must engage in the groove drawing. The probe is attracted by the magnet. The completion of probe insertion is sig­naled by an audible click.
Position of the pin The pin position in the adapter plate receptacle depends on the
probe system.
VAST: pin on the left side.
1 Pin in the probe head adapter plate receptacle 2 Groove in the adapter plate.

Removing the probe (manually)

The procedure for removing a probe is basically the same for all probe systems. However, several special points must be observed. First, the procedure must be initiated by the measuring software.
Please read the corresponding chapters in the operating instruc­tions for the measuring software.
1 Initiate the procedure via the measuring software.
With the VAST probe system, the delay required until the probe is released from the adapter plate receptacle must be specified in the measuring software.
Set the delay so that enough time always remains to grasp the probe after initiating probe removal. This is essential to prevent damage to probes, workpieces, or the measuring table caused by falling probes.
2 Remove the probe.
6-17CenterMax Operating Instructions61211-1020202
Start-up
Special features of probe systems
VAST: Hold the probe; it falls off automatically.
Automatic probe change
Information on automatic probe change is provided in the chapter
Automatic probe change on page 7-30.
6-18
61211-1020202 CenterMax Operating Instructions

Start-up checklist

Clamping the workpiece
Is the workpiece in the correct position? page 6-3.
Is it possible to measure all workpiece dimensions in one cycle?
Is measurement possible without collision? E.g. collision with the
probe rack.
– Is the workpiece clamped? page 6-3.
Switching on the CMM
Has the CMM been switched on in the correct order? page 6-5.
Is the power supply connected?
Control cabinet switched on?
Start-up checklist
Drives switched on?
Operating mode selected?
Computer switched on?
Measuring software started?
Protective circuit for the driving forces
Has the protective circuit been checked? page 6-8.
Probe assembly
Have the criteria for probe assembly been observed? page 6-13.
Have the limiting values for the probe been observed?
page 6-14.
6-19CenterMax Operating Instructions61211-1020202
Start-up
6-20
61211-1020202 CenterMax Operating Instructions
Loading...