Knowledge of this manual is required for the operation of the instrument. Therefore please make
yourself familiar with the contents of this manual and pay special attention to hints concerning the
safe operation of the instrument.
The specifications are subject to change; the manual is not covered by an update service.
Unless expressly authorized, forwarding and duplication of this document, and the utilization and
communication of its contents are not permitted. Violations will entail an obligation to pay
compensation.
All rights reserved in the event of granting of patents or registration of a utility model.
Carl ZeissGeneral Notes, Instrument Safety and EMCAxiostar plus
General Notes on Safety
Before starting up the microscope, please
ensure that you have familiarized yourself
thoroughly with the contents of this manual.
Further information can be obtained from our
service department or from authorized
representatives.
The Axiostar plus microscope is an optical
precision instrument which may be impaired in
its performance or damaged if handled
improperly.
The microscope must be operated by trained
personnel only who must be aware of the
possible problems involved with microscopy and
who are trained in the relevant application.
CAUTION
Hot surface; there is a risk of burns.
CAUTION!
If the safety notes are not observed, there is a
risk that the instrument will be destroyed.
☞
Notes that must be observed when working
with the microscope.
NOTE!
!
To ensure that the working procedure is safe
and that the microscope functions without
problems, the notes and warnings included in
the operating instructions must be observed.
These are highlighted in the text by the
following symbols:
CAUTION!
If the safety notes are not observed there is a
risk for the user.
CAUTION!
Disconnect the instrument from the line.
Instrument Safety and EMC
The Axiostar plus microscope was designed,
produced and tested in compliance with DIN EN
61010-1 (IEC 1010-1) “Safety requirements for
electrical measuring, control and laboratory
instruments“. It left the factory in a perfectly
safe condition.
The Axiostar plus microscope meets the
requirements of the EC directives 73/23/ECappendix 1 and 89/336/EC and the EMC
legislation of 09/18/1998:
Radio interference suppression in accordance
with EN 55011 class B
Resistance to jamming in accordance with EN
50082-2
Conformity with the above EC directives is
documented by the CE symbol.
Please observe the following safety notes when
unpacking, transporting and storing the
Axiostar plus microscope:
– In accordance with standard practice, the
microscope is supplied in a plastic container
with cardboard packaging; only use the
original packaging for transport.
– Retain the packaging for longer storage
periods or to return the instrument to the
manufacturer.
– When unpacking, use the delivery note to
ensure that all configurations and modules
are present.
– Observe transport and storage tempera-
tures in accordance with the technical data.
– Place the microscope on a stable work
surface.
Operation
Please observe the following safety notes when
using the Axiostar plus microscope:
– The microscope can be operated with a line
voltage of 100 − 240 V.
– The microscope must be connected only to
a properly installed outlet featuring a
grounding contact; the grounding effect
should not be made ineffective with an
extension cable that does not have a
protective ground wire.
– If it is established that the protection meas-
ures are no longer effective:
⇒
switch off the microscope
⇒
safeguard against inadvertent operation
⇒
contact an authorized workshop or the
manufacturer for repairs.
– Keep optical surfaces free of fingerprints.
Disposal
When disposing of the Axiostar plus
microscope, please observe the following safety
notes:
– Defective microscopes should not be placed
in ordinary domestic waste;
they should be disposed of in accordance
with the relevant statutory provisions.
– Always disconnect the instrument from the
line before opening the instrument,
changing the lamps or changing the fuses!
– Only use fuses for the required rated
power!
– The use of makeshift fuses and the short-
circuiting of the fuse holders are not
permitted!
Be careful of light-emitting diode radiation
when using the LED illuminator. Do not look
directly into the illuminator with your bare eyes
from a short distance. When adjustments are
required, the intensity must be reduced
accordingly via the potentiometer, or a suitable
attenuator must be used.
Carl ZeissCare and Maintenance, Warranty NotesAxiostar plus
Care and Maintenance
Please observe the following safety notes when
caring for and maintaining the Axiostar plus
microscope:
– With the exception of the work specified in
Chapter 4 of this manual “Care,
Maintenance and Troubleshooting“, no
maintenance or repair may be performed by
the user. We expressly indicate that any
other repairs must only be carried out by
our authorized personnel.
Damaged instruments or components must
–
only be repaired and maintained by our
service department.
– Protect the microscope against dirt, dust
and moisture; these influences can impair
the performance of the instrument.
– If the microscope is not in use, protect it
with the dust cover and make sure that it
has been switched off.
– Lamps are to be changed in accordance
with the steps in Chapter 4 “Care,
Maintenance and Troubleshooting“.
Notes on Warranty
The Axiostar plus microscope, including its
original accessories, may only be used for the
microscope techniques described in this
manual. The manufacturer cannot assume any
liability for any other applications.
Please observe the following warranty notes for
the Axiostar plus microscope:
– The manufacturer guarantees that the
instrument has no material and production
defects when delivered.
– You must inform us of any defects
immediately and do everything possible to
minimize the damage.
– If the manufacturer is informed of such a
defect, he is obligated to remove it; it is his
decision whether he does this by repairing
the instrument or by delivering an
instrument free of any defect.
– No guarantee is provided for defects caused
by natural wear (wearing parts in particular)
or improper use.
– The manufacturer is not liable for damage
caused by faulty operation, negligence or
any other tampering with the instrument, in
particular as a result of removing and
exchanging microscope components or the
use of accessories from other
manufacturers.
Axiostar plusIntended application, Main featuresCarl Zeiss
1 INSTRUMENT DESCRIPTION
1.1 Name and intended application
Manufacturer‘s name:
Axiostar plus microscope
The Axiostar plus is a transmitted light
microscope for the visualization of fine
structures and forms in biology and medicine.
Typical applications of the Axiostar plus include:
– Histology, pathology,
– Doctors’ offices,
– Training (schools and universities),
– Routine laboratory microscopy,
– Immunofluorescence and FISH methods,
– Field use with LED illumination.
Key to Fig. 1-1
1Eyepieces
2Binocular tube
3Microscope stand
4Knurled screw for tube locking
5Brightness control
6On/off switch with integrated signal lamp
7Fine focusing drive (two-way)
8Coarse focusing drive (two-way)
9Drive for adjusting mechanical stage
in X direction
10Drive for adjusting mechanical stage in Y direction
11Luminous field diaphragm
12Condenser carrier
13Lever for adjusting aperture diaphragm
14Condenser
15Centering screw for condenser (two-way)
16Mechanical stage with specimen holder
17Objective
184- or 5-position nosepiece
1.2 Description and main features
Thanks to its pyramid design, the Axiostar plus
is a compact transmitted-light microscope.
In addition to high-resolution ICS objectives and
the major brightfield, darkfield, phase contrast,
polarization contrast and epi-fluorescence
techniques, an optional camera port for photo
and video documentation is also available to
the user.
Major instrument features:
– Sturdy and convenient stand in the pyramid
design
– Ergonomic base plate with hand rests and
adjustable viewing height and angle
– User-friendly coaxial coarse and fine drive
– Mechanical stage 75 × 30 R/L with ceramic-
coated stage surface and specimen holder;
selectable stage drive right or left
– Space-saving and continuously adjustable
integrated illuminator with long-life 6 V,
20 W halogen lamp
– 0.9/1.25 condenser for brightfield, darkfield
and phase contrast Ph 1, 2, 3
– Ball-bearing, 4- or 5-position objective
revolver with W 0.8 thread
– ICS objectives in the price/performance
categories CP-Achromat, A-Plan and
Achroplan in finely graduated rows
– Binocular tube 45°/20 ICS with a viewing
height of 425 − 470 mm and binocular
phototube 45°/20 ICS with folding prism
100% vis / 100% doc
– Binocular tube 30°/20 ICS with a viewing
height of 430 – 475 mm and binocular
phototube 30°/20 ICS with folding prism
100 % vis / 100 % doc
– 10× eyepieces for the field numbers 20 or
18, suitable for eyeglass wearers, fixed and
adjustable
The objectives are the optical centerpiece of
the microscope.
The following is an example of how objectives
can be labeled:
CP-ACHROMAT 10×/0.25 ∞/-.
where:
10× = objective magnification, with a defined
color ring on the objective
allocated to each magnification step
(Carl Zeiss color code)
0.25 = numerical aperture
∞ = infinite tube length
− = can be used with cover slip thickness
D = 0 or 0.17 mm
or
0.17 = can be used with cover slip thickness
D = 0.17 mm
and
Oil =oil immersion objective
Ph 2 = phase contrast objective with green
objective labeling and annular
diaphragm Ph 2
The objective magnification multiplied by the
eyepiece magnification (usually 10×) results in
the visual overall magnification:
e. g. 10 × 10 = 100×.
☞
The objective labeling "∞" indicates that these
objectives may only be used with microscopes
featuring infinite optics (ICS) and not with
instruments whose objectives are marked with
"160" as their mechanical tube length (in
mm).
The exact observance of the cover slip thickness
of 0.17 mm is all the more necessary the
higher the numerical aperture of the
objective. Therefore, so-called "Corr"
objectives can be set for different cover slip
thickness via a correction ring. For this, a
specimen area is searched, and the position of
the correction ring where optimum focus and
image contrast are obtained is determined
(refocusing is always required).
Immersion objectives are always insensitive to
differences in cover slip thickness.
When immersion objectives are used, the air
between the cover slip and the objective is
replaced with a liquid, which is immersion oil
in most cases. The plastic oiler containing
20 ml of 581 N Immersol is particularly
suitable for this purpose.
Due to their short working distance, objectives
25× and higher feature resilient mounts
(specimen protection).
