ba76071e01 02/2012
ADA-WA 1
ADA-WA 4
Installation instructions
ADA-WA1ADA-WA1
ADA-WA4ADA-WA4
Sensor adapter for installation in the
retractable armatures WA700/2 and WA 700/10
ADA-WA 1; ADA-WA 4
2
ba76071e01 02/2012
Important information for using this adapter
together with a SensoLyt armature:
Observe the following note when you want to use a
SensoLyt measuring armature with the aid of the adapter:
Warning
For safety reasons, use only electrodes especially
specified for high pressure and temperature loads (for
example pH combination electrode SensoLyt SEA-HP,
order no. 109 118). For more information, refer to the
YSI catalog or the Internet. YSI assumes no liability for
damages caused by using different electrodes.
ADA-WA 1; ADA-WA 4 Installation recommendations
3
ba76071e01 02/2012
Installation recommendations
General installation recommendations
Basic types of
installation
The optimum installation location must be individually determined for the type of sensor and the application.
Influential factors
The table on the following pages contains recommendations
and the special characteristics of the individual sensors.
Flow
direction
Minimum distance: 80 mm
A
45 ° 45 °
BC
Good
Sufficient flow of the
sensor element
A, B
Burdening of the measuring medium with
long fibers - avoidance of them getting
caught on the sensor (rake effect)
C
Burdening of the sample with small stones
and abrasive particles - avoidance of any
damage to the sensor element
C
Avoidance of malfunctions of the optical
measurement (turbidity/total suspended solids) by captured air bubbles and particles
A, B
Avoidance of malfunctions of the optical
measurement (turbidity/total suspended solids) by wall surfaces
A, C
Installation recommendations ADA-WA 1; ADA-WA 4
4
ba76071e01 02/2012
Overview of the recommended installation locations
Note
Always follow the I
NSTALLATION chapter of the respective sensor operating manual.
Sensor type Installation
location
(see
diagram on
page 3)
Advantages (+) /
Disadvantages (-)
Conclusion / Comments
TriOxmatic 700 IQ,
TriOxmatic 701 IQ,
TriOxmatic 702 IQ
FDO 700 IQ
B + Good flow
+ Low susceptibility to contamina-
tion
+ Low risk of membrane damage
TetraCon 700 IQ B + Good flow
+ Low susceptibility to contamina-
tion
Check and adjust cell constants
after installation (see sensor operating manual)
SensoLyt 700 IQ C + Good flow
+ Low susceptibility to contamina-
tion
- Danger of small stones damaging
the electrode
The enclosed fixing ring instead
of the protective hood improves
the incident flow.
Attention
: It is no protection
against small stones!
The best incident flow is
achieved when the diaphragm
points against the flow direction.
TFK 700 IQ C + Good flow
+ Low susceptibility to contamina-
tion
The enclosed fixing ring instead
of the protective hood improves
the incident flow.