YSI 95 User Manual

Page 1
Y
SI MODEL 95
Y
Y
Y
SI MODEL 95
SI MODEL 95
SI MODEL 95
Handheld
Handheld
HandheldHandheld Dissolved Oxygen
Dissolved Oxygen
Dissolved OxygenDissolved Oxygen
nd
nd
nd
nd
Temperature System
Temperature System
Temperature SystemTemperature System
Operations Manual
Page 2
Page 3
CONTENTS
CONTENTS
CONTENTSCONTENTS
SECTION 1 INTRODUCTION.........................................................................................................................................1
SECTION 2 PREPARING THE METER ........................................................................................................................2
2.1 U
NPACKING.................................................................................................................................................................2
2.2 W
ARRANTY CARD ......................................................................................................................................................2
2.3 B
ATTERIES ..................................................................................................................................................................2
2.4 C
ALIBRATIO N /S TORAGE CHAMBER ............................................................................................................................3
2.5 H
AND STRAP...............................................................................................................................................................3
2.6 T
HE METER CASE .......................................................................................................................................................3
SECTION 3 PREPARING THE PROBE.........................................................................................................................4
3.1 C
HOOSING THE RIGHT MEMBRANE CAP.....................................................................................................................4
3.2 M
EMBRANE CAP INSTALL A TION.................................................................................................................................4
SECTION 4 OPERATION.................................................................................................................................................6
4.1 T
URNING THE INSTRUMENT ON .................................................................................................................................7
4.2 C
ALIBRATIO N..............................................................................................................................................................7
4.3 M
AKING MEASUREMENTS ..........................................................................................................................................8
4.4 S
TIRRING.....................................................................................................................................................................9
4.5 S
AVING DATA..............................................................................................................................................................9
4.6 R
ECALLING STORED DATA..........................................................................................................................................9
4.7 E
RASIN G STORED DATA............................................................................................................................................10
4.8 T
OGGLING BETWEEN %-AIR SATURATION AND MG/L...............................................................................................11
4.9 D
ISSOLVED OXYGEN FILTER.....................................................................................................................................11
4.10 D
ISPLAY BACKLIGHT ..............................................................................................................................................11
SECTION 5 PRINCIPLES OF OPERATI ON.............................................................................................................12
5.1 MEA C
5.2 DO R
5.3 F
5.4 F
LARK OXYGEN SENSOR ................................................................................................................................12
EADINGS FROM THE CAT HODE REDUCTI ON.....................................................................................................13
ORMATION OF AGCL AT THE ANODE ......................................................................................................................13
UNCTION OF THE EL ECTROLYTE .............................................................................................................................13
SECTION 6 MAINTENANCE OF THE MEA SENSOR...........................................................................................14
6.1 A
NODE SERVICE........................................................................................................................................................14
6.2 C
ATHODE SERVICE....................................................................................................................................................14
6.3 D
ISSOLVED OXYGEN PROBE PRECAUTI ONS.............................................................................................................15
6.4 P
ROBE STORAGE .......................................................................................................................................................15
SECTION 7 DISCUSSION OF MEASUREMENT ERRORS...................................................................................16
SECTION 8 TROUBLESHOOTI NG............................................................................................................................17
SECTION 9 WARRANTY AND REPAIR...................................................................................................................19
APPENDIX A - GENERAL SPECIFICAT I ONS..........................................................................................................24
APPENDIX B - REQUIRED NOTICE...........................................................................................................................26
APPENDIX C - ACCESSORIES AND REPLACEMENT PARTS............................................................................27
APPENDIX D - UNIT CONVERSION...........................................................................................................................28
APPENDIX E - OXYGEN SOLUBILI T Y T ABL E.......................................................................................................29
APPENDIX F - CALIBRATION VALUES TABLE.....................................................................................................31
i
Page 4
ii
Page 5
SECTION 1
SECTION 1 INTRODUCTION
SECTION 1SECTION 1
The YSI Model 95 Handheld Dissolved Oxygen and Temperature Sy stem is a rugged, micro­processor based, digital meter with an attached YSI m icroelectrode array (MEA) dissolv ed oxygen probe. The MEA sensor eliminates the need for stirring in most environmental applications allowing measurement of DO in the field without an external stirring dev ice. I t also reduces measurement errors caused by insufficient or inconsistent stirring .
The YSI Model 95 has the following features:
Microprocessor control
Minimal stirring dependence
Low maintenance MEA DO probe
Push-button calibration
Cap membranes for easy membrane replacement
Simultaneous display of temperature and DO in % air saturation or mg/L
Automatic salinity com pensation w ith m anual entering of salinity value
Automatic temperature com p ensation
Data storage for 50 sets of readings with on screen recall
Waterproof case (IP65)
The YSI Model 95 has a non-detachable, combination sensor available with cable lengths of 10, 25, 50 or 100 feet. The Model 95D has a detachable cable. The probe utilizes easy to install cap membranes and the probe body has been manufactured with stainless steel to add rugged durability and sinking weight. The silver anode of the MEA DO sensor requires no servicing for up to 10,000 hours (four years) of operation under normal operating conditions (no sulfite contam ination).
INTRODUCTION
INTRODUCTIONINTRODUCTION
The Model 95’s micro-processor allows the system to be easily calibrated with the press of a few buttons. Additionally, the micro-processor performs a self-diagnostic routine each time the instrument is turned on. The self-diagnostic routine provides useful information about the function of the instrument and probe. For a list of these diagnostic codes, see Section 8, Troubleshooting.
A probe calibration/storage chamber is built into the instrument case. A small moist sponge in the chamber provides a water-saturated air environment that is ideal for air calibration of the dissolved oxygen probe. This chamber also provides a convenient place to store the probe when the system is not in use. The Model 95 case is waterproof (rated to IP65) allowing operation in the rain without damage to the instrument.
Six AA-sized alkaline batteries power the Model 95. A new set of alkaline batteries will provide approximately 150 hours of continuous operation. When batteries need to be replaced, the LCD will display a “LO BAT” message.
The YSI Model 95 is designed for use in environmental, aquaculture, and industrial applications where accurate dissolved oxygen and tem perature measurements are desired with minim al stirring .
YSI Incorporated Model 95 1
Page 6
SECTION 2
SECTION 2 PREPARING THE METER
SECTION 2SECTION 2
2.1
2.1 UNPACKING
2.12.1
When you unpack your new YSI Model 95 Handheld Dissolv ed Oxy g en and Tem perature Sy stem for the first time, compare the packing list with the contents of the shipping box. If there is anything missing or damaged, call the dealer from whom you purchased the Model 95. If y ou do not k now which authorized dealers sold the system to you, call YSI Customer Service at 800-765- 4974 or 937-767-7241, and we'll be happy to help y ou.
2.2
2.2 WARRANTY CARD
2.22.2
Please complete the Warranty Card and return it to YSI. The warranty card allows the entry of your purchase of this instrument in our computer system. Once your purchase is recorded, you will receive prompt, efficient service if any part of your YSI Model 95 needs repair during the warranty period.