The numerical aperture × 1000,
e. g. 0.25 × 1000 = 250×, is the highest
useful magnification, i.e. no further details are
resolved above that limit.
Carl ZeissEyepieces, Stage micrometers, Eyepiece reticlesAxiostar plus
Crossline disk / d = 26 mm474064-0000-000
0
1
2
3
4
5
6
7
8
9
10
Crossline micrometer disk 10:100 / d =
26 mm
gradation length = 10 mm
increments = 0.1 mm
gradation tolerance ≤ 0.001 mm
Net micrometer
474066-9901-000
474068-0000-000
12.5××××12.5/5;10 / d = 26 mm
area 12.5 ×××× 12.5 mm, divided in fields of 5 ×××× 5 or
10 ×××× 10
Photo reticle MC 2.5×××× / d = 26 mm
454075-0000-000
for 35mm photography with an additional
magnification of 2.5×××× or for large-format
photography with a 10×××× additional magnification
Fixed pointer for focusing eyepiece /
000000-1095-418
d = 26 mm
To point toward objects in the specimen
☞
If an eyepiece reticle is used, the binocular tube or the phototube must be equipped with two
foc. eyepieces containing an adjustable eye lens, into one of which the eyepiece reticle is
mounted.
Average life .................................................................................................................1000 h
Luminous area ...................................................................................................2.0 × 2.0 mm
LED Illumination
Constant, brightness independent color temperature of................................................7480 K
Homogenous image field illumination ................................................... up to 20 mm diameter
Suitable for objectives with magnification of .......................................................2.5 x to 100 x
Analog brightness control of ...................................................................approx. 15 to 100 %
Operation with 9 V alkaline battery / 9 V NiCd Accublock:
Operating voltage.......................................................................... approx. 4.5 to 10 V
Current at maximum brightness depending on operating voltage of:
Operating voltage 4.5 V.................................................................................... 90 mA
Operating voltage 9 V....................................................................................... 51 mA
Operating voltage 10 V..................................................................................... 45 mA
Average life with 9 V alkaline battery (depending on battery capacity):
Maximum intensity................................................................................. approx. 6.5 h
65 % of maximum intensity..................................................................... approx. 16 h
50 % of maximum intensity..................................................................... approx. 21 h
Operation with 12 V battery / Accu (voltage between 9-13.5 V):
Operating voltage 9 V....................................................................................... 88 mA
Operating voltage 12 V..................................................................................... 60 mA
Operating voltage 13.5 V.................................................................................. 50 mA
Fluorescent Lamp HBO 50
Lamp voltage.............................................................................. L1: 39 45 V / L2: 34 to 39 V
Lamp current ....................................................................................... L1: 1.30 A / L2: 1.45 A
Power ........................................................................................................................... 50 W
Average light current .................................................................................................2000 lm
Average life ...................................................................................................................100 h
Illumination surface................................................................................................ 0.3x1 mm
Power supply mbq52ac-z for HBO 50
Line voltage ................... 100,110,120,127,230, 240 V, ±10 % AC, can be switched externally
Line frequency ............................................................ 50 to 60 Hz, can be switched externally
Power consumption. ...........................................................................................max. 350 VA
1
2
1
For photography using artificial light color reversal film for 3200 K, the conversion filter CB 3 (467852)
produces the correct color temperature in the light path.
Axiostar plusAttaching and adjusting ergonomic base plateCarl Zeiss
2 START-UP
☞
Before installing and starting-up the Axiostar
plus, carefully read through and observe the
notes on safety (page VI ff.).
To prevent fingerprints, do not touch optical
surfaces when unpacking!
2.1 Unpacking the instrument
The Axiostar plus microscope, including
accessories, is delivered in standard packaging.
• Remove the microscope from the transport
case and place it on the work table.
• Keep the packaging so that the instrument
can be stored for a longer period of time or
returned to the manufacturer.
2.2 Attaching and setting
ergonomic base plate
Using the Axiostar plus with ergonomic base
plate allows you to set the microscope to your
individual ergonomic requirements.
An experienced technician can attach the
Axiostar plus (000000-1169-149, -150, -151) or
Axiostar (000000-1122-098, -099, -100) to the
ergonomic base plate on his own. However, we
recommend asking another person to help.
Before attaching the ergonomic base plate,
unplug the microscope.
Carl ZeissAttaching and adjusting ergonomic base plateAxiostar plus
2.2.2 Adjusting the ergonomic base plate
Adjusting viewing height
The viewing height of the microscope can be
adjusted by turning the feet (2-3/3) on the base
plate in four levels. The height difference is 10
mm per level (total lift 30 mm).
When the viewing height is changed, the slope
of the hand rest on the ergonomic base plate is
changed automatically. The adjustment range is
approximately 6.5°.
When setting the viewing height, there is a risk
of getting your hands caught between the table
and ergonomic base plate or adjustment feet;
therefore, do not place your hands under the
ergonomic base plate or adjustment feet when
setting it down.
Setting viewing angle (microscope slope)
The slope of the microscope in relation to the
table surface or hand rest can be set according
to your needs and application. There are four
settings (adjustment range 6.5° in steps of
2.2°).
In one setting, the microscope is basically
parallel to the table surface.
• Grasp the microscope at the top of the
stand (2-3/1)and lift it by tilting it back
slightly; while doing so, press the locking
button (2-3/4) until it unlocks.
• Tilt the microscope backward or forward
until you reach the desired angle.
• Let go the the locking button when you
reach the desired position. When the
button engages, let go of the stand.
• To set the height, lift the entire unit with
the handle on the back of the ergonomic
base plate (2-3/2). The adjustment feet
(2-3/3) will disengage and can be adjusted.
• Select desired height by turning the
adjustment feet, then lower the entire unit.
The adjustment feet will engage in position
automatically.
One knurled screw and one hexagonal screw
each are supplied with the mount. Any of these
screws can be used to clamp the tube
(phototube) to the stand.
• Screw the required screw into the threaded
hole (2-4/6) in the stand.
2.3.1 Attaching binocular tube 45°/20 ICS
• Remove dust cover (2-4/1) from tube
underside. Unscrew the knurled screw
(2-4/5) or the hexagonal screw (2-4/4) from
the stand.
• Press the binocular tube (2-4/2) in a slightly
inclined position to the right into the spring
of the stand.
1Dust cap
2Binocular tube
3Binocular phototube
4Hexagonal screw
5Knurled screw
6Threaded hole for knurled or hexagonal screw
Fig. 2-4Attaching binocular tube / phototube 45°
• Then set the tube in the dovetail of the
stand by lowering the left side. The tube is
now secured from falling out.
• Turn the tube to the desired observation
position; both the illustrated rotation and
the 180° backward rotation are possible.
• Tighten the knurled screw (2-4/5) or the
hexagonal screw (2-4/4).
2.3.2 Attaching binocular phototube
45°/20
ICS
• Remove dust caps (2-4/1); loosen knurled
screw (2-4/5) or hexagonal screw (2-4/4).
• Attach the binocular phototube (2-4/3) to
the stand mount in the same way as the
binocular tube.
Using the binocular tube/phototube 30° ICS
with the ergonomic base plate offers the best
ergonomic conditions for microscope use. The
adapter for 30° tubes is required to attach the
binocular tube 30°
first remove the 45°
ICS to the stand. If necessary,
ICS tube from the stand.
• Remove dust cap from tube underside.
Loosen knurled screw (2-5/4) or hexagonal
screw on the stand.
• Press the adapter for 30° tubes (2-5/2)
slightly to the right into the spring of the
stand.
• Then set the adapter in the dovetail of the
stand by lowering the left side. This secures
the adapter from falling out.
• Turn the adapter to align the outer edges
parallel to the stand; the clamping screw for
the 30° tube should point to the right.
• Tighten knurled screw (2-5/4) or hexagonal
screw.
• Place the 30°
ICS tube (2-5/1) in the adapter
(2-5/2), align the outer edges and tighten
clamping screws (2-5/3) on the adapter.
Proceed the same way when attaching the
phototube 30°
The eyepieces PL 10×/18 Br. foc. and
E-PL 10×/20 Br. foc. are intended for use with
eyepiece reticles (see chapter 1.6).
The slight image shift caused by the additional
path through the glass is taken into account
on the diopter scale by the fact that the zero
point position is indicated not by the white
dot (2-8/W) but by the red dot (2-8/R).
The eyepiece reticles (2-8/3) have been
attached to screw-in mounts (2-8/4) to allow
easy replacement. Complete mounts with
eyepiece reticles can be obtained directly from
Zeiss. When reordering reticles for already
available eyepieces 444232-9902-000, you
must also order the mount 444232-0206-000.
For eyepieces 444132-9901-000, order the
mount 444132-0206-000. The eyepiece
reticle will be delivered already in the mount.
To replace the mount:
• Unscrew the mount containing the
eyepiece reticle from the eyepiece.
• Screw in the new mount containing the
required eyepiece reticle.
RZero point position of diopter scale with
eyepiece reticle
WZero point position of diopter scale without
eyepiece reticle
1Eyecup
2Spectacle protection ring
3Eyepiece reticle
4Mount
Fig. 2-8Inserting eyepiece reticle
2.5.2 Compensation of ametropia
when eyepiece reticles are
used
The correct use of an eyepiece reticle requires
two focusing eyepieces, e. g. PL 10×/18
Br. foc., to make it possible to compensate for
possible ametropia of the observer's eyes.
• Use the eyelens of the focusing eyepiece
to focus on the line figure of the eyepiece
reticle.
• Focus on the microscope image of a
specimen via the focusing drive by looking
through the eyepiece with reticle.
• When the image and the eyepiece reticle
are in focus in this eyepiece, focus the
image for the second eye via the focusing
eyelens of the second eyepiece.
Both microscope images and plates should be
focused.
Now focusing should occur only with the
focusing drive.
Carl ZeissSetting interpupillary distance and viewing heightAxiostar plus
2.5.3 Attaching folding eyecups
The eyepieces are equipped with rubber rings
to protect eyeglasses from scratches. These
rings can be replaced with folding eyecups, if
desired.
• Remove the rubber rings (2-8/2) from the
eyepieces and attach the eyecups (2-8/1).
Sometimes, the spectacle protection rings fit
very tightly in the eyepiece groove so that a dull
object (e.g. tooth pick) is required to press
them out.
2.6 Setting interpupillary distance
and viewing height
• The eyepiece distance is matched to the
individual interpupillary distance of the
observer by swinging the eyepiece tubes
symmetrically towards one another
(Fig. 2-9).
Fig. 2-9Setting interpupillary distance
Fig. 2-10 Setting viewing height
The interpupillary distance is correct if you see
only one round image when you look in both
eyepieces!
• The viewing height can be matched to the
requirements of the individual user by
swinging the eyepiece tubes up
(2-10/A) or down (2-10/B).
Attachment is made in the same way as the
Abbe condenser 0.9/1.25.
• Loosen both centering screws (2-14/3) on
the condenser carrier (2-14/1).
• Press condenser with dovetail against the
spring pin (2-14/2) in the condenser carrier.
Make sure that the orientation screw
(2-14/5) on the condenser engages in the
groove (2-14/4) of the condenser carrier.
• Tighten both centering screws (2-14/3) on
the condenser carrier (2-14/1) until they
engage in the dovetail and hold the
condenser (2-14/7) in position.
condenser carrier (2-15/1) in the same way
as the Abbe condensers.
• Tighten both centering screws (2-15/2) until
the LED illuminator lies approximately in the
center of the beam path.
• Use condenser drive (2-15/4) to move the
condenser carrier into the topmost position.
Power to the LED illuminator can be supplied
either via the line power unit or, independent
of the line, via a commercially available 9 V
Alkaline battery (9 V / Ni Cd battery block) in
the integrated battery compartment, or via an
external voltage source ranging from 11 to
14 V DC (e.g. 12 V car battery or battery
pack C). The external voltage source must
permit a constant current of up to approx.
120 mA to be loaded.
The LED illuminator provides pleasant, colorneutral white light with a constant color
temperature (approx. 7500 K) independent of
the set brightness. It is suitable for objectives
ranging from 2.5x to 100x. The light-emitting
diodes have a long life and cause only minor
operating costs.
Operation with power unit
Be careful of light-emitting diode radiation
when using the LED illuminator. Do not look
directly into the illuminator with your bare eyes
from a short distance. When adjustments are
required, the intensity must be reduced
accordingly via the potentiometer, or a suitable
attenuator must be used.
☞
The operation life with battery operation
• Insert country-specific adapter (2-16/10)
into the power unit (2-16/9).
• Insert angled connector of the power unit
(2-16/9) into the +12 V socket (2-16/2) of
the LED illuminator (2-16/1).
• Connect power unit to the line.
• Attach the cable to the back of the stand
with the self-sticking cable ties included.
strongly depends on the brightness set and the
loading capacity of the battery.
Operation with 9 V Alkaline battery
(9 V / Ni Cd battery block)
• Turn holding screw (2-16/8) on the cover
(2-16/7) of the battery compartment to the
left and remove the cover.
• Insert 9 V Alkaline battery (2-16/6) into the
battery compartment (2-16/5) with the
terminals in front (plus pole points
upwards).
• Attach cover, press it on and fix it in
position by turning the holding screw to the
right.
Operation with external voltage supply
• Insert angled connector of the 12 V battery-
adapter cable (2-16/11) into the +12 V
socket (2-16/2) of the LED illuminator.
• Connect cable clamps of the 12 V battery-
adapter cable (2-16/11) to the used 12 V
battery:
red clamp to plus pole (+),
black clamp to minus pole (-).
Depending on the brightness set via the control
(2-16/3), the integrated current control
guarantees constant brightness as long as there
is sufficient loading capacity from the battery. If
the loading capacity is exhausted, the LED
illuminator will go out or drop to minimum
brightness.
• After connection to the voltage source,
switch the LED illuminator on or off via the
toggle switch (2-16/4).
• Match the illumination intensity via the
1LED illuminator
2+12 V socket
3Control
4Toggle switch
5Battery compartment
69 V Alkaline battery
7Cover
8Holding screw
9Power unit
10Country-specific adapter
1112 V battery-adapter cable (4 m long)
12Cable holder
control (2-16/3).
Fig. 2-16 Connecting LED illuminator to power
source
Operation with battery pack C
• Use the included connection cables to
connect a 12 V connection box on the
battery pack C (cigarette lighter) with the
12V connection box on the LED illuminator.
Make absolutely sure to avoid moistening of
the 12 V battery-adapter cable with liquids.
For further details regarding the battery pack C,
please see the applicable operating manual.
Carl ZeissAttaching colored disk carrier, polarizer or wide-angle unitAxiostar plus
2.9 Attaching colored disk carrier,
polarizer or wide-angle unit
The polarizer must be installed if you want to
use polarization contrast. The wide-angle unit
must be used with objectives having a
magnification < 4x so that the entire visible
field is illuminated.
• Hold colored disk carrier (2-18/2) parallel to
the underside of the condenser carrier
(2-18/1) and screw the holding bolts of the
colored disk carrier (2-18/6) into the front
thread on the bottom left of the condenser
carrier (2-18/1, position a) using the 90°-
offset hexagonal key SW 2 (2-18/5) until it
engages.
• Screw stop bolt (2-18/3) with adjusting lever
(2-18/4) into the rear thread on the
condenser carrier (2-18/1, position b) until it
engages.
1Condenser carrier
2Colored disk carrier
3Stop bolt
4Adjusting lever
5Hexagonal key SW 2
6Holding bolt
7Slotted screwdriver
8Wide angle unit or swiveling polarizer
Fig. 2-18 Attaching colored disk carrier, polarizer or
wide angle unit
• Screw wide-angle unit or swiveling polarizer
(2-18/8) with exceeding threaded bolt
(2-18/8, position a) into the front threaded
hole of the condenser carrier (2-18/1,
position a) by turning the entire wide-angle
unit (or the polarizer). Screw in the unit
until the threaded hole of the holding angle
(2-18/8 position b) lies below the rear
threaded hole (2-18/1, psition b) of the
condenser carrier and contacts the
underside of the condenser.
• Then use the slotted screwdriver to screw
the holding angle to the underside of the
condenser carrier (2-18/1, position b) with
the supplied slotted screw.
Axiostar plus Inserting white-balance filterCarl Zeiss
2.10 Inserting white-balance filter
The white-balance filter can be used to balance
differing, subjective color perceptions when
observing a specimen and to increase contrast.
☞
The white-balance filter is only for subjective
balancing for the user. For microphotography,
remove the white-balance filter and use an
appropriate conversion filter (e.g. conversion
filter 3200 ... 5500 K).
• Carefully hold the white-balance filter
(2-19/2) on the edge without smudging it
and place it in the luminous field diaphragm
opening (2-19/4).
• To make sure the filter does not fall out,
push the filter clamp (2-19/1) in the groove
(2-19/3) until it engages. A maximum of
two filters can be secured with the filter
clamp.
• There are depressions (2-19/5) on the top of
the luminous field diaphragm (front and
back) to make it easier to remove the filter
clamp and white-balance filter. This allows
you to pry out the filter clamp or filter with
your fingers.
1Filter clamp
2White-balance filter
3Groove
4Luminous field diaphragm
5Depression in the luminous field diaphragm
Fig. 2-19 Inserting white-balance filter
☞
If a yellow tone appears when using objectives
5x and 2.5x even though a white-balance filter
is inserted, use an additional neutral filter 25 %
(467849-0000-000) on the luminous field
diaphragm and increase the lamp intensity.
2.11.1 Filter stop carrier for epi-fluorescence illuminator
The epi-fluorescence illuminator has two slots for filter stop carriers. The carriers can be inserted and
accessed from the right or left.
Filter stop carrier 1 (designated with one dot) has three positions (a, b, c) for filters with a 25 mm
diameter. O-rings are used to hold the filters in the stop carrier.
Filter stop carrier 2 (designated with two dots) also has three positions. Position 1 (adjust) is used to
adjust the fluorescent lamp (see section 2.11.4). Position 2 (open) is a free passage, but a 25 mm
diameter filter can be used. Position 3 (closed) is the locking position for the fluorescent lamp.
• When inserting the filter stop carrier, make sure that the markings on the top and right side of
the carrier match the markings and the epi-fluorescence illuminator.
Carl ZeissAttaching epi-fluorescence illuminatorAxiostar plus
• Hold the new burner on the labeled metal
base and place it in the heat sink so that
the reflective portion of the burner points
down when placed in the lamp frame. If
both sides are reflective, make sure that the
The heat sink must be parallel to the lamp base.
To aling, press spring handle and turn heat sink
with burner in the lamp frame.
electrode labeled “UP“ is at the top. If there
is a sideways reflective melting position on
the burner, it must point to the side in
relation to the beam path.
• Replace lamp housing and tighten clamping
screws (2-23/1). Make a note of the
number on the operational hours counter.
• Carefully tighten clamping screw on the
heat sink.
Change the burner after reaching the expected
life of 100 hours.
Avoid getting fingerprints on the glass
components of the burner. If fingerprints do
occur, remove immediately.