2.3
2.3 BATTERIES
2.32.3
UNPACKING
UNPACKINGUNPACKING
WARRANTY CARD
WARRANTY CARDWARRANTY CARD
BATTERIES
BATTERIESBATTERIES
PREPARING THE METER
PREPARING THE METERPREPARING THE METER
There are a few things you must do to prepare your YSI Model 95 for use. First, locate the six AA­sized alkaline batteries that were included. Use a screwdriver or a small coin to remove the thumbscrew on the bottom of the instrument (see figure below). This thumbscrew holds the battery­chamber cover in place. The battery-chamber cover is marked with the words "OPEN" and "CLOSE."
NOTE: On some models, the battery cover thumbscrew may be unscrewed by hand (a screwdriver may not be required).
Battery chamber cover
Thumb screw
Polarity marking
O-rings
Hand strap
YSI Incorporated Model 95 2
Page 7
Preparing the Meter Section 2
There is a small molded insert inside each of the two battery-chamber sleeves. These labels illustrate the correct way to install the batteries into each sleeve of the battery- chamber.
CAUTION: It is very important that the batteries be installed ONLY as illustrated. The instrument will not function and may be damaged if the batteries are installed incorrectly.
Turn the instrument on by pressing and releasing the ON/OFF button on the front of the instrument. The liquid crystal display (LCD) should come on. Allow a few seconds for the instrument to complete its diagnostic routine. If the instrument does not operate, consult Section 8, Troubleshooting.
You may also want to take the instrument into a dark room and with the instrument ON, hold down the LIGHT button. The instrument back light should illuminate the LCD so that the display can be easily read.
2.4
2.4 CALIBRATION/STORAGE CHAMBER
2.42.4
The Model 95 has a convenient calibration/storage chamber built into the instrument’s side. This chamber provides an ideal storage area for the probe during transport and extended non-use. If you look into the chamber, you should notice a small round sponge in the bottom. Carefully put about 10 drops of clean water into the sponge. Turn the instrument over and allow any excess water to drain out of the chamber. The wet sponge creates a 100% water saturated air environment for the probe that is ideal for dissolved oxygen calibration.
CALIBRATION/STORAGE CHAMBER
CALIBRATION/STORAGE CHAMBERCALIBRATION/STORAGE CHAMBER
2.5
2.5 HAND STRAP
2.52.5
The hand strap (see figure on previous page) is designed to allow comfortable operation of the Model 95 with minimum effort. If the hand strap is adjusted correctly, it is unlikely that the instrument will be easily dropped or bumped from your hand.
To adjust the hand strap on the back of the meter, unsnap the vinyl cover and pull the two Velcro strips apart. Place your hand between the meter and the strap and adjust the strap length so that your hand is snugly held in place. Press the two Velcro strips back together and snap the vinyl cover back into place.
2.6
2.6 METER CASE
2.62.6
The meter case is sealed at the factory and is not intended to be opened, except by authorized service technicians. Do not attempt to separate the two halves of the meter case as this may
damage the instrument, break the water-proof seal, and may void the manufacturer's warranty.
YSI Incorporated Model 95 3
HAND STRAP
HAND STRAPHAND STRAP
METER CASE
METER CASEMETER CASE
Figure 1
Page 8
SECTION 3
SECTION 3 PREPARING THE PROBE
SECTION 3SECTION 3
The YSI Model 95 dissolved oxygen probe is shipped wet with a shipping membrane installed. This protective membrane cap on the probe tip must be removed and replaced with a new membrane cap filled with MEA probe solution before using the probe. Follow the instructions below to install the new membrane cap.
3.1
3.1 CHOOSING THE CORRECT MEMBRANE CAP
3.13.1
Two different membrane caps are available for the Model 95. The YSI Model 9501 Membrane Cap Kit is supplied with the Model 95. This kit contains six 0.5 mil (.0005") membrane caps and a bottle of MEA probe solution (KCl).
NOTE: YSI 9501 Membrane Caps offer the fastest response to changes in DO and are recommended by YSI for most applications.
For conditions with low flow or stagnant water, a 1 mil (.001") mem brane is av ailable (YSI Model 9502 Membrane Cap Kit). This membrane requires less stirring than the 9501, but has a m uch slower response. Use this membrane when minimal stirring (<2"/sec) is available.
3.2
3.2 MEMBRANE CAP INSTALLATION
3.23.2
CHOOSING THE CORRECT MEMBRANE CAP
CHOOSING THE CORRECT MEMBRANE CAPCHOOSING THE CORRECT MEMBRANE CAP
MEMBRANE CAP INSTALLATION
MEMBRANE CAP INSTALLATIONMEMBRANE CAP INSTALLATION
PREPARING THE PROBE
PREPARING THE PROBEPREPARING THE PROBE
WARNING: Use only YSI MEA probe solution in the m em brane cap. Any other solution will
damage the MEA sensor.
To install a new membrane cap on your YSI Model 95 dissolv ed oxy g en probe:
1. Unscrew and remove the probe sensor guard (see Figure 2 ).
2. Unscrew and remove the old mem brane cap.
3. Thoroughly rinse the sensor tip with distilled water.
4. Hold the membrane cap and add 8 to 9 drops of MEA probe solution (about half full).
5. Tap the bottom of the cap with your finger a few times to remove any trapped air bubbles.
CAUTION: Do not touch the membrane surface.
6. Screw the membrane cap onto the probe tightly by hand (to prevent leakage of electrolyte). A
small amount of probe solution should overflow.
7. Shake off any excess probe solution and rinse the stainless steel thoroughly with distilled water
to prevent corrosion.
YSI Incorporated Model 95 4
Page 9
Preparing the Probe Section 3
Fill new membrane cap with 8-9 drops of MEA probe solution.
Unscrew guard
Unscrew cap
Tap cap with finger to remove bubbles.
Screw cap on tightly by hand
Screw guard on tightly by hand
Figure 2
WARNING: Use only YSI MEA probe solution in the membrane cap. Any other solution will damage the MEA sensor.
YSI Incorporated Model 95 5
Page 10
SECTION 4
SECTION 4 OPERATION
SECTION 4SECTION 4
The following diagram is an overview of the operation of the Model 95. See the following sections for details of operation.
OPERATION
OPERATIONOPERATION
Figure 3
YSI Incorporated Model 95 6
Page 11
Operation Section 4
4.1
4.1 TURNING THE
4.14.1
With the batteries installed correctly, press the ON/OFF button. The instrument will activate all segments of the display for a few seconds, which will be followed by a self test procedure which will last for several more seconds. During this power on self test sequence, the instrument’s microprocessor is verifying that the system is working properly. If the instrument were to detect a problem, a continuous error message would be displayed. See the section entitled Troubleshooting for a list of error messages.