• Press down spring handle (2-23/4) and
place burner with heat sink into the lamp
frame (2-23/5). Make sure to touch only the
heat sink.
Since the Axiostar plus epi-fluorescence
illuminator has an adjustment guide (filter stop
carrier 2, 2-24/5), the HBO 50 can remain on
the illuminator while adjusting the burner.
• Turn on completely connected fluorescent
lamp via the power supply unit. The burner
will ignite automatically.
If the image of the light arc and the mirror
image are superimposed, the burner will have a
greater thermal load. This will reduce burner
life.
•Move filter stop carrier 2 (2-24/5) to the
adjust position. The light arc of the burner
(lighter) and its mirror image (somewhat
darker) can be seen the filter stop carrier 2
window (2-24/6).
• Use the knob (2-24/4) to focus the collector
so both light arcs are sharply focused.
• Use adjustment screw (2-24/1) to adjust
1Axial burner adjustment
2Height adjustment for burner
3Side adjustment for burner
4Knob for collector adjustment
5Filter stop carrier 2 with adjustment guide
6Image from light arc and mirror image of
burner in the adjustment guide
Fig. 2-24 Adjusting burner
burner in an axial direction to the mirror so
that both light arcs appear the same size in
the adjustment guide window (see Fig.
2-24/6).
• Use the adjustment screws for height
(2-24/2) and side adjustment (2-24/3) to
position the light arc and mirror image
centered and parallel to each other in the
adjustment circle (2-24/6). The light arc and
mirror image should not be superimposed.
• When you are finished making the
adjustments, use the knob to refocus again
and move the filter stop carrier 2 (2-24/5) to
the open or closed position.
index finger as shown in Fig. 2-26 and
insert at a slant from the top with the right
and left holding supports (2-26/5) into the
bottom spring clamps (2-26/2) in the
reflector carriage (2-26/3).
• Press on the top of the module until it
engages in the top spring clamps (2-26/1).
Removing a P&C reflector module:
• Slightly tip the module and disengage first
from the top spring clamps (2-26/1) and
then from the bottom spring clamps
(2-26/2) to remove completely.
Inserting the reflector carriage:
• Turn the reflector carriage (2-25/4) to the
correct insertion position and place in the
epi-fluorescence illuminator. Hold the lever
(2-25/5) to the right and push in the
reflector carriage until it engages.
• Turn lock (2-25/3) clockwise to lock
position.
• Push the reflector carriage lever to the left
or middle position an insert the protective
cap into the groove from above.
1Top spring clamps
2Bottom spring clamps
3Reflector carriage
4P&C reflector module
5Holding supports
Fig. 2-26 Removing/inserting P&C reflector module
☞
If the reflector carriage is not locked correctly,
the imaging quality will be affected negatively.
• The last step is to label the three reflector
carriage positions (2-25/6) on the epifluorescence illuminator with the included
labels (000000-1100-994) for filter
combination and pigment.
Carl ZeissAttaching epi-fluorescence illuminatorAxiostar plus
2.11.6 Changing filter set in P&C
reflector module
You can combine and attach the filter sets for
the P&C reflector module as needed. Filter sets
or completely equipped P&C reflector modules
can be ordered from Carl Zeiss.
• Remove P&C reflector module (2-27/3) from
the reflector revolver and place in a secure
position.
(2-27/2) and excitation filter (2-27/5) and
secure with the fastening (2-27/1).
Fig. 2-27 Changing filter set in P&C reflector module
The emission filter and excitation filter will be
labeled with an inscription or arrow on the
perimeter. The arrow indicates the installation
direction for the filter into the reflector module
and should point in (see arrow in Fig. 2-27).
To minimize image shift during multiple
fluorescence procedures, the emission filter can
have an additional marking showing the
position of the wedge angle.
The marking should be aligned to the
orientation groove (2-27/4) when inserting an
emission filter in the reflector module. This
guarantees that the wedge angle of the
emission filter has the same defined position in
the reflector modules used. This compensates
for or minimizes the already slight image shift
when Zeiss filter sets are used between
modules.
We recommend the following procedure if you
need to install filters without direction markings
(arrow):
Filters with reflective, dielectric layers should be
installed so that the reflective layer
– Points out for an excitation filter (in relation
to the reflector module)
and
– Points in for an emission filter (Fig. 2-28).
2.11.7 Changing color splitter in P&C
reflector module
Extreme care is required when attaching the
filter and color splitter to avoid damage to the
optical elements and to keep them clean.
We recommend ordering completely equipped
P&C reflector modules since changing the color
splitter has higher requirements.
To replace a color splitter:
• Remove P&C reflector module from
reflector revolver.
• Loosen both slotted screws
(2-29/1) with a screwdriver.
• Hold together both halves of the reflector
module, turn against the installation
position and set aside.
1Locking filter (emission filter)
2Reflective layer on emission filter
3Color splitter
4Reflective layer on the color splitter
5Excitation filter
6Reflective layer on the excitation filter
7Illumination/Imaging beam path
Carl ZeissAttaching epi-fluorescence illuminatorAxiostar plus
• Tip the top module half up (Excitation)
(2-30/1) and remove the bottom module
half (Emission) from the holding supports
(2-30/5b).
• Remove color splitter (2-30/2) and spring
frame (2-30/3) from bottom module half.
• Remove old color splitter and carefully place
the new one on the spring frame (2-30/3)
with the reflective side pointing up. Place
both parts together in the bottom module
half. Make sure that the side catch of the
spring frame engages in the correct portion
of the bottom module half.
☞
The reflective (layered) side of the color splitter
has an angled edge or corner (Fig. 2-31).
Axiostar plusConnecting the instrument to the lineCarl Zeiss
2.12 Connecting the instrument to
the line
The microscope can be operated using line
voltages of 100 - 240 V without conversion.
• Connect the line cable with connector
(2-32/2) to the instrument socket (2-32/1)
and connect the earth-contact plug (2-32/3)
to the line.
• Switch on the instrument via the on/off
switch (2-32/4) on the left-hand side of the
instrument.
• The green LED integrated into the on/off
switch lights up to indicate that the
instrument is ready for operation (switch in
"I" position). The integrated halogen lamp
6 V, 20 W must also be on.
Axiostar plusSwitching on the instrument Carl Zeiss
3 OPERATION
3.1 Switching on the instrument
• Switch on the instrument via the On/Off
switch (3-1/1).
• The green LED integrated in the on/off
switch lights up to indicate that the
instrument is ready for operation (switch in
"I" position). The integrated 6 V 20 W longlife halogen lamp must also be on.
☞
The Axiostar plus microscope is supplied with
factory-aligned illumination. The illumination
need not be adjusted even when the lamp is
exchanged by the customer.
• Set the required brightness via the
brightness control (3-1/2).
1 On/off switch with integrated control lamp
2 Brightness control
Fig. 3-1Switching on the instrument
• Depending on the application, place one or
several of the following dia. 32 filters
(3-2/2) on the dust-protection glass of the
luminous-field diaphragm (3-2/1) , e.g.
– Interference wide-band filter, green, 32x 4,
for contrast enhancement in b/w
photography of stained sections and for
phase contrast.
– CB 3 conversion filter, d = 32 x 2, to
generate the correct color temperature of
3200 K when artificial light color reversal
film is used.
– CB 3, d = 32 x 2, and CB 12, d = 32 x 2,
conversion filters for use with daylight color
film.
• If necessary, use the filter clamp (3-2/3) to
secure the filter.
When the LED illuminator is used instead of a
condenser, it must be switched on via the
toggle switch, and the illumination intensity
must be set via the control. Setting Köhler
illumination is not required on account of the
homogeneous field illumination up to diameter
20 mm. Only when objective 2.5x is used, it
may be necessary to center the LED illuminator
and to slightly lower the condenser carrier.
Carl ZeissTransmitted-light brightfieldAxiostar plus
3.2 Using the transmitted-light
brightfield technique
3.2.1 General principle
Transmitted-light brightfield microscopy is the
most usual of all the optical microscopy
techniques, since it allows the easy and fast
viewing of high-contrast or stained specimens
(e.g. blood smears).
In order to obtain optimum resolution with full
illumination of the field, the condenser, the
luminous-field diaphragm and the aperture
diaphragm must be set in accordance with the
rules of the KÖHLER illumination principle.
Here, the illumination cone is adapted to the
objective's opening cone. In this way, the
numerical aperture of the optical system is
used, and "superfluous" light, which can cause
interference in the form of scattered light, is
prevented.
3.2.2 Transmitted-light brightfield
configuration
– Each Axiostar plus microscope can be
configured to permit the transmitted-light
brightfield technique.
– When using the Abbe condenser with 5-
position turret, set the brightfield position
(H) via the turret disk.
– If a slider for phase contrast is available with
the 0.9/1.25 Abbe condenser, it must be
removed when using the transmitted-light
brightfield technique.
3.2.3 Setting transmitted-light
brightfield for KÖHLER
illumination
Requirement:
– As described in section 2, the microscope is
ready for operation and switched on
according to section 3.1.
1Brightness control
2Aperture diaphragm lever
3Condenser drive
Fig. 3-3Transmitted-light brightfield for KÖHLER
illumination, preparation
Settings:
• First, place a high-contrast specimen with
0.17 mm cover slip on top of the
mechanical stage 75x30 R/L.
• Set image brightness using the brightness
control (3-3/1) on the microscope stand.
• Move the Abbe condenser 0.9/1.25 to the
upper stop position via the condenser drive
(3-3/3) and move aperture diaphragm lever
(3-3/2) to the center position.
☞
If the microscope is equipped with a mechanical
stage 75 x 30 R (stage drive on the right side),
then the condenser drive will be attached on
the left in the factory.
The condenser drive will be located on the right
side of the microscope if the microscope is
equipped with a mechanical stage 75 x 30 L.
binocular tube (3-4/6) first and focus on the
object via the focusing drive (3-4/1).
• Then set the focus for the other eye by
turning the eyelens of the focusing eyepiece
if necessary.
• Close luminous-field diaphragm (3-4/2) until
it is visible in the field of view, even if not in
focus (3-4/A).
• Use the condenser (3-3/3) to adjust the
condenser until the luminous-field
diaphragm is sufficiently in focus (3-4/B)
• Center luminous-field diaphragm (3-4/C)
using both centering screws (3-4/4) and
then open it until the edge of the
diaphragm just disappears from the field of
(3-4/D).
• For aperture diaphragm setting (contrast),
remove one eyepiece from the tube and
look into the tube with your naked eye. Use
lever (3-4/3) to set the aperture diaphragm
to approx. 2/3 ... 4/5 of the diameter of the
objective exit pupil (3-4/E). In most
applications, this setting of the aperture
diaphragm provides optimum contrast at
almost full resolution and therefore the best
compromise for the human eye.
• Replace eyepiece into the tube.
1Focusing drive
2Luminous field diaphragm
3Lever for aperture diaphragm setting
4Centering screws for condenser
5Knurled ring - nosepiece
6Binocular tube
Fig. 3-4Setting transmitted-light brighfield for
KÖHLER illumination
☞
The size of the field of view and objective
aperture changes every time an objective is
switched; therefore, you must reset field of
view and aperture settings for optimal results.
For objectives < 4x, the swivel wide-angle unit
must be moved into the light path (see
section 2.9).
Carl ZeissTransmitted-light phase contrastAxiostar plus
3.3 Using transmitted-light phase
contrast
3.3.1 General principle
The phase contrast technique is ideal for
examinations of thin, unstained specimens, e.g.
culture cells.
The phase contrast technique uses the optical
modulators "phase stop and phase ring" and
the interference procedures during the
formation of the intermediate image to change
the small phase differences in intensity and
color differences which are visible to the human
eye.
☞
Optimum phase contrast requires a high level of
cleanliness! Therefore, clean the front lens of
the objective used, the visible condenser
surfaces, the upper cover slip surface and the
lower carrier plate surface of the specimen with
particular care and carefully remove grease.
1Eyelens – auxiliary microscope
2Knurled ring – auxiliary microscope
3Auxiliary microscope
4Binocular tube/phototube
5Hexagonal key SW 2
6Centering screws for annular diaphragm centering
7Luminous field diaphragm
8Slider for phase contrast
9Lever for aperture diaphragm
10Phase contrast objective
11Diopter
Fig. 3-5Setting transmitted-light phase contrast
3.3.2 Transmitted-light phase contrast
configuration
– Abbe condenser 0.9/1.25 and slider for
phase contrast, e.g. Ph 2.
– When using the Abbe condenser with 5-
position turret, set the relevant phase stop
position (Ph1, Ph2 or Ph3) via the turret
disk.
– Phase contrast objectives with phase rings
Ph 1, Ph 2 or Ph 3 for different average
numeric apertures which can also be used
in brightfield without any restriction.
– The name of the phase stop on the slider
for phase contrast must correspond to the
relevant name on the objective, e.g. Ph 1.
ready for operation and switched on
according to section 3.1.
– The microscope must be set for
transmitted-light brightfield, as described in
section 3.2.
Settings:
• Swivel phase contrast objective, e.g.
40×/0.65 Ph 2 (3-5/10), into the beam path.
• Open luminous-field diaphragm (3-5/7) and
aperture diaphragm (3-5/9) or (3-6/3) on
the Abbe condenser with turret.
If a phase stop position (or darkfield position) is
set on the turret condenser, a green dot
appears in the window of the aperture
diaphragm indicating that the stop must be
opened completely (green dot on the aperture
diaphragm and in the window (3-6/4) are in the
same row.
• On the condenser, insert the slider for
1Turret disk
2Window for turret disk position
3Aperture diaphragm lever
4Window for aperture diaphragm marking
5Adjusting screws for phase stops
Fig. 3-6Setting Abbe condenser with
5-position turret disk
phase contrast (3-5/8) with the same name
as on the phase contrast objective, e.g.
Ph 2, or set the relevant position on the
turret disk (3-6/2) of the Abbe condenser
with 5-position turret.
• Adjust the brightness.
• Check whether phase stop centering
complies with the figure (3-7). For this,
remove one eyepiece and replace it with the
diopter (3-5/11) or a auxiliary telescope
(3-5/3). Focus on the phase ring by pulling
out or pushing in the 3-5/1) of the auxiliary
telescope via the knurled ring (3-5/2).
Fig. 3-7Centering phase stop
• If required, center the phase stop (3-7/A)
via the two adjusting screws (3-5/6 or
(3-6/5) using a 90° hexagonal key SW 1.5.
• Then replace the diopter or the auxiliary
telescope with the eyepiece.
☞
Complete phase contrast is only achieved if the
bright phase stop (in the condenser) and the
dark phase ring (in the objective) are exactly
congruent in the illumination beam path
(3-7/B).
Carl ZeissTransmitted-light darkfieldAxiostar plus
3.4 Using the transmitted-light
darkfield technique
3.4.1 General principle
Darkfield is mainly used for small or minute
objects such as bacteria, but also for emulsions
or unstained objects in watery solutions.
In transmitted-light brightfield, unstained
biological specimens, such as bacteria or living
cell cultures, are often barely identifiable – if at
all – on account of their light transmission. The
situation changes significantly if such specimens
are viewed in transmitted-light darkfield. In
principle, the specimen is illuminated with an
illumination aperture which is larger than that
of the objective used.
Only the diffracted and scattered light
components which are so important for image
production reach the objective, while the
directly reflecting light bundles are guided past
the objective.
This is one of the reasons why even fine
structures can be resolved and appear bright on
a dark background although they partially lie
below the resolving power of the light
microscope.
3.4.2 Transmitted-light darkfield
configuration
– Condenser with slider for darkfield.
– When using the Abbe condenser with
5-position turret, set the darkfield position
(DF, 0.65 / 0.9) via the turret disk.
– ICS objectives with a numerical aperture
smaller than that of the used darkfield stop
can be used.
– For further details, please see the following
table:
Condenser with slider for contrasting techniquesSuitable objectives
Slider for darkfield 0.25/0.9 (DF 10)CP-Achromat 5x/0.12
DF setting on condenser revolver (0.65/0.9)CP-Achromat 10x/0.25
A-Plan 5x/0.12
A-Plan 10x/0.25
Achroplan 4x/0.10
Plan-Neofluar 2.5x/0.075
Slider for darkfield 0 .65/0.9 (DF 40)CP-Achromat 10x/0.25
DF setting on condenser revolver (0.65/0.9)CP-Achromat 40x/0.65
ready for operation and switched on
according to section 3.1.
Settings:
• Set KÖHLER illumination in the same way as
for transmitted-light brightfield, e.g. with
objective 10x.
• Insert the relevant slider (3-8/3) for darkfield
− see table on page 3-8 − into the opening
on the condenser until stop.
• Open the luminous-field diaphragm (3-8/2).
• Increase the brightness setting on the
potentiometer.
• If the image center is too bright or too dark,
sensitively correct the condenser height
until the field of view appears
homogeneously dark or features
homogeneous brightness distribution.
☞
Darkfield specimens require a considerably
higher level of cleanliness than specimens for
other methods; fingerprints, dirt and dust in
particular brighten the background and reduce
the contrast of the object image.
For darkfield setting, it is also possible to use
the slider for phase contrast 3 and the
following objectives instead of the slider for
darkfield 0.25/0.9: CP-Achromat, A-Plan and
Plan-Neofluar 2.5x – 10x and Achroplan 4x.
Darkfield condensers can be placed directly in
the Axiostar plus when you use the darkfield
condenser carrier. Chapter 2 offers an overview
of applicable types and recommended
objectives.
1Focusing drive
2Luminous field diaphragm
3Dark field slider
4Aperture diaphragm lever
5Centering screws for condenser
6Nosepiece
7Binocular tube
Carl ZeissTransmitted-light darkfieldAxiostar plus
Setting darkfield contrast with dry
darkfield condensers
• If necessary, swivel open the wide-angle
unit, colored disk carrier, polarizer or λ
plate.
• Move condenser carrier down until it makes
contact.
• Place dry darkfield condenser in darkfield
condenser carrier (see section 2).
• Place darkfield condenser carrier in
condenser carrier and center approximately
so that when the condenser carrier moves
the condenser moves into the gap of the
mechanical stage without making contact.
• Move condenser up until it makes contact.
• Place specimen.
• Set lamp brightness to maximum.
• Swivel in objective with a small
magnification (e.g. 5x or 10x) and focus the
specimen with the focusing drive.
• Reset focus level with Z drive. Focus the
luminous field diaphragm with the
condenser drive. Open the luminous field
diaphragm just above the viewing field
edge.
• Optimize the contrast with the condenser
drive, if necessary.
Additional information:
Darkfield specimens require a considerably
higher level of cleanliness than specimens for
other methods; fingerprints, dirt and dust in
particular brighten the background.
• Since you can only see the specimen where
small components illuminate when using
darkfield illumination, set the specimen so
that an even detail distribution can be seen.
This will make the imaging of the luminous
field diaphragm easier to identify.
• Close luminous field diaphragm by turning
luminous field diaphragm ring
counterclockwise until it makes contact.