NOTE: It is normal for an error to be displayed for a second or two when the system is first
4.2
4.2 CALIBRATION
4.24.2
Dissolved oxygen calibration must be done in an environment with a known oxygen content. Since the amount of oxygen in the atm osphere is k nown, it makes an excellent environment for calibration (at 100% relative humidity). The calibration/storage chamber contains a moist sponge to create a 100% water saturated air environment.
TURNING THE INSTRUMENT ON
TURNING THE TURNING THE
turned on.
CALIBRATION
CALIBRATIONCALIBRATION
INSTRUMENT ON
INSTRUMENT ONINSTRUMENT ON
Before calibrating the YSI Model 95, complete the procedures discussed in the Preparing the Meter and Preparing the Probe sections of this manual.
To accurately calibrate the YSI Model 95 you will need to k now the follow ing inform ation:
The approximate altitude of the region in which you are located.
The approximate salinity of the samples that you will be measuring. Fresh water has a salinity
of approximately zero. Sea water has a salinity of approximately 35 (parts per thousand, ppt). If you are not certain what the salinity of the sample water is, use a YSI Model 30 Salinity­Conductivity-Temperature system to determine it.
1. Ensure that the sponge inside the instrument's
calibration chamber is wet. Insert the probe into the calibration chamber.
2. Turn the instrument on by pressing the ON/OFF
button. Wait for the dissolved oxygen and temperature readings to stabilize (usually 15 minutes is required after turning the instrument on). If the instrument was already on, press the MODE button until dissolved oxygen is displayed in mg/L or % air saturation. To enter the calibration menu, use two fingers to press and release both the UP ARROW and DOWN ARROW buttons at the same time (DOWN AR R OW slightly ahead).
Figure 4
YSI Incorporated Model 95 7
Page 12
Operation Section 4
3. The LCD will prompt you to enter the local altitude in hundreds of feet. Use the arrow keys to
increase or decrease the altitude. When the proper altitude appears on the LCD, press the ENTER button once.
EXAMPLE: Entering the number 12 here indicates 1200 feet.
4. The LCD will prompt you to enter the salinity of the sample(s) that you will be measuring. You
can enter any number from 0 to 80 (ppt). Use the arrow keys to increase or decrease the salinity setting. When the proper salinity appears on the LCD (zero for fresh water), press the ENTER button.
5. The Model 95 should now display CAL in the lower left of the display, the calibration value
should be displayed in the lower right of the display and the current DO reading (before calibration) should be on the main display. Make sure that the DO reading (large display) is stable, then press the ENTER button. The display should read SAVE then should return to the Normal Operation Mode.
For best results:
Each time the Model 95 is turned off, re-calibrate before taking m easurements.
Calibrate at a temperature within ±10°C of the sam ple temperature.
4.3
4.3 MAKING MEASURMENTS
4.34.3
MAKING MEASURMENTS
MAKING MEASURMENTSMAKING MEASURMENTS
The Model 95 has four modes: Ø Dissolved Oxygen % -- A measurement of oxygen in percent of air saturation (partial
pressure).
Ø Dissolved Oxygen mg/L -- A measurement of oxygen solubility in mg/L. Ø Recall -- Allows previously stored data to be displayed. Ø Erase all -- Allows ALL previously stored data to be deleted.
Temperature is displayed in both dissolved oxyg en m odes.
NOTE: When you turn the Model 95 off, it will “remember” which DO mode you used last and will return to that mode the next time the instrum ent is turned on.
To change between the Model 95 modes, simply press and release the MODE button. The Model 95 will cycle through the modes as follows:
Dissolved Oxygen
in % with°C
Dissolved Oxygen
in mg/L with °C
Recall Erase all
YSI Incorporated Model 95 8
Page 13
Operation Section 4
4.4
4.4 STIRRING
4.44.4
It is important to realize that even a small amount of stirring will improve the DO and temperature response times in stagnant water, because the transfer process of heat and oxygen will be facilitated by convection. Also, the MEA dissolved oxygen probe is not totally stirring independent due to the consumption of oxygen at the sensor tip during measurement. When taking dissolved oxygen measurements in totally stagnant samples, the probe must be moved through the sample at a rate of 2 inches per second to provide adequate stirring.
4.5
4.5 SSSSAVING DATA
4.54.5
The Model 95 is equipped with a non-volatile memory that is capable of storing up to 50 different sets of readings. Non-volatile means that you do not need to worry that your data will be lost due to a power failure or interruption, such as when the batteries are removed. Each set consists of dissolved oxygen in percent, dissolved oxygen in mg/L and temperature. The Model 95 will also assign a site identity number to each set of readings to allow easy review of the data. This feature is useful in situations where transcribing data is difficult or not available.
While dissolved oxygen is displayed on the screen (in % or mg/L), depress the ENTER button and hold it for approximately 2 seconds. The meter will flash SAVE on the display along with the current site identity (1 through 50) being used.
STIRRING
STIRRINGSTIRRING
AVING DATA
AVING DATAAVING DATA
When all 50 sites are full, the display will flash FULL on the screen. This message will remain on the screen (even after power down) until a button is pushed.
Once you have acknowledged the memory is full, any subsequent saved data will begin overwriting existing data starting with site #1. No additional warning will be displayed.
4.6
4.6 RECALLING STORED DATA
4.64.6
1. To put the Model 95 into the RECALL mode,
2. Depress the ENTER button to review the last set
RECALLING STORED DATA
RECALLING STORED DATARECALLING STORED DATA
depress the MODE button repeatedly until “rcl” is displayed on the screen along with the site ID number in the lower right corner.
of data that was saved. The Model 95 will display the dissolved oxygen in % air saturation and temperature. Another press of the ENTER button will display the dissolved oxygen in mg/L and the temperature.
SAVE
01
rcl
01
3. Depress the UP ARROW button to move up
YSI Incorporated Model 95 9
Page 14
Operation Section 4
through the saved sets of data.
4. Depress the DOWN ARROW button to move down through the saved sets of data.
5. When you have finished recalling data, press MODE two tim es to return to normal operation. NOTE: The Model 95 will recall data as a list. When the UP ARROW is depressed the Model 95
will display the Site ID# for the previously recorded date. For example: If you are reviewing Site ID# 5 and the UP ARROW is depressed the Model 95 will display Site ID#4. If you are reviewing Site ID# 5 and Site ID# 5 was the last set of data stored the DOWN ARROW button will display Site ID# 1.
Here is an example of the Model 95 memory . Site ID #1 Site ID #2 Site ID #3 If the UP ARROW button was pressed the Model 95 would display Site ID #2 Site ID #4 Site ID #5
4.7
4.7 ERASING STORED DATA
4.74.7
1. To erase the data that is stored in the Model 95’s
2. Depress and hold the DOWN ARROW and
3. When the Model 95 has successfully erased data,
ERASING STORED DATA
ERASING STORED DATAERASING STORED DATA
memory, depress the MOD E button until the Model 95 displays ErAS on the screen.
ENTER buttons simultaneously for approximately 5 seconds.
the display reads DONE for 1 to 2 seconds. The instrument will automatically change to normal operation after completion and the next saved data will be stored in site ID# 1.