• Lower the condenser until the edge of the
luminous field diaphragm is displayed
clearly (luminous field diaphragm focus
level). There will be an increasing or
decreasing light ring above or below the
focus level of the luminous field diaphragm
(circular “breathing” of the luminous field
diaphragm depiction).
Setting darkfield contrast with immersion
oil darkfield condensers
• If necessary, swivel open the wide-angle
unit, colored disk carrier, polarizer or λ
plate.
• Move condenser carrier down until it makes
contact.
• Place immersion oil darkfield condenser in
darkfield condenser carrier (see section 2).
• Place darkfield condenser carrier in
condenser carrier and center approximately
so that when the condenser carrier moves
the condenser moves into the gap of the
mechanical stage without making contact.
• Move condenser up until it makes contact.
• Place a drop of immersion oil (without
bubbles, if possible) on the center of the
condenser.
• Place specimen. The immersion oil will
disperse due to the capillary effect between
the top of the condenser and the bottom of
the specimen holder.
• Slightly move the mechanical stage back
and forth to dissipate any air bubbles in the
immersion oil.
• Set lamp brightness to maximum and open
luminous field diaphragm completely.
• Swivel in objective with a small
magnification (e.g. 10x) and focus the
specimen with the focusing drive. Then step
by step swivel in the next largest dry
objectives and focus the specimens with the
focusing drive.
• Close luminous field diaphragm by turning
luminous field diaphragm ring
counterclockwise until it makes contact.
• Lower the condenser until the edge of the
luminous field diaphragm is displayed
clearly.
• Recenter the luminous field diaphragm with
the adjustment screws on the condenser
carrier.
The luminous field diaphragm appears only as a
circle segment on the edge of the viewing field
due to the high magnification of the immersion
oil objective. As a result, focusing and centering
of the luminous field diaphragm must be
repeated. If necessary, the luminous field
objective should be opened slightly if the light
intensity is too small.
The luminous field diaphragm is centered when
the edge of the luminous field diaphragm is
centered or equidistant from the viewing field
edge.
• For a sharply focused specimen, open the
sharply set luminous field diaphragm until
just above the viewing field edge.
• You can improve the contrast of the
microscope image by slightly adjusting the
focus level of the condenser with the
condenser drive.
• The last criteria is an equally dark
background of the eyepiece image.
• For immersion oil objectives with an iris
diaphragm, the contrast can be further
optimized by turning the adjustment of the
iris diaphragm.
• Center the luminous field diaphragm on the
condenser carrier with the adjustment
screws and focus with the condenser drive.
• Place one drop of immersion oil on the
specimen location, swivel in immersion oil
objective and focus specimen.
Additional information:
Darkfield specimens require a considerably
higher level of cleanliness than specimens for
other methods; fingerprints, dirt and dust in
particular brighten the background.
Carl ZeissTransmitted-light polarizationAxiostar plus
3.5 Using transmitted-light
polarization contrast
3.5.1 General principle
The transmitted-light polarization technique is
used with specimens that change the
polarization state of light. Such specimens, e.g.
crystals, minerals or polymers, are termed as
birefringent. If birefringent substances are
viewed between crossed polarizers (polarizer ⊥
analyzer), they appear bright while their
surrounding remains dark.
Birefringent substances are recognized by the
fact that they display 4 bright and 4 dark
positions between crossed polarizers after
rotation of the specimen around 360°.
Depending on the level of birefringence,
thickness and orientation of the object,
interference colors will occur from gray (usually
in biological objects) to white, yellow, red and
blue. These interference colors can be of 1
st
or
higher order.
3.5.2 Transmitted-light polarization
contrast configuration
– Polarizer, oriented in EAST-WEST direction,
mounted on swivel-out carrier (3-9/1).
– Analyzer, oriented in NORTH-SOUTH
direction (3-9/2).
1Polarizer on swivel-out carrier
2Analyzer
Fig. 3-9Setting transmitted-light polarization
contrast
3.5.3 Setting transmitted-light
polarization contrast
Before starting:
– As described in section 2, the microscope is
ready for operation and switched on
according to section 3.1.
– The microscope must be set for
transmitted-light brightfield, as described in
section 3.2.
Settings:
• Screw NORTH-SOUTH-oriented analyzer
(3-9/2) into the tube (removed from the
stand) from below. NORTH-SOUTH
orientation is available when the marking
on the analyzer is aligned along the
longitudinal axis of the stand.
• Swivel polarizer (3-9/1) into the beam path;
the field of view appears dark because of
the crossed polarizers.
lambda plate is available (e.g. 4452260000-000), set the oscillation direction to
45° (γ, stop position).
• Select crystal needles that are oriented in
the gamma direction (see marking on the
lambda plate).
Analysis:
• If the crystal needles oriented parallel to the
gamma direction of the lambda plate are
yellow, and the crystal needles lying at a
right angle to the gamma direction are
blue, the crystals are monosodium urate
crystals (gout).
• If the crystal needles oriented parallel to the
gamma direction of the lambda plate are
blue, and the crystal needles lying at a right
angle to the gamma direction are yellow,
the crystals are calcium pyrophosphate
crystals (pseudogout).
This analysis is also possible using a polarizer
with cemented lambda plate which can be
placed on the luminous-field diaphragm. In that
case, the lambda plate needs not be rotated.
The epi-fluorescence method displays
fluorescent substances with strong contrast in
typical fluorescent colors. In an epi-fluorescence
microscope, this occurs through a light
generated via a heat protection filter on the
excitation filter (bandpass). The filtered shortwave excitation beam is reflected by a
dichromatic beam splitter and focused on the
specimen via the objective. The specimen
absorbs the short-wave beam and emits a longwave fluorescence beam (Stoke’s Law), which is
registered by the objective and let through by
the dichromatic beam splitter. Then the beams
pass through a blocking filter
(lowpass/bandpass) that only allows long-wave
beams emitted from the specimen to pass.
The excitation and blocking filters must be
calibrated spectrally; they are both in the
reflector module FL P&C, along with the
dichromatic beam splitter.
You can get an overview of filter sets and
fluorochromes available from Zeiss at:
www.zeiss.de/micro
“Fluorescent microscopy” under
“Techniques of Microscopy”
3.6.2 Epi-fluorescence configuration
• Recommended objectives Plan-Neofluar or
Fluar (UV excitation).
• Epi-fluorescence illumination with reflector
module FL P&C.
• Mercury vapor short-arc lamp HBO 50 for
incident illumination.
☞
The mercury vapor short-arc lamp must be
adjusted with the adjustment guide before
using it for the epi-fluorescence procedure. If
necessary, you must readjust the setting
depending on the amount of use.
3.6.3 Setting epi-fluorescence
Before starting:
– The microscope is ready for use as outlined
in section 2.
Settings:
• Turn on the halogen lamp with the line
switch on the microscope.
• Swivel the desired objective into place.
• Search for the specimen location to be
observed in the transmitted light. If the 5position Abbe condenser with turret is to be
used, set the turret to position H
transmitted-light brightfield (or phase
contrast).
The first epi-fluorescence setting is significantly
easier if you use the Plan-Neofluar 20x/0.50
objective and a strong fluorescent specimen.
Demonstration specimens can be used first.
• First leave the illumination beam closed;
move the filter stop carrier 2 (3-10/2) to the
closed position.
• Turn on the HBO 50 with the power supply
and let it warm up to operating
temperature for about 15 minutes.
• Use the adjustment handle (3-10/3) to
move the desired fluorescence filter
combination on the reflector carrier into the
beam path.
• Turn off the halogen lamp (on microscope)
and open illumination beam path. Move the
filter stop carrier 2 to the open position.
• Refocus on the specimen and, if necessary,
optimize the collector setting with the knob
1Knob for collector adjustment
2Filter stop carrier 2
3Adjustment handle
Fig. 3-10 Setting epi-fluorescence
(3-10/1) on the HBO 50. Set the collector so
that the illumination from the field of view
appears as equal as possible to the shortwave excitation from the reflector module.
Correction of the collector position is no
longer necessary for modules with longwave excitation.
The measurement of lengths using the Axiostar
plus requires the following, for example:
– stage micrometer, positive 5 + 100/100 y
D = 0.17 mm
– eyepiece crossline micrometer 10 : 100,
d = 26 mm
– An overview of available stage micrometers
and eyepiece reticles is provided in
chapter 1.6.
☞
The distance to be measured should be ≥ 5 mm
in the intermediate eyepiece image in order to
keep the influence of random measuring
deviations as low as possible.
Other measuring errors can occur if the
eyepiece has not been inserted into the tube
until stop.
Settings:
Before the length measurement using the
microscope can be performed, the micrometer
or scale value of the used objective / eyepiece
reticle combination must be determined. This
scale value is exactly that distance in the
specimen which complies to one interval of the
used eyepiece crossline micrometer.
For calibration, align the scales of the stage
micrometer and the crossline micrometer
parallel to each other by turning the eyepiece,
and make the zero lines of both scales exactly
congruent.
If, for example, 99 increments (of 10 µm each)
of the stage micrometer correspond to exactly
100 increments of the crossline micrometer, as
in Fig. (3-11), the resulting scale value k’ for the
used objective / eyepiece reticle combination
(A-Plan 10x/0.25 and crossline micrometer
10:100) is
99
k‘ = x× 10 µm = 9.9 µm
100
Fig. 3-11 Length measurement
After exchange of the stage micrometer for the
specimen to be measured, the measuring
distance of interest L results from the number
of increments of the eyepiece crossline
micrometer (tenth estimated) multiplied with
the scale value k’:
L = 35.5 × 9.9 µm = 351.5 µm.
Particularly large object structures can also be
determined by using the vernier scale
gradations (0.1 mm) on the mechanical stage.
Here, it might be necessary to determine the
distance to be measured through calculation
from a combined x and y measurement
(Pythagoras).
right in a slightly inclined position into the
spring of the stand mount.
• Then set the double tube
ICS by lowering
the left side into the dovetail of the stand
mound. This will secure the double tube
ICS
from falling out.
• Turn the
ICS double tube into the required
viewing direction and use the knurled screw
(3-12/4 or hexagonal screw) to clamp it into
the threaded opening (3-12/5). Both the
illustrated rotation and the 90° backward
rotation are possible. Accordingly, the
controls of the
ICS double tube are
accessible to the user from the side or from
the front.
• Remove dust caps from the eyepiece tubes
and insert two eyepieces each (3-12/2) with
the same field number 18 or 20.
• Insert country-specific adapter (3-12/9) into
the power unit (3-12/8).
• Insert angled connector (3-12/10) of the
power unit into the 12 V 3-12/3) of the ICS
double tube.
• Use the enclosed self-adhesive cable holders
(3-12/11) to suitably attach the cable to the
stand and guide it to the back.
• Connect power unit to the line.
The ICS double tube is operated via:
• Adjusting lever (3-12/7) for the positioning
of the light pointer in the microscope image
• Control (3-12/6) to set the brightness of the
light pointer.
1ICS double tube
2Eyepieces
312 V socket
4Knurled screw (or hexagonal screw)
5Threaded opening for knurled screw or hexagonal
screw
6 Control for intensity of light pointer
7Adjusting lever for light pointer
8Power unit
9Country-specific adapter
10Angled connector
11Cable holder
Carl ZeissAttaching photomicrography equipmentAxiostar plus
3.9 Attaching photomicrography
☞
For the use of the photomicrography equipment, please observe the relevant separate
manuals in addition to the information provided
in this manual.
The Axiostar plus microscope with phototube
can be changed from observation to photomicrography via the control knob (3-13/7 and
3-14/8) attached to both sides of the
equipment
phototube (photomicrography position: turned
counterclockwise until stop: 100 % doc). In the
other position of the knob, the beam path is
directed to the eyepieces (100 % vis). Special
adapters allow commercially available 35 mm
SLR cameras and special microscope cameras
(e.g. MC 80
DX) to be attached to the camera
port of the Axiostar plus.
If focusing is not to be made via the viewfinder
of the camera, the component with the
eyepiece reticle must be screwed in the
eyepieces (see sections 1.6 and 2.5.1).
• For focusing, set the eyepiece until the
reticle is imaged in focus.
1Cable release
2Camera housing
3T2-Adapter for CONTAX bayonet
42.5× connector for T2
5Phototube
6Hexagonal screw
7Control
8Dust protection cap
• Then focus until both the microscope image
and the reticle are in focus.
3.9.1 Attaching SLR camera,
e.g. CONTAX ARIA
• Screw T-2 adapter (3-13/3) for the CONTAX
bayonet on the 2.5x connector for T2
(3-13/4) (456005-0000-000).
• Attach camera housing (3-13/2) and cable
release (3-13/1), if required.
• Loosen three hexagonal screws (3-13/6)
remove the dust cover (3-13/8) from the
phototube (3-13/5) and insert the
premounted unit (3-13/A) in the phototube.
• Align the camera unit in the required
position and tighten the three hexagonal
screws (3-13/6).
• For photomicrography, turn control (3-13/7)
counterclockwise until stop. When using
the phototube 30°
ICS, pull out the lever for
light path switching completely (100 %
light to camera port).
When artificial light color reversal film is used, the CB 3 conversion filter provides the correct color
temperature of 3200 K. The filter must be placed on the dust cover of the luminous-field diaphragm
(3-2/2), as mentioned in chapter 3.1.
For daylight color reversal film, the CB 12 conversion filter must be used in addition to the CB 3
conversion filter.
Various T2 adapters for SLR cameras are listed below:
T2 adapters for SLR camera housingsCat. No.
T2 adapter for CONTAX (CONTAX bayonet)416010-0000-000
T2 adapter for OLYMPUS OM (OM bayonet)416002-0000-000
T2 adapter for MINOLTA (SR bayonet)416003-0000-000
T2 adapter for CANON (FD bayonet)416004-0000-000
T2 adapter for NIKON (F bayonet)416009-0000-000
T2 adapter for PENTAX (KA bayonet)416011-0000-000
☞
For detailed information on SLR cameras, please see the new operating manual B 40-046 e,
“Photomicrography using 35 mm SLR cameras”, from Carl Zeiss.
remove dust protection cap (3-14/9) from
the phototube.
• Insert adapter 60 for microscope camera
(3-14/5) (456006-0000-000) into phototube
(3-14/6) and fix it using three hexagonal
screws (3-14/7).
• Insert P 2.5× projection lens (3-14/4) into
adapter 60 for microscope cameras
(3-14/5).
135 mm Mot DX film cassette
2MC 80
3Clamping ring
4P 2.5× projection lens
5Adapter 60 for microscope cameras
6Phototube
7Hexagonal screws
8Control
9Dust protection cap
Fig. 3-14 Attaching MC 80
DX basic body
DX microscope
camera
• Attach MC 80
DX basic body (3-14/2) onto
adapter 60 for microscope cameras until
stop, align it and fix it by turning clamping
ring (3-14/3) anti-clockwise.
• Attach 35 mm Mot DX film cassette
(3-14/1) to the basic body in such a way
that the contact pins firmly engage in the
relevant sockets.
• For microphotography, turn control knob
(3-14/8) counterclockwise until stop. When
using the phototube 30°
ICS, pull out the
lever for light path switching completely
(100 % light to camera port).
When artificial light color reversal film is used,
the CB 3 conversion filter provides the correct
color temperature of 3200 K. The filter must be
placed on the dust cover of the luminous-field
(3-2/2), as mentioned in chapter 3.1.
For daylight color reversal film, the CB 12
conversion filter must also be used. The filter
must be placed on the dust cover of the
luminous-field diaphragm (3-2/2), as described
in chapter 3.1.
☞
For detailed information on the MC 80 DX,
please see manual B 40-036-d, “Microscope
Camera MC 80
Carl ZeissAttaching adapters for video camerasAxiostar plus
The connecting piece 60 - 44 also allows video
adapters with 44 mm interface to be used with
Video adapter
(Cat. No.)
456140
456140-0000-000
Connecting piece 60 - 44
Microscopes with interface 60 mm and
all video adapters for 44 mm interface.
Suitable forComments
☞
The instructions of the camera manufacturer
must also be observed when operating the
video camera.
Attachment and settings:
• Loosen three hexagonal screws and remove
dust cover from the camera tube of the
Axiostar plus
• Screw video adapter or video zoom adapter
with C-mount thread into the video
camera.
• Insert video adapter or video zoom adapter
in ENG 2/3" or ENG 1/2“ bayonet of the
video camera and clamp it tight.
• Insert premounted unit (video camera with
video adapter or video zoom adapter) in
camera tube of the Axiostar plus, align it
and fix it using the three hexagonal screws.
.
the phototube of the Axiostar plus with 60 mm
interface.
Connects video adapters for 44 mm
interface to microscopes with 60 mm
interface.
• Insert eyepiece with photo reticle in the
binocular tube and align photo reticle
parallel to the camera.
• Swivel out folding prism on the binocular
phototube 45° ICS to direct 100 % of the
light to the port. When using the
phototube 30° ICS, pull out the lever for
light path switching completely (100 %
light to camera port).
• Set the required zoom magnification factor
via the wheel of the video zoom adapter.
• If required, adjust image brightness on the
monitor by changing the lamp brightness
on the microscope stand.
Carl ZeissAttaching adapters for digital compact camerasAxiostar plus
3.11.1 Digital cameras with 37 mm
filter thread, e.g. SONY DCRPC100 / SONY DSC – S50
When delivered, the sliding mount with thread
M37 (3-15/2) and the lens mount (3-15/4) are
premounted. The drawing on the right shows
the disassembled components.
• Remove threaded adapter ring M37/M52
(3-16/3).
• Plug connector 60 (3-15/6) onto the
binocular phototube (3-15/7) and tighten
the 3 set screws (3-15/8).
• Screw the unit, consisting of sliding mount
with thread M37 (3-15/2) and lens mount
(3-15/4), into the M37 filter thread of the
camera (3-15/1).
• With the lens mount (3-15/4) pointing
forward, plug the unit onto connector 60
(3-15/6) until stop and tighten the set
screw (3-15/5).
1SONY DCR-PC100 camera
2Sliding mount with thread M37
3Set screw
4Lens mount
5Set screw
6Connector 60 for microscope camera
7Binocular phototube
8Set screw
Fig. 3-15 Attaching Sony "Digital Handycam DCR-
PC100" camera
Depending on the microscope configuration
and the camera used, the distance between
the camera lens and the lens mount (3-15/4)
must perhaps be optimized (see double arrow).
This is required in particular if an untrimmed
image cannot be achieved in any zoom
position of the camera lens.
Make the following settings on the camera:
• Switch off the autofocus.
• Set the distance to ∞.
• Set the automatic exposure control to time
priority.
• Set the aperture as wide as possible (i.e.
select a small f-stop number!).
Not all cameras feature these possibilities.
Please see the operating instructions of the
camera used.
• Loosen the set screw (3-15/3).
• Vary the distance between the camera lens
and the lens mount in steps, i.e. move the
sliding mount with camera on the lens
mount by defined steps.
Axiostar plusAttaching adapters for digital compact camerasCarl Zeiss
• Zoom camera lens from wide-angle
position (W) through to tele position (T).
• Perform test until image is format-filling
without trimming or vignetting.
• Tighten the set screw (3-15/3) again.
No untrimmed image may be obtained when a
camera adapter combination is used which is
not recommended by Carl Zeiss.
3.11.2 Digital cameras with 52 mm
filter thread,
e.g. SONY DSC–S70, DSC-S75,
DSC-S85
When delivered, the sliding (3-16/4), the
threaded adapter ring M37/M52 (3-16/3) and
the lens mount (3-16/6) are premounted. The
drawing on the right shows the disassembled
status.
Furthermore, the threaded ring adapter
M37/52 (3-16/3) can be unscrewed from the
sliding mount (3-16/4) (not shown), i.e.
cameras with M37 can also be attached as
described in the previous chapter.
• Screw SONY VAD-S70 adapter ring
(3-16/2) into the objective/filter thread of
the DSC-S70 camera (3-16/1) until stop.
• Plug connector 60 (3-16/8) onto the
binocular phototube (3-16/9) and tighten
the 3 set screws (3-16/10).
• Screw the unit, consisting of sliding mount
(3-16/4), the threaded adapter ring
M37/M52 (3-16/3) and the lens mount
(3-16/6) into the SONY VAD-S70 adapter
ring (3-16/2).
• With the lens mount (3-16/6) pointing
forward, plug the unit onto connector 60
(3-16/8) until stop and tighten the set
screw (3-16/7).
1SONY DSC-S70 camera
2SONY VAD-S70 adapter ring
3M37/M52 threaded adapter ring
4Sliding mount
5Set screw
6Lens mount
7Set screw
8Connector 60 for microscope camera
9Binocular phototube
10Set screw
Fig. 3-16 Attaching Sony "Digital Still Camera DSC-
Carl ZeissAttaching adapters for digital compact camerasAxiostar plus
Depending on the microscope configuration or
camera used, the distance between the
camera lens and the lens mount (3-16/6) must
perhaps be optimized (see double arrow). This
is required in particular if an untrimmed image
cannot be achieved in any zoom position of
the camera lens.
Make the following settings on the camera:
• Switch off the autofocus.
• Set the distance to ∞.
• Set the automatic exposure control to time
priority.
• Set the aperture as wide as possible (i.e.
select a small f-stop number!).
Not all cameras feature these possibilities.
Please see the operating instructions of the
camera used.
• Loosen the set screw (3-16/5).
3.11.3 Attaching cameras with other
filter thread sizes
In principle, it is also possible to connect
cameras with a filter thread other than M37 or
M52. Suitable filter adapters or reduction rings
are available in photo shops. As mentioned
above, only a test can clarify whether such
cameras are compatible with the digital
adapters.
3.11.4 Sony DSC-S70, DC-S75, DSC-S85
on digital camera adapter 44
M52x0.75, 000000-1108-984
This adapter has been particularly designed for
the Sony DSC-S70. Combined with “Plan”
objectives, e.g. “Plan-Neofluar” objectives,
increased edge definition and lower distortion
can be achieved. Should future cameras prove
to be suitable, too, our sales staff will be
informed accordingly.
• Vary the distance between the camera lens
and the lens mount in steps, i.e. move the
sliding mount with camera on the lens
mount by defined steps.
• Zoom camera lens from wide-angle
position (W) through to tele position (T).
• Perform test until image is format-filling
without trimming or vignetting.
• Tighten the set screw (3-16/5) again.
No untrimmed image may be obtained when a
camera adapter combination is used which is
not recommended by Carl Zeiss.
Attachment of the camera requires the SONY
VAD S 70 adapter ring (Fig. 3-16/2). Furthermore, the connecting piece 60-44 (4561400000-000) must be inserted between the
camera adapter and the binocular tube.
The portable case for the Axiostar plus can be
used to transport the basic Axiostar plus
equipment (stand with binocular tube 45° ICS,
Abbe condenser, mechanical stage, line cable,
dust cover).
• Before inserting the Axiostar plus into the
case, loosen the knurled screw on the stand
(3-17/9) by turning it by ¼, turn the
binocular tube 45°
ICS (3-17/6) around 180°
(eyepieces pointing backwards) and tighten
the knurled screw again.
• Open the four locks of the case and pull the
case cover (3-17/1) upwards to remove it
apart and slide the rubber foam component
(3-17/7) between the eyepiece tubes until
stop. It might be necessary to apply some
pressure on the rubber foam component
until the holding flaps (3-17/8) are flush
with the upper stand part.
• Line cable, dust cover and any small
accessories can be placed on the case floor
beside the stand and fixed in position using
the Velcro fasteners (3-17/3).
• Attach the case cover, close the locks with a
snap and, if required, lock them with the
enclosed key.
from the case floor (3-17/2).
• Insert microscope into the recess in the case
floor.
• Push rubber foam component (3-17/4) over
the condenser (3-17/5) and move the
condenser upwards until it contacts the
rubber foam component.
Carl ZeissCare and maintenance of the instrumentAxiostar plus
4 CARE, MAINTENANCE AND
TROUBLE-SHOOTING
4.1 Care and maintenance of the
instrument
Care of the Axiostar plus is limited to the
following operations:
• Cover the instrument with the dust cover
after every use.
• Do not set up the instrument in a damp
room, i.e. max. humidity < 85 %.
• Cover open tubes with dust protection
caps.
• Remove dust and dirt from visible optical
surfaces using a brush, airblower, Q-tip,
optics cleaning paper or a cotton cloth.
• Remove water-soluble dirt (coffee, Coke,
etc.) after breathing on it and wiping it off
with a dust-free cotton cloth or a
moistened cloth. A mild cleaning agent can
also be added to the water.
•Remove stubborn, oily or greasy dirt
(immersion oils, fingerprints) with a wad of
cotton or a dust-free cotton cloth dipped in
the optics cleaning mixture L.
This cleaning mixture is produced of
90 Vol% benzoline and 10 Vol%
isopropanol (IPA). The various components
are also known under the following
synonyms:
benzoline:medical alcohol,
petroleum ether
Isopropanol2-propanol,
dimethyl carbinol,
2-hydroxypropane
When using the Axiostar plus in humid climatic
zones, proceed as follows:
• Store the Axiostar plus in bright, dry and
well-ventilated rooms with a humidity of
less than 85 %; store particularly sensitive
components and accessories, such as
objectives and eyepieces, in a dry closet.
The risk of growth of fungus on optomechanical instruments always exists in the
following conditions:
− Relative humidity of more than 75% and
temperatures between +15 °C and +35 °C
for more than three days.
− Installation in dark rooms without air
ventilation.
− Dust deposits and fingerprints on optical
surfaces.
Clean the optical surface by moving in circles
starting in the middle. Slight pressure should be
exerted on the optics during cleaning.
Do not touch the lamp bulb with your bare
hands; if required, clean the bulb with clean
alcohol before switching it on for the first time
to prevent dirt from burning in.
The following procedure is required to
exchange the lamp:
1Halogen lamp
2Cover
Fig. 4-2 Changing the 6 V, 20 W halogen lamp
• Switch off lamp supply via on/off switch
and allow lamp to cool down for approx.
15 minutes.
☞
☞
☞☞
When changing the lamp, it is not necessary to
remove the microscope from the ergonomic
base plate (in the event the base plate is being
used).
• Disconnect the earth-contact plug from the
line and remove the flat plug from the
instrument connector.
• Place the disconnected instrument on its
back to make the underside of the pyramid
stand accessible.
• Fold down cover (4-2/2) and pull out
defective halogen lamp (4-2/1).
• Use the protective cover to hold the new
6 V, 20 W halogen lamp and insert both
lamp pins carefully into the receptacles.
☞
The Axiostar plus microscope is supplied
with factory-aligned illumination. In general,
the illumination need not be adjusted even
when the lamp is exchanged by the
customer.
• Fold up cover again, return the stand to the
upright position and reconnect the stand to
the line.
for color photography using daylight
color films and artificial light color
reversal films
for visual observation and b/w
photography with transmission factor
information
for white light adjustment and contrast
enhancement for visual observation
for photography without color distortion,
with transmission factor information
All repairs of mechanical, optical or electronic
components inside the instrument and of the
electrical components of the Axiostar plus may
only be performed by Carl Zeiss service staff or
specially authorized personnel.
To ensure the optimum setting and trouble-free
function of your microscope over a longer
period of time, we would recommend that you
enter into a service/maintenance contract with
Carl Zeiss.
In the case of subsequent orders or when
service is required, please get in touch with
your local Carl Zeiss agency.
Within Germany, you can reach Carl Zeiss
Microscopy Service as follows:
Telephone:++49 180 333 6 333
Fax:++49 7364-20 4939
E-mail:Med-Mikro-Service@Zeiss.de
Further information is also available in the
Internet at the following address:
ddiameter
Ddark field or cover slip thickness
DINDeutsches Institut für Normung (German standards association)
docdocumentation
DXcoding system for the storage of electronically legible information, e. g. film speed
EGEuropean Community
EMVelectromagnetic compatibility
ENEuropean standards
ENGElectronic N
ews Gathering
E-PLname of eyepiece type with aspheric lens and flat field of view
EWGEuropean Economic Community
FISHFluorescence In Situ Hybridization
foc.Focusing
Hbright field
HALhalogen lamp
HBOmercury vapor short-arc lamp
ICSInfinity C
olor corrected System
IECInternational Electrotechnical Commission
IPInternational Protection (protection type)
ISOInternational Standard Organization
MCMicroscope C
amera
Nneutral-density filter
N.A.numerical aperture
Phphase contrast
PLplan
Rright (control on the right of the mechanical stage)