ErAS
IMPORTANT: Using the erase function forever and completely erases data in all 50 site
ID’s. Do not use the erase function until all recorded data has been transcribed to an archive outside the Model 95.
YSI Incorporated Model 95 10
Page 15
Operation Section 4
4.8
4.8 TOGGLING BETWEEN %-AIR SATURATION AND MG/L
4.84.8
The UP ARROW key allows quick and convenient switching between the two DO parameters without going through the instrument’s four modes (using the MODE key). Press the UP ARROW key to toggle the DO reading betw een %- air saturation and mg/L.
4.9
4.9 DISSOLVED OXYGEN FILTER
4.94.9
The Model 95 is equipped with a DO filter to help filter out instability and high frequency noise. This feature is useful when measuring dissolved oxygen in an unstable environment such as a fast moving stream or an aeration tank . The default option for the filter is off. To activ ate the filter, enter the filter option menu by pressing both the DOWN ARROW and MODE keys togethe r (Ú key slightly ahead). Operation procedures:
Press both the DOWN ARROW and MODE keys
Press the UP ARROW or DOWN ARROW ke y to
TOGGLING BETWEEN %-AIR SATURATION AND MG/L
TOGGLING BETWEEN %-AIR SATURATION AND MG/LTOGGLING BETWEEN %-AIR SATURATION AND MG/L
DISSOLVED OXYGEN FILTER
DISSOLVED OXYGEN FILTERDISSOLVED OXYGEN FILTER
together (ÚÚÚÚ key slightly ahead). The current status of the filter is displayed, On or OFF, with a smaller “FIL” displayed in the bottom right corner of the screen.
On
change the current status of the filter option.
Press ENTER to confirm the change.
To abort any changes and exit the filter menu, press
the MODE key (instead of ENTER).
4.10
4.10 DISPLAY BACKLIGHT
4.104.10
At times it may be necessary to tak e measurements with the Model 95 in dark or poorly lit areas. To help in this situation, the Model 95 comes equipped with a backlight that will illuminate the display so that it can be easily read. To activate the backlight, press and hold the LIGHT button. The display will remain lit as long as the button is depressed. When y ou let it up, the lig ht g oes out to preserve battery life.
DISPLAY BACKLIGHT
DISPLAY BACKLIGHTDISPLAY BACKLIGHT
FIL
YSI Incorporated Model 95 11
Page 16
SECTION 5
SECTION 5 PRINCIPLES OF O
SECTION 5SECTION 5
5.1
5.1 MEA CLARK OXYGEN SENSOR
5.15.1
The MEA (microelectrrode array) is a steady-state Clark type polarographic (voltammetric) dissolved oxygen sensor. The sensor is made of a silv er anode and a gold cathode (consisting of 100 very small electrodes, each measuring approximately 8 micrometers in diameter) and is separated from the measured medium by a semi-permeable Teflon membrane. The small dimensions of each individual micro surface consume a very small amount of oxygen. Large spacing between adjacent microsurfaces allows for minimal overlap of diffusion layers from adjacent cathode surfaces. This design produces the minimal stirring dependence of the MEA probe. The temperature sensing
MEA CLARK OXYGEN SENSOR
MEA CLARK OXYGEN SENSORMEA CLARK OXYGEN SENSOR
PRINCIPLES OF OPERATION
PRINCIPLES OF OPRINCIPLES OF O
PERATION
PERATIONPERATION
Temperature sensor
Anode (silver)
MEA Cathode (gold)
Figure 5
element (thermistor assembly) is mounted next to the oxygen sensor vertically (see Figure 1), providing temperature reading s for the DO sy stem.
The membrane selectively allows oxygen to permeate into the sensor, but prevents most interfering molecules and fouling materials from entering. Upon permeating through the membrane, oxygen is reduced at the gold cathode. The current resulting from this reduction is diffusion-limited and is proportional to the partial pressure of oxygen in the sample. The counter reaction is the oxidation of silver at the anode/reference electrode that completes the overall electrolytic reaction in the chloride medium (KCl electrolyte) behind the membrane. These reactions, at the cathode and the anode, are as follows:
Cathode reaction: O Anode reaction: Ag + Cl- ==> AgCl + e
+ 2H2O + 4e- ==> 4OH
2
-
-
YSI Incorporated Model 95 12
Page 17
Principals of Operation Section 5
5.2
5.2 DO READINGS FROM THE CATHODE REDUCTION
5.25.2
The oxygen reduction current is sampled and processed, by the meter, and displayed as either %-air saturation or mg/L. While the parameter of %-air (partial pressure) is independent of temperature and salinity, mg/L (solubility of oxygen) is a function of temperature and salinity. The same %-air reading (same partial pressure) would give a higher mg/L reading at a lower temperature than at a higher temperature. The higher the salinity, the lower the solubility (mg/L) is for the same %-air reading at the same temperature.
5.3
5.3 FORMATION OF AgCl AT THE ANODE
5.35.3
While the oxygen reduction current passes through the internal circuit to be reported as the DO reading, it also passes through the anode oxidizing the silver and forming a thin layer of silver chloride. Furthermore, the oxidation of silver at the chloride medium provides a stable potential that the cathode potential is referenced to (for instance, the polarization potential of the cathode is -1.0 V versus the potential of the Ag/AgCl redox couple at the silver anode). Since the current of the MEA sensor is so small, there should not be any significant accumulation of AgCl at the anode for 3 to 4 years.
DO READINGS FROM THE CATHODE REDUCTION
DO READINGS FROM THE CATHODE REDUCTIONDO READINGS FROM THE CATHODE REDUCTION
FORMATION OF AgCl AT THE ANODE
FORMATION OF AgCl AT THE ANODEFORMATION OF AgCl AT THE ANODE
5.4
5.4 FUNCTION OF THE EL
5.45.4
There are two main functions for the electrolyte:
Under normal operating conditions, such as measuring oxygen around 100%-air saturation (8.27 mg/L) at 25°C, the electrolyte should last up to 500 hours. This translates into about 62.5 working days at 8 hours per day operation. The actual electrolyte life, however, may be shorter since, in most environmental applications, membrane fouling determines the life of the electrolyte/membrane.
FUNCTION OF THE ELECTROLYTE
FUNCTION OF THE ELFUNCTION OF THE EL
1. Supply the chloride (Cl
oxygen reduction at the cathode.
2. Provide the ionic conduction of electricity inside the cell, especially in the thin layer
between the gold cathode and the membrane.
ECTROLYTE
ECTROLYTEECTROLYTE
-
) to the anode/reference electrode for the counter reaction of the
YSI Incorporated Model 95 13
Page 18
SECTION 6
SECTION 6 MAINTENANCE OF THE MEA SENSOR
SECTION 6SECTION 6
6.1
6.1 ANODE SERVICE
6.16.1
Warning: Under no circumstances should ammonium hydroxide be used to clean the silver anode. Ammonium hydroxide will permanently damage the condition of the MEA surface.
The MEA oxygen sensor is, in principle, the same as the conventional Clark oxygen sensor in that the sensor is made of a silver anode and a gold cathode, but the cathode is a microelectrode array. Since the current of the MEA oxygen sensor is so much sm aller (on av erag e 100 tim es sm aller) than the current of YSI conventional oxygen sensors, consumption of the silver anode, due to the formation of AgCl, is minimal during the lifetime of the probe. There should not be any significant build-up of silver chloride at the surface of the anode for 3 to 4 years, therefore, the anode should not require chemical cleaning. Howev er, if the surface of the silv er anode has become fouled, gently wet sand it using 400 grit wet/dry sandpaper, rinse thoroughly with deionized or distilled water and wipe with a wet paper towel until the dark layer is removed. The directions are as follows:
Anode Cleaning Procedures (See figure 6)
ü Rinse the sensor thoroughly after removing the membrane cap. ü Use wet 400 grit sandpaper to sand away the top layer of the anode by wrapping the sandpaper
ANODE SERVICE
ANODE SERVICEANODE SERVICE
around the anode and gently rotating it until the dark lay er is removed.
MAINTENANCE OF THE MEA SENSOR
MAINTENANCE OF THE MEA SENSORMAINTENANCE OF THE MEA SENSOR
ü Rinse the anode thoroughly with deionized or distilled water and wipe with a w et paper towel. ü Rinse the anode again with deionized or distilled water.
Anode
Wet
microcloth
Figure 6
6.2
6.2 CATHODE SERVICE
6.26.2
Warning: Under no circumstances should the gold cathode surface (the MEA surface) be sanded. Sanding will permanently damage the condition of the MEA surface.
If the MEA oxygen sensor exhibits erratic behavior, such as a current rise at a rate of 1%/hour or very jumpy readings, it can be serviced by buffing. You can use a few light twists against the surface with the wet microcloth mounted on the buffing tool provided in the 9503 reconditioning kit.
CATHODE SERVICE
CATHODE SERVICECATHODE SERVICE
Cathode
Buffing Tool
Press lightly against sensor surface
YSI Incorporated Model 95 14
Page 19
Maintenance of the MEA Sensor Section 6
Note: The MEA sensor does not require buffing (cleaning) every time the membrane cap is changed. Under normal operating conditions, the MEA sensor should be buffed no more than two times per year.
Cathode Cleaning Procedures
ü Remove the mem brane cap and rinse the sensor thoroug hly with deionized or distilled water. ü Place the microcloth on the buffing tool (self-adhesive). Wet the microcloth thoroughly with
deionized or distilled water.
ü Twist the buffing tool back and forth three times in opposite directions while lightly pressing
the buffing tool against the sensor surface (see figure 6).
ü Rinse the sensor surface well with deionized or distilled water after buffing.
6.3
6.3 DISSOLVED OXYGEN PROBE PRECAUTIONS
6.36.3
Membrane life depends on usage. If the probe is properly maintained, one membrane cap should last two to four weeks depending on how often the probe is used and the type of samples measured. It is recommended that m embrane caps not be re-used.
1. To keep the electrolyte from drying out, store the probe in a moist environment, such as the
DISSOLVED OXYGEN PROBE PRECAUTIONS
DISSOLVED OXYGEN PROBE PRECAUTIONSDISSOLVED OXYGEN PROBE PRECAUTIONS
calibration chamber with the wet sponge inside.
2. Erratic readings are a result of loose, wrinkled, damaged, or fouled membranes, or from large
(more than 1/4 of the circumference of the probe) bubbles in the electrolyte reservoir. If erratic readings or evidence of membrane damage occurs, you should replace the membrane cap and the KCl solution. The average replacement interval is tw o to four weeks.
3. If the membrane is coated with oxygen consuming (e.g. bacteria) or oxygen evolving organisms
(e.g. algae), erroneous readings may occur.
4. Chlorine, sulfur dioxide, nitric oxide, and nitrous oxide can affect readings by behaving like
oxygen at the probe. If you suspect erroneous readings, it may be necessary to determine if these gases are the cause.
5. Avoid any environment that contains substances that may attack the probe materials. Some of
these substances are concentrated acids, caustics, and strong solvents. The probe materials that come in contact with the sample include FEP Teflon, stainless steel, epoxy, polyetherimide and the polyurethane cable covering.
6. Do not allow the probe to strike hard objects. The membrane or sensor inside may be damaged.
6.4
6.4 PROBE STORAGE
6.46.4
For long term storage (4 weeks), remove the membrane cap, thoroughly rinse the MEA sensor with deionized or distilled water and install a new membrane cap filled with MEA probe solution. Store the sensor in a humid environment such as the calibration chamber with the wet sponge inside. Do NOT store the probe dry.
PROBE STORAGE
PROBE STORAGEPROBE STORAGE
YSI Incorporated Model 95 15
Page 20
SECTION 7 DISCUSSION OF MEASUREMENT ERRORS
SECTION 7 DISCUSSION OF MEASUREMENT ERRORS
SECTION 7 DISCUSSION OF MEASUREMENT ERRORSSECTION 7 DISCUSSION OF MEASUREMENT ERRORS
There are three basic types of dissolved oxygen errors. Type 1 errors are related to limitations of instrument design and tolerances of instrument components. These are primarily the meter linearity and the resistor tolerances. Type 2 errors are due to basic probe accuracy tolerances, mainly background signal, probe linearity, and variations in membrane temperature coefficient. Type 3 errors are related to the operator's ability to determine the conditions at the time of calibration. If calibration is performed against more accurately known conditions, type 3 errors are appropriately reduced.
Type 1 Errors
A. Meter linearity error: ±0.5% of full scale reading, or ±0.04 mg/l at 25° C whichev er is g reater. B. Component and circuitry error: ±0.04 mg/l
Type 2 Errors
A. DO errors caused by temperature compensation for measurements at ±10°C from calibration
temperature: ±1% of 25° C (±0.08 mg/l)
B. DO errors caused by temperature measurement errors: A maximum ±0.2°C temperature error is
equal to ±0.5% (0.04mg/L at 25°C).
Type 3 Errors
A. Altitude: The maximum DO error caused by calibrating to altitude in increments of 100 feet:
±0.18% (< 0.015 mg/l at 25°C)
B. Humidity: Errors occur if calibration is performed at less than 100% humidity. The worst
possible case would be calibration at 0% humidity. The error varies with the calibration temperature as follows:
Temperature Calibration Error at 0% humidity
0oC 0.09 mg/l 10oC 0.14 mg/l 20oC 0.21 mg/l 30oC 0.33 mg/l 40oC 0.50 mg/l
Approximating The Error
It is unlikely that the actual error in any measurement will be the maximum possible error. A better error approximation is obtained using a root mean squared (r.m .s.) calculation:
2
r.m.s. error = ±[1a
NOTE: This calculation is for a near extreme set of conditions.
If the probe is calibrated in water-saturated air, then type 3B errors (humidity), the largest error of all types, is virtually eliminated and the maximum possible error is in the order of 0.1 mg/L for the case of calibrating around 25°C.
YSI Incorporated Model 95 16
+ 1b2 + 2a2 + 2b2 + 3a2 + 3b2]½ mg/l
Page 21
SECTION 8
SECTION 8 TROUBLESHOOTING
SECTION 8SECTION 8
TROUBLESHOOTING
TROUBLESHOOTINGTROUBLESHOOTING
Symptom Possible Cause Action
1. Instrument will not turn on A. Low battery voltage B. Batteries installed wrong C. Meter requires service
2. Instrument will not calibrate
3. Instrument "loc k s up" A. Instrument ha s r e c'd a shock
4. Dissolved Oxygen readings are
inaccurate
A. Membrane is fouled or damaged B. Probe anode is fouled or da rk C. Probe cathode is fouled D. System requires service
B. Batteries are low or damaged C. System requires service
A. Cal altitude is incorrect B. P robe not in 100% water saturated
air during Cal proce dure
C. Membrane fouled or damage d D. Probe anode is fouled or dark E. Probe ca t hode is fouled F. System requires service
A. Replace batteries (Section 2) B. Check battery pola rity. (Section 2) C. Return system for service (Section 9)
A. Replace membrane cap (Section 3) B. Clean anode (Sec tion 6) C. Buff cathode ( Se c tion 6) D. Return system for service (Section 9)
A & B. Remove battery lid, wait 15 seconds for res e t, re place lid. (Section 2)
C. Return system for service (Section 9)
A. Recalibrate w/correct value (Section
4)
B. Moisten sponge & place in Cal
chamber w/ probe & Recal (Section 4)
C. Replace membrane cap (Section 3) D. Clean anode (Section 6) E. Buff cathode (Sec tion 6) F. Return system for service (Section 9)
5. LCD displays "LO BAT" A. Batteries are low or damaged A. Replace batteries (Section 2)
6. Main Display reads “OVEr”
(Secondary displa y reads “ovr”) (Secondary displa y reads “udr”)
7. Main display reads “PErr” A. Incorrect sequence of keystrokes. A. Refer to manual section for step by step
8. Main display reads “Err”
(Secondary displa y reads “ra”)
9. Main display reads “Err”
(Secondary displa y reads “ro”)
A. Temperature reading is >45°C B. Temperature reading is <-5°C C. DO temperature is >45°C D. DO % saturation is > 500% E. DO concentration is >50 m g /L F. Probe current too high to calibrate
A. System has failed its RAM test check procedure.
A. System has failed its ROM test check procedure.
In all cases, check calibration values and procedures. (Section 4)
If each of these were done correctly, return instrument for service. (Section 9)
instruction for the function you are attempting.
A. Turn instrument OFF and back ON again.
B. Return the system for service (Section 9) A.Turn instrument OFF and back ON
again. B. Return the system for service (Section 9)
YSI Incorporated Model 95 17
Page 22
Troubleshooting Section 8
10. Main reads “Undr” Probe current too low to calibrate
10. Main display reads “ FAIL”
(Secondary displa y reads “eep”)
A. EEPROM has failed to respond in time.
A. Return the system for service (Section 9)
11. Readings on main display don’t
change
A. Meter is in recall mode . A. Press MODE button to return to
Normal Operation ( Se c tion 4)
YSI Incorporated Model 95 18
Page 23
SECTION 9
SECTION 9 WARRANTY AND REPAIR
SECTION 9SECTION 9
YSI Model 95 Dissolved Oxygen Meters are warranted for two years from date of purchase by the end user against defects in materials and workmanship. YSI Model 95 probes and cables are warranted for one year from date of purchase by the end user against defects in material and workmanship. Within the warranty period, YSI will repair or replace, at its sole discretion, free of charge, any product that YSI determ ines to be cov ered by this warranty .
To exercise this warranty, write or call your local YSI representative, or contact YSI Customer Service in Yellow Springs, Ohio. Send the product and proof of purchase, transportation prepaid, to the Authorized Service Center selected by YSI. Repair or replacement will be made and the product returned, transportation prepaid. Repaired or replaced products are warranted for the balance of the original warranty period, or at least 90 days from date of repair or replacement.
Limitation of Warranty
This Warranty does not apply to any YSI product damage or failure caused by (i) failure to install, operate or use the product in accordance with YSI’s written instructions, (ii) abuse or misuse of the product, (iii) failure to maintain the product in accordance with YSI’s written instructions or standard industry procedure, (iv) any improper repairs to the product, (v) use by you of defective or improper components or parts in servicing or repairing the product, or (vi) modification of the product in any way not expressly authorized by YSI.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. YSI’s LIABILITY UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF THE PRODUCT, AND THIS SHALL BE YOUR SOLE AND EXCLUSIVE REMEDY FOR ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY. IN NO EVENT SHALL YSI BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM ANY DEFECTIVE PRODUCT COVERED BY THIS WARRANTY.
WARRANTY AND REPAIR
WARRANTY AND REPAIRWARRANTY AND REPAIR
YSI,Incorporated Model 95 19
Page 24
Warranty and Repair Section 9
AUTHORIZED U.S. SERVICE CENTERS
AUTHORIZED U.S. SERVICE CENTERS
AUTHORIZED U.S. SERVICE CENTERSAUTHORIZED U.S. SERVICE CENTERS
North and East Region
YSI Incorporated • Repair Center 1725 Brannum Lane Yellow Springs, Ohio • 45387 • Phone: (800) 765-4974 (937) 767-7241 E -Mail: info@ysi.com
YSI Incorporated Model 95 20
Page 25
Warranty and Repair Section 9
INTERNATIONAL SERVICE CENTERS
INTERNATIONAL SERVICE CENTERS
INTERNATIONAL SERVICE CENTERSINTERNATIONAL SERVICE CENTERS
YSI Incorporated • Repair Center 1725 Brannum Lane Yellow Springs, Ohio • 45387 • Phone: (937) 767-7241 E-Mail: info@ y si.com
SPECIALTY SERVICE CENTERS
SPECIALTY SERVICE CENTERS
SPECIALTY SERVICE CENTERSSPECIALTY SERVICE CENTERS
YSI Incorporated Model 95 21
Page 26
Warranty and Repair Section 9
CLEANING INSTRUCTIONS
CLEANING INSTRUCTIONS
CLEANING INSTRUCTIONSCLEANING INSTRUCTIONS
NOTE: Before they can be serviced, equipment exposed to biological, radioactive, or toxic materials must be cleaned and disinfected. Biological contamination is presumed for any
instrument, probe, or other device that has been used with body fluids or tissues, or with waste water. Radioactive contamination is presumed for any instrument, probe or other device that has been used near any radioactive source.
If an instrument, probe, or other part is returned or presented for service without a Cleaning Certificate, and if in our opinion it represents a potential biological or radioactive hazard, our service personnel reserve the right to withhold service until appropriate cleaning, decontamination, and certification has been completed. We will contact the sender for instructions as to the disposition of the equipment. Disposition costs will be the responsibility of the sender.
When service is required, either at the user's facility or at YSI, the following steps must be taken to insure the safety of our service personnel.
1. In a manner appropriate to each device, decontaminate all exposed surfaces, including any
containers. 70% isopropyl alcohol or a solution of 1/4 cup bleach to 1 gallon tap water are suitable for most disinfecting. Instruments used with waste water may be disinfected with .5% Lysol if this is more convenient to the user.
2. The user shall take normal precautions to prevent radioactive contamination and must use
appropriate decontamination procedures should exposure occur.
3. If exposure has occurred, the customer must certify that decontamination has been
accomplished and that no radioactivity is detectable by survey equipment.
4. Any product being returned to the YSI Repair Center, should be packed securely to prevent
damage.
5. Cleaning must be completed and certified on any product before returning it to YSI.
PACKING INSTRUCTIONS
PACKING INSTRUCTIONS
PACKING INSTRUCTIONSPACKING INSTRUCTIONS
1. Clean and decontaminate items to insure the safety of the handler.
2. Complete and include the Cleaning Certificate.
3. Place the product in a plastic bag to keep out dirt and packing m aterial.
4. Use a large carton, preferably the original, and surround the product completely with packing
material.
5. Insure for the replacement value of the product.
YSI Incorporated Model 95 22
Page 27
Warranty and Repair Section 9
Cleaning Certificate
Organization ________________________________ Department _________________________________ Address ___________________________________ City _______________ State ______ Zip ________ Country __________________ Model No. of Device ______ Lot Number _________ Contaminant (if known) _____________________ Cleaning Agent(s) used _____________________ Radioactive Decontamination Certified? (Answer only if there has been radioactive exposure) ___ Yes ___ No
Cleaning Certified By _________________________
Name Date
YSI Incorporated Model 95 23
Page 28
APPENDIX A
APPENDIX A GENERAL SPECIFICATIONS
APPENDIX AAPPENDIX A
Materials: ABS, Stainless Steel, and other materials
Dimensions:
Height: 9.5 inches (24.13 cm) Thickness: 2.2 inches (5.6 cm ) Width: 3.5 inches max. (8.89 cm) Weight: 1.7 pounds (w/ 10’ cable) (.77 kg) Display: 2.3”W x 1.5”L (5.8 cm W x 3.8 cm L)
Power: 6 AA-size Alkaline Batteries (included)
Approximately 150 hours operation from each new set of batteries Automatic shutoff after 4 hours without a key press
Water Tightness: Meets or exceeds IP65 standards
Operating Environment
Medium: fresh, sea, or polluted water and most other liquid solutions.
GENERAL SPECIFICATIONS
GENERAL SPECIFICATIONSGENERAL SPECIFICATIONS
Temperature: -5 to +45 °C Depth: 0 to 10, 0 to 25, 0 to 50, or 0 to 100 feet (depending on cable length)
Storage Temperature: -5 to +55 °C
System Performance Specifications:
Measurement Range Resolution Accuracy Temperature -5 to +45 °C 0.1 °C
Dissolved Oxygen 0 to 500 % Air Sat.
0 to 50 mg/L
DO Response Time: Application dependent. Typically 60 sec for 95% of the
change at 25°C
Temperature Response Time: Typically 95% of the chang e
Temperature Compensation: Automatic
0.1% Air Saturation
0.01 mg/L
± 0.2 °C ± 2% or ± 0.2 mg/L,
whichever is greater. 0 to 200% air. + 5% for 200% to 500% air
YSI Incorporated Model 95 24
Page 29
Salinity Compensation Range: 0 to 80 ppt (compensation between 40 and 80 is based on
extrapolation)
Stirring Dependence: 0.5-mil membrane: <5% error in stagnant water or <2% at a
flow rate of 2 in/sec at 120 nA nominal current 1-mil membrane: <3% error in stagnant water or <1% at a
flow rate of 2 in/sec at 100 nA nominal current
Data Storage: 50 points with ID number.
YSI Incorporated Model 95 25
Page 30
APPENDIX B
APPENDIX B REQUIRED NOTICE
APPENDIX BAPPENDIX B
The Federal Communications Commission defines this product as a computing device and requires the following notice:
This equipment generates and uses radio frequency energy and if not installed and used properly, may cause interference to radio and television reception. There is no guarantee that interference will not occur in a particular installation. If this equipment does cause interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
re-orient the receiving antenna
relocate the computer with respect to the receiver
move the computer away from the receiver
plug the computer into a different outlet so that the computer and receiver are on
different branch circuits.
If necessary, the user should consult the dealer or an experienced radio/television technician for additional suggestions. The user may find the following booklet, prepared by the Federal Communications Commission, helpful: "How to Identify and Resolve Radio-TV Interference Problems." This booklet is available from the U.S. Government Printing Office, Washington, DC 20402, Stock No. 0004-000-00345- 4.
REQUIRED NOTICE
REQUIRED NOTICEREQUIRED NOTICE
YSI Incorporated Model 95 26
Page 31
APPENDIX C
APPENDIX C ACCESSORIES AND REPLACEMENT PARTS
APPENDIX CAPPENDIX C
The following parts and accessories are available from YSI or any Franchise Dealer authorized by YSI.
YSI Order Number Description
9501 Replacement Membrane Cap Kit, 0.5 mil ( 6 each ) 9502 Replacement Membrane Cap Kit, 1 mil ( 6 each ) 9503 MEA Probe Reconditioning Kit 5520 Carryin g C a s e, s mall (95-10 or 95-25 only) 5050 Carryin g C a s e, medium (95- 50 or 95- 100) 059965 Replacement Probe & Cable As s embly (10 feet) 059981 Replacement Probe & Cable As s embly (25 feet) 059984 Replacement Probe & Cable As s embly (50 feet) 059987 Replacement Probe and Cable As s embly (100 feet) 059971 Replacement Front Cas e C over
ACCESSORIES AND REPLACEMENT PARTS
ACCESSORIES AND REPLACEMENT PARTSACCESSORIES AND REPLACEMENT PARTS
055242 Replacement Rear Case C ov er 055244 Replacement Battery C over Kit 055204 Replacement Case Gask et an d S crew 055219 Calibration Chamber Sponge 030157 Main Board Assembly
YSI Incorporated Model 95 27
Page 32
APPENDIX D
APPENDIX D UNIT CONVERSION
APPENDIX DAPPENDIX D
UNIT CONVERSION
UNIT CONVERSIONUNIT CONVERSION
Conversion Chart
To Convert From To Equation
Feet Meters Multiply by 0.3048 Meters Feet Multiply by 3.2808399 Degrees Celsius Degrees Fahrenheit (9/5oC)+32 Degrees Fahrenheit Degrees Celsius 5/9(oF-32) Milligrams per liter (mg /l) Parts per million (ppm) Multiply by 1
YSI Incorporated Model 95 28
Page 33
APPENDIX E
APPENDIX E OXYGEN SOLUBILITY TABLE
APPENDIX EAPPENDIX E
OXYGEN SOLUBILITY TABLE
OXYGEN SOLUBILITY TABLEOXYGEN SOLUBILITY TABLE
Solubility of Oxygen in m g /l in Water Exposed to Water-Saturated Air at 760 mm Hg Pressure. Salinity = Measure of quantity of dissolved salts in water.
Chlorinity = Measure of chloride content, by mass, of water.
0
S(
/00) = 1.80655 x Chlorinity (0/00)
Temp
o
C
0.0 14.62 13.73 12.89 12.10 11.36 10.66
1.0 14.22 13.36 12.55 11.78 11.07 10.39
2.0 13.83 13.00 12.22 11.48 10.79 10.14
3.0 13.46 12.66 11.91 11.20 10.53 9.90
4.0 13.11 12.34 11.61 10.92 10.27 9.66
5.0 12.77 12.02 11.32 10.66 10.03 9.44
6.0 12.45 11.73 11.05 10.40 9.80 9.23
7.0 12.14 11.44 10.78 10.16 9.58 9.02
8.0 11.84 11.17 10.53 9.93 9.36 8.83
9.0 11.56 10.91 10.29 9.71 9.16 8.64
10.0 11.29 10.66 10.06 9.49 8.96 8.45
11.0 11.03 10.42 9.84 9.29 8.77 8.28
12.0 10.78 10.18 9.62 9.09 8.59 8.11
Chlorinity:0
Salinity:0
5.0 ppt
9.0 ppt
10.0 ppt
18.1 ppt
15.0 ppt
27.1 ppt
20.0 ppt
36.1 ppt
25.0 ppt
45.2 ppt
13.0 10.54 9.96 9.42 8.90 8.41 7.95
14.0 10.31 9.75 9.22 8.72 8.24 7.79
15.0 10.08 9.54 9.03 8.54 8.08 7.64
16.0 9.87 9.34 8.84 8.37 7.92 7.50
17.0 9.67 9.15 8.67 8.21 7.77 7.36
18.0 9.47 8.97 8.50 8.05 7.62 7.22
19.0 9.28 8.79 8.33 7.90 7.48 7.09
20.0 9.09 8.62 8.17 7.75 7.35 6.96
21.0 8.92 8.46 8.02 7.61 7.21 6.84
22.0 8.74 8.30 7.87 7.47 7.09 6.72
23.0 8.58 8.14 7.73 7.34 6.96 6.61
YSI Incorporated Model 95 29
Page 34
Oxygen Solubility Table Appendix E
Temp
o
C
Chlorinity:0
Salinity:0
5.0 ppt
9.0 ppt
10.0 ppt
18.1 ppt
15.0 ppt
27.1 ppt
20.0 ppt
36.1 ppt
24.0 8.42 7.99 7.59 7.21 6.84 6.50
25.0 8.26 7.85 7.46 7.08 6.72 6.39
26.0 8.11 7.71 7.33 6.96 6.62 6.28
27.0 7.97 7.58 7.20 6.85 6.51 6.18
28.0 7.83 7.44 7.08 6.73 6.40 6.09
29.0 7.69 7.32 6.96 6.62 6.30 5.99
30.0 7.56 7.19 6.85 6.51 6.20 5.90
31.0 7.43 7.07 6.73 6.41 6.10 5.81
32.0 7.31 6.96 6.62 6.31 6.01 5.72
33.0 7.18 6.84 6.52 6.21 5.91 5.63
34.0 7.07 6.73 6.42 6.11 5.82 5.55
35.0 6.95 6.62 6.31 6.02 5.73 5.46
36.0 6.84 3.52 6.22 5.93 5.65 5.38
25.0 ppt
45.2 ppt
37.0 6.73 6.42 6.12 5.84 5.56 5.31
38.0 6.62 6.32 6.03 5.75 5.48 5.23
39.0 6.52 6.22 5.98 5.66 5.40 5.15
40.0 6.41 6.12 5.84 5.58 5.32 5.08
41.0 6.31 6.03 5.75 5.49 5.24 5.01
42.0 6.21 5.93 5.67 5.41 5.17 4.93
43.0 6.12 5.84 5.58 5.33 5.09 4.86
44.0 6.02 5.75 5.50 5.25 5.02 4.79
45.0 5.93 5.67 5.41 5.17 4.94 4.72
* This table is provided for your information only. It is NOT required when calibrating the Model 95 in accordance with the instructions outlined in the section entitled Calibration
YSI Incorporated Model 95 30
Page 35
APPENDIX F
APPENDIX F CALIBRATION VALUES TABLE
APPENDIX FAPPENDIX F
Calibration values for various atmospheric pressures and altitudes. Note: This table is for your information only. I t is not required for calibration.
CALIBRATION VALUES TABLE
CALIBRATION VALUES TABLECALIBRATION VALUES TABLE
Pressure Inches of Hg
30.23 768 102.3 -276 -84 101
29.92 760 101.3 0 0 100
29.61 752 100.3 278 85 99
29.33 745 99.3 558 170 98
29.02 737 98.3 841 256 97
28.74 730 97.3 1126 343 96
28.43 722 96.3 1413 431 95
28.11 714 95.2 1703 519 94
27.83 707 94.2 1995 608 93
27.52 699 93.2 2290 698 92
27.24 692 92.2 2587 789 91
26.93 684 91.2 2887 880 90
26.61 676 90.2 3190 972 89
26.34 669 89.2 3496 1066 88
26.02 661 88.2 3804 1160 87
25.75 654 87.1 4115 1254 86
25.43 646 86.1 4430 1350 85
25.12 638 85.1 4747 1447 84
24.84 631 84.1 5067 1544 83
24.53 623 83.1 5391 1643 82
24.25 616 82.1 5717 1743 81
23.94 608 81.1 6047 1843 80
23.62 600 80.0 6381 1945 79
23.35 593 79.0 6717 2047 78
23.03 585 78.0 7058 2151 77
22.76 578 77.0 7401 2256 76
22.44 570 76.0 7749 2362 75
22.13 562 75.0 8100 2469 74
21.85 555 74.0 8455 2577 73
21.54 547 73.0 8815 2687 72
21.26 540 71.9 9178 2797 71
20.94 532 70.9 9545 2909 70
20.63 524 69.9 9917 3023 69
20.35 517 68.9 10293 3137 68
Pressure mm Hg
Pressure kPA
Altitude in feet
Altitude in meters
Calibration Value in %
YSI Incorporated Model 95 31
Page 36
1700/1725 Brannum Lane Yellow Springs, Ohio 45387 USA (800) 765-4974 (937) 767-7241 FAX: (937) 767-9320 Website: http://www.ysi.com E-MAIL: environmental@ y si.com
ITEM # 059976
DRW # A59976D
October 1998
Loading